SlideShare a Scribd company logo
Automation in the Clinical Laboratory
Course: Clinical Laboratory Principle (SIMS-443)
ZA School of Medical Technology
1
Dr. Ali Raza
Senior Lecturer
SIMS-SIUT
Automation in the clinical laboratory
 Introduction
Automation of the Analytical Process
Unit Operations
Specimen identification
Specimen preparation
Specimen delivery
Specimen loading and aspiration
Specimen processing
Sample induction and internal transport
Reagent handling and storage
Chemical reaction phase
Measurement approaches
Signal processing, data handling and process control
Applications
2
Automation in the clinical laboratory
3
Automation
in the clinical laboratory
Definition:
“ The process whereby an analytical instrument performs
many tests with only minimal involvement of an analyst”
“ Controlled operation of an apparatus, process, or system
by mechanical or electronic devices without human
intervention”
4
Automation in the clinical laboratory
Intelligent automation :
• Built in systems to self-monitor and respond
appropriately to changing conditions.
• Instrument performs a repetitive task by itself,
• Instrument performs a variety of different tasks.
5
Advantages of automation
Automated instruments enables laboratories to
 Process much larger workloads
 Reduce number of staff
 Reduction in the variability of results and errors of
analysis
 Significant improvement in the quality of lab tests
 Cost reduction
6
Advantages of Automation
• Assist the laboratory technologist in test performance
(1) Processing and transport of specimens
(2) Loading of specimens into automated analyzers
(3) Assessment of the results of the performed tests.
7
Principle: Automation in the clinical laboratory
 Automated analyzers generally incorporates mechanized
version of basic manual laboratory techniques and
procedures
 Modern instrumentation is packaged in a wide variety of
configurations.
 Common configuration is the Random-Access Analyzer.
8
Random-Access Analysis
Analyses are performed on a collection of specimens sequentially,
with each specimen analyzed for a different selection of tests.
 This approach permits measurement of a variable number and
variety of analytes in each specimen
9
Random-access analysis
 Profiles or groups of tests are defined for a specimen at
the time the tests to be performed are entered into the
analyzer
(1) A keyboard
(2) by instruction from a laboratory information system
(3) Conjunction with bar coding on the specimen tube
(4) by operator selection of appropriate reagent packs
10
Unit Operations
(11) Steps required to complete an analysis are referred to
collectively as unit operations
11
1-Specimen Identification
• The identifying link (identifier) between Patient and
specimen is made at the patient's bedside
• Maintenance of identifier should remains throughout
(1) Transport of the specimen to the laboratory,
(2) Specimen Analysis,
3) Preparation of a Report
12
1- Specimen Identification
Automatic Identification and Data Collection (AIDC)
 Electronically detect a unique characteristic or unique data
string associated with a physical object.
Example of identifiers:
(1) Serial number
(2) Part number
(3) Colour
(4) Manufacturer
(5) Patient number
13
Automatic Identification and Data Collection
14
Labeling
15
Bar Coding
• Incorporation of bar coding
technology into analytical systems.
• Initiating bar code identification at a
patient's bedside ensures greater
integrity of the specimen's identity in
an analyzer.
16
Labeling
• Electronic entry of a test order for a uniquely
identified patient generates a specimen label bearing
a unique laboratory Accession Number.
 Laboratory station
 Nursing station
• The unique label is fixed to the specimen
collection tube when the blood is drawn.
17
labeling
A- Primary labeling
B-Secondary labeling
18
A- Primary labeling
 Arrival of the Specimen (log-in)
 Patient Identification and Collection Information
 Laboratory Requisition Form
 An Accession Number
19
A- Primary labeling
Proper alignment of the label on the collection tube
20
B-Secondary labeling:
• Bearing essential
information from the
original label must be
affixed to any secondary
tubes created.
• E.g:
Serum removal from the
original tube
21
A number may be handwritten on the specimen cup,
or a coded label may be affixed to the original tube or
to a specimen cup.
Bar Coding System
Composition:
 Bar code Printer
 Bar code Reader/ Scanner.
22
Bar code Reader
Bar code Printer
Bar Coding System
Symbology :
• describe as the rules specifying the way the data are
encoded into the bars and spaces.
 Specification for that symbology.:
• The width of the bars and spaces,
• The number of each bar
Different combinations of the bars and spaces
represent different characters.
23
Bar Coding System
24
Bar Coding System
• When a bar code scanner is passed over the bar code, the
light beam from the scanner is absorbed by the dark bars
and not reflected;
• But the beam is reflected by the light spaces.
• A photocell detector in the scanner receives the reflected
light
• converts that light into an electrical signal that then is digitized.
25
Bar Coding System
 One-dimensional bar coding systems
 Two-dimensional bar coding systems
26
One-dimensional bar coding systems
is an array of rectangular bars and spaces arranged in a
predetermined pattern following unambiguous rules to
represent elements of data referred to as characters.
27
Two-dimensional bar coding systems
 The data is encoded based on both the vertical and horizontal
arrangement of the pattern, thus it is read in two dimensions.
 doesn't just encode alphanumeric information.
 use patterns of squares, hexagons, dots, and other shapes to
encode data.
 Example:
 Data Matrix
 QR Code
 PDF417
 or
28
PDF417
Choice for automatic identification: Bar Code Technology
“Decrease in identification Errors”
29
Errors
“The state or condition of being wrong in conduct or
judgement.”
“ A measure of the estimated difference between the
observed or calculated value of a quantity and its true
value.”
30
Errors
• Human misreading of either specimen label or
loading list may cause
• misplacement of specimens, calibrators, or
controls.
31
Identification Errors
 Error risks begin at the bedside
 Compounded with each specimen processing step
between collection and analysis.
• High risks with hand transcription :
• Accessioning,
• labeling and relabeling,
• Creation of load lists.
32
2- Specimen Preparation
33
2- Specimen Preparation
• Manually Specimen Preparation process results in a delay
• Examples :
 Clotting of blood
 Centrifugation
 Transfer of serum to secondary tubes
34
2- Specimen Preparation
• Whole blood assay system: specimen preparation time
essentially is eliminated.
• Automated or semi automated ion-selective electrodes:
Measure ion activity in whole blood rather than Ion
concentration
35
3- Specimen delivery
36
3- Specimen Delivery
Methods are used to deliver specimens to the lab
1) Courier Service
2) Pneumatic tube systems
3) Electric track vehicles:
4) Mobile robots
37
Courier Service
 Collection site to the lab and between lab
 At a given pick up point and specified time.
 Arrangement of immediate pick up adds cost to the analytical
process
 Specimen breakage
38
Pneumatic tube systems
 Propel cylindrical containers through networks
 By compressed air or by partial vacuum.
used for transporting solid objects
39
Electric track vehicles
• Conveyor system for light goods transport.
• utilizes independently driven vehicles
• traveling on a monorail track network
40
Mobile Robots
41
Successful to transport lab
specimen both within lab and
outside lab.
Various sizes and shapes of
specimen containers
Programmable
Cost effective
4- Specimen loading and Aspiration
42
4- Specimen Loading and Aspiration
• Automatic analyzer directly analyzed serum/plasma from
primary collection tube
• Serum transferred from the specimen tubes to cups
CUP features:
• Permit required volume for testing
• Made of inert material
• Disposable
• Minimize cost
• Minimize Evaporation
43
4- Specimen Loading and Aspiration
• Specimens may undergo
• Evaporation
• Degradation
a) Thermo-labile Analytes: Temperatures
b) Photo labile Analytes : Photo degradation. E.g.: Bilirubin
• Specimens and calibrators are held at refrigerated loading zone
for Thermo-labile Analytes
• Reduced Photo-degradation by
• Semi-opaque cups
• smoke- or orange-colored plastic covers
The loading zone: area in which specimens are held in the instrument before they are
analyzed.
44
4- Specimen Loading and Aspiration
 Contamination
• Splatter of serum
• Stoppers of primary containers are opened
• Decant serum into specimen cups
45
4- Specimen Loading and Aspiration
 Contamination
 Closed container sampling systems
• Initially penetrates the primary container's rubber stopper,
followed by
• the specimen probe passes through a hollow needle
• Prevents damage or plugging of the specimen probe
• After the specimen probe is withdrawn, the outer hollow
needle also is withdrawn so that the stopper reseals and no
specimen escapes.
• Examples: Automated hematology and chemistry analyzers.
46
5- Specimen Processing
47
5- Specimen Processing
 Automation of analytical procedures requires removal of
• Proteins
• Other interferants
 To automate this separation step, several automated
immunoassay analyzers use bound antibodies or
proteins in a solid phase format.
48
5- Specimen Processing
• Binding of antigens and antibodies occurs on a solid
surface to which the antibodies or reactive proteins
have been adsorbed or chemically bonded.
 Solid phases are
(1) Beads
(2) coated tubes
(3) Microtiter plates
(4) Magnetic & Nonmagnetic Microparticles
(5) Fiber matrices
49
6- Sample introduction and internal transport
50
6- Sample Introduction and Internal Transport
• Sample introduction into the analyzer and its subsequent
transport within the analyzer
A) Continuous-flow Systems
B) Discrete Processing Systems
51
Continuous-flow Systems
A type of sample analysis in which each specimen in a
batch passes through the same continuous stream at the
same rate and is subjected to the same analytical
reactions
52
Discrete Processing Systems
A type of analysis in which each specimen in a batch has
its own physical and chemical space separate from other
specimen.
53
A) Continuous-flow Systems:
• Peristaltic Pump
• Sample is aspirated through the sample probe
• into a stream of flowing liquid, whereby transported to
analytical stations in the instrument
• To ensure proportionality between calibrators, controls, and
specimens, the pump and roller speed must remain constant.
Peristaltic pump
54
B) Discrete Processing Systems
• Positive-liquid-displacement pipettes
• Specimens, calibrators, and controls are delivered
by a single pipette to the next stage in the analytical process.
• A positive-displacement pipette designed for
1- to dispense only aspirated sample into the reaction receptacle
2- to flush out sample together with diluent.
55
6- Sample Introduction and Internal Transport
• Carry-Over:
Transport of a quantity of analytes or reagent from one specimen
reaction into and contaminating a subsequent one.
56
Carry-Over
• Minimized by
• Adequate flush-to-specimen ratio and incorporating wash
stations for the sample probe.
• Wiping the outside of the sample probe to prevent transfer of
a portion of the previous specimen into the next specimen
cup.
• Using New pipette tip for each pipetting
57
7- Reagent handling and storage
58
7- Reagent Handling and Storage
• Reagent Handling:
Labels on reagent containers include information such as
(1) Reagent identification
(2) Volume of the contents or number of tests
(3) Expiration date
(4) Lot number
Storage
• Plastic or glass containers used for reagents storage
59
Automated Analyzers are classified as
 Open-system analyzer
 Closed-system analyzer
60
Open Versus Closed Systems
Open system
• change the parameters
related to an analysis
• prepare "in-house" reagents
• use reagents from a variety
of suppliers.
• Less expansive
• Longer stability
Closed system
• reagent to be in a
unique container or
• format provided by the
manufacturer.
• Expansive
• Shorter stability
61
8- Reagent delivery
62
8- Reagent Delivery
• Liquid reagents are acquired and delivered to mixing
and reaction chambers either by
• Pumps (through tubes)
• Positive-displacement syringe devices
63
9- Chemical reaction phase
64
9- Chemical Reaction Phase
• Sample and reagents react in the chemical reaction
phase.
• Factors are important in this phase
(1) Vessel in which the reaction occurs
(2) Cuvet in which the reaction is monitored
(3) Timing of the reaction(s)
(4) Mixing and transport of reactants
(5) Thermal conditioning of fluids.
65
9- Chemical Reaction Phase
 Reaction vessels: Reused in many instruments.
• Time before reusable must be replaced depends on their
