SlideShare a Scribd company logo
1 of 75
VISUAL FIELD EXAMINATION
AND
INTERPRETATION OF AUTOMATED
PERIMETRY
PRESENTER: Dr DIVYA KESARWANI
MODERATOR: Dr VAIBHAV JAIN
DEPARTMENT OF OPHTHALMOLOGY, SGPGIMS, LUCKNOW
29/04/2019
EXAMINATION METHODOLOGY
KINETIC
 Test object of particular size and intensity is passed from
non seeing area to seeing area along a particular meridian
at the rate of 3 – 5 deg per sec
 Repeated every 15 – 30 deg
STATIC
 The location, size and duration of stimulus is kept constant and the
luminance is gradually increased until seen
 Actual estimation of sensitivity ( THRESHOLD ) of each point is
made out
 SUPRA THRESHOLD stimulus used for screening
-------------------------------------------------------------------------------
IMPORTANT :
 One eye is tested at a time, other is occluded.
 Fixation of the patient has to be steady and is monitored
throughout the test
---------------------------------------------------------------------------------
EQUIPMENTS
PERIMETRY VS CAMPIMETRY
o Perimetry – measurement of visual field by projecting
targets onto a curved surface.
o Campimetry – measurement of visual field by
projecting targets onto a flat surface.
BJERRUM’S SCREEN ( CAMPIMETRY)
 Patient sits at 1 or 2 m from flat screen
 Kinetic and static
 For central 30 deg only
 Done under subdued lighting
GOLDMANN’S PERIMETER
 Bowl type
 Standardization
 Both kinetic and static
 Peripheral as well as central
AUTOMATED PERIMETRY
HUMPHREY FIELD ANALYZER
OCTOPUS
 STATIC perimetry
 Measurement of threshold values
 STATPAC (HFA)- Comparison to normative data
 Inbuilt program for analysis – diagnosis and progression
AUTOMATED PERIMETRY
 Automated threshold static perimetry quantifies the sensitivity
of a patient’s central vision using efficient and standardized
testing algorithms.
PROGRAMS / PATTERNS
o 30-2 – gold standard
o 24-2
o 10-2
o MACULAR
o Nasal step program – additional 12 locations upto 50 deg nasal
peripheral 60 and 60-4 prog
o Estermann test – for binocular 120 deg field
MACULA PROGRAM :16 locations
within the central 5° with 2° spacing.
Each location is tested three times
TERMINOLOGIES
 Isopter
 Line on a visual field representation connecting points with
same visible threshold.
 Scotoma
 It is an area of reduced(‘relative’) or total(‘absolute’)loss of vision that
is surrounded by a field of normal or a relatively well-preserved
vision.
 Blind spot
 Physiological scotoma corresponding to optic nerve head
 b/w 10-20 ˚ .It is located temporally and slightly inferiorly.
 Threshold :
 Physiological capacity to detect a stimulus at a given location, that
stimulus intensity that has a 50%probability of being seen.
SWEDISH INTERACTIVE TESTING ALGORITHM (SITA)
1. SITA STANDARD ( Bracketing strategy based)
2. SITA FAST ( FASTPAC based)
o Analyzes patients response and responds accordingly
o Decreases overall no of stimuli presented, hence test duration
o Doesn’t estimate Short term Fluctuations
THRESHOLD
BRACKETING STRATEGY( staircase)
– GOLD STANDARD
–It is useful to measure the foveal threshold at the very beginning of the
test .
 If the patient is not properly focused on the bowl ,foveal sensitivity will
be reduced along with the remainder of the field.
 Visual Acuity-BCVA should be 6/36 or better for the visual field to be
tested.
 Apostilbs:
 Absolute units of light intensity.
 Luminance of test target.
 1 asb=.3183 candela/m2
 Decibels:
 Relative scale, created by manufacturers.
 Attenuation of light by neutral density filters.
 1dB=1/10 log units of attenuation of max. stimulus.
 Measures sensitivity at each point.0to40dB.
 Background intensity = 31.6 asb
 Higher intensity stimulus has higher asb value, this is
inverse of retinal sensitivity.
 Stimulus  Maximum intensity = 0 dB
Minimum intensity = 40 dB
 Selection of adequate test
 Proper environment
 Comfortable sitting position
 Adequate size of pupil >3mm
 Adequate Near correction
 Proper explanation – running of demonstration
 Reassurance – not all points will be seen
- test can be paused by keeping the response button
pressed
INTERPRETATION OF
VISUAL FIELD PRINTOUT.
Patient data
 Name, DOB, eye
 Vision, refraction,
 Pupil diameter
Test data
 Date and time
 Program and strategy
 Background illumination
 Test size, color, duration,
interval
ZONE 1 : REPRODUCIBILITY
REFRACTIVE ERROR-
o Refractive blur reduces visual sensitivity to perimetric stimuli.
o One dioptre of refractive blur in undilated patient will produce a
little more than 1dB of depression of field of vision when testing
with Goldmann Size III stimulus.
o Hence near vision must be properly corrected , otherwise the visual
field will show a generalized depression .
Constricted field d/t miosis
Pupil Diameter: 1.0mm
Repeat Perimetry with
Pupil Diameter: 2.5 mm
ZONE 2 : RELIABILITY
 Fixation monitor
 Fixation target
 Reliability indices
Fixation losses < 20%
Gaze tracking
False positives < 15%
(trigger happy)
False negatives < 15%
Foveal threshold
 Fixation monitor may be chosen as blind spot and/or gaze.
 Fixation target may be central or large diamond, the latter being
chosen in patients with central visual loss.
 Fixation losses: The fixation can be tested in different machines using
blind spot monitoring with Heijl Krakau method, gaze monitoring
with gaze tracker or manually by observation of eye monitor.
 When the fixation monitoring test parameter is set to blind spot
(Heijl-Krakau) mode, proper fixation is checked by projecting 5% of
stimuli at the presumed location of the physiological blind spot.
 If fixation losses exceed 20%, “XX” will be printed after the score.
•False positive response-
- FP >15% is strongly
associated with
compromised test.
-With a high FP score grey
scale appears abnormally pale.
High false positive responses :
 Seen in patients who are anxious about not seeing all test points and
might be trigger happy.
 Such a field would show in addition to false positives, numeric
sensitivities above 40 dB, white areas on grayscale, a pattern
deviation plot worse than total deviation, positive mean deviation
and abnormally high sensitivity on GHT.
 False negative response-
- In general >15% of FN is
considered abnormal
-Grey scale pattern in high FN
cases tends to have a clover-
leaf pattern.
High values occur if:
 The patient is fatigued and falling asleep
 Has changed personal criteria for response,
 There is presence of true visual field loss where sensitivities are
variable.
 May be normally present in cases with advanced defect as the
visibility may be highly variable in glaucomatous visual fields
While false negative errors and fixation loss scores may not make a
field invalid, high false positive errors may warrant rejection of the
test results.
