FLIP CROSS CYLINDER
“……far more useful and far more
used” than any other lenses in clinical
refraction.
by- Dr Edward Jackson
…. OBJECTIVE OF PRESENTATION
•To understand the optics and
procedure of Jackson Cross
Cylinder in clinical refraction.
CONTENTS
• Discovery of JCC
• Introduction to JCC
• Optics of JCC
• Detection of Astigmatism by using JCC
• Refinement of cylindrical axis
• Refinement of cylindrical power
• Sources of error
• Summary
• Conclusion
• References
DISCOVERY OF JCC
• The original concept of JCC was first described by Stokes in 1849.
• 1855- The stokes lens was used in variation of the present technique
by Dennet
Crisp brought it to the worldwide attention, and it has become
known as JCC technique.
The present day
technique of JCC that
was used in clinical
practice was designed
and described by
Dr. Edward Jackson for
the determination of
cylinder power in 1887
and for axis in 1907.
INRODUCTION
• The Jackson cross cylinder is a combination of two cylinders of equal strength but with
opposite signs placed with their axis at right angles to each other and mounted in a
handle.
OR
JCC is a spherocylindrical lens having a spherical power component that combined with a
cylinder power component which is twice the power of sphere and of opposite sign.
E.g.- +0.25 DS combined with -0.50 DC. This result in the net power of +0.25 DC in
one axis and -0.25 DC in other axis .
The spherical equivalent of JCC IS 0 . (+0.25 DS/ -0.50 DC @ 90)
• The two principle axis of cross cylinder lens exhibit same
power but of opposite sign.
MARKINGS OF JCC
• In typical JCC the axis of + cylinder is
lie on the – power meridian while the
axis of – cylinder lie on the + cylinder
meridian.
• The Red sign represent the – meridian
while Green sign represent the +
meridian.
• The axis of JCC which is 45 degree
align to the power meridian.Axis of + cylinder
Axis of - cylinderxis
of
PROCEDURE OF JCC
Adjust the cylinder by subjective
refraction
ADJUSTMENT OF SPHERE
• Find the spherical power by using the retinoscopy.
• End point of retinoscopy ins the starting point of JCC
refinement.
• Always remember conoid of strum during correction, so that
circle of least confusion fall on retina.
PRESENCE OF ASTIGMATISM
Place JCC at 90 and 180 degree and flip in front of
eye to detect whether astigmatism is present or
not.
If patient see any change in visual acuity in
either of position than astigmatism is present.
If no change in visual acuity is detected at
either of position than place jcc at 45 and 135
degree to confirm astigmatism is present or
not.
1- Refinement
of Axis
2- Refinement
of Power
Reason- Correct axis can be found in the presence of wrong
power but full correct power cannot be Found in wrong axis.
REFINEMENT OF AXIS
1-
• Place a cylinder lens in a trial frame identify during retinoscopy.
2- • Occlude one eye of patient and ask him to see on the Snellen’s chart.
3-
• Place the JCC in front of eye in such a way that handle of JCC is parallel to the axis of
cylinder lens discovered during retinoscopy.
4-
• Flip the JCC in such a way that both the marking are placed in front of eye.
5-
• Ask the patient is there any change in visual acuity occur by flipping the JCC.
6-
• If the letters on Snellen’s chart are equally clear or equally blurred in either of two position
than the axis that is discovered by retinoscopy is correct.
7-
• If patient recognized any change in visual acuity in either of two position.
8-
• “Minus correcting cylinder should be rotated in the direction of minus cylinder component (
red) of the JCC.
9-
• “Plus correcting cylinder should be rotated in the direction of plus cylinder component ( green
) of the JCC.
10-
• Repeat the procedure until correct axis can be discovered
AXIS MOVEMENT IN JCC
• Del Priore and Guyton
gave the guidelines for
the initial change in
the correcting axis
position relative to the
power of correcting
cylinder while
checking the axis.
Source- William j Benjamin,2006 , Borish refraction, Butterworth Hineman Elsevier 20:818
REFINEMENT OF POWER
1-
•Place the cylinder lens in trial frame that is discovered by retinoscopy.
2-
• Occlude one eye and ask the patient to look at clear line on Snellen's chart with
other eye.
3-
• Place the cylinder lens that is identified by retinoscopy in trial lens on the axis
that is discovered by JCC.
4-
•Place the JCC parallel to power to axis.
5-
•Flip the JCC in front of patient eye and ask the patient to tell about any change in
visual acuity.
6-
•If patient does not recognized any change in visual acuity in either of
position i.e. the power placed in trial frame is correct.
7-
•If patient recognize any change in visual acuity in either of two position
than add cylinder power on the basis of patient exceptance.
8-
• If patient see clear from minus side of JCC than add minus cylinder power
to the number that is discovered by subjective refraction.
9
•If patient see clear from plus side of JCC than add plus cylinder power to
the number that is discovered by subjective refraction.

