This document discusses various types of artifacts that can occur in MRI images and their causes and remedies. It covers technique-related artifacts such as chemical shift artifacts, Gibbs artifacts, aliasing artifacts, and magic angle artifacts. It also discusses patient-related artifacts like motion artifacts, metal artifacts, and flow artifacts. System-related artifacts discussed include shimming artifacts, gradient artifacts, and radiofrequency-related artifacts. For each type of artifact, the etiology, manifestation, and tips for remedying the artifact are provided. The document uses images to demonstrate examples of artifacts and their effects on MRI scans.
Mri spin echo pulse sequences its variations andYashawant Yadav
MRI spin echo pulse sequences its variation and applications , in this slide collection principle of spine echo pulse sequences is described with physic behind it ,, this slide also coves the inverse recovery pulse sequences and types ,,,, image weighting and parameters are explained .. hope it may be help ful.
Mri spin echo pulse sequences its variations andYashawant Yadav
MRI spin echo pulse sequences its variation and applications , in this slide collection principle of spine echo pulse sequences is described with physic behind it ,, this slide also coves the inverse recovery pulse sequences and types ,,,, image weighting and parameters are explained .. hope it may be help ful.
The gradient echo pulse sequence is the simplest type of MRI sequence.
The major purposes behind the gradient technique is a significant reduction in scan time. Small variable flip angle are employed , usually less than 90 degrees. which in turn allow very short repetition time thus decreasing the scan time.
Gradient echo pulse sequence differ from spin echo pulse sequence . There is no 180 degree pulse in GRE. T2 relaxation in GRE is called as T2* relaxation. Gradient can be used to either dephase or rephase the magnetic moments of nuclei.
The gradient echo pulse sequence is the simplest type of MRI sequence.
The major purposes behind the gradient technique is a significant reduction in scan time. Small variable flip angle are employed , usually less than 90 degrees. which in turn allow very short repetition time thus decreasing the scan time.
Gradient echo pulse sequence differ from spin echo pulse sequence . There is no 180 degree pulse in GRE. T2 relaxation in GRE is called as T2* relaxation. Gradient can be used to either dephase or rephase the magnetic moments of nuclei.
A review of advances in Brachytherapy treatment planning and delivery in last decade or so, with main focus on brachytherapy for Prostate cancer, Breast cancer and Cervical cancer
Geometric distortion of functional MR images especially of DWI hinders the visualization of the frontal & temporal regions. This is a simple & reliable approach to get rid of. This needs no special software or hardware & applies to old systems.
Magnetic resonance imaging (MRI) is a medical imaging technique that uses a magnetic field and computer-generated radio waves to create detailed images of the organs and tissues in your body.
Most MRI machines are large, tube-shaped magnets. When you lie inside an MRI machine, the magnetic field temporarily realigns water molecules in your body. Radio waves cause these aligned atoms to produce faint signals, which are used to create cross-sectional MRI images — like slices in a loaf of bread.
The MRI machine can also produce 3D images that can be viewed from different angles.
Products & Services
Sign up for Email: Get Your Free Resource – Coping with Cancer
Why it's done
MRI is a noninvasive way for your doctor to examine your organs, tissues and skeletal system. It produces high-resolution images of the inside of the body that help diagnose a variety of problems.
MRI of the brain and spinal cord
MRI is the most frequently used imaging test of the brain and spinal cord. It's often performed to help diagnose:
Aneurysms of cerebral vessels
Disorders of the eye and inner ear
Multiple sclerosis
Spinal cord disorders
Stroke
Tumors
Brain injury from trauma
A special type of MRI is the functional MRI of the brain (fMRI). It produces images of blood flow to certain areas of the brain. It can be used to examine the brain's anatomy and determine which parts of the brain are handling critical functions.
This helps identify important language and movement control areas in the brains of people being considered for brain surgery. Functional MRI can also be used to assess damage from a head injury or from disorders such as Alzheimer's disease.
MRI of the heart and blood vessels
MRI that focuses on the heart or blood vessels can assess:
Size and function of the heart's chambers
Thickness and movement of the walls of the heart
Extent of damage caused by heart attacks or heart disease
Structural problems in the aorta, such as aneurysms or dissections
Inflammation or blockages in the blood vessels
This presentation reviews the common artifacts in EEG, their identification and rectification. Examples of various artifacts are provided in the presentation.
This presentation looks at EEG signal generation, pyramidal cells, recording of EEG, source localisation, polarity, analysis of dipole, derivations, montages,
Magnetic resonance imaging (MRI) is a medical imaging technique that uses a magnetic field and computer-generated radio waves to create detailed images of the organs and tissues in your body.
