 As per Atomic Energy (Radiation Protection)
Rules 2004 Section (3) a License / Registration
is necessary for establishing a radiation
installation.
 This process involves 3 steps.
 (i) Site Approval by AERB
 (ii) Quality Assurance Test by
manufacturer/supplier under intimation to DRS
(iii) License/Registration by AERB .
Guideline 1
 Submit completely filled and duly signed “Application
for Site Approval of X-Ray Installation Layout” along
with layout plan of the diagnostic x-ray room, drawn
to scale 1:50 showing details as described in the
application form, in original with two copies for
approval to the Director, Directorate of Radiation
Safety,AERB , Niyamak Bhavan, Anusaktinagar ,
Mumbai-400094.
 Separate form should be submitted for each
medical diagnostic x-ray unit installation. Mobile
units do not require plan approval (mobile units
should not be used in a room with table) but
need Registration Certificate.
 Enclose fees in the form of Demand Draft drawn
in favor of the Director of Radiation Safety
payable at AERB, at the rates prescribed by the
Government.
Guideline 2
 The room housing an x-ray unit shall not be less
than 18m2 for general-purpose radiography and
conventional fluoroscopy equipment and no single
dimension of the x-ray room shall be less than 4.0m.
 The room housing the gantry of the computed
tomography (CT Scan) unit shall not be less than 25
m 2 and no single dimension of the gantry of CT
Scan room shall be less than 4m.
 The room housing an orthopan tomography
(OPG) units shall not be less that 15 m2 and no
single dimension of the OPG room shall be less
than 3.5m.
 The room housing a mammography x-ray unit
shall not be less than 10 m2 and no single
dimension of x-ray room shall be less than 3m.
Also, not more than one unit of any type shall be
installed in the same room.
Cont
 The minimum wall thickness of the x-ray room
should be 9 inch or 23cm solid brick or
equivalent.
 There shall be no windows and ventilations shall
be above 2m from the floor outside. Table shall
be oriented such that the primary (chest stand) is
not directed towards the main door or darkroom.
 The primary wall thickness shall be 35 cm of
solid brick or 32 cm solid brick + 1.5 mm LEAD
Guideline 3
 The users should install only Atomic Energy
Regulatory Board (AERB) “Type Approved”
unit or unit with valid NOC for medical
diagnostic purpose and obtain a copy of the
AERB “Type Approval” certificate from the
supplier/manufacturer of the x-ray unit prior
to the procurement of the unit.
 User may verify the Type Approval
Certificate Number from AERB in case of
any doubt.
The users should procure and use
protective accessories such as mobile
protective barrier, lead rubber aprons, lead
rubber flaps, gloves, thyroid shields, etc.
 The employer shall provide Personnel Monitoring
Badges to all radiation workers of the institute.
For procuring personnel monitoring badges,
please contact Avanttec Laboratories Private
Limited, #31, Kamaraj Street, Srinivasa Nagar,
Padi, Chennai-600 050.
 The users should follow the guidelines given in AERB
Safety Code for Medical Diagnostic X-Ray Equipment
and Installations [Code No.AERB/SC/MED-2 (Rev.-1)].
 The same can be purchased from AERB or downloaded
from AERB website www.aerb.gov.in.
 The user shall comply with other regulatory documents
related to the use and operation of x-ray units that are
available in AERB website or DRS website.
 i Atomic Energy (Radiation Protection) Rules, 2004
(G.S.R.303)
ii Radiation Surveillance Procedures for Medical
Applications of Radiation, 1989 (G.S.R.No.388)
iii Performa for “Registration of Diagnostic X-Ray
installations and Licence of Computed Tomography
(CT)/Cathlab unit Installations”.
iv Documents (i) & (ii) can be purchased from bookstalls
selling government publications or can be downloaded
from DRS website. (iii) shall be given by the Directorate
of Radiation Safety on Site Approval
 The users should request the supplier/manufacturer
of the diagnostic x-ray units to conduct Quality
Assurance (QA) tests after the installation of x-ray
unit/computed tomography unit/mammography/
cathlab unit prior to using it for patient diagnosis,
under intimation to the DRS.
 Preserve the QA tests report in the
hospital/institution record for further use and submit
a copy of QA tests report to the DRS. The unit
should be used for patient diagnosis only if its
performance is found satisfactory.
 After installation, the users should send the completely filled
and duly signed “Application for Registration of Diagnostic X-
Ray installations” / “Application for Licence of Computed
Tomography (CT) unit / Interventional Radiology (Cathlab)
Installations” (as applicable) to the Director, Directorate of
Radiation Safety, AERB for onward transmission with
recommendation to the Atomic Energy Regulatory Board for
obtaining Registration Certificate for diagnostic x-ray units
and/or License Certificate for CT Scan/Cathlab units for
operation.
The employer shall employ only qualified
Radiologist and Medical imaging
Technologist. The minimum qualification is
given in AERB Safety Code for Medical
Diagnostic X-Ray Equipment and
Installations [Code No.AERB/SC/MED-
2(Rev-1)]
AERB issues notice every year in all
leading newspapers in the country (both
Hindi and English) regarding “Statutory
Requirements for Safe Operation of
Medical diagnostic X-ray units by
Hospitals, Clinics and other Medical
Institutions in India”.
Design Certificate.
Apporoval type/ NOC
Layout approval and registration.
Certification of RSO/ Service engineers
e-LORA is a web based I&CT application
implemented by AERB to facilitate
licensing of radiation facilities.
e-governance initiative by AERB.
Objective:
Enhanced efficiency and transparency in
the regulatory processes of AERB.
Aimed at achieving paperless licensing.
Step : 1 creation of account in e-LORAfor
user D & password.
Step 2 : Approval letter.
Step 3: QA report for applying license.
Step 4 : TLD badges for doctor, operator.
Step 5 : Qualified Technicians.
Step 6: Approved RSO.
Step 7 : Radiation License certificate.
Aerb guidelines for Room installation
Aerb guidelines for Room installation
Aerb guidelines for Room installation
Aerb guidelines for Room installation