composition . E.g.:
• 1 month for plastic
• 2 years for standard glass vessels
• Not replaced unless physically damaged. Pyrex glass
 Cuvet: disposable cuvets
• simplified automation
• eliminated carryover and maintenance of flow cells.
• development of improved plastics ( acrylic and polyvinylchloride)
and manufacturing technology.
66
 Timing of the reaction(s):
• The time allowed for a reaction to occur depends on a variety of
factors.
• Reaction time depends on the rate of transport of reaction mixture
through the system to the measurement station.
 Mixing and transport of Reactants
1. Forceful dispensing
2. Magnetic stirring
3. Vigorous lateral displacement
4. A rotating paddle
5. Use of ultrasonic energy
67
 Thermal Regulation:
 Establishment of a controlled temperature environment
in close contact with the reaction container
 Efficient heat transfer from the environment to the
reaction mixture.
68
10- Measurement Appoaches
69
10- Measurement Approaches
 Photometers
 Spectrophotometer
 Fluorometers
 Luminometers
70
10- Measurement Approaches
(Photomety or Spectrophotomety)
The measurement of absorbance requires the following three basic
components
1. An optical source: Radiant energy sources used in automated
systems
• E.g: Tungsten, quartz-halogen, deuterium, mercury, xenon lamps,
and lasers.
• Spectrum wavelengths 300 to 700 nm.
71
2. Spectral Isolation
• Spectral isolation is achieved by Interference filters.
• Filters have peak transmissions of 30 - 80% and bandwidths
of 5 to 15 nm
• Filters are mounted in a filter wheel,
• Appropriate filter is moved into place under command of the
system's computer
72
Spectral Isolation
• Monochromators with gratings and slits provide a continuous
choice of wavelengths.
• Coupled with a stationary photodiode array, to isolate the
spectrum.
• These two elements also are coupled with fiber-optic light
guides to transfer the passage of light energy through cuvets
at locations convenient for mechanization.
73
3. A detector
• Photometric Detectors:
• Photodiodes used as detectors in many automated
systems
• Provide a high signal to noise ratio and fast detector
response times for fluorescent and chemiluminescent
measurements.
74
3. A detector
• Notes:
• Proper alignment of cuvets with the light path(s) is important
in both automated and manual analyzers.
• Stray energy and internal reflections must be kept to acceptable
levels.
• If the light path is not perpendicular to the cuvet, inaccuracy and
imprecision may occur, particularly in kinetic analyses.
75
76
Reflectance Photometry
• In reflectance photometry diffuse reflected light is measured.
• The reflected light results from illumination, with diffused
light, of a reaction mixture in a carrier or from the diffusion
of light by a reaction mixture in an illuminated carrier.
• The intensity of the reflected light from the reagent carrier is
compared with that reflected from a reference surface.
77
Fluorometry
• emission of electromagnetic radiation by a species that has
absorbed exciting radiation from an outside source.
• Intensity of emitted (fluorescent) light is directly proportional to
concentration of the excited species
• used widely for automated immunoassay.
• It is approximately 1000 times more sensitive than comparable
absorbance spectrophotometry,
• but background interference due to fluorescence of native
serum is a major problem.
78
Turbidimetry and Nephelometry
• Turbidimetry and nephelometry are optical techniques
• Are applicable to methods measuring the precipitate
formation in antigen-antibody reactions
• These techniques are used to measure plasma proteins
and for therapeutic drug monitoring.
79
Chemiluminescence and Bioluminescence
• Chemiluminescence and bioluminescence differ from fluorometry
in that the excitation event is caused by a chemical or
electrochemical reaction and not by photo-luminescence
• The applications of chemiluminescence and bioluminescence
have increased significantly with the development of automated
instrumentation and several new reagent systems.
• Because of their attamole-to-zeptomole detection limits,
chemiluminescence and bioluminescence reactions have been
used widely as direct and indicator labels in the development of
immunoassays.
80
Electrochemical
• The most widely used electrochemical approach involves ion-
selective electrodes.
• These electrodes have replaced flame photometry in the
determination of sodium and potassium.
• Electrochemical detectors also have been used for the
measurement of other electrolytes and indirect application in the
analysis of several other serum constituents
• The relationship between ion activity and the concentration of
ions in the specimens must be established with calibrating
solutions, and such electrodes need to be recalibrated frequently
to compensate for alterations of electrode response.
81
11- Signal Processing, Data handling and Process
Control
82
11- Signal Processing, Data handling and Process Control
• The interfacing and integration of computers into automated
analyzers and analytical systems has had a major impact on the
acquisition and processing of analytical data.
• Analogue signals are converted to digital forms by analog-to-
digital converters.
• The computer and resident software then process the digital data
into useful and meaningful output.
• Data processing has allowed automation of such procedures as
nonisotopic immunoassays and reflectance spectrometry because
computer algorithms readily transform complex, nonlinear
standard responses into linear calibration curves.
83
11- Signal Processing, Data handling and Process Control
Several functions performed by integrated computers in
automated analyzers
• Command and phase the electromechanical operation
of the analyzer are performed
• Uniformly
• Repeatable
• Correct Sequence
• Control of operational features of automated equipment,
• calculation of results,
• monitoring of operation contribute to the increased
reproducibility of results. 84
11- Signal Processing, Data handling and Process Control
• Computers acquire, assess, process, and store operational
data from the analyzers.
• Monitor instrument functions for correct execution and react
to improper function by recording the site and nature of the
malfunction.
• Computers enable communication interactions between
the analyzer and operator.
• Diagnostic computer messages to the user describing the site
and type of problem enable quick identification of problems
and prompt correction.
• Graphical displays provide detailed and interactive
troubleshooting guidance to instrument operators and visual
display of the status of each specimen and associated quality
control data. 85
• Permit interactive communication between computer
systems in the modem laboratory analyzer and the
Laboratory Information System (LIS).
• Instrument manufacturers have been developing ethernet
interfaces for networked connections with TCP/IP
(Transmission Control Protocol/lnternet Protocol).
86
Workstation
(1)Serves as the point of interaction with the instrument
operator
(2)Accepts test orders
(3)Monitors the testing process
(4) Assists with analysis of process quality
(5) Provides facilities for review and verification of test results
87
Workstation
The workstation is usually directly interfaced with the LIS
host, accepting downloaded test orders, and uploading
test results.
Most workstations have facilities to
(1) display Levy-Jennings quality control charts,
(2) monitor the progress of each test order, and
(3) troubleshoot the analyze
88
Integrated automation for clinical laboratory
1. Chemistry
2. Hematology
3. Immunoassay
4. Coagulation
5. Microbiology
6. Nucleic acid testing
• Provide efficient and cost-effective operation with a
minimum of operator input.
89
Instrument Cluster
To reduce labor costs, instrument manufacturers are
developing approaches that will allow a single
technologist to simultaneously control and monitor the
functions of several instruments.
90
Automation
Processes have been automated and used in the clinical
laboratory.
1. Urine analyzers
2. Cell counters
3. Nucleic acid analyzers
4. Microtiter plate systems
5. Automated pipetting stations
6. Point of- care testing analyzers.
91
Automation
Urine Analyzers
 Many of the same analytical principles are used for the
quantification of serum and urine constituents.
 It is more difficult to automate testing of urine than serum
because of the broad range of concentrations of many urine
constituents.
 This requires a low limit of detection to measure low
concentrations, and expanded linearity to permit measurements
of high concentrations without dilution.
 This requirement, together with the relatively low demand for
urine tests compared with that for serum tests, has restricted
the development of analyzers designed specifically for urine
constituents. 92
1. Cell Counters:
 Analyzers that perform a complete blood count have been
automated through the use of the "Coulter principle,'' which
is based on
(1) Cell conductivity
(2) light scatter
(3) Flow cytometry.
93
1. Cell Counters
The Coulter principle is based on changes in electrical impedance produced by
nonconductive particles suspended in an electrolyte as they pass through a small
aperture between electrodes.
 In the sensing zone of the aperture, the volume of electrolyte displaced by the particle
(cell) is measured as a change in voltage that is proportional to the volume of the
particle.
By carefully controlling the quantity of electrolyte drawn through the aperture, several
thousand particles per second are counted and sized individually.
Red blood cells, white blood cells, and platelets are identified by their sizes.
Alternating current in the radiofrequency range short-circuits the bipolar lipid layer of
the cell membrane, allowing energy to penetrate the cell.
 Information about intracellular structure, including chemical composition and nuclear
volume, is collected with this technique.
94
Flow Cytometry
 Cells stained with a fluorescent dye that travel in suspension one
by one past a laser light source. (Unstained cells also are
measured.)
 Scattered light and emitted light are collected in front of the light
source and at right angles, respectively.
 Information derived through measurement of light scatter when
a cell is struck by the laser beam is then used to estimate
(1) Cell Shape
(2) Size
(3) Cellular granularity
(4) Nuclear lobularity
(5) Cell surface structure
95
Automation
Nucleic Acid Analyzers:
 Automation of the analysis of nucleic acids developed
rapidly as an outgrowth of the Human Genome Project.“
 Several manufacturers have developed automation to
assist with the isolation of nucleic acids and with
analysis of nucleic acids using several amplification
schemes and nucleic acid sequencing.
 Many of these techniques have been miniaturized using
chip technology Microfluidic chip
96
Microtiter Plate Systems
 Commonly used in immunoassays and nucleic acid analyses.
 As used for enzyme-linked immunosorbent assay (ELISA) assays,
microtiter plates usually are made of polystyrene and have 48 or
96 wells coated with antibody
specific for the antigen of interest.
 After incubation of serum in the microtiter plate well, the well is
washed to remove unbound antigen, and a second antibody with
conjugated indicator
enzyme is added.
97
Microtiter plate systems
 After a second incubation period, the well is washed to remove the unbound
conjugate.
 A color producing product is developed by the addition of enzyme substrate and
the reaction is terminated at a specific time.
 With the development of automated pipetting stations, the liquid handling steps
required for microtiter plate assays have been fully automated to make
microtiter plate assays a viable technology for carrying out large numbers of
immunoassays.
 Automated pipetting stations have a cartesian robot with a pipette fixed to the
end of a probe that moves about a rectangular space.
 The probe is capable of moving in the X, Y, and Z axes. Liquids may be aspirated
and dispensed in any location within the rectangular space.
98
Automatic Pipetting stations
• used to automate an analytical procedure for which an
automated analyzer does not exist or cannot be justified.
Pipetting robots are
(1) Easy to program
(2) Rarely malfunction
(3) delivering aliquots with precision and accuracy.
(4) Multiple-channel pipetting Robots:
allow parallel processing of specimens with 8- or 12-
channel probes to handle microtiter plates.
99
Point-of-Care Testing Analyzers
(POCT Analyzers )
• known by a variety of names
 "near-patient"
 "decentralized"
 "off-site" testing
• Rapidly growing component of laboratory testing
100
101
Reference:
Tietz Fundamentals of Clinical Chemistry,
Sixth Edition. Automation in the clinical laboratory,
Chapter 11, pg. 171-187
102
Thank You

