K SPACE
AND
PARALLEL IMAGING
ASHIK E H
DAMIT,SCTIMST
02/08/2018
OBJECTIVE
oProperties
oTrajectories of filling
oParallel imaging
oApplication
K STANDS FOR
• Wave number =k=1/Wavelength
• the number of waves or cycles per unit distance.
• kayser (K), where 1 K = 1 cm-1
• Greek word =krustallos kruos
K SPACE
•X Axis = Frequency encodings axis
•Y Axis = phase encodings axis
•Determine FOV
•Central line = seperate the + &- ve
half of k space
•Above and Below=determine + &- ve
half
•Stored-pfile
K SPACE
kx
ky
K SPACE
FT
PROPERTIES OF K SPACE
oCONTRAST
oRESOLUTION
oPOINT SYMMETRY
oFOV
CENTRE OF K-SPACE
oLow spatial frequency
oImage contrast, brightness, and general shapes
K-SPACE (ONLY CENTER FILLED ))
FT
PERIPHERY OF K-SPACE
oHigh spatial frequency
oedges, details, sharp transitions-resolution
FT
POINT SYMMETRY
Phase-conjugate symmetry lead to
Siemens - "Half Fourier
Philips and Hitachi -"Half scan".
GE-½-NEX" or "Fractional NEX"
FOV
o FOV refers the distance ( cm/mm) over which an MR
image is acquired or displayed.
o Δk = 1/FOV.
oΔk =spacing between samples in k space
oΔw = 1/kFOV. Δw is the pixel width
FT
K SPACE FILLING
oCartesian trajectories
oNon Cartesian trajectories.
oEcho planar.
Cartesian trajectories
 Uses a line by line rectilinear trajectory.
 One line is get filled for each TR.
 This trajectory can be done with
SE,GRE,IR,and other pulse sequences.
LINEAR REORDERING
CENTERIC REORDERING
REVERSE CENTERIC REORDERING
FAST K SPACE FILLING TECHNIQUES
 Partial Fourier imaging
 Radial imaging
 Spiral imaging
 Keyhole imaging
 EPI imaging
HALF K-SPACE” ACQUISITION
 Only part of k-space is measured.
 The unmeasured points are estimated by conjugate symmetry
through the center of k-space.
Siemens - "Half Fourier
Philips and Hitachi -"Half scan".
GE-½-NEX" or "Fractional NEX"
PARTIAL FOURIER IMAGING
PARTIAL FOURIER IMAGING
PARTIAL FOURIER IMAGING
The spatial information in either half of the k-space
is identical
If only half of the data along the Ky is acquired , we
can reduce the imaging time
It reduce the SNR by a factor of 2,but no loss in
spatial resolution
Routinely used for time-resolved perfusion & MRA
Commonly used in conjunction with HASTE &
Balanced SSFP
Radial trajectory
The first MR images reconstructed - radial trajectory
Samples k space on radial lines passing through the center.
These lines of k space is sampled more densely in the center
than in the periphery.
So the center of k space is filled almost immediately after
excitation.
The motion artifacts propagate radially rather than along
one axis.
Radial trajectory
PROPELLER/BLADE/MULTIVANE
 Use -
Uncooperative patients that cannot
remain motionless during exam
 Benefit -
Reduces artifact from in-plane
motion
Other names for similar sequence
BLADE (Sie)
Reduces artifact from in-plane motion
SPIRAL IMAGING
Two orthogonal gradients are applied simultaneously
& oscillated during read out
Producing spiral trajectory through the the k space
Data collection begins at the center, then curved
towards its edges
Center of the k-space is sampled more densely than
the periphery
Multiple interleaved spirals are needed to collect
enough data for an image of moderate resolution
SPIRAL IMAGING
 Very shot echo time.
 Employed in several applications like f MRI of brain, ASL
coronary MR angiography etc.
Advantages & applications
 Usually multiple interleaved spirals are required to collect
enough data for image of moderate resolution.
K-Space Representation of
Spiral Image Acquisition
EPI
•Very fast-capable to acquire an entire MR image only
in a fraction of second
•Proposed by Mansfield and pykett in 1977
•All the lines in k-space are filled by multiple gradient
reversals, producing multiple echoes after a single RF
excitation
•sequence highly vulnerable to susceptibility artifacts
Types of EPI
(a) single shot
(b) multi shot
Single short EPI
• All the spatial encoding data of an image can be obtained
after a single RF excitation
• Frequency-encoding gradient oscillates rapidly from a + to - a
negative amplitude, forming a train of gradient echoes .
•Each echo is phase encoded differently by phase-encoding
blips on the phase
•Reduced imaging time, decreased motion artifact, and the
ability to image rapid physiologic processes of the human
body
Filling K-space for EPI
1. Because of the alternating
positive/negative readout
gradients,
k-space is filled in a “Zig-zag”
pattern.
2. Because this is a long train of
gradient echoes, we get
geometric distortions in the
image. This happens
whenever we take too long to
fill k-space for EPI. To
reduce this problem, we often
use “half k-space”
acquisition.
We measure these
We estimate these
(w/ the complex conjugates)
Multishot EPI
•Also called segmental EPI
•Readout is divided into multiple shots or segments
•Only a portion of k-space data is acquired with each
shot
•Shots are repeated until a full set of data is collected
Clinical applications of EPI
Diffusion imaging of the brain
Dynamic perfusion studies
Cardiac and abdominal imaging free of motion
artifact
Imaging of coronary arteries free of cardiac motion
artifact
Cine cardiac imaging with a single heart beat
KEYHOLE -IMAGING
Arrival of contrast
injection
Fast dynamic
scanning
Reconstruction
ky
kz
ky
kz
ky
kz
ky
kz1 2 3
ky
kz
mask
Centric Keyhole (4D TRAK - Philips)
• 2 regions only
• Centric version of keyhole
• Oval central & peripheries
Semi randomized Centric Keyhole (TWIST – Siemens)
• 2 regions only
• Central k-space sampled alternately with
incomplete peripheral k-space
• Peripheral k-space is acquired over many cycles
Multi-Centric Keyhole (TRICKS – GE)
4 regions (3 central + 1 peripheral)
• A-BA-CA-DRepeat
• Central > intermediate > peripheral k-spaces sampled with
decremental frequency
A
D
B
c
LIMITATION OF KEYHOLE IMAGING
• If any motion between the baseline higher resolution
images& dynamic contrast enhanced images, it will
produce misregistration between the contrast & fine
detail of the image
• Dramatic changes in signal intensities of acquired and
replaced k space data will create ghost artifacts
PARALLEL
IMAGING
Basic Requirements For Pl
High field MRI magnet
Phased array coil
High fast computer reconstruction system
NON PARALLEL/REGULAR IMAGING
Single/multiple surface coils may be used to detect
the MR signal
but their individual outputs are combined into one
aggregate complex signal that is digitized and
processed into the final image


PARALLEL IMAGING –FULL FOV
 Consider image with the field of view (FOV) as shown
and two
coils with the following sensitivities


Full FOV


Half FOV
Acceleration factor (R)
Which reduce the aquisition time
If we have M coil elements covering the FOV, we can skip up to [M-1]
lines for each line in k-space .
