SlideShare a Scribd company logo
Radiation Detection &
   Measurement I
    Types of detectors
Types of detectors
• Gas-filled detectors consist of a volume of gas
  between two electrodes
• In scintillation detectors, the interaction of
  ionizing radiation produces UV and/or visible
  light
• Semiconductor detectors are especially pure
  crystals of silicon, germanium, or other materials
  to which trace amounts of impurity atoms have
  been added so that they act as diodes
Types of detectors (cont.)
• Detectors may also be classified by the type of
  information produced:
   – Detectors, such as Geiger-Mueller (GM) detectors, that
     indicate the number of interactions occurring in the
     detector are called counters
   – Detectors that yield information about the energy
     distribution of the incident radiation, such as NaI
     scintillation detectors, are called spectrometers
   – Detectors that indicate the net amount of energy
     deposited in the detector by multiple interactions are
     called dosimeters
Modes of operation
• In pulse mode, the signal from each
  interaction is processed individually
• In current mode, the electrical signals from
  individual interactions are averaged
  together, forming a net current signal
Interaction rate
• Main problem with detectors in pulse mode is that
  two interactions must be separated by a finite
  amount of time if they are to produce distinct
  signals
• This interval is called the dead time of the system
• If a second interaction occurs in this interval, its
  signal will be lost; if it occurs close enough to the
  first interaction, it may distort the signal from the
  first interaction
Dead time
• Dead time of a detector system largely determined
  by the component in the series with the longest
  dead time
   – Detector has longest dead time in GM counter systems
   – In multichannel analyzer systems the analog-to-digital
     converter often has the longest dead time
• GM counters have dead times ranging from tens to
  hundreds of microseconds, most other systems
  have dead times of less than a few microseconds
Paralyzable or nonparalyzable
• In a paralyzable system, an interaction that
  occurs during the dead time after a previous
  interaction extends the dead time
• In a nonparalyzable system, it does not
• At very high interaction rates, a paralyzable
  system will be unable to detect any
  interactions after the first, causing the
  detector to indicate a count rate of zero
Current mode operation
• In current mode, all information regarding
  individual interactions is lost
• If the amount of electrical charge collected from
  each interaction is proportional to the energy
  deposited by that interaction, then the net current
  is proportional to the dose rate in the detector
  material
• Used for detectors subjected to very high
  interaction rates
Spectroscopy
• Most spectrometers operated in pulse mode
• Amplitude of each pulse is proportional to the
  energy deposited in the detector by the interaction
  causing that pulse
• The energy deposited by an interaction is not
  always the total energy of the incident particle or
  photon
• A pulse height spectrum is usually depicted as a
  graph of the number of interactions depositing a
  particular amount of energy in the spectrometer as
  a function of energy
Detection efficiency
• The efficiency (sensitivity) of a detector is a
  measure of its ability to detect radiation
• Efficiency of a detection system operated in
  pulse mode is defined as the probability that
  a particle or photon emitted by a source will
  be detected
Number detected
Efficiency =
             Number emitted
             Number reaching detector
Efficiency =                            ×
                 Number emitted
                     Number detected
                Number reaching detector
Efficiency = Geometric efficiency × Intrinsic efficiency
Intrinsic efficiency
• Often called the quantum detection efficiency or
  QDE
• Determined by the energy of the photons and the
  atomic number, density, and thickness of the
  detector
• For a a parallel beam of monoenergetic photons
  incident on a detector of uniform thickness:
                                        − µx
         Intrinsic efficiency = 1 - e
Gas-filled detectors
• A gas-filled detector consists of a volume of gas
  between two electrodes, with an electrical
  potential difference (voltage) applied between the
  electrodes
• Ionizing radiation produces ion pairs in the gas
• Positive ions (cations) attracted to negative
  electrode (cathode); electrons or anions attracted
  to positive electrode (anode)
• In most detectors, cathode is the wall of the
  container that holds the gas and anode is a wire
  inside the container
Types of gas-filled detectors
• Three types of gas-filled detectors in common use:
   – Ionization chambers
   – Proportional counters
   – Geiger-Mueller (GM) counters
• Type determined primarily by the voltage applied
  between the two electrodes
• Ionization chambers have wider range of physical
  shape (parallel plates, concentric cylinders, etc.)
• Proportional counters and GM counters must have
  thin wire anode
Ionization chambers
• If gas is air and walls of chamber are of a material
  whose effective atomic number is similar to air,
  the amount of current produced is proportional to
  the exposure rate
• Air-filled ion chambers are used in portable
  survey meters, for performing QA testing of
  diagnostic and therapeutic x-ray machines, and are
  the detectors in most x-ray machine phototimers
• Low intrinsic efficiencies because of low densities
  of gases and low atomic numbers of most gases
Proportional counters
• Must contain a gas with specific properties
• Commonly used in standards laboratories,
  health physics laboratories, and for physics
  research
• Seldom used in medical centers
GM counters
• GM counters also must contain gases with specific
  properties
• Gas amplification produces billions of ion pairs
  after an interaction – signal from detector requires
  little amplification
• Often used for inexpensive survey meters
• In general, GM survey meters are inefficient
  detectors of x-rays and gamma rays
• Over-response to low energy x-rays – partially
  corrected by placing a thin layer of higher atomic
  number material around the detector
GM counters (cont.)
• GM detectors suffer from extremely long dead
  times – seldom used when accurate measurements
  are required of count rates greater than a few
  hundred counts per second
• Portable GM survey meter may become paralyzed
  in a very high radiation field – should always use
  ionization chamber instruments for measuring
  such fields
Scintillation detectors
• Scintillators are used in conventional film-screen
  radiography, many digital radiographic receptors,
  fluoroscopy, scintillation cameras, most CT
  scanners, and PET scanners
• Scintillation detectors consist of a scintillator and
  a device, such as a PMT, that converts the light
  into an electrical signal
Scintillators
• Desirable properties:
   – High conversion efficiency
   – Decay times of excited states should be short
   – Material transparent to its own emissions
   – Color of emitted light should match spectral sensitivity
     of the light receptor
   – For x-ray and gamma-ray detectors, µ should be large –
     high detection efficiencies
   – Rugged, unaffected by moisture, and inexpensive to
     manufacture
Scintillators (cont.)
• Amount of light emitted after an interaction
  increases with energy deposited by the
  interaction
• May be operated in pulse mode as
  spectrometers
• High conversion efficiency produces
  superior energy resolution
Materials
• Sodium iodide activated with thallium
  [NaI(Tl)], coupled to PMTs and operated in
  pulse mode, is used for most nuclear
  medicine applications
  – Fragile and hygroscopic
• Bismuth germanate (BGO) is coupled to
  PMTs and used in pulse mode as detectors
  in most PET scanners
Photomultiplier tubes
• PMTs perform two functions:
  – Conversion of ultraviolet and visible light
    photons into an electrical signal
  – Signal amplification, on the order of millions to
    billions
• Consists of an evacuated glass tube
  containing a photocathode, typically 10 to
  12 electrodes called dynodes, and an anode
Dynodes
• Electrons emitted by the photocathode are
  attracted to the first dynode and are accelerated to
  kinetic energies equal to the potential difference
  between the photocathode and the first dynode
• When these electrons strike the first dynode, about
  5 electrons are ejected from the dynode for each
  electron hitting it
• These electrons are attracted to the second dynode,
  and so on, finally reaching the anode
PMT amplification
• Total amplification of the PMT is the product of
  the individual amplifications at each dynode
• If a PMT has ten dynodes and the amplification at
  each stage is 5, the total amplification will be
  approximately 10,000,000
• Amplification can be adjusted by changing the
  voltage applied to the PMT

