SlideShare a Scribd company logo
FOLLOW UP AND MANAGEMENT OF
PACEMAKER PROGRAMMING AND
TROUBLESHOOTING
SOUMYA KANTI DUTTA.
 Pacemaker components
 Battery(Lithium iodide or Vanadium silver)
 Pacing impedance
 Pulse generator
1. Output circuit
2. Sensing circuit
3. Timing circuit
4. Rate adaptive sensor
5. Modes and mode switching
 Pacing impedance :
 Pacing impedance refers to the opposition to
current flow.
 Sources that contribute to pacing impedance:
1. Pacing lead conductor coil
2. Electrode-tissue interface
 Normal lead impedance vary from 250-
1200ohms.
 Single impedance value may be of little use
with out previous values for comparison.
Pulse generator output circuit
 Capture threshold or Pacing threshold .
 Minimum amount of energy required to constantly
cause depolarization
 Voltage(mA) and pulse duration(mSec)
 V=IR
2.0 v 1.5 v 1 v
1. Pulse generator output circuit
1. Pulse generator output
circuit
 High Pacemaker Output can
cause
◦ Reduce longevity
◦ Diaphragmatic stimulation
◦ Muscle Sti. in Unipolar
pacemakers
◦ Patient may “feel” heart
beat
Electrical Testing Of Pacemaker
Electrical Testing Of Pacemaker
2. Pulse generator sensing circuit :
◦ Ability of the device to detect intrinsic beat of
the heart
◦ Measured - peak to peak magnitude (mv) &
slew rate(mv/ms)
Pulse generator timing circuit :
a. Lower rate limit (LRL)
b. Hysteresis rate
c. Refractory and blanking periods
d. Ventricular safety pacing interval .
e. Upper rate response .
Electrical Testing Of Pacemaker
Pulse generator timing circuit :
 Lower rate interval - lowest rate that the pacemaker
will pace .
 Escape interval:sensed beat to next paced beat.
 A paced or non-refractory sensed event restarts the rate
timer at the programmed rate.
Condition LRL (beats/mt)
Infrequent pauses 40-50
Chronic persistent bradycardia 60-70
Relative bradycardia
detrimental (long QT)
70-80
Detrimental fast heart rates
(angina)
50-60
VVI 60-70
Pulse generator timing circuit :
Pulse generator timing circuit :
 Hysteresis :
 Hysteresis allows the rate to drop below the programmed
pacing LRL.
 Advantages of hysteresis :
1. Encourages native rhythm – maintain AV sync in VVI ,
prolong battery life
2. Prevent retrograde conduction – avoids pacemaker
syndrome
 AV delay (AVI) – pacemaker equivalent of PR interval.
 Sensed vs paced AVI – paced AVI is programmed at 125-
200ms , sensed AV interval is programmed at 20-50ms
shorter than paced.
 Dynamic/adaptive AV delay :shorter AVI with increase in
heart rate,allow pacemaker to respond to exercise and
have better haemodynamics.
Pulse generator timing circuit :
sAVI – 150ms
pAVI – 200ms
 AV delay (AVI)
 Longer AVI :
Good AV conduction – maintains AV synchrony ,
long battery life
Achieved by following methods :Programming
longer AVI , managed ventricular pacing , AV delay
hysteresis .
 Shorter AVI:
HOCM – RV apical pacing decreases HOCM gradient
CRT – usually 80-120ms , for 100%ventricular pacing
and optimize CO
Physiological response to faster heart rates can be
answered
Pulse generator timing circuit :
 Refractory and
blanking
periods :
 Refractory
period –
sensing present
but no action
 Blanking period
- sensing
absent and
hence no action
Pulse generator timing circuit :
Blanking
period
Time Importance
Atrial blanking
period
50-100ms Non programmable ,
Avoid atrial sensing of its own
paced beat
Post ventricular
atrial blanking
period
220ms Avoid sensing of ventricular
beat
Long PVAB decreases
detection of AF,AFL
Ventricular
blanking period
50-100ms Non programmable,
Avoid ventricular sensing of
its paced beat
Post atrial
ventricular
blanking period
28ms if the PAvB period is too long,
R on T - ventricular
tachyarrhythmia.
Refractory period Importance
Ventricular refractory period
(VRP)
Prevent sensing of T wave
.
Atrial refractory period
(ARP)
AVI (120-200ms) .
Post ventricular atrial
refractory period
Avoid sensing retrograde P
waves (PMT) , far field R
waves .
Pulse generator timing circuit :
Pulse generator timing circuit :
Ventricular safety pacing/ventricular
triggered period/cross talk sensing
window :
Atrial pacing in DDD
Trigger ventricular sensing
PAVB -pAVI
False inhibition of
ventricular pacing circuit
Asystole
 Rate responsive pacing refer to ability of pacemaker
to increase its lower rate in response to physiological
stimulus
 Upper rate limit:important to prevent tracking of
rapid atrial activity like in AF
 HRR should start with in 10s of exercise , peak at 90
– 120s and should return to baseline with in 60 –
120s after exercise.
 If intrinsic atrial rate exceeds URR then wenckebach
or 2:1 AVB
 Choosing URR : young patients (150b/mt) , old
angina (<110b/mt).
 Various sensors (activity , minute ventilation , QT)
 Rate smoothing: allows variation of R-R interval <6%
Pulse generator rate responsive pacing:
Pulse generator rate responsive pacing:
Wenckebach
2:1 AVB
PACING MODES
 BEPG(British pacing and electrophysiology
group)
 NASPE(North American society of pacing
and electrophysiology)
 DD I/T R
 III-D(I+R) :dual tracking of atrial activity while
inhibited by ventricular activity
 5th position: antitachycardia function(pacing/shock)
5. Pulse generator modes switching:
Electrical Testing Of Pacemaker
DDD / VDD
Atial tachyarrythmias
Sensed atrial events
Trigger fast ventricular rates
Palpitations. Dyspnoea. And Fatigue.
DDIR /
VVIR
Pulse generator modes switching:
Programming mode switching Mode switching
occurs when the sensed atrial rate exceeds a programmed
atrial tachycardia detection rate. By definition, this value
must be faster than the URL (maximum tracking rate).
Atrial tachycardia detection is typically programmed to
175-l88bpm or thereabouts.AMS base rate is higher than
LRI.
Electrical Testing Of Pacemaker
Pacemaker follow up guidelines:
Single chambered
pacing
Dual chambered
pacing
1st month q 2 week 1st month q 2 weeks
2nd -48th month q 12
week
2nd – 30th month q
12weeks
49th – 72nd month q 8
week
31st – 48th month q 8
weeks
73rd month and later q 4
weeks
49th month and later q
4week
Pacemaker complications
Pocket
complications
Pocket hematoma
Infection
Erosion
Wound pain
Allergic reactions
Pacemaker
complications
Lead dislodgement
Pneumothorax /air
embolism
Cardiac perforation
Extracardiac stimulation
Venous thrombosis
Coronary sinus
dissection
Twidller syndrome
Pacemaker malfunction
Pocket hematoma :
 The risk of haematoma is increased in patients
taking antithrombotic or anticoagulant drugs
(Goldstein et al., 1998).
 Most small hematomas can be managed
conservatively with cold compress and withdrawal
of antiplatelet or antithrombotic agents.
 Occasionally, large hematomas that compromise
the suture line or skin integrity may have to be
surgically evacuated.
 Needle aspiration increases risk of infection and
should not be done.
Pocket hematoma :
 In patients requiring oral anticoagulants (warfarin), to
take INR of about 2.0 at the time of implantation is
safe (Belott & Reynolds, 2000).
 Unfractionated heparin or low-molecular-weight
heparin are always discontinued prior to device implant
and ideally avoided for a minimum of 24 hours post
implantation.
 Administration of anticoagulants can be resumed
within 48-72 h after implantation if there is no
evidence of substantial hematoma formation.
Device-related infections :
 The reported incidence of pacemaker-related
infection ranges from 0.5% to 6% in early series.
 The use of prophylactic antibiotics has decreased
