MAPLESON SYSTEM
Dr. Padmaja Pallavi Pandey
ERA’s Lucknow Medical College & Hospital
2014 Batch
CONTENTS
 History of Breathing System
 Characteristics
 Classification
 Components
 Lacks
 Mapleson A System
 Configurations
 Classic form
 Lack Modification
CONTENTS
 Techniques of Use
 Functional Analysis
 Spontaneous Respiration
 Controlled or Assisted Ventilation
 Hazards
 Preuse Checks
 Mapleson B System
 Techniques of Use
CONTENTS
 Functional Analysis
 Spontaneous Respiration
 Controlled or Assisted Ventilation
 Mapleson C System
 Techniques of Use
 Functional Analysis
 Mapleson D System
 Configurations
 Classic form
CONTENTS
 Bain Modification
 Techniques of Use
 Functional Analysis
 Spontaneous Breathing
 Controlled Ventilation
 Bain System Hazards
 Preuse Checks
 Continuous Positive Airway Pressure
CONTENTS
 Mapleson E System
 Techniques of Use
 Functional Analysis
 Rebreathing
 Air Dilution
 Hazards
 Mapleson F System
 Techniques of Use
 Functional Analysis
CONTENTS
 Hazards
 Respiratory Gas Monitoring with the Mapleson
Systems
 Advantages of the Mapleson Systems
 Disadvantages of the Mapleson System
“NECESSITY IS MOTHER OF INVENTION”
 Earlier circuits were simple, differing in the type of
anesthetic agent administered.
 The purpose of breathing systems that have evolved
in anesthetic practice is to deliver Gas & Vapor to the
patient in an appropriate, controlled & efficient
manner.
 1846 Sir W.T.G
Morton did public
demonstration with
Ether.
 1876  Clover`s Inhaler
developed by J.T Clover.
 1907 Barth used it to
administer N₂O.
 1909 Teter`s apparatus
developed.
 1909-13 F.W.Hewitts
developed Hewitt`s
apparatus.
 1913 Gwathemy
Apparatus developed.
 1917 Boyle`s Apparatus
developed.
 1928 Magill`s Circuit was
developed.
 1937 Philip Ayre
introduced T piece.
HISTORY
 1972 J.A Bain & W.E Spoerel introduced Bain`s
Circuit.
 1975 Dr Gordon Jackson Rees developed
Mapleson F system.
 Humphrey Davy, Brock & Downing  developed
combined ADE system.
History
In 1954, Prof. WW Mapleson from University of
Wales, Cardiff, classified the several breathing
systems around depending on what components they
contained and what position they took in the system,
Analyzed Five different anaesthetic breathing
systems & referred as Mapleson A – E.
In 1975 Willis et al described F system & added to
above.
In 1976 – Lack circuit .
HISTORY
.
.
CHARACTERISTICS
 Absence of unidirectional valves to direct gases to or
from the patient.
 FGF must wash Co2 out of the circuit, because there is
no device for absorbing Co2. So, known as “Co2
Washout Circuits or Flow-Controlled Breathing
Systems”.
 No clear seperation of Inspired & Expired gases,
therefore, Rebreathing will occur, Inspiratory flow >
FGF.
CHARACTERISTICS
 Composition of the inspired mixture will depend on
how much Rebreathing takes place.
 Monitoring End-Tidal Carbon-Dioxide is the best
method to determine the optimal FGF.
 With Rebreathing , arterial Carbon-Dioxide to End-
Tidal Carbon-Dioxide gradient decreases.
CLASSIFICATION
Six types :-
 Mapleson A – Magill & Lack System
 Mapleson B
 Mapleson C Bagging System
 Mapleson D or Bain System
 Mapleson E
 Mapleson F or Jackson-Rees modification of the T-
Piece
COMPONENTS
 Reservoir Bag
 Corrugated Tubing
 APL Valve
 Fresh gas inlet
 Patient Connection
LACKS
 Carbon-Dioxide absorbers
 Unidirectional Valves
 Separate Inspiratory & Expiratory limbs.
For better understanding of functional analysis they have
been classified as -
1) Afferent Reservoir System (ARS)
2) Enclosed Afferent Reservoir System
3) Efferent Reservoir System
4) Combined System
The efficiency of a system is determined in terms of CO₂
elimination & FGF utilization.
• Afferent limb is that part of the breathing system
which delivers the fresh gas from the machine to the
patient.
• If the reservoir is placed in this limb as in Mapleson A,
B, C and Lack’s systems they are called as afferent
reservoir system.
• Efferent limb is that part of the breathing system
which carries the expired gas from the patient and
vents it to the atmosphere through the expiratory
valve/port.
• If the reservoir is placed in this limb as in Mapleson D,
E, F and Bain systems they are called efferent reservoir
system
Mapleson postulates (1954)
 Mapleson has analyzed these bi-directional flow
systems & few basic assumptions have been made
which are of historical interest.
 Gases move En-bloc i.e they maintain their identity as
fresh gas, dead space gas & alveolar gas. There is no
mixing of these gases.
• Reservoir bags continues to fill up, without offering
any resistance till it is full.
• The expiratory valve opens as soon as the reservoir
bag is full & pressure inside the system goes above
the atmospheric pressure.
• The valve remains open throughout the expiratory
phase without offering any resistance to gas flow &
closes at the start of next inspiration.
MAPLESON A SYSTEM
• Originally described by
Evan Magill.
• Length of breathing tube
 110-180 cms.
• FGF  from machine
end.
• APL  close to patient.
• Sampling ports to be
placed between APL
valve & the tube.
MAPLESON A SYSTEM
CONFIGURATIONS :-
1) CLASSIC FORM –
 Also k/a Magill attachment or system
 Fresh gas does not enter the system near the patient
connection but enters at the other end of the system
near the reservoir bag.
 A corrugated tubing connects the bag to the
adjustable pressure limiting(APL) Valve at the
patient end of the system.
MAPLESON A
MAPLESON A SYSTEM
MAPLESON A SYSTEM
 A sensor for a nondiverting respiratory gas monitor or
the sampling site for a diverting monitor may be
placed between the :-
a) APL Valve & the Corrugated tubing.
b) APL Valve & the patient - result in excessive dead
space in small patients.
c) Neck of the bag & its mount
d) Bag & the Corrugated tubing.
e) In the fresh gas supply tube.
MAPLESON A SYSTEM
2) LACK MODIFICATION –
 Has an added “Expiratory” Limb, which runs from the
patient connection to the APL Valve at the machine end
of the system, makes it easier to adjust the valve &
facilitates scavenging excess gases, increases the work
of breathing slightly.
 Available in both-
 Dual (Parallel) tube arrangement
 Tube-Within-a-tube (Coaxial) Configuration, the
Expiratory limb runs concentrically inside the outer
inspiratory limb.
Lack system
 Co-axial Mapleson A.
 Outer tube 30mm in
diameter.
 Inner tube 14mm in
diameter.
 APL valve placed near
machine end.
LACK’S MODIFICATION
MAPLESON A (LACK) SYSTEM
MAPLESON A SYSTEM
TECHNIQUES OF USE :-
SPONTANEOUS VENTILATION –
 APL Valve is kept in the fully open position.
 During late phase of exhalation, excess gas exits
through it.
MAPLESON A SYSTEM
CONTROLLED OR ASSISTED VENTILATION –
 Intermittent positive pressure is applied to the bag.
 The APL Valve is partially closed so that when the bag
is squeezed, sufficient pressure to inflate the lungs is
achieved.
 The APL Valve opens during inspiration.
MAPLESON A SYSTEM
FUNCTIONAL ANALYSIS :-
1) SPONTANEOUS RESPIRATION –
 Patient exhales-
 First “Dead Space Gas” & then “Alveolar Gases” flow
into the Corrugated tubing toward the bag
 “Fresh Gas” flows into the bag.
 Bag is full , pressure in the system rises until the APL
Valve opens.
 First gas vented will be “Alveolar Gas”.
 Spontaneous
Breathing
3 phases identified 
 Inspiratory
 Expiratory
 Expiratory Pause.
FGF Dead space gas Alveolar gas
MAPLESON A- spontaneous respiration
MAPLESON A SYSTEM
 Remainder of exhalation i.e. only “Alveolar Gas”, exhausts
through the open APL Valve.
 In the Corrugated tubing, the continuing inflow of “Fresh
Gas” reverses the flow of exhaled gases.