More Related Content

What's hot

Computed radiography
Computed radiographyComputed radiography
Computed radiography
Anjan Dangal
 
Qc in xray
Qc in xrayQc in xray
Qc in xray
Anupam Niraula
 
Pediatric radiography
Pediatric radiographyPediatric radiography
Pediatric radiographyJulie Parsons
 
Information regarding e lora for AERB registration
Information regarding e lora for AERB registrationInformation regarding e lora for AERB registration
Information regarding e lora for AERB registration
Dr Jitu Lal Meena
 
NCRP National Council On Radiation Protection
NCRP  National Council On Radiation ProtectionNCRP  National Council On Radiation Protection
NCRP National Council On Radiation Protection
Chandan Prasad
 
Ultrasound contrast agents
Ultrasound contrast agentsUltrasound contrast agents
Ultrasound contrast agents
Naseem Abu hawash
 
CT angiography
CT angiographyCT angiography
CT angiography
Jaya Yadav
 
Production of xrays
Production of xraysProduction of xrays
Production of xrays
priyanka rana
 
Post processing of computed tomography
Post processing of computed tomographyPost processing of computed tomography
Post processing of computed tomography
BeuniquewithNehaSing
 
Emergency radiography
Emergency radiography Emergency radiography
Emergency radiography
milan timilsina
 
X ray test machine and its calibration
X ray test machine and its calibrationX ray test machine and its calibration
X ray test machine and its calibration
Istanbul Technical University
 
CT fluoroscopy1 (1).pptx
CT fluoroscopy1 (1).pptxCT fluoroscopy1 (1).pptx
CT fluoroscopy1 (1).pptx
VANI PUSHPA MUDAVATH
 
Ct instrumentation and types of detector configuration
Ct instrumentation and types of detector configurationCt instrumentation and types of detector configuration
Ct instrumentation and types of detector configuration
SUJAN KARKI
 
Fluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptxFluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptx
Rohit Bansal
 
Computed radiography
Computed radiographyComputed radiography
Computed radiographyLirih Kuu
 
Emegency drugs in radiology department
Emegency drugs in radiology department Emegency drugs in radiology department
Emegency drugs in radiology department
AbubakarMustaphaAman
 
OPERATION THEATRE RADIOGRAPHY (3)-1.pptx
OPERATION THEATRE RADIOGRAPHY (3)-1.pptxOPERATION THEATRE RADIOGRAPHY (3)-1.pptx
OPERATION THEATRE RADIOGRAPHY (3)-1.pptx
VanshikaGarg76
 
Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das
AkashDas169
 
X ray generator, circuits
X  ray generator, circuitsX  ray generator, circuits
X ray generator, circuits
Anjan Dangal
 

What's hot (20)

Computed radiography
Computed radiographyComputed radiography
Computed radiography
 
Qc in xray
Qc in xrayQc in xray
Qc in xray
 
Pediatric radiography
Pediatric radiographyPediatric radiography
Pediatric radiography
 
Information regarding e lora for AERB registration
Information regarding e lora for AERB registrationInformation regarding e lora for AERB registration
Information regarding e lora for AERB registration
 
NCRP National Council On Radiation Protection
NCRP  National Council On Radiation ProtectionNCRP  National Council On Radiation Protection
NCRP National Council On Radiation Protection
 
Ultrasound contrast agents
Ultrasound contrast agentsUltrasound contrast agents
Ultrasound contrast agents
 
CT angiography
CT angiographyCT angiography
CT angiography
 
Production of xrays
Production of xraysProduction of xrays
Production of xrays
 
Post processing of computed tomography
Post processing of computed tomographyPost processing of computed tomography
Post processing of computed tomography
 
Emergency radiography
Emergency radiography Emergency radiography
Emergency radiography
 
X ray test machine and its calibration
X ray test machine and its calibrationX ray test machine and its calibration
X ray test machine and its calibration
 
Basics in ct
Basics in ctBasics in ct
Basics in ct
 
CT fluoroscopy1 (1).pptx
CT fluoroscopy1 (1).pptxCT fluoroscopy1 (1).pptx
CT fluoroscopy1 (1).pptx
 
