SlideShare a Scribd company logo
DR DHARMENDRA SINGH
DR REJAKUL ISLAM
DR DHARMENDRA SINGH
DR REJAKUL ISLAM
They are used to treat patients with beams of electrons or X- rays
following interactions of electrons in a suitable target.
The term MV is typically used to describe photon beams (e.g., 6
MV) whereas the term MeV is typically used to describe an
electron beam (e.g., 6 MeV).
Irrespective of the accelerator type, two basic conditions must
be met for particle acceleration:
● The particle to be accelerated must be charged;
● An electric field must be provided in the direction of particle
acceleration.
LINEAR ACCELERATOR ● Medium energy photons (10–15 MV)
● Low energy photons (4–8 MV)
● High energy photons (18–25 MV)
ACCELERATOR : ELECTROSTATIC AND CYCLIC
ELECTROSTATIC
ACCELERATOR
the particles are accelerated by applying an
electrostatic electric field through a voltage
difference, constant in time, whose value fixes
the value of the final kinetic energy of the
particle.
The energy that an electrostatic accelerator
can reach is limited by the discharges that
occur between the high voltage terminal and
the walls of the accelerator chamber when
the voltage drop exceeds a certain critical
value (typically 1 MV).
e.g: orthovoltage X ray tubes.
Electric fields used in cyclic accelerators are
variable and associated with a variable magnetic
field and resulting in some close paths along
which the kinetic energy gained by the particle
differs from zero.
If the particle is made to follow such a closed path
many times over, one obtains a process of gradual
acceleration that is not limited to the maximum
voltage drop existing in the accelerator.
Thus the final kinetic energy of the particle is obtained
by submitting the charged particle to the same,
relatively small, potential difference a large number of
times, each cycle adding a small amount of energy to
the kinetic energy of the particle.
example of a cyclic accelerator is the linac
CYCLIC
ACCELERATOR :
Accelerator Building Blocks
• Power source
• Electron source
• Accelerator structure
• Beam bending system
• Beam Conditioning
• Collimation system
• Control system
• Support structure
POWER SOURCE
Before discussing about power source or details of LINAC
Some basic principles of X-ray production
Why??? Crook’s tube can not produce high energy electrons for medical purpose.
Cockcroft-Walton Generator 665 kV
Early Challenges
• High voltage breakdown
• Reliable vacuum systems
• Low beam currents
• Zero diagnostics
• Small industrial base
• Scale – big science emerges
All accelerators are based on the same
principle. A charged particle
accelerates between a gap between
two electrodes when there is a
potential difference between them.
E= QV
RESONANT ACCELERATOR CONCEPT
RESONANT ACCELERATOR CONCEPT
THE ACCELERATION OCCURS IN THE ELECTRIC FIELD BETWEEN
CYLINDRICAL DRIFT TUBES.
THE RF POWER MUST BE SYNCHRONISED WITH THE MOTION OF THE
ELECTRONS, SO THAT ACCELERATION OCCURS IN EVERY CAVITY.
THIS NATURALLY PRODUCES BUNCHES OF ELECTRONS
ELECTRON SOURCE
DRIFT TUBES
When the first electrode is oppositely charged to the entering charged
particle (i.e. an electron or proton ), the particle accelerates towards it.
There is no electric field inside the tubes as they are hollow conductors.
When the particle enters the first tube , the voltage is switched so that
the next tube is oppositely charged, therefore it accelerates to next
electrode.
Each time the same magnitude of voltage is applied and so the energy
of the Particle E=n Q V , is built up in steps without the needing to
increase the voltage.
When the first electrode is oppositely charged to the entering charged
particle (i.e. an electron or proton ), the particle accelerates towards it.
There is no electric field inside the tubes as they are hollow conductors.
When the particle enters the first tube , the voltage is switched so that
the next tube is oppositely charged, therefore it accelerates to next
electrode.
Each time the same magnitude of voltage is applied and so the energy
of the Particle E=n Q V , is built up in steps without the needing to
increase the voltage.
When the first electrode is oppositely charged to the entering charged
particle (i.e. an electron or proton ), the particle accelerates towards it.
There is no electric field inside the tubes as they are hollow conductors.
When the particle enters the first tube , the voltage is switched so that
the next tube is oppositely charged, therefore it accelerates to next
electrode.
Each time the same magnitude of voltage is applied and so the energy
of the Particle E=n Q V , is built up in steps without the needing to
increase the voltage.
POWER SOURCE
● An RF power source;
● A pulsed modulator.
—A magnetron is a source of high power RF required for
electron acceleration, while a klystron is an RF power amplifier
that amplifies the low power RF generated by an RF oscillator
commonly called the RF driver.
Microwaves
Microwaves are sometimes considered to be very short radio waves
(high frequency and high-energy radio waves).
Some important properties of microwaves are:
They are reflected by metal surfaces.
They heat materials if they can make atoms or molecules in the
material vibrate. The amount of heating depends on the intensity of
the microwave radiation, and the time that the material is exposed to
the radiation.
They pass through glass and plastics.
They pass through the atmosphere.
They pass through the ionosphere without being reflected.
They are absorbed by water molecules, how well depends on the
frequency (energy) of the microwaves.
MAGNETRON Produces Microwaves also called RF waves.
The magnetron is a high-powered vacuum
tube, that works as self-excited microwave
oscillator. Crossed electron and magnetic fields
are used in the magnetron to produce the
high-power output required in radar
equipment. These multi-cavity devices works
at frequencies ranging from approximately 600
to 30,000 megahertz. The relatively simple
construction has the disadvantage that the
Magnetron usually can work only on a
constructively fixed frequency.
Microwaves are produced by vacuum tubes devices that operate
on the ballistic motion of electron controlled by magnetic or
electric fields. Some different kinds of microwave emitters are
the cavity magnetron, the klystron, the traveling-wave
tube(TWT), the gyrotron.
ELECTRON
GUN
PULSE
MODULATOR
MICROWAVE
POWER SOURCE
(MAGNETRON)
CONTROL
UNIT
POWER SUPPLY
GAS
PRESSURE
SYSTEM
Bending
magnets
MLC Component
Primary collimator
Flattening filter
Dual ionisation
chamber
Secondary collimator
Ceramic windowAnode
Vacuum Pump
BeamtransportsystemLINACHead
Thick X-ray target
Accelerating Wave guide
Patient support assembly
Steering coil
Cathode
ELECTRON SOURCE
It is essentially a simple electrostatic accelerator
called an electron gun.
Two types of electron gun are in use as sources of electrons in
medical linacs
— Diode type;
— Triode type.
Both electron gun types contain a heated filament cathode and a
perforated grounded anode; in addition, the triode electron gun
also incorporates a grid.
Anode
Cathode
ELECTRON GUN
Electrons are thermionically emitted from
the heated cathode, focused into a pencil
beam by a curved focusing electrode and
accelerated towards the perforated anode
through which they drift to enter the
accelerating waveguide.
ACCELERATOR STRUCTURE
The simplest kind of accelerating waveguide is obtained from a
cylindrical uniform waveguide by adding a series of discs (irises) with
circular holes at the centre, placed at equal distances along the tube.
These discs divide the waveguide into a series of cylindrical cavities
that form the basic structure of the accelerating waveguide in a linac.
The accelerating waveguide is evacuated to allow free propagation of
electrons.
The cavities of the accelerating waveguide serve two purposes:
- To couple and distribute microwave power between adjacent cavities;
- To provide a suitable electric field pattern for the acceleration of
electrons.
Two types of accelerating waveguide have been developed for the
acceleration of electrons:
