STOP FLOW METHOD
PRESENDED BY;
B.MURUGAVENI
I MSC BIOCHEMISTRY
ROLL NO: 23PSBCH11
REG NO: 23701511
CONTENT:
INTRODUCTION
HISTORY OF STOP FLOW METHOD
STOP FLOW METHOD SUMMARY
REACTANT SYRINGE
MIXING CHAMBER
DEAD TIME
OBSERVATION CELL
STOPPING
SINGLE MIXING
MULTI MIXING STOP FLOW
APPLICATION OF STOP FLOW METHOD
INTRODUCTION:
Stopped-flow is an experimental technique for studying chemical reactions
with a half time of the order of 1 ms, introduced by Britton Chance and
extended by Quentin Gibson.
History of stopped flow method:
The stopped-flow technique was invented in the 1950s by professor
Britton Chance as a result of a need for instrumentation to follow
enzymatic reactions in the tens of millisecond time range. In 1983 Yves
Dupont from the French research agency, the CNRS (Centre National de la
Recherche Scientifique) developed his own stopped-flow for his laboratory
for fast fluorescence and radioactive labeling measurements.
STOP FLOW METHOD SUMMARY:
Stopped-flow spectrometry allows chemical kinetics of fast reactions
(with half times of the order of milliseconds) to be studied in
solution.
It was first used primarily to study enzyme-catalyzed reactions.
Then the stopped-flow rapidly found its place in almost all
biochemistry, biophysics, and chemistry laboratories with a need to
follow chemical reactions in the millisecond time scale.
In its simplest form, a stopped-flow mixes two solutions. Small
volumes of solutions are rapidly and continuously driven into a high-
efficiency mixer. This mixing process then initiates an extremely fast
reaction.
 The newly mixed solution travels to the observation cell and pushes out
the contents of the cell (the solution remaining from the previous
experiment or from necessary washing steps).
The time required for this solution to pass from the mixing point to the
observation point is known as dead time.
The minimum injection volume will depend on the volume of the mixing
cell.
Once enough solution has been injected to completely remove the
previous solution, the instrument reaches a stationary state and the flow
can be stopped.
Depending on the syringe drive technology, the flow stop is achieved
by using a stop valve called the hard-stop or by using a stop syringe.
The stopped-flow also sends a ‘start signal’ to the detector called the
trigger so the reaction can be observed.
The timing of the trigger is usually software controlled so the user can
trigger at the same time the flow stops or a few milliseconds before the
stop to check the stationary state has been reached.
REACTANT SYRINGES:
Two syringes are filled with solutions that do not undergo a
chemical reaction until mixed together.
These have pistons that are driven by a single drive piston or
by independent stepping motors, so that they are coupled
together and their contents are forced out simultaneously into
a mixing device.
MIXING CHAMBER:
Once the two solutions are
forced out of their syringes they
enter a mixing system that has
to ensure complete mixing,
with turbulent flow rather
baffles than laminar flow,
which would allow the two
solutions to flow side by side
with incomplete mixing.
DEAD TIME:
The dead time is the time for the solutions to go from the mixing
point to the observation point, it is the part of the kinetics which
cannot be observed. So the lower the dead time, the more information
the user can get.
In older instruments this could be of the order of 1 ms, but
improvements now allow a dead time of about 0.3 ms.
OBSERVATION CELL:
The mixed reactants pass an observation cell that allows the reaction to be followed
spectrophotometrically, typically by ultraviolet spectroscopy, fluorescence
spectroscopy, circular dichroism or light scattering, and it is now common to
combine several of these.
Observation cuvette with a short light path (0.75 to 1.5mm) are usually preferred
for fluorescence measurements to reduce self-absorption effects.
