1) The document discusses various hyphenated flow injection systems coupled with analytical techniques for the determination of heavy metal ions in water samples.
2) Flow injection analysis coupled with atomic absorption spectrometry, inductively coupled plasma techniques, UV-Vis spectrophotometry, and atomic fluorescence spectrometry have been used to develop sensitive and selective methods for heavy metal analysis.
3) Hyphenating flow injection with detection techniques provides benefits like preconcentration of analytes, reduced analysis time, lower reagent consumption, and improved detection limits and selectivity compared to using detection techniques alone.
Hyphenated Techniques - coupling of a separation technique and an on-line spectroscopic detection technology.
Advantages of hyphenated techniques;
1. Fast and accurate analysis.
2. Higher degree of automation.
3. Higher sample throughput.
4. Better reproducibility.
5. Reduction of contamination due to its closed system.
6. Separation and quantification achieved at same time.
The hyphenated technique is a combination or coupling of two analytical techniques with the help of a proper interface.
The aim of the coupling is to obtain an information-rich detection for both identification and quantification compared to that with a single analytical technique.
Hyphenated Techniques - coupling of a separation technique and an on-line spectroscopic detection technology.
Advantages of hyphenated techniques;
1. Fast and accurate analysis.
2. Higher degree of automation.
3. Higher sample throughput.
4. Better reproducibility.
5. Reduction of contamination due to its closed system.
6. Separation and quantification achieved at same time.
The hyphenated technique is a combination or coupling of two analytical techniques with the help of a proper interface.
The aim of the coupling is to obtain an information-rich detection for both identification and quantification compared to that with a single analytical technique.
Hyphenated technique is a combination or coupling of two analytical techniques with the help of proper interface.
In this presentation Hyphenated techniques-LC-MS/MS, GC-MS/MS, HPTLC-MS has been discussed
Hyphenated techniques have received ever-increasing attention as the principal means to solve complex analytical problems.
Hyphenated techniques are widely used in chemistry and biochemistry and used for both quantitative and qualitative analysis of unknown compounds in complex natural product extracts or fraction and estimation of protein samples also.
an overview of lcms and gcms and its applications....
LC-MS:
It is the combination of liquid chromatography and the mass spectrometry.
* In LC-MS we are removing the detector from the column of LC and fitting the column to interface of MS.
* In the most of the cases the interface used in LC-MS are ionization source
INTERFACE
Apart from being an inlet system for the
MS, an LC–MS interface is also the coupling
of a detector (MS) to a chromatograph.
The choice of LC–MS interface strongly
influences the characteristics of the MS as
a detector for LC. Therefore, we should
keep in mind what characteristics are ideal
for an LC detector
1. Direct liquid introduction (DLI):
The DLI interface was developed in order to solve the problems with in-capillary evaporation in the capillary inlet.
In a DLI interface, the column effluent is nebulized by the disintegration into small droplets of a liquid jet formed at a small diaphragm After desolvation of the droplets in a desolvation chamber, the analytes can be analysed using solvent-mediated CI with the LC solvents as the reagent gas.
Advantages:
• No heat is applied to the interface and it is therefore able to deal with thermally labile materials better than the moving-belt interface.
• The interface contains no moving parts and is cheap and simple to construct and operate and is inherently more reliable than the moving-belt interface.
• Both positive- and negative-ion CI spectra can be generated and the interface provides molecular weight information, plus it can also be used for sensitive quantitative and semi-quantitative procedures.
Disadvantages:
• Involatile compounds are not usually ionized with good efficiency.
• The pinhole is prone to blockage and therefore the system must be kept completely free of solid materials.
• Only a small proportion of the flow from a conventional HPLC column is able to enter the source of the mass spectrometer and sensitivity is consequently low
2. Moving belt/wire:
In a moving-belt interface (MBI), the column effluent is deposited onto an endless moving belt from which the solvent is evaporated by means of gentle heating and efficiently exhausting the solvent vapours. After removal of the solvents, the analyte molecules are (thermally) desorbed from the belt into the ion source and mass analysed.
The MBI for LC.MS was used in a wide variety of applications, including the analysis of drugs and their metabolites, pesticides, steroids, alkaloids, polycyclic, aromatic hydrocarbons and others.
Advantages:
• The interface can be used with a wide range of HPLC conditions.
• The analyst does have some choice of the ionization method to be used; EI, CI and FAB are available, subject to certain limitations, and thus both molecular weight and structural information may be obtained from the analyte(s) under investigation.
1. It is one of the type of Hyphenated technique.
2. It is a combination of gas chromatographic technique and spectroscopic technique.
3. It is having a high resolution capacity.
4. It is used has volatile and Non-volatile compounds.
5. It is used for qualitative and quantitative analysis.
This ppt consist of basic principle of GC-MS, instrumentation of GC-MS, components of GC-MS ,Advantages and disadvantages of GC-MS and application of GC-MS
Hyphenated technique is a combination or coupling of two analytical techniques with the help of proper interface.
In this presentation Hyphenated techniques-LC-MS/MS, GC-MS/MS, HPTLC-MS has been discussed
Hyphenated techniques have received ever-increasing attention as the principal means to solve complex analytical problems.
Hyphenated techniques are widely used in chemistry and biochemistry and used for both quantitative and qualitative analysis of unknown compounds in complex natural product extracts or fraction and estimation of protein samples also.
an overview of lcms and gcms and its applications....
LC-MS:
It is the combination of liquid chromatography and the mass spectrometry.
* In LC-MS we are removing the detector from the column of LC and fitting the column to interface of MS.
* In the most of the cases the interface used in LC-MS are ionization source
INTERFACE
Apart from being an inlet system for the
MS, an LC–MS interface is also the coupling
of a detector (MS) to a chromatograph.
