The document discusses magnetic resonance imaging (MRI) hardware components and artifacts. It describes the basic principles of MRI including proton imaging, spin and precession. It explains the main hardware components including the magnet, gradients, radiofrequency coils and system components. It then discusses various types of artifacts including patient-related, hardware-related and processing artifacts. Corrective measures for common artifacts like motion, chemical shift and aliasing are provided. The document emphasizes the importance of understanding artifacts to avoid confusing them with pathology.
MRI artifacts remains a big challenge to get a diagnostic image. This represents a practical comprehensive approach to understand MRI artifacts & how to get rid of.
MRI artifacts remains a big challenge to get a diagnostic image. This represents a practical comprehensive approach to understand MRI artifacts & how to get rid of.
Geometric distortion of functional MR images especially of DWI hinders the visualization of the frontal & temporal regions. This is a simple & reliable approach to get rid of. This needs no special software or hardware & applies to old systems.
Magnetic resonance imaging (MRI) is a medical imaging technique that uses a magnetic field and computer-generated radio waves to create detailed images of the organs and tissues in your body.
Most MRI machines are large, tube-shaped magnets. When you lie inside an MRI machine, the magnetic field temporarily realigns water molecules in your body. Radio waves cause these aligned atoms to produce faint signals, which are used to create cross-sectional MRI images — like slices in a loaf of bread.
The MRI machine can also produce 3D images that can be viewed from different angles.
Products & Services
Sign up for Email: Get Your Free Resource – Coping with Cancer
Why it's done
MRI is a noninvasive way for your doctor to examine your organs, tissues and skeletal system. It produces high-resolution images of the inside of the body that help diagnose a variety of problems.
MRI of the brain and spinal cord
MRI is the most frequently used imaging test of the brain and spinal cord. It's often performed to help diagnose:
Aneurysms of cerebral vessels
Disorders of the eye and inner ear
Multiple sclerosis
Spinal cord disorders
Stroke
Tumors
Brain injury from trauma
A special type of MRI is the functional MRI of the brain (fMRI). It produces images of blood flow to certain areas of the brain. It can be used to examine the brain's anatomy and determine which parts of the brain are handling critical functions.
This helps identify important language and movement control areas in the brains of people being considered for brain surgery. Functional MRI can also be used to assess damage from a head injury or from disorders such as Alzheimer's disease.
MRI of the heart and blood vessels
MRI that focuses on the heart or blood vessels can assess:
Size and function of the heart's chambers
Thickness and movement of the walls of the heart
Extent of damage caused by heart attacks or heart disease
Structural problems in the aorta, such as aneurysms or dissections
Inflammation or blockages in the blood vessels
Everything regarding the physics of MRA is given along with flow charts and images. Also have covered new advances and refrences taken from MR made easy and some articles related to MRI
Echo planar imaging (EPI) is the method of rapid magnetic resonance imaging (MRI), overcoming one of the significant disadvantage of MRI concerning with slow imaging time. However, EPI-MRI imaging comes with it's own unique imaging artifacts.
Geometric distortion of functional MR images especially of DWI hinders the visualization of the frontal & temporal regions. This is a simple & reliable approach to get rid of. This needs no special software or hardware & applies to old systems.
Magnetic resonance imaging (MRI) is a medical imaging technique that uses a magnetic field and computer-generated radio waves to create detailed images of the organs and tissues in your body.
Most MRI machines are large, tube-shaped magnets. When you lie inside an MRI machine, the magnetic field temporarily realigns water molecules in your body. Radio waves cause these aligned atoms to produce faint signals, which are used to create cross-sectional MRI images — like slices in a loaf of bread.
The MRI machine can also produce 3D images that can be viewed from different angles.
Products & Services
Sign up for Email: Get Your Free Resource – Coping with Cancer
Why it's done
MRI is a noninvasive way for your doctor to examine your organs, tissues and skeletal system. It produces high-resolution images of the inside of the body that help diagnose a variety of problems.