the rate of acute infections following pacemaker
implantations to <1 to 2 percent in most series.
 The mortality of persistent infection when infected
leads are not removed can be as high as 66%.
 DM, malignancy, operator inexperience, advanced
age, corticosteroid use, anticoagulation, recent
device manipulation, CRF, and bacteremia from a
distant focus of infection.
Device related infection :
2007;49;1851-1859 J. Am. Co
Twiddler syndrome:
Obese women with loose, fatty subcutaneous tissue
Small size of the implanted generator with a large pocket
Twisting of pulse generator in long axis
Lead dislodgement and lead fracture
Failure to capture
 The prevelance of this
syndrome is 0.07% (Gungor
et al., 2009)
 Rotated along the transverse
axis it is referred by us as
the reel syndrome.
 Pocket should be revised.
 Avoid by
◦ Limit the pocket size,
◦ Suture the device to the
fascia
◦ The patients not to
manipulate their device
pocket
Twiddler syndrome:
 Failure to capture
 Failure to output
 Sensing abnormalities(under and over
sensing)
 Specific mode complications
1. Pacemaker related tachycardia
2. Pacemaker syndrome
PACEMAKER MALFUNCTION
Failure to capture:
 Pacing artifact present but no evoked potential .
 Causes
1. Lead dislodgement or perforation
2. Lead maturation(inflammation/fibrosis)(exit
block)
3. Battery depletion
4. Circuit failure(coil fracture , insulation defect)
5. Inappropriate programming(low output /pulse
width)
6. Pseudo malfunction(pacing during refractory
period)
7. Functional non capture(oversensing).
8. Metabolic , drugs , cardiomyopathies
Medication effect on Capture
Drugs that increase
capture threshold
Drugs that decrease
capture threshold
Amiodarone
Flecainide
Propafenone
Sotalol
Procainamide
lidocaine
Atropine
Epinephrine
Isoproterenol
corticosteroids
Electrocardiographic tracing from a patient with a DDDR pacemaker. All ventricular
pacing artifacts but one failed to result in ventricular depolarization—
that is, failure to capture
Failure to capture:
Failure to capture:
Pacing threshold
Normal Increased
Battery depletion
Functional non
capture
Impedance
Normal
Dislodgement
Exit block
Decreased
Insulation
failure/
break
Increased
Lead
fracture
Loose screw
Failure to output:
 Absence of pacing stimuli and hence no capture .
 Causes
1. Pseudo malfunction - hysteresis , PMT
termination , sleep rate
2. Over sensing - EMI ; T P R over sensing ;
Myopotential/diaphragmatic ; Cross talk .
3. Open circuit - lead fracture , loose screw ,
incompatible lead .
4. Battery depletion
5. Recording artifact.
 Failure to output:
VVIR pacemaker This patient had a pacemaker programmed to a
unipolar sensing configuration. The sensing of myopotentials led to
symptomatic pauses, and reprogramming the pacemaker to a
bipolar sensing configuration prevented subsequent myopotential
over sensing.
Application of magnet
Failure to output:
Eliminates pauses Pauses persistent
Impedance
Normal
Decreased
Insulation
failure/break
Increased
Lead fracture
Loose screw
Battery
depletionOver sensing
Pseudo malfunction
Battery depletion :
 Elective replacement indicators (ERI)
1. Low voltage(2.1-2.4V)
2. Low pacing rate on magnet application
3. Elevated battery impedance
4. Increased pulse width duration
5. Restricted programmability
6. Change to simpler pacing mode
 End of life (EOL)
1. Low voltage(≤2.1vol)
Pacemaker undersensing :
 Pulse artifact or intrinsic complex present but no
sensing(sensed beat doesn’t reset escape interval or
cycle).
 Causes are
1. Defect in signal production – scar /fibrosis
following MI , ectopic , cardioversion,defibrillation ,
metabolic.
2. Defect in signal transmission – lead fracture
/dislodgement , insulation failure , partial open
circuit.
3. Defect in pacemaker – battery depletion , sensing
circuit abnormalities , committed DVI.
Pacemaker over sensing :
 Inappropriate inhibition as
failure to pace
 Causes
1. EMI
2. T , P , R over sensing .
3. Far field R wave
sensing(atrial).
4. Cross talk
5. Myopotential
(unipolar)
6. Make break signals
Cross talk :
High atrial output
High ventricular sensitivity
Low VBP
Ventricular sensing of paced
atrial impulse
Ventricular Asystole
Pacemaker Syndrome
“An assortment of symptoms
related to the adverse
hemodynamic impact from the
loss of AV synchrony.”
Pacemaker Syndrome
Symptoms include:
 Dizziness
 Presyncope
 Chest tightness
 Shortness of breath
 Neck pulsations(cannon A wave)
 Apprehension/malaise
 Fatigue
Pacemaker Syndrome May Be Caused By:
 Loss of capture, sensing
 A-V intervals of long duration
 Onset of 2:1 block
 Single chamber system(VVI)
 Absence of rate increase with exercise
 ECG:Retrogradely conducted inverted P over T
 CORRECTIVE ACTION:
◦ Reduce pacing rate.
◦ Hysteresis on to allow more sinus rhythm
◦ Shift to dual chamber pacing.
Pacemaker-Mediated Tachycardia
(PMT)
 A rapid paced rhythm that can occur
with atrial tracking pacemakers
PMT is the Result of:
 Also known as pacemaker re-entry
tachycardia or endless-loop tachycardia
(like other re-entry dysrhythmias,
pacemaker is now a part of the re-entry
circuit,Mostly anterograde limb of
reentrant tachycardia)
 Retrograde conduction
 PVARP
 Tracking fast atrial rates (physiologic or
non-physiologic)
 only in patients with dual chamber devices
 The ventricular impulse is retrogradely
conducted via the AV node to the atria, now the
atrial lead senses the atrial impulse and
generates a ventriculair pulse (completed loop)
 This pacemaker re-entry tachycardia will not
exceed the programmed upper rate, however
can be significant enough to cause symptoms
 magnet therapy.
 Pacemakers atrial sensing tresholds need to be
adjusted
 Mode switch can cause the pacemaker to
discontinue atrial tracking when the atrial upper
rate is exceeded
Rate Drop Response
 Delivers pacing at high rate when
episodic drop in rate occurs
 Pacing therapy indicated for patients
with neurocardiogenic syncope
Rate Drop Response
Venous thrombosis :
 Venous thrombosis occurs in 30% to 50% of patients
and only 1-3% of patients become symptomatic.
 Manifestations vary from usually asymptomatic,
acute symptomatic thrombosis, and even SVCS .
 Early or late after pacemaker implantation.
 Predictors of severe stenosis are multiple pacemaker
leads , previous pacing , double coils , hormone
therapy .
 Asymptomatic (no treatment) , symptomatic
(anticoagulants – endovascular stents – surgical
correction ).
Extracardiac stimulation
 The diaphragm or pectoral or intercostal muscles
 Diaphragmatic stimulation - direct stimulation of the
diaphragm (left) or stimulation of the phrenic nerve
(right).
 Early postimplantation period , dislodgment of the
pacing lead.
 MC in patients with LV coronary vein branch lead
placement for CRT
 Output pacing importance (testing and treatment)
 Pectoral stimulation - incorrect orientation of the
pacemaker or a current leak from a lead insulation
failure or exposed connector.
PERIOPERATIVE MANAGEMENT
 Require perioperative determination of pacemaker
dependency.
 An external defibrillator should be close by.
 Intraoperative management:
Application of magnet.
Perioperative reprogramming to
assynchronous mode.
Rate adaptive sensors should be disabled.
Oversensing of electrocautery is more common with
unipolar leads.
 When external cardioversion is required ,defibrillation pads
should be placed at least 8 inches from the pulse
generator.
THANK YOU