The choice of LC–MS interface strongly
influences the characteristics of the MS as
a detector for LC. Therefore, we should
keep in mind what characteristics are ideal
for an LC detector
1. Direct liquid introduction (DLI):
The DLI interface was developed in order to solve the problems with in-capillary evaporation in the capillary inlet.
In a DLI interface, the column effluent is nebulized by the disintegration into small droplets of a liquid jet formed at a small diaphragm After desolvation of the droplets in a desolvation chamber, the analytes can be analysed using solvent-mediated CI with the LC solvents as the reagent gas.
Advantages:
• No heat is applied to the interface and it is therefore able to deal with thermally labile materials better than the moving-belt interface.
• The interface contains no moving parts and is cheap and simple to construct and operate and is inherently more reliable than the moving-belt interface.
• Both positive- and negative-ion CI spectra can be generated and the interface provides molecular weight information, plus it can also be used for sensitive quantitative and semi-quantitative procedures.
Disadvantages:
• Involatile compounds are not usually ionized with good efficiency.
• The pinhole is prone to blockage and therefore the system must be kept completely free of solid materials.
• Only a small proportion of the flow from a conventional HPLC column is able to enter the source of the mass spectrometer and sensitivity is consequently low
2. Moving belt/wire:
In a moving-belt interface (MBI), the column effluent is deposited onto an endless moving belt from which the solvent is evaporated by means of gentle heating and efficiently exhausting the solvent vapours. After removal of the solvents, the analyte molecules are (thermally) desorbed from the belt into the ion source and mass analysed.
The MBI for LC.MS was used in a wide variety of applications, including the analysis of drugs and their metabolites, pesticides, steroids, alkaloids, polycyclic, aromatic hydrocarbons and others.
Advantages:
• The interface can be used with a wide range of HPLC conditions.
• The analyst does have some choice of the ionization method to be used; EI, CI and FAB are available, subject to certain limitations, and thus both molecular weight and structural information may be obtained from the analyte(s) under investigation.
1. It is one of the type of Hyphenated technique.
2. It is a combination of gas chromatographic technique and spectroscopic technique.
3. It is having a high resolution capacity.
4. It is used has volatile and Non-volatile compounds.
5. It is used for qualitative and quantitative analysis.
This ppt consist of basic principle of GC-MS, instrumentation of GC-MS, components of GC-MS ,Advantages and disadvantages of GC-MS and application of GC-MS
Reagentless iron detection in water based on unclad fiber optical sensorTELKOMNIKA JOURNAL
A simple and low-cost fiber based optical sensor for iron detection is demonstrated in this paper. The sensor head consist of an unclad optical fiber with the unclad length of 1 cm and it has a straight structure. Results obtained shows a linear relationship between the output light intensity and iron concentration, illustrating the functionality of this iron optical sensor. Based on the experimental results, the sensitivity and linearity are achieved at 0.0328/ppm and 0.9824 respectively at the wavelength of 690 nm. With the same wavelength, other performance parameters are also studied. Resolution and limit of detection (LOD) are found to be 0.3049 ppm and 0.0755 ppm correspondingly. This iron sensor is advantageous in that it does not require any reagent for detection, enabling it to be simpler and cost-effective in the implementation of the iron sensing.
Heavy Metals Testing: Health Risks, Methods, Applications, Challenges | GQ Re...GQ Research
This article delves into the importance, methods, applications, and challenges of heavy metals testing, highlighting why it's crucial for safety and compliance in various industries.
Industrial Water Treatment Optimization at Kelly Air Force Basenjcnews777
Co-precipitation gains ground as a popular cost-effective alternative to conventional
treatment of industrial wastewater (Parts 1 & II Environmental Technology
Journal, January to April 1999)
Wastewater from industrial processes such as electroplating, painting, degreasing, and
engine testing and maintenance activities typically contain heavy metals that must be
removed prior to discharge in order to meet NPDES requirements. Although several
technologies are available for heavy metals treatment, many of these systems employ
treatment techniques that are not cost effective. An innovative technology that uses a
process known as co-precipitation has been increasingly recognized as an alternative to
conventional treatment of industrial waste waters, resulting in significant cost benefits to
plant owners and operators.
Co-precipitation occurs when ferrous iron is added to metallic waste streams and
subsequently oxidized in an aerated reactor. The oxidized iron, which is insoluble,
precipitates along with other metallic contaminants present in the waste stream, thereby
enhancing metals removal. The precipitates can then be separated from the treated water
by chemical coagulation, flocculation and clarification processes. The clarified effluent
is subsequently filtered to remove any residual solids prior to discharge.
Optimization of light source wavelength for ammonia detection in waterTELKOMNIKA JOURNAL
Optimization of light source wavelength for ammonia detection in surface water is presented in this work. For the ammonia detection, the surface water sample is mixed with sodium chloride and nessler reagent, whereas the sensor head consists of unclad plastic optical fiber. The unclad region has a length of 1 cm and the cladding is removed by immersing it in acetone solution. Experimental results indicate that the output light intensity of the sensor has linear relationship with the ammonia concentration. At the wavelength of 510 nm, the output light increases linearly as the ammonia concentration varies from 0.07 mg/L to 8.97 mg/L. At the same wavelength, the proposed sensor achieves the sensitivity of 0.0139 (mg/L)-1, accuracy of 99.59% and resolution of 0.72 µg/L. The analysis of light source wavelength reveals that a wavelength range from 450 nm to 580 nm produces the optimized performances. Within this wavelength range, the proposed sensor achieves sensitivity of higher than 0.01 (mg/L)-1, accuracy of higher than 99% and resolution of less than 1 µg/L.
