SlideShare a Scribd company logo
GRIDS
MUHAMMED ASLAM BS
BSC MIT
OBJECTIVES
1. DEFINITION
2. CONSTRUCTION OF GRIDS
3. PATTERNS OF GRIDS
4. TYPES OF GRIDS
5. EFFICIENCY OF GRID PERFORMANCE
6. GRID CUT-OFF
DEFINITION
• A radiographic grid is a device that is
placed between the anatomic area of
interest and the image receptor to absorb
scatter radiation exiting the patient.
• Invented in 1913 by Gustave Bucky
• Most effective means for limiting the
amount of scatter radiation
GRIDS
• Scatter radiation is detrimental to image
quality and decreases radiographic
contrast.
• Grids are used only when the anatomic
part is 10 cm or greater in thickness and
more than 60 kVp is needed for
examination.
GRID CONSTRUCTION
• Grids contain thin lead strips or lines
separated by x-ray transparent spacers.
• Interspace material is made up of
aluminium.
• Lead lines and interspace material are
covered by an aluminium front and back
panel.
• Main purpose of inerspace material is to
support the thin lead strips.
WHY ALUMINIUM?
• Aluminium :1.structurally stronger
2. absorbs both primary
radiation, thus increasing patient dose,
however it also absorbs more secondary
radiation thereby improving contrast.
• GRID RATIO:
it is the ratio of height of the lead strips
to the distance between the strips.
r = h/D
Where r is the grid ratio
h is the height of the lead strips
D is the width of the spaces
between the strips.
Example :if the lead strips are 2mm high and
the space separating them is 0.4 mm,grid
ratio is 2/0.4=5:1
GRID RATIO
• Grid ratio range from 4:1 to 16:1,where the
first number is the actual ratio and second
number is always 1.
• Higher the grid ratio straighter the rays
have to be to get through the interspaces,
better is the absorption of scattered
radiation and better is the radiographic
contrast.
• GRID FREQUENCY:
expresses the number of
lead lines per unit length, in inches ,
centimetres or both.
• Can range in value from 25-80lines/cm.
• Lower grid frequency (thicker strips) will
be more efficient in removing scattered
radiation and improving contrast , because
of higher lead content.
• But more primary radiation is also
absorbed and hence it requires larger
exposure for the same film density.
• As the grid frequency increases ,the grid
ratio must also be increased to maintain
same efficiency.
• Lead content = indicator of performance of
the grid.
GRID PATTERN
• Refers to orientation of the lead strips in
the longitudinal axis.
• Two types:
1.Linear grid
2.Crossed or cross-hatched grid
LINEAR GRIDS
• Lead lines run in only one direction. They are
parallel to each other in their longitudinal axis.
• They allow angulation of the x-ray tube along
the length of lead lines
LINEAR GRID
CROSSED GRID
• Lead lines run at right angle to one another.
• Crossed grids remove more scattered
photons than linear grids because they
contain more lead strips, oriented in two
directions.
Disadvantage: x-ray tube cannot be angled in
any direction without producing grid cut-off
CROSSED GRID
GRID FOCUS
• Refers to orientation of lead lines to one
another.
• Two types-
1.Parallel grid(non-focussed)
2.Focussed grid
PARALLEL GRIDS
• Lead lines run parallel to one another.
• Used primarily in fluoroscopy and mobile
imaging
FOCUSSED GRIDS
• A focussed grid has lead lines that are
angled, or canted, to match approximately
the angle of divergence of the primary
beam.
• These allow more transmitted photons to
reach the image receptor than parallel
grids
FOCUSSED GRIDS
CONVERGENT POINT:
If imaginary lines were drawn from each
of the lead lines in a focussed grid, these
lines would meet to form an imaginary
point called convergent point.
CONVERGENT LINE:
• If points were connected along the length
of the grid , they would form an imaginary
line called convergent line.
• FOCAL DISTANCE (GRID RADIUS):
It is the distance between the grid and the
convergent line or point.
• FOCAL RANGE:
Recommended range of SIDs that can be
used with focussed grid.
• In practice , grids have a focussing range
that indicates the distance within which the
grid can be used without significant loss
of primary radiation.
CONVEREGRNT POINT AND LINE
FOCAL RANGE
• The convergent line or point always falls in
the focal range.
• The focussing range is fairly wide for a low
grid ratio and narrow for a high grid ratio.
• Focal range is indicated on top of the grids
by manufacturers.
TYPES
STATIONARY MOVING
STATIONARY GRIDS
1. Wafer grids: matches the size of the
cassette and is used by placing it on top
of the image receptor.
2. Grid cassette : image receptor that has a
grid permanently mounted to its front
surface.
3. Grid cap: permanently mounted grid and
allows the image receptor to slide in
behind it.
MOVING GRIDS
• Invented by Dr. Hollis E . Potter in 1920.
• Also known as reciprocating grids or
Potter-Bucky diaphragm.
• Located directly below the radiographic
tabletop , the grid is found just above the
tray that holds the image receptor.
• These grids continuously move 1-3 cm
back and forth throughout the exposure.
DIRECTION OF GRID TRAVEL
• While moving parallel to its surface , the grid
must move perpendicular to the long axis of
its lead strips.
• LONG DIMENSION GRIDS
Linear grids having lead strips running
parallel to the long axis of the grid.
• SHORT DIMENSION GRIDS
Linear grids having strips running
perpendicular to the long axis of the grid.
useful for examination where it is difficult
to centre the central ray correctly for long
dimension grid
GRID PERFORMNCE
1. Primary transmission(Tp).
2. Bucky factor or Grid Conversion factor(B)
