SlideShare a Scribd company logo
1 of 44
FUNDAMENTAL
S OF IMAGING
RAD 206
Control of Scatter Radiation
Contrast changes with the use of a grid
Less scatter radiation & less radiographic
noise – shorter scale = “better contrast”
With Grid No Grid
GRIDS
CONTRAST &
CONTRAST
RESOLUTION
Two devices are used to reduce Compton effect beam-
restricting devices and radiographic grids
Beam-restricting devices effects what reaches the patient.
Grids effect the remnant beam
CONTRAST &
CONTRAST
RESOLUTION
Contrast = the comparison of areas of light, dark and shades
of gray on the image
Contrast Resolution = the ability to image adjacent similar
tissues
BEAM-RESTRICTING
DEVICES
Are helpful to improve contrast resolution however the
inherent problem is they are placed between the source and
the patient. Even under the most favorable conditions, most
if the remnant x-rays are scattered.
EFFECTS OF SCATTER
RADIATION ON IMAGE
CONTRAST
Contrast is the degree of difference in OD between areas of
an image
If you could only capture transmitted, unscattered x-rays, the
image would be very sharp
The corresponding bone-soft tissue interface, would be very
abrupt, and therefore the image contrast would be high
GRIDS
Are very effective device for reducing scatter radiation
The grid is a series of sections of radiopaque material (grid
strips) alternating with sections of radiolucent material
(interspace material)
The grid is designed to transmit only x-rays that are traveling
in a straight line from the source to the IR
GRIDS “CLEAN UP” SCATTER
RADIATION
A high quality
grid can attenuate 80 –
90 percent of scatter
radiation
GRID STRIPS
Should be very thin and have high photon absorption
properties
Lead is most common
Tungsten, platinum, gold, and uranium have been tried but
Pb is still most desirable
INTERSPACE
MATERIAL
Used to maintain precise separation between the delicate
lead strips
Aluminum or Plastic Fiber
Grid Casing = covered completely by thin aluminum to
provide rigidity and to seal out moisture. Yuck!
GRID RATIO
3 important dimensions on a grid: The thickness of the grid
strips, the width of the interspace material, and the height of
the grid
The grid ratio is the HEIGHT of the grid divided by the
INTERSPACE WIDTH:
Grid ratio = h
D
H = HEIGHT OF THE GRID,
T = THICKNESS OF THE GRID STRIP,
D = WIDTH OF THE INTERSPACE MATERIAL
GRID RATIO
High-ratio grids are more effective in cleaning up scatter
radiation than low-ratio grids
The angle of deviation is smaller for high-ratio grids. (the
photon must be traveling in a straighter line to make it
through the grid)
However, the higher the ratio the more radiation exposure
necessary to get a sufficient number of x-rays through the
grid to the IR
THE HIGHER THE RATIO THE STRAIGHTER THE
PHOTON MUST TRAVEL TO REACH THE IR
Grid ratios range
from 5:1 to 16:1
Most common
8:1 to 10:1
A 5:1 grid will
clean up 85%
16:1 clean up 97%
GRID FREQUENCY
The number of grid strips or grid lines per inch or centimeter
The higher the frequency the more strips and less interspace
material and the higher the grid ratio
As grid frequency increases, patient does is increase
because more scatter will be absorbed
GRID FREQUENCY
Some grids reduce the thickness of the strips to reduce the
exposure to the patient, this over all reduces the grid clean
up
Grids have frequencies in the range of 25 to 45 lines per
centimeter (60 to 110 lines per inch)
HIGHER FREQUENCY WITH THE SAME
INTERSPACE DISTANCE REDUCES THE GRID
EFFECTIVENESS
GRID PERFORMANCE
The principal function of a grid is to improve
image contrast
Contrast Improvement Factor (k) = the ratio of
the contrast of a radiograph made with a grid
to the contrast of the radiograph made without
a grid. A contrast improvement factor of 1
indicates no improvements
The higher the grid ratio & frequency the
higher the k
Using grids require more patient dose. Why is this?
When a grid is used technique must be increased to maintain
OD
The amount of increase is given by the Bucky factor (B) or
grid factor
BUCKY FACTOR OR GRID FACTOR
BUCKY FACTOR OR GRID FACTOR
The higher the grid ratio or frequency the higher the bucky
factor
The Bucky factor increases with increasing kVp
Pg 235: We will use the average values for calculations.
Selectivity or ability to “clean up”
the heavier the grid the more Pb it contains
GRID CUTOFF
Distance to
cutoff
SID
Grid ratio
With decreasing
SID more
potential
for grid cutoff
IR size will also
GRID TYPES
PARALLEL GRID
simplest type of grid
All the lead strips are parallel
Only clean up scatter in one direction (along the axis of the
grid)
Inexpensive, Easy to make, however can cause noticeable
grid cutoff with short SID’s (the greater the SID the lesser
artifact seen) .
GRID CUTOFF FROM PARALLEL GRID
THE HIGHER-RATIO THE MORE CUTOFF POTENTIAL
FOCUSED GRID
Designed to minimize grid cutoff
Lead strips are aligned with the divergence of the x-ray beam
Each focused grid must be identified with the appropriate
SID
Wrong SID = Grid cutoff
Focused grid have a little SID latitude or tolerance
(e.g. 100cm Focused grid would produce artifacts
if used at less than 90cm – or more than 110cm)
CROSSED GRID
AKA crosshatch
Have lead strips running along the long and short axes of the
grid
Made by placing two parallel grid on top of each other
CROSSED GRID
Have twice the grid ratio as Parallel (linear) grids
However, CR vs grid placement
is critical. The CR must align
with the center of the grid and
the grid and CR must be exactly
parallel or grid cutoff will occur
e.g.
MOVING GRIDS
All stationary grids will give you grid lines on your
radiograph. Thinner Pb strips will give you less noticeable
lines. However, thinner strips have less Pb content not
“cleaning up” as well
Grid Lines are made when primary x-rays are absorbed in the
grid strips.
FOCUSED GRIDS ARE USUALLY
USED AS MOVING GRIDS
The grid is placed in a holding mechanism that begins
moving just before the x-ray exposure and continues moving
after the exposure ends
2 types of movement Reciprocating & Oscillating
Reciprocating: moving back and forth about 2 cm
Oscillating: moving in a circular motion 2-3 cm
GRID MOTION
Reciprocating = moves several times about 2cm back and
forth during the exposure
Oscillating = moves several times about 2 – 3 cm in a circular
pattern
Most grids are moving. Except for portable imaging
GRID PROBLEMS
Increased OID, especially with moving grids
The biggest problem with grids is misalignment
GRID PROBLEMS
RESULT IN:
UNDEREXPOSED IMAGE
OR UNDEREXPOSED
EDGES OF IMAGE
https://www.youtube.com/watch?
v=mZPD_gLs5Dw
GRID PROBLEMS – OFF LEVEL
GRID PROBLEMS – OFF CENTER
A problem with focused & crossed grids
GRID PROBLEMS – OFF FOCUS
(WRONG SID)
GRID PROBLEMS – UPSIDE-DOWN
A problem with focused & crossed grids
GRID SELECTION
Patient Dose
• Pg 241 – mAs changes
Exam
Detail required
Part thickness
Desired technique (kVp)
Equipment availability
Contrast Improvement by Using an Air Gap
QUESTIONS….?

