The gradient echo pulse sequence is the simplest type of MRI sequence.
The major purposes behind the gradient technique is a significant reduction in scan time. Small variable flip angle are employed , usually less than 90 degrees. which in turn allow very short repetition time thus decreasing the scan time.
Gradient echo pulse sequence differ from spin echo pulse sequence . There is no 180 degree pulse in GRE. T2 relaxation in GRE is called as T2* relaxation. Gradient can be used to either dephase or rephase the magnetic moments of nuclei.
2. GRADIENTS
Gradients are generated by coils of wire situated within the bore of the
magnet.
This magnetic field interacts with the main magnetic field so that the
magnetic field strength along the axis of gradient coil is altered in a
linear way.
The middle of the axis of the gradient remains at the field strength of
main magnetic field.
This is called magnetic isocentre.
3. Apart from localization, gradients are also useful in the gradient echo
sequences.
Used for spoiling or rewinding TM.
Rephasing of TM is done by gradients thus they eliminate 180 degree
pulse and make GRE sequences much faster.
4. Basic three differences between SE and
GRE sequences
1. There is no 180 degree pulse in GRE. Rephasing of TM in GRE is done by gradients;
particularly by reversal of frequency encoding gradient. Since rephasing by gradient gives
signal this sequence is called as Gradient echo sequence.
2. Flip angle in GRE are smaller, usually less than 90 degree. Since flip angle is smaller there
will be early recovery of LM such that TR can be reduced hence scanning time.
3. Transverse relaxation in SE sequence is caused by combination of two mechanisms
A. Irreversible dephasing of TM - SPIN SPIN RELAXATION.
B. Dephasing caused by magnetic field inhomogeneity.
In SE sequence the dephasing caused by inhomogeneity is eliminated by 180 degree pulse.
Hence there is ‘true’ relaxation in SE sequence
6. INTRODUCTION TO CONVENTIONAL
GRADIENT ECHO PULSE SEQUENCE
The gradient echo pulse sequence is the simplest type of MRI sequence.
The major purposes behind the gradient technique is a significant reduction
in scan time.
Small variable flip angle are employed , usually less than 90 degrees. which in
turn allow very short repetition time thus decreasing the scan time.
Gradient echo pulse sequence differ from spin echo pulse sequence .
There is no 180 degree pulse in GRE.
T2 relaxation in GRE is called as T2* relaxation.
Gradient can be used to either dephase or rephase the magnetic moments of
nuclei.
8. TYPICAL VALUES IN GRE IMAGING
Long TR: 100ms+
Short TR: less than 50ms
Short TE: 15-25ms
Low flip angles: 5-20 degree
Large flip angles: 70-80 degree
9. USES:
Acquire T2 * , T1 and proton density weighting.
Reduction in the scan time.
Single - slice or volume breath - hold acquisitions in the
abdomen, and for dynamic contrast enhancement
Used to produce angiographic - type images (TOF).
10. WEIGHTING AND CONTRAST IN GRADIENT
ECHO
Three different processes affect weighting in gradient echo pulse
sequences.
1. Extrinsic parameters ( TR, TE and flip angle affect image
weighting and contrast).
2. The steady state
3. Residual transverse magnetization
11. STEADY STATE
It is defined as the stable condition that does not change over time.
The overall energy of hydrogen remains constant as the energy ‘ in’ as
determined by the flip angle equals the energy ‘ out ’ as determined by
the TR.
Energy in = RF excitation pulse.(FLIP ANGLE)
Energy out= energy lost through spin lattice energy transfer (TR)
Stage where the TR is shorter than the T1 and T2 times of the tissue.
Flip angle of 30-45 degree and TR of 20- 50 ms achieves this state.
There are, therefore, critical values of flip angle and TR to maintain the
steady state.
The steady state is a state in which not only LM and TM are present at the
same time but also their value or magnitude is maintained steady during
data acquisition.
13. RESIDUAL TRANSVERSE MAGNETISATION
The transverse magnetisation produced as a result of previous
excitations is called the residual transverse
magnetization(RTM).
but remains over several TR periods in the transverse plane. It
is called the residual transverse magnetization and it affects
image contrast as it induces a voltage in the receiver coil.
