SlideShare a Scribd company logo
Pulse Oximeter Sensor
Our Goal
We hope to quantify the dynamics of an active
muscle.
LED spectroscopy provides a way to track fatigue
and oxygen trends of an area of interest.
Properties of the Red and IR LED
• Near-IR light has much better penetration into
biological tissue than visible light
• NIR light absorption affected by: hemoglobin,
myoglobin and cytochrome c oxidase molecules
(90%Hb)
• To track specifically muscle activity, it appears
that Infrared light will be the most appropriate
signal to track
IR signal
We know from Beer-Lambert’s law that the absorption of
an LED is a function of each absorbed substance given
their:
1. extinction factor
2. concentration
3. path length.
Our IR light detects mostly Hb/Mb-deoxygenation in
muscle tissue but we cannot account for path length
using 1 LED.
Path Length
For a fixed optode (forehead) we could assume
path length to be constant.
However, we are interested in quantifying
muscle movement within exercise, and our
signal includes an initial contraction of muscle
(change of path length)
Accounting for path length
The optical path length was accounted for by
measuring the absorbance changes obtained during
cuff ischemia.
The percent deoxygenation during exercise is then
calculated from the maximal deoxygenation of the
ischemic response.
A way of normalizing NIRS signals across subjects.
Oxygen saturations
Boushel claims one must use multiple
wavelengths to generate algorithms for NIRS.
• Coefficients used at each wavelength
deconvolute the overlapping absorption
spectra.
• Algorithms are used for either trend
monitoring or to assess relative
concentrations.
Below are wavelengths and coefficients used in
the laboratory at Duke for assessing relative
concentrations.
Measurements with Two
Wavelengths
Other NIRS applications use two near-IR LEDs
(760 and 850nm) and use the difference
between the signals to determine relative
oxygen saturation. The sum of the signals to
provide a reference to blood volume changes.
Red-IR= relative O2
While I am using a
Red LED and IR LED,
the data shows
plausible oxygen
trends in that the
bicep oxygen content
desaturates with
increased load in
isometric exercise.
Infrared Trends
We changed the experiment set up
Infrared signals have shown a promising fatigue
trend in almost all of our samples.
Analyzing IR trends
Static load in some types of activity compromise
blood flow to working muscle.
Are our IR decline trends due to decreased
blood flow, or is the use of oxygen in the muscle
exceeding a continuous supply?
Moving Forward
Ultrasound & IR signal of Nellcor sensor in
cooperation. Separate muscle influence and oxygen
Classify experiment activity, static load or active
Use multiple NIRS LEDs?
• Overlapping spectra= trend monitoring/ find
relative concentrations of each molecule.

More Related Content

What's hot

Pulse oximetry
Pulse oximetryPulse oximetry
Pulse oximetry
subhasish karmakar
 
Capnography
Capnography Capnography
Capnography
Mahtab Ansari
 
Pulse oximeter
Pulse oximeterPulse oximeter
Pulse oximeter
Lajpat Rai
 
Pulseoximetry fianl
Pulseoximetry fianlPulseoximetry fianl
Pulseoximetry fianl
Biswas Lohani
 
Pulse oximeter
Pulse oximeterPulse oximeter
Pulse oximeter
Mahesh Sivaji
 
How Pulse oximeter works- Principle and Limitations I SpO2 I Respiratory Phys...
How Pulse oximeter works- Principle and Limitations I SpO2 I Respiratory Phys...How Pulse oximeter works- Principle and Limitations I SpO2 I Respiratory Phys...
How Pulse oximeter works- Principle and Limitations I SpO2 I Respiratory Phys...
HM Learnings
 
Et co2
Et co2Et co2
Et co2
Vijay Anand
 
PULSE OXIMETRY
PULSE OXIMETRYPULSE OXIMETRY
PULSE OXIMETRY
MAHESWARI JAIKUMAR
 
Cardiac output monitoring
Cardiac output monitoringCardiac output monitoring
Cardiac output monitoring
mauryaramgopal
 
Pulse oximetry & capnography
Pulse oximetry & capnographyPulse oximetry & capnography
Pulse oximetry & capnography
Sapan Jena
 
cardiac output pptx
cardiac output pptxcardiac output pptx
cardiac output pptx
ananya nanda
 
Capnography
Capnography  Capnography
Capnography
Anwar Basha
 
Capnography vs plethysmography
Capnography vs plethysmographyCapnography vs plethysmography
Capnography vs plethysmography
Sourav Mondal
 