More Related Content

What's hot

Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionparthajyotidas11
 
Diffusion Weighted Imaging- Avinesh Shrestha
Diffusion Weighted Imaging- Avinesh ShresthaDiffusion Weighted Imaging- Avinesh Shrestha
Diffusion Weighted Imaging- Avinesh ShresthaAvinesh Shrestha
 
Chapter 6 image quality in ct
Chapter 6 image quality in ct Chapter 6 image quality in ct
Chapter 6 image quality in ct Muntaser S.Ahmad
 
Factors affecting x-ray beam quality and quantity
Factors affecting x-ray beam quality and quantityFactors affecting x-ray beam quality and quantity
Factors affecting x-ray beam quality and quantityravi christian
 
Dose reduction in Conventional Radiography and Fluoroscopy
Dose reduction in Conventional Radiography and FluoroscopyDose reduction in Conventional Radiography and Fluoroscopy
Dose reduction in Conventional Radiography and FluoroscopyTarun Goyal
 
CT RADIATION DOSE REDUCTION
CT RADIATION DOSE REDUCTION CT RADIATION DOSE REDUCTION
CT RADIATION DOSE REDUCTION Ganesan Yogananthem
 
Filters and beam restrictors
Filters  and beam restrictorsFilters  and beam restrictors
Filters and beam restrictorsvishwanath0908
 
X ray attenuation and scatter radiation
X ray attenuation and scatter radiationX ray attenuation and scatter radiation
X ray attenuation and scatter radiationNISHANT RAJ
 
Digital breast tomosynthesis
Digital breast tomosynthesisDigital breast tomosynthesis
Digital breast tomosynthesisMAMTA PANDA
 
soft tissue radiography
soft tissue radiographysoft tissue radiography
soft tissue radiographyDinesh Darshana
 
Image Contrast, Noise, Resolution
Image Contrast, Noise, ResolutionImage Contrast, Noise, Resolution
Image Contrast, Noise, ResolutionMiami Cancer Institute
 