Most MRI machines are large, tube-shaped magnets. When you lie inside an MRI machine, the magnetic field temporarily realigns water molecules in your body. Radio waves cause these aligned atoms to produce faint signals, which are used to create cross-sectional MRI images — like slices in a loaf of bread.
The MRI machine can also produce 3D images that can be viewed from different angles.
Products & Services
Sign up for Email: Get Your Free Resource – Coping with Cancer
Why it's done
MRI is a noninvasive way for your doctor to examine your organs, tissues and skeletal system. It produces high-resolution images of the inside of the body that help diagnose a variety of problems.
MRI of the brain and spinal cord
MRI is the most frequently used imaging test of the brain and spinal cord. It's often performed to help diagnose:
Aneurysms of cerebral vessels
Disorders of the eye and inner ear
Multiple sclerosis
Spinal cord disorders
Stroke
Tumors
Brain injury from trauma
A special type of MRI is the functional MRI of the brain (fMRI). It produces images of blood flow to certain areas of the brain. It can be used to examine the brain's anatomy and determine which parts of the brain are handling critical functions.
This helps identify important language and movement control areas in the brains of people being considered for brain surgery. Functional MRI can also be used to assess damage from a head injury or from disorders such as Alzheimer's disease.
MRI of the heart and blood vessels
MRI that focuses on the heart or blood vessels can assess:
Size and function of the heart's chambers
Thickness and movement of the walls of the heart
Extent of damage caused by heart attacks or heart disease
Structural problems in the aorta, such as aneurysms or dissections
Inflammation or blockages in the blood vessels
Graphic record heart sound - Phonogram.
Recording the sounds connected with the pumping action of heart.
Sound from heart – phonocardiogram
Instrument to measure this – phonocardiograph
Basic function – to pick up the different heart sound,filter the required and display.
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journeygreendigital
Tom Selleck, an enduring figure in Hollywood. has captivated audiences for decades with his rugged charm, iconic moustache. and memorable roles in television and film. From his breakout role as Thomas Magnum in Magnum P.I. to his current portrayal of Frank Reagan in Blue Bloods. Selleck's career has spanned over 50 years. But beyond his professional achievements. fans have often been curious about Tom Selleck Health. especially as he has aged in the public eye.
Follow us on: Pinterest
Introduction
Many have been interested in Tom Selleck health. not only because of his enduring presence on screen but also because of the challenges. and lifestyle choices he has faced and made over the years. This article delves into the various aspects of Tom Selleck health. exploring his fitness regimen, diet, mental health. and the challenges he has encountered as he ages. We'll look at how he maintains his well-being. the health issues he has faced, and his approach to ageing .
Early Life and Career
Childhood and Athletic Beginnings
Tom Selleck was born on January 29, 1945, in Detroit, Michigan, and grew up in Sherman Oaks, California. From an early age, he was involved in sports, particularly basketball. which played a significant role in his physical development. His athletic pursuits continued into college. where he attended the University of Southern California (USC) on a basketball scholarship. This early involvement in sports laid a strong foundation for his physical health and disciplined lifestyle.
Transition to Acting
Selleck's transition from an athlete to an actor came with its physical demands. His first significant role in "Magnum P.I." required him to perform various stunts and maintain a fit appearance. This role, which he played from 1980 to 1988. necessitated a rigorous fitness routine to meet the show's demands. setting the stage for his long-term commitment to health and wellness.
Fitness Regimen
Workout Routine
Tom Selleck health and fitness regimen has evolved. adapting to his changing roles and age. During his "Magnum, P.I." days. Selleck's workouts were intense and focused on building and maintaining muscle mass. His routine included weightlifting, cardiovascular exercises. and specific training for the stunts he performed on the show.
Selleck adjusted his fitness routine as he aged to suit his body's needs. Today, his workouts focus on maintaining flexibility, strength, and cardiovascular health. He incorporates low-impact exercises such as swimming, walking, and light weightlifting. This balanced approach helps him stay fit without putting undue strain on his joints and muscles.
