This document provides competency assessment guidelines for staff working in a Radioimmunoassay (RIA) Laboratory. It outlines 11 key areas of training including radiation safety, quality control procedures, assay protocols, equipment operation and maintenance. It also describes the 5 main assays performed - aldosterone, 17-hydroxyprogesterone, human growth hormone, active renin, and 11-deoxycortisol. Staff are expected to demonstrate proficiency in all laboratory standard operating procedures (SOPs) and understanding of RIA principles, techniques, calculations and quality assurance.
Electron microscope, principle and applicationKAUSHAL SAHU
Introduction
History
Resolution &Magnification of
Electron microscope
Types of electron microscope
1) Transmission electron microscope (TEM)
- Structural parts of TEM
- Principle & Working of TEM
- Sample preparation for TEM
- Advantages & disadvantages of TEM
Scanning electron microscope (SEM)
- Structural parts of SEM
- Principle & Working of SEM
- Sample preparation for SEM
- Advantages & disadvantages of SEM
3) Scanning transmission electron microscope (STEM)
Applications of electron microscope
Conclusion
References
Sepration of molecules on the basis of applied Electric Field
Categorized into 1) Zone Electrophoresis 2) Moving Boundary Electrophoresis
We can seprate macromolecules (DNA , RNA, PROTEINS )on the basis of their charge, size shape & molecular weight
Electron microscope, principle and applicationKAUSHAL SAHU
Introduction
History
Resolution &Magnification of
Electron microscope
Types of electron microscope
1) Transmission electron microscope (TEM)
- Structural parts of TEM
- Principle & Working of TEM
- Sample preparation for TEM
- Advantages & disadvantages of TEM
Scanning electron microscope (SEM)
- Structural parts of SEM
- Principle & Working of SEM
- Sample preparation for SEM
- Advantages & disadvantages of SEM
3) Scanning transmission electron microscope (STEM)
Applications of electron microscope
Conclusion
References
Sepration of molecules on the basis of applied Electric Field
Categorized into 1) Zone Electrophoresis 2) Moving Boundary Electrophoresis
We can seprate macromolecules (DNA , RNA, PROTEINS )on the basis of their charge, size shape & molecular weight
The following presentation contains helpful information regarding Radioimmunoassay (RIA) and Enzyme-Linked Immunosorbent Assay (ELISA), including their history, introduction, advantages, procedures and applications.
A scintillation counter is an instrument for detecting and measuring ionizing radiation by using the excitation effect of incident radiation on a scintillating material, and detecting the resultant light pulses or it can be defined as it is used to detect gamma rays and the presence of a particle. It can also measure the radiation in the scintillating medium, the energy loss, or the energy gain. The medium can be solid and liquid.
The phenomenon in which the nucleus of the atom of an element undergoes spontaneous and uncontrollable disintegration or decay and emit alpha, beta, or gamma rays
It is the property of some unstable atoms to spontaneously emit nuclear radiation to gain stability.
The heavy elements are called radioactive elements and rays emitted these elements are called radioactive rays.
The phenomenon of radioactivity is discovered by HENRI BACQUEREL IN 1896.
Radioactive decay (also known as nuclear decay, radioactivity, radioactive disintegration, or nuclear disintegration) is the process by which an unstable atomic nucleus loses energy by radiation. A material containing unstable nuclei is considered radioactive. Three of the most common types of decay are alpha decay (α-decay), beta decay (β-decay), and gamma decay (γ-decay), all of which involve emitting one or more particles.
A Presentation on Solid and Liquid ScintillationAnshdhaNANDRA1
Solid and liquid scintillation are fundamental techniques in radiation detection, vital across scientific, medical, and industrial domains. Solid scintillation utilizes materials such as crystals or plastics doped with scintillating compounds. When ionizing radiation interacts with these materials, they emit photons, producing flashes of light proportional to the radiation's energy. Photomultiplier tubes or photodiodes then detect and amplify these light signals for analysis, enabling precise measurement of radiation levels and identification of radioactive isotopes.
Liquid scintillation, meanwhile, involves dissolving radioactive samples in organic solvents containing scintillating molecules. Radiation interactions within the liquid generate photons, which are similarly detected and analyzed for radiation quantification and isotopic identification. Liquid scintillation is particularly useful in radiometric dating, environmental monitoring, and biochemical assays due to its versatility and sensitivity.
Both solid and liquid scintillation techniques offer unparalleled sensitivity and efficiency in detecting various types of radiation, from alpha and beta particles to gamma rays. Their widespread application continues to drive advancements in radiation detection, supporting diverse fields from nuclear physics to medical imaging and beyond.
An isotope is one of two or more atoms having the same atomic number but different mass numbers.
Unstable isotopes are called Radioisotopes.
uses of radioisotopes are many which are discussed in this slide.
Radiopharmaceutical is topic of subject Pharmaceutical inorganic Chemistry for B. Pharmacy First year students. This slide is presented with an aim to enable the students to easily understand and grasp unfamiliar concept of this topic
Thermoluminescence in Pure and Nd+3 + K+ Doped Lead Germanate Single CrystalsAI Publications
Pure lead germanate was grown by Zochralski’stechnique.Nd+3 + K+ doped lead germanate was grown for comparative studies. Thermoluminescence studies were conducted on pure lead germanate and Nd+3 + K+ doped lead germanate. Thermoluminescence in pure lead germanate single crystals (LG) and Nd+3 + K+ doped lead germanate single crystals (DDLG) has been measured in terms of photomultiplier output current on Thermoluminescence Set Up. Thermoluminescence studies of LG revealed an increasing trend of Thermoluminescence output, with increasing temperature and sharp consistent maxima at 50°C and 100°C. TL studies of DDLG revealed a decreasing trend of Thermoluminescence output, with increasing temperature and a sharp consistent peak at 100°C and 120°C. It was further noted that doping increased Thermoluminescence output. Thermoluminescence in ferroelectric lead germanate was attributed to change in polarization with change in temperature and creation of defect states.