Ct radiation dose & safety lecture 1
Ct radiation dose & safety lecture 1Ct radiation dose & safety lecture 1
Ct radiation dose & safety lecture 1Gamal Mahdaly
 
ct image quality
ct image qualityct image quality
ct image qualitydypradio
 
MACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxMACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxThejaTej6
 
Radiology day 3 mediastinal anatomy
Radiology day 3   mediastinal anatomyRadiology day 3   mediastinal anatomy
Radiology day 3 mediastinal anatomyVibhay Pareek
 
Basic principle of ct and ct generations
Basic principle of ct and ct generationsBasic principle of ct and ct generations
Basic principle of ct and ct generationsTarunGoyal66
 
Mammography technique
Mammography techniqueMammography technique
Mammography techniquevishwanath0908
 
MDCT AN INTRODUCTION TO CLINICAL APPLICATIONS
MDCT AN INTRODUCTION TO CLINICAL APPLICATIONSMDCT AN INTRODUCTION TO CLINICAL APPLICATIONS
MDCT AN INTRODUCTION TO CLINICAL APPLICATIONShazem youssef
 

What's hot (20)

Magnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortionMagnification(macro and micro radiography), distortion
Magnification(macro and micro radiography), distortion
 
Diffusion Weighted Imaging- Avinesh Shrestha
Diffusion Weighted Imaging- Avinesh ShresthaDiffusion Weighted Imaging- Avinesh Shrestha
Diffusion Weighted Imaging- Avinesh Shrestha
 
Filters.ppt
Filters.pptFilters.ppt
Filters.ppt
 
Chapter 6 image quality in ct
Chapter 6 image quality in ct Chapter 6 image quality in ct
Chapter 6 image quality in ct
 
Factors affecting x-ray beam quality and quantity
Factors affecting x-ray beam quality and quantityFactors affecting x-ray beam quality and quantity
Factors affecting x-ray beam quality and quantity
 
Dose reduction in Conventional Radiography and Fluoroscopy
Dose reduction in Conventional Radiography and FluoroscopyDose reduction in Conventional Radiography and Fluoroscopy
Dose reduction in Conventional Radiography and Fluoroscopy
 
CT RADIATION DOSE REDUCTION
CT RADIATION DOSE REDUCTION CT RADIATION DOSE REDUCTION
CT RADIATION DOSE REDUCTION
 
Filters and beam restrictors
Filters  and beam restrictorsFilters  and beam restrictors
Filters and beam restrictors
 
X ray attenuation and scatter radiation
X ray attenuation and scatter radiationX ray attenuation and scatter radiation
X ray attenuation and scatter radiation
 
Digital breast tomosynthesis
Digital breast tomosynthesisDigital breast tomosynthesis
Digital breast tomosynthesis
 
soft tissue radiography
soft tissue radiographysoft tissue radiography
soft tissue radiography
 
Image Contrast, Noise, Resolution
Image Contrast, Noise, ResolutionImage Contrast, Noise, Resolution
Image Contrast, Noise, Resolution
 
Ct radiation dose & safety lecture 1
Ct radiation dose & safety lecture 1Ct radiation dose & safety lecture 1
Ct radiation dose & safety lecture 1
 
ct image quality
ct image qualityct image quality
ct image quality
 
MACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptxMACRORADIOGRAPHY.pptx
MACRORADIOGRAPHY.pptx
 
Radiology day 3 mediastinal anatomy
Radiology day 3   mediastinal anatomyRadiology day 3   mediastinal anatomy
Radiology day 3 mediastinal anatomy
 
Basic principle of ct and ct generations
Basic principle of ct and ct generationsBasic principle of ct and ct generations
Basic principle of ct and ct generations
 
Ct of the larynx
Ct of the larynxCt of the larynx
Ct of the larynx
 
Mammography technique
Mammography techniqueMammography technique
Mammography technique
 
MDCT AN INTRODUCTION TO CLINICAL APPLICATIONS
MDCT AN INTRODUCTION TO CLINICAL APPLICATIONSMDCT AN INTRODUCTION TO CLINICAL APPLICATIONS
MDCT AN INTRODUCTION TO CLINICAL APPLICATIONS
 

Similar to Fundamentals of Imaging Rad 206: Contrast Improvement

grids,films.pptx
grids,films.pptxgrids,films.pptx
grids,films.pptxrohanjohnjacob
 
Grids
GridsGrids
Gridsaslam bs
 
Beam restriction devices in x rays by sushil pattar
Beam restriction devices in x rays by sushil pattarBeam restriction devices in x rays by sushil pattar
Beam restriction devices in x rays by sushil pattarSushilPattar
 