Importance of Flexibility and Mobility
In recent years, Selleck has emphasized the importance of flexibility and mobility in his fitness regimen. Understanding the natural decline in muscle mass and joint flexibility with age. he includes stretching and yoga in his routine. These practices help prevent injuries, improve posture, and maintain mobilit
263778731218 Abortion Clinic /Pills In Harare ,sisternakatoto
263778731218 Abortion Clinic /Pills In Harare ,ABORTION WOMEN’S CLINIC +27730423979 IN women clinic we believe that every woman should be able to make choices in her pregnancy. Our job is to provide compassionate care, safety,affordable and confidential services. That’s why we have won the trust from all generations of women all over the world. we use non surgical method(Abortion pills) to terminate…Dr.LISA +27730423979women Clinic is committed to providing the highest quality of obstetrical and gynecological care to women of all ages. Our dedicated staff aim to treat each patient and her health concerns with compassion and respect.Our dedicated group ABORTION WOMEN’S CLINIC +27730423979 IN women clinic we believe that every woman should be able to make choices in her pregnancy. Our job is to provide compassionate care, safety,affordable and confidential services. That’s why we have won the trust from all generations of women all over the world. we use non surgical method(Abortion pills) to terminate…Dr.LISA +27730423979women Clinic is committed to providing the highest quality of obstetrical and gynecological care to women of all ages. Our dedicated staff aim to treat each patient and her health concerns with compassion and respect.Our dedicated group of receptionists, nurses, and physicians have worked together as a teamof receptionists, nurses, and physicians have worked together as a team wwww.lisywomensclinic.co.za/
Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...VarunMahajani
Disruption of blood supply to lung alveoli due to blockage of one or more pulmonary blood vessels is called as Pulmonary thromboembolism. In this presentation we will discuss its causes, types and its management in depth.
Title: Sense of Smell
Presenter: Dr. Faiza, Assistant Professor of Physiology
Qualifications:
MBBS (Best Graduate, AIMC Lahore)
FCPS Physiology
ICMT, CHPE, DHPE (STMU)
MPH (GC University, Faisalabad)
MBA (Virtual University of Pakistan)
Learning Objectives:
Describe the primary categories of smells and the concept of odor blindness.
Explain the structure and location of the olfactory membrane and mucosa, including the types and roles of cells involved in olfaction.
Describe the pathway and mechanisms of olfactory signal transmission from the olfactory receptors to the brain.
Illustrate the biochemical cascade triggered by odorant binding to olfactory receptors, including the role of G-proteins and second messengers in generating an action potential.
Identify different types of olfactory disorders such as anosmia, hyposmia, hyperosmia, and dysosmia, including their potential causes.
Key Topics:
Olfactory Genes:
3% of the human genome accounts for olfactory genes.
400 genes for odorant receptors.
Olfactory Membrane:
Located in the superior part of the nasal cavity.
Medially: Folds downward along the superior septum.
Laterally: Folds over the superior turbinate and upper surface of the middle turbinate.
Total surface area: 5-10 square centimeters.
Olfactory Mucosa:
Olfactory Cells: Bipolar nerve cells derived from the CNS (100 million), with 4-25 olfactory cilia per cell.
Sustentacular Cells: Produce mucus and maintain ionic and molecular environment.
Basal Cells: Replace worn-out olfactory cells with an average lifespan of 1-2 months.
Bowman’s Gland: Secretes mucus.
Stimulation of Olfactory Cells:
Odorant dissolves in mucus and attaches to receptors on olfactory cilia.
Involves a cascade effect through G-proteins and second messengers, leading to depolarization and action potential generation in the olfactory nerve.
Quality of a Good Odorant:
Small (3-20 Carbon atoms), volatile, water-soluble, and lipid-soluble.
Facilitated by odorant-binding proteins in mucus.
Membrane Potential and Action Potential:
Resting membrane potential: -55mV.
Action potential frequency in the olfactory nerve increases with odorant strength.
Adaptation Towards the Sense of Smell:
Rapid adaptation within the first second, with further slow adaptation.
Psychological adaptation greater than receptor adaptation, involving feedback inhibition from the central nervous system.
Primary Sensations of Smell:
Camphoraceous, Musky, Floral, Pepperminty, Ethereal, Pungent, Putrid.
Odor Detection Threshold:
Examples: Hydrogen sulfide (0.0005 ppm), Methyl-mercaptan (0.002 ppm).
Some toxic substances are odorless at lethal concentrations.
Characteristics of Smell:
Odor blindness for single substances due to lack of appropriate receptor protein.
Behavioral and emotional influences of smell.
Transmission of Olfactory Signals:
From olfactory cells to glomeruli in the olfactory bulb, involving lateral inhibition.
Primitive, less old, and new olfactory systems with different path
Prix Galien International 2024 Forum ProgramLevi Shapiro
June 20, 2024, Prix Galien International and Jerusalem Ethics Forum in ROME. Detailed agenda including panels:
- ADVANCES IN CARDIOLOGY: A NEW PARADIGM IS COMING
- WOMEN’S HEALTH: FERTILITY PRESERVATION
- WHAT’S NEW IN THE TREATMENT OF INFECTIOUS,
ONCOLOGICAL AND INFLAMMATORY SKIN DISEASES?