More Related Content

What's hot

Cath hemodynamics vir
Cath hemodynamics virCath hemodynamics vir
Cath hemodynamics vir
Dr Virbhan Balai
 
Echo assessment of cardiomyopathy pdf
Echo assessment of cardiomyopathy pdfEcho assessment of cardiomyopathy pdf
Echo assessment of cardiomyopathy pdf
Nizam Uddin
 
OCT in coronary PCI
OCT in coronary PCIOCT in coronary PCI
OCT in coronary PCI
AhmedElBorae1
 
Coronary lesion assessment
Coronary lesion assessmentCoronary lesion assessment
Coronary lesion assessment
Uday Prashant
 
Diagnostic catheters for coronary angiography
Diagnostic catheters for coronary angiography Diagnostic catheters for coronary angiography
Diagnostic catheters for coronary angiography
Aswin Rm
 
Cardiac Resynchronization therapy.pptx
Cardiac Resynchronization therapy.pptxCardiac Resynchronization therapy.pptx
Cardiac Resynchronization therapy.pptx
desktoppc
 
cath Lab Hemoduhynamic
cath Lab Hemoduhynamiccath Lab Hemoduhynamic
cath Lab Hemoduhynamic
Muhammad Naveed Saeed
 
Coronary artery dissection and perforation
Coronary artery dissection and perforationCoronary artery dissection and perforation
Coronary artery dissection and perforationFuad Farooq
 
Trans septal puncture
Trans septal punctureTrans septal puncture
Trans septal puncture
Satyam Rajvanshi
 
Pressure, Damping and Ventricularization_Crimson Publishers
Pressure, Damping and Ventricularization_Crimson PublishersPressure, Damping and Ventricularization_Crimson Publishers
Pressure, Damping and Ventricularization_Crimson Publishers
crimsonpublishersOJCHD
 
LEFT HEART CATHETERIZATION
LEFT HEART CATHETERIZATIONLEFT HEART CATHETERIZATION
LEFT HEART CATHETERIZATIONPraveen Nagula
 
Fraction flow reserve
Fraction flow reserveFraction flow reserve
Fraction flow reserve
Priyanka Thakur
 
Understanding pacemakers
Understanding pacemakersUnderstanding pacemakers
Understanding pacemakers
dibufolio
 
VSD devices
VSD devicesVSD devices
VSD devices
Navin Agrawal
 
Step by Step Rotational Athrectomy
Step by Step Rotational AthrectomyStep by Step Rotational Athrectomy
Step by Step Rotational Athrectomy
Dr Virbhan Balai
 
Bifurcation stenting
Bifurcation stentingBifurcation stenting
Bifurcation stenting
Dr Virbhan Balai
 
CORONARY BALLOONS PRACTICAL ASPECTS.pptx
CORONARY BALLOONS PRACTICAL ASPECTS.pptxCORONARY BALLOONS PRACTICAL ASPECTS.pptx
CORONARY BALLOONS PRACTICAL ASPECTS.pptx
RohitWalse2
 
Approch to bifurcation lesion
Approch to bifurcation lesionApproch to bifurcation lesion
Approch to bifurcation lesion
Ramachandra Barik
 
FRACTIONAL FLOW RESERVE
FRACTIONAL FLOW RESERVEFRACTIONAL FLOW RESERVE
FRACTIONAL FLOW RESERVE
Vishwanath Hesarur
 

What's hot (20)

Cath hemodynamics vir
Cath hemodynamics virCath hemodynamics vir
Cath hemodynamics vir
 
Echo assessment of cardiomyopathy pdf
Echo assessment of cardiomyopathy pdfEcho assessment of cardiomyopathy pdf
Echo assessment of cardiomyopathy pdf
 
OCT in coronary PCI
OCT in coronary PCIOCT in coronary PCI
OCT in coronary PCI
 
Coronary lesion assessment
Coronary lesion assessmentCoronary lesion assessment
Coronary lesion assessment
 
Diagnostic catheters for coronary angiography
Diagnostic catheters for coronary angiography Diagnostic catheters for coronary angiography
Diagnostic catheters for coronary angiography
 
Cardiac Resynchronization therapy.pptx
Cardiac Resynchronization therapy.pptxCardiac Resynchronization therapy.pptx
Cardiac Resynchronization therapy.pptx
 
cath Lab Hemoduhynamic
cath Lab Hemoduhynamiccath Lab Hemoduhynamic
cath Lab Hemoduhynamic
 
Coronary artery dissection and perforation
Coronary artery dissection and perforationCoronary artery dissection and perforation
Coronary artery dissection and perforation
 
Trans septal puncture
Trans septal punctureTrans septal puncture
Trans septal puncture
 
Left ventricular angiogram (1)
Left ventricular angiogram (1)Left ventricular angiogram (1)
Left ventricular angiogram (1)
 
Pressure, Damping and Ventricularization_Crimson Publishers
Pressure, Damping and Ventricularization_Crimson PublishersPressure, Damping and Ventricularization_Crimson Publishers
Pressure, Damping and Ventricularization_Crimson Publishers
 
LEFT HEART CATHETERIZATION
LEFT HEART CATHETERIZATIONLEFT HEART CATHETERIZATION
LEFT HEART CATHETERIZATION
 
Fraction flow reserve
Fraction flow reserveFraction flow reserve
Fraction flow reserve
 