More Related Content

What's hot

What's hot (20)

Design's of pal's
Design's of pal'sDesign's of pal's
Design's of pal's
 
04 prism
04 prism 04 prism
04 prism
 
amslergrid.pdf
amslergrid.pdfamslergrid.pdf
amslergrid.pdf
 
RAMJI PANDEY keratometry PPT
RAMJI PANDEY keratometry PPTRAMJI PANDEY keratometry PPT
RAMJI PANDEY keratometry PPT
 
Optics of RGP contact lens
Optics of RGP contact lensOptics of RGP contact lens
Optics of RGP contact lens
 
Retinoscopy PART 1
Retinoscopy PART 1Retinoscopy PART 1
Retinoscopy PART 1
 
Subjective refraction
Subjective refractionSubjective refraction
Subjective refraction
 
Keratometry & autorefraction
Keratometry & autorefractionKeratometry & autorefraction
Keratometry & autorefraction
 
Vertex distance and power
Vertex distance and powerVertex distance and power
Vertex distance and power
 
JCC -Jackson Cross Cylinder
JCC -Jackson Cross CylinderJCC -Jackson Cross Cylinder
JCC -Jackson Cross Cylinder
 
Visiontherapy
VisiontherapyVisiontherapy
Visiontherapy
 
IPD Measurement
IPD MeasurementIPD Measurement
IPD Measurement
 
OPTICS OF CONTACT LENSES
OPTICS OF CONTACT LENSESOPTICS OF CONTACT LENSES
OPTICS OF CONTACT LENSES
 
Sturm's Conoid
Sturm's ConoidSturm's Conoid
Sturm's Conoid
 
Frame measurementsand markings
Frame measurementsand markingsFrame measurementsand markings
Frame measurementsand markings
 
Standard alignment
Standard alignmentStandard alignment
Standard alignment
 
Aberrometry
AberrometryAberrometry
Aberrometry
 
DUOCHROM TEST.pptx
DUOCHROM TEST.pptxDUOCHROM TEST.pptx
DUOCHROM TEST.pptx
 
Subjective refraction
Subjective refractionSubjective refraction
Subjective refraction
 
keratometry
 keratometry keratometry
keratometry
 

Similar to Technique of Jackson cross cylinder

Astigmatism, presbyopia and aphakia
Astigmatism, presbyopia and aphakiaAstigmatism, presbyopia and aphakia
Astigmatism, presbyopia and aphakia
tutsimundi
 

Similar to Technique of Jackson cross cylinder (20)

Jcc .pptx
Jcc .pptxJcc .pptx
Jcc .pptx
 
Jackson cross cylinder
Jackson cross cylinderJackson cross cylinder
Jackson cross cylinder
 
Cross Cylinder 4th Yr.pptx
Cross Cylinder 4th Yr.pptxCross Cylinder 4th Yr.pptx
Cross Cylinder 4th Yr.pptx
 
lecture(3) phoropter.ppt
lecture(3) phoropter.pptlecture(3) phoropter.ppt
lecture(3) phoropter.ppt
 
Retinoscopy dr.asif siddiqui
Retinoscopy dr.asif siddiquiRetinoscopy dr.asif siddiqui
Retinoscopy dr.asif siddiqui
 
REFRACTION: OBJECTIVE RETINOSCOPY AND SUBJECTIVE ACCEPTANCE
REFRACTION: OBJECTIVE RETINOSCOPY AND SUBJECTIVE ACCEPTANCEREFRACTION: OBJECTIVE RETINOSCOPY AND SUBJECTIVE ACCEPTANCE
REFRACTION: OBJECTIVE RETINOSCOPY AND SUBJECTIVE ACCEPTANCE
 
Astigmatism, anisometropia, anisekonia.
Astigmatism, anisometropia, anisekonia.Astigmatism, anisometropia, anisekonia.
Astigmatism, anisometropia, anisekonia.
 