(i) Travelling wave structure (ii) Standing wave structure.
Electron gun RF in
Accelerating cavity
Accelerating wave guide
RF out
Electron gun
RF in Coupling cavity
Travelling wave structure
Standing wave structure
Electron gun RF in
Accelerating cavity
Accelerating wave guide
RF out
Electron gun
RF in Coupling cavity
λ
λ'
λ‘=λ/2
In the travelling wave structure the microwaves enter the
accelerating waveguide on the gun side and propagate towards the
high energy end of the waveguide
where they either are absorbed without any reflection or exit the waveguide to
be absorbed in a resistive load or to be fed back to the input end of the
accelerating waveguide.
In the standing wave structure each end of the accelerating
waveguide is terminated with a conducting disc to reflect the
microwave power ,resulting in a buildup of standing waves in the
waveguide.
These cavities therefore serve only as coupling cavities and can be moved out to the
side of the waveguide structure, effectively shortening the accelerating waveguide
by 50%.
Bending magnets are used in linacs operating at energies above 6
MeV, where the accelerating waveguides are too long for straight-
through mounting.
The accelerating waveguide is usually mounted parallel to the gantry
rotation axis and the electron beam must be bent to make it strike the
X - ray target or be able to exit through the beam exit window.
BEAM BENDING SYSTEM
In low energy linacs the target is embedded in the accelerating waveguide and no
beam transport between the accelerating waveguide and target is required
COLLIMATION SYSTEM
The important components found in a typical head of linac include:
—Retractable X ray targets;
—Flattening filters and electron scattering foils (also called scattering
filters);
—Primary and adjustable secondary collimators;
—Dual transmission ionization chambers;
—A field defining light and a range finder;
—retractable wedges;
— MLC.
The photon beam collimation is achieved with two or three
collimator devices:
1. A primary collimator;
2. Secondary movable beam defining collimators;
3. An MLC (optional).
The primary collimator defines the
largest available circular field size and is
a conical opening machined into a
tungsten shielding block, with the sides
of the conical opening projecting on to
edges of the target on one end of the
block and on to the flattening filter on
the other end
The secondary beam defining collimators consist of four blocks,
two forming the upper and two forming the lower jaws of the
collimator. They can provide rectangular or square fields at the
linac isocentre, with sides of the order of few mm up to 40 cm.
The secondary beam defining collimators consist of four blocks,
two forming the upper and two forming the lower jaws of the
collimator. They can provide rectangular or square fields at the
linac isocentre, with sides of the order of few mm up to 40 cm.
A multileaf collimator (MLC) for photon beams consists of a large
number of collimating blocks or leaves that can be driven
automatically, independent of each other, to generate a field of any
shape
The individual leaf has a width of 1 cm or less as projected at the
isocenter. The leaves are made of tungsten alloy (r = 17.0 to 18.5
g/cm3) and have thickness along the beam direction ranging from
6 cm to 7.5 cm, depending on the type of accelerator. The leaf
thickness is sufficient to provide primary x-ray transmission
through the leaves of less than 2%
A multileaf collimator (MLC) for photon beams consists of a large
number of collimating blocks or leaves that can be driven
automatically, independent of each other, to generate a field of any
shape
The individual leaf has a width of 1 cm or less as projected at the
isocenter. The leaves are made of tungsten alloy (r = 17.0 to 18.5
g/cm3) and have thickness along the beam direction ranging from
6 cm to 7.5 cm, depending on the type of accelerator. The leaf
thickness is sufficient to provide primary x-ray transmission
through the leaves of less than 2%
BEAM CONDITIONING
Flattening filter
X-Ray
The photon dose distribution produced
by a linac is strongly forward peaked.
To make beam intensity uniform across a
field , a flattening filter is inserted in the beam.
The filter is usually made of Pb , although tungsten,
steel, Al or combination has also been suggested.
BEAM CONDITIONING
Flattening filter
The photon dose distribution produced
by a linac is strongly forward peaked.
To make beam intensity uniform across a
field , a flattening filter is inserted in the beam.
The filter is usually made of Pb , although tungsten,
steel, Al or combination has also been suggested.
X-Ray
Scattering foil
Electron Beam
The electron pencil beam exits the
evacuated beam transport system
through a thin window usually
made of beryllium, which, with its
low atomic number Z, minimizes
the pencil beam scattering and X-
ray production
The scattering of the electron pencil beam over the relatively large
area used in radiotherapy (up to 25 × 25 cm2) is achieved by placing
thin foils of high Z material (copper or lead) into the pencil beam at
the level of the flattening filter in the X -ray mode.
Scattering foil
The electron pencil beam exits the
evacuated beam transport system
through a thin window usually
made of beryllium, which, with its
low atomic number Z, minimizes
the pencil beam scattering and X-
ray production
The scattering of the electron pencil beam over the relatively large
area used in radiotherapy (up to 25 × 25 cm2) is achieved by placing
thin foils of high Z material (copper or lead) into the pencil beam .
Electron Beam
Dual ionisation chamber
Dual ionisation chamber
Dual ionisation chamber
Dual ionisation chamber
Dual ionisation chamber
Dual ionisation chamber
Termination of irradiation
Most common dose monitors in linacs
are transmission ionization chambers
permanently imbedded in the linac
clinical photon and electron beams to
monitor the beam output continuously
during patient treatment
Dual ionisation chamber
Properties of the ionization chamber monitors
 Chambers must have a minimal effect on clinical photon and
electron radiation beams;
Chamber response should be independent of ambient
temperature and pressure (most linacs use sealed ionization
chambers to satisfy this condition);
Chambers should be operated under saturation conditions.
The primary ionization chamber measures MUs. Typically, the
sensitivity of the chamber electrometer circuitry is adjusted in
such a way that 1 MU corresponds to a dose of 1 cGy delivered in
a water phantom at the depth of dose maximum on the central
beam axis when irradiated with a 10 × 10 cm2 field at a source to
surface distance (SSD) of 100 cm.
Once the operator preset number of MUs has been
reached, the primary ionization chamber circuitry shuts the
linac down and terminates the dose delivery to the patient.
Before a new irradiation can be initiated, it is necessary to
reset the MU displays to zero.
Furthermore, irradiation is not possible until a new
selection of MUs has been made.
The dose monitoring system also monitors other operating
parameters such as the beam energy, flatness and symmetry.
Field lights and lasers
It is a field localizing device.
Used to display the position of the radiation field on the patient
skin.
The accuracy of lasers guides in determining isocentre
position.
A combination of mirror and a light source located in the space
between the chambers and the jaws projects a light beam as if
emitting from the x-ray focal spot.
Pulse modulator
A power supply provides direct current (DC) power to the
modulator, which includes the pulse-forming network and a switch
tube known as hydrogen thyratron. High-voltage pulses from the
modulator section are flat-topped DC pulses of a few microseconds
in duration. These pulses are delivered to the magnetron or
klystron and simultaneously to the electron gun.
Control console
For monitoring and controlling linac
All interlocks checks to allow beam started.
To provide digital display of monitor units
Mechanical beam parameters.
Support structure
Thank you

