SlideShare a Scribd company logo
Lecture No. 08
Course title:
Atomic spectroscopy
Topic: Spectral Interferences, background absorption
Course instructor: Dr. Salma Amir
GFCW Peshawar
Spectral Interferences
Spectral interferences are those in which the measured light
absorption is erroneously high due to absorption by a
species other than the analyte element.
Absorption due to the presence of light-absorbing molecules
in the flame and light dimming due to the presence of small
particles in the flame are much more common spectral
interferences. Such phenomena are referred to as
background absorption.
The only cures for direct atomic spectral interference are
(1) to choose an alternate analytical wavelength
(2) to remove the interfering element from the sample
Background Absorption
 A common occurrence that results in spectral interference is absorption of
the HCL radiation by molecules or polyatomic species in the atomizer. This
is called “background absorption”; it occurs because not all of the molecules
from the sample matrix and the flame gases are completely atomized. This
type of interference is more commonplace at short wavelengths (<250 nm)
where many compounds absorb.
 Since atoms have extremely narrow absorption lines, there are few
problems involving interferences where one element absorbs at the
wavelength of another. Even when an absorbing wavelength of another
element falls within the spectral bandwidth used, no absorption can occur
unless the light source produces light at that wavelength, i.e., that element
is also present in the light source. However, undissociated molecular forms
of matrix materials may have broadband absorption spectra, and tiny solid
particles in the flame may scatter light over a wide wavelength region.
When this type of nonspecific absorption overlaps the atomic absorption
wavelength of the analyte, background absorption occurs.
Causes of Background Absorption
 Incomplete combustion of organic molecules in the atomizer can cause
serious background interference problems.
 If a flame atomizer is used, incomplete combustion of the solvent may
take place, particularly if the flame is too reducing (fuel rich). The
extent of the interference depends on flame conditions (reducing or
oxidizing), the flame temperature used, the solvent used, and the
sample aspiration rate.
 Background interference is much more severe when graphite furnace
atomizers are used because the pyrolysis step is limited to a maximum
temperature that does not volatilize the analyte. Consequently, many
matrix molecules are not thermally decomposed. They then volatilize
into the light path as the higher atomization temperature is reached
and absorb significant amounts of the source radiation.
Background absorption correction
 To compensate for this problem, the background absorption
must be measured and subtracted from the total measured
absorption to determine the true atomic absorption
component. Various Background correction techniques
employed are
 1. Two line background correction
 2. Continuum source background correction
 3. Zeeman background correction
 4. Smith–Hieftje Background Corrector
1. Two-Line Background Correction
An early manual method of measuring background and applying a correction
for it used the absorption of a nearby nonresonance line. The emission
spectrum from a hollow cathode is quite rich in emission lines and contains
the resonance lines of the element of interest plus many other nonresonance
emission lines from the element and the filler gas. The analyte atoms do not
absorb these nonresonance lines; however, the broad molecular background
absorbs them.
Two measurements are made.
 The absorbance at a nonresonance line close to the analyte resonance line
but far enough away that atomic absorption does not occur is measured.
 Then the absorbance at the resonance line of the analyte is measured.
The absorbance at the nonresonance
line is due only to background
absorption. The absorbance at the
resonance line is due to both atomic
and background absorption. The
difference in absorbance between
the resonance and nonresonance
lines is the net atomic absorption.
SUMMARY:
Resonance line= analyte + background
Non-resonance line= background
Net absorption due to analyte = Resonance - Non-resonance
2. Continuum source background
correction
 Continuum source background correction is a technique for automatically
measuring and compensating for any background component which might be
present in an atomic absorption measurement. This method incorporates a
continuum light source in a modified optical system
 In the continuum source background correction method, when an HCL source
is used, the absorption measured is the total of the atomic and background
absorptions. When a continuum lamp source is used, only the background
absorption is measured. The continuum lamp, which emits light over a range
or band of wavelengths, is placed into the spectrometer system as shown in
Figure 3.7. This setup allows radiation from both the HCL and the continuum
lamp to follow the same path to reach the detector. The detector observes
each source alternately in time, either through the use of mirrored choppers
or through pulsing of the lamp currents.
When the HCL lamp is in position as the source, the emission line from the HCL is
only about 0.002 nm wide; in other words, it fills only about 1% of the spectral
window as shown in the upper left-hand-side drawing in Figure 3.8. When the HCL
radiation passes through the flame, both the free atoms at the resonance
line and the broadband absorbing
molecules will absorb the line. This
results in significant attenuation of the
light reaching the detector, as shown in
the upper center and right-hand-side
drawing in Figure 3.8. When the
continuum source is in place, the
continuum lamp emission fills the entire
spectral window as shown in the lower
left-hand-side drawing in Figure 3.8.
Any absorption of the radiation from the
continuum lamp observed is broadband
background absorption, since it will be
absorbed over the entire 0.2 nm as seen
in the lower right-hand-side drawing in
Figure 3.8
 The absorbance of the continuum source is therefore an accurate
measure of background absorption. An advantage of this method is
that the background is measured at the same nominal wavelength as
the resonance line, resulting in more accurate correction.
3. Zeeman Background correction
 Atomic absorption lines occur at discrete wavelengths because the transition
that gives rise to the absorption is between two discrete energy levels.
However, when a vapor-phase atom is placed in a strong magnetic field, the
electronic energy levels split. This gives rise to several absorption lines for a
given transition in place of the single absorption line in the absence of a
magnetic field. This occurs in all atomic spectra and is called Zeeman splitting
or the Zeeman effect.
 In the simplest case, the Zeeman effect splits an absorption line into two
components. The first component is the π-component, at the same wavelength
as before (unshifted); the second is the σ-component, which undergoes both a
positive and negative shift in wavelength, resulting in two equally spaced lines
on either side of the original line. This splitting pattern is presented in Figure
6.27. The splitting results in lines that are separated by approximately 0.01 nm
or less depending on the field strength. The strength of the magnetic field used
is between 7 and 15 kG. Background absorption and scatter are usually not
affected by a magnetic field.
The π- and σ-components respond differently to polarized light. The π-
component absorbs light polarized in the direction parallel to the magnetic
field. The σ-components absorb only radiation polarized 90° to the applied
field. The combination of splitting and polarization differences can be used
to measured total absorbance (atomic plus background) and background
only, permitting the net atomic absorption to be determined.
A permanent magnet can be placed around the furnace to split the energy levels.
A rotating polarizer is used in front of the HCL or EDL. During that portion of
the cycle when the light is polarized parallel to the magnetic field, both atomic
and background absorptions occur. No atomic absorption occurs when the light
is polarized perpendicular
to the field, but background
absorption still occurs.
The difference between
the two is the net atomic
absorption. Such a system
is a DC Zeeman correction
system.
 Alternately, a fixed polarizer can be placed in front of the light source and an
electromagnet can be placed around the furnace. By making absorption
measurements with the magnetic field off (atomic plus background) and with
the magnetic field on (background only), the net atomic absorption signal can
be determined. This is a transverse AC Zeeman correction system. The AC
electromagnet can also be oriented so that the field is along the light path (a
longitudinal AC Zeeman system) rather than across the light path. No
polarizer is required in a longitudinal AC Zeeman system.
 Zeeman correction can also be achieved by having an alternating magnetic
field surround the hollow cathode, causing the emission line to be split and
then not split as the field is turned on and off. By tuning the amplifier to this
frequency, it is possible to discriminate between the split and unsplit
radiation. A major difficulty with the technique is that the magnetic field used
to generate Zeeman splitting also interacts with the ions in the hollow
cathode. This causes the emission from the hollow cathode to be erratic,
which in turn introduces imprecision into the measurement. Most commercial
instruments with Zeeman background correctors put the magnet around the
atomizer.
4. Smith–Hieftje Background Corrector
 It will be remembered that the HCL functions by the creation of excited
atoms that radiate at the desired resonance wavelengths. After radiating, the
atoms form a cloud of neutral atoms that, unless removed, will absorb
resonance radiation from other emitting atoms. If the HCL is run at a high
current, an abundance of free atoms form. These free atoms absorb at
precisely the resonance lines the hollow cathode is intended to emit; an
example is shown in Figure 6.28, with the free atoms absorbing exactly at λ
over an extremely narrow bandwidth (Figure 6.28b). The result is that the
line emitted from the HCL is as depicted in Figure 6.28c instead of the
desired emission line depicted in Figure 6.28a. The phenomenon of
absorption of the central portion of the emission line by free atoms in the
lamp is called “self-reversal.” Such absorption is not easily detectable,
because it is at the very center of the emitted resonance line and very
difficult to resolve.
It is, of course, exactly the radiation that is most easily absorbed by the
atoms of the sample. In practice, if the HCL is operated at too high a current,
the self-reversal decreases the sensitivity of the analysis by removing
absorbable light. It also shortens the life of the lamp significantly.
Figure 6.28 distortion of spectral line shape in an HCL due to self-reversal.
(a) shape of the spectral line
emitted by the HCL.
(b) shape of the spectral
energy band absorbed by
cool atoms inside the HCL.
(c) shape of the net signal
emerging from an HCL,
showing self-absorption
of the center of the
emission band.
 The Smith–Hieftje background corrector has taken advantage of this self-
reversal phenomenon by pulsing the lamp, alternating between high current
and low current. At low current, a normal resonance line is emitted and the
sample undergoes normal atomic absorption. When the HCL is pulsed to a
high current, the center of the emission line is self-absorbed, leaving only
the wings of the emitted resonance line; the emitted line would look similar
to Figure 6.28c. The atoms of the sample do not significantly absorb such a
line. The broad background, however, absorbs the wings of the line.
Consequently, the absorption of the wings is a direct measurement of the
background absorption at the wavelength of the atomic resonance
absorption. The Smith–Hieftje technique can be used to automatically
correct for background. In practice, the low current is run for a fairly short
period of time and the absorbance measured. The high current is run for a
very short, sharp burst, liberating intense emission and free atoms inside
the hollow cathode and the absorbance is again measured. The difference
between the measurements is the net atomic absorption. There is then a
delay time to disperse the free atoms within the lamp before the cycle is
started again.

More Related Content

What's hot

ATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPY
nadeem akhter
 
Graphite furnace atomic absorption spectroscopy
Graphite furnace atomic absorption spectroscopyGraphite furnace atomic absorption spectroscopy
Graphite furnace atomic absorption spectroscopy
Anuradha Verma
 

What's hot (20)

AAS interference.pptx
AAS interference.pptxAAS interference.pptx
AAS interference.pptx
 
Atomic absorption spectrcopy
Atomic absorption spectrcopyAtomic absorption spectrcopy
Atomic absorption spectrcopy
 
DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.
DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.
DETECTORS USED IN GAS CHROMATOGRAPHY AND HPLC BY P.RAVISANKAR.
 
Gas chromatography detectors
Gas chromatography detectorsGas chromatography detectors
Gas chromatography detectors
 
Atomic emission spectroscopy PPT
Atomic emission spectroscopy PPTAtomic emission spectroscopy PPT
Atomic emission spectroscopy PPT
 
Atomic absorption spectroscopy
Atomic absorption spectroscopy Atomic absorption spectroscopy
Atomic absorption spectroscopy
 
ATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPYATOMIC ABSORPTION SPECTROSCOPY
ATOMIC ABSORPTION SPECTROSCOPY
 
Atomic emission spectroscopy
Atomic emission spectroscopyAtomic emission spectroscopy
Atomic emission spectroscopy
 
Uv visible spectroscopy
Uv visible spectroscopyUv visible spectroscopy
Uv visible spectroscopy
 
Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry  Flame and atomic abosrption spectrophometry
Flame and atomic abosrption spectrophometry
 
Graphite furnace atomic absorption spectroscopy
Graphite furnace atomic absorption spectroscopyGraphite furnace atomic absorption spectroscopy
Graphite furnace atomic absorption spectroscopy
 
Stripping voltammetry
Stripping voltammetryStripping voltammetry
Stripping voltammetry
 
ATOMIC ABSORPTION SPECTROPHOTOMETRY
ATOMIC ABSORPTION SPECTROPHOTOMETRYATOMIC ABSORPTION SPECTROPHOTOMETRY
ATOMIC ABSORPTION SPECTROPHOTOMETRY
 
Atomic Fluorescence Spectroscopy (AFS)
Atomic Fluorescence Spectroscopy (AFS)Atomic Fluorescence Spectroscopy (AFS)
Atomic Fluorescence Spectroscopy (AFS)
 
Atomic absorption Spectrophotometry
Atomic absorption Spectrophotometry Atomic absorption Spectrophotometry
Atomic absorption Spectrophotometry
 
Atomic absorption spectroscopy, History, atomization techniques, and instrume...
Atomic absorption spectroscopy, History, atomization techniques, and instrume...Atomic absorption spectroscopy, History, atomization techniques, and instrume...
Atomic absorption spectroscopy, History, atomization techniques, and instrume...
 
Proton NMR
Proton NMRProton NMR
Proton NMR
 
Atomic Spectroscopy: Basic Principles and Instruments
Atomic Spectroscopy: Basic Principles and InstrumentsAtomic Spectroscopy: Basic Principles and Instruments
Atomic Spectroscopy: Basic Principles and Instruments
 
Applications of ICPAES
Applications of ICPAESApplications of ICPAES
Applications of ICPAES
 
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPYINDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
INDUCTIVELY COUPLED PLASMA -ATOMIC EMISSION SPECTROSCOPY
 

Similar to Lecture 08; spectral interferences and background absorption by Dr. Salma Amir

PHYSICS GROUP- 5 UV Spectroscopy for nanomaterials.pptx
PHYSICS GROUP- 5 UV Spectroscopy for nanomaterials.pptxPHYSICS GROUP- 5 UV Spectroscopy for nanomaterials.pptx
PHYSICS GROUP- 5 UV Spectroscopy for nanomaterials.pptx
dbhansali57
 
Dario Scotto - Detecting and Imaging Magneto-Optically Trapped Rubidium-85 Io...
Dario Scotto - Detecting and Imaging Magneto-Optically Trapped Rubidium-85 Io...Dario Scotto - Detecting and Imaging Magneto-Optically Trapped Rubidium-85 Io...
Dario Scotto - Detecting and Imaging Magneto-Optically Trapped Rubidium-85 Io...
Dario Scotto
 
Atomic Absorption Spectroscopy, Principles and Applications.pptx
Atomic Absorption Spectroscopy, Principles and Applications.pptxAtomic Absorption Spectroscopy, Principles and Applications.pptx
Atomic Absorption Spectroscopy, Principles and Applications.pptx
Ahnaf maznun
 

Similar to Lecture 08; spectral interferences and background absorption by Dr. Salma Amir (20)

Atomic Absorption spectroscopy.pptx
Atomic Absorption spectroscopy.pptxAtomic Absorption spectroscopy.pptx
Atomic Absorption spectroscopy.pptx
 
UV visible Specrosopy and applications
UV visible Specrosopy and applicationsUV visible Specrosopy and applications
UV visible Specrosopy and applications
 
UV Visible Spectrometry
UV Visible SpectrometryUV Visible Spectrometry
UV Visible Spectrometry
 
APA seminar
APA seminarAPA seminar
APA seminar
 
PHYSICS GROUP- 5 UV Spectroscopy for nanomaterials.pptx
PHYSICS GROUP- 5 UV Spectroscopy for nanomaterials.pptxPHYSICS GROUP- 5 UV Spectroscopy for nanomaterials.pptx
PHYSICS GROUP- 5 UV Spectroscopy for nanomaterials.pptx
 
Instru booklet (1)
Instru booklet (1)Instru booklet (1)
Instru booklet (1)
 
Dario Scotto - Detecting and Imaging Magneto-Optically Trapped Rubidium-85 Io...
Dario Scotto - Detecting and Imaging Magneto-Optically Trapped Rubidium-85 Io...Dario Scotto - Detecting and Imaging Magneto-Optically Trapped Rubidium-85 Io...
Dario Scotto - Detecting and Imaging Magneto-Optically Trapped Rubidium-85 Io...
 
UV-Visible Spectroscopy
UV-Visible SpectroscopyUV-Visible Spectroscopy
UV-Visible Spectroscopy
 
AAS.pptx
AAS.pptxAAS.pptx
AAS.pptx
 
Atomic_Absorption_Emission.ppt
Atomic_Absorption_Emission.pptAtomic_Absorption_Emission.ppt
Atomic_Absorption_Emission.ppt
 
Infrared / IR spectroscopy
Infrared / IR spectroscopyInfrared / IR spectroscopy
Infrared / IR spectroscopy
 
Radiology.pptx
Radiology.pptxRadiology.pptx
Radiology.pptx
 
Atomic Absorption Techniques & Applications
Atomic Absorption  Techniques & ApplicationsAtomic Absorption  Techniques & Applications
Atomic Absorption Techniques & Applications
 
Uv spectroscopy sybsc
Uv spectroscopy sybscUv spectroscopy sybsc
Uv spectroscopy sybsc
 
Instru booklet
Instru bookletInstru booklet
Instru booklet
 
Basics of Infrared Spectroscopy : Theory, principles and applications
Basics of Infrared Spectroscopy : Theory, principles and applicationsBasics of Infrared Spectroscopy : Theory, principles and applications
Basics of Infrared Spectroscopy : Theory, principles and applications
 
Atomic Absorption Spectroscopy, Principles and Applications.pptx
Atomic Absorption Spectroscopy, Principles and Applications.pptxAtomic Absorption Spectroscopy, Principles and Applications.pptx
Atomic Absorption Spectroscopy, Principles and Applications.pptx
 
Atomic spectroscopy
Atomic spectroscopy Atomic spectroscopy
Atomic spectroscopy
 
Uv-visible spectroscopy
Uv-visible spectroscopyUv-visible spectroscopy
Uv-visible spectroscopy
 
Prabhakar singh ii sem-paper v-colorimeter &amp; spectrophotometer
Prabhakar singh  ii sem-paper v-colorimeter &amp; spectrophotometerPrabhakar singh  ii sem-paper v-colorimeter &amp; spectrophotometer
Prabhakar singh ii sem-paper v-colorimeter &amp; spectrophotometer
 

More from salmaamir2

More from salmaamir2 (12)

Lecture 06; atomization by Dr. Salma Amir
Lecture 06; atomization by Dr. Salma AmirLecture 06; atomization by Dr. Salma Amir
Lecture 06; atomization by Dr. Salma Amir
 
Lecture 05; atomic absortion spectroscopy , general introduction by Dr. Salma...
Lecture 05; atomic absortion spectroscopy , general introduction by Dr. Salma...Lecture 05; atomic absortion spectroscopy , general introduction by Dr. Salma...
Lecture 05; atomic absortion spectroscopy , general introduction by Dr. Salma...
 
Lecture 04; spectral lines and broadening by Dr. Salma Amir
Lecture 04; spectral lines and broadening by Dr. Salma AmirLecture 04; spectral lines and broadening by Dr. Salma Amir
Lecture 04; spectral lines and broadening by Dr. Salma Amir
 
Lecture 03; Boltzmann equation by Dr. Salma Amir
Lecture 03; Boltzmann equation by Dr. Salma AmirLecture 03; Boltzmann equation by Dr. Salma Amir
Lecture 03; Boltzmann equation by Dr. Salma Amir
 
Lecture 02.; spectroscopic notations by Dr. Salma Amir
Lecture 02.; spectroscopic notations by Dr. Salma AmirLecture 02.; spectroscopic notations by Dr. Salma Amir
Lecture 02.; spectroscopic notations by Dr. Salma Amir
 
Lecture 01; atomic spectroscopy by Dr. Salma Amir
Lecture 01; atomic spectroscopy by Dr. Salma AmirLecture 01; atomic spectroscopy by Dr. Salma Amir
Lecture 01; atomic spectroscopy by Dr. Salma Amir
 
Transition metal polymers by Dr. Salma Amir
Transition metal polymers by Dr. Salma AmirTransition metal polymers by Dr. Salma Amir
Transition metal polymers by Dr. Salma Amir
 
Polyphosphazenes... preparation and properties by Dr. Salma Amir
Polyphosphazenes... preparation and properties by Dr. Salma AmirPolyphosphazenes... preparation and properties by Dr. Salma Amir
Polyphosphazenes... preparation and properties by Dr. Salma Amir
 
Polysiloxanes,... preparation and properties by Dr. Salma Amir
Polysiloxanes,... preparation and properties by Dr. Salma AmirPolysiloxanes,... preparation and properties by Dr. Salma Amir
Polysiloxanes,... preparation and properties by Dr. Salma Amir
 
Polymerization process or synthesis by Dr. Salma Amir
Polymerization process or synthesis by Dr. Salma AmirPolymerization process or synthesis by Dr. Salma Amir
Polymerization process or synthesis by Dr. Salma Amir
 
Classification of inorganic polymers by Dr. Salma Amir
Classification of inorganic polymers by Dr. Salma AmirClassification of inorganic polymers by Dr. Salma Amir
Classification of inorganic polymers by Dr. Salma Amir
 
Introduction to inorganic polymers by Dr. Salma AMIR
Introduction to inorganic polymers by Dr. Salma AMIRIntroduction to inorganic polymers by Dr. Salma AMIR
Introduction to inorganic polymers by Dr. Salma AMIR
 

Recently uploaded

The Roman Empire A Historical Colossus.pdf
The Roman Empire A Historical Colossus.pdfThe Roman Empire A Historical Colossus.pdf
The Roman Empire A Historical Colossus.pdf
kaushalkr1407
 
Additional Benefits for Employee Website.pdf
Additional Benefits for Employee Website.pdfAdditional Benefits for Employee Website.pdf
Additional Benefits for Employee Website.pdf
joachimlavalley1
 

Recently uploaded (20)

NLC-2024-Orientation-for-RO-SDO (1).pptx
NLC-2024-Orientation-for-RO-SDO (1).pptxNLC-2024-Orientation-for-RO-SDO (1).pptx
NLC-2024-Orientation-for-RO-SDO (1).pptx
 
Palestine last event orientationfvgnh .pptx
Palestine last event orientationfvgnh .pptxPalestine last event orientationfvgnh .pptx
Palestine last event orientationfvgnh .pptx
 
Phrasal Verbs.XXXXXXXXXXXXXXXXXXXXXXXXXX
Phrasal Verbs.XXXXXXXXXXXXXXXXXXXXXXXXXXPhrasal Verbs.XXXXXXXXXXXXXXXXXXXXXXXXXX
Phrasal Verbs.XXXXXXXXXXXXXXXXXXXXXXXXXX
 
Cambridge International AS A Level Biology Coursebook - EBook (MaryFosbery J...
Cambridge International AS  A Level Biology Coursebook - EBook (MaryFosbery J...Cambridge International AS  A Level Biology Coursebook - EBook (MaryFosbery J...
Cambridge International AS A Level Biology Coursebook - EBook (MaryFosbery J...
 
Jose-Rizal-and-Philippine-Nationalism-National-Symbol-2.pptx
Jose-Rizal-and-Philippine-Nationalism-National-Symbol-2.pptxJose-Rizal-and-Philippine-Nationalism-National-Symbol-2.pptx
Jose-Rizal-and-Philippine-Nationalism-National-Symbol-2.pptx
 
The Roman Empire A Historical Colossus.pdf
The Roman Empire A Historical Colossus.pdfThe Roman Empire A Historical Colossus.pdf
The Roman Empire A Historical Colossus.pdf
 
Extraction Of Natural Dye From Beetroot (Beta Vulgaris) And Preparation Of He...
Extraction Of Natural Dye From Beetroot (Beta Vulgaris) And Preparation Of He...Extraction Of Natural Dye From Beetroot (Beta Vulgaris) And Preparation Of He...
Extraction Of Natural Dye From Beetroot (Beta Vulgaris) And Preparation Of He...
 
Additional Benefits for Employee Website.pdf
Additional Benefits for Employee Website.pdfAdditional Benefits for Employee Website.pdf
Additional Benefits for Employee Website.pdf
 
2024.06.01 Introducing a competency framework for languag learning materials ...
2024.06.01 Introducing a competency framework for languag learning materials ...2024.06.01 Introducing a competency framework for languag learning materials ...
2024.06.01 Introducing a competency framework for languag learning materials ...
 
50 ĐỀ LUYỆN THI IOE LỚP 9 - NĂM HỌC 2022-2023 (CÓ LINK HÌNH, FILE AUDIO VÀ ĐÁ...
50 ĐỀ LUYỆN THI IOE LỚP 9 - NĂM HỌC 2022-2023 (CÓ LINK HÌNH, FILE AUDIO VÀ ĐÁ...50 ĐỀ LUYỆN THI IOE LỚP 9 - NĂM HỌC 2022-2023 (CÓ LINK HÌNH, FILE AUDIO VÀ ĐÁ...
50 ĐỀ LUYỆN THI IOE LỚP 9 - NĂM HỌC 2022-2023 (CÓ LINK HÌNH, FILE AUDIO VÀ ĐÁ...
 
Home assignment II on Spectroscopy 2024 Answers.pdf
Home assignment II on Spectroscopy 2024 Answers.pdfHome assignment II on Spectroscopy 2024 Answers.pdf
Home assignment II on Spectroscopy 2024 Answers.pdf
 
Unit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdfUnit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdf
 
NCERT Solutions Power Sharing Class 10 Notes pdf
NCERT Solutions Power Sharing Class 10 Notes pdfNCERT Solutions Power Sharing Class 10 Notes pdf
NCERT Solutions Power Sharing Class 10 Notes pdf
 
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
 
Solid waste management & Types of Basic civil Engineering notes by DJ Sir.pptx
Solid waste management & Types of Basic civil Engineering notes by DJ Sir.pptxSolid waste management & Types of Basic civil Engineering notes by DJ Sir.pptx
Solid waste management & Types of Basic civil Engineering notes by DJ Sir.pptx
 
Template Jadual Bertugas Kelas (Boleh Edit)
Template Jadual Bertugas Kelas (Boleh Edit)Template Jadual Bertugas Kelas (Boleh Edit)
Template Jadual Bertugas Kelas (Boleh Edit)
 
Supporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptxSupporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptx
 
The approach at University of Liverpool.pptx
The approach at University of Liverpool.pptxThe approach at University of Liverpool.pptx
The approach at University of Liverpool.pptx
 
How libraries can support authors with open access requirements for UKRI fund...
How libraries can support authors with open access requirements for UKRI fund...How libraries can support authors with open access requirements for UKRI fund...
How libraries can support authors with open access requirements for UKRI fund...
 
How to Split Bills in the Odoo 17 POS Module
How to Split Bills in the Odoo 17 POS ModuleHow to Split Bills in the Odoo 17 POS Module
How to Split Bills in the Odoo 17 POS Module
 

Lecture 08; spectral interferences and background absorption by Dr. Salma Amir

  • 1. Lecture No. 08 Course title: Atomic spectroscopy Topic: Spectral Interferences, background absorption Course instructor: Dr. Salma Amir GFCW Peshawar
  • 2. Spectral Interferences Spectral interferences are those in which the measured light absorption is erroneously high due to absorption by a species other than the analyte element. Absorption due to the presence of light-absorbing molecules in the flame and light dimming due to the presence of small particles in the flame are much more common spectral interferences. Such phenomena are referred to as background absorption. The only cures for direct atomic spectral interference are (1) to choose an alternate analytical wavelength (2) to remove the interfering element from the sample
  • 3. Background Absorption  A common occurrence that results in spectral interference is absorption of the HCL radiation by molecules or polyatomic species in the atomizer. This is called “background absorption”; it occurs because not all of the molecules from the sample matrix and the flame gases are completely atomized. This type of interference is more commonplace at short wavelengths (<250 nm) where many compounds absorb.  Since atoms have extremely narrow absorption lines, there are few problems involving interferences where one element absorbs at the wavelength of another. Even when an absorbing wavelength of another element falls within the spectral bandwidth used, no absorption can occur unless the light source produces light at that wavelength, i.e., that element is also present in the light source. However, undissociated molecular forms of matrix materials may have broadband absorption spectra, and tiny solid particles in the flame may scatter light over a wide wavelength region. When this type of nonspecific absorption overlaps the atomic absorption wavelength of the analyte, background absorption occurs.
  • 4. Causes of Background Absorption  Incomplete combustion of organic molecules in the atomizer can cause serious background interference problems.  If a flame atomizer is used, incomplete combustion of the solvent may take place, particularly if the flame is too reducing (fuel rich). The extent of the interference depends on flame conditions (reducing or oxidizing), the flame temperature used, the solvent used, and the sample aspiration rate.  Background interference is much more severe when graphite furnace atomizers are used because the pyrolysis step is limited to a maximum temperature that does not volatilize the analyte. Consequently, many matrix molecules are not thermally decomposed. They then volatilize into the light path as the higher atomization temperature is reached and absorb significant amounts of the source radiation.
  • 5. Background absorption correction  To compensate for this problem, the background absorption must be measured and subtracted from the total measured absorption to determine the true atomic absorption component. Various Background correction techniques employed are  1. Two line background correction  2. Continuum source background correction  3. Zeeman background correction  4. Smith–Hieftje Background Corrector
  • 6. 1. Two-Line Background Correction An early manual method of measuring background and applying a correction for it used the absorption of a nearby nonresonance line. The emission spectrum from a hollow cathode is quite rich in emission lines and contains the resonance lines of the element of interest plus many other nonresonance emission lines from the element and the filler gas. The analyte atoms do not absorb these nonresonance lines; however, the broad molecular background absorbs them. Two measurements are made.  The absorbance at a nonresonance line close to the analyte resonance line but far enough away that atomic absorption does not occur is measured.  Then the absorbance at the resonance line of the analyte is measured.
  • 7. The absorbance at the nonresonance line is due only to background absorption. The absorbance at the resonance line is due to both atomic and background absorption. The difference in absorbance between the resonance and nonresonance lines is the net atomic absorption. SUMMARY: Resonance line= analyte + background Non-resonance line= background Net absorption due to analyte = Resonance - Non-resonance
  • 8. 2. Continuum source background correction  Continuum source background correction is a technique for automatically measuring and compensating for any background component which might be present in an atomic absorption measurement. This method incorporates a continuum light source in a modified optical system  In the continuum source background correction method, when an HCL source is used, the absorption measured is the total of the atomic and background absorptions. When a continuum lamp source is used, only the background absorption is measured. The continuum lamp, which emits light over a range or band of wavelengths, is placed into the spectrometer system as shown in Figure 3.7. This setup allows radiation from both the HCL and the continuum lamp to follow the same path to reach the detector. The detector observes each source alternately in time, either through the use of mirrored choppers or through pulsing of the lamp currents.
  • 9.
  • 10. When the HCL lamp is in position as the source, the emission line from the HCL is only about 0.002 nm wide; in other words, it fills only about 1% of the spectral window as shown in the upper left-hand-side drawing in Figure 3.8. When the HCL radiation passes through the flame, both the free atoms at the resonance line and the broadband absorbing molecules will absorb the line. This results in significant attenuation of the light reaching the detector, as shown in the upper center and right-hand-side drawing in Figure 3.8. When the continuum source is in place, the continuum lamp emission fills the entire spectral window as shown in the lower left-hand-side drawing in Figure 3.8. Any absorption of the radiation from the continuum lamp observed is broadband background absorption, since it will be absorbed over the entire 0.2 nm as seen in the lower right-hand-side drawing in Figure 3.8
  • 11.  The absorbance of the continuum source is therefore an accurate measure of background absorption. An advantage of this method is that the background is measured at the same nominal wavelength as the resonance line, resulting in more accurate correction.
  • 12. 3. Zeeman Background correction  Atomic absorption lines occur at discrete wavelengths because the transition that gives rise to the absorption is between two discrete energy levels. However, when a vapor-phase atom is placed in a strong magnetic field, the electronic energy levels split. This gives rise to several absorption lines for a given transition in place of the single absorption line in the absence of a magnetic field. This occurs in all atomic spectra and is called Zeeman splitting or the Zeeman effect.  In the simplest case, the Zeeman effect splits an absorption line into two components. The first component is the π-component, at the same wavelength as before (unshifted); the second is the σ-component, which undergoes both a positive and negative shift in wavelength, resulting in two equally spaced lines on either side of the original line. This splitting pattern is presented in Figure 6.27. The splitting results in lines that are separated by approximately 0.01 nm or less depending on the field strength. The strength of the magnetic field used is between 7 and 15 kG. Background absorption and scatter are usually not affected by a magnetic field.
  • 13. The π- and σ-components respond differently to polarized light. The π- component absorbs light polarized in the direction parallel to the magnetic field. The σ-components absorb only radiation polarized 90° to the applied field. The combination of splitting and polarization differences can be used to measured total absorbance (atomic plus background) and background only, permitting the net atomic absorption to be determined.
  • 14. A permanent magnet can be placed around the furnace to split the energy levels. A rotating polarizer is used in front of the HCL or EDL. During that portion of the cycle when the light is polarized parallel to the magnetic field, both atomic and background absorptions occur. No atomic absorption occurs when the light is polarized perpendicular to the field, but background absorption still occurs. The difference between the two is the net atomic absorption. Such a system is a DC Zeeman correction system.
  • 15.  Alternately, a fixed polarizer can be placed in front of the light source and an electromagnet can be placed around the furnace. By making absorption measurements with the magnetic field off (atomic plus background) and with the magnetic field on (background only), the net atomic absorption signal can be determined. This is a transverse AC Zeeman correction system. The AC electromagnet can also be oriented so that the field is along the light path (a longitudinal AC Zeeman system) rather than across the light path. No polarizer is required in a longitudinal AC Zeeman system.  Zeeman correction can also be achieved by having an alternating magnetic field surround the hollow cathode, causing the emission line to be split and then not split as the field is turned on and off. By tuning the amplifier to this frequency, it is possible to discriminate between the split and unsplit radiation. A major difficulty with the technique is that the magnetic field used to generate Zeeman splitting also interacts with the ions in the hollow cathode. This causes the emission from the hollow cathode to be erratic, which in turn introduces imprecision into the measurement. Most commercial instruments with Zeeman background correctors put the magnet around the atomizer.
  • 16. 4. Smith–Hieftje Background Corrector  It will be remembered that the HCL functions by the creation of excited atoms that radiate at the desired resonance wavelengths. After radiating, the atoms form a cloud of neutral atoms that, unless removed, will absorb resonance radiation from other emitting atoms. If the HCL is run at a high current, an abundance of free atoms form. These free atoms absorb at precisely the resonance lines the hollow cathode is intended to emit; an example is shown in Figure 6.28, with the free atoms absorbing exactly at λ over an extremely narrow bandwidth (Figure 6.28b). The result is that the line emitted from the HCL is as depicted in Figure 6.28c instead of the desired emission line depicted in Figure 6.28a. The phenomenon of absorption of the central portion of the emission line by free atoms in the lamp is called “self-reversal.” Such absorption is not easily detectable, because it is at the very center of the emitted resonance line and very difficult to resolve.
  • 17. It is, of course, exactly the radiation that is most easily absorbed by the atoms of the sample. In practice, if the HCL is operated at too high a current, the self-reversal decreases the sensitivity of the analysis by removing absorbable light. It also shortens the life of the lamp significantly. Figure 6.28 distortion of spectral line shape in an HCL due to self-reversal. (a) shape of the spectral line emitted by the HCL. (b) shape of the spectral energy band absorbed by cool atoms inside the HCL. (c) shape of the net signal emerging from an HCL, showing self-absorption of the center of the emission band.
  • 18.  The Smith–Hieftje background corrector has taken advantage of this self- reversal phenomenon by pulsing the lamp, alternating between high current and low current. At low current, a normal resonance line is emitted and the sample undergoes normal atomic absorption. When the HCL is pulsed to a high current, the center of the emission line is self-absorbed, leaving only the wings of the emitted resonance line; the emitted line would look similar to Figure 6.28c. The atoms of the sample do not significantly absorb such a line. The broad background, however, absorbs the wings of the line. Consequently, the absorption of the wings is a direct measurement of the background absorption at the wavelength of the atomic resonance absorption. The Smith–Hieftje technique can be used to automatically correct for background. In practice, the low current is run for a fairly short period of time and the absorbance measured. The high current is run for a very short, sharp burst, liberating intense emission and free atoms inside the hollow cathode and the absorbance is again measured. The difference between the measurements is the net atomic absorption. There is then a delay time to disperse the free atoms within the lamp before the cycle is started again.