It affects image contrast as it results in tissues with long T2
times (such as water), appearing bright on the image
14. How is steady state achieved?
Steady state is achieved by keeping TR shorter than T1 and T2
times of tissues .Since TR is shorter than T2 there is no time for
TM to decay completely, before next RF pulse excitation. So, there
will be some residual TM left over.
Second factor contributing for SS is low flip angle usually 30-45
degree. Since NMV is flipped by less than 90 degrees there is
always some residual LM left.
15. How magnitudes of LM and TM are
maintained steady?
Residual TM is flipped by gradients through 180 degree from along positive
side of Y-axis to negative side of Y-axis. When next RF excitation is done this
residual TM along negative side of Y-axis is moved towards Z-axis. This adds
up into magnitude of residual LM, which now increases. At the same time
some part of the residual LM is flipped by same flip angle so that TM is
formed along the positive side of Y-axis. This cycle is repeated with every
RF excitation and values of LM and TM are maintained.
16. Why do we want steady state?
With steady state shortest TR and scan time are achieved.
It also affects image contrast as it causes tissues with long
T2 appear more bright on the image.
17. Types of Steady State
Depending on what is done to residual TM after gradient
echo is received, SS can be of three types-
1. Coherent (in-phase) residual TM
2. Incoherent (spoiled) residual TM
3. Steady state free precession (SSFP)
18. Coherent (in-phase) residual TM
Uses variable flip angle excitation pulses followed by gradient
repahsing to produce a gradient echo.
These sequences keep this residual magnetization coherent by a
process known as rewinding.
Rewinding is achieved by reversing the slope of the phase encoding
gradient after readout. This results in the residual magnetization
rephasing, so that it is in phase at the beginning of the next
repetition.
Since residual TM is rephased by gradients and it is in phase, it is
called as coherent residual TM.
Examples—FISP, GRASS
20. The rewinder gradient rephases all transverse magnetization
regardless of when it was created.
Therefore the resultant echo contains information from the FID
and the stimulated echo.
These sequences can therefore be used to achieve T1 or T2 *
weighted images.
USES
This pulse sequence produce T2* weighted image.
As fluid is hyper intense ,they give an angiographic ,myelographic
or orthographic effect.
Can be acquired slice by slice or in a 3D volume Acquisition.
As the TR is short, a slice can be acquired in a single breath hold.
21. Advantages
Very fast scans, breath - holding possible
Very sensitive to flow (angiography)
can be acquired in a volume acquisition
Disadvantages
Reduced SNR in 2D acquisitions
Magnetic susceptibility increases
Loud gradient noise
23. INCOHERENT OR SPOILED GRADIENT ECHO
Uses variable flip angle excitation pulses followed by gradient
repahsing to produce a gradient echo.
The steady state is maintained so that residual transverse
magnetization is left over from previous repetitions. These sequences
dephase or spoil this magnetizati on so that its effect on image contrast
is minimal.
Spoiling can be done by RF pulses or gradients.
RF spoiling
Gradient spoiling
25. USES
This pulse sequences produce T1 or proton density
weighted images.
Image contrast is mainly influenced by the FID
that contributes T1 and proton density contrast.
26. Advantages
can be acquired in a volume or 2D
breath holding possible
good SNR and anatomical detail in volume
can be used after gadolinium contrast injection
Disadvantages
SNR poor in 2D
loud gradient noise
28. Steady state free precession (SSFP)
In gradient echo sequences the TE is not long enough to measure the T2 time of tissues as
a TE of at least 70ms is required for this. In addition, gradient rephasing is so inefficient
that any echo is dominated by T2 * effects and therefore true T2 weighting cannot be
achieved. The SSFP sequence overcomes this problem to obtain images that have a
sufficiently long TE and less T2* than in other steady state sequences.
However, in SSFP we need to digitize frequencies only from this stimulated echo and not
from the FID. To do this, the stimulated echo must be repositioned so that it does not
occur at the same time as the subsequent excitation pulse. This is achieved by applying a
rewinder gradient, which speeds up the rephasing so that the stimulated echo occurs
sooner
Example: PSIF
30. By this way, the stimulated echo can be received
and data from it is collected and which is used to
form the image.
In reverse echo gradient echo ,there are usually
two TEs;
1. THE ACTUAL TE- the time between the peak of
the gradient echo and the next RF excitation
pulse.