Monitoring Modality in anesthesia
Monitoring Modality  in anesthesiaMonitoring Modality  in anesthesia
Monitoring Modality in anesthesia
Dr.RMLIMS lucknow
 
Pulse Oxymetry
Pulse OxymetryPulse Oxymetry
Pulse Oxymetry
MR. JAGDISH SAMBAD
 
CAPNOGRAPHY
CAPNOGRAPHYCAPNOGRAPHY
CAPNOGRAPHY
MAHESWARI JAIKUMAR
 
Pulseoximetry
PulseoximetryPulseoximetry
Pulseoximetrydrriyas03
 
Uses of Ultrasound in Anesthesiology
Uses of Ultrasound in AnesthesiologyUses of Ultrasound in Anesthesiology
Uses of Ultrasound in Anesthesiology
Saneesh P J
 
Multipara monitor
Multipara monitorMultipara monitor
Multipara monitor
RiyaBaghele
 

What's hot (20)

Pulse oximetry
Pulse oximetryPulse oximetry
Pulse oximetry
 
Capnography
Capnography Capnography
Capnography
 
Pulse oximeter
Pulse oximeterPulse oximeter
Pulse oximeter
 
Pulse oximetry
Pulse oximetryPulse oximetry
Pulse oximetry
 
Pulseoximetry fianl
Pulseoximetry fianlPulseoximetry fianl
Pulseoximetry fianl
 
Pulse oximeter
Pulse oximeterPulse oximeter
Pulse oximeter
 
How Pulse oximeter works- Principle and Limitations I SpO2 I Respiratory Phys...
How Pulse oximeter works- Principle and Limitations I SpO2 I Respiratory Phys...How Pulse oximeter works- Principle and Limitations I SpO2 I Respiratory Phys...
How Pulse oximeter works- Principle and Limitations I SpO2 I Respiratory Phys...
 
Et co2
Et co2Et co2
Et co2
 
PULSE OXIMETRY
PULSE OXIMETRYPULSE OXIMETRY
PULSE OXIMETRY
 
Cardiac output monitoring
Cardiac output monitoringCardiac output monitoring
Cardiac output monitoring
 
Pulse oximetry & capnography
Pulse oximetry & capnographyPulse oximetry & capnography
Pulse oximetry & capnography
 
cardiac output pptx
cardiac output pptxcardiac output pptx
cardiac output pptx
 
Capnography
Capnography  Capnography
Capnography
 
Capnography vs plethysmography
Capnography vs plethysmographyCapnography vs plethysmography
Capnography vs plethysmography
 
Monitoring Modality in anesthesia
Monitoring Modality  in anesthesiaMonitoring Modality  in anesthesia
Monitoring Modality in anesthesia
 
Pulse Oxymetry
Pulse OxymetryPulse Oxymetry
Pulse Oxymetry
 
CAPNOGRAPHY
CAPNOGRAPHYCAPNOGRAPHY
CAPNOGRAPHY
 
Pulseoximetry
PulseoximetryPulseoximetry
Pulseoximetry
 
Uses of Ultrasound in Anesthesiology
Uses of Ultrasound in AnesthesiologyUses of Ultrasound in Anesthesiology
Uses of Ultrasound in Anesthesiology
 
Multipara monitor
Multipara monitorMultipara monitor
Multipara monitor
 

Viewers also liked

Pulse oximeter - SpO2
Pulse oximeter - SpO2Pulse oximeter - SpO2
Pulse oximeter - SpO2
Nguyễn Linh
 
Pulse Oximetry
Pulse OximetryPulse Oximetry
Pulse Oximetry
The Apprentice Doctor
 
Diagnosis of testicular torsion using near infrared spectroscopy
Diagnosis of testicular torsion using near infrared spectroscopyDiagnosis of testicular torsion using near infrared spectroscopy
Diagnosis of testicular torsion using near infrared spectroscopy
National Institute Of Child Health (N.I.C.H) Karachi
 
Monitoring of bladder filling to capacity
Monitoring of bladder filling to capacityMonitoring of bladder filling to capacity
Monitoring of bladder filling to capacity
Keerthi Kancharla
 
