SlideShare a Scribd company logo
Bioanalytical method and its
validation
VALIDATION
PRODUCT
PROCESS
SYSTEM
ACTIVITY
PROCEDURE
BIOANALYTICAL
METHOD
COLLECTING
PROCESSING
STORAGE
ANALYSIS
 Literature search for drugs
 Identification of analytical techniques and
optimization
 Reference standard preparation
 Selection of internal standard
 Sample pre- treatment (Extraction procedure)
 Sample storage
 Analysis through suitable technique
BIOANLYTICAL
METHOD
VALIDATION
PRAMETERS
SENSITIVITY
REPRODUCI
BILITY
RECOVERY
PRECISION
ACCURACY
STABILITY
SELECTIVITY
CALIBRATION
CURVE
Chromatography
 A blank biological matrix is spiked with the analytes of
interest using solutions of reference standard(s) to prepare
calibration standards, QC sample and stability QC
sample.
 The reference standard used should be obtained from an
authentic and traceable source.
 The reference standard should be identical to the
analyte.
 A certificate of analysis (CoA) is required to
ensure quality .
Selectivity:
 Blank samples are obtained from at least 6
individual sources/lots.
 Responses detected should not be more than 20% of
the analyte response at the LLOQ.
 For the investigation of selectivity in lipaemic and haem
 olysed matrices at least one source of matrix should be
used.
Specificity
 If the presence of related substances is anticipated in
the biological matrix, the impact of such substances
should be evaluated during method validation.
 Responses detected should not be more than 20% of
the analyte response at the LLOQ
Matrix effect
 The matrix effect should be evaluated by analysing at least 3
replicates of low and high QCs, each prepared using matrix from
at least 6 different sources/lots.
 Use of fewer sources/lots is acceptable in the case of rare
matrices.
 The accuracy should be within ±15% of the nominal
concentration and the precision percent coefficient of
variation (CV) should not be greater than 15% in all
individual matrix sources/lots
Calibration curve and range
 A calibration curve should be generated with a
blank sample, a zero sample, and at least 6
concentration levels of calibration standards,
including the LLOQ and the ULOQ.
 The calibration curve parameters should be
reported (slope and intercept in the case of a
linear model).
Accuracy and precision
Preparation of quality control sample
 The QCs are intended to mimic study samples and
should be prepared by spiking matrix with a known
quantity of analyte, stored under the conditions
anticipated for study samples and analysed to assess
the validity of the analytical method.
 During method validation the QCs should be
prepared at a minimum of 4 concentration levels
within the calibration curve range.
Evaluation of accuracy and precision
Accuracy and precision should be determined by
analysing the QCs. Accuracy and precision
should be evaluated using the data.
Carry over
 Carry-over should be minimised during method
development.
 Carry-over in the blank samples following the
highest calibration standard should not be greater
than 20% of the analyte response at the LLOQ
Stability
 Stability evaluations should be carried out to ensure that every
step taken during sample preparation, processing and analysis as
well as the storage conditions used do not affect the
concentration of the analyte.
 The storage and analytical conditions applied to the
stability tests, such as the sample storage times and
temperatures, sample matrix, anticoagulant and container
materials, should reflect those used for the study samples.
Reinjection reproducibility
 Reproducibility of the method is assessed by replicate
measurements of the QCs and is usually included in the
assessment of precision and accuracy.
 However, if samples could be reinjected, reinjection
reproducibility should be evaluated.
 These are included in the Validation Report or
provided in the Bioanalytical Report of the study
where it was conducted.
 The analysis of study samples can be carried out after
validation has been completed, however, it is
understood that some parameters may be completed at a
later stage.
 By the time the data are submitted to a regulatory authority, the
bioanalytical method validation should have been completed.
 The study samples, QCs and calibration standards should be
processed in accordance with the validated analytical method
Ligand binding assay
 The reference standard should be well characterized and
documented.
 A biological drug has a highly complex structure and its
reactivity with binding reagents for bioanalysis may be
influenced by a change in the manufacturing process of
the drug substance.
Specificity
 The response of blank samples spiked with related
molecules should be below the LLOQ.
 The accuracy of the target analyte in presence of related
molecules should be within ±25% of the nominal values.
Selectivity
 Selectivity is evaluated using blank samples
obtained from at least 10 individual sources and
by spiking the individual blank matrices at the
LLOQ.
 The accuracy should be within ±25% at the
LLOQ of the nominal concentration.
Calibration and range
 A calibration curve should be generated with at
least 6 concentration levels of calibration
standards, including LLOQ and ULOQ
standards.
 The blank sample should not be included in the
calculation of calibration curve parameters
 The accuracy and precision of concentrations of each
calibration standard should be within ±25% of the
nominal concentration at the LLOQ and ULOQ.
Accuracy and Precision
Preparation of quality control sample
 During method validation the QCs should be prepared at a
minimum of 4 concentration levels within the calibration curve
range.
Evaluation of accuracy and precision
 The accuracy at each concentration level should be
within ±20% of the nominal values, except for the
LLOQ and ULOQ, which should be within ±25% of
the nominal value.
 Precision of the QC concentrations determined at
each level should not exceed 20%, except at the
LLOQ and ULOQ, where it should not exceed 25%.
Stability
Study Sample Analysis
 General and specific SOPs and good record
keeping are essential to a properly validated
analytical method.
 The data generated for bio analytical method
validation should be documented
 Should be available during data audit and
inspection.
 ICH, Validation of analytical procedure, International
conference on Harmonization, IFPMA, Q2B Validation of
Analytical Procedures: Methodology
 ICH, International conference on Harmonization, IFPMA,
M10 Bioanalytical method validation
 Bioanalytical Method Development and Validation: from
the USFDA 2001 to the USFDA 2018 Guidance for
Industry