- ARTIFICIAL INTELLIGENCE AND ETHICS
- GENE THERAPY
- BEYOND BORDERS: GLOBAL INITIATIVES FOR DEMOCRATIZING LIFE SCIENCE TECHNOLOGIES AND PROMOTING ACCESS TO HEALTHCARE
- ETHICAL CHALLENGES IN LIFE SCIENCES
- Prix Galien International Awards Ceremony
micro teaching on communication m.sc nursing.pdfAnurag Sharma
Microteaching is a unique model of practice teaching. It is a viable instrument for the. desired change in the teaching behavior or the behavior potential which, in specified types of real. classroom situations, tends to facilitate the achievement of specified types of objectives.
Ozempic: Preoperative Management of Patients on GLP-1 Receptor Agonists Saeid Safari
Preoperative Management of Patients on GLP-1 Receptor Agonists like Ozempic and Semiglutide
ASA GUIDELINE
NYSORA Guideline
2 Case Reports of Gastric Ultrasound
These lecture slides, by Dr Sidra Arshad, offer a quick overview of physiological basis of a normal electrocardiogram.
Learning objectives:
1. Define an electrocardiogram (ECG) and electrocardiography
2. Describe how dipoles generated by the heart produce the waveforms of the ECG
3. Describe the components of a normal electrocardiogram of a typical bipolar leads (limb II)
4. Differentiate between intervals and segments
5. Enlist some common indications for obtaining an ECG
Study Resources:
1. Chapter 11, Guyton and Hall Textbook of Medical Physiology, 14th edition
2. Chapter 9, Human Physiology - From Cells to Systems, Lauralee Sherwood, 9th edition
3. Chapter 29, Ganong’s Review of Medical Physiology, 26th edition
4. Electrocardiogram, StatPearls - https://www.ncbi.nlm.nih.gov/books/NBK549803/
5. ECG in Medical Practice by ABM Abdullah, 4th edition
6. ECG Basics, http://www.nataliescasebook.com/tag/e-c-g-basics
NVBDCP.pptx Nation vector borne disease control programSapna Thakur
NVBDCP was launched in 2003-2004 . Vector-Borne Disease: Disease that results from an infection transmitted to humans and other animals by blood-feeding arthropods, such as mosquitoes, ticks, and fleas. Examples of vector-borne diseases include Dengue fever, West Nile Virus, Lyme disease, and malaria.
Ethanol (CH3CH2OH), or beverage alcohol, is a two-carbon alcohol
that is rapidly distributed in the body and brain. Ethanol alters many
neurochemical systems and has rewarding and addictive properties. It
is the oldest recreational drug and likely contributes to more morbidity,
mortality, and public health costs than all illicit drugs combined. The
5th edition of the Diagnostic and Statistical Manual of Mental Disorders
(DSM-5) integrates alcohol abuse and alcohol dependence into a single
disorder called alcohol use disorder (AUD), with mild, moderate,
and severe subclassifications (American Psychiatric Association, 2013).
In the DSM-5, all types of substance abuse and dependence have been
combined into a single substance use disorder (SUD) on a continuum
from mild to severe. A diagnosis of AUD requires that at least two of
the 11 DSM-5 behaviors be present within a 12-month period (mild
AUD: 2–3 criteria; moderate AUD: 4–5 criteria; severe AUD: 6–11 criteria).
The four main behavioral effects of AUD are impaired control over
drinking, negative social consequences, risky use, and altered physiological
effects (tolerance, withdrawal). This chapter presents an overview
of the prevalence and harmful consequences of AUD in the U.S.,
the systemic nature of the disease, neurocircuitry and stages of AUD,
comorbidities, fetal alcohol spectrum disorders, genetic risk factors, and
pharmacotherapies for AUD.