 Observation cuvette with longer light path (0.5 cm to 1 cm) are preferred for
absorbance measurements.
Modern stopped-flow can accommodate different models of cells and it is possible
to change the cuvette between two experiments. For stopped-flow X-ray
measurements, a quartz capillary with thin wall is used to minimize quartz
absorption. Simultaneous x-ray and absorbance measurements are possible in the
same capillary.
STOPPING:
Once through the observation cell the mixture enters a third syringe
that contains a piston that is driven by the flow to activate a switch to
stop the flow and activate the observation.
SINGLE MIXING:
In its simplest form, a stopped-flow mixes two solutions. Small volumes
of solutions are rapidly and continuously driven into a high-efficiency Ball
mixer, so mixing is completed in just a few microseconds.
This mixing process then initiates an extremely fast reaction. The newly
mixed solution travels to the observation cell and pushes out the
contents of the cell (the solution remaining from the previous experiment
or from necessary washing steps).
The time required for this solution to pass from the mixing point to the
observation point is known as dead time. The minimum injection
volume will depend on the volume of the mixing cell.
MULTI MIXING STOPPED FLOW:
A three syringe-stopped flow has 2 mixers that work consecutively.
Syringes 1 and 2 mix first, then the mixture mixes with syringe 3.
Two types of experiments can be performed with a multi-mixing
stopped-flow.
1. DOUBLE MIXING EXPERINMENT
2. FOUR SYRINGE STOPPED FLOW
DOUBLE MIXING EXPERINMENT:
A double mixing experiment (also called sequential mixing) is an
application that generate one reactant by mixing two solutions because
this reactant is not stable and could therefore not be loaded in a syringe for
a single mixing application.
 So, in this scenario, a first reaction is generated in mixer 1. This mixture
is allowed to age, from 2ms to several seconds, in a delay line before being
mixed with a third sample. The second reaction is then followed optically.
 The delay line is also called the aging line, aging loop, or retardation
loop but all of these terms refer to the same component.
Such experiments are performed to generate an unstable reactant that
could not be directly loaded into a two-syringe instrument.
The aging time is the time it takes for the solution to pass from mixer
1 to mixer 2. For short aging times (<200ms) the stopped-flow works
in continuous flow mode, so the aging time is defined as the ratio of
the volume of the delay line divided by the flow rate of solutions
through the line.
The user can change the delay line itself or can change the flow rate in
the software. For longer aging times, the stopped-flow uses an
interrupted flow method (push-pause-push) where the flow is stopped
between the two mixing events.
FOUR SYRINGE STOPPED FLOW:
A four-syringe stopped-flow has
three consecutive mixers and four
independent motors so it can carry
out triple mixing experiments, for
example by combining the double
mixing and the concentration
dependence.
APPLICATION OF STOP FLOW METHOD:
Stopped-Flows are used across a wide range of academic and
industrial applications including drug binding processes and protein
stability.
Stopped-Flows are key instruments used to increase researchers'
understanding of rapid kinetics and are especially useful for
chemists and biochemists.
Stopped flow method allows the evaluation of the degradation
kinetics in milliseconds and avoids the use of additional analytical
methodologies; this method allows determining the reaction times on
line.
REFERENCE:
1. https://www.biologic.net/topics/
what-are-stopped-flows-and-
how-do-they-help-chemists-and-
biochemists-understand-high-
speed-chemical-reactions/
2. https://en.wikipedia.org/wiki/Sto
pped-flow
STOPPED FLOW METHOD & APPLICATION MURUGAVENI B.pptx