This can be fractional as well:
no of phase-encodes to cover k-space
R = –––––––––––––––––––––––––
no of phase-encodes used in acquisition
iPAT factor (Siemens)
ASSET factor (GE)
SENSE factor (Philips)
K-space
#Profiles
Time
R=2
cut to half both
• scan duration
• FOV
FT
FT
+
unwrap
• Under sampled k space resulting in aliasing artifact
• More the lines skipped more will be the aliasing
Specific reconstruction algorithm are used to locate
the aliased part and under go PI Imaging
reconstruction
• So, fundamental problem is PI is how to "unfold“
/unwrap the wrap around
SIGNAL TO NOISE RATIO
parallel
PHASED ARRAY COIL ELEMENTS
 It has multiple no :of coil element
 Uneven distribution of SNR throughout
volume i . e
- Very high SNR at the edge
- Lower in the middle
+ + + =
Sensitivity map
sensitivity map = The spatial sensitivity of each coil element
 A calibration scan is usually done to get sensitivity map
S2S1
Uses of phased array coil
In regular /Non PI imaging only problem related to spatial
sensitivity
Conventional use of
phased array [unaliased]
Uses of phased array coil [PI]
•Spatial sensitivity varies for each coil element
•Along conjunction with undersampling
FT
+
unwrap
Two Families of PAT
Image-based PAT k-space-based PAT
SENSE [Philips] SMASH
ASSET [GE] ARC [GE]
mSENSE[Siemens] GRAPPA [Siemens]
RAPID [Hitachi] iPAT [Siemens]
SPEEDER [Toshiba] SPEEDER [Toshiba]
K-SPACE BASED pMRI
Reconstruct images from each
elements, then untangle
IMAGE BASED pMRI
Untangle data to create fully
filled k space, then
reconstruct image
SENSE/ASSET methods reconstruct, then correct.
GRAPPA/ARC methods correct, then reconstruct.
SENSE
Introduced by Pruessmann et al
 first commercially available PI method (by PHILIPS)
 1 -sensitivity maps for each coil element short,low
resolution calibration scan.
 2-A reduced k space is MR data formed by fewer phase
encoding gradient steps in conjunction with phased
array coil
 3. Reconstruct partial FOV images from each coil
 [Sub sampled k space shows aliasing]
 4. Unfold/Combine partial FOV image by matrix
inversion
FFT
FFT
SENSE
RECON
FULL FOV
IMAGE
r FOV
IMAGES
COILCALIBRATION
COIL ELEMENT 1
COIL ELEMENT 2
UNDER SAMPLING IN
K-SPACE
SENSE (IMAGE BASED)
SPATIAL SENSITIVITY MEASURED BY
A]Acquire quick images from each element of coil
B]Reconstruct the full image using all elements
C]Image (a) divided by (b) gives a noisy
sensitivity map[C]
=A
B
[C]
D]Filtering smoothes out the noise, yielding
our sensitivity map [D] [D]
SENSE Reduces SNR
SENSE factor =1.0 SENSE factor =3.0
As SENSE factor increases, noisiness of the image
increases
SENSE
Artifact
Acquire , Reconstruct and Unfold
• In pre scan calibration- calculate the point by point sensitivity
of each elements of coil
• sensitivities of Coils 1 and 2 @A =S1A and S2A
• sensitivities of Coils 1 and 2 @B =S1B and S2B
• Pixel values from Coils 1 and 2 = P1 and P2
• P1 = A•S1A + B•S1B
P2 = A•S2A + B•S2B
• 2]MATRIX INVERSION TECHNIQUE-similalar to it
• 2]MATRIX INVERSION TECHNIQUE-similalar to it
• I = (SHψ-1S + λ-1)-1SHψ-1P
• S = coarse sensitivity profiles from the individual coils,
normalized for uniform signal by the body coil
• ψ = noise covariance matrix, representing noise increase
due to patient-specific interactions between coil elements.
• P = partial FOV images with aliasing
• I = final image at full FOV
Applications of SENSE
1. Cardiovascular-breath hold, real-time
2. CEMRA
3. Brain-f MRI
mSENSE
•Modified SENSE is a version of SENSE
•Which does not require a separate calibration scan
•Additional lines are acquired at the centre of k space
during diagnostic scan
•Central lines are extracted and reconstruct low
resolution,unaliased images
•This images used to provide sensitivity map
•Reduction factor R cannot achieve-additional k space
lines are required for calibrations.
SMASH
•Described by sodickson and manning
•Requires a prior estimation of the individual coil
sensitivities of the receiver array
•A linear combination of these estimated coil
sensitivities can directly generate missing phase
encoding steps,which would normally be performed by
using phase encoding magnetic field gradient
•The missing k space lines are restored prior to the FT
Requires a prior estimation of the individual coil
sensitivities of the receiver array
 the sensitivity values Ck(x,y) are combined with
appropriate linear weights nk (m) to generate composite
sensitivity profiles with sinusoidal spatial sensitivity
variations of the order m
 sensitivity maps are fit to the desired harmonic
modulations.[m=-0,1]
Unser sampled data with desired harmonic generate full
FOV K Space
Fourier transformed to yield the full-FOV reconstructed
image
SMASH reconstruction process.
GRAPPA
• Variation of smash
• Less dependant on coil geometry than SMASH
• Better SNR than SMASH
FOUR MAJOR STEPS IN GRAPPA [ARC]
• Data Acquisition
• Estimation of Missing Lines
• Generate Individual Coil Images
• Combine images of each coil
1- DATA ACQUISITION
• Acquired MR signals are digitized,
demodulated and used to fill k space
• center of k-space with ACS caliberation
• [ACS are used to calculate weighting factors for
each coil ]
• others k space are under sampled
2 Estimation of Missing Lines
• weighting factors for each coil-
reflect how each coil distorts,
smears, and displaces spatial
frequencies within the full FOV k-
space data
• combined with local known data
for each small region (known a
block or kernel
• 3-Generate Individual Coil Images
• 4-Combine images of each coil
ARC-
• ARC uses a full 3D kernel to synthesize missing target data
from neighbouring source data from all three imaging
directions.