More Related Content

What's hot

Pre analytic and postanalytic test management
Pre analytic and postanalytic test managementPre analytic and postanalytic test management
Pre analytic and postanalytic test management
Varsha Shahane
 

What's hot (20)

Quality Control in Laboratory
Quality Control in LaboratoryQuality Control in Laboratory
Quality Control in Laboratory
 
Automation in hematology part 1
Automation in hematology part 1Automation in hematology part 1
Automation in hematology part 1
 
Quality control in clinical laboratory
Quality control in clinical laboratoryQuality control in clinical laboratory
Quality control in clinical laboratory
 
Errors related to clinical laboratory
Errors related to clinical laboratoryErrors related to clinical laboratory
Errors related to clinical laboratory
 
Full automation
Full automationFull automation
Full automation
 
Quality Control in a Medical Testing Laboratory
Quality Control in a Medical Testing LaboratoryQuality Control in a Medical Testing Laboratory
Quality Control in a Medical Testing Laboratory
 
Analytical and post analytical errors in laboratory
Analytical and post analytical errors in laboratoryAnalytical and post analytical errors in laboratory
Analytical and post analytical errors in laboratory
 
Urine preservative
Urine preservative Urine preservative
Urine preservative
 
Laboratory errors.pdf
Laboratory errors.pdfLaboratory errors.pdf
Laboratory errors.pdf
 
PRE AND POST ANALYTICAL ERRORS
PRE AND POST ANALYTICAL ERRORSPRE AND POST ANALYTICAL ERRORS
PRE AND POST ANALYTICAL ERRORS
 
Automation in Clinical Biochemistry.pptx
Automation in Clinical Biochemistry.pptxAutomation in Clinical Biochemistry.pptx
Automation in Clinical Biochemistry.pptx
 
Pre analytic and postanalytic test management
Pre analytic and postanalytic test managementPre analytic and postanalytic test management
Pre analytic and postanalytic test management
 
Hematology auto analyzer
Hematology auto analyzerHematology auto analyzer
Hematology auto analyzer
 
Quality assurance in the department of clinical biochemistry
Quality assurance in the department of clinical biochemistryQuality assurance in the department of clinical biochemistry
Quality assurance in the department of clinical biochemistry
 
Automation
Automation Automation
Automation
 
Point of care testing
Point of care testingPoint of care testing
Point of care testing
 
Serum iron estimation & total iron
Serum iron estimation & total ironSerum iron estimation & total iron
Serum iron estimation & total iron
 
Hemolysis
HemolysisHemolysis
Hemolysis
 
Clinical laboratory
Clinical laboratoryClinical laboratory
Clinical laboratory
 
WESTGARD RULES
WESTGARD RULESWESTGARD RULES
WESTGARD RULES
 

Similar to Automation in the Clinical Lab

Lecture 2 diagnostic molecular microbiology bls
Lecture 2 diagnostic molecular microbiology blsLecture 2 diagnostic molecular microbiology bls
Lecture 2 diagnostic molecular microbiology bls
Bruno Mmassy
 
automation-151203160733-lva1-app6892.pdf
automation-151203160733-lva1-app6892.pdfautomation-151203160733-lva1-app6892.pdf
automation-151203160733-lva1-app6892.pdf
Mohamed Afifi
 
Lecture 2 diagnostic molecular microbiology bls
Lecture 2 diagnostic molecular microbiology blsLecture 2 diagnostic molecular microbiology bls
Lecture 2 diagnostic molecular microbiology bls
Bruno Mmassy
 
Design and development of lateral flow assays for field use
Design and development of lateral flow assays for field useDesign and development of lateral flow assays for field use
Design and development of lateral flow assays for field use
Brendan O'Farrell
 
3 TOPIC3 Forensic Laboratory Techniques SUN.pptx
3 TOPIC3 Forensic Laboratory Techniques SUN.pptx3 TOPIC3 Forensic Laboratory Techniques SUN.pptx
3 TOPIC3 Forensic Laboratory Techniques SUN.pptx
zenAlkaff
 

Similar to Automation in the Clinical Lab (20)

Lecture 2 diagnostic molecular microbiology bls
Lecture 2 diagnostic molecular microbiology blsLecture 2 diagnostic molecular microbiology bls
Lecture 2 diagnostic molecular microbiology bls
 
automation-151203160733-lva1-app6892.pdf
automation-151203160733-lva1-app6892.pdfautomation-151203160733-lva1-app6892.pdf
automation-151203160733-lva1-app6892.pdf
 
Lecture 2 diagnostic molecular microbiology bls
Lecture 2 diagnostic molecular microbiology blsLecture 2 diagnostic molecular microbiology bls
Lecture 2 diagnostic molecular microbiology bls
 
laboratoryservicesinhospitalbyihmr-b-120818024937-phpapp01.pdf
laboratoryservicesinhospitalbyihmr-b-120818024937-phpapp01.pdflaboratoryservicesinhospitalbyihmr-b-120818024937-phpapp01.pdf
laboratoryservicesinhospitalbyihmr-b-120818024937-phpapp01.pdf
 
Laboratory services in hospital by ihmr b
Laboratory services in hospital by ihmr bLaboratory services in hospital by ihmr b
Laboratory services in hospital by ihmr b
 
Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...
Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...
Chromatography Data System: Getting It “Right First Time” Seminar Series – Pa...
 