3. Contrast Improvement Factor(K)
PRIMARY TRANSMISSION
• Primary transmission is the measurement
of the percentage of primary radiation
transmitted through the grid.
Tp = Ip/I’p X 100
Where Tp is the primary transmission
Ip is the intensity with grid
I’pis the intensity without grid
APPARATUS TO MEASURE Tp
BUCKY FACTOR
• It is the ratio of incident radiation falling on
the grid to the transmitted radiation
passing through the grid.
• Bucky factor can be used to determine the
adjustment in mAs needed when changing
from using a grid to non grid (or vice
versa)
• GCF or Bucky Factor = mAs with grid
mAs without grid
• Generally ,high ratio grids absorb more
scatter radiation and have larger Bucky
Factors than low ratio grids.
APPARATUS TO MEASURE BUCKY
FACTOR
BUCKY FACTOR
GRID RATIO GCF
NO GRID 1
5:1 2
6:1 3
8:1 4
12:1 5
16:1 6
CONTRAST IMPROVEMENT
FACTOR
• It is the ratio of the contrast with the grid to
the contrast without the grid.
• This is the measure of grid’s ability to
improve contrast , which is its primary
function.
• K factor depends on kVp,field size and
part thickness
• To permit comparison of different grids ,the
contrast improvement factor is usually
determined at 100kVp with a large field
and a phantom of 20 cm thick.
SELECTIVITY
• Selectivity = ratio of primary radiation
transmitted through the grid to the
scattered radiation transmitted through the
grid.
• Larger the ratio , greater will be the
selectivity of the grid.
GRID CUT-OFF
• Grid cut-off refers to decrease in the
number of transmitted photons that reach
the image receptor because of some
misallignment of the grid.
• Higher grid ratio and short grid –focus
distance results in more potential grid cut-
off.
• Grid cut-off is a result of poor geometric
relationship between the primary beam
and lead foil strips of the grid.
• When the projected images of the lead
strips are thicker than the width of the
interspaces , cut-off is complete and no
primary radiation reaches the film.
• The resultant radiograph will be light in the
area in which cut-off occurs.
TYPES OF GRID CUT-OFF
ERRORS
1. Upside –down focussed grid
2. Lateral decentering(off-centre)
3. Focus-grid distance decentering(off-
focus)
4. Combined lateral and focus-grid distance
decentering.
UPSIDDE DOWN FOCUSSED
GRID
• This occurs when a focussed grid is
placed upside down on the image recptor ,
resulting in the grid lines going opposite
the angle of divergence of the x-ray beam.
• Radiographically , there is significant loss
of exposure along the edges of the image.
• Photons easily pass through the centre of
the grid because the lead lines are
perpendicular to the image receptor
surface.
• The tilted strips near the edges of the grid
will absorb progressively more radiation,
causing severe peripheral cut-off.
• However with parallel grids either side
may face the tube or film.
LATERAL DECENTERING
• It occurs when the central ray of x-ray
beam is not aligned from side to side with
the centre of focussed grid.
• All the lead strips cut off the same amount
of primary radiation.
• Hence, there is uniform loss of radiation
over the entire surface of the grid
,producing a uniformly light radiograph.
• FACTORS AFFECTING MAGNITUDE OF
CUT –OFF:
1. Grid ratio
2.Focal distance
3. Amount of decentering
L = rb/foX100
where r is the grid ratio
b is the lateral decentering
distance
fo is the focal distance of grid
• Amount of cut-off increases as the grid
ratio and decentering distance increases,
and cut-off decreases as the focal
distance increases.
• In portable radiography, low ratio grids and
long focal distances should be used .
OFF-LEVEL GRIDS
• Off level grid cut off results when the x-ray beam is
angled across the lead strips.
• Occur from either the tube or the grid being angled.
• Often seen with portable radiograph or horizontal
beam examinations.
• Only type that occurs with bth focussed and parallel
grids
OFF-LEVEL GRID
• IMPORTANT RELATIONSHIP:
Angling the x-ray tube across the
grid lines or angling the grid itself during
exposure produces an overall decrease in
exposure on the radiograph.
FOCUS-GRID DISTANCE DECENTERING
• Target of the x-ray tube is correctly centered
to the grid,but it is positioned above or below
the convergent line.
• Far focus-grid distance decentering-target is
above
• Near focus-grid distance decentering-target is
below
NEAR FOCUS GRID DISTANCE
DECENTERING
FAR FOCUS GRID DISTANCE
DECENTERING
• The loss of primary radiation increases as the
grid ratio increases and as the distance from
the centre of the grid increases.
• Loss is greater with near than far-focus grid
distance.
COMBINED
LATERAL AND FOCUS-GRID DISTANCE
DECENTERING
• Easy to recognise
• Causes an uneven exposure , resulting in a
film that is dark on one side and light on the
other side.
PATIENT PROTECTION ALERT
• GRID SELECTION:
In order to keep the patient exposure as
low as possible , grids should be used only
when appropriate , and the grid ratio
should be lowest that would provide
sufficient contrast improvement .
• below 90 kVp-8:1,above 90 kVp 12:1
CONCLUSION
• Construction of the grid-lead strips separated by
x-ray transparent spacers.
• They are used to absorb scatter radiation and
improve radiographic image contrast
• Generally high ratio grids with high lead content
have the highest contrast improvement factors.
• The amount of grid cut-off with a particular
decentering error is greatest with high ratio grids
and short focussing distance .
• Grid selection involves compromise between film
quality and patient exposure.
THANK YOU