More Related Content

What's hot

What's hot (20)

GLYCOGEN METABOLISM
GLYCOGEN METABOLISMGLYCOGEN METABOLISM
GLYCOGEN METABOLISM
 
hmp-shunt
 hmp-shunt hmp-shunt
hmp-shunt
 
URONIC ACID PATHWAY
URONIC ACID PATHWAYURONIC ACID PATHWAY
URONIC ACID PATHWAY
 
Regulation of fattyacid metabolism
Regulation of fattyacid metabolismRegulation of fattyacid metabolism
Regulation of fattyacid metabolism
 
HEXOSE MONOPHOSPHATE SHUNT
HEXOSE MONOPHOSPHATE SHUNTHEXOSE MONOPHOSPHATE SHUNT
HEXOSE MONOPHOSPHATE SHUNT
 
Citric Acid Cycle
Citric Acid CycleCitric Acid Cycle
Citric Acid Cycle
 
Glycogen metabolism and it’s regulation
Glycogen metabolism and it’s regulationGlycogen metabolism and it’s regulation
Glycogen metabolism and it’s regulation
 
Citric acid cycle (TCA cycle) by Dr. Anurag Yadav
Citric acid cycle (TCA cycle) by Dr. Anurag YadavCitric acid cycle (TCA cycle) by Dr. Anurag Yadav
Citric acid cycle (TCA cycle) by Dr. Anurag Yadav
 
Glucogenolysis
GlucogenolysisGlucogenolysis
Glucogenolysis
 
Class 4 gluconeogenesis
Class 4 gluconeogenesisClass 4 gluconeogenesis
Class 4 gluconeogenesis
 
6 shuttles
6 shuttles6 shuttles
6 shuttles
 
Nucleotide metabolism (purine and pyrimidine synthesis)
Nucleotide metabolism (purine and pyrimidine synthesis)Nucleotide metabolism (purine and pyrimidine synthesis)
Nucleotide metabolism (purine and pyrimidine synthesis)
 
Glycolysis
GlycolysisGlycolysis
Glycolysis
 
Carbohydrate metabolism minor pathways
Carbohydrate metabolism minor pathwaysCarbohydrate metabolism minor pathways
Carbohydrate metabolism minor pathways
 
Glycogenesis and glycogenolysis
Glycogenesis and glycogenolysisGlycogenesis and glycogenolysis
Glycogenesis and glycogenolysis
 
Gluconeogenesis and glycogenolysis
Gluconeogenesis and glycogenolysisGluconeogenesis and glycogenolysis
Gluconeogenesis and glycogenolysis
 
Gluconeogenesis
GluconeogenesisGluconeogenesis
Gluconeogenesis
 
carbohydrate metabolism, Glycolysis, metabolic process of carbohydrates, EMP ...
carbohydrate metabolism, Glycolysis, metabolic process of carbohydrates, EMP ...carbohydrate metabolism, Glycolysis, metabolic process of carbohydrates, EMP ...
carbohydrate metabolism, Glycolysis, metabolic process of carbohydrates, EMP ...
 
Enzyme inhibition - Competitive, Non- Competitive, Uncompetitive, Allosteric
Enzyme inhibition - Competitive, Non- Competitive, Uncompetitive, Allosteric Enzyme inhibition - Competitive, Non- Competitive, Uncompetitive, Allosteric
Enzyme inhibition - Competitive, Non- Competitive, Uncompetitive, Allosteric
 
Glycolysis ppt
Glycolysis pptGlycolysis ppt
Glycolysis ppt
 

Similar to STOPPED FLOW METHOD & APPLICATION MURUGAVENI B.pptx

HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
Vandana Devesh Sharma
 
1957992 635169139917595000
1957992 6351691399175950001957992 635169139917595000
1957992 635169139917595000
MANJU461988
 
Research Project Poster
Research Project PosterResearch Project Poster
Research Project Poster
Joshua Arucan
 
Hplc..
Hplc..Hplc..
Hplc..
VN Raj
 

Similar to STOPPED FLOW METHOD & APPLICATION MURUGAVENI B.pptx (20)

Determining the rate of a chemical reaction(Stopped-flow techniques)
Determining the rate of a chemical reaction(Stopped-flow techniques)Determining the rate of a chemical reaction(Stopped-flow techniques)
Determining the rate of a chemical reaction(Stopped-flow techniques)
 
