Ultrasound is used for many reasons, including to:
View the uterus and ovaries during pregnancy and monitor the developing baby's health
Diagnose gallbladder disease
Evaluate blood flow
Guide a needle for biopsy or tumor treatment
Examine a breast lump
Check the thyroid gland
Find genital and prostate problems
Assess joint inflammation (synovitis)
Evaluate metabolic bone disease
Before your ultrasound begins, you may be asked to do the following:
Remove any jewelry from the area being examined.
Remove or reposition some or all of your clothing.
Change into a gown.
You'll be asked to lie on an examination table.
During the procedure
Gel is applied to your skin over the area being examined. It helps prevent air pockets, which can block the sound waves that create the images. This safe, water-based gel is easy to remove from skin and, if needed, clothing.
A trained technician (sonographer) presses a small, hand-held device (transducer) against the area being studied and moves it as needed to capture the images. The transducer sends sound waves into your body, collects the ones that bounce back and sends them to a computer, which creates the images.
Sometimes, ultrasounds are done inside your body. In this case, the transducer is attached to a probe that's inserted into a natural opening in your body. Examples include:
Transesophageal echocardiogram. A transducer, inserted into the esophagus, obtains heart images. It's usually done while under sedation.
Transrectal ultrasound. This test creates images of the prostate by placing a special transducer into the rectum.
Transvaginal ultrasound. A special transducer is gently inserted into the vagina to look at the uterus and ovaries.
Ultrasound is usually painless. However, you may experience mild discomfort as the sonographer guides the transducer over your body, especially if you're required to have a full bladder, or inserts it into your body.
A typical ultrasound exam takes from 30 minutes to an hour.
Results
When your exam is complete, a doctor trained to interpret imaging studies (radiologist) analyzes the images and sends a report to your doctor. Your doctor will share the results with you.
You should be able to return to normal activities immediately after an ultrasound.
Definition of Side lobes and the principle behind its production during ultrasound imaging. Side lobes artifact and its result on image. Explanation of harmonic imaging, its production and the techniques use to eliminate fundamental frequency to produce optimal harmonic images.
Definition of Side lobes and the principle behind its production during ultrasound imaging. Side lobes artifact and its result on image. Explanation of harmonic imaging, its production and the techniques use to eliminate fundamental frequency to produce optimal harmonic images.
Ultrasound Physics Made easy - By Dr Chandni WadhwaniChandni Wadhwani
History of ultrasound, Principle of Ultrasound.
Ultrasound wave and its interactions
Ultrasound machine and its parts, Image display, Artifacts and their clinical importance
what is Doppler ultrasound, Elastography and Recent advances in field of ultrasound.
Safety issues in ultrasound.
A detailed description of ct coronary angiography and calcium scoring with various aspects regarding the preparation, procedure, limitations and a short review regarding post CABG imaging.
Usg transducer and basic principles of ultrasound Doppler, this slide describe the basic physics of ultrasound transducer and Doppler , must know thing is given in this presentaion. Good review for radiology resident. Thanks.
Ultrasound Physics Made easy - By Dr Chandni WadhwaniChandni Wadhwani
History of ultrasound, Principle of Ultrasound.
Ultrasound wave and its interactions
Ultrasound machine and its parts, Image display, Artifacts and their clinical importance
what is Doppler ultrasound, Elastography and Recent advances in field of ultrasound.
Safety issues in ultrasound.
A detailed description of ct coronary angiography and calcium scoring with various aspects regarding the preparation, procedure, limitations and a short review regarding post CABG imaging.
Usg transducer and basic principles of ultrasound Doppler, this slide describe the basic physics of ultrasound transducer and Doppler , must know thing is given in this presentaion. Good review for radiology resident. Thanks.
Magnetic resonance imaging (MRI) is a medical imaging technique that uses a magnetic field and computer-generated radio waves to create detailed images of the organs and tissues in your body.
Most MRI machines are large, tube-shaped magnets. When you lie inside an MRI machine, the magnetic field temporarily realigns water molecules in your body. Radio waves cause these aligned atoms to produce faint signals, which are used to create cross-sectional MRI images — like slices in a loaf of bread.
The MRI machine can also produce 3D images that can be viewed from different angles.
Products & Services
Sign up for Email: Get Your Free Resource – Coping with Cancer
Why it's done
MRI is a noninvasive way for your doctor to examine your organs, tissues and skeletal system. It produces high-resolution images of the inside of the body that help diagnose a variety of problems.