Ct instrumentation and types of detector configuration
Ct instrumentation and types of detector configurationCt instrumentation and types of detector configuration
Ct instrumentation and types of detector configuration
 
Fluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptxFluoroscopy-Rohit.pptx
Fluoroscopy-Rohit.pptx
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
 
Emegency drugs in radiology department
Emegency drugs in radiology department Emegency drugs in radiology department
Emegency drugs in radiology department
 
OPERATION THEATRE RADIOGRAPHY (3)-1.pptx
OPERATION THEATRE RADIOGRAPHY (3)-1.pptxOPERATION THEATRE RADIOGRAPHY (3)-1.pptx
OPERATION THEATRE RADIOGRAPHY (3)-1.pptx
 
Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das Flat Panel Digital Fluoroscopy by Akash Das
Flat Panel Digital Fluoroscopy by Akash Das
 
X ray generator, circuits
X  ray generator, circuitsX  ray generator, circuits
X ray generator, circuits
 

Similar to Aerb guidelines for Room installation

Aerb guidelines for x ray and ct installation
Aerb guidelines for x ray and ct installationAerb guidelines for x ray and ct installation
Aerb guidelines for x ray and ct installation
Aabid Rahiman
 
Aerb Guidlines for starting , decommissioning, replacing source and responsib...
Aerb Guidlines for starting , decommissioning, replacing source and responsib...Aerb Guidlines for starting , decommissioning, replacing source and responsib...
Aerb Guidlines for starting , decommissioning, replacing source and responsib...
Nithin Sadanandan
 
New Microsoft PowerPoint Presentation.pptx
New Microsoft PowerPoint Presentation.pptxNew Microsoft PowerPoint Presentation.pptx
New Microsoft PowerPoint Presentation.pptx
TaushifulHoque
 
AERB registration for a dental RVG x-ray machine.pdf
AERB registration for a dental RVG x-ray machine.pdfAERB registration for a dental RVG x-ray machine.pdf
AERB registration for a dental RVG x-ray machine.pdf
Jagriti Agarwal
 
Setting a new radiotherapy facility-steps involved
Setting a new radiotherapy facility-steps involvedSetting a new radiotherapy facility-steps involved
Setting a new radiotherapy facility-steps involved
Anil Gupta
 
Qualification of INFARED SPECTROSCOPY BY HIMAJA.D
Qualification of INFARED SPECTROSCOPY BY HIMAJA.DQualification of INFARED SPECTROSCOPY BY HIMAJA.D
Qualification of INFARED SPECTROSCOPY BY HIMAJA.D
himaja donthula
 
37/2008 Rules for importing, exporting
37/2008 Rules for importing, exporting37/2008 Rules for importing, exporting
37/2008 Rules for importing, exporting
traoman
 
Atomic energy regulatory board (aerb)
Atomic energy regulatory board (aerb)Atomic energy regulatory board (aerb)
Atomic energy regulatory board (aerb)
Chandan Prasad
 
Clinicals In The Uk 26 January 2009
Clinicals In The Uk   26 January 2009Clinicals In The Uk   26 January 2009
Clinicals In The Uk 26 January 2009
Colin Rylett
 
RESNET: Interim Guidelines for Thermographic Inspections of Buildings
RESNET: Interim Guidelines for Thermographic Inspections of BuildingsRESNET: Interim Guidelines for Thermographic Inspections of Buildings
RESNET: Interim Guidelines for Thermographic Inspections of Buildings
RESNET - Make Your Home More Energy Efficient
 
IECEx: Conformity Assessment Solution for the Ex field
IECEx: Conformity Assessment Solution for the Ex fieldIECEx: Conformity Assessment Solution for the Ex field
IECEx: Conformity Assessment Solution for the Ex field
Statybos Akademija
 
Guide testing-and-tagging-portable-electrical-equipment-and-residual-current-...
Guide testing-and-tagging-portable-electrical-equipment-and-residual-current-...Guide testing-and-tagging-portable-electrical-equipment-and-residual-current-...
Guide testing-and-tagging-portable-electrical-equipment-and-residual-current-...
دكتور تامر عبدالله شراكى
 
Mobile Payment Successes and Challenges of the Industry and the Consumer’s Pe...
Mobile Payment Successes and Challenges of the Industry and the Consumer’s Pe...Mobile Payment Successes and Challenges of the Industry and the Consumer’s Pe...
Mobile Payment Successes and Challenges of the Industry and the Consumer’s Pe...
Nurul Yakin
 
싸이퍼랩PDA CipherLab RS30 2D 4.7인치 안드로이드PDA 산업용PDA 기업용모바일컴퓨터 바코드PDA 매뉴얼
싸이퍼랩PDA CipherLab RS30 2D 4.7인치 안드로이드PDA 산업용PDA 기업용모바일컴퓨터 바코드PDA 매뉴얼싸이퍼랩PDA CipherLab RS30 2D 4.7인치 안드로이드PDA 산업용PDA 기업용모바일컴퓨터 바코드PDA 매뉴얼
싸이퍼랩PDA CipherLab RS30 2D 4.7인치 안드로이드PDA 산업용PDA 기업용모바일컴퓨터 바코드PDA 매뉴얼
HION IT
 
biomedical instrumentation
biomedical  instrumentationbiomedical  instrumentation
biomedical instrumentation
samiharahman16
 