Low-cost colorimetric setup for concentration measurement of manganese ions b...IJECEIAES
This study presents a cost-effective setup for measuring the concentration of Mn2+ ions using colorimetry. The current method involves a calibration curve created with expensive and large commercial laboratory-based instruments, limiting its use in financially constrained situations. To address this issue, the study proposes a low-cost setup consisting of a light-emitting diode and photodiode that utilizes colorimetric and absorbance effects for Mn2+ concentration measurement. Mn2+ colorimetric samples were prepared using the 1-(2-pyridylazo)-2-naphthol (PAN) method with concentrations ranging from 0.2 to 1.0 mg/L. The samples were tested using the proposed setup, followed by a spectrophotometry test to determine the optimal configuration for the setup. The validity of the setup was confirmed by measuring the voltage and calculating the optical absorbance, which exhibited a good correlation with the concentration, consistent with the initial expectation. The correlation coefficient for voltage and absorbance against Mn2+ concentration was found to be 0.9976 and 0.9987, respectively, indicating good linearity and suitability as a calibration curve for Mn2+ detection and measurement. Consequently, the study’s objectives were successfully achieved, and the proposed setup is considered a viable platform for more complex applications, such as real-time monitoring activities.
Plasmonic wave assessment via optomechatronics system for biosensor applicationIJECEIAES
Transduction biosensor (mass-based, optical and electrochemical) involves analysis, recognition and amplification in the acquired sample. In this work, the plasmonic-based biosensor was employed without using tags. It is crucial to determine angles of Brewster (Ɵb) and critical (Ɵc) for generating plasmonic resonance (Ɵr). The objective is to verify a cost-effective plasmonic biosensor through Fresnel simulation and experimentation of a developed optomechatronics system. The borosilicate glass, Au and Air layers were simulated with the Winspall 3.02 simulator. The optomechatronics system consists of: 1-optics (650 nm laser, slit, polarizer, photodiode), 2-mechanical (bipolar stepper motors, gears, stages) and 3-electronics (PIC18F4550, liquid crystal display (LCD) and drivers). Later, the software performs angular interrogation by reading the reflected beam from a rotating prism at 0.1125. Experimentation to simulation accuracy indicates that percentage differences for Ɵr and Ɵc are 1% and 0.2%, respectively. In conclusion, excellence verification was successfully achieved between experimentation and simulation. It proved that the lowcost optomechatronics system is capable and reliable to be deployed for the biosensor application.
Comparative studies on flotation of kasolite using cationic and anionic surfa...eSAT Publishing House
IJRET : International Journal of Research in Engineering and Technology is an international peer reviewed, online journal published by eSAT Publishing House for the enhancement of research in various disciplines of Engineering and Technology. The aim and scope of the journal is to provide an academic medium and an important reference for the advancement and dissemination of research results that support high-level learning, teaching and research in the fields of Engineering and Technology. We bring together Scientists, Academician, Field Engineers, Scholars and Students of related fields of Engineering and Technology
Laser Ablation Molecular Isotopic Spectrometry for rare isotopes of the light...Alexander Bolshakov
Laser ablation molecular isotopic spectrometry (LAMIS) involves measuring isotope-resolved molecular emission. Measurements of several key isotopes (hydrogen, boron, carbon, nitrogen, oxygen, and chlorine) in laser ablation plumes were demonstrated. Requirements for spectral resolution of the optical detection system could be significantly relaxed when the isotopic ratio was determined using chemometric regression models. Multiple applications of LAMIS are anticipated in the nuclear power industry, medical diagnostics and therapies, forensics, carbon sequestration, and agronomy studies.
Ion Beam Analytical Technique PIXE for Pollution Study at Dhaka Van de Graaff...iosrjce
IOSR Journal of Applied Physics (IOSR-JAP) is a double blind peer reviewed International Journal that provides rapid publication (within a month) of articles in all areas of physics and its applications. The journal welcomes publications of high quality papers on theoretical developments and practical applications in applied physics. Original research papers, state-of-the-art reviews, and high quality technical notes are invited for publications.
Natural farming @ Dr. Siddhartha S. Jena.pptxsidjena70
A brief about organic farming/ Natural farming/ Zero budget natural farming/ Subash Palekar Natural farming which keeps us and environment safe and healthy. Next gen Agricultural practices of chemical free farming.
Artificial Reefs by Kuddle Life Foundation - May 2024punit537210
Situated in Pondicherry, India, Kuddle Life Foundation is a charitable, non-profit and non-governmental organization (NGO) dedicated to improving the living standards of coastal communities and simultaneously placing a strong emphasis on the protection of marine ecosystems.
One of the key areas we work in is Artificial Reefs. This presentation captures our journey so far and our learnings. We hope you get as excited about marine conservation and artificial reefs as we are.
Please visit our website: https://kuddlelife.org
Our Instagram channel:
@kuddlelifefoundation
Our Linkedin Page:
https://www.linkedin.com/company/kuddlelifefoundation/
and write to us if you have any questions:
info@kuddlelife.org
Characterization and the Kinetics of drying at the drying oven and with micro...Open Access Research Paper
The objective of this work is to contribute to valorization de Nephelium lappaceum by the characterization of kinetics of drying of seeds of Nephelium lappaceum. The seeds were dehydrated until a constant mass respectively in a drying oven and a microwawe oven. The temperatures and the powers of drying are respectively: 50, 60 and 70°C and 140, 280 and 420 W. The results show that the curves of drying of seeds of Nephelium lappaceum do not present a phase of constant kinetics. The coefficients of diffusion vary between 2.09.10-8 to 2.98. 10-8m-2/s in the interval of 50°C at 70°C and between 4.83×10-07 at 9.04×10-07 m-8/s for the powers going of 140 W with 420 W the relation between Arrhenius and a value of energy of activation of 16.49 kJ. mol-1 expressed the effect of the temperature on effective diffusivity.
UNDERSTANDING WHAT GREEN WASHING IS!.pdfJulietMogola
Many companies today use green washing to lure the public into thinking they are conserving the environment but in real sense they are doing more harm. There have been such several cases from very big companies here in Kenya and also globally. This ranges from various sectors from manufacturing and goes to consumer products. Educating people on greenwashing will enable people to make better choices based on their analysis and not on what they see on marketing sites.