More Related Content

What's hot

Flattening filter Free
Flattening filter FreeFlattening filter Free
Electron beam therapy
Electron beam therapyElectron beam therapy
Electron beam therapy
Kiran Ramakrishna
 
Linear Accelerators
Linear AcceleratorsLinear Accelerators
Linear Accelerators
Velliangiri P
 
EXTERNAL PHOTON BEAMS THERAPY (PART 2)
EXTERNAL PHOTON BEAMS THERAPY (PART 2)EXTERNAL PHOTON BEAMS THERAPY (PART 2)
EXTERNAL PHOTON BEAMS THERAPY (PART 2)
Nik Noor Ashikin Nik Ab Razak
 
LINAC- LINEAR ACCELERATOR
LINAC- LINEAR ACCELERATORLINAC- LINEAR ACCELERATOR
LINAC- LINEAR ACCELERATOR
Mariya Kikali
 
Dosimetry
DosimetryDosimetry
Dosimetry
studious girl
 
Cobalt 60
Cobalt 60Cobalt 60
Cobalt 60
Ovais Ur Rehman
 
Linear accelerator vinay
Linear accelerator vinayLinear accelerator vinay
Linear accelerator vinay
Vinay Desai
 
Linac
LinacLinac
Vmat technique for Breast, Head and Neck, Brain and Craniospinal irradiation ...
Vmat technique for Breast, Head and Neck, Brain and Craniospinal irradiation ...Vmat technique for Breast, Head and Neck, Brain and Craniospinal irradiation ...
Vmat technique for Breast, Head and Neck, Brain and Craniospinal irradiation ...
Biplab Sarkar
 
Linear Accelerator
Linear AcceleratorLinear Accelerator
Linear Accelerator
sailakshmi pullookkara
 
Beam modification
Beam modification Beam modification
Beam modification
LAKSHMI DEEPTHI GEDELA
 
Image guided radiation therapy (IGRT)
Image guided radiation therapy (IGRT)Image guided radiation therapy (IGRT)
Image guided radiation therapy (IGRT)
Anees Muhammed
 
ELECTRON BEAM THERAPY
ELECTRON BEAM THERAPYELECTRON BEAM THERAPY
ELECTRON BEAM THERAPY
Sathish Kumar
 
Linac- A Practical Approach
Linac- A Practical Approach Linac- A Practical Approach
Linac- A Practical Approach
Aaditya Sinha
 
Isotopic Teletherapy Machines
Isotopic Teletherapy MachinesIsotopic Teletherapy Machines
Isotopic Teletherapy Machines
Arnab Bose
 
Quality assurance of linear accelerator DHX
Quality assurance of linear accelerator DHXQuality assurance of linear accelerator DHX
Quality assurance of linear accelerator DHX
Sohail Qureshi
 
Epid
EpidEpid
TRS 398 (Technical Report Series)
TRS 398 (Technical Report Series)TRS 398 (Technical Report Series)
TRS 398 (Technical Report Series)
Vinay Desai
 
Dose volume histogram
Dose volume histogramDose volume histogram
Dose volume histogram
Sasikumar Sambasivam
 

What's hot (20)

Flattening filter Free
Flattening filter FreeFlattening filter Free
Flattening filter Free
 
Electron beam therapy
Electron beam therapyElectron beam therapy
Electron beam therapy
 
Linear Accelerators
Linear AcceleratorsLinear Accelerators
Linear Accelerators
 
EXTERNAL PHOTON BEAMS THERAPY (PART 2)
EXTERNAL PHOTON BEAMS THERAPY (PART 2)EXTERNAL PHOTON BEAMS THERAPY (PART 2)
EXTERNAL PHOTON BEAMS THERAPY (PART 2)
 
LINAC- LINEAR ACCELERATOR
LINAC- LINEAR ACCELERATORLINAC- LINEAR ACCELERATOR
LINAC- LINEAR ACCELERATOR
 
Dosimetry
DosimetryDosimetry
Dosimetry
 
Cobalt 60
Cobalt 60Cobalt 60
Cobalt 60
 
Linear accelerator vinay
Linear accelerator vinayLinear accelerator vinay
Linear accelerator vinay
 
Linac
LinacLinac
Linac
 
Vmat technique for Breast, Head and Neck, Brain and Craniospinal irradiation ...
Vmat technique for Breast, Head and Neck, Brain and Craniospinal irradiation ...Vmat technique for Breast, Head and Neck, Brain and Craniospinal irradiation ...
Vmat technique for Breast, Head and Neck, Brain and Craniospinal irradiation ...
 
Linear Accelerator
Linear AcceleratorLinear Accelerator
Linear Accelerator
 
Beam modification
Beam modification Beam modification
Beam modification
 
Image guided radiation therapy (IGRT)
Image guided radiation therapy (IGRT)Image guided radiation therapy (IGRT)
Image guided radiation therapy (IGRT)
 
ELECTRON BEAM THERAPY
ELECTRON BEAM THERAPYELECTRON BEAM THERAPY
ELECTRON BEAM THERAPY
 
Linac- A Practical Approach
Linac- A Practical Approach Linac- A Practical Approach
Linac- A Practical Approach
 
Isotopic Teletherapy Machines
Isotopic Teletherapy MachinesIsotopic Teletherapy Machines
Isotopic Teletherapy Machines
 
Quality assurance of linear accelerator DHX
Quality assurance of linear accelerator DHXQuality assurance of linear accelerator DHX
Quality assurance of linear accelerator DHX
 
Epid
EpidEpid
Epid
 
TRS 398 (Technical Report Series)
TRS 398 (Technical Report Series)TRS 398 (Technical Report Series)
TRS 398 (Technical Report Series)
 
Dose volume histogram
Dose volume histogramDose volume histogram
Dose volume histogram
 

Viewers also liked

LINEAR ACCELARATOR
LINEAR ACCELARATORLINEAR ACCELARATOR
LINEAR ACCELARATOR
Waqar Ahmed
 
Linear accelator
Linear accelatorLinear accelator
Linear accelator
SUST\RICK
 
Linear accelerator linac
Linear accelerator linacLinear accelerator linac
Linear accelerator linac
Syed Hammad .
 
linear accelerator
linear acceleratorlinear accelerator
linear accelerator
Anwar Hussain
 
Electron Beam Therapy
Electron Beam TherapyElectron Beam Therapy
Electron Beam Therapy
jyotimannath
 
Brachytherapy
BrachytherapyBrachytherapy
Brachytherapy
Joann Vargas
 
Fluoroscopy ppt
Fluoroscopy pptFluoroscopy ppt
Fluoroscopy ppt
Brittany Jansante
 
Beam Modification in Radiotherapy
Beam Modification in RadiotherapyBeam Modification in Radiotherapy
Beam Modification in Radiotherapy
Santam Chakraborty
 
Basics Of MRI
Basics Of MRIBasics Of MRI
Basics Of MRI
Keshav Kulkarni
 
Basics of mri
Basics of mriBasics of mri
Basics of mri
Ganesh Nair
 
Breast cancer ppt
Breast cancer pptBreast cancer ppt
Breast cancer ppt
drizsyed
 
linear accelerator
linear acceleratorlinear accelerator
linear accelerator
Anwar Hussain
 
08.[슬라이드]호흡동조 v20160224
08.[슬라이드]호흡동조 v2016022408.[슬라이드]호흡동조 v20160224
08.[슬라이드]호흡동조 v20160224
Samsung Medical Center
 
