Quality Assurance Programme in Computed TomographyRamzee Small
Introduction to Computed Tomography
Basic description of the components of a CT System
Introduction to Quality Assurance
Quality Assurance and Quality Control Tests in Computed Tomography base on frequency
Objective of QA/QC Test
Quality Assurance Programme in Computed TomographyRamzee Small
Introduction to Computed Tomography
Basic description of the components of a CT System
Introduction to Quality Assurance
Quality Assurance and Quality Control Tests in Computed Tomography base on frequency
Objective of QA/QC Test
This slide best explains the introduction of CT, basis and types of CT image reconstructions with detailed explanation about Interpolation, convolution, Fourier slice theorem, Fourier transformation and brief explanation about the image domain i.e digital image processing.
Mri spin echo pulse sequences its variations andYashawant Yadav
MRI spin echo pulse sequences its variation and applications , in this slide collection principle of spine echo pulse sequences is described with physic behind it ,, this slide also coves the inverse recovery pulse sequences and types ,,,, image weighting and parameters are explained .. hope it may be help ful.
This slide best explains the introduction of CT, basis and types of CT image reconstructions with detailed explanation about Interpolation, convolution, Fourier slice theorem, Fourier transformation and brief explanation about the image domain i.e digital image processing.
Mri spin echo pulse sequences its variations andYashawant Yadav
MRI spin echo pulse sequences its variation and applications , in this slide collection principle of spine echo pulse sequences is described with physic behind it ,, this slide also coves the inverse recovery pulse sequences and types ,,,, image weighting and parameters are explained .. hope it may be help ful.
Development of a receiver circuit for medium frequency shift keying signals.inventionjournals
Frequency shift keying (fsk) mode of digital signal information transfer switches between two predetermined frequencies of the carrier wave, either by modulating one sine wave oscillator or by switching between two oscillators.The need for a receiver to decode an fsk signal along the transmitting medium from a digital source code within about 5 kilometer radius for security monitoring of environment informed this work. The design of a receiver circuit at a frequency of 500 kHzfor an input frequency shift keying (fsk) signal from a transmitter is presented. The receiver is to receive an RF signal, amplify it, filter it to remove unwanted signals, and recover the desired base band information. It consists of an amplifier, tuned circuitsand mixers which filters the base-band information. A comparator circuit is incorporated, to detect the digital signal received. The output from the comparators is the digital equivalent of the coded signals sent by the transmitter circuit, and transferred to a microcontroller circuit, to act as a coded signal representing information from the transmitting end. The bode-plot response of the receiver to the incoming signals using a FET tuned circuit, shows that only frequencies above 470kHz, and below 495kHz are allowed to pass through the network with a resonant frequency of 483.553 kHz and a gain of 27.734dB, while others are totally attenuated. The reliability of the designed receiver circuit was evaluated for a 1 year continuous operating period and was found to be 74.7%.Area of application of this work include electronic policing of a defined environment with good success
As a communication engineer our princple interest is signal and their measuring instrument; Thus Spectrum Analyzer is an instrument which graphically provides the energy distribution of a signal as a function of the frequency on its CRT.
The spectrum analyzer provides information about all these things, by displaying the signal in the frequency domain.
The measurement of dominant frequencies and their responses.
The component levels and energy strength
Frequency stability
Bandwidth and Spectral purity
Modulation index and attenuation
Harmonic and intermodulation distortion
Various signal generation and so on
Which is not easy to measure at time domain
micro teaching on communication m.sc nursing.pdfAnurag Sharma
Microteaching is a unique model of practice teaching. It is a viable instrument for the. desired change in the teaching behavior or the behavior potential which, in specified types of real. classroom situations, tends to facilitate the achievement of specified types of objectives.
HOT NEW PRODUCT! BIG SALES FAST SHIPPING NOW FROM CHINA!! EU KU DB BK substit...GL Anaacs
Contact us if you are interested:
Email / Skype : kefaya1771@gmail.com
Threema: PXHY5PDH
New BATCH Ku !!! MUCH IN DEMAND FAST SALE EVERY BATCH HAPPY GOOD EFFECT BIG BATCH !
Contact me on Threema or skype to start big business!!