• Thus, improved reconstruction accuracy with fewer required
calibration lines
• The end result is highly accelerated MR data acquisition with
improved image quality and fewer artifacts.
• .
CAIPIRINHA
• Controlled aliasing in parallel imaging results in higher
acceleration
• PI offered by Siemens with R≥4
• used primarily for 3D breath-hold abdominal imaging
• Aliasing controlled by providing individual slice with
different phase cycle by means of altering multiband RF
pulse.
phase offset improves the accuracy of reconstruction while reducing noise and aliasing.
Aliasing artifacts are still present but they are shifted to the corners of image space and
are less likely to become concentrated in a few locations.
Choosing Of PI Method [SENSE/GRAPPA]
 FACTORS
 Total Imaging Time/Speed. GRAPPA/ARC is a somewhat
longer sequence than SENSE/ASSET becoz :extra time for the
self-calibration of k-space lines
• Signal-to-noise SENSE/ASSET provides slightly higher SNR
• [ But same SNR @ R=0 ]
• Body Region accurate coil sensitivity maps may difficult to
obtain for SENSE/ASSET ;SO, GRAPPA/ARC is preffered
Choosing Of PI Method [SENSE/GRAPPA]
 FACTORS
 Motion: SENSE/ASSET may do poorly preferred due to
motion /recontruction artifact arise b/w calibration and
acquisition scans
 difficulty suspending respiration to exactly the same degree
between SENSE/ASSET calibration and imaging
 Field-of-view (FOV) GRAPPA/ARC is more tolerant toward
small FOVs
 SENSE/ASSET produce aliasing/wrap around if
full FOV < OBJECT Size
 Use in Single-Shot EPI : GRAPPA/ARC prefered becoz of less
distortion during reconstruction
Clinical Applications
• wide applicability: PI can be applied to nearly any pulse
sequence
• gains in acquisition speed can be used in a no: of
different ways
MR Angiography:
• Because of high acquisition speed, SNR, minimal
artifacts, and relatively high spatial resolution-it become
standard technique
Patient cooperative to hold breath [reduced aqu - time]
thus also reduce motion artifact
Clinical Applications
Not only reduce motion artifact ,but also also
diminish venous contamination, particularly in regions
where there is rapid venous return, such as the renal
and carotid arteries
Clinical Applications
• Time-resolved pulmonary MR angiography
Performed with a high temporal resolution
• The gains in speed by PI can alternatively be applied to
improve spatial resolution [so additional space encoding
steps is acquired]
• Helps to reduce artifacts
• PI optimizing the length of the arterial phase
• There by, characterize lesions that enhance in the
arterial phase and rapidly wash out, such as small
hepatocellular carcinomas, certain metastatic lesions
microadenoma
Cardiac Imaging
• reduce the total number and/or length of breath holds
required in an examination and thereby significantly
reduce the total examination time
Cardiac Imaging
• PI Technique helps to improve spatial and temporal
resolution of Coronal cardiac images
Advantage of parallel imaging…
 To reduce scan time
 To speed up single shot MRI method
 To reduce TE on long echo train methods
 To mitigate susceptibility , chemical shift and
other artifacts
 To decrease RF heating [SAR] by minimizing
number of RF pulse
Thank you

K space and parallel imaging

  • 1.
    K SPACE AND PARALLEL IMAGING ASHIKE H DAMIT,SCTIMST 02/08/2018
  • 2.
  • 3.
    K STANDS FOR •Wave number =k=1/Wavelength • the number of waves or cycles per unit distance. • kayser (K), where 1 K = 1 cm-1 • Greek word =krustallos kruos
  • 4.
    K SPACE •X Axis= Frequency encodings axis •Y Axis = phase encodings axis •Determine FOV •Central line = seperate the + &- ve half of k space •Above and Below=determine + &- ve half •Stored-pfile
  • 5.
  • 6.
  • 7.
    PROPERTIES OF KSPACE oCONTRAST oRESOLUTION oPOINT SYMMETRY oFOV
  • 8.
    CENTRE OF K-SPACE oLowspatial frequency oImage contrast, brightness, and general shapes K-SPACE (ONLY CENTER FILLED )) FT
  • 9.
    PERIPHERY OF K-SPACE oHighspatial frequency oedges, details, sharp transitions-resolution FT
  • 11.
    POINT SYMMETRY Phase-conjugate symmetrylead to Siemens - "Half Fourier Philips and Hitachi -"Half scan". GE-½-NEX" or "Fractional NEX"
  • 12.
    FOV o FOV refersthe distance ( cm/mm) over which an MR image is acquired or displayed. o Δk = 1/FOV. oΔk =spacing between samples in k space oΔw = 1/kFOV. Δw is the pixel width FT
  • 13.
    K SPACE FILLING oCartesiantrajectories oNon Cartesian trajectories. oEcho planar.
  • 14.
    Cartesian trajectories  Usesa line by line rectilinear trajectory.  One line is get filled for each TR.  This trajectory can be done with SE,GRE,IR,and other pulse sequences.
  • 15.
  • 16.
  • 17.
  • 18.
    FAST K SPACEFILLING TECHNIQUES  Partial Fourier imaging  Radial imaging  Spiral imaging  Keyhole imaging  EPI imaging
  • 19.
    HALF K-SPACE” ACQUISITION Only part of k-space is measured.  The unmeasured points are estimated by conjugate symmetry through the center of k-space.
  • 20.
    Siemens - "HalfFourier Philips and Hitachi -"Half scan". GE-½-NEX" or "Fractional NEX"
  • 21.
    PARTIAL FOURIER IMAGING PARTIALFOURIER IMAGING PARTIAL FOURIER IMAGING The spatial information in either half of the k-space is identical If only half of the data along the Ky is acquired , we can reduce the imaging time It reduce the SNR by a factor of 2,but no loss in spatial resolution Routinely used for time-resolved perfusion & MRA Commonly used in conjunction with HASTE & Balanced SSFP
  • 22.
    Radial trajectory The firstMR images reconstructed - radial trajectory Samples k space on radial lines passing through the center. These lines of k space is sampled more densely in the center than in the periphery. So the center of k space is filled almost immediately after excitation. The motion artifacts propagate radially rather than along one axis.
  • 23.
  • 24.
    PROPELLER/BLADE/MULTIVANE  Use - Uncooperativepatients that cannot remain motionless during exam  Benefit - Reduces artifact from in-plane motion Other names for similar sequence BLADE (Sie) Reduces artifact from in-plane motion
  • 25.