Different Types of Radioactive Counters or detectors used in analyzing low or high penetrating power radiation or particles are explained briefly with their advantages and disadvantages.
Medha Thakur (M.Sc Chemistry)
radiopharmaceuticals introduction isotopes types of radioisotopes measurement of radioactivity handling and storage of radioactive material applications
general introduction of radioactivity, it include discovery of radioactivity, types of radiation, isotopes and radioactive isotopes difference, half life, prevention and precaution from radiation. detecting devices used in laboreatory for radiation spillage and protection.
UiPath Test Automation using UiPath Test Suite series, part 4DianaGray10
Welcome to UiPath Test Automation using UiPath Test Suite series part 4. In this session, we will cover Test Manager overview along with SAP heatmap.
The UiPath Test Manager overview with SAP heatmap webinar offers a concise yet comprehensive exploration of the role of a Test Manager within SAP environments, coupled with the utilization of heatmaps for effective testing strategies.
Participants will gain insights into the responsibilities, challenges, and best practices associated with test management in SAP projects. Additionally, the webinar delves into the significance of heatmaps as a visual aid for identifying testing priorities, areas of risk, and resource allocation within SAP landscapes. Through this session, attendees can expect to enhance their understanding of test management principles while learning practical approaches to optimize testing processes in SAP environments using heatmap visualization techniques
What will you get from this session?
1. Insights into SAP testing best practices
2. Heatmap utilization for testing
3. Optimization of testing processes
4. Demo
Topics covered:
Execution from the test manager
Orchestrator execution result
Defect reporting
SAP heatmap example with demo
Speaker:
Deepak Rai, Automation Practice Lead, Boundaryless Group and UiPath MVP
Generative AI Deep Dive: Advancing from Proof of Concept to ProductionAggregage
Join Maher Hanafi, VP of Engineering at Betterworks, in this new session where he'll share a practical framework to transform Gen AI prototypes into impactful products! He'll delve into the complexities of data collection and management, model selection and optimization, and ensuring security, scalability, and responsible use.
Essentials of Automations: The Art of Triggers and Actions in FMESafe Software
In this second installment of our Essentials of Automations webinar series, we’ll explore the landscape of triggers and actions, guiding you through the nuances of authoring and adapting workspaces for seamless automations. Gain an understanding of the full spectrum of triggers and actions available in FME, empowering you to enhance your workspaces for efficient automation.
We’ll kick things off by showcasing the most commonly used event-based triggers, introducing you to various automation workflows like manual triggers, schedules, directory watchers, and more. Plus, see how these elements play out in real scenarios.
Whether you’re tweaking your current setup or building from the ground up, this session will arm you with the tools and insights needed to transform your FME usage into a powerhouse of productivity. Join us to discover effective strategies that simplify complex processes, enhancing your productivity and transforming your data management practices with FME. Let’s turn complexity into clarity and make your workspaces work wonders!
zkStudyClub - Reef: Fast Succinct Non-Interactive Zero-Knowledge Regex ProofsAlex Pruden
This paper presents Reef, a system for generating publicly verifiable succinct non-interactive zero-knowledge proofs that a committed document matches or does not match a regular expression. We describe applications such as proving the strength of passwords, the provenance of email despite redactions, the validity of oblivious DNS queries, and the existence of mutations in DNA. Reef supports the Perl Compatible Regular Expression syntax, including wildcards, alternation, ranges, capture groups, Kleene star, negations, and lookarounds. Reef introduces a new type of automata, Skipping Alternating Finite Automata (SAFA), that skips irrelevant parts of a document when producing proofs without undermining soundness, and instantiates SAFA with a lookup argument. Our experimental evaluation confirms that Reef can generate proofs for documents with 32M characters; the proofs are small and cheap to verify (under a second).
Paper: https://eprint.iacr.org/2023/1886
PHP Frameworks: I want to break free (IPC Berlin 2024)Ralf Eggert
In this presentation, we examine the challenges and limitations of relying too heavily on PHP frameworks in web development. We discuss the history of PHP and its frameworks to understand how this dependence has evolved. The focus will be on providing concrete tips and strategies to reduce reliance on these frameworks, based on real-world examples and practical considerations. The goal is to equip developers with the skills and knowledge to create more flexible and future-proof web applications. We'll explore the importance of maintaining autonomy in a rapidly changing tech landscape and how to make informed decisions in PHP development.
This talk is aimed at encouraging a more independent approach to using PHP frameworks, moving towards a more flexible and future-proof approach to PHP development.
Enhancing Performance with Globus and the Science DMZGlobus
ESnet has led the way in helping national facilities—and many other institutions in the research community—configure Science DMZs and troubleshoot network issues to maximize data transfer performance. In this talk we will present a summary of approaches and tips for getting the most out of your network infrastructure using Globus Connect Server.
A tale of scale & speed: How the US Navy is enabling software delivery from l...sonjaschweigert1
Rapid and secure feature delivery is a goal across every application team and every branch of the DoD. The Navy’s DevSecOps platform, Party Barge, has achieved:
- Reduction in onboarding time from 5 weeks to 1 day
- Improved developer experience and productivity through actionable findings and reduction of false positives
- Maintenance of superior security standards and inherent policy enforcement with Authorization to Operate (ATO)
Development teams can ship efficiently and ensure applications are cyber ready for Navy Authorizing Officials (AOs). In this webinar, Sigma Defense and Anchore will give attendees a look behind the scenes and demo secure pipeline automation and security artifacts that speed up application ATO and time to production.