Hot-sale products:
NEW HOT EUTYLONE WHITE CRYSTAL!!
5cl-adba precursor (semi finished )
5cl-adba raw materials
ADBB precursor (semi finished )
ADBB raw materials
APVP powder
5fadb/4f-adb
Jwh018 / Jwh210
Eutylone crystal
Protonitazene (hydrochloride) CAS: 119276-01-6
Flubrotizolam CAS: 57801-95-3
Metonitazene CAS: 14680-51-4
Payment terms: Western Union,MoneyGram,Bitcoin or USDT.
Deliver Time: Usually 7-15days
Shipping method: FedEx, TNT, DHL,UPS etc.Our deliveries are 100% safe, fast, reliable and discreet.
Samples will be sent for your evaluation!If you are interested in, please contact me, let's talk details.
We specializes in exporting high quality Research chemical, medical intermediate, Pharmaceutical chemicals and so on. Products are exported to USA, Canada, France, Korea, Japan,Russia, Southeast Asia and other countries.
Ozempic: Preoperative Management of Patients on GLP-1 Receptor Agonists Saeid Safari
Preoperative Management of Patients on GLP-1 Receptor Agonists like Ozempic and Semiglutide
ASA GUIDELINE
NYSORA Guideline
2 Case Reports of Gastric Ultrasound
Report Back from SGO 2024: What’s the Latest in Cervical Cancer?bkling
Are you curious about what’s new in cervical cancer research or unsure what the findings mean? Join Dr. Emily Ko, a gynecologic oncologist at Penn Medicine, to learn about the latest updates from the Society of Gynecologic Oncology (SGO) 2024 Annual Meeting on Women’s Cancer. Dr. Ko will discuss what the research presented at the conference means for you and answer your questions about the new developments.
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...i3 Health
i3 Health is pleased to make the speaker slides from this activity available for use as a non-accredited self-study or teaching resource.
This slide deck presented by Dr. Kami Maddocks, Professor-Clinical in the Division of Hematology and
Associate Division Director for Ambulatory Operations
The Ohio State University Comprehensive Cancer Center, will provide insight into new directions in targeted therapeutic approaches for older adults with mantle cell lymphoma.
STATEMENT OF NEED
Mantle cell lymphoma (MCL) is a rare, aggressive B-cell non-Hodgkin lymphoma (NHL) accounting for 5% to 7% of all lymphomas. Its prognosis ranges from indolent disease that does not require treatment for years to very aggressive disease, which is associated with poor survival (Silkenstedt et al, 2021). Typically, MCL is diagnosed at advanced stage and in older patients who cannot tolerate intensive therapy (NCCN, 2022). Although recent advances have slightly increased remission rates, recurrence and relapse remain very common, leading to a median overall survival between 3 and 6 years (LLS, 2021). Though there are several effective options, progress is still needed towards establishing an accepted frontline approach for MCL (Castellino et al, 2022). Treatment selection and management of MCL are complicated by the heterogeneity of prognosis, advanced age and comorbidities of patients, and lack of an established standard approach for treatment, making it vital that clinicians be familiar with the latest research and advances in this area. In this activity chaired by Michael Wang, MD, Professor in the Department of Lymphoma & Myeloma at MD Anderson Cancer Center, expert faculty will discuss prognostic factors informing treatment, the promising results of recent trials in new therapeutic approaches, and the implications of treatment resistance in therapeutic selection for MCL.
Target Audience
Hematology/oncology fellows, attending faculty, and other health care professionals involved in the treatment of patients with mantle cell lymphoma (MCL).
