Image reconstruction in CT is mostly a mathematical process however, this presentation tries to explain the complicated process of image reconstruction in a visual way, mainly focusing om Filtered back projection, Iterative Reconstruction and AI based image reconstruction.
This slide best explains the introduction of CT, basis and types of CT image reconstructions with detailed explanation about Interpolation, convolution, Fourier slice theorem, Fourier transformation and brief explanation about the image domain i.e digital image processing.
This slide best explains the introduction of CT, basis and types of CT image reconstructions with detailed explanation about Interpolation, convolution, Fourier slice theorem, Fourier transformation and brief explanation about the image domain i.e digital image processing.
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaAvinesh Shrestha
CT is one of the frequently used diagnostic imaging modalities in Radiology. Knowledge about image quality and artifacts is essential when diagnosing a patient with the help of CT images. Moreover, Radiology Technologist's should be very well aware about the ways to identify and eliminate or minimize the artifacts in CT for better image quality.
Quality Assurance Programme in Computed TomographyRamzee Small
Introduction to Computed Tomography
Basic description of the components of a CT System
Introduction to Quality Assurance
Quality Assurance and Quality Control Tests in Computed Tomography base on frequency
Objective of QA/QC Test
MDCT Principles and Applications- Avinesh ShresthaAvinesh Shrestha
Multidetector CT (MDCT) is one of the most commonly used imaging modality in the field of Radiology. Development and advancement in MDCT has made it's application as a major component in diagnosis and treatment planning of multitude of disease across the planet. This presentation briefly describes its basic principle and it's wide variety of application in medical imaging.
Basic physics of multidetector computed tomography ( CT Scan) - how ct scan works, different generations of ct, how image is generated and displayed and image artifacts related to CT Scan.
it includes generations and advancement in CT. In generations fifth generation CT is described in detail.
UFC detector, stellar detectors and gemstone detector is also described
straton x-ray tube, MRC, LIMAX and aquillion one xray tube
different techniques used in CT
dual energy CT is also described
Image Quality, Artifacts and it's Remedies in CT-Avinesh ShresthaAvinesh Shrestha
CT is one of the frequently used diagnostic imaging modalities in Radiology. Knowledge about image quality and artifacts is essential when diagnosing a patient with the help of CT images. Moreover, Radiology Technologist's should be very well aware about the ways to identify and eliminate or minimize the artifacts in CT for better image quality.
Quality Assurance Programme in Computed TomographyRamzee Small
Introduction to Computed Tomography
Basic description of the components of a CT System
Introduction to Quality Assurance
Quality Assurance and Quality Control Tests in Computed Tomography base on frequency
Objective of QA/QC Test
MDCT Principles and Applications- Avinesh ShresthaAvinesh Shrestha
Multidetector CT (MDCT) is one of the most commonly used imaging modality in the field of Radiology. Development and advancement in MDCT has made it's application as a major component in diagnosis and treatment planning of multitude of disease across the planet. This presentation briefly describes its basic principle and it's wide variety of application in medical imaging.
Basic physics of multidetector computed tomography ( CT Scan) - how ct scan works, different generations of ct, how image is generated and displayed and image artifacts related to CT Scan.
it includes generations and advancement in CT. In generations fifth generation CT is described in detail.
UFC detector, stellar detectors and gemstone detector is also described
straton x-ray tube, MRC, LIMAX and aquillion one xray tube
different techniques used in CT
dual energy CT is also described
Non-Blind Deblurring Using Partial Differential Equation MethodEditor IJCATR
In this paper, a new idea for two dimensional image deblurring algorithm is introduced which uses basic concepts of PDEs... The various methods to estimate the degradation function (PSF is known in prior called non-blind deblurring) for use in restoration are observation, experimentation and mathematical modeling. Here, PDE based mathematical modeling is proposed to model the degradation and recovery process. Several restoration methods such as Weiner Filtering, Inverse Filtering [1], Constrained Least Squares, and Lucy -Richardson iteration remove the motion blur either using Fourier Transformation in frequency domain or by using optimization techniques. The main difficulty with these methods is to estimate the deviation of the restored image from the original image at individual points that is due to the mechanism of these methods as processing in frequency domain .Another method, the travelling wave de-blurring method is a approach that works in spatial domain.PDE type of observation model describes well several physical mechanisms, such as relative motion between the camera and the subject (motion blur), bad focusing (defocusing blur), or a number of other mechanisms which are well modeled by a convolution. In last PDE method is compared with the existing restoration techniques such as weiner filters, median filters [2] and the results are compared on the basis of calculated PSNR for various noises
Two Dimensional Image Reconstruction Algorithmsmastersrihari
Convolution Back-Projection (CBP) Algorithm was used for the reconstruction of the image. The performance was compared by implementing the algorithm by using RAM- LAK filter, Shepp- Logan filter and also No filter being used.
Adaptive lifting based image compression scheme using interactive artificial ...csandit
This paper presents image compression method using Interactive Artificial Bee Colony (IABC) optimization algorithm. The proposed method reduces storage and facilitates data transmission by reducing transmission costs. To get the finest quality of compressed image, utilizing local search, IABC determines different update coefficient, and the best update coefficient is chosen
optimally. By using local search in the update step, we alter the center pixels with the coefficient in 8-different directions with a considerable window size, to produce the compressed image, expressed in terms of both PSNR and compression ratio. The IABC brings in the idea of
universal gravitation into the consideration of the affection between onlooker bees and the employed bees. By passing on different values of the control parameter, the universal gravitation involved in the IABC has various quantities of the single onlooker bee and employed bees. As a result when compared to existing methods, the proposed work gives better PSNR.
Contourlet Transform Based Method For Medical Image DenoisingCSCJournals
Noise is an important factor of the medical image quality, because the high noise of medical imaging will not give us the useful information of the medical diagnosis. Basically, medical diagnosis is based on normal or abnormal information provided diagnose conclusion. In this paper, we proposed a denoising algorithm based on Contourlet transform for medical images. Contourlet transform is an extension of the wavelet transform in two dimensions using the multiscale and directional filter banks. The Contourlet transform has the advantages of multiscale and time-frequency-localization properties of wavelets, but also provides a high degree of directionality. For verifying the denoising performance of the Contourlet transform, two kinds of noise are added into our samples; Gaussian noise and speckle noise. Soft thresholding value for the Contourlet coefficients of noisy image is computed. Finally, the experimental results of proposed algorithm are compared with the results of wavelet transform. We found that the proposed algorithm has achieved acceptable results compared with those achieved by wavelet transform.
Initial Introduction of Image processing is included in these slides which contain 1. Introduction of Image Processing
2.Elements of visual perception
3. Image sensing and Quantization
4.A simple image formation model
5.Basic concept of Sampling and Quantization
Reader will find it easy to understand the topics described here in slides . A detailed description of each topic illustrated here.
Please read and if you like do comments also.... Thanks
Principle of Radiation Protection- Avinesh ShresthaAvinesh Shrestha
Radiation protection is the science whose aim is to minimize the risks generated by the use of ionizing radiation. Briefly discusses The ICRP System of Radiological Protection, STRUCTURAL SHIELDING OF
IMAGING FACILITIES, APPLICATION OF INDIVIDUAL DOSE LIMTS, RADIATION EXPOSURE IN PREGNANCY, Diagnostic reference level, Personnel Protection in
Medical X-ray Imaging, Dose Optimization in CT, Radiation Protection in Nuclear Medicine.
Brief discussion on the Quality Assurance and Quality Control in Magnetic Resonance Imaging department.
Quality assurance in MRI is a comprehensive concept that comprises all of the management practices developed by the MR imaging team.
Radiation Dose Units and Dose Limits- Avinesh ShresthaAvinesh Shrestha
Describes different units of radiation dose and the dose limits in diagnostic radiology imaging. Discuses different radiation units described by ICRU. Describes different radiation dose limits given by different organizations like ICRP, NCRP, AERB.
The history and fundamental of radiation biology us presented in this presentation. Topics including basic human biology and response to radiation, Law of Bergonie and Tribondeau,
Physical factors that affect radiation response,
Biologic factors that affect radiation response,
Radiation dose-response relationships and 6R's of radiobiology are discussed in this presentation. Topics like Linear energy transfer (LET),
Relative Biologic Effectiveness (RBE),
Protraction and fractionation, age, oxygen enhancement, hormesis, repair, repopulation, re-oxygenation, redistribution, remote (bystander) cellular effects
radio sensitivity etc. are included in concise and comprehensive manner.
Magnetic Resonance Elastography is an advanced imaging technique in MRI. This method is a method of "virtual palpation" of internal organs with the help of MRI.
Echo planar imaging (EPI) is the method of rapid magnetic resonance imaging (MRI), overcoming one of the significant disadvantage of MRI concerning with slow imaging time. However, EPI-MRI imaging comes with it's own unique imaging artifacts.
CT is one of the highest contributor for medical radiation exposure to patients. Some common CT dose descriptors and dose optimizations methods are briefly described in this presentation.
Application of Perfusion imaging in radiology is increasing with advancement in technology. This presentation briefly describes different perfusion modalities including Computed Tomography, Magnetic Resonance Imaging and Nuclear Medicine. Some of the aspects of perfusion imaging are described in this presentation. This topic was Presented in Radiology department, Institute of Medicine, Maharajgunj.
Venography is a radiological procedure for the evaluation of the veins by the help of intravenous radiological contrast media. It is also known as phlebography. Contrast venography is the gold standard for judging diagnostic imaging methods for deep venous thrombosis; although, because of its cost, invasiveness, the increased sensitivity of sonography to demonstrate pathology and other limitations this test is rarely performed.
This presentation discusses briefly about the anatomy of neck and about different protocols used for CT examination of neck. Also, some pathology are shown in the presentation.
The presentation describes basic anatomy of shoulder and focuses on different radiographic projections used for the evaluation of shoulder. Also, it shows some problems that can be identified in the shoulder radiograph.
Ivu is a radiological investigation for visualization and assessment of the urinary tract.This presentation covers brief anatomy of urinary tract, indication and contraindication,contrast media dose and administration, routine and modified ivu procedure,its complication,ctivu and some abnormalities in the urinary tract.
Intensifying screens are major component of the image receptor used in conventional radiography.Its function is to convert the X-rays into visible light through the process of fluorescence.
Modern medical imaging has been digitized using various technologies which are described here in this presentation.Presented in Department of radiology, ,B.Sc Medical Imaging technology,Institute of Medicine, Nepal.
Telehealth Psychology Building Trust with Clients.pptxThe Harvest Clinic
Telehealth psychology is a digital approach that offers psychological services and mental health care to clients remotely, using technologies like video conferencing, phone calls, text messaging, and mobile apps for communication.
Navigating the Health Insurance Market_ Understanding Trends and Options.pdfEnterprise Wired
From navigating policy options to staying informed about industry trends, this comprehensive guide explores everything you need to know about the health insurance market.
Welcome to Secret Tantric, London’s finest VIP Massage agency. Since we first opened our doors, we have provided the ultimate erotic massage experience to innumerable clients, each one searching for the very best sensual massage in London. We come by this reputation honestly with a dynamic team of the city’s most beautiful masseuses.
Global launch of the Healthy Ageing and Prevention Index 2nd wave – alongside...ILC- UK
The Healthy Ageing and Prevention Index is an online tool created by ILC that ranks countries on six metrics including, life span, health span, work span, income, environmental performance, and happiness. The Index helps us understand how well countries have adapted to longevity and inform decision makers on what must be done to maximise the economic benefits that comes with living well for longer.
Alongside the 77th World Health Assembly in Geneva on 28 May 2024, we launched the second version of our Index, allowing us to track progress and give new insights into what needs to be done to keep populations healthier for longer.
The speakers included:
Professor Orazio Schillaci, Minister of Health, Italy
Dr Hans Groth, Chairman of the Board, World Demographic & Ageing Forum
Professor Ilona Kickbusch, Founder and Chair, Global Health Centre, Geneva Graduate Institute and co-chair, World Health Summit Council
Dr Natasha Azzopardi Muscat, Director, Country Health Policies and Systems Division, World Health Organisation EURO
Dr Marta Lomazzi, Executive Manager, World Federation of Public Health Associations
Dr Shyam Bishen, Head, Centre for Health and Healthcare and Member of the Executive Committee, World Economic Forum
Dr Karin Tegmark Wisell, Director General, Public Health Agency of Sweden
QA Paediatric dentistry department, Hospital Melaka 2020Azreen Aj
QA study - To improve the 6th monthly recall rate post-comprehensive dental treatment under general anaesthesia in paediatric dentistry department, Hospital Melaka
Navigating Challenges: Mental Health, Legislation, and the Prison System in B...Guillermo Rivera
This conference will delve into the intricate intersections between mental health, legal frameworks, and the prison system in Bolivia. It aims to provide a comprehensive overview of the current challenges faced by mental health professionals working within the legislative and correctional landscapes. Topics of discussion will include the prevalence and impact of mental health issues among the incarcerated population, the effectiveness of existing mental health policies and legislation, and potential reforms to enhance the mental health support system within prisons.
Anatomy and Physiology Chapter-16_Digestive-System.pptx
CT Image Reconstruction- Avinesh Shrestha
1.
2. Introduction
The objective of computed tomography (CT) is to reconstruct two-dimensional (2D) or three-
dimensional (3D) images of internal structures from collected signals through an object.
Steps for CT Image Formation: All CT systems use a three step process
Scan or Data Acquisition GET DATA
Image Reconstruction USE DATA
Image Display DISPLAY DATA
7/11/2021 2
3. History
History of reconstruction algorithms that can be traced back to as early as 1917, when J. Radon, an
Austrian mathematician, first presented a mathematics solution for reconstruction as Radon
Transform
William H. Oldendorf (1963) developed a direct back projection method.
Later, the idea of filtered back projection was first proposed by Bracewell and Riddle (1967),
probably the most influential development in this area.
Gordon et al. (1970) proposed the algebraic reconstruction technique (ART), which could produce a
good reconstruction when projections are not uniformly distributed or limited
Special algorithms (convolution back-projection algorithms) were soon introduced. These algorithms
were developed by Ramachandran and Lakshminarayanan (1971) and later used by Shepp and Logan
(1974) to improve image quality and processing time.
Originally, CT images were reconstructed with an iterative method called algebraic reconstruction
technique (ART). Due to lack of computational power, this technique was quickly replaced by simple
analytic methods such as filtered back projection (FBP). FBP was the method of choice for decades,
until the first iterative reconstruction (IR) technique was clinically introduced in 2009
7/11/2021 3
4. Prior to CT reconstruction
Various preprocessing procedures are applied to the actual acquired projection data prior to CT
image reconstruction
air calibration scans: influence of the bow tie filter is characterized, characterize differences in
individual detector response
A dead pixel correction algorithm: replaces dead pixel data with interpolated data from
surrounding pixels
Scatter correction algorithms: Adaptive noise filtration methods algorithms; to reduce the impact
of noise
7/11/2021 4
7. Total linear Attenuation
n
i
n
t
t
t e
N
e
N
N 1
2
1
(
0
)
..
..........
0
n
i
n
......
..........
2
1 =
Where,
8. Image Reconstruction Algorithm
In CT, reconstruction algorithms are used by the computer to solve the many mathematical equations
necessary for information from the detector array to be converted to information suitable for image
display
The process of using the raw data to create an image is called image reconstruction. It is a
mathematical process that generates tomographic images from X-ray projection data acquired at many
different angles around the patient. This process is a computer-intensive task and one of the most
crucial steps in the CT imaging process.
Out of many algorithms used in image reconstruction, these algorithms can be divided into three
mathematical methods of image reconstruction:
Back-projection
Analytical Methods: Fourier analysis and filtered back projection
Iterative Methods: Simultaneous Reconstruction, Ray by Ray Correction, Point by Point correction,
Algebraic reconstruction technique
7/11/2021 8
9. Few Terminologies
Raw Data
Image Data
Sinogram
Convolution
Fourier Transform
7/11/2021 9
10. RAW DATA VERSUS IMAGE DATA
All of the thousands of bits of data acquired by the system with each scan are called raw data.
Once raw data have been processed, i.e. reconstructed, so that each pixel is assigned a Hounsfield
unit value, an image can be created; the data included in the image are now referred to as image
data
The reconstruction that is automatically produced during scanning is often called prospective
reconstruction.
The same raw data may be used later to generate new images. This process is referred to as
retrospective reconstruction
7/11/2021 10
11. SINOGRAM
“Sinogram”, which is simply the 2-D array of
data containing the projections in differenty
angles
Fig. shows the cross section of a body phantom
(b) and its sinogram (a). The two air pockets
located near the center of the phantom (b) are
clearly visible in the sinogram (a) as two dark
sinusoidal curves near the center, and the five
high-density ribs near the periphery of the
phantom (b) are depicted as bright sinusoidal
curves (a).
7/11/2021 11
18. CONVOLUTION
Convolution, a general-purpose algorithm, is a technique of filtering in the space domain. This will become
clear during the discussion of the filtered back-projection algorithm.
Spatial Frequency Filtering: High-Pass Filtering: aka; edge enhancement or sharpness; intended to sharpen
an input image in the spatial domain that appears blurred. First the spatial location image is converted into
spatial frequencies by using the FT, followed by the use of a high-pass filter, which suppresses the low
spatial frequencies to produce a sharper output image.
Spatial Frequency Filtering: Low-Pass Filtering: uses a low-pass filter to operate on the input image with
the goal of smoothing. The output image will appear blurred. Smoothing is intended to reduce noise and the
displayed brightness levels of pixels; however, image detail is compromised
7/11/2021 18
19. Fourier Transform
The Fourier transform, which was developed by the mathematician Baron Jean-Baptiste-Joseph Fourier
in 1807, is widely used in science and engineering.
In radiology, the Fourier transform is used to reconstruct images of a patient’s anatomy in CT and also in
magnetic resonance imaging (MRI) and various other modalities.
In the analysis of time-based signals, the Fourier transform converts from the time domain (x-axis
labeled in seconds) to the temporal frequency domain (x-axis labeled in sec−1). While time-series
analysis has applications in radiological imaging in ultrasound Doppler and magnetic resonance imaging,
here we focus on input signals that are in the spatial domain (x-axis labeled typically in mm), which is
relevant to any anatomical medical image. The Fourier transform of a spatial domain signal converts it
to the spatial frequency domain (x-axis labeled in mm−1).
The Fourier transform is an algorithm that decomposes a spatial or time domain signal into a series of
sine waves that, when summed, replicate that signal. Once a spatial domain signal is Fourier
transformed, the resulting data are considered to be in the frequency domain.
7/11/2021 19
21. Fourier Slice Theorem
7/11/2021 22
One of the most fundamental concepts in CT image reconstruction is the
“Central-slice” theorem. This theorem states that the 1-D FT of the
projection of an object is the same as the values of the 2-D FT of the
object along a line drawn through the 2-D FT plane at same angle.
22. Fourier Slice Theorem
7/11/2021 23
The 1-D projection of the object, measured at certain angle , is the same
as the profile through the 2D FT of the object, at the same angle
23. Fourier Slice Theorem
7/11/2021 24
The 1-D projection of the object, measured at certain angle , is the same
as the profile through the 2D FT of the object, at the same angle
24. Simple Back Projection
Back-projection, sometimes called the "summation method,“ is
the oldest means of image reconstruction. None of the
commercial CT scanners uses simple back-projection, but it is
the easiest method to describe
Using simple backprojection, the reconstructed image has a
characteristic 1/r blurring that results from the geometry of
backprojection. To correct for this, a mathematical filtering
operation is required, and that leads to the discussion of
filtered backprojection in the next section.
7/11/2021 25
27. Analytical Method
A mathematical technique known as convolution or filtering
Technique employs a spatial filter for remove blurring artifacts.
2 types of method
Filtered back projection
Fourier filtering
7/11/2021 28
28. Filtered back projection
Similar to back projection except the image is filtered or modified to exactly counter balance the
density which causes blurring(Star-pattern) in simple back projection
The Mathematical filtering step involve convolving the projection data with a convolution “KERNEL”
With FBP, CT slices are reconstructed from projection data (sinograms) by applying a high pass filter
followed by a backward projection step
Convolution filter refers to a mathematical filtering of the data designed to change the appearance of
the image.
7/11/2021 29
31. Filtered back projection
CONVOLUTION/FILTER:
It is a mathematical filtering process of projected data by the mathematical filter (Kernel) to
reduce the blurring effect of the projections.
7/11/2021 32
Ram Lak
/Ramp
Filter
Shepp-
Logan
Filter
Hamming
Filter
Ideal
Reconstruction
Filter in the
absence of Noise.
It has even more
Pronounced High
frequency roll-off, with
Better higher frequency
noise suppression
It incorporates some roll
off at HF, reduction in
amplification at HF has a
profound influence in
terms of reducing HF noise
in the final CT image
40. Limitations of filtered back projection
In FBP reconstruction, it is assumed that
the x-rays travel in straight lines,
the x-ray photons all have the same energy, and
the x-ray intensity attenuates exponentially in the body (Beer's Law)
the x-ray source is an infinitely small focal spot, and that the x-ray interactions occur along a line
between the focal spot and the geometric center of the detector element, rather than
continuously throughout the patient slab being imaged and the whole detector element.
7/11/2021 41
41. Disadvantage of FBP
The disadvantages of FBP are that it assumes that the sinogram represents a perfect representation
of the object being imaged, and the filtering step amplifies noise in the acquired signal, producing
the characteristic mottled axial images.
Any deviation in the acquisition data can produce dramatic image artefacts, such as streak or
banding artefacts.
Degradation of IQ by high noise, poor low-contrast detectability, and possibly by artifacts resulting
from photon starvation (e.g. streaking) if photon statistics at the detector are very low, e.g. when
imaging obese patients or at low radiation dose.
Therefore, FBP offers negligible dose reduction potential.
7/11/2021 42
42. Iterative Reconstruction
Filtered back projection was the primary method for reconstructing CT images for many decades;
however, advancements in algorithm design coupled with fast computer hardware have led to the use
of clinically useful iterative reconstruction algorithms for CT images.
Unlike the FBP algorithm, which is based on the theoretical inversion of the Radon transform, IR starts
with an initial guess of the object and iteratively improves on it by comparing the synthesized
projection from the object estimate with the acquired projection data and making an incremental
change to the previous guess.
It was first described in 1970 as promising method to generate clinical image with low noise.
7/11/2021 43
43. Iterative Reconstruction
Hounsfield used this technique in his first EMI brain Scanner.
IR for MDCT can achieve superior image quality to its analytical counterparts, such as FBP, and at a
similar or even lower radiation dose level.
IR techniques have been available since the advent of the first CT scanners, but these initial algebraic
reconstruction techniques (ART), and simultaneous iterative reconstruction techniques (SIRT) were
quickly dropped in favor of FBP due to the excessively large computation time required to reconstruct
images.
7/11/2021 44
44. Iterative Reconstruction
Generic Steps:
Assumption : (for example all points in the matrix have same value)
Comparison: (with the measured values)
Correction(to bring the two into agreement)
Repetition(of the process until the assumed and measured values are the same or within acceptable limit
7/11/2021 45
46. Iterative Reconstruction
This method is also called successive approximation method or correction method and image is formed
by 3 ways
Simultaneous reconstruction
Ray by Ray correction
Point-point correction
7/11/2021 47
47. Iterative Reconstruction
Simultaneous reconstruction:
all projections of the entire matrix are calculated at the beginning of the iteration and all
corrections are made simultaneously for each iteration.
Ray –by-Ray Correction:
one ray sum is calculated and corrected and these corrections are incorporated into future ray
sums , with the process being repeated for every ray in each iteration.
Point-by-point correction:
The calculations and corrections are made for all rays passing through one point , and these
corrections are used in ensuring calculations , again with the process being repeated for every
point.
7/11/2021 48
48. Iterative Reconstruction
7/11/2021 49
TYPES
• Iterative algorithm without statistical modelling (Hybrid)
• used originally by Godfrey Hounsfield, however not commercially used due to the inherent limitations of
microprocessors at that time
• Characterized by iterative filtration of data separately performed in projection space and/or in image space.
• The transition from projection to image space, i.e. the actual image reconstruction, usually relies on FBP
.
• only implements modeling of photon statistics
• Iterative algorithm with statistical modelling (Model Based)
• takes into account
• geometry of the CT system, (distances between X-ray tube, iso-center, and detector, and of its components,
(shape and size of the focal spot and the detector elements, as well as to beam and detector geometry).
• noise (photon statistics, photon spectrum emitted by the X-ray tube, the statistical distribution of photons
data acquisition geometry
• object (radiation attenuation)
50. Iterative Reconstruction
While clinical scanners operated with FBP, the CT research community spent a significant effort into
the development of advanced IR algorithms, with the goal to enable low-dose CT with high
diagnostic quality.
Studies showed improved image quality and diagnostic value with IR compared to FBP. Radiation
dose can be reduced with IR by 23 to 76% without compromising on image quality.
Recent review of 1616 articles (2015) dealing with clinical use of iterative reconstruction concluded
that both subjective and objective measures of image quality were the same or improved without
reported diagnostic compromise compared to older techniques. However, radiologists needed time
working with iterative reconstruction images to become accustomed to the different look and to
gain confidence in the diagnostic capability. Typically, after about 90 days, many radiologists hardly
notice the difference in image appearance.
Radiation dose can be reduced further with model based IR compared to hybrid IR and FBP
7/11/2021 51
51. Iterative Reconstruction
In 2009, the first IR algorithm ;IRIS (iterative reconstruction in image space, Siemens
Healthineers) received FDA clearance.
Within 2 years, four more advanced IR algorithms received FDA clearance: ASIR (adaptive
statistical iterative reconstruction, GE Healthcare), SAFIRE (sinogram-affirmed iterative
reconstruction, Siemens Healthineers), iDose4 (Philips Healthcare), and Veo (GE Healthcare). The
first three methods are so-called hybrid IR algorithms
Veo was the first clinical fully iterative IR algorithm, which is one of the most advanced
algorithms so far.
7/11/2021 52
52. Iterative Reconstruction
Advantage :
It has many advantage compare with FBP , important physical factors including focal spot and
detector geometry, photon statistic, beam spectrum and scattering can be more accurately
incorporated into IR.
artifact correction :higher order beam hardening correction
:cone beam artifact correction
:scatter correction
It is better than FBP in metal artifact reduction
Better image with reduced dose is possible
Low image noise
7/11/2021 53
53. Iterative Reconstruction
Disadvantage
It requires immense computing power . longer recon time
Larger number of iteration is required so time consuming
7/11/2021 54
54. Iterative Reconstruction
7/11/2021 55
The process of backward and forward projection is repeated until the difference between
true and artificial raw data is minimized.
55. Iterative Reconstruction
7/11/2021 56
An example of the clearer images enabled by iterative reconstruction using Philips iDose4
software. The software can enhance the sharpness of the stent and reduce metal and calcium
blooming artifacts to give a clearer view of the coronary artery lumen.
56. 7/11/2021 57
Standard FBP (a) and iterative (b) reconstructions at the same level of the ascending aorta.
Image noise expressed as the standard deviation of the attenuation (HU) in the region of
interest was significantly lower in images reconstructed using IR (SAFIRE; Siemens)
(circle in b) than in those reconstructed using FBP (circle in a)
60. Parallel vs Fan-beam
Every ray-sum in fan-beam “sinogram” has equivalent
point in parallel-beam sinogram
The red ray in both the parallel and diverging
configurations are the same, and therefore occupy the
same point in sinogram space.
This behaviour allows us to interpolate the diverging-ray
data into a parallel ray sinogram.
For fan beam reconstruction, the previous parallel beam
algorithms may be applied after we reformat fan beam
projections into parallel beam projections. This
reformatting process is called rebinning
Interpolate div ray projections into parallel-beam
sinogram
Perform reconstruction as if data were collected in
parallel-beam geometry
7/11/2021 61
61. IMAGE RECONSTRUCTION IN SPIRAL CT:
Reconstruction Of spiral CT image is the same as that for conventional CT except for interpolation.
Interpolation is the computation of an unknown value using known values on either side.
A transverse planar image can be reconstruction at any position along the axis of the pt. i.e. z- axis.
Data interpolation is performed by a special computer program called an interpolation algorithm.
7/11/2021 62
Z-AXIS
KNOWN
DATA
KNOWN
DATA
INTERPOLATED
DATA
62. IMAGE RECONSTRUCTION IN SPIRAL CT:
During Spiral or Helical CT image data are received continuously but when an image is reconstructed,
the plane of image does not contain enough data for image reconstruction so the data needed for
image reconstruction estimated by a special computation method known as ‘Interpolation Algorithm’
which meant the estimation of value b/w two known value.
The ability to reconstruct an image at any Z- Axis position is due to interpolation.
Most modern scanners operate in a helical or spiral mode where the x-ray tube and detector system
rotate continuously during data acquisition as the patient table moves through the scanner. Under
these conditions, the projections are not collected on a slice-by-slice basis, and so the reconstruction
techniques described earlier cannot be used directly. However, virtual projections, (or a virtual
sinogram) can be constructed for each required reconstructed slice by suitable interpolation from the
adjacent projections.
7/11/2021 63
64. INTERPOLATION ALGORITHM
7/11/2021 65
180ºlinear interpolation
360º linear interpolation
Results in improved Z-axis
resolution & greatly improved
reformatted image
Blurring in the reformatted
image
Interpolation of the value
separated by180º,half a
revolution of the x-ray tube
The plane of reconstructed
Image interpolated from
Data acquired in one
revolution apart
Interpolation Algorithm
65. INTERPOLATION ALGORITHM
7/11/2021 66
1800 LINEAR INTERPOLATION:
• Interpolation of the value separated
by180º,half a revolution of the x-ray tube
Results in improved Z-axis resolution
and greatly improved reformatted
image.
• It results thinner slice than 3600
interpolation
It results more noiser image than 3600
interpolation.
• 20% higher noise than conventional
CT.
Allows scanning at higher pitch.
• Less Broadens sensitivity profile.
3600 LINEAR INTERPOLATION:
• The plane of reconstructed image
interpolated from data acquired in
one revolution apart
Blurring in the reformatted image
• Thicker slice than 1800
interpolation.
Less noiser image than 1800
interpolation..
• 20% less noise than conventional
CT.
Allows scanning at lower pitch than
1800 interpolation.
• Broadens sensitivity profile.
68. Cone-Beam Algorithms
Fan beam approximation algorithms require that the data be consistent, that is, the x-ray beam from
the tube to the detector and the section being imaged must be in the same plane.
This is no longer the case for large cone angles characteristic of MSCT systems with larger than four
detector rows
The fan beam approximation algorithms are not very
accurate used with the new generation of MSCT
scanners, so other image reconstruction algorithms
are needed. These algorithms are called cone-beam
algorithms, and they have been developed to eliminate
the cone-beam artifacts
7/11/2021 69
70. Cone-Beam Algorithms
Several cone-beam algorithms have become available for use with the new generation of MSCT
scanners, and they basically fall into two classes:
exact cone-beam algorithms and
approximate cone beam algorithms
Exact algorithms for cone-beam data have not been successful in recent times, they are also
“computationally complex and difficult to implement”
However, approximate cone-beam algorithms are successful and fall into two categories: 3D and 2D
algorithms. Two such algorithms that have become common place in MSCT scanners are as follows:
1. Feldkamp-Davis-Kress (FDK), also simply referred to as the Feldkamp-type 3D algorithm
2. Advanced single-slice rebinning (ASSR) 2D approiximate algorithm
7/11/2021 71
71. Cone-Beam Algorithms
Feldkamp-Davis-Kress (FDK) Algorithm:
FDK algorithm simply is an extension of the 2D FBP for fan-beam geometry into a 3D FBP for cone-
beam geometry
This algorithm is more extensive than the 2D FBP algorithm, and it is more commonplace for use with
cone-beam CT scanners
7/11/2021 72
72. Cone-Beam Algorithms
Two-Dimensional Approximate Algorithm:
The main goal of the 2D approximate algorithms “is to provide an image quality close to that of a
three dimensional reconstruction algorithm using two dimensional and back projection methods
It is based on the notion of rebinning, a term used to describe the resorting of the 3D data collected
from the cone-beam acquisition (geometry) to a set of 2D fan-beam projection data and,
subsequently, use the conventional 2D FBP algorithm to reconstruct transaxial images
7/11/2021 73
73. Cone-Beam Algorithms
Two-Dimensional Approximate Algorithm:
Some of the algorithms in this category are:
Adaptive multiplane reconstruction (AMPR) algorithm
weighted hyperplane reconstruction algorithm
7/11/2021 74
77. 7/11/2021 78
Validation of deep
learning–based
reconstruction (DLR)
applied to human
upper abdominal (a,
b) and pelvic
images (c, d). (a) 120
kVp, 35 mAs. (b) 120
kVp, 52 mAs. (c) 120
kVp, 157 mAs. (d) 120
kVp, 175 mAs. Window
level/window width
(WL/WW) = 40/400
HU.
78. 7/11/2021 79
Comparison of filtered back
projection (FBP) and deep
learning–based
reconstruction (DLR) images
of metal and truncation
artifacts. The window
level/window width (WL/WW)
= 60/350 HU for metal and
40/400 HU for truncation
artifacts.
Allan M. Cormack (1963) solved the problem of how to reconstruct images by using a finite number of projections
An algorithm is a precise set of steps to be performed in a specific order to solve a problem.
“Sinogram”, which is simply the 2-D array of data containing the projections in differenty angles
Radon transform data. The high-intensity curves near the middle (left to right) of the sinogram correspond to the projections formed by the phantom itself, and the low-intensity curves across the left-right span of the sinogram are formed by the projections of the table underneath the phantom.
In Figure 4-12, let the solid black line be a trace of gray scale as a function of position across an image. Nineteenth-century French mathematician Joseph Fourier developed a method for decomposing a function such as this gray scale profile into the sum of a number of sine waves. Each sine wave has three parameters that characterize its shape: amplitude (a), frequency (f), and phase (), where
g(x )a sin(2π fx +y ) [4-6]
Figure 4-12 illustrates the sum of four different sine waves (solid black line), which approximates the shape of two rectangular functions (dashed lines). Only four
Demonstrates a two-dimensional reconstruction of a cross cut from the center of a solid block. The block is scanned (Fig. 19-13A) from both the top and left sides by a moving x-ray beam to produce the image profile shown in Figure 19-13B. The image pro-files look like steps. The height of the steps is proportional to the amount of radiation that passed through the block. The center transmitted the most radiation, so it is the highest step in the image profile. The steps are then assigned to a gray scale density that is proportional to their height. These densities are arranged in rows, called "rays“.
When the rays from the two projections are superimposed, or backprojected, they produce a crude reproduction of the original object (Fig. 19-13D). In practice many more projections would be added to improve image quality, but the principle is the same.
A high frequency convolution filter suppress high frequency signals, causing the image to have a smooth appearance and possible improvement in contrast resolution.
A low frequency convolution filter suppresses low frequency signal , resulting in edge enhancement & improvement of spatial resolution.
Intensify, aggravate
It is also called algebraic method , in this method reconstruction problem is seen differently and no longer refer to the radon transform equation.
FIGURE 10-55 The logic behind iterative CT image reconstruction is illustrated in this figure. The reconstruction process begins with an initial estimate of what the object may “look like.” This guess is used to compute forward projection data sets, which are then compared with the measured projection data sets. The difference between the mathematical projections and the physical measurements creates an error matrix for each projection. Each type of iterative algorithm for reconstruction uses the error matrix and then updates
the next iteration of the image. The process continues for a number of iterations until the error matrix is minimized, and the CT image at that point should be a good estimate of the actual object that was scanned.
There are multiple benefits of IR algorithms. First, they allow modeling of the x-ray source and detector, which can improve reconstruction accuracy and spatial resolution. Second, photon statistics are readily considered, allowing the algorithm to more highly consider lower-noise projections and devalue high-noise projections, thereby reducing artifacts and improving dose efficiency. Third, generally true assumptions about physical objects, such as that objects tend to change smoothly except at edges, allow IR algorithms to reduce image noise while preserving the sharpness of anatomic boundaries. Finally, IR algorithms can readily handle nontraditional scanning geometries, such as when the data are not acquired in an axial or helical trajectory.