    SPIRAL IMAGING Two orthogonalgradients are applied simultaneously & oscillated during read out Producing spiral trajectory through the the k space Data collection begins at the center, then curved towards its edges Center of the k-space is sampled more densely than the periphery Multiple interleaved spirals are needed to collect enough data for an image of moderate resolution
  • 26.
  • 27.
     Very shotecho time.  Employed in several applications like f MRI of brain, ASL coronary MR angiography etc. Advantages & applications  Usually multiple interleaved spirals are required to collect enough data for image of moderate resolution. K-Space Representation of Spiral Image Acquisition
  • 28.
    EPI •Very fast-capable toacquire an entire MR image only in a fraction of second •Proposed by Mansfield and pykett in 1977 •All the lines in k-space are filled by multiple gradient reversals, producing multiple echoes after a single RF excitation •sequence highly vulnerable to susceptibility artifacts Types of EPI (a) single shot (b) multi shot
  • 29.
    Single short EPI •All the spatial encoding data of an image can be obtained after a single RF excitation • Frequency-encoding gradient oscillates rapidly from a + to - a negative amplitude, forming a train of gradient echoes . •Each echo is phase encoded differently by phase-encoding blips on the phase •Reduced imaging time, decreased motion artifact, and the ability to image rapid physiologic processes of the human body
  • 30.
    Filling K-space forEPI 1. Because of the alternating positive/negative readout gradients, k-space is filled in a “Zig-zag” pattern. 2. Because this is a long train of gradient echoes, we get geometric distortions in the image. This happens whenever we take too long to fill k-space for EPI. To reduce this problem, we often use “half k-space” acquisition. We measure these We estimate these (w/ the complex conjugates)
  • 31.
    Multishot EPI •Also calledsegmental EPI •Readout is divided into multiple shots or segments •Only a portion of k-space data is acquired with each shot •Shots are repeated until a full set of data is collected
  • 32.
    Clinical applications ofEPI Diffusion imaging of the brain Dynamic perfusion studies Cardiac and abdominal imaging free of motion artifact Imaging of coronary arteries free of cardiac motion artifact Cine cardiac imaging with a single heart beat
  • 33.
    KEYHOLE -IMAGING Arrival ofcontrast injection Fast dynamic scanning Reconstruction ky kz ky kz ky kz ky kz1 2 3 ky kz mask
  • 34.
    Centric Keyhole (4DTRAK - Philips) • 2 regions only • Centric version of keyhole • Oval central & peripheries
  • 35.
    Semi randomized CentricKeyhole (TWIST – Siemens) • 2 regions only • Central k-space sampled alternately with incomplete peripheral k-space • Peripheral k-space is acquired over many cycles
  • 36.
    Multi-Centric Keyhole (TRICKS– GE) 4 regions (3 central + 1 peripheral) • A-BA-CA-DRepeat • Central > intermediate > peripheral k-spaces sampled with decremental frequency A D B c
  • 37.
    LIMITATION OF KEYHOLEIMAGING • If any motion between the baseline higher resolution images& dynamic contrast enhanced images, it will produce misregistration between the contrast & fine detail of the image • Dramatic changes in signal intensities of acquired and replaced k space data will create ghost artifacts
  • 38.
  • 39.
    Basic Requirements ForPl High field MRI magnet Phased array coil High fast computer reconstruction system
  • 40.
    NON PARALLEL/REGULAR IMAGING Single/multiplesurface coils may be used to detect the MR signal but their individual outputs are combined into one aggregate complex signal that is digitized and processed into the final image  
  • 41.
    PARALLEL IMAGING –FULLFOV  Consider image with the field of view (FOV) as shown and two coils with the following sensitivities   Full FOV   Half FOV
  • 42.
    Acceleration factor (R) Whichreduce the aquisition time If we have M coil elements covering the FOV, we can skip up to [M-1] lines for each line in k-space . This can be fractional as well: no of phase-encodes to cover k-space R = ––––––––––––––––––––––––– no of phase-encodes used in acquisition iPAT factor (Siemens) ASSET factor (GE) SENSE factor (Philips)
  • 43.
    K-space #Profiles Time R=2 cut to halfboth • scan duration • FOV
  • 44.
  • 45.
    • Under sampledk space resulting in aliasing artifact • More the lines skipped more will be the aliasing Specific reconstruction algorithm are used to locate the aliased part and under go PI Imaging reconstruction • So, fundamental problem is PI is how to "unfold“ /unwrap the wrap around
  • 46.
    SIGNAL TO NOISERATIO parallel
  • 47.
    PHASED ARRAY COILELEMENTS  It has multiple no :of coil element  Uneven distribution of SNR throughout volume i . e - Very high SNR at the edge - Lower in the middle + + + =
  • 48.
    Sensitivity map sensitivity map= The spatial sensitivity of each coil element  A calibration scan is usually done to get sensitivity map S2S1
  • 49.
    Uses of phasedarray coil In regular /Non PI imaging only problem related to spatial sensitivity Conventional use of phased array [unaliased]
  • 50.
    Uses of phasedarray coil [PI] •Spatial sensitivity varies for each coil element •Along conjunction with undersampling
  • 51.
  • 52.
    Two Families ofPAT Image-based PAT k-space-based PAT SENSE [Philips] SMASH ASSET [GE] ARC [GE] mSENSE[Siemens] GRAPPA [Siemens] RAPID [Hitachi] iPAT [Siemens] SPEEDER [Toshiba] SPEEDER [Toshiba]
  • 53.
    K-SPACE BASED pMRI Reconstructimages from each elements, then untangle IMAGE BASED pMRI Untangle data to create fully filled k space, then reconstruct image SENSE/ASSET methods reconstruct, then correct. GRAPPA/ARC methods correct, then reconstruct.
  • 54.
    SENSE Introduced by Pruessmannet al  first commercially available PI method (by PHILIPS)  1 -sensitivity maps for each coil element short,low resolution calibration scan.  2-A reduced k space is MR data formed by fewer phase encoding gradient steps in conjunction with phased array coil  3. Reconstruct partial FOV images from each coil  [Sub sampled k space shows aliasing]  4. Unfold/Combine partial FOV image by matrix inversion
  • 55.
    FFT FFT SENSE RECON FULL FOV IMAGE r FOV IMAGES COILCALIBRATION COILELEMENT 1 COIL ELEMENT 2 UNDER SAMPLING IN K-SPACE SENSE (IMAGE BASED)
  • 56.
    SPATIAL SENSITIVITY MEASUREDBY A]Acquire quick images from each element of coil B]Reconstruct the full image using all elements C]Image (a) divided by (b) gives a noisy sensitivity map[C] =A B [C] D]Filtering smoothes out the noise, yielding our sensitivity map [D] [D]
  • 57.