Learning Objectives
1.) Identify clinical and biological prognostic factors that can guide treatment decision making for older adults with MCL
2.) Evaluate emerging data on targeted therapeutic approaches for treatment-naive and relapsed/refractory MCL and their applicability to older adults
3.) Assess mechanisms of resistance to targeted therapies for MCL and their implications for treatment selection
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...Oleg Kshivets
RESULTS: Overall life span (LS) was 2252.1±1742.5 days and cumulative 5-year survival (5YS) reached 73.2%, 10 years – 64.8%, 20 years – 42.5%. 513 LCP lived more than 5 years (LS=3124.6±1525.6 days), 148 LCP – more than 10 years (LS=5054.4±1504.1 days).199 LCP died because of LC (LS=562.7±374.5 days). 5YS of LCP after bi/lobectomies was significantly superior in comparison with LCP after pneumonectomies (78.1% vs.63.7%, P=0.00001 by log-rank test). AT significantly improved 5YS (66.3% vs. 34.8%) (P=0.00000 by log-rank test) only for LCP with N1-2. Cox modeling displayed that 5YS of LCP significantly depended on: phase transition (PT) early-invasive LC in terms of synergetics, PT N0—N12, cell ratio factors (ratio between cancer cells- CC and blood cells subpopulations), G1-3, histology, glucose, AT, blood cell circuit, prothrombin index, heparin tolerance, recalcification time (P=0.000-0.038). Neural networks, genetic algorithm selection and bootstrap simulation revealed relationships between 5YS and PT early-invasive LC (rank=1), PT N0—N12 (rank=2), thrombocytes/CC (3), erythrocytes/CC (4), eosinophils/CC (5), healthy cells/CC (6), lymphocytes/CC (7), segmented neutrophils/CC (8), stick neutrophils/CC (9), monocytes/CC (10); leucocytes/CC (11). Correct prediction of 5YS was 100% by neural networks computing (area under ROC curve=1.0; error=0.0).
CONCLUSIONS: 5YS of LCP after radical procedures significantly depended on: 1) PT early-invasive cancer; 2) PT N0--N12; 3) cell ratio factors; 4) blood cell circuit; 5) biochemical factors; 6) hemostasis system; 7) AT; 8) LC characteristics; 9) LC cell dynamics; 10) surgery type: lobectomy/pneumonectomy; 11) anthropometric data. Optimal diagnosis and treatment strategies for LC are: 1) screening and early detection of LC; 2) availability of experienced thoracic surgeons because of complexity of radical procedures; 3) aggressive en block surgery and adequate lymph node dissection for completeness; 4) precise prediction; 5) adjuvant chemoimmunoradiotherapy for LCP with unfavorable prognosis.
TEST BANK for Operations Management, 14th Edition by William J. Stevenson, Ve...kevinkariuki227
TEST BANK for Operations Management, 14th Edition by William J. Stevenson, Verified Chapters 1 - 19, Complete Newest Version.pdf
TEST BANK for Operations Management, 14th Edition by William J. Stevenson, Verified Chapters 1 - 19, Complete Newest Version.pdf
These simplified slides by Dr. Sidra Arshad present an overview of the non-respiratory functions of the respiratory tract.
Learning objectives:
1. Enlist the non-respiratory functions of the respiratory tract
2. Briefly explain how these functions are carried out
3. Discuss the significance of dead space
4. Differentiate between minute ventilation and alveolar ventilation
5. Describe the cough and sneeze reflexes
Study Resources:
1. Chapter 39, Guyton and Hall Textbook of Medical Physiology, 14th edition
2. Chapter 34, Ganong’s Review of Medical Physiology, 26th edition
3. Chapter 17, Human Physiology by Lauralee Sherwood, 9th edition
4. Non-respiratory functions of the lungs https://academic.oup.com/bjaed/article/13/3/98/278874
MANAGEMENT OF ATRIOVENTRICULAR CONDUCTION BLOCK.pdfJim Jacob Roy
Cardiac conduction defects can occur due to various causes.
Atrioventricular conduction blocks ( AV blocks ) are classified into 3 types.
This document describes the acute management of AV block.
The prostate is an exocrine gland of the male mammalian reproductive system
It is a walnut-sized gland that forms part of the male reproductive system and is located in front of the rectum and just below the urinary bladder
Function is to store and secrete a clear, slightly alkaline fluid that constitutes 10-30% of the volume of the seminal fluid that along with the spermatozoa, constitutes semen
A healthy human prostate measures (4cm-vertical, by 3cm-horizontal, 2cm ant-post ).