More Related Content

What's hot

Radiation detection & measurement
Radiation detection & measurementRadiation detection & measurement
Radiation detection & measurement
mahbubul hassan
 
Scintillation Detector.pptx
Scintillation Detector.pptxScintillation Detector.pptx
Scintillation Detector.pptx
MandeepKaur528541
 
Chapter3 1 basics_dosimetry
Chapter3 1 basics_dosimetryChapter3 1 basics_dosimetry
Chapter3 1 basics_dosimetry
Nelis Van Peteghem
 
Ionization chamber
Ionization chamberIonization chamber
Ionization chamberAnas Yess
 
Gas filled detectors
Gas filled detectorsGas filled detectors
Gas filled detectors
Amara Usman
 
1 radiation detection and measurement
1 radiation detection and measurement 1 radiation detection and measurement
1 radiation detection and measurement Shahid Younas
 
Solid state detector mamita
Solid state detector mamitaSolid state detector mamita
Solid state detector mamita
Mamita Sakhakarmi
 
Nuclear radiation detector
Nuclear radiation detectorNuclear radiation detector
Nuclear radiation detector
siddharth gupta
 
Ionization Chambers
Ionization ChambersIonization Chambers
Ionization Chambers
Dilshad Kottuparamban
 
Radiation detectors
Radiation detectorsRadiation detectors
Radiation detectors
Kiran Ramakrishna
 
RADIATION DETECTION AND MEASUREMENT ppt 2.pptx
RADIATION DETECTION AND MEASUREMENT ppt 2.pptxRADIATION DETECTION AND MEASUREMENT ppt 2.pptx
RADIATION DETECTION AND MEASUREMENT ppt 2.pptx
Srinath Chowdary
 
Nuclear i nstrumentation
Nuclear i nstrumentationNuclear i nstrumentation
Nuclear i nstrumentation
shalet kochumuttath Shaji
 
Scintillation Counter and Semiconductor Detector
Scintillation Counter and Semiconductor DetectorScintillation Counter and Semiconductor Detector
Radioactive sources
Radioactive sourcesRadioactive sources
Radioactive sources
Kiran Ramakrishna
 
Proportional counter
Proportional counterProportional counter
Proportional counter
Harish kumawat
 