Automation in cytology.
Automation in cytology.Automation in cytology.
Automation in cytology.
 
Point of care testing ( POCT)
Point of care testing ( POCT)Point of care testing ( POCT)
Point of care testing ( POCT)
 
ICP QC protocol
ICP  QC  protocolICP  QC  protocol
ICP QC protocol
 
Lab Automation Final version.pptx
Lab Automation Final version.pptxLab Automation Final version.pptx
Lab Automation Final version.pptx
 
Lab Automation Final version.pptx
Lab Automation Final version.pptxLab Automation Final version.pptx
Lab Automation Final version.pptx
 
pdf .pdf
pdf .pdfpdf .pdf
pdf .pdf
 
Clinical laboratory workflow - 9.pptx
Clinical laboratory workflow - 9.pptxClinical laboratory workflow - 9.pptx
Clinical laboratory workflow - 9.pptx
 
Automation.pptx
Automation.pptxAutomation.pptx
Automation.pptx
 
Biopharmaceutical Formulation Development CM3 Implementation and Initial Testing
Biopharmaceutical Formulation Development CM3 Implementation and Initial TestingBiopharmaceutical Formulation Development CM3 Implementation and Initial Testing
Biopharmaceutical Formulation Development CM3 Implementation and Initial Testing
 
1.qualitycontrol final
1.qualitycontrol final1.qualitycontrol final
1.qualitycontrol final
 
Design and development of lateral flow assays for field use
Design and development of lateral flow assays for field useDesign and development of lateral flow assays for field use
Design and development of lateral flow assays for field use
 
3 TOPIC3 Forensic Laboratory Techniques SUN.pptx
3 TOPIC3 Forensic Laboratory Techniques SUN.pptx3 TOPIC3 Forensic Laboratory Techniques SUN.pptx
3 TOPIC3 Forensic Laboratory Techniques SUN.pptx
 
Validation of gc instrument
Validation of gc instrumentValidation of gc instrument
Validation of gc instrument
 
Quality control clia
Quality control cliaQuality control clia
Quality control clia
 

More from Ali Raza Ph.D

Chlorine -minerals and trace elements
Chlorine -minerals and trace elements Chlorine -minerals and trace elements
Chlorine -minerals and trace elements
Ali Raza Ph.D
 
Calcium -minerals and trace elements
Calcium -minerals and trace elements Calcium -minerals and trace elements
Calcium -minerals and trace elements
Ali Raza Ph.D
 
Irons -minerals and trace elements
Irons -minerals and trace elements Irons -minerals and trace elements
Irons -minerals and trace elements
Ali Raza Ph.D
 
Sodium -minerals and trace elements
Sodium -minerals and trace elements Sodium -minerals and trace elements
Sodium -minerals and trace elements
Ali Raza Ph.D
 
Potassium -minerals and trace elements
Potassium -minerals and trace elements Potassium -minerals and trace elements
Potassium -minerals and trace elements
Ali Raza Ph.D
 

More from Ali Raza Ph.D (20)

Principles of clinical enzymology
Principles of clinical enzymologyPrinciples of clinical enzymology
Principles of clinical enzymology
 
Optical Techniques sims 2021
Optical Techniques sims 2021Optical Techniques sims 2021
Optical Techniques sims 2021
 
Clinical lab principles, chapter 2 introduction to principles of lab analyses...
Clinical lab principles, chapter 2 introduction to principles of lab analyses...Clinical lab principles, chapter 2 introduction to principles of lab analyses...
Clinical lab principles, chapter 2 introduction to principles of lab analyses...
 
Lecture 1 introduction to nucleic acid,sims 443, 2021
Lecture 1 introduction to nucleic acid,sims 443, 2021Lecture 1 introduction to nucleic acid,sims 443, 2021
Lecture 1 introduction to nucleic acid,sims 443, 2021
 
Chlorine -minerals and trace elements
Chlorine -minerals and trace elements Chlorine -minerals and trace elements
Chlorine -minerals and trace elements
 
Calcium -minerals and trace elements
Calcium -minerals and trace elements Calcium -minerals and trace elements
Calcium -minerals and trace elements
 
Irons -minerals and trace elements
Irons -minerals and trace elements Irons -minerals and trace elements
Irons -minerals and trace elements
 
Sodium -minerals and trace elements
Sodium -minerals and trace elements Sodium -minerals and trace elements
Sodium -minerals and trace elements
 
Potassium -minerals and trace elements
Potassium -minerals and trace elements Potassium -minerals and trace elements
Potassium -minerals and trace elements
 
Minerals and trace elements introduction
Minerals and trace elements introductionMinerals and trace elements introduction
Minerals and trace elements introduction
 
Vitamin-A -- SIMS 2020
Vitamin-A -- SIMS 2020Vitamin-A -- SIMS 2020
Vitamin-A -- SIMS 2020
 
Vitamin k --sims 2020
Vitamin k --sims 2020Vitamin k --sims 2020
Vitamin k --sims 2020
 
Vitamin-E --sims 2020
Vitamin-E --sims 2020Vitamin-E --sims 2020
Vitamin-E --sims 2020
 
Vitamin-D--sims 2020
Vitamin-D--sims 2020Vitamin-D--sims 2020
Vitamin-D--sims 2020
 
Water soluble vitamins --sims 2020
Water soluble vitamins --sims 2020Water soluble vitamins --sims 2020
Water soluble vitamins --sims 2020
 
Classification of vitamins --sims 2020
Classification of vitamins --sims 2020Classification of vitamins --sims 2020
Classification of vitamins --sims 2020
 
Introduction -vitamins
Introduction -vitaminsIntroduction -vitamins
Introduction -vitamins
 
Reverse Transcription PCR
Reverse Transcription PCRReverse Transcription PCR
Reverse Transcription PCR
 
Real-Time PCR
Real-Time PCRReal-Time PCR
Real-Time PCR
 
Nucleic acid detection Techniques
Nucleic acid detection TechniquesNucleic acid detection Techniques
Nucleic acid detection Techniques
 

Recently uploaded

Accounting and finance exit exam 2016 E.C.pdf
Accounting and finance exit exam 2016 E.C.pdfAccounting and finance exit exam 2016 E.C.pdf
Accounting and finance exit exam 2016 E.C.pdf
YibeltalNibretu
 
Industrial Training Report- AKTU Industrial Training Report
Industrial Training Report- AKTU Industrial Training ReportIndustrial Training Report- AKTU Industrial Training Report
Industrial Training Report- AKTU Industrial Training Report
Avinash Rai
 

Recently uploaded (20)

Accounting and finance exit exam 2016 E.C.pdf
Accounting and finance exit exam 2016 E.C.pdfAccounting and finance exit exam 2016 E.C.pdf
Accounting and finance exit exam 2016 E.C.pdf
 
Home assignment II on Spectroscopy 2024 Answers.pdf
Home assignment II on Spectroscopy 2024 Answers.pdfHome assignment II on Spectroscopy 2024 Answers.pdf
Home assignment II on Spectroscopy 2024 Answers.pdf
 
Gyanartha SciBizTech Quiz slideshare.pptx
Gyanartha SciBizTech Quiz slideshare.pptxGyanartha SciBizTech Quiz slideshare.pptx
Gyanartha SciBizTech Quiz slideshare.pptx
 
Mattingly "AI & Prompt Design: Limitations and Solutions with LLMs"
Mattingly "AI & Prompt Design: Limitations and Solutions with LLMs"Mattingly "AI & Prompt Design: Limitations and Solutions with LLMs"
Mattingly "AI & Prompt Design: Limitations and Solutions with LLMs"
 