MRI of the brain and spinal cord
MRI is the most frequently used imaging test of the brain and spinal cord. It's often performed to help diagnose:
Aneurysms of cerebral vessels
Disorders of the eye and inner ear
Multiple sclerosis
Spinal cord disorders
Stroke
Tumors
Brain injury from trauma
A special type of MRI is the functional MRI of the brain (fMRI). It produces images of blood flow to certain areas of the brain. It can be used to examine the brain's anatomy and determine which parts of the brain are handling critical functions.
This helps identify important language and movement control areas in the brains of people being considered for brain surgery. Functional MRI can also be used to assess damage from a head injury or from disorders such as Alzheimer's disease.
MRI of the heart and blood vessels
MRI that focuses on the heart or blood vessels can assess:
Size and function of the heart's chambers
Thickness and movement of the walls of the heart
Extent of damage caused by heart attacks or heart disease
Structural problems in the aorta, such as aneurysms or dissections
Inflammation or blockages in the blood vessels
Oral and maxillofacial imaging is no exception. As a specialty that deals with uncommon lesions and complex
anatomy, both students and practicing dental clinicians
may benefit from this simplistic, pattern-based approach.
This presentation describes a compendium of the classic signs in oral and maxillofacial radiology.
These lecture slides, by Dr Sidra Arshad, offer a quick overview of physiological basis of a normal electrocardiogram.
Learning objectives:
1. Define an electrocardiogram (ECG) and electrocardiography
2. Describe how dipoles generated by the heart produce the waveforms of the ECG
3. Describe the components of a normal electrocardiogram of a typical bipolar leads (limb II)
4. Differentiate between intervals and segments
5. Enlist some common indications for obtaining an ECG
Study Resources:
1. Chapter 11, Guyton and Hall Textbook of Medical Physiology, 14th edition
2. Chapter 9, Human Physiology - From Cells to Systems, Lauralee Sherwood, 9th edition
3. Chapter 29, Ganong’s Review of Medical Physiology, 26th edition
4. Electrocardiogram, StatPearls - https://www.ncbi.nlm.nih.gov/books/NBK549803/
5. ECG in Medical Practice by ABM Abdullah, 4th edition
6. ECG Basics, http://www.nataliescasebook.com/tag/e-c-g-basics
Flu Vaccine Alert in Bangalore Karnatakaaddon Scans
As flu season approaches, health officials in Bangalore, Karnataka, are urging residents to get their flu vaccinations. The seasonal flu, while common, can lead to severe health complications, particularly for vulnerable populations such as young children, the elderly, and those with underlying health conditions.
Dr. Vidisha Kumari, a leading epidemiologist in Bangalore, emphasizes the importance of getting vaccinated. "The flu vaccine is our best defense against the influenza virus. It not only protects individuals but also helps prevent the spread of the virus in our communities," he says.
This year, the flu season is expected to coincide with a potential increase in other respiratory illnesses. The Karnataka Health Department has launched an awareness campaign highlighting the significance of flu vaccinations. They have set up multiple vaccination centers across Bangalore, making it convenient for residents to receive their shots.
To encourage widespread vaccination, the government is also collaborating with local schools, workplaces, and community centers to facilitate vaccination drives. Special attention is being given to ensuring that the vaccine is accessible to all, including marginalized communities who may have limited access to healthcare.
Residents are reminded that the flu vaccine is safe and effective. Common side effects are mild and may include soreness at the injection site, mild fever, or muscle aches. These side effects are generally short-lived and far less severe than the flu itself.
Healthcare providers are also stressing the importance of continuing COVID-19 precautions. Wearing masks, practicing good hand hygiene, and maintaining social distancing are still crucial, especially in crowded places.
Protect yourself and your loved ones by getting vaccinated. Together, we can help keep Bangalore healthy and safe this flu season. For more information on vaccination centers and schedules, residents can visit the Karnataka Health Department’s official website or follow their social media pages.
Stay informed, stay safe, and get your flu shot today!
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdfAnujkumaranit
Artificial intelligence (AI) refers to the simulation of human intelligence processes by machines, especially computer systems. It encompasses tasks such as learning, reasoning, problem-solving, perception, and language understanding. AI technologies are revolutionizing various fields, from healthcare to finance, by enabling machines to perform tasks that typically require human intelligence.
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...i3 Health
i3 Health is pleased to make the speaker slides from this activity available for use as a non-accredited self-study or teaching resource.
