4. Stuart Allen 06
Contains a battery that provides
the energy for sending electrical
impulses to the heart
Houses the circuitry that
controls pacemaker operations Circuitry
Battery
The Pulse Generator
5. THE PULSE GENERATOR……
Common battery compositions
include:
Lithium-Iodine
Lithium silver vanadium
oxide with carbon
monoflouride
Starting battery voltage will
vary depending on
composition
Longevity
Dependent on impedance
and output
Commonly ranges from 6-12
years
14. CONDUCTORS -
MP-35N – Alloy of nickel , cobalt, chromium, molybdenum
advantage – high strength and resistance to corrosion
disadvantage – high electrical resistance
This is overcome by incorporating low resistance metals –
silver ,titanium , platinum .
These alloy can be constructed in two ways -
17. LEAD INSULATION -
POLYURETHANE –
High tensile strength
resistance to mechanical abrasion
has excellent lubricity with low frictional coefficient
Thin layer of insulation can be used so reducing overall lead
dimension
Main disadvantage – in vivo polymer degradation
18. Two mechanism –
ESC (environmental stress cracking ) at point of
mechanical stress , at anchoring sleevs , venous entry
point
Metal ion oxidation (MIO)- at conductor insulator
interface
19.
20. SILICON INSULATION
Completely inert and biostable
More flexible – less risk of cardiac perforation
Main disadvantage – low tensile strength susceptibility to
abrasion and tear
So thicker layer of insulation needed which lead to bulky
design
Cold flow phenomenon
21.
22. Lead Insulation May Be Silicone or
Polyurethane
Advantages of Silicone-
Insulated Leads
• Inert
• Biocompatible
• Biostable
• Repairable with
medical adhesive
• Historically very
reliable
Advantages of
Polyurethane-
Insulated Leads
• Biocompatible
• High tear
strength
• Low friction
coefficient
• Smaller lead
diameter
23. Epicardial Leads
Leads applied directly to
the surface of the heart
Utilized in pediatric
patients and patients
contraindicated for
transvenous leads
Fixation mechanisms
include:
Epicardial stab-in
Myocardial screw-in
Suture-on
Applied via sternotomy,
thoroscopy, or limited
thoracotomy
24. Flows through the tip electrode
(cathode)
Stimulates the heart
Returns through body fluid and
tissue to the PG (anode)
A UNIPOLAR PACING SYSTEM
CONTAINS A LEAD WITH AN ELECTRODE IN THE HEART
Cathode
Anode
-
+
25. Anode
Flows through the tip
electrode located at the
end of the lead wire
Stimulates the heart
Returns to the ring
electrode above the lead
tip
A BIPOLAR PACING SYSTEM
CONTAINS A LEAD WITH 2 ELECTRODES IN THE HEART
Cathode
26. Unipolar leads
Unipolar leads have a smaller diameter than bipolar leads
Unipolar leads exhibit larger pacing artifacts on the surface
ECG
• One electrode on the tip & one conductor coil
• Conductor coil may consist of multiple strands - (multifilar
leads)
27. Bipolar leads
Bipolar leads are less susceptible to oversensing
noncardiac signals (myopotentials and EMI)
Coaxial Lead Design
• Circuit is tip electrode to ring electrode
• Two conductor coils (one inside the other)
• Inner layer of insulation
• Bipolar leads are typically thicker than unipolar leads
28. Unipolar Bipolar
Advantages Smaller diameter
Easier to implant
Large spike
No pocket stimulation
Less susceptible to EMI
Programming flexibility
Disadvantages Pocket stimulation
Far-field oversensing
No programming flexibility
Larger diameter
Stiffer lead body
Small spike
Higher impedance
Voltage threshold is 30%
higher
32. VOLTAGE
Voltage is the force, or “push,” that
causes electrons to move through a
circuit
In a pacing system, voltage is:
Measured in volts (V)
Represented by the letter “V”
Provided by the pacemaker battery
Often referred to as amplitude or pulse
amplitude
33. CURRENT
The flow of electrons in a completed
circuit
In a pacing system, current is:
Measured in milliamps (mA)
Represented by the letter “I”
Determined by the amount of electrons that
move through a circuit
34. IMPEDANCE
The opposition to current flow
In a pacing system, impedance is:
Measured in ohms (W)
Represented by the letter “R”
The measurement of the sum of all resistance to the flow
of current
Impedance is a function of the characteristics of the conductor
(wire), the electrode (tip), and the myocardium
35. Impedance is the sum of all resistance to the flow of
current. The resistive factors to a pacing system
include:
• Lead conductor resistance
• The resistance to current flow from the electrode to the
myocardium
• POLARIZATION IMPEDANCE, which is the accumulation
of charges of opposite polarity in the myocardium at the
electrode-tissue interface.