We will cover:
- How to remove silos in DevSecOps
- How to build efficient development pipeline roles and component templates
- How to deliver security artifacts that matter for ATO’s (SBOMs, vulnerability reports, and policy evidence)
- How to streamline operations with automated policy checks on container images
Securing your Kubernetes cluster_ a step-by-step guide to success !KatiaHIMEUR1
Today, after several years of existence, an extremely active community and an ultra-dynamic ecosystem, Kubernetes has established itself as the de facto standard in container orchestration. Thanks to a wide range of managed services, it has never been so easy to set up a ready-to-use Kubernetes cluster.
However, this ease of use means that the subject of security in Kubernetes is often left for later, or even neglected. This exposes companies to significant risks.
In this talk, I'll show you step-by-step how to secure your Kubernetes cluster for greater peace of mind and reliability.
Le nuove frontiere dell'AI nell'RPA con UiPath Autopilot™UiPathCommunity
In questo evento online gratuito, organizzato dalla Community Italiana di UiPath, potrai esplorare le nuove funzionalità di Autopilot, il tool che integra l'Intelligenza Artificiale nei processi di sviluppo e utilizzo delle Automazioni.
📕 Vedremo insieme alcuni esempi dell'utilizzo di Autopilot in diversi tool della Suite UiPath:
Autopilot per Studio Web
Autopilot per Studio
Autopilot per Apps
Clipboard AI
GenAI applicata alla Document Understanding
👨🏫👨💻 Speakers:
Stefano Negro, UiPath MVPx3, RPA Tech Lead @ BSP Consultant
Flavio Martinelli, UiPath MVP 2023, Technical Account Manager @UiPath
Andrei Tasca, RPA Solutions Team Lead @NTT Data
The Art of the Pitch: WordPress Relationships and SalesLaura Byrne
Clients don’t know what they don’t know. What web solutions are right for them? How does WordPress come into the picture? How do you make sure you understand scope and timeline? What do you do if sometime changes?
All these questions and more will be explored as we talk about matching clients’ needs with what your agency offers without pulling teeth or pulling your hair out. Practical tips, and strategies for successful relationship building that leads to closing the deal.
Welcome to the first live UiPath Community Day Dubai! Join us for this unique occasion to meet our local and global UiPath Community and leaders. You will get a full view of the MEA region's automation landscape and the AI Powered automation technology capabilities of UiPath. Also, hosted by our local partners Marc Ellis, you will enjoy a half-day packed with industry insights and automation peers networking.
📕 Curious on our agenda? Wait no more!
10:00 Welcome note - UiPath Community in Dubai
Lovely Sinha, UiPath Community Chapter Leader, UiPath MVPx3, Hyper-automation Consultant, First Abu Dhabi Bank
10:20 A UiPath cross-region MEA overview
Ashraf El Zarka, VP and Managing Director MEA, UiPath
10:35: Customer Success Journey
Deepthi Deepak, Head of Intelligent Automation CoE, First Abu Dhabi Bank
11:15 The UiPath approach to GenAI with our three principles: improve accuracy, supercharge productivity, and automate more
Boris Krumrey, Global VP, Automation Innovation, UiPath
12:15 To discover how Marc Ellis leverages tech-driven solutions in recruitment and managed services.
Brendan Lingam, Director of Sales and Business Development, Marc Ellis
1. RADIOIMMUNOASSAY LABORATORY:
COMPETENCY ASSESSMENT 2008
DIVISION OF CHEMICAL PATHOLOGY
GROOTE SCHUUR HOSPITAL
C17 NHLS
Author: David Haarburger
Supervisor: Judy King
February 2008
This document is intended for the use of medical technologists, registrars and scientists, and
forms part of the training required when using the RIA lab.
2. RADIOIMMUNOASSAY LABORATORY: COMPETENCY ASSESSMENT 2008
CHEMICAL PATHOLOGY C17 NHLS
1. Radiation Safety/Radiation Protection Training Course
See SOP 1
2. RIA Spillage Monitoring
See SOP 2
3. RIA Gamma Counter Operation/Assay Protocols
See SOP 3
4. RIA Gamma Counter Calibration and Maintenance
See SOP 4
5. RIA Internal Quality Control/Assay Parameters
See SOP 5
6. Radioactive Waste Disposal
See SOP 7
7. RIA Package Inserts
See SOP 8
8. RIA Turnaround Times
See SOP 9
9. RIA Specimen and Kit Storage
See SOP 10
10. RIA Procurement
See SOP 11
11. Describe briefly the assays performed routinely in the RIA Laboratory. Include:
clinical background, type of assay, sample collection, brief methodology, rationale
for each step, sensitivity, specificity, intra- and inter-assay coefficient of variation,
reference range.
Five assays, radioimmuno- (RIA) and immuno-radiometric assays (IRMAs), are routinely
run in this RIA lab, all employing 125I-labeled material. Four (Aldosterone, 17-OH-
Progesterone, Human Growth Hormone and Active Renin) are batched and run bi-
weekly, while one (11-Desoxycortisol) is processed quarterly. Internal quality control is
managed within the RIA lab, with 3 levels of Bio-Rad lyphochek samples analysed in
every run of all assays (except Active Renin – controls come with kit).
Details can be found in the kit package inserts.