Biomedical Instrumentation auto analyser.pptx
Biomedical Instrumentation auto analyser.pptxBiomedical Instrumentation auto analyser.pptx
Biomedical Instrumentation auto analyser.pptx
 
Royal ppt hplc
Royal ppt hplcRoyal ppt hplc
Royal ppt hplc
 
Ultra centrifuge
Ultra centrifugeUltra centrifuge
Ultra centrifuge
 
High Performance Liquid Chromatography
High Performance Liquid ChromatographyHigh Performance Liquid Chromatography
High Performance Liquid Chromatography
 
Hplc instrumentation
Hplc instrumentationHplc instrumentation
Hplc instrumentation
 
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...HPLC,introduction, theory, instrumentation, advantage, limitation,application...
HPLC,introduction, theory, instrumentation, advantage, limitation,application...
 
1957992 635169139917595000
1957992 6351691399175950001957992 635169139917595000
1957992 635169139917595000
 
Anu
AnuAnu
Anu
 
New microsoft office power point presentation
New microsoft office power point presentationNew microsoft office power point presentation
New microsoft office power point presentation
 
HPLC AND ITS PHARMACEUTICAL APPLICATIONS
HPLC AND ITS PHARMACEUTICAL APPLICATIONSHPLC AND ITS PHARMACEUTICAL APPLICATIONS
HPLC AND ITS PHARMACEUTICAL APPLICATIONS
 
Flow injection analysis
Flow injection analysisFlow injection analysis
Flow injection analysis
 
instrumentation of HPLC
instrumentation of HPLCinstrumentation of HPLC
instrumentation of HPLC
 
High performance liquid chromatography
High performance liquid chromatographyHigh performance liquid chromatography
High performance liquid chromatography
 
Hplc doc
Hplc docHplc doc
Hplc doc
 
High throughput nanoparticle synthesis and treatment with combinatorial robot
High throughput nanoparticle synthesis and treatment with combinatorial robotHigh throughput nanoparticle synthesis and treatment with combinatorial robot
High throughput nanoparticle synthesis and treatment with combinatorial robot
 
Compendial methods of dissolution testing
Compendial methods of dissolution testing Compendial methods of dissolution testing
Compendial methods of dissolution testing
 
Research Project Poster
Research Project PosterResearch Project Poster
Research Project Poster
 
Hplc..
Hplc..Hplc..
Hplc..
 
Auto analyzer
Auto analyzerAuto analyzer
Auto analyzer
 

Recently uploaded

Warming the earth and the atmosphere.pptx
Warming the earth and the atmosphere.pptxWarming the earth and the atmosphere.pptx
Warming the earth and the atmosphere.pptx
GlendelCaroz
 
Quantifying Artificial Intelligence and What Comes Next!
Quantifying Artificial Intelligence and What Comes Next!Quantifying Artificial Intelligence and What Comes Next!
Quantifying Artificial Intelligence and What Comes Next!
University of Hertfordshire
 

Recently uploaded (20)

Film Coated Tablet and Film Coating raw materials.pdf
Film Coated Tablet and Film Coating raw materials.pdfFilm Coated Tablet and Film Coating raw materials.pdf
Film Coated Tablet and Film Coating raw materials.pdf
 
Lubrication System in forced feed system
Lubrication System in forced feed systemLubrication System in forced feed system
Lubrication System in forced feed system
 
Heads-Up Multitasker: CHI 2024 Presentation.pdf
Heads-Up Multitasker: CHI 2024 Presentation.pdfHeads-Up Multitasker: CHI 2024 Presentation.pdf
Heads-Up Multitasker: CHI 2024 Presentation.pdf
 
MODERN PHYSICS_REPORTING_QUANTA_.....pdf
MODERN PHYSICS_REPORTING_QUANTA_.....pdfMODERN PHYSICS_REPORTING_QUANTA_.....pdf
MODERN PHYSICS_REPORTING_QUANTA_.....pdf
 
Information science research with large language models: between science and ...
Information science research with large language models: between science and ...Information science research with large language models: between science and ...
Information science research with large language models: between science and ...
 