It surrounds the urethra just below the urinary bladder. It has anterior, median, posterior and two lateral lobes
It’s work is regulated by androgens which are responsible for male sex characteristics
Generalised disease of the prostate due to hormonal derangement which leads to non malignant enlargement of the gland (increase in the number of epithelial cells and stromal tissue)to cause compression of the urethra leading to symptoms (LUTS
Title: Sense of Smell
Presenter: Dr. Faiza, Assistant Professor of Physiology
Qualifications:
MBBS (Best Graduate, AIMC Lahore)
FCPS Physiology
ICMT, CHPE, DHPE (STMU)
MPH (GC University, Faisalabad)
MBA (Virtual University of Pakistan)
Learning Objectives:
Describe the primary categories of smells and the concept of odor blindness.
Explain the structure and location of the olfactory membrane and mucosa, including the types and roles of cells involved in olfaction.
Describe the pathway and mechanisms of olfactory signal transmission from the olfactory receptors to the brain.
Illustrate the biochemical cascade triggered by odorant binding to olfactory receptors, including the role of G-proteins and second messengers in generating an action potential.
Identify different types of olfactory disorders such as anosmia, hyposmia, hyperosmia, and dysosmia, including their potential causes.
Key Topics:
Olfactory Genes:
3% of the human genome accounts for olfactory genes.
400 genes for odorant receptors.
Olfactory Membrane:
Located in the superior part of the nasal cavity.
Medially: Folds downward along the superior septum.
Laterally: Folds over the superior turbinate and upper surface of the middle turbinate.
Total surface area: 5-10 square centimeters.
Olfactory Mucosa:
Olfactory Cells: Bipolar nerve cells derived from the CNS (100 million), with 4-25 olfactory cilia per cell.
Sustentacular Cells: Produce mucus and maintain ionic and molecular environment.
Basal Cells: Replace worn-out olfactory cells with an average lifespan of 1-2 months.
Bowman’s Gland: Secretes mucus.
Stimulation of Olfactory Cells:
Odorant dissolves in mucus and attaches to receptors on olfactory cilia.
Involves a cascade effect through G-proteins and second messengers, leading to depolarization and action potential generation in the olfactory nerve.
Quality of a Good Odorant:
Small (3-20 Carbon atoms), volatile, water-soluble, and lipid-soluble.
Facilitated by odorant-binding proteins in mucus.
Membrane Potential and Action Potential:
Resting membrane potential: -55mV.
Action potential frequency in the olfactory nerve increases with odorant strength.
Adaptation Towards the Sense of Smell:
Rapid adaptation within the first second, with further slow adaptation.
Psychological adaptation greater than receptor adaptation, involving feedback inhibition from the central nervous system.
Primary Sensations of Smell:
Camphoraceous, Musky, Floral, Pepperminty, Ethereal, Pungent, Putrid.
Odor Detection Threshold:
Examples: Hydrogen sulfide (0.0005 ppm), Methyl-mercaptan (0.002 ppm).
Some toxic substances are odorless at lethal concentrations.
Characteristics of Smell:
Odor blindness for single substances due to lack of appropriate receptor protein.
Behavioral and emotional influences of smell.
Transmission of Olfactory Signals:
From olfactory cells to glomeruli in the olfactory bulb, involving lateral inhibition.
Primitive, less old, and new olfactory systems with different path
2. Introduction
• Most of the radiation detectors used in nuclear
medicine are operated in a “pulse mode”; that
is, they generate pulses of electrical charge or
current that are counted to determine the
number of radiation events detected.
• In addition, by analyzing the amplitude of
pulses from the detector, it is possible with
energy-sensitive detectors, such as scintillation
and semiconductor detectors and proportional
counters, to determine the energy of each
radiation event detected.
3. • Selection of a narrow energy range for
counting permits discrimination against
events other than those of the energy of
interest, such as scattered radiation and
background radiation or the multiple
emissions from a mixture of radionuclides.
4. • schematic form the basic electronic
components of a nuclear radiation-counting
instrument. These components are present
In systems ranging from the most simple
sample counters to complex imaging
instruments.
5. Preamplifiers
• The pulse output characteristics of detectors
used in nuclear medicine. Most of them
produce pulse signals of relatively small
amplitude. In addition, most of the detectors
listed have relatively high output impedance,
that is, a high internal resistance to the flow of
electrical current. In handling electronic
signals, it is important that the impedance
levels of successive components be matched to
one another, or electronic interferences that
distort the pulse signals may develop and
system performance will be degraded.
