THE MULTIPHYSIOLOGICAL PARAMETER MONITORING SYSTEM
Roshinie D1, Swathy S2, Sangeetha V3, Sriba R4
1,2Roshinie D, Student, Department of BME, Dr.N.G.P. Institute of Technology, Tamil Nadu, India
-----------------------------------------------------------------------***--------------------------------------------------------------------
Abstract:- The Physiological Monitoring System, which is
commonly known as Patient Monitors, in hospital
environment, finds a great use in modern era. These type
of monitoring systems facilitate the measurement or
monitoring of human vital signs such as ECG, Heart rate,
Temperature, NIBP and SpO2. These monitors are mainly
used in all Intensive Care Units, Critical Care Units,
Operation Theatres, Post-operative wards, Emergency
wards, etc., In additional to these parameters, there are
also some essential parameters to be monitored every
instant, in order to eliminate sudden complications and
illness. Though there are rising demands in need of patient
monitors at almost all departments, there are also some
special needs by some departments. The existing monitors
measure only these parameters irrespective of the
department being used. The monitors which are used in
ICUs are similar to where they are used in wards. To
design special monitors, some parameters can be added.
Here are some of the most required parameters, in which
the patient must be regularly monitored: Glucose, pCo2
and Hemoglobin. An additional advantage is that, these
parameters are to be measured in Non-Invasive
technique, to satisfy patient comfort.
Index terms: Non-invasive, Near Infra-red, Micro
controller, Glucose, Hemoglobin, pCo2.
1. INTRODUCTION:
The existing patient monitoring systems finds various
uses in all departments of a hospital, as they possess
some good features such as: they are easily portable,
they possess less weight, all vitals of a person can be
measured less than a minute. The patient monitor
measure a person’s ECG, Heart rate, NIBP, Respiration
rate, Temperature and SpO2 levels. According to the
target parameters, these monitors can be modified and
can be used for various purposes.
1.1. Cardiac monitoring
Involves the measurement of cardiac related parameters
such as ECG, Heart rhythm, Heart rate continuously for a
particular period of time, and is capable of recording the
results. This kind of equipment is commercially known
as ‘Holter monitor’, a small monitor which can be worn
by an ambulatory patient. Cardiac output can also be
measured by an invasive method, with the help of
SwanGanz catheter.
1.2. Hemodynamic monitoring
Involves the measurement of blood pressure and blood
flow. The blood pressure can be measured by both
invasive and non-invasive techniques. In invasive
method, a blood pressure transducer assembly is used
and in non-invasive method, an inflatable blood pressure
cuff is used.
1.3. Respiratory monitoring
Involves the measurement of SpO2, ETCO2 and
Respiratory rate. The SpO2 can be measured non-
invasively via an Infrared sensor. ETCO2 can be
measured invasively, which is popularly known as
capnography. The Respiratory rate can be measured
non-invasively with the help of a thoracic transducer
belt.
1.4. Neurological monitoring
Involves the measurement of intra-cranial pressure,
anaesthetic concentrations, bi-spectral index and most
importantly the EEG. The first three measurements are
incorporated into anaesthesia machines. A typical
neurological unit must contain EEG, as it plays a major
role in determining the activities of brain at various
states.
3. EXISTING SYSTEM:
A general block diagram below, represents the overall
functioning of a physiological monitor.
Fig.1 Block diagram representing the functioning of a
patient monitor.
As discussed before, a Patient monitor can measure 5
vital parameters:
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072
© 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1766
3.1 ECG:
ECG refers to the recording of the electrical activity of
heart for a particular period of time, with the help of
electrodes/leads placed on skin. The leads are placed
according to Einthoven’s principle: Einthoven’s triangle
is the imaginary formation of a triangle, in which, the
two sides belong to the right and the left shoulders and
the third side belong to the pubis. Thus, the shape forms
an inverted triangle, with the heart at its centre.
Generally, both limb electrodes and surface electrodes
are used to measure ECG, but the existing patient
monitors use only 6 (V1 to V6) surface electrodes. V1 is
placed at the right sternal border of 4th intercoastal
space, V2 is placed at the left sternal border of 4th
intercoastal space, V3 is placed at the halfway between V2
and V4, V4 is placed at the 5th intercoastal space at the
mid-clavicular line, V5 is placed at the left anterior
axillary line on the same horizontal plane similar to V4,
V6 is placed at the left mid-axillary line on the same
horizontal plane as V5.
z
Fig.2 Lead placement based on Einthoven triangle to
record ECG.
The ECG wave consists of PQRST peaks.
P wave represents atrial depolarization, QRS complex
represents ventricular depolarization, T wave represents
ventricular repolarization.
Fig.3 Normal ECG waveform.
3.2 Heart Rate:
Heart rate can also be calculated with the help of ECG
waveform, with the help of a formula:
Rate = 60/(R-R interval) bpm.
The R-R interval is calculated from the ECG waveform.
The normal range for heart rate is 60-100bpm. If the
heart rate is above 100 bpm, it indicates Tachycardia and
if the heart rate is below 60 bpm, it indicates
Bradycardia.
3.3 NIBP:
Non-invasive Blood Pressure monitoring enables the
continuous measurement of blood pressure. High blood
pressure does not usually give any warning signs.
Though you experience high blood pressure, you might
feel perfectly well. The only way to determine whether
you have high blood pressure is to have it checked
regularly. Non-invasive blood pressure measurement
usually involves the use of an inflatable-cuff which is
wrapped around the limb of a test subject. The cuff is
inflated and deflated at a controlled rate and hence the
physical parameters are observed. The auscultatory and
oscillometric techniques are well known non-invasive
methods. These methods are indirect because they do
not couple directly to the artery. Modern NIBP monitors
now use oscillometric technique to measure pressure.
The oscillometric technique monitors the changes in cuff
pressure caused by the flow of blood through the artery.
The monitor inflates the cuff to a pressure that occludes
the artery. Even when the artery is occluded, the
pumping of the heart against the artery can cause small
pressure pulses in the cuff baseline pressure. The
monitor lowers cuff pressure at a controlled rate. As the
cuff pressure decreases, blood starts to flow through the
artery. The increasing blood flow causes the amplitude of
the pressure pulses in the cuff to increase. These
pressure pulses continue to increase in amplitude with
decreasing cuff pressure until they reach maximum
amplitude, at which point they begin to decrease with
decreasing cuff pressure. The cuff pressure at which the
pressure pulse amplitude is the greatest is known as
Mean Arterial Pressure (MAP).
The manner in which the pulse amplitudes vary is often
referred to as a pulse envelope. The envelope is an
imaginary line that connects the peak of each pressure
pulse and forms an outline. The shape of the envelope is
observed by different monitors using a variety of
techniques to determine the diastolic and systolic blood
pressures.
The normal range of NIBP is 120/80 mmHg. The top,
larger number is called the systolic pressure. This is the
pressure generated when the heart contracts. It reflects
the pressure of the blood against arterial walls. The
bottom, smaller number is called the diastolic pressure.
This reflects the pressure in the arteries while the heart
is filling and resting between heartbeats.
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072
© 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1767
3.4 Pleth or SpO2:
SpO2 is measured with the help of pulseoximeter. Pulse
oximetry is a simple non-invasive method of monitoring
the percentage of haemoglobin (Hb), which is saturated
with oxygen. The pulse oximeter consists of a probe
attached to the patient’s finger, toe, ear lobe or forehead,
which is linked to a computerised unit. The unit displays
the percentage of Hb saturated with oxygen together
with an audible signal for each pulse beat, and calculated
heart rate. A pulse wave related to flow is displayed
graphically.
Fig.4 Normal pleth waveform.
The measurement of SpO2 is done with the help of a
probe, which consists of two beams of light: red light of
wavelength 600-750 nm and infrared light of wavelength
850-1000 nm. It is purely based on the concept that,
oxygenated hemoglobin absorbs more infrared light and
allows more red light to pass through.
Fig.5 SpO2 measurement probe assembly.
The red/infrared ratio is calculated and then compared
with the look-up table and finally converted to SpO2
value. The normal SpO2 value ranges from 95-100%.
3.5 Temperature:
The body temperature plays a vital role in maintaining
the hemostasis, as temperature plays a vital role in
biochemical reactions. The normal temperature of a
human is 97 to 99 F. The temperature above the normal
range indicates Hyperthermia and the temperature
below normal range indicates Hypothermia.
These parameters make up together into a complete
module.
Fig. 6 Internal Diagram of a patient monitor.
4. UNSATISFACTORINESS IN THE EXISTING SYSTEM:
Though, these monitors provide a continuous
measurement of most of the vital parameters, there are
also some vitals, which must be measured continuously,
in order to avoid risks.
A patient’s vitals are measured mostly in critical
departments, where the patients are at high risk, such as
Intensive Care Units, Critical Care Units, Post-operative
wards, Operation theatres, Emergency departments,
Maternity wards etc., where, not only the above
discussed vitals are considered important. There are also
some parameters, which are not continuously measured,
but at regular time intervals and also involves invasive
method, which leads to patient discomfort and
sometimes pain, which could not be tolerated.
Glucose, pCO2 and Hemoglobin values are also necessary
for proper functioning of the body. Hence it must also be
monitored continuously and also by invasive method, for
satisfying patient’s comfort.
5. PROPOSED MODEL:
In order to solve the above mentioned issues or in other
words, to satisfy the physician’s needs, we have
considered 3 main parameters, Glucose, pCO2 and
Hemoglobin, which can be continuously measured by
non-invasive technique.
5.1 WHY THESE PARAMETERS ARE CONSIDERED?
Here are some reasons, for considering the above
mentioned parameters to be added:
1. Any patient, who is recovering from a surgery,
may face an alternating blood glucose level, no
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072
© 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1768
matter, whether the patient is a diabetic or non-
diabetic. Sometimes, the blood glucose level may
reach high or too low. If the glucose values are
left unnoticed, it may lead to serious ill effects.
In addition to that, the blood glucose level is
measured by invasive method, which would be
uncomfortable to the patient. Hence, glucose
level must be monitored both continuously and
non-invasively.
2. A patient, who has undergone a lung surgery,
may face with periodic respiratory acidosis or
respiratory alkalosis. If these changes are left
unnoticed, then it leads to lung failure or even
death. To determine the proper functioning of
the lungs, pO2, pCO2 and pH must be measured.
Here, ABG machine, also known as blood gas
machine aids in measuring these parameters. It
is a painful and invasive technique, which