RAW DATA AND GRAYSCALE
 Simple threshold sensitivities are measured in decibels and given in
the numeric plot.
 Values above 40 dB are generally not expected and may be indicative
of a trigger happy field.
 The actual numeric values may not be concentrated upon for day to
day interpretation of field but only with specific purposes.
 Grayscale is an artists interpolation of the raw data and we do not base
our diagnosis on it.
 The raw data is the exact retinal sensitivity in dB units of the selected
points calculated by field analyser.
 In raw data 0 dB = absolute scotoma.
 40 dB is the highest retinal sensitivity.
 In same patient the raw data calculated by different strategies is not
exactly similar .
ZONE 4 : GREY SCALE
 Based on actual threshold values at each location
 General identification
 Patient information
 It can though typically highlight certain artifactual fields
 Lid artifact –showing defect in superior part due to a droopy eyelid
 Rim artifact- showing concentric constriction due to rim of trial
lens which is either decentered or placed away from the eye
 Clover leaf-typical pattern of initial response around four primary
points in each quadrant followed by depressed or no response as seen
in a new and poorly instructed patient.
 Trigger happy fields-patches of white tones in grayscale seen in
overanxious patient with high false positive responses.
ZONE 5&6 :TOTAL DEVIATION PLOT
 Numerical plot – indicates by how much decibels
is each point depressed compared to mean value
in normal population of similar age.
 Probability plot- grey scale indicates the
probability of occurrence of the deviation in
normal population.
Generalized depression due to media opacities,
refractive error, miosis may hamper appearance
of a pattern
 P<5% indicates the retinal sensitivity of that
point is seen in < 5% of normal population.
 Darker the symbol, the greater the probability
of abnormality as indicated by P value.
 Higher the P value lesser the chances of field
being abnormal
ZONE 7&8 : PATTERN DEVIATION
PLOT
 Numerical - calculated by adjustment for
generalized depression or elevation of visual
field
COMPARISON OF TOTAL AND PATTERN
DEVIATION PLOTS
 Uniformly depressed total deviation plot and normal looking pattern
deviation plot: cataract or any generalized depression of fi eld
 Both plots look more or less the same: little or no generalized loss
 Normal total deviation and abnormal looking pattern deviation:
trigger happy patient
ZONE-9 GLOBAL INDICES
 The global indices summarise the field.
 The indices available in SITA strategies are:
 Mean Deviation: how much the total field departs from normal
 Pattern Standard Deviation: irregularities in the field.
 Visual Field Index: visual field status as a percent of a normal field.
 Short term fluctuation (SF)
 Corrected pattern standard deviation(CPSD)
• MD gives an indication of the
overall sensitivity of the field.
• The positive value  better
overall retinal sensitivity than
normal observer.
• Negative value  worse than
normal.
 Negative value would be increased in presence of:
 Generalized depression of entire field
 Multiple localized depressions
 Deep localized depression
 Positive mean deviation is seen in:
 Hypersensitive or trigger happy patient
 Normal patients having sensitivity values higher than those in normative
database.
• Measure of focal loss or variability
within the field taking into account any
generalized depression in the hill of
vision.
 More specific indicator of
glaucomatous damage than MD.
 Measures irregularity by summing the absolute value of the difference between
the threshold value for each point and the average visual field sensitivity at
each point (equal to the normal value for each point + the MD).
 Low PSD indicates:
 Smooth hill of vision
 Near normal field
 Severely depressed field
 High PSD indicates:
 Variability in field response
 Irregular field due to localized deep defects
KEY NOTE POINTS
 The diagnosis of glaucoma should not be based on global indices.
 Probability plots and GHT are more useful for diagnosis while indices may
help in staging disease and follow up.
 So in follow up of a patient if a worsening MD is accompanied by a stable
PSD it could be due to increasing generalized loss like in cataract.
 On the other hand a nearly stable or slightly lowering MD accompanied by
a worsening PSD could be indicative of worsening localized loss as in
glaucoma
VISUAL FIELD INDEX
 New index which expresses the visual field status as a percent of a
normal age-adjusted visual field.
 This PSD based index was devised mainly to detect progression of
glaucomatous defect.
 Greater weight is given to points closer to fi xation to adjust for
gangion cell density and visual function.
 The index may also be less sensitive to cataract and media changes.
ZONE 10 : GLAUCOMA HEMIFIELD TEST
 Comparison of 5 clusters of points in superior
hemifield with mirror images in inferior hemifield
POSSIBLE OUTCOMES ARE:
 Outside normal limits. The GHT is described as “outside normal limits” when
matched pairs of zones differ by an amount greater than that seen in 99%
normal subjects, or when both members of a pair of zones are more abnormal
than 99.5% of the individuals with the normative population.
 Borderline. When matched pairs of zones differ by an amount greater than
that seen in 97% normal subjects but not reaching the level of outside normal
limits
 General reduction of sensitivity. When the best region of the VF is depressed
to a level seen in only 0.5% of the normative database but both conditions for
“outside normal limits” are not met
 Abnormally high sensitivity. When the best region of the VF is elevated to a
level seen in only 0.5% of the normative database.
 Within normal limits. When none of the above conditions are met.
ARTEFACTS
 OBSTRUCTION
RIM ARTEFACTS
PTOSIS
MEDIA OPACITIES
 MIOSIS
 REFRACTION ARTEFACTS
 High power plus and minus lenses
ANDERSON & PATELLA CRITERIA
 3 or more congrous ‘non edge points’ in typical arcuate area on 30-2
program
depressed @ p< 5 % with at least one point @ p<1 %
PSD / CPSD @ p< 5%
GHT – outside normal limits
 Must be demonstrated on 2 field tests
Nasal step
scotoma
ARCUATE
SCOTOMA
SEVERE DEFECT
RECOGNITION OF PROGRESSIVE DAMAGE
 Compare baseline field to each individual field
 Study the entire series.
OVERVIEW
READING A SAMPLE PRINTOUT
TAKE HOME MESSAGE
 The field defect must be reproducible as is confirmed by repeat
field examination of the same patient
 Clinical correlation is of prime importance .
 Interpreting a single analysis may not be difficult if we go in a
stepwise manner.
 Important thing to remember is that presence of field defect does
not always mean presence of disease.
THANK YOU