36. Impedance
Pacing lead impedance typically stated in broad ranges,
i.e. 300 to 1500 Ω
Factors that can influence impedance
Resistance of the conductor coils
Tissue between anode and cathode
The electrode/myocardial interface
Size of the electrode’s surface area
Size and shape of the tip electrode
37. Voltage, Current, and Impedance are
Interdependent
The interrelationship of the three components is
analogous to the flow of water through a hose
Voltage represents the force with which . . .
Current (water) is delivered through . . .
A hose, where each component represents the total
impedance:
The nozzle, representing the electrode
The tubing, representing the lead wire
38. Voltage and Current Flow
Electrical Analogies
Spigot (voltage) turned up, lots
of water flows (high current
drain)
Spigot (voltage) turned low, little flow
(low current drain)
Water pressure in system
is analogous to voltage –
providing the force to
move the current
39. Resistance and Current Flow
Electrical Analogies
• Normal resistance – friction caused by the hose and nozzle
More water discharges, but is all of it going
to the nozzle?
• High resistance – a knot results in low total current flow
• Low resistance – leaks in the hose reduce the resistance
40.
41.
42. POLARIZATION
After an output pulse, positively charged particles
gather near the electrode.
The amount of positive charge is
Directly proportional to pulse duration
Inversely proportional to the functional electrode
size
(i.e. smaller electrodes offer higher polarization)
Polarization effect can represent 30–40% of the total pacing
impedance As high as 70% for smooth surface, small surface area
electrodes
43. Within the electrode, current flow is due to movement of electrons
(e−).
At the electrode–tissue interface, the current flow becomes ionic &
(-) vely charged ions (Cl−, OH−) flow into the tissues toward the
anode leaving behind oppositely charged particles attracted by the
emerging electrons.
It is this capacitance effect at the electrode tissue interface, that is
the basis of polarization
44.
45.
46.
47.
48. Lead wire
fracture
Increased
resistance
High Impedance Conditions
A Fractured Conductor
A fractured wire can cause
Impedance values to rise
Current flow from the battery may
be too low to be effective
Impedance values may exceed
3,000 W
Other reason for high impedance: Lead not seated properly
in pacemaker header (usually an acute problem).
49. Low Impedance Conditions
An Insulation Break
Insulation breaks can
cause impedance values
to fall
Current drain is high and can
lead to more rapid battery
depletion
Current can drain through
the insulation break into the
body or other lead wire, not
through myocardium
Impedance values may
be less than 300 W
50. Ohm’s Law
Describes the relationship
between voltage, current,
and resistance
V = I X R
I = V / R
R = V / I
V
I R
V
R
I
V
R
I
R
V
I =
=
=
X
51. Ohm’s law tells us:
1. If the impedance remains constant, and the voltage
decreases, the current decreases.
1. If the voltage is constant, and the impedance decreases,
the current increases.
52.