6. Typical signal output and pulse duration of
various radiation detectors.
Detector Signal (V)
Pulse Duration
(µsec)
Sodium iodide scintillator with
photomultiplier tube
10
−1
-1 0.23*
Lutetium oxyorthosilicate scintillator with
photomultiplier tube
10
−1
-1 0.04*
Liquid scintillator with photomultiplier
tube
10
−2
-10
−1
10
−2
*
Lutetium oxyorthosilicate scintillator with
avalanche photodiode
10
−5
-10
−4
0.04*
Direct semiconductor detector 10
−4
-10
−3
10
−1
-1
Gas proportional counter 10
−3
-10
−2
10
−1
-1
Geiger-Müller counter
1-10
50-300
7. The purposes of a preamplifier are :
1. To amplify, if necessary, the relatively small
signals produced by the radiation detector,
2. To match impedance levels between the
detector and subsequent components in
the system.
3. To shape the signal pulse for optimal signal
processing by the subsequent components.
8. There are two main types of preamplifier
configurations used with radiation
detectors:
The voltage-sensitive preamplifier
The charge-sensitive preamplifier.
9. Simplified circuit diagram of a voltage-sensitive preamplifier. The output voltage is
determined by the amount of charge from the radiation detector, the input capacitance Ci,
and the resistances R1 and R2. B, Simplified circuit diagram of a charge-sensitive
preamplifier. The output voltage is determined by the charge from the radiation .
10. With energy-sensitive detectors, the amount of charge, Q, and thus the
amplitude of the voltage Vi are proportional to the energy of the radiation event
detected. The output voltage, Vo, in this configuration is approximately
in which R1 and R2 are as shown . The minus sign indicates that the polarity of
the pulse has been changed.
In semiconductor detectors, the input capacitance of the detector is sensitive to
operating conditions, particularly temperature. Therefore the proportionality
between charge and the voltage seen at the preamp input may not be stable.
The charge-sensitive preamplifier overcomes this undesirable feature by using a
feedback capacitor of capacitance Cf to integrate the charge from the radiation
detector.
11. Amplifiers
Amplification and Pulse-Shaping Functions
The amplifier component of a nuclear counting
instrument has two major functions:
1. To amplify the still relatively small pulses from
the preamp to sufficient amplitude (volts) to
drive auxiliary equipment (pulse-height
analyzers, scalers, etc.)
2. To reshape the slow decaying pulse from the
preamp into a narrow one to avoid the problem
of pulse pile-up at high counting rates and to
improve the electronic SNR.
12. • The gain factor on an amplifier may range from
×1 to ×1000. Usually it is adjustable, first by a
coarse adjustment (i.e., ×2, ×4, ×8) and then by a
fine gain adjustment providing gain factors
between the coarse steps. The coarse gain
adjustment permits amplification of pulses over
a wide range of amplitudes from different
detectors and preamplifiers to the maximum
output capability of the amplifier. The fine gain
adjustment permits precise calibration of the
relationship between amplifier output pulse
amplitude (volts) and radiation energy absorbed
(keV or MeV). For example, a convenient ratio
might be 10 V of pulse amplitude per 1 MeV of
radiation energy absorbed in the detector.
13. Pulse shaping—i.e., pulse shortening—is an essential function of the amplifier. The
output of the preamp is a sharply rising pulse that decays with a time constant of about
50 µsec, returning to baseline after approximately 500 µsec. Thus if a second pulse
occurs within 500 µsec, it rides on the tail of the previous pulse, providing incorrect
amplitude information . The system could not operate at counting rates exceeding a
few hundred events per second without introducing this type of amplitude distortion.
14. • The pulse-shaping circuits of the amplifier must
provide an output of cleanly separated pulses, even
though the output pulses from the preamp overlap.
It must do this without distorting the information in
the preamplifier signal, which is, mainly,
1. Pulse amplitude (proportional to the energy of
radiation event for energy-sensitive detectors)
2. Rise time (time at which the radiation event was
detected). An additional function of the pulse-
shaping circuits is to discriminate against
electronic noise signals, such as microphonic
pickup and 50- to 120-Hz power line frequency
15. • The most common methods for amplifier
pulse shaping are resistor-capacitor (RC),
gaussian, and delay-line methods. The RC
technique, commonly referred to as RC
shaping, is described to illustrate the basic
principles.