More Related Content

What's hot

Binocular anomalies What we should know?
Binocular anomalies What we should know?Binocular anomalies What we should know?
Binocular anomalies What we should know?Anis Suzanna Mohamad
 
Hess chart and it's Interpretation
Hess chart and it's InterpretationHess chart and it's Interpretation
Hess chart and it's InterpretationDrAzmat Ali
 
Goldman Applanation Tonometer
Goldman Applanation TonometerGoldman Applanation Tonometer
Goldman Applanation TonometerAnisur Rahman
 
Superior oblique palsy
Superior oblique palsySuperior oblique palsy
Superior oblique palsysiraj safi
 
A scan anu101 updated
A scan anu101 updatedA scan anu101 updated
A scan anu101 updatedAnurag Shukla
 
Slit lamp biomicroscopy and illumination techniques
Slit lamp biomicroscopy and illumination techniquesSlit lamp biomicroscopy and illumination techniques
Slit lamp biomicroscopy and illumination techniquesLoknath Goswami
 
Automated perimetry
Automated perimetryAutomated perimetry
Automated perimetryarmaan ahmed
 
Binocular refraction techniques, binocular balancing &amp; binocular
Binocular refraction techniques, binocular balancing &amp; binocularBinocular refraction techniques, binocular balancing &amp; binocular
Binocular refraction techniques, binocular balancing &amp; binocularsabina paudel
 
Glaucoma drainage devices
Glaucoma drainage devicesGlaucoma drainage devices
Glaucoma drainage devicesvaishusmail
 
SOFT TORIC CONTACT LENS FITTING.pptx
SOFT TORIC CONTACT LENS FITTING.pptxSOFT TORIC CONTACT LENS FITTING.pptx
SOFT TORIC CONTACT LENS FITTING.pptxBipin Koirala
 
Ophthalmic Prisms: Prismatic Effects and Decentration
Ophthalmic Prisms: Prismatic Effects and DecentrationOphthalmic Prisms: Prismatic Effects and Decentration
Ophthalmic Prisms: Prismatic Effects and DecentrationRabindraAdhikary
 
Contrast sensitivity
Contrast sensitivityContrast sensitivity
Contrast sensitivitySuman Sen
 
Prosthetic Contact Lens (Grand round)
Prosthetic Contact Lens (Grand round)Prosthetic Contact Lens (Grand round)
Prosthetic Contact Lens (Grand round)Suraj Chaurasiya
 

What's hot (20)

Pentacam
Pentacam Pentacam
Pentacam
 
Binocular anomalies What we should know?
Binocular anomalies What we should know?Binocular anomalies What we should know?
Binocular anomalies What we should know?
 
Hess chart and it's Interpretation
Hess chart and it's InterpretationHess chart and it's Interpretation
Hess chart and it's Interpretation
 
Goldman Applanation Tonometer
Goldman Applanation TonometerGoldman Applanation Tonometer
Goldman Applanation Tonometer
 
SYNAPTOPHORE.pptx
SYNAPTOPHORE.pptxSYNAPTOPHORE.pptx
SYNAPTOPHORE.pptx
 
Gonioscopy
GonioscopyGonioscopy
Gonioscopy
 
Bruckner test
Bruckner testBruckner test
Bruckner test
 
Superior oblique palsy
Superior oblique palsySuperior oblique palsy
Superior oblique palsy
 
A scan anu101 updated
A scan anu101 updatedA scan anu101 updated
A scan anu101 updated
 
Slit lamp biomicroscopy and illumination techniques
Slit lamp biomicroscopy and illumination techniquesSlit lamp biomicroscopy and illumination techniques
Slit lamp biomicroscopy and illumination techniques
 
Automated perimetry
Automated perimetryAutomated perimetry
Automated perimetry
 
Binocular refraction techniques, binocular balancing &amp; binocular
Binocular refraction techniques, binocular balancing &amp; binocularBinocular refraction techniques, binocular balancing &amp; binocular
Binocular refraction techniques, binocular balancing &amp; binocular
 
Glaucoma drainage devices
Glaucoma drainage devicesGlaucoma drainage devices
Glaucoma drainage devices
 
SOFT TORIC CONTACT LENS FITTING.pptx
SOFT TORIC CONTACT LENS FITTING.pptxSOFT TORIC CONTACT LENS FITTING.pptx
SOFT TORIC CONTACT LENS FITTING.pptx
 
Ophthalmic Prisms: Prismatic Effects and Decentration
Ophthalmic Prisms: Prismatic Effects and DecentrationOphthalmic Prisms: Prismatic Effects and Decentration
Ophthalmic Prisms: Prismatic Effects and Decentration
 
Contrast sensitivity
Contrast sensitivityContrast sensitivity
Contrast sensitivity
 