Sha'Carri Richardson Presentation 202345
Sha'Carri Richardson Presentation 202345Sha'Carri Richardson Presentation 202345
Sha'Carri Richardson Presentation 202345
 
Basic_QTL_Marker-assisted_Selection_Sourabh.ppt
Basic_QTL_Marker-assisted_Selection_Sourabh.pptBasic_QTL_Marker-assisted_Selection_Sourabh.ppt
Basic_QTL_Marker-assisted_Selection_Sourabh.ppt
 
Industrial Training Report- AKTU Industrial Training Report
Industrial Training Report- AKTU Industrial Training ReportIndustrial Training Report- AKTU Industrial Training Report
Industrial Training Report- AKTU Industrial Training Report
 
NLC-2024-Orientation-for-RO-SDO (1).pptx
NLC-2024-Orientation-for-RO-SDO (1).pptxNLC-2024-Orientation-for-RO-SDO (1).pptx
NLC-2024-Orientation-for-RO-SDO (1).pptx
 
How to Create Map Views in the Odoo 17 ERP
How to Create Map Views in the Odoo 17 ERPHow to Create Map Views in the Odoo 17 ERP
How to Create Map Views in the Odoo 17 ERP
 
Jose-Rizal-and-Philippine-Nationalism-National-Symbol-2.pptx
Jose-Rizal-and-Philippine-Nationalism-National-Symbol-2.pptxJose-Rizal-and-Philippine-Nationalism-National-Symbol-2.pptx
Jose-Rizal-and-Philippine-Nationalism-National-Symbol-2.pptx
 
The Art Pastor's Guide to Sabbath | Steve Thomason
The Art Pastor's Guide to Sabbath | Steve ThomasonThe Art Pastor's Guide to Sabbath | Steve Thomason
The Art Pastor's Guide to Sabbath | Steve Thomason
 
Basic phrases for greeting and assisting costumers
Basic phrases for greeting and assisting costumersBasic phrases for greeting and assisting costumers
Basic phrases for greeting and assisting costumers
 
Operations Management - Book1.p - Dr. Abdulfatah A. Salem
Operations Management - Book1.p  - Dr. Abdulfatah A. SalemOperations Management - Book1.p  - Dr. Abdulfatah A. Salem
Operations Management - Book1.p - Dr. Abdulfatah A. Salem
 
Synthetic Fiber Construction in lab .pptx
Synthetic Fiber Construction in lab .pptxSynthetic Fiber Construction in lab .pptx
Synthetic Fiber Construction in lab .pptx
 
Instructions for Submissions thorugh G- Classroom.pptx
Instructions for Submissions thorugh G- Classroom.pptxInstructions for Submissions thorugh G- Classroom.pptx
Instructions for Submissions thorugh G- Classroom.pptx
 
Basic Civil Engg Notes_Chapter-6_Environment Pollution & Engineering
Basic Civil Engg Notes_Chapter-6_Environment Pollution & EngineeringBasic Civil Engg Notes_Chapter-6_Environment Pollution & Engineering
Basic Civil Engg Notes_Chapter-6_Environment Pollution & Engineering
 
NCERT Solutions Power Sharing Class 10 Notes pdf
NCERT Solutions Power Sharing Class 10 Notes pdfNCERT Solutions Power Sharing Class 10 Notes pdf
NCERT Solutions Power Sharing Class 10 Notes pdf
 
Phrasal Verbs.XXXXXXXXXXXXXXXXXXXXXXXXXX
Phrasal Verbs.XXXXXXXXXXXXXXXXXXXXXXXXXXPhrasal Verbs.XXXXXXXXXXXXXXXXXXXXXXXXXX
Phrasal Verbs.XXXXXXXXXXXXXXXXXXXXXXXXXX
 
[GDSC YCCE] Build with AI Online Presentation
[GDSC YCCE] Build with AI Online Presentation[GDSC YCCE] Build with AI Online Presentation
[GDSC YCCE] Build with AI Online Presentation
 
Introduction to Quality Improvement Essentials
Introduction to Quality Improvement EssentialsIntroduction to Quality Improvement Essentials
Introduction to Quality Improvement Essentials
 