Distribution Code (Framework).pdf
Distribution Code (Framework).pdfDistribution Code (Framework).pdf
Distribution Code (Framework).pdf
SandeepGuhaNiyogi2
 
RADIATION SAFETY AND REGULATORY REQUIREMENT IN NUCLEAR GAUGING FACILITIES
RADIATION SAFETY AND REGULATORY REQUIREMENT IN NUCLEAR GAUGING FACILITIESRADIATION SAFETY AND REGULATORY REQUIREMENT IN NUCLEAR GAUGING FACILITIES
RADIATION SAFETY AND REGULATORY REQUIREMENT IN NUCLEAR GAUGING FACILITIESMuhammad Bello Gusau
 
Qualification of analytical instrument of FTIR
Qualification of analytical instrument of FTIR Qualification of analytical instrument of FTIR
Qualification of analytical instrument of FTIR
KUNDLAJAYALAKSHMI
 

Similar to Aerb guidelines for Room installation (20)

Aerb guidelines for x ray and ct installation
Aerb guidelines for x ray and ct installationAerb guidelines for x ray and ct installation
Aerb guidelines for x ray and ct installation
 
Aerb Guidlines for starting , decommissioning, replacing source and responsib...
Aerb Guidlines for starting , decommissioning, replacing source and responsib...Aerb Guidlines for starting , decommissioning, replacing source and responsib...
Aerb Guidlines for starting , decommissioning, replacing source and responsib...
 
New Microsoft PowerPoint Presentation.pptx
New Microsoft PowerPoint Presentation.pptxNew Microsoft PowerPoint Presentation.pptx
New Microsoft PowerPoint Presentation.pptx
 
AERB registration for a dental RVG x-ray machine.pdf
AERB registration for a dental RVG x-ray machine.pdfAERB registration for a dental RVG x-ray machine.pdf
AERB registration for a dental RVG x-ray machine.pdf
 
Setting a new radiotherapy facility-steps involved
Setting a new radiotherapy facility-steps involvedSetting a new radiotherapy facility-steps involved
Setting a new radiotherapy facility-steps involved
 
Qualification of INFARED SPECTROSCOPY BY HIMAJA.D
Qualification of INFARED SPECTROSCOPY BY HIMAJA.DQualification of INFARED SPECTROSCOPY BY HIMAJA.D
Qualification of INFARED SPECTROSCOPY BY HIMAJA.D
 
37/2008 Rules for importing, exporting
37/2008 Rules for importing, exporting37/2008 Rules for importing, exporting
37/2008 Rules for importing, exporting
 
Atomic energy regulatory board (aerb)
Atomic energy regulatory board (aerb)Atomic energy regulatory board (aerb)
Atomic energy regulatory board (aerb)
 
Clinicals In The Uk 26 January 2009
Clinicals In The Uk   26 January 2009Clinicals In The Uk   26 January 2009
Clinicals In The Uk 26 January 2009
 
RESNET: Interim Guidelines for Thermographic Inspections of Buildings
RESNET: Interim Guidelines for Thermographic Inspections of BuildingsRESNET: Interim Guidelines for Thermographic Inspections of Buildings
RESNET: Interim Guidelines for Thermographic Inspections of Buildings
 
IECEx: Conformity Assessment Solution for the Ex field
IECEx: Conformity Assessment Solution for the Ex fieldIECEx: Conformity Assessment Solution for the Ex field
IECEx: Conformity Assessment Solution for the Ex field
 
Guide testing-and-tagging-portable-electrical-equipment-and-residual-current-...
Guide testing-and-tagging-portable-electrical-equipment-and-residual-current-...Guide testing-and-tagging-portable-electrical-equipment-and-residual-current-...
Guide testing-and-tagging-portable-electrical-equipment-and-residual-current-...
 
Mobile Payment Successes and Challenges of the Industry and the Consumer’s Pe...
Mobile Payment Successes and Challenges of the Industry and the Consumer’s Pe...Mobile Payment Successes and Challenges of the Industry and the Consumer’s Pe...
Mobile Payment Successes and Challenges of the Industry and the Consumer’s Pe...
 