Monitoring the injured brain
Monitoring the injured brainMonitoring the injured brain
Monitoring the injured brain
Barbara Stanley
 
neurophysiologic monitoring final
neurophysiologic monitoring finalneurophysiologic monitoring final
Near Infrared Spectroscopy In Off Line Biomass Monitoring Of Candida Utilis C...
Near Infrared Spectroscopy In Off Line Biomass Monitoring Of Candida Utilis C...Near Infrared Spectroscopy In Off Line Biomass Monitoring Of Candida Utilis C...
Near Infrared Spectroscopy In Off Line Biomass Monitoring Of Candida Utilis C...
bengreenman
 
Pulse oxymetry
Pulse oxymetryPulse oxymetry
Project Pelican Pulse Oximeter Presentations
Project Pelican Pulse Oximeter PresentationsProject Pelican Pulse Oximeter Presentations
Project Pelican Pulse Oximeter PresentationsLéah Chung
 
Pulse oximetry slidefinal abc
Pulse oximetry slidefinal abcPulse oximetry slidefinal abc
Pulse oximetry slidefinal abc
Jibi Life
 

Viewers also liked (12)

Pulse oximeter - SpO2
Pulse oximeter - SpO2Pulse oximeter - SpO2
Pulse oximeter - SpO2
 
Pulse Oximetry
Pulse OximetryPulse Oximetry
Pulse Oximetry
 
use of pulse oximetry
use of pulse oximetryuse of pulse oximetry
use of pulse oximetry
 
Diagnosis of testicular torsion using near infrared spectroscopy
Diagnosis of testicular torsion using near infrared spectroscopyDiagnosis of testicular torsion using near infrared spectroscopy
Diagnosis of testicular torsion using near infrared spectroscopy
 
Monitoring of bladder filling to capacity
Monitoring of bladder filling to capacityMonitoring of bladder filling to capacity
Monitoring of bladder filling to capacity
 
Monitoring the injured brain
Monitoring the injured brainMonitoring the injured brain
Monitoring the injured brain
 
Cerebral oximetry
Cerebral oximetryCerebral oximetry
Cerebral oximetry
 
neurophysiologic monitoring final
neurophysiologic monitoring finalneurophysiologic monitoring final
neurophysiologic monitoring final
 
Near Infrared Spectroscopy In Off Line Biomass Monitoring Of Candida Utilis C...
Near Infrared Spectroscopy In Off Line Biomass Monitoring Of Candida Utilis C...Near Infrared Spectroscopy In Off Line Biomass Monitoring Of Candida Utilis C...
Near Infrared Spectroscopy In Off Line Biomass Monitoring Of Candida Utilis C...
 
Pulse oxymetry
Pulse oxymetryPulse oxymetry
Pulse oxymetry
 
Project Pelican Pulse Oximeter Presentations
Project Pelican Pulse Oximeter PresentationsProject Pelican Pulse Oximeter Presentations
Project Pelican Pulse Oximeter Presentations
 
Pulse oximetry slidefinal abc
Pulse oximetry slidefinal abcPulse oximetry slidefinal abc
Pulse oximetry slidefinal abc
 

Similar to pulse oximetry final

New Microsoft PowerPoint Presentation.pptx
New Microsoft PowerPoint Presentation.pptxNew Microsoft PowerPoint Presentation.pptx
New Microsoft PowerPoint Presentation.pptx
AniketKumar125504
 
Oximeter
OximeterOximeter
Oximeter
Adil Khan
 
Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based...
Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based...Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based...
Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based...
IOSRJEEE
 
2_sensor_oximter.pdf
2_sensor_oximter.pdf2_sensor_oximter.pdf
2_sensor_oximter.pdf
aliamjd
 
photoplethysmogram
photoplethysmogram photoplethysmogram
photoplethysmogram
SIVAKUMAR745527
 
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPGProblems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
IJSRD
 
Non-Invasive Blood Glucose Measurement
Non-Invasive Blood Glucose MeasurementNon-Invasive Blood Glucose Measurement
Non-Invasive Blood Glucose Measurement
ijceronline
 
Co32559562
Co32559562Co32559562
Co32559562
IJERA Editor
 
U34129131
U34129131U34129131
U34129131
IJERA Editor
 
Ed4301791797
Ed4301791797Ed4301791797
Ed4301791797
IJERA Editor
 
International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER) International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER)
ijceronline
 
An Introduction to Near Infared Spectroscopy (NIRS)
An Introduction to Near Infared Spectroscopy (NIRS)An Introduction to Near Infared Spectroscopy (NIRS)
An Introduction to Near Infared Spectroscopy (NIRS)
TeamMitchell1
 