2. THE EFFECTIVE TE- the time from the peak of
the gradient echo to a previous RF excitation
pulse.
31. Advantages
can be acquired in a volume and in 2D
true T2 weighting achieved than in conventional GE
Disadvantages
susceptible to artefacts
image quality can be poor
loud gradient noise
34. BALANCED GRADIENT ECHO
This sequence is a modification of the coherent gradient echo sequence that uses a
balanced gradient system to correct for phase errors in flowing blood and CSF both
the FID and the spin echo are collected within a single readout.
This results in images where fat and water produce a higher signal, greater SNR and
fewer flow artefacts than coherent gradient echo in shorter scan times.
Balanced gradient -echo is a steady state sequences in which longitudinal during the
acquisition, thereby preventing saturation.
As the area of the gradient under the line equals that above the line, moving spins
accumulate a zero phase change as they pass along the gradients. As a result, spins in
blood and CSF are coherent and have a high signal intensity. This gradient formation is
the same as flow compensation or gradient moment rephasing .
35. BALANCED GRADIENT ECHO
In balanced gradient echo this gradient is applied in the slice and frequency axes.
In addition, higher flip angles and shorter TRs are used than in coherent gradient echo,
producing a higher SNR and shorter scan times
This is achieved by selecting a flip angle of 90° , for example, but in the first TR period only
applying half of this, i.e. 45° . In successive TRs the full flip angle is applied but with
alternating polarity so that the resultant transverse magnetization is created at a different
phase every TR
The resultant images display high SNR, good CNR between fat, water and surrounding
tissues, fewer flow voids and in very short scan times.
A balanced gradient scheme is used to correct for flow artifacts.
37. USES
Balanced gradient echo was developed initially for imaging the heart
and grerat vessels .
It is also sometimes used in joint and abdominal imaging.
38. Advantages Disadvantages
Fast shorter scan times
Reduced artifacts from
flow
Good SNR and anatomical
details in 3D
Images demonstrate good
contrast
Reduced SNR in 2D
acquisitions
Loud gradient noise
Susceptible to artifacts
Requires high
performance gradients
39. FAST GRADIENT ECHO
Fast gradient echo pulse sequences acquire a volume in a single
breath hold.
Ultrafast scanning techniques can be SE, GRE or combination of
both. Major ultrafast scanning techniques include
RARE,
STEAM (SE)
FLASH and
EPI (GRE).
40. FLASH (fast low angle shot)
FLASH is possible because of availability of high gradient
strengths of more than 25 mT/m. With this gradient strength TRs
of 2-5 ms and TEs of 1-3 ms are easily achieved so that k-space
can be filled very fast with one gradient echo per TR.
FLASH can be used in body imaging where physiological motion
needs to be overcome,
e.g. peristaltic motions.
cardiac perfusion.
41. ECHO PLANNER IMAGING
Echo planner imaging is an MR acquisition method that either fills all the
lines of k-space in a single repetition(single shot-SS) or in multiple
section( multishot-MS).
In EPI, gradient echoes are typically generated by oscillation of the
readout gradient.
There are many types of EPI;
1. GE-EPI uses a variable flip angle followed by EPI readout in K-space.
2. SE-EPI uses a 90-180 degree followed by EPI readout in K-space
K- SPACE:- K-space is an imaginary space which represents raw data
matrix.
42. USES
Diffusion weighted imaging
Functional imaging
Real time cardiac imaging
Perfusion imaging
Interventional techniques
Breath hold techniques
43. Advantages Disadvantages
Fast shorter scan times
Reduced artifacts from
respiratory and cardiac
motion
All three types of weighting
can be achived
Functional information
acquired
Scan time savings can be
used to improve phase
resolution.
Chemical shift artefact is
common
Peripheral nerve stimulation
due to fast switching of
gradients
Susceptible to artifacts
44. ADVANTAGES DISADVANTAGES
OF GRE
Much shorter scan times than
spin echo pulse sequences.
Minium TE is much shorter than
in spin echo pulse sequences.
TR can also be decreased
because other than 90 degree
are used.
No comparision for magnetic
field inhomogeneities.
Very susceptible to magnetic
field inhomogeneities
Contain magnetic susceptibility
artefact.