Freshnel prism final
Freshnel prism finalFreshnel prism final
Freshnel prism final
 
Accommodation
Accommodation Accommodation
Accommodation
 
Worth 4 dot test
Worth 4 dot testWorth 4 dot test
Worth 4 dot test
 
Prosthetic Contact Lens (Grand round)
Prosthetic Contact Lens (Grand round)Prosthetic Contact Lens (Grand round)
Prosthetic Contact Lens (Grand round)
 

Similar to VISUAL FIELD EXAMINATION AND INTERPRETATION OF AUTOMATED PERIMETRY

Perimetry by dr.ricky
Perimetry by dr.rickyPerimetry by dr.ricky
Perimetry by dr.rickyRicky Mittal
 
Visual Field presentation.nagla.ppt
Visual Field presentation.nagla.pptVisual Field presentation.nagla.ppt
Visual Field presentation.nagla.pptnanoAly
 
Humphreys visual field analysis powerpoint.pptx
Humphreys visual field analysis powerpoint.pptxHumphreys visual field analysis powerpoint.pptx
Humphreys visual field analysis powerpoint.pptxShineNS2
 
Peripheral visual field
Peripheral visual fieldPeripheral visual field
Peripheral visual fieldNishan Faiyaz
 
Humphrey visual field analyser (HVFA)
Humphrey visual field analyser (HVFA)Humphrey visual field analyser (HVFA)
Humphrey visual field analyser (HVFA)Azizul Islam
 
perimetry-130707155722-phpapp01.pdf
perimetry-130707155722-phpapp01.pdfperimetry-130707155722-phpapp01.pdf
perimetry-130707155722-phpapp01.pdfVidhiMadrecha
 
Visual Field Examination
Visual Field ExaminationVisual Field Examination
Visual Field ExaminationPaavan Kalra
 
Perimetry & Humprey visual field assay (HVFA)
Perimetry & Humprey visual field assay (HVFA)Perimetry & Humprey visual field assay (HVFA)
Perimetry & Humprey visual field assay (HVFA)Rashed-Ul-Hasan Rasu
 
Diagnostic procedures
Diagnostic proceduresDiagnostic procedures
Diagnostic proceduresMeghna Verma
 
Visual field assessment,optic nerve changes and retinal changes
Visual field assessment,optic nerve changes and retinal changesVisual field assessment,optic nerve changes and retinal changes
Visual field assessment,optic nerve changes and retinal changesBipin Bista
 
Automated perimetry
Automated  perimetryAutomated  perimetry
Automated perimetryKishor Badhe
 
PPT on BASIC CONCEPTION ON HUMPHERY AUTOMATED PERIMETRY
PPT on BASIC CONCEPTION ON HUMPHERY AUTOMATED PERIMETRY PPT on BASIC CONCEPTION ON HUMPHERY AUTOMATED PERIMETRY
PPT on BASIC CONCEPTION ON HUMPHERY AUTOMATED PERIMETRY Nalin Nayan
 

Similar to VISUAL FIELD EXAMINATION AND INTERPRETATION OF AUTOMATED PERIMETRY (20)

Visual field Analysis .ppt
Visual field Analysis .pptVisual field Analysis .ppt
Visual field Analysis .ppt
 
Perimetry by dr.ricky
Perimetry by dr.rickyPerimetry by dr.ricky
Perimetry by dr.ricky
 
Visual Field presentation.nagla.ppt
Visual Field presentation.nagla.pptVisual Field presentation.nagla.ppt
Visual Field presentation.nagla.ppt
 
Visual Field | Humphrey Perimetry
Visual Field | Humphrey PerimetryVisual Field | Humphrey Perimetry
Visual Field | Humphrey Perimetry
 
Automated perimetry
Automated perimetryAutomated perimetry
Automated perimetry
 
Humphreys visual field analysis powerpoint.pptx
Humphreys visual field analysis powerpoint.pptxHumphreys visual field analysis powerpoint.pptx
Humphreys visual field analysis powerpoint.pptx
 
Peripheral visual field
Peripheral visual fieldPeripheral visual field
Peripheral visual field
 
Humphrey visual field analyser (HVFA)
Humphrey visual field analyser (HVFA)Humphrey visual field analyser (HVFA)
Humphrey visual field analyser (HVFA)
 
Visual field examination
Visual field examinationVisual field examination
Visual field examination
 
perimetry-130707155722-phpapp01.pdf
perimetry-130707155722-phpapp01.pdfperimetry-130707155722-phpapp01.pdf
perimetry-130707155722-phpapp01.pdf
 
Visual Field Examination
Visual Field ExaminationVisual Field Examination
Visual Field Examination
 
Perimetry & Humprey visual field assay (HVFA)
Perimetry & Humprey visual field assay (HVFA)Perimetry & Humprey visual field assay (HVFA)
Perimetry & Humprey visual field assay (HVFA)
 
Diagnostic procedures
Diagnostic proceduresDiagnostic procedures
Diagnostic procedures
 
Visual field assessment,optic nerve changes and retinal changes
Visual field assessment,optic nerve changes and retinal changesVisual field assessment,optic nerve changes and retinal changes
Visual field assessment,optic nerve changes and retinal changes
 
Hvfa humphrey visual field
Hvfa humphrey visual fieldHvfa humphrey visual field
Hvfa humphrey visual field
 
Automated perimetry
Automated  perimetryAutomated  perimetry
Automated perimetry
 
Hfa
HfaHfa
Hfa
 
Refraction
RefractionRefraction
Refraction
 
Automated perimetry
Automated perimetryAutomated perimetry
Automated perimetry
 
PPT on BASIC CONCEPTION ON HUMPHERY AUTOMATED PERIMETRY
PPT on BASIC CONCEPTION ON HUMPHERY AUTOMATED PERIMETRY PPT on BASIC CONCEPTION ON HUMPHERY AUTOMATED PERIMETRY
PPT on BASIC CONCEPTION ON HUMPHERY AUTOMATED PERIMETRY
 