Radiographic grid Swastik
Radiographic grid   SwastikRadiographic grid   Swastik
Radiographic grid SwastikBipul Poudel
 
grids-160225101450 (1).pptx
grids-160225101450 (1).pptxgrids-160225101450 (1).pptx
grids-160225101450 (1).pptxVishnuDutt40
 
Xray filters beam restrictors and Grids Radiodiagnosis.ppt
Xray filters beam restrictors and Grids Radiodiagnosis.pptXray filters beam restrictors and Grids Radiodiagnosis.ppt
Xray filters beam restrictors and Grids Radiodiagnosis.pptNagasai Pelala
 
Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.Arya Prasad
 
XRAY GRIDS (1).pptx
XRAY GRIDS (1).pptxXRAY GRIDS (1).pptx
XRAY GRIDS (1).pptxDRVISHAL12
 
Filters grids and collimators
Filters grids and collimatorsFilters grids and collimators
Filters grids and collimatorsPrajwith Rai
 
Radiographic Image contrast & image resolution
Radiographic Image contrast & image resolutionRadiographic Image contrast & image resolution
Radiographic Image contrast & image resolutionNitish Virmani
 
Bandwidth enhancement patch antenna
Bandwidth enhancement patch antennaBandwidth enhancement patch antenna
Bandwidth enhancement patch antennaAnurag Anupam
 
MicroStrip Antenna
MicroStrip AntennaMicroStrip Antenna
MicroStrip AntennaTarek Nader
 
X-RAY GRID BUCKY
X-RAY GRID BUCKYX-RAY GRID BUCKY
X-RAY GRID BUCKYDEEPAK KUMAR
 

Similar to Fundamentals of Imaging Rad 206: Contrast Improvement (20)

grids,films.pptx
grids,films.pptxgrids,films.pptx
grids,films.pptx
 
GRIDS.ppt
GRIDS.pptGRIDS.ppt
GRIDS.ppt
 
Grids
GridsGrids
Grids
 
GRIDS.ppt
GRIDS.pptGRIDS.ppt
GRIDS.ppt
 
Beam restriction devices in x rays by sushil pattar
Beam restriction devices in x rays by sushil pattarBeam restriction devices in x rays by sushil pattar
Beam restriction devices in x rays by sushil pattar
 
Radiographic grid Swastik
Radiographic grid   SwastikRadiographic grid   Swastik
Radiographic grid Swastik
 
Grids JP .ppt
Grids JP .pptGrids JP .ppt
Grids JP .ppt
 
Neelotpal Mitra
Neelotpal MitraNeelotpal Mitra
Neelotpal Mitra
 
grids-160225101450 (1).pptx
grids-160225101450 (1).pptxgrids-160225101450 (1).pptx
grids-160225101450 (1).pptx
 
Grids
GridsGrids
Grids
 
Xray filters beam restrictors and Grids Radiodiagnosis.ppt
Xray filters beam restrictors and Grids Radiodiagnosis.pptXray filters beam restrictors and Grids Radiodiagnosis.ppt
Xray filters beam restrictors and Grids Radiodiagnosis.ppt
 
Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.Filters used in radiology.ppt.radiology.
Filters used in radiology.ppt.radiology.
 
XRAY GRIDS (1).pptx
XRAY GRIDS (1).pptxXRAY GRIDS (1).pptx
XRAY GRIDS (1).pptx
 
Filters grids and collimators
Filters grids and collimatorsFilters grids and collimators
Filters grids and collimators
 
Radiographic Image contrast & image resolution
Radiographic Image contrast & image resolutionRadiographic Image contrast & image resolution
Radiographic Image contrast & image resolution
 
Bandwidth enhancement patch antenna
Bandwidth enhancement patch antennaBandwidth enhancement patch antenna
Bandwidth enhancement patch antenna
 
MicroStrip Antenna
MicroStrip AntennaMicroStrip Antenna
MicroStrip Antenna
 
X-RAY GRID BUCKY
X-RAY GRID BUCKYX-RAY GRID BUCKY
X-RAY GRID BUCKY
 
Tearhertz Sub-Nanometer Sub-Surface Imaging of 2D Materials
Tearhertz Sub-Nanometer Sub-Surface Imaging of 2D MaterialsTearhertz Sub-Nanometer Sub-Surface Imaging of 2D Materials
Tearhertz Sub-Nanometer Sub-Surface Imaging of 2D Materials
 
