SlideShare a Scribd company logo
A Minor Project Report On
“REAL TIME PULSE DETECTOR”
by: SHATRUGHAN PATEL
DEPARTMENT OF ELECTRICAL & ELECTRONICS ENGINEERING
BRANCH: ELECTRONICS & COMMUNICATIONS
INTRODUCATION
• The heart is the organ that responsible for pumping blood throughout
the body. It is located in the middle of the thorax, slightly offset to the
left and surrounded by the lungs basically; the human heart is
composed of four chambers which are two atriums and two
ventricles. The right atrium receives blood returning to the heart from
the whole body. That blood passes through the right ventricle and is
pumped to the lungs where it is oxygenated and goes back to the
heart through the left atrium, and then the blood passes through the
left ventricle and is pumped again to be distributed to the entire body
through the arteries
MATERIAL AND TOOLS
• The fingertip sensor - The heart rate monitor builds to test the value of users’ heart rate
currently. The circuit will base on the non-invasive PPG sensor which detects the variation of the
blood flow in finger when the mechanical contraction of heart. The sensor contains an infrared
LED as an IR transmitter and photodiode as an IR receiver.
• ATMEGA32 Microcontroller - The high-performance, low-power Atmel 8-bit AVR RISC-based
microcontroller combines 16KB of programmable flash memory, 2KB SRAM, 1KB EEPROM, an
8channel 10-bit A/D converter, 2 capture/compare/PWM functions, 3 timers, a synchronous serial
port that can be configured as either 3-wire SPI or 2-wire I2C bus, a USART, and a Parallel Slave
Port. The device supports throughput of 16 MIPS at 16 MHz and operates between 4.5-5.5 volts
ATMEGA32 Microcontroller PIN DIAGRAM -
Fingertip sensor - Use of light to measure heart rate is a field of study where abundant research has been
done in the past few decades. Fingertip sensor relies on measurement of a physiological signal called
Photoplethysmogrpahy (PPG) [14.], which is an optical measurement of the change in blood volume in
the arteries. Fingertip sensor acquires PPG signals by irradiating wavelength of light through the tissue,
and compares the light absorption characteristics of blood under these wavelengths
High level design -
By high level design we mean to see the project from an
overall view point, then to get down and consider the
project as a system of main parts.
Fingertip sensor is used in the device as transducer converts
the physical PPG signal to into an electrical signal. This
electrical signal is applied to the microcontroller which is
the main brain of the heart rate monitor. The
microcontroller, according to the
CIRCUIT
DIAGRAM
• This project used LM358 two-class amplifier to adjust the voltage value to 5V for being received by ATMEGA32.
Because the signal acquisition is extremely weak and vulnerable to interference by the external circuit, high-pass
filter and low-pass filter are used for normal heart beat signals.
• The amplifier (see figure 4.9) uses an LM358 dual op amp to provide two identical broadly-tuned low pass stages
with the gain of each filter stage is set to 101 and a cut-off frequency of about 2.34 Hz. So the maximum HR that
could be read is approximately equal 150. The equations for calculating gain and cut-off frequency of the active
low pass filter are expressed as
LCD
• In this project, a 16×2 LCD display will be used for displaying the value of heart rate.
There are 16 pins in whole in LCD. The pin diagram of LCD is shown as below. From the
datasheet, it is known that pin 1 & pin 2 and pin 15 & pin 16 both functions as power
supply ports. However, just pin 1 & pin 2 needed, and the lightness of LCD should be
controlled by a 10kΩ variable resistor.
RESULT
gender Age HR on display HR on oscilloscope Error rate (%)
Male 22 97 96 1.03
Male 22 83 81 2.41
Male 20 78 78 0
Male 22 90 87 3.33
Male 20 80 79 1.25
Female 22 77 77 0
Female 22 104 103 0.96
Female 19 75 75 0
Female 20 69 71 2.81
The accuracy of the device depends on the testing performed is about
1.414. In general, the HR of females is higher than males. But this is not
true where difference in weight, daily life (i.e. kind of work, health,
inheritance of heart disease, doing exercises… etc).
CONCLUSION
Infrared sensor instead of pressure sensor was used in Heart Rate
Monitor, it has higher sensitivity, and the output signal is more stability.
The infrared sensor depends on the fact that the speed of blood is
proportional to heart rate. That is to say, when people are in high active,
the heart rate will change. As the same theory as previous research, heart
rate changes when people are excited. The HR value displayed on LCD
shows the information that when people took exercise, the heart rate will
increase; sometimes it is out of healthy range. When people take a break,
the normal heart rate will recover.
In totally, heart rate is physiological indicators to show what is going on in body.
People need to care about their healthy condition with the help of Heart Rate Monitor.
References
[1] Sokwoo Rhee “DESIGN AND ANALYSIS ARTIFACTRESISTIVEFINGER
PHOTOPLETHYSMOGRAPHIC SENSORS FOR VITAL SIGN MONITORING”
Massachusetts,U.S, 2001, chapter1,P9.
[2] World health organization, Mortality country fact sheet 2006, Sudan. .
WEB URL-
[3] Wikipedia – http://en.wikipedia.org/wiki/Main_Page.
THE END
THANK YOU