More Related Content

What's hot

Qualification of instrumets
Qualification of instrumetsQualification of instrumets
Qualification of instrumets
ChowdaryPavani
 
Sample preparation and protocols in metabolite identification
Sample preparation and protocols in metabolite identificationSample preparation and protocols in metabolite identification
Sample preparation and protocols in metabolite identification
Durgadevi Ganesan
 
General considerations and method development in ce,
General considerations and method development in ce,General considerations and method development in ce,
General considerations and method development in ce,
ChowdaryPavani
 
Analytical Method Validation as per ICH vs USP
Analytical Method Validation as per ICH vs USPAnalytical Method Validation as per ICH vs USP
Analytical Method Validation as per ICH vs USP
Kushal Shah
 
M10 bioanalytical method validation
M10 bioanalytical method validationM10 bioanalytical method validation
M10 bioanalytical method validation
sawantanil
 
bio analytical method validation usfda guidlines
bio analytical method validation usfda guidlinesbio analytical method validation usfda guidlines
bio analytical method validation usfda guidlines
chandu chatla
 
45160177 forced-degradation
45160177 forced-degradation45160177 forced-degradation
45160177 forced-degradationAmit Shah
 
Bioanlytical method development
Bioanlytical method developmentBioanlytical method development
Bioanlytical method development
Sagar Savale
 
Impurity profiling and degradent characterization {presented by shameer m.pha...
Impurity profiling and degradent characterization {presented by shameer m.pha...Impurity profiling and degradent characterization {presented by shameer m.pha...
Impurity profiling and degradent characterization {presented by shameer m.pha...
ShameerAbid
 
ICH-Q2 AMV
ICH-Q2 AMVICH-Q2 AMV
ICH-Q2 AMV
NIHASULTANA2
 
Related Substances-Method Validation-PPT_slide
Related Substances-Method Validation-PPT_slideRelated Substances-Method Validation-PPT_slide
Related Substances-Method Validation-PPT_slideBhanu Prakash N
 
Human liver microsomes & rat liver microsomes
Human liver microsomes & rat liver microsomesHuman liver microsomes & rat liver microsomes
Human liver microsomes & rat liver microsomes
gaurav sharma
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instruments
Sanathoiba Singha
 
Analytical method validation as per USP
Analytical method validation as per USPAnalytical method validation as per USP
Bioanalytical method validation
Bioanalytical method validationBioanalytical method validation
Bioanalytical method validation
Shikha Popali
 
Immunoassay and its appication
Immunoassay and its appicationImmunoassay and its appication
Immunoassay and its appication
Jp Prakash
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
Tausif Momin
 