WRI’s brand new “Food Service Playbook for Promoting Sustainable Food Choices” gives food service operators the very latest strategies for creating dining environments that empower consumers to choose sustainable, plant-rich dishes. This research builds off our first guide for food service, now with industry experience and insights from nearly 350 academic trials.
Summary of the Climate and Energy Policy of Australia
A review on application of hyphenated
1. 692 | P a g e
A REVIEW ON APPLICATION OF HYPHENATED
FLOW INJECTION SYSTEM FOR DETERMINATION
OF HEAVY METAL IONS IN WATER SAMPLES
Shelja Tiwari1
, Niharika Sharma2
, Reena Saxena3
1,2,3
Department of Chemistry, Kirori Mal College, University of Delhi, (India)
ABSTRACT
The presence of heavy metal ions even in trace amounts pose an imminent danger to environment and once
these enter the food chain they become threat for both mammals and for aquatic organisms. The industrial
effluents are one of the major reasons responsible for the contamination of environmental system. The direct
determination of trace metal ions using a singular detection technique is a challenging task. Flow injection
analysis has become a popular tool for solving analytical problems. The flow injection system is simple, fast and
a useful technique which is applied in quantitative chemical analysis. This offers many advantages like
minimum reagent consumption, reproducible results, and high sampling frequency. The system is automated
and can be easily coupled to many analytical techniques. The aim of this review is to discuss various methods
based on the hyphenation of flow injection system with detection techniques for the determination of heavy
metal ions in different water samples.
Keywords: Heavy Metal Ions, Flow Analysis, Hyphenation, Flow Injection
I. INTRODUCTION
Heavy metal toxicity is one of the major environmental problems. The toxicity of these metals is associated to
the fact that they are not important from biological aspect but their presence is harmful for proper functioning of
human body. Due to their long existence in environment the treatment of these ions has become extremely
important. Few of these are capable of replacing essential metal ions in human body [1]. The build-up of heavy
metal ions in soil has raise a serious concern in agricultural production as they adversely affects crop growth by
interfering with the biochemical processes of the plants [2]. This is the reason that recently a lot of stress is put
in developing methods for removal of these ions from wastewater. WHO has decided permissible limit for
various metal ions in drinking water [3]. So far many spectrometric techniques have been extensively used for
the detection of metal ions at trace levels. Electrothermal atomic absorption spectrometry (ETAAS) [4] and
flame absorption spectrometry (FAAS) offers better selectivity and are cost effective. Inductively coupled
plasma atomic optical spectrometry (ICP-OES) and inductively coupled plasma mass spectrometry (ICP-MS)
allows simultaneous determination of more than one element and also gives low detection limits [5]. Other than
these techniques UV-Visible spectrophotometry (UV-Vis) [6] and Atomic fluorescence spectrometry (AFS) [7]
have also been used for determination of metal ions. In spite of the several advantages associated to these
techniques operational factors, matrix interferences reduce the effectiveness of these systems. The batch method
2. 693 | P a g e
of detection involves large volume of reagents and is also time consuming. In this method there is a risk of
contamination too. These disadvantages can be overcome by introducing a step before the detection step which
simultaneously allows the enrichment of metal ions such as hyphenation with flow injection preconcentration
techniques. A hyphenated technique is developed by coupling or combining a separation technique with a
spectroscopic technique. This technique involves various combinations like separation-separation,
identification-identification or separation-identification. The advantages of hyphenation are better analysis,
improved sensitivity and selectivity [8]. This technique is automated, fast, requires less volume of reagents,
lowers the risk of contamination and can easily be coupled to various detection techniques [9]. The flow
injection analysis was proposed by Ruzicka and Hansen in1975 [10].
Flow injection (FI) systems are considered as a very important tool which enhances the ability of analytical
techniques. The presence of microcolumn in the flow injection system provides better selectivity and sensitivity
to the system after the optimising various flow and chemical variables. The FI separation methods can be
differentiated on the basis of the type of interface across which the transfer is taking place. The types of
separation systems are given in Fig. 1 [11].
Fig.1 Classification of FI separation techniques
The FI separation technique in conjunction with a detection technique provides a mean to overcome the
difficulties and enhances the selectivity of a singular detection technique. Various analytical techniques
hyphenated to FI have been routinely used for the determination of metal ions in different environmental
matrices (Fig. 2). The aim of this review is to discuss latest research based on the hyphenation of FI system with
various detection techniques.
3. 694 | P a g e
Fig. 2 Hyphenation of Flow Injection with Different Analytical Techniques
II. FLOW INJECTION SYSTEM HYPHENATED TO ATOMIC ABSORPTION
SPECTROMETRIC TECHNIQUES
Atomic absorption spectrometry is a widely used technique for the quantitative determination of metal ions at
trace levels. The sample here is generally volatilized either by furnace or by flame. In this a specific wavelength
of light is adsorbed by an element which corresponds to the energy required to promote the electrons from one
energy level to another higher level.