More Related Content

What's hot

Filters and beam restrictors
Filters  and beam restrictorsFilters  and beam restrictors
Filters and beam restrictors
vishwanath0908
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
parthajyotidas11
 
Filters
FiltersFilters
Filters
aslam bs
 
Anode heel effect, line focus principle,
Anode heel effect, line focus principle,Anode heel effect, line focus principle,
Anode heel effect, line focus principle,
InosRagan
 
Intensifying screens
Intensifying screensIntensifying screens
Intensifying screensEddy Rumhadi
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
Anjan Dangal
 
digital radiography
digital radiographydigital radiography
digital radiography
shokoofeh mousavi
 
Mobile & portable radiography
Mobile & portable radiographyMobile & portable radiography
Mobile & portable radiography
Priyanka Parimala
 
X ray filters
X ray filtersX ray filters
X ray filters
SaruGosain
 
Digital radiography
Digital radiographyDigital radiography
Digital radiography
haribudke
 
MACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxMACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptx
ThejaTej6
 
Intensifying screen avinesh shrestha
Intensifying screen avinesh shresthaIntensifying screen avinesh shrestha
Intensifying screen avinesh shrestha
Avinesh Shrestha
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image qualityRad Tech
 
X ray tube
X ray tubeX ray tube
X ray tube
Bipin Joseph
 
Faults in x ray tube and its care
Faults in x ray tube and its careFaults in x ray tube and its care
Faults in x ray tube and its care
SaruGosain
 
X ray beam rest
X ray beam restX ray beam rest
X ray beam rest
Buandari
 
Line focus principle.
Line focus principle.Line focus principle.
Line focus principle.
mohammed althaf
 
Beam restriction
Beam restrictionBeam restriction
Beam restriction
aslam bs
 
Filters grids and collimators
Filters grids and collimatorsFilters grids and collimators
Filters grids and collimators
Prajwith Rai
 

What's hot (20)

Filters and beam restrictors
Filters  and beam restrictorsFilters  and beam restrictors
Filters and beam restrictors
 
Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
 
Filters
FiltersFilters
Filters
 
Anode heel effect, line focus principle,
Anode heel effect, line focus principle,Anode heel effect, line focus principle,
Anode heel effect, line focus principle,
 
Intensifying screens
Intensifying screensIntensifying screens
Intensifying screens
 
Computed radiography
Computed radiographyComputed radiography
Computed radiography
 
digital radiography
digital radiographydigital radiography
digital radiography
 
TLD
TLDTLD
TLD
 
Mobile & portable radiography
Mobile & portable radiographyMobile & portable radiography
Mobile & portable radiography
 
X ray filters
X ray filtersX ray filters
X ray filters
 
Digital radiography
Digital radiographyDigital radiography
Digital radiography
 
MACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxMACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptx
 
Intensifying screen avinesh shrestha
Intensifying screen avinesh shresthaIntensifying screen avinesh shrestha
Intensifying screen avinesh shrestha
 
Radiographic exposure and image quality
Radiographic exposure and image qualityRadiographic exposure and image quality
Radiographic exposure and image quality
 
X ray tube
X ray tubeX ray tube
X ray tube
 
Faults in x ray tube and its care
Faults in x ray tube and its careFaults in x ray tube and its care
Faults in x ray tube and its care
 
X ray beam rest
X ray beam restX ray beam rest
X ray beam rest
 
Line focus principle.
Line focus principle.Line focus principle.
Line focus principle.
 
Beam restriction
Beam restrictionBeam restriction
Beam restriction
 
Filters grids and collimators
Filters grids and collimatorsFilters grids and collimators
Filters grids and collimators
 

Similar to Grids

grids-160225101450 (1).pptx
grids-160225101450 (1).pptxgrids-160225101450 (1).pptx
grids-160225101450 (1).pptx
VishnuDutt40
 
GRIDS IS PRESENTATION IN RADIO PHYSICS MUST IMPORTANT
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANTGRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT
GRIDS IS PRESENTATION IN RADIO PHYSICS MUST IMPORTANT
Raghavendarkishore
 
GRIDS IS PRESENTATION IN RADIO PHYSICS MUST IMPORTANT
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANTGRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT
GRIDS IS PRESENTATION IN RADIO PHYSICS MUST IMPORTANT
Raghavendarkishore
 
grids,films.pptx
grids,films.pptxgrids,films.pptx
grids,films.pptx
rohanjohnjacob
 
Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.
Arya Prasad
 
Filters grids and collimators
Filters grids and collimatorsFilters grids and collimators
Filters grids and collimators
Georgina George
 