Alternative method of dissolution in-vitro in-vivo correlation and dissolutio...
Alternative method of dissolution in-vitro in-vivo correlation and dissolutio...Alternative method of dissolution in-vitro in-vivo correlation and dissolutio...
Alternative method of dissolution in-vitro in-vivo correlation and dissolutio...
 
Molecular and Cellular Mechanism of Action of Hormones such as Growth Hormone...
Molecular and Cellular Mechanism of Action of Hormones such as Growth Hormone...Molecular and Cellular Mechanism of Action of Hormones such as Growth Hormone...
Molecular and Cellular Mechanism of Action of Hormones such as Growth Hormone...
 
In-pond Race way systems for Aquaculture (IPRS).pptx
In-pond Race way systems for Aquaculture (IPRS).pptxIn-pond Race way systems for Aquaculture (IPRS).pptx
In-pond Race way systems for Aquaculture (IPRS).pptx
 
Introduction and significance of Symbiotic algae
Introduction and significance of  Symbiotic algaeIntroduction and significance of  Symbiotic algae
Introduction and significance of Symbiotic algae
 
Biochemistry and Biomolecules - Science - 9th Grade by Slidesgo.pptx
Biochemistry and Biomolecules - Science - 9th Grade by Slidesgo.pptxBiochemistry and Biomolecules - Science - 9th Grade by Slidesgo.pptx
Biochemistry and Biomolecules - Science - 9th Grade by Slidesgo.pptx
 
GBSN - Microbiology (Unit 4) Concept of Asepsis
GBSN - Microbiology (Unit 4) Concept of AsepsisGBSN - Microbiology (Unit 4) Concept of Asepsis
GBSN - Microbiology (Unit 4) Concept of Asepsis
 
Efficient spin-up of Earth System Models usingsequence acceleration
Efficient spin-up of Earth System Models usingsequence accelerationEfficient spin-up of Earth System Models usingsequence acceleration
Efficient spin-up of Earth System Models usingsequence acceleration
 
ABHISHEK ANTIBIOTICS PPT MICROBIOLOGY // USES OF ANTIOBIOTICS TYPES OF ANTIB...
ABHISHEK ANTIBIOTICS PPT MICROBIOLOGY  // USES OF ANTIOBIOTICS TYPES OF ANTIB...ABHISHEK ANTIBIOTICS PPT MICROBIOLOGY  // USES OF ANTIOBIOTICS TYPES OF ANTIB...
ABHISHEK ANTIBIOTICS PPT MICROBIOLOGY // USES OF ANTIOBIOTICS TYPES OF ANTIB...
 
GBSN - Microbiology (Unit 5) Concept of isolation
GBSN - Microbiology (Unit 5) Concept of isolationGBSN - Microbiology (Unit 5) Concept of isolation
GBSN - Microbiology (Unit 5) Concept of isolation
 
Mining Activity and Investment Opportunity in Myanmar.pptx
Mining Activity and Investment Opportunity in Myanmar.pptxMining Activity and Investment Opportunity in Myanmar.pptx
Mining Activity and Investment Opportunity in Myanmar.pptx
 
WASP-69b’s Escaping Envelope Is Confined to a Tail Extending at Least 7 Rp
WASP-69b’s Escaping Envelope Is Confined to a Tail Extending at Least 7 RpWASP-69b’s Escaping Envelope Is Confined to a Tail Extending at Least 7 Rp
WASP-69b’s Escaping Envelope Is Confined to a Tail Extending at Least 7 Rp
 
Warming the earth and the atmosphere.pptx
Warming the earth and the atmosphere.pptxWarming the earth and the atmosphere.pptx
Warming the earth and the atmosphere.pptx
 