involves drawing of a small amount of blood
from the radial artery or femoral artery and
then the values are estimated. This again leads
to patient discomfort. Hence, in our proposed
model, pCO2 value is estimated both
continuously and non-invasively.
3. A patient may face with decreasing hemoglobin
levels, after a surgery or after the chemotherapy
(for oncologic patients). The decrease in
hemoglobin usually lead to fatigue, anaemia and
several other complications. The hemoglobin
level in blood is also estimated only periodically
at regular time intervals and also involves
invasive technique. Hence, in our proposed
model, the hemoglobin value is estimated both
continuously and non-invasively.
5.2 Principle behind the model:
The whole module is built based on the principle of Near
Infra Red (NIR) Spectroscopic Technique.
Fig.7 Electromagnetic spectrum.
Near-infrared spectroscopic method uses the near-
infrared region of the electromagnetic spectrum which
ranges from 780 nm to 2500 nm. Some applications
include medical and physiological diagnostics and
research including blood sugar, pulse oximetry,
functional neuroimaging, sports medicine, elite sports
training, ergonomics, rehabilitation, neonatal research,
brain computer interface, urology, and neurology.
Fig.8 A simple representation of NIR spectroscopy.
5.3 COMPONENTS USED FOR BUILDING UP THE
MODULE:
The major components required are as follows:
1. Arduino Uno R3
2. Infrared LEDs
3. Quadripple Operational Amplifier (LM324N)
4. Operational Amplifier (TL082)
The above mentioned components play a vital role in
estimating the proposed parameters. In addition to
these, the following components are also required:
5. Phototransistors
6. Photodiode
7. Resistors and Capacitors
8. USB cable
5.3.1 ARDUINO UNO R3:
The Arduino is a microcontroller board based on
ATmega328 microcontroller. It consists of 14
input/output pins, of which, 6 pins can be used as PWM
inputs, 6 analog inputs, a 16MHz crystal oscillator, a USB
connection, a power jack, an ICSP header and a reset
button.
Fig.9 Arduino Uno microcontroller.
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072
© 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1769
The Arduino Uno can be powered via the USB connection
or with an external power supply. The power source is
selected automatically. External power can come either
from an AC-to-DC adapter or battery.
The board can operate on an external supply of 6 to 20
volts. The recommended range is 7 to 12 volts. Out of the
above mentioned 14 pins, some pins have specific
functions as listed below:
Serial Pins 0 (Rx) and 1 (Tx): Rx and Tx pins are used
to receive and transmit TTL serial data. They are
connected with the ATmega328P USB to TTL serial chip.
External Interrupt Pins 2 and 3: These pins can be
configured to trigger an interrupt on a low value, a rising
or falling edge, or a change in value.
PWM Pins 3, 5, 6, 9 and 11: These pins provide an 8-bit
PWM output by using analogWrite() function.
SPI Pins: 10 (SS), 11 (MOSI), 12 (MISO) and 13 (SCK):
These pins are used for SPI communication.
In-built LED Pin 13: This pin is connected with an built-
in LED, when pin 13 is HIGH, LED is on and when pin 13
is LOW, it is off.
Power pins:
1. VIN refers to the input voltage to the
Arduino board when it is using an external
power source.
2. 5V: This pin outputs a regulated 5V from
the regulator on the board. The board can
be supplied with power either from the DC
power jack (7 - 12V), the USB connector
(5V), or the VIN pin of the board (7-12V).
Supplying voltage via the 5V or 3.3V pins
bypasses the regulator, and can damage
your board.
3. 3.3V: A 3.3 volt supply generated by the on-
board regulator. Maximum current draw is
50 mA.
4. GND: Ground pins.
4.3.2 Infrared LEDs:
For detection of Glucose, 700-1100nm of light is capable
of absorbing glucose molecules in blood. Similarly, for
the detection of pCO2, 1200-1500nm of light is capable of
absorbing pCO2 present in the blood. The detection of
hemoglobin doesn’t need any light source, rather, it can
be detected with the help of arduino program.
5.3.3 Quadripple Operational Amplifier: (LM324N)
The Quadripple operational amplifier is used here, as a
part of the circuit, which estimates the blood glucose
value.
The LM324 series are low−cost, quad operational
amplifiers with true differential inputs. They have
several distinct advantages over standard operational
amplifier types in single supply applications. The quad
amplifier can operate at supply voltages as low as 3.0 V
or as high as 32 V.
Fig.10 LM324N
Features:
1. Internally frequency compensated for
unity gain.
2. Large DC voltage gain 100 dB.
3. Wide bandwidth 1 MHz.
4. Wide power supply range: Single supply
3V to 32V.
5. Essentially independent of supply
voltage.
Fig.11 Pin configuration of LM324N.
Pin functions:
Pin No. Function
1 Output of first comparator
2 Inverting input of first
comparator
3 Non-Inverting input of first
comparator
4 Supply voltage 5V
5 Non-Inverting input of second
comparator
6 Inverting input of second
comparator
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072
© 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1770
7 Output of second comparator
8 Output of third comparator
9 Inverting input of third
comparator
10 Non-Inverting input of third
comparator
11 Ground
12 Non-Inverting input of fourth
comparator
13 Inverting input of fourth
comparator
14 Output of fourth comparator
5.3.4 Operational Amplifier (TL082):
This amplifier is used as a part of the circuit, which
estimates pCO2. TL082 is a low cost, high speed, dual
JFET input operational amplifiers with an internally
trimmed input offset voltage. They require low supply
current yet maintain a large gain bandwidth product and
fast slew rate. In addition, well matched high voltage
JFET input devices provide very low input bias and offset
currents. These amplifiers may be used in applications
such as high speed integrators, fast D/A converters,
sample and hold circuits and many other circuits
requiring low input offset voltage, low input bias current,
high input impedance, high slew rate and wide
bandwidth.
The devices also exhibit low noise and offset voltage
drift.
Fig.12 TL082.
Features of TL082:
1. Low Power Consumption: 1.4 mA
2. Wide Common-Mode and Differential Voltage
range.
3. Low Input Bias Current: 30 pA
4. Low Input Offset Current: 5 pA
5. Output Short-Circuit Protection.
6.
Pin Configuration:
Fig.13 TL082 pin configuration.
5.3.5 Photodiode (OPT101):
The OPT101 is a monolithic photodiode with on-chip
trans-impedance amplifier. Output voltage increases
linearly with light intensity. The amplifier is designed for
single or dual power-supply operation, making it ideal
for batteryoperated equipment. The integrated
combination of photodiode and transimpedance
amplifier on a single chip eliminates the problems
commonly encountered in discrete designs such as
leakage current errors, noise pick-up, and gain peaking
due to stray capacitance. The 0.09 x 0.09 inch
photodiode is operated in the photoconductive mode for
excellent linearity and low dark current.
The OPT101 operates from +2.7V to +36V supplies and
quiescent current is only 120µA. It is available in clear
plastic 8-pin DIP, and J-formed DIP for surface mounting.
Temperature range is 0°C to +70°C.
Fig.14. OPT101
Pin configuration:
Fig.15 Pin configuration of OPT101.
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072
© 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1771
7. BLOCK DIAGRAM:
Fig. 16 Block diagram for proposed system.
7. ESTIMATION OF GLUCOSE:
Some of the optical properties of glucose help us to
determine that the NIR wavelength of 950 nm is suitable
to estimate the blood glucose level.
Block diagram for glucose:
Fig.17 Block diagram for detection of glucose.
The IR source and the detector are hence fixed in the
board, and it is covered to eliminate the interference of
the environmental light sources.
Fig. 18. Assembly of source and detector.
The source and detector are connected to the Arduino
Uno for power supply and to read the analog inputs.
Fig.19 Circuit diagram for estimation of glucose.
The amount of IR source, absorbed by the glucose in the
blood is determined by the detector, which is sent as an
analog signal to the Arduino Uno. The analog signal is
then converted into a digital voltage, by means of
algorithm. According to Beer Lambert’s law, the
absorption of light by the solution is directly
proportional to the voltage developed.
Fig.20. Experimental setup for the detection of glucose.
The normal range of glucose is 100 to 125 mg/dL.
8. ESTIMATION OF PCO2:
It is experimentally proven by most of the research
scholars that, the CO2 can be absorbed by the near
infrared light source of wavelength ranging from 800-
2500 nm. Here, a NIR source of wavelength 800 nm is
used to estimate the amount of pCO2 present in blood.
The circuit diagram to estimate pCO2 concentration is as
follows:
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072
© 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1772
Fig. 21. Circuit diagram for the estimation of pCO2.
OPT101 plays a major role here to estimate pCO2. The
amount of NIR light, absorbed by pCO2 molecules, is
detected by means of a photodetector, which converts
the absorbed light into its corresponding analog voltage.
The analog voltage is then converted into digital voltage,
which follows Beer Lambert’s law.
The normal range of pCO2 is 38 to 45 mmHg.
Fig.22 Experimental setup for the detection of pCO2.
9. ESTIMATION OF HEMOGLOBIN:
The hemoglobin can be estimated by means of simple
algorithm, which requires the values of pCO2 and SpO2.
The common formula to estimate Hemoglobin
concentration in blood is:
RESULTS:
All the 3 proposed parameters are combined together
and the final setup is given as follows:
Fig.23 Final circuit for the proposed system.
It has been checked with our faculties and student
members, the results displayed are as follows:
The Multiphysiological Parameter Monitoring
SystemResult 1.png
The Multiphysiological Parameter Monitoring
SystemResult 2.png
The Multiphysiological Parameter Monitoring
SystemResult 3.png
10. CONCLUSION:
Thus, among all the invasive parameters, glucose, pCO2
and hemoglobin were chosen, to detect them purely non-
invasive and tried to resolve patient discomfort. This will
find a greater advantage in all the emergency and critical
care departments. The future works will deal with
developing this circuit into a complete module and
testing among the general public, thus ensuring that, this
module will be safe that they can be used in hospitals.
11. REFERENCES:
1. Shyqyri Haxha, Jaspreet Jhoja et.al, OPTICAL
BASED NON-INVASIVE GLUCOSE MONITORING
SENSOR, IEEE Photonics Journal, Volume: 8
Issue: 6.
2. Carl D Machoff et al., A NOVEL NON-INVASIVE
BLOOD GLUCOSE MONITOR, American Diabetes
Association in Emerging Treatment and
Technology, 2002 Dec; 25(12): 2268-2275.
3. Masab Ahmad et al., NON-INVASIVE BLOOD
GLUCOSE MONITORING USING NEAR
INFRARED SPECTROSCOPY, EDN Journal,
October 16, 2013.
4. N.Joglekar et al., A REVIEW OF BLOOD GAS
SENSORS AND A PROPOSED PORTABLE
SOLUTION, IEEE Journal,
10.1109/ISICIR.2007.4441843.
5. Jens Kratil et al., NON INVASIVE MEASUREMENT
FOR AN INVIVO HEMOGLOBIN MEASUREMENT,
IEEE Journal, 10.1109/ICSENS.2011.6126982
6. Md Kamrul Hasan et al., ANALYSING THE
EXISTING NON-INVASIVE HEMOGLOBIN
International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056
Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072
© 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1773
MEASUREMENT TECHNIQUES, IEEE Journal,
10.1109/UEMCON.2017.8249080.
7. Raid Saleem Al-Baradie et al., PORTABLE
SMART NON-INVASIVE HEMOGLOBIN
MEASUREMENT SYSTEM, IEEE Journal,
10.1109/SSD.2013.6564149.
8. U. Timm et al., SENSOR SYSTEM FOR NON-
INVASIVE OPTICAL HEMOGLOBIN
DETERMINATION, IEEE Journal,
10.1109/ICSENS.2009.5398321.