16. Resistor-Capacitor Shaping
• Basic RC pulse-shaping circuits are shown. When a
sharply rising pulse of relatively long duration (e.g.,
preamplifier output pulse) is applied to a capacitor-
resistor (CR), or differentiation circuit the output is a
rapidly rising pulse that decays with a time constant
τd determined by the RC product of the circuit
components . The amplitude of the output pulse
depends on the amplitude of the sharply rising portion
of the input pulse and is insensitive to the “tail” of any
preceding pulse. Note that a CR differentiation circuit
also is used for pulse shaping in the preamplifier;
however, the time constants used in the preamplifier
circuits are much longer than those used in the
amplifier. This also illustrates how the CR circuit
discriminates against low-frequency noise signals.
17.
18. • shows an RC, or integration circuit. (Note
that differentiation and integration differ
only by the interchanging of the resistor R
and the capacitor C.) When a sharply rising
pulse is applied to this circuit, the output is
a pulse with a shape described by
where Vi is the amplitude of the input pulse
and RC = τi is the integration time constant
of the circuit. This circuit discriminates
effectively against high-frequency noise.
19. Baseline Shift and Pulse Pile-Up
• Baseline shift and pulse pile-up are two
practical problems that occur in all amplifiers
at high counting rates. Baseline shift is caused
by the negative component that occurs at the
end of the amplifier output pulse. A second
pulse occurring during this component will be
slightly depressed in amplitude (Fig. 8-6A).
Inaccurate pulse amplitude and an apparent
shift (decrease) in energy of the detected
radiation event are the result.
20.
21. • Special circuitry has been developed to minimize
baseline shift. This is called pole zero
cancellation, or baseline restoration. This type of
circuitry is employed in modern scintillation
cameras to provide a high counting rate
capability, particularly for cardiac studies.
• At high counting rates, amplifier pulses can occur
so close together that they fall on top of each
other. This is referred to as pulse pile-up When
this happens, two pulses sum together and
produce a single pulse with an amplitude that is
not representative of either. Pulse pile-up distorts
energy information and also contributes to
counting losses (dead time) of the detection
system, because two pulses are counted as one.
22. • Both baseline shift and pulse pile-up can be
decreased by decreasing the width of the
amplifier pulse (i.e., the time constant of
the amplifier); however, shortening of the
time constant usually produces poorer SNR
and energy resolution. It is generally true
that all the factors that provide high count
rate capabilities in amplifiers also degrade
energy resolution.
23. PULSE-HEIGHT ANALYZERS
Basic Functions
• When an energy-sensitive detector is used [e.g., NaI(Tl):PM
tube or a semiconductor detector], the amplitude of the voltage
pulse from the amplifier is proportional to the amount of energy
deposited in the detector by the detected radiation event.
• When an energy-sensitive detector is used [e.g., NaI(Tl):PM
tube or a semiconductor detector], the amplitude of the voltage
pulse from the amplifier is proportional to the amount of energy
deposited in the detector by the detected radiation event.
24. Single-Channel Analyzers
• An SCA is used to select for counting only those pulses from the amplifier
that fall within a selected voltage amplitude range. At this stage in the
system voltage amplitude is proportional to radiation energy deposited in
the detector, so it is equivalent to selecting an energy range for counting.
Modern amplifiers produce output pulses with amplitudes in the range of
0-10╯V. Therefore the voltage selection provided by most SCAs is also in
the 0- to 10-V range.
• The LLD and ULD establish voltage levels in electronic circuits called
comparators. As their name implies, these circuits compare the amplitude
of an input pulse with the LLD and ULD voltages. They produce an output
pulse only when these voltages are exceeded. Pulses from the comparator
circuits are then sent to the anticoincidence circuit, which produces an
output pulse when only one (LLD) but not both (ULD and LLD) pulses
are Present.
25.