Automation in the Clinical Lab

  • 1. Automation in the Clinical Laboratory Course: Clinical Laboratory Principle (SIMS-443) ZA School of Medical Technology 1 Dr. Ali Raza Senior Lecturer SIMS-SIUT
  • 2. Automation in the clinical laboratory  Introduction Automation of the Analytical Process Unit Operations Specimen identification Specimen preparation Specimen delivery Specimen loading and aspiration Specimen processing Sample induction and internal transport Reagent handling and storage Chemical reaction phase Measurement approaches Signal processing, data handling and process control Applications 2
  • 3. Automation in the clinical laboratory 3
  • 4. Automation in the clinical laboratory Definition: “ The process whereby an analytical instrument performs many tests with only minimal involvement of an analyst” “ Controlled operation of an apparatus, process, or system by mechanical or electronic devices without human intervention” 4
  • 5. Automation in the clinical laboratory Intelligent automation : • Built in systems to self-monitor and respond appropriately to changing conditions. • Instrument performs a repetitive task by itself, • Instrument performs a variety of different tasks. 5
  • 6. Advantages of automation Automated instruments enables laboratories to  Process much larger workloads  Reduce number of staff  Reduction in the variability of results and errors of analysis  Significant improvement in the quality of lab tests  Cost reduction 6
  • 7. Advantages of Automation • Assist the laboratory technologist in test performance (1) Processing and transport of specimens (2) Loading of specimens into automated analyzers (3) Assessment of the results of the performed tests. 7
  • 8. Principle: Automation in the clinical laboratory  Automated analyzers generally incorporates mechanized version of basic manual laboratory techniques and procedures  Modern instrumentation is packaged in a wide variety of configurations.  Common configuration is the Random-Access Analyzer. 8
  • 9. Random-Access Analysis Analyses are performed on a collection of specimens sequentially, with each specimen analyzed for a different selection of tests.  This approach permits measurement of a variable number and variety of analytes in each specimen 9
  • 10. Random-access analysis  Profiles or groups of tests are defined for a specimen at the time the tests to be performed are entered into the analyzer (1) A keyboard (2) by instruction from a laboratory information system (3) Conjunction with bar coding on the specimen tube (4) by operator selection of appropriate reagent packs 10
  • 11. Unit Operations (11) Steps required to complete an analysis are referred to collectively as unit operations 11
  • 12. 1-Specimen Identification • The identifying link (identifier) between Patient and specimen is made at the patient's bedside • Maintenance of identifier should remains throughout (1) Transport of the specimen to the laboratory, (2) Specimen Analysis, 3) Preparation of a Report 12
  • 13. 1- Specimen Identification Automatic Identification and Data Collection (AIDC)  Electronically detect a unique characteristic or unique data string associated with a physical object. Example of identifiers: (1) Serial number (2) Part number (3) Colour (4) Manufacturer (5) Patient number 13
  • 14. Automatic Identification and Data Collection 14
  • 16. Bar Coding • Incorporation of bar coding technology into analytical systems. • Initiating bar code identification at a patient's bedside ensures greater integrity of the specimen's identity in an analyzer. 16
  • 17. Labeling • Electronic entry of a test order for a uniquely identified patient generates a specimen label bearing a unique laboratory Accession Number.  Laboratory station  Nursing station • The unique label is fixed to the specimen collection tube when the blood is drawn. 17
  • 19. A- Primary labeling  Arrival of the Specimen (log-in)  Patient Identification and Collection Information  Laboratory Requisition Form  An Accession Number 19
  • 20. A- Primary labeling Proper alignment of the label on the collection tube 20
  • 21. B-Secondary labeling: • Bearing essential information from the original label must be affixed to any secondary tubes created. • E.g: Serum removal from the original tube 21 A number may be handwritten on the specimen cup, or a coded label may be affixed to the original tube or to a specimen cup.
  • 22. Bar Coding System Composition:  Bar code Printer  Bar code Reader/ Scanner. 22 Bar code Reader Bar code Printer
  • 23. Bar Coding System Symbology : • describe as the rules specifying the way the data are encoded into the bars and spaces.  Specification for that symbology.: • The width of the bars and spaces, • The number of each bar Different combinations of the bars and spaces represent different characters. 23
  • 25. Bar Coding System • When a bar code scanner is passed over the bar code, the light beam from the scanner is absorbed by the dark bars and not reflected; • But the beam is reflected by the light spaces. • A photocell detector in the scanner receives the reflected light • converts that light into an electrical signal that then is digitized. 25
  • 26. Bar Coding System  One-dimensional bar coding systems  Two-dimensional bar coding systems 26
  • 27. One-dimensional bar coding systems is an array of rectangular bars and spaces arranged in a predetermined pattern following unambiguous rules to represent elements of data referred to as characters. 27
  • 28. Two-dimensional bar coding systems  The data is encoded based on both the vertical and horizontal arrangement of the pattern, thus it is read in two dimensions.  doesn't just encode alphanumeric information.  use patterns of squares, hexagons, dots, and other shapes to encode data.  Example:  Data Matrix  QR Code  PDF417  or 28 PDF417
  • 29. Choice for automatic identification: Bar Code Technology “Decrease in identification Errors” 29
  • 30. Errors “The state or condition of being wrong in conduct or judgement.” “ A measure of the estimated difference between the observed or calculated value of a quantity and its true value.” 30
  • 31. Errors • Human misreading of either specimen label or loading list may cause • misplacement of specimens, calibrators, or controls. 31
  • 32. Identification Errors  Error risks begin at the bedside  Compounded with each specimen processing step between collection and analysis. • High risks with hand transcription : • Accessioning, • labeling and relabeling, • Creation of load lists. 32
  • 34. 2- Specimen Preparation • Manually Specimen Preparation process results in a delay • Examples :  Clotting of blood  Centrifugation  Transfer of serum to secondary tubes 34
  • 35. 2- Specimen Preparation • Whole blood assay system: specimen preparation time essentially is eliminated. • Automated or semi automated ion-selective electrodes: Measure ion activity in whole blood rather than Ion concentration 35
  • 37. 3- Specimen Delivery Methods are used to deliver specimens to the lab 1) Courier Service 2) Pneumatic tube systems 3) Electric track vehicles: 4) Mobile robots 37
  • 38. Courier Service  Collection site to the lab and between lab  At a given pick up point and specified time.  Arrangement of immediate pick up adds cost to the analytical process  Specimen breakage 38
  • 39. Pneumatic tube systems  Propel cylindrical containers through networks  By compressed air or by partial vacuum. used for transporting solid objects 39
  • 40. Electric track vehicles • Conveyor system for light goods transport. • utilizes independently driven vehicles • traveling on a monorail track network 40
  • 41. Mobile Robots 41 Successful to transport lab specimen both within lab and outside lab. Various sizes and shapes of specimen containers Programmable Cost effective
  • 42. 4- Specimen loading and Aspiration 42
  • 43. 4- Specimen Loading and Aspiration • Automatic analyzer directly analyzed serum/plasma from primary collection tube • Serum transferred from the specimen tubes to cups CUP features: • Permit required volume for testing • Made of inert material • Disposable • Minimize cost • Minimize Evaporation 43
  • 44. 4- Specimen Loading and Aspiration • Specimens may undergo • Evaporation • Degradation a) Thermo-labile Analytes: Temperatures b) Photo labile Analytes : Photo degradation. E.g.: Bilirubin • Specimens and calibrators are held at refrigerated loading zone for Thermo-labile Analytes • Reduced Photo-degradation by • Semi-opaque cups • smoke- or orange-colored plastic covers The loading zone: area in which specimens are held in the instrument before they are analyzed. 44
  • 45. 4- Specimen Loading and Aspiration  Contamination • Splatter of serum • Stoppers of primary containers are opened • Decant serum into specimen cups 45
  • 46. 4- Specimen Loading and Aspiration  Contamination  Closed container sampling systems • Initially penetrates the primary container's rubber stopper, followed by • the specimen probe passes through a hollow needle • Prevents damage or plugging of the specimen probe • After the specimen probe is withdrawn, the outer hollow needle also is withdrawn so that the stopper reseals and no specimen escapes. • Examples: Automated hematology and chemistry analyzers. 46
  • 48. 5- Specimen Processing  Automation of analytical procedures requires removal of • Proteins • Other interferants  To automate this separation step, several automated immunoassay analyzers use bound antibodies or proteins in a solid phase format. 48
  • 49. 5- Specimen Processing • Binding of antigens and antibodies occurs on a solid surface to which the antibodies or reactive proteins have been adsorbed or chemically bonded.  Solid phases are (1) Beads (2) coated tubes (3) Microtiter plates (4) Magnetic & Nonmagnetic Microparticles (5) Fiber matrices 49
  • 50. 6- Sample introduction and internal transport 50
  • 51. 6- Sample Introduction and Internal Transport • Sample introduction into the analyzer and its subsequent transport within the analyzer A) Continuous-flow Systems B) Discrete Processing Systems 51
  • 52. Continuous-flow Systems A type of sample analysis in which each specimen in a batch passes through the same continuous stream at the same rate and is subjected to the same analytical reactions 52
  • 53. Discrete Processing Systems A type of analysis in which each specimen in a batch has its own physical and chemical space separate from other specimen. 53
  • 54. A) Continuous-flow Systems: • Peristaltic Pump • Sample is aspirated through the sample probe • into a stream of flowing liquid, whereby transported to analytical stations in the instrument • To ensure proportionality between calibrators, controls, and specimens, the pump and roller speed must remain constant. Peristaltic pump 54
  • 55. B) Discrete Processing Systems • Positive-liquid-displacement pipettes • Specimens, calibrators, and controls are delivered by a single pipette to the next stage in the analytical process. • A positive-displacement pipette designed for 1- to dispense only aspirated sample into the reaction receptacle 2- to flush out sample together with diluent. 55
  • 56. 6- Sample Introduction and Internal Transport • Carry-Over: Transport of a quantity of analytes or reagent from one specimen reaction into and contaminating a subsequent one. 56
  • 57. Carry-Over • Minimized by • Adequate flush-to-specimen ratio and incorporating wash stations for the sample probe. • Wiping the outside of the sample probe to prevent transfer of a portion of the previous specimen into the next specimen cup. • Using New pipette tip for each pipetting 57
  • 58. 7- Reagent handling and storage 58
  • 59. 7- Reagent Handling and Storage • Reagent Handling: Labels on reagent containers include information such as (1) Reagent identification (2) Volume of the contents or number of tests (3) Expiration date (4) Lot number Storage • Plastic or glass containers used for reagents storage 59
  • 60. Automated Analyzers are classified as  Open-system analyzer  Closed-system analyzer 60
  • 61. Open Versus Closed Systems Open system • change the parameters related to an analysis • prepare "in-house" reagents • use reagents from a variety of suppliers. • Less expansive • Longer stability Closed system • reagent to be in a unique container or • format provided by the manufacturer. • Expansive • Shorter stability 61
  • 63. 8- Reagent Delivery • Liquid reagents are acquired and delivered to mixing and reaction chambers either by • Pumps (through tubes) • Positive-displacement syringe devices 63
  • 65. 9- Chemical Reaction Phase • Sample and reagents react in the chemical reaction phase. • Factors are important in this phase (1) Vessel in which the reaction occurs (2) Cuvet in which the reaction is monitored (3) Timing of the reaction(s) (4) Mixing and transport of reactants (5) Thermal conditioning of fluids. 65