Optical fibre
Optical fibreOptical fibre
Optical fibre
 

More from sehlawi

Rad 206 p13
Rad 206 p13Rad 206 p13
Rad 206 p13sehlawi
 
Rad 206 p11 Fundamentals of Imaging - Control of Scatter Radiation
Rad 206 p11 Fundamentals of Imaging - Control of Scatter RadiationRad 206 p11 Fundamentals of Imaging - Control of Scatter Radiation
Rad 206 p11 Fundamentals of Imaging - Control of Scatter Radiationsehlawi
 
Rad 206 p10 Fundamentals of Imaging -
Rad 206 p10 Fundamentals of Imaging - Rad 206 p10 Fundamentals of Imaging -
Rad 206 p10 Fundamentals of Imaging - sehlawi
 
Rad 206 p05 Fundamentals of Imaging - Fluoroscopy
Rad 206 p05 Fundamentals of Imaging - FluoroscopyRad 206 p05 Fundamentals of Imaging - Fluoroscopy
Rad 206 p05 Fundamentals of Imaging - Fluoroscopysehlawi
 
Rad 206 p03
Rad 206 p03Rad 206 p03
Rad 206 p03sehlawi
 
Rad 206 p09
Rad 206 p09Rad 206 p09
Rad 206 p09sehlawi
 
Rad 206 p08
Rad 206 p08Rad 206 p08
Rad 206 p08sehlawi
 
Rad 206 p04
Rad 206 p04Rad 206 p04
Rad 206 p04sehlawi
 
Rad 206 introduction
Rad 206 introductionRad 206 introduction
Rad 206 introductionsehlawi
 
Rad 206 p02
Rad 206 p02Rad 206 p02
Rad 206 p02sehlawi
 
Rad 206 p01
Rad 206 p01Rad 206 p01
Rad 206 p01sehlawi
 
Rad 104 hospital practice and care of patients 8 types of catheters 2016
Rad 104 hospital practice and care of patients 8  types of catheters 2016Rad 104 hospital practice and care of patients 8  types of catheters 2016
Rad 104 hospital practice and care of patients 8 types of catheters 2016sehlawi
 
Rad 104 hospital practice and care of patients 7 safe patient movement and h...
Rad 104 hospital practice and care of patients 7  safe patient movement and h...Rad 104 hospital practice and care of patients 7  safe patient movement and h...
Rad 104 hospital practice and care of patients 7 safe patient movement and h...sehlawi
 
Rad 104 hospital practice and care of patients 6 infection control 2016
Rad 104 hospital practice and care of patients 6  infection control 2016Rad 104 hospital practice and care of patients 6  infection control 2016
Rad 104 hospital practice and care of patients 6 infection control 2016sehlawi
 
Rad 104 hospital practice and care of patients 5 anesthesia 2016
Rad 104 hospital practice and care of patients 5  anesthesia 2016Rad 104 hospital practice and care of patients 5  anesthesia 2016
Rad 104 hospital practice and care of patients 5 anesthesia 2016sehlawi
 
Rad 104 hospital practice and care of patients 3 drugs & contrast 2016
Rad 104 hospital practice and care of patients 3  drugs & contrast 2016Rad 104 hospital practice and care of patients 3  drugs & contrast 2016
Rad 104 hospital practice and care of patients 3 drugs & contrast 2016sehlawi
 
Rad 104 hospital practice and care of patients 2 terminology p2 - 2016
Rad 104 hospital practice and care of patients 2  terminology p2 - 2016Rad 104 hospital practice and care of patients 2  terminology p2 - 2016
Rad 104 hospital practice and care of patients 2 terminology p2 - 2016sehlawi
 
Rad 104 hospital practice and care of patients 1 objectives & terminolog...
Rad 104 hospital practice and care of patients 1  objectives & terminolog...Rad 104 hospital practice and care of patients 1  objectives & terminolog...
Rad 104 hospital practice and care of patients 1 objectives & terminolog...sehlawi
 
Peer observation of teaching ( mahdi) 2016
Peer observation of teaching ( mahdi) 2016Peer observation of teaching ( mahdi) 2016
Peer observation of teaching ( mahdi) 2016sehlawi
 

More from sehlawi (19)

Rad 206 p13
Rad 206 p13Rad 206 p13
Rad 206 p13
 
Rad 206 p11 Fundamentals of Imaging - Control of Scatter Radiation
Rad 206 p11 Fundamentals of Imaging - Control of Scatter RadiationRad 206 p11 Fundamentals of Imaging - Control of Scatter Radiation
Rad 206 p11 Fundamentals of Imaging - Control of Scatter Radiation
 
Rad 206 p10 Fundamentals of Imaging -
Rad 206 p10 Fundamentals of Imaging - Rad 206 p10 Fundamentals of Imaging -
Rad 206 p10 Fundamentals of Imaging -
 