4DCT with Optical Tracking Systems
4DCT with Optical Tracking Systems4DCT with Optical Tracking Systems
4DCT with Optical Tracking Systems
Az.Ospedaliero-Universitaria di Modena
 
CHARGE PARTICLE ACCELERATORCharge particle accelerator
CHARGE PARTICLE ACCELERATORCharge particle acceleratorCHARGE PARTICLE ACCELERATORCharge particle accelerator
CHARGE PARTICLE ACCELERATORCharge particle accelerator
SYED SHAHEEN SHAH
 
Magnetron power generator
Magnetron power generator Magnetron power generator
Magnetron power generator
Alexander Frolov
 
Mi ppt fi print this
Mi ppt fi print thisMi ppt fi print this
Mi ppt fi print this
Ruta Angel
 
Mr fluoroscopy
Mr fluoroscopyMr fluoroscopy
Mr fluoroscopy
shajitha khan
 
Klystron
KlystronKlystron
Klystron
Parsa Nayeem
 
Particle accelerator
Particle acceleratorParticle accelerator
Particle accelerator
Chen Siyuan
 

Viewers also liked (20)

LINEAR ACCELARATOR
LINEAR ACCELARATORLINEAR ACCELARATOR
LINEAR ACCELARATOR
 
Linear accelator
Linear accelatorLinear accelator
Linear accelator
 
Linear accelerator linac
Linear accelerator linacLinear accelerator linac
Linear accelerator linac
 
linear accelerator
linear acceleratorlinear accelerator
linear accelerator
 
Electron Beam Therapy
Electron Beam TherapyElectron Beam Therapy
Electron Beam Therapy
 
Brachytherapy
BrachytherapyBrachytherapy
Brachytherapy
 
Fluoroscopy ppt
Fluoroscopy pptFluoroscopy ppt
Fluoroscopy ppt
 
Beam Modification in Radiotherapy
Beam Modification in RadiotherapyBeam Modification in Radiotherapy
Beam Modification in Radiotherapy
 
Basics Of MRI
Basics Of MRIBasics Of MRI
Basics Of MRI
 
Basics of mri
Basics of mriBasics of mri
Basics of mri
 
Breast cancer ppt
Breast cancer pptBreast cancer ppt
Breast cancer ppt
 
linear accelerator
linear acceleratorlinear accelerator
linear accelerator
 
08.[슬라이드]호흡동조 v20160224
08.[슬라이드]호흡동조 v2016022408.[슬라이드]호흡동조 v20160224
08.[슬라이드]호흡동조 v20160224
 
4DCT with Optical Tracking Systems
4DCT with Optical Tracking Systems4DCT with Optical Tracking Systems
4DCT with Optical Tracking Systems
 
CHARGE PARTICLE ACCELERATORCharge particle accelerator
CHARGE PARTICLE ACCELERATORCharge particle acceleratorCHARGE PARTICLE ACCELERATORCharge particle accelerator
CHARGE PARTICLE ACCELERATORCharge particle accelerator
 
Magnetron power generator
Magnetron power generator Magnetron power generator
Magnetron power generator
 
Mi ppt fi print this
Mi ppt fi print thisMi ppt fi print this
Mi ppt fi print this
 
Mr fluoroscopy
Mr fluoroscopyMr fluoroscopy
Mr fluoroscopy
 
Klystron
KlystronKlystron
Klystron
 
Particle accelerator
Particle acceleratorParticle accelerator
Particle accelerator
 

Similar to Linac presentation

CLINICAL RADIATION GENERATORS
CLINICAL RADIATION GENERATORS CLINICAL RADIATION GENERATORS
CLINICAL RADIATION GENERATORS
Sathish Kumar
 
Linear accelerator
Linear acceleratorLinear accelerator
Linear accelerator
Neena John
 
Linear Accelerator working and construction
Linear Accelerator working and constructionLinear Accelerator working and construction
Linear Accelerator working and construction
BDRRISHIKESHVLOGS
 
Linear Accelerators.pptx
Linear Accelerators.pptxLinear Accelerators.pptx
Linear Accelerators.pptx
SaikatChakraborty80
 
LINAC by Dr. Musaib Mushtaq.pptx
LINAC by Dr. Musaib Mushtaq.pptxLINAC by Dr. Musaib Mushtaq.pptx
LINAC by Dr. Musaib Mushtaq.pptx
MusaibMushtaq
 
4 microwave device
4 microwave device4 microwave device
4 microwave device
AymanMahmoud66
 
MICMicrowave Tubes – klystron, reflex klystron, magnetron and TWT.
MICMicrowave Tubes – klystron, reflex klystron, magnetron and TWT.MICMicrowave Tubes – klystron, reflex klystron, magnetron and TWT.
MICMicrowave Tubes – klystron, reflex klystron, magnetron and TWT.
jeronimored
 
Data communications and computer networking slides
Data communications and computer networking slidesData communications and computer networking slides
Data communications and computer networking slides
MohammedAbbas653737
 
Witricity
WitricityWitricity
circuit diagram.pptx
circuit diagram.pptxcircuit diagram.pptx
circuit diagram.pptx
FAYAZ KHAWAJA
 
Role of Particle accelerators in Radiotherapy
Role of Particle accelerators in RadiotherapyRole of Particle accelerators in Radiotherapy
Role of Particle accelerators in Radiotherapy
sangeethamani26
 
Linac
LinacLinac
Linac
salimsha
 
X Ray Production
X Ray ProductionX Ray Production
X Ray Production
Anees Muhammed
 
UNIT V RADAR TRANSMITTERS AND RECEIVERS 14.11.23.pdf
UNIT V RADAR TRANSMITTERS AND RECEIVERS 14.11.23.pdfUNIT V RADAR TRANSMITTERS AND RECEIVERS 14.11.23.pdf
UNIT V RADAR TRANSMITTERS AND RECEIVERS 14.11.23.pdf
devicaf983
 
Clinical Generators in Radiotherapy by Dr.Avilash.pptx
Clinical Generators in Radiotherapy by Dr.Avilash.pptxClinical Generators in Radiotherapy by Dr.Avilash.pptx
Clinical Generators in Radiotherapy by Dr.Avilash.pptx
AbhilashBanerjee3
 
Short wave diathermy.pptx
Short wave diathermy.pptxShort wave diathermy.pptx
Short wave diathermy.pptx
Aashish Gho Shrestha
 
Production of xrays
Production of xraysProduction of xrays
Production of xrays
priyanka rana
 
Microwave generators
Microwave generatorsMicrowave generators
Microwave generators
Shankar Gangaju
 
Serena barbanotti INFN milano
Serena barbanotti INFN milanoSerena barbanotti INFN milano
Serena barbanotti INFN milano
adamas comunicazione
 
UNIT II-microwave sources (1).pptx
UNIT II-microwave sources (1).pptxUNIT II-microwave sources (1).pptx
UNIT II-microwave sources (1).pptx
PavanVangapally
 

Similar to Linac presentation (20)

CLINICAL RADIATION GENERATORS
CLINICAL RADIATION GENERATORS CLINICAL RADIATION GENERATORS
CLINICAL RADIATION GENERATORS
 
Linear accelerator
Linear acceleratorLinear accelerator
Linear accelerator
 
Linear Accelerator working and construction
Linear Accelerator working and constructionLinear Accelerator working and construction
Linear Accelerator working and construction
 
Linear Accelerators.pptx
Linear Accelerators.pptxLinear Accelerators.pptx
Linear Accelerators.pptx
 
LINAC by Dr. Musaib Mushtaq.pptx
LINAC by Dr. Musaib Mushtaq.pptxLINAC by Dr. Musaib Mushtaq.pptx
LINAC by Dr. Musaib Mushtaq.pptx
 
4 microwave device
4 microwave device4 microwave device
4 microwave device
 
MICMicrowave Tubes – klystron, reflex klystron, magnetron and TWT.
MICMicrowave Tubes – klystron, reflex klystron, magnetron and TWT.MICMicrowave Tubes – klystron, reflex klystron, magnetron and TWT.
MICMicrowave Tubes – klystron, reflex klystron, magnetron and TWT.
 