Refraction using a phoropter
Refraction using a phoropterRefraction using a phoropter
Refraction using a phoropter
 
14 cross cylinder
14 cross cylinder14 cross cylinder
14 cross cylinder
 
Cylinder prescription guidelines
Cylinder prescription guidelinesCylinder prescription guidelines
Cylinder prescription guidelines
 
Corneal curvature
Corneal curvature Corneal curvature
Corneal curvature
 
subjective verification of refraction
subjective verification of refractionsubjective verification of refraction
subjective verification of refraction
 
Astigmatism, presbyopia and aphakia
Astigmatism, presbyopia and aphakiaAstigmatism, presbyopia and aphakia
Astigmatism, presbyopia and aphakia
 
Retinoscopy.pptx
Retinoscopy.pptxRetinoscopy.pptx
Retinoscopy.pptx
 
Astigmatism, presbyopia and aphakia
Astigmatism, presbyopia and aphakiaAstigmatism, presbyopia and aphakia
Astigmatism, presbyopia and aphakia
 
IOL CALCULATIONS IN SPECIAL CASES
IOL CALCULATIONS IN SPECIAL CASES IOL CALCULATIONS IN SPECIAL CASES
IOL CALCULATIONS IN SPECIAL CASES
 
Keratometry and Dynamic Retinoscopy
Keratometry and Dynamic RetinoscopyKeratometry and Dynamic Retinoscopy
Keratometry and Dynamic Retinoscopy
 
Methods of subjective refraction
Methods of subjective refractionMethods of subjective refraction
Methods of subjective refraction
 
Iol
IolIol
Iol
 
Subjective refraction
Subjective refractionSubjective refraction
Subjective refraction
 

Recently uploaded

Detectability of Solar Panels as a Technosignature
Detectability of Solar Panels as a TechnosignatureDetectability of Solar Panels as a Technosignature
Detectability of Solar Panels as a Technosignature
Sérgio Sacani
 
The importance of continents, oceans and plate tectonics for the evolution of...
The importance of continents, oceans and plate tectonics for the evolution of...The importance of continents, oceans and plate tectonics for the evolution of...
The importance of continents, oceans and plate tectonics for the evolution of...
Sérgio Sacani
 
Aerodynamics. flippatterncn5tm5ttnj6nmnynyppt
Aerodynamics. flippatterncn5tm5ttnj6nmnynypptAerodynamics. flippatterncn5tm5ttnj6nmnynyppt
Aerodynamics. flippatterncn5tm5ttnj6nmnynyppt
sreddyrahul
 
THYROID-PARATHYROID medical surgical nursing
THYROID-PARATHYROID medical surgical nursingTHYROID-PARATHYROID medical surgical nursing
THYROID-PARATHYROID medical surgical nursing
Jocelyn Atis
 
Mammalian Pineal Body Structure and Also Functions
Mammalian Pineal Body Structure and Also FunctionsMammalian Pineal Body Structure and Also Functions
Mammalian Pineal Body Structure and Also Functions
YOGESH DOGRA
 
platelets- lifespan -Clot retraction-disorders.pptx
platelets- lifespan -Clot retraction-disorders.pptxplatelets- lifespan -Clot retraction-disorders.pptx
platelets- lifespan -Clot retraction-disorders.pptx
muralinath2
 
Seminar on Halal AGriculture and Fisheries.pptx
Seminar on Halal AGriculture and Fisheries.pptxSeminar on Halal AGriculture and Fisheries.pptx
Seminar on Halal AGriculture and Fisheries.pptx
RUDYLUMAPINET2
 

Recently uploaded (20)

Hemoglobin metabolism: C Kalyan & E. Muralinath
Hemoglobin metabolism: C Kalyan & E. MuralinathHemoglobin metabolism: C Kalyan & E. Muralinath
Hemoglobin metabolism: C Kalyan & E. Muralinath
 
Hemoglobin metabolism_pathophysiology.pptx
Hemoglobin metabolism_pathophysiology.pptxHemoglobin metabolism_pathophysiology.pptx
Hemoglobin metabolism_pathophysiology.pptx
 
Astronomy Update- Curiosity’s exploration of Mars _ Local Briefs _ leadertele...
Astronomy Update- Curiosity’s exploration of Mars _ Local Briefs _ leadertele...Astronomy Update- Curiosity’s exploration of Mars _ Local Briefs _ leadertele...
Astronomy Update- Curiosity’s exploration of Mars _ Local Briefs _ leadertele...
 