2.1 Fi-Faas
The flame atomic absorption spectrometer (FAAS) comprises a flow-through detector which fed the sample
directly into the nebulizer-burner system via suction process. The characteristics of the FAAS system can be
significantly enhanced by introducing sample by the FI mode. The hyphenation of FI-FAAS leads to decrease in
sample consumption and preconcentrates the analyte of interest. The other advantages associated to this coupled
technique are lower time of analysis, removal of matrix ions at preconcentration stage, direct introduction of
sample to nebulizer from FI system there by lowering the risk of contamination. Recently, several FI-FAAS
systems have been proposed for heavy metal preconcentration in industrial water samples. Most of these
methods generally use solid phase extractant as a method of preconcentration. In 2012, xylenol orange
immobilized on Amberlite XAD-2 was used in a FI-FAAS system to perconcentrate cadmium in industrial
water samples [12]. The method suggested here showed good results in terms of sensitivity, selectivity in
comparison to other reported methods. Recently an Innovative flow injection flame atomic absorption
spectrometric method was suggested for the determination of Cu(II) in industrial samples [13]. Amberlite XAD-
2 modified with dithizone was used as a complexing agent. The reported system was very simple and was
capable of detecting Cu(II) in presence of high concentration of interfering ions. The suggested method was
automated and offered better sensitivity in terms of detection limit. This FI-FAAS method has been also
4. 695 | P a g e
exploited for the speciation of metal ions. The importance of this can be highlighted by the fact that for certain
metal ions one of their oxidation state is considered to be essential while other can be harmful to living system.
In this direction, a method for the speciation of chromium species was proposed using FI-FAAS [14]. This paper
has described a system for the speciation and preconcentration of chromium species by synthesizing dithizone
anchored Dowex Optipore L493 system. The suggested method was applied for the determination of both
Cr(III) and Cr(VI) in industrial water samples. The reported method was fast and required very low volume of
reagents and has offered better concentration efficiency as well. A schematic representation for the working of
FI-FAAS preconcentration system is given in Fig. 3 [14]. Other than these many FI-FAAS systems have been
recently developed for heavy metal preconcentration in different water samples [15, 16]. Various ion exchange
resins can also be directly used in FI-FAAS system for the removal of metal ions [17]. It is not necessary to use
polymeric support for preconcentration, many systems with silica [18, 19], alumina [20], nano materials [21]
etc. has also been used with FI-FAAS system for heavy metal analysis in different water samples.
Fig.3 A Schematic Representation for the Working of FI-FAAS Preconcentration System
2.2 Fi-Gfaas
Graphite furnace atomic absorption spectrometry which is also known as Electrothermal Atomic Absorption
Spectrometry is also a very sensitive spectroscopic technique. It is a single element technique used for trace
metal analysis in various solid and aqueous samples. The coupling of FI system with GFAAS results in
improvement in the overall performance of this technique in terms of sensitivity and selectivity [22]. A system
was proposed for the preconcentration of palladium using chelating resin microcolumn [1,5-Bis(2-pyridyl)-3-
sulphophenylmethylene thiocarbonohydrazide immobilized on an anion-exchange resin (Dowex 1 X8-200)]
[23]. This paper has shown the advantages of hyphenated system over singular technique. A fully automated
5. 696 | P a g e
system without any complex hardware was used. The method reported was fast, selective and was relatively free
from interferences making it appropriate for palladium determination in different samples. Recently, a
biosorption system was proposed for the preconcentration of cadmium in environmental samples [24]. In this
work metallothionein isolated from rabbit liver and recombined cyanobacteria metallothionein were anchored
on spherical SiO2 particles. The biosorption behaviours of these for the sorption of Cd(II) were studied. The
results obtained with recombined cyanobacteria metallothionein were better. For a range of 5-100 ng L -1
and
sample volume of 1 mL, enrichment factor of 13.8 was observed. Other than this excellent detection limit of 1.4
ng L-1
was achieved.
III. FLOW INJECTION SYSTEM HYPHENATED TO INDUCTIVE COUPLED PLASMA
TECHNIQUES
In Inductive coupled plasma gas plasma provides a very high temperature excitation source for atomic
spectrometry. The Inductive coupled plasma (ICP) spectrometer are similar to FAAS system in its sample
introduction system. The optimum uptake rate of ICP is less than FAAS systems. The advantage of ICP system
over FAAS is its capability for multi-elemental detection. Inductively couples plasma-mass spectrometry (ICP-
MS) and inductively coupled plasma-optical emission spectrometry (ICP-OES) are most frequently used for
multielemental detection. In both ICP-OES and ICP-MS, the sample is first nebulised to an aerosol and then it is
atomised and ionised in plasma. ICP-OES distinguishes between different elements by the specific light
emission which is induces as a result of the excitation of atom ions in the plasma while a mass spectrometer is
used in ICP-MS for separation of ions of different elements. When combined to FI ICP becomes a more
versatile technique. This combination of FI-ICP minimizes time of analysis and labour and lowers risk of
contamination. In 2012, a review was published on the recent advancements in the area of on-line solid-phase
pre-concentration for inductively-coupled plasma techniques for the determination of mineral elements [25].
This review has discussed the advantages of the hyphenated SPE system over singular system in terms of
selectivity, sensitivity, sampling frequency etc.
3.1 FI-ICP-MS
ICP-MS is a great analytical technique which is capable of providing simultaneous multielement/isotope
analysis with good sensitivity. Still, the direct determination of trace metal ions in different forms in aqueous
system is difficult. To overcome this, a preconcentration step is required which can be done by coupling it with
a FI system [26]. In 2011, an FI-ICP-MS system was proposed for the speciation of heavy metal ions in fresh
water samples [27]. A fully automated approach was suggested for discrimination in dissolved labile and stable
metal-complex ions. Better detection limits in the range of from 0.005 to 0.186 μg L−1
were observed. The
method was found to be promising for the online determination of metal-organic complexes in fresh water
samples. In another automated FI-ICP-MS system based on solid phase extraction was suggested for the
preconcentration of metal ions in sea-water and other water samples [28]. This method was based on the
sorption of analytes onto a minicolumn filled with a chelating resin, 1,5-bis (2-pyridyl)-3-sulphophenyl
methylene thiocarbonohydrazide immobilized on aminopropyl-controlled pore glass which was placed in the
injection valve of FI manifold. The method reported was fast, automatic and showed good sensitivity. In other
6. 697 | P a g e
method for speciation of Sb(III) and Sb(V) in seawater by FI-ICP-MS was reported [29]. Here, a chelating resin
[1,5-bis(2-pyridyl)-3-sulfophenyl methylene] thio carbonohydrazide immobilised on amino propyl-controlled
pore glass (550 Å, PSTH-cpg) and an anion exchanger (Amberlite IRA-910) were used for the speciation of the
two species. The proposed method was free of complex treatment procedures with organic solvents which can
change the concentration of antimony species. No contamination of analytes took place as entire system was
closed.