XRAY GRIDS (1).pptx
XRAY GRIDS (1).pptxXRAY GRIDS (1).pptx
XRAY GRIDS (1).pptx
DRVISHAL12
 
Rad 206 p12 Fundamentals of Imaging - Control of Scatter Radiation
Rad 206 p12 Fundamentals of Imaging - Control of Scatter RadiationRad 206 p12 Fundamentals of Imaging - Control of Scatter Radiation
Rad 206 p12 Fundamentals of Imaging - Control of Scatter Radiation
sehlawi
 
Grids JP .ppt
Grids JP .pptGrids JP .ppt
Grids JP .ppt
jayapandiyan Paraman
 
grid.pptx
grid.pptxgrid.pptx
grid.pptx
VishnuDutt40
 
GRIDS.ppt
GRIDS.pptGRIDS.ppt
GRIDS.ppt
sudheendrapv
 
GRIDS.ppt
GRIDS.pptGRIDS.ppt
MicroStrip Antenna
MicroStrip AntennaMicroStrip Antenna
MicroStrip Antenna
Tarek Nader
 
Xray filters beam restrictors and Grids Radiodiagnosis.ppt
Xray filters beam restrictors and Grids Radiodiagnosis.pptXray filters beam restrictors and Grids Radiodiagnosis.ppt
Xray filters beam restrictors and Grids Radiodiagnosis.ppt
Nagasai Pelala
 
Filters.ppt
Filters.pptFilters.ppt
Neelotpal Mitra
Neelotpal MitraNeelotpal Mitra
Neelotpal Mitra
Neelotpal Mitra
 
Novel band microstrip antenna
Novel band microstrip antennaNovel band microstrip antenna
Novel band microstrip antenna
Raghuvardhan Kumar
 
Micro strip Antenna
Micro strip AntennaMicro strip Antenna
Micro strip Antenna
subash_kotha
 
Millimeter wave circular microstrip Patch antenna for 5 g applications
Millimeter wave circular microstrip Patch antenna for 5 g applicationsMillimeter wave circular microstrip Patch antenna for 5 g applications
Millimeter wave circular microstrip Patch antenna for 5 g applications
Gana U Kumar
 
Optical Communications Systems Lecture 1(1).pptx
Optical Communications Systems Lecture 1(1).pptxOptical Communications Systems Lecture 1(1).pptx
Optical Communications Systems Lecture 1(1).pptx
SamanZebari1
 

Similar to Grids (20)

grids-160225101450 (1).pptx
grids-160225101450 (1).pptxgrids-160225101450 (1).pptx
grids-160225101450 (1).pptx
 
GRIDS IS PRESENTATION IN RADIO PHYSICS MUST IMPORTANT
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANTGRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT
GRIDS IS PRESENTATION IN RADIO PHYSICS MUST IMPORTANT
 
GRIDS IS PRESENTATION IN RADIO PHYSICS MUST IMPORTANT
GRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANTGRIDS IS PRESENTATION IN RADIO PHYSICS  MUST IMPORTANT
GRIDS IS PRESENTATION IN RADIO PHYSICS MUST IMPORTANT
 
grids,films.pptx
grids,films.pptxgrids,films.pptx
grids,films.pptx
 
Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.
 
Filters grids and collimators
Filters grids and collimatorsFilters grids and collimators
Filters grids and collimators
 
XRAY GRIDS (1).pptx
XRAY GRIDS (1).pptxXRAY GRIDS (1).pptx
XRAY GRIDS (1).pptx
 
Rad 206 p12 Fundamentals of Imaging - Control of Scatter Radiation
Rad 206 p12 Fundamentals of Imaging - Control of Scatter RadiationRad 206 p12 Fundamentals of Imaging - Control of Scatter Radiation
Rad 206 p12 Fundamentals of Imaging - Control of Scatter Radiation
 
Grids JP .ppt
Grids JP .pptGrids JP .ppt
Grids JP .ppt
 
grid.pptx
grid.pptxgrid.pptx
grid.pptx
 
GRIDS.ppt
GRIDS.pptGRIDS.ppt
GRIDS.ppt
 
GRIDS.ppt
GRIDS.pptGRIDS.ppt
GRIDS.ppt
 
MicroStrip Antenna
MicroStrip AntennaMicroStrip Antenna
MicroStrip Antenna
 
Xray filters beam restrictors and Grids Radiodiagnosis.ppt
Xray filters beam restrictors and Grids Radiodiagnosis.pptXray filters beam restrictors and Grids Radiodiagnosis.ppt
Xray filters beam restrictors and Grids Radiodiagnosis.ppt
 
Filters.ppt
Filters.pptFilters.ppt
Filters.ppt
 
Neelotpal Mitra
Neelotpal MitraNeelotpal Mitra
Neelotpal Mitra
 
Novel band microstrip antenna
Novel band microstrip antennaNovel band microstrip antenna
Novel band microstrip antenna
 
Micro strip Antenna
Micro strip AntennaMicro strip Antenna
Micro strip Antenna
 
Millimeter wave circular microstrip Patch antenna for 5 g applications
Millimeter wave circular microstrip Patch antenna for 5 g applicationsMillimeter wave circular microstrip Patch antenna for 5 g applications
Millimeter wave circular microstrip Patch antenna for 5 g applications
 