Understanding pacemakers
Understanding pacemakersUnderstanding pacemakers
Understanding pacemakers
 
VSD devices
VSD devicesVSD devices
VSD devices
 
Step by Step Rotational Athrectomy
Step by Step Rotational AthrectomyStep by Step Rotational Athrectomy
Step by Step Rotational Athrectomy
 
Bifurcation stenting
Bifurcation stentingBifurcation stenting
Bifurcation stenting
 
CORONARY BALLOONS PRACTICAL ASPECTS.pptx
CORONARY BALLOONS PRACTICAL ASPECTS.pptxCORONARY BALLOONS PRACTICAL ASPECTS.pptx
CORONARY BALLOONS PRACTICAL ASPECTS.pptx
 
Approch to bifurcation lesion
Approch to bifurcation lesionApproch to bifurcation lesion
Approch to bifurcation lesion
 
FRACTIONAL FLOW RESERVE
FRACTIONAL FLOW RESERVEFRACTIONAL FLOW RESERVE
FRACTIONAL FLOW RESERVE
 

Viewers also liked

Pacemaker
PacemakerPacemaker
Management of a patient with pacemaker
Management of a patient with pacemakerManagement of a patient with pacemaker
Management of a patient with pacemaker
salman habeeb
 
Pacemaker
PacemakerPacemaker
Pacemaker
JANARDANCNMC
 
Pacemaker Rhythms
Pacemaker RhythmsPacemaker Rhythms
Pacemaker Rhythms
Michael-Joseph Agbayani
 
A D V A N C E D P A C I N G
A D V A N C E D  P A C I N GA D V A N C E D  P A C I N G
A D V A N C E D P A C I N G
Sherry Knowles
 
Cardiac pacemakerspart iii
Cardiac pacemakerspart iiiCardiac pacemakerspart iii
Cardiac pacemakerspart iii
salah_atta
 
Dr hardik temporary pacemaker preview (1)
Dr hardik temporary pacemaker  preview (1)Dr hardik temporary pacemaker  preview (1)
Dr hardik temporary pacemaker preview (1)
Hardik Patel
 
Cardiac pace makerspart 1
Cardiac pace makerspart 1Cardiac pace makerspart 1
Cardiac pace makerspart 1
salah_atta
 
Pacemaker ECGs. Yasmeen Kamal
Pacemaker ECGs. Yasmeen KamalPacemaker ECGs. Yasmeen Kamal
Pacemaker ECGs. Yasmeen Kamal
Tanta Rhythm Group
 
Pacemakers
PacemakersPacemakers
Pacemakers
Saurabh Gole
 
Pacemaker basics
Pacemaker basicsPacemaker basics
Pacemaker basics
drpraveen1986
 
Cardiac pacemakers part ii
Cardiac pacemakers part iiCardiac pacemakers part ii
Cardiac pacemakers part ii
salah_atta
 
Pacemakers ppt1
Pacemakers  ppt1Pacemakers  ppt1
Pacemakers ppt1
tulasi27
 
Temporary Pacemaker Slides
Temporary Pacemaker SlidesTemporary Pacemaker Slides
Temporary Pacemaker Slidesmaltiaziz24
 
pacemaker
pacemakerpacemaker
pacemaker
Fatima Ahmad
 

Viewers also liked (16)

Pacemaker
PacemakerPacemaker
Pacemaker
 
Management of a patient with pacemaker
Management of a patient with pacemakerManagement of a patient with pacemaker
Management of a patient with pacemaker
 
Pacemaker
PacemakerPacemaker
Pacemaker
 
Pacemaker Rhythms
Pacemaker RhythmsPacemaker Rhythms
Pacemaker Rhythms
 
A D V A N C E D P A C I N G
A D V A N C E D  P A C I N GA D V A N C E D  P A C I N G
A D V A N C E D P A C I N G
 
Cardiac pacemakerspart iii
Cardiac pacemakerspart iiiCardiac pacemakerspart iii
Cardiac pacemakerspart iii
 
Dr hardik temporary pacemaker preview (1)
Dr hardik temporary pacemaker  preview (1)Dr hardik temporary pacemaker  preview (1)
Dr hardik temporary pacemaker preview (1)
 
Cardiac pace makerspart 1
Cardiac pace makerspart 1Cardiac pace makerspart 1
Cardiac pace makerspart 1
 
Pacemaker ECGs. Yasmeen Kamal
Pacemaker ECGs. Yasmeen KamalPacemaker ECGs. Yasmeen Kamal
Pacemaker ECGs. Yasmeen Kamal
 
Pacemakers
PacemakersPacemakers
Pacemakers
 
Pacemaker basics
Pacemaker basicsPacemaker basics
Pacemaker basics
 
Cardiac pacemakers part ii
Cardiac pacemakers part iiCardiac pacemakers part ii
Cardiac pacemakers part ii
 
Pacemakers ppt1
Pacemakers  ppt1Pacemakers  ppt1
Pacemakers ppt1
 
Temporary Pacemaker Slides
Temporary Pacemaker SlidesTemporary Pacemaker Slides
Temporary Pacemaker Slides
 
Cardiac Pacemaker
Cardiac PacemakerCardiac Pacemaker
Cardiac Pacemaker
 
pacemaker
pacemakerpacemaker
pacemaker
 

Similar to Follow up and management of pacemaker programming and

Pacemaker & cabg
Pacemaker & cabgPacemaker & cabg
Pacemaker & cabg
Ehab Khairy
 
Patients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implicationsPatients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implications
Gowri Shankar
 
PACEMAKER by Dr.Sravani Vishnubhatla
PACEMAKER by Dr.Sravani VishnubhatlaPACEMAKER by Dr.Sravani Vishnubhatla
PACEMAKER by Dr.Sravani Vishnubhatla
DrSravaniVishnubhatl
 
Anesthesia and cardiac pacemaker
Anesthesia  and   cardiac pacemakerAnesthesia  and   cardiac pacemaker
Anesthesia and cardiac pacemaker
Vishal Cb
 
Anethesia and cardiac implantable electronic devices
Anethesia and cardiac implantable electronic devicesAnethesia and cardiac implantable electronic devices
Anethesia and cardiac implantable electronic devicesAmr Moustafa Kamel
 
Pacer ppt
Pacer pptPacer ppt
Pacer ppt
EM OMSB
 
心臟植入性電子儀器(CIED)護理照護指引-Cathroom Troubleshooting_20131019南區
心臟植入性電子儀器(CIED)護理照護指引-Cathroom Troubleshooting_20131019南區心臟植入性電子儀器(CIED)護理照護指引-Cathroom Troubleshooting_20131019南區
心臟植入性電子儀器(CIED)護理照護指引-Cathroom Troubleshooting_20131019南區Taiwan Heart Rhythm Society
 
basicsofecg-150707084411-lva1-app6892.ppt
basicsofecg-150707084411-lva1-app6892.pptbasicsofecg-150707084411-lva1-app6892.ppt
basicsofecg-150707084411-lva1-app6892.ppt
SheliDuya2
 