Radiation detection and measurement
Radiation detection and measurement Radiation detection and measurement
Radiation detection and measurement Shahid Younas
 
Interaction of radiation with matter
Interaction of radiation with matterInteraction of radiation with matter
Interaction of radiation with matter
Sabari Kumar
 
nuclear radiation detector unit V
nuclear radiation detector unit Vnuclear radiation detector unit V
nuclear radiation detector unit V
Dr. Vishal Jain
 
Basic dosimetric principle and dosimeters
Basic dosimetric principle and dosimetersBasic dosimetric principle and dosimeters
Basic dosimetric principle and dosimeters
Vinay Desai
 
Irradiation products of water & Fricks dosimetry
Irradiation products of water & Fricks dosimetryIrradiation products of water & Fricks dosimetry
Irradiation products of water & Fricks dosimetryGopi krishna Giri
 

What's hot (20)

Radiation detection & measurement
Radiation detection & measurementRadiation detection & measurement
Radiation detection & measurement
 
Scintillation Detector.pptx
Scintillation Detector.pptxScintillation Detector.pptx
Scintillation Detector.pptx
 
Chapter3 1 basics_dosimetry
Chapter3 1 basics_dosimetryChapter3 1 basics_dosimetry
Chapter3 1 basics_dosimetry
 
Ionization chamber
Ionization chamberIonization chamber
Ionization chamber
 
Gas filled detectors
Gas filled detectorsGas filled detectors
Gas filled detectors
 
1 radiation detection and measurement
1 radiation detection and measurement 1 radiation detection and measurement
1 radiation detection and measurement
 
Solid state detector mamita
Solid state detector mamitaSolid state detector mamita
Solid state detector mamita
 
Nuclear radiation detector
Nuclear radiation detectorNuclear radiation detector
Nuclear radiation detector
 
Ionization Chambers
Ionization ChambersIonization Chambers
Ionization Chambers
 
Radiation detectors
Radiation detectorsRadiation detectors
Radiation detectors
 
RADIATION DETECTION AND MEASUREMENT ppt 2.pptx
RADIATION DETECTION AND MEASUREMENT ppt 2.pptxRADIATION DETECTION AND MEASUREMENT ppt 2.pptx
RADIATION DETECTION AND MEASUREMENT ppt 2.pptx
 
Nuclear i nstrumentation
Nuclear i nstrumentationNuclear i nstrumentation
Nuclear i nstrumentation
 
Scintillation Counter and Semiconductor Detector
Scintillation Counter and Semiconductor DetectorScintillation Counter and Semiconductor Detector
Scintillation Counter and Semiconductor Detector
 
Radioactive sources
Radioactive sourcesRadioactive sources
Radioactive sources
 
Proportional counter
Proportional counterProportional counter
Proportional counter
 
Radiation detection and measurement
Radiation detection and measurement Radiation detection and measurement
Radiation detection and measurement
 
Interaction of radiation with matter
Interaction of radiation with matterInteraction of radiation with matter
Interaction of radiation with matter
 
nuclear radiation detector unit V
nuclear radiation detector unit Vnuclear radiation detector unit V
nuclear radiation detector unit V
 
Basic dosimetric principle and dosimeters
Basic dosimetric principle and dosimetersBasic dosimetric principle and dosimeters
Basic dosimetric principle and dosimeters
 
Irradiation products of water & Fricks dosimetry
Irradiation products of water & Fricks dosimetryIrradiation products of water & Fricks dosimetry
Irradiation products of water & Fricks dosimetry
 

Viewers also liked

Module 7 radiation detection, american fork fire rescue
Module 7 radiation detection, american fork fire rescueModule 7 radiation detection, american fork fire rescue
Module 7 radiation detection, american fork fire rescuejhendrickson1983
 
Geiger muller counting system
Geiger muller counting systemGeiger muller counting system
Geiger muller counting systemGaurav Bhati
 
Radiation detection devices
Radiation detection devicesRadiation detection devices
Radiation detection deviceswaynelowell
 
Geiger muller counter
Geiger muller counterGeiger muller counter
Geiger muller counter
Britto Samuel
 
Geiger–Müller Counter
Geiger–Müller CounterGeiger–Müller Counter
Geiger–Müller Counter
ABHISEK KUMAR SINGH
 
Non-Imaging Devices in Nuclear Medicine
Non-Imaging Devices in Nuclear MedicineNon-Imaging Devices in Nuclear Medicine
Non-Imaging Devices in Nuclear MedicinePawitra Masa-ah
 
Nuclear Medicine Instrumentation and quality control presentation
 Nuclear Medicine Instrumentation and quality control presentation Nuclear Medicine Instrumentation and quality control presentation
Nuclear Medicine Instrumentation and quality control presentation
Mumba Chilimboyi
 
79035bb0 a6be-4cc1-8f52-b943b8d5af78 (1)
79035bb0 a6be-4cc1-8f52-b943b8d5af78 (1)79035bb0 a6be-4cc1-8f52-b943b8d5af78 (1)
79035bb0 a6be-4cc1-8f52-b943b8d5af78 (1)
GGS Medical College/Baba Farid Univ.of Health Sciences.
 