Analytical method validation, ICH Q2 guideline
Analytical method validation, ICH Q2 guidelineAnalytical method validation, ICH Q2 guideline
Analytical method validation, ICH Q2 guideline
Abhishek Soni
 
analytical method validation and validation of hplc
analytical method validation and validation of hplcanalytical method validation and validation of hplc
analytical method validation and validation of hplc
venkatesh thota
 
Impurities in Drug Substance & in Drug Product
Impurities in Drug Substance & in Drug ProductImpurities in Drug Substance & in Drug Product
Impurities in Drug Substance & in Drug ProductKamal Ambalia
 

What's hot (20)

Qualification of instrumets
Qualification of instrumetsQualification of instrumets
Qualification of instrumets
 
Sample preparation and protocols in metabolite identification
Sample preparation and protocols in metabolite identificationSample preparation and protocols in metabolite identification
Sample preparation and protocols in metabolite identification
 
General considerations and method development in ce,
General considerations and method development in ce,General considerations and method development in ce,
General considerations and method development in ce,
 
Analytical Method Validation as per ICH vs USP
Analytical Method Validation as per ICH vs USPAnalytical Method Validation as per ICH vs USP
Analytical Method Validation as per ICH vs USP
 
M10 bioanalytical method validation
M10 bioanalytical method validationM10 bioanalytical method validation
M10 bioanalytical method validation
 
bio analytical method validation usfda guidlines
bio analytical method validation usfda guidlinesbio analytical method validation usfda guidlines
bio analytical method validation usfda guidlines
 
45160177 forced-degradation
45160177 forced-degradation45160177 forced-degradation
45160177 forced-degradation
 
Bioanlytical method development
Bioanlytical method developmentBioanlytical method development
Bioanlytical method development
 
Impurity profiling and degradent characterization {presented by shameer m.pha...
Impurity profiling and degradent characterization {presented by shameer m.pha...Impurity profiling and degradent characterization {presented by shameer m.pha...
Impurity profiling and degradent characterization {presented by shameer m.pha...
 
ICH-Q2 AMV
ICH-Q2 AMVICH-Q2 AMV
ICH-Q2 AMV
 
Related Substances-Method Validation-PPT_slide
Related Substances-Method Validation-PPT_slideRelated Substances-Method Validation-PPT_slide
Related Substances-Method Validation-PPT_slide
 
Human liver microsomes & rat liver microsomes
Human liver microsomes & rat liver microsomesHuman liver microsomes & rat liver microsomes
Human liver microsomes & rat liver microsomes
 
Qualification of analytical instruments
Qualification of analytical instrumentsQualification of analytical instruments
Qualification of analytical instruments
 
Analytical method validation as per USP
Analytical method validation as per USPAnalytical method validation as per USP
Analytical method validation as per USP
 
Bioanalytical method validation
Bioanalytical method validationBioanalytical method validation
Bioanalytical method validation
 
Immunoassay and its appication
Immunoassay and its appicationImmunoassay and its appication
Immunoassay and its appication
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Analytical method validation, ICH Q2 guideline
Analytical method validation, ICH Q2 guidelineAnalytical method validation, ICH Q2 guideline
Analytical method validation, ICH Q2 guideline
 
analytical method validation and validation of hplc
analytical method validation and validation of hplcanalytical method validation and validation of hplc
analytical method validation and validation of hplc
 
Impurities in Drug Substance & in Drug Product
Impurities in Drug Substance & in Drug ProductImpurities in Drug Substance & in Drug Product
Impurities in Drug Substance & in Drug Product
 

Similar to Bioanalytical methods and its validation

Bioanalytical method validation guidelines for ligand binding assays
Bioanalytical method validation guidelines for ligand binding assaysBioanalytical method validation guidelines for ligand binding assays
Bioanalytical method validation guidelines for ligand binding assays
SHANE_LOBO145
 
method validation as per emea.pptx
method validation as per emea.pptxmethod validation as per emea.pptx
method validation as per emea.pptx
anumalagundam sreekanth
 
Validation of Analytical Methods.pdf
Validation of Analytical Methods.pdfValidation of Analytical Methods.pdf
Validation of Analytical Methods.pdf
anjaneyulu49
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
Devipriya Viswambharan
 
usfda guidelines for bioanalytical method validation.pptx
usfda guidelines for bioanalytical method validation.pptxusfda guidelines for bioanalytical method validation.pptx
usfda guidelines for bioanalytical method validation.pptx
anumalagundam sreekanth
 