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...Oleg Kshivets
RESULTS: Overall life span (LS) was 2252.1±1742.5 days and cumulative 5-year survival (5YS) reached 73.2%, 10 years – 64.8%, 20 years – 42.5%. 513 LCP lived more than 5 years (LS=3124.6±1525.6 days), 148 LCP – more than 10 years (LS=5054.4±1504.1 days).199 LCP died because of LC (LS=562.7±374.5 days). 5YS of LCP after bi/lobectomies was significantly superior in comparison with LCP after pneumonectomies (78.1% vs.63.7%, P=0.00001 by log-rank test). AT significantly improved 5YS (66.3% vs. 34.8%) (P=0.00000 by log-rank test) only for LCP with N1-2. Cox modeling displayed that 5YS of LCP significantly depended on: phase transition (PT) early-invasive LC in terms of synergetics, PT N0—N12, cell ratio factors (ratio between cancer cells- CC and blood cells subpopulations), G1-3, histology, glucose, AT, blood cell circuit, prothrombin index, heparin tolerance, recalcification time (P=0.000-0.038). Neural networks, genetic algorithm selection and bootstrap simulation revealed relationships between 5YS and PT early-invasive LC (rank=1), PT N0—N12 (rank=2), thrombocytes/CC (3), erythrocytes/CC (4), eosinophils/CC (5), healthy cells/CC (6), lymphocytes/CC (7), segmented neutrophils/CC (8), stick neutrophils/CC (9), monocytes/CC (10); leucocytes/CC (11). Correct prediction of 5YS was 100% by neural networks computing (area under ROC curve=1.0; error=0.0).
CONCLUSIONS: 5YS of LCP after radical procedures significantly depended on: 1) PT early-invasive cancer; 2) PT N0--N12; 3) cell ratio factors; 4) blood cell circuit; 5) biochemical factors; 6) hemostasis system; 7) AT; 8) LC characteristics; 9) LC cell dynamics; 10) surgery type: lobectomy/pneumonectomy; 11) anthropometric data. Optimal diagnosis and treatment strategies for LC are: 1) screening and early detection of LC; 2) availability of experienced thoracic surgeons because of complexity of radical procedures; 3) aggressive en block surgery and adequate lymph node dissection for completeness; 4) precise prediction; 5) adjuvant chemoimmunoradiotherapy for LCP with unfavorable prognosis.
Recomendações da OMS sobre cuidados maternos e neonatais para uma experiência pós-natal positiva.
Em consonância com os ODS – Objetivos do Desenvolvimento Sustentável e a Estratégia Global para a Saúde das Mulheres, Crianças e Adolescentes, e aplicando uma abordagem baseada nos direitos humanos, os esforços de cuidados pós-natais devem expandir-se para além da cobertura e da simples sobrevivência, de modo a incluir cuidados de qualidade.
Estas diretrizes visam melhorar a qualidade dos cuidados pós-natais essenciais e de rotina prestados às mulheres e aos recém-nascidos, com o objetivo final de melhorar a saúde e o bem-estar materno e neonatal.
Uma “experiência pós-natal positiva” é um resultado importante para todas as mulheres que dão à luz e para os seus recém-nascidos, estabelecendo as bases para a melhoria da saúde e do bem-estar a curto e longo prazo. Uma experiência pós-natal positiva é definida como aquela em que as mulheres, pessoas que gestam, os recém-nascidos, os casais, os pais, os cuidadores e as famílias recebem informação consistente, garantia e apoio de profissionais de saúde motivados; e onde um sistema de saúde flexível e com recursos reconheça as necessidades das mulheres e dos bebês e respeite o seu contexto cultural.
Estas diretrizes consolidadas apresentam algumas recomendações novas e já bem fundamentadas sobre cuidados pós-natais de rotina para mulheres e neonatos que recebem cuidados no pós-parto em unidades de saúde ou na comunidade, independentemente dos recursos disponíveis.
É fornecido um conjunto abrangente de recomendações para cuidados durante o período puerperal, com ênfase nos cuidados essenciais que todas as mulheres e recém-nascidos devem receber, e com a devida atenção à qualidade dos cuidados; isto é, a entrega e a experiência do cuidado recebido. Estas diretrizes atualizam e ampliam as recomendações da OMS de 2014 sobre cuidados pós-natais da mãe e do recém-nascido e complementam as atuais diretrizes da OMS sobre a gestão de complicações pós-natais.
O estabelecimento da amamentação e o manejo das principais intercorrências é contemplada.
Recomendamos muito.
Vamos discutir essas recomendações no nosso curso de pós-graduação em Aleitamento no Instituto Ciclos.
Esta publicação só está disponível em inglês até o momento.
Prof. Marcus Renato de Carvalho
www.agostodourado.com
3. Technique Related Artifacts
Fat & water
protons exist in
the same Voxel
)Chemical Shift(
Fat & water
protons exist
in
•The same
Voxel
)Gibbs –Fine
Line or
Truncation(
Planning Fault
)Magic Angle &
Cross Talk(
FOV smaller
than the
anatomy
inside the
receive coil
)Aliasing(
Experta Medica
Experta Medica
4. Artifacts
Wrap around, Fold over,
Aliasing or Back folding :
Etiology :
Anatomy inside the receive coil is greater than the prescribed FOV.