Lab report on liquid viscosity of glycerin
Lab report on liquid viscosity of glycerinLab report on liquid viscosity of glycerin
Lab report on liquid viscosity of glycerin
 
PRESENTATION ABOUT PRINCIPLE OF COSMATIC EVALUATION
PRESENTATION ABOUT PRINCIPLE OF COSMATIC EVALUATIONPRESENTATION ABOUT PRINCIPLE OF COSMATIC EVALUATION
PRESENTATION ABOUT PRINCIPLE OF COSMATIC EVALUATION
 
Microbial Type Culture Collection (MTCC)
Microbial Type Culture Collection (MTCC)Microbial Type Culture Collection (MTCC)
Microbial Type Culture Collection (MTCC)
 
SAMPLING.pptx for analystical chemistry sample techniques
SAMPLING.pptx for analystical chemistry sample techniquesSAMPLING.pptx for analystical chemistry sample techniques
SAMPLING.pptx for analystical chemistry sample techniques
 
Detectability of Solar Panels as a Technosignature
Detectability of Solar Panels as a TechnosignatureDetectability of Solar Panels as a Technosignature
Detectability of Solar Panels as a Technosignature
 
NuGOweek 2024 Ghent - programme - final version
NuGOweek 2024 Ghent - programme - final versionNuGOweek 2024 Ghent - programme - final version
NuGOweek 2024 Ghent - programme - final version
 
The importance of continents, oceans and plate tectonics for the evolution of...
The importance of continents, oceans and plate tectonics for the evolution of...The importance of continents, oceans and plate tectonics for the evolution of...
The importance of continents, oceans and plate tectonics for the evolution of...
 
SCHIZOPHRENIA Disorder/ Brain Disorder.pdf
SCHIZOPHRENIA Disorder/ Brain Disorder.pdfSCHIZOPHRENIA Disorder/ Brain Disorder.pdf
SCHIZOPHRENIA Disorder/ Brain Disorder.pdf
 
electrochemical gas sensors and their uses.pptx
electrochemical gas sensors and their uses.pptxelectrochemical gas sensors and their uses.pptx
electrochemical gas sensors and their uses.pptx
 
Aerodynamics. flippatterncn5tm5ttnj6nmnynyppt
Aerodynamics. flippatterncn5tm5ttnj6nmnynypptAerodynamics. flippatterncn5tm5ttnj6nmnynyppt
Aerodynamics. flippatterncn5tm5ttnj6nmnynyppt
 
THYROID-PARATHYROID medical surgical nursing
THYROID-PARATHYROID medical surgical nursingTHYROID-PARATHYROID medical surgical nursing
THYROID-PARATHYROID medical surgical nursing
 
In silico drugs analogue design: novobiocin analogues.pptx
In silico drugs analogue design: novobiocin analogues.pptxIn silico drugs analogue design: novobiocin analogues.pptx
In silico drugs analogue design: novobiocin analogues.pptx
 
GBSN - Biochemistry (Unit 5) Chemistry of Lipids
GBSN - Biochemistry (Unit 5) Chemistry of LipidsGBSN - Biochemistry (Unit 5) Chemistry of Lipids
GBSN - Biochemistry (Unit 5) Chemistry of Lipids
 
Mammalian Pineal Body Structure and Also Functions
Mammalian Pineal Body Structure and Also FunctionsMammalian Pineal Body Structure and Also Functions
Mammalian Pineal Body Structure and Also Functions
 
platelets- lifespan -Clot retraction-disorders.pptx
platelets- lifespan -Clot retraction-disorders.pptxplatelets- lifespan -Clot retraction-disorders.pptx
platelets- lifespan -Clot retraction-disorders.pptx
 
Seminar on Halal AGriculture and Fisheries.pptx
Seminar on Halal AGriculture and Fisheries.pptxSeminar on Halal AGriculture and Fisheries.pptx
Seminar on Halal AGriculture and Fisheries.pptx
 
INSIGHT Partner Profile: Tampere University
INSIGHT Partner Profile: Tampere UniversityINSIGHT Partner Profile: Tampere University
INSIGHT Partner Profile: Tampere University
 