Bioanalytical method validation usfda
Bioanalytical method validation usfdaBioanalytical method validation usfda
Bioanalytical method validation usfda
gopinathannsriramachandraeduin
 
method validation as per USFDA.pptx
method validation as per USFDA.pptxmethod validation as per USFDA.pptx
method validation as per USFDA.pptx
anumalagundam sreekanth
 
Analytical Method Validation.pptx
Analytical Method Validation.pptxAnalytical Method Validation.pptx
Analytical Method Validation.pptx
Bholakant raut
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
SUBHASISH DAS
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
roshankhetade2
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
Sai Praveen Reddy
 
QA QC Program for Waste Water Analysis ppt
QA QC Program for Waste Water Analysis pptQA QC Program for Waste Water Analysis ppt
QA QC Program for Waste Water Analysis ppt
KrisnaBagtasos1
 
Analytical Method Validation
Analytical Method ValidationAnalytical Method Validation
Analytical Method Validation
Soumyadarshi Choudhury
 
Analytical method validation 2020
Analytical method validation 2020Analytical method validation 2020
Analytical method validation 2020
PRANJAY PATIL
 
Validation methods.pptx
Validation methods.pptxValidation methods.pptx
Validation methods.pptx
BrindhaLakshmi2
 
Validation of Analytical method.ppt
Validation of Analytical method.pptValidation of Analytical method.ppt
Validation of Analytical method.ppt
Priyanka Yadav
 
HPLC validation.ppt
HPLC validation.pptHPLC validation.ppt
HPLC validation.ppt
Priyanka Yadav
 
Seminar by prakash on validation
Seminar by prakash on validationSeminar by prakash on validation
Seminar by prakash on validation
prakash64742
 

Similar to Bioanalytical methods and its validation (20)

Bioanalytical method validation guidelines for ligand binding assays
Bioanalytical method validation guidelines for ligand binding assaysBioanalytical method validation guidelines for ligand binding assays
Bioanalytical method validation guidelines for ligand binding assays
 
method validation as per emea.pptx
method validation as per emea.pptxmethod validation as per emea.pptx
method validation as per emea.pptx
 
Validation of Analytical Methods.pdf
Validation of Analytical Methods.pdfValidation of Analytical Methods.pdf
Validation of Analytical Methods.pdf
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
usfda guidelines for bioanalytical method validation.pptx
usfda guidelines for bioanalytical method validation.pptxusfda guidelines for bioanalytical method validation.pptx
usfda guidelines for bioanalytical method validation.pptx
 
Bioanalytical method validation usfda
Bioanalytical method validation usfdaBioanalytical method validation usfda
Bioanalytical method validation usfda
 
method validation as per USFDA.pptx
method validation as per USFDA.pptxmethod validation as per USFDA.pptx
method validation as per USFDA.pptx
 
Analytical Method Validation.pptx
Analytical Method Validation.pptxAnalytical Method Validation.pptx
Analytical Method Validation.pptx
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
QA QC Program for Waste Water Analysis ppt
QA QC Program for Waste Water Analysis pptQA QC Program for Waste Water Analysis ppt
QA QC Program for Waste Water Analysis ppt
 
BMV CDER Nishant
BMV CDER NishantBMV CDER Nishant
BMV CDER Nishant
 
Analytical Method Validation
Analytical Method ValidationAnalytical Method Validation
Analytical Method Validation
 
Analytical method validation 2020
Analytical method validation 2020Analytical method validation 2020
Analytical method validation 2020
 
Validation methods.pptx
Validation methods.pptxValidation methods.pptx
Validation methods.pptx
 
Validation of Analytical method.ppt
Validation of Analytical method.pptValidation of Analytical method.ppt
Validation of Analytical method.ppt
 
HPLC validation.ppt
HPLC validation.pptHPLC validation.ppt
HPLC validation.ppt
 
Seminar by prakash on validation
Seminar by prakash on validationSeminar by prakash on validation
Seminar by prakash on validation
 