More Related Content

What's hot

Design and Implementation of programmable Cardiac Pacemaker Using VHDL
Design and Implementation of programmable Cardiac Pacemaker Using VHDLDesign and Implementation of programmable Cardiac Pacemaker Using VHDL
Design and Implementation of programmable Cardiac Pacemaker Using VHDLIJERA Editor
 
Cuffless blood pressure monitoring project
Cuffless blood pressure monitoring projectCuffless blood pressure monitoring project
Cuffless blood pressure monitoring projectVipul Chavan
 
Design and fabrication of cost-effective heart-rate pulse monitoring sensor s...
Design and fabrication of cost-effective heart-rate pulse monitoring sensor s...Design and fabrication of cost-effective heart-rate pulse monitoring sensor s...
Design and fabrication of cost-effective heart-rate pulse monitoring sensor s...TELKOMNIKA JOURNAL
 
Multipara monitor -Application and Maintenance
Multipara monitor -Application and MaintenanceMultipara monitor -Application and Maintenance
Multipara monitor -Application and Maintenanceshashi sinha
 
Improving the Design of Blood Pressure and Blood Saturation Monitors
Improving the Design of Blood Pressure and Blood Saturation MonitorsImproving the Design of Blood Pressure and Blood Saturation Monitors
Improving the Design of Blood Pressure and Blood Saturation MonitorsLuca Parisi
 
Blood pressure and blood saturation monitors
Blood pressure and blood saturation monitorsBlood pressure and blood saturation monitors
Blood pressure and blood saturation monitorsLuca Parisi
 
Pacemaker | Implantable Cardiac Devices For Heart Failures
Pacemaker | Implantable Cardiac Devices For Heart FailuresPacemaker | Implantable Cardiac Devices For Heart Failures
Pacemaker | Implantable Cardiac Devices For Heart FailuresYashodaHospitals
 
high blood pressure 8
high blood pressure 8high blood pressure 8
high blood pressure 8giddyegg4296
 
Postoperative cardiovascular complications
Postoperative cardiovascular complicationsPostoperative cardiovascular complications
Postoperative cardiovascular complicationsRiyaBaghele
 

What's hot (20)

Design and Implementation of programmable Cardiac Pacemaker Using VHDL
Design and Implementation of programmable Cardiac Pacemaker Using VHDLDesign and Implementation of programmable Cardiac Pacemaker Using VHDL
Design and Implementation of programmable Cardiac Pacemaker Using VHDL
 
Cuffless blood pressure monitoring project
Cuffless blood pressure monitoring projectCuffless blood pressure monitoring project
Cuffless blood pressure monitoring project
 
Holter monitoring
Holter monitoringHolter monitoring
Holter monitoring
 
Design and fabrication of cost-effective heart-rate pulse monitoring sensor s...
Design and fabrication of cost-effective heart-rate pulse monitoring sensor s...Design and fabrication of cost-effective heart-rate pulse monitoring sensor s...
Design and fabrication of cost-effective heart-rate pulse monitoring sensor s...
 
Hcl hh20
Hcl hh20Hcl hh20
Hcl hh20
 
Holter monitor and information does it provide
Holter monitor and information does it provideHolter monitor and information does it provide
Holter monitor and information does it provide
 
Multipara monitor -Application and Maintenance
Multipara monitor -Application and MaintenanceMultipara monitor -Application and Maintenance
Multipara monitor -Application and Maintenance
 
Holter monitoring
Holter monitoringHolter monitoring
Holter monitoring
 
Improving the Design of Blood Pressure and Blood Saturation Monitors
Improving the Design of Blood Pressure and Blood Saturation MonitorsImproving the Design of Blood Pressure and Blood Saturation Monitors
Improving the Design of Blood Pressure and Blood Saturation Monitors
 
Hawzhin
HawzhinHawzhin
Hawzhin
 
Cardiac monitors - Medical Equipment
Cardiac monitors - Medical EquipmentCardiac monitors - Medical Equipment
Cardiac monitors - Medical Equipment
 
Blood pressure and blood saturation monitors
Blood pressure and blood saturation monitorsBlood pressure and blood saturation monitors
Blood pressure and blood saturation monitors
 
Pacemaker | Implantable Cardiac Devices For Heart Failures
Pacemaker | Implantable Cardiac Devices For Heart FailuresPacemaker | Implantable Cardiac Devices For Heart Failures
Pacemaker | Implantable Cardiac Devices For Heart Failures
 
high blood pressure 8
high blood pressure 8high blood pressure 8
high blood pressure 8
 
Pacemakers
PacemakersPacemakers
Pacemakers
 
Cardiac electrophysiology master class
Cardiac electrophysiology master classCardiac electrophysiology master class
Cardiac electrophysiology master class
 
high blood pressure 8
high blood pressure 8high blood pressure 8
high blood pressure 8
 
Pace maker professor md toufiqur rahman
Pace maker professor md  toufiqur rahman Pace maker professor md  toufiqur rahman
Pace maker professor md toufiqur rahman
 
Postoperative cardiovascular complications
Postoperative cardiovascular complicationsPostoperative cardiovascular complications
Postoperative cardiovascular complications
 
2015 tutorial 11 ECG
2015 tutorial 11 ECG2015 tutorial 11 ECG
2015 tutorial 11 ECG
 