III. Migrations
III. MigrationsIII. Migrations
III. Migrationsaldelaitre
 
Great Leaders – Alfred Sloan
Great Leaders – Alfred SloanGreat Leaders – Alfred Sloan
Great Leaders – Alfred Sloan
Shiv Shivakumar
 
Micro autoradiography pptx
Micro autoradiography pptxMicro autoradiography pptx
Micro autoradiography pptx
ravi kiran
 
Haz Mat Monitoring And You!!!
Haz Mat Monitoring And You!!!Haz Mat Monitoring And You!!!
Haz Mat Monitoring And You!!!
disastermaster
 
Radiology day 1 mammography
Radiology day 1 mammographyRadiology day 1 mammography
Radiology day 1 mammography
Vibhay Pareek
 
Interactions of radiation_with_matter
Interactions of radiation_with_matterInteractions of radiation_with_matter
Interactions of radiation_with_matterFernando Nainggolan
 
Protein structure determination from hybrid NMR data.
Protein structure determination from hybrid NMR data.Protein structure determination from hybrid NMR data.
Protein structure determination from hybrid NMR data.
Mark Berjanskii
 

Viewers also liked (16)

Module 7 radiation detection, american fork fire rescue
Module 7 radiation detection, american fork fire rescueModule 7 radiation detection, american fork fire rescue
Module 7 radiation detection, american fork fire rescue
 
Geiger muller counting system
Geiger muller counting systemGeiger muller counting system
Geiger muller counting system
 
Radiation detection devices
Radiation detection devicesRadiation detection devices
Radiation detection devices
 
Geiger muller counter
Geiger muller counterGeiger muller counter
Geiger muller counter
 
Geiger–Müller Counter
Geiger–Müller CounterGeiger–Müller Counter
Geiger–Müller Counter
 
Non-Imaging Devices in Nuclear Medicine
Non-Imaging Devices in Nuclear MedicineNon-Imaging Devices in Nuclear Medicine
Non-Imaging Devices in Nuclear Medicine
 
Nuclear Medicine Instrumentation and quality control presentation
 Nuclear Medicine Instrumentation and quality control presentation Nuclear Medicine Instrumentation and quality control presentation
Nuclear Medicine Instrumentation and quality control presentation
 
79035bb0 a6be-4cc1-8f52-b943b8d5af78 (1)
79035bb0 a6be-4cc1-8f52-b943b8d5af78 (1)79035bb0 a6be-4cc1-8f52-b943b8d5af78 (1)
79035bb0 a6be-4cc1-8f52-b943b8d5af78 (1)
 
III. Migrations
III. MigrationsIII. Migrations
III. Migrations
 
Great Leaders – Alfred Sloan
Great Leaders – Alfred SloanGreat Leaders – Alfred Sloan
Great Leaders – Alfred Sloan
 
Micro autoradiography pptx
Micro autoradiography pptxMicro autoradiography pptx
Micro autoradiography pptx
 
Haz Mat Monitoring And You!!!
Haz Mat Monitoring And You!!!Haz Mat Monitoring And You!!!
Haz Mat Monitoring And You!!!
 
Radiology day 1 mammography
Radiology day 1 mammographyRadiology day 1 mammography
Radiology day 1 mammography
 
Interactions of radiation_with_matter
Interactions of radiation_with_matterInteractions of radiation_with_matter
Interactions of radiation_with_matter
 
Protein structure determination from hybrid NMR data.
Protein structure determination from hybrid NMR data.Protein structure determination from hybrid NMR data.
Protein structure determination from hybrid NMR data.
 
Radioactivity
RadioactivityRadioactivity
Radioactivity
 

Similar to 25 -radiation_detection_&_measurement_i

Neutron detection in medical industry and hsopital
Neutron detection in medical industry and hsopitalNeutron detection in medical industry and hsopital
Neutron detection in medical industry and hsopital
medicalHunter2
 
Radio-active Counters
Radio-active CountersRadio-active Counters
Radio-active Counters
Nizam Ashraf
 
instrumenation all.pptx
instrumenation all.pptxinstrumenation all.pptx
instrumenation all.pptx
reshadnuredin1
 
FOURIER TRANSFORM - INFRARED SPECTROSCOPY
FOURIER TRANSFORM - INFRARED SPECTROSCOPYFOURIER TRANSFORM - INFRARED SPECTROSCOPY
FOURIER TRANSFORM - INFRARED SPECTROSCOPY
SalmanLatif14
 
upload RADIATION DETECTION AND MEASUREMENT.pptx
upload RADIATION DETECTION AND MEASUREMENT.pptxupload RADIATION DETECTION AND MEASUREMENT.pptx
upload RADIATION DETECTION AND MEASUREMENT.pptx
Preethi Natarajan
 
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptxLec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Ali Ayaz
 