12. What types of particles are emitted during radioactive decay?
There are three types of decay. In alpha decay, the nucleus emits an alpha particle
( ); in beta decay, the nucleus emits an electron (or positron), and in gamma decay,
the nucleus emits gamma particles (high-energy photons).
He4
2
13. List the radioisotopes commonly used in biomedical work. Draw up a table
showing the half-life and type of particle emitted for each isotope.
Radioisotope Half-life Decay type Decay equation
3
H 12,3y β-
e0
1
3
2
3
1 HeH −+⎯→⎯
14
C 5730y β-
e0
1
14
7
14
6 NC −+⎯→⎯
32
P 14,3d β-
e0
1
32
16
32
15 SP −+⎯→⎯
35
S 87d β-
e0
1
35
17
35
16 ClS −+⎯→⎯
51
Cr 27,7d EC, γ γ+⎯→⎯+− VCr 51
23
0
1
51
24 e
57
Co 272d EC, γ γ+⎯→⎯+− FeCo 57
26
0
1
57
27 e
58
Co 71d EC, β+
, γ γ+⎯→⎯+− FeCo 58
26
0
1
58
27 e
3. γ++⎯→⎯ p0
1
58
26
58
27 FeCo
59
Fe 45d β-
, γ γ++⎯→⎯ − e0
1
59
27
59
26 CoFe
99
Mo 66h β-
, γ γ++⎯→⎯ − e0
1
99
43
99
42 TcMo
99m
Tc 6,0h γ γ+⎯→⎯ TcTc 99
43
m99
43
125
I 60h EC, γ γ+⎯→⎯+− TeI 125
52
0
1
125
53 e
131
I 8,04d β, γ γ++⎯→⎯ − e0
1
131
54
131
53 XeI
14. How is radioactivity measured? What does specific activity mean?
Autoradiography, gas ionization detectors and fluorescent scintillation can be used to
measure radiation.
Autoradiography
Autoradiography is a procedure for localizing and recording a radiolabeled compound
within a solid sample, which involves the production of an image in a photographic
emulsion. The solid sample often consists of size-fractionated DNA or protein samples
that are embedded within a dried gel, fixed to the surface of a dried nylon membrane or
nitrocellulose filter, or located within fixed chromatin or tissue samples mounted on a
glass slide. The photographic material consists of an emulsion layer sandwiched between
two gelatin layers. One provides adhesion and the other protection. The photosensitive
emulsion layer contains minute crystals of silver halides ranging from 0,07 to 0,40μm in
diameter suspended in gelatin. Following passage through the emulsion of a β-particle or
a γ-ray emitted by a radionuclide, the Ag+
ions are converted to Ag atoms. This process
repeats until a metallic silver grain of increasing size and stability is formed resulting in a
latent image. During development, the silver halide is reduced to metallic silver but the
process proceeds faster in crystals with latent image silver, hence amplifying the image.
Fixing is then done to remove any unexposed silver halide crystals, giving an
autoradiographic image which provides a two-dimensional representation of the
distribution of the radiolabel in the original sample.
Gas ionization detectors
This is the most common type of instrument. This instrument works on the principle that
as radiation passes through air or a specific gas, it gives off energy to orbital electrons,
causing ionization and excitation of the gas atoms. When a high voltage is placed
between two areas of the gas filled space, the positive ions will be attracted to the
cathode of the detector and the free electrons will travel to the anode. These charges are
collected by the anode and cathode which then form a very small current in the wires
going to the detector. By placing a very sensitive current measuring device between the
wires from the cathode and anode, the small current is measured and displayed as a
signal. The more radiation which enters the chamber, the more current displayed by the
instrument. Many types of gas-filled detectors exist, but the two most common are the
ion chamber used for measuring large amounts of radiation and the Geiger-Muller
detector used to measure very small amounts of radiation.
Fluorescent scintillation
In the scintillation process, the absorbed energy produces a flash of light. When a
particle passes through the material it collides with atomic electrons, exciting them to
higher energy levels. After a very short period of time the electrons fall back to their
natural levels, causing emission of light. Two common scintillation detectors are the
sodium iodide crystal scintillation detector (γ-counter) and the organic liquid scintillation
detector (β-counter).
The crystal scintillation detector commonly occurs as a well detector which has a hole
drilled in the end of the cylindrical crystal to accept a test tube. Because it is hygroscopic,
the crystal is sealed in an aluminium can with a transparent quartz window at one end
4. through which the blue-violet (420nm) scintillations are detected. The photos of gamma
emitters in the sample easily penetrate the specimen tube and the thin, low-density can
and enter the crystal where they are absorbed in the thick, high density sodium iodide. A
well counter is not suitable for measuring β-radiation, because it can not penetrate the
sample container or aluminium lining of the wall.
The crystal is usually a circular cylinder machined from a single crystal of sodium iodide,
to which a small amount of thallium is added to improve performance. The high atomic
number of iodine and the high density of sodium iodide (3,7g/cm3
) favour the absorption
of γ-radiation. For this reason, a well counter is often referred to as a γ-counter. For a
typical well detector, the counting efficiency for 125
I is approximately 70%.
A liquid scintillation detector measures radioactivity by recording scintillations occurring
within a transparent vial that contains the unknown sample and liquid scintillator.
Because the radionuclide is mixed with the liquid scintillator the technique is ideal for pure
β-emitters. Counting efficiencies range from between 60% for 3
H to 90% for 14
C. The
liquid scintillator is known as the scintillation cocktail and contains two components, the
primary solvent and the primary scintillator. The primary solvent is usually inexpensive
and is chosen for its efficiency in absorbing and transferring radiation energy. It is usually
one of the aromatic hydrocarbons: toluene, or xylene. The primary scintillator absorbs
energy from the primary solvent and converts it into light. A common primary scintillator
is 2,5-diphenyl oxazole which emits ultraviolet light of 380nm. Other components may be
added to the liquid scintillator such as a secondary solvent to improve solubility or a
surfactant to stabilise or emulsify the sample. A secondary scintillator may be added to
absorb the ultraviolet photons of the primary scintillator and reemit the energy at a longer
wavelength.