Technique of Jackson cross cylinder

  • 2. “……far more useful and far more used” than any other lenses in clinical refraction. by- Dr Edward Jackson
  • 3. …. OBJECTIVE OF PRESENTATION •To understand the optics and procedure of Jackson Cross Cylinder in clinical refraction.
  • 4. CONTENTS • Discovery of JCC • Introduction to JCC • Optics of JCC • Detection of Astigmatism by using JCC • Refinement of cylindrical axis • Refinement of cylindrical power • Sources of error • Summary • Conclusion • References
  • 5. DISCOVERY OF JCC • The original concept of JCC was first described by Stokes in 1849. • 1855- The stokes lens was used in variation of the present technique by Dennet Crisp brought it to the worldwide attention, and it has become known as JCC technique.
  • 6. The present day technique of JCC that was used in clinical practice was designed and described by Dr. Edward Jackson for the determination of cylinder power in 1887 and for axis in 1907.
  • 7. INRODUCTION • The Jackson cross cylinder is a combination of two cylinders of equal strength but with opposite signs placed with their axis at right angles to each other and mounted in a handle. OR JCC is a spherocylindrical lens having a spherical power component that combined with a cylinder power component which is twice the power of sphere and of opposite sign. E.g.- +0.25 DS combined with -0.50 DC. This result in the net power of +0.25 DC in one axis and -0.25 DC in other axis . The spherical equivalent of JCC IS 0 . (+0.25 DS/ -0.50 DC @ 90)
  • 8. • The two principle axis of cross cylinder lens exhibit same power but of opposite sign.
  • 9. MARKINGS OF JCC • In typical JCC the axis of + cylinder is lie on the – power meridian while the axis of – cylinder lie on the + cylinder meridian. • The Red sign represent the – meridian while Green sign represent the + meridian. • The axis of JCC which is 45 degree align to the power meridian.Axis of + cylinder Axis of - cylinderxis of
  • 10. PROCEDURE OF JCC Adjust the cylinder by subjective refraction
  • 11. ADJUSTMENT OF SPHERE • Find the spherical power by using the retinoscopy. • End point of retinoscopy ins the starting point of JCC refinement. • Always remember conoid of strum during correction, so that circle of least confusion fall on retina.
  • 12.
  • 13.
  • 14. PRESENCE OF ASTIGMATISM Place JCC at 90 and 180 degree and flip in front of eye to detect whether astigmatism is present or not. If patient see any change in visual acuity in either of position than astigmatism is present. If no change in visual acuity is detected at either of position than place jcc at 45 and 135 degree to confirm astigmatism is present or not.
  • 15. 1- Refinement of Axis 2- Refinement of Power Reason- Correct axis can be found in the presence of wrong power but full correct power cannot be Found in wrong axis.
  • 16. REFINEMENT OF AXIS 1- • Place a cylinder lens in a trial frame identify during retinoscopy. 2- • Occlude one eye of patient and ask him to see on the Snellen’s chart. 3- • Place the JCC in front of eye in such a way that handle of JCC is parallel to the axis of cylinder lens discovered during retinoscopy. 4- • Flip the JCC in such a way that both the marking are placed in front of eye. 5- • Ask the patient is there any change in visual acuity occur by flipping the JCC. 6- • If the letters on Snellen’s chart are equally clear or equally blurred in either of two position than the axis that is discovered by retinoscopy is correct. 7- • If patient recognized any change in visual acuity in either of two position.
  • 17. 8- • “Minus correcting cylinder should be rotated in the direction of minus cylinder component ( red) of the JCC. 9- • “Plus correcting cylinder should be rotated in the direction of plus cylinder component ( green ) of the JCC. 10- • Repeat the procedure until correct axis can be discovered
  • 18. AXIS MOVEMENT IN JCC • Del Priore and Guyton gave the guidelines for the initial change in the correcting axis position relative to the power of correcting cylinder while checking the axis. Source- William j Benjamin,2006 , Borish refraction, Butterworth Hineman Elsevier 20:818
  • 19. REFINEMENT OF POWER 1- •Place the cylinder lens in trial frame that is discovered by retinoscopy. 2- • Occlude one eye and ask the patient to look at clear line on Snellen's chart with other eye. 3- • Place the cylinder lens that is identified by retinoscopy in trial lens on the axis that is discovered by JCC. 4- •Place the JCC parallel to power to axis. 5- •Flip the JCC in front of patient eye and ask the patient to tell about any change in visual acuity.
  • 20. 6- •If patient does not recognized any change in visual acuity in either of position i.e. the power placed in trial frame is correct. 7- •If patient recognize any change in visual acuity in either of two position than add cylinder power on the basis of patient exceptance. 8- • If patient see clear from minus side of JCC than add minus cylinder power to the number that is discovered by subjective refraction. 9 •If patient see clear from plus side of JCC than add plus cylinder power to the number that is discovered by subjective refraction.