53. Stimulation Threshold
Pacing Voltage Threshold
The minimum pacing voltage
at any given pulse width
required to consistently stimulate the heart
outside the myocardial refractory period causing it to
contract
54. E = Energy delivered by Pulse to the Pacing
Circuit
and Cardiac Tissue
- V . I . t
I2Rt V2t/R
55.
56. Rheobase- (the lowest point on the curve) by definition is the lowest
voltage that results in myocardial depolarization at infinitely long
pulse duration
Chronaxie(pulse duration time ) by definition, the chronaxie is the
threshold pulse duration at twice the rheobase voltage
61. Effect of Lead Design on Capture
Lead maturation
Fibrotic “capsule” develops around the electrode
following lead implantation
May gradually raise threshold
Usually no measurable effect on impedance
62. STEROID ELUTING LEADS
Steroid eluting leads reduce
the inflammatory process
Exhibit little to no acute
stimulation threshold
peaking
Leads maintain low
chronic thresholds
64. Evolution of Pacing Threshold
Voltage
Threshold
(V)
Observation Time (weeks)
Acute Phase
Chronic Phase
Safety Margin
6
s
4
3
2
1
0 4 8 12 16
65.
66. INJURY CURRENT…….
Due to pressure exerted by electrode on myocardium
Seen in both active and passive leads(less pronounced)
Lack of injury current – denotes poor contact / scarred myocardium
Disappears over minutes to hours
this may leads to high threshold soon after lead placement which
subside at end of procedure
67.
68.
69.
70. Magnet mode
I. Activates magnetic reed switch
II. Model dependent behavior with magnet
I. Asynchronous pacing – most common
II. No apparent rate or rhythm change
III. Brief asynchronous pacing and then return to programmed
value
IV. Continuous or transient loss of pacing
III. Magnets will cause most DDDs to convert to DOO at about 85
with a BOL (beginning of life) battery, and to VOO at a rate of
about 65 at ERI (effective replacement interval) is reached to
conserve power.
71. The Pacing Pulse
t
Pacing Pulse
Pulse Duration (Width)
Output
Voltage
V = Pulse Amplitude in Volts (V) (say 2.5 V)
t = Pulse Duration or Width in milliseconds (ms)
(say 0.5 ms)
R = Impedance of Pacing Circuit (ohms)
(say 500 ohms)
I = V/R = Current through pacing circuit (mA)
= 2.5 V/ 500 ohms = 0.005 A = 5 mA
V
t
72. Why measure stimulation threshold?
To enable programming stimulus voltage amplitude and
pulse width such that
Consistent capture & Patient Safety is ensured
Battery drain minimized, Pacemaker longevity maximized
Good thresholds
Ventricle - <1V @ 0.5ms
Atrium - <1.5V @ 0.5ms
73. SENSING OF INTRINSIC HEARTBEATS
Sensing is the ability of the pacemaker to “see” when an
intrinsic depolarization is occurring
Pacemakers record the Intracardiac Electrogram (EGM)
by constantly recording the potential difference
between the cathode and anode
Depolarization Wave
Processed by
Device
74.
75.
76.
77.
78.
79. Sensitivity Setting
Sensitivity settings less than 2.5 mv – High sensitivity – can lead
to oversensing
Sensitivity settings greater than 2.5 mV – Low sensitivity – can
lead to undersensing
Amplitude
(mV)
Amplitude
(mV)
Time Time
5.0
2.5
1.25
5.0
2.5
1.25
80. Undersensing . . .
Pacemaker does not “see” the intrinsic beat, and
therefore does not respond appropriately
Intrinsic beat
not sensed
Scheduled pace
delivered
VVI / 60
81. Oversensing
An electrical signal other than the intended P or R
wave is detected
Pacing is inhibited
Marker channel
shows intrinsic
activity...
...though no
activity is
present
99. HYSTERESIS……
Goal is to encourage own intrinsic activity .
One of the cause of rate variation in paced ECG
Hysteresis work best in patient with intrinsic rate near the
programmed base rate.