This slide deck presented by Dr. Kami Maddocks, Professor-Clinical in the Division of Hematology and
Associate Division Director for Ambulatory Operations
The Ohio State University Comprehensive Cancer Center, will provide insight into new directions in targeted therapeutic approaches for older adults with mantle cell lymphoma.
STATEMENT OF NEED
Mantle cell lymphoma (MCL) is a rare, aggressive B-cell non-Hodgkin lymphoma (NHL) accounting for 5% to 7% of all lymphomas. Its prognosis ranges from indolent disease that does not require treatment for years to very aggressive disease, which is associated with poor survival (Silkenstedt et al, 2021). Typically, MCL is diagnosed at advanced stage and in older patients who cannot tolerate intensive therapy (NCCN, 2022). Although recent advances have slightly increased remission rates, recurrence and relapse remain very common, leading to a median overall survival between 3 and 6 years (LLS, 2021). Though there are several effective options, progress is still needed towards establishing an accepted frontline approach for MCL (Castellino et al, 2022). Treatment selection and management of MCL are complicated by the heterogeneity of prognosis, advanced age and comorbidities of patients, and lack of an established standard approach for treatment, making it vital that clinicians be familiar with the latest research and advances in this area. In this activity chaired by Michael Wang, MD, Professor in the Department of Lymphoma & Myeloma at MD Anderson Cancer Center, expert faculty will discuss prognostic factors informing treatment, the promising results of recent trials in new therapeutic approaches, and the implications of treatment resistance in therapeutic selection for MCL.
Target Audience
Hematology/oncology fellows, attending faculty, and other health care professionals involved in the treatment of patients with mantle cell lymphoma (MCL).
Learning Objectives
1.) Identify clinical and biological prognostic factors that can guide treatment decision making for older adults with MCL
2.) Evaluate emerging data on targeted therapeutic approaches for treatment-naive and relapsed/refractory MCL and their applicability to older adults
3.) Assess mechanisms of resistance to targeted therapies for MCL and their implications for treatment selection
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...Oleg Kshivets
RESULTS: Overall life span (LS) was 2252.1±1742.5 days and cumulative 5-year survival (5YS) reached 73.2%, 10 years – 64.8%, 20 years – 42.5%. 513 LCP lived more than 5 years (LS=3124.6±1525.6 days), 148 LCP – more than 10 years (LS=5054.4±1504.1 days).199 LCP died because of LC (LS=562.7±374.5 days). 5YS of LCP after bi/lobectomies was significantly superior in comparison with LCP after pneumonectomies (78.1% vs.63.7%, P=0.00001 by log-rank test). AT significantly improved 5YS (66.3% vs. 34.8%) (P=0.00000 by log-rank test) only for LCP with N1-2. Cox modeling displayed that 5YS of LCP significantly depended on: phase transition (PT) early-invasive LC in terms of synergetics, PT N0—N12, cell ratio factors (ratio between cancer cells- CC and blood cells subpopulations), G1-3, histology, glucose, AT, blood cell circuit, prothrombin index, heparin tolerance, recalcification time (P=0.000-0.038). Neural networks, genetic algorithm selection and bootstrap simulation revealed relationships between 5YS and PT early-invasive LC (rank=1), PT N0—N12 (rank=2), thrombocytes/CC (3), erythrocytes/CC (4), eosinophils/CC (5), healthy cells/CC (6), lymphocytes/CC (7), segmented neutrophils/CC (8), stick neutrophils/CC (9), monocytes/CC (10); leucocytes/CC (11). Correct prediction of 5YS was 100% by neural networks computing (area under ROC curve=1.0; error=0.0).
CONCLUSIONS: 5YS of LCP after radical procedures significantly depended on: 1) PT early-invasive cancer; 2) PT N0--N12; 3) cell ratio factors; 4) blood cell circuit; 5) biochemical factors; 6) hemostasis system; 7) AT; 8) LC characteristics; 9) LC cell dynamics; 10) surgery type: lobectomy/pneumonectomy; 11) anthropometric data. Optimal diagnosis and treatment strategies for LC are: 1) screening and early detection of LC; 2) availability of experienced thoracic surgeons because of complexity of radical procedures; 3) aggressive en block surgery and adequate lymph node dissection for completeness; 4) precise prediction; 5) adjuvant chemoimmunoradiotherapy for LCP with unfavorable prognosis.