  • 66. 9- Chemical Reaction Phase  Reaction vessels: Reused in many instruments. • Time before reusable must be replaced depends on their composition . E.g.: • 1 month for plastic • 2 years for standard glass vessels • Not replaced unless physically damaged. Pyrex glass  Cuvet: disposable cuvets • simplified automation • eliminated carryover and maintenance of flow cells. • development of improved plastics ( acrylic and polyvinylchloride) and manufacturing technology. 66
  • 67.  Timing of the reaction(s): • The time allowed for a reaction to occur depends on a variety of factors. • Reaction time depends on the rate of transport of reaction mixture through the system to the measurement station.  Mixing and transport of Reactants 1. Forceful dispensing 2. Magnetic stirring 3. Vigorous lateral displacement 4. A rotating paddle 5. Use of ultrasonic energy 67
  • 68.  Thermal Regulation:  Establishment of a controlled temperature environment in close contact with the reaction container  Efficient heat transfer from the environment to the reaction mixture. 68
  • 70. 10- Measurement Approaches  Photometers  Spectrophotometer  Fluorometers  Luminometers 70
  • 71. 10- Measurement Approaches (Photomety or Spectrophotomety) The measurement of absorbance requires the following three basic components 1. An optical source: Radiant energy sources used in automated systems • E.g: Tungsten, quartz-halogen, deuterium, mercury, xenon lamps, and lasers. • Spectrum wavelengths 300 to 700 nm. 71
  • 72. 2. Spectral Isolation • Spectral isolation is achieved by Interference filters. • Filters have peak transmissions of 30 - 80% and bandwidths of 5 to 15 nm • Filters are mounted in a filter wheel, • Appropriate filter is moved into place under command of the system's computer 72
  • 73. Spectral Isolation • Monochromators with gratings and slits provide a continuous choice of wavelengths. • Coupled with a stationary photodiode array, to isolate the spectrum. • These two elements also are coupled with fiber-optic light guides to transfer the passage of light energy through cuvets at locations convenient for mechanization. 73
  • 74. 3. A detector • Photometric Detectors: • Photodiodes used as detectors in many automated systems • Provide a high signal to noise ratio and fast detector response times for fluorescent and chemiluminescent measurements. 74
  • 75. 3. A detector • Notes: • Proper alignment of cuvets with the light path(s) is important in both automated and manual analyzers. • Stray energy and internal reflections must be kept to acceptable levels. • If the light path is not perpendicular to the cuvet, inaccuracy and imprecision may occur, particularly in kinetic analyses. 75
  • 76. 76
  • 77. Reflectance Photometry • In reflectance photometry diffuse reflected light is measured. • The reflected light results from illumination, with diffused light, of a reaction mixture in a carrier or from the diffusion of light by a reaction mixture in an illuminated carrier. • The intensity of the reflected light from the reagent carrier is compared with that reflected from a reference surface. 77
  • 78. Fluorometry • emission of electromagnetic radiation by a species that has absorbed exciting radiation from an outside source. • Intensity of emitted (fluorescent) light is directly proportional to concentration of the excited species • used widely for automated immunoassay. • It is approximately 1000 times more sensitive than comparable absorbance spectrophotometry, • but background interference due to fluorescence of native serum is a major problem. 78
  • 79. Turbidimetry and Nephelometry • Turbidimetry and nephelometry are optical techniques • Are applicable to methods measuring the precipitate formation in antigen-antibody reactions • These techniques are used to measure plasma proteins and for therapeutic drug monitoring. 79
  • 80. Chemiluminescence and Bioluminescence • Chemiluminescence and bioluminescence differ from fluorometry in that the excitation event is caused by a chemical or electrochemical reaction and not by photo-luminescence • The applications of chemiluminescence and bioluminescence have increased significantly with the development of automated instrumentation and several new reagent systems. • Because of their attamole-to-zeptomole detection limits, chemiluminescence and bioluminescence reactions have been used widely as direct and indicator labels in the development of immunoassays. 80
  • 81. Electrochemical • The most widely used electrochemical approach involves ion- selective electrodes. • These electrodes have replaced flame photometry in the determination of sodium and potassium. • Electrochemical detectors also have been used for the measurement of other electrolytes and indirect application in the analysis of several other serum constituents • The relationship between ion activity and the concentration of ions in the specimens must be established with calibrating solutions, and such electrodes need to be recalibrated frequently to compensate for alterations of electrode response. 81
  • 82. 11- Signal Processing, Data handling and Process Control 82
  • 83. 11- Signal Processing, Data handling and Process Control • The interfacing and integration of computers into automated analyzers and analytical systems has had a major impact on the acquisition and processing of analytical data. • Analogue signals are converted to digital forms by analog-to- digital converters. • The computer and resident software then process the digital data into useful and meaningful output. • Data processing has allowed automation of such procedures as nonisotopic immunoassays and reflectance spectrometry because computer algorithms readily transform complex, nonlinear standard responses into linear calibration curves. 83
  • 84. 11- Signal Processing, Data handling and Process Control Several functions performed by integrated computers in automated analyzers • Command and phase the electromechanical operation of the analyzer are performed • Uniformly • Repeatable • Correct Sequence • Control of operational features of automated equipment, • calculation of results, • monitoring of operation contribute to the increased reproducibility of results. 84
  • 85. 11- Signal Processing, Data handling and Process Control • Computers acquire, assess, process, and store operational data from the analyzers. • Monitor instrument functions for correct execution and react to improper function by recording the site and nature of the malfunction. • Computers enable communication interactions between the analyzer and operator. • Diagnostic computer messages to the user describing the site and type of problem enable quick identification of problems and prompt correction. • Graphical displays provide detailed and interactive troubleshooting guidance to instrument operators and visual display of the status of each specimen and associated quality control data. 85
  • 86. • Permit interactive communication between computer systems in the modem laboratory analyzer and the Laboratory Information System (LIS). • Instrument manufacturers have been developing ethernet interfaces for networked connections with TCP/IP (Transmission Control Protocol/lnternet Protocol). 86
  • 87. Workstation (1)Serves as the point of interaction with the instrument operator (2)Accepts test orders (3)Monitors the testing process (4) Assists with analysis of process quality (5) Provides facilities for review and verification of test results 87
  • 88. Workstation The workstation is usually directly interfaced with the LIS host, accepting downloaded test orders, and uploading test results. Most workstations have facilities to (1) display Levy-Jennings quality control charts, (2) monitor the progress of each test order, and (3) troubleshoot the analyze 88
  • 89. Integrated automation for clinical laboratory 1. Chemistry 2. Hematology 3. Immunoassay 4. Coagulation 5. Microbiology 6. Nucleic acid testing • Provide efficient and cost-effective operation with a minimum of operator input. 89
  • 90. Instrument Cluster To reduce labor costs, instrument manufacturers are developing approaches that will allow a single technologist to simultaneously control and monitor the functions of several instruments. 90
  • 91. Automation Processes have been automated and used in the clinical laboratory. 1. Urine analyzers 2. Cell counters 3. Nucleic acid analyzers 4. Microtiter plate systems 5. Automated pipetting stations 6. Point of- care testing analyzers. 91
  • 92. Automation Urine Analyzers  Many of the same analytical principles are used for the quantification of serum and urine constituents.  It is more difficult to automate testing of urine than serum because of the broad range of concentrations of many urine constituents.  This requires a low limit of detection to measure low concentrations, and expanded linearity to permit measurements of high concentrations without dilution.  This requirement, together with the relatively low demand for urine tests compared with that for serum tests, has restricted the development of analyzers designed specifically for urine constituents. 92
  • 93. 1. Cell Counters:  Analyzers that perform a complete blood count have been automated through the use of the "Coulter principle,'' which is based on (1) Cell conductivity (2) light scatter (3) Flow cytometry. 93
  • 94. 1. Cell Counters The Coulter principle is based on changes in electrical impedance produced by nonconductive particles suspended in an electrolyte as they pass through a small aperture between electrodes.  In the sensing zone of the aperture, the volume of electrolyte displaced by the particle (cell) is measured as a change in voltage that is proportional to the volume of the particle. By carefully controlling the quantity of electrolyte drawn through the aperture, several thousand particles per second are counted and sized individually. Red blood cells, white blood cells, and platelets are identified by their sizes. Alternating current in the radiofrequency range short-circuits the bipolar lipid layer of the cell membrane, allowing energy to penetrate the cell.  Information about intracellular structure, including chemical composition and nuclear volume, is collected with this technique. 94
  • 95. Flow Cytometry  Cells stained with a fluorescent dye that travel in suspension one by one past a laser light source. (Unstained cells also are measured.)  Scattered light and emitted light are collected in front of the light source and at right angles, respectively.  Information derived through measurement of light scatter when a cell is struck by the laser beam is then used to estimate (1) Cell Shape (2) Size (3) Cellular granularity (4) Nuclear lobularity (5) Cell surface structure 95
  • 96. Automation Nucleic Acid Analyzers:  Automation of the analysis of nucleic acids developed rapidly as an outgrowth of the Human Genome Project.“  Several manufacturers have developed automation to assist with the isolation of nucleic acids and with analysis of nucleic acids using several amplification schemes and nucleic acid sequencing.  Many of these techniques have been miniaturized using chip technology Microfluidic chip 96
  • 97. Microtiter Plate Systems  Commonly used in immunoassays and nucleic acid analyses.  As used for enzyme-linked immunosorbent assay (ELISA) assays, microtiter plates usually are made of polystyrene and have 48 or 96 wells coated with antibody specific for the antigen of interest.  After incubation of serum in the microtiter plate well, the well is washed to remove unbound antigen, and a second antibody with conjugated indicator enzyme is added. 97
  • 98. Microtiter plate systems  After a second incubation period, the well is washed to remove the unbound conjugate.  A color producing product is developed by the addition of enzyme substrate and the reaction is terminated at a specific time.  With the development of automated pipetting stations, the liquid handling steps required for microtiter plate assays have been fully automated to make microtiter plate assays a viable technology for carrying out large numbers of immunoassays.  Automated pipetting stations have a cartesian robot with a pipette fixed to the end of a probe that moves about a rectangular space.  The probe is capable of moving in the X, Y, and Z axes. Liquids may be aspirated and dispensed in any location within the rectangular space. 98
  • 99. Automatic Pipetting stations • used to automate an analytical procedure for which an automated analyzer does not exist or cannot be justified. Pipetting robots are (1) Easy to program (2) Rarely malfunction (3) delivering aliquots with precision and accuracy. (4) Multiple-channel pipetting Robots: allow parallel processing of specimens with 8- or 12- channel probes to handle microtiter plates. 99
  • 100. Point-of-Care Testing Analyzers (POCT Analyzers ) • known by a variety of names  "near-patient"  "decentralized"  "off-site" testing • Rapidly growing component of laboratory testing 100
  • 101. 101 Reference: Tietz Fundamentals of Clinical Chemistry, Sixth Edition. Automation in the clinical laboratory, Chapter 11, pg. 171-187