    SENSE Reduces SNR SENSEfactor =1.0 SENSE factor =3.0 As SENSE factor increases, noisiness of the image increases SENSE Artifact
  • 59.
  • 60.
    • In prescan calibration- calculate the point by point sensitivity of each elements of coil • sensitivities of Coils 1 and 2 @A =S1A and S2A • sensitivities of Coils 1 and 2 @B =S1B and S2B • Pixel values from Coils 1 and 2 = P1 and P2 • P1 = A•S1A + B•S1B P2 = A•S2A + B•S2B • 2]MATRIX INVERSION TECHNIQUE-similalar to it
  • 61.
    • 2]MATRIX INVERSIONTECHNIQUE-similalar to it • I = (SHψ-1S + λ-1)-1SHψ-1P • S = coarse sensitivity profiles from the individual coils, normalized for uniform signal by the body coil • ψ = noise covariance matrix, representing noise increase due to patient-specific interactions between coil elements. • P = partial FOV images with aliasing • I = final image at full FOV
  • 62.
    Applications of SENSE 1.Cardiovascular-breath hold, real-time 2. CEMRA 3. Brain-f MRI
  • 63.
    mSENSE •Modified SENSE isa version of SENSE •Which does not require a separate calibration scan •Additional lines are acquired at the centre of k space during diagnostic scan •Central lines are extracted and reconstruct low resolution,unaliased images •This images used to provide sensitivity map •Reduction factor R cannot achieve-additional k space lines are required for calibrations.
  • 64.
    SMASH •Described by sodicksonand manning •Requires a prior estimation of the individual coil sensitivities of the receiver array •A linear combination of these estimated coil sensitivities can directly generate missing phase encoding steps,which would normally be performed by using phase encoding magnetic field gradient •The missing k space lines are restored prior to the FT
  • 65.
    Requires a priorestimation of the individual coil sensitivities of the receiver array  the sensitivity values Ck(x,y) are combined with appropriate linear weights nk (m) to generate composite sensitivity profiles with sinusoidal spatial sensitivity variations of the order m  sensitivity maps are fit to the desired harmonic modulations.[m=-0,1] Unser sampled data with desired harmonic generate full FOV K Space Fourier transformed to yield the full-FOV reconstructed image SMASH reconstruction process.
  • 68.
    GRAPPA • Variation ofsmash • Less dependant on coil geometry than SMASH • Better SNR than SMASH
  • 69.
    FOUR MAJOR STEPSIN GRAPPA [ARC] • Data Acquisition • Estimation of Missing Lines • Generate Individual Coil Images • Combine images of each coil
  • 70.
    1- DATA ACQUISITION •Acquired MR signals are digitized, demodulated and used to fill k space • center of k-space with ACS caliberation • [ACS are used to calculate weighting factors for each coil ] • others k space are under sampled
  • 71.
    2 Estimation ofMissing Lines • weighting factors for each coil- reflect how each coil distorts, smears, and displaces spatial frequencies within the full FOV k- space data • combined with local known data for each small region (known a block or kernel • 3-Generate Individual Coil Images • 4-Combine images of each coil
  • 73.
    ARC- • ARC usesa full 3D kernel to synthesize missing target data from neighbouring source data from all three imaging directions. • Thus, improved reconstruction accuracy with fewer required calibration lines • The end result is highly accelerated MR data acquisition with improved image quality and fewer artifacts. • .
  • 75.
    CAIPIRINHA • Controlled aliasingin parallel imaging results in higher acceleration • PI offered by Siemens with R≥4 • used primarily for 3D breath-hold abdominal imaging • Aliasing controlled by providing individual slice with different phase cycle by means of altering multiband RF pulse.
  • 76.
    phase offset improvesthe accuracy of reconstruction while reducing noise and aliasing. Aliasing artifacts are still present but they are shifted to the corners of image space and are less likely to become concentrated in a few locations.
  • 77.
    Choosing Of PIMethod [SENSE/GRAPPA]  FACTORS  Total Imaging Time/Speed. GRAPPA/ARC is a somewhat longer sequence than SENSE/ASSET becoz :extra time for the self-calibration of k-space lines • Signal-to-noise SENSE/ASSET provides slightly higher SNR • [ But same SNR @ R=0 ] • Body Region accurate coil sensitivity maps may difficult to obtain for SENSE/ASSET ;SO, GRAPPA/ARC is preffered
  • 78.
    Choosing Of PIMethod [SENSE/GRAPPA]  FACTORS  Motion: SENSE/ASSET may do poorly preferred due to motion /recontruction artifact arise b/w calibration and acquisition scans  difficulty suspending respiration to exactly the same degree between SENSE/ASSET calibration and imaging  Field-of-view (FOV) GRAPPA/ARC is more tolerant toward small FOVs  SENSE/ASSET produce aliasing/wrap around if full FOV < OBJECT Size  Use in Single-Shot EPI : GRAPPA/ARC prefered becoz of less distortion during reconstruction
  • 79.
    Clinical Applications • wideapplicability: PI can be applied to nearly any pulse sequence • gains in acquisition speed can be used in a no: of different ways MR Angiography: • Because of high acquisition speed, SNR, minimal artifacts, and relatively high spatial resolution-it become standard technique Patient cooperative to hold breath [reduced aqu - time] thus also reduce motion artifact
  • 80.
    Clinical Applications Not onlyreduce motion artifact ,but also also diminish venous contamination, particularly in regions where there is rapid venous return, such as the renal and carotid arteries
  • 81.
    Clinical Applications • Time-resolvedpulmonary MR angiography Performed with a high temporal resolution
  • 82.
    • The gainsin speed by PI can alternatively be applied to improve spatial resolution [so additional space encoding steps is acquired]
  • 83.
    • Helps toreduce artifacts
  • 84.
    • PI optimizingthe length of the arterial phase • There by, characterize lesions that enhance in the arterial phase and rapidly wash out, such as small hepatocellular carcinomas, certain metastatic lesions microadenoma
  • 85.
    Cardiac Imaging • reducethe total number and/or length of breath holds required in an examination and thereby significantly reduce the total examination time
  • 86.
    Cardiac Imaging • PITechnique helps to improve spatial and temporal resolution of Coronal cardiac images
  • 87.
    Advantage of parallelimaging…  To reduce scan time  To speed up single shot MRI method  To reduce TE on long echo train methods  To mitigate susceptibility , chemical shift and other artifacts  To decrease RF heating [SAR] by minimizing number of RF pulse
  • 88.