Optical Communications Systems Lecture 1(1).pptx
Optical Communications Systems Lecture 1(1).pptxOptical Communications Systems Lecture 1(1).pptx
Optical Communications Systems Lecture 1(1).pptx
 

Recently uploaded

Overview on Edible Vaccine: Pros & Cons with Mechanism
Overview on Edible Vaccine: Pros & Cons with MechanismOverview on Edible Vaccine: Pros & Cons with Mechanism
Overview on Edible Vaccine: Pros & Cons with Mechanism
DeeptiGupta154
 
How to Break the cycle of negative Thoughts
How to Break the cycle of negative ThoughtsHow to Break the cycle of negative Thoughts
How to Break the cycle of negative Thoughts
Col Mukteshwar Prasad
 
Supporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptxSupporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptx
Jisc
 
Thesis Statement for students diagnonsed withADHD.ppt
Thesis Statement for students diagnonsed withADHD.pptThesis Statement for students diagnonsed withADHD.ppt
Thesis Statement for students diagnonsed withADHD.ppt
EverAndrsGuerraGuerr
 
Instructions for Submissions thorugh G- Classroom.pptx
Instructions for Submissions thorugh G- Classroom.pptxInstructions for Submissions thorugh G- Classroom.pptx
Instructions for Submissions thorugh G- Classroom.pptx
Jheel Barad
 
CLASS 11 CBSE B.St Project AIDS TO TRADE - INSURANCE
CLASS 11 CBSE B.St Project AIDS TO TRADE - INSURANCECLASS 11 CBSE B.St Project AIDS TO TRADE - INSURANCE
CLASS 11 CBSE B.St Project AIDS TO TRADE - INSURANCE
BhavyaRajput3
 
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
Nguyen Thanh Tu Collection
 
Operation Blue Star - Saka Neela Tara
Operation Blue Star   -  Saka Neela TaraOperation Blue Star   -  Saka Neela Tara
Operation Blue Star - Saka Neela Tara
Balvir Singh
 
Sectors of the Indian Economy - Class 10 Study Notes pdf
Sectors of the Indian Economy - Class 10 Study Notes pdfSectors of the Indian Economy - Class 10 Study Notes pdf
Sectors of the Indian Economy - Class 10 Study Notes pdf
Vivekanand Anglo Vedic Academy
 
special B.ed 2nd year old paper_20240531.pdf
special B.ed 2nd year old paper_20240531.pdfspecial B.ed 2nd year old paper_20240531.pdf
special B.ed 2nd year old paper_20240531.pdf
Special education needs
 
How to Split Bills in the Odoo 17 POS Module
How to Split Bills in the Odoo 17 POS ModuleHow to Split Bills in the Odoo 17 POS Module
How to Split Bills in the Odoo 17 POS Module
Celine George
 
Unit 8 - Information and Communication Technology (Paper I).pdf
Unit 8 - Information and Communication Technology (Paper I).pdfUnit 8 - Information and Communication Technology (Paper I).pdf
Unit 8 - Information and Communication Technology (Paper I).pdf
Thiyagu K
 
Additional Benefits for Employee Website.pdf
Additional Benefits for Employee Website.pdfAdditional Benefits for Employee Website.pdf
Additional Benefits for Employee Website.pdf
joachimlavalley1
 
ESC Beyond Borders _From EU to You_ InfoPack general.pdf
ESC Beyond Borders _From EU to You_ InfoPack general.pdfESC Beyond Borders _From EU to You_ InfoPack general.pdf
ESC Beyond Borders _From EU to You_ InfoPack general.pdf
Fundacja Rozwoju Społeczeństwa Przedsiębiorczego
 
How to Make a Field invisible in Odoo 17
How to Make a Field invisible in Odoo 17How to Make a Field invisible in Odoo 17
How to Make a Field invisible in Odoo 17
Celine George
 
How to Create Map Views in the Odoo 17 ERP
How to Create Map Views in the Odoo 17 ERPHow to Create Map Views in the Odoo 17 ERP
How to Create Map Views in the Odoo 17 ERP
Celine George
 
Phrasal Verbs.XXXXXXXXXXXXXXXXXXXXXXXXXX
Phrasal Verbs.XXXXXXXXXXXXXXXXXXXXXXXXXXPhrasal Verbs.XXXXXXXXXXXXXXXXXXXXXXXXXX
Phrasal Verbs.XXXXXXXXXXXXXXXXXXXXXXXXXX
MIRIAMSALINAS13
 
1.4 modern child centered education - mahatma gandhi-2.pptx
1.4 modern child centered education - mahatma gandhi-2.pptx1.4 modern child centered education - mahatma gandhi-2.pptx
1.4 modern child centered education - mahatma gandhi-2.pptx
JosvitaDsouza2
 
Mule 4.6 & Java 17 Upgrade | MuleSoft Mysore Meetup #46
Mule 4.6 & Java 17 Upgrade | MuleSoft Mysore Meetup #46Mule 4.6 & Java 17 Upgrade | MuleSoft Mysore Meetup #46
Mule 4.6 & Java 17 Upgrade | MuleSoft Mysore Meetup #46
MysoreMuleSoftMeetup
 