Qaqc Course Total
Qaqc Course TotalQaqc Course Total
Qaqc Course Total
 

Recently uploaded

ANATOMY AND PHYSIOLOGY OF URINARY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF URINARY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF URINARY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF URINARY SYSTEM.pptx
Swetaba Besh
 
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdf
ARTIFICIAL INTELLIGENCE IN  HEALTHCARE.pdfARTIFICIAL INTELLIGENCE IN  HEALTHCARE.pdf
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdf
Anujkumaranit
 
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptxPharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
Dr. Rabia Inam Gandapore
 
ARTHROLOGY PPT NCISM SYLLABUS AYURVEDA STUDENTS
ARTHROLOGY PPT NCISM SYLLABUS AYURVEDA STUDENTSARTHROLOGY PPT NCISM SYLLABUS AYURVEDA STUDENTS
ARTHROLOGY PPT NCISM SYLLABUS AYURVEDA STUDENTS
Dr. Vinay Pareek
 
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
i3 Health
 
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Savita Shen $i11
 
ACUTE SCROTUM.....pdf. ACUTE SCROTAL CONDITIOND
ACUTE SCROTUM.....pdf. ACUTE SCROTAL CONDITIONDACUTE SCROTUM.....pdf. ACUTE SCROTAL CONDITIOND
ACUTE SCROTUM.....pdf. ACUTE SCROTAL CONDITIOND
DR SETH JOTHAM
 
Flu Vaccine Alert in Bangalore Karnataka
Flu Vaccine Alert in Bangalore KarnatakaFlu Vaccine Alert in Bangalore Karnataka
Flu Vaccine Alert in Bangalore Karnataka
addon Scans
 
263778731218 Abortion Clinic /Pills In Harare ,
263778731218 Abortion Clinic /Pills In Harare ,263778731218 Abortion Clinic /Pills In Harare ,
263778731218 Abortion Clinic /Pills In Harare ,
sisternakatoto
 
KDIGO 2024 guidelines for diabetologists
KDIGO 2024 guidelines for diabetologistsKDIGO 2024 guidelines for diabetologists
KDIGO 2024 guidelines for diabetologists
د.محمود نجيب
 
How to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for DoctorsHow to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for Doctors
LanceCatedral
 
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #GirlsFor Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
Savita Shen $i11
 
Ocular injury ppt Upendra pal optometrist upums saifai etawah
Ocular injury  ppt  Upendra pal  optometrist upums saifai etawahOcular injury  ppt  Upendra pal  optometrist upums saifai etawah
Ocular injury ppt Upendra pal optometrist upums saifai etawah
pal078100
 
Novas diretrizes da OMS para os cuidados perinatais de mais qualidade
Novas diretrizes da OMS para os cuidados perinatais de mais qualidadeNovas diretrizes da OMS para os cuidados perinatais de mais qualidade
Novas diretrizes da OMS para os cuidados perinatais de mais qualidade
Prof. Marcus Renato de Carvalho
 
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 UpakalpaniyaadhyayaCharaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Dr KHALID B.M
 
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.GawadHemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
NephroTube - Dr.Gawad
 
BRACHYTHERAPY OVERVIEW AND APPLICATORS
BRACHYTHERAPY OVERVIEW  AND  APPLICATORSBRACHYTHERAPY OVERVIEW  AND  APPLICATORS
BRACHYTHERAPY OVERVIEW AND APPLICATORS
Krishan Murari
 
heat stroke and heat exhaustion in children
heat stroke and heat exhaustion in childrenheat stroke and heat exhaustion in children
heat stroke and heat exhaustion in children
SumeraAhmad5
 
24 Upakrama.pptx class ppt useful in all
24 Upakrama.pptx class ppt useful in all24 Upakrama.pptx class ppt useful in all
24 Upakrama.pptx class ppt useful in all
DrSathishMS1
 
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
Oleg Kshivets
 

Recently uploaded (20)

ANATOMY AND PHYSIOLOGY OF URINARY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF URINARY SYSTEM.pptxANATOMY AND PHYSIOLOGY OF URINARY SYSTEM.pptx
ANATOMY AND PHYSIOLOGY OF URINARY SYSTEM.pptx
 
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdf
ARTIFICIAL INTELLIGENCE IN  HEALTHCARE.pdfARTIFICIAL INTELLIGENCE IN  HEALTHCARE.pdf
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdf
 