Rad 206 p05 Fundamentals of Imaging - Fluoroscopy
Rad 206 p05 Fundamentals of Imaging - FluoroscopyRad 206 p05 Fundamentals of Imaging - Fluoroscopy
Rad 206 p05 Fundamentals of Imaging - Fluoroscopy
 
Rad 206 p03
Rad 206 p03Rad 206 p03
Rad 206 p03
 
Rad 206 p09
Rad 206 p09Rad 206 p09
Rad 206 p09
 
Rad 206 p08
Rad 206 p08Rad 206 p08
Rad 206 p08
 
Rad 206 p04
Rad 206 p04Rad 206 p04
Rad 206 p04
 
Rad 206 introduction
Rad 206 introductionRad 206 introduction
Rad 206 introduction
 
Rad 206 p02
Rad 206 p02Rad 206 p02
Rad 206 p02
 
Rad 206 p01
Rad 206 p01Rad 206 p01
Rad 206 p01
 
Rad 104 hospital practice and care of patients 8 types of catheters 2016
Rad 104 hospital practice and care of patients 8  types of catheters 2016Rad 104 hospital practice and care of patients 8  types of catheters 2016
Rad 104 hospital practice and care of patients 8 types of catheters 2016
 
Rad 104 hospital practice and care of patients 7 safe patient movement and h...
Rad 104 hospital practice and care of patients 7  safe patient movement and h...Rad 104 hospital practice and care of patients 7  safe patient movement and h...
Rad 104 hospital practice and care of patients 7 safe patient movement and h...
 
Rad 104 hospital practice and care of patients 6 infection control 2016
Rad 104 hospital practice and care of patients 6  infection control 2016Rad 104 hospital practice and care of patients 6  infection control 2016
Rad 104 hospital practice and care of patients 6 infection control 2016
 
Rad 104 hospital practice and care of patients 5 anesthesia 2016
Rad 104 hospital practice and care of patients 5  anesthesia 2016Rad 104 hospital practice and care of patients 5  anesthesia 2016
Rad 104 hospital practice and care of patients 5 anesthesia 2016
 
Rad 104 hospital practice and care of patients 3 drugs & contrast 2016
Rad 104 hospital practice and care of patients 3  drugs & contrast 2016Rad 104 hospital practice and care of patients 3  drugs & contrast 2016
Rad 104 hospital practice and care of patients 3 drugs & contrast 2016
 
Rad 104 hospital practice and care of patients 2 terminology p2 - 2016
Rad 104 hospital practice and care of patients 2  terminology p2 - 2016Rad 104 hospital practice and care of patients 2  terminology p2 - 2016
Rad 104 hospital practice and care of patients 2 terminology p2 - 2016
 
Rad 104 hospital practice and care of patients 1 objectives & terminolog...
Rad 104 hospital practice and care of patients 1  objectives & terminolog...Rad 104 hospital practice and care of patients 1  objectives & terminolog...
Rad 104 hospital practice and care of patients 1 objectives & terminolog...
 
Peer observation of teaching ( mahdi) 2016
Peer observation of teaching ( mahdi) 2016Peer observation of teaching ( mahdi) 2016
Peer observation of teaching ( mahdi) 2016
 

Recently uploaded

Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfHemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfMedicoseAcademics
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableNehru place Escorts
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Serviceparulsinha
 
Call Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service SuratCall Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service Suratnarwatsonia7
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girlsnehamumbai
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...Miss joya
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Servicesonalikaur4
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photosnarwatsonia7
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Gabriel Guevara MD
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Miss joya
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...narwatsonia7
 
Call Girls Budhwar Peth 7001305949 All Area Service COD available Any Time
Call Girls Budhwar Peth 7001305949 All Area Service COD available Any TimeCall Girls Budhwar Peth 7001305949 All Area Service COD available Any Time
Call Girls Budhwar Peth 7001305949 All Area Service COD available Any Timevijaych2041
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.MiadAlsulami
 
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Modelssonalikaur4
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Availablenarwatsonia7
 
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowSonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowRiya Pathan
 

Recently uploaded (20)

Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdfHemostasis Physiology and Clinical correlations by Dr Faiza.pdf
Hemostasis Physiology and Clinical correlations by Dr Faiza.pdf
 
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls AvailableVip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
Vip Call Girls Anna Salai Chennai 👉 8250192130 ❣️💯 Top Class Girls Available
 
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort ServiceCall Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
Call Girls Service In Shyam Nagar Whatsapp 8445551418 Independent Escort Service
 
Call Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service SuratCall Girl Surat Madhuri 7001305949 Independent Escort Service Surat
Call Girl Surat Madhuri 7001305949 Independent Escort Service Surat
 
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Servicesauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
sauth delhi call girls in Bhajanpura 🔝 9953056974 🔝 escort Service
 
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy GirlsCall Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
Call Girls In Andheri East Call 9920874524 Book Hot And Sexy Girls
 
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
 
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls ServiceCall Girls Thane Just Call 9910780858 Get High Class Call Girls Service
Call Girls Thane Just Call 9910780858 Get High Class Call Girls Service
 
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original PhotosCall Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
Call Girl Service Bidadi - For 7001305949 Cheap & Best with original Photos
 
Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024Asthma Review - GINA guidelines summary 2024
Asthma Review - GINA guidelines summary 2024
 
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
Low Rate Call Girls Pune Esha 9907093804 Short 1500 Night 6000 Best call girl...
 