The coil, then receives signal from the anatomy outside the FOV.
The signal of these data oversampled is plotted during readout in
the opposite direction because the system fills the k-space lines
in one direction. It fills the K-space lines first with the prescribed
FOV. Then the oversampled area is then plotted on the already
filled first lines.
Manifested as :
Image data out side (FOV) wraps around it (in opposite side of
the image)
Occurs in any of the 3 encoding axes, hence the names phase-
wrap, frequency-wrap & slice wrap. Slice-wrap occurs only in
Fourier 3D acquisitions.
Experta Medica
6. This case study shows a 3Dacquisition technique with aliasing )in the red circle(
in the slice selection direction. The image of the upper leg wraps into the image of
the lower leg.
Experta Medica
7. Remedy Tips:
For frequency-wrap,
Use larger FOV .
Oversample the frequency.
Use digital filters to eliminate signal with frequency out
side the (FOV).
For Phase-wrap,
Use larger FOV- Phase oversampling- No Phase Wrap
Option.
For 3D or Slice-wrap,
Decrease the number of slices/slab, Use a larger FOV.
Experta Medica
8. Chemical shift - Chemical Mis-registration –India
Ink Artifact
Etiology:
Occurrence of water & fat Protons in the same Voxel as occurs in :
Abdomen – Spine – Orbits where in low TE’s , fat & water protons
precess at nearby frequencies resulting in spin- spin resonance
causing positional shift (Mis-registration) of fat signal along the
frequency axis. This is especially evident in long TR sequences
especially when TE is short ( PD WI).
Manifested as :
Bright & Dark demarcation at fat/ water interface.
Remedy Tips:
Increase the band width to help separate the peaks of water & fat
signals.
Decrease the ETL as every Echo contains its innate chemical shift &
increasing the ETL accumulates the shifts.
Use a Lower field strength.
Increase the TE & reduce the TR.
Experta Medica
10. Chemical shift - Chemical Mis-registration - Odd
Presented-Complicated with Susceptibility
Etiology:
Dental fillings in the frequency direction as occurs in the brain where
in low TE’s , fat & water protons precess at nearby frequencies
resulting in spin- spin resonance causing positional shift (Mis-
registration) of fat signal along the frequency axis. This is further
complicated by different magnetic susceptibility of dentures. This is
especially evident in long TR sequences especially when TE is short
( PD WI & FLAIR).
Manifested as :
Bright crescent near the middle of some images near the skull base.
Remedy Tips:
Increase the band width to help separate the peaks of water & fat
signals.
Swap the frequency/phase axes.
Decrease the ETL as every Echo contains its innate chemical shift &
increasing the ETL accumulates the shifts.
Use a Lower field strength. Experta Medica
12. Gibbs, Fine line or Truncation :
Etiology:
Periodic brain & visceral motion produces different phases registered
as fine lines parallel to the anatomy borders along the phase
encoding axis.
Manifested as :
- Bright and dark lines that may be seen parallel and adjacent
to boarders of the anatomy in areas containing both fat &
water as the CSF/ spinal cord interface
The artifact occurs in the phase encoding direction
Remedy Tips:
Increase the matrix & use an isotropic pixel .
Use a fine line filter
Swap the Frequency/ Phase axes.
Decrease the Echo Sampling time.
Use even NEX.
Experta Medica
15. Magic angle artifacts:
Manifested as:
Bright signal commonly seen in tendons & ligaments of the knee
& shoulder joints that are oriented at magic angle (55°) to the
Bo.
Etiology :
Normally, signal from water molecules associated with the
tendon collagen fibers is not seen due to dipolar interactions
that result in very short T2 Times…..But at an angle of about 55°
to the main magnetic field, the dipolar interactions become
zero, resulting in an increase of the T2 Times about 100 folds.
This results in signal being visible in tendons with ordinary
pulse sequences especially at short TE long TR SE & FSE
sequences i.e. PD WI.
Remedy Tips:
*Change the positioning and/or the planning angle.
*Scan in multiple planes & multiple pulse sequences.
*Increase the TE & reduce the TR in PD sequences.
Experta Medica
17. Cross Talk artifacts:
Manifested as:
Dark signal commonly seen at areas of intersection as in
axial angulated spine.
Etiology:
At the areas of intersection as in axial angulated spine, the
excitation pulse of the 2nd
image set saturates the signal
of the spins at the areas of intersection which were already
excited in the first image set.
Remedy Tips:
Change the planning angle.
Use FSE & fr-FSE.
Increase the TE & reduce the TR in PD sequences.
Use interleaved rather than sequential image acquisition.