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptxPharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
 
ARTHROLOGY PPT NCISM SYLLABUS AYURVEDA STUDENTS
ARTHROLOGY PPT NCISM SYLLABUS AYURVEDA STUDENTSARTHROLOGY PPT NCISM SYLLABUS AYURVEDA STUDENTS
ARTHROLOGY PPT NCISM SYLLABUS AYURVEDA STUDENTS
 
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
 
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
 
ACUTE SCROTUM.....pdf. ACUTE SCROTAL CONDITIOND
ACUTE SCROTUM.....pdf. ACUTE SCROTAL CONDITIONDACUTE SCROTUM.....pdf. ACUTE SCROTAL CONDITIOND
ACUTE SCROTUM.....pdf. ACUTE SCROTAL CONDITIOND
 
Flu Vaccine Alert in Bangalore Karnataka
Flu Vaccine Alert in Bangalore KarnatakaFlu Vaccine Alert in Bangalore Karnataka
Flu Vaccine Alert in Bangalore Karnataka
 
263778731218 Abortion Clinic /Pills In Harare ,
263778731218 Abortion Clinic /Pills In Harare ,263778731218 Abortion Clinic /Pills In Harare ,
263778731218 Abortion Clinic /Pills In Harare ,
 
KDIGO 2024 guidelines for diabetologists
KDIGO 2024 guidelines for diabetologistsKDIGO 2024 guidelines for diabetologists
KDIGO 2024 guidelines for diabetologists
 
How to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for DoctorsHow to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for Doctors
 
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #GirlsFor Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
For Better Surat #ℂall #Girl Service ❤85270-49040❤ Surat #ℂall #Girls
 
Ocular injury ppt Upendra pal optometrist upums saifai etawah
Ocular injury  ppt  Upendra pal  optometrist upums saifai etawahOcular injury  ppt  Upendra pal  optometrist upums saifai etawah
Ocular injury ppt Upendra pal optometrist upums saifai etawah
 
Novas diretrizes da OMS para os cuidados perinatais de mais qualidade
Novas diretrizes da OMS para os cuidados perinatais de mais qualidadeNovas diretrizes da OMS para os cuidados perinatais de mais qualidade
Novas diretrizes da OMS para os cuidados perinatais de mais qualidade
 
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 UpakalpaniyaadhyayaCharaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
 
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.GawadHemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
 
BRACHYTHERAPY OVERVIEW AND APPLICATORS
BRACHYTHERAPY OVERVIEW  AND  APPLICATORSBRACHYTHERAPY OVERVIEW  AND  APPLICATORS
BRACHYTHERAPY OVERVIEW AND APPLICATORS
 
heat stroke and heat exhaustion in children
heat stroke and heat exhaustion in childrenheat stroke and heat exhaustion in children
heat stroke and heat exhaustion in children
 
24 Upakrama.pptx class ppt useful in all
24 Upakrama.pptx class ppt useful in all24 Upakrama.pptx class ppt useful in all
24 Upakrama.pptx class ppt useful in all
 