Editor's Notes

  • #6 In 2DFT imaging, each row in k-space corresponds to the echo data obtained from a single application of the phase-encoding gradient.  By convention, rows near the center of the k-space grid are defined to correspond to low-order phase-encode steps, whereas those rows near the top and bottom correspond to higher-order phase-encodings.  Since echo amplitudes are larger at the low-order phase-encode steps (there is less gradient-induced dephasing), the values of k-space will be greater near the center of the grid. 
  • #7 In 2DFT imaging, each row in k-space corresponds to the echo data obtained from a single application of the phase-encoding gradient.  By convention, rows near the center of the k-space grid are defined to correspond to low-order phase-encode steps, whereas those rows near the top and bottom correspond to higher-order phase-encodings.  Since echo amplitudes are larger at the low-order phase-encode steps (there is less gradient-induced dephasing), the values of k-space will be greater near the center of the grid. 
  • #15 In 2DFT imaging, each row in k-space corresponds to the echo data obtained from a single application of the phase-encoding gradient.  By convention, rows near the center of the k-space grid are defined to correspond to low-order phase-encode steps, whereas those rows near the top and bottom correspond to higher-order phase-encodings.  Since echo amplitudes are larger at the low-order phase-encode steps (there is less gradient-induced dephasing), the values of k-space will be greater near the center of the grid. 
  • #16 After aquring the mr signal we will put into the k space matrix… one std way of filling is such a way that one hrizondal line after the other…from bottam of k space to top… this method is called liner reodering
  • #17 In centric reordeing method first we will fill the cnter part of the k space… then we fill towards top and bottam
  • #18 In revers e reordering first we fill the peripheral part of the k space…. And then toward the central part
  • #19 For decresing the scan time we using fst k space filling methods…. That includes
  • #20 This is spatial type of k space filling method.. In this only a part of the k space is measured… and the unmeasured ponints are estimates by callclating the complex… and the conjugate of the measured points are symmetric through the center of k space
  • #22 Half k space filling is mainly used for ticks angiogram and for some localizer seqence ie;HASTE,SSFSE,balanced SSFP ect… in this method only half of the data aqired along phace encoding direction.tn we can reduse the time… wn we use this method SNR redsed by facre 2.. Bt will not lost in spatal resoltion Half-Fourier Acquisition Single-shot Turbo spin Echo imaging=siemens Ssfse=ge Single-shot fast spin echo,  Both are ep fsecho
  • #23 In this method k space lines are sampled more densily in center rather than peripheral part.. So center of k space is filled immidediatly aftr exitation. By using radial trajectory we can reduse the motion artifacts.. First mr image reconstructed by using radial trajectory  Today, radial and spiral readouts are becoming the norm because they offer lower intrinsic sensitivity to motion and permit shorter TE values. The spirals may be "spiral in", "spiral out", or combined. Since there are not separate frequency-and phase-encode directions in radial and spiral techniques, the artifacts are different (they include curvilinear bands and ring-shaped blurring). Spiral scans are also more affected by incorrect gradient timing and concomitant field gradients. Image reconstruction is also more complicated, and may require "gridding" or other methods to warp the measured k-space points into a rectangular matrix for fast Fourier transform numerical processing.
  • #24 In this method k space lines are sampled more densily in center rather than peripheral part.. So center of k space is filled immidediatly aftr exitation. By using radial trajectory we can reduse the motion artifacts.. First mr image reconstructed by using radial trajectory
  • #25 Some of the example for radial imaging is blade,proppaller,multivane
  • #26 In this method the use of orthogonal gradient will prdsed spiral trajectory through k space… here also the center of k space is filled more densly then peripheral part… and the main application of spiral imaging for imaging coronary arteries
  • #27 This is the schematic represntaion of the spiral imaging techinque
  • #29 Choosing among the various k-space traversal strategies depends on the application. The choice is perhaps most critical for echo-planar imaging (EPI), where the long readout train makes the sequence highly vulnerable to susceptibility artifacts and rapid switching produces unwanted eddy current effects. The usual analytic approach for understanding these effects decomposes them along the three standard imaging axes (phase-encoding, frequency-encoding, and slice-select).
  • #33 Shorter excitation RF pulse (200 μs) accelerated with ASSET
  • #34  An initial pre contrast mask is obtained at full resolution After contrast injection,centre of K-space is sampled ,while periphery is sampled occasionally. Final images are then reconstructed using the peripheral k space data from reference data set and central K space from keyhole sets.
  • #35 Vendor-specific nomenclature: GE   “TRICKS“ (TIME RESOLVED IMAGES IN CONTRAST KINETICS) Siemens “TWIST" (TIME RESOLVED ANGIOGRAPHY USING STOCHASTIC STRATEGY Philips  4D TRAK (“4D TIME RESOLVED ANGIOGRAPHY USING KEYHOLE"), Hitachi  TRAQ  (“TIME RESOLVED ACQUISITION), & Toshiba  FREEZE FRAME 
  • #39 In conventional mri aqusition time is a challenging for patient who were unable to lied for a long time…. So to over come this lots of fst imaging sequences are come…for eg FSE sequences which is a varient form of SE. then some gradient sqenses introdused.. Still these is hving its on limitation due to techneical and physiological prblms asssociated with rf as well as time warriying magnetic field..to over come this limitation new techiqe is come which is called pmri
  • #40 In pmri the major requirment is dedicated reciver coil which contain more then one elemnt, called multichannel coil.. In parellel imaging we r using the spatial information frm all the elements of rf coil and we will sample the data in parellel form
  • #41 In "regular" (non-parallel) imaging, multiple surface coils may be used to detect the MR signal, but their individual outputs are combined into one aggregate complex signal that is digitized and processed into the final imageIn pmri the major requirment is dedicated reciver coil which contain more then one elemnt, called multichannel coil.. In parellel imaging we r using the spatial information frm all the elements of rf coil and we will sample the data in parellel form
  • #42  In parallel imaging, conversely, the signals from individual coils are amplified, digitized, and processed simultaneously "in parallel" along separate channels, retaining their identities until near the end.In pmri the major requirment is dedicated reciver coil which contain more then one elemnt, called multichannel coil.. In parellel imaging we r using the spatial information frm all the elements of rf coil and we will sample the data in parellel form
  • #46 Undersampling is the decrease in data to increase image acquisition speed (shorter scan times without loss of quality - increased productivity - reduced cost of equipment).