Recently uploaded

TOPIC 8 Temperature and Heat.pdf physics
TOPIC 8 Temperature and Heat.pdf physicsTOPIC 8 Temperature and Heat.pdf physics
TOPIC 8 Temperature and Heat.pdf physicsssuserddc89b
 
Call Girls in Majnu Ka Tilla Delhi 🔝9711014705🔝 Genuine
Call Girls in Majnu Ka Tilla Delhi 🔝9711014705🔝 GenuineCall Girls in Majnu Ka Tilla Delhi 🔝9711014705🔝 Genuine
Call Girls in Majnu Ka Tilla Delhi 🔝9711014705🔝 Genuinethapagita
 
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptx
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptxRESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptx
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptxFarihaAbdulRasheed
 
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)riyaescorts54
 
STOPPED FLOW METHOD & APPLICATION MURUGAVENI B.pptx
STOPPED FLOW METHOD & APPLICATION MURUGAVENI B.pptxSTOPPED FLOW METHOD & APPLICATION MURUGAVENI B.pptx
STOPPED FLOW METHOD & APPLICATION MURUGAVENI B.pptxMurugaveni B
 
Pests of castor_Binomics_Identification_Dr.UPR.pdf
Pests of castor_Binomics_Identification_Dr.UPR.pdfPests of castor_Binomics_Identification_Dr.UPR.pdf
Pests of castor_Binomics_Identification_Dr.UPR.pdfPirithiRaju
 
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)User Guide: Pulsar™ Weather Station (Columbia Weather Systems)
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)Columbia Weather Systems
 
Speech, hearing, noise, intelligibility.pptx
Speech, hearing, noise, intelligibility.pptxSpeech, hearing, noise, intelligibility.pptx
Speech, hearing, noise, intelligibility.pptxpriyankatabhane
 
Microteaching on terms used in filtration .Pharmaceutical Engineering
Microteaching on terms used in filtration .Pharmaceutical EngineeringMicroteaching on terms used in filtration .Pharmaceutical Engineering
Microteaching on terms used in filtration .Pharmaceutical EngineeringPrajakta Shinde
 
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdfPests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdfPirithiRaju
 
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.PraveenaKalaiselvan1
 
Environmental Biotechnology Topic:- Microbial Biosensor
Environmental Biotechnology Topic:- Microbial BiosensorEnvironmental Biotechnology Topic:- Microbial Biosensor
Environmental Biotechnology Topic:- Microbial Biosensorsonawaneprad
 
Pests of soyabean_Binomics_IdentificationDr.UPR.pdf
Pests of soyabean_Binomics_IdentificationDr.UPR.pdfPests of soyabean_Binomics_IdentificationDr.UPR.pdf
Pests of soyabean_Binomics_IdentificationDr.UPR.pdfPirithiRaju
 
User Guide: Orion™ Weather Station (Columbia Weather Systems)
User Guide: Orion™ Weather Station (Columbia Weather Systems)User Guide: Orion™ Weather Station (Columbia Weather Systems)
User Guide: Orion™ Weather Station (Columbia Weather Systems)Columbia Weather Systems
 
Grafana in space: Monitoring Japan's SLIM moon lander in real time
Grafana in space: Monitoring Japan's SLIM moon lander  in real timeGrafana in space: Monitoring Japan's SLIM moon lander  in real time
Grafana in space: Monitoring Japan's SLIM moon lander in real timeSatoshi NAKAHIRA
 
User Guide: Capricorn FLX™ Weather Station
User Guide: Capricorn FLX™ Weather StationUser Guide: Capricorn FLX™ Weather Station
User Guide: Capricorn FLX™ Weather StationColumbia Weather Systems
 
Pests of jatropha_Bionomics_identification_Dr.UPR.pdf
Pests of jatropha_Bionomics_identification_Dr.UPR.pdfPests of jatropha_Bionomics_identification_Dr.UPR.pdf
Pests of jatropha_Bionomics_identification_Dr.UPR.pdfPirithiRaju
 
Harmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms PresentationHarmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms Presentationtahreemzahra82
 
Solution chemistry, Moral and Normal solutions
Solution chemistry, Moral and Normal solutionsSolution chemistry, Moral and Normal solutions
Solution chemistry, Moral and Normal solutionsHajira Mahmood
 

Recently uploaded (20)

TOPIC 8 Temperature and Heat.pdf physics
TOPIC 8 Temperature and Heat.pdf physicsTOPIC 8 Temperature and Heat.pdf physics
TOPIC 8 Temperature and Heat.pdf physics
 
Call Girls in Majnu Ka Tilla Delhi 🔝9711014705🔝 Genuine
Call Girls in Majnu Ka Tilla Delhi 🔝9711014705🔝 GenuineCall Girls in Majnu Ka Tilla Delhi 🔝9711014705🔝 Genuine
Call Girls in Majnu Ka Tilla Delhi 🔝9711014705🔝 Genuine
 
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptx
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptxRESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptx
RESPIRATORY ADAPTATIONS TO HYPOXIA IN HUMNAS.pptx
 
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)
(9818099198) Call Girls In Noida Sector 14 (NOIDA ESCORTS)
 
STOPPED FLOW METHOD & APPLICATION MURUGAVENI B.pptx
STOPPED FLOW METHOD & APPLICATION MURUGAVENI B.pptxSTOPPED FLOW METHOD & APPLICATION MURUGAVENI B.pptx
STOPPED FLOW METHOD & APPLICATION MURUGAVENI B.pptx
 
Pests of castor_Binomics_Identification_Dr.UPR.pdf
Pests of castor_Binomics_Identification_Dr.UPR.pdfPests of castor_Binomics_Identification_Dr.UPR.pdf
Pests of castor_Binomics_Identification_Dr.UPR.pdf
 
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)User Guide: Pulsar™ Weather Station (Columbia Weather Systems)
User Guide: Pulsar™ Weather Station (Columbia Weather Systems)
 
Speech, hearing, noise, intelligibility.pptx
Speech, hearing, noise, intelligibility.pptxSpeech, hearing, noise, intelligibility.pptx
Speech, hearing, noise, intelligibility.pptx
 
Microteaching on terms used in filtration .Pharmaceutical Engineering
Microteaching on terms used in filtration .Pharmaceutical EngineeringMicroteaching on terms used in filtration .Pharmaceutical Engineering
Microteaching on terms used in filtration .Pharmaceutical Engineering
 
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdfPests of Blackgram, greengram, cowpea_Dr.UPR.pdf
Pests of Blackgram, greengram, cowpea_Dr.UPR.pdf
 
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
BIOETHICS IN RECOMBINANT DNA TECHNOLOGY.
 