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
 

Bioanalytical methods and its validation

  • 1. Bioanalytical method and its validation
  • 4.
  • 5.  Literature search for drugs  Identification of analytical techniques and optimization  Reference standard preparation  Selection of internal standard  Sample pre- treatment (Extraction procedure)  Sample storage  Analysis through suitable technique
  • 6.
  • 9.  A blank biological matrix is spiked with the analytes of interest using solutions of reference standard(s) to prepare calibration standards, QC sample and stability QC sample.  The reference standard used should be obtained from an authentic and traceable source.
  • 10.  The reference standard should be identical to the analyte.  A certificate of analysis (CoA) is required to ensure quality .
  • 11. Selectivity:  Blank samples are obtained from at least 6 individual sources/lots.  Responses detected should not be more than 20% of the analyte response at the LLOQ.
  • 12.  For the investigation of selectivity in lipaemic and haem  olysed matrices at least one source of matrix should be used.
  • 13. Specificity  If the presence of related substances is anticipated in the biological matrix, the impact of such substances should be evaluated during method validation.  Responses detected should not be more than 20% of the analyte response at the LLOQ
  • 14. Matrix effect  The matrix effect should be evaluated by analysing at least 3 replicates of low and high QCs, each prepared using matrix from at least 6 different sources/lots.  Use of fewer sources/lots is acceptable in the case of rare matrices.
  • 15.  The accuracy should be within ±15% of the nominal concentration and the precision percent coefficient of variation (CV) should not be greater than 15% in all individual matrix sources/lots
  • 16. Calibration curve and range  A calibration curve should be generated with a blank sample, a zero sample, and at least 6 concentration levels of calibration standards, including the LLOQ and the ULOQ.
  • 17.  The calibration curve parameters should be reported (slope and intercept in the case of a linear model).
  • 18. Accuracy and precision Preparation of quality control sample  The QCs are intended to mimic study samples and should be prepared by spiking matrix with a known quantity of analyte, stored under the conditions anticipated for study samples and analysed to assess the validity of the analytical method.
  • 19.  During method validation the QCs should be prepared at a minimum of 4 concentration levels within the calibration curve range.
  • 20. Evaluation of accuracy and precision Accuracy and precision should be determined by analysing the QCs. Accuracy and precision should be evaluated using the data.
  • 21. Carry over  Carry-over should be minimised during method development.  Carry-over in the blank samples following the highest calibration standard should not be greater than 20% of the analyte response at the LLOQ
  • 22. Stability  Stability evaluations should be carried out to ensure that every step taken during sample preparation, processing and analysis as well as the storage conditions used do not affect the concentration of the analyte.
  • 23.  The storage and analytical conditions applied to the stability tests, such as the sample storage times and temperatures, sample matrix, anticoagulant and container materials, should reflect those used for the study samples.
  • 24. Reinjection reproducibility  Reproducibility of the method is assessed by replicate measurements of the QCs and is usually included in the assessment of precision and accuracy.  However, if samples could be reinjected, reinjection reproducibility should be evaluated.
  • 25.  These are included in the Validation Report or provided in the Bioanalytical Report of the study where it was conducted.
  • 26.  The analysis of study samples can be carried out after validation has been completed, however, it is understood that some parameters may be completed at a later stage.
  • 27.  By the time the data are submitted to a regulatory authority, the bioanalytical method validation should have been completed.  The study samples, QCs and calibration standards should be processed in accordance with the validated analytical method
  • 29.  The reference standard should be well characterized and documented.  A biological drug has a highly complex structure and its reactivity with binding reagents for bioanalysis may be influenced by a change in the manufacturing process of the drug substance.
  • 30. Specificity  The response of blank samples spiked with related molecules should be below the LLOQ.  The accuracy of the target analyte in presence of related molecules should be within ±25% of the nominal values.
  • 31. Selectivity  Selectivity is evaluated using blank samples obtained from at least 10 individual sources and by spiking the individual blank matrices at the LLOQ.
  • 32.  The accuracy should be within ±25% at the LLOQ of the nominal concentration.
  • 33. Calibration and range  A calibration curve should be generated with at least 6 concentration levels of calibration standards, including LLOQ and ULOQ standards.  The blank sample should not be included in the calculation of calibration curve parameters
  • 34.  The accuracy and precision of concentrations of each calibration standard should be within ±25% of the nominal concentration at the LLOQ and ULOQ.
  • 35. Accuracy and Precision Preparation of quality control sample  During method validation the QCs should be prepared at a minimum of 4 concentration levels within the calibration curve range.
  • 36. Evaluation of accuracy and precision  The accuracy at each concentration level should be within ±20% of the nominal values, except for the LLOQ and ULOQ, which should be within ±25% of the nominal value.  Precision of the QC concentrations determined at each level should not exceed 20%, except at the LLOQ and ULOQ, where it should not exceed 25%.
  • 38.  General and specific SOPs and good record keeping are essential to a properly validated analytical method.  The data generated for bio analytical method validation should be documented  Should be available during data audit and inspection.
  • 39.  ICH, Validation of analytical procedure, International conference on Harmonization, IFPMA, Q2B Validation of Analytical Procedures: Methodology  ICH, International conference on Harmonization, IFPMA, M10 Bioanalytical method validation  Bioanalytical Method Development and Validation: from the USFDA 2001 to the USFDA 2018 Guidance for Industry