Experta Medica
18. Regional Fat Sat Failure
Manifested as:
Areas where fat appears bright in fat sat images
(inhomogeneous fat saturation).
Etiology:
Mal positioning of the shim volume.
Using too large FOV leading to improper shimming.
Improper use of the receivers of phase array coils.
Remedy Tips:
Centralize the shim volume to the anatomy of interest.
Decrease the FOV.
Review the active receiver s of phased array coil .
Slightly shift the central frequency or transmit gain .
Change the chemical shift tuning factor (CSTUN). Experta Medica
19. Failed Fat Saturation
SAG FSE + Fat Sat: Autoshim On, without moving the Autoshim volume.
Autoshim failed: Default shim values = 1 0 0
Experta Medica
20. Annefact
Manifested as:
Striated bands along the S/I axis in sag images.
Etiology:
Mal positioning of the FOV to the used receivers of phased
array coil.
Improper selection of the receivers of phase array coils.
Remedy Tips:
Centralize the S/I FOV in front of the active receiver set.
Use the adequate number of receivers suitable for the
FOV.
22. Patient Related Artifacts
Motion (Ghosting)
Artifacts
Metal (Susceptibility) Artifact
Physical & PhysiologicalPhysical & Physiological
Motion of the patientMotion of the patient
produces multipleproduces multiple
time-resolved phasestime-resolved phases
that are manifested asthat are manifested as
GhostsGhosts
Due to (microscopic
gradients) or variations
in the magnetic field
strength that occurs
near the interfaces of
substance of different
magnetic susceptibility
Experta Medica
23. Motion (Ghosting) Artifacts:
Etiology :
Patient physical motion .
Periodic physiological motions like:
Cardiac motion ,Respiratory motion ,Swallowing, Vascular & CSF
pulsations & Peristalsis.
Manifested as:
Repeating contour figures oriented in the phase direction as
every motion creates a new different phase spatially shifted
slightly apart from the preceding one.
Remedy Tips:
Stabilize your Patient.
Trigger the Pulse Sequence to the physiological motion (Respiratory,
cardiac & peripheral gating).
Reduce the ETL in FSE to minimize phase shifting.
Use antispasmodic drugs to scan the abdomen.
Use breath hold pulse sequences.
Use Flow Compensation.
Use Spatial pre-saturation.
Swap the Phase/ Frequency axes.
Put the source of artifact in the phase direction.
Experta Medica
27. Vessel Stepping or Banding
Etiology :
Intra- voxel spin de-phasing results in low signal at the
slice or slab interface especially from the moving spins
in the vessels. Vascular Pulsations exaggerate this
effect.
Manifested as:
Bands or demarcations at the slice or slab
interface.
Remedy Tips:
Use Flow Compensation, cardiac or peripheral gating &
ramp RF pulse.
Experta Medica
29. Flow Artifacts
Etiology :
Fully magnetized spins entering the slice in the vessels may
possess brighter signal than the tissue spins ( Flow
Enhancement) or
Fresh non excited spins may enter the imaging volume during
readout producing signal void. Also, the excited protons may
flow outside the imaging volume before read producing a
signal void. These two states are known as TOF artifact.
Position encoding of the Voxel containing the vessel in the
phase direction at a time factor of TE/2 before frequency
encoding.
Manifested as:
Either Flow enhancement or void or as ghosting.
Remedy Tips:
Use Flow Compensation, cardiac or peripheral gating.
Increase TE.
Use SPGR.
Use S/I Spatial saturation outside the FOV. Experta Medica
33. Susceptibility Artifacts
Etiology :
Due to (microscopic gradients) or variations in the magnetic field
strength that occur near the interfaces of substance of different
magnetic susceptibility. Large susceptibility artifacts are
commonly seen surrounding ferromagnetic objects inside of
diamagnetic materials (such as the human body). These
gradients cause de-phasing of spins and frequency shifts of the
surrounding tissues.
Manifested as:
Bright and dark areas with spatial distortion of surrounding
anatomy especially with Long TE SE & GRE pulse sequences.
Remedy Tips:
Use Fast Sequences ( fr-FSE, f-GRE & f-SPGR, FIESTA,SWAN,
SWIFT).
Reduce the TE.
Swap the Frequency/ Phase axes.
Use S/I Spatial saturation outside the FOV or at the different
susceptibility area.
Experta Medica
36. N/2 & Geometric
Distortion ArtifactsEtiology :
Due to high B values used in DWI, the MPG pulse duration
increases & hence minimum TE increases. With the following
consequences:
Increase of gradient amplitude/Voxel increasing Microscopic
Gradients & thereby increasing the magnetic susceptibility
artifacts.