MRI of the brain and spinal cord
MRI is the most frequently used imaging test of the brain and spinal cord. It's often performed to help diagnose:
Aneurysms of cerebral vessels
Disorders of the eye and inner ear
Multiple sclerosis
Spinal cord disorders
Stroke
Tumors
Brain injury from trauma
A special type of MRI is the functional MRI of the brain (fMRI). It produces images of blood flow to certain areas of the brain. It can be used to examine the brain's anatomy and determine which parts of the brain are handling critical functions.
This helps identify important language and movement control areas in the brains of people being considered for brain surgery. Functional MRI can also be used to assess damage from a head injury or from disorders such as Alzheimer's disease.
MRI of the heart and blood vessels
MRI that focuses on the heart or blood vessels can assess:
Size and function of the heart's chambers
Thickness and movement of the walls of the heart
Extent of damage caused by heart attacks or heart disease
Structural problems in the aorta, such as aneurysms or dissections
Inflammation or blockages in the blood vessels
Everything regarding the physics of MRA is given along with flow charts and images. Also have covered new advances and refrences taken from MR made easy and some articles related to MRI
Echo planar imaging (EPI) is the method of rapid magnetic resonance imaging (MRI), overcoming one of the significant disadvantage of MRI concerning with slow imaging time. However, EPI-MRI imaging comes with it's own unique imaging artifacts.
mass spectrometry for pesticides residue analysis- L4sherif Taha
This is the fourth and the last lecture in series of lectures on mass spectrometry for pesticides residue analysis. this lecture present the commonly used mass to charge analyzer for pesticides residue analysis.
MRI Scanner, Instrumentation. MDIRT ST. Louis Bamenda, Nchanji Nkeh KenethNchanji Nkeh Keneth
The MRI Scanner; historical facts, Understanding Nuclear Magnetic Resonance, The Scanner componenets, Radiofrequency coils, the casing, types of MRI magnets, understanding the principle of superconductivity. MRI applications
Magnetic resonance imaging (MRI) is a medical imaging technique that uses a magnetic field and computer-generated radio waves to create detailed images of the organs and tissues in your body.
Most MRI machines are large, tube-shaped magnets. When you lie inside an MRI machine, the magnetic field temporarily realigns water molecules in your body. Radio waves cause these aligned atoms to produce faint signals, which are used to create cross-sectional MRI images — like slices in a loaf of bread.
The MRI machine can also produce 3D images that can be viewed from different angles.
Products & Services
Sign up for Email: Get Your Free Resource – Coping with Cancer
Why it's done
MRI is a noninvasive way for your doctor to examine your organs, tissues and skeletal system. It produces high-resolution images of the inside of the body that help diagnose a variety of problems.
MRI of the brain and spinal cord
MRI is the most frequently used imaging test of the brain and spinal cord. It's often performed to help diagnose:
Aneurysms of cerebral vessels
Disorders of the eye and inner ear
Multiple sclerosis
Spinal cord disorders
Stroke
Tumors
Brain injury from trauma
A special type of MRI is the functional MRI of the brain (fMRI). It produces images of blood flow to certain areas of the brain. It can be used to examine the brain's anatomy and determine which parts of the brain are handling critical functions.
This helps identify important language and movement control areas in the brains of people being considered for brain surgery. Functional MRI can also be used to assess damage from a head injury or from disorders such as Alzheimer's disease.
MRI of the heart and blood vessels
MRI that focuses on the heart or blood vessels can assess:
Size and function of the heart's chambers
Thickness and movement of the walls of the heart
Extent of damage caused by heart attacks or heart disease
Structural problems in the aorta, such as aneurysms or dissections
Inflammation or blockages in the blood vessels
Explaining all the difficult concepts with precise and accurate points, 3D models, animations and smart art graphics.
Principle
The NMR phenomenon
Theory
Precessional frequency (ν)
Chemical shift
Spin-spin interactions
Interpretation of NMR
Chemical shift (δ)
Multiplicity of the signal
Coupling constant
Instrumentation
Fourier NMR
Continuous wave NMR
Applications
Identification testing
Assay of drugs
Nuclear magnetic resonance spectroscopy, most commonly known as NMR spectroscopy or magnetic resonance spectroscopy, is a spectroscopic technique to observe local magnetic fields around atomic nuclei.