Specific activity refers to the radioactivity per unit mass or unit volume of a substance.
The maximum specific activity attainable for each radionuclide is that for the pure
radionuclide. However, usually the pure radionuclide is unavailable and only makes up a
small fraction of the substance it represents.
15. Units of radioactivity: what are Ci (Curie), cpm, dpm, Sievert, and Becquerels?
The Becquerel (Bq) is the SI unit of activity, defined as one decay per second (dps). The
Curie was originally defined as the radioactivity of one gram of pure radium, but has been
redefined as exactly 37GBq.
However, not all decays are capable of affecting the scintillator and being recorded.
Some photons do not reach the scintillator or the detector, and those that do, may not
interact with it. The number of decays detected by the detector is called counts and they
are related by the equation: Counting efficiency = Count Rate / Decay Rate. The counts
are usually expressed as counts per minute.
Radiation carries energy, and when it is absorbed by matter, the matter receives this
energy. The radiation dose is the amount of energy deposited per unit of mass. The SI
unit of radiation dose is the Gray (Gy), which is defined as the dose of one joule of energy
absorbed per kilogram of matter.
Various kinds of radiation have different effects on living tissue, so a simple measurement
of dose as energy received, stated in grays, does not give a clear indication of the
probable biological effects of the radiation. The equivalent dose, which is measured in
sieverts, is equal to the actual dose, in grays, multiplied by a ‘quality factor’ which is larger
for more dangerous forms of radiation. An effective dose of one sievert requires 1 gray of
beta or gamma radiation but only 0,05 gray of alpha radiation or 0,1 gray of neutron
radiation.
16. How much radioactivity is commonly used in a RIA (say, 50 samples)? What dose
of I-131 is usually given to patients with Grave’s disease?
In RIAs I-125 is commonly used. One vial of labelled I-125 ligand may have a total
radioactivity of 130kBq, which can be used for 50 samples. To treat a patient with Grave’s
disease, doses between 370MBq and 550MBq of I-131 are given.
5. 17. Describe and illustrate the principles of: Competitive RIA, Double antibody
(sandwich type) IRMA, ELISA.
In a competitive radioimmunoassay, an antibody and a radiolabelled antigen are used to
measure an analyte. The analyte in the patient’s serum is mixed with the radiolabelled
analyte and allowed to compete for a limited amount of the antibody for a fixed time
period. The unbound analyte is then removed in a wash step, and the bound antibody-
analyte is then measured, usually in a gamma counter. The concentration of the analyte
is then determined from a (decreasing sigmoid) calibration curve.
In an immunoradiometric assay, the analyte is incubated with a solid-phase antibody and
a second radiolabelled antibody. An excess of both antibodies should always be present.
This is left for a fixed time period, after which all unbound radiolabelled antibody is
removed in a wash step. The antibody-analyte-radiolablled antibody complex is then
measured with a gamma counter. The concentration of the analyte is then determined
from a (linear) calibration curve.
An enzyme-linked immunosorbent assay is similar to an immunoradiometric assay except
that the second antibody is not radiolabelled but instead is covalently bound to an
enzyme. The two antibodies (fixed and enzyme-bound) and the sample are allowed to
incubate, after which the unbound antibody is washed off. A substrate of the enzyme is
then added which is converted to a detectable product. The product may be a coloured-
dye, a fluorescent product, or a chemical that undergoes chemiluminescence. The
amount of product is measured, and the concentration of the analyte can then be
determined from a (linear) calibration curve.
18. What factors determine the stability of hormones in plasma? Discuss the specimen
handling precautions which may be necessary. What are the functions of trasylol
and EDTA? What is cryoactivation, and why does it affect the Active Renin IRMA?
The stability of hormones in samples is determined by the type of hormone, the anti-
coagulant used and the storage temperature. As a general rule, steroid hormones are
more stable than peptide hormones, and storing at colder temperatures is better than
warmer temperatures. Exceptions to this are aldosterone, a steroid which degrades
quickly at room temperature, and C-peptide which is stable for more than 5 days
throughout a range of laboratory-used temperatures. EDTA is generally considered the
best anti-coagulant to use as it has some anti-proteolytic properties. However, with the
exception of ACTH, this effect is usually non-significant.
Aprotinin (trasylol) is a polypeptide derived from bovine lung tissue that inhibits serine
proteases such as trypsin, chymotrypsin, plasmin and kallikrein. It is often used as an
anti-proteolytic when collecting blood for unstable peptide hormones such as ACTH or
glucagon.
Approximately tenfold more prorenin than renin normally circulates in human plasma, with
almost 100 times more being seen in some low-renin patients. At temperatures below
25°C, prorenin develops intrinsic renin activity, and the prosequence becomes vulnerable
to cleavage by plasma enzymes, resulting in irreversible formation of renin in vitro. This is
called cryoactivation. The lower the temperature (short of freezing) the more likely these
processes are to occur. Because renin is remarkably stable in plasma at room
temperature, to avoid cryoactivation of prorenin, blood samples for renin should be
processed at room temperature. Prorenin does not cryoactivate in frozen plasma, or
during rapid freezing and thawing. It is for this reason that plasma must be thawed in a
37°C water bath, and not thawed gently on a bench as for most other analytes. The
antibody used in the renin IRMA binds to active renin. If cryoactivation occurs, more of
the prorenin will be converted to active renin, and a falsely high renin value will be
obtained.