MEASUREMENT OF RADIATION DOSE.pptx
MEASUREMENT OF RADIATION DOSE.pptxMEASUREMENT OF RADIATION DOSE.pptx
MEASUREMENT OF RADIATION DOSE.pptx
VishnuDutt47
 
Microwave measurements in detail
Microwave measurements in detailMicrowave measurements in detail
Microwave measurements in detail
Hedayath Basha Shaik
 
3ok industrial industrialinstructions.pdf
3ok industrial industrialinstructions.pdf3ok industrial industrialinstructions.pdf
3ok industrial industrialinstructions.pdf
qweewq2302
 
Radiation/Particle Detectors
Radiation/Particle DetectorsRadiation/Particle Detectors
Radiation/Particle Detectors
Erich Wanzek
 
Pharmaceutical Inorganic chemistry UNIT-V Radiopharmaceutical.pptx
Pharmaceutical Inorganic chemistry UNIT-V Radiopharmaceutical.pptxPharmaceutical Inorganic chemistry UNIT-V Radiopharmaceutical.pptx
Pharmaceutical Inorganic chemistry UNIT-V Radiopharmaceutical.pptx
Ms. Pooja Bhandare
 
Radiopharmaceutical
Radiopharmaceutical Radiopharmaceutical
Radiopharmaceutical
HimanshuJoshi255
 
UV / VIS
UV / VIS UV / VIS
COMPARISION OF PHOTODETECTORS.pptx
COMPARISION OF PHOTODETECTORS.pptxCOMPARISION OF PHOTODETECTORS.pptx
COMPARISION OF PHOTODETECTORS.pptx
tamilselvan sivasamy
 
Sensor and transducers lect 1
Sensor and transducers lect 1Sensor and transducers lect 1
Sensor and transducers lect 1
prashant tripathi
 
Sensors
SensorsSensors
Sensors
Bipin Mishra
 

Similar to 25 -radiation_detection_&_measurement_i (20)

Neutron detection in medical industry and hsopital
Neutron detection in medical industry and hsopitalNeutron detection in medical industry and hsopital
Neutron detection in medical industry and hsopital
 
Radio-active Counters
Radio-active CountersRadio-active Counters
Radio-active Counters
 
Uv vis-instrument-1
Uv vis-instrument-1Uv vis-instrument-1
Uv vis-instrument-1
 
instrumenation all.pptx
instrumenation all.pptxinstrumenation all.pptx
instrumenation all.pptx
 
FOURIER TRANSFORM - INFRARED SPECTROSCOPY
FOURIER TRANSFORM - INFRARED SPECTROSCOPYFOURIER TRANSFORM - INFRARED SPECTROSCOPY
FOURIER TRANSFORM - INFRARED SPECTROSCOPY
 
upload RADIATION DETECTION AND MEASUREMENT.pptx
upload RADIATION DETECTION AND MEASUREMENT.pptxupload RADIATION DETECTION AND MEASUREMENT.pptx
upload RADIATION DETECTION AND MEASUREMENT.pptx
 
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptxLec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
Lec2 Ali 5.Lecture 5 - CT Scan Data Acquisition System.pptx
 
MEASUREMENT OF RADIATION DOSE.pptx
MEASUREMENT OF RADIATION DOSE.pptxMEASUREMENT OF RADIATION DOSE.pptx
MEASUREMENT OF RADIATION DOSE.pptx
 
Microwave measurements in detail
Microwave measurements in detailMicrowave measurements in detail
Microwave measurements in detail
 
Presentation1
Presentation1Presentation1
Presentation1
 
3ok industrial industrialinstructions.pdf
3ok industrial industrialinstructions.pdf3ok industrial industrialinstructions.pdf
3ok industrial industrialinstructions.pdf
 
Radiation/Particle Detectors
Radiation/Particle DetectorsRadiation/Particle Detectors
Radiation/Particle Detectors
 
Pharmaceutical Inorganic chemistry UNIT-V Radiopharmaceutical.pptx
Pharmaceutical Inorganic chemistry UNIT-V Radiopharmaceutical.pptxPharmaceutical Inorganic chemistry UNIT-V Radiopharmaceutical.pptx
Pharmaceutical Inorganic chemistry UNIT-V Radiopharmaceutical.pptx
 
Radiopharmaceutical
Radiopharmaceutical Radiopharmaceutical
Radiopharmaceutical
 
UV / VIS
UV / VIS UV / VIS
UV / VIS
 
COMPARISION OF PHOTODETECTORS.pptx
COMPARISION OF PHOTODETECTORS.pptxCOMPARISION OF PHOTODETECTORS.pptx
COMPARISION OF PHOTODETECTORS.pptx
 
Sensors
SensorsSensors
Sensors
 
Sensor and transducers lect 1
Sensor and transducers lect 1Sensor and transducers lect 1
Sensor and transducers lect 1
 