Similar to Navneet.mr hardware & artifacts related to it (20)
Model Attribute Check Company Auto PropertyCeline George
In Odoo, the multi-company feature allows you to manage multiple companies within a single Odoo database instance. Each company can have its own configurations while still sharing common resources such as products, customers, and suppliers.
This is a presentation by Dada Robert in a Your Skill Boost masterclass organised by the Excellence Foundation for South Sudan (EFSS) on Saturday, the 25th and Sunday, the 26th of May 2024.
He discussed the concept of quality improvement, emphasizing its applicability to various aspects of life, including personal, project, and program improvements. He defined quality as doing the right thing at the right time in the right way to achieve the best possible results and discussed the concept of the "gap" between what we know and what we do, and how this gap represents the areas we need to improve. He explained the scientific approach to quality improvement, which involves systematic performance analysis, testing and learning, and implementing change ideas. He also highlighted the importance of client focus and a team approach to quality improvement.
Unit 8 - Information and Communication Technology (Paper I).pdfThiyagu K
This slides describes the basic concepts of ICT, basics of Email, Emerging Technology and Digital Initiatives in Education. This presentations aligns with the UGC Paper I syllabus.
We all have good and bad thoughts from time to time and situation to situation. We are bombarded daily with spiraling thoughts(both negative and positive) creating all-consuming feel , making us difficult to manage with associated suffering. Good thoughts are like our Mob Signal (Positive thought) amidst noise(negative thought) in the atmosphere. Negative thoughts like noise outweigh positive thoughts. These thoughts often create unwanted confusion, trouble, stress and frustration in our mind as well as chaos in our physical world. Negative thoughts are also known as “distorted thinking”.
Synthetic Fiber Construction in lab .pptxPavel ( NSTU)
Synthetic fiber production is a fascinating and complex field that blends chemistry, engineering, and environmental science. By understanding these aspects, students can gain a comprehensive view of synthetic fiber production, its impact on society and the environment, and the potential for future innovations. Synthetic fibers play a crucial role in modern society, impacting various aspects of daily life, industry, and the environment. ynthetic fibers are integral to modern life, offering a range of benefits from cost-effectiveness and versatility to innovative applications and performance characteristics. While they pose environmental challenges, ongoing research and development aim to create more sustainable and eco-friendly alternatives. Understanding the importance of synthetic fibers helps in appreciating their role in the economy, industry, and daily life, while also emphasizing the need for sustainable practices and innovation.
Instructions for Submissions thorugh G- Classroom.pptxJheel Barad
This presentation provides a briefing on how to upload submissions and documents in Google Classroom. It was prepared as part of an orientation for new Sainik School in-service teacher trainees. As a training officer, my goal is to ensure that you are comfortable and proficient with this essential tool for managing assignments and fostering student engagement.
The French Revolution, which began in 1789, was a period of radical social and political upheaval in France. It marked the decline of absolute monarchies, the rise of secular and democratic republics, and the eventual rise of Napoleon Bonaparte. This revolutionary period is crucial in understanding the transition from feudalism to modernity in Europe.
For more information, visit-www.vavaclasses.com
Palestine last event orientationfvgnh .pptxRaedMohamed3
An EFL lesson about the current events in Palestine. It is intended to be for intermediate students who wish to increase their listening skills through a short lesson in power point.
How to Make a Field invisible in Odoo 17Celine George
It is possible to hide or invisible some fields in odoo. Commonly using “invisible” attribute in the field definition to invisible the fields. This slide will show how to make a field invisible in odoo 17.
3. MR imaging is based on proton imaging.
Most of the signals on MR images comes from water
molecules are mostly composed of hydrogen(H+).
Spin : rotation of a proton around its axis.
Precession : rotation of the axis itself under the influence
of external magnetic field.
Larmours equation
4.