Introduction To Electrophysiology
Introduction To ElectrophysiologyIntroduction To Electrophysiology
Introduction To Electrophysiology
jmlafroscia
 
Pacemaker and anaesthesia.pptx
Pacemaker and anaesthesia.pptxPacemaker and anaesthesia.pptx
Pacemaker and anaesthesia.pptx
TulikaSingh60
 
Pacemakers and implantable cardiac defibrillators
Pacemakers and implantable cardiac defibrillatorsPacemakers and implantable cardiac defibrillators
Pacemakers and implantable cardiac defibrillators
AndrewCrofton
 
Trouble shoooting ICD AND CRT
Trouble shoooting ICD AND CRTTrouble shoooting ICD AND CRT
Trouble shoooting ICD AND CRT
Malleswara rao Dangeti
 
Anaesthesia for patient with pacemaker
Anaesthesia for patient with pacemakerAnaesthesia for patient with pacemaker
Anaesthesia for patient with pacemaker
HASSAN RASHID
 
TACHY ADITYA-2.pptx
TACHY ADITYA-2.pptxTACHY ADITYA-2.pptx
TACHY ADITYA-2.pptx
AdityaRahane7
 
Defibrillators
DefibrillatorsDefibrillators
Defibrillators
Sapan Jena
 
Pacemaker
PacemakerPacemaker
Pacemaker
Rajee Ravindran
 
Pacemaker and anaesthetic implications
Pacemaker and anaesthetic implicationsPacemaker and anaesthetic implications
Pacemaker and anaesthetic implications
Rahul Chauhan
 
pace1.ppt
pace1.pptpace1.ppt
Jp's pacemaker
Jp's pacemakerJp's pacemaker
Jp's pacemaker
Jyothis Prakash
 
Lvad
LvadLvad

Similar to Follow up and management of pacemaker programming and (20)

Pacemaker & cabg
Pacemaker & cabgPacemaker & cabg
Pacemaker & cabg
 
Patients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implicationsPatients with pacemaker anaesthetic implications
Patients with pacemaker anaesthetic implications
 
PACEMAKER by Dr.Sravani Vishnubhatla
PACEMAKER by Dr.Sravani VishnubhatlaPACEMAKER by Dr.Sravani Vishnubhatla
PACEMAKER by Dr.Sravani Vishnubhatla
 
Anesthesia and cardiac pacemaker
Anesthesia  and   cardiac pacemakerAnesthesia  and   cardiac pacemaker
Anesthesia and cardiac pacemaker
 
Anethesia and cardiac implantable electronic devices
Anethesia and cardiac implantable electronic devicesAnethesia and cardiac implantable electronic devices
Anethesia and cardiac implantable electronic devices
 
Pacer ppt
Pacer pptPacer ppt
Pacer ppt
 
心臟植入性電子儀器(CIED)護理照護指引-Cathroom Troubleshooting_20131019南區
心臟植入性電子儀器(CIED)護理照護指引-Cathroom Troubleshooting_20131019南區心臟植入性電子儀器(CIED)護理照護指引-Cathroom Troubleshooting_20131019南區
心臟植入性電子儀器(CIED)護理照護指引-Cathroom Troubleshooting_20131019南區
 
basicsofecg-150707084411-lva1-app6892.ppt
basicsofecg-150707084411-lva1-app6892.pptbasicsofecg-150707084411-lva1-app6892.ppt
basicsofecg-150707084411-lva1-app6892.ppt
 
Introduction To Electrophysiology
Introduction To ElectrophysiologyIntroduction To Electrophysiology
Introduction To Electrophysiology
 
Pacemaker and anaesthesia.pptx
Pacemaker and anaesthesia.pptxPacemaker and anaesthesia.pptx
Pacemaker and anaesthesia.pptx
 
Pacemakers and implantable cardiac defibrillators
Pacemakers and implantable cardiac defibrillatorsPacemakers and implantable cardiac defibrillators
Pacemakers and implantable cardiac defibrillators
 
Trouble shoooting ICD AND CRT
Trouble shoooting ICD AND CRTTrouble shoooting ICD AND CRT
Trouble shoooting ICD AND CRT
 
Anaesthesia for patient with pacemaker
Anaesthesia for patient with pacemakerAnaesthesia for patient with pacemaker
Anaesthesia for patient with pacemaker
 
TACHY ADITYA-2.pptx
TACHY ADITYA-2.pptxTACHY ADITYA-2.pptx
TACHY ADITYA-2.pptx
 
Defibrillators
DefibrillatorsDefibrillators
Defibrillators
 
Pacemaker
PacemakerPacemaker
Pacemaker
 
Pacemaker and anaesthetic implications
Pacemaker and anaesthetic implicationsPacemaker and anaesthetic implications
Pacemaker and anaesthetic implications
 
pace1.ppt
pace1.pptpace1.ppt
pace1.ppt
 
Jp's pacemaker
Jp's pacemakerJp's pacemaker
Jp's pacemaker
 
Lvad
LvadLvad
Lvad
 

Recently uploaded

PART A. Introduction to Costumer Service
PART A. Introduction to Costumer ServicePART A. Introduction to Costumer Service
PART A. Introduction to Costumer Service
PedroFerreira53928
 
Operation Blue Star - Saka Neela Tara
Operation Blue Star   -  Saka Neela TaraOperation Blue Star   -  Saka Neela Tara
Operation Blue Star - Saka Neela Tara
Balvir Singh
 
Unit 8 - Information and Communication Technology (Paper I).pdf
Unit 8 - Information and Communication Technology (Paper I).pdfUnit 8 - Information and Communication Technology (Paper I).pdf
Unit 8 - Information and Communication Technology (Paper I).pdf
Thiyagu K
 
The Challenger.pdf DNHS Official Publication
The Challenger.pdf DNHS Official PublicationThe Challenger.pdf DNHS Official Publication
The Challenger.pdf DNHS Official Publication
Delapenabediema
 
Supporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptxSupporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptx
Jisc
 
Basic phrases for greeting and assisting costumers
Basic phrases for greeting and assisting costumersBasic phrases for greeting and assisting costumers
Basic phrases for greeting and assisting costumers
PedroFerreira53928
 
special B.ed 2nd year old paper_20240531.pdf
special B.ed 2nd year old paper_20240531.pdfspecial B.ed 2nd year old paper_20240531.pdf
special B.ed 2nd year old paper_20240531.pdf
Special education needs
 
Unit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdfUnit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdf
Thiyagu K
 
Introduction to Quality Improvement Essentials
Introduction to Quality Improvement EssentialsIntroduction to Quality Improvement Essentials
Introduction to Quality Improvement Essentials
Excellence Foundation for South Sudan
 
Chapter 3 - Islamic Banking Products and Services.pptx
Chapter 3 - Islamic Banking Products and Services.pptxChapter 3 - Islamic Banking Products and Services.pptx
Chapter 3 - Islamic Banking Products and Services.pptx
Mohd Adib Abd Muin, Senior Lecturer at Universiti Utara Malaysia
 