Introduction to Quality Improvement Essentials
Introduction to Quality Improvement EssentialsIntroduction to Quality Improvement Essentials
Introduction to Quality Improvement Essentials
Excellence Foundation for South Sudan
 

Recently uploaded (20)

Overview on Edible Vaccine: Pros & Cons with Mechanism
Overview on Edible Vaccine: Pros & Cons with MechanismOverview on Edible Vaccine: Pros & Cons with Mechanism
Overview on Edible Vaccine: Pros & Cons with Mechanism
 
How to Break the cycle of negative Thoughts
How to Break the cycle of negative ThoughtsHow to Break the cycle of negative Thoughts
How to Break the cycle of negative Thoughts
 
Supporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptxSupporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptx
 
Thesis Statement for students diagnonsed withADHD.ppt
Thesis Statement for students diagnonsed withADHD.pptThesis Statement for students diagnonsed withADHD.ppt
Thesis Statement for students diagnonsed withADHD.ppt
 
Instructions for Submissions thorugh G- Classroom.pptx
Instructions for Submissions thorugh G- Classroom.pptxInstructions for Submissions thorugh G- Classroom.pptx
Instructions for Submissions thorugh G- Classroom.pptx
 
CLASS 11 CBSE B.St Project AIDS TO TRADE - INSURANCE
CLASS 11 CBSE B.St Project AIDS TO TRADE - INSURANCECLASS 11 CBSE B.St Project AIDS TO TRADE - INSURANCE
CLASS 11 CBSE B.St Project AIDS TO TRADE - INSURANCE
 
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
 
Operation Blue Star - Saka Neela Tara
Operation Blue Star   -  Saka Neela TaraOperation Blue Star   -  Saka Neela Tara
Operation Blue Star - Saka Neela Tara
 
Sectors of the Indian Economy - Class 10 Study Notes pdf
Sectors of the Indian Economy - Class 10 Study Notes pdfSectors of the Indian Economy - Class 10 Study Notes pdf
Sectors of the Indian Economy - Class 10 Study Notes pdf
 
special B.ed 2nd year old paper_20240531.pdf
special B.ed 2nd year old paper_20240531.pdfspecial B.ed 2nd year old paper_20240531.pdf
special B.ed 2nd year old paper_20240531.pdf
 
How to Split Bills in the Odoo 17 POS Module
How to Split Bills in the Odoo 17 POS ModuleHow to Split Bills in the Odoo 17 POS Module
How to Split Bills in the Odoo 17 POS Module
 
Unit 8 - Information and Communication Technology (Paper I).pdf
Unit 8 - Information and Communication Technology (Paper I).pdfUnit 8 - Information and Communication Technology (Paper I).pdf
Unit 8 - Information and Communication Technology (Paper I).pdf
 
Additional Benefits for Employee Website.pdf
Additional Benefits for Employee Website.pdfAdditional Benefits for Employee Website.pdf
Additional Benefits for Employee Website.pdf
 
ESC Beyond Borders _From EU to You_ InfoPack general.pdf
ESC Beyond Borders _From EU to You_ InfoPack general.pdfESC Beyond Borders _From EU to You_ InfoPack general.pdf
ESC Beyond Borders _From EU to You_ InfoPack general.pdf
 
How to Make a Field invisible in Odoo 17
How to Make a Field invisible in Odoo 17How to Make a Field invisible in Odoo 17
How to Make a Field invisible in Odoo 17
 
How to Create Map Views in the Odoo 17 ERP
How to Create Map Views in the Odoo 17 ERPHow to Create Map Views in the Odoo 17 ERP
How to Create Map Views in the Odoo 17 ERP
 
Phrasal Verbs.XXXXXXXXXXXXXXXXXXXXXXXXXX
Phrasal Verbs.XXXXXXXXXXXXXXXXXXXXXXXXXXPhrasal Verbs.XXXXXXXXXXXXXXXXXXXXXXXXXX
Phrasal Verbs.XXXXXXXXXXXXXXXXXXXXXXXXXX
 
1.4 modern child centered education - mahatma gandhi-2.pptx
1.4 modern child centered education - mahatma gandhi-2.pptx1.4 modern child centered education - mahatma gandhi-2.pptx
1.4 modern child centered education - mahatma gandhi-2.pptx
 
Mule 4.6 & Java 17 Upgrade | MuleSoft Mysore Meetup #46
Mule 4.6 & Java 17 Upgrade | MuleSoft Mysore Meetup #46Mule 4.6 & Java 17 Upgrade | MuleSoft Mysore Meetup #46
Mule 4.6 & Java 17 Upgrade | MuleSoft Mysore Meetup #46
 
Introduction to Quality Improvement Essentials
Introduction to Quality Improvement EssentialsIntroduction to Quality Improvement Essentials
Introduction to Quality Improvement Essentials
 