The prostate is an exocrine gland of the male mammalian reproductive system
It is a walnut-sized gland that forms part of the male reproductive system and is located in front of the rectum and just below the urinary bladder
Function is to store and secrete a clear, slightly alkaline fluid that constitutes 10-30% of the volume of the seminal fluid that along with the spermatozoa, constitutes semen
A healthy human prostate measures (4cm-vertical, by 3cm-horizontal, 2cm ant-post ).
It surrounds the urethra just below the urinary bladder. It has anterior, median, posterior and two lateral lobes
It’s work is regulated by androgens which are responsible for male sex characteristics
Generalised disease of the prostate due to hormonal derangement which leads to non malignant enlargement of the gland (increase in the number of epithelial cells and stromal tissue)to cause compression of the urethra leading to symptoms (LUTS
Ozempic: Preoperative Management of Patients on GLP-1 Receptor Agonists Saeid Safari
Preoperative Management of Patients on GLP-1 Receptor Agonists like Ozempic and Semiglutide
ASA GUIDELINE
NYSORA Guideline
2 Case Reports of Gastric Ultrasound
2. Artifact is used to describe any part of an image that
does not accurately represent the anatomic structures
present within the subject being evaluated.
In ultrasonography (US), artifacts may cause structures
to appear in an image that are not present
anatomically or a structure that is present anatomically
may be missing from the image. US artifacts may also
show structures as present but incorrect in location,
size, or brightness.
US is prone to numerous imaging artifacts, and these
are commonly encountered in clinical practice.
Artifacts have the potential to interfere with image
interpretation.
3. ARTIFACT
PROPAGATION
VELOCITY ARTIFACT
ATTENUATION ERRORS
PATH OF SOUND-
RELATED
GAS-RELATED
ARTIFACTS
BEAM PROFILE–
RELATED
ARTIFACTS
DOPPLER
IMAGING
ARTIFACTS
Shadowing
Increased Through
Transmission
Mirror Image Artifact
Comet-Tail Artifact
Refraction
Anisotropy
Reverberation Artifact
Reverberation Artifact
Ring-Down Artifact
Dirty Shadowing Artifact
Side Lobe & Grating Lobe
Artifacts
Partial Volume Averaging
Loss or Distortion of
Doppler
Information
Artifactual Vascular Flow
Tissue Vibration Artifact
Aliasing and Velocity
Scale Errors
Spectral Broadening
Blooming Artifact
Twinkle Artifact
4. PROPAGATION VELOCITY ARTIFACT
ATTENUATION ERRORS
SHADOWING
Shadowing results when there is a marked reduction in the
intensity of the ultrasound deep to a strong reflector, attenuator,
or refractor. Clean dark shadows will be seen behind calcified
objects when the focal zone is at or just below the structure
Gallstone With Shadowing Shadowing From Hernia Repair Mesh
5. Shadowing of Intrauterine Device Coned down view of distal neonatal spine shows shadowing from vertebral bodies that
interrupt the linear appearance of the nerve roots
Shadowing From Ovarian Fibroma. This is a slightly less
“clean” shadow since the fibroma is not as dense as a calcified stone.
Edge shadowing artifact behind the posterior aspect of the fetal skull.
The beam bends at curved surface and loses intensity, producing a
shadow
6. PROPAGATION VELOCITY ARTIFACT
ATTENUATION ERRORS
Increased Through Transmission
The hepatic parenchyma distal to the cysts is falsely displayed as increased in echogenicity (arrows)
secondary to increased through-transmission artifact. The cyst attenuates 7 dB less than the normal tissue, and time gain curve correction for
normal tissue results in over amplification of the signals deep to the cyst, producing increased through transmission of these tissues
Increased through transmission occurs when an
object (such as a cyst) attenuates the sound waves
less than the surrounding tissues.
Nebothian cyst
7. ARTIFACT
PROPAGATION
VELOCITY ARTIFACT
ATTENUATION ERRORS
PATH OF SOUND-
RELATED
GAS-RELATED
ARTIFACTS
BEAM PROFILE–
RELATED
ARTIFACTS
DOPPLER
IMAGING
ARTIFACTS
Shadowing
Increased Through
Transmission
Mirror Image Artifact
Comet-Tail Artifact
Refraction
Anisotropy
Reverberation Artifact
Reverberation Artifact
Ring-Down Artifact
Dirty Shadowing Artifact
Side Lobe & Grating Lobe
Artifacts
Partial Volume Averaging
Loss or Distortion of
Doppler
Information
Artifactual Vascular Flow
Tissue Vibration Artifact
Aliasing and Velocity
Scale Errors
Spectral Broadening
Blooming Artifact
Twinkle Artifact
8. PATH OF SOUND-RELATED ARTIFACTS
Mirror Image Artifact
Mirror image artifact in sonography is seen when there is
a highly reflective surface in the path of the primary
beam. (e.g. diaphragm)
The primary beam reflects from such a surface (e.g diaphragm ) but instead of directly being received by
the transducer, it encounters another structure (e.g. a nodular lesion) in its path and is reflected back to the highly
reflective surface (e.g. diaphragm). It then again reflects back towards the transducer.