 Some alveolar gas that bypassed the APL Valve, returns
back & exits through it.
 FGF is :-
High –Force the dead space gas out.
Intermediate –Some dead space gas will be retained in the
system.
Low –More alveolar gas will be retained.
MAPLESON A SYSTEM
 Start of Inspiration-
 First gas inhaled will be from
“Dead Space Gas” between the
patient & the APL Valve.
 Next gas will be either
a) Alveolar Gas (if the FGF is
low)
b) Dead Space Gas (if the FGF
is intermediate)
c) Fresh Gas (if the FGF is high)
 Changes in respiratory pattern
have little effect on
Rebreathing.
MAPLESON A SYSTEM
 Rebreathing begins when the FGF is reduced to 56-
82 ml/kg/min or 58% - 83% of minute volume.
 FGF to 51-85 ml/kg/min & 42% - 88% of minute
volume have been recommended to avoid
rebreathing.
 .
 .
.
MAPLESON A SPONTANEOUS
MAPLESON A SYSTEM
2) CONTROLLED OR ASSISTED VENTILATION –
 The pattern of gas flow changes.
 Exhalation-
 Pressure in the system will remain low
 No gas will escape through the APL Valve, unless the
bag becomes distended.
 Exhaled gases i.e. both “Dead Space Gas” & “Alveolar
Gas” remain in the Corrugated tubing.
 “Alveolar Gas” is nearest to the patient.
MAPLESON A- controlled ventilation
MAPLESON A - CONTROLLED
MAPLESON A SYSTEM
 Some “Alveolar Gas” may enter the bag, if the Tidal
Volume is large.
 Early Inspiration –
 Gases in the tubing flow to the patient.
 “Alveolar Gas” will be inhaled first because it
occupies the space nearest to the patient.
 As the pressure in the system rises, the APL Valve
opens, so that gas, both exits through the APL Valve &
flows to the patient.
MAPLESON A SYSTEM
 Late Inspiration –
 All the exhaled gas has been driven from the tube.
 “Fresh Gas” fills the tubing.
 Some “Fresh Gas” enters the patient & some is vented
through the valve.
 Rebreathing of “Alveolar Gases” & venting of “Fresh
Gases”.
 The composition of the inspired gas mixture depends
on the respiratory pattern.
MAPLESON A SYSTEM
 As the expiratory phase is prolonged, system becomes
more efficient.
 During Inspiration, if the APL Valve does not vent
gas, the Mapleson A System can be as efficient as the
Mapleson D System.
 During Assisted Ventilation, it is less efficient than
with Spontaneous Ventilation but is more efficient than
with Controlled Ventilation.
MAPLESON A SYSTEM
HAZARDS :-
 A Mechanical Ventilator that vents excess gases should
not be used because the entire system then becomes
“Dead Space”.
TESTING FOR LEAKS/MAGILL
PREUSE CHECKS :-
 Tested for leaks by occluding the patient end of the
system, closing the APL Valve & pressurizing the
system.
 Opening the APL Valve will confirm proper
functioning of that component.
 The user or a patient should breathe through the
system.
TESTING FOR LEAKS/LACK
 The coaxial lack system requires additional testing to
confirm the integrity of the inner tube –
A) To attach a tracheal tube to the inner tubing at the
patient end of the system. Blowing down the tube
with the APL Valve closed will produce movement of
the bag, if there is a leak between the two limbs.
B) To occlude both limbs at the patient connection with
the APL Valve open & then squeeze the bag. Leak in
the inner limb, gas will escape through the APL Valve
& the bag will collapse.
MAPLESON B SYSTEM
 The fresh gas inlet & APL
Valve are both located
near the patient port.
 The reservoir bag is at the
patient end of the system,
seperated from the fresh
gas inlet by corrugated
tubing.
MAPLESON B SYSTEM
MAPLESON B SYSTEM
FUNCTIONAL ANALYSIS :-
1) SPONTANEOUS RESPIRATION –
 Exhalation
 “Dead Space Gas” will pass down the corrugated
tubing, along with “Fresh Gas”.
 The tubing near the patient will be filled with “Fresh
Gas” & some “Alveolar Gas”.
 When the bag reaches full capacity, the APL Valve
opens & both “Fresh Gas” and “Alveolar Gas” will exit
from the system.
MAPLESON B SYSTEM
 Inspiration –
 The APL Valve closes
 The patient inhales “Fresh Gas” and gas from the
tubing.
 No gas will be inhaled from the bag, if the volume of
the tubing exceeds the tidal volume.
 To avoid Rebreathing, the FGF must be equal to peak
inspiratory flow rate (normally 20-25l/min)
MAPLESON B SYSTEM
 A FGF more than double Minute Volume has been
recommended, but flows as low as 0.8 -1.2 times
Minute Volume may be sufficient.
2) CONTROLLED OR ASSISTED VENTILATION –
 Behavior is similar to that of the Maplson A
 During the EXPIRATORY PAUSE, “Fresh Gas”
accumulates at the patient end of the tubing. So, more
efficient.
MAPLESON B SYSTEM
 The composition of the inspired gas is influenced by
the ventilatory pattern
 Has variable performance.
 A FGF of 2-2.5 times Minute Volume has been
recommended.
MAPLESON C SYSTEM
 Identical to the Mapleson
B System except that the
Corrugated tubing is
omitted.
TECHNIQUES OF USE :-
Similar to that described for
the Mapleson B System.
FUNCTIONAL ANALYSIS
Behaves similarly to the
Mapleson B System.
MAPLESON C (BAGGING SYSTEM)
MAPLESON C SYSTEM
MAPLESON C SYSTEM
 Also k/a “Westminster face piece”.
1) SPONTANEOUS VENTILATION –
 When the EXPIRATORY PAUSE is minimal, almost
as efficient as the Mapleson A but becomes less
efficient as the EXPIRATORY PAUSE increases.
 A FGFof 2 times Minute Volume has been
recommended for spontaneous breathing.
2) CONTROLLED VENTILATION :-
A FGF of 2-2.5 times Minute Volume is recommended.
MAPLESON A,B,C
MAPLESON D SYSTEM
 Have a T-piece near the patient & function similarly.
 The T-piece is a three way tubular connector with a –
a) Patient connection port
b) Fresh Gas port
c) Port for connection to a corrugated tubing.
 Popular because excess gas scavenging is relatively
easy.
 Most efficient during Controlled Ventilation
MAPLESON D
MAPLESON D SYSTEM
CONFIGURATION :-
1) CLASSIC FORM –
 A length of tubing connects the T-piece at the patient
end to the APL Valve and the reservoir bag adjacent to
it.
 The length of the tubing determines the distance the
user can be from the patient but has minimal effects
on Ventilation.
MAPLESON D SYSTEM
 The sensor or sampling site for a respiratory gas
monitor may be placed between the –
a) Bag & its mount
b) Corrugated tubing & the T-piece
c) Corrugated tubing & the APL Valve
d) T-piece & the patient
MAPLESON D SYSTEM
BIDIRECTIONAL PEEP
VALVE
 Placed between the
corrugated tubing & the
APL Valve.
 Permits PEEP to be
administered during
manual or mechanical
ventilation.
 Spontaneous breathing is
impossible because when
a negative pressure is
applied, it closes.
UNIDIRECTIONAL PEEP
VALVE
 Used at the bag attachment
site by using a special
connectors.
 Allows PEEP to be applied
during spontaneous or
mechanical but not manual
ventilation.
MAPLESON D SYSTEM
BIDIRECTIONAL PEEP
VALVE
 Placed in the hose
leading to anesthesia
ventilator, thus effective
only during Mechanical
Ventilation.
UNIDIRECTIONAL PEEP
VALVE
MAPLESON D SYSTEM
2) BAIN MODIFICATION –
 The “Fresh Gas” supply tube runs coaxially inside the
corrugated tubing, ends at the point where the “Fresh
Gas” would enter if the classic Mapleson D form were
used.
 The outer tube is clear so that the inner tube can be
inspected.
 The outer tube is narrow.
 Available with a metal head with channels drilled into
it, provides a fixed position for the reservoir bag & APL
Valve and attachment of corrugated tubing.
MAPLESON D BAIN SYSTEM
Bain modification of Mapleson D
system
 Originally modified by Bain
& Spoerel in 1972.
 Is co-axial system.
 Usual length is 180cm.
 Outer tube 
 Diameter -22mm.
 Carries exhaled gas.
 Inner tube 
 Diameter-7mm.
 Carries fresh gas.
MAPLESON D SYSTEM
 Some heads also have a “Pressure Manometer”.
 Long Version- used for remote anesthesia in MRI.
 Static Compliance is increased with a reduction in
peak inspiratory pressure & Tidal Volume with the
same ventilator settings.
 PEEP is increased.
 Longer, increased resistance to spontaneous breathing.
BAIN’S CIRCUIT
MAPLESON D SYSTEM
TECHNIQUES OF USE :-
SPONTANEOUS RESPIRATION –
 APL Valve is left open.
 Excess gases are vented during expiration.
MANUALLY CONTROLLED/ASSISTED VENTILATION
 Partial closure of APL Valve
 Squeezing the bag.
 Excess gases are vented during inspiration.
MAPLESON D SYSTEM
MECHANICALLY CONTROLLED VENTILATION –
 Connecting the hose from a ventilator in place of the
reservoir bag.
 Closing the APL Valve.
 Excess gases are vented through the ventilator spill
valve.
MAPLESON D SYSTEM
FUNCTIONAL ANALYSIS :-
1) SPONTANEOUS BREATHING –
 Exhalation-
 Exhaled gases mix with fresh gas, move through the
corrugated tube, towards the bag.
 Bag has filled, gas exits via the APL Valve.
 Expiratory Pause –
 FG pushes exhaled gases down the corrugated tubing.
MAPLESON D - Spontaneous respiration
MAPLESON D SYSTEM
 Inspiration –
 Patient will inhale gas from
the FG inlet & the corrugated
tubing.
 FG flow is
a)High- All the gases drawn
from the corrugated tube will
be FG.
b)Low- Exhaled gas containing
Co2 will be inhaled,
ventilatory pattern will help
to determine the amount of
Rebreathing.
MAPLESON D SYSTEM
 Factors that decreases Rebreathing –
 High Inspiratory:Expiratory (I:E) time ratio
 A slow rise in Inspiratory flow rate.
 A low flow rate during the last part of exhalation.
 A long Expiratory pause, having the greatest effect.
 The End-Tidal Co2 will rise, when gas containing Co2
is inhaled.
 The End-Tidal Co2 will ”fall” while inspired Co2 will
increase, when patient’s spontaneous respiration
increases.
MAPLESON D SYSTEM
 End-Tidal Co2 reaches “Plateau” no matter how hard
the patient works, the End-Tidal Co2 cannot be
lowered further.
 End-Tidal Co2 will “rise”, when the patient’s
respiration is depressed.
 End-Tidal Co2 depends on both the ratio of Minute
Volume & FGF and their absolute values.
 Expired Volume > FGF, end-tidal Co2 will be
determined mainly by FGF.
MAPLESON D SYSTEM
 FGF > Minute Volume, end-tidal Co2 will be
determined mainly by Minute Volume.
 Recommendations for FGF based on body weight
vary from 100-300 ml/kg/min.
 FGF should be 1.5 to 3.0 times the Minute Volume.
 FGF = Total ventilation (Adequate).
 FGF of 4000-4700 ml/m2/min have been
recommended.
MAPLESON D-controlled ventilation
MAPLESON D SYSTEM
2) CONTROLLED VENTILATION –
 Exhalation-
 Gases flow from the patient down to the corrugated
tubing.
 FG enters the tubing.
 Expiratory Pause-
 FGF continues, pushes exhaled gases down to the
tubing.
BAIN’S CONTROLLED.
MAPLESON D SYSTEM
 Inspiration –
 FG & gas from the corrugated tubing enter the
patient.
 FGF is “low”, exhaled gases may be inhaled.
 Increased Rebreathing-
 Inspiratory time is prolonged
 Respiratory Rate is increased
 Adding an Inspiratory Plateau
MAPLESON D SYSTEM
 Decreased Rebreathing –
 Long Expiratory Pause, FG can flush exhaled gases
from the tubing.
 FGF is “high”-
 Less rebreathing
 End-Tidal Co2 is determined mainly by Minute
Ventilation, Tidal Volume, the volume of the expiratory
limb & Expiratory Resistance.
 Minute Volume > FGF , FGF is the main factor
controlling Co2 elimination.
Recommendations by Bain & Spoerel
• 2L/min FGF in patients <10kg.
• 3.5L/min FGF in patients between 10-50 kg.
• 70ml/kg/min FGF in patients more than 60kg.
• Tidal volume to be set at 10ml/kg.
• Respiratory rate at 12-16 breaths/min.
MAPLESON D SYSTEM
 The higher the FGF, the lower the End-Tidal Co2.
 A series of “Curves” can be constructed when we
combine FGF, Minute Volume & Arterial Co2 levels.
 To produce a given PaCo2, an infinite number of
combinations of FGF & Minute Volume can be used.
 High FGF & low Minute Volumes or high Minute
Volumes & low FGF or combinations in between can
be used.
MINUTE VENTILATION
F
R
E
S
H
G
A
S
F
L
O
W
PaCO2
Almost same PaCO2- for FGF from
100 to 240ml/kg/min
MAPLESON D SYSTEM
 At the Left, with a “High” FGF, Circuit is a
nonrebreathing and end-tidal Co2 depends only on
Ventilation. Uneconomical, associated with lost heat and
humidity.
 End-Tidal Co2 depends on Minute Volume.
 On the Right, region of hyperventilation and partial
rebreathing is present.
 End-Tidal Co2 is regulated by adjusting the FGF.
MAPLESON D SYSTEM
 Lower FGF and increased rebreathing are associated
with :-
1) Higher humidity
2) Less heat loss
3) Greater fresh gas economy
 Hyperventilation can be used without inducing
hypocarbia.
 Individual differences in Dead Space: Tidal Volume are
minimized at high levels of Minute Volume. So, it is
useful to aim for the right side of the graph.
MAPLESON D SYSTEM
 In patients with Stiff Lungs, poor cardiac performance
or Hypovolemia , using the “Left” side of the graph and
a relatively small total ventilation with a high FGF may
be better.
 Formulas to predict FGF requirements have been based
on-
1)Body Weight
2)Minute Volume
3)Body Surface Area
 With Assisted Ventilation, the efficacy is intermediate
between that for spontaneous & controlled ventilation
MAPLESON D SYSTEM
BAIN SYSTEM HAZARDS :-
 Inner tube becomes detached from its connections at
either end or develops a leak at the machine end.
 FG supply tube becomes kinked or twisted.
 System is incorrectly assembled(standard corrugated
tubing).
 A defect in the metal head , FG & Exhaled gas mix.
The entire limb becomes Dead Space.
MAPLESON D SYSTEM
PREUSE CHECKS:-
 Tested for “Leaks” by occluding the patient end,
closing the APL Valve & pressurizing the system.
 APL Valve is then opened.
 Bag should deflate easily, if the valve & Scavenging
system are working properly.
 Either the user or a patient should breathe through the
system to detect obstructions
MAPLESON D SYSTEM
 To confirm the integrity of the inner tube in Bain
modification –
1) Performed by setting a low flow on the oxygen
flowmeter & occluding the inner tube(with a finger or
the barrel of a small syringe) at the patient end while
observing the flowmeter indicator. Intact inner tube &
correctly connected, indicator will fall.
2) Activating the oxygen flush & observing the bag. A
venturi effect caused by the high flow at the patient
end will create a negative pressure in the outer
exhalation tubing , bag deflate. Non-intact inner tube,
bag inflate slightly- PETHICK TEST
CPAP
CONTINUOUS POSITIVE AIRWAY PRESSURE :-
 One lung ventilation, using double lumen tube , a
modified Mapleson D System attached to the lumen,
leading to the nondependent lung is used to apply
CPAP to that lung.
 Configurations –
a) A source of oxygen is connected to the system.
b) APL Valve is set to maintain the desired pressure.
c) PEEP Valve, added to function as a high pressure
relief device.
MAPLESON E SYSTEM
 Also k/a “T-Piece”.
 A length of tubing may be attached
to the T-Piece to form a reservoir.
 Does not have a bag.
 Expiratory port enclosed in a
chamber, excess gases are
evacuated.
 Sensor or sampling site for the
respiratory gas monitor may be
placed between the-
a) Expiratory port & the expiratory
tubing.
b) T-Piece & the patient.
MAPLESON E SYSTEM
 Modifications –
a) FG Inlet, extending inside the body of the T-Piece,
towards the patient connection to minimize Dead
Space.
b) A pressure-limiting device
 Its uses has decreased because of the difficulty in
Scavenging excess gases.
 Commonly used to administer oxygen or humidified
gas to patients breathing spontaneously.
 Used initially for pediatric patients undergoing palate
repair & intracranial surgery.
MAPLESON E SYSTEM
TECHNIQUES OF USE :-
1) SPONTANEOUS VENTILATION-
 Expiratory limb is open to the atmosphere.
2) CONTROLLED VENTILATION –
 Intermittently occluding the expiratory limb, allowing
the FGF to inflate the lings.
 Assisted respiration is difficult to perform.
MAPLESON E SYSTEM
FUNCTIONAL ANALYSIS –
 The presence or absence & the amount of Rebreathing
or Air Dilution will depend on the –
a) Fresh Gas flow
b) Patient’s Minute Volume
c) Volume of the exhalation limb
d) Type of Ventilation( Spontaneous or Controlled)
e) Respiratory Pattern
MAPLESON E SYSTEM
REBREATHING –
1) Sponatneous Ventilation –
 No rebreathing can occur,
if there is no exhalation
limb.
 Expiratory limb, FGF
needed to prevent
rebreathing.
2) Controlled Ventilation –
 No rebreathing because
only FG will inflate the
lungs
MAPLESON E SYSTEM
AIR DILUTION –
1) Controlled Ventilation-
 No air dilution
2) Spontaneous Ventilation-
 Volume of the tubing > patient’s Tidal Volume, no air
dilution occur.
 Air dilution can be prevented by providing a FGF that
exceeds the peak inspiratory flow rate( normally 3-5
times the Minute Volume) , if expiratory limb is absent
or if the volume of the limb < patient’s Tidal Volume.
MAPLESON E SYSTEM
 Air Dilution can be prevented by
FGF of two times Minute
Volume & a reservoir volume ,
one-third of the Tidal Volume.
HAZARDS –
 Controlling Ventilation by
intermittently occluding the
expiratory limb may lead to –
1) Overinflation
2) Barotrauma
 The Pressure-Buffering effect
of the bag is absent.
 No APL Valve to moderate the
pressure in the lungs.
MAPLESON F SYSTEM
 Also k/a Jackson-Rees, Rees,
Jackson-Rees modification of
the T-Piece
 Has a bag with a mechanism
for venting excess gases, like
a “Hole” in the tail or side of
the bag , occluded by using a
finger to provide pressure.
 May be fitted with a device to
prevent the bag from
collapsing and allowing
excess gases to escape.
 An anesthesia Ventilator may
be used in place of the bag.
JACKSON REES MODIFICATION
MAPLESON F WITH APL VALVE
MAPLESON F SYSTEM
 An APL Valve near the patient connection to provide
protection from high pressure.
 Scavenging can be performed by -
A) Enclosing the bag in a chamber from which waste
gases are suctioned.
B) Attaching various devices to the relief mechanism in
the bag.
MAPLESON F SYSTEM
TECHNIQUES OF USE –
1) SPONTANEOUS RESPIRATION-
 The relief mechanism is left fully open.
2) ASSISTED/CONTROLLED RESPIRATION-
 The relief mechanism is occluded sufficiently to
distend the bag.
 Respiration can be controlled or assisted by squeezing
the bag.
 Inspiration, Hole in the bag can be occluded by the
user’s finger.
MAPLESON F SYSTEM
3) MECHANICAL VENTILATION –
 Bag is replaced by the hose from a ventilator.
 HME can be used either by –
a) Inserting it between the patient & the T-Piece
b) Using the gas sampling port on the HME, as the FG
inlet.
 Spontaneous respiration, most of the FG being vented
from the distal end of the expiratory limb. To prevent
this –
a) Expiratory limb can be partially or totally occluded
b) FG flow is increased + HME not used
MAPLESON F SYSTEM
FUNCTIONALANALYSIS-
 Functions much like the Mapleson D System.
 Flows required to prevent rebreathing during spontaneous
& controlled respiration are the same as those required with
the Mapleson D system.
 Less Work Of Breathing.
 Controlled Ventilation, PEEP does not affect end-tidal Co2.
 Spontaneous breathing, PEEP increases end-tidal Co2 ,
FGF are less than three times minute volume.
 PEEP should not be applied by using an undwerwater seal.
MAPLESON F SYSTEM
 HME , during an inhalation induction – Increaed
resistance will result in more of the FGF entering the
expiratory limb & delaying induction.
HAZARDS –
 Because of a bag in the system, excessive pressure is
less likely to develop.
 Same as those described for the Mapleson E System.
 If a ventilator that uses a ram of oxygen to produce
inspiration is used, a disconnection at the common gas
outlet may not be detected by an airway pressure
monitor due to the high resistance of the FG tubing.
Mapleson Systems
What FGF’s are needed?
Mapleson Systems Uses FGF SV FGF IPPV
A Magill
Lack
Spontaneous
Gen Anaesthesia
70-100 ml/kg/min Min 3 x MV
B Very uncommon,
not in use today
C Resuscitation
Bagging
Min 15/pm
D Bain Spontaneous
IPPV, Gen. Anaes
150-200
ml/kg/min
70-100 ml/kg/min
E Ayres T Piece Very uncommon,
not in use today
F Jackson Rees Paediatric
<25 Kg
2.5 – 3 x MV
Min 4 /pm
 For spontaneous ventilation in the order of efficiency –
ADCB (All Dogs Can Bite).
 For controlled ventilation – DBCA (Dead Bodies
Can’t Argue)
 Here D includes E, F and Bain`s system
RESPIRATORY GAS MONITORING
WITH THE MAPLESON SYSTEMS
 All of the Mapleson systems except the A System have
the FG inlet near the patient connection port, make it
difficult to get a reliable sample of exhaled gases.
 Four Sampling sites at the –
1) Junction of the breathing system & elbow connector
2) The corner of the elbow connector
3) 2 cm distal in the elbow connector
4) The Tracheal tube connector
RESPIRATORY GAS MONITORING
WITH THE MAPLESON SYSTEMS
 If sampling were carried out at the two sites, closest to
the patient, values were accurate.
 Significant errors were noted when –
a) Samples were taken from the corner of the elbow
connector, only if high FGF was used.
b) Sampling was performed at the junction of the
breathing system & elbow connector, only if low FGF
were used.
 A cannula that projects into the airway can be used to
improve sampling.
RESPIRATORY GAS MONITORING
WITH THE MAPLESON SYSTEMS
 In infants and children –
a) Sampling at the junction of the tracheal tube and
breathing system resulted in “falsely low end-tidal
Co2” values in patients weighing less than 8kg.
b) The accuracy of measurements can be improved by
inserting a small HME between the breathing system
and the tracheal tube connector.
ADVANTAGES
 Simple, Inexpensive and rugged.
 No moving parts except the APL Valve.
 Components are easy to disassemble and can be
disinfected or sterilized in a variety of ways.
 A popular choice to provide positive pressure
ventilation in emergencies.
 Variations in Minute Volume affect end-tidal Co2.
 In coaxial systems(Lack, Bain), the inspiratory limb is
heated by the warm exhaled gas in the coaxial
expiratory tubing.
ADVANTAGES
 Resistance is usually low at flows .
 Work Of Breathing during spontaneous ventilation is
less but not always. Sometimes WOB increases, if the
APL Valve is not oriented properly.
 Lightweight systems and not bulky.
 Not cause “Drag” on the mask or Tracheal tube or
accidental extubation.
 Easy to position conveniently.
 A long Mapleson D System with an aluminium APL
Valve may be used to ventilate a patient in the MRI
Unit.
ADVANTAGES
 Compression and Compliance volume losses are less.
 Changes in FG concentrations result in rapid changes in
Inspiratory gas composition.
 No Co2 absorbent, no production of possibly toxic
products such as Carbon Mononoxide and Compound
A.
DISADVANTAGES
 Require high gas flows .
 Higher costs, increased atmospheric pollution and
difficulty assessing spontaneous ventilation.
 Inspired heat and humidity tend to be low, unless a
humidification device is used.
 Optimum FGF may be difficult to determine.
 Necessary to change the flow when changing from
spontaneous to controlled ventilation or vice versa.
 Anything that causes FGF to be lowered, presents a
hazard, because rebreathing may occur.
DISADVANTAGES
 APL Valve is located close to the patient in Mapleson
A,B and C Systems ; inaccessible to the user.
Scavenging is awkward. Can be overcome by using the
Lack modification of the Mapleson A.
 Mapleson E and F Systems are difficult to scavenge &
Air Dilution can occur with the Mapleson E System.
 Not suitable for patients with malignant hyperthermia
because FGF is not enough to remove the increased
Co2 Load.
THANK YOU
.

Mapleson system

  • 1.
    MAPLESON SYSTEM Dr. PadmajaPallavi Pandey ERA’s Lucknow Medical College & Hospital 2014 Batch
  • 2.
    CONTENTS  History ofBreathing System  Characteristics  Classification  Components  Lacks  Mapleson A System  Configurations  Classic form  Lack Modification
  • 3.
    CONTENTS  Techniques ofUse  Functional Analysis  Spontaneous Respiration  Controlled or Assisted Ventilation  Hazards  Preuse Checks  Mapleson B System  Techniques of Use
  • 4.
    CONTENTS  Functional Analysis Spontaneous Respiration  Controlled or Assisted Ventilation  Mapleson C System  Techniques of Use  Functional Analysis  Mapleson D System  Configurations  Classic form
  • 5.
    CONTENTS  Bain Modification Techniques of Use  Functional Analysis  Spontaneous Breathing  Controlled Ventilation  Bain System Hazards  Preuse Checks  Continuous Positive Airway Pressure
  • 6.
    CONTENTS  Mapleson ESystem  Techniques of Use  Functional Analysis  Rebreathing  Air Dilution  Hazards  Mapleson F System  Techniques of Use  Functional Analysis
  • 7.
    CONTENTS  Hazards  RespiratoryGas Monitoring with the Mapleson Systems  Advantages of the Mapleson Systems  Disadvantages of the Mapleson System
  • 8.
    “NECESSITY IS MOTHEROF INVENTION”  Earlier circuits were simple, differing in the type of anesthetic agent administered.  The purpose of breathing systems that have evolved in anesthetic practice is to deliver Gas & Vapor to the patient in an appropriate, controlled & efficient manner.
  • 9.
     1846 SirW.T.G Morton did public demonstration with Ether.
  • 10.
     1876 Clover`s Inhaler developed by J.T Clover.
  • 11.
     1907 Barthused it to administer N₂O.  1909 Teter`s apparatus developed.  1909-13 F.W.Hewitts developed Hewitt`s apparatus.
  • 12.
     1913 Gwathemy Apparatusdeveloped.  1917 Boyle`s Apparatus developed.  1928 Magill`s Circuit was developed.  1937 Philip Ayre introduced T piece.
  • 13.
    HISTORY  1972 J.ABain & W.E Spoerel introduced Bain`s Circuit.  1975 Dr Gordon Jackson Rees developed Mapleson F system.  Humphrey Davy, Brock & Downing  developed combined ADE system.
  • 14.
    History In 1954, Prof.WW Mapleson from University of Wales, Cardiff, classified the several breathing systems around depending on what components they contained and what position they took in the system, Analyzed Five different anaesthetic breathing systems & referred as Mapleson A – E. In 1975 Willis et al described F system & added to above. In 1976 – Lack circuit .
  • 15.
  • 16.
  • 17.
    CHARACTERISTICS  Absence ofunidirectional valves to direct gases to or from the patient.  FGF must wash Co2 out of the circuit, because there is no device for absorbing Co2. So, known as “Co2 Washout Circuits or Flow-Controlled Breathing Systems”.  No clear seperation of Inspired & Expired gases, therefore, Rebreathing will occur, Inspiratory flow > FGF.
  • 18.
    CHARACTERISTICS  Composition ofthe inspired mixture will depend on how much Rebreathing takes place.  Monitoring End-Tidal Carbon-Dioxide is the best method to determine the optimal FGF.  With Rebreathing , arterial Carbon-Dioxide to End- Tidal Carbon-Dioxide gradient decreases.
  • 19.
    CLASSIFICATION Six types :- Mapleson A – Magill & Lack System  Mapleson B  Mapleson C Bagging System  Mapleson D or Bain System  Mapleson E  Mapleson F or Jackson-Rees modification of the T- Piece
  • 21.
    COMPONENTS  Reservoir Bag Corrugated Tubing  APL Valve  Fresh gas inlet  Patient Connection
  • 22.
    LACKS  Carbon-Dioxide absorbers Unidirectional Valves  Separate Inspiratory & Expiratory limbs.
  • 23.
    For better understandingof functional analysis they have been classified as - 1) Afferent Reservoir System (ARS) 2) Enclosed Afferent Reservoir System 3) Efferent Reservoir System 4) Combined System The efficiency of a system is determined in terms of CO₂ elimination & FGF utilization.
  • 24.
    • Afferent limbis that part of the breathing system which delivers the fresh gas from the machine to the patient. • If the reservoir is placed in this limb as in Mapleson A, B, C and Lack’s systems they are called as afferent reservoir system. • Efferent limb is that part of the breathing system which carries the expired gas from the patient and vents it to the atmosphere through the expiratory valve/port. • If the reservoir is placed in this limb as in Mapleson D, E, F and Bain systems they are called efferent reservoir system
  • 25.
    Mapleson postulates (1954) Mapleson has analyzed these bi-directional flow systems & few basic assumptions have been made which are of historical interest.  Gases move En-bloc i.e they maintain their identity as fresh gas, dead space gas & alveolar gas. There is no mixing of these gases.
  • 26.
    • Reservoir bagscontinues to fill up, without offering any resistance till it is full. • The expiratory valve opens as soon as the reservoir bag is full & pressure inside the system goes above the atmospheric pressure. • The valve remains open throughout the expiratory phase without offering any resistance to gas flow & closes at the start of next inspiration.
  • 27.
    MAPLESON A SYSTEM •Originally described by Evan Magill. • Length of breathing tube  110-180 cms. • FGF  from machine end. • APL  close to patient. • Sampling ports to be placed between APL valve & the tube.
  • 28.
    MAPLESON A SYSTEM CONFIGURATIONS:- 1) CLASSIC FORM –  Also k/a Magill attachment or system  Fresh gas does not enter the system near the patient connection but enters at the other end of the system near the reservoir bag.  A corrugated tubing connects the bag to the adjustable pressure limiting(APL) Valve at the patient end of the system.
  • 29.
  • 30.
  • 31.
    MAPLESON A SYSTEM A sensor for a nondiverting respiratory gas monitor or the sampling site for a diverting monitor may be placed between the :- a) APL Valve & the Corrugated tubing. b) APL Valve & the patient - result in excessive dead space in small patients. c) Neck of the bag & its mount d) Bag & the Corrugated tubing. e) In the fresh gas supply tube.
  • 32.
    MAPLESON A SYSTEM 2)LACK MODIFICATION –  Has an added “Expiratory” Limb, which runs from the patient connection to the APL Valve at the machine end of the system, makes it easier to adjust the valve & facilitates scavenging excess gases, increases the work of breathing slightly.  Available in both-  Dual (Parallel) tube arrangement  Tube-Within-a-tube (Coaxial) Configuration, the Expiratory limb runs concentrically inside the outer inspiratory limb.
  • 33.
    Lack system  Co-axialMapleson A.  Outer tube 30mm in diameter.  Inner tube 14mm in diameter.  APL valve placed near machine end.
  • 34.
  • 35.
  • 36.
    MAPLESON A SYSTEM TECHNIQUESOF USE :- SPONTANEOUS VENTILATION –  APL Valve is kept in the fully open position.  During late phase of exhalation, excess gas exits through it.
  • 37.
    MAPLESON A SYSTEM CONTROLLEDOR ASSISTED VENTILATION –  Intermittent positive pressure is applied to the bag.  The APL Valve is partially closed so that when the bag is squeezed, sufficient pressure to inflate the lungs is achieved.  The APL Valve opens during inspiration.
  • 38.
    MAPLESON A SYSTEM FUNCTIONALANALYSIS :- 1) SPONTANEOUS RESPIRATION –  Patient exhales-  First “Dead Space Gas” & then “Alveolar Gases” flow into the Corrugated tubing toward the bag  “Fresh Gas” flows into the bag.  Bag is full , pressure in the system rises until the APL Valve opens.  First gas vented will be “Alveolar Gas”.
  • 39.
     Spontaneous Breathing 3 phasesidentified   Inspiratory  Expiratory  Expiratory Pause. FGF Dead space gas Alveolar gas
  • 40.
  • 41.
    MAPLESON A SYSTEM Remainder of exhalation i.e. only “Alveolar Gas”, exhausts through the open APL Valve.  In the Corrugated tubing, the continuing inflow of “Fresh Gas” reverses the flow of exhaled gases.  Some alveolar gas that bypassed the APL Valve, returns back & exits through it.  FGF is :- High –Force the dead space gas out. Intermediate –Some dead space gas will be retained in the system. Low –More alveolar gas will be retained.
  • 42.
    MAPLESON A SYSTEM Start of Inspiration-  First gas inhaled will be from “Dead Space Gas” between the patient & the APL Valve.  Next gas will be either a) Alveolar Gas (if the FGF is low) b) Dead Space Gas (if the FGF is intermediate) c) Fresh Gas (if the FGF is high)  Changes in respiratory pattern have little effect on Rebreathing.
  • 43.
    MAPLESON A SYSTEM Rebreathing begins when the FGF is reduced to 56- 82 ml/kg/min or 58% - 83% of minute volume.  FGF to 51-85 ml/kg/min & 42% - 88% of minute volume have been recommended to avoid rebreathing.  .  . .
  • 44.
  • 45.
    MAPLESON A SYSTEM 2)CONTROLLED OR ASSISTED VENTILATION –  The pattern of gas flow changes.  Exhalation-  Pressure in the system will remain low  No gas will escape through the APL Valve, unless the bag becomes distended.  Exhaled gases i.e. both “Dead Space Gas” & “Alveolar Gas” remain in the Corrugated tubing.  “Alveolar Gas” is nearest to the patient.
  • 46.
  • 47.
    MAPLESON A -CONTROLLED
  • 48.
    MAPLESON A SYSTEM Some “Alveolar Gas” may enter the bag, if the Tidal Volume is large.  Early Inspiration –  Gases in the tubing flow to the patient.  “Alveolar Gas” will be inhaled first because it occupies the space nearest to the patient.  As the pressure in the system rises, the APL Valve opens, so that gas, both exits through the APL Valve & flows to the patient.
  • 49.
    MAPLESON A SYSTEM Late Inspiration –  All the exhaled gas has been driven from the tube.  “Fresh Gas” fills the tubing.  Some “Fresh Gas” enters the patient & some is vented through the valve.  Rebreathing of “Alveolar Gases” & venting of “Fresh Gases”.  The composition of the inspired gas mixture depends on the respiratory pattern.
  • 50.
    MAPLESON A SYSTEM As the expiratory phase is prolonged, system becomes more efficient.  During Inspiration, if the APL Valve does not vent gas, the Mapleson A System can be as efficient as the Mapleson D System.  During Assisted Ventilation, it is less efficient than with Spontaneous Ventilation but is more efficient than with Controlled Ventilation.
  • 51.
    MAPLESON A SYSTEM HAZARDS:-  A Mechanical Ventilator that vents excess gases should not be used because the entire system then becomes “Dead Space”.
  • 52.
    TESTING FOR LEAKS/MAGILL PREUSECHECKS :-  Tested for leaks by occluding the patient end of the system, closing the APL Valve & pressurizing the system.  Opening the APL Valve will confirm proper functioning of that component.  The user or a patient should breathe through the system.
  • 53.
    TESTING FOR LEAKS/LACK The coaxial lack system requires additional testing to confirm the integrity of the inner tube – A) To attach a tracheal tube to the inner tubing at the patient end of the system. Blowing down the tube with the APL Valve closed will produce movement of the bag, if there is a leak between the two limbs. B) To occlude both limbs at the patient connection with the APL Valve open & then squeeze the bag. Leak in the inner limb, gas will escape through the APL Valve & the bag will collapse.
  • 54.
    MAPLESON B SYSTEM The fresh gas inlet & APL Valve are both located near the patient port.  The reservoir bag is at the patient end of the system, seperated from the fresh gas inlet by corrugated tubing.
  • 55.
  • 56.
    MAPLESON B SYSTEM FUNCTIONALANALYSIS :- 1) SPONTANEOUS RESPIRATION –  Exhalation  “Dead Space Gas” will pass down the corrugated tubing, along with “Fresh Gas”.  The tubing near the patient will be filled with “Fresh Gas” & some “Alveolar Gas”.  When the bag reaches full capacity, the APL Valve opens & both “Fresh Gas” and “Alveolar Gas” will exit from the system.
  • 57.
    MAPLESON B SYSTEM Inspiration –  The APL Valve closes  The patient inhales “Fresh Gas” and gas from the tubing.  No gas will be inhaled from the bag, if the volume of the tubing exceeds the tidal volume.  To avoid Rebreathing, the FGF must be equal to peak inspiratory flow rate (normally 20-25l/min)
  • 58.
    MAPLESON B SYSTEM A FGF more than double Minute Volume has been recommended, but flows as low as 0.8 -1.2 times Minute Volume may be sufficient. 2) CONTROLLED OR ASSISTED VENTILATION –  Behavior is similar to that of the Maplson A  During the EXPIRATORY PAUSE, “Fresh Gas” accumulates at the patient end of the tubing. So, more efficient.
  • 59.
    MAPLESON B SYSTEM The composition of the inspired gas is influenced by the ventilatory pattern  Has variable performance.  A FGF of 2-2.5 times Minute Volume has been recommended.
  • 60.
    MAPLESON C SYSTEM Identical to the Mapleson B System except that the Corrugated tubing is omitted. TECHNIQUES OF USE :- Similar to that described for the Mapleson B System. FUNCTIONAL ANALYSIS Behaves similarly to the Mapleson B System.
  • 61.
  • 62.
  • 63.
    MAPLESON C SYSTEM Also k/a “Westminster face piece”. 1) SPONTANEOUS VENTILATION –  When the EXPIRATORY PAUSE is minimal, almost as efficient as the Mapleson A but becomes less efficient as the EXPIRATORY PAUSE increases.  A FGFof 2 times Minute Volume has been recommended for spontaneous breathing. 2) CONTROLLED VENTILATION :- A FGF of 2-2.5 times Minute Volume is recommended.
  • 64.
  • 65.
    MAPLESON D SYSTEM Have a T-piece near the patient & function similarly.  The T-piece is a three way tubular connector with a – a) Patient connection port b) Fresh Gas port c) Port for connection to a corrugated tubing.  Popular because excess gas scavenging is relatively easy.  Most efficient during Controlled Ventilation
  • 66.
  • 67.
    MAPLESON D SYSTEM CONFIGURATION:- 1) CLASSIC FORM –  A length of tubing connects the T-piece at the patient end to the APL Valve and the reservoir bag adjacent to it.  The length of the tubing determines the distance the user can be from the patient but has minimal effects on Ventilation.
  • 68.
    MAPLESON D SYSTEM The sensor or sampling site for a respiratory gas monitor may be placed between the – a) Bag & its mount b) Corrugated tubing & the T-piece c) Corrugated tubing & the APL Valve d) T-piece & the patient
  • 69.
    MAPLESON D SYSTEM BIDIRECTIONALPEEP VALVE  Placed between the corrugated tubing & the APL Valve.  Permits PEEP to be administered during manual or mechanical ventilation.  Spontaneous breathing is impossible because when a negative pressure is applied, it closes. UNIDIRECTIONAL PEEP VALVE  Used at the bag attachment site by using a special connectors.  Allows PEEP to be applied during spontaneous or mechanical but not manual ventilation.
  • 70.
    MAPLESON D SYSTEM BIDIRECTIONALPEEP VALVE  Placed in the hose leading to anesthesia ventilator, thus effective only during Mechanical Ventilation. UNIDIRECTIONAL PEEP VALVE
  • 72.
    MAPLESON D SYSTEM 2)BAIN MODIFICATION –  The “Fresh Gas” supply tube runs coaxially inside the corrugated tubing, ends at the point where the “Fresh Gas” would enter if the classic Mapleson D form were used.  The outer tube is clear so that the inner tube can be inspected.  The outer tube is narrow.  Available with a metal head with channels drilled into it, provides a fixed position for the reservoir bag & APL Valve and attachment of corrugated tubing.
  • 73.
  • 74.
    Bain modification ofMapleson D system  Originally modified by Bain & Spoerel in 1972.  Is co-axial system.  Usual length is 180cm.  Outer tube   Diameter -22mm.  Carries exhaled gas.  Inner tube   Diameter-7mm.  Carries fresh gas.
  • 75.
    MAPLESON D SYSTEM Some heads also have a “Pressure Manometer”.  Long Version- used for remote anesthesia in MRI.  Static Compliance is increased with a reduction in peak inspiratory pressure & Tidal Volume with the same ventilator settings.  PEEP is increased.  Longer, increased resistance to spontaneous breathing.
  • 76.
  • 77.
    MAPLESON D SYSTEM TECHNIQUESOF USE :- SPONTANEOUS RESPIRATION –  APL Valve is left open.  Excess gases are vented during expiration. MANUALLY CONTROLLED/ASSISTED VENTILATION  Partial closure of APL Valve  Squeezing the bag.  Excess gases are vented during inspiration.
  • 78.
    MAPLESON D SYSTEM MECHANICALLYCONTROLLED VENTILATION –  Connecting the hose from a ventilator in place of the reservoir bag.  Closing the APL Valve.  Excess gases are vented through the ventilator spill valve.
  • 79.
    MAPLESON D SYSTEM FUNCTIONALANALYSIS :- 1) SPONTANEOUS BREATHING –  Exhalation-  Exhaled gases mix with fresh gas, move through the corrugated tube, towards the bag.  Bag has filled, gas exits via the APL Valve.  Expiratory Pause –  FG pushes exhaled gases down the corrugated tubing.
  • 80.
    MAPLESON D -Spontaneous respiration
  • 81.
    MAPLESON D SYSTEM Inspiration –  Patient will inhale gas from the FG inlet & the corrugated tubing.  FG flow is a)High- All the gases drawn from the corrugated tube will be FG. b)Low- Exhaled gas containing Co2 will be inhaled, ventilatory pattern will help to determine the amount of Rebreathing.
  • 82.
    MAPLESON D SYSTEM Factors that decreases Rebreathing –  High Inspiratory:Expiratory (I:E) time ratio  A slow rise in Inspiratory flow rate.  A low flow rate during the last part of exhalation.  A long Expiratory pause, having the greatest effect.  The End-Tidal Co2 will rise, when gas containing Co2 is inhaled.  The End-Tidal Co2 will ”fall” while inspired Co2 will increase, when patient’s spontaneous respiration increases.
  • 83.
    MAPLESON D SYSTEM End-Tidal Co2 reaches “Plateau” no matter how hard the patient works, the End-Tidal Co2 cannot be lowered further.  End-Tidal Co2 will “rise”, when the patient’s respiration is depressed.  End-Tidal Co2 depends on both the ratio of Minute Volume & FGF and their absolute values.  Expired Volume > FGF, end-tidal Co2 will be determined mainly by FGF.
  • 84.
    MAPLESON D SYSTEM FGF > Minute Volume, end-tidal Co2 will be determined mainly by Minute Volume.  Recommendations for FGF based on body weight vary from 100-300 ml/kg/min.  FGF should be 1.5 to 3.0 times the Minute Volume.  FGF = Total ventilation (Adequate).  FGF of 4000-4700 ml/m2/min have been recommended.
  • 85.
  • 86.
    MAPLESON D SYSTEM 2)CONTROLLED VENTILATION –  Exhalation-  Gases flow from the patient down to the corrugated tubing.  FG enters the tubing.  Expiratory Pause-  FGF continues, pushes exhaled gases down to the tubing.
  • 87.
  • 88.
    MAPLESON D SYSTEM Inspiration –  FG & gas from the corrugated tubing enter the patient.  FGF is “low”, exhaled gases may be inhaled.  Increased Rebreathing-  Inspiratory time is prolonged  Respiratory Rate is increased  Adding an Inspiratory Plateau
  • 89.
    MAPLESON D SYSTEM Decreased Rebreathing –  Long Expiratory Pause, FG can flush exhaled gases from the tubing.  FGF is “high”-  Less rebreathing  End-Tidal Co2 is determined mainly by Minute Ventilation, Tidal Volume, the volume of the expiratory limb & Expiratory Resistance.  Minute Volume > FGF , FGF is the main factor controlling Co2 elimination.
  • 90.
    Recommendations by Bain& Spoerel • 2L/min FGF in patients <10kg. • 3.5L/min FGF in patients between 10-50 kg. • 70ml/kg/min FGF in patients more than 60kg. • Tidal volume to be set at 10ml/kg. • Respiratory rate at 12-16 breaths/min.
  • 91.
    MAPLESON D SYSTEM The higher the FGF, the lower the End-Tidal Co2.  A series of “Curves” can be constructed when we combine FGF, Minute Volume & Arterial Co2 levels.  To produce a given PaCo2, an infinite number of combinations of FGF & Minute Volume can be used.  High FGF & low Minute Volumes or high Minute Volumes & low FGF or combinations in between can be used.
  • 92.
    MINUTE VENTILATION F R E S H G A S F L O W PaCO2 Almost samePaCO2- for FGF from 100 to 240ml/kg/min
  • 93.
    MAPLESON D SYSTEM At the Left, with a “High” FGF, Circuit is a nonrebreathing and end-tidal Co2 depends only on Ventilation. Uneconomical, associated with lost heat and humidity.  End-Tidal Co2 depends on Minute Volume.  On the Right, region of hyperventilation and partial rebreathing is present.  End-Tidal Co2 is regulated by adjusting the FGF.
  • 94.
    MAPLESON D SYSTEM Lower FGF and increased rebreathing are associated with :- 1) Higher humidity 2) Less heat loss 3) Greater fresh gas economy  Hyperventilation can be used without inducing hypocarbia.  Individual differences in Dead Space: Tidal Volume are minimized at high levels of Minute Volume. So, it is useful to aim for the right side of the graph.
  • 95.
    MAPLESON D SYSTEM In patients with Stiff Lungs, poor cardiac performance or Hypovolemia , using the “Left” side of the graph and a relatively small total ventilation with a high FGF may be better.  Formulas to predict FGF requirements have been based on- 1)Body Weight 2)Minute Volume 3)Body Surface Area  With Assisted Ventilation, the efficacy is intermediate between that for spontaneous & controlled ventilation
  • 96.
    MAPLESON D SYSTEM BAINSYSTEM HAZARDS :-  Inner tube becomes detached from its connections at either end or develops a leak at the machine end.  FG supply tube becomes kinked or twisted.  System is incorrectly assembled(standard corrugated tubing).  A defect in the metal head , FG & Exhaled gas mix. The entire limb becomes Dead Space.
  • 97.
    MAPLESON D SYSTEM PREUSECHECKS:-  Tested for “Leaks” by occluding the patient end, closing the APL Valve & pressurizing the system.  APL Valve is then opened.  Bag should deflate easily, if the valve & Scavenging system are working properly.  Either the user or a patient should breathe through the system to detect obstructions
  • 98.
    MAPLESON D SYSTEM To confirm the integrity of the inner tube in Bain modification – 1) Performed by setting a low flow on the oxygen flowmeter & occluding the inner tube(with a finger or the barrel of a small syringe) at the patient end while observing the flowmeter indicator. Intact inner tube & correctly connected, indicator will fall. 2) Activating the oxygen flush & observing the bag. A venturi effect caused by the high flow at the patient end will create a negative pressure in the outer exhalation tubing , bag deflate. Non-intact inner tube, bag inflate slightly- PETHICK TEST
  • 100.
    CPAP CONTINUOUS POSITIVE AIRWAYPRESSURE :-  One lung ventilation, using double lumen tube , a modified Mapleson D System attached to the lumen, leading to the nondependent lung is used to apply CPAP to that lung.  Configurations – a) A source of oxygen is connected to the system. b) APL Valve is set to maintain the desired pressure. c) PEEP Valve, added to function as a high pressure relief device.
  • 101.
    MAPLESON E SYSTEM Also k/a “T-Piece”.  A length of tubing may be attached to the T-Piece to form a reservoir.  Does not have a bag.  Expiratory port enclosed in a chamber, excess gases are evacuated.  Sensor or sampling site for the respiratory gas monitor may be placed between the- a) Expiratory port & the expiratory tubing. b) T-Piece & the patient.
  • 102.
    MAPLESON E SYSTEM Modifications – a) FG Inlet, extending inside the body of the T-Piece, towards the patient connection to minimize Dead Space. b) A pressure-limiting device  Its uses has decreased because of the difficulty in Scavenging excess gases.  Commonly used to administer oxygen or humidified gas to patients breathing spontaneously.  Used initially for pediatric patients undergoing palate repair & intracranial surgery.
  • 103.
    MAPLESON E SYSTEM TECHNIQUESOF USE :- 1) SPONTANEOUS VENTILATION-  Expiratory limb is open to the atmosphere. 2) CONTROLLED VENTILATION –  Intermittently occluding the expiratory limb, allowing the FGF to inflate the lings.  Assisted respiration is difficult to perform.
  • 104.
    MAPLESON E SYSTEM FUNCTIONALANALYSIS –  The presence or absence & the amount of Rebreathing or Air Dilution will depend on the – a) Fresh Gas flow b) Patient’s Minute Volume c) Volume of the exhalation limb d) Type of Ventilation( Spontaneous or Controlled) e) Respiratory Pattern
  • 105.
    MAPLESON E SYSTEM REBREATHING– 1) Sponatneous Ventilation –  No rebreathing can occur, if there is no exhalation limb.  Expiratory limb, FGF needed to prevent rebreathing. 2) Controlled Ventilation –  No rebreathing because only FG will inflate the lungs
  • 106.
    MAPLESON E SYSTEM AIRDILUTION – 1) Controlled Ventilation-  No air dilution 2) Spontaneous Ventilation-  Volume of the tubing > patient’s Tidal Volume, no air dilution occur.  Air dilution can be prevented by providing a FGF that exceeds the peak inspiratory flow rate( normally 3-5 times the Minute Volume) , if expiratory limb is absent or if the volume of the limb < patient’s Tidal Volume.
  • 107.
    MAPLESON E SYSTEM Air Dilution can be prevented by FGF of two times Minute Volume & a reservoir volume , one-third of the Tidal Volume. HAZARDS –  Controlling Ventilation by intermittently occluding the expiratory limb may lead to – 1) Overinflation 2) Barotrauma  The Pressure-Buffering effect of the bag is absent.  No APL Valve to moderate the pressure in the lungs.
  • 108.
    MAPLESON F SYSTEM Also k/a Jackson-Rees, Rees, Jackson-Rees modification of the T-Piece  Has a bag with a mechanism for venting excess gases, like a “Hole” in the tail or side of the bag , occluded by using a finger to provide pressure.  May be fitted with a device to prevent the bag from collapsing and allowing excess gases to escape.  An anesthesia Ventilator may be used in place of the bag.
  • 109.
  • 110.
    MAPLESON F WITHAPL VALVE
  • 111.
    MAPLESON F SYSTEM An APL Valve near the patient connection to provide protection from high pressure.  Scavenging can be performed by - A) Enclosing the bag in a chamber from which waste gases are suctioned. B) Attaching various devices to the relief mechanism in the bag.
  • 112.
    MAPLESON F SYSTEM TECHNIQUESOF USE – 1) SPONTANEOUS RESPIRATION-  The relief mechanism is left fully open. 2) ASSISTED/CONTROLLED RESPIRATION-  The relief mechanism is occluded sufficiently to distend the bag.  Respiration can be controlled or assisted by squeezing the bag.  Inspiration, Hole in the bag can be occluded by the user’s finger.
  • 113.
    MAPLESON F SYSTEM 3)MECHANICAL VENTILATION –  Bag is replaced by the hose from a ventilator.  HME can be used either by – a) Inserting it between the patient & the T-Piece b) Using the gas sampling port on the HME, as the FG inlet.  Spontaneous respiration, most of the FG being vented from the distal end of the expiratory limb. To prevent this – a) Expiratory limb can be partially or totally occluded b) FG flow is increased + HME not used
  • 114.
    MAPLESON F SYSTEM FUNCTIONALANALYSIS- Functions much like the Mapleson D System.  Flows required to prevent rebreathing during spontaneous & controlled respiration are the same as those required with the Mapleson D system.  Less Work Of Breathing.  Controlled Ventilation, PEEP does not affect end-tidal Co2.  Spontaneous breathing, PEEP increases end-tidal Co2 , FGF are less than three times minute volume.  PEEP should not be applied by using an undwerwater seal.
  • 115.
    MAPLESON F SYSTEM HME , during an inhalation induction – Increaed resistance will result in more of the FGF entering the expiratory limb & delaying induction. HAZARDS –  Because of a bag in the system, excessive pressure is less likely to develop.  Same as those described for the Mapleson E System.  If a ventilator that uses a ram of oxygen to produce inspiration is used, a disconnection at the common gas outlet may not be detected by an airway pressure monitor due to the high resistance of the FG tubing.
  • 116.
  • 118.
    What FGF’s areneeded? Mapleson Systems Uses FGF SV FGF IPPV A Magill Lack Spontaneous Gen Anaesthesia 70-100 ml/kg/min Min 3 x MV B Very uncommon, not in use today C Resuscitation Bagging Min 15/pm D Bain Spontaneous IPPV, Gen. Anaes 150-200 ml/kg/min 70-100 ml/kg/min E Ayres T Piece Very uncommon, not in use today F Jackson Rees Paediatric <25 Kg 2.5 – 3 x MV Min 4 /pm
  • 119.
     For spontaneousventilation in the order of efficiency – ADCB (All Dogs Can Bite).  For controlled ventilation – DBCA (Dead Bodies Can’t Argue)  Here D includes E, F and Bain`s system
  • 120.
    RESPIRATORY GAS MONITORING WITHTHE MAPLESON SYSTEMS  All of the Mapleson systems except the A System have the FG inlet near the patient connection port, make it difficult to get a reliable sample of exhaled gases.  Four Sampling sites at the – 1) Junction of the breathing system & elbow connector 2) The corner of the elbow connector 3) 2 cm distal in the elbow connector 4) The Tracheal tube connector
  • 121.
    RESPIRATORY GAS MONITORING WITHTHE MAPLESON SYSTEMS  If sampling were carried out at the two sites, closest to the patient, values were accurate.  Significant errors were noted when – a) Samples were taken from the corner of the elbow connector, only if high FGF was used. b) Sampling was performed at the junction of the breathing system & elbow connector, only if low FGF were used.  A cannula that projects into the airway can be used to improve sampling.
  • 122.
    RESPIRATORY GAS MONITORING WITHTHE MAPLESON SYSTEMS  In infants and children – a) Sampling at the junction of the tracheal tube and breathing system resulted in “falsely low end-tidal Co2” values in patients weighing less than 8kg. b) The accuracy of measurements can be improved by inserting a small HME between the breathing system and the tracheal tube connector.
  • 123.
    ADVANTAGES  Simple, Inexpensiveand rugged.  No moving parts except the APL Valve.  Components are easy to disassemble and can be disinfected or sterilized in a variety of ways.  A popular choice to provide positive pressure ventilation in emergencies.  Variations in Minute Volume affect end-tidal Co2.  In coaxial systems(Lack, Bain), the inspiratory limb is heated by the warm exhaled gas in the coaxial expiratory tubing.
  • 124.
    ADVANTAGES  Resistance isusually low at flows .  Work Of Breathing during spontaneous ventilation is less but not always. Sometimes WOB increases, if the APL Valve is not oriented properly.  Lightweight systems and not bulky.  Not cause “Drag” on the mask or Tracheal tube or accidental extubation.  Easy to position conveniently.  A long Mapleson D System with an aluminium APL Valve may be used to ventilate a patient in the MRI Unit.
  • 125.
    ADVANTAGES  Compression andCompliance volume losses are less.  Changes in FG concentrations result in rapid changes in Inspiratory gas composition.  No Co2 absorbent, no production of possibly toxic products such as Carbon Mononoxide and Compound A.
  • 126.
    DISADVANTAGES  Require highgas flows .  Higher costs, increased atmospheric pollution and difficulty assessing spontaneous ventilation.  Inspired heat and humidity tend to be low, unless a humidification device is used.  Optimum FGF may be difficult to determine.  Necessary to change the flow when changing from spontaneous to controlled ventilation or vice versa.  Anything that causes FGF to be lowered, presents a hazard, because rebreathing may occur.
  • 127.
    DISADVANTAGES  APL Valveis located close to the patient in Mapleson A,B and C Systems ; inaccessible to the user. Scavenging is awkward. Can be overcome by using the Lack modification of the Mapleson A.  Mapleson E and F Systems are difficult to scavenge & Air Dilution can occur with the Mapleson E System.  Not suitable for patients with malignant hyperthermia because FGF is not enough to remove the increased Co2 Load.
  • 128.