Similar to IRJET- The Multiphysiological Parameter Monitoring System

An Integrated Approach of Blood Pressure and Heart Rate Measurement Systems
An Integrated Approach of Blood Pressure and Heart Rate Measurement SystemsAn Integrated Approach of Blood Pressure and Heart Rate Measurement Systems
An Integrated Approach of Blood Pressure and Heart Rate Measurement SystemsIRJET Journal
 
Design and Implementation of programmable Cardiac Pacemaker Using VHDL
Design and Implementation of programmable Cardiac Pacemaker Using VHDLDesign and Implementation of programmable Cardiac Pacemaker Using VHDL
Design and Implementation of programmable Cardiac Pacemaker Using VHDLIJERA Editor
 
Insight into Monitoring of the surgical patient
Insight into Monitoring of the surgical patientInsight into Monitoring of the surgical patient
Insight into Monitoring of the surgical patientlmdm11
 
IRJET-Intelligent ICU Monitoring System
IRJET-Intelligent ICU Monitoring SystemIRJET-Intelligent ICU Monitoring System
IRJET-Intelligent ICU Monitoring SystemIRJET Journal
 
anaesthesia.Monitoring 2(dr.amr)
anaesthesia.Monitoring 2(dr.amr)anaesthesia.Monitoring 2(dr.amr)
anaesthesia.Monitoring 2(dr.amr)student
 
4. 7770 8115-1-pb
4. 7770 8115-1-pb4. 7770 8115-1-pb
4. 7770 8115-1-pbIAESIJEECS
 
Heamodynamic Monitoring System Using Photoplethysmography
Heamodynamic Monitoring System Using PhotoplethysmographyHeamodynamic Monitoring System Using Photoplethysmography
Heamodynamic Monitoring System Using Photoplethysmographyabhijith prabha
 
Patient monitoring system
Patient monitoring systemPatient monitoring system
Patient monitoring systemSuni Reddy
 
Design of Radial Pulse Detector
Design of Radial Pulse DetectorDesign of Radial Pulse Detector
Design of Radial Pulse DetectorIRJET Journal
 
Cardiovascular monitoring Part I
Cardiovascular monitoring Part ICardiovascular monitoring Part I
Cardiovascular monitoring Part ISiddhanta Choudhury
 
IRJET - Sleep Apnea Detection using Physiological Signals
IRJET -  	  Sleep Apnea Detection using Physiological SignalsIRJET -  	  Sleep Apnea Detection using Physiological Signals
IRJET - Sleep Apnea Detection using Physiological SignalsIRJET Journal
 
Measurement of Multiparameters using Anaesthesia Injector Based on Arm Processor
Measurement of Multiparameters using Anaesthesia Injector Based on Arm ProcessorMeasurement of Multiparameters using Anaesthesia Injector Based on Arm Processor
Measurement of Multiparameters using Anaesthesia Injector Based on Arm Processorijsrd.com
 
An Introduction To Surgical Icu
An Introduction To Surgical IcuAn Introduction To Surgical Icu
An Introduction To Surgical IcuDang Thanh Tuan
 
cardiac output monitoring
cardiac output monitoringcardiac output monitoring
cardiac output monitoringmadhu chaitanya
 
SMART HOSPITAL TECHNOLOGY
SMART HOSPITAL TECHNOLOGYSMART HOSPITAL TECHNOLOGY
SMART HOSPITAL TECHNOLOGYhiij
 
Merged document 12
Merged document 12Merged document 12
Merged document 12V!vEk@nAnD S
 

Similar to IRJET- The Multiphysiological Parameter Monitoring System (20)

An Integrated Approach of Blood Pressure and Heart Rate Measurement Systems
An Integrated Approach of Blood Pressure and Heart Rate Measurement SystemsAn Integrated Approach of Blood Pressure and Heart Rate Measurement Systems
An Integrated Approach of Blood Pressure and Heart Rate Measurement Systems
 
Design and Implementation of programmable Cardiac Pacemaker Using VHDL
Design and Implementation of programmable Cardiac Pacemaker Using VHDLDesign and Implementation of programmable Cardiac Pacemaker Using VHDL
Design and Implementation of programmable Cardiac Pacemaker Using VHDL
 
Anesthesia 5th year, 6th & 7th lectures (Dr. Gona)
Anesthesia 5th year, 6th & 7th lectures (Dr. Gona)Anesthesia 5th year, 6th & 7th lectures (Dr. Gona)
Anesthesia 5th year, 6th & 7th lectures (Dr. Gona)
 
Non invasive
Non invasiveNon invasive
Non invasive
 
Insight into Monitoring of the surgical patient
Insight into Monitoring of the surgical patientInsight into Monitoring of the surgical patient
Insight into Monitoring of the surgical patient
 
IRJET-Intelligent ICU Monitoring System
IRJET-Intelligent ICU Monitoring SystemIRJET-Intelligent ICU Monitoring System
IRJET-Intelligent ICU Monitoring System
 
IJET-V3I1P26
IJET-V3I1P26IJET-V3I1P26
IJET-V3I1P26
 
anaesthesia.Monitoring 2(dr.amr)
anaesthesia.Monitoring 2(dr.amr)anaesthesia.Monitoring 2(dr.amr)
anaesthesia.Monitoring 2(dr.amr)
 
4. 7770 8115-1-pb
4. 7770 8115-1-pb4. 7770 8115-1-pb
4. 7770 8115-1-pb
 
Heamodynamic Monitoring System Using Photoplethysmography
Heamodynamic Monitoring System Using PhotoplethysmographyHeamodynamic Monitoring System Using Photoplethysmography
Heamodynamic Monitoring System Using Photoplethysmography
 
Patient monitoring system
Patient monitoring systemPatient monitoring system
Patient monitoring system
 
Design of Radial Pulse Detector
Design of Radial Pulse DetectorDesign of Radial Pulse Detector
Design of Radial Pulse Detector
 
Cardiovascular monitoring Part I
Cardiovascular monitoring Part ICardiovascular monitoring Part I
Cardiovascular monitoring Part I
 
IRJET - Sleep Apnea Detection using Physiological Signals
IRJET -  	  Sleep Apnea Detection using Physiological SignalsIRJET -  	  Sleep Apnea Detection using Physiological Signals
IRJET - Sleep Apnea Detection using Physiological Signals
 
Measurement of Multiparameters using Anaesthesia Injector Based on Arm Processor
Measurement of Multiparameters using Anaesthesia Injector Based on Arm ProcessorMeasurement of Multiparameters using Anaesthesia Injector Based on Arm Processor
Measurement of Multiparameters using Anaesthesia Injector Based on Arm Processor
 
NIBP
NIBPNIBP
NIBP
 
An Introduction To Surgical Icu
An Introduction To Surgical IcuAn Introduction To Surgical Icu
An Introduction To Surgical Icu
 
cardiac output monitoring
cardiac output monitoringcardiac output monitoring
cardiac output monitoring
 
SMART HOSPITAL TECHNOLOGY
SMART HOSPITAL TECHNOLOGYSMART HOSPITAL TECHNOLOGY
SMART HOSPITAL TECHNOLOGY
 
Merged document 12
Merged document 12Merged document 12
Merged document 12
 

More from IRJET Journal

TUNNELING IN HIMALAYAS WITH NATM METHOD: A SPECIAL REFERENCES TO SUNGAL TUNNE...
TUNNELING IN HIMALAYAS WITH NATM METHOD: A SPECIAL REFERENCES TO SUNGAL TUNNE...TUNNELING IN HIMALAYAS WITH NATM METHOD: A SPECIAL REFERENCES TO SUNGAL TUNNE...
TUNNELING IN HIMALAYAS WITH NATM METHOD: A SPECIAL REFERENCES TO SUNGAL TUNNE...IRJET Journal
 
STUDY THE EFFECT OF RESPONSE REDUCTION FACTOR ON RC FRAMED STRUCTURE
STUDY THE EFFECT OF RESPONSE REDUCTION FACTOR ON RC FRAMED STRUCTURESTUDY THE EFFECT OF RESPONSE REDUCTION FACTOR ON RC FRAMED STRUCTURE
STUDY THE EFFECT OF RESPONSE REDUCTION FACTOR ON RC FRAMED STRUCTUREIRJET Journal
 
A COMPARATIVE ANALYSIS OF RCC ELEMENT OF SLAB WITH STARK STEEL (HYSD STEEL) A...
A COMPARATIVE ANALYSIS OF RCC ELEMENT OF SLAB WITH STARK STEEL (HYSD STEEL) A...A COMPARATIVE ANALYSIS OF RCC ELEMENT OF SLAB WITH STARK STEEL (HYSD STEEL) A...
A COMPARATIVE ANALYSIS OF RCC ELEMENT OF SLAB WITH STARK STEEL (HYSD STEEL) A...IRJET Journal
 
Effect of Camber and Angles of Attack on Airfoil Characteristics
Effect of Camber and Angles of Attack on Airfoil CharacteristicsEffect of Camber and Angles of Attack on Airfoil Characteristics
Effect of Camber and Angles of Attack on Airfoil CharacteristicsIRJET Journal
 