3.2 Fi-Icp-Aes/Oes
A solid phase extraction system was proposed for the determination of zinc in natural water using FI-ICP-OES.
This paper has discussed the importance of FI preconcentration system in terms of low consumption of samples
and reagents, high analytical throughput and minimal waste production [30]. The use of solid phase extractant
for preconcentration of analyte is also highlighted. In another system SPE based fractionation method using FI-
ICP-AES system was proposed for speciation of metal ions in aqueous samples [31]. The use of FI system has
provided preconcentration of different species resulting in increase in sensitivity thus improving the detection
limits. The coupling of FI-ICP-AES system has improved the detection limits drastically in comparison to
conventional ICP-AES system. The automated control of FI system has also lowered the risk of contamination
resulting in good reproducibility of results. Another advantage discussed here is as the whole process of
separation and preconcentration is taking place in single step the sample throughput is also improved. Carbon
nanotubes (CNTs) in a FI preconcentration system along with ICP-AES have also been used for heavy metal
detection [32]. The sorption behaviour of CNTs was compared for Cd(II) preconcentration. To enhance the
efficiency of system the authors have used ultrasonic nebulisation (USN). In some FI systems CNTs generate
high back pressure. This can be solved by using non-compacted minicolumns. The coupling with FI has made
the system fast, increased sample throughput and enhanced systems analytical performance.
IV. FLOW INJECTION SYSTEM HYPHENATED TO UV-VISIBLE
SPECTROPHOTOMETRIC TECHNIQUE
In UV-Visible absorbance depends on both light absorption and diffraction. It varies with the wavelength of
light and size of the particles. In organic molecules it depends on the composition of the chromophore [33]. UV-
Visible spectrophotometers are most frequently coupled with FI system as a detector. To hyphenate UV-Vis
with FI the intensity the cuvette is substituted with a flow through cell [6]. A novel procedure for cadmium
determination was proposed [34]. This system was based on the complex formation between Cd(II), iodide ions,
and 2-[2-(4-dimethylamino-phenyl)-vinyl]-1-ethyl-quinolinium iodide (R), commonly known as Quinaldine red.
The reported method was quick and precise and can be applied to real samples with complex matrices. The
advantages of the proposed system were that no preconcentration step such as solvent or solid phase extraction
was required and no surfactant was added for solubilisation. The consumptions of reagents were less and more
reproducible results were achieved with the FI system. In another recently developed system a dispersive liquid-
liquid microextraction procedure was proposed for determination of copper in environmental water samples
[35]. The concentration of copper in water sample is low so a preconcentration step was required. Other than the
development of the preconcentration procedure, focus was drawn on the automation of the system. The
7. 698 | P a g e
hyphenation with FI system was done to make the process automated. By introducing FI technique solutions can
be injected as well as re-directed and use of columns can introduce selectivity too. The entire system is
computer controlled from solution injection to transport step so this increases the reproducibility of the
analytical system. The developed system was based on the complex formation between copper and
bathocuprine. The suggested system was fast and sensitive in terms of sensitivity, reagent consumption and
linear range.
V. FLOW INJECTION SYSTEM HYPHENATED TO ATOMIC FLOURESCENCE
SPECTROMETRIC TECHNIQUE
Cold vapour generation atomic fluorescence spectrometric (CVGSFS) technique is mostly used for the
determination of mercury speciation in various samples. It’s a selective, low cost and a very sensitive technique
for the determination of mercury. Before entering in the cold vapour stage the organomercury is converted to
Hg(II) by using an oxidation system. This system is a new green aspect as it avoids use of carcinogenic
chemicals for mercury digestion [36]. In 2015, an analytical system was described for the speciation of mercury
at ppq (pg L-1
) level [37]. The system was very sensitive and selective for mercury determination in water
samples. Other advantages of the reported method were that it was matrix independent, cost effective,
automated and high preconcentration factor was achieved. A schematic representation of FI-HPLC-CVGAFS
system is given in Fig. 4. In another system, a novel analytical method was proposed using hyphenation of FI
system with AFS for the speciation of mercury at ultratrace level [38]. The method was reliable, cost-effective
and green. This system has photochemical vapour generation which provides system more stability and
enhances the analytical characteristics as it is free from interferences. Here, mercury was first reacted with
diethyldithiocarbamate forming a hydrophobic compound which is then preconcentrated using FI system. Then
it was in situ converted into cold vapour using UV irradiation and then these cold vapors are passed to AFS for
detection. This method can be applied for the speciation of Hg2+
and MeHg+
respectively.
Fig. 4 A Schematic Representation of FI-HPLC-CVGAFS System.
8. 699 | P a g e
VI. CONCLUSION
The determination of heavy metal ions in environmental samples is a crucial task. The toxic nature of these
species poses a threat for both mammals and environment. The need of a sensitive and selective analytical
method thus has become demand of time and nature. Few of the oxidation states of many elements are essential
to us as well thus, a technique with the property of speciation will be extremely important. Many singular
techniques have found application in the trace analysis of metal ions. However the increase in interferences
arising from matrix ions or extremely low concentration of analyte makes the direct determination of these ions
a challenging task. The hyphenation with FI system has proven to solve these issues and has made these
techniques extremely selective and sensitive. The sensitivity of these techniques has reached the level of ppq (pg
L-1
) as a result of coupling to FI system. In Table 1 a comparison of detection limits between singular and
hyphenated techniques is given. As it can be clearly seen that the detection limit improves significantly after
coupling the singular analytical system with FI system. Not just the sensitivity is enhanced but the system has
also become automated, computer controlled program makes it even more efficient and less vulnerable to errors
as well. The green aspect can also be introduced as it lowers the consumption of reagents. Thus, the hyphenation
of analytical techniques to FI system has become one of the most important systems in ultratrace analysis of
metal ions in various matrices.