싸이퍼랩PDA CipherLab RS30 2D 4.7인치 안드로이드PDA 산업용PDA 기업용모바일컴퓨터 바코드PDA 매뉴얼
싸이퍼랩PDA CipherLab RS30 2D 4.7인치 안드로이드PDA 산업용PDA 기업용모바일컴퓨터 바코드PDA 매뉴얼싸이퍼랩PDA CipherLab RS30 2D 4.7인치 안드로이드PDA 산업용PDA 기업용모바일컴퓨터 바코드PDA 매뉴얼
싸이퍼랩PDA CipherLab RS30 2D 4.7인치 안드로이드PDA 산업용PDA 기업용모바일컴퓨터 바코드PDA 매뉴얼
 
Rmn rpr-2004
Rmn rpr-2004Rmn rpr-2004
Rmn rpr-2004
 
biomedical instrumentation
biomedical  instrumentationbiomedical  instrumentation
biomedical instrumentation
 
Distribution Code (Framework).pdf
Distribution Code (Framework).pdfDistribution Code (Framework).pdf
Distribution Code (Framework).pdf
 
RADIATION SAFETY AND REGULATORY REQUIREMENT IN NUCLEAR GAUGING FACILITIES
RADIATION SAFETY AND REGULATORY REQUIREMENT IN NUCLEAR GAUGING FACILITIESRADIATION SAFETY AND REGULATORY REQUIREMENT IN NUCLEAR GAUGING FACILITIES
RADIATION SAFETY AND REGULATORY REQUIREMENT IN NUCLEAR GAUGING FACILITIES
 
National radio regulations
National radio regulationsNational radio regulations
National radio regulations
 
Qualification of analytical instrument of FTIR
Qualification of analytical instrument of FTIR Qualification of analytical instrument of FTIR
Qualification of analytical instrument of FTIR
 

More from Self

Image formation of Radiology from Radiation
Image formation of Radiology from RadiationImage formation of Radiology from Radiation
Image formation of Radiology from Radiation
Self
 
Tomographic equipment
Tomographic equipmentTomographic equipment
Tomographic equipment
Self
 
Multisection tomography and Transaxial Tomography
Multisection tomography and Transaxial TomographyMultisection tomography and Transaxial Tomography
Multisection tomography and Transaxial Tomography
Self
 
Image quality in fluroscopy
Image quality in fluroscopyImage quality in fluroscopy
Image quality in fluroscopy
Self
 
Qa of ct and mamography
Qa of ct and mamographyQa of ct and mamography
Qa of ct and mamography
Self
 
Mri coils
Mri coilsMri coils
Mri coils
Self
 
Barium swallow examination
Barium swallow examinationBarium swallow examination
Barium swallow examination
Self
 
Roles and responsibility of Radiology professional
Roles and responsibility of Radiology professionalRoles and responsibility of Radiology professional
Roles and responsibility of Radiology professional
Self
 
Dexa and its aplication
Dexa and its aplicationDexa and its aplication
Dexa and its aplication
Self
 
Various Pelvic lines
Various Pelvic linesVarious Pelvic lines
Various Pelvic lines
Self
 
Chest decubitus
Chest decubitusChest decubitus
Chest decubitus
Self
 
Pelvimetry
PelvimetryPelvimetry
Pelvimetry
Self
 
Planning and organization of Radiology Department
Planning and organization of Radiology DepartmentPlanning and organization of Radiology Department
Planning and organization of Radiology Department
Self
 
Aims of rp
Aims of rpAims of rp
Aims of rp
Self
 

More from Self (14)

Image formation of Radiology from Radiation
Image formation of Radiology from RadiationImage formation of Radiology from Radiation
Image formation of Radiology from Radiation
 
Tomographic equipment
Tomographic equipmentTomographic equipment
Tomographic equipment
 
Multisection tomography and Transaxial Tomography
Multisection tomography and Transaxial TomographyMultisection tomography and Transaxial Tomography
Multisection tomography and Transaxial Tomography
 
Image quality in fluroscopy
Image quality in fluroscopyImage quality in fluroscopy
Image quality in fluroscopy
 
Qa of ct and mamography
Qa of ct and mamographyQa of ct and mamography
Qa of ct and mamography
 
Mri coils
Mri coilsMri coils
Mri coils
 
Barium swallow examination
Barium swallow examinationBarium swallow examination
Barium swallow examination
 
Roles and responsibility of Radiology professional
Roles and responsibility of Radiology professionalRoles and responsibility of Radiology professional
Roles and responsibility of Radiology professional
 
Dexa and its aplication
Dexa and its aplicationDexa and its aplication
Dexa and its aplication
 
Various Pelvic lines
Various Pelvic linesVarious Pelvic lines
Various Pelvic lines
 
Chest decubitus
Chest decubitusChest decubitus
Chest decubitus
 
Pelvimetry
PelvimetryPelvimetry
Pelvimetry
 
Planning and organization of Radiology Department
Planning and organization of Radiology DepartmentPlanning and organization of Radiology Department
Planning and organization of Radiology Department
 
Aims of rp
Aims of rpAims of rp
Aims of rp
 

Recently uploaded

Welcome to TechSoup New Member Orientation and Q&A (May 2024).pdf
Welcome to TechSoup   New Member Orientation and Q&A (May 2024).pdfWelcome to TechSoup   New Member Orientation and Q&A (May 2024).pdf
Welcome to TechSoup New Member Orientation and Q&A (May 2024).pdf
TechSoup
 
Overview on Edible Vaccine: Pros & Cons with Mechanism
Overview on Edible Vaccine: Pros & Cons with MechanismOverview on Edible Vaccine: Pros & Cons with Mechanism
Overview on Edible Vaccine: Pros & Cons with Mechanism
DeeptiGupta154
 