Home assignment II on Spectroscopy 2024 Answers.pdf
Home assignment II on Spectroscopy 2024 Answers.pdfHome assignment II on Spectroscopy 2024 Answers.pdf
Home assignment II on Spectroscopy 2024 Answers.pdf
Tamralipta Mahavidyalaya
 
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
Nguyen Thanh Tu Collection
 
Digital Tools and AI for Teaching Learning and Research
Digital Tools and AI for Teaching Learning and ResearchDigital Tools and AI for Teaching Learning and Research
Digital Tools and AI for Teaching Learning and Research
Vikramjit Singh
 
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
siemaillard
 
The approach at University of Liverpool.pptx
The approach at University of Liverpool.pptxThe approach at University of Liverpool.pptx
The approach at University of Liverpool.pptx
Jisc
 
The Art Pastor's Guide to Sabbath | Steve Thomason
The Art Pastor's Guide to Sabbath | Steve ThomasonThe Art Pastor's Guide to Sabbath | Steve Thomason
The Art Pastor's Guide to Sabbath | Steve Thomason
Steve Thomason
 
2024.06.01 Introducing a competency framework for languag learning materials ...
2024.06.01 Introducing a competency framework for languag learning materials ...2024.06.01 Introducing a competency framework for languag learning materials ...
2024.06.01 Introducing a competency framework for languag learning materials ...
Sandy Millin
 
Polish students' mobility in the Czech Republic
Polish students' mobility in the Czech RepublicPolish students' mobility in the Czech Republic
Polish students' mobility in the Czech Republic
Anna Sz.
 
Thesis Statement for students diagnonsed withADHD.ppt
Thesis Statement for students diagnonsed withADHD.pptThesis Statement for students diagnonsed withADHD.ppt
Thesis Statement for students diagnonsed withADHD.ppt
EverAndrsGuerraGuerr
 
Model Attribute Check Company Auto Property
Model Attribute  Check Company Auto PropertyModel Attribute  Check Company Auto Property
Model Attribute Check Company Auto Property
Celine George
 

Recently uploaded (20)

PART A. Introduction to Costumer Service
PART A. Introduction to Costumer ServicePART A. Introduction to Costumer Service
PART A. Introduction to Costumer Service
 
Operation Blue Star - Saka Neela Tara
Operation Blue Star   -  Saka Neela TaraOperation Blue Star   -  Saka Neela Tara
Operation Blue Star - Saka Neela Tara
 
Unit 8 - Information and Communication Technology (Paper I).pdf
Unit 8 - Information and Communication Technology (Paper I).pdfUnit 8 - Information and Communication Technology (Paper I).pdf
Unit 8 - Information and Communication Technology (Paper I).pdf
 
The Challenger.pdf DNHS Official Publication
The Challenger.pdf DNHS Official PublicationThe Challenger.pdf DNHS Official Publication
The Challenger.pdf DNHS Official Publication
 
Supporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptxSupporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptx
 
Basic phrases for greeting and assisting costumers
Basic phrases for greeting and assisting costumersBasic phrases for greeting and assisting costumers
Basic phrases for greeting and assisting costumers
 
special B.ed 2nd year old paper_20240531.pdf
special B.ed 2nd year old paper_20240531.pdfspecial B.ed 2nd year old paper_20240531.pdf
special B.ed 2nd year old paper_20240531.pdf
 
Unit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdfUnit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdf
 
Introduction to Quality Improvement Essentials
Introduction to Quality Improvement EssentialsIntroduction to Quality Improvement Essentials
Introduction to Quality Improvement Essentials
 
Chapter 3 - Islamic Banking Products and Services.pptx
Chapter 3 - Islamic Banking Products and Services.pptxChapter 3 - Islamic Banking Products and Services.pptx
Chapter 3 - Islamic Banking Products and Services.pptx
 
Home assignment II on Spectroscopy 2024 Answers.pdf
Home assignment II on Spectroscopy 2024 Answers.pdfHome assignment II on Spectroscopy 2024 Answers.pdf
Home assignment II on Spectroscopy 2024 Answers.pdf
 
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
 
Digital Tools and AI for Teaching Learning and Research
Digital Tools and AI for Teaching Learning and ResearchDigital Tools and AI for Teaching Learning and Research
Digital Tools and AI for Teaching Learning and Research
 
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
 
The approach at University of Liverpool.pptx
The approach at University of Liverpool.pptxThe approach at University of Liverpool.pptx
The approach at University of Liverpool.pptx
 
The Art Pastor's Guide to Sabbath | Steve Thomason
The Art Pastor's Guide to Sabbath | Steve ThomasonThe Art Pastor's Guide to Sabbath | Steve Thomason
The Art Pastor's Guide to Sabbath | Steve Thomason
 
2024.06.01 Introducing a competency framework for languag learning materials ...
2024.06.01 Introducing a competency framework for languag learning materials ...2024.06.01 Introducing a competency framework for languag learning materials ...
2024.06.01 Introducing a competency framework for languag learning materials ...
 
Polish students' mobility in the Czech Republic
Polish students' mobility in the Czech RepublicPolish students' mobility in the Czech Republic
Polish students' mobility in the Czech Republic
 
Thesis Statement for students diagnonsed withADHD.ppt
Thesis Statement for students diagnonsed withADHD.pptThesis Statement for students diagnonsed withADHD.ppt
Thesis Statement for students diagnonsed withADHD.ppt
 
Model Attribute Check Company Auto Property
Model Attribute  Check Company Auto PropertyModel Attribute  Check Company Auto Property
Model Attribute Check Company Auto Property
 