Table 1 Comparison of Detection Limits of Singular and Hyphenated Techniques for Various
Metal Determination
Detection
technique
Metal ion
Detection Limit
(µg L-1
) References
Singular technique Hyphenated technique
FAAS Cr 60 0.13 14
GFAAS Pd 50 2 23
ICP-MS Cd 0.07 0.004 28
ICP-OES Zn 0.1 0.08 31
AFS Hg 0.003 4 x 10-5
37
VII. ACKNOWLEDGEMENTS
The authors are thankful to Department of Science and Technology (DST) and University Grants
Commission (UGC), New Delhi, India for financial support.
9. 700 | P a g e
REFERENCES
[1] M. Jaishankar, T.Tseten, N. Anbalagan, B.B. Mathew and K.N. Beeregowda, Toxicity, mechanism and
health effects of some heavy metals, Interdisciplinary Toxicology, 7(2), 2014, 60–72.
[2] P. C. Nagajyoti, K. D. Lee and T. V. M. Sreekanth, Heavy metals, occurrence and toxicity for plants: a
review, Environment Chemistry Letters, 8, 2010, 199–216.
[3] World Health Organisation, Guidelines for Drinking-water Quality, 4th edn, 2011.
[4] J.D. Butcher, Advances in electrothermal atomization atomic absorption spectrometry: Instrumentation,
methods, and applications, Applied Spectroscopy Reviews, 4, 2006, 15–34.
[5] B. Hu, S. Li, G. Xiang, M. He, and Z. Jiang, Recent progress in electrothermal vaporization–inductively
coupled plasma atomic emission spectrometry and inductively coupled plasma mass spectrometry, Applied
Spectroscopy Reviews, 42, 2007, 203–234.
[6] Z. L. Fang, Flow Injection Atomic Absorption Spectrometry (John Wiley & Sons: Chichester, UK, 1995).
[7] X. Wen, Q.Yang, Z. Yan and Q. Deng, Determination of cadmium and copper in water and food samples
by dispersive liquid–liquid microextraction combined with UV–vis spectrophotometry, Microchemical
Journal, 97(2), 2011, 249–254.
[8] P. Phalke and S. Kavade, Review on Hyphenated Techniques, International Journal of Chemical Studies,
1(3), 2013, 157-165.
[9] T-L. Denga, Y-W. Chenb and N. Belzileb, Antimony speciation at ultra trace levels using hydride
generation atomic fluorescence spectrometry and 8-hydroxyquinoline as an efficient masking agent,
Analytica Chimica Acta, 432(2), 2001, 293–302
[10] Ruzicka, J., and E. H. Hansen. 1975. Flow injection analyses 1. New concept of fast continuous-flow
analysis, Analytica Chimica Acta, 78(1), 1975, 145–157.
[11] Z. Fang, Flow Injection Separation and Preconcentration (VCH Publishers, Inc., New York, 1973).
[12] R. Saxena and S. Saxena, Flow injection preconcentration system using a new functionalized resin for
determination of cadmium (II) by flame atomic absorption spectroscopy, Indian Journal of Chemistry,
51A, 2012, 1567-1573.
[13] R. Saxena, P. L. Meena and S.Tiwari, Determination of Copper in Industrial Water by Innovative Flow
Injection Flame Atomic Absorption Spectrometry, Instrumentation Science and Technology, DOI:
10.1080/10739149.2015.1089275.
[14] R. Saxena, S. Tiwari and N. Sharma, Flow-injection solid phase extraction using Dowex Optipore L493
loaded with dithizone for preconcentration of chromium species from industrial waters and determination
by FAAS, RSC Advances, 5, 2015, 69196–69204.
[15] R. Saxena and P.L. Meena, Flow injection online solid phase extraction system using Amberlite XAD-16
functionalized with 8-hydroxyquinoline for copper and zinc determination by flame atomic absorption
spectrometry, RSC Advances, 4, 2014, 20216-20225.
[16] T .Çetina, Ş. Tokalıoğlua, A. Ülgena, S. Şahana, İ. Özentürk and C. Soykan, Synthesis/characterization of
a new chelating resin and on-line solid phase extraction for the determination of Ag(I) and Pd(II) from
10. 701 | P a g e
water, cream, anode slime and converter samples by flow injection flame atomic absorption spectrometry,
Talanta, 105, 2013, 340–346.
[17] T. Daşbaşı, Ş. Saçmacı, S. Şahan, A. Ülgen and Ş. Kartal, An on-line separation and preconcentration
system coupled with flame atomic absorption spectrometry for the determination of lead, Analytical
Methods, 5, 2013, 3307-3313.
[18] S-Y. Zhou, N. Song, S-X. Liu, D-X. Chen, Q. Jia and Y-W. Yang, Separation and preconcentration of
gold and palladium ions with a carboxylated pillar [5] arene derived sorbent prior to their determination by
flow injection FAAS, Microchimica Acta, 181, 2014, 1551-1556.
[19] J.Waluvanaruk, W. Aeungmaitrepirom, T. Tuntulani and P.Ngamukot, Preconcentration and
Determination of Trace Silver Ion Using Benzothiazole Calix[4]arene Modified Silica by Flow Injection
Flame Atomic Absorption Spectrometry, Analytical Sciences, 30(3) , 2014, 389-395.