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
Call Girls Service in Bommanahalli - 7001305949 with real photos and phone nu...
 
Call Girls Budhwar Peth 7001305949 All Area Service COD available Any Time
Call Girls Budhwar Peth 7001305949 All Area Service COD available Any TimeCall Girls Budhwar Peth 7001305949 All Area Service COD available Any Time
Call Girls Budhwar Peth 7001305949 All Area Service COD available Any Time
 
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
Artifacts in Nuclear Medicine with Identifying and resolving artifacts.
 
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCREscort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
Escort Service Call Girls In Sarita Vihar,, 99530°56974 Delhi NCR
 
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Hosur Just Call 7001305949 Top Class Call Girl Service Available
 
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking ModelsMumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
Mumbai Call Girls Service 9910780858 Real Russian Girls Looking Models
 
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
Russian Call Girls in Delhi Tanvi ➡️ 9711199012 💋📞 Independent Escort Service...
 
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service AvailableCall Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
Call Girls Jayanagar Just Call 7001305949 Top Class Call Girl Service Available
 
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call NowSonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
Sonagachi Call Girls Services 9907093804 @24x7 High Class Babes Here Call Now
 

Fundamentals of Imaging Rad 206: Contrast Improvement

  • 1. FUNDAMENTAL S OF IMAGING RAD 206 Control of Scatter Radiation
  • 2.
  • 3. Contrast changes with the use of a grid Less scatter radiation & less radiographic noise – shorter scale = “better contrast” With Grid No Grid
  • 5. CONTRAST & CONTRAST RESOLUTION Two devices are used to reduce Compton effect beam- restricting devices and radiographic grids Beam-restricting devices effects what reaches the patient. Grids effect the remnant beam
  • 6. CONTRAST & CONTRAST RESOLUTION Contrast = the comparison of areas of light, dark and shades of gray on the image Contrast Resolution = the ability to image adjacent similar tissues
  • 7. BEAM-RESTRICTING DEVICES Are helpful to improve contrast resolution however the inherent problem is they are placed between the source and the patient. Even under the most favorable conditions, most if the remnant x-rays are scattered.
  • 8. EFFECTS OF SCATTER RADIATION ON IMAGE CONTRAST Contrast is the degree of difference in OD between areas of an image If you could only capture transmitted, unscattered x-rays, the image would be very sharp The corresponding bone-soft tissue interface, would be very abrupt, and therefore the image contrast would be high
  • 9. GRIDS Are very effective device for reducing scatter radiation The grid is a series of sections of radiopaque material (grid strips) alternating with sections of radiolucent material (interspace material) The grid is designed to transmit only x-rays that are traveling in a straight line from the source to the IR
  • 10. GRIDS “CLEAN UP” SCATTER RADIATION A high quality grid can attenuate 80 – 90 percent of scatter radiation
  • 11. GRID STRIPS Should be very thin and have high photon absorption properties Lead is most common Tungsten, platinum, gold, and uranium have been tried but Pb is still most desirable
  • 12. INTERSPACE MATERIAL Used to maintain precise separation between the delicate lead strips Aluminum or Plastic Fiber Grid Casing = covered completely by thin aluminum to provide rigidity and to seal out moisture. Yuck!
  • 13.
  • 14. GRID RATIO 3 important dimensions on a grid: The thickness of the grid strips, the width of the interspace material, and the height of the grid The grid ratio is the HEIGHT of the grid divided by the INTERSPACE WIDTH: Grid ratio = h D
  • 15. H = HEIGHT OF THE GRID, T = THICKNESS OF THE GRID STRIP, D = WIDTH OF THE INTERSPACE MATERIAL
  • 16. GRID RATIO High-ratio grids are more effective in cleaning up scatter radiation than low-ratio grids The angle of deviation is smaller for high-ratio grids. (the photon must be traveling in a straighter line to make it through the grid) However, the higher the ratio the more radiation exposure necessary to get a sufficient number of x-rays through the grid to the IR