19. In a RIA, what is meant by the following: Total Counts, Zero Binding, Nonspecific
Binding, Percentage Bound (%B0), %NSB, ED50, ED20, ED80? What type of curve is
used?
6. Total counts - are tubes that represent the total amount of radioactivity added in an RIA
tube. These tubes are not decanted in the separation step. They represent the total
amount of tracer aliquoted per tube. When the assay is counted, these tubes will have the
highest CPMs. These counts are not used as part of the dose estimate calculation for
unknowns, but rather as a quality control comparison to the counts in the B0 tubes.
Because the amount of antibody is limiting and tracer is in excess, total count tubes are
included to guarantee and document this excess. The degree of excess is expressed as a
percent of CPMs in the B0 tubes divided by the CPMs in the total count tubes, often
referred to as the %B/T or Bound/Total. This %B/T value should be between thirty and
fifty percent.
Zero binding (B0) - are tubes that contain labelled antigen, the limiting antibody, possibly
assay buffer and the precipitant, but do not have any unlabeled antigen such as unknown
samples or standards (except zero standard). After separating the free from the bound
fraction, these tubes will have the highest CPMs, other than the total count tubes.
Non-specific binding - are tubes that contain labelled antigen, sometimes assay buffer,
zero standard or precipitant, but they never have any antibody. When the assay is
counted, these tubes will have the lowest CPMs in a radioimmunoassay system. These
counts serve as a record of binding which is not due to the antibody. For example, the
labelled antigen may bind to elements of the buffer or to the tube wall. (Generally, plastic
or polystyrene tubes absorb more label than glass tubes, which in turn, absorb more label
than polypropylene tubes.) These counts are subtracted from the counts of all the other
tubes to obtain a more accurate estimate of counts in the bound fraction.
Percentage bound of B0 – Data are generally expressed as standards and unknowns as a
percentage of the maximum possible bound (B0).
Percent non-specific binding (%NSB) – This is the non-specific binding count divided by
the total counts and tells you how much of the total radioactivity added, gets bound to
non-specific binding sites. Ideally this should be as low as possible.
Effective dose ED50, ED20, ED80 – This is the concentration of analyte that corresponds to
50% / 20% / 80% B/B0. These data are often derived from a plot of B/B0 against the
analyte concentration.
In RIA, data are often presented as a graph with the B/B0 on the y-axis and log of the
concentration on the x-axis. This gives a sigmoid shaped graph from where unknown
data points can be extrapolated. To make the graph linear, the y-axis is often replaced
with the logit B/B0 which makes extrapolating the data easier.
20. In an IRMA, what is meant by the following: Total Counts, Nonspecific Binding?
What type of curve is used?
Total counts - are tubes that represent the total amount of radioactivity added in an IRMA
tube. These tubes are not decanted in the separation step. They represent the total
amount of tracer aliquoted per tube. When the assay is counted, these tubes will have the
highest CPMs. These counts are not used as part of the dose estimate calculation for
unknowns, but are rather compared to the counts obtained in the highest standard tubes
as a means of quality control. Because the amount of analyte is limiting and tracer
(antibody) is in excess, total count tubes are included to guarantee and document this
excess.
Non-specific binding - are tubes that contain labelled antibody, sometimes assay buffer,
zero standard, but no analyte. When the assay is counted, these tubes will have the
lowest CPMs in the immunoradiometric assay system. These counts serve as a record of
antibody binding to sites which are not on the analyte. For example, the labelled antibody
may bind to proteins in the buffer or directly to the tube wall. These counts are subtracted
7. from the counts of all the other tubes to obtain a more accurate estimate of counts in the
bound fraction. This tube is equivalent to the zero standard counts.
In IRMA data analysis, either the counts per minute or the counts/total counts (B/T) is
plotted against the concentration. This should give a linear graph when plotted on either
linear or log-log paper.
21. How is cross-reactivity measured?
Cross-reacting substances are those substances which affect binding of antigen by
competing for the specific binding site on the antibody. The cross-reactivity of a
substance can be reported several ways. General guidelines on interference testing
recommend reporting interferences as the maximum effect expected from the interfering
substance at a specified concentration of the interfering substance at the medical
decision point of the analyte. This method has been accepted as the most useful way of
reporting interferences. However, for immunoassays, interferences are often reported as
percent cross-reactivity – defined as the mass ratio of analyte to interfering substance,
each at 50% displacement of label (ED50). Cross-reactivity can also be calculated at
other levels of displacement such as 20% or the ED20. Depending on the slope and shape
of the response curve the percent cross-reactivity may be different at different
displacement levels. Another method to report cross-reactivity may be to simply report
the concentration of cross-reactant required to displace a given amount of labelled
antigen. For example, one might report the concentration of cross-reactant required to
displace 50% of the label i.e. the ED50.
Cross-reactivity Cross-reactivity
Substance Cross-reactivity,
%
hCG added (IU/l) Apparent increase
in TSH
concentration
(IU/L)
hCG 5x10
-5
1000 <0,03
10 000 0,4
100 000 5
Two methods of reporting cross-reactivity
Example calculation of cross-reactivity of hCG in TSH assay. ED50 TSH =
10IU/l. ED50 hCG = 100IU/l. hCG cross-reactivity = 10 / 100 x 100% = 10%
8. 22. Describe the terms “sensitivity” and “specificity” in the context of immunoassays.