Environmental Biotechnology Topic:- Microbial Biosensor
Environmental Biotechnology Topic:- Microbial BiosensorEnvironmental Biotechnology Topic:- Microbial Biosensor
Environmental Biotechnology Topic:- Microbial Biosensor
 
Pests of soyabean_Binomics_IdentificationDr.UPR.pdf
Pests of soyabean_Binomics_IdentificationDr.UPR.pdfPests of soyabean_Binomics_IdentificationDr.UPR.pdf
Pests of soyabean_Binomics_IdentificationDr.UPR.pdf
 
User Guide: Orion™ Weather Station (Columbia Weather Systems)
User Guide: Orion™ Weather Station (Columbia Weather Systems)User Guide: Orion™ Weather Station (Columbia Weather Systems)
User Guide: Orion™ Weather Station (Columbia Weather Systems)
 
Grafana in space: Monitoring Japan's SLIM moon lander in real time
Grafana in space: Monitoring Japan's SLIM moon lander  in real timeGrafana in space: Monitoring Japan's SLIM moon lander  in real time
Grafana in space: Monitoring Japan's SLIM moon lander in real time
 
User Guide: Capricorn FLX™ Weather Station
User Guide: Capricorn FLX™ Weather StationUser Guide: Capricorn FLX™ Weather Station
User Guide: Capricorn FLX™ Weather Station
 
Volatile Oils Pharmacognosy And Phytochemistry -I
Volatile Oils Pharmacognosy And Phytochemistry -IVolatile Oils Pharmacognosy And Phytochemistry -I
Volatile Oils Pharmacognosy And Phytochemistry -I
 
Pests of jatropha_Bionomics_identification_Dr.UPR.pdf
Pests of jatropha_Bionomics_identification_Dr.UPR.pdfPests of jatropha_Bionomics_identification_Dr.UPR.pdf
Pests of jatropha_Bionomics_identification_Dr.UPR.pdf
 
Harmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms PresentationHarmful and Useful Microorganisms Presentation
Harmful and Useful Microorganisms Presentation
 
Solution chemistry, Moral and Normal solutions
Solution chemistry, Moral and Normal solutionsSolution chemistry, Moral and Normal solutions
Solution chemistry, Moral and Normal solutions
 

