Call Girl In Zirakpur ❤️♀️@ 9988299661 Zirakpur Call Girls Near Me ❤️♀️@ Sexy...
patient dose management in angiography king saud unversity.pdf
1. Patient dose management in
angiography
Dr. Naima SENHOU
King Saud bin Abdulaziz University
for Health Sciences
Ministry of National Guard Health
Affairs KAMC-Riyadh
12/11/2019
2.
3. Radiation risks (Patients)
Skin burn is a well-documented radiation induced effect
•But there are several other important
deterministic effects than can occur:
• Lens opacities and cataract
• Affected fertility
• Reduced count in White Blood Cells
4. Exposure: Ionization (charge) produced
by photons per unit mass in air
SI units are C/kg
Conventional is the Roentgen R
1 R = 2.58×10−4 C/kg
if the exposure = 1 R then the dose is ~
8.76 mGy In Air
Radiation Units
5. Absorbed Dose: Energy absorbed per
unit mass of the irradiated material.
SI units are Gray (Gy), (1 Gy = 1 J/kg).
Old units were rad (radiation absorbed
dose)
Not all types of radiation cause the same
biological damage
Radiation Units
6. Equivalent Dose: Measure the biological effect of specific type
of radiation
SI units are Sievert (Sv)
Old units were rem (radiation equivalent man)
To get the Equivalent dose we multiply the absorbed dose by a
radiation weighting factor (WR)
Radiation Units
WR for X-ray, Gamma = 1
WR for Alpha ray = 20. The biological effect of Alpha rays is 20
times greater than that of X-Ray
DEQ = WR x D(Absorbed)
8. Effective Dose: (Whole body exposure)
SI units are Sievert (Sv)
Old units were rem (radiation equivalent man)
DEffec = WT x WR x D(Absorbed)
WR: Radiation Weighting Factor
WT: Tissue Weighting Factor
Radiation Units
10. Dose Area Product (DAP)
DAP= Dose (Gy) x Exposed Area (cm2)
DAP provides a good estimation of the total
radiation energy delivered to a patient during a
procedure.
Radiation Units
11. DAP: (Dose Area Product)
Used with Fluoroscopy machines
DAP Units
mGy x cm2
mRad x cm2
Gy = 100 rad
mGy = 0.001 Gy
mGy = 100 mRad
Radiation Units
12. where μ is the linear attenuation coefficient and X is the distance the photon has to travel
through the tissue.
The amount of radiation that passes through tissue is given by the Beer-Lambert law,
which says that
Intensity(out) = Intensity(in) * e -(μ * X)
Attenuation and Dose
The thicker the patient or body part, the fewer photons escape to the other side. We
need a certain number of photons to produce an image
13. The use of iodine compounds was initially related to low toxicity and excellent radio-opacity
rather than physical considerations. However, it was also fortunate that iodine compounds
possess physical properties which make them better contrast agents than compounds with
higher atomic number.
The K-edge of iodine is 33.2 keV. At photon energies slightly above this, iodine actually has
greater attenuating properties for x-rays than the same mass of lead.
iodine compounds as contract agent
14. Iodine k-edge
The k-edge represents the
energy needed to eject a K-
shell electron (the innermost
and most strongly bound
electrons).
Outer shell electrons have
absorption edges but these
are much too low energy to
be relevant.
15. Tissue Contrast
If all different types of tissue stopped x-rays in the same way, then we would have no picture
In soft tissues, the dominant elements (e.g. C, H, O, and N) have very low K-edges, in the
range of a few keV. While these elements do contribute to the photoelectric effect and
attenuate low energy x-rays, there is no relevant k-edge with its substantial change in
attenuation.
However, the elements iodine and barium have K-edges around 30-40 keV, right in the
middle of the x-ray beam spectrum. Thus, soft tissues with even a small amount of iodine
will have a much stronger x-ray stopping power than those without.
The presence of materials with high k-edges such as iodine or barium improve the
contrast at higher kV, but contrast is still better at lower kV.