Sensitivity
The term sensitivity can have different meanings depending on the context. In its strictest
sense, the sensitivity is the change in the response of a measuring instrument divided by
the corresponding change in the stimulus, so for an IRMA, the sensitivity would be the
change in cpm divided by the change in analyte concentration. The sensitivity can also
refer to the detection limit of the assay and can be defined as that concentration of
antigen which can be distinguished from zero concentration with a stated degree of
probability. Sensitivity is affected by the titre, affinity, and specificity of the reference
antibody used in the assay. As such it can be affected by possible differences in antibody
affinity for unlabelled and labelled antigen, and the presence of cross-reactive antigens,
interfering substances or conditions in the test sample, as well as separation artefacts
generated by experimental technique.
Specificity
Specificity is a characteristic of a laboratory test which describes its ability to distinguish
between true (or specific) and non-specific results. With immunoassay methods,
interferences which affect specificity can be categorized into two major classes: 1) those
which affect the binding event between the antibody and an antigen in a general way,
such as pH or ionic strength; or 2) those substances which affect binding of antigen by
competing for the specific binding site on the antibody. These ‘specific’ interferences are
often referred to as ‘cross-reactants’. The specificity of an immunoassay may be
characterized by adding increasing amounts of a potential cross reacting substance to a
sample and measuring the response in the immunoassay. See previous question.
23. Describe the terms “accuracy” and “precision” in measuring hormone levels.
Accuracy is usually used to denote the ability of an assay system to generate the correct
result. It is defined as the closeness of agreement between the result of a test and its
true value. Unfortunately, the true value for any individual test result produced is usually
unknowable, so this definition of accuracy exists mainly as a theoretical concept.
Although accuracy and trueness are often used synonymously, there is a difference.
Accuracy strictly refers to the correctness of a single result whereas trueness refers to the
correctness of the mean of a number of results. This difference is important because a
result’s accuracy is influenced by bias and imprecision whereas trueness is only
influenced by bias. Practically, accuracy is estimated from a ‘comparison of methods’
experiment where the average difference between results by the method of interest and a
reference or comparative method is calculated. Other tests used to give an indication of
accuracy include recovery and dilution testing. Immunoassay manufacturers are required
to make a claim for the accuracy of their analytic measurement products and that claim
typically is based on results from a comparison of methods experiment. Also, laboratories
are required to verify a manufacturer’s claim for accuracy, which again would typically be
done by data from a ‘comparison of methods’ experiment.
Precision is defined as the closeness of agreement between independent test results
under prescribed conditions. The degree of precision is usually expressed on the basis of
statistical measures of impression such as the standard deviation or CV. The precision is
solely related to the random error of measurements and has no relation to the trueness of
measurements. Manufacturers are required to make a claim for precision and typically
provide estimates within a single run and for many runs performed over a period of one
month. Laboratories are required to verify the manufacturer’s claims, which again is
typically by performing 20 measurements within a single run and 20 measurements over
20 different runs over 20 different days. Again, both manufacturers and laboratories are
customers who have practical applications in characterizing and verifying this
performance characteristic.
24. For endocrine hormones, how do you convert IU to mass or moles? How do you
convert mass to moles? List the relevant conversion factors for the five routine
RIA/IRMA assays.
9. Wherever an anaylte has been chemically well defined and can be easily synthesised or
isolated, it is preferable to express its concentration in Système International (SI)
recommended units of moles per litre. However, conventionally in some parts of the
world, mass units are in common usage and concentrations are often expressed in grams
per litre. Where the analyte is well characterised, it is easy to convert from mass units to
mole units by dividing by the molecular weight (MW).
When an analyte can not be chemically well defined (for example it occurs in various
isoforms or glycosylation states) or the analyte is actually a group of similar but different
analytes, then it is preferable to express its concentration in international units. The
international unit (IU) is the unitage assigned by the world health organisation (WHO) to
an international biological standard. This standard is a reference preparation, prepared
by the best methods available at the time, and distributed world wide to be used as a
reference calibrant. International units can be converted to mass units by a conversion
factor release by the WHO. The WHO reference preparations are updated periodically
and conversion factors do change; the conversion factor used must therefore always
match the reference samples that the manufactures used to prepare the calibrants.
Conversion factors for analytes measured in this lab are given in the table below:
Analyte Convention
unit
Conversion
factor
International
recommended
unit
Conversion
standard
Active Renin pg/ml (ng/l) x 1,8 mIU/l WHO 68/356
Growth Hormone ng/ml (μg/l) x 3 mIU/l WHO IS 98/574
Aldosterone pg/ml (ng/l) x 2,774 pmol/l MW: 360,444020
17OH-Progesterone ng/ml (μg/l) x 3,026 nmol/l MW: 330,4611
11-Deoxycortisol ng/ml (μg/l) x 2,886 nmol/l MW: 346,46050
25. What are inter- and intra-assay coefficients of variation (CV), and how are they
determined?
Two indicators of precision are the repeatability and reproducibility. Repeatability is
defined as the closeness of agreement between results of successive measurements
carried out under the same conditions. It is evaluated by performing twenty
measurements, within a single run, and calculating the coefficient of variation (the mean
divided by the standard deviation). This result is called the intra-assay CV.
Reproducibility is the closeness of agreement between results of measurements
performed under changed conditions of measurements (for example: time, operators,
calibrators and reagent lots). This is determined by performing twenty measurements
over twenty different runs over twenty different days and calculating the coefficient of
variation. The result of this calculation is called the inter-assay CV.
26. Describe and illustrate a Scatchard plot of binding data.
Scatchard analysis is a standard method for analysing the equilibrium binding parameters
of a radiolabelled ligand with its receptor. The binding data are derived from a ratio of
specifically bound to specifically free antigen, plotted against the concentration of
specifically bound antigen. This plot gives an estimate of binding affinity and the number
of available binding sites per volume.