  • #48 As I told bfor the major reqirment for pmri is phased arrey coil..the major disadvantage of phased arry coil is it produse uneven snr through ot the volume ie;very high snr at the edges and lower in the middile
  • #49 So for reconstruction process for removing wrappa rond will need a coil sensitivity map.. This tell the spatial sensitivity of each coil…
  • #50 Wht is the role of phased array coil in pmri…? Since the coil having multiple element… and each coil collect the signal frm perticular portion. So the spatial sensitivity d/f for each element.by comabaining this property image with aliased data image we will get a aliased image.. Later we will unwrapp the image with recunstrution method
  • #51 Wht is the role of phased array coil in pmri…? Since the coil having multiple element… and each coil collect the signal frm perticular portion. So the spatial sensitivity d/f for each element.by comabaining this property image with aliased data image we will get a aliased image.. Later we will unwrapp the image with recunstrution method
  • #53 SENSE (SENSitivity Encoding) and ASSET (Array coil Spatial Sensitivity Encoding) are among the most widely used parallel imaging methods. These techniques are primarily performed in image space after reconstruction of data from the individual coils. (This contrasts with GRAPPA/ARC methods which operate primarily on k-space data before image reconstruction).  Each major MR vendor offers some version of the SENSE technique under different trade names: Siemens (mSENSE), GE (ASSET), Philips (SENSE), Hitachi (RAPID - "Rapid Acquisition through Parallel Imaging Design"), and Canon (SPEEDER).
  • #56 Let as see the working of sense..assume tht this head coil hav 2 coil elements.. Intialy we will take a calibration scan which provide a sensitivity profile.. Then image aqusition made with undersampled data.. Then we will ft the k space..resultent image in redsed fov image with wrapp arround artifact.. Then we will nfold the data by compaining the coil calibrtaion images and wrapp artifact image.. Finaly we will get a original image with ot artifacts
  • #57 In short, with this method, one must acquire a map before running a parallel imaging sequence. Takes a minute or so. If one uses the summed image from all elements as a reference, this technique is called “Auto-SENSE”. Calculating coil sensitivities are the initial and most important step in the SENSE process. Low-resolution images are acquired separately from each surface coil at full field-of-view. These surface coil images are normalized by dividing them by a low-resolution body coil image. Filtering, thresholding, and point estimation are then applied to the data to generate coil sensitivity maps (shown right). These maps quantify the relative weighting of signals from different points of origin within the reception area of each coil. The calculation of coils sensitivities may be obtained as a separate ~20 second acquisition before actual imaging begins (GE uses this method for ASSET). Alternatively, automated coil calibration may be integrated into the pulse sequence itself (Siemens' mSENSE). The latter method has the advantage of being less sensitive to motion that occurs between the time of calibration and the beginning of the scan proper. Once coil sensitivity maps have been calculated, the MR pulse sequence begins. For a PI acceleration factor of 2, alternate lines of k-space are skipped, resulting in a ½-FOV images obtained from each coil with aliasing (wrap-around). A matrix inversion process is used to unfold and combine the aliased images from each coil. How this inversion process works is not as complicated as it may at first appear, so please try to follow the simple 2-pixel example below:
  • #61 During the prescan calibration step the scanner has calculated point-by-point sensitivities for each surface coil, now stored in memory as a big array of numbers. For an MR signal arising from point A in the patient, the sensitivities of Coils 1 and 2 for detecting that signal will be denoted S1A and S2Arespectively. Similarly, the coil sensitivities for any other point B are also known and will be denoted S1Band S2B. When the data from each coil are reconstructed into images, significant wrap-around (fold-over) artifact is present. This phenomenon, known as aliasing, has occurred because an insufficient number of frequency components have been sampled during the imaging process to uniquely distinguish all spatial locations. Each pixel (P) in the ½-FOV images has a signal that is the sum of contributions from two points (A and B) in the patient. Denoting these pixel values from Coils 1 and 2 by P1 and P2, we can write P1 = A•S1A + B•S1B P2 = A•S2A + B•S2B Since the Pi's and Si's are all known, the true signals (A and B) can be calculated by simple algebraic methods for solving 2 simultaneous equations with 2 unknowns. In the MR scanner a similar process is performed for all data points in the image using a matrix inversion technique, but the idea is the same. Hopefully this example will remove some of the mysteries surrounding the PI reconstruction process for SENSE.
  • #62 During the prescan calibration step the scanner has calculated point-by-point sensitivities for each surface coil, now stored in memory as a big array of numbers. For an MR signal arising from point A in the patient, the sensitivities of Coils 1 and 2 for detecting that signal will be denoted S1A and S2Arespectively. Similarly, the coil sensitivities for any other point B are also known and will be denoted S1Band S2B. When the data from each coil are reconstructed into images, significant wrap-around (fold-over) artifact is present. This phenomenon, known as aliasing, has occurred because an insufficient number of frequency components have been sampled during the imaging process to uniquely distinguish all spatial locations. Each pixel (P) in the ½-FOV images has a signal that is the sum of contributions from two points (A and B) in the patient. Denoting these pixel values from Coils 1 and 2 by P1 and P2, we can write P1 = A•S1A + B•S1B P2 = A•S2A + B•S2B Since the Pi's and Si's are all known, the true signals (A and B) can be calculated by simple algebraic methods for solving 2 simultaneous equations with 2 unknowns. In the MR scanner a similar process is performed for all data points in the image using a matrix inversion technique, but the idea is the same. Hopefully this example will remove some of the mysteries surrounding the PI reconstruction process for SENSE. NEW FROM REFFERENCE disadvantage of the pixel-by-pixel approach, however, is that in regions of low actual or apparent coil sensitivity, the matrix C may be poorly conditioned, and error propagation through the inverse may amplify the effects both of noise and of sensitivity miscalibrations
  • #65 sinusoidal modulations, or `spatial harmonics', are generated by manipulations of component coil sensitivities
  • #66 the sensitivity values Ck(x,y) are combined with appropriate linear weights nk (m) to generate composite sensitivity profiles Cm comp with sinusoidal spatial sensitivity variations of the order m
  • #67 a small number of reference k-space lines are added to the acquisition, and the usual MR signal data lines are used to `train' SMASH reconstructions directly in k-space. Maximum achievable SMASH acceleration factor M is equal to the number of independent component coils in the array. SMASH---HASTE ,EPI,GRE,FSE
  • #68 Schematic representation of the SMASH reconstruction procedure (left: k-space cartoon, right: corresponding phantom images). (A) Acquisition of data with reduced phase encoding. (B) Formation of shifted data sets using spatial harmonic combinations. (C) Interleaving of shifted data sets to generate a full signal matrix, corresponding to an image with full FOV
  • #71 .  Data Acquisition. The acquired MR signals are digitized, demodulated, and used to fill the k-space matrix for each coil. Because multiple phase-encoding steps have been skipped, many k-space lines will be missing. Lines through the center of k-space, however, are fully sampled and constitute the autocalibration signal (ACS) region. These extra ACS lines are interspersed with image acquisition itself (hence the term, "autocalibrating").      2.  Estimation of Missing Lines.  Known data from the ACS are used to calculate weighting factorsfor each coil. These weighting factors reflect how each coil distorts, smears, and displaces spatial frequencies within the full FOV k-space data. Missing k-space points are estimated in an iterative fashion using these global weighting factors combined with local known data for each small region (known a block or kernel). It should be noted that weighting factors and known data fromall coils are used to estimate missing data for each coil.       3.  Generate Individual Coil Images.  With the missing lines of k-space now filled, Fourier transformation is performed to create individual images from each coil. Unlike the coil images in SENSE/ASSET, these GRAPPA/ARC images are free from aliasing.      4.  Combine. The individual coil images are at last combined using a sum of squares method into the final magnitude image.