Follow up and management of pacemaker programming and

  • 1. FOLLOW UP AND MANAGEMENT OF PACEMAKER PROGRAMMING AND TROUBLESHOOTING SOUMYA KANTI DUTTA.
  • 2.  Pacemaker components  Battery(Lithium iodide or Vanadium silver)  Pacing impedance  Pulse generator 1. Output circuit 2. Sensing circuit 3. Timing circuit 4. Rate adaptive sensor 5. Modes and mode switching
  • 3.  Pacing impedance :  Pacing impedance refers to the opposition to current flow.  Sources that contribute to pacing impedance: 1. Pacing lead conductor coil 2. Electrode-tissue interface  Normal lead impedance vary from 250- 1200ohms.  Single impedance value may be of little use with out previous values for comparison.
  • 4. Pulse generator output circuit  Capture threshold or Pacing threshold .  Minimum amount of energy required to constantly cause depolarization  Voltage(mA) and pulse duration(mSec)  V=IR
  • 5. 2.0 v 1.5 v 1 v 1. Pulse generator output circuit
  • 6. 1. Pulse generator output circuit  High Pacemaker Output can cause ◦ Reduce longevity ◦ Diaphragmatic stimulation ◦ Muscle Sti. in Unipolar pacemakers ◦ Patient may “feel” heart beat Electrical Testing Of Pacemaker
  • 7. Electrical Testing Of Pacemaker 2. Pulse generator sensing circuit : ◦ Ability of the device to detect intrinsic beat of the heart ◦ Measured - peak to peak magnitude (mv) & slew rate(mv/ms)
  • 8. Pulse generator timing circuit : a. Lower rate limit (LRL) b. Hysteresis rate c. Refractory and blanking periods d. Ventricular safety pacing interval . e. Upper rate response . Electrical Testing Of Pacemaker
  • 9. Pulse generator timing circuit :  Lower rate interval - lowest rate that the pacemaker will pace .  Escape interval:sensed beat to next paced beat.  A paced or non-refractory sensed event restarts the rate timer at the programmed rate.
  • 10. Condition LRL (beats/mt) Infrequent pauses 40-50 Chronic persistent bradycardia 60-70 Relative bradycardia detrimental (long QT) 70-80 Detrimental fast heart rates (angina) 50-60 VVI 60-70 Pulse generator timing circuit :
  • 11. Pulse generator timing circuit :  Hysteresis :  Hysteresis allows the rate to drop below the programmed pacing LRL.  Advantages of hysteresis : 1. Encourages native rhythm – maintain AV sync in VVI , prolong battery life 2. Prevent retrograde conduction – avoids pacemaker syndrome
  • 12.  AV delay (AVI) – pacemaker equivalent of PR interval.  Sensed vs paced AVI – paced AVI is programmed at 125- 200ms , sensed AV interval is programmed at 20-50ms shorter than paced.  Dynamic/adaptive AV delay :shorter AVI with increase in heart rate,allow pacemaker to respond to exercise and have better haemodynamics. Pulse generator timing circuit : sAVI – 150ms pAVI – 200ms
  • 13.  AV delay (AVI)  Longer AVI : Good AV conduction – maintains AV synchrony , long battery life Achieved by following methods :Programming longer AVI , managed ventricular pacing , AV delay hysteresis .  Shorter AVI: HOCM – RV apical pacing decreases HOCM gradient CRT – usually 80-120ms , for 100%ventricular pacing and optimize CO Physiological response to faster heart rates can be answered Pulse generator timing circuit :
  • 14.  Refractory and blanking periods :  Refractory period – sensing present but no action  Blanking period - sensing absent and hence no action Pulse generator timing circuit :
  • 15. Blanking period Time Importance Atrial blanking period 50-100ms Non programmable , Avoid atrial sensing of its own paced beat Post ventricular atrial blanking period 220ms Avoid sensing of ventricular beat Long PVAB decreases detection of AF,AFL Ventricular blanking period 50-100ms Non programmable, Avoid ventricular sensing of its paced beat Post atrial ventricular blanking period 28ms if the PAvB period is too long, R on T - ventricular tachyarrhythmia.
  • 16. Refractory period Importance Ventricular refractory period (VRP) Prevent sensing of T wave . Atrial refractory period (ARP) AVI (120-200ms) . Post ventricular atrial refractory period Avoid sensing retrograde P waves (PMT) , far field R waves . Pulse generator timing circuit :
  • 17. Pulse generator timing circuit : Ventricular safety pacing/ventricular triggered period/cross talk sensing window : Atrial pacing in DDD Trigger ventricular sensing PAVB -pAVI False inhibition of ventricular pacing circuit Asystole
  • 18.  Rate responsive pacing refer to ability of pacemaker to increase its lower rate in response to physiological stimulus  Upper rate limit:important to prevent tracking of rapid atrial activity like in AF  HRR should start with in 10s of exercise , peak at 90 – 120s and should return to baseline with in 60 – 120s after exercise.  If intrinsic atrial rate exceeds URR then wenckebach or 2:1 AVB  Choosing URR : young patients (150b/mt) , old angina (<110b/mt).  Various sensors (activity , minute ventilation , QT)  Rate smoothing: allows variation of R-R interval <6% Pulse generator rate responsive pacing:
  • 19. Pulse generator rate responsive pacing: Wenckebach 2:1 AVB
  • 20. PACING MODES  BEPG(British pacing and electrophysiology group)  NASPE(North American society of pacing and electrophysiology)  DD I/T R  III-D(I+R) :dual tracking of atrial activity while inhibited by ventricular activity  5th position: antitachycardia function(pacing/shock)
  • 21. 5. Pulse generator modes switching: Electrical Testing Of Pacemaker DDD / VDD Atial tachyarrythmias Sensed atrial events Trigger fast ventricular rates Palpitations. Dyspnoea. And Fatigue. DDIR / VVIR
  • 22. Pulse generator modes switching: Programming mode switching Mode switching occurs when the sensed atrial rate exceeds a programmed atrial tachycardia detection rate. By definition, this value must be faster than the URL (maximum tracking rate). Atrial tachycardia detection is typically programmed to 175-l88bpm or thereabouts.AMS base rate is higher than LRI.
  • 23. Electrical Testing Of Pacemaker Pacemaker follow up guidelines: Single chambered pacing Dual chambered pacing 1st month q 2 week 1st month q 2 weeks 2nd -48th month q 12 week 2nd – 30th month q 12weeks 49th – 72nd month q 8 week 31st – 48th month q 8 weeks 73rd month and later q 4 weeks 49th month and later q 4week
  • 24.
  • 25. Pacemaker complications Pocket complications Pocket hematoma Infection Erosion Wound pain Allergic reactions Pacemaker complications Lead dislodgement Pneumothorax /air embolism Cardiac perforation Extracardiac stimulation Venous thrombosis Coronary sinus dissection Twidller syndrome Pacemaker malfunction
  • 26. Pocket hematoma :  The risk of haematoma is increased in patients taking antithrombotic or anticoagulant drugs (Goldstein et al., 1998).  Most small hematomas can be managed conservatively with cold compress and withdrawal of antiplatelet or antithrombotic agents.  Occasionally, large hematomas that compromise the suture line or skin integrity may have to be surgically evacuated.  Needle aspiration increases risk of infection and should not be done.
  • 27. Pocket hematoma :  In patients requiring oral anticoagulants (warfarin), to take INR of about 2.0 at the time of implantation is safe (Belott & Reynolds, 2000).  Unfractionated heparin or low-molecular-weight heparin are always discontinued prior to device implant and ideally avoided for a minimum of 24 hours post implantation.  Administration of anticoagulants can be resumed within 48-72 h after implantation if there is no evidence of substantial hematoma formation.