VISUAL FIELD EXAMINATION AND INTERPRETATION OF AUTOMATED PERIMETRY

  • 1. VISUAL FIELD EXAMINATION AND INTERPRETATION OF AUTOMATED PERIMETRY PRESENTER: Dr DIVYA KESARWANI MODERATOR: Dr VAIBHAV JAIN DEPARTMENT OF OPHTHALMOLOGY, SGPGIMS, LUCKNOW 29/04/2019
  • 3. KINETIC  Test object of particular size and intensity is passed from non seeing area to seeing area along a particular meridian at the rate of 3 – 5 deg per sec  Repeated every 15 – 30 deg
  • 4. STATIC  The location, size and duration of stimulus is kept constant and the luminance is gradually increased until seen  Actual estimation of sensitivity ( THRESHOLD ) of each point is made out  SUPRA THRESHOLD stimulus used for screening ------------------------------------------------------------------------------- IMPORTANT :  One eye is tested at a time, other is occluded.  Fixation of the patient has to be steady and is monitored throughout the test ---------------------------------------------------------------------------------
  • 6. PERIMETRY VS CAMPIMETRY o Perimetry – measurement of visual field by projecting targets onto a curved surface. o Campimetry – measurement of visual field by projecting targets onto a flat surface.
  • 7. BJERRUM’S SCREEN ( CAMPIMETRY)  Patient sits at 1 or 2 m from flat screen  Kinetic and static  For central 30 deg only  Done under subdued lighting
  • 8. GOLDMANN’S PERIMETER  Bowl type  Standardization  Both kinetic and static  Peripheral as well as central
  • 9. AUTOMATED PERIMETRY HUMPHREY FIELD ANALYZER OCTOPUS  STATIC perimetry  Measurement of threshold values  STATPAC (HFA)- Comparison to normative data  Inbuilt program for analysis – diagnosis and progression
  • 10.
  • 11. AUTOMATED PERIMETRY  Automated threshold static perimetry quantifies the sensitivity of a patient’s central vision using efficient and standardized testing algorithms.
  • 12.
  • 13. PROGRAMS / PATTERNS o 30-2 – gold standard o 24-2 o 10-2 o MACULAR o Nasal step program – additional 12 locations upto 50 deg nasal peripheral 60 and 60-4 prog o Estermann test – for binocular 120 deg field
  • 14. MACULA PROGRAM :16 locations within the central 5° with 2° spacing. Each location is tested three times
  • 15. TERMINOLOGIES  Isopter  Line on a visual field representation connecting points with same visible threshold.
  • 16.  Scotoma  It is an area of reduced(‘relative’) or total(‘absolute’)loss of vision that is surrounded by a field of normal or a relatively well-preserved vision.  Blind spot  Physiological scotoma corresponding to optic nerve head  b/w 10-20 ˚ .It is located temporally and slightly inferiorly.  Threshold :  Physiological capacity to detect a stimulus at a given location, that stimulus intensity that has a 50%probability of being seen.
  • 17.
  • 18. SWEDISH INTERACTIVE TESTING ALGORITHM (SITA) 1. SITA STANDARD ( Bracketing strategy based) 2. SITA FAST ( FASTPAC based) o Analyzes patients response and responds accordingly o Decreases overall no of stimuli presented, hence test duration o Doesn’t estimate Short term Fluctuations
  • 20. –It is useful to measure the foveal threshold at the very beginning of the test .  If the patient is not properly focused on the bowl ,foveal sensitivity will be reduced along with the remainder of the field.  Visual Acuity-BCVA should be 6/36 or better for the visual field to be tested.
  • 21.  Apostilbs:  Absolute units of light intensity.  Luminance of test target.  1 asb=.3183 candela/m2  Decibels:  Relative scale, created by manufacturers.  Attenuation of light by neutral density filters.  1dB=1/10 log units of attenuation of max. stimulus.  Measures sensitivity at each point.0to40dB.
  • 22.
  • 23.
  • 24.
  • 25.  Background intensity = 31.6 asb  Higher intensity stimulus has higher asb value, this is inverse of retinal sensitivity.  Stimulus  Maximum intensity = 0 dB Minimum intensity = 40 dB
  • 26.  Selection of adequate test  Proper environment  Comfortable sitting position  Adequate size of pupil >3mm  Adequate Near correction  Proper explanation – running of demonstration  Reassurance – not all points will be seen - test can be paused by keeping the response button pressed
  • 28.
  • 29.
  • 30. Patient data  Name, DOB, eye  Vision, refraction,  Pupil diameter Test data  Date and time  Program and strategy  Background illumination  Test size, color, duration, interval ZONE 1 : REPRODUCIBILITY
  • 31. REFRACTIVE ERROR- o Refractive blur reduces visual sensitivity to perimetric stimuli. o One dioptre of refractive blur in undilated patient will produce a little more than 1dB of depression of field of vision when testing with Goldmann Size III stimulus. o Hence near vision must be properly corrected , otherwise the visual field will show a generalized depression .
  • 32. Constricted field d/t miosis Pupil Diameter: 1.0mm
  • 33. Repeat Perimetry with Pupil Diameter: 2.5 mm
  • 34. ZONE 2 : RELIABILITY  Fixation monitor  Fixation target  Reliability indices Fixation losses < 20% Gaze tracking False positives < 15% (trigger happy) False negatives < 15% Foveal threshold
  • 35.  Fixation monitor may be chosen as blind spot and/or gaze.  Fixation target may be central or large diamond, the latter being chosen in patients with central visual loss.  Fixation losses: The fixation can be tested in different machines using blind spot monitoring with Heijl Krakau method, gaze monitoring with gaze tracker or manually by observation of eye monitor.  When the fixation monitoring test parameter is set to blind spot (Heijl-Krakau) mode, proper fixation is checked by projecting 5% of stimuli at the presumed location of the physiological blind spot.  If fixation losses exceed 20%, “XX” will be printed after the score.
  • 36.
  • 37. •False positive response- - FP >15% is strongly associated with compromised test. -With a high FP score grey scale appears abnormally pale.
  • 38. High false positive responses :  Seen in patients who are anxious about not seeing all test points and might be trigger happy.  Such a field would show in addition to false positives, numeric sensitivities above 40 dB, white areas on grayscale, a pattern deviation plot worse than total deviation, positive mean deviation and abnormally high sensitivity on GHT.
  • 39.  False negative response- - In general >15% of FN is considered abnormal -Grey scale pattern in high FN cases tends to have a clover- leaf pattern.
  • 40. High values occur if:  The patient is fatigued and falling asleep  Has changed personal criteria for response,  There is presence of true visual field loss where sensitivities are variable.  May be normally present in cases with advanced defect as the visibility may be highly variable in glaucomatous visual fields
  • 41. While false negative errors and fixation loss scores may not make a field invalid, high false positive errors may warrant rejection of the test results.
  • 42. RAW DATA AND GRAYSCALE  Simple threshold sensitivities are measured in decibels and given in the numeric plot.  Values above 40 dB are generally not expected and may be indicative of a trigger happy field.  The actual numeric values may not be concentrated upon for day to day interpretation of field but only with specific purposes.  Grayscale is an artists interpolation of the raw data and we do not base our diagnosis on it.
  • 43.  The raw data is the exact retinal sensitivity in dB units of the selected points calculated by field analyser.  In raw data 0 dB = absolute scotoma.  40 dB is the highest retinal sensitivity.  In same patient the raw data calculated by different strategies is not exactly similar .
  • 44. ZONE 4 : GREY SCALE  Based on actual threshold values at each location  General identification  Patient information
  • 45.  It can though typically highlight certain artifactual fields  Lid artifact –showing defect in superior part due to a droopy eyelid  Rim artifact- showing concentric constriction due to rim of trial lens which is either decentered or placed away from the eye  Clover leaf-typical pattern of initial response around four primary points in each quadrant followed by depressed or no response as seen in a new and poorly instructed patient.  Trigger happy fields-patches of white tones in grayscale seen in overanxious patient with high false positive responses.
  • 46.
  • 47. ZONE 5&6 :TOTAL DEVIATION PLOT  Numerical plot – indicates by how much decibels is each point depressed compared to mean value in normal population of similar age.  Probability plot- grey scale indicates the probability of occurrence of the deviation in normal population. Generalized depression due to media opacities, refractive error, miosis may hamper appearance of a pattern
  • 48.  P<5% indicates the retinal sensitivity of that point is seen in < 5% of normal population.  Darker the symbol, the greater the probability of abnormality as indicated by P value.  Higher the P value lesser the chances of field being abnormal
  • 49. ZONE 7&8 : PATTERN DEVIATION PLOT  Numerical - calculated by adjustment for generalized depression or elevation of visual field
  • 50. COMPARISON OF TOTAL AND PATTERN DEVIATION PLOTS  Uniformly depressed total deviation plot and normal looking pattern deviation plot: cataract or any generalized depression of fi eld  Both plots look more or less the same: little or no generalized loss  Normal total deviation and abnormal looking pattern deviation: trigger happy patient
  • 51.
  • 52. ZONE-9 GLOBAL INDICES  The global indices summarise the field.  The indices available in SITA strategies are:  Mean Deviation: how much the total field departs from normal  Pattern Standard Deviation: irregularities in the field.  Visual Field Index: visual field status as a percent of a normal field.  Short term fluctuation (SF)  Corrected pattern standard deviation(CPSD)
  • 53. • MD gives an indication of the overall sensitivity of the field. • The positive value  better overall retinal sensitivity than normal observer. • Negative value  worse than normal.
  • 54.  Negative value would be increased in presence of:  Generalized depression of entire field  Multiple localized depressions  Deep localized depression  Positive mean deviation is seen in:  Hypersensitive or trigger happy patient  Normal patients having sensitivity values higher than those in normative database.
  • 55. • Measure of focal loss or variability within the field taking into account any generalized depression in the hill of vision.  More specific indicator of glaucomatous damage than MD.
  • 56.  Measures irregularity by summing the absolute value of the difference between the threshold value for each point and the average visual field sensitivity at each point (equal to the normal value for each point + the MD).  Low PSD indicates:  Smooth hill of vision  Near normal field  Severely depressed field  High PSD indicates:  Variability in field response  Irregular field due to localized deep defects
  • 57. KEY NOTE POINTS  The diagnosis of glaucoma should not be based on global indices.  Probability plots and GHT are more useful for diagnosis while indices may help in staging disease and follow up.  So in follow up of a patient if a worsening MD is accompanied by a stable PSD it could be due to increasing generalized loss like in cataract.  On the other hand a nearly stable or slightly lowering MD accompanied by a worsening PSD could be indicative of worsening localized loss as in glaucoma
  • 58. VISUAL FIELD INDEX  New index which expresses the visual field status as a percent of a normal age-adjusted visual field.  This PSD based index was devised mainly to detect progression of glaucomatous defect.  Greater weight is given to points closer to fi xation to adjust for gangion cell density and visual function.  The index may also be less sensitive to cataract and media changes.
  • 59. ZONE 10 : GLAUCOMA HEMIFIELD TEST  Comparison of 5 clusters of points in superior hemifield with mirror images in inferior hemifield
  • 60. POSSIBLE OUTCOMES ARE:  Outside normal limits. The GHT is described as “outside normal limits” when matched pairs of zones differ by an amount greater than that seen in 99% normal subjects, or when both members of a pair of zones are more abnormal than 99.5% of the individuals with the normative population.  Borderline. When matched pairs of zones differ by an amount greater than that seen in 97% normal subjects but not reaching the level of outside normal limits  General reduction of sensitivity. When the best region of the VF is depressed to a level seen in only 0.5% of the normative database but both conditions for “outside normal limits” are not met  Abnormally high sensitivity. When the best region of the VF is elevated to a level seen in only 0.5% of the normative database.  Within normal limits. When none of the above conditions are met.
  • 61. ARTEFACTS  OBSTRUCTION RIM ARTEFACTS PTOSIS MEDIA OPACITIES  MIOSIS  REFRACTION ARTEFACTS  High power plus and minus lenses
  • 62.
  • 63.
  • 64.
  • 65. ANDERSON & PATELLA CRITERIA  3 or more congrous ‘non edge points’ in typical arcuate area on 30-2 program depressed @ p< 5 % with at least one point @ p<1 % PSD / CPSD @ p< 5% GHT – outside normal limits  Must be demonstrated on 2 field tests
  • 66.
  • 70.
  • 71. RECOGNITION OF PROGRESSIVE DAMAGE  Compare baseline field to each individual field  Study the entire series.
  • 73. READING A SAMPLE PRINTOUT
  • 74. TAKE HOME MESSAGE  The field defect must be reproducible as is confirmed by repeat field examination of the same patient  Clinical correlation is of prime importance .  Interpreting a single analysis may not be difficult if we go in a stepwise manner.  Important thing to remember is that presence of field defect does not always mean presence of disease.

Editor's Notes

  1. To find points in the visual field of equal sensitivities – ISOPTER (Groenouw) marking Intensity and size of stimulus is varied to mark various isopters Thus 2 D Contour map of the hill of vision is made Extent of scotomas and blind spot marked from inside out
  2. Screening: glaucoma, full field test, central30 deg, peripheral -68.
  3. Intensity of stimulus increased by 4 dbs steps until threshold is crossed. Threshold is then redetermined by decreasing the intensity by 2 dbs steps. Most accurate method of monitoring glvf defects.
  4. Inverted logarithmic scale with 0dB=brightest stimulus. Not standardized.
  5. Recognise artifacts: Reliability. Assessment of damage
  6. losses-The fixation loss rate measures patients gaze stability. During the test 5% of the stimulus will be presented on the blind spot. The patients response to this stimulus presentation will tell that the patient is not gazing straight or is looking from side to side during test.
  7. These are detected when stimulus is accompanied by a sound.If the sound alone is presented and the patient still responds,a false positive is recorded.
  8. These are detected by presenting a stimulus much (9dB)brighter than threshold at a location where the threshold has already been determined. If the patient fails to respond a false negative is recorded
  9. The diagnosis of glaucoma should not be based on global indices. Probability plots and GHT are more useful for diagnosis while indices may help in staging disease and follow up. So in follow up of a patient if a worsening MD is accompanied by a stable PSD it could be due to increasing generalized loss like in cataract. On the other hand a nearly stable or slightly lowering MD accompanied by a worsening PSD could be indicative of worsening localized loss as in glaucoma.
  10. GHT zone pattern optimised for glaucoma though can identify most abnormal fi elds so one should not be tempted to rely on this for analyzing non glaucomatous fi eld defects. A central scotoma caused by retinal pathology as in this case would be classifi ed as within normal limits due to absence of hemifi eld asymmetry (Figure 7).