No change with water and calcium
16. The interventional reference point (IRP) was introduced by the International
Electrotechnical Commission (IEC) in 2000. It is defined to be located at 15 cm from the
isocentre of the C-arm toward the X-ray tube (IEC, 2000)
This point is fixed relative to the X-ray tube and is independent of the patient variation.
The IRP is fixed relative to the X-ray tube as it is always at a constant distance from the
isocentre, which may represent a position on the patient’s skin or a point inside or outside
of the patient’s body The IRP rarely represents the exact location of the patient’s skin over
clinical procedures.
Using accumulated doses at IRP locations tends to result in higher than actual
peak skin doses and thus overestimates the radiation risk to patients.
The interventional reference point (IRP)
17. 17
Factors that influence Patient Absorbed
Dose
• Equipment-related factors
– Movement capabilities of C-arm, X ray source, image
receptor
– Field-of-view size
– Collimator position
– Beam filtration
– Angiography pulse rate and acquisition frame rate
– Angiography and acquisition input dose rates
– Automatic dose-rate control including beam energy
management options
– X ray photon energy spectra
– Software image filters
– Preventive maintenance and calibration
– Quality control
18. 18
Factors that influence Patient
Absorbed Dose
Minimise beam-on time
Maximise source to patient distance
Optimize kVp
Collimate to region of interest
Use last image hold and pulsed angiograohy
Review real time image recordings
Minimise use of magnification modes
Methods to minimise radiation dose to patients during a angiography
procedure are:
19. Patient Position
• The patient should be placed away from the radiation source to lower the skin dose
• The image detector should be placed close to the patient to lower the necessary
dose
less geometric blurring
•The tube should always be positioned under the table (under couch configuration)
less scatter towards the personnel’s eyes
•Keep the patient’s extremities out of the beam
20. Tube Angulation
• The use of steep angulations should be minimized and used only if necessary
• Each 3 cm additional tissue doubles the dose to the patient (factor of ~10
each 10 cm)
Using small angulations actually has the potential to reduce the peak skin dose of the patient
• Moving the X-ray beam to different areas of the patient’s body during the procedures lowers
the peak skin dose
• Good practice: every 30 minutes, change the beam angle
21. Magnification will increase the patient dose (beam energy is condensed to a smaller
area)
• Use the largest Field of view (FOV) possible for the procedure being performed.
The difference in skin dose between the larger FOVs and the smallest is a factor of ~4
•In modern systems, there is a "Live Zoom“ feature without additional radiation and
without significant degradation of the image.
Magnification
22. Collimation
A ‘collimator’ is an adjustable lead shutter located at the exit of the X-ray source that can be
closed down to limit the irradiated area
• Appropriate collimation should be used for the imaging task
this lowers both the patient’s dose and the staff’s
• Modern systems should be equipped with virtual collimation
23. System Settings
• It is important to get to know your system and its options
• Make sure to use the lowest dose setting when possible
• Talk with the manufacturer/technician to set the initial dose setting to ‘low dose’
increase when necessary instead of lowering when possible (less chance to forget)
• Use the proper frame rate, lower where possible
• Use the proper acquisition program
• Make sure when buying a new system that it has proper filtration
Cu filters out the low energy X-rays, lowering skin dose
24. Dose estimation dose management
How can we estimate the
dose received by the
patient???
25. National Council on Radiation Protection & Measurements( NCRP) 168
Fluoroscopically guided
interventional (FGI)
procedures
kerma-area product (PKA)
26.
27.
28.
29.
30. This highlights the need for dose standardisation and to further extend dose
reduction, while maintaining diagnostic quality of the X-ray images or procedures.
DRLs can be set based on local dosimetry or practice data, or can be based on
published values that are appropriate for the local circumstances.
Diagnostic reference levels (DRLs) are a practical tool to promote optimization.
DRLs are general guideline for clinical operations and do not apply directly to
individual patients and examinations.
DRLs should be examination/procedure specific
Different DRLs can also be set for each hospital, district, region or country.
National DRLs (NDRLs) are usually set based on a wide scale survey of the median
doses representing typical practice for specific patient groups (e.g. adults or
children).
Diagnostic reference levels (DRLs