  • 28. Device-related infections :  The reported incidence of pacemaker-related infection ranges from 0.5% to 6% in early series.  The use of prophylactic antibiotics has decreased the rate of acute infections following pacemaker implantations to <1 to 2 percent in most series.  The mortality of persistent infection when infected leads are not removed can be as high as 66%.  DM, malignancy, operator inexperience, advanced age, corticosteroid use, anticoagulation, recent device manipulation, CRF, and bacteremia from a distant focus of infection.
  • 29. Device related infection : 2007;49;1851-1859 J. Am. Co
  • 30. Twiddler syndrome: Obese women with loose, fatty subcutaneous tissue Small size of the implanted generator with a large pocket Twisting of pulse generator in long axis Lead dislodgement and lead fracture Failure to capture
  • 31.  The prevelance of this syndrome is 0.07% (Gungor et al., 2009)  Rotated along the transverse axis it is referred by us as the reel syndrome.  Pocket should be revised.  Avoid by ◦ Limit the pocket size, ◦ Suture the device to the fascia ◦ The patients not to manipulate their device pocket Twiddler syndrome:
  • 32.  Failure to capture  Failure to output  Sensing abnormalities(under and over sensing)  Specific mode complications 1. Pacemaker related tachycardia 2. Pacemaker syndrome PACEMAKER MALFUNCTION
  • 33. Failure to capture:  Pacing artifact present but no evoked potential .  Causes 1. Lead dislodgement or perforation 2. Lead maturation(inflammation/fibrosis)(exit block) 3. Battery depletion 4. Circuit failure(coil fracture , insulation defect) 5. Inappropriate programming(low output /pulse width) 6. Pseudo malfunction(pacing during refractory period) 7. Functional non capture(oversensing). 8. Metabolic , drugs , cardiomyopathies
  • 34. Medication effect on Capture Drugs that increase capture threshold Drugs that decrease capture threshold Amiodarone Flecainide Propafenone Sotalol Procainamide lidocaine Atropine Epinephrine Isoproterenol corticosteroids
  • 35. Electrocardiographic tracing from a patient with a DDDR pacemaker. All ventricular pacing artifacts but one failed to result in ventricular depolarization— that is, failure to capture Failure to capture:
  • 36. Failure to capture: Pacing threshold Normal Increased Battery depletion Functional non capture Impedance Normal Dislodgement Exit block Decreased Insulation failure/ break Increased Lead fracture Loose screw
  • 37. Failure to output:  Absence of pacing stimuli and hence no capture .  Causes 1. Pseudo malfunction - hysteresis , PMT termination , sleep rate 2. Over sensing - EMI ; T P R over sensing ; Myopotential/diaphragmatic ; Cross talk . 3. Open circuit - lead fracture , loose screw , incompatible lead . 4. Battery depletion 5. Recording artifact.
  • 38.  Failure to output: VVIR pacemaker This patient had a pacemaker programmed to a unipolar sensing configuration. The sensing of myopotentials led to symptomatic pauses, and reprogramming the pacemaker to a bipolar sensing configuration prevented subsequent myopotential over sensing.
  • 39. Application of magnet Failure to output: Eliminates pauses Pauses persistent Impedance Normal Decreased Insulation failure/break Increased Lead fracture Loose screw Battery depletionOver sensing Pseudo malfunction
  • 40. Battery depletion :  Elective replacement indicators (ERI) 1. Low voltage(2.1-2.4V) 2. Low pacing rate on magnet application 3. Elevated battery impedance 4. Increased pulse width duration 5. Restricted programmability 6. Change to simpler pacing mode  End of life (EOL) 1. Low voltage(≤2.1vol)
  • 41. Pacemaker undersensing :  Pulse artifact or intrinsic complex present but no sensing(sensed beat doesn’t reset escape interval or cycle).  Causes are 1. Defect in signal production – scar /fibrosis following MI , ectopic , cardioversion,defibrillation , metabolic. 2. Defect in signal transmission – lead fracture /dislodgement , insulation failure , partial open circuit. 3. Defect in pacemaker – battery depletion , sensing circuit abnormalities , committed DVI.
  • 42. Pacemaker over sensing :  Inappropriate inhibition as failure to pace  Causes 1. EMI 2. T , P , R over sensing . 3. Far field R wave sensing(atrial). 4. Cross talk 5. Myopotential (unipolar) 6. Make break signals Cross talk : High atrial output High ventricular sensitivity Low VBP Ventricular sensing of paced atrial impulse Ventricular Asystole
  • 43. Pacemaker Syndrome “An assortment of symptoms related to the adverse hemodynamic impact from the loss of AV synchrony.”
  • 44. Pacemaker Syndrome Symptoms include:  Dizziness  Presyncope  Chest tightness  Shortness of breath  Neck pulsations(cannon A wave)  Apprehension/malaise  Fatigue
  • 45. Pacemaker Syndrome May Be Caused By:  Loss of capture, sensing  A-V intervals of long duration  Onset of 2:1 block  Single chamber system(VVI)  Absence of rate increase with exercise  ECG:Retrogradely conducted inverted P over T  CORRECTIVE ACTION: ◦ Reduce pacing rate. ◦ Hysteresis on to allow more sinus rhythm ◦ Shift to dual chamber pacing.
  • 46. Pacemaker-Mediated Tachycardia (PMT)  A rapid paced rhythm that can occur with atrial tracking pacemakers
  • 47. PMT is the Result of:  Also known as pacemaker re-entry tachycardia or endless-loop tachycardia (like other re-entry dysrhythmias, pacemaker is now a part of the re-entry circuit,Mostly anterograde limb of reentrant tachycardia)  Retrograde conduction  PVARP  Tracking fast atrial rates (physiologic or non-physiologic)
  • 48.  only in patients with dual chamber devices  The ventricular impulse is retrogradely conducted via the AV node to the atria, now the atrial lead senses the atrial impulse and generates a ventriculair pulse (completed loop)  This pacemaker re-entry tachycardia will not exceed the programmed upper rate, however can be significant enough to cause symptoms  magnet therapy.  Pacemakers atrial sensing tresholds need to be adjusted  Mode switch can cause the pacemaker to discontinue atrial tracking when the atrial upper rate is exceeded
  • 49. Rate Drop Response  Delivers pacing at high rate when episodic drop in rate occurs  Pacing therapy indicated for patients with neurocardiogenic syncope Rate Drop Response
  • 50. Venous thrombosis :  Venous thrombosis occurs in 30% to 50% of patients and only 1-3% of patients become symptomatic.  Manifestations vary from usually asymptomatic, acute symptomatic thrombosis, and even SVCS .  Early or late after pacemaker implantation.  Predictors of severe stenosis are multiple pacemaker leads , previous pacing , double coils , hormone therapy .  Asymptomatic (no treatment) , symptomatic (anticoagulants – endovascular stents – surgical correction ).
  • 51. Extracardiac stimulation  The diaphragm or pectoral or intercostal muscles  Diaphragmatic stimulation - direct stimulation of the diaphragm (left) or stimulation of the phrenic nerve (right).  Early postimplantation period , dislodgment of the pacing lead.  MC in patients with LV coronary vein branch lead placement for CRT  Output pacing importance (testing and treatment)  Pectoral stimulation - incorrect orientation of the pacemaker or a current leak from a lead insulation failure or exposed connector.
  • 52.
  • 53. PERIOPERATIVE MANAGEMENT  Require perioperative determination of pacemaker dependency.  An external defibrillator should be close by.  Intraoperative management: Application of magnet. Perioperative reprogramming to assynchronous mode. Rate adaptive sensors should be disabled. Oversensing of electrocautery is more common with unipolar leads.  When external cardioversion is required ,defibrillation pads should be placed at least 8 inches from the pulse generator.