26. Timing Methods
• Accurate time placement of the radiation event is
important in some nuclear medicineapplications. For
example, in the scintillation camera ,accurate timing is
required to identify the multiple phototubes involved in
detecting individual radiation events striking the NaI(Tl)
crystal
• Most SCAs used in nuclear medicine employ leading-edge
timing. With this method, as shown in Figure, the analyzer
output pulse occurs at a fixed time TD following the
instant at which the rising portion of the input pulse
triggers the LLD.
27.
28. Multichannel Analyzers
• Some applications of pulse-height analysis require simultaneous
recording of events in multiple voltage or energy windows. One
approach is to use many SCAs, each with its own voltage window.
• For example, some imaging devices have two or three independent SCAs
to record simultaneously the multiple γ-ray energies emitted by
nuclides such as 67Ga; however, this approach is unsatisfactory when
tens or even thousands of different windows are required.
• A practical solution is provided by an MCA. Figure demonstrates the
basic principles. The heart of the MCA is an analog-to-digital converter
(ADC), which measures and sorts out the incoming pulses according to
their amplitudes.
29.
30. DIGITAL COUNTERS AND RATE
METERS
1. Scalers, Timers, and Counters
• Digital counters are used to count output signals from radiation
detectors after pulseheight analysis of the signals. A device that counts
only pulses is called a scaler. An auxiliary device that controls the
scaler counting time is called a timer. An instrument that incorporates
both functions in a single unit is called a scaler-timer.
• These devices are often referred to under the generic name of counters.
The number of counts recorded and the elapsed counting time may be
displayed on a visual readout, or, more commonly, the output of the
scaler-timer may be interfaced to a personal computer automated data
processing.
31.
32. Analog Rate Meters
• An analog rate meter is used to determine the average number
of events (e.g., SCA output pulses) occurring per unit of time.
• The average is determined continuously, rather than during
discrete counting intervals, as would be the case with a scaler-
timer. The output of a rate meter is a continuously varying
voltage level proportional to the average rate at which pulses are
received at the rate meter input.
• The output voltage can be displayed on a front-panel meter or
interfaced through a continuously sampling ADC to a personal
computer. Rate meters are commonly used in radiation
monitors.
33.
34. HIGH-VOLTAGE POWER SUPPLIES
• The high-voltage (HV) power supply provides
the charge collection voltage for semiconductor,
gas proportional, and GM detectors and
the accelerating voltage for electron multiplication
in the PM tubes used with scintillation
detectors such as NaI(Tl) and liquid scintillators.
The HV power supply converts the alternating
current voltage provided by the line
source into a constant or direct current (DC)
voltage.
35. OSCILLOSCOPES
Analog Oscilloscope
A typical analog oscilloscope consists of a
CRT, a signal amplifier for the vertical deflection
plate of the CRT, and a time-sweep generator.
An amplifier is provided so that small
voltage inputs can be amplified and applied
to the vertical deflection plate to display the
amplitude of the input signals.
36. The time sweep generator is connected to the
horizontal deflection plates of the CRT to
sweep the electron beam across the screen at
a constant speed and repetition rate. The
horizontal sweep rate usually can be varied
from nanoseconds (10−9╯sec) to seconds per
centimeter by a calibrated selector switch on
the front panel of the oscilloscope.
37. Digital Oscilloscope
Most modern oscilloscopes are digital,
employing fast ADCs that digitize the amplified
waveforms prior to display, and some form of
microprocessor that allows pulses to be analyzed
and manipulated. The CRT screen is typically
replaced with a flat-panel liquid crystal display.
Digital oscilloscopes have the advantage that
pulses can be stored in computer memory for
further analysis and are ideal for studying
repetitive, regular pulses.
38. • The disadvantage of using a digital
oscilloscope to look at the pulses from γ-ray
detectors is that individual pulses generally
are of a different amplitude (reflecting
differing energies deposited in the
detector), and that they arrive randomly in
time.
39. REFERENCES
• Physics in Nuclear medicine Simon R
Cherry , James A Sorenson , Michael A
Phelps.
• Knoll GF: Radiation Detection and
Measurement, ed 4, New York, 2010, John
Wiley. Leo WR: Techniques for Nuclear and
Particle Physics Experiments, ed 2, New
York, 1994, Springer-Verlag.