Merged document 12
Merged document 12Merged document 12
Merged document 12
V!vEk@nAnD S
 
Infrared-Spectroscopy-ppt.pptx
Infrared-Spectroscopy-ppt.pptxInfrared-Spectroscopy-ppt.pptx
Infrared-Spectroscopy-ppt.pptx
Nagen87
 
Ndir gas sensor
Ndir gas sensorNdir gas sensor
Ndir gas sensor
Rishav Pandey
 
Measuring fatigue of soldiers in wireless body area networks
Measuring fatigue of soldiers in wireless body area networksMeasuring fatigue of soldiers in wireless body area networks
Measuring fatigue of soldiers in wireless body area networks
Shah Faisal
 
PulseOxymetry_AN_Finalv3p1
PulseOxymetry_AN_Finalv3p1PulseOxymetry_AN_Finalv3p1
PulseOxymetry_AN_Finalv3p1Ahmed Ghouri
 
Biosensors
BiosensorsBiosensors
Biosensors
Mateen Shahid
 
pulseoximeter-160518140401.pptx
pulseoximeter-160518140401.pptxpulseoximeter-160518140401.pptx
pulseoximeter-160518140401.pptx
Subi Babu
 

Similar to pulse oximetry final (20)

New Microsoft PowerPoint Presentation.pptx
New Microsoft PowerPoint Presentation.pptxNew Microsoft PowerPoint Presentation.pptx
New Microsoft PowerPoint Presentation.pptx
 
Oximeter
OximeterOximeter
Oximeter
 
Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based...
Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based...Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based...
Design of the Pulse Oximetry Measurement Circuit and Its Sensing System Based...
 
2_sensor_oximter.pdf
2_sensor_oximter.pdf2_sensor_oximter.pdf
2_sensor_oximter.pdf
 
photoplethysmogram
photoplethysmogram photoplethysmogram
photoplethysmogram
 
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPGProblems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
Problems in Getting Ambulatory Blood Pressure Monitoring using Infrared PPG
 
Non-Invasive Blood Glucose Measurement
Non-Invasive Blood Glucose MeasurementNon-Invasive Blood Glucose Measurement
Non-Invasive Blood Glucose Measurement
 
Co32559562
Co32559562Co32559562
Co32559562
 
U34129131
U34129131U34129131
U34129131
 
Ed4301791797
Ed4301791797Ed4301791797
Ed4301791797
 
International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER) International Journal of Computational Engineering Research(IJCER)
International Journal of Computational Engineering Research(IJCER)
 
An Introduction to Near Infared Spectroscopy (NIRS)
An Introduction to Near Infared Spectroscopy (NIRS)An Introduction to Near Infared Spectroscopy (NIRS)
An Introduction to Near Infared Spectroscopy (NIRS)
 
Merged document 12
Merged document 12Merged document 12
Merged document 12
 
Infrared-Spectroscopy-ppt.pptx
Infrared-Spectroscopy-ppt.pptxInfrared-Spectroscopy-ppt.pptx
Infrared-Spectroscopy-ppt.pptx
 
Ndir gas sensor
Ndir gas sensorNdir gas sensor
Ndir gas sensor
 
Measuring fatigue of soldiers in wireless body area networks
Measuring fatigue of soldiers in wireless body area networksMeasuring fatigue of soldiers in wireless body area networks
Measuring fatigue of soldiers in wireless body area networks
 
Medical dissection lab pulse oximeter
Medical dissection lab pulse oximeterMedical dissection lab pulse oximeter
Medical dissection lab pulse oximeter
 
PulseOxymetry_AN_Finalv3p1
PulseOxymetry_AN_Finalv3p1PulseOxymetry_AN_Finalv3p1
PulseOxymetry_AN_Finalv3p1
 
Biosensors
BiosensorsBiosensors
Biosensors
 
pulseoximeter-160518140401.pptx
pulseoximeter-160518140401.pptxpulseoximeter-160518140401.pptx
pulseoximeter-160518140401.pptx
 

pulse oximetry final

  • 2. Our Goal We hope to quantify the dynamics of an active muscle. LED spectroscopy provides a way to track fatigue and oxygen trends of an area of interest.
  • 3.
  • 4.
  • 5.
  • 6. Properties of the Red and IR LED • Near-IR light has much better penetration into biological tissue than visible light • NIR light absorption affected by: hemoglobin, myoglobin and cytochrome c oxidase molecules (90%Hb) • To track specifically muscle activity, it appears that Infrared light will be the most appropriate signal to track
  • 7. IR signal We know from Beer-Lambert’s law that the absorption of an LED is a function of each absorbed substance given their: 1. extinction factor 2. concentration 3. path length. Our IR light detects mostly Hb/Mb-deoxygenation in muscle tissue but we cannot account for path length using 1 LED.
  • 8. Path Length For a fixed optode (forehead) we could assume path length to be constant. However, we are interested in quantifying muscle movement within exercise, and our signal includes an initial contraction of muscle (change of path length)
  • 9. Accounting for path length The optical path length was accounted for by measuring the absorbance changes obtained during cuff ischemia. The percent deoxygenation during exercise is then calculated from the maximal deoxygenation of the ischemic response. A way of normalizing NIRS signals across subjects.
  • 10. Oxygen saturations Boushel claims one must use multiple wavelengths to generate algorithms for NIRS. • Coefficients used at each wavelength deconvolute the overlapping absorption spectra. • Algorithms are used for either trend monitoring or to assess relative concentrations.
  • 11. Below are wavelengths and coefficients used in the laboratory at Duke for assessing relative concentrations.
  • 12. Measurements with Two Wavelengths Other NIRS applications use two near-IR LEDs (760 and 850nm) and use the difference between the signals to determine relative oxygen saturation. The sum of the signals to provide a reference to blood volume changes.
  • 13. Red-IR= relative O2 While I am using a Red LED and IR LED, the data shows plausible oxygen trends in that the bicep oxygen content desaturates with increased load in isometric exercise.
  • 14. Infrared Trends We changed the experiment set up Infrared signals have shown a promising fatigue trend in almost all of our samples.
  • 15.
  • 16. Analyzing IR trends Static load in some types of activity compromise blood flow to working muscle. Are our IR decline trends due to decreased blood flow, or is the use of oxygen in the muscle exceeding a continuous supply?
  • 17. Moving Forward Ultrasound & IR signal of Nellcor sensor in cooperation. Separate muscle influence and oxygen Classify experiment activity, static load or active Use multiple NIRS LEDs? • Overlapping spectra= trend monitoring/ find relative concentrations of each molecule.

Editor's Notes

  1. The basis for our sensor are two LED signals, Red and Infrared. They have wavelengths of 660 and 910nm, respectively.
  2.   When there is high oxygen saturation, the IR pulse amplitude is higher, and Red pulse amplitude greater for lower oxygen saturation. A red-to-infrared pulse modulation ratio is needed to determine Oxygen saturation.
  3. We have a Nellcor SpO2 Forehead Sensor that was designed to track poor pulse perfusion as fast as possible as arterial blood reaches the head before the fingers. Nellcor products have monitors that account for calibration curves that correlate the pulse Modulation Ratio to oxygen saturation. Using that system, oxygen content for an area of interest is easily monitored. However, we are using a TI SP02-FE EVM dac that allows us to capture raw Infrared and Red LED signals from the sensor.
  4. A chromophore is the part of a molecule responsible for its color. The color arises when a molecule absorbs certain wavelengths of visible light and transmits or reflects others.
  5. Cut off the supply of oxygen to the area of interest and use the max deoxygenation measurement to calculate the oxygenation during exercise
  6. (scaled to the maximum and minimum values). However, without accounting for the influence of tissue and the change within the muscle, there exists the problem of normalizing the data between people, and determining true oxygen saturation changes. You can also see some activity in the signal with the contraction in the first few seconds, and it doesn’t seem intuitive to assume that is due to oxygen saturation changes over milliseconds.
  7. The experiments were taken over 5 minutes with the subject maintaining a steady force of 30% of their maximum contraction against a table top. They were seated with their arm at about 90 degrees, without any support and the maximum force was calibrated for each person. HOPES: to track fatigue over a longer time period, paying attention to IR and less on initial contraction fluctuations in the data— put aside path length
  8. We hope to get an ultrasound that will be able to track muscle diameter in cooperation with the Nellcor sensor’s Infrared signal. Ideally, the ultrasound will be a useful tool to provide a real-time picture of the muscle, and I think we will be able to better account for the muscle thickness in the individual signal. Something we may want to consider is using multiple Near-Infrared LEDs as they penetrate to the muscular level well. We could explore how to deconvolute the overlapping absorption spectra to trend monitoring or to assess relative concentrations. Keep in mind the type of activity we are tracking– keep in mind the intramuscular pressure and it’s affect on blood flow.