Data communications and computer networking slides
Data communications and computer networking slidesData communications and computer networking slides
Data communications and computer networking slides
 
Witricity
WitricityWitricity
Witricity
 
circuit diagram.pptx
circuit diagram.pptxcircuit diagram.pptx
circuit diagram.pptx
 
Role of Particle accelerators in Radiotherapy
Role of Particle accelerators in RadiotherapyRole of Particle accelerators in Radiotherapy
Role of Particle accelerators in Radiotherapy
 
Linac
LinacLinac
Linac
 
X Ray Production
X Ray ProductionX Ray Production
X Ray Production
 
UNIT V RADAR TRANSMITTERS AND RECEIVERS 14.11.23.pdf
UNIT V RADAR TRANSMITTERS AND RECEIVERS 14.11.23.pdfUNIT V RADAR TRANSMITTERS AND RECEIVERS 14.11.23.pdf
UNIT V RADAR TRANSMITTERS AND RECEIVERS 14.11.23.pdf
 
Clinical Generators in Radiotherapy by Dr.Avilash.pptx
Clinical Generators in Radiotherapy by Dr.Avilash.pptxClinical Generators in Radiotherapy by Dr.Avilash.pptx
Clinical Generators in Radiotherapy by Dr.Avilash.pptx
 
Short wave diathermy.pptx
Short wave diathermy.pptxShort wave diathermy.pptx
Short wave diathermy.pptx
 
Production of xrays
Production of xraysProduction of xrays
Production of xrays
 
Microwave generators
Microwave generatorsMicrowave generators
Microwave generators
 
Serena barbanotti INFN milano
Serena barbanotti INFN milanoSerena barbanotti INFN milano
Serena barbanotti INFN milano
 
UNIT II-microwave sources (1).pptx
UNIT II-microwave sources (1).pptxUNIT II-microwave sources (1).pptx
UNIT II-microwave sources (1).pptx
 

Recently uploaded

Novas diretrizes da OMS para os cuidados perinatais de mais qualidade
Novas diretrizes da OMS para os cuidados perinatais de mais qualidadeNovas diretrizes da OMS para os cuidados perinatais de mais qualidade
Novas diretrizes da OMS para os cuidados perinatais de mais qualidade
Prof. Marcus Renato de Carvalho
 
Best Ayurvedic medicine for Gas and Indigestion
Best Ayurvedic medicine for Gas and IndigestionBest Ayurvedic medicine for Gas and Indigestion
Best Ayurvedic medicine for Gas and Indigestion
Swastik Ayurveda
 
#cALL# #gIRLS# In Dehradun ꧁❤8107221448❤꧂#cALL# #gIRLS# Service In Dehradun W...
#cALL# #gIRLS# In Dehradun ꧁❤8107221448❤꧂#cALL# #gIRLS# Service In Dehradun W...#cALL# #gIRLS# In Dehradun ꧁❤8107221448❤꧂#cALL# #gIRLS# Service In Dehradun W...
#cALL# #gIRLS# In Dehradun ꧁❤8107221448❤꧂#cALL# #gIRLS# Service In Dehradun W...
chandankumarsmartiso
 
REGULATION FOR COMBINATION PRODUCTS AND MEDICAL DEVICES.pptx
REGULATION FOR COMBINATION PRODUCTS AND MEDICAL DEVICES.pptxREGULATION FOR COMBINATION PRODUCTS AND MEDICAL DEVICES.pptx
REGULATION FOR COMBINATION PRODUCTS AND MEDICAL DEVICES.pptx
LaniyaNasrink
 
The Best Ayurvedic Antacid Tablets in India
The Best Ayurvedic Antacid Tablets in IndiaThe Best Ayurvedic Antacid Tablets in India
The Best Ayurvedic Antacid Tablets in India
Swastik Ayurveda
 
Identification and nursing management of congenital malformations .pptx
Identification and nursing management of congenital malformations .pptxIdentification and nursing management of congenital malformations .pptx
Identification and nursing management of congenital malformations .pptx
MGM SCHOOL/COLLEGE OF NURSING
 
CHEMOTHERAPY_RDP_CHAPTER 3_ANTIFUNGAL AGENT.pdf
CHEMOTHERAPY_RDP_CHAPTER 3_ANTIFUNGAL AGENT.pdfCHEMOTHERAPY_RDP_CHAPTER 3_ANTIFUNGAL AGENT.pdf
CHEMOTHERAPY_RDP_CHAPTER 3_ANTIFUNGAL AGENT.pdf
rishi2789
 
ABDOMINAL TRAUMA in pediatrics part one.
ABDOMINAL TRAUMA in pediatrics part one.ABDOMINAL TRAUMA in pediatrics part one.
ABDOMINAL TRAUMA in pediatrics part one.
drhasanrajab
 
Integrating Ayurveda into Parkinson’s Management: A Holistic Approach
Integrating Ayurveda into Parkinson’s Management: A Holistic ApproachIntegrating Ayurveda into Parkinson’s Management: A Holistic Approach
Integrating Ayurveda into Parkinson’s Management: A Holistic Approach
Ayurveda ForAll
 
Top 10 Best Ayurvedic Kidney Stone Syrups in India
Top 10 Best Ayurvedic Kidney Stone Syrups in IndiaTop 10 Best Ayurvedic Kidney Stone Syrups in India
Top 10 Best Ayurvedic Kidney Stone Syrups in India
Swastik Ayurveda
 
OCT Training Course for clinical practice Part 1
OCT Training Course for clinical practice Part 1OCT Training Course for clinical practice Part 1
OCT Training Course for clinical practice Part 1
KafrELShiekh University
 
Journal Article Review on Rasamanikya
Journal Article Review on RasamanikyaJournal Article Review on Rasamanikya
Journal Article Review on Rasamanikya
Dr. Jyothirmai Paindla
 
Light House Retreats: Plant Medicine Retreat Europe
Light House Retreats: Plant Medicine Retreat EuropeLight House Retreats: Plant Medicine Retreat Europe
Light House Retreats: Plant Medicine Retreat Europe
Lighthouse Retreat
 
CHEMOTHERAPY_RDP_CHAPTER 2 _LEPROSY.pdf1
CHEMOTHERAPY_RDP_CHAPTER 2 _LEPROSY.pdf1CHEMOTHERAPY_RDP_CHAPTER 2 _LEPROSY.pdf1
CHEMOTHERAPY_RDP_CHAPTER 2 _LEPROSY.pdf1
rishi2789
 
Efficacy of Avartana Sneha in Ayurveda
Efficacy of Avartana Sneha in AyurvedaEfficacy of Avartana Sneha in Ayurveda
Efficacy of Avartana Sneha in Ayurveda
Dr. Jyothirmai Paindla
 