The ultrasound machine makes a false assumption that the returning echo has been reflected once and hence the
delayed echoes are judged as if being returned from a deeper structure, thus giving a mirror artifact on the other
side of the reflective surface.
It is a friendly artifact that allows the sonographer to exclude pleural effusion by the reflection of the liver image
through the diaphragm.
Examples:
reflection of a liver lesion into the thorax (the commonest example)
reflection of abdominal ascites mimicking pleural effusion
duplication of gestational sac (either ghost twin or heterotopic pregnancy) 3
duplication of the uterus
To avoid this artifact, change the position and angle of scanning to change the angel of insonation of the primary
ultrasound beam.
9. Mirror Image Artifact. Soft tissue–gas interfaces, such as the diaphragm, are excellent
reflectors of the sound beam owing to the large difference in the acoustic impedance
between the two materials. Therefore hepatic structures are frequently seen mirrored
beyond the diaphragm onto the lungs. These two figures show liver parenchyma, hepatic
veins in part A, and a hepatic hemangioma in part B all inversely projected onto lung.
10. PROPAGATION VELOCITY ARTIFACT
ATTENUATION ERRORS
Comet-Tail Artifact
The comet tail artifact seen when small calcific / crystalline /
highly reflective objects are interrogated and is believed to
be a special form of reverberation artifact.
Small renal or ureteric calculi
Small common bile duct stones
Adenomyomatosis of the gallbladder
Pancreatic calcifications of chronic pancreatitis
Testicular microlithiasis (sometimes)
Thyroid colloid nodules
Identification of foreign bodies
Surgical clips
Catheter, Needle tips
Debris/glass/metal
Gallbladder adenomyomatosis
12. PROPAGATION VELOCITY ARTIFACT
ATTENUATION ERRORS
REFRACTION
A change in velocity of the ultrasound beam as it
travels through two adjacent tissues with different
density and elastic properties may produce a
refraction artifact such as from muscle to fat.
13. Second gestational sac due to refraction artifact “ghosting”
Refraction artifact (arrowheads) is caused by the interfaces between the ruptured tendon edges and hemorrhage.
14. PROPAGATION VELOCITY ARTIFACT
ATTENUATION ERRORS
Anisotrophy
Anisotropy is an artifact caused by structures that are
composed of bundles of highly reflective fibers
running parallel to each other, such as tendons and
ligaments
The insertion of supraspinatus tendon
artifactually appears hypoechoic due to anisotropy.
“Peritrigonal blush” region of the white matte
(ellipse) appears relatively echogenic
15. PROPAGATION VELOCITY ARTIFACT
ATTENUATION ERRORS
Reverberation Artifact
Reverberation artifact occurs when a strong
reflector runs perpendicular to the direction of the
beam (i.e., parallel to the probe surface), usually
close to the skin surface.
The sound waves may then get partially “trapped” between the reflector and skin, reverberating back and forth and
causing the appearance of multiple regular lines when some of the waves eventually.
Reverberation artifacts arise when the ultrasound signal reflects repeatedly between highly reflective interfaces near
the transducer, resulting in delayed echo return to the transducer. This appears in the image as a series of regularly
spaced echoes at increasing depth reach the probe. Often the reflectors are the skin and subcutaneous
fascia.Reverberation artifact is angle dependent, so moving the transducer slightly will cause a decrease in or
elimination of the artifact.
Reverberation can be helpful during biopsies, to see a needle. If it is distracting, try changing the angle of insonation,
using a different window, or decreasing the gain.
The highly reflective and mirror like surface of the pacemaker (arrow) results in repeated bouncing of sound wave among the probe
surface, pacemaker and skin.