A Review on the Progress and Challenges of Aluminum-Based Metal Matrix Compos...
A Review on the Progress and Challenges of Aluminum-Based Metal Matrix Compos...A Review on the Progress and Challenges of Aluminum-Based Metal Matrix Compos...
A Review on the Progress and Challenges of Aluminum-Based Metal Matrix Compos...IRJET Journal
 
Dynamic Urban Transit Optimization: A Graph Neural Network Approach for Real-...
Dynamic Urban Transit Optimization: A Graph Neural Network Approach for Real-...Dynamic Urban Transit Optimization: A Graph Neural Network Approach for Real-...
Dynamic Urban Transit Optimization: A Graph Neural Network Approach for Real-...IRJET Journal
 
Structural Analysis and Design of Multi-Storey Symmetric and Asymmetric Shape...
Structural Analysis and Design of Multi-Storey Symmetric and Asymmetric Shape...Structural Analysis and Design of Multi-Storey Symmetric and Asymmetric Shape...
Structural Analysis and Design of Multi-Storey Symmetric and Asymmetric Shape...IRJET Journal
 
A Review of “Seismic Response of RC Structures Having Plan and Vertical Irreg...
A Review of “Seismic Response of RC Structures Having Plan and Vertical Irreg...A Review of “Seismic Response of RC Structures Having Plan and Vertical Irreg...
A Review of “Seismic Response of RC Structures Having Plan and Vertical Irreg...IRJET Journal
 
A REVIEW ON MACHINE LEARNING IN ADAS
A REVIEW ON MACHINE LEARNING IN ADASA REVIEW ON MACHINE LEARNING IN ADAS
A REVIEW ON MACHINE LEARNING IN ADASIRJET Journal
 
Long Term Trend Analysis of Precipitation and Temperature for Asosa district,...
Long Term Trend Analysis of Precipitation and Temperature for Asosa district,...Long Term Trend Analysis of Precipitation and Temperature for Asosa district,...
Long Term Trend Analysis of Precipitation and Temperature for Asosa district,...IRJET Journal
 
P.E.B. Framed Structure Design and Analysis Using STAAD Pro
P.E.B. Framed Structure Design and Analysis Using STAAD ProP.E.B. Framed Structure Design and Analysis Using STAAD Pro
P.E.B. Framed Structure Design and Analysis Using STAAD ProIRJET Journal
 
A Review on Innovative Fiber Integration for Enhanced Reinforcement of Concre...
A Review on Innovative Fiber Integration for Enhanced Reinforcement of Concre...A Review on Innovative Fiber Integration for Enhanced Reinforcement of Concre...
A Review on Innovative Fiber Integration for Enhanced Reinforcement of Concre...IRJET Journal
 
Survey Paper on Cloud-Based Secured Healthcare System
Survey Paper on Cloud-Based Secured Healthcare SystemSurvey Paper on Cloud-Based Secured Healthcare System
Survey Paper on Cloud-Based Secured Healthcare SystemIRJET Journal
 
Review on studies and research on widening of existing concrete bridges
Review on studies and research on widening of existing concrete bridgesReview on studies and research on widening of existing concrete bridges
Review on studies and research on widening of existing concrete bridgesIRJET Journal
 
React based fullstack edtech web application
React based fullstack edtech web applicationReact based fullstack edtech web application
React based fullstack edtech web applicationIRJET Journal
 
A Comprehensive Review of Integrating IoT and Blockchain Technologies in the ...
A Comprehensive Review of Integrating IoT and Blockchain Technologies in the ...A Comprehensive Review of Integrating IoT and Blockchain Technologies in the ...
A Comprehensive Review of Integrating IoT and Blockchain Technologies in the ...IRJET Journal
 
A REVIEW ON THE PERFORMANCE OF COCONUT FIBRE REINFORCED CONCRETE.
A REVIEW ON THE PERFORMANCE OF COCONUT FIBRE REINFORCED CONCRETE.A REVIEW ON THE PERFORMANCE OF COCONUT FIBRE REINFORCED CONCRETE.
A REVIEW ON THE PERFORMANCE OF COCONUT FIBRE REINFORCED CONCRETE.IRJET Journal
 
Optimizing Business Management Process Workflows: The Dynamic Influence of Mi...
Optimizing Business Management Process Workflows: The Dynamic Influence of Mi...Optimizing Business Management Process Workflows: The Dynamic Influence of Mi...
Optimizing Business Management Process Workflows: The Dynamic Influence of Mi...IRJET Journal
 
Multistoried and Multi Bay Steel Building Frame by using Seismic Design
Multistoried and Multi Bay Steel Building Frame by using Seismic DesignMultistoried and Multi Bay Steel Building Frame by using Seismic Design
Multistoried and Multi Bay Steel Building Frame by using Seismic DesignIRJET Journal
 
Cost Optimization of Construction Using Plastic Waste as a Sustainable Constr...
Cost Optimization of Construction Using Plastic Waste as a Sustainable Constr...Cost Optimization of Construction Using Plastic Waste as a Sustainable Constr...
Cost Optimization of Construction Using Plastic Waste as a Sustainable Constr...IRJET Journal
 

More from IRJET Journal (20)

TUNNELING IN HIMALAYAS WITH NATM METHOD: A SPECIAL REFERENCES TO SUNGAL TUNNE...
TUNNELING IN HIMALAYAS WITH NATM METHOD: A SPECIAL REFERENCES TO SUNGAL TUNNE...TUNNELING IN HIMALAYAS WITH NATM METHOD: A SPECIAL REFERENCES TO SUNGAL TUNNE...
TUNNELING IN HIMALAYAS WITH NATM METHOD: A SPECIAL REFERENCES TO SUNGAL TUNNE...
 
STUDY THE EFFECT OF RESPONSE REDUCTION FACTOR ON RC FRAMED STRUCTURE
STUDY THE EFFECT OF RESPONSE REDUCTION FACTOR ON RC FRAMED STRUCTURESTUDY THE EFFECT OF RESPONSE REDUCTION FACTOR ON RC FRAMED STRUCTURE
STUDY THE EFFECT OF RESPONSE REDUCTION FACTOR ON RC FRAMED STRUCTURE
 
A COMPARATIVE ANALYSIS OF RCC ELEMENT OF SLAB WITH STARK STEEL (HYSD STEEL) A...
A COMPARATIVE ANALYSIS OF RCC ELEMENT OF SLAB WITH STARK STEEL (HYSD STEEL) A...A COMPARATIVE ANALYSIS OF RCC ELEMENT OF SLAB WITH STARK STEEL (HYSD STEEL) A...
A COMPARATIVE ANALYSIS OF RCC ELEMENT OF SLAB WITH STARK STEEL (HYSD STEEL) A...
 
Effect of Camber and Angles of Attack on Airfoil Characteristics
Effect of Camber and Angles of Attack on Airfoil CharacteristicsEffect of Camber and Angles of Attack on Airfoil Characteristics
Effect of Camber and Angles of Attack on Airfoil Characteristics
 
A Review on the Progress and Challenges of Aluminum-Based Metal Matrix Compos...
A Review on the Progress and Challenges of Aluminum-Based Metal Matrix Compos...A Review on the Progress and Challenges of Aluminum-Based Metal Matrix Compos...
A Review on the Progress and Challenges of Aluminum-Based Metal Matrix Compos...
 
Dynamic Urban Transit Optimization: A Graph Neural Network Approach for Real-...
Dynamic Urban Transit Optimization: A Graph Neural Network Approach for Real-...Dynamic Urban Transit Optimization: A Graph Neural Network Approach for Real-...
Dynamic Urban Transit Optimization: A Graph Neural Network Approach for Real-...
 
Structural Analysis and Design of Multi-Storey Symmetric and Asymmetric Shape...
Structural Analysis and Design of Multi-Storey Symmetric and Asymmetric Shape...Structural Analysis and Design of Multi-Storey Symmetric and Asymmetric Shape...
Structural Analysis and Design of Multi-Storey Symmetric and Asymmetric Shape...
 
A Review of “Seismic Response of RC Structures Having Plan and Vertical Irreg...
A Review of “Seismic Response of RC Structures Having Plan and Vertical Irreg...A Review of “Seismic Response of RC Structures Having Plan and Vertical Irreg...
A Review of “Seismic Response of RC Structures Having Plan and Vertical Irreg...
 
A REVIEW ON MACHINE LEARNING IN ADAS
A REVIEW ON MACHINE LEARNING IN ADASA REVIEW ON MACHINE LEARNING IN ADAS
A REVIEW ON MACHINE LEARNING IN ADAS
 
Long Term Trend Analysis of Precipitation and Temperature for Asosa district,...
Long Term Trend Analysis of Precipitation and Temperature for Asosa district,...Long Term Trend Analysis of Precipitation and Temperature for Asosa district,...
Long Term Trend Analysis of Precipitation and Temperature for Asosa district,...
 
P.E.B. Framed Structure Design and Analysis Using STAAD Pro
P.E.B. Framed Structure Design and Analysis Using STAAD ProP.E.B. Framed Structure Design and Analysis Using STAAD Pro
P.E.B. Framed Structure Design and Analysis Using STAAD Pro
 
A Review on Innovative Fiber Integration for Enhanced Reinforcement of Concre...
A Review on Innovative Fiber Integration for Enhanced Reinforcement of Concre...A Review on Innovative Fiber Integration for Enhanced Reinforcement of Concre...
A Review on Innovative Fiber Integration for Enhanced Reinforcement of Concre...
 
Survey Paper on Cloud-Based Secured Healthcare System
Survey Paper on Cloud-Based Secured Healthcare SystemSurvey Paper on Cloud-Based Secured Healthcare System
Survey Paper on Cloud-Based Secured Healthcare System
 
Review on studies and research on widening of existing concrete bridges
Review on studies and research on widening of existing concrete bridgesReview on studies and research on widening of existing concrete bridges
Review on studies and research on widening of existing concrete bridges
 
React based fullstack edtech web application
React based fullstack edtech web applicationReact based fullstack edtech web application
React based fullstack edtech web application
 
A Comprehensive Review of Integrating IoT and Blockchain Technologies in the ...
A Comprehensive Review of Integrating IoT and Blockchain Technologies in the ...A Comprehensive Review of Integrating IoT and Blockchain Technologies in the ...
A Comprehensive Review of Integrating IoT and Blockchain Technologies in the ...
 
A REVIEW ON THE PERFORMANCE OF COCONUT FIBRE REINFORCED CONCRETE.
A REVIEW ON THE PERFORMANCE OF COCONUT FIBRE REINFORCED CONCRETE.A REVIEW ON THE PERFORMANCE OF COCONUT FIBRE REINFORCED CONCRETE.
A REVIEW ON THE PERFORMANCE OF COCONUT FIBRE REINFORCED CONCRETE.
 
Optimizing Business Management Process Workflows: The Dynamic Influence of Mi...
Optimizing Business Management Process Workflows: The Dynamic Influence of Mi...Optimizing Business Management Process Workflows: The Dynamic Influence of Mi...
Optimizing Business Management Process Workflows: The Dynamic Influence of Mi...
 
Multistoried and Multi Bay Steel Building Frame by using Seismic Design
Multistoried and Multi Bay Steel Building Frame by using Seismic DesignMultistoried and Multi Bay Steel Building Frame by using Seismic Design
Multistoried and Multi Bay Steel Building Frame by using Seismic Design
 
Cost Optimization of Construction Using Plastic Waste as a Sustainable Constr...
Cost Optimization of Construction Using Plastic Waste as a Sustainable Constr...Cost Optimization of Construction Using Plastic Waste as a Sustainable Constr...
Cost Optimization of Construction Using Plastic Waste as a Sustainable Constr...
 

Recently uploaded

UNIT-V FMM.HYDRAULIC TURBINE - Construction and working
UNIT-V FMM.HYDRAULIC TURBINE - Construction and workingUNIT-V FMM.HYDRAULIC TURBINE - Construction and working
UNIT-V FMM.HYDRAULIC TURBINE - Construction and workingrknatarajan
 
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130Suhani Kapoor
 
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Serviceranjana rawat
 
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...Call Girls in Nagpur High Profile
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCollege Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 
SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )Tsuyoshi Horigome
 
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...ranjana rawat
 
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝soniya singh
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Dr.Costas Sachpazis
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escortsranjana rawat
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
Processing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptxProcessing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptxpranjaldaimarysona
 
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).pptssuser5c9d4b1
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations120cr0395
 
(TARA) Talegaon Dabhade Call Girls Just Call 7001035870 [ Cash on Delivery ] ...
(TARA) Talegaon Dabhade Call Girls Just Call 7001035870 [ Cash on Delivery ] ...(TARA) Talegaon Dabhade Call Girls Just Call 7001035870 [ Cash on Delivery ] ...
(TARA) Talegaon Dabhade Call Girls Just Call 7001035870 [ Cash on Delivery ] ...ranjana rawat
 

Recently uploaded (20)

UNIT-V FMM.HYDRAULIC TURBINE - Construction and working
UNIT-V FMM.HYDRAULIC TURBINE - Construction and workingUNIT-V FMM.HYDRAULIC TURBINE - Construction and working
UNIT-V FMM.HYDRAULIC TURBINE - Construction and working
 
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
VIP Call Girls Service Kondapur Hyderabad Call +91-8250192130
 
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
(RIA) Call Girls Bhosari ( 7001035870 ) HI-Fi Pune Escorts Service
 
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...Booking open Available Pune Call Girls Koregaon Park  6297143586 Call Hot Ind...
Booking open Available Pune Call Girls Koregaon Park 6297143586 Call Hot Ind...
 
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(PRIYA) Rajgurunagar Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service NashikCollege Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
College Call Girls Nashik Nehal 7001305949 Independent Escort Service Nashik
 
SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )
 
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
(SHREYA) Chakan Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Esc...
 
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
Model Call Girl in Narela Delhi reach out to us at 🔝8264348440🔝
 
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
Structural Analysis and Design of Foundations: A Comprehensive Handbook for S...
 
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINEDJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
DJARUM4D - SLOT GACOR ONLINE | SLOT DEMO ONLINE
 
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
(MEERA) Dapodi Call Girls Just Call 7001035870 [ Cash on Delivery ] Pune Escorts
 
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
Processing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptxProcessing & Properties of Floor and Wall Tiles.pptx
Processing & Properties of Floor and Wall Tiles.pptx
 
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
 
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations
 
(TARA) Talegaon Dabhade Call Girls Just Call 7001035870 [ Cash on Delivery ] ...
(TARA) Talegaon Dabhade Call Girls Just Call 7001035870 [ Cash on Delivery ] ...(TARA) Talegaon Dabhade Call Girls Just Call 7001035870 [ Cash on Delivery ] ...
(TARA) Talegaon Dabhade Call Girls Just Call 7001035870 [ Cash on Delivery ] ...
 
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
9953056974 Call Girls In South Ex, Escorts (Delhi) NCR.pdf
 