5. Four basic steps are involved in MR
imaging :
1. Placing the patient in the magnet.
2. Sending the radiofrequency pulse by coil.
3. Receiving signals from patient by coil.
4. Transformation of signals into images by Fourier
transformation.
6. Main magnetic fields are oriented in space along x ,y and z-
axis.
- Longitudinal magnetisation.
- Transverse magnetisation
7. 3 Gradient fields :
- These have different strength in varying locations .
- produced by gradient coils.
1. Slice selection gradient
2. Frequency encoding gradient
3. Phase encoding gradient
9. All MRI scanners include several essential components
1. Polarizing magnetic field (B0)/Magnet
2. Secondary magnetic fields
Gradients
Radio frequency (RF) irradiation (B1)
10.
11.
12.
13. Magnet types used in MRI may be classified into
three categories:
Permanent.
Resistive.
Superconducting.
14. Composed of one or more pieces of iron or magnetizable
alloy
Can be open or closed cylindrical geometry type
Good spatial homogeneity, but they are susceptible to
temporal changes in field strength.
Maximum field strength approximately 0.3 T.
Weight - 5 - 25 tons
They consume no electric power, dissipate no heat, and are
very stable
Inexpensive to maintain
15. Generate their field by the conduction of electricity
through loops of wire.
Resistive or superconducting depending upon whether
the wire loops have finite or zero electrical resistance.
Lighter in weight.
Produce an strong magnetic field.
Used in open MRI.
16. Magnets are manufactured using wire composed of
Nb/Ti or Nb/Sn alloys.
Have the property of zero resistance when cooled down
to 10 K(–263°C) .
Cooled by a bath of liquid helium.
Current continues in the closed loop of the coil for years.
Magnetic field is always present.
18. Permanent Iron Core
Low Field “Open”
Resistive Electromagnet
Up to 0.2T
Superconducting Magnet
Cools wire coil with cryogens
0.5T to 35T
20. Dedicated electromagnetic coils (shim coils) are provided
to optimize the B0 field homogeneity within the design
of the main.
In a superconducting electromagnet, superconducting
shims are additional coils of superconducting wire
wound coaxially with the main coil in such a way as to
generate specific field gradients.
23. Used to produce deliberate variations in the main
magnetic field.
There are usually three sets of gradient coils, one for each
direction.
The variation in the magnetic field permits localization of
image slices as well as Frequency and Phase Encoding.
24. The set of gradient coils for the z axis are helmholtz pairs,
and for the x and y axis paired saddle coils.
Gradient fields are produced by passing current through
a set of wire coils located inside the magnet bore.
Provide linear gradations of the magnetic field means
one end of the bore of the magnet has a lesser strength
and the other a greater strength.
25.
26.
27.
28. “Antenna" of the MRI system.
Transmit the RF signal and receives the return signal.
Act as both transmitter and receiver
They are simply a loop of wire either circular or
rectangular.
29. RF TRANSMITTER
Generates RF energy in the form of RF pulses.
Applied to coil and transmitted to patient’s body.
Absorbed by the tissues.
RF RECEIVER
Short time after RF pulse transmission resonating
tissue will respond by returning Signal.
Provide data from which image is reconstructed.
Resulting image is display of RF signal.
35. Data acquisition control
Acquisition of RF signal from patient body . Sequence of
RF pulse is transmitted to the body.
Image reconstruction
computer use collected data during acquisition process to
create or construct image by Fourier transform.
Image storage
Image are stored in the computer for future viewing.
36.
37.
38. A structure not normally present.
But visible as a result of a limitation or malfunction.
Affect the quality of the MRI exam.
May be confused with pathology.
The knowledge of MRI artifacts and noise producing
factors is important.
39. Depending on their origin, artifacts are typically
classified as
Patient-related.
Signal processing dependent.
Hardware (machine)-related.
40. Patient related artefacts :
Motion artefacts.
Flow.
Metal artefacts.