Costs to heap leach gold ore tailings in Karamoja region of Uganda
Costs to heap leach gold ore tailings in Karamoja region of UgandaCosts to heap leach gold ore tailings in Karamoja region of Uganda
Costs to heap leach gold ore tailings in Karamoja region of Uganda
 
Manganese‐RichSandstonesasanIndicatorofAncientOxic LakeWaterConditionsinGale...
Manganese‐RichSandstonesasanIndicatorofAncientOxic  LakeWaterConditionsinGale...Manganese‐RichSandstonesasanIndicatorofAncientOxic  LakeWaterConditionsinGale...
Manganese‐RichSandstonesasanIndicatorofAncientOxic LakeWaterConditionsinGale...
 
Quantifying Artificial Intelligence and What Comes Next!
Quantifying Artificial Intelligence and What Comes Next!Quantifying Artificial Intelligence and What Comes Next!
Quantifying Artificial Intelligence and What Comes Next!
 

STOPPED FLOW METHOD & APPLICATION MURUGAVENI B.pptx

  • 1. STOP FLOW METHOD PRESENDED BY; B.MURUGAVENI I MSC BIOCHEMISTRY ROLL NO: 23PSBCH11 REG NO: 23701511
  • 2. CONTENT: INTRODUCTION HISTORY OF STOP FLOW METHOD STOP FLOW METHOD SUMMARY REACTANT SYRINGE MIXING CHAMBER DEAD TIME OBSERVATION CELL STOPPING SINGLE MIXING MULTI MIXING STOP FLOW APPLICATION OF STOP FLOW METHOD
  • 3. INTRODUCTION: Stopped-flow is an experimental technique for studying chemical reactions with a half time of the order of 1 ms, introduced by Britton Chance and extended by Quentin Gibson. History of stopped flow method: The stopped-flow technique was invented in the 1950s by professor Britton Chance as a result of a need for instrumentation to follow enzymatic reactions in the tens of millisecond time range. In 1983 Yves Dupont from the French research agency, the CNRS (Centre National de la Recherche Scientifique) developed his own stopped-flow for his laboratory for fast fluorescence and radioactive labeling measurements.
  • 4. STOP FLOW METHOD SUMMARY: Stopped-flow spectrometry allows chemical kinetics of fast reactions (with half times of the order of milliseconds) to be studied in solution. It was first used primarily to study enzyme-catalyzed reactions. Then the stopped-flow rapidly found its place in almost all biochemistry, biophysics, and chemistry laboratories with a need to follow chemical reactions in the millisecond time scale. In its simplest form, a stopped-flow mixes two solutions. Small volumes of solutions are rapidly and continuously driven into a high- efficiency mixer. This mixing process then initiates an extremely fast reaction.
  • 5.  The newly mixed solution travels to the observation cell and pushes out the contents of the cell (the solution remaining from the previous experiment or from necessary washing steps). The time required for this solution to pass from the mixing point to the observation point is known as dead time. The minimum injection volume will depend on the volume of the mixing cell. Once enough solution has been injected to completely remove the previous solution, the instrument reaches a stationary state and the flow can be stopped.
  • 6. Depending on the syringe drive technology, the flow stop is achieved by using a stop valve called the hard-stop or by using a stop syringe. The stopped-flow also sends a ‘start signal’ to the detector called the trigger so the reaction can be observed. The timing of the trigger is usually software controlled so the user can trigger at the same time the flow stops or a few milliseconds before the stop to check the stationary state has been reached.
  • 7.
  • 8.
  • 9. REACTANT SYRINGES: Two syringes are filled with solutions that do not undergo a chemical reaction until mixed together. These have pistons that are driven by a single drive piston or by independent stepping motors, so that they are coupled together and their contents are forced out simultaneously into a mixing device.
  • 10. MIXING CHAMBER: Once the two solutions are forced out of their syringes they enter a mixing system that has to ensure complete mixing, with turbulent flow rather baffles than laminar flow, which would allow the two solutions to flow side by side with incomplete mixing.
  • 11. DEAD TIME: The dead time is the time for the solutions to go from the mixing point to the observation point, it is the part of the kinetics which cannot be observed. So the lower the dead time, the more information the user can get. In older instruments this could be of the order of 1 ms, but improvements now allow a dead time of about 0.3 ms.
