SlideShare a Scribd company logo
1 of 33
Quality assurance in PET/CT
Dr.Vishnukumar R, MD Nuclear Medicine
Jawaharlal Institute of Post graduate Medical Education and Research (JIPMER),
Puducherry, India.
• Quality Assurance- General concept of taking actions
to ensure that delivered products or services meet
performance requirements.
• QMS- Defines steps has to be taken to ensure that
the desired level of care.
• Quality control- Specific set of measurements
focused on monitoring the performance of installed
imaging equipment relative to image quality and
dose on a periodic basis, for example, monthly or
daily.
Quality management system
Define actions
Set rules
Record
Train your staff
• Defined actions regarding quality assurance (QA).
• Documents illustrating correct use of the imaging
equipment, test objects, phantoms and sources.
• Detailing test modalities and procedures to follow in
the case of abnormal results.
• Records of all tests, calibrations and corrective
actions performed.
• Proper training of all the staff involved in handling
PET, QC and QA.
Acceptance testing
Optimal parameters
Regulatory body
• Acceptance testing:
– Done at the time of installation, end of warranty
period and post service.
– Serves to set reference values.
– Sets optimal operating parameters.
– Ensures buyer and regulatory bodies to meet their
specifications.
• Quality control:
– Routine procedures done at different frequencies.
– To ensure there is no deviation from the reference
values.
• Spatial resolution.
• Sensitivity.
• Scatter fraction.
• Image quality.
Spatial resolution
• Purpose:
– PMT and other components age over a period of
time.
• Materials:
– 3 point sources 1mm in diameter.
– 1MBq F18
– Source holder
• Data acquisition:
– 1cm represents centre of
the axial FOV.
– 20 cm: edge effects.
• Pixel size- 1.5mm/pixel
• 1,00,000 cpm
• Filtered back-projection
with a ramp filter and no
further smoothing should
be applied
• Data analysis:
– FWHM and FWTM will be calculated for all 3
positions (Radial, sagittal and transaxial).
– Values should meet specifications quoted by
vendor.
– Ratio of FWHM to FWTM-1.8 (1.8-2.0)
– Tolerance level
Sensitivity
• Aim:
– To determine the rate of detected true coincidence
events per unit of radioactivity concentration for a
standard source configuration.
• Materials:
– Line source required (700±5mm)
– 700 mm long Aluminum sleeves (5 sleeves).
– 5 MBq F18
• Data acquisition:
– Centre of transaxial FOV and 10 cm from centre FOV.
– 10000 true events per slice should be recorded.
– After each recording increase wall thickness by adding next
sleeve and complete all measurement associated with all 5
sleeves.
• Data analysis:
– Single slice rebinning should be used.
– Count rate of each slice associated with each sleeve
measurement has to be noted.
– Measured count rate has to be decay corrected and
attenuation corrected.
• System sensitivity is obtained by
• Tolerance level
Scatter fraction
• To estimate count
losses due to random
and scatter events.
• However due to
improvement in PET
systems random event
loss is negligible.
• The scatter fraction (SF)
is a measure of scatter
coincidence data in the
overall counts recorded.
• Materials:
– Polyethylene
phantom.
– Plastic line source
of 3.2 mm in
diameter.
– Activity should not
interfere with
system dead time.
• Data acquisition and
processing:
– Tomographic images are
acquired.
– Summed projection data of
all projections within
central 24 cm FOV will be
considered.
– Before summing, LOR in
each projection will be
normalized to central pixel.
– SF will be obtained from
Trapezoidal area and
counts beyond 24 cm from
central FOV
Before normalization
After normalization
Tolerance level
Image quality
• Different parameters like scanner sensitivity,
uniformity, contrast and spatial resolution, etc,..
Determines image quality.
• To provide an overall assessment of the imaging
capabilities of the system under similar
conditions as a clinical whole body scan.
• To compare different systems, to check the
reproducibility.
• NEMA recommends to use a phantom mimicking
patient chest.
Materials
• Body compartment- 18cm in
length.
• Six hollow spheres with internal
diameters of 1.0, 1.3, 1.7, 2.2,
2.8 and 3.7 cm with wall
thickness of not more that 1
mm.
• Cylindrical insert to mimic lung-
5 cm in diameter, filled with
low atomic number material.
• It contains styroform beads and
water.
• For out of field radioactivity,
another phantom similar to
scatter phantom is used
• Body compartment:
– Fill with an F18 solution of 5.3 ± 0.27 kBq/mL radioactivity
concentration.
• Spheres:
– 2.8 cm and 3.7 cm spheres are filled with water to mimic cold
lesions.
– 1.0, 1.3, 1.7 and 2.2 cm spheres are to be filled with an F18
solution that has either 4 or 8 times higher radioactivity
concentration than the background.
• Line source:
– Filled with 116 MBq of F18 to mimic an out-of-field radioactivity
with an effective concentration equal to the background
radioactivity concentration.
• Data acquisition:
– Should mimic whole body protocol.
– Scan speed- 100cm in 60 min
• Analysis:
– Choose a centre slice in which the hot and cold spheres are visualized with the
highest contrast.
– Circular ROI’s are drawn over hot and cold sphere areas.
– ROI size should equal to size of the respective spheres.
– 12 background ROI if 3.7 cm should be drawn 15 mm away from spheres.
– ROIs of the same sizes as the smaller spheres (10, 13, 17, 22 and 28 mm)
should be drawn concentric to each of the 37 mm ROIs on the background
region.
– 3 cm ROI is drawn over lung insert.
– Similar set of ROI are drawn over 4 slices surrounding center slice.
ROI
Image contrast:
– Contrast for hot sphere:
– Contrast for cold sphere:
Accuracy of attenuation and scatter correction:
– Check accuracy in uniform medium.
– Analysed in lung insert, if accurate values should
be as close to 0.
• Accuracy of radioactivity quantification:
– Measure in background area.
– Calculate true concentration after decay
correction.
– Calculate concentration from 12 background ROI.
Summary
• Quality assurance- process related; Quality
control- product related.
• Acceptance testing- done at installation, end
of warranty period and post service.
• Quality control- to ensure product works
according to the set standards.
• Acceptance testing- Spatial resolution,
sensitivity, scatter correction and image
quality.
Quality assurance in pet

More Related Content

What's hot

planning systems in radiotherapy
 planning systems in radiotherapy planning systems in radiotherapy
planning systems in radiotherapy
fondas vakalis
 

What's hot (20)

Epid
EpidEpid
Epid
 
ICRU reports 50 and 62
ICRU reports 50 and 62ICRU reports 50 and 62
ICRU reports 50 and 62
 
Brachytherapy dosimetry
Brachytherapy dosimetryBrachytherapy dosimetry
Brachytherapy dosimetry
 
Icrp60
Icrp60Icrp60
Icrp60
 
Radioisotopes and dose rates used for brachytherapy
Radioisotopes and dose rates used for brachytherapyRadioisotopes and dose rates used for brachytherapy
Radioisotopes and dose rates used for brachytherapy
 
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEXTISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
TISSUE PHANTOM RATIO - THE PHOTON BEAM QUALITY INDEX
 
planning systems in radiotherapy
 planning systems in radiotherapy planning systems in radiotherapy
planning systems in radiotherapy
 
EPID AND CBCT ON RADIATION THERAPY
EPID AND CBCT ON RADIATION THERAPYEPID AND CBCT ON RADIATION THERAPY
EPID AND CBCT ON RADIATION THERAPY
 
Treatment verification and set up errors
Treatment verification and set up errorsTreatment verification and set up errors
Treatment verification and set up errors
 
Rrecent advances in linear accelerators [MR linac]
Rrecent advances in linear accelerators [MR linac]Rrecent advances in linear accelerators [MR linac]
Rrecent advances in linear accelerators [MR linac]
 
EXTERNAL PHOTON BEAMS THERAPY (PART 2)
EXTERNAL PHOTON BEAMS THERAPY (PART 2)EXTERNAL PHOTON BEAMS THERAPY (PART 2)
EXTERNAL PHOTON BEAMS THERAPY (PART 2)
 
Respiration motion management
Respiration motion managementRespiration motion management
Respiration motion management
 
Measurement of absorbed dose
Measurement of absorbed doseMeasurement of absorbed dose
Measurement of absorbed dose
 
Icru 50,62,83 volume deliniation
Icru 50,62,83 volume deliniationIcru 50,62,83 volume deliniation
Icru 50,62,83 volume deliniation
 
Manchester system_Dr.Supriya
Manchester system_Dr.SupriyaManchester system_Dr.Supriya
Manchester system_Dr.Supriya
 
Target volume definition
Target volume definitionTarget volume definition
Target volume definition
 
CTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose IndexCTDI (Computed Tomography Dose Index
CTDI (Computed Tomography Dose Index
 
Patient Positioning and Immobilization Devices In Radiotherapy Planning
Patient Positioning and Immobilization Devices In Radiotherapy PlanningPatient Positioning and Immobilization Devices In Radiotherapy Planning
Patient Positioning and Immobilization Devices In Radiotherapy Planning
 
Quality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed TomographyQuality Assurance Programme in Computed Tomography
Quality Assurance Programme in Computed Tomography
 
Imrt&vmat
Imrt&vmatImrt&vmat
Imrt&vmat
 

Similar to Quality assurance in pet

Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjQualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Pratik434909
 
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxQualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Pratik434909
 
Optimal fuzzy rule based pulmonary nodule detection
Optimal fuzzy rule based pulmonary nodule detectionOptimal fuzzy rule based pulmonary nodule detection
Optimal fuzzy rule based pulmonary nodule detection
Wookjin Choi
 
Metrology and Quality Engineering.pptx
Metrology and Quality Engineering.pptxMetrology and Quality Engineering.pptx
Metrology and Quality Engineