This document provides an overview of capnography for EMS personnel. It defines capnography as the measurement of end-tidal carbon dioxide (ETCO2) and analysis of the respiratory cycle waveform. The document discusses the types of capnography devices, normal ETCO2 values, waveform phases, and uses of capnography to confirm endotracheal tube placement and diagnose various respiratory conditions. It also outlines how capnography can be used as a monitoring tool during medical procedures like intubation and resuscitation.
Inhalational Anesthetics; Isoflurane and Sevoflurane.pptxMahmood Hasan Taha
Isoflurane (Furane) 1979, Sevoflurane (Ultane) 1990s
general description ,physical properties and anesthetic properties .
Effects on organ system, contraindications, drug interaction.
Inhalational Anesthetics; Isoflurane and Sevoflurane.pptxMahmood Hasan Taha
Isoflurane (Furane) 1979, Sevoflurane (Ultane) 1990s
general description ,physical properties and anesthetic properties .
Effects on organ system, contraindications, drug interaction.
Development Of Online Sensor For CO2 Measurement In Exhaled Air & Signature A...idescitation
This method is used for the online measurement of
carbon dioxide in the exhaled air. The transit time for the
ultrasonic waves to travel from the transmitter to the receiver,
located at the two ends of the Kundt’s tube, is a function of the
density of the medium present in the tube. The variation in
the CO 2 concentration in various patients causes change in
the density of the medium, ultimately causing the change in
the transit time of ultrasonic waves in the tube. Thus, the
different transit times help in giving various concentrations
of carbon dioxide in exhaled air.
2024 03 Monumental Mistakes in EMS BAD EMS v0.2.pdfRobert Cole
(note: This presentation contained videos not included in this slide deck)
Describe the elements of Negligence
Describe the concept of vicarious liability
Describe the role of anchor bias, fatigue, anger and fear in EMS decision making
Review the case of Kyle Vess
Review the case of Paul Tarashuk
Review the case of Crystal Galloway
Introductory/onboarding training for Video Laryngeoscopy, specifically for the MacGrath VL.
NOTE: This is meant to be part of a larger educational endeavor including online, hands on, and team based training.
2018 Effect of Bag-Mask Ventilation vs Endotracheal Intubation During Cardiop...Robert Cole
Bag-mask ventilation (BMV) is a less complex technique than endotracheal
intubation (ETI) for airway management during the advanced cardiac life support phase of
cardiopulmonary resuscitation of patients with out-of-hospital cardiorespiratory arrest.
It has been reported as superior in terms of survival.
1963 COWLEY Clinical Shock: A study of the Biochemical Response in Man.pdfRobert Cole
Accession Number: AD0427998
Title: CLINICAL SHOCK; A STUDY OF THE BIOCHEMICAL RESPONSE TO INJURY IN MAN
Descriptive Note: Annual progress rept. 1 Jan-31 Dec 1963
Corporate Author: MARYLAND UNIV BALTIMORE SCHOOL OF MEDICINE
Personal Author(s): Crowley, R. A.
Report Date: 1963-12-31
Pagination or Media Count: 226.0
Abstract: Traumatic shock is associated usually with severe injury and characterized principally by inability to maintain an adequate circulation. This study focuses on the total problem - the reaction of the body to injury, maintenance of life, and repair of injury. Studies currently in progress and those proposed are aimed primarily to understanding the biochemical response to injury in man. Provisions have been made for careful metabolic studies in the shocked patient without interfering with obvious life saving measures. Such extensive studies have required the assembly of a considerable staff - professional and technical - to support a C.S.U. on a 24-hour basis. Experimental problems relevant to establishment of such a unit evolved from two major factors 1 original nature of the study a scientific study of shock in man and 2 an unprecedented design of this study. Solutions to these problems are described. Since inception of the contract January, 1962, some 200 patients have been studied as they have undergone resuscitation measures. Final organization of the unit now permits more complex studies into the physio-biochemical response to injury in man.
Descriptors: *ENDOTOXIC SHOCK BACTERIA ENZYMES METABOLISM AMMONIA THERAPY HYPOXIA PHYSIOLOGY WOUNDS AND INJURIES IMMUNOLOGY CARDIOVASCULAR SYSTEM HYPOTHERMIA TOXINS AND ANTITOXINS HEMORRHAGE BLOOD COAGULATION
Subject Categories: Stress Physiology
Distribution Statement: APPROVED FOR PUBLIC RELEASE
Proposal to establish a new training center for Multi Agency EMS Training v1....Robert Cole
Vision
The Joint Emergency Medical Services training Center (JEMSTC) is a multi-use campus
and facilities dedicated to the provision of EMS and public safety education in the Ada
County-City Emergency Medical Services System. It would serve as a locus of collaboration and
effort in EMS education, providing not simply classroom space, but a relevant, dynamic,
realistic, and effective learning capacity, ultimately affecting the provision of all EMS services in
a positive way.
The JEMSTC would provide facilities for 24 /7 EMS education, vehicle operation, skills
practice, and credentialing. The facilities would be able to accommodate both EMS and Fire
apparatus in all climates for a diverse array of educational activities. This JEMSTC would meet
all the EMS (and related operational) training for the ACCESS system.
This document from • The Centers for Medicare & Medicaid Services shows that refusing to accept reports or parking EMS patients on the wall may be an EMTALA violation.
Hospitals and administrators do not want line EMS providers to know this, but this is ammo against abuse of EMS systems by ER Staff.
Improving Drug Calculation Performance in Paramedics Practicing in an Emergen...Robert Cole
This literature review will examine the scope of the problem and challenges with mathematical proficiency in out-of-hospital care. It will also explore interventions targeted at improving performance in the out-of-hospital environment, and how they may be applied in initial and continuing education models. The author hopes that improvement in drug calculations will result in fewer medical errors and improved patient care.
Recomendações da OMS sobre cuidados maternos e neonatais para uma experiência pós-natal positiva.
Em consonância com os ODS – Objetivos do Desenvolvimento Sustentável e a Estratégia Global para a Saúde das Mulheres, Crianças e Adolescentes, e aplicando uma abordagem baseada nos direitos humanos, os esforços de cuidados pós-natais devem expandir-se para além da cobertura e da simples sobrevivência, de modo a incluir cuidados de qualidade.
Estas diretrizes visam melhorar a qualidade dos cuidados pós-natais essenciais e de rotina prestados às mulheres e aos recém-nascidos, com o objetivo final de melhorar a saúde e o bem-estar materno e neonatal.
Uma “experiência pós-natal positiva” é um resultado importante para todas as mulheres que dão à luz e para os seus recém-nascidos, estabelecendo as bases para a melhoria da saúde e do bem-estar a curto e longo prazo. Uma experiência pós-natal positiva é definida como aquela em que as mulheres, pessoas que gestam, os recém-nascidos, os casais, os pais, os cuidadores e as famílias recebem informação consistente, garantia e apoio de profissionais de saúde motivados; e onde um sistema de saúde flexível e com recursos reconheça as necessidades das mulheres e dos bebês e respeite o seu contexto cultural.
Estas diretrizes consolidadas apresentam algumas recomendações novas e já bem fundamentadas sobre cuidados pós-natais de rotina para mulheres e neonatos que recebem cuidados no pós-parto em unidades de saúde ou na comunidade, independentemente dos recursos disponíveis.
É fornecido um conjunto abrangente de recomendações para cuidados durante o período puerperal, com ênfase nos cuidados essenciais que todas as mulheres e recém-nascidos devem receber, e com a devida atenção à qualidade dos cuidados; isto é, a entrega e a experiência do cuidado recebido. Estas diretrizes atualizam e ampliam as recomendações da OMS de 2014 sobre cuidados pós-natais da mãe e do recém-nascido e complementam as atuais diretrizes da OMS sobre a gestão de complicações pós-natais.
O estabelecimento da amamentação e o manejo das principais intercorrências é contemplada.
Recomendamos muito.
Vamos discutir essas recomendações no nosso curso de pós-graduação em Aleitamento no Instituto Ciclos.
Esta publicação só está disponível em inglês até o momento.
Prof. Marcus Renato de Carvalho
www.agostodourado.com
Local Advanced Lung Cancer: Artificial Intelligence, Synergetics, Complex Sys...Oleg Kshivets
Overall life span (LS) was 1671.7±1721.6 days and cumulative 5YS reached 62.4%, 10 years – 50.4%, 20 years – 44.6%. 94 LCP lived more than 5 years without cancer (LS=2958.6±1723.6 days), 22 – more than 10 years (LS=5571±1841.8 days). 67 LCP died because of LC (LS=471.9±344 days). AT significantly improved 5YS (68% vs. 53.7%) (P=0.028 by log-rank test). Cox modeling displayed that 5YS of LCP significantly depended on: N0-N12, T3-4, blood cell circuit, cell ratio factors (ratio between cancer cells-CC and blood cells subpopulations), LC cell dynamics, recalcification time, heparin tolerance, prothrombin index, protein, AT, procedure type (P=0.000-0.031). Neural networks, genetic algorithm selection and bootstrap simulation revealed relationships between 5YS and N0-12 (rank=1), thrombocytes/CC (rank=2), segmented neutrophils/CC (3), eosinophils/CC (4), erythrocytes/CC (5), healthy cells/CC (6), lymphocytes/CC (7), stick neutrophils/CC (8), leucocytes/CC (9), monocytes/CC (10). Correct prediction of 5YS was 100% by neural networks computing (error=0.000; area under ROC curve=1.0).
- Video recording of this lecture in English language: https://youtu.be/kqbnxVAZs-0
- Video recording of this lecture in Arabic language: https://youtu.be/SINlygW1Mpc
- Link to download the book free: https://nephrotube.blogspot.com/p/nephrotube-nephrology-books.html
- Link to NephroTube website: www.NephroTube.com
- Link to NephroTube social media accounts: https://nephrotube.blogspot.com/p/join-nephrotube-on-social-media.html
Basavarajeeyam is an important text for ayurvedic physician belonging to andhra pradehs. It is a popular compendium in various parts of our country as well as in andhra pradesh. The content of the text was presented in sanskrit and telugu language (Bilingual). One of the most famous book in ayurvedic pharmaceutics and therapeutics. This book contains 25 chapters called as prakaranas. Many rasaoushadis were explained, pioneer of dhatu druti, nadi pareeksha, mutra pareeksha etc. Belongs to the period of 15-16 century. New diseases like upadamsha, phiranga rogas are explained.
- Video recording of this lecture in English language: https://youtu.be/lK81BzxMqdo
- Video recording of this lecture in Arabic language: https://youtu.be/Ve4P0COk9OI
- Link to download the book free: https://nephrotube.blogspot.com/p/nephrotube-nephrology-books.html
- Link to NephroTube website: www.NephroTube.com
- Link to NephroTube social media accounts: https://nephrotube.blogspot.com/p/join-nephrotube-on-social-media.html
New Drug Discovery and Development .....NEHA GUPTA
The "New Drug Discovery and Development" process involves the identification, design, testing, and manufacturing of novel pharmaceutical compounds with the aim of introducing new and improved treatments for various medical conditions. This comprehensive endeavor encompasses various stages, including target identification, preclinical studies, clinical trials, regulatory approval, and post-market surveillance. It involves multidisciplinary collaboration among scientists, researchers, clinicians, regulatory experts, and pharmaceutical companies to bring innovative therapies to market and address unmet medical needs.
CDSCO and Phamacovigilance {Regulatory body in India}NEHA GUPTA
The Central Drugs Standard Control Organization (CDSCO) is India's national regulatory body for pharmaceuticals and medical devices. Operating under the Directorate General of Health Services, Ministry of Health & Family Welfare, Government of India, the CDSCO is responsible for approving new drugs, conducting clinical trials, setting standards for drugs, controlling the quality of imported drugs, and coordinating the activities of State Drug Control Organizations by providing expert advice.
Pharmacovigilance, on the other hand, is the science and activities related to the detection, assessment, understanding, and prevention of adverse effects or any other drug-related problems. The primary aim of pharmacovigilance is to ensure the safety and efficacy of medicines, thereby protecting public health.
In India, pharmacovigilance activities are monitored by the Pharmacovigilance Programme of India (PvPI), which works closely with CDSCO to collect, analyze, and act upon data regarding adverse drug reactions (ADRs). Together, they play a critical role in ensuring that the benefits of drugs outweigh their risks, maintaining high standards of patient safety, and promoting the rational use of medicines.
These simplified slides by Dr. Sidra Arshad present an overview of the non-respiratory functions of the respiratory tract.
Learning objectives:
1. Enlist the non-respiratory functions of the respiratory tract
2. Briefly explain how these functions are carried out
3. Discuss the significance of dead space
4. Differentiate between minute ventilation and alveolar ventilation
5. Describe the cough and sneeze reflexes
Study Resources:
1. Chapter 39, Guyton and Hall Textbook of Medical Physiology, 14th edition
2. Chapter 34, Ganong’s Review of Medical Physiology, 26th edition
3. Chapter 17, Human Physiology by Lauralee Sherwood, 9th edition
4. Non-respiratory functions of the lungs https://academic.oup.com/bjaed/article/13/3/98/278874
1. Capnography : An Overview for EMS personnel
By Robert S. Cole, CCEMT-P
About the Author: Steve Cole has been involved in EMS and EMS education since 1990. He has worked
for a variety of EMS agencies including volunteer fire, military, hospital based, private, and third service.
He is currently employed by Ada County Paramedics (Boise, Idaho),a top-tier and often besieged Third
Service EMS.
He has no current conflicts of interest with any portion of this article, and has not received from any
commercial interest for its production. He would like to mention that he is open for some money, should a
sufficiently insane commercial enterprise wish to invest in his expansive, eclectic, and somewhat
haphazard quest for EMS excellence.
Comments are welcome. He can be reached by email at: colemedic@hotmail.com
“CO2 is the smoke from the flames of
metabolism.”
– Ray Fowler, M.D. Dallas, Street Doc’s Society
Introduction
Capnography, while seen as a relatively new tool for EMS, was first introduced in 1943.
Using infrared technology, EMS has seen the advancement of many tools and
technologies that have promised revolution in medicine. Ultrasound, cardiac markers,
CO-oximetry, and pulse oximetry to name a few. Some have panned out and some have
not. One of the more promising technologies to break into EMS practice has been End
Tidal Capnography (ETCO2). Originally conceived as a method of monitoring
ventilation, it has gained wide acceptance in EMS for airway maintenance and
monitoring in intubated patients. The availability of waveform capnography, as well as
recent breakthroughs in the understanding of capnography as a marker for overall
physiologic function, have broadened the role of capnography in EMS further. This
article will attempt to discuss both the standard and the new roles for ETCO2.
Types of Capnography
Most ETCO2 devices usually work on the principle that CO2 absorbs infra-red radiation.
A beam of infra-red light is passed across the gas sample to fall onto a sensor. The
presence of CO2 in the sample leads to a reduction in the amount of light falling on the
sensor. The analysis is typically rapid, accurate, and in EMS, fairly accurate. In HAZ-
MAT and anesthesia situations, the presence of high concentrations of other gasses (such
as nitrous oxide) will adversely effect the accuracy of the ETCO2.
There are three major types of ETCO2 devices, side stream and main stream sampling.
2. Colorimetric ETCO2
This is better described as an ETCO2 detector, rather than capnography. These devices
provide CO2 detection based on PH changes in airflow. Colorimetric ETCO2 is a
disposable, single use ETOC2 strictly for use in tube confirmation. In this role is cost
effective, easy, and quick. It does have some limitations. It does not provide any type of
waveform, quantitative analysis, or even changes specific to CO2. It loses its
effectiveness rapidly. It cannot be used to determine different types of respiratory
abnormalities. Presence of some fluids and moisture in the ETT will give false readings
as well.
“End Tidal CO2 reading without a waveform is like a heart rate without an
ECG recording.” – Bob Page “Riding the Waves”
Main-stream ETCO2
With mainstream capnography,
an infrared sensor is placed
directly over the passageway of
gas to reflect light across the gas
sample. The most common use
of this method is monitoring the
placement and function of an
endotracheal tube (ETT). While
uncommon, it is possible to hook
up mainstream capnography to a
breathing patient.
Side-stream ETCO2.
Side-stream capnography is more versatile than
mainstream, as it can be used in a variety of situations
besides placement and monitoring of an ETT. In side-
stream ETCO2, a sample of gas is extracted from the
pathway of gas, and sampled remotely from the
airway. It is also quicker to use, requires less
calibration by the user, and is more durable than
external sensors.
Side-stream technology can be used to monitor ETT,
as well as respiratory function in patients without an ETT. The most common example of
this is the nasal cannula ETCO2, which can be used to monitor patients with various
respiratory conditions, or those receiving treatments that might have respiratory effects.
The two main disadvantages of this method are (1) that it is considered slightly less
accurate, and (2) there is a slight delay in time for the sample to proceed up the tube to
the monitor. Neither of these concerns is considered clinically significant in EMS (yet).
3. Micro-stream ETCO2
Micro-stream technology is a new variation of side-
stream technology. The main difference of this
technology is that only very small amounts of the gas
sample are required to accurately detect and analyze
CO2 readings.
This technology was originally developed for use in
neonates with small tidal volumes, but has shown
promise in other patients as well, such as those with a
decreased tidal volume or in respiratory extremis.
Accuracy of types of ETCO2
Generally speaking, mainstream ETCO2 is considered slightly more accurate, but takes
longer to set up (“warm up”) and requires frequent calibration. The differences in
accuracy have only been demonstrated in spontaneously breathing patients, and are not
considered significant. In addition, both methods are considered accurate for monitoring
ETT placement. Micro-stream ETCO2 technology shows promise in overcoming these
differences in accuracy.
Many things can “spoof” ETCO2: The most common is excessive water vapor in the
ETT, which can cause sensors (especially main stream sensors) to provide “false highs”.
Leaky ETCO2 tubing can give false lows. Finally, repeated use of epinephrine (such as
seen in cardiac arrest) by ET or IV routes, can cause ETCO2 to drop. Typically a
decrease of about 5 torr has been seen. This is not constant through out the study group,
and varies by patient. The clinical significance of this beyond its effect on ETCO2
monitoring is the subject of several ongoing studies.
Basics of Capnography
Simply put it is a measurement of the byproduct of cellular metabolism in the human
body. CO2 is produced by cellular metabolization through the process known as the Kreb
Cycle. It is then exhaled through the respiratory process where we can measure it. The
point where it is measured is at the end of the airway, therefore it is called end tidal
measurement. ETCO2 can be measured in either torr, or mmHg.
Basic capnography is not only the measurement of ETCO2, but analysis of the phases
of exhalation (measured indirectly by monitoring ETCO2) in the respiration cycle.
It is also affected by a variety of respiratory states and perfusion conditions; which can be
indirectly assessed by detailed analysis of ETCO2.
Key Point
Basic capnography is not only the measurement of ETCO2, but analysis of the phases of
exhalation (measured indirectly by monitoring ETCO2) in the respiration cycle.
What is normal?
ETCO2 30-45 mm Hg is the normal value for capnography. Many iatrogenic factors can
alter this reading. For example, imperfect positioning of nasal cannula capnofilters may
cause distorted readings. Unique nasal anatomy, obstructed nares and mouth breathers
may skew results and/or require repositioning of cannula. In addition, oxygen by mask
4. may lower the reading by 10% or more. Finally, as discussed elsewhere, use of nitrous
oxide or excessive water vapor may cause false high readings.
What is the waveform?
Similar to the ECG, the waveform in ETOC2 describes function over time and volume.
The waveform can be divided into inspiratory (phase 0) and expiratory segments. The
expiratory segment, is divided into phases I, II and III, and occasionally, phase IV, which
represents the terminal rise in CO2 concentration. The angle between phase II and phase
III is the alpha angle. The nearly 90 degree angle between phase III and the descending
limb is the beta angle.
A waveform illustrating phase 4 of the capnogram:
Another view of the representative parts of the waveform (using different terminology):
5. When looking at ETCO2, 5 things are measured:
• Frequency: ETCO2 provides an objective analysis of
the rate of breathing.
• Rhythm: By assessing the rhythm in ETCO2
periodic apnea, cheyne stokes, boyts, and other
respiratory patters become easily identifiable.
• Baseline: Evaluation of the baseline can give clues
regarding inter-thoracic pressure and other resistance
to ventilation issues.
• Shape: The shape of the waveform gives clues
regarding the exhalation cycle itself. An asthmatic with an acute asthma attack,
for example, will present with a “shark fin” shape to their waveform.
• Height: The height of a waveform is representative of the level of CO@ being
measured. Various factors result in either increased or decreased/absent ETCO2.
ETCO2 increased :
Technical errors
CO2 output Pulmonary perfusion Alveolar Ventilation
Machine faults
Fever
Malignant Hypoventilation Inadequate O2 flows
hyperpyrexia Increased cardiac Bronchial intubation Leaks in breathing
Sodium bicarbonate output Partial airway system
use Increased blood obstruction Faulty ventilator
Tourniquet release in pressure Rebreathing of airway Faulty valves
compression injuries gas
ETCO2 decreased:
Technical errors
CO2 output Pulmonary perfusion Alveolar Ventilation
Machine faults
Hypothermia Reduced cardiac output Hyperventilation Circuit disconnection
Epinephrine Use Hypotension Apnea Sampling tube leak
Hypovolemia Total airway obstruction
Pulmonary embolism Partial airway obstruction
6. Accidental tracheal
Cardiac arrest extubation
Ventilation/Perfusion mismatch and ETCO2
If ventilation or perfusion are unstable, a Ventilation/Perfusion (V/Q) mismatch can
occur. This will alter the correlation between Arterial CO2 (PaC02) and ETCO2.
This V/Q mismatch can be caused by blood shunting such as occurs during atelectasis
(perfusing unventilated lung area) or by dead space in the lungs (Ventilating unperfused
lung area) such as occurs with a pulmonary embolism or hypovolemia.
Uses of capnography
Capnography and intubation
“When exhaled CO2 is detected (positive reading for CO2) in cardiac arrest,
it is usually a reliable indicator of tube position in the trachea.” - The
American Heart Association 2005 CPR and ECG Guidelines
A relatively new concept, ETCO2 monitoring to confirm and monitor ETT placement,
was first reported in 1991. One early use was to prevent inadvertent placement of a
gastric tube into the airway instead of the esophagus. Since then, it has evolved to
become the standard of care for the confirmation of tube placement.
While most research has focused on maintenance of ETT, the ETCO2 has been shown
useful in studies involving the LMA. In theory it should also be acceptable in double
lumen airway devices such as the PTLA and Combitube, as well as new airway devices
like the King LTD.
ETCO2 is considered mandatory in many services for ETT. Therefore, EMS providers
should document their use of continuous ETCO2 monitoring and when possible attach
wave form strips to their PCRs. This provides an objective and concrete documentation
of tube placement. This is especially important if tube placement is later called into
question by other agencies or the ER. This will timestamp and document your tube as
good.
A good tube will produce a clearly identifiable waveform, but the details of the waveform
will depend on the underlying medical conditions of the patient. Vukmir et al reported a
sensitivity and specificity of 100% for endotracheal tube localization by capnography.
7. A Properly Placed ETT
An esophageal placement (AKA “Gut tube”) should not produce any waveform. Example
(A) below shows sudden dislodgement. Example (b) shows a gut tube from the start.
A:
B:
A misplaced ETT
In low cardiac output states (like shock, cardiac arrest or inadequate chest compressions)
ETCO2 may not be detected. If a patient has had carbonated beverages, or if mouth to
mouth ventilation has been attempted, CO2 may be detected after esophageal intubation
(false positive). The EtCO2 should rapidly decrease to zero (within 3-6 breaths) in this
situation, and the wave form will not be ‘Normal’ looking.
Capnography as a diagnostic tool
Broncheospasm/Asthma: ETCO2 has proven diagnostic value in the determination of
broncheospasm in reactive airway states (as opposed to CHF, hyperventilation, and other
states). This has increasing importance given that recent research calls into doubt the
ability of providers (and even physicians) to adequately differentiate between pneumonia,
asthma/COPD, and CHF based on assessment alone.
ETCO2 values change with the severity of the broncheospasm. With a mild asthma, the
CO2 will drop (below 35) as the patient hyperventilates to compensate. As the asthma
worsens, the C02 levels will rise to normal. When the asthma becomes severe, and the
patient is tiring and has little air movement, the C02 numbers will rise to dangerous levels
(above 60).
The characteristic waveform of broncheospasm is the “Shark Fin”. The Shark Fin is term
for the classic shape of the waveform seen in acute broncheospasm. It is manifested by a
more acute upward of phase II and III than is normally seen, making the ALPHA angle
less acute and the BETA angle more acute.
8. Successful treatment will lessen or eliminate the shark fin shape and return the ETCO2 to
normal range. Therefore, documenting ETCO2 before and after a breathing treatment is
useful in documenting treatment.
It is important to note that a recent study published in the Annals of Emergency Medicine
failed to show the efficacy of ETCO2 in detecting asthma in pediatrics. Future studies are
pending.
9. KEY POINT:
ETCO2 is not an effective diagnostic tool for asthma in pediatrics.
Hypoventilation: This may be due to over sedation, head injury, or other issues.
A real example of hypoventilation. Note the normal shape of the waveform, but the
elevated levels of CO2.
Hyperventilation: This may be due to anxiety, a panic attack, or well compensated
respiratory distress.
Return of spontaneous ventilation during paralysis: ETCO2 is useful for monitoring a
patients level of paralysis, and may detect the need for further paralysis or sedation
before other methods commonly used in EMS. This is usually evident by “notches” in
the waveform.
10. In this capnogram, the notches are notable between the normal waveforms.
Air trapping/overventilation: Mechanical over ventilation in some patients causes a
number of issues effecting coronary perfusion, ROSC, and barotraumas. This will
manifest itself by an elevation of the baseline over time. When this is seen, then
prolonged exhalation times are needed.
Equipment Malfunction: The ETCO2 can detect a failure in the ventilation circuit,
particularly a leaky cuff, detached BVM, or failing one-way valve.
11. Detection of sudden cardiac arrest: While uncommon, the sudden loss of pulses, while
an electrical rhythm remains (such as seen in a tension pneumo-thorax) does happen. This
is even harder to detect in an intubated and paralyzed patient.
Return of Spontaneous Circulation: A sudden increase in the height in an ETCO2
waveform may indicate a sudden increase in blood flow in cardiac arrest, usually from
ROSC.
PEA vs. Severe Shock: White R.D. et al used EtCO2 measurement in out-of hospital
cardiac arrest. They concluded that capnography can detect the presence of pulmonary
blood flow even in the absence of major pulses (pseudo-pulseless electrical activity-
PEA) and also can rapidly indicate changes in pulmonary blood flow (cardiac output)
caused by alterations in cardiac rhythm. This can be helpful in determining the difference
between a true low-flow state and actual PEA.
Capnography as a monitoring tool
ETCO2, like any “tool” is not, and should not, be considered the ultimate monitoring
device. It is however very useful in select situations. Its effectiveness is increased when
used with other monitoring devices, such as SPO2. One anesthesia study showed that use
of both ETCO2 and SPO2 together would have prevented 93% of anesthesia “mishaps”.
This is not to imply that SPO2 is accurate or reliable by itself either, and as a general rule
ETCO2 is more “timely” in its detection of problems than SPO2.
Another study used ETCO2 as a detector for compressor fatigue in cardiac arrest. This is
based on the thought that quality compressions produce better ETCO2 readings. When
the quality of compressions fall, then ETCO2 will fall as well. Therefore as a matter of
troubleshooting, when ETCO2 decreases, switching compressors should be considered.
What does the literature say?
In a study of 291 intubated head injured patients, 144 had ETCO2 monitoring. Patients
with ETCO2 monitoring had lower incidence of inadvertant severe hyperventilation
(5.6%) than those without ETCO2 monitoring (13.4%). Patients in both groups with
12. severe hyperventilation had significantly higher mortality (56%) than those without
(30%). –Davis, The Use of Quantitative End-Tidal Capnometry to Avoid Inadvertant
Severe Hyperventilation in Patients with Head Injury After Paramedic Rapid Sequence
Intubation,
-Journal of Trauma, April 2004
Trauma - A 2004 study of blunt trauma patients requiring RSI showed that only 5 percent
of patients with ETCO2 below 26.25 mm Hg after 20 minutes survived to discharge. The
median ETCO2 for survivors was 30.75.
- Deakin CD, Sado DM, Coats TJ, Davies G.
“Prehospital end-tidal carbon dioxide concentration and outcome in major trauma.”
Journal of Trauma. 2004;57:65-68.
Use of ETCO2 in intubated patients is not just useful in tube confirmation. It can be used
as a guide to ventilation rates. A target ETCO2 of 30-35 is acceptable and desirable.
Decreasing the ventilatory rate in an intubated patient is acceptable to achieve this goal.
Key Point
An ETCO2 of 30-35 is a desirable goal of ventilation
Capnography as a predictor of survival in SCA
Several studies have looked at ETOC2 as a predictor of survival of cardiac arrest. One of
larger studies was conducted overseas with a study population of nearly 400 patients. In
this study, the researchers looked at both initial and final ETOC2. They found that the
average initial ETCO2 in the ROSC was 18 torr vs 6.75 torr in the non-ROSC group. In
addition, the final ETCO2 for ROSC group was on average 26 torr vs 7.5 for the non-
ROSC group. The most important finding was that no patient with an initial ETCO2 of 9
or less survived to admission.
Another study noted that "An end-tidal carbon dioxide level of 10 mmHg or less
measured 20 minutes after the initiation of advanced cardiac life support accurately
predicts death in patients with cardiac arrest associated with electrical activity but no
pulse. Cardiopulmonary resuscitation may reasonably be terminated in such patients.”
(Levine, 1997) It should be noted than most normal arrests that fail to be resuscitated
after 20 minutes survive, regardless of ETCO2.
Several other studies have found dramatic differences in ETCO2 in survivors and non
survivors of SCA, but no consensus has yet been reached on what constitutes a survivable
ETCO2. In addition, it remains unclear if ETCO2 is truly an independent predictor, or
simply a result of other factors (such a quality of CPR) that affect ROSC.
Therefore, while future studies will be needed before ETCO2 alone can be used as
justification to terminate resuscitation (especially in special resuscitation groups like
hypothermia), use of this technology does hold promise, and should be considered
important in the termination thought process and included in discussion with medical
control.