  • 12. OBSERVATION CELL: The mixed reactants pass an observation cell that allows the reaction to be followed spectrophotometrically, typically by ultraviolet spectroscopy, fluorescence spectroscopy, circular dichroism or light scattering, and it is now common to combine several of these. Observation cuvette with a short light path (0.75 to 1.5mm) are usually preferred for fluorescence measurements to reduce self-absorption effects.  Observation cuvette with longer light path (0.5 cm to 1 cm) are preferred for absorbance measurements. Modern stopped-flow can accommodate different models of cells and it is possible to change the cuvette between two experiments. For stopped-flow X-ray measurements, a quartz capillary with thin wall is used to minimize quartz absorption. Simultaneous x-ray and absorbance measurements are possible in the same capillary.
  • 13.
  • 14. STOPPING: Once through the observation cell the mixture enters a third syringe that contains a piston that is driven by the flow to activate a switch to stop the flow and activate the observation.
  • 15. SINGLE MIXING: In its simplest form, a stopped-flow mixes two solutions. Small volumes of solutions are rapidly and continuously driven into a high-efficiency Ball mixer, so mixing is completed in just a few microseconds. This mixing process then initiates an extremely fast reaction. The newly mixed solution travels to the observation cell and pushes out the contents of the cell (the solution remaining from the previous experiment or from necessary washing steps). The time required for this solution to pass from the mixing point to the observation point is known as dead time. The minimum injection volume will depend on the volume of the mixing cell.
  • 16.
  • 17. MULTI MIXING STOPPED FLOW: A three syringe-stopped flow has 2 mixers that work consecutively. Syringes 1 and 2 mix first, then the mixture mixes with syringe 3. Two types of experiments can be performed with a multi-mixing stopped-flow. 1. DOUBLE MIXING EXPERINMENT 2. FOUR SYRINGE STOPPED FLOW
  • 18. DOUBLE MIXING EXPERINMENT: A double mixing experiment (also called sequential mixing) is an application that generate one reactant by mixing two solutions because this reactant is not stable and could therefore not be loaded in a syringe for a single mixing application.  So, in this scenario, a first reaction is generated in mixer 1. This mixture is allowed to age, from 2ms to several seconds, in a delay line before being mixed with a third sample. The second reaction is then followed optically.  The delay line is also called the aging line, aging loop, or retardation loop but all of these terms refer to the same component. Such experiments are performed to generate an unstable reactant that could not be directly loaded into a two-syringe instrument.
  • 19. The aging time is the time it takes for the solution to pass from mixer 1 to mixer 2. For short aging times (<200ms) the stopped-flow works in continuous flow mode, so the aging time is defined as the ratio of the volume of the delay line divided by the flow rate of solutions through the line. The user can change the delay line itself or can change the flow rate in the software. For longer aging times, the stopped-flow uses an interrupted flow method (push-pause-push) where the flow is stopped between the two mixing events.
  • 20.
  • 21. FOUR SYRINGE STOPPED FLOW: A four-syringe stopped-flow has three consecutive mixers and four independent motors so it can carry out triple mixing experiments, for example by combining the double mixing and the concentration dependence.
  • 22. APPLICATION OF STOP FLOW METHOD: Stopped-Flows are used across a wide range of academic and industrial applications including drug binding processes and protein stability. Stopped-Flows are key instruments used to increase researchers' understanding of rapid kinetics and are especially useful for chemists and biochemists. Stopped flow method allows the evaluation of the degradation kinetics in milliseconds and avoids the use of additional analytical methodologies; this method allows determining the reaction times on line.