A Classical Text Review on Basavarajeeyam
A Classical Text Review on BasavarajeeyamA Classical Text Review on Basavarajeeyam
A Classical Text Review on Basavarajeeyam
Dr. Jyothirmai Paindla
 
Vestibulocochlear Nerve by Dr. Rabia Inam Gandapore.pptx
Vestibulocochlear Nerve by Dr. Rabia Inam Gandapore.pptxVestibulocochlear Nerve by Dr. Rabia Inam Gandapore.pptx
Vestibulocochlear Nerve by Dr. Rabia Inam Gandapore.pptx
Dr. Rabia Inam Gandapore
 
Muscles of Mastication by Dr. Rabia Inam Gandapore.pptx
Muscles of Mastication by Dr. Rabia Inam Gandapore.pptxMuscles of Mastication by Dr. Rabia Inam Gandapore.pptx
Muscles of Mastication by Dr. Rabia Inam Gandapore.pptx
Dr. Rabia Inam Gandapore
 
Hemodialysis: Chapter 4, Dialysate Circuit - Dr.Gawad
Hemodialysis: Chapter 4, Dialysate Circuit - Dr.GawadHemodialysis: Chapter 4, Dialysate Circuit - Dr.Gawad
Hemodialysis: Chapter 4, Dialysate Circuit - Dr.Gawad
NephroTube - Dr.Gawad
 
CHEMOTHERAPY_RDP_CHAPTER 6_Anti Malarial Drugs.pdf
CHEMOTHERAPY_RDP_CHAPTER 6_Anti Malarial Drugs.pdfCHEMOTHERAPY_RDP_CHAPTER 6_Anti Malarial Drugs.pdf
CHEMOTHERAPY_RDP_CHAPTER 6_Anti Malarial Drugs.pdf
rishi2789
 

Recently uploaded (20)

Novas diretrizes da OMS para os cuidados perinatais de mais qualidade
Novas diretrizes da OMS para os cuidados perinatais de mais qualidadeNovas diretrizes da OMS para os cuidados perinatais de mais qualidade
Novas diretrizes da OMS para os cuidados perinatais de mais qualidade
 
Best Ayurvedic medicine for Gas and Indigestion
Best Ayurvedic medicine for Gas and IndigestionBest Ayurvedic medicine for Gas and Indigestion
Best Ayurvedic medicine for Gas and Indigestion
 
#cALL# #gIRLS# In Dehradun ꧁❤8107221448❤꧂#cALL# #gIRLS# Service In Dehradun W...
#cALL# #gIRLS# In Dehradun ꧁❤8107221448❤꧂#cALL# #gIRLS# Service In Dehradun W...#cALL# #gIRLS# In Dehradun ꧁❤8107221448❤꧂#cALL# #gIRLS# Service In Dehradun W...
#cALL# #gIRLS# In Dehradun ꧁❤8107221448❤꧂#cALL# #gIRLS# Service In Dehradun W...
 
REGULATION FOR COMBINATION PRODUCTS AND MEDICAL DEVICES.pptx
REGULATION FOR COMBINATION PRODUCTS AND MEDICAL DEVICES.pptxREGULATION FOR COMBINATION PRODUCTS AND MEDICAL DEVICES.pptx
REGULATION FOR COMBINATION PRODUCTS AND MEDICAL DEVICES.pptx
 
The Best Ayurvedic Antacid Tablets in India
The Best Ayurvedic Antacid Tablets in IndiaThe Best Ayurvedic Antacid Tablets in India
The Best Ayurvedic Antacid Tablets in India
 
Identification and nursing management of congenital malformations .pptx
Identification and nursing management of congenital malformations .pptxIdentification and nursing management of congenital malformations .pptx
Identification and nursing management of congenital malformations .pptx
 
CHEMOTHERAPY_RDP_CHAPTER 3_ANTIFUNGAL AGENT.pdf
CHEMOTHERAPY_RDP_CHAPTER 3_ANTIFUNGAL AGENT.pdfCHEMOTHERAPY_RDP_CHAPTER 3_ANTIFUNGAL AGENT.pdf
CHEMOTHERAPY_RDP_CHAPTER 3_ANTIFUNGAL AGENT.pdf
 
ABDOMINAL TRAUMA in pediatrics part one.
ABDOMINAL TRAUMA in pediatrics part one.ABDOMINAL TRAUMA in pediatrics part one.
ABDOMINAL TRAUMA in pediatrics part one.
 
Integrating Ayurveda into Parkinson’s Management: A Holistic Approach
Integrating Ayurveda into Parkinson’s Management: A Holistic ApproachIntegrating Ayurveda into Parkinson’s Management: A Holistic Approach
Integrating Ayurveda into Parkinson’s Management: A Holistic Approach
 
Top 10 Best Ayurvedic Kidney Stone Syrups in India
Top 10 Best Ayurvedic Kidney Stone Syrups in IndiaTop 10 Best Ayurvedic Kidney Stone Syrups in India
Top 10 Best Ayurvedic Kidney Stone Syrups in India
 
OCT Training Course for clinical practice Part 1
OCT Training Course for clinical practice Part 1OCT Training Course for clinical practice Part 1
OCT Training Course for clinical practice Part 1
 
Journal Article Review on Rasamanikya
Journal Article Review on RasamanikyaJournal Article Review on Rasamanikya
Journal Article Review on Rasamanikya
 
Light House Retreats: Plant Medicine Retreat Europe
Light House Retreats: Plant Medicine Retreat EuropeLight House Retreats: Plant Medicine Retreat Europe
Light House Retreats: Plant Medicine Retreat Europe
 
CHEMOTHERAPY_RDP_CHAPTER 2 _LEPROSY.pdf1
CHEMOTHERAPY_RDP_CHAPTER 2 _LEPROSY.pdf1CHEMOTHERAPY_RDP_CHAPTER 2 _LEPROSY.pdf1
CHEMOTHERAPY_RDP_CHAPTER 2 _LEPROSY.pdf1
 
Efficacy of Avartana Sneha in Ayurveda
Efficacy of Avartana Sneha in AyurvedaEfficacy of Avartana Sneha in Ayurveda
Efficacy of Avartana Sneha in Ayurveda
 
A Classical Text Review on Basavarajeeyam
A Classical Text Review on BasavarajeeyamA Classical Text Review on Basavarajeeyam
A Classical Text Review on Basavarajeeyam
 
Vestibulocochlear Nerve by Dr. Rabia Inam Gandapore.pptx
Vestibulocochlear Nerve by Dr. Rabia Inam Gandapore.pptxVestibulocochlear Nerve by Dr. Rabia Inam Gandapore.pptx
Vestibulocochlear Nerve by Dr. Rabia Inam Gandapore.pptx
 
Muscles of Mastication by Dr. Rabia Inam Gandapore.pptx
Muscles of Mastication by Dr. Rabia Inam Gandapore.pptxMuscles of Mastication by Dr. Rabia Inam Gandapore.pptx
Muscles of Mastication by Dr. Rabia Inam Gandapore.pptx
 
Hemodialysis: Chapter 4, Dialysate Circuit - Dr.Gawad
Hemodialysis: Chapter 4, Dialysate Circuit - Dr.GawadHemodialysis: Chapter 4, Dialysate Circuit - Dr.Gawad
Hemodialysis: Chapter 4, Dialysate Circuit - Dr.Gawad
 