  • #72  2.  Estimation of Missing Lines.  Known data from the ACS are used to calculate weighting factorsfor each coil. These weighting factors reflect how each coil distorts, smears, and displaces spatial frequencies within the full FOV k-space data. Missing k-space points are estimated in an iterative fashion using these global weighting factors combined with local known data for each small region (known a block or kernel). It should be noted that weighting factors and known data fromall coils are used to estimate missing data for each coil.
  • #76 CAIPIRINHA pattern with acceleration in two directions plus phase shift of alternate rows
  • #79 SENSE/ASSET may do poorly if motion occurs between the calibration and acquisition scans, resulting in reconstruction artifacts. One obvious example is gross patient motion between sequences, even though the patient may hold perfectly still during the acquisition of each individual sequence separately. A second more subtle example applies to breath-hold liver imaging. Here, a patient may have difficulty suspending respiration to exactly the same degree between SENSE/ASSET calibration and imaging. Again, an autocalibrating technique like GRAPPA/ARC may be preferred (provided the entire sequence can be performed in the time of a single breathold). It should be noted that some newer SENSE variants (mSENSE, RAPID) use autocalibration, so this distinction disappears. GRAPPA/ARC is more tolerant toward small FOVs. SENSE/ASSET may produce aliasing/wraparound in the phase-encode direction if the full FOV is smaller than the imaged object. Conversely, GRAPPA/ARC allows a smaller FOV to be selected without significant artifact. GRAPPA/ARC holds a definite advantage over SENSE/ASSET in that susceptibility-induced field distortions are less likely to affect the reconstruction process. Some more advanced follow-up questions and explanations appear below: Why is GRAPPA better than SENSE for highly inhomogeneous regions like the lung and abdomen? The answer is that SENSE requires the generation of accurate sensitivity maps from each coil, which may be difficult to obtain. In GRAPPA, estimation of missing k-space lines is more of a global process involving data from all three coils simultaneously plus the autocalibration region; it is not as affected by local inhomogeneities. Acquiring extra lines at the center of k-space also improves signal-to-noise and estimation accuracy. Why is GRAPPA better for single-shot echo-planar imaging (EPI)? Here, coil sensitivity maps are the problem. EPI methods suffer image distortions originating from differences in local resonant frequencies due to susceptibility effects. With SENSE image distortions due to susceptibility effects and coil sensitivity maps may be different. For reasons described above, in GRAPPA these image distortions have relatively little effect on the estimation of missing k-space lines. Why is GRAPPA more tolerant to small FOVs? For some PI applications (like cardiac MRI), relatively small FOVs are needed to view the anatomy of interest. By reducing the full FOV, gross aliasing occurs beyond that induced by the PI process itself. In SENSE/ASSET this aliasing cannot be corrected by the usual PI unfolding algorithms. A "SENSE Ghost" artifact in the center for the image is commonly observed. Conversely, GRAPPA/ARC is able to generate partially aliased full FOV images without any modifications to its reconstruction algorithm. Recall that a small FOV corresponds to a greater than usual spacing between k-space lines. Simply increasing line spacing does not significantly affect the GRAPPA/ARC estimation of missing lines. Hence GRAPPA/ARC methods are preferred for small FOV applications.
  • #81 Partial volume maximum intensity projection images from renal MR angiography of a patient with fibromuscular dysplasia, obtained without (a) and with (b) SENSE (acceleration factor 2, applied in the inplane right-left phase-encoding direction). In the standard image, early filling of the left renal vein limits visualization of the left renal artery. Venous contamination is eliminated in the SENSE image, which was obtained with the acquisition time reduced by half. Collecting system activity in the SENSE image is due to excretion of contrast material from the first non-SENSE acquisition
  • #82  Maximum intensity projection images from successive frames show minimal contrast material in the right ventricle and outflow tract (a), optimal demonstration of the pulmonary arteries (b), mixed signal from the pulmonary arteries and veins (c), optimal demonstration of the pulmonary veins (d), and the aorta (e).
  • #83 imaging can alternatively be applied to improve spatial resolution. If the acquisition time is kept constant, additional phase-encoding steps can be acquired, thereby increasing resolution in the phase-encoding direction (Fig 7). However, the combination of parallel imaging and increased spatial resolution will usually result in a significant reduction in SNR. Therefore, it is important to ascertain that the signal intensity is high enough in the reconstructed images to achieve a diagnostic examination. \7. Image from 3D contrast-enhanced MR angiography of the abdominal aorta performed with SENSE (acceleration factor 2, applied in the inplane right-left direction) to improve spatial resolution. Forty 1.6-mm-thick sections were obtained in 21 seconds with an in-plane matrix of 320 256 and a 32-cm FOV, yielding a high spatial resolution of 1.0 1.25 1.6 mm.
  • #84 (Note that the phase ghosting artifact due to aortic pulsation in the standard image has been significantly reduced in the SENSE images.
  • #85 (Note that the phase ghosting artifact due to aortic pulsation in the standard image has been significantly reduced in the SENSE images.
  • #87 Coronal cardiac images showing the use of SMASH for increased spatial resolution (2D segmented-k-space FLASH, 9 echoes per segment, TR 14.4 ms, TE 7.3 ms, custom-designed four-element array, Siemens Vision scanner). The reference image has a matrix size of 144256, corresponding to an in-plane resolution of 2.2 mm1.2 mm. The SMASH image has double the spatial resolution in both dimensions (288512 matrix size, in-plane resolution 1.1 mm0.6 mm). Both images were obtained in a breath-hold lasting 16 cardiac cycles. A long segment of the right coronary artery (thick black arrows) may be seen in both images but is notably sharper in the higher-resolution SMASH image. Branches of the left coronary system (thick white arrow) may also be discerned in the SMASH image, whereas they are not seen in the reference image. Finally, internal mammary arteries (thin white arrow), invisible in the reference image, may just be discerned running down the center of the SMASH image