Grids

  • 2. OBJECTIVES 1. DEFINITION 2. CONSTRUCTION OF GRIDS 3. PATTERNS OF GRIDS 4. TYPES OF GRIDS 5. EFFICIENCY OF GRID PERFORMANCE 6. GRID CUT-OFF
  • 3. DEFINITION • A radiographic grid is a device that is placed between the anatomic area of interest and the image receptor to absorb scatter radiation exiting the patient. • Invented in 1913 by Gustave Bucky • Most effective means for limiting the amount of scatter radiation
  • 5. • Scatter radiation is detrimental to image quality and decreases radiographic contrast. • Grids are used only when the anatomic part is 10 cm or greater in thickness and more than 60 kVp is needed for examination.
  • 6. GRID CONSTRUCTION • Grids contain thin lead strips or lines separated by x-ray transparent spacers. • Interspace material is made up of aluminium. • Lead lines and interspace material are covered by an aluminium front and back panel. • Main purpose of inerspace material is to support the thin lead strips.
  • 7. WHY ALUMINIUM? • Aluminium :1.structurally stronger 2. absorbs both primary radiation, thus increasing patient dose, however it also absorbs more secondary radiation thereby improving contrast.
  • 8. • GRID RATIO: it is the ratio of height of the lead strips to the distance between the strips. r = h/D Where r is the grid ratio h is the height of the lead strips D is the width of the spaces between the strips. Example :if the lead strips are 2mm high and the space separating them is 0.4 mm,grid ratio is 2/0.4=5:1
  • 10.
  • 11. • Grid ratio range from 4:1 to 16:1,where the first number is the actual ratio and second number is always 1. • Higher the grid ratio straighter the rays have to be to get through the interspaces, better is the absorption of scattered radiation and better is the radiographic contrast.
  • 12. • GRID FREQUENCY: expresses the number of lead lines per unit length, in inches , centimetres or both. • Can range in value from 25-80lines/cm. • Lower grid frequency (thicker strips) will be more efficient in removing scattered radiation and improving contrast , because of higher lead content.
  • 13. • But more primary radiation is also absorbed and hence it requires larger exposure for the same film density. • As the grid frequency increases ,the grid ratio must also be increased to maintain same efficiency. • Lead content = indicator of performance of the grid.
  • 14. GRID PATTERN • Refers to orientation of the lead strips in the longitudinal axis. • Two types: 1.Linear grid 2.Crossed or cross-hatched grid
  • 15. LINEAR GRIDS • Lead lines run in only one direction. They are parallel to each other in their longitudinal axis. • They allow angulation of the x-ray tube along the length of lead lines
  • 17. CROSSED GRID • Lead lines run at right angle to one another. • Crossed grids remove more scattered photons than linear grids because they contain more lead strips, oriented in two directions. Disadvantage: x-ray tube cannot be angled in any direction without producing grid cut-off
  • 19. GRID FOCUS • Refers to orientation of lead lines to one another. • Two types- 1.Parallel grid(non-focussed) 2.Focussed grid
  • 20. PARALLEL GRIDS • Lead lines run parallel to one another. • Used primarily in fluoroscopy and mobile imaging
  • 21.
  • 22. FOCUSSED GRIDS • A focussed grid has lead lines that are angled, or canted, to match approximately the angle of divergence of the primary beam. • These allow more transmitted photons to reach the image receptor than parallel grids
  • 24.
  • 25. CONVERGENT POINT: If imaginary lines were drawn from each of the lead lines in a focussed grid, these lines would meet to form an imaginary point called convergent point. CONVERGENT LINE: • If points were connected along the length of the grid , they would form an imaginary line called convergent line.
  • 26. • FOCAL DISTANCE (GRID RADIUS): It is the distance between the grid and the convergent line or point. • FOCAL RANGE: Recommended range of SIDs that can be used with focussed grid. • In practice , grids have a focussing range that indicates the distance within which the grid can be used without significant loss of primary radiation.
  • 29. • The convergent line or point always falls in the focal range. • The focussing range is fairly wide for a low grid ratio and narrow for a high grid ratio. • Focal range is indicated on top of the grids by manufacturers.
  • 31. STATIONARY GRIDS 1. Wafer grids: matches the size of the cassette and is used by placing it on top of the image receptor. 2. Grid cassette : image receptor that has a grid permanently mounted to its front surface. 3. Grid cap: permanently mounted grid and allows the image receptor to slide in behind it.
  • 32. MOVING GRIDS • Invented by Dr. Hollis E . Potter in 1920. • Also known as reciprocating grids or Potter-Bucky diaphragm. • Located directly below the radiographic tabletop , the grid is found just above the tray that holds the image receptor. • These grids continuously move 1-3 cm back and forth throughout the exposure.
  • 33. DIRECTION OF GRID TRAVEL • While moving parallel to its surface , the grid must move perpendicular to the long axis of its lead strips.
  • 34. • LONG DIMENSION GRIDS Linear grids having lead strips running parallel to the long axis of the grid. • SHORT DIMENSION GRIDS Linear grids having strips running perpendicular to the long axis of the grid. useful for examination where it is difficult to centre the central ray correctly for long dimension grid
  • 35.
  • 36. GRID PERFORMNCE 1. Primary transmission(Tp). 2. Bucky factor or Grid Conversion factor(B) 3. Contrast Improvement Factor(K)