KEY POINT:
13. LOW ETCO2 alone is not indication to terminate resuscitation, but it is an
important consideration. High ETCO2 may be an indicator to continue
prolonged resuscitation.
Pitfalls of capnography
The greatest potential for failure is focusing on the ETCo2 and not the patients overall
clinical picture. The ages old mantra “Treat the patient, not the monitor” is more true than
ever.
Other pitfalls include not using ETCO2 on ETT patients. This is now considered by many
experts as the “gold standard” and failing to do so opens up the medic to accusations of
tube displacement.
Summary
Basic capnography is not only the measurement of ETCO2, but analysis of the phases of
exhalation (measured indirectly by monitoring ETCO2) in the respiration cycle. In this, it
is as useful, and as complex, as electrocardiography. It has been proven accurate and in
some cases definitive in many clinical concerns facing the EMS community today. While
still an emerging technique, it is likely that future paramedics will be stressing over
capnography classes as much as they do over cardiology or pharmacology today.
14. Glossary
Capnography – A graphic representation of the measurement of carbon dioxide (CO2) in
exhaled breath.
Capnometer – The numeric measurement of CO2.
Capnogram – CO2 waveforms which can be of two types: FCO2 can be plotted against
expired volume (SBTCO2 curve /volume capnogram/ CO2 expirogram) or against time
(time capnogram) during a respiratory cycle. Most EMS monitors are TIME capnograms.
End Tidal CO2 (ETCO2 or PetCO2) - the level of (partial pressure of) carbon dioxide
released at end of expiration.
15. References
Lisa Evered, MD;Francine Ducharme, MD;G. Michael Davis, MD; Martin Pusic, MD,
Can we assess asthma severity using expiratory capnography in a pediatric emergency
department? Can J Emerg Med 2003;5(3):169-170
Levine R, End-tidal Carbon Dioxide and Outcome of Out-of-Hospital Cardiac Arrest,
New England Journal of Medicine, July 1997
Bhavani Shankar K, Moseley H, Kumar AY, Delph Y. Capnometry and
anaesthesia. Review article. Canadian J Anaesth 1992;39:6:617-32.
Ornato J P et al: Multicenter study of a portable, hand-size, colorimetric end-tidal
carbondioxide detection device. Ann Emerg Med, May 1992; 21: 518-523.
Garnett A R, Germin C A, Germin A S: Capnographic wavefoms in esophageal
intubations: effect of carbonated beverages. Ann Emerg Med 1989; 18:387-90.
Vukmir R B, Heller M B, Stein K L: Confirmation of endotracheal tube placement:
A miniaturized infrared qualitative carbondioxide detector. Ann Emerg Med July
1991; 20:726-729.
Tobias JD, Flanagan JF, Wheeler TJ, Garrett JS, Burney C. Noninvasive
monitoring of end-tidal CO2 via nasal cannulas in spontaneously breathing children
during the perioperative period. Crit Care Med 1994;22:11:1805-8.
Ornato J P, Garnett A R, Glauser F L, Virginia R: Relationship between cardiac
output and the end-tidal carbondioxide tension. Ann Emerg med 1990, 19: 1104-
1106.
White R D, Asplin B R: Out of hospital quantitative monitoring of end-tidal
carbondioxide pressure during CPR. Ann Emerg Med January 1994; 23;25-30.
Lewis L M, Sthothert J, Standeven J et al : Correlation of end-tidal carbondioxide
to cerebral perfusion during CPR. Ann Emerg Med 1992; 21:1131-4.
16. Callaham M, Barton C: Prediction of outcome of CPR from end-tidal
carbondioxide concentration. Crit Care Med 1990; 18:358-362.
Chase P B, Kern k B, Sanders A B et al: Effects of graded doses of epinephrine
on both non invasive and invasive measures of myocardial perfusion and blood
flow during cardio-pulmonary resuscitation. Crit care Med 1993; 21:413-9.
Callaham M, Barton C, Mathay M: Effect of epinephrine on the ability of end-tidal
carbondioxide readings to predict initial resuscitation from cardiac arrest. Crit
Care Med 1992; 20:337-343.
Ward k R, Yealy D M: End-tidal carbondioxide monitoring in emergency
medicine, part2: Clinical applications. Academic Emergency Med 1998; 5: 637-
646.
Kalenda Z: Capnogram as a guide to the efficacy of cardiac massage.
Resuscitation 1978; 6:259-63.