CACJapan - GROUP Presentation 1- Wk 4.pdf
CACJapan - GROUP Presentation 1- Wk 4.pdfCACJapan - GROUP Presentation 1- Wk 4.pdf
CACJapan - GROUP Presentation 1- Wk 4.pdf
camakaiclarkmusic
 
How libraries can support authors with open access requirements for UKRI fund...
How libraries can support authors with open access requirements for UKRI fund...How libraries can support authors with open access requirements for UKRI fund...
How libraries can support authors with open access requirements for UKRI fund...
Jisc
 
Guidance_and_Counselling.pdf B.Ed. 4th Semester
Guidance_and_Counselling.pdf B.Ed. 4th SemesterGuidance_and_Counselling.pdf B.Ed. 4th Semester
Guidance_and_Counselling.pdf B.Ed. 4th Semester
Atul Kumar Singh
 
Adversarial Attention Modeling for Multi-dimensional Emotion Regression.pdf
Adversarial Attention Modeling for Multi-dimensional Emotion Regression.pdfAdversarial Attention Modeling for Multi-dimensional Emotion Regression.pdf
Adversarial Attention Modeling for Multi-dimensional Emotion Regression.pdf
Po-Chuan Chen
 
Home assignment II on Spectroscopy 2024 Answers.pdf
Home assignment II on Spectroscopy 2024 Answers.pdfHome assignment II on Spectroscopy 2024 Answers.pdf
Home assignment II on Spectroscopy 2024 Answers.pdf
Tamralipta Mahavidyalaya
 
special B.ed 2nd year old paper_20240531.pdf
special B.ed 2nd year old paper_20240531.pdfspecial B.ed 2nd year old paper_20240531.pdf
special B.ed 2nd year old paper_20240531.pdf
Special education needs
 
Palestine last event orientationfvgnh .pptx
Palestine last event orientationfvgnh .pptxPalestine last event orientationfvgnh .pptx
Palestine last event orientationfvgnh .pptx
RaedMohamed3
 
Synthetic Fiber Construction in lab .pptx
Synthetic Fiber Construction in lab .pptxSynthetic Fiber Construction in lab .pptx
Synthetic Fiber Construction in lab .pptx
Pavel ( NSTU)
 
Unit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdfUnit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdf
Thiyagu K
 
Honest Reviews of Tim Han LMA Course Program.pptx
Honest Reviews of Tim Han LMA Course Program.pptxHonest Reviews of Tim Han LMA Course Program.pptx
Honest Reviews of Tim Han LMA Course Program.pptx
timhan337
 
Unit 8 - Information and Communication Technology (Paper I).pdf
Unit 8 - Information and Communication Technology (Paper I).pdfUnit 8 - Information and Communication Technology (Paper I).pdf
Unit 8 - Information and Communication Technology (Paper I).pdf
Thiyagu K
 
The approach at University of Liverpool.pptx
The approach at University of Liverpool.pptxThe approach at University of Liverpool.pptx
The approach at University of Liverpool.pptx
Jisc
 
Additional Benefits for Employee Website.pdf
Additional Benefits for Employee Website.pdfAdditional Benefits for Employee Website.pdf
Additional Benefits for Employee Website.pdf
joachimlavalley1
 
The Accursed House by Émile Gaboriau.pptx
The Accursed House by Émile Gaboriau.pptxThe Accursed House by Émile Gaboriau.pptx
The Accursed House by Émile Gaboriau.pptx
DhatriParmar
 
How to Make a Field invisible in Odoo 17
How to Make a Field invisible in Odoo 17How to Make a Field invisible in Odoo 17
How to Make a Field invisible in Odoo 17
Celine George
 
Thesis Statement for students diagnonsed withADHD.ppt
Thesis Statement for students diagnonsed withADHD.pptThesis Statement for students diagnonsed withADHD.ppt
Thesis Statement for students diagnonsed withADHD.ppt
EverAndrsGuerraGuerr
 
1.4 modern child centered education - mahatma gandhi-2.pptx
1.4 modern child centered education - mahatma gandhi-2.pptx1.4 modern child centered education - mahatma gandhi-2.pptx
1.4 modern child centered education - mahatma gandhi-2.pptx
JosvitaDsouza2
 
Instructions for Submissions thorugh G- Classroom.pptx
Instructions for Submissions thorugh G- Classroom.pptxInstructions for Submissions thorugh G- Classroom.pptx
Instructions for Submissions thorugh G- Classroom.pptx
Jheel Barad
 

Recently uploaded (20)

Welcome to TechSoup New Member Orientation and Q&A (May 2024).pdf
Welcome to TechSoup   New Member Orientation and Q&A (May 2024).pdfWelcome to TechSoup   New Member Orientation and Q&A (May 2024).pdf
Welcome to TechSoup New Member Orientation and Q&A (May 2024).pdf
 
Overview on Edible Vaccine: Pros & Cons with Mechanism
Overview on Edible Vaccine: Pros & Cons with MechanismOverview on Edible Vaccine: Pros & Cons with Mechanism
Overview on Edible Vaccine: Pros & Cons with Mechanism
 
CACJapan - GROUP Presentation 1- Wk 4.pdf
CACJapan - GROUP Presentation 1- Wk 4.pdfCACJapan - GROUP Presentation 1- Wk 4.pdf
CACJapan - GROUP Presentation 1- Wk 4.pdf
 
How libraries can support authors with open access requirements for UKRI fund...
How libraries can support authors with open access requirements for UKRI fund...How libraries can support authors with open access requirements for UKRI fund...
How libraries can support authors with open access requirements for UKRI fund...
 