CHEMOTHERAPY_RDP_CHAPTER 6_Anti Malarial Drugs.pdf
CHEMOTHERAPY_RDP_CHAPTER 6_Anti Malarial Drugs.pdfCHEMOTHERAPY_RDP_CHAPTER 6_Anti Malarial Drugs.pdf
CHEMOTHERAPY_RDP_CHAPTER 6_Anti Malarial Drugs.pdf
 

Linac presentation

  • 1. DR DHARMENDRA SINGH DR REJAKUL ISLAM
  • 2. DR DHARMENDRA SINGH DR REJAKUL ISLAM
  • 3.
  • 4.
  • 5.
  • 6.
  • 7.
  • 8.
  • 9.
  • 10.
  • 11.
  • 12.
  • 13.
  • 14.
  • 15.
  • 16.
  • 17.
  • 18.
  • 19.
  • 20.
  • 21.
  • 22.
  • 23. They are used to treat patients with beams of electrons or X- rays following interactions of electrons in a suitable target. The term MV is typically used to describe photon beams (e.g., 6 MV) whereas the term MeV is typically used to describe an electron beam (e.g., 6 MeV). Irrespective of the accelerator type, two basic conditions must be met for particle acceleration: ● The particle to be accelerated must be charged; ● An electric field must be provided in the direction of particle acceleration. LINEAR ACCELERATOR ● Medium energy photons (10–15 MV) ● Low energy photons (4–8 MV) ● High energy photons (18–25 MV)
  • 24. ACCELERATOR : ELECTROSTATIC AND CYCLIC ELECTROSTATIC ACCELERATOR the particles are accelerated by applying an electrostatic electric field through a voltage difference, constant in time, whose value fixes the value of the final kinetic energy of the particle. The energy that an electrostatic accelerator can reach is limited by the discharges that occur between the high voltage terminal and the walls of the accelerator chamber when the voltage drop exceeds a certain critical value (typically 1 MV). e.g: orthovoltage X ray tubes.
  • 25. Electric fields used in cyclic accelerators are variable and associated with a variable magnetic field and resulting in some close paths along which the kinetic energy gained by the particle differs from zero. If the particle is made to follow such a closed path many times over, one obtains a process of gradual acceleration that is not limited to the maximum voltage drop existing in the accelerator. Thus the final kinetic energy of the particle is obtained by submitting the charged particle to the same, relatively small, potential difference a large number of times, each cycle adding a small amount of energy to the kinetic energy of the particle. example of a cyclic accelerator is the linac CYCLIC ACCELERATOR :
  • 26. Accelerator Building Blocks • Power source • Electron source • Accelerator structure • Beam bending system • Beam Conditioning • Collimation system • Control system • Support structure
  • 27. POWER SOURCE Before discussing about power source or details of LINAC Some basic principles of X-ray production Why??? Crook’s tube can not produce high energy electrons for medical purpose.
  • 28. Cockcroft-Walton Generator 665 kV Early Challenges • High voltage breakdown • Reliable vacuum systems • Low beam currents • Zero diagnostics • Small industrial base • Scale – big science emerges All accelerators are based on the same principle. A charged particle accelerates between a gap between two electrodes when there is a potential difference between them. E= QV
  • 30. RESONANT ACCELERATOR CONCEPT THE ACCELERATION OCCURS IN THE ELECTRIC FIELD BETWEEN CYLINDRICAL DRIFT TUBES. THE RF POWER MUST BE SYNCHRONISED WITH THE MOTION OF THE ELECTRONS, SO THAT ACCELERATION OCCURS IN EVERY CAVITY. THIS NATURALLY PRODUCES BUNCHES OF ELECTRONS ELECTRON SOURCE DRIFT TUBES
  • 31. When the first electrode is oppositely charged to the entering charged particle (i.e. an electron or proton ), the particle accelerates towards it. There is no electric field inside the tubes as they are hollow conductors. When the particle enters the first tube , the voltage is switched so that the next tube is oppositely charged, therefore it accelerates to next electrode. Each time the same magnitude of voltage is applied and so the energy of the Particle E=n Q V , is built up in steps without the needing to increase the voltage.
  • 32. When the first electrode is oppositely charged to the entering charged particle (i.e. an electron or proton ), the particle accelerates towards it. There is no electric field inside the tubes as they are hollow conductors. When the particle enters the first tube , the voltage is switched so that the next tube is oppositely charged, therefore it accelerates to next electrode. Each time the same magnitude of voltage is applied and so the energy of the Particle E=n Q V , is built up in steps without the needing to increase the voltage.
  • 33. When the first electrode is oppositely charged to the entering charged particle (i.e. an electron or proton ), the particle accelerates towards it. There is no electric field inside the tubes as they are hollow conductors. When the particle enters the first tube , the voltage is switched so that the next tube is oppositely charged, therefore it accelerates to next electrode. Each time the same magnitude of voltage is applied and so the energy of the Particle E=n Q V , is built up in steps without the needing to increase the voltage.
  • 34. POWER SOURCE ● An RF power source; ● A pulsed modulator. —A magnetron is a source of high power RF required for electron acceleration, while a klystron is an RF power amplifier that amplifies the low power RF generated by an RF oscillator commonly called the RF driver.
  • 35. Microwaves Microwaves are sometimes considered to be very short radio waves (high frequency and high-energy radio waves). Some important properties of microwaves are: They are reflected by metal surfaces. They heat materials if they can make atoms or molecules in the material vibrate. The amount of heating depends on the intensity of the microwave radiation, and the time that the material is exposed to the radiation. They pass through glass and plastics. They pass through the atmosphere. They pass through the ionosphere without being reflected. They are absorbed by water molecules, how well depends on the frequency (energy) of the microwaves. MAGNETRON Produces Microwaves also called RF waves.
  • 36. The magnetron is a high-powered vacuum tube, that works as self-excited microwave oscillator. Crossed electron and magnetic fields are used in the magnetron to produce the high-power output required in radar equipment. These multi-cavity devices works at frequencies ranging from approximately 600 to 30,000 megahertz. The relatively simple construction has the disadvantage that the Magnetron usually can work only on a constructively fixed frequency. Microwaves are produced by vacuum tubes devices that operate on the ballistic motion of electron controlled by magnetic or electric fields. Some different kinds of microwave emitters are the cavity magnetron, the klystron, the traveling-wave tube(TWT), the gyrotron.
  • 37. ELECTRON GUN PULSE MODULATOR MICROWAVE POWER SOURCE (MAGNETRON) CONTROL UNIT POWER SUPPLY GAS PRESSURE SYSTEM Bending magnets MLC Component Primary collimator Flattening filter Dual ionisation chamber Secondary collimator Ceramic windowAnode Vacuum Pump BeamtransportsystemLINACHead Thick X-ray target Accelerating Wave guide Patient support assembly Steering coil Cathode
  • 38. ELECTRON SOURCE It is essentially a simple electrostatic accelerator called an electron gun. Two types of electron gun are in use as sources of electrons in medical linacs — Diode type; — Triode type. Both electron gun types contain a heated filament cathode and a perforated grounded anode; in addition, the triode electron gun also incorporates a grid. Anode Cathode ELECTRON GUN Electrons are thermionically emitted from the heated cathode, focused into a pencil beam by a curved focusing electrode and accelerated towards the perforated anode through which they drift to enter the accelerating waveguide.
  • 39. ACCELERATOR STRUCTURE The simplest kind of accelerating waveguide is obtained from a cylindrical uniform waveguide by adding a series of discs (irises) with circular holes at the centre, placed at equal distances along the tube. These discs divide the waveguide into a series of cylindrical cavities that form the basic structure of the accelerating waveguide in a linac. The accelerating waveguide is evacuated to allow free propagation of electrons. The cavities of the accelerating waveguide serve two purposes: - To couple and distribute microwave power between adjacent cavities; - To provide a suitable electric field pattern for the acceleration of electrons. Two types of accelerating waveguide have been developed for the acceleration of electrons: (i) Travelling wave structure (ii) Standing wave structure.