Sensors
SensorsSensors
Sensors
 
Ei unit 2
Ei unit 2Ei unit 2
Ei unit 2
 

Recently uploaded

Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Savita Shen $i11
 
ARTHROLOGY PPT NCISM SYLLABUS AYURVEDA STUDENTS
ARTHROLOGY PPT NCISM SYLLABUS AYURVEDA STUDENTSARTHROLOGY PPT NCISM SYLLABUS AYURVEDA STUDENTS
ARTHROLOGY PPT NCISM SYLLABUS AYURVEDA STUDENTS
Dr. Vinay Pareek
 
The Normal Electrocardiogram - Part I of II
The Normal Electrocardiogram - Part I of IIThe Normal Electrocardiogram - Part I of II
The Normal Electrocardiogram - Part I of II
MedicoseAcademics
 
Antiulcer drugs Advance Pharmacology .pptx
Antiulcer drugs Advance Pharmacology .pptxAntiulcer drugs Advance Pharmacology .pptx
Antiulcer drugs Advance Pharmacology .pptx
Rohit chaurpagar
 
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
Oleg Kshivets
 
THOA 2.ppt Human Organ Transplantation Act
THOA 2.ppt Human Organ Transplantation ActTHOA 2.ppt Human Organ Transplantation Act
THOA 2.ppt Human Organ Transplantation Act
DrSathishMS1
 
Ocular injury ppt Upendra pal optometrist upums saifai etawah
Ocular injury  ppt  Upendra pal  optometrist upums saifai etawahOcular injury  ppt  Upendra pal  optometrist upums saifai etawah
Ocular injury ppt Upendra pal optometrist upums saifai etawah
pal078100
 
Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...
Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...
Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...
VarunMahajani
 
How to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for DoctorsHow to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for Doctors
LanceCatedral
 
Physiology of Chemical Sensation of smell.pdf
Physiology of Chemical Sensation of smell.pdfPhysiology of Chemical Sensation of smell.pdf
Physiology of Chemical Sensation of smell.pdf
MedicoseAcademics
 
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness JourneyTom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
greendigital
 
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 UpakalpaniyaadhyayaCharaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Dr KHALID B.M
 
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.GawadHemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
NephroTube - Dr.Gawad
 
Prix Galien International 2024 Forum Program
Prix Galien International 2024 Forum ProgramPrix Galien International 2024 Forum Program
Prix Galien International 2024 Forum Program
Levi Shapiro
 
Surgical Site Infections, pathophysiology, and prevention.pptx
Surgical Site Infections, pathophysiology, and prevention.pptxSurgical Site Infections, pathophysiology, and prevention.pptx
Surgical Site Infections, pathophysiology, and prevention.pptx
jval Landero
 
Superficial & Deep Fascia of the NECK.pptx
Superficial & Deep Fascia of the NECK.pptxSuperficial & Deep Fascia of the NECK.pptx
Superficial & Deep Fascia of the NECK.pptx
Dr. Rabia Inam Gandapore
 
KDIGO 2024 guidelines for diabetologists
KDIGO 2024 guidelines for diabetologistsKDIGO 2024 guidelines for diabetologists
KDIGO 2024 guidelines for diabetologists
د.محمود نجيب
 
Couples presenting to the infertility clinic- Do they really have infertility...
Couples presenting to the infertility clinic- Do they really have infertility...Couples presenting to the infertility clinic- Do they really have infertility...
Couples presenting to the infertility clinic- Do they really have infertility...
Sujoy Dasgupta
 
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdf
ARTIFICIAL INTELLIGENCE IN  HEALTHCARE.pdfARTIFICIAL INTELLIGENCE IN  HEALTHCARE.pdf
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdf
Anujkumaranit
 
Charaka Samhita Sutra Sthana 9 Chapter khuddakachatuspadadhyaya
Charaka Samhita Sutra Sthana 9 Chapter khuddakachatuspadadhyayaCharaka Samhita Sutra Sthana 9 Chapter khuddakachatuspadadhyaya
Charaka Samhita Sutra Sthana 9 Chapter khuddakachatuspadadhyaya
Dr KHALID B.M
 

Recently uploaded (20)

Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
Phone Us ❤85270-49040❤ #ℂall #gIRLS In Surat By Surat @ℂall @Girls Hotel With...
 
ARTHROLOGY PPT NCISM SYLLABUS AYURVEDA STUDENTS
ARTHROLOGY PPT NCISM SYLLABUS AYURVEDA STUDENTSARTHROLOGY PPT NCISM SYLLABUS AYURVEDA STUDENTS
ARTHROLOGY PPT NCISM SYLLABUS AYURVEDA STUDENTS
 
The Normal Electrocardiogram - Part I of II
The Normal Electrocardiogram - Part I of IIThe Normal Electrocardiogram - Part I of II
The Normal Electrocardiogram - Part I of II
 
Antiulcer drugs Advance Pharmacology .pptx
Antiulcer drugs Advance Pharmacology .pptxAntiulcer drugs Advance Pharmacology .pptx
Antiulcer drugs Advance Pharmacology .pptx
 
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
 
THOA 2.ppt Human Organ Transplantation Act
THOA 2.ppt Human Organ Transplantation ActTHOA 2.ppt Human Organ Transplantation Act
THOA 2.ppt Human Organ Transplantation Act
 
Ocular injury ppt Upendra pal optometrist upums saifai etawah
Ocular injury  ppt  Upendra pal  optometrist upums saifai etawahOcular injury  ppt  Upendra pal  optometrist upums saifai etawah
Ocular injury ppt Upendra pal optometrist upums saifai etawah
 
Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...
Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...
Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...
 