16. ARTIFACT
PROPAGATION
VELOCITY ARTIFACT
ATTENUATION ERRORS
PATH OF SOUND-
RELATED
GAS-RELATED
ARTIFACTS
BEAM PROFILE–
RELATED
ARTIFACTS
DOPPLER
IMAGING
ARTIFACTS
Shadowing
Increased Through
Transmission
Mirror Image Artifact
Comet-Tail Artifact
Refraction
Anisotropy
Reverberation Artifact
Reverberation Artifact
Ring-Down Artifact
Dirty Shadowing Artifact
Side Lobe & Grating Lobe
Artifacts
Partial Volume Averaging
Loss or Distortion of
Doppler
Information
Artifactual Vascular Flow
Tissue Vibration Artifact
Aliasing and Velocity
Scale Errors
Spectral Broadening
Blooming Artifact
Twinkle Artifact
17. GAS-RELATED ARTIFACTS The physical pressure of the probe on the soft
tissues, such as the abdominal wall, flattens their
shape in the near field, making them perpendicular
to the surface of the probe and the incident
sound beam. Because gases conform to the shape of
their containers, they too will form a linear and
perpendicular interface to the incident sound beam,
especially when in the form of a large bubble. The
highly reflective gas interface will act like a
mirror,resulting in reverberation artifacts.
Reverberation Artifact
Small amount of gas within the
bladder is recognizable by the
reverberation artifact
18. GAS-RELATED ARTIFACTS
Ring-Down Artifact
Ring-down artifact has been thought to be a variant of
comet tailartifact. This assumption was based on the
often similar appearance
of the two artifacts.
In ring-down artifact, the transmitted ultrasound
energy causes resonant vibrations within fluid trapped
between a tetrahedron of air bubbles. These vibrations
create a continuous sound wave that is transmitted
back to the transducer .
This phenomenon is displayed as a line or series of
parallel bands extending posterior to a gas collection.
Despite the similar sonographic appearance, these two
artifacts have separate mechanisms
19. Ring-Down Artifact- Pneumobilia causing ring-down artifact in the biliary tree (A) and gallbladder (B).
A B
Ring-down artifact-peripheral ducts of the liver(C) Iatrogenic air in the bladder introduced at cystoscopy
appears as a nondependent bright region (D)
C D
20. GAS-RELATED ARTIFACTS
Dirty Shadowing Artifact
Dirty shadowing results from a combination of
reflection,reverberation, and ring-down artifacts arising
from multiple variably sized bubbles in foam.
21. Dirty Shadowing Posterior to Emphysematous Cholecystitis
Dirty Shadowing in the Kidney Due to Emphysematous Pyelonephritis
22. ARTIFACT
PROPAGATION
VELOCITY ARTIFACT
ATTENUATION ERRORS
PATH OF SOUND-
RELATED
GAS-RELATED
ARTIFACTS
BEAM PROFILE–
RELATED
ARTIFACTS
DOPPLER
IMAGING
ARTIFACTS
Shadowing
Increased Through
Transmission
Mirror Image Artifact
Comet-Tail Artifact
Refraction
Anisotropy
Reverberation Artifact
Reverberation Artifact
Ring-Down Artifact
Dirty Shadowing Artifact
Side Lobe & Grating Lobe
Artifacts
Partial Volume Averaging
Loss or Distortion of
Doppler
Information
Artifactual Vascular Flow
Tissue Vibration Artifact
Aliasing and Velocity
Scale Errors
Spectral Broadening
Blooming Artifact
Twinkle Artifact
23. BEAM PROFILE–RELATED ARTIFACTS
Side Lobe & Grating Lobe Artifacts
Side lobe artifacts occur where side lobes reflect sound from
a strong reflector that is outside of the central beam, and
where the echoes are displayed as if they originated from
within the central beam. Side lobe beams are low-intensity
beams that surround the central beam
Side lobe artifacts are echogenic, linear or curvilinear
artifacts. Strong reflectors include bowel gas adjacent to the
gallbladder or urinary bladder.
24. Pseudosludge
Echoes seen in the nondependent portion of
the bladder. Note also the shadowing behind
the bladder stone
25. BEAM PROFILE–RELATED ARTIFACTS
Partial Volume Averaging
The ultrasound beam not only has a complex shape in the imaging
plane but also has a real profile in the third dimension, called
the “elevation plane” or “Z plane.” The ultrasound image appears
as a flat two-dimensional image, but the brightness of each pixel
is representative of the average sum of all the echoes received
within the thickness of the beam in the elevation plane. This
results in echoes being projected in structures that should be
anechoic.