Enhancing of ghosting resulting from Periodic Brain Motion.
Exaggerating the Phase Error Buildup due to Off- Resonance
Spins
These effects are further complicated by:
Partial Volume Averaging due to occurrence of multiple signals
in the same Voxel.
Eddy Currents due to high amplitude gradients with resultant
Feed through artifacts.
Obligatory A/P phase direction to prevent Peripheral Nerve
stimulation.
Narrow Receive Bandwidth slows down the sampling rate
thereby enhancing Ghosting & increasing Chemical Shift.
Occurrence of tissues of different susceptibilities in the Frontal
& Temporal regions.
Experta Medica
39. N/2 & Geometric Distortion
Artifacts
Manifested as:
Bright and dark areas with huge spatial distortion of the
anatomy especially in the frontal area obscuring parts of
the frontal & temporal lobes.
Remedy Tips:
Use S/I Spatial saturation outside the FOV or at the different
susceptibility area. This increases the dielectric effect but
reduces Susceptibility effects. ( Tradeoff: SAR increases).
Reduce the Echo Sampling Time by:
1-Decreasing B value with trade off of Poor Diff weighting &
appearance of T2 Shine through artifact.
2- Widening of RBW with trade off of low SNR.
3-Optimizing TE to allow maximum gradient amplitude at
minimum possible TE.
4-Ramp Sampling allowing echo sampling during ramp ups &
downs of the frequency encoding gradient, thereby reducing
ech0spacing & allowing faster sampling with tradeoff of wide
RBW & hence reduced SNR. Experta Medica
40. 5-Using an additional refocusing pulse to reduce the eddy
currents with a tradeoff of increasing the minimum TE.
6-Using Parallel Imaging techniques making use of spatial
information related to the spatially varying sensitivity of
different receive channels of the receive coil. Using these pre-
measured data, sensitivity profiles of the coil helps reconstruct
the K-Space trajectory missing lines which is under sampled
during the actual scan .This Tech is called ASSET or SMASH.
• SENSE is another parallel imaging technique where K-Space is
under sampled by widening the K-Space lines . The overflowing
data / reduced number of lines leads to aliasing. Each pixel of
an aliased receiver channel image reflects the signal from
different spatial origins. The premeasured sensitivity profile
helps allocate each signal to its spatial origin. The trade off is
low SNR due to lossy K-Space.
7-Using PROPELLER where K-Space trajectory is rotated across
the time domain to define the areas of K-Space where distortion
occurred & subtract them from the data set with a tradeoff of
limited SNR because of lossy K-Space.
Experta Medica
43. Shimming Related Artifacts
Etiology :
Defective shimming causes interaction with unwanted
waves of external sources that may cause spatial or
intensity distortions or even both.
Manifested as:
Spatial ( geometric ) or intensity distortions or both.
Zebra or totally clipped images due to faulty K-Space
Pixel.
Remedy Tips:
Use auto shim at the first sequence of the exam.
Filter the currents in the field vicinity.
Active Shimming may be advisable if problems
persist.
47. Gradient Related Artifacts
Etiology :
Faulty gradients cause local spatial or intensity distortions
or even both in the direction of the faulty gradient. Moiré
Fringes are due to B0 inhomogeneity from one side of the
body to the other where signals of different phases
superimpose . The in-phase signals add to each other
where the off-phase signals subtract.
Manifested as:
Spatial ( geometric ) or intensity distortions or both. Moiré
fringes are manifested as 2 window image i.e. half the
image being bright & the other half being dark.
Remedy Tips:
Swap Freq/ Phase axes.
Filter the currents in the field vicinity.
Use FSE/ fr-FSE rather than SE or GRE.
Use surface coils.
50. The arrows point to the signal
differences in this T2 weighted
image with fat suppression (SPIR(.
Imaging the edge of an object causes
static noise. Also the small part
provides no enough signal.
51. RF-Related Artifacts
Etiology :
Faulty RF- Amplifier or Receiver.
Manifested as:
Zipper (RF Feed through) artifacts in cases of faulty RF
amplifier. Normally, the RF amplifier should stop working
when read out gradient is applied. When faulty, it works
while readout( F. encoding gradient) is applied. Residual
FID stimulated echoes may also produce zippers.
Bright spot of increasing intensity in the center of the
image in cases of constant offset of the DC in the signal
perception channels of the receiver coil. Narrow band
noise appears perpendicular to the frequency encoding
axis while broad band noise disfigures the image over a
wide zone.
Remedy Tips:
Coil tuning.
Filter the currents in the field vicinity.