[20] A. M.H. Shabani, S. Dadfarnia and Z. Dehghani, On-line solid phase extraction system using 1, 10-
phenanthroline immobilized on surfactant coated alumina for the flame atomic absorption spectrometric
determination of copper and cadmium, Talanta, 79(4), 2009, 1066–1070.
[21] S. Dadfarnia, A.M.H. Shabani, E. Kazemi, S. A. H. Khormizi and F. Tammadon, Synthesis of Nano-Pore
Size Ag(I)-Imprinted Polymer for the Extraction and Preconcentration of Silver Ions Followed by Its
Determination with Flame Atomic Absorption Spectrometry and Spectrophotometry Using Localized
Surface Plasmon Resonance Peak of Silver Nanoparticles, Journal of Brazilian Chemical Society, 26(6),
2015, 1180-1190.
[22] Y-H. Zhang, M. Wang, X-G. Su, T. Zheng, H-Q Zhang and Q-H. Jin, Flow Injection Semi-online
Preconcentration Graphite Furnace Atomic Absorption Spectrometry for Determination of Cadmium,
Copper and Manganese, Chemical Research in Chinese Universities, 18(1), 2002, 1-7.
[23] C. B. Ojeda, F.S. Rojas, and J. M.C. Pavòn On-line preconcentration of palladium(II) using a
microcolumn packed with a chelating resin, and its subsequent determination by graphite furnace atomic
absorption spectrometry, Microchimica Acta, 158, 2007, 103–110.
[24] T. Yang, L-Y. Ma, M-Li. Chen and J-H. Wang, Metallothionein isoforms for selective biosorption and
preconcentration of cadmium at ultra-trace levels, Journal of Analytical Atomic Spectrometry, 30, 2015,
929-935.
[25] D. Das, M. Dutta, M. L. Cervera and M. de la Guardia, Recent advances in on-line solid-phase pre-
concentration for inductively-coupled plasma techniques for determination of mineral elements, Trends in
Analytical Chemistry, 33, 2012 35-45.
[26] J.A. Caruso, B. Klaue, B. Michalke and D.M. Rocke, Group assessment: elemental speciation,
Ecotoxicology and Environmental Safety, 56, 2003, 32–44.
[27] T-T. Shih, W-Yu.Tseng, K-H. Tsai, W-Yu. Chen, M-W. Tsai and Y-C. Sun, Online coupling of ultraviolet
titanium dioxide film reactor with poly(methyl methacrylate) solid phase extraction–inductively coupled
plasma mass spectrometry for speciation of trace heavy metals in freshwater, Microchemical Journal, 99,
2011, 260–266.
11. 702 | P a g e
[28] I. S.Trujillo, Elisa V. Alonso, A. G. de Torres and J. M.C. Pavón, Development of a solid phase extraction
method for the multielement determination of trace metals in natural waters including sea-water by FI-
ICP-MS, Microchemical Journal, 101, 2012, 87–94.
[29] A. C. Fornieles, A. G. de Torres, E. V. Alonso, M. T. S. Cordero and J. M. C. Pavón, Speciation of
antimony(III) and antimony(V) in seawater by flow injection solid phase extraction coupled with online
hydride generation inductively coupled plasma mass spectrometry, Journal of Analytical Atomic
Spectrometry, 26, 2011, 1619-1626.
[30] L. A. Escudero , L. D. Martinez, J. A. Salonia and J. A. Gasquez, Determination of Zn(II) in natural
waters by ICP-OES with on-line preconcentration using a simple solid phase extraction system,
Microchemical Journal, 95, 2010, 164–168.
[31] C. Puls and A. Limbeck, Determination of trace metal fractionation in aqueous solutions using a solid
phase extraction flow injection system on-line coupled to ICP-AES, Journal of Analytical Atomic
Spectrometry, 24, 2009, 1434–1440.
[32] B. P. M. Savio , L. D. Martinez, R. A. Gil and P. Smichowski, Study of carbon nanotubes and
functionalized-carbon nanotubes as substrates for flow injection solid phase extraction associated to
inductively coupled plasma with ultrasonic nebulisation Application to Cd monitoring in solid
environmental samples, Microchemical Journal, 98 , 2011, 225–230.
[33] M. Baalousha, B. Stolpe and J.R. Lead, Flow field-flow fractionation for the analysis and characterization
of natural colloids and manufactured nanoparticles in environmental systems: A critical review, Journal of
Chromatography A, 1218 , 2011, 4078– 4103.
[34] J. Škrlíková, V. Andruch, H. Sklenáŕová, P. Solich, I. S. Balogh, and F. Billes, A novel non-extractive
sequential injection procedure for determination of cadmium, Analytical Letters, 44, 2011, 431–445.
[35] B. Horstkotte, M. Alexović, F. Maya, C. M. Duarte, V. Andruch and V. Cerdá, Automatic determination
of copper by in-syringe dispersive liquid–liquid microextraction of its bathocuproine-complex using long
path-length spectrophotometric detection, Talanta, 99, 2012, 349-356.
[36] V. Angeli, S. Biagi, S. Ghimenti, M. Onor, A. D'Ulivo and E. Bramanti, Flow injection-chemical vapor
generation atomic fluorescence spectrometry hyphenated system for organic mercury determination: A
step forward, Spectrochimica Acta Part B, 66, 2011, 799–804.
[37] C-C. Brombach, B. Chen, W.T. Corns, J. Feldmann and E. M. Krupp, Methylmercury in water samples at
the pg/L level by online preconcentration liquid chromatography cold vapor-atomic fluorescence
spectrometry, Spectrochimica Acta Part B, 105, 2015, 103–108.
[38] Y. Gao, W. Yang, C. Zheng, X. Hou and L.Wu, On-line preconcentration and in situ photochemical vapor
generation in coiled reactor for speciation analysis of mercury and methylmercury by atomic fluorescence
spectrometry, Journal of Analytical Atomic Spectrometry, 26, 2011, 126-132.