Signal processing dependent artifacts:
Chemical shift artefact
Wrap around
Gibbs phenomenon (ringing artefact)
41. Machine/hardware-related artefacts:
Magnetic susceptibility artifacts.
Magnetic field in homogeneities.
Shading artefacts.
Cross excitation and cross talk artefacts.
parallel imaging artefacts.
43. Gradients applied at a very high duty cycle (eg, those in
echo-planar imaging)
Caused by bad data points, or a spike of noise, in k-space.
Seen as regularly spaced straight lines through MR image.
44.
45.
46. Due to loose electrical connections and breakdown of
interconnections in RF coil.
The location of spike and its distance from center of k
space determines the angulation b/w lines and distance
between them.
48. It is a line with alternating bright and dark pixels
propagating along the frequency encoding direction.
Mainly due to RF leakage, as in defective farady cage,
FM radio station, few electronic eqipments.
- may occur in either frequency/phase encoding direction.
49.
50. Corrective measures :
Make sure the MR scanner room-door is shut during
imaging.
Remove all electronic devices from the patient prior to
imaging.
51. If the artifact persists despite all nearby electronic
equipment being turned off, it is possible that the RF
shielding is compromised.
this usually occurs at the contacts between the door and
the jam and may need to be cleaned or repaired.
the penetration panel where the cables enter the room
is another site to be checked.
52. Ghosting and smearing, common artifacts produced by
voluntary or involuntary motion of the patient.
Random motion (patient motion) produces smearing.
Periodic motion (respiratory/cardiac/vascular) produces
discrete , well defined ghosts.
Motion related artefacts are produced in phase encoding
direction.
53.
54.
55. Esophageal contraction and vascular pulsation
during head and neck imaging.
Respiration and cardiac activity during thoracic
and abdominal imaging.
Bowel peristalsis during abdominal and pelvic imaging.
56. Phase encoding axis swap i.e, changing the phase encoding
direction.
Spatial presaturation bands placed over moving tissues.
Spatial presaturation bands placed outside the FOV.
Scanning prone to reduce abdominal excursion
Cardiac/Respiratory gating
Shorten the scan time when motion is from patient moving.
57.
58. Flow can manifest as altered intravascular signal (flow
enhancement or flow-related signal loss).
When unsaturated spins in blood first enter in to a slice or
slices.
Characterized by bright signal in a blood vessel at the first
slice.
High velocity flow - protons do not contribute to the
echo and are registered as a signal void or flow-related
signal loss.
59. GRE sequences are much more susceptible to flow
artifacts than are SE sequences.
SE sequences – flow appears dark
GRE imaging -- in-flow effect produces the bright blood
phenomenon.
Confused with thrombosis
60.
61. Use spatial saturation bands before the first and after
the last
62. Susceptibility (χ) is a measure of the extent a substance
becomes magnetized when placed in an external magnetic
field.
MSA results from local magnetic field in homogeneities
introduced by the metallic object.
Materials that disperse the main field are called diamagnetic.
63. Materials that concentrate the field are called
paramagnetic, super paramagnetic or ferromagnetic.
Ferromagnetic materials have the largest susceptibilities.
So field distortions and MR artifacts are more prominent
around metal objects and implants.
Axis : frequency and phase encoding
64.
65.
66.
67.
68. Use of spin echo sequence. (Avoid echo planar imaging)
Remove all metals.
Increasing band width .
Using thin slice and higher matrix.
Use of parallel imaging.
Using radial k-space sampling.
69. Good effects :
Used to diagnose hemorrhage , hemosiderin deposition
and calcification.
Forms the basis of post contrast T2* weighted MR
perfusion studies.
Used to quantify myocardial
and liver iron overload.
70. Chemical shift is due to the differences between resonance
(Larmour) frequencies of fat and water.
Which leads to shift in the detected anatomy.
At 1.5 T, protons from fat resonate at a point approximately
220 Hz downfield from the water.
If this occurs, there will be a slight misregistration of the fat
content on images, because of the slight shift in the frequency
of the fat protons.
71. There are two types of chemical shift artefacts:
1. Chemical shift misregistration artefacts.
2. Interference from chemical shift (in-phase/out-
phase).