More Related Content

What's hot

IRJET- The Multiphysiological Parameter Monitoring System
IRJET-  	  The Multiphysiological Parameter Monitoring SystemIRJET-  	  The Multiphysiological Parameter Monitoring System
IRJET- The Multiphysiological Parameter Monitoring System
IRJET Journal
 
Heart rate monitoring system using 8051
 Heart rate monitoring system using 8051 Heart rate monitoring system using 8051
Heart rate monitoring system using 8051
Eldhose George
 
Cardiac pacemaker
Cardiac pacemakerCardiac pacemaker
Cardiac pacemaker
priyajpr
 
Medical Instrumentation- body parameters measurement
Medical Instrumentation- body parameters measurementMedical Instrumentation- body parameters measurement
Medical Instrumentation- body parameters measurement
Poornima D
 
Holter,ilr,elr
Holter,ilr,elrHolter,ilr,elr
Holter,ilr,elr
Shriya Shankar
 
Pacemaker | Implantable Cardiac Devices For Heart Failures
Pacemaker | Implantable Cardiac Devices For Heart FailuresPacemaker | Implantable Cardiac Devices For Heart Failures
Pacemaker | Implantable Cardiac Devices For Heart Failures
YashodaHospitals
 
Cardiac monitors - Medical Equipment
Cardiac monitors - Medical EquipmentCardiac monitors - Medical Equipment
Cardiac monitors - Medical Equipment
AtheenaPandian Enterprises
 
Pacemaker
Pacemaker Pacemaker
Pacemaker
MahmutKAMALAK
 
photoplethysmogram
photoplethysmogram photoplethysmogram
photoplethysmogram
SIVAKUMAR745527
 
Pacemaker
PacemakerPacemaker
Pacemaker
Pallavi Lokhande
 
Holter monitoring
Holter monitoringHolter monitoring
Holter monitoring
JeroneTirados
 
Ecg123 (1)
Ecg123 (1)Ecg123 (1)
Ecg123 (1)
Himanshu Gawande
 
The pacemaker
The pacemakerThe pacemaker
The pacemaker
Mahmoud Ebada
 
ECG
ECGECG
Senior Project Students' Presentation on ECG Monitoring
Senior Project Students' Presentation on ECG MonitoringSenior Project Students' Presentation on ECG Monitoring
Senior Project Students' Presentation on ECG Monitoring
Md Kafiul Islam
 
Electrocardiogram
ElectrocardiogramElectrocardiogram
ElectrocardiogramTarek Selim
 
Pace maker professor md toufiqur rahman
Pace maker professor md  toufiqur rahman Pace maker professor md  toufiqur rahman
Pace maker professor md toufiqur rahman
PROFESSOR DR. MD. TOUFIQUR RAHMAN
 
Hawzhin
HawzhinHawzhin
Hawzhin
naz9386
 
Medical Instrumentation- EMG,ERG,EOG,PCG
Medical Instrumentation- EMG,ERG,EOG,PCGMedical Instrumentation- EMG,ERG,EOG,PCG
Medical Instrumentation- EMG,ERG,EOG,PCG
Poornima D
 

What's hot (20)

IRJET- The Multiphysiological Parameter Monitoring System
IRJET-  	  The Multiphysiological Parameter Monitoring SystemIRJET-  	  The Multiphysiological Parameter Monitoring System
IRJET- The Multiphysiological Parameter Monitoring System
 
Heart rate monitoring system using 8051
 Heart rate monitoring system using 8051 Heart rate monitoring system using 8051
Heart rate monitoring system using 8051
 