Partial Volume Averaging. The bladder wall is indistinct
in region of this image. It is unclear if there are masses or if this is due
to grating lobe artifact or partial volume artifact or both. Moving the
patient or transducer to avoid bowel gas and putting the focal zone in
the region in question should aid in evaluation. For the area closer to
the patient’s anterior abdominal wall, a higher-frequency transducer
might be helpful
26. ARTIFACT
PROPAGATION
VELOCITY ARTIFACT
ATTENUATION ERRORS
PATH OF SOUND-
RELATED
GAS-RELATED
ARTIFACTS
BEAM PROFILE–
RELATED
ARTIFACTS
DOPPLER
IMAGING
ARTIFACTS
Shadowing
Increased Through
Transmission
Mirror Image Artifact
Comet-Tail Artifact
Refraction
Anisotropy
Reverberation Artifact
Reverberation Artifact
Ring-Down Artifact
Dirty Shadowing Artifact
Side Lobe & Grating Lobe
Artifacts
Partial Volume Averaging
Loss or Distortion of
Doppler
Information
Artifactual Vascular Flow
Tissue Vibration Artifact
Aliasing and Velocity
Scale Errors
Spectral Broadening
Blooming Artifact
Twinkle Artifact
27. DOPPLER IMAGING ARTIFACTS
Loss or Distortion of Doppler
Information
The appearance of color Doppler signal is affected by several
variables including color-write priority, gray-scale gain, and
pulse repetition frequency. If the color gain is too low, flow
might not be visualized. If the gain is too high, artifactual flow
may be seen in adjacent soft tissues, and thrombus within the
vessel might be missed.
Incorrect gain, wall-filter, and velocity scale can all lead to
loss or distortion of Doppler signal.
Frequency, the rule of thumb is to use a Doppler angle less than
60 degrees (but not 0). Wall filters eliminate low-frequency
noise, but a high setting can lead to loss of signal.
In general, wall filters should be kept at the lowest practical
level, typically in the range of 50 to 100 Hz.
28. A (PRF = 700 Hz) B (PRF = 4500 Hz)
Artifactual Lack of Flow. Color Doppler image of a carotid artery and jugular vein. In (A), the pulse repetition frequency
(PRF) is 700 Hz and there is aliasing in the carotid artery, but slow flow in the jugular vein is seen. In (B), the PRF is 4500 Hz,
eliminating aliasing in the artery but also suppressing the display of the low Doppler frequencies in the internal jugular vein.
29. Artifactual Lack of Flow. (A) Color Doppler image shows flow in the portal vein. (B) When a color image is taken to assess
the common duct, no flow is visible in the portal vein. Note the difference in overall gain in image B.
A B
30. DOPPLER IMAGING ARTIFACTS
Artifactual Vascular Flow
The Doppler effect (shift) is not specific to vascular flow and
occurs with movement of any reflector toward or away from the
ultrasound beam. Fluid or solid tissue motion can mimic
vascular flow. Transmitted pulsations, especially close to the
heart or major arteries, can therefore result in artifactual
appearance of flow within thrombosed veins or avascular areas.
Artifactual vascular flow can also be visualized if the color gain
is too high or the color-write priority is too high.
Color Doppler jets in the bladder use color motion
to indicate that there is flow from each ureteral
orifice
31. Color flow Doppler signal in pleural fluid with debris. (A) Gray-scale sonogram shows left pleural fluid
with much debris. (B) Color flow Doppler signal
A B
32. Color Streaking Artifact. (A) This complicated cyst contains floating punctate echoes that move posteriorly
while being scanned,creating a scintillating appearance on gray-scale sonography. (B) Power Doppler
ultrasound pushes the echoes posteriorly faster and with more energy than does the gray-scale beam. The
echoes move fast enough that color persistence creates the appearance of “color streaking” artifact
33. Flash Artifact From Pulsation. In this image of the left portal vein, note how
transmitted pulsation from the heart causes artifactual flow in the
adjacent liver owing to motion.
34. DOPPLER IMAGING ARTIFACTS
Tissue Vibration Artifact
The color bruit or tissue vibration artifact is a type of color
Doppler ultrasound artifact which results in color signal
overflowing to the perivascular tissues most often caused by
stenosis, AV fistulas, or shunts. Thus, this artifact is useful by
pinpointing areas of potentially pathological blood flow.