  • 17. THE HIGHER THE RATIO THE STRAIGHTER THE PHOTON MUST TRAVEL TO REACH THE IR Grid ratios range from 5:1 to 16:1 Most common 8:1 to 10:1 A 5:1 grid will clean up 85% 16:1 clean up 97%
  • 18. GRID FREQUENCY The number of grid strips or grid lines per inch or centimeter The higher the frequency the more strips and less interspace material and the higher the grid ratio As grid frequency increases, patient does is increase because more scatter will be absorbed
  • 19. GRID FREQUENCY Some grids reduce the thickness of the strips to reduce the exposure to the patient, this over all reduces the grid clean up Grids have frequencies in the range of 25 to 45 lines per centimeter (60 to 110 lines per inch)
  • 20. HIGHER FREQUENCY WITH THE SAME INTERSPACE DISTANCE REDUCES THE GRID EFFECTIVENESS
  • 21. GRID PERFORMANCE The principal function of a grid is to improve image contrast Contrast Improvement Factor (k) = the ratio of the contrast of a radiograph made with a grid to the contrast of the radiograph made without a grid. A contrast improvement factor of 1 indicates no improvements The higher the grid ratio & frequency the higher the k
  • 22. Using grids require more patient dose. Why is this? When a grid is used technique must be increased to maintain OD The amount of increase is given by the Bucky factor (B) or grid factor BUCKY FACTOR OR GRID FACTOR
  • 23. BUCKY FACTOR OR GRID FACTOR The higher the grid ratio or frequency the higher the bucky factor The Bucky factor increases with increasing kVp Pg 235: We will use the average values for calculations.
  • 24. Selectivity or ability to “clean up” the heavier the grid the more Pb it contains
  • 25. GRID CUTOFF Distance to cutoff SID Grid ratio With decreasing SID more potential for grid cutoff IR size will also
  • 26. GRID TYPES PARALLEL GRID simplest type of grid All the lead strips are parallel Only clean up scatter in one direction (along the axis of the grid) Inexpensive, Easy to make, however can cause noticeable grid cutoff with short SID’s (the greater the SID the lesser artifact seen) .
  • 27. GRID CUTOFF FROM PARALLEL GRID THE HIGHER-RATIO THE MORE CUTOFF POTENTIAL
  • 28. FOCUSED GRID Designed to minimize grid cutoff Lead strips are aligned with the divergence of the x-ray beam Each focused grid must be identified with the appropriate SID Wrong SID = Grid cutoff
  • 29. Focused grid have a little SID latitude or tolerance (e.g. 100cm Focused grid would produce artifacts if used at less than 90cm – or more than 110cm)
  • 30. CROSSED GRID AKA crosshatch Have lead strips running along the long and short axes of the grid Made by placing two parallel grid on top of each other
  • 31. CROSSED GRID Have twice the grid ratio as Parallel (linear) grids However, CR vs grid placement is critical. The CR must align with the center of the grid and the grid and CR must be exactly parallel or grid cutoff will occur e.g.
  • 32. MOVING GRIDS All stationary grids will give you grid lines on your radiograph. Thinner Pb strips will give you less noticeable lines. However, thinner strips have less Pb content not “cleaning up” as well Grid Lines are made when primary x-rays are absorbed in the grid strips.
  • 33. FOCUSED GRIDS ARE USUALLY USED AS MOVING GRIDS The grid is placed in a holding mechanism that begins moving just before the x-ray exposure and continues moving after the exposure ends 2 types of movement Reciprocating & Oscillating Reciprocating: moving back and forth about 2 cm Oscillating: moving in a circular motion 2-3 cm
  • 34. GRID MOTION Reciprocating = moves several times about 2cm back and forth during the exposure Oscillating = moves several times about 2 – 3 cm in a circular pattern Most grids are moving. Except for portable imaging
  • 35. GRID PROBLEMS Increased OID, especially with moving grids The biggest problem with grids is misalignment GRID PROBLEMS RESULT IN: UNDEREXPOSED IMAGE OR UNDEREXPOSED EDGES OF IMAGE https://www.youtube.com/watch? v=mZPD_gLs5Dw
  • 37. GRID PROBLEMS – OFF CENTER A problem with focused & crossed grids
  • 38. GRID PROBLEMS – OFF FOCUS (WRONG SID)
  • 39. GRID PROBLEMS – UPSIDE-DOWN A problem with focused & crossed grids
  • 40. GRID SELECTION Patient Dose • Pg 241 – mAs changes Exam Detail required Part thickness Desired technique (kVp) Equipment availability
  • 41. Contrast Improvement by Using an Air Gap
  • 42.
  • 43.

Editor's Notes

  1. How many important words can you see Whare are you in this sentence Where am I