The plot is achieved as follows: various concentrations of labelled antigen are prepared.
Three tubes are prepared for each concentration of labelled antigen: total tubes (T)
containing the labelled antigen; non-specific binding tubes (NSB) containing the labelled
antigen, but none of the antibody in question; and total binding (TB) tubes containing the
labelled antigen and the antibody, which are left to equilibrate. After equilibrium (in the
10. TB and NSB tubes only) the bound fraction is precipitated and separated. Once the
assay is completed, three sets of results should be available for each concentration of
radioligand – The total counts (T), the non-specific binding counts (NSB) and the total
binding counts (TB).
The following can now be calculated:
Specifically bound (B): B = TB – NSB
Free ligand (F): F = T - B
These can be converted from counts to concentration units as follows:
)/(activityspecific
1
)/(1022,2
1
)/(effeciency
)/(counts
)/(ionConcentrat 12
molCiCidpmdpmcpm
lcpm
lmol ⋅
⋅
⋅=
The bound can now be plotted against the free to achieve a graph as such:
The data from the saturation experiment can be plotted with Bound/Free on the Y axis
and Bound on the X axis. This data can be analyzed by linear regression to give a
straight line. This is called a Rosenthal Plot.
11. The equilibrium constant of the antibody-antigen reaction (Kd) can be calculated from the
negative reciprocal of the slope of the graph, whereas the x-intercept of the graph gives
the total concentration of binding sites (BMax) measurable under assay conditions. The
presence of a curve (as opposed to a line) indicates the presence of a mixed population
of binding sites.
It should be noted that it is more accurate to do binding analysis on the saturation curve
analyzed by non-linear regression analysis, than by linear analysis on the Rosenthal plot.
This is because the Rosenthal Plot contains the bound on both the x- and y-axes. Since
this is the variable containing the greatest error, a larger error will be distributed in both
directions.
27. In centrifugation, how do you convert rpm to Xg force? What is the conversion
factor for the centrifuge in the RIA Laboratory?
The relative centrifugal force (XG) can be calculated by the equation:
2
2
)()(
00090
rpmspeedrotorcmradius
g
Xg ⋅⋅=
π
where g is the standard gravity equal to exactly 9,80665m/s2
.
The centrifuge in the RIA laboratory has a radius of 24,5cm so the relative centrifugal
force can be calculated using:
6503
)( 2
cmspeedrotor
Xg =
28. Summarise the principle of the plasma renin activity assay, and compare it to the
active renin assay.
Renin is a proteolytic enzyme involved in blood pressure control. Renin cleaves
angiotensinogen to produce the decapeptide angiotensin I. Angiotensin I is then rapidily
cleaved to the biologically active angiotensin II by the angiotensin-converting enzyme.
Two methods of estimating the amount of active renin are available - an enzyme kinetic
assay and a mass assay.
12. Plasma renin activity
This method measures the renin (angiotensinogen proteolytic) activity in serum.
Plasma is mixed with a buffer (pH 6,0) and a angiotensin-converting enzyme inhibitor
(phenylmethylsulfonyl fluoride). The sample is then split into two tubes. One tube is kept
at 37°C to allow for the generation of angiotensin I, the other is kept at 4°C to act as a
blank. After 90 minutes the reaction is stopped (by cooling to 4°C) and the amount of
angiotensin I is measured in each tube using a radioimmunoassay. The angiotensin I
generated is then calculated as shown below. This is the plasma renin activity (PRA) and
is usually expressed in ng/ml/hr.
Time
factorDilution
PRA VolumePlasma
InAngiotensiInAngiotensi 437
⋅
=
°° −
Active renin assay
Historically, the problem with measuring renin directly was that renin exists in both active
and inactive forms. The inactive from (prorenin) can account for up to 90% of the total
renin in the circulation, making its measurement worthless. Since the discovery of
antibodies which are selective for active renin, direct measurements are possible. The
active renin assay is a immunoradiometric assay which uses two antibodies. The first
antibody is a solid phase monoclonal antibody that recognises both active and inactive
renin. The second antibody is a 125 iodine-labelled monoclonal antibody specific for
active renin. The active renin is usually expressed in ng/l.
13. References[1-5]
1. Deupree, J.D.; Tutorial in Receptor Binding Techniques;
http://www.unmc.edu/Pharmacology/receptortutorial/; Lurz, Matthew J.; 1998
2. Edwards, R., S. Blincko, and I. Howes, Principles of immunodiagnostic tests
and their development; with specific use of radioisotopes as tracers, in
Immunodiagnostics : a practical approach, The Practical approach series ;
206, R. Edwards, Editor. 1999, Oxford University Press: Oxford ; New York.
p. cm.
3. Fuentes-Arderiu, X.; Glossary of ISO Metrological and Realted Terms and
Definitions Relevant to Clinical Laboratory Sciences;
http://www.westgard.com/isoglossary.htm; Westgard, James O; 1999
4. Kricka, L.J., Principles of Immunochemical Techniques, in Tietz textbook of
clinical chemistry and molecular diagnostics, C.A. Burtis, et al., Editors.
2006, Elsevier Saunders: Philadelphia. p. XXXVI, 2412 s.
5. Linnet, K. and J.C. Boyd, Selection and Analytical Evaluation of Methods -
With Statistical Techniques, in Tietz textbook of clinical chemistry and
molecular diagnostics, C.A. Burtis, et al., Editors. 2006, Elsevier Saunders:
Philadelphia. p. XXXVI, 2412 s.