Tissue Vibration Artifact From Common Femoral Artery
to Common Femoral Vein Arteriovenous Fistula. Turbulent
flow through the fistula may affect surrounding tissues,
causing a tissue vibration artifact, which may be the first
clue that an arteriovenous fistula (AVF) is present. Common
femoral artery to common femoral vein
AVF. (A)
Color Doppler shows common femoral artery to common
femoral vein AVF. Note the adjacent tissue vibration artifact
(arrowheads).
Turbulent flow causes vibration of the vessel wall and
perivascular soft tissues, resulting in a detectable doppler
shift which typically results in a mixture of red and blue
speckles .
The color bruit is a beneficial artifact that helps with
localizing areas of critical stenosis or turbulent flow due to
other reasons
35. Tissue Vibration Artifact. Postbiopsy arteriovenous fistula. Color Doppler image shows markedly increased lower-pole
blood flow (arrows) with adjacent soft tissue color artifact related to tissue vibration (arrowheads)
36. DOPPLER IMAGING ARTIFACTS
Aliasing and Velocity Scale Errors
Aliasing is an artifact due to undersampling of Doppler
signal, resulting in incorrect estimation of flow velocity
To measure the Doppler frequency shift appropriately by
pulsed Doppler, at least two samples from the cycle are
required (Nyquist limit). If the pulse repetition frequency is
less than twice the maximum
frequency shift produced by movement of the target, aliasing
results. The Doppler signal wraps around either the spectrum
(spectral Doppler) or the color scale (color Doppler). The most
common methods for correcting for aliasing are to shift the
baseline down or up, or increase the pulse repetition
frequency (or velocity scale). A lower-frequency transducer
can also be used to help in correcting for aliasing. As
angulation approaches 90 degrees, directional ambiguity can
occur, suggesting bidirectional flow. Errors in Doppler angle
correction can also lead to misleading assessments of flow
velocity.
Tissue Vibration Artifact. Spectral waveform at site of
superior mesenteric artery stenosis shows very high systolic
velocity, and diastolic velocities of greater than 500 cm/sec.
Color portion of the image shows color bruit consisting of
color outside the vessel near the stenosis site. Color bruit is
believed to be caused by tissue vibration aliasing prevents
precise calculation of the maximum systolic velocity
37. A B
Aliasing (A) appears in the spectral display as a “wraparound” of the higher frequencies to display below
the baseline. (B) In color Doppler display, aliasing results in a wraparound of the frequency color map
from one flow direction to the opposite of unsaturated color.
Aliasing in Superior Mesenteric Artery (SMA) Due to
High-Velocity Flow
38. DOPPLER IMAGING ARTIFACTS
Spectral Broadening
Spectral broadening occurs when there are multiple
different velocities of flow within a vessel. This can be a sign
of stenosis.
However, artifactual spectral broadening can occur owing to
improper positioning of the sample volume near the vessel
wall, use of an excessively large sample volume, or excessive
system gain.
Spectral Broadening. The range of velocities
detected at a given time in the pulse cycle is
reflected in the Doppler spectrum
as spectral broadening. (A) Normal spectrum.
Spectral broadening may arise from turbulent
flow in association with vessel stenosis.
(B and C)
Artifactual spectral broadening may be
produced by improper positioning
of the sample volume near the vessel wall, use
of an excessively large sample volume (B), or
excessive system gain (C)
40. DOPPLER IMAGING ARTIFACTS
Blooming Artifact
Blooming artifact occurs when color reaches beyond the
vessel wall, making the vessels look larger than expected. It
is gain dependent in that lowering the gain will decrease
blooming. It is also wall filter and color-overwrite
dependent, in that increasing the wall filter will decrease the
artifact
Two images of the same carotid artery. In (A) the gain and wall filter are appropriate to show the flow centrallyin the vessel. In (B) the
gain and wall filter have been changed (in a manner beyond what would be used in clinical imaging) to demonstrate how these can
make the vessel “bloom” and appear larger
41. DOPPLER IMAGING ARTIFACTS
Twinkle Artifact
Twinkle artifact occurs as a focus of alternating
colors on Doppler signal behind a reflective object
(such as a calculus), which gives the appearance of
turbulent blood flow 2. It appears with or without an
associated color comet tail artifact.
The underlying mechanism of this artifact is thought
to be a result of inherent noise within the ultrasound
scanner, specifically phase (a.k.a. clock) jitter within
the Doppler electronics
Twinkling artifact is more sensitive for detection of
small stones (e.g. urolithiasis,cholelithiasis ) than is
acoustic shadowing.