  • 40. Electron gun RF in Accelerating cavity Accelerating wave guide RF out Electron gun RF in Coupling cavity Travelling wave structure Standing wave structure
  • 41. Electron gun RF in Accelerating cavity Accelerating wave guide RF out Electron gun RF in Coupling cavity λ λ' λ‘=λ/2
  • 42. In the travelling wave structure the microwaves enter the accelerating waveguide on the gun side and propagate towards the high energy end of the waveguide where they either are absorbed without any reflection or exit the waveguide to be absorbed in a resistive load or to be fed back to the input end of the accelerating waveguide. In the standing wave structure each end of the accelerating waveguide is terminated with a conducting disc to reflect the microwave power ,resulting in a buildup of standing waves in the waveguide. These cavities therefore serve only as coupling cavities and can be moved out to the side of the waveguide structure, effectively shortening the accelerating waveguide by 50%.
  • 43. Bending magnets are used in linacs operating at energies above 6 MeV, where the accelerating waveguides are too long for straight- through mounting. The accelerating waveguide is usually mounted parallel to the gantry rotation axis and the electron beam must be bent to make it strike the X - ray target or be able to exit through the beam exit window. BEAM BENDING SYSTEM In low energy linacs the target is embedded in the accelerating waveguide and no beam transport between the accelerating waveguide and target is required
  • 44. COLLIMATION SYSTEM The important components found in a typical head of linac include: —Retractable X ray targets; —Flattening filters and electron scattering foils (also called scattering filters); —Primary and adjustable secondary collimators; —Dual transmission ionization chambers; —A field defining light and a range finder; —retractable wedges; — MLC.
  • 45.
  • 46. The photon beam collimation is achieved with two or three collimator devices: 1. A primary collimator; 2. Secondary movable beam defining collimators; 3. An MLC (optional). The primary collimator defines the largest available circular field size and is a conical opening machined into a tungsten shielding block, with the sides of the conical opening projecting on to edges of the target on one end of the block and on to the flattening filter on the other end
  • 47. The secondary beam defining collimators consist of four blocks, two forming the upper and two forming the lower jaws of the collimator. They can provide rectangular or square fields at the linac isocentre, with sides of the order of few mm up to 40 cm.
  • 48. The secondary beam defining collimators consist of four blocks, two forming the upper and two forming the lower jaws of the collimator. They can provide rectangular or square fields at the linac isocentre, with sides of the order of few mm up to 40 cm.
  • 49. A multileaf collimator (MLC) for photon beams consists of a large number of collimating blocks or leaves that can be driven automatically, independent of each other, to generate a field of any shape The individual leaf has a width of 1 cm or less as projected at the isocenter. The leaves are made of tungsten alloy (r = 17.0 to 18.5 g/cm3) and have thickness along the beam direction ranging from 6 cm to 7.5 cm, depending on the type of accelerator. The leaf thickness is sufficient to provide primary x-ray transmission through the leaves of less than 2%
  • 50. A multileaf collimator (MLC) for photon beams consists of a large number of collimating blocks or leaves that can be driven automatically, independent of each other, to generate a field of any shape The individual leaf has a width of 1 cm or less as projected at the isocenter. The leaves are made of tungsten alloy (r = 17.0 to 18.5 g/cm3) and have thickness along the beam direction ranging from 6 cm to 7.5 cm, depending on the type of accelerator. The leaf thickness is sufficient to provide primary x-ray transmission through the leaves of less than 2%
  • 51. BEAM CONDITIONING Flattening filter X-Ray The photon dose distribution produced by a linac is strongly forward peaked. To make beam intensity uniform across a field , a flattening filter is inserted in the beam. The filter is usually made of Pb , although tungsten, steel, Al or combination has also been suggested.
  • 52. BEAM CONDITIONING Flattening filter The photon dose distribution produced by a linac is strongly forward peaked. To make beam intensity uniform across a field , a flattening filter is inserted in the beam. The filter is usually made of Pb , although tungsten, steel, Al or combination has also been suggested. X-Ray
  • 53. Scattering foil Electron Beam The electron pencil beam exits the evacuated beam transport system through a thin window usually made of beryllium, which, with its low atomic number Z, minimizes the pencil beam scattering and X- ray production The scattering of the electron pencil beam over the relatively large area used in radiotherapy (up to 25 × 25 cm2) is achieved by placing thin foils of high Z material (copper or lead) into the pencil beam at the level of the flattening filter in the X -ray mode.
  • 54. Scattering foil The electron pencil beam exits the evacuated beam transport system through a thin window usually made of beryllium, which, with its low atomic number Z, minimizes the pencil beam scattering and X- ray production The scattering of the electron pencil beam over the relatively large area used in radiotherapy (up to 25 × 25 cm2) is achieved by placing thin foils of high Z material (copper or lead) into the pencil beam . Electron Beam
  • 61. Termination of irradiation Most common dose monitors in linacs are transmission ionization chambers permanently imbedded in the linac clinical photon and electron beams to monitor the beam output continuously during patient treatment Dual ionisation chamber
  • 62. Properties of the ionization chamber monitors  Chambers must have a minimal effect on clinical photon and electron radiation beams; Chamber response should be independent of ambient temperature and pressure (most linacs use sealed ionization chambers to satisfy this condition); Chambers should be operated under saturation conditions.
  • 63. The primary ionization chamber measures MUs. Typically, the sensitivity of the chamber electrometer circuitry is adjusted in such a way that 1 MU corresponds to a dose of 1 cGy delivered in a water phantom at the depth of dose maximum on the central beam axis when irradiated with a 10 × 10 cm2 field at a source to surface distance (SSD) of 100 cm. Once the operator preset number of MUs has been reached, the primary ionization chamber circuitry shuts the linac down and terminates the dose delivery to the patient. Before a new irradiation can be initiated, it is necessary to reset the MU displays to zero. Furthermore, irradiation is not possible until a new selection of MUs has been made. The dose monitoring system also monitors other operating parameters such as the beam energy, flatness and symmetry.
  • 64. Field lights and lasers It is a field localizing device. Used to display the position of the radiation field on the patient skin. The accuracy of lasers guides in determining isocentre position. A combination of mirror and a light source located in the space between the chambers and the jaws projects a light beam as if emitting from the x-ray focal spot.
  • 65. Pulse modulator A power supply provides direct current (DC) power to the modulator, which includes the pulse-forming network and a switch tube known as hydrogen thyratron. High-voltage pulses from the modulator section are flat-topped DC pulses of a few microseconds in duration. These pulses are delivered to the magnetron or klystron and simultaneously to the electron gun. Control console For monitoring and controlling linac All interlocks checks to allow beam started. To provide digital display of monitor units Mechanical beam parameters.