Editor's Notes

  1. Molecular diagnostics based on nucleic acids, represents one of the most rapidly developing areas in laboratory medicine. Advances in the field have been made possible by our improved understanding of molecular biology and genetics and of their relationships with human diseases, and the development of powerful technologies for the analysis of nucleic acids.
  2. AIDC, Auto ID, automatic data capture and automatic data collection) is a family of technologies that identify, verify, record, communicate and store information on discrete, packaged or containerized items. Because the process is automated (rather than reliant on pen, paper and people), information is gathered quickly and accurately. The most common technologies used to identify and capture data are barcodes, handheld and fixed-position scanners and imagers, radio frequency identification (RFID) tags and readers, and voice recognition, weighing and cubing devices. Typical applications include receiving and putaway, inventory picking, order fulfillment, determination of weight and volume, and tracking and tracing throughout the supply chain.
  3. Optical character recognition or optical character reader (OCR) is the electronic or mechanical conversion of images of typed, handwritten or printed text into machine-encoded text, whether from a scanned document, a photo of a document, a scene-photo (for example the text on signs and billboards in a landscape photo) ...
  4. n official order laying claim to the use of property or materials.
  5. After accessioning specimens begin the technical handling processes
  6. The risks begin at the bedside Compounded with each processing step a specimen undergoes between collection from the patient and anaiysis by the instrument. The risks are particularly great when hand transcription is invoked for accessioning, labeling and relabeling, and creation of load lists. Human misreading of either specimen label or loading list may cause misplacement of specimens, calibrators, or controls.
  7. The risks begin at the bedside Compounded with each processing step a specimen undergoes between collection from the patient and anaiysis by the instrument. The risks are particularly great when hand transcription is invoked for accessioning, labeling and relabeling, and creation of load lists. Human misreading of either specimen label or loading list may cause misplacement of specimens, calibrators, or controls.
  8. The risks begin at the bedside Compounded with each processing step a specimen undergoes between collection from the patient and anaiysis by the instrument. The risks are particularly great when hand transcription is invoked for accessioning, labeling and relabeling, and creation of load lists. Human misreading of either specimen label or loading list may cause misplacement of specimens, calibrators, or controls.
  9. Manually Specimen Preparation process results in a delay in the preparation of a specimen for analysis. clotting of blood in specimen collection tubes, their subsequent centrifugation, and the transfer of serum to secondary tubes requires a finite time to complete. whole blood is used in an assay system, specimen preparation time essentially is eliminated. Automated or semi automated ion-selective electrodes, which measure ion activity in whole blood rather than ion concentration,
  10. Manually Specimen Preparation process results in a delay in the preparation of a specimen for analysis. clotting of blood in specimen collection tubes, their subsequent centrifugation, and the transfer of serum to secondary tubes requires a finite time to complete. whole blood is used in an assay system, specimen preparation time essentially is eliminated. Automated or semi automated ion-selective electrodes, which measure ion activity in whole blood rather than ion concentration,
  11. Manually Specimen Preparation process results in a delay in the preparation of a specimen for analysis. clotting of blood in specimen collection tubes, their subsequent centrifugation, and the transfer of serum to secondary tubes requires a finite time to complete. whole blood is used in an assay system, specimen preparation time essentially is eliminated. Automated or semi automated ion-selective electrodes, which measure ion activity in whole blood rather than ion concentration,
  12. Manually Specimen Preparation process results in a delay in the preparation of a specimen for analysis. clotting of blood in specimen collection tubes, their subsequent centrifugation, and the transfer of serum to secondary tubes requires a finite time to complete. whole blood is used in an assay system, specimen preparation time essentially is eliminated. Automated or semi automated ion-selective electrodes, which measure ion activity in whole blood rather than ion concentration,
  13. Pneumatic tube systems: Systems that propel cylindrical containers through networks of tubes by compressed air or by partial vacuum. used for transporting solid objects Electric track vehicles: is a conveyor system for light goods transport. The system utilizes independently driven vehicles traveling on a monorail track network, consisting of straight track elements, bends, curves and transfer-units for changing of travel direction.
  14. Pneumatic tube systems: Systems that propel cylindrical containers through networks of tubes by compressed air or by partial vacuum. used for transporting solid objects Electric track vehicles: is a conveyor system for light goods transport. The system utilizes independently driven vehicles traveling on a monorail track network, consisting of straight track elements, bends, curves and transfer-units for changing of travel direction.
  15. Pneumatic tube systems: Systems that propel cylindrical containers through networks of tubes by compressed air or by partial vacuum. used for transporting solid objects Electric track vehicles: is a conveyor system for light goods transport. The system utilizes independently driven vehicles traveling on a monorail track network, consisting of straight track elements, bends, curves and transfer-units for changing of travel direction.
  16. Pneumatic tube systems: Systems that propel cylindrical containers through networks of tubes by compressed air or by partial vacuum. used for transporting solid objects Electric track vehicles: is a conveyor system for light goods transport. The system utilizes independently driven vehicles traveling on a monorail track network, consisting of straight track elements, bends, curves and transfer-units for changing of travel direction.
  17. Pneumatic tube systems: Systems that propel cylindrical containers through networks of tubes by compressed air or by partial vacuum. used for transporting solid objects Electric track vehicles: is a conveyor system for light goods transport. The system utilizes independently driven vehicles traveling on a monorail track network, consisting of straight track elements, bends, curves and transfer-units for changing of travel direction.
  18. Manually Specimen Preparation process results in a delay in the preparation of a specimen for analysis. clotting of blood in specimen collection tubes, their subsequent centrifugation, and the transfer of serum to secondary tubes requires a finite time to complete. whole blood is used in an assay system, specimen preparation time essentially is eliminated. Automated or semi automated ion-selective electrodes, which measure ion activity in whole blood rather than ion concentration,
  19. the specimen for automatic analysis is serum. contain separator material Many analyzers also sample from cups or tubes filled with serum transferred from the original specimen tubes excess serum that must be present in a cup to permit aspiration of the full volume required for testing. Cups must be made of inert material so that they do not interact with the analytes being measured. Specimen cups also should be disposable to minimize cost, and their shape should, even without a cap, minimize evaporation.
  20. Manually Specimen Preparation process results in a delay in the preparation of a specimen for analysis. clotting of blood in specimen collection tubes, their subsequent centrifugation, and the transfer of serum to secondary tubes requires a finite time to complete. whole blood is used in an assay system, specimen preparation time essentially is eliminated. Automated or semi automated ion-selective electrodes, which measure ion activity in whole blood rather than ion concentration,
  21. Automation of analytical procedures requires removal of Proteins other interferants from some specimens and to separate free and bound fractions of heterogeneous immunoassays.
  22. Automation of analytical procedures requires removal of Proteins other interferants from some specimens and to separate free and bound fractions of heterogeneous immunoassays.
  23. Manually Specimen Preparation process results in a delay in the preparation of a specimen for analysis. clotting of blood in specimen collection tubes, their subsequent centrifugation, and the transfer of serum to secondary tubes requires a finite time to complete. whole blood is used in an assay system, specimen preparation time essentially is eliminated. Automated or semi automated ion-selective electrodes, which measure ion activity in whole blood rather than ion concentration,
  24. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  25. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  26. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  27. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  28. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  29. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  30. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  31. Manually Specimen Preparation process results in a delay in the preparation of a specimen for analysis. clotting of blood in specimen collection tubes, their subsequent centrifugation, and the transfer of serum to secondary tubes requires a finite time to complete. whole blood is used in an assay system, specimen preparation time essentially is eliminated. Automated or semi automated ion-selective electrodes, which measure ion activity in whole blood rather than ion concentration,
  32. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  33. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  34. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  35. Manually Specimen Preparation process results in a delay in the preparation of a specimen for analysis. clotting of blood in specimen collection tubes, their subsequent centrifugation, and the transfer of serum to secondary tubes requires a finite time to complete. whole blood is used in an assay system, specimen preparation time essentially is eliminated. Automated or semi automated ion-selective electrodes, which measure ion activity in whole blood rather than ion concentration,
  36. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  37. Manually Specimen Preparation process results in a delay in the preparation of a specimen for analysis. clotting of blood in specimen collection tubes, their subsequent centrifugation, and the transfer of serum to secondary tubes requires a finite time to complete. whole blood is used in an assay system, specimen preparation time essentially is eliminated. Automated or semi automated ion-selective electrodes, which measure ion activity in whole blood rather than ion concentration,
  38. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  39. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  40. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  41. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  42. Manually Specimen Preparation process results in a delay in the preparation of a specimen for analysis. clotting of blood in specimen collection tubes, their subsequent centrifugation, and the transfer of serum to secondary tubes requires a finite time to complete. whole blood is used in an assay system, specimen preparation time essentially is eliminated. Automated or semi automated ion-selective electrodes, which measure ion activity in whole blood rather than ion concentration,
  43. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  44. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  45. A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).
  46. The slit function is the monochromatic image of the entrance slit of the spectrometer on the detector
  47. Manually Specimen Preparation process results in a delay in the preparation of a specimen for analysis. clotting of blood in specimen collection tubes, their subsequent centrifugation, and the transfer of serum to secondary tubes requires a finite time to complete. whole blood is used in an assay system, specimen preparation time essentially is eliminated. Automated or semi automated ion-selective electrodes, which measure ion activity in whole blood rather than ion concentration,