IRJET- The Multiphysiological Parameter Monitoring System

  • 1. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072 © 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1765 THE MULTIPHYSIOLOGICAL PARAMETER MONITORING SYSTEM Roshinie D1, Swathy S2, Sangeetha V3, Sriba R4 1,2Roshinie D, Student, Department of BME, Dr.N.G.P. Institute of Technology, Tamil Nadu, India -----------------------------------------------------------------------***-------------------------------------------------------------------- Abstract:- The Physiological Monitoring System, which is commonly known as Patient Monitors, in hospital environment, finds a great use in modern era. These type of monitoring systems facilitate the measurement or monitoring of human vital signs such as ECG, Heart rate, Temperature, NIBP and SpO2. These monitors are mainly used in all Intensive Care Units, Critical Care Units, Operation Theatres, Post-operative wards, Emergency wards, etc., In additional to these parameters, there are also some essential parameters to be monitored every instant, in order to eliminate sudden complications and illness. Though there are rising demands in need of patient monitors at almost all departments, there are also some special needs by some departments. The existing monitors measure only these parameters irrespective of the department being used. The monitors which are used in ICUs are similar to where they are used in wards. To design special monitors, some parameters can be added. Here are some of the most required parameters, in which the patient must be regularly monitored: Glucose, pCo2 and Hemoglobin. An additional advantage is that, these parameters are to be measured in Non-Invasive technique, to satisfy patient comfort. Index terms: Non-invasive, Near Infra-red, Micro controller, Glucose, Hemoglobin, pCo2. 1. INTRODUCTION: The existing patient monitoring systems finds various uses in all departments of a hospital, as they possess some good features such as: they are easily portable, they possess less weight, all vitals of a person can be measured less than a minute. The patient monitor measure a person’s ECG, Heart rate, NIBP, Respiration rate, Temperature and SpO2 levels. According to the target parameters, these monitors can be modified and can be used for various purposes. 1.1. Cardiac monitoring Involves the measurement of cardiac related parameters such as ECG, Heart rhythm, Heart rate continuously for a particular period of time, and is capable of recording the results. This kind of equipment is commercially known as ‘Holter monitor’, a small monitor which can be worn by an ambulatory patient. Cardiac output can also be measured by an invasive method, with the help of SwanGanz catheter. 1.2. Hemodynamic monitoring Involves the measurement of blood pressure and blood flow. The blood pressure can be measured by both invasive and non-invasive techniques. In invasive method, a blood pressure transducer assembly is used and in non-invasive method, an inflatable blood pressure cuff is used. 1.3. Respiratory monitoring Involves the measurement of SpO2, ETCO2 and Respiratory rate. The SpO2 can be measured non- invasively via an Infrared sensor. ETCO2 can be measured invasively, which is popularly known as capnography. The Respiratory rate can be measured non-invasively with the help of a thoracic transducer belt. 1.4. Neurological monitoring Involves the measurement of intra-cranial pressure, anaesthetic concentrations, bi-spectral index and most importantly the EEG. The first three measurements are incorporated into anaesthesia machines. A typical neurological unit must contain EEG, as it plays a major role in determining the activities of brain at various states. 3. EXISTING SYSTEM: A general block diagram below, represents the overall functioning of a physiological monitor. Fig.1 Block diagram representing the functioning of a patient monitor. As discussed before, a Patient monitor can measure 5 vital parameters:
  • 2. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072 © 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1766 3.1 ECG: ECG refers to the recording of the electrical activity of heart for a particular period of time, with the help of electrodes/leads placed on skin. The leads are placed according to Einthoven’s principle: Einthoven’s triangle is the imaginary formation of a triangle, in which, the two sides belong to the right and the left shoulders and the third side belong to the pubis. Thus, the shape forms an inverted triangle, with the heart at its centre. Generally, both limb electrodes and surface electrodes are used to measure ECG, but the existing patient monitors use only 6 (V1 to V6) surface electrodes. V1 is placed at the right sternal border of 4th intercoastal space, V2 is placed at the left sternal border of 4th intercoastal space, V3 is placed at the halfway between V2 and V4, V4 is placed at the 5th intercoastal space at the mid-clavicular line, V5 is placed at the left anterior axillary line on the same horizontal plane similar to V4, V6 is placed at the left mid-axillary line on the same horizontal plane as V5. z Fig.2 Lead placement based on Einthoven triangle to record ECG. The ECG wave consists of PQRST peaks. P wave represents atrial depolarization, QRS complex represents ventricular depolarization, T wave represents ventricular repolarization. Fig.3 Normal ECG waveform. 3.2 Heart Rate: Heart rate can also be calculated with the help of ECG waveform, with the help of a formula: Rate = 60/(R-R interval) bpm. The R-R interval is calculated from the ECG waveform. The normal range for heart rate is 60-100bpm. If the heart rate is above 100 bpm, it indicates Tachycardia and if the heart rate is below 60 bpm, it indicates Bradycardia. 3.3 NIBP: Non-invasive Blood Pressure monitoring enables the continuous measurement of blood pressure. High blood pressure does not usually give any warning signs. Though you experience high blood pressure, you might feel perfectly well. The only way to determine whether you have high blood pressure is to have it checked regularly. Non-invasive blood pressure measurement usually involves the use of an inflatable-cuff which is wrapped around the limb of a test subject. The cuff is inflated and deflated at a controlled rate and hence the physical parameters are observed. The auscultatory and oscillometric techniques are well known non-invasive methods. These methods are indirect because they do not couple directly to the artery. Modern NIBP monitors now use oscillometric technique to measure pressure. The oscillometric technique monitors the changes in cuff pressure caused by the flow of blood through the artery. The monitor inflates the cuff to a pressure that occludes the artery. Even when the artery is occluded, the pumping of the heart against the artery can cause small pressure pulses in the cuff baseline pressure. The monitor lowers cuff pressure at a controlled rate. As the cuff pressure decreases, blood starts to flow through the artery. The increasing blood flow causes the amplitude of the pressure pulses in the cuff to increase. These pressure pulses continue to increase in amplitude with decreasing cuff pressure until they reach maximum amplitude, at which point they begin to decrease with decreasing cuff pressure. The cuff pressure at which the pressure pulse amplitude is the greatest is known as Mean Arterial Pressure (MAP). The manner in which the pulse amplitudes vary is often referred to as a pulse envelope. The envelope is an imaginary line that connects the peak of each pressure pulse and forms an outline. The shape of the envelope is observed by different monitors using a variety of techniques to determine the diastolic and systolic blood pressures. The normal range of NIBP is 120/80 mmHg. The top, larger number is called the systolic pressure. This is the pressure generated when the heart contracts. It reflects the pressure of the blood against arterial walls. The bottom, smaller number is called the diastolic pressure. This reflects the pressure in the arteries while the heart is filling and resting between heartbeats.
  • 3. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072 © 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1767 3.4 Pleth or SpO2: SpO2 is measured with the help of pulseoximeter. Pulse oximetry is a simple non-invasive method of monitoring the percentage of haemoglobin (Hb), which is saturated with oxygen. The pulse oximeter consists of a probe attached to the patient’s finger, toe, ear lobe or forehead, which is linked to a computerised unit. The unit displays the percentage of Hb saturated with oxygen together with an audible signal for each pulse beat, and calculated heart rate. A pulse wave related to flow is displayed graphically. Fig.4 Normal pleth waveform. The measurement of SpO2 is done with the help of a probe, which consists of two beams of light: red light of wavelength 600-750 nm and infrared light of wavelength 850-1000 nm. It is purely based on the concept that, oxygenated hemoglobin absorbs more infrared light and allows more red light to pass through. Fig.5 SpO2 measurement probe assembly. The red/infrared ratio is calculated and then compared with the look-up table and finally converted to SpO2 value. The normal SpO2 value ranges from 95-100%. 3.5 Temperature: The body temperature plays a vital role in maintaining the hemostasis, as temperature plays a vital role in biochemical reactions. The normal temperature of a human is 97 to 99 F. The temperature above the normal range indicates Hyperthermia and the temperature below normal range indicates Hypothermia. These parameters make up together into a complete module. Fig. 6 Internal Diagram of a patient monitor. 4. UNSATISFACTORINESS IN THE EXISTING SYSTEM: Though, these monitors provide a continuous measurement of most of the vital parameters, there are also some vitals, which must be measured continuously, in order to avoid risks. A patient’s vitals are measured mostly in critical departments, where the patients are at high risk, such as Intensive Care Units, Critical Care Units, Post-operative wards, Operation theatres, Emergency departments, Maternity wards etc., where, not only the above discussed vitals are considered important. There are also some parameters, which are not continuously measured, but at regular time intervals and also involves invasive method, which leads to patient discomfort and sometimes pain, which could not be tolerated. Glucose, pCO2 and Hemoglobin values are also necessary for proper functioning of the body. Hence it must also be monitored continuously and also by invasive method, for satisfying patient’s comfort. 5. PROPOSED MODEL: In order to solve the above mentioned issues or in other words, to satisfy the physician’s needs, we have considered 3 main parameters, Glucose, pCO2 and Hemoglobin, which can be continuously measured by non-invasive technique. 5.1 WHY THESE PARAMETERS ARE CONSIDERED? Here are some reasons, for considering the above mentioned parameters to be added: 1. Any patient, who is recovering from a surgery, may face an alternating blood glucose level, no