72.
73. Based on the difference in frequency between fat and water
(about 220 Hz).
So the fat and water signals are in phase every 1/220 Hz.
In other words, they are in phase at 4.45 msec, 8.9 msec, and
so on.
At the in-phase echo times, the water and fat signals are
summed up
whereas at opposed-phase echo times, the signal in these
pixels is canceled out, leading to the appearance of a dark
band at the fat-water interface
IN-PHASE/OUT- PHASE
74.
75. Chemical shift misregistration Interference from chemical
shift(IPOP)
Along frequency encoding axis Along phase encoding axis
Dark edge at the interface between
fat and water
Dark edge around certain organs
Corrected by :
1.Fat suppression
2.Increasing band width
1.Using spin echo sequences
2. Selecting TE that is multiple of 4.5
at 1.5T.
Good effects:
Forms the basis of MR spectroscopy In phase and out phase imaging is used
to detect fat in lesion. (Eg – Fatty
infiltration)
76. Chemical shift is inversely proportional to bandwidth.
Increasing the bandwidth reduces the chemical shift.
Fat suppressed imaging can be used to eliminate the
chemical shift misregistration.
Use of a spin echo sequence instead of a gradient echo can
eliminate the black boundary artifact.
77. When the imaging field of view is smaller than the
anatomy being imaged.
Anatomy that appears outside the FOV appears within
image and on opposite side.
Aliasing can occur along any axis – frequency encoding
axis/ phase encoding axis.
Basis of aliasing lies in analog to digital conversion.
78.
79.
80. Enlarging the field of view (FOV).
Using pre-saturation bands on areas outside the FOV.
Anti-aliasing software.
Switching the phase and frequency directions.
Use a surface coil to reduce the signal outside of the area of
interest.
81. Occurs at high contrast boundaries.
Due to truncation(omission) of sampled signals.
Commonly seen at the low signal intensity spinal cord with
high signal intensity CSF on T2WI of the spine.
Occurs due to the omission of some data in k-space.
Which cause the signal intensity of a given pixel to vary
from its ideal signal intensity.
Appearance
Bright and dark lines.
82.
83. Increasing the matrix size.
Use of smoothing filters.
If fat is one of the boundaries, use of fat suppression.
84.
85. Patients who are very large promote this artifact.
The image will have uneven contrast with loss of signal
intensity in one part of the image.
Frequency and phase encoding direction.
- Causes :
1. Uneven excitation of nuclei within patient .
2. Abnormal loading of coil/coupling of coil
3. Inhomogeneity of magnetic field.
4. Over flow of analog to digital converter.
86.
87. Load the coil correctly.
Shimming to reduce inhomogenecity of field.
Acquire the images with less amplification to avoid ADC
overflow.
88.
89. During slice selection there is some degree of excitation
of adjacent slices as well.
Partial saturation of the signal in that section leads to a
lower signal intensity.
If that slice is imaged soon after without living a gap
there will be a loss of signal.
90.
91.
92. Same as cross excitation.
Produced when RF pulse is switched off.
Due to dissipation of energy to nuclei from neighboring
nuclei.
Seen in slice selection gradient.
93. Increase inter slice gap.
Scanning with interleaved slices with odd and even
numbered slice respectively.
Optimized RF pulse that have a more rectangular slice
profile can be implemented.
94. Parallel imaging allows for faster image acquisition and
potentially improved image quality by under sampling k-
space.
The reduced acquisition time can be used to improve spatial
or temporal resolution.
High acceleration factor and small FOV is used.
Graininess in the image.
Can be reduced by using lower acceleration factor and
increasing the FOV.
97. Seen when doing gradient echo images with the body
coil.
Seen in heavy pt.
Seen in coronal imaging when large FOV is used
which is still smaller than the object in view.
Because of lack of perfect homogeneity of the main
magnetic field from one side of the body to the other
(at edges).
98. Aliasing of one side of the body to the other results in
superimposition of signals of different phases that
alternatively add and cancel.
