This document discusses the advancement of mammographic equipment. It begins by introducing the components and purpose of mammography equipment. Key components discussed in detail include the x-ray tube, compressor, anti-scatter grid, cassette holder, and digital detectors. The document then covers recent advancements, such as digital mammography technologies like computed radiography, full-field digital mammography, and digital breast tomosynthesis, which uses 3D imaging to improve cancer detection rates.
I have include all the contain about mammography like introduction,principle,anatomy,general views ,mammography physics (x-ray tube, housing,filter ,collimator and generator) and different advance technology about mammography.
Hope it will help your queries.
Thank you....!!
I have include all the contain about mammography like introduction,principle,anatomy,general views ,mammography physics (x-ray tube, housing,filter ,collimator and generator) and different advance technology about mammography.
Hope it will help your queries.
Thank you....!!
this power-point slide presentation includes lots of information like how MRI coil works. what is shimming, magnet, fringe, and design of mri coil and also magnet. this will help a lot for radiologist and technician radiographers.. thanks.
Mammograhy is a radiological modality to examine the breast tissue.
The new technology is gradually replacing the conventional mammography by Digital Breast Tomosynthesis. Tomosynthesis or “3D” mammography is a new type of digital x-ray mammogram which
creates 2D and 3D-like pictures of the breasts.
During a “3D” exam, an X-ray arm sweeps in a slight arc over your breast, taking multiple
low dose x-ray images.
Then, a computer produces synthetic 2D and “3D” images of your breast tissue.
The images include thin one millimeter slices, enabling the radiologist to scroll through
images of the entire breast.
We can always opt for other modalities to rule out diagnosis, like Ultrasound in case of breast. But there is always some pros and cons for a particular investigation.
Mammography is a gold standard investigation for breast over 40 years of age. As there is some limitation of age in mammography, due to the biological effects of radiation.
As we know breasts are one of the radiosensitive organ and USG uses sound waves which is harmless to human body so below the age of 40 years doctors recommed to go for USG first to rule out any type of pathology.
Mammography uses low kvp and high mAs thats why the dose in ammography is significantly low.
this power-point slide presentation includes lots of information like how MRI coil works. what is shimming, magnet, fringe, and design of mri coil and also magnet. this will help a lot for radiologist and technician radiographers.. thanks.
Mammograhy is a radiological modality to examine the breast tissue.
The new technology is gradually replacing the conventional mammography by Digital Breast Tomosynthesis. Tomosynthesis or “3D” mammography is a new type of digital x-ray mammogram which
creates 2D and 3D-like pictures of the breasts.
During a “3D” exam, an X-ray arm sweeps in a slight arc over your breast, taking multiple
low dose x-ray images.
Then, a computer produces synthetic 2D and “3D” images of your breast tissue.
The images include thin one millimeter slices, enabling the radiologist to scroll through
images of the entire breast.
We can always opt for other modalities to rule out diagnosis, like Ultrasound in case of breast. But there is always some pros and cons for a particular investigation.
Mammography is a gold standard investigation for breast over 40 years of age. As there is some limitation of age in mammography, due to the biological effects of radiation.
As we know breasts are one of the radiosensitive organ and USG uses sound waves which is harmless to human body so below the age of 40 years doctors recommed to go for USG first to rule out any type of pathology.
Mammography uses low kvp and high mAs thats why the dose in ammography is significantly low.
Mammography and recent advances dr avinashAvinashDahatre
mammography and recent advances includes some physics regarding x ray mammography with different views taken. then some recent advances in mammography like optical mammo and dual energy etc refrence taken from Yochum rowe essential of skeletal radiology, christensen radiology.
Similar to Mammographic equipment and its advancement (20)
This presentation consist MR procedure of pelvis and hip joint , anatomy and MR planning is shown by picture with positioning block and parameters are included ,it includes basic sequence of both procedures
Positron emission tomography pet scan and its applicationsYashawant Yadav
Slides contains physic about the PET scan that is positron emission tomography , its principle , detector configuration types , clinical application of PET Scan and advancement with CT and MRI
This slide contain application of ultrasound and biological effects of ultrasound , ppt contains many GIF files and notes , which may not be accessible here ,,
computed tomography intravenous urography protocol and advancements ,,, slides coves urinary system anatomy glance ,, contrast media used in procedure , radiation doses and some pathological findings
Mri spin echo pulse sequences its variations andYashawant Yadav
MRI spin echo pulse sequences its variation and applications , in this slide collection principle of spine echo pulse sequences is described with physic behind it ,, this slide also coves the inverse recovery pulse sequences and types ,,,, image weighting and parameters are explained .. hope it may be help ful.
Technique of thoracic and lumber spine radiographyYashawant Yadav
The technique of thoracic spine and lumbar spine, do check out not section and most of the slide only content picture because this presentation is verbally explained,
This is a presentation by Dada Robert in a Your Skill Boost masterclass organised by the Excellence Foundation for South Sudan (EFSS) on Saturday, the 25th and Sunday, the 26th of May 2024.
He discussed the concept of quality improvement, emphasizing its applicability to various aspects of life, including personal, project, and program improvements. He defined quality as doing the right thing at the right time in the right way to achieve the best possible results and discussed the concept of the "gap" between what we know and what we do, and how this gap represents the areas we need to improve. He explained the scientific approach to quality improvement, which involves systematic performance analysis, testing and learning, and implementing change ideas. He also highlighted the importance of client focus and a team approach to quality improvement.
Operation “Blue Star” is the only event in the history of Independent India where the state went into war with its own people. Even after about 40 years it is not clear if it was culmination of states anger over people of the region, a political game of power or start of dictatorial chapter in the democratic setup.
The people of Punjab felt alienated from main stream due to denial of their just demands during a long democratic struggle since independence. As it happen all over the word, it led to militant struggle with great loss of lives of military, police and civilian personnel. Killing of Indira Gandhi and massacre of innocent Sikhs in Delhi and other India cities was also associated with this movement.
Students, digital devices and success - Andreas Schleicher - 27 May 2024..pptxEduSkills OECD
Andreas Schleicher presents at the OECD webinar ‘Digital devices in schools: detrimental distraction or secret to success?’ on 27 May 2024. The presentation was based on findings from PISA 2022 results and the webinar helped launch the PISA in Focus ‘Managing screen time: How to protect and equip students against distraction’ https://www.oecd-ilibrary.org/education/managing-screen-time_7c225af4-en and the OECD Education Policy Perspective ‘Students, digital devices and success’ can be found here - https://oe.cd/il/5yV
How to Split Bills in the Odoo 17 POS ModuleCeline George
Bills have a main role in point of sale procedure. It will help to track sales, handling payments and giving receipts to customers. Bill splitting also has an important role in POS. For example, If some friends come together for dinner and if they want to divide the bill then it is possible by POS bill splitting. This slide will show how to split bills in odoo 17 POS.
Ethnobotany and Ethnopharmacology:
Ethnobotany in herbal drug evaluation,
Impact of Ethnobotany in traditional medicine,
New development in herbals,
Bio-prospecting tools for drug discovery,
Role of Ethnopharmacology in drug evaluation,
Reverse Pharmacology.
The Art Pastor's Guide to Sabbath | Steve ThomasonSteve Thomason
What is the purpose of the Sabbath Law in the Torah. It is interesting to compare how the context of the law shifts from Exodus to Deuteronomy. Who gets to rest, and why?
Model Attribute Check Company Auto PropertyCeline George
In Odoo, the multi-company feature allows you to manage multiple companies within a single Odoo database instance. Each company can have its own configurations while still sharing common resources such as products, customers, and suppliers.
We all have good and bad thoughts from time to time and situation to situation. We are bombarded daily with spiraling thoughts(both negative and positive) creating all-consuming feel , making us difficult to manage with associated suffering. Good thoughts are like our Mob Signal (Positive thought) amidst noise(negative thought) in the atmosphere. Negative thoughts like noise outweigh positive thoughts. These thoughts often create unwanted confusion, trouble, stress and frustration in our mind as well as chaos in our physical world. Negative thoughts are also known as “distorted thinking”.
How to Make a Field invisible in Odoo 17Celine George
It is possible to hide or invisible some fields in odoo. Commonly using “invisible” attribute in the field definition to invisible the fields. This slide will show how to make a field invisible in odoo 17.
Palestine last event orientationfvgnh .pptxRaedMohamed3
An EFL lesson about the current events in Palestine. It is intended to be for intermediate students who wish to increase their listening skills through a short lesson in power point.
2024.06.01 Introducing a competency framework for languag learning materials ...Sandy Millin
http://sandymillin.wordpress.com/iateflwebinar2024
Published classroom materials form the basis of syllabuses, drive teacher professional development, and have a potentially huge influence on learners, teachers and education systems. All teachers also create their own materials, whether a few sentences on a blackboard, a highly-structured fully-realised online course, or anything in between. Despite this, the knowledge and skills needed to create effective language learning materials are rarely part of teacher training, and are mostly learnt by trial and error.
Knowledge and skills frameworks, generally called competency frameworks, for ELT teachers, trainers and managers have existed for a few years now. However, until I created one for my MA dissertation, there wasn’t one drawing together what we need to know and do to be able to effectively produce language learning materials.
This webinar will introduce you to my framework, highlighting the key competencies I identified from my research. It will also show how anybody involved in language teaching (any language, not just English!), teacher training, managing schools or developing language learning materials can benefit from using the framework.
2. Abstract / Outlines
•Introduction of Equipment
•Components of mammographic equipment
•Mammographic Cassette
•Advancement in mammography
•References
3.
4. Introduction
The goal of mammography is the
early detection of breast cancer,
typically through detection of
characteristic masses or
microcalcifications.
Basic projection :- CC and MLO
5. Risk factors
• Age
• Hormonal History
• Family History
• Genetics presence of BRCA1 or, BRCA2 gene
• Breast architecture (density)
• Menstruation (before 12yr)
• Menopause (after 55 yr)
• Late age of birth of first child or no children
• Never breast fed
6. Anatomy
• Breast is composed of fatty tissue,
glandular tissue, and connective tissue.
• Normal and cancerous tissues in the
breast have small x-ray attenuation
differences between them
• Need x-ray equipment specifically
designed to optimize breast cancer
detection
7. • Younger breasts are make imaging difficult because it is highly dense
due glandular tissue , and older breasts are more fatty so easier to
imaging .
• Mass density and atomic number of soft tissue of breast are similar
that’s why conventional radiography is useless. (????)
• At 70-100 kVp Compton scattering is predominant than photoelectric
effect which cause low differential absorption with in soft tissue finally
the contrast and resolution is decreased.
• Low kVp is used to maximize the PE effect which enhance the
differential absorption and improve the contrast resolution .
8. Best differential between the tissues is obtained at low x-ray
energies
Mammography equipment
• High contrast sensitivity
• high resolution
• low dose
9. Components of equipment (dedicated)
• Generator
• Support system
• X-ray tube
• Collimator / cone
• Breast compressor
• Anti scatter grids
• Cassette holder
• Automatic exposure control
• Photo timers
• Detector systems
10. Generator
• In mammography we are dealing with soft
tissue of breast.
• 20-40 kvp of x ray energy is required for good
contrast.
• Max mA-200mA
• For homogenous x-ray beam, HFG is
required.
• 3 -10 kw tube rating.
• 3-5 lakh HU-heat loading capacity
12. X –ray tube
Anode or target materials
• Molybdenum
• Rhodium
• Tungsten
Characteristic radiation produced by Mo & Rh are effective for breast
imaging.
Molybdenum is the most common anode target material used in
mammography x-ray tubes.
With digital detectors, W is becoming the target of choice.
13. Cathode
The mammography x-ray tube is configured with dual filaments in the
focusing cup to produce 0.3- and 0.1-mm focal spot sizes
• Mammographic x ray tubes have rotating anode.
• Anode angle- 0 to 16 degrees.
• Tube tilt - 6 -24 degree.
• Source-to-image receptor distance is about 60-65 cm. (80)
• Effective anode angle :- (the actual anode angle plus the physical
tube tilt) is at least 22-24 degrees for coverage of the 24 × 30-cm
field area.
14.
15. • Be(beryllium) window: is used as it has
low atomic no.(4) and lower absorption of
less than 5 kVp x-rays.1mm of Be is
equivalent to 0.1mm Al material(inherent
filter)
• HEEL EFFECT:-Positioning the cathode
over the chest wall of the patient and the
anode over the anterior portion (nipple)
achieves better uniformity of
the transmitted x-rays through the breast .
• Orientation of the tube in this way also
decreases the equipment bulk near the
patient’s head.
16.
17. •FOCAL SPOT SIZE :- 0.3 to 0.4
mm for contact mammography (breast
compressed against the grid and image
receptor) .
• 0.10 to 0.15 mm for magnification
imaging .
• It reduce geometric blurring so that
micro calcifications can be resolved.
18. Target and filter combination
• The filters used in mammography are based on the "k edge" principle
and attenuate the radiation above the k-edge energy of the specific filter
material, either molybdenum or rhodium
• Mo target with 0.03mm(30µm)Mo filter (42)
• Rh target with 0.025mm (25µm)Rh filter (45)
• W target with 0.05mm (50µm) Rh filter (74)
19.
20.
21.
22. Molybdenum target with molybdenum filter
combinations are appropriate for breasts less than 5
cm thick
Molybdenum target with rhodium filters or rhodium
target with rhodium filters may be appropriate for
breasts greater than 5 cm thick
23. Collimator / cones
• Cones are required for proper direction of x ray beam, proper
collimation , for rejecting scatter radiation and protection of chest
wall and lungs from unnecessary radiation.
THREE DESIGN ARE CURRENTLY USED :-
Fixed aperture.
A set of interchangeable cones for various sizes of breast.
An internal set of collimating blades.
24. • for magnification and spot
compression studies, manually
adjusted shutters allow the x-ray field
to be more closely matched to the
volume being imaged.
25. Anti scattered grid
The anti scattered grid must be a moving grid.
Carbon fiber is the preferred interspace material.
Focused grid, with a grid ratio of 3:1 or 5:1.
Grid frequencies of 30-40 lines/cm.
26. • Use of such grid doesn’t compromise the spatial resolution but it does
increase the patient dose .
• Use of 4:1 ratio grid approximately double the pt. dose as compare to
non grid contact mammography
• A unique grid is developed specially for the mammography (HTC) high
transmission cellular grid. (crossed type )
• It reduce the scattered radiation in two direction, HTC grid has a copper
as a grid stripe and air as a interspace , physical dimension grid ratio
(3.8:1)
27. Compressor devices
• COMPRESSION PLATE: flat
surfaced plate used for uniform
compression.
1. In Conventional compression
plate anterior portion of breast is
not compressed properly.
2. Newer slanted S.O.F.T paddles
are now available for equal
compression to all areas of the
breast .
28. ADVANTAGE :-
Provides immobilization of the part during exposure.
Produces radiograph of uniform density.
Reduces geometric unsharpness due to closer object to IR distance.
Reduces scatter and improves contrast.
Reduces tissue overlap and improves resolution.
Enhances recognition of architectural distortion produced by tumors.
DISADVANTAGE – is only patient discomfort
SPOT COMPRESSION –it is used to magnify small area of breast for
better visualization.
30. Image receptor
FILM –SCREEN COMBINATION
single-coated film –screen combination is commonly used today
(GADOLINIUM OXYSULPHITE) provide higher film contrast and
require significantly lower radiation exposure as compared to direct
exposure film.
• Currently in mammography single emulsion film are used in
combination with a single back screen.
• MAMMOGRAPHIC CASSETEE ;These are dedicated cassette used
for mammography only (18X24/ 24X30 cm)
31.
32. Automatic exposure control
• Phototimers are design not only to measure x –ray intensity at image
receptor but also for x –ray quality .
• Phototimers are also called AEC device . And it is place after the image
receptor .
• and consists of a single ionization chamber or an array of three or more
semiconductor diodes.
• x-rays transmitted through the breast, antiscatter grid (if present), and
the image receptor generate a signal in the detector.
33. • The signal is accumulated (integrated) and, when the
accumulated signal reaches a preset value, the exposure is
terminated.
• The preset value corresponds to a specified signal-to-noise ratio
(SNR) in a digital mammography unit or an acceptable optical
density (OD) if a film-screen system is used.
34. Digital mammography
Used with existing Film
Mammography unit
Film cassettes are
replaced with CR
imaging plates with
photo-stimulable
phosphor (BaFBr:Eu2)
Image reader or ADC
Image Processor or the
computer system for
post -processing
Also called full-field
digital mammography
(FFDM),
The x-ray film is
replaced by solid-state
detectors (aSi or a
selenium) that convert x-
rays into electrical
signals.
The electrical signals are
used to produce images,
can be seen on a
computer or printed on
Special film
CR MAMMOGRAPHY DR MAMMOGRAPHY
37. Advantages
• Post acquisition manipulation possible
• Enhances detection of anomalies
• Selectively increases magnification
• Distinguishes between malignant and benign lesions.
• High contrast resolution
38. • Potential in reduction of radiation dose.
• Storage and transfer of image possible.
• Lower image noise.
• Quicker than conventional mammography.
39. Advancement in digital mammography
1. Tomosynthesis
2. Photon counting
3. CAD computer aided detection
40. Tomosynthesis
• Digital Breast Tomosynthesis (DBT) is a 3D imaging technology that
acquires a series of low-dose projection images of the compressed or
uncompressed breast at different angles.
• 3D captures multiple slices of the breast, all at different angles. The
images are brought together to create crystal clear 3D reconstruction of
the breast.
• The radiologist is then able to review reconstruction, one slide at a
time, almost like turning pages in a book.
• This makes it easier for doctors to see if there’s anything to be
concerned about.
42. Benefits
• Minimal pressure is needed, just enough to hold the breast in place.
• A lower dose of radiation is required, up to 50% reduction for dense breasts.
• The cost of DTS is expected to drop below the cost a of traditional
mammogram.
• Tomosynthesis allow easier differentiation between benign and malignant
lesions
• Early detection of small breast cancers that may be hidden during 2D
mammography
• Greater accuracy in pinpointing size, shape and location of abnormalities
43. • Typical exposure parameters are 29 kVp and 44 mAs
(recorded during imaging of an American College of
Radiology phantom), which would result in a total radiation
dose of 1.45mGy to a breast thickness of 4.2 cm.
44. Photon counting
Photon-counting mammography is one of the
newer forms of X-ray-based mammography
technologies.
In 2003, the Micro Dose Mammography (MDM)
system was introduced by Sectra Mamea.
Photon-counting systems employ a slot-
scanning design in which the system performs a
lateral scan of the breast with a slot collimator.
The main principle is that smaller the detector
used, lower the risk that a scattered photon will
interact and increase image quality.
No grid is needed.
45. • The detector system consists of a large
number of crystalline silicon strip detectors, .
• Silicon has excellent electronic properties
and strip detectors are placed in edge-on
geometry with their long axis parallel to the
direction of the X-ray beam.
• In this way, a sufficient absorption length is
created, resulting in high quantum efficiency
for the full energy spectrum used in
mammography.
46. Technology
• The x-ray beam is collimated to a fan beam
matching the pre-collimator. It remove that photons
which are not initially directed towards the detector
• The pre-collimator transforms the beam into
several equidistant line beams.
• Beneath the breast support there is a detector box
containing a post-collimator and the x-ray detector.
Si-strip detectors matching the line beams exiting
the breast. The number of counted photons in each
channel is accumulated during a single sampling
interval before being sent to the workstation for
image reconstruction and display,
47. Computer Aided detection
• Radiologists misdiagnose 10-30% of the malignant cases
• Of the cases sent for surgical biopsy, only 10-20% are actually
malignant
• The CAD technology basically works like a second pair of eyes,
reviewing a patient's mammogram .
Can detect stage 1 cancers CAD systems
can assist
radiologists to
Reduce these
problems
48. Contd..
Highlights suspicious areas
• marks clusters of calcifications
with a small triangle and breast
masses with a star
• 100% sensitivity,
• decrease false negative
readings.
• Requires digital input.
49. aSi flat panel
X-ray Photon
Scintillator
Amplification
light
Capacitor charge
Storage
analog signal
A/D – Converter
Digital signal
X-ray Photon
aSe
X-ray into electrons
Electron
electron
Capacitor charge
Storage
analog signal
Analog signal
A/D – Converter
Digital signal
Photon Counting
X-ray Photon
MicroDose
Detector
Digital signal
5 (00000000000101)
Se flat panel
COMPARISON
49
50. Limitation of mammography
1. Not always certain between benign and malignant tumours.
2. Interpretations of mammograms can be difficult because a normal
breast can appear differently for each woman.
3. Appearance of an image may be compromised if there is powder or
cream on the breasts or post operative cases.
4. Not all types of cancers of the breast can be seen on mammography.
51. Other modalities of mammography
• Although mammography is the most sensitive exam available for
detecting small breast ca. False Negative Rate is 5-10% .
• USG
• DEDICATED BREAST CT
• BREAST MRI
52.
53.
54. Question for discussions
1. Why subject contrast is high in conventional radiography?
2. In conventional mammography intensifying screen is kept below the film
why?
3. Incidence of breast cancer is higher in upper lateral quadrant why?
4. What is double tack anode ?
5. What is the ratio of radiation induced breast cancer resulting from x ray
mammography ? Benefit VS risk
6. Can we perform mammogram with general x - ray ?
Editor's Notes
all dedicated mammography systems utilize a “half-field” x-ray beam geometry, which is achieved by fixed collimation at the x-ray tube head. As a result, the central axis of the x-ray beam is directed at the chest wall edge of the receptor and perpendicular to the plane of the image receptor.