  • 37. PRIMARY TRANSMISSION • Primary transmission is the measurement of the percentage of primary radiation transmitted through the grid. Tp = Ip/I’p X 100 Where Tp is the primary transmission Ip is the intensity with grid I’pis the intensity without grid
  • 39. BUCKY FACTOR • It is the ratio of incident radiation falling on the grid to the transmitted radiation passing through the grid. • Bucky factor can be used to determine the adjustment in mAs needed when changing from using a grid to non grid (or vice versa)
  • 40. • GCF or Bucky Factor = mAs with grid mAs without grid • Generally ,high ratio grids absorb more scatter radiation and have larger Bucky Factors than low ratio grids.
  • 41. APPARATUS TO MEASURE BUCKY FACTOR
  • 42. BUCKY FACTOR GRID RATIO GCF NO GRID 1 5:1 2 6:1 3 8:1 4 12:1 5 16:1 6
  • 43. CONTRAST IMPROVEMENT FACTOR • It is the ratio of the contrast with the grid to the contrast without the grid. • This is the measure of grid’s ability to improve contrast , which is its primary function.
  • 44. • K factor depends on kVp,field size and part thickness • To permit comparison of different grids ,the contrast improvement factor is usually determined at 100kVp with a large field and a phantom of 20 cm thick.
  • 45. SELECTIVITY • Selectivity = ratio of primary radiation transmitted through the grid to the scattered radiation transmitted through the grid. • Larger the ratio , greater will be the selectivity of the grid.
  • 46. GRID CUT-OFF • Grid cut-off refers to decrease in the number of transmitted photons that reach the image receptor because of some misallignment of the grid. • Higher grid ratio and short grid –focus distance results in more potential grid cut- off.
  • 47. • Grid cut-off is a result of poor geometric relationship between the primary beam and lead foil strips of the grid. • When the projected images of the lead strips are thicker than the width of the interspaces , cut-off is complete and no primary radiation reaches the film. • The resultant radiograph will be light in the area in which cut-off occurs.
  • 48. TYPES OF GRID CUT-OFF ERRORS 1. Upside –down focussed grid 2. Lateral decentering(off-centre) 3. Focus-grid distance decentering(off- focus) 4. Combined lateral and focus-grid distance decentering.
  • 49. UPSIDDE DOWN FOCUSSED GRID • This occurs when a focussed grid is placed upside down on the image recptor , resulting in the grid lines going opposite the angle of divergence of the x-ray beam. • Radiographically , there is significant loss of exposure along the edges of the image.
  • 50. • Photons easily pass through the centre of the grid because the lead lines are perpendicular to the image receptor surface. • The tilted strips near the edges of the grid will absorb progressively more radiation, causing severe peripheral cut-off. • However with parallel grids either side may face the tube or film.
  • 51.
  • 52. LATERAL DECENTERING • It occurs when the central ray of x-ray beam is not aligned from side to side with the centre of focussed grid. • All the lead strips cut off the same amount of primary radiation. • Hence, there is uniform loss of radiation over the entire surface of the grid ,producing a uniformly light radiograph.
  • 53. • FACTORS AFFECTING MAGNITUDE OF CUT –OFF: 1. Grid ratio 2.Focal distance 3. Amount of decentering L = rb/foX100 where r is the grid ratio b is the lateral decentering distance fo is the focal distance of grid
  • 54. • Amount of cut-off increases as the grid ratio and decentering distance increases, and cut-off decreases as the focal distance increases. • In portable radiography, low ratio grids and long focal distances should be used .
  • 55. OFF-LEVEL GRIDS • Off level grid cut off results when the x-ray beam is angled across the lead strips. • Occur from either the tube or the grid being angled. • Often seen with portable radiograph or horizontal beam examinations. • Only type that occurs with bth focussed and parallel grids
  • 57. • IMPORTANT RELATIONSHIP: Angling the x-ray tube across the grid lines or angling the grid itself during exposure produces an overall decrease in exposure on the radiograph.
  • 58.
  • 59. FOCUS-GRID DISTANCE DECENTERING • Target of the x-ray tube is correctly centered to the grid,but it is positioned above or below the convergent line. • Far focus-grid distance decentering-target is above • Near focus-grid distance decentering-target is below
  • 60. NEAR FOCUS GRID DISTANCE DECENTERING
  • 61. FAR FOCUS GRID DISTANCE DECENTERING
  • 62. • The loss of primary radiation increases as the grid ratio increases and as the distance from the centre of the grid increases. • Loss is greater with near than far-focus grid distance.
  • 63.
  • 64. COMBINED LATERAL AND FOCUS-GRID DISTANCE DECENTERING • Easy to recognise • Causes an uneven exposure , resulting in a film that is dark on one side and light on the other side.
  • 65.
  • 66.
  • 67. PATIENT PROTECTION ALERT • GRID SELECTION: In order to keep the patient exposure as low as possible , grids should be used only when appropriate , and the grid ratio should be lowest that would provide sufficient contrast improvement . • below 90 kVp-8:1,above 90 kVp 12:1
  • 68. CONCLUSION • Construction of the grid-lead strips separated by x-ray transparent spacers. • They are used to absorb scatter radiation and improve radiographic image contrast • Generally high ratio grids with high lead content have the highest contrast improvement factors. • The amount of grid cut-off with a particular decentering error is greatest with high ratio grids and short focussing distance . • Grid selection involves compromise between film quality and patient exposure.