  • 4. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072 © 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1768 matter, whether the patient is a diabetic or non- diabetic. Sometimes, the blood glucose level may reach high or too low. If the glucose values are left unnoticed, it may lead to serious ill effects. In addition to that, the blood glucose level is measured by invasive method, which would be uncomfortable to the patient. Hence, glucose level must be monitored both continuously and non-invasively. 2. A patient, who has undergone a lung surgery, may face with periodic respiratory acidosis or respiratory alkalosis. If these changes are left unnoticed, then it leads to lung failure or even death. To determine the proper functioning of the lungs, pO2, pCO2 and pH must be measured. Here, ABG machine, also known as blood gas machine aids in measuring these parameters. It is a painful and invasive technique, which involves drawing of a small amount of blood from the radial artery or femoral artery and then the values are estimated. This again leads to patient discomfort. Hence, in our proposed model, pCO2 value is estimated both continuously and non-invasively. 3. A patient may face with decreasing hemoglobin levels, after a surgery or after the chemotherapy (for oncologic patients). The decrease in hemoglobin usually lead to fatigue, anaemia and several other complications. The hemoglobin level in blood is also estimated only periodically at regular time intervals and also involves invasive technique. Hence, in our proposed model, the hemoglobin value is estimated both continuously and non-invasively. 5.2 Principle behind the model: The whole module is built based on the principle of Near Infra Red (NIR) Spectroscopic Technique. Fig.7 Electromagnetic spectrum. Near-infrared spectroscopic method uses the near- infrared region of the electromagnetic spectrum which ranges from 780 nm to 2500 nm. Some applications include medical and physiological diagnostics and research including blood sugar, pulse oximetry, functional neuroimaging, sports medicine, elite sports training, ergonomics, rehabilitation, neonatal research, brain computer interface, urology, and neurology. Fig.8 A simple representation of NIR spectroscopy. 5.3 COMPONENTS USED FOR BUILDING UP THE MODULE: The major components required are as follows: 1. Arduino Uno R3 2. Infrared LEDs 3. Quadripple Operational Amplifier (LM324N) 4. Operational Amplifier (TL082) The above mentioned components play a vital role in estimating the proposed parameters. In addition to these, the following components are also required: 5. Phototransistors 6. Photodiode 7. Resistors and Capacitors 8. USB cable 5.3.1 ARDUINO UNO R3: The Arduino is a microcontroller board based on ATmega328 microcontroller. It consists of 14 input/output pins, of which, 6 pins can be used as PWM inputs, 6 analog inputs, a 16MHz crystal oscillator, a USB connection, a power jack, an ICSP header and a reset button. Fig.9 Arduino Uno microcontroller.
  • 5. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072 © 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1769 The Arduino Uno can be powered via the USB connection or with an external power supply. The power source is selected automatically. External power can come either from an AC-to-DC adapter or battery. The board can operate on an external supply of 6 to 20 volts. The recommended range is 7 to 12 volts. Out of the above mentioned 14 pins, some pins have specific functions as listed below: Serial Pins 0 (Rx) and 1 (Tx): Rx and Tx pins are used to receive and transmit TTL serial data. They are connected with the ATmega328P USB to TTL serial chip. External Interrupt Pins 2 and 3: These pins can be configured to trigger an interrupt on a low value, a rising or falling edge, or a change in value. PWM Pins 3, 5, 6, 9 and 11: These pins provide an 8-bit PWM output by using analogWrite() function. SPI Pins: 10 (SS), 11 (MOSI), 12 (MISO) and 13 (SCK): These pins are used for SPI communication. In-built LED Pin 13: This pin is connected with an built- in LED, when pin 13 is HIGH, LED is on and when pin 13 is LOW, it is off. Power pins: 1. VIN refers to the input voltage to the Arduino board when it is using an external power source. 2. 5V: This pin outputs a regulated 5V from the regulator on the board. The board can be supplied with power either from the DC power jack (7 - 12V), the USB connector (5V), or the VIN pin of the board (7-12V). Supplying voltage via the 5V or 3.3V pins bypasses the regulator, and can damage your board. 3. 3.3V: A 3.3 volt supply generated by the on- board regulator. Maximum current draw is 50 mA. 4. GND: Ground pins. 4.3.2 Infrared LEDs: For detection of Glucose, 700-1100nm of light is capable of absorbing glucose molecules in blood. Similarly, for the detection of pCO2, 1200-1500nm of light is capable of absorbing pCO2 present in the blood. The detection of hemoglobin doesn’t need any light source, rather, it can be detected with the help of arduino program. 5.3.3 Quadripple Operational Amplifier: (LM324N) The Quadripple operational amplifier is used here, as a part of the circuit, which estimates the blood glucose value. The LM324 series are low−cost, quad operational amplifiers with true differential inputs. They have several distinct advantages over standard operational amplifier types in single supply applications. The quad amplifier can operate at supply voltages as low as 3.0 V or as high as 32 V. Fig.10 LM324N Features: 1. Internally frequency compensated for unity gain. 2. Large DC voltage gain 100 dB. 3. Wide bandwidth 1 MHz. 4. Wide power supply range: Single supply 3V to 32V. 5. Essentially independent of supply voltage. Fig.11 Pin configuration of LM324N. Pin functions: Pin No. Function 1 Output of first comparator 2 Inverting input of first comparator 3 Non-Inverting input of first comparator 4 Supply voltage 5V 5 Non-Inverting input of second comparator 6 Inverting input of second comparator
  • 6. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072 © 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1770 7 Output of second comparator 8 Output of third comparator 9 Inverting input of third comparator 10 Non-Inverting input of third comparator 11 Ground 12 Non-Inverting input of fourth comparator 13 Inverting input of fourth comparator 14 Output of fourth comparator 5.3.4 Operational Amplifier (TL082): This amplifier is used as a part of the circuit, which estimates pCO2. TL082 is a low cost, high speed, dual JFET input operational amplifiers with an internally trimmed input offset voltage. They require low supply current yet maintain a large gain bandwidth product and fast slew rate. In addition, well matched high voltage JFET input devices provide very low input bias and offset currents. These amplifiers may be used in applications such as high speed integrators, fast D/A converters, sample and hold circuits and many other circuits requiring low input offset voltage, low input bias current, high input impedance, high slew rate and wide bandwidth. The devices also exhibit low noise and offset voltage drift. Fig.12 TL082. Features of TL082: 1. Low Power Consumption: 1.4 mA 2. Wide Common-Mode and Differential Voltage range. 3. Low Input Bias Current: 30 pA 4. Low Input Offset Current: 5 pA 5. Output Short-Circuit Protection. 6. Pin Configuration: Fig.13 TL082 pin configuration. 5.3.5 Photodiode (OPT101): The OPT101 is a monolithic photodiode with on-chip trans-impedance amplifier. Output voltage increases linearly with light intensity. The amplifier is designed for single or dual power-supply operation, making it ideal for batteryoperated equipment. The integrated combination of photodiode and transimpedance amplifier on a single chip eliminates the problems commonly encountered in discrete designs such as leakage current errors, noise pick-up, and gain peaking due to stray capacitance. The 0.09 x 0.09 inch photodiode is operated in the photoconductive mode for excellent linearity and low dark current. The OPT101 operates from +2.7V to +36V supplies and quiescent current is only 120µA. It is available in clear plastic 8-pin DIP, and J-formed DIP for surface mounting. Temperature range is 0°C to +70°C. Fig.14. OPT101 Pin configuration: Fig.15 Pin configuration of OPT101.
  • 7. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072 © 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1771 7. BLOCK DIAGRAM: Fig. 16 Block diagram for proposed system. 7. ESTIMATION OF GLUCOSE: Some of the optical properties of glucose help us to determine that the NIR wavelength of 950 nm is suitable to estimate the blood glucose level. Block diagram for glucose: Fig.17 Block diagram for detection of glucose. The IR source and the detector are hence fixed in the board, and it is covered to eliminate the interference of the environmental light sources. Fig. 18. Assembly of source and detector. The source and detector are connected to the Arduino Uno for power supply and to read the analog inputs. Fig.19 Circuit diagram for estimation of glucose. The amount of IR source, absorbed by the glucose in the blood is determined by the detector, which is sent as an analog signal to the Arduino Uno. The analog signal is then converted into a digital voltage, by means of algorithm. According to Beer Lambert’s law, the absorption of light by the solution is directly proportional to the voltage developed. Fig.20. Experimental setup for the detection of glucose. The normal range of glucose is 100 to 125 mg/dL. 8. ESTIMATION OF PCO2: It is experimentally proven by most of the research scholars that, the CO2 can be absorbed by the near infrared light source of wavelength ranging from 800- 2500 nm. Here, a NIR source of wavelength 800 nm is used to estimate the amount of pCO2 present in blood. The circuit diagram to estimate pCO2 concentration is as follows:
  • 8. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072 © 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1772 Fig. 21. Circuit diagram for the estimation of pCO2. OPT101 plays a major role here to estimate pCO2. The amount of NIR light, absorbed by pCO2 molecules, is detected by means of a photodetector, which converts the absorbed light into its corresponding analog voltage. The analog voltage is then converted into digital voltage, which follows Beer Lambert’s law. The normal range of pCO2 is 38 to 45 mmHg. Fig.22 Experimental setup for the detection of pCO2. 9. ESTIMATION OF HEMOGLOBIN: The hemoglobin can be estimated by means of simple algorithm, which requires the values of pCO2 and SpO2. The common formula to estimate Hemoglobin concentration in blood is: RESULTS: All the 3 proposed parameters are combined together and the final setup is given as follows: Fig.23 Final circuit for the proposed system. It has been checked with our faculties and student members, the results displayed are as follows: The Multiphysiological Parameter Monitoring SystemResult 1.png The Multiphysiological Parameter Monitoring SystemResult 2.png The Multiphysiological Parameter Monitoring SystemResult 3.png 10. CONCLUSION: Thus, among all the invasive parameters, glucose, pCO2 and hemoglobin were chosen, to detect them purely non- invasive and tried to resolve patient discomfort. This will find a greater advantage in all the emergency and critical care departments. The future works will deal with developing this circuit into a complete module and testing among the general public, thus ensuring that, this module will be safe that they can be used in hospitals. 11. REFERENCES: 1. Shyqyri Haxha, Jaspreet Jhoja et.al, OPTICAL BASED NON-INVASIVE GLUCOSE MONITORING SENSOR, IEEE Photonics Journal, Volume: 8 Issue: 6. 2. Carl D Machoff et al., A NOVEL NON-INVASIVE BLOOD GLUCOSE MONITOR, American Diabetes Association in Emerging Treatment and Technology, 2002 Dec; 25(12): 2268-2275. 3. Masab Ahmad et al., NON-INVASIVE BLOOD GLUCOSE MONITORING USING NEAR INFRARED SPECTROSCOPY, EDN Journal, October 16, 2013. 4. N.Joglekar et al., A REVIEW OF BLOOD GAS SENSORS AND A PROPOSED PORTABLE SOLUTION, IEEE Journal, 10.1109/ISICIR.2007.4441843. 5. Jens Kratil et al., NON INVASIVE MEASUREMENT FOR AN INVIVO HEMOGLOBIN MEASUREMENT, IEEE Journal, 10.1109/ICSENS.2011.6126982 6. Md Kamrul Hasan et al., ANALYSING THE EXISTING NON-INVASIVE HEMOGLOBIN
  • 9. International Research Journal of Engineering and Technology (IRJET) e-ISSN: 2395-0056 Volume: 06 Issue: 03 | Mar 2019 www.irjet.net p-ISSN: 2395-0072 © 2019, IRJET | Impact Factor value: 7.211 | ISO 9001:2008 Certified Journal | Page 1773 MEASUREMENT TECHNIQUES, IEEE Journal, 10.1109/UEMCON.2017.8249080. 7. Raid Saleem Al-Baradie et al., PORTABLE SMART NON-INVASIVE HEMOGLOBIN MEASUREMENT SYSTEM, IEEE Journal, 10.1109/SSD.2013.6564149. 8. U. Timm et al., SENSOR SYSTEM FOR NON- INVASIVE OPTICAL HEMOGLOBIN DETERMINATION, IEEE Journal, 10.1109/ICSENS.2009.5398321.