How to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for DoctorsHow to Give Better Lectures: Some Tips for Doctors
How to Give Better Lectures: Some Tips for Doctors
 
Physiology of Chemical Sensation of smell.pdf
Physiology of Chemical Sensation of smell.pdfPhysiology of Chemical Sensation of smell.pdf
Physiology of Chemical Sensation of smell.pdf
 
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness JourneyTom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
 
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 UpakalpaniyaadhyayaCharaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
 
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.GawadHemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
Hemodialysis: Chapter 3, Dialysis Water Unit - Dr.Gawad
 
Prix Galien International 2024 Forum Program
Prix Galien International 2024 Forum ProgramPrix Galien International 2024 Forum Program
Prix Galien International 2024 Forum Program
 
Surgical Site Infections, pathophysiology, and prevention.pptx
Surgical Site Infections, pathophysiology, and prevention.pptxSurgical Site Infections, pathophysiology, and prevention.pptx
Surgical Site Infections, pathophysiology, and prevention.pptx
 
Superficial & Deep Fascia of the NECK.pptx
Superficial & Deep Fascia of the NECK.pptxSuperficial & Deep Fascia of the NECK.pptx
Superficial & Deep Fascia of the NECK.pptx
 
KDIGO 2024 guidelines for diabetologists
KDIGO 2024 guidelines for diabetologistsKDIGO 2024 guidelines for diabetologists
KDIGO 2024 guidelines for diabetologists
 
Couples presenting to the infertility clinic- Do they really have infertility...
Couples presenting to the infertility clinic- Do they really have infertility...Couples presenting to the infertility clinic- Do they really have infertility...
Couples presenting to the infertility clinic- Do they really have infertility...
 
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdf
ARTIFICIAL INTELLIGENCE IN  HEALTHCARE.pdfARTIFICIAL INTELLIGENCE IN  HEALTHCARE.pdf
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdf
 
Charaka Samhita Sutra Sthana 9 Chapter khuddakachatuspadadhyaya
Charaka Samhita Sutra Sthana 9 Chapter khuddakachatuspadadhyayaCharaka Samhita Sutra Sthana 9 Chapter khuddakachatuspadadhyaya
Charaka Samhita Sutra Sthana 9 Chapter khuddakachatuspadadhyaya
 

25 -radiation_detection_&_measurement_i

  • 1. Radiation Detection & Measurement I Types of detectors
  • 2. Types of detectors • Gas-filled detectors consist of a volume of gas between two electrodes • In scintillation detectors, the interaction of ionizing radiation produces UV and/or visible light • Semiconductor detectors are especially pure crystals of silicon, germanium, or other materials to which trace amounts of impurity atoms have been added so that they act as diodes
  • 3. Types of detectors (cont.) • Detectors may also be classified by the type of information produced: – Detectors, such as Geiger-Mueller (GM) detectors, that indicate the number of interactions occurring in the detector are called counters – Detectors that yield information about the energy distribution of the incident radiation, such as NaI scintillation detectors, are called spectrometers – Detectors that indicate the net amount of energy deposited in the detector by multiple interactions are called dosimeters
  • 4. Modes of operation • In pulse mode, the signal from each interaction is processed individually • In current mode, the electrical signals from individual interactions are averaged together, forming a net current signal
  • 5. Interaction rate • Main problem with detectors in pulse mode is that two interactions must be separated by a finite amount of time if they are to produce distinct signals • This interval is called the dead time of the system • If a second interaction occurs in this interval, its signal will be lost; if it occurs close enough to the first interaction, it may distort the signal from the first interaction
  • 6. Dead time • Dead time of a detector system largely determined by the component in the series with the longest dead time – Detector has longest dead time in GM counter systems – In multichannel analyzer systems the analog-to-digital converter often has the longest dead time • GM counters have dead times ranging from tens to hundreds of microseconds, most other systems have dead times of less than a few microseconds
  • 7. Paralyzable or nonparalyzable • In a paralyzable system, an interaction that occurs during the dead time after a previous interaction extends the dead time • In a nonparalyzable system, it does not • At very high interaction rates, a paralyzable system will be unable to detect any interactions after the first, causing the detector to indicate a count rate of zero
  • 8.
  • 9. Current mode operation • In current mode, all information regarding individual interactions is lost • If the amount of electrical charge collected from each interaction is proportional to the energy deposited by that interaction, then the net current is proportional to the dose rate in the detector material • Used for detectors subjected to very high interaction rates
  • 10. Spectroscopy • Most spectrometers operated in pulse mode • Amplitude of each pulse is proportional to the energy deposited in the detector by the interaction causing that pulse • The energy deposited by an interaction is not always the total energy of the incident particle or photon • A pulse height spectrum is usually depicted as a graph of the number of interactions depositing a particular amount of energy in the spectrometer as a function of energy
  • 11.
  • 12. Detection efficiency • The efficiency (sensitivity) of a detector is a measure of its ability to detect radiation • Efficiency of a detection system operated in pulse mode is defined as the probability that a particle or photon emitted by a source will be detected
  • 13. Number detected Efficiency = Number emitted Number reaching detector Efficiency = × Number emitted Number detected Number reaching detector Efficiency = Geometric efficiency × Intrinsic efficiency
  • 14.
  • 15. Intrinsic efficiency • Often called the quantum detection efficiency or QDE • Determined by the energy of the photons and the atomic number, density, and thickness of the detector • For a a parallel beam of monoenergetic photons incident on a detector of uniform thickness: − µx Intrinsic efficiency = 1 - e
  • 16. Gas-filled detectors • A gas-filled detector consists of a volume of gas between two electrodes, with an electrical potential difference (voltage) applied between the electrodes • Ionizing radiation produces ion pairs in the gas • Positive ions (cations) attracted to negative electrode (cathode); electrons or anions attracted to positive electrode (anode) • In most detectors, cathode is the wall of the container that holds the gas and anode is a wire inside the container
  • 17.
  • 18. Types of gas-filled detectors • Three types of gas-filled detectors in common use: – Ionization chambers – Proportional counters – Geiger-Mueller (GM) counters • Type determined primarily by the voltage applied between the two electrodes • Ionization chambers have wider range of physical shape (parallel plates, concentric cylinders, etc.) • Proportional counters and GM counters must have thin wire anode
  • 19.
  • 20. Ionization chambers • If gas is air and walls of chamber are of a material whose effective atomic number is similar to air, the amount of current produced is proportional to the exposure rate • Air-filled ion chambers are used in portable survey meters, for performing QA testing of diagnostic and therapeutic x-ray machines, and are the detectors in most x-ray machine phototimers • Low intrinsic efficiencies because of low densities of gases and low atomic numbers of most gases
  • 21.
  • 22. Proportional counters • Must contain a gas with specific properties • Commonly used in standards laboratories, health physics laboratories, and for physics research • Seldom used in medical centers
  • 23. GM counters • GM counters also must contain gases with specific properties • Gas amplification produces billions of ion pairs after an interaction – signal from detector requires little amplification • Often used for inexpensive survey meters • In general, GM survey meters are inefficient detectors of x-rays and gamma rays • Over-response to low energy x-rays – partially corrected by placing a thin layer of higher atomic number material around the detector
  • 24.
  • 25.
  • 26. GM counters (cont.) • GM detectors suffer from extremely long dead times – seldom used when accurate measurements are required of count rates greater than a few hundred counts per second • Portable GM survey meter may become paralyzed in a very high radiation field – should always use ionization chamber instruments for measuring such fields
  • 27. Scintillation detectors • Scintillators are used in conventional film-screen radiography, many digital radiographic receptors, fluoroscopy, scintillation cameras, most CT scanners, and PET scanners • Scintillation detectors consist of a scintillator and a device, such as a PMT, that converts the light into an electrical signal
  • 28. Scintillators • Desirable properties: – High conversion efficiency – Decay times of excited states should be short – Material transparent to its own emissions – Color of emitted light should match spectral sensitivity of the light receptor – For x-ray and gamma-ray detectors, µ should be large – high detection efficiencies – Rugged, unaffected by moisture, and inexpensive to manufacture
  • 29. Scintillators (cont.) • Amount of light emitted after an interaction increases with energy deposited by the interaction • May be operated in pulse mode as spectrometers • High conversion efficiency produces superior energy resolution
  • 30. Materials • Sodium iodide activated with thallium [NaI(Tl)], coupled to PMTs and operated in pulse mode, is used for most nuclear medicine applications – Fragile and hygroscopic • Bismuth germanate (BGO) is coupled to PMTs and used in pulse mode as detectors in most PET scanners
  • 31. Photomultiplier tubes • PMTs perform two functions: – Conversion of ultraviolet and visible light photons into an electrical signal – Signal amplification, on the order of millions to billions • Consists of an evacuated glass tube containing a photocathode, typically 10 to 12 electrodes called dynodes, and an anode
  • 32.
  • 33.
  • 34. Dynodes • Electrons emitted by the photocathode are attracted to the first dynode and are accelerated to kinetic energies equal to the potential difference between the photocathode and the first dynode • When these electrons strike the first dynode, about 5 electrons are ejected from the dynode for each electron hitting it • These electrons are attracted to the second dynode, and so on, finally reaching the anode
  • 35. PMT amplification • Total amplification of the PMT is the product of the individual amplifications at each dynode • If a PMT has ten dynodes and the amplification at each stage is 5, the total amplification will be approximately 10,000,000 • Amplification can be adjusted by changing the voltage applied to the PMT