Guidance_and_Counselling.pdf B.Ed. 4th Semester
Guidance_and_Counselling.pdf B.Ed. 4th SemesterGuidance_and_Counselling.pdf B.Ed. 4th Semester
Guidance_and_Counselling.pdf B.Ed. 4th Semester
 
Adversarial Attention Modeling for Multi-dimensional Emotion Regression.pdf
Adversarial Attention Modeling for Multi-dimensional Emotion Regression.pdfAdversarial Attention Modeling for Multi-dimensional Emotion Regression.pdf
Adversarial Attention Modeling for Multi-dimensional Emotion Regression.pdf
 
Home assignment II on Spectroscopy 2024 Answers.pdf
Home assignment II on Spectroscopy 2024 Answers.pdfHome assignment II on Spectroscopy 2024 Answers.pdf
Home assignment II on Spectroscopy 2024 Answers.pdf
 
special B.ed 2nd year old paper_20240531.pdf
special B.ed 2nd year old paper_20240531.pdfspecial B.ed 2nd year old paper_20240531.pdf
special B.ed 2nd year old paper_20240531.pdf
 
Palestine last event orientationfvgnh .pptx
Palestine last event orientationfvgnh .pptxPalestine last event orientationfvgnh .pptx
Palestine last event orientationfvgnh .pptx
 
Synthetic Fiber Construction in lab .pptx
Synthetic Fiber Construction in lab .pptxSynthetic Fiber Construction in lab .pptx
Synthetic Fiber Construction in lab .pptx
 
Unit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdfUnit 2- Research Aptitude (UGC NET Paper I).pdf
Unit 2- Research Aptitude (UGC NET Paper I).pdf
 
Honest Reviews of Tim Han LMA Course Program.pptx
Honest Reviews of Tim Han LMA Course Program.pptxHonest Reviews of Tim Han LMA Course Program.pptx
Honest Reviews of Tim Han LMA Course Program.pptx
 
Unit 8 - Information and Communication Technology (Paper I).pdf
Unit 8 - Information and Communication Technology (Paper I).pdfUnit 8 - Information and Communication Technology (Paper I).pdf
Unit 8 - Information and Communication Technology (Paper I).pdf
 
The approach at University of Liverpool.pptx
The approach at University of Liverpool.pptxThe approach at University of Liverpool.pptx
The approach at University of Liverpool.pptx
 
Additional Benefits for Employee Website.pdf
Additional Benefits for Employee Website.pdfAdditional Benefits for Employee Website.pdf
Additional Benefits for Employee Website.pdf
 
The Accursed House by Émile Gaboriau.pptx
The Accursed House by Émile Gaboriau.pptxThe Accursed House by Émile Gaboriau.pptx
The Accursed House by Émile Gaboriau.pptx
 
How to Make a Field invisible in Odoo 17
How to Make a Field invisible in Odoo 17How to Make a Field invisible in Odoo 17
How to Make a Field invisible in Odoo 17
 
Thesis Statement for students diagnonsed withADHD.ppt
Thesis Statement for students diagnonsed withADHD.pptThesis Statement for students diagnonsed withADHD.ppt
Thesis Statement for students diagnonsed withADHD.ppt
 
1.4 modern child centered education - mahatma gandhi-2.pptx
1.4 modern child centered education - mahatma gandhi-2.pptx1.4 modern child centered education - mahatma gandhi-2.pptx
1.4 modern child centered education - mahatma gandhi-2.pptx
 
Instructions for Submissions thorugh G- Classroom.pptx
Instructions for Submissions thorugh G- Classroom.pptxInstructions for Submissions thorugh G- Classroom.pptx
Instructions for Submissions thorugh G- Classroom.pptx
 

Aerb guidelines for Room installation

  • 1.
  • 2.  As per Atomic Energy (Radiation Protection) Rules 2004 Section (3) a License / Registration is necessary for establishing a radiation installation.  This process involves 3 steps.  (i) Site Approval by AERB  (ii) Quality Assurance Test by manufacturer/supplier under intimation to DRS (iii) License/Registration by AERB .
  • 3. Guideline 1  Submit completely filled and duly signed “Application for Site Approval of X-Ray Installation Layout” along with layout plan of the diagnostic x-ray room, drawn to scale 1:50 showing details as described in the application form, in original with two copies for approval to the Director, Directorate of Radiation Safety,AERB , Niyamak Bhavan, Anusaktinagar , Mumbai-400094.
  • 4.
  • 5.  Separate form should be submitted for each medical diagnostic x-ray unit installation. Mobile units do not require plan approval (mobile units should not be used in a room with table) but need Registration Certificate.  Enclose fees in the form of Demand Draft drawn in favor of the Director of Radiation Safety payable at AERB, at the rates prescribed by the Government.
  • 6. Guideline 2  The room housing an x-ray unit shall not be less than 18m2 for general-purpose radiography and conventional fluoroscopy equipment and no single dimension of the x-ray room shall be less than 4.0m.  The room housing the gantry of the computed tomography (CT Scan) unit shall not be less than 25 m 2 and no single dimension of the gantry of CT Scan room shall be less than 4m.
  • 7.
  • 8.  The room housing an orthopan tomography (OPG) units shall not be less that 15 m2 and no single dimension of the OPG room shall be less than 3.5m.  The room housing a mammography x-ray unit shall not be less than 10 m2 and no single dimension of x-ray room shall be less than 3m. Also, not more than one unit of any type shall be installed in the same room.
  • 9.
  • 10. Cont  The minimum wall thickness of the x-ray room should be 9 inch or 23cm solid brick or equivalent.  There shall be no windows and ventilations shall be above 2m from the floor outside. Table shall be oriented such that the primary (chest stand) is not directed towards the main door or darkroom.  The primary wall thickness shall be 35 cm of solid brick or 32 cm solid brick + 1.5 mm LEAD
  • 11. Guideline 3  The users should install only Atomic Energy Regulatory Board (AERB) “Type Approved” unit or unit with valid NOC for medical diagnostic purpose and obtain a copy of the AERB “Type Approval” certificate from the supplier/manufacturer of the x-ray unit prior to the procurement of the unit.  User may verify the Type Approval Certificate Number from AERB in case of any doubt.
  • 12.
  • 13. The users should procure and use protective accessories such as mobile protective barrier, lead rubber aprons, lead rubber flaps, gloves, thyroid shields, etc.
  • 14.
  • 15.  The employer shall provide Personnel Monitoring Badges to all radiation workers of the institute. For procuring personnel monitoring badges, please contact Avanttec Laboratories Private Limited, #31, Kamaraj Street, Srinivasa Nagar, Padi, Chennai-600 050.
  • 16.
  • 17.  The users should follow the guidelines given in AERB Safety Code for Medical Diagnostic X-Ray Equipment and Installations [Code No.AERB/SC/MED-2 (Rev.-1)].  The same can be purchased from AERB or downloaded from AERB website www.aerb.gov.in.  The user shall comply with other regulatory documents related to the use and operation of x-ray units that are available in AERB website or DRS website.
  • 18.  i Atomic Energy (Radiation Protection) Rules, 2004 (G.S.R.303) ii Radiation Surveillance Procedures for Medical Applications of Radiation, 1989 (G.S.R.No.388) iii Performa for “Registration of Diagnostic X-Ray installations and Licence of Computed Tomography (CT)/Cathlab unit Installations”. iv Documents (i) & (ii) can be purchased from bookstalls selling government publications or can be downloaded from DRS website. (iii) shall be given by the Directorate of Radiation Safety on Site Approval
  • 19.  The users should request the supplier/manufacturer of the diagnostic x-ray units to conduct Quality Assurance (QA) tests after the installation of x-ray unit/computed tomography unit/mammography/ cathlab unit prior to using it for patient diagnosis, under intimation to the DRS.  Preserve the QA tests report in the hospital/institution record for further use and submit a copy of QA tests report to the DRS. The unit should be used for patient diagnosis only if its performance is found satisfactory.
  • 20.
  • 21.  After installation, the users should send the completely filled and duly signed “Application for Registration of Diagnostic X- Ray installations” / “Application for Licence of Computed Tomography (CT) unit / Interventional Radiology (Cathlab) Installations” (as applicable) to the Director, Directorate of Radiation Safety, AERB for onward transmission with recommendation to the Atomic Energy Regulatory Board for obtaining Registration Certificate for diagnostic x-ray units and/or License Certificate for CT Scan/Cathlab units for operation.
  • 22. The employer shall employ only qualified Radiologist and Medical imaging Technologist. The minimum qualification is given in AERB Safety Code for Medical Diagnostic X-Ray Equipment and Installations [Code No.AERB/SC/MED- 2(Rev-1)]
  • 23.
  • 24. AERB issues notice every year in all leading newspapers in the country (both Hindi and English) regarding “Statutory Requirements for Safe Operation of Medical diagnostic X-ray units by Hospitals, Clinics and other Medical Institutions in India”.
  • 25.
  • 26. Design Certificate. Apporoval type/ NOC Layout approval and registration. Certification of RSO/ Service engineers
  • 27.
  • 28. e-LORA is a web based I&CT application implemented by AERB to facilitate licensing of radiation facilities. e-governance initiative by AERB. Objective: Enhanced efficiency and transparency in the regulatory processes of AERB. Aimed at achieving paperless licensing.
  • 29.
  • 30. Step : 1 creation of account in e-LORAfor user D & password. Step 2 : Approval letter. Step 3: QA report for applying license. Step 4 : TLD badges for doctor, operator. Step 5 : Qualified Technicians. Step 6: Approved RSO. Step 7 : Radiation License certificate.