This causes the banding appearance and is similar to the
effect of looking though two screen windows.
99. • Use Surface coil
• Shimming
• Arms by the side of
pt. so that they are
within the FOV.
100. An artifactual T2 bright signal seen in tendons and
ligaments that are oriented at about a 55 degree angle to
the main magnetic field.
A bright signal from this artifact is commonly seen in the
rotator cuff and distal patellar tendon.
- Not to be mistaken for tear.
- Tends to occur only on short TE sequences (e.g. T1, GRE, PD).
101.
102.
103. Lengthen TE
Use T1 weighted sequences since T1 relaxation is
unaffected by this.
104. A focal dot of increased or decreased signal in the
center of an image.
Caused by a constant offset of the DC voltage in the
amplifiers.
Remedy
Requires recalibration by engineer
Maintain a constant temperature in equipment room
for amplifiers.
105.
106. MR has achieved widespread clinical acceptance as A
noninvasive, radiation–free alternative to
radiography and computed tomography.
MR artefacts hinders the image quality.
MR artefacts are confused with pathology.
Identifying them and correcting them by radiologist.
Editor's Notes
Frequency of precession = magnetic field strength x gyromagnetic ratio
Conversion of time domain into frequency domain.
By passing a particular RF pulse we introduce transverse magnetization to obtain signal, which is done in slice selction.
Frequency encoding is required to locate position within the slice.
Niobium titanium tin
Unlike superconducting shims, passive shims do not rely upon the flow of electrical current through a coil to generate a field gradient. Instead, they are pieces of ferromagnetic metal of a size and shape designed to improve B0 homogeneity when they are inserted into the magnet.
The knowledge of MRI artifacts and noise producing factors is important for continuing maintenance of image quality.
But visible as a result of a limitation or malfunction.
in the hardware or software in the MRI device.
Spike artifact. Bad data points in k-space (arrow in b) result in band artifacts on the MR image in a. The location of the baddata points, and their distance from the center of k-space, determine the angulation of the bands and the distance between them. The intensity of the spike determines the severity of the artifact.
k-space is the 2D or 3D Fourier transform of the MR image measured
Saturation pulses involve the application of RF energy to suppress the MR signal from moving tissue outside the imaged volume to reduce/eliminate the motion artifacts.
(a, c) Images acquired
without compensation show respiration-induced artifact. (b, d) Images acquired with compensation
are unaffected by respiratory motion.
Figure
This results from local magnetic field inhomogeneities introduced by the metallic object into the otherwise homogeneous external magnetic field B0
Echo-planar images show magnetic susceptibility artifacts. (a) Left-right phase encoding causes severe distortion of the signal (arrows). (b) Anterior-posterior phase encoding minimizes the distortion, since the phase axis is symmetric around the susceptibility gradients.
This signal loss is especially severe at air-tissue or bone–soft tissue boundaries, because air and bone have much lower magnetic susceptibility than do most tissues
Chemical shift artifact at echo-planar imaging. (a) Image shows severe chemical shift artifact from insufficient fat suppression. (b) Image obtained
with fat saturation shows minimization of the chemical shift artifact or off-resonance effect.
In-phase MR image acquired with an echo time of 2.2 msec. (b) Opposed-phase MR image acquired with an echo time of 4.4 msec. Both images were acquired at 1.5 T.
Bandwidth is the range of frequencies involved in the transmission or reception of an electronic signal.
Aliasing artifact caused by phase errors at both sides of the magnet. Image shows a moire´ artifact produced by the addition and cancellation of signals
2.(as with large pt whom touches one side of the coil)
Real RF pulse is a truncated
Parallel imaging artifacts resulting from excessively small fields of view. A comparison of images acquired with a field of view of 22 cm (a), 18 cm(b), and 16 cm (c) and with a constant acceleration factor of two shows increasing severity of aliasing artifacts, as well as increasing noise at the image
center, in b and c.
Figure 24 shows increased
ghosting and noise in the center of the image with
progressive reductions in the field of view