Cardiac pacemaker
Cardiac pacemakerCardiac pacemaker
Cardiac pacemaker
 
Medical Instrumentation- body parameters measurement
Medical Instrumentation- body parameters measurementMedical Instrumentation- body parameters measurement
Medical Instrumentation- body parameters measurement
 
Holter,ilr,elr
Holter,ilr,elrHolter,ilr,elr
Holter,ilr,elr
 
Pacemaker | Implantable Cardiac Devices For Heart Failures
Pacemaker | Implantable Cardiac Devices For Heart FailuresPacemaker | Implantable Cardiac Devices For Heart Failures
Pacemaker | Implantable Cardiac Devices For Heart Failures
 
Cardiac monitors - Medical Equipment
Cardiac monitors - Medical EquipmentCardiac monitors - Medical Equipment
Cardiac monitors - Medical Equipment
 
Pacemaker
Pacemaker Pacemaker
Pacemaker
 
photoplethysmogram
photoplethysmogram photoplethysmogram
photoplethysmogram
 
Pacemaker
PacemakerPacemaker
Pacemaker
 
Holter monitoring
Holter monitoringHolter monitoring
Holter monitoring
 
Ecg123 (1)
Ecg123 (1)Ecg123 (1)
Ecg123 (1)
 
The pacemaker
The pacemakerThe pacemaker
The pacemaker
 
ECG
ECGECG
ECG
 
Senior Project Students' Presentation on ECG Monitoring
Senior Project Students' Presentation on ECG MonitoringSenior Project Students' Presentation on ECG Monitoring
Senior Project Students' Presentation on ECG Monitoring
 
Electrocardiogram
ElectrocardiogramElectrocardiogram
Electrocardiogram
 
Pace maker professor md toufiqur rahman
Pace maker professor md  toufiqur rahman Pace maker professor md  toufiqur rahman
Pace maker professor md toufiqur rahman
 
Hawzhin
HawzhinHawzhin
Hawzhin
 
Cardiac Pacemaker
Cardiac PacemakerCardiac Pacemaker
Cardiac Pacemaker
 
Medical Instrumentation- EMG,ERG,EOG,PCG
Medical Instrumentation- EMG,ERG,EOG,PCGMedical Instrumentation- EMG,ERG,EOG,PCG
Medical Instrumentation- EMG,ERG,EOG,PCG
 

Similar to Pulse detector

micro controller based heart rate monitoring system
micro controller based heart rate monitoring systemmicro controller based heart rate monitoring system
micro controller based heart rate monitoring system
Eldhose George
 
Heart Attack Detection VIA HEART Beat Monitor System
Heart Attack Detection VIA HEART Beat Monitor System Heart Attack Detection VIA HEART Beat Monitor System
Heart Attack Detection VIA HEART Beat Monitor System
Utkarsh Lokpriye
 
Electronics Equipments Used in Hospital
Electronics Equipments Used in HospitalElectronics Equipments Used in Hospital
Electronics Equipments Used in Hospital
Alok Singh
 
Arduino based heartbeat monitoring system.
Arduino based heartbeat monitoring system.Arduino based heartbeat monitoring system.
Arduino based heartbeat monitoring system.
Arkadeep Dey
 
Patient monitoring system to monitor physiological parameters.pptx
Patient  monitoring system to monitor physiological parameters.pptxPatient  monitoring system to monitor physiological parameters.pptx
Patient monitoring system to monitor physiological parameters.pptx
RevathiJ10
 
DIAGNOSIS OF BRADYCARDIA ARRHYTHMIA USING MEMD AND CONVOLUTIONAL NEURAL NETWORKS
DIAGNOSIS OF BRADYCARDIA ARRHYTHMIA USING MEMD AND CONVOLUTIONAL NEURAL NETWORKSDIAGNOSIS OF BRADYCARDIA ARRHYTHMIA USING MEMD AND CONVOLUTIONAL NEURAL NETWORKS
DIAGNOSIS OF BRADYCARDIA ARRHYTHMIA USING MEMD AND CONVOLUTIONAL NEURAL NETWORKS
IRJET Journal
 
Heartbeat monitor USING IC PIC16F72
Heartbeat monitor USING IC PIC16F72Heartbeat monitor USING IC PIC16F72
Heartbeat monitor USING IC PIC16F72
Girish Kumar
 
Non invasive
Non invasiveNon invasive
Non invasive
Amarjith Valanchery
 
Measurement of Multiparameters using Anaesthesia Injector Based on Arm Processor
Measurement of Multiparameters using Anaesthesia Injector Based on Arm ProcessorMeasurement of Multiparameters using Anaesthesia Injector Based on Arm Processor
Measurement of Multiparameters using Anaesthesia Injector Based on Arm Processor
ijsrd.com
 
Report on Automatic Heart Rate monitoring using Arduino Uno
Report on Automatic Heart Rate monitoring using Arduino UnoReport on Automatic Heart Rate monitoring using Arduino Uno
Report on Automatic Heart Rate monitoring using Arduino Uno
Ashfaqul Haque John
 
Heart beat monitoring system
Heart beat monitoring systemHeart beat monitoring system
Heart beat monitoring system
Vipul Tripathi
 
Cardiac monitoring.pdf
Cardiac monitoring.pdfCardiac monitoring.pdf
PPG, ECG and Blood Pressure Circuitry
PPG, ECG and Blood Pressure CircuitryPPG, ECG and Blood Pressure Circuitry
PPG, ECG and Blood Pressure Circuitry
mgoutham kumarreddy
 
Case study
Case studyCase study
Case study
Mrinal Das
 
Iaetsd emergency recovery control unit using microcontroller
Iaetsd emergency recovery control unit using microcontrollerIaetsd emergency recovery control unit using microcontroller
Iaetsd emergency recovery control unit using microcontroller
Iaetsd Iaetsd
 
Evaluating ECG Capturing Using Sound-Card of PC/Laptop
Evaluating ECG Capturing Using Sound-Card of PC/LaptopEvaluating ECG Capturing Using Sound-Card of PC/Laptop
Evaluating ECG Capturing Using Sound-Card of PC/Laptop
ijics
 
Android based Fail Safe Dual Chamber Cardiac Integrated Pacemaker Device usin...
Android based Fail Safe Dual Chamber Cardiac Integrated Pacemaker Device usin...Android based Fail Safe Dual Chamber Cardiac Integrated Pacemaker Device usin...
Android based Fail Safe Dual Chamber Cardiac Integrated Pacemaker Device usin...
Editor IJMTER
 
Merged document 12
Merged document 12Merged document 12
Merged document 12
V!vEk@nAnD S
 
4. 7770 8115-1-pb
4. 7770 8115-1-pb4. 7770 8115-1-pb
4. 7770 8115-1-pb
IAESIJEECS
 

Similar to Pulse detector (20)

micro controller based heart rate monitoring system
micro controller based heart rate monitoring systemmicro controller based heart rate monitoring system
micro controller based heart rate monitoring system
 
Heart Attack Detection VIA HEART Beat Monitor System
Heart Attack Detection VIA HEART Beat Monitor System Heart Attack Detection VIA HEART Beat Monitor System
Heart Attack Detection VIA HEART Beat Monitor System
 
Electronics Equipments Used in Hospital
Electronics Equipments Used in HospitalElectronics Equipments Used in Hospital
Electronics Equipments Used in Hospital
 
Arduino based heartbeat monitoring system.
Arduino based heartbeat monitoring system.Arduino based heartbeat monitoring system.
Arduino based heartbeat monitoring system.
 
heartbeatsensor
heartbeatsensorheartbeatsensor
heartbeatsensor
 
Patient monitoring system to monitor physiological parameters.pptx
Patient  monitoring system to monitor physiological parameters.pptxPatient  monitoring system to monitor physiological parameters.pptx
Patient monitoring system to monitor physiological parameters.pptx
 
DIAGNOSIS OF BRADYCARDIA ARRHYTHMIA USING MEMD AND CONVOLUTIONAL NEURAL NETWORKS
DIAGNOSIS OF BRADYCARDIA ARRHYTHMIA USING MEMD AND CONVOLUTIONAL NEURAL NETWORKSDIAGNOSIS OF BRADYCARDIA ARRHYTHMIA USING MEMD AND CONVOLUTIONAL NEURAL NETWORKS
DIAGNOSIS OF BRADYCARDIA ARRHYTHMIA USING MEMD AND CONVOLUTIONAL NEURAL NETWORKS
 
Heartbeat monitor USING IC PIC16F72
Heartbeat monitor USING IC PIC16F72Heartbeat monitor USING IC PIC16F72
Heartbeat monitor USING IC PIC16F72
 
Non invasive
Non invasiveNon invasive
Non invasive
 
Measurement of Multiparameters using Anaesthesia Injector Based on Arm Processor
Measurement of Multiparameters using Anaesthesia Injector Based on Arm ProcessorMeasurement of Multiparameters using Anaesthesia Injector Based on Arm Processor
Measurement of Multiparameters using Anaesthesia Injector Based on Arm Processor
 
Report on Automatic Heart Rate monitoring using Arduino Uno
Report on Automatic Heart Rate monitoring using Arduino UnoReport on Automatic Heart Rate monitoring using Arduino Uno
Report on Automatic Heart Rate monitoring using Arduino Uno
 
Heart beat monitoring system
Heart beat monitoring systemHeart beat monitoring system
Heart beat monitoring system
 
Cardiac monitoring.pdf
Cardiac monitoring.pdfCardiac monitoring.pdf
Cardiac monitoring.pdf
 
PPG, ECG and Blood Pressure Circuitry
PPG, ECG and Blood Pressure CircuitryPPG, ECG and Blood Pressure Circuitry
PPG, ECG and Blood Pressure Circuitry
 
Case study
Case studyCase study
Case study
 
Iaetsd emergency recovery control unit using microcontroller
Iaetsd emergency recovery control unit using microcontrollerIaetsd emergency recovery control unit using microcontroller
Iaetsd emergency recovery control unit using microcontroller
 
Evaluating ECG Capturing Using Sound-Card of PC/Laptop
Evaluating ECG Capturing Using Sound-Card of PC/LaptopEvaluating ECG Capturing Using Sound-Card of PC/Laptop
Evaluating ECG Capturing Using Sound-Card of PC/Laptop
 
Android based Fail Safe Dual Chamber Cardiac Integrated Pacemaker Device usin...
Android based Fail Safe Dual Chamber Cardiac Integrated Pacemaker Device usin...Android based Fail Safe Dual Chamber Cardiac Integrated Pacemaker Device usin...
Android based Fail Safe Dual Chamber Cardiac Integrated Pacemaker Device usin...
 
Merged document 12
Merged document 12Merged document 12
Merged document 12
 
4. 7770 8115-1-pb
4. 7770 8115-1-pb4. 7770 8115-1-pb
4. 7770 8115-1-pb
 

Recently uploaded

Nuclear Power Economics and Structuring 2024
Nuclear Power Economics and Structuring 2024Nuclear Power Economics and Structuring 2024
Nuclear Power Economics and Structuring 2024
Massimo Talia
 
Water Industry Process Automation and Control Monthly - May 2024.pdf
Water Industry Process Automation and Control Monthly - May 2024.pdfWater Industry Process Automation and Control Monthly - May 2024.pdf
Water Industry Process Automation and Control Monthly - May 2024.pdf
Water Industry Process Automation & Control
 
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdfAKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
SamSarthak3
 
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
AJAYKUMARPUND1
 
The Role of Electrical and Electronics Engineers in IOT Technology.pdf
The Role of Electrical and Electronics Engineers in IOT Technology.pdfThe Role of Electrical and Electronics Engineers in IOT Technology.pdf
The Role of Electrical and Electronics Engineers in IOT Technology.pdf
Nettur Technical Training Foundation
 
Water billing management system project report.pdf
Water billing management system project report.pdfWater billing management system project report.pdf
Water billing management system project report.pdf
Kamal Acharya
 
Harnessing WebAssembly for Real-time Stateless Streaming Pipelines
Harnessing WebAssembly for Real-time Stateless Streaming PipelinesHarnessing WebAssembly for Real-time Stateless Streaming Pipelines
Harnessing WebAssembly for Real-time Stateless Streaming Pipelines
Christina Lin
 
CW RADAR, FMCW RADAR, FMCW ALTIMETER, AND THEIR PARAMETERS
CW RADAR, FMCW RADAR, FMCW ALTIMETER, AND THEIR PARAMETERSCW RADAR, FMCW RADAR, FMCW ALTIMETER, AND THEIR PARAMETERS
CW RADAR, FMCW RADAR, FMCW ALTIMETER, AND THEIR PARAMETERS
veerababupersonal22
 
一比一原版(SFU毕业证)西蒙菲莎大学毕业证成绩单如何办理
一比一原版(SFU毕业证)西蒙菲莎大学毕业证成绩单如何办理一比一原版(SFU毕业证)西蒙菲莎大学毕业证成绩单如何办理
一比一原版(SFU毕业证)西蒙菲莎大学毕业证成绩单如何办理
bakpo1
 
Immunizing Image Classifiers Against Localized Adversary Attacks
Immunizing Image Classifiers Against Localized Adversary AttacksImmunizing Image Classifiers Against Localized Adversary Attacks
Immunizing Image Classifiers Against Localized Adversary Attacks
gerogepatton
 
在线办理(ANU毕业证书)澳洲国立大学毕业证录取通知书一模一样
在线办理(ANU毕业证书)澳洲国立大学毕业证录取通知书一模一样在线办理(ANU毕业证书)澳洲国立大学毕业证录取通知书一模一样
在线办理(ANU毕业证书)澳洲国立大学毕业证录取通知书一模一样
obonagu
 
MCQ Soil mechanics questions (Soil shear strength).pdf
MCQ Soil mechanics questions (Soil shear strength).pdfMCQ Soil mechanics questions (Soil shear strength).pdf
MCQ Soil mechanics questions (Soil shear strength).pdf
Osamah Alsalih
 
Gen AI Study Jams _ For the GDSC Leads in India.pdf
Gen AI Study Jams _ For the GDSC Leads in India.pdfGen AI Study Jams _ For the GDSC Leads in India.pdf
Gen AI Study Jams _ For the GDSC Leads in India.pdf
gdsczhcet
 
Steel & Timber Design according to British Standard
Steel & Timber Design according to British StandardSteel & Timber Design according to British Standard
Steel & Timber Design according to British Standard
AkolbilaEmmanuel1
 
一比一原版(IIT毕业证)伊利诺伊理工大学毕业证成绩单专业办理
一比一原版(IIT毕业证)伊利诺伊理工大学毕业证成绩单专业办理一比一原版(IIT毕业证)伊利诺伊理工大学毕业证成绩单专业办理
一比一原版(IIT毕业证)伊利诺伊理工大学毕业证成绩单专业办理
zwunae
 
Hierarchical Digital Twin of a Naval Power System
Hierarchical Digital Twin of a Naval Power SystemHierarchical Digital Twin of a Naval Power System
Hierarchical Digital Twin of a Naval Power System
Kerry Sado
 
Investor-Presentation-Q1FY2024 investor presentation document.pptx
Investor-Presentation-Q1FY2024 investor presentation document.pptxInvestor-Presentation-Q1FY2024 investor presentation document.pptx
Investor-Presentation-Q1FY2024 investor presentation document.pptx
AmarGB2
 
14 Template Contractual Notice - EOT Application
14 Template Contractual Notice - EOT Application14 Template Contractual Notice - EOT Application
14 Template Contractual Notice - EOT Application
SyedAbiiAzazi1
 
Tutorial for 16S rRNA Gene Analysis with QIIME2.pdf
Tutorial for 16S rRNA Gene Analysis with QIIME2.pdfTutorial for 16S rRNA Gene Analysis with QIIME2.pdf
Tutorial for 16S rRNA Gene Analysis with QIIME2.pdf
aqil azizi
 
Building Electrical System Design & Installation
Building Electrical System Design & InstallationBuilding Electrical System Design & Installation
Building Electrical System Design & Installation
symbo111
 

Recently uploaded (20)

Nuclear Power Economics and Structuring 2024
Nuclear Power Economics and Structuring 2024Nuclear Power Economics and Structuring 2024
Nuclear Power Economics and Structuring 2024
 
Water Industry Process Automation and Control Monthly - May 2024.pdf
Water Industry Process Automation and Control Monthly - May 2024.pdfWater Industry Process Automation and Control Monthly - May 2024.pdf
Water Industry Process Automation and Control Monthly - May 2024.pdf
 
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdfAKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
 
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
Pile Foundation by Venkatesh Taduvai (Sub Geotechnical Engineering II)-conver...
 
The Role of Electrical and Electronics Engineers in IOT Technology.pdf
The Role of Electrical and Electronics Engineers in IOT Technology.pdfThe Role of Electrical and Electronics Engineers in IOT Technology.pdf
The Role of Electrical and Electronics Engineers in IOT Technology.pdf
 
Water billing management system project report.pdf
Water billing management system project report.pdfWater billing management system project report.pdf
Water billing management system project report.pdf
 
Harnessing WebAssembly for Real-time Stateless Streaming Pipelines
Harnessing WebAssembly for Real-time Stateless Streaming PipelinesHarnessing WebAssembly for Real-time Stateless Streaming Pipelines
Harnessing WebAssembly for Real-time Stateless Streaming Pipelines
 
CW RADAR, FMCW RADAR, FMCW ALTIMETER, AND THEIR PARAMETERS
CW RADAR, FMCW RADAR, FMCW ALTIMETER, AND THEIR PARAMETERSCW RADAR, FMCW RADAR, FMCW ALTIMETER, AND THEIR PARAMETERS
CW RADAR, FMCW RADAR, FMCW ALTIMETER, AND THEIR PARAMETERS
 
一比一原版(SFU毕业证)西蒙菲莎大学毕业证成绩单如何办理
一比一原版(SFU毕业证)西蒙菲莎大学毕业证成绩单如何办理一比一原版(SFU毕业证)西蒙菲莎大学毕业证成绩单如何办理
一比一原版(SFU毕业证)西蒙菲莎大学毕业证成绩单如何办理
 
Immunizing Image Classifiers Against Localized Adversary Attacks
Immunizing Image Classifiers Against Localized Adversary AttacksImmunizing Image Classifiers Against Localized Adversary Attacks
Immunizing Image Classifiers Against Localized Adversary Attacks
 
在线办理(ANU毕业证书)澳洲国立大学毕业证录取通知书一模一样
在线办理(ANU毕业证书)澳洲国立大学毕业证录取通知书一模一样在线办理(ANU毕业证书)澳洲国立大学毕业证录取通知书一模一样
在线办理(ANU毕业证书)澳洲国立大学毕业证录取通知书一模一样
 
MCQ Soil mechanics questions (Soil shear strength).pdf
MCQ Soil mechanics questions (Soil shear strength).pdfMCQ Soil mechanics questions (Soil shear strength).pdf
MCQ Soil mechanics questions (Soil shear strength).pdf
 
Gen AI Study Jams _ For the GDSC Leads in India.pdf
Gen AI Study Jams _ For the GDSC Leads in India.pdfGen AI Study Jams _ For the GDSC Leads in India.pdf
Gen AI Study Jams _ For the GDSC Leads in India.pdf
 
Steel & Timber Design according to British Standard
Steel & Timber Design according to British StandardSteel & Timber Design according to British Standard
Steel & Timber Design according to British Standard
 
一比一原版(IIT毕业证)伊利诺伊理工大学毕业证成绩单专业办理
一比一原版(IIT毕业证)伊利诺伊理工大学毕业证成绩单专业办理一比一原版(IIT毕业证)伊利诺伊理工大学毕业证成绩单专业办理
一比一原版(IIT毕业证)伊利诺伊理工大学毕业证成绩单专业办理
 
Hierarchical Digital Twin of a Naval Power System
Hierarchical Digital Twin of a Naval Power SystemHierarchical Digital Twin of a Naval Power System
Hierarchical Digital Twin of a Naval Power System
 
Investor-Presentation-Q1FY2024 investor presentation document.pptx
Investor-Presentation-Q1FY2024 investor presentation document.pptxInvestor-Presentation-Q1FY2024 investor presentation document.pptx
Investor-Presentation-Q1FY2024 investor presentation document.pptx
 
14 Template Contractual Notice - EOT Application
14 Template Contractual Notice - EOT Application14 Template Contractual Notice - EOT Application
14 Template Contractual Notice - EOT Application
 
Tutorial for 16S rRNA Gene Analysis with QIIME2.pdf
Tutorial for 16S rRNA Gene Analysis with QIIME2.pdfTutorial for 16S rRNA Gene Analysis with QIIME2.pdf
Tutorial for 16S rRNA Gene Analysis with QIIME2.pdf
 
Building Electrical System Design & Installation
Building Electrical System Design & InstallationBuilding Electrical System Design & Installation
Building Electrical System Design & Installation
 

Pulse detector

  • 1. A Minor Project Report On “REAL TIME PULSE DETECTOR” by: SHATRUGHAN PATEL DEPARTMENT OF ELECTRICAL & ELECTRONICS ENGINEERING BRANCH: ELECTRONICS & COMMUNICATIONS
  • 2. INTRODUCATION • The heart is the organ that responsible for pumping blood throughout the body. It is located in the middle of the thorax, slightly offset to the left and surrounded by the lungs basically; the human heart is composed of four chambers which are two atriums and two ventricles. The right atrium receives blood returning to the heart from the whole body. That blood passes through the right ventricle and is pumped to the lungs where it is oxygenated and goes back to the heart through the left atrium, and then the blood passes through the left ventricle and is pumped again to be distributed to the entire body through the arteries
  • 3. MATERIAL AND TOOLS • The fingertip sensor - The heart rate monitor builds to test the value of users’ heart rate currently. The circuit will base on the non-invasive PPG sensor which detects the variation of the blood flow in finger when the mechanical contraction of heart. The sensor contains an infrared LED as an IR transmitter and photodiode as an IR receiver. • ATMEGA32 Microcontroller - The high-performance, low-power Atmel 8-bit AVR RISC-based microcontroller combines 16KB of programmable flash memory, 2KB SRAM, 1KB EEPROM, an 8channel 10-bit A/D converter, 2 capture/compare/PWM functions, 3 timers, a synchronous serial port that can be configured as either 3-wire SPI or 2-wire I2C bus, a USART, and a Parallel Slave Port. The device supports throughput of 16 MIPS at 16 MHz and operates between 4.5-5.5 volts
  • 5. Fingertip sensor - Use of light to measure heart rate is a field of study where abundant research has been done in the past few decades. Fingertip sensor relies on measurement of a physiological signal called Photoplethysmogrpahy (PPG) [14.], which is an optical measurement of the change in blood volume in the arteries. Fingertip sensor acquires PPG signals by irradiating wavelength of light through the tissue, and compares the light absorption characteristics of blood under these wavelengths High level design - By high level design we mean to see the project from an overall view point, then to get down and consider the project as a system of main parts. Fingertip sensor is used in the device as transducer converts the physical PPG signal to into an electrical signal. This electrical signal is applied to the microcontroller which is the main brain of the heart rate monitor. The microcontroller, according to the
  • 7. • This project used LM358 two-class amplifier to adjust the voltage value to 5V for being received by ATMEGA32. Because the signal acquisition is extremely weak and vulnerable to interference by the external circuit, high-pass filter and low-pass filter are used for normal heart beat signals. • The amplifier (see figure 4.9) uses an LM358 dual op amp to provide two identical broadly-tuned low pass stages with the gain of each filter stage is set to 101 and a cut-off frequency of about 2.34 Hz. So the maximum HR that could be read is approximately equal 150. The equations for calculating gain and cut-off frequency of the active low pass filter are expressed as
  • 8. LCD • In this project, a 16×2 LCD display will be used for displaying the value of heart rate. There are 16 pins in whole in LCD. The pin diagram of LCD is shown as below. From the datasheet, it is known that pin 1 & pin 2 and pin 15 & pin 16 both functions as power supply ports. However, just pin 1 & pin 2 needed, and the lightness of LCD should be controlled by a 10kΩ variable resistor.
  • 9. RESULT gender Age HR on display HR on oscilloscope Error rate (%) Male 22 97 96 1.03 Male 22 83 81 2.41 Male 20 78 78 0 Male 22 90 87 3.33 Male 20 80 79 1.25 Female 22 77 77 0 Female 22 104 103 0.96 Female 19 75 75 0 Female 20 69 71 2.81 The accuracy of the device depends on the testing performed is about 1.414. In general, the HR of females is higher than males. But this is not true where difference in weight, daily life (i.e. kind of work, health, inheritance of heart disease, doing exercises… etc).
  • 10. CONCLUSION Infrared sensor instead of pressure sensor was used in Heart Rate Monitor, it has higher sensitivity, and the output signal is more stability. The infrared sensor depends on the fact that the speed of blood is proportional to heart rate. That is to say, when people are in high active, the heart rate will change. As the same theory as previous research, heart rate changes when people are excited. The HR value displayed on LCD shows the information that when people took exercise, the heart rate will increase; sometimes it is out of healthy range. When people take a break, the normal heart rate will recover. In totally, heart rate is physiological indicators to show what is going on in body. People need to care about their healthy condition with the help of Heart Rate Monitor.
  • 11. References [1] Sokwoo Rhee “DESIGN AND ANALYSIS ARTIFACTRESISTIVEFINGER PHOTOPLETHYSMOGRAPHIC SENSORS FOR VITAL SIGN MONITORING” Massachusetts,U.S, 2001, chapter1,P9. [2] World health organization, Mortality country fact sheet 2006, Sudan. . WEB URL- [3] Wikipedia – http://en.wikipedia.org/wiki/Main_Page.