SlideShare a Scribd company logo
Mammographic Equipment and
its Advancement
Yashawant ku. Yadav
B.Sc MIT 2ND YEAR
NAMS (Bir Hospital)
Abstract / Outlines
•Introduction of Equipment
•Components of mammographic equipment
•Mammographic Cassette
•Advancement in mammography
•References
Introduction
The goal of mammography is the
early detection of breast cancer,
typically through detection of
characteristic masses or
microcalcifications.
Basic projection :- CC and MLO
Risk factors
• Age
• Hormonal History
• Family History
• Genetics presence of BRCA1 or, BRCA2 gene
• Breast architecture (density)
• Menstruation (before 12yr)
• Menopause (after 55 yr)
• Late age of birth of first child or no children
• Never breast fed
Anatomy
• Breast is composed of fatty tissue,
glandular tissue, and connective tissue.
• Normal and cancerous tissues in the
breast have small x-ray attenuation
differences between them
• Need x-ray equipment specifically
designed to optimize breast cancer
detection
• Younger breasts are make imaging difficult because it is highly dense
due glandular tissue , and older breasts are more fatty so easier to
imaging .
• Mass density and atomic number of soft tissue of breast are similar
that’s why conventional radiography is useless. (????)
• At 70-100 kVp Compton scattering is predominant than photoelectric
effect which cause low differential absorption with in soft tissue finally
the contrast and resolution is decreased.
• Low kVp is used to maximize the PE effect which enhance the
differential absorption and improve the contrast resolution .
Best differential between the tissues is obtained at low x-ray
energies
Mammography equipment
• High contrast sensitivity
• high resolution
• low dose
Components of equipment (dedicated)
• Generator
• Support system
• X-ray tube
• Collimator / cone
• Breast compressor
• Anti scatter grids
• Cassette holder
• Automatic exposure control
• Photo timers
• Detector systems
Generator
• In mammography we are dealing with soft
tissue of breast.
• 20-40 kvp of x ray energy is required for good
contrast.
• Max mA-200mA
• For homogenous x-ray beam, HFG is
required.
• 3 -10 kw tube rating.
• 3-5 lakh HU-heat loading capacity
Patient supports
• Screen
• Lead apron
• Seating table
• Holding support
• Free space
X –ray tube
Anode or target materials
• Molybdenum
• Rhodium
• Tungsten
Characteristic radiation produced by Mo & Rh are effective for breast
imaging.
Molybdenum is the most common anode target material used in
mammography x-ray tubes.
With digital detectors, W is becoming the target of choice.
Cathode
The mammography x-ray tube is configured with dual filaments in the
focusing cup to produce 0.3- and 0.1-mm focal spot sizes
• Mammographic x ray tubes have rotating anode.
• Anode angle- 0 to 16 degrees.
• Tube tilt - 6 -24 degree.
• Source-to-image receptor distance is about 60-65 cm. (80)
• Effective anode angle :- (the actual anode angle plus the physical
tube tilt) is at least 22-24 degrees for coverage of the 24 × 30-cm
field area.
• Be(beryllium) window: is used as it has
low atomic no.(4) and lower absorption of
less than 5 kVp x-rays.1mm of Be is
equivalent to 0.1mm Al material(inherent
filter)
• HEEL EFFECT:-Positioning the cathode
over the chest wall of the patient and the
anode over the anterior portion (nipple)
achieves better uniformity of
the transmitted x-rays through the breast .
• Orientation of the tube in this way also
decreases the equipment bulk near the
patient’s head.
•FOCAL SPOT SIZE :- 0.3 to 0.4
mm for contact mammography (breast
compressed against the grid and image
receptor) .
• 0.10 to 0.15 mm for magnification
imaging .
• It reduce geometric blurring so that
micro calcifications can be resolved.
Target and filter combination
• The filters used in mammography are based on the "k edge" principle
and attenuate the radiation above the k-edge energy of the specific filter
material, either molybdenum or rhodium
• Mo target with 0.03mm(30µm)Mo filter (42)
• Rh target with 0.025mm (25µm)Rh filter (45)
• W target with 0.05mm (50µm) Rh filter (74)
Molybdenum target with molybdenum filter
combinations are appropriate for breasts less than 5
cm thick
Molybdenum target with rhodium filters or rhodium
target with rhodium filters may be appropriate for
breasts greater than 5 cm thick
Collimator / cones
• Cones are required for proper direction of x ray beam, proper
collimation , for rejecting scatter radiation and protection of chest
wall and lungs from unnecessary radiation.
THREE DESIGN ARE CURRENTLY USED :-
Fixed aperture.
A set of interchangeable cones for various sizes of breast.
An internal set of collimating blades.
• for magnification and spot
compression studies, manually
adjusted shutters allow the x-ray field
to be more closely matched to the
volume being imaged.
Anti scattered grid
The anti scattered grid must be a moving grid.
Carbon fiber is the preferred interspace material.
 Focused grid, with a grid ratio of 3:1 or 5:1.
 Grid frequencies of 30-40 lines/cm.
• Use of such grid doesn’t compromise the spatial resolution but it does
increase the patient dose .
• Use of 4:1 ratio grid approximately double the pt. dose as compare to
non grid contact mammography
• A unique grid is developed specially for the mammography (HTC) high
transmission cellular grid. (crossed type )
• It reduce the scattered radiation in two direction, HTC grid has a copper
as a grid stripe and air as a interspace , physical dimension grid ratio
(3.8:1)
Compressor devices
• COMPRESSION PLATE: flat
surfaced plate used for uniform
compression.
1. In Conventional compression
plate anterior portion of breast is
not compressed properly.
2. Newer slanted S.O.F.T paddles
are now available for equal
compression to all areas of the
breast .
ADVANTAGE :-
 Provides immobilization of the part during exposure.
 Produces radiograph of uniform density.
 Reduces geometric unsharpness due to closer object to IR distance.
 Reduces scatter and improves contrast.
 Reduces tissue overlap and improves resolution.
 Enhances recognition of architectural distortion produced by tumors.
DISADVANTAGE – is only patient discomfort
SPOT COMPRESSION –it is used to magnify small area of breast for
better visualization.
Amount of compression is
between 25 and 40 pounds
of pressure
Image receptor
FILM –SCREEN COMBINATION
single-coated film –screen combination is commonly used today
(GADOLINIUM OXYSULPHITE) provide higher film contrast and
require significantly lower radiation exposure as compared to direct
exposure film.
• Currently in mammography single emulsion film are used in
combination with a single back screen.
• MAMMOGRAPHIC CASSETEE ;These are dedicated cassette used
for mammography only (18X24/ 24X30 cm)
Automatic exposure control
• Phototimers are design not only to measure x –ray intensity at image
receptor but also for x –ray quality .
• Phototimers are also called AEC device . And it is place after the image
receptor .
• and consists of a single ionization chamber or an array of three or more
semiconductor diodes.
• x-rays transmitted through the breast, antiscatter grid (if present), and
the image receptor generate a signal in the detector.
• The signal is accumulated (integrated) and, when the
accumulated signal reaches a preset value, the exposure is
terminated.
• The preset value corresponds to a specified signal-to-noise ratio
(SNR) in a digital mammography unit or an acceptable optical
density (OD) if a film-screen system is used.
Digital mammography
 Used with existing Film
Mammography unit
 Film cassettes are
replaced with CR
imaging plates with
photo-stimulable
phosphor (BaFBr:Eu2)
 Image reader or ADC
 Image Processor or the
computer system for
post -processing
Also called full-field
digital mammography
(FFDM),
The x-ray film is
replaced by solid-state
detectors (aSi or a
selenium) that convert x-
rays into electrical
signals.
The electrical signals are
used to produce images,
can be seen on a
computer or printed on
Special film
CR MAMMOGRAPHY DR MAMMOGRAPHY
DIRECT” conversion
• Amorphous selenium (direct conversion) using (TFT) flat panel
technology (Siemens, Hologic)
• ~70 micron pixels
“INDIRECT” conversion
• Scintillating phosphor (CsI columns) on an array of amorphous silicon
photodiodes using thin-film transistor (TFT) flat panel technology
• ~100 micron pixels
• Computed radiography
• ~50 micron pixels
• Slot scanning CCD
• ~27 micron pixels
Digital mammography
Pixel size
pixel size decreases:
• Spatial resolution improves
• Noise increase
• Signal-to-noise decreases
Advantages
• Post acquisition manipulation possible
• Enhances detection of anomalies
• Selectively increases magnification
• Distinguishes between malignant and benign lesions.
• High contrast resolution
• Potential in reduction of radiation dose.
• Storage and transfer of image possible.
• Lower image noise.
• Quicker than conventional mammography.
Advancement in digital mammography
1. Tomosynthesis
2. Photon counting
3. CAD computer aided detection
Tomosynthesis
• Digital Breast Tomosynthesis (DBT) is a 3D imaging technology that
acquires a series of low-dose projection images of the compressed or
uncompressed breast at different angles.
• 3D captures multiple slices of the breast, all at different angles. The
images are brought together to create crystal clear 3D reconstruction of
the breast.
• The radiologist is then able to review reconstruction, one slide at a
time, almost like turning pages in a book.
• This makes it easier for doctors to see if there’s anything to be
concerned about.
https://www.youtube.com/watch?v=EvpGVww4mzU
Benefits
• Minimal pressure is needed, just enough to hold the breast in place.
• A lower dose of radiation is required, up to 50% reduction for dense breasts.
• The cost of DTS is expected to drop below the cost a of traditional
mammogram.
• Tomosynthesis allow easier differentiation between benign and malignant
lesions
• Early detection of small breast cancers that may be hidden during 2D
mammography
• Greater accuracy in pinpointing size, shape and location of abnormalities
• Typical exposure parameters are 29 kVp and 44 mAs
(recorded during imaging of an American College of
Radiology phantom), which would result in a total radiation
dose of 1.45mGy to a breast thickness of 4.2 cm.
Photon counting
Photon-counting mammography is one of the
newer forms of X-ray-based mammography
technologies.
In 2003, the Micro Dose Mammography (MDM)
system was introduced by Sectra Mamea.
Photon-counting systems employ a slot-
scanning design in which the system performs a
lateral scan of the breast with a slot collimator.
The main principle is that smaller the detector
used, lower the risk that a scattered photon will
interact and increase image quality.
No grid is needed.
• The detector system consists of a large
number of crystalline silicon strip detectors, .
• Silicon has excellent electronic properties
and strip detectors are placed in edge-on
geometry with their long axis parallel to the
direction of the X-ray beam.
• In this way, a sufficient absorption length is
created, resulting in high quantum efficiency
for the full energy spectrum used in
mammography.
Technology
• The x-ray beam is collimated to a fan beam
matching the pre-collimator. It remove that photons
which are not initially directed towards the detector
• The pre-collimator transforms the beam into
several equidistant line beams.
• Beneath the breast support there is a detector box
containing a post-collimator and the x-ray detector.
Si-strip detectors matching the line beams exiting
the breast. The number of counted photons in each
channel is accumulated during a single sampling
interval before being sent to the workstation for
image reconstruction and display,
Computer Aided detection
• Radiologists misdiagnose 10-30% of the malignant cases
• Of the cases sent for surgical biopsy, only 10-20% are actually
malignant
• The CAD technology basically works like a second pair of eyes,
reviewing a patient's mammogram .
Can detect stage 1 cancers CAD systems
can assist
radiologists to
Reduce these
problems
Contd..
Highlights suspicious areas
• marks clusters of calcifications
with a small triangle and breast
masses with a star
• 100% sensitivity,
• decrease false negative
readings.
• Requires digital input.
aSi flat panel
X-ray Photon
Scintillator
Amplification
light
Capacitor charge
Storage
analog signal
A/D – Converter
Digital signal
X-ray Photon
aSe
X-ray into electrons
Electron
electron
Capacitor charge
Storage
analog signal
Analog signal
A/D – Converter
Digital signal
Photon Counting
X-ray Photon
MicroDose
Detector
Digital signal
5 (00000000000101)
Se flat panel
COMPARISON
49
Limitation of mammography
1. Not always certain between benign and malignant tumours.
2. Interpretations of mammograms can be difficult because a normal
breast can appear differently for each woman.
3. Appearance of an image may be compromised if there is powder or
cream on the breasts or post operative cases.
4. Not all types of cancers of the breast can be seen on mammography.
Other modalities of mammography
• Although mammography is the most sensitive exam available for
detecting small breast ca. False Negative Rate is 5-10% .
• USG
• DEDICATED BREAST CT
• BREAST MRI
Question for discussions
1. Why subject contrast is high in conventional radiography?
2. In conventional mammography intensifying screen is kept below the film
why?
3. Incidence of breast cancer is higher in upper lateral quadrant why?
4. What is double tack anode ?
5. What is the ratio of radiation induced breast cancer resulting from x ray
mammography ? Benefit VS risk
6. Can we perform mammogram with general x - ray ?

More Related Content

What's hot

Mobile & portable radiography
Mobile & portable radiographyMobile & portable radiography
Mobile & portable radiography
Priyanka Parimala
 
MRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pkMRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pk
Dr pradeep Kumar
 
Fluoroscopy
FluoroscopyFluoroscopy
Fluoroscopy
SUJAN KARKI
 
Post processing of computed tomography
Post processing of computed tomographyPost processing of computed tomography
Post processing of computed tomography
BeuniquewithNehaSing
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
Illyasmk
 
Contrast
ContrastContrast
Contrastmr_koky
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x ray
SushilPattar
 
Exposure factors2
Exposure factors2Exposure factors2
Exposure factors2mr_koky
 
Soft tissue radiography.pptx
Soft tissue radiography.pptxSoft tissue radiography.pptx
Soft tissue radiography.pptx
Ritupanta1
 
Gamma Camera
Gamma CameraGamma Camera
Gamma Camera
Muhammad Musaddiq
 
Mammography
MammographyMammography
Mammography
Bribin Bright
 
Image quality
Image qualityImage quality
Image quality
airwave12
 
CT RADIATION DOSE REDUCTION
CT RADIATION DOSE REDUCTION CT RADIATION DOSE REDUCTION
CT RADIATION DOSE REDUCTION
Ganesan Yogananthem
 
Faults in x ray tube and its care
Faults in x ray tube and its careFaults in x ray tube and its care
Faults in x ray tube and its care
SaruGosain
 
Anode heel effect, line focus principle,
Anode heel effect, line focus principle,Anode heel effect, line focus principle,
Anode heel effect, line focus principle,
InosRagan
 
CR & DR
CR & DRCR & DR
CT artifact
CT artifact CT artifact
CT artifact
Ganesan Yogananthem
 
Qc in xray
Qc in xrayQc in xray
Qc in xray
Anupam Niraula
 
CT Generation (Generation of CT)
CT Generation (Generation of CT)CT Generation (Generation of CT)
CT Generation (Generation of CT)
Upakar Paudel
 
Dual Energy CT
Dual Energy CTDual Energy CT
Dual Energy CT
Sudil Paudyal
 

What's hot (20)

Mobile & portable radiography
Mobile & portable radiographyMobile & portable radiography
Mobile & portable radiography
 
MRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pkMRI Coil and Gradient power-point slide pk
MRI Coil and Gradient power-point slide pk
 
Fluoroscopy
FluoroscopyFluoroscopy
Fluoroscopy
 
Post processing of computed tomography
Post processing of computed tomographyPost processing of computed tomography
Post processing of computed tomography
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
 
Contrast
ContrastContrast
Contrast
 
Beam restricted device and filter used in x ray
Beam restricted device and filter used in x rayBeam restricted device and filter used in x ray
Beam restricted device and filter used in x ray
 
Exposure factors2
Exposure factors2Exposure factors2
Exposure factors2
 
Soft tissue radiography.pptx
Soft tissue radiography.pptxSoft tissue radiography.pptx
Soft tissue radiography.pptx
 
Gamma Camera
Gamma CameraGamma Camera
Gamma Camera
 
Mammography
MammographyMammography
Mammography
 
Image quality
Image qualityImage quality
Image quality
 
CT RADIATION DOSE REDUCTION
CT RADIATION DOSE REDUCTION CT RADIATION DOSE REDUCTION
CT RADIATION DOSE REDUCTION
 
Faults in x ray tube and its care
Faults in x ray tube and its careFaults in x ray tube and its care
Faults in x ray tube and its care
 
Anode heel effect, line focus principle,
Anode heel effect, line focus principle,Anode heel effect, line focus principle,
Anode heel effect, line focus principle,
 
CR & DR
CR & DRCR & DR
CR & DR
 
CT artifact
CT artifact CT artifact
CT artifact
 
Qc in xray
Qc in xrayQc in xray
Qc in xray
 
CT Generation (Generation of CT)
CT Generation (Generation of CT)CT Generation (Generation of CT)
CT Generation (Generation of CT)
 
Dual Energy CT
Dual Energy CTDual Energy CT
Dual Energy CT
 

Similar to Mammographic equipment and its advancement

Mammography.pptx
Mammography.pptxMammography.pptx
Mammography.pptx
HeenaParveen57
 
Mammography
MammographyMammography
Mammography
Subodhraj Dhungana
 
MAMMOGRAPHIC EQUIPMENT.pptx//////////////
MAMMOGRAPHIC EQUIPMENT.pptx//////////////MAMMOGRAPHIC EQUIPMENT.pptx//////////////
MAMMOGRAPHIC EQUIPMENT.pptx//////////////
justinfan550
 
Mammography.pdf
Mammography.pdfMammography.pdf
Mammography.pdf
AfaqAhmed849251
 
FINAL MAMMO.pptx
FINAL MAMMO.pptxFINAL MAMMO.pptx
FINAL MAMMO.pptx
JayaAditya6
 
Breast mammography and tomosynthesis
Breast mammography and tomosynthesisBreast mammography and tomosynthesis
Breast mammography and tomosynthesis
shamimhyder161
 
25632789 01-basics1-advanced-mammo-system
25632789 01-basics1-advanced-mammo-system25632789 01-basics1-advanced-mammo-system
25632789 01-basics1-advanced-mammo-system
Klaus19
 
MAMMOGRAPHY.pptx
MAMMOGRAPHY.pptxMAMMOGRAPHY.pptx
MAMMOGRAPHY.pptx
rohanjohnjacob
 
Mammography
Mammography Mammography
Mammography
عبدالله فهد
 
Mammography -A ppt bt J K PATIL, Prof,dept of radiology
Mammography -A ppt bt J K PATIL, Prof,dept of radiologyMammography -A ppt bt J K PATIL, Prof,dept of radiology
Mammography -A ppt bt J K PATIL, Prof,dept of radiology
dypradio
 
Mammography-Physics-and-Technique-PDF.pdf
Mammography-Physics-and-Technique-PDF.pdfMammography-Physics-and-Technique-PDF.pdf
Mammography-Physics-and-Technique-PDF.pdf
UmaMaheswariC4
 
mammophysics.ppt
mammophysics.pptmammophysics.ppt
mammophysics.ppt
dido_pegio
 
allisonmammophysics2018.ppt
allisonmammophysics2018.pptallisonmammophysics2018.ppt
allisonmammophysics2018.ppt
PHCSADRAWADI
 
Radiotherapy Breast Cancer
Radiotherapy Breast CancerRadiotherapy Breast Cancer
Radiotherapy Breast Cancer
Mohamed Ali Morsy
 
Mammography
MammographyMammography
Mammography
Akari Kyaw
 
The innovation of breast scanning
The innovation of breast scanningThe innovation of breast scanning
The innovation of breast scanning
Ktn Aye
 
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfQUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
Angelina Protik
 
Digital breast tomosynthesis
Digital breast tomosynthesisDigital breast tomosynthesis
Digital breast tomosynthesis
MAMTA PANDA
 
Mammography and recent advances dr avinash
Mammography and recent advances dr avinashMammography and recent advances dr avinash
Mammography and recent advances dr avinash
AvinashDahatre
 

Similar to Mammographic equipment and its advancement (20)

Mammography.pptx
Mammography.pptxMammography.pptx
Mammography.pptx
 
Mammography
MammographyMammography
Mammography
 
MAMMOGRAPHIC EQUIPMENT.pptx//////////////
MAMMOGRAPHIC EQUIPMENT.pptx//////////////MAMMOGRAPHIC EQUIPMENT.pptx//////////////
MAMMOGRAPHIC EQUIPMENT.pptx//////////////
 
Mammography.pdf
Mammography.pdfMammography.pdf
Mammography.pdf
 
FINAL MAMMO.pptx
FINAL MAMMO.pptxFINAL MAMMO.pptx
FINAL MAMMO.pptx
 
Breast mammography and tomosynthesis
Breast mammography and tomosynthesisBreast mammography and tomosynthesis
Breast mammography and tomosynthesis
 
25632789 01-basics1-advanced-mammo-system
25632789 01-basics1-advanced-mammo-system25632789 01-basics1-advanced-mammo-system
25632789 01-basics1-advanced-mammo-system
 
MAMMOGRAPHY.pptx
MAMMOGRAPHY.pptxMAMMOGRAPHY.pptx
MAMMOGRAPHY.pptx
 
Mammography
Mammography Mammography
Mammography
 
Mammography -A ppt bt J K PATIL, Prof,dept of radiology
Mammography -A ppt bt J K PATIL, Prof,dept of radiologyMammography -A ppt bt J K PATIL, Prof,dept of radiology
Mammography -A ppt bt J K PATIL, Prof,dept of radiology
 
Mammography-Physics-and-Technique-PDF.pdf
Mammography-Physics-and-Technique-PDF.pdfMammography-Physics-and-Technique-PDF.pdf
Mammography-Physics-and-Technique-PDF.pdf
 
mammophysics.ppt
mammophysics.pptmammophysics.ppt
mammophysics.ppt
 
allisonmammophysics2018.ppt
allisonmammophysics2018.pptallisonmammophysics2018.ppt
allisonmammophysics2018.ppt
 
Radiotherapy Breast Cancer
Radiotherapy Breast CancerRadiotherapy Breast Cancer
Radiotherapy Breast Cancer
 
Mammography
MammographyMammography
Mammography
 
The innovation of breast scanning
The innovation of breast scanningThe innovation of breast scanning
The innovation of breast scanning
 
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfQUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
 
Mammography
Mammography Mammography
Mammography
 
Digital breast tomosynthesis
Digital breast tomosynthesisDigital breast tomosynthesis
Digital breast tomosynthesis
 
Mammography and recent advances dr avinash
Mammography and recent advances dr avinashMammography and recent advances dr avinash
Mammography and recent advances dr avinash
 

More from Yashawant Yadav

Mr procedure of pelvis and hip joint
Mr procedure of pelvis and hip jointMr procedure of pelvis and hip joint
Mr procedure of pelvis and hip joint
Yashawant Yadav
 
Positron emission tomography pet scan and its applications
Positron emission tomography pet scan and its applicationsPositron emission tomography pet scan and its applications
Positron emission tomography pet scan and its applications
Yashawant Yadav
 
Application of ultrasound
Application of ultrasoundApplication of ultrasound
Application of ultrasound
Yashawant Yadav
 
Ptc )percutaneous transhepatic cholangiography
Ptc )percutaneous transhepatic cholangiographyPtc )percutaneous transhepatic cholangiography
Ptc )percutaneous transhepatic cholangiography
Yashawant Yadav
 
Ct coronary angiography edited 1st
Ct   coronary  angiography edited 1stCt   coronary  angiography edited 1st
Ct coronary angiography edited 1st
Yashawant Yadav
 
Ct protocol for ivu
Ct protocol for ivuCt protocol for ivu
Ct protocol for ivu
Yashawant Yadav
 
Mri spin echo pulse sequences its variations and
Mri spin echo pulse sequences its variations andMri spin echo pulse sequences its variations and
Mri spin echo pulse sequences its variations and
Yashawant Yadav
 
Technique of thoracic and lumber spine radiography
Technique  of thoracic and lumber spine radiographyTechnique  of thoracic and lumber spine radiography
Technique of thoracic and lumber spine radiography
Yashawant Yadav
 
Digital breast tomosynthesis
Digital breast tomosynthesisDigital breast tomosynthesis
Digital breast tomosynthesis
Yashawant Yadav
 
Ptc )percutaneous transhepatic cholangiography
Ptc )percutaneous transhepatic cholangiographyPtc )percutaneous transhepatic cholangiography
Ptc )percutaneous transhepatic cholangiography
Yashawant Yadav
 
High voltage transformer and high frequency generator
High voltage transformer and high frequency generatorHigh voltage transformer and high frequency generator
High voltage transformer and high frequency generator
Yashawant Yadav
 
Radiological contrast media
Radiological contrast mediaRadiological contrast media
Radiological contrast media
Yashawant Yadav
 
Procedure of ercp and t tube cholangiography
Procedure of ercp and t tube cholangiographyProcedure of ercp and t tube cholangiography
Procedure of ercp and t tube cholangiography
Yashawant Yadav
 
Presentation1.pptx thoraccic and lumber spine
Presentation1.pptx thoraccic and lumber spinePresentation1.pptx thoraccic and lumber spine
Presentation1.pptx thoraccic and lumber spine
Yashawant Yadav
 
Technique of dental radiographic
Technique of dental radiographicTechnique of dental radiographic
Technique of dental radiographic
Yashawant Yadav
 
Presentation2.pptx technique chest 2
Presentation2.pptx technique chest 2Presentation2.pptx technique chest 2
Presentation2.pptx technique chest 2
Yashawant Yadav
 
Image characteristics of x ray film
Image characteristics of x ray filmImage characteristics of x ray film
Image characteristics of x ray film
Yashawant Yadav
 
Procedure of upper gi
Procedure of upper giProcedure of upper gi
Procedure of upper gi
Yashawant Yadav
 
Dacrocystography and sialography
Dacrocystography and sialographyDacrocystography and sialography
Dacrocystography and sialography
Yashawant Yadav
 
Ct image quality artifacts and it remedy
Ct image quality artifacts and it remedyCt image quality artifacts and it remedy
Ct image quality artifacts and it remedy
Yashawant Yadav
 

More from Yashawant Yadav (20)

Mr procedure of pelvis and hip joint
Mr procedure of pelvis and hip jointMr procedure of pelvis and hip joint
Mr procedure of pelvis and hip joint
 
Positron emission tomography pet scan and its applications
Positron emission tomography pet scan and its applicationsPositron emission tomography pet scan and its applications
Positron emission tomography pet scan and its applications
 
Application of ultrasound
Application of ultrasoundApplication of ultrasound
Application of ultrasound
 
Ptc )percutaneous transhepatic cholangiography
Ptc )percutaneous transhepatic cholangiographyPtc )percutaneous transhepatic cholangiography
Ptc )percutaneous transhepatic cholangiography
 
Ct coronary angiography edited 1st
Ct   coronary  angiography edited 1stCt   coronary  angiography edited 1st
Ct coronary angiography edited 1st
 
Ct protocol for ivu
Ct protocol for ivuCt protocol for ivu
Ct protocol for ivu
 
Mri spin echo pulse sequences its variations and
Mri spin echo pulse sequences its variations andMri spin echo pulse sequences its variations and
Mri spin echo pulse sequences its variations and
 
Technique of thoracic and lumber spine radiography
Technique  of thoracic and lumber spine radiographyTechnique  of thoracic and lumber spine radiography
Technique of thoracic and lumber spine radiography
 
Digital breast tomosynthesis
Digital breast tomosynthesisDigital breast tomosynthesis
Digital breast tomosynthesis
 
Ptc )percutaneous transhepatic cholangiography
Ptc )percutaneous transhepatic cholangiographyPtc )percutaneous transhepatic cholangiography
Ptc )percutaneous transhepatic cholangiography
 
High voltage transformer and high frequency generator
High voltage transformer and high frequency generatorHigh voltage transformer and high frequency generator
High voltage transformer and high frequency generator
 
Radiological contrast media
Radiological contrast mediaRadiological contrast media
Radiological contrast media
 
Procedure of ercp and t tube cholangiography
Procedure of ercp and t tube cholangiographyProcedure of ercp and t tube cholangiography
Procedure of ercp and t tube cholangiography
 
Presentation1.pptx thoraccic and lumber spine
Presentation1.pptx thoraccic and lumber spinePresentation1.pptx thoraccic and lumber spine
Presentation1.pptx thoraccic and lumber spine
 
Technique of dental radiographic
Technique of dental radiographicTechnique of dental radiographic
Technique of dental radiographic
 
Presentation2.pptx technique chest 2
Presentation2.pptx technique chest 2Presentation2.pptx technique chest 2
Presentation2.pptx technique chest 2
 
Image characteristics of x ray film
Image characteristics of x ray filmImage characteristics of x ray film
Image characteristics of x ray film
 
Procedure of upper gi
Procedure of upper giProcedure of upper gi
Procedure of upper gi
 
Dacrocystography and sialography
Dacrocystography and sialographyDacrocystography and sialography
Dacrocystography and sialography
 
Ct image quality artifacts and it remedy
Ct image quality artifacts and it remedyCt image quality artifacts and it remedy
Ct image quality artifacts and it remedy
 

Recently uploaded

The approach at University of Liverpool.pptx
The approach at University of Liverpool.pptxThe approach at University of Liverpool.pptx
The approach at University of Liverpool.pptx
Jisc
 
Introduction to Quality Improvement Essentials
Introduction to Quality Improvement EssentialsIntroduction to Quality Improvement Essentials
Introduction to Quality Improvement Essentials
Excellence Foundation for South Sudan
 
Operation Blue Star - Saka Neela Tara
Operation Blue Star   -  Saka Neela TaraOperation Blue Star   -  Saka Neela Tara
Operation Blue Star - Saka Neela Tara
Balvir Singh
 
CLASS 11 CBSE B.St Project AIDS TO TRADE - INSURANCE
CLASS 11 CBSE B.St Project AIDS TO TRADE - INSURANCECLASS 11 CBSE B.St Project AIDS TO TRADE - INSURANCE
CLASS 11 CBSE B.St Project AIDS TO TRADE - INSURANCE
BhavyaRajput3
 
Students, digital devices and success - Andreas Schleicher - 27 May 2024..pptx
Students, digital devices and success - Andreas Schleicher - 27 May 2024..pptxStudents, digital devices and success - Andreas Schleicher - 27 May 2024..pptx
Students, digital devices and success - Andreas Schleicher - 27 May 2024..pptx
EduSkills OECD
 
Unit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdfUnit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdf
Thiyagu K
 
How to Split Bills in the Odoo 17 POS Module
How to Split Bills in the Odoo 17 POS ModuleHow to Split Bills in the Odoo 17 POS Module
How to Split Bills in the Odoo 17 POS Module
Celine George
 
Ethnobotany and Ethnopharmacology ......
Ethnobotany and Ethnopharmacology ......Ethnobotany and Ethnopharmacology ......
Ethnobotany and Ethnopharmacology ......
Ashokrao Mane college of Pharmacy Peth-Vadgaon
 
The Art Pastor's Guide to Sabbath | Steve Thomason
The Art Pastor's Guide to Sabbath | Steve ThomasonThe Art Pastor's Guide to Sabbath | Steve Thomason
The Art Pastor's Guide to Sabbath | Steve Thomason
Steve Thomason
 
Language Across the Curriculm LAC B.Ed.
Language Across the  Curriculm LAC B.Ed.Language Across the  Curriculm LAC B.Ed.
Language Across the Curriculm LAC B.Ed.
Atul Kumar Singh
 
Model Attribute Check Company Auto Property
Model Attribute  Check Company Auto PropertyModel Attribute  Check Company Auto Property
Model Attribute Check Company Auto Property
Celine George
 
PART A. Introduction to Costumer Service
PART A. Introduction to Costumer ServicePART A. Introduction to Costumer Service
PART A. Introduction to Costumer Service
PedroFerreira53928
 
How to Break the cycle of negative Thoughts
How to Break the cycle of negative ThoughtsHow to Break the cycle of negative Thoughts
How to Break the cycle of negative Thoughts
Col Mukteshwar Prasad
 
How to Make a Field invisible in Odoo 17
How to Make a Field invisible in Odoo 17How to Make a Field invisible in Odoo 17
How to Make a Field invisible in Odoo 17
Celine George
 
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
siemaillard
 
1.4 modern child centered education - mahatma gandhi-2.pptx
1.4 modern child centered education - mahatma gandhi-2.pptx1.4 modern child centered education - mahatma gandhi-2.pptx
1.4 modern child centered education - mahatma gandhi-2.pptx
JosvitaDsouza2
 
Palestine last event orientationfvgnh .pptx
Palestine last event orientationfvgnh .pptxPalestine last event orientationfvgnh .pptx
Palestine last event orientationfvgnh .pptx
RaedMohamed3
 
Fish and Chips - have they had their chips
Fish and Chips - have they had their chipsFish and Chips - have they had their chips
Fish and Chips - have they had their chips
GeoBlogs
 
Mule 4.6 & Java 17 Upgrade | MuleSoft Mysore Meetup #46
Mule 4.6 & Java 17 Upgrade | MuleSoft Mysore Meetup #46Mule 4.6 & Java 17 Upgrade | MuleSoft Mysore Meetup #46
Mule 4.6 & Java 17 Upgrade | MuleSoft Mysore Meetup #46
MysoreMuleSoftMeetup
 
2024.06.01 Introducing a competency framework for languag learning materials ...
2024.06.01 Introducing a competency framework for languag learning materials ...2024.06.01 Introducing a competency framework for languag learning materials ...
2024.06.01 Introducing a competency framework for languag learning materials ...
Sandy Millin
 

Recently uploaded (20)

The approach at University of Liverpool.pptx
The approach at University of Liverpool.pptxThe approach at University of Liverpool.pptx
The approach at University of Liverpool.pptx
 
Introduction to Quality Improvement Essentials
Introduction to Quality Improvement EssentialsIntroduction to Quality Improvement Essentials
Introduction to Quality Improvement Essentials
 
Operation Blue Star - Saka Neela Tara
Operation Blue Star   -  Saka Neela TaraOperation Blue Star   -  Saka Neela Tara
Operation Blue Star - Saka Neela Tara
 
CLASS 11 CBSE B.St Project AIDS TO TRADE - INSURANCE
CLASS 11 CBSE B.St Project AIDS TO TRADE - INSURANCECLASS 11 CBSE B.St Project AIDS TO TRADE - INSURANCE
CLASS 11 CBSE B.St Project AIDS TO TRADE - INSURANCE
 
Students, digital devices and success - Andreas Schleicher - 27 May 2024..pptx
Students, digital devices and success - Andreas Schleicher - 27 May 2024..pptxStudents, digital devices and success - Andreas Schleicher - 27 May 2024..pptx
Students, digital devices and success - Andreas Schleicher - 27 May 2024..pptx
 
Unit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdfUnit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdf
 
How to Split Bills in the Odoo 17 POS Module
How to Split Bills in the Odoo 17 POS ModuleHow to Split Bills in the Odoo 17 POS Module
How to Split Bills in the Odoo 17 POS Module
 
Ethnobotany and Ethnopharmacology ......
Ethnobotany and Ethnopharmacology ......Ethnobotany and Ethnopharmacology ......
Ethnobotany and Ethnopharmacology ......
 
The Art Pastor's Guide to Sabbath | Steve Thomason
The Art Pastor's Guide to Sabbath | Steve ThomasonThe Art Pastor's Guide to Sabbath | Steve Thomason
The Art Pastor's Guide to Sabbath | Steve Thomason
 
Language Across the Curriculm LAC B.Ed.
Language Across the  Curriculm LAC B.Ed.Language Across the  Curriculm LAC B.Ed.
Language Across the Curriculm LAC B.Ed.
 
Model Attribute Check Company Auto Property
Model Attribute  Check Company Auto PropertyModel Attribute  Check Company Auto Property
Model Attribute Check Company Auto Property
 
PART A. Introduction to Costumer Service
PART A. Introduction to Costumer ServicePART A. Introduction to Costumer Service
PART A. Introduction to Costumer Service
 
How to Break the cycle of negative Thoughts
How to Break the cycle of negative ThoughtsHow to Break the cycle of negative Thoughts
How to Break the cycle of negative Thoughts
 
How to Make a Field invisible in Odoo 17
How to Make a Field invisible in Odoo 17How to Make a Field invisible in Odoo 17
How to Make a Field invisible in Odoo 17
 
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
 
1.4 modern child centered education - mahatma gandhi-2.pptx
1.4 modern child centered education - mahatma gandhi-2.pptx1.4 modern child centered education - mahatma gandhi-2.pptx
1.4 modern child centered education - mahatma gandhi-2.pptx
 
Palestine last event orientationfvgnh .pptx
Palestine last event orientationfvgnh .pptxPalestine last event orientationfvgnh .pptx
Palestine last event orientationfvgnh .pptx
 
Fish and Chips - have they had their chips
Fish and Chips - have they had their chipsFish and Chips - have they had their chips
Fish and Chips - have they had their chips
 
Mule 4.6 & Java 17 Upgrade | MuleSoft Mysore Meetup #46
Mule 4.6 & Java 17 Upgrade | MuleSoft Mysore Meetup #46Mule 4.6 & Java 17 Upgrade | MuleSoft Mysore Meetup #46
Mule 4.6 & Java 17 Upgrade | MuleSoft Mysore Meetup #46
 
2024.06.01 Introducing a competency framework for languag learning materials ...
2024.06.01 Introducing a competency framework for languag learning materials ...2024.06.01 Introducing a competency framework for languag learning materials ...
2024.06.01 Introducing a competency framework for languag learning materials ...
 

Mammographic equipment and its advancement

  • 1. Mammographic Equipment and its Advancement Yashawant ku. Yadav B.Sc MIT 2ND YEAR NAMS (Bir Hospital)
  • 2. Abstract / Outlines •Introduction of Equipment •Components of mammographic equipment •Mammographic Cassette •Advancement in mammography •References
  • 3.
  • 4. Introduction The goal of mammography is the early detection of breast cancer, typically through detection of characteristic masses or microcalcifications. Basic projection :- CC and MLO
  • 5. Risk factors • Age • Hormonal History • Family History • Genetics presence of BRCA1 or, BRCA2 gene • Breast architecture (density) • Menstruation (before 12yr) • Menopause (after 55 yr) • Late age of birth of first child or no children • Never breast fed
  • 6. Anatomy • Breast is composed of fatty tissue, glandular tissue, and connective tissue. • Normal and cancerous tissues in the breast have small x-ray attenuation differences between them • Need x-ray equipment specifically designed to optimize breast cancer detection
  • 7. • Younger breasts are make imaging difficult because it is highly dense due glandular tissue , and older breasts are more fatty so easier to imaging . • Mass density and atomic number of soft tissue of breast are similar that’s why conventional radiography is useless. (????) • At 70-100 kVp Compton scattering is predominant than photoelectric effect which cause low differential absorption with in soft tissue finally the contrast and resolution is decreased. • Low kVp is used to maximize the PE effect which enhance the differential absorption and improve the contrast resolution .
  • 8. Best differential between the tissues is obtained at low x-ray energies Mammography equipment • High contrast sensitivity • high resolution • low dose
  • 9. Components of equipment (dedicated) • Generator • Support system • X-ray tube • Collimator / cone • Breast compressor • Anti scatter grids • Cassette holder • Automatic exposure control • Photo timers • Detector systems
  • 10. Generator • In mammography we are dealing with soft tissue of breast. • 20-40 kvp of x ray energy is required for good contrast. • Max mA-200mA • For homogenous x-ray beam, HFG is required. • 3 -10 kw tube rating. • 3-5 lakh HU-heat loading capacity
  • 11. Patient supports • Screen • Lead apron • Seating table • Holding support • Free space
  • 12. X –ray tube Anode or target materials • Molybdenum • Rhodium • Tungsten Characteristic radiation produced by Mo & Rh are effective for breast imaging. Molybdenum is the most common anode target material used in mammography x-ray tubes. With digital detectors, W is becoming the target of choice.
  • 13. Cathode The mammography x-ray tube is configured with dual filaments in the focusing cup to produce 0.3- and 0.1-mm focal spot sizes • Mammographic x ray tubes have rotating anode. • Anode angle- 0 to 16 degrees. • Tube tilt - 6 -24 degree. • Source-to-image receptor distance is about 60-65 cm. (80) • Effective anode angle :- (the actual anode angle plus the physical tube tilt) is at least 22-24 degrees for coverage of the 24 × 30-cm field area.
  • 14.
  • 15. • Be(beryllium) window: is used as it has low atomic no.(4) and lower absorption of less than 5 kVp x-rays.1mm of Be is equivalent to 0.1mm Al material(inherent filter) • HEEL EFFECT:-Positioning the cathode over the chest wall of the patient and the anode over the anterior portion (nipple) achieves better uniformity of the transmitted x-rays through the breast . • Orientation of the tube in this way also decreases the equipment bulk near the patient’s head.
  • 16.
  • 17. •FOCAL SPOT SIZE :- 0.3 to 0.4 mm for contact mammography (breast compressed against the grid and image receptor) . • 0.10 to 0.15 mm for magnification imaging . • It reduce geometric blurring so that micro calcifications can be resolved.
  • 18. Target and filter combination • The filters used in mammography are based on the "k edge" principle and attenuate the radiation above the k-edge energy of the specific filter material, either molybdenum or rhodium • Mo target with 0.03mm(30µm)Mo filter (42) • Rh target with 0.025mm (25µm)Rh filter (45) • W target with 0.05mm (50µm) Rh filter (74)
  • 19.
  • 20.
  • 21.
  • 22. Molybdenum target with molybdenum filter combinations are appropriate for breasts less than 5 cm thick Molybdenum target with rhodium filters or rhodium target with rhodium filters may be appropriate for breasts greater than 5 cm thick
  • 23. Collimator / cones • Cones are required for proper direction of x ray beam, proper collimation , for rejecting scatter radiation and protection of chest wall and lungs from unnecessary radiation. THREE DESIGN ARE CURRENTLY USED :- Fixed aperture. A set of interchangeable cones for various sizes of breast. An internal set of collimating blades.
  • 24. • for magnification and spot compression studies, manually adjusted shutters allow the x-ray field to be more closely matched to the volume being imaged.
  • 25. Anti scattered grid The anti scattered grid must be a moving grid. Carbon fiber is the preferred interspace material.  Focused grid, with a grid ratio of 3:1 or 5:1.  Grid frequencies of 30-40 lines/cm.
  • 26. • Use of such grid doesn’t compromise the spatial resolution but it does increase the patient dose . • Use of 4:1 ratio grid approximately double the pt. dose as compare to non grid contact mammography • A unique grid is developed specially for the mammography (HTC) high transmission cellular grid. (crossed type ) • It reduce the scattered radiation in two direction, HTC grid has a copper as a grid stripe and air as a interspace , physical dimension grid ratio (3.8:1)
  • 27. Compressor devices • COMPRESSION PLATE: flat surfaced plate used for uniform compression. 1. In Conventional compression plate anterior portion of breast is not compressed properly. 2. Newer slanted S.O.F.T paddles are now available for equal compression to all areas of the breast .
  • 28. ADVANTAGE :-  Provides immobilization of the part during exposure.  Produces radiograph of uniform density.  Reduces geometric unsharpness due to closer object to IR distance.  Reduces scatter and improves contrast.  Reduces tissue overlap and improves resolution.  Enhances recognition of architectural distortion produced by tumors. DISADVANTAGE – is only patient discomfort SPOT COMPRESSION –it is used to magnify small area of breast for better visualization.
  • 29. Amount of compression is between 25 and 40 pounds of pressure
  • 30. Image receptor FILM –SCREEN COMBINATION single-coated film –screen combination is commonly used today (GADOLINIUM OXYSULPHITE) provide higher film contrast and require significantly lower radiation exposure as compared to direct exposure film. • Currently in mammography single emulsion film are used in combination with a single back screen. • MAMMOGRAPHIC CASSETEE ;These are dedicated cassette used for mammography only (18X24/ 24X30 cm)
  • 31.
  • 32. Automatic exposure control • Phototimers are design not only to measure x –ray intensity at image receptor but also for x –ray quality . • Phototimers are also called AEC device . And it is place after the image receptor . • and consists of a single ionization chamber or an array of three or more semiconductor diodes. • x-rays transmitted through the breast, antiscatter grid (if present), and the image receptor generate a signal in the detector.
  • 33. • The signal is accumulated (integrated) and, when the accumulated signal reaches a preset value, the exposure is terminated. • The preset value corresponds to a specified signal-to-noise ratio (SNR) in a digital mammography unit or an acceptable optical density (OD) if a film-screen system is used.
  • 34. Digital mammography  Used with existing Film Mammography unit  Film cassettes are replaced with CR imaging plates with photo-stimulable phosphor (BaFBr:Eu2)  Image reader or ADC  Image Processor or the computer system for post -processing Also called full-field digital mammography (FFDM), The x-ray film is replaced by solid-state detectors (aSi or a selenium) that convert x- rays into electrical signals. The electrical signals are used to produce images, can be seen on a computer or printed on Special film CR MAMMOGRAPHY DR MAMMOGRAPHY
  • 35. DIRECT” conversion • Amorphous selenium (direct conversion) using (TFT) flat panel technology (Siemens, Hologic) • ~70 micron pixels “INDIRECT” conversion • Scintillating phosphor (CsI columns) on an array of amorphous silicon photodiodes using thin-film transistor (TFT) flat panel technology • ~100 micron pixels • Computed radiography • ~50 micron pixels • Slot scanning CCD • ~27 micron pixels Digital mammography
  • 36. Pixel size pixel size decreases: • Spatial resolution improves • Noise increase • Signal-to-noise decreases
  • 37. Advantages • Post acquisition manipulation possible • Enhances detection of anomalies • Selectively increases magnification • Distinguishes between malignant and benign lesions. • High contrast resolution
  • 38. • Potential in reduction of radiation dose. • Storage and transfer of image possible. • Lower image noise. • Quicker than conventional mammography.
  • 39. Advancement in digital mammography 1. Tomosynthesis 2. Photon counting 3. CAD computer aided detection
  • 40. Tomosynthesis • Digital Breast Tomosynthesis (DBT) is a 3D imaging technology that acquires a series of low-dose projection images of the compressed or uncompressed breast at different angles. • 3D captures multiple slices of the breast, all at different angles. The images are brought together to create crystal clear 3D reconstruction of the breast. • The radiologist is then able to review reconstruction, one slide at a time, almost like turning pages in a book. • This makes it easier for doctors to see if there’s anything to be concerned about.
  • 42. Benefits • Minimal pressure is needed, just enough to hold the breast in place. • A lower dose of radiation is required, up to 50% reduction for dense breasts. • The cost of DTS is expected to drop below the cost a of traditional mammogram. • Tomosynthesis allow easier differentiation between benign and malignant lesions • Early detection of small breast cancers that may be hidden during 2D mammography • Greater accuracy in pinpointing size, shape and location of abnormalities
  • 43. • Typical exposure parameters are 29 kVp and 44 mAs (recorded during imaging of an American College of Radiology phantom), which would result in a total radiation dose of 1.45mGy to a breast thickness of 4.2 cm.
  • 44. Photon counting Photon-counting mammography is one of the newer forms of X-ray-based mammography technologies. In 2003, the Micro Dose Mammography (MDM) system was introduced by Sectra Mamea. Photon-counting systems employ a slot- scanning design in which the system performs a lateral scan of the breast with a slot collimator. The main principle is that smaller the detector used, lower the risk that a scattered photon will interact and increase image quality. No grid is needed.
  • 45. • The detector system consists of a large number of crystalline silicon strip detectors, . • Silicon has excellent electronic properties and strip detectors are placed in edge-on geometry with their long axis parallel to the direction of the X-ray beam. • In this way, a sufficient absorption length is created, resulting in high quantum efficiency for the full energy spectrum used in mammography.
  • 46. Technology • The x-ray beam is collimated to a fan beam matching the pre-collimator. It remove that photons which are not initially directed towards the detector • The pre-collimator transforms the beam into several equidistant line beams. • Beneath the breast support there is a detector box containing a post-collimator and the x-ray detector. Si-strip detectors matching the line beams exiting the breast. The number of counted photons in each channel is accumulated during a single sampling interval before being sent to the workstation for image reconstruction and display,
  • 47. Computer Aided detection • Radiologists misdiagnose 10-30% of the malignant cases • Of the cases sent for surgical biopsy, only 10-20% are actually malignant • The CAD technology basically works like a second pair of eyes, reviewing a patient's mammogram . Can detect stage 1 cancers CAD systems can assist radiologists to Reduce these problems
  • 48. Contd.. Highlights suspicious areas • marks clusters of calcifications with a small triangle and breast masses with a star • 100% sensitivity, • decrease false negative readings. • Requires digital input.
  • 49. aSi flat panel X-ray Photon Scintillator Amplification light Capacitor charge Storage analog signal A/D – Converter Digital signal X-ray Photon aSe X-ray into electrons Electron electron Capacitor charge Storage analog signal Analog signal A/D – Converter Digital signal Photon Counting X-ray Photon MicroDose Detector Digital signal 5 (00000000000101) Se flat panel COMPARISON 49
  • 50. Limitation of mammography 1. Not always certain between benign and malignant tumours. 2. Interpretations of mammograms can be difficult because a normal breast can appear differently for each woman. 3. Appearance of an image may be compromised if there is powder or cream on the breasts or post operative cases. 4. Not all types of cancers of the breast can be seen on mammography.
  • 51. Other modalities of mammography • Although mammography is the most sensitive exam available for detecting small breast ca. False Negative Rate is 5-10% . • USG • DEDICATED BREAST CT • BREAST MRI
  • 52.
  • 53.
  • 54. Question for discussions 1. Why subject contrast is high in conventional radiography? 2. In conventional mammography intensifying screen is kept below the film why? 3. Incidence of breast cancer is higher in upper lateral quadrant why? 4. What is double tack anode ? 5. What is the ratio of radiation induced breast cancer resulting from x ray mammography ? Benefit VS risk 6. Can we perform mammogram with general x - ray ?

Editor's Notes

  1. all dedicated mammography systems utilize a “half-field” x-ray beam geometry, which is achieved by fixed collimation at the x-ray tube head. As a result, the central axis of the x-ray beam is directed at the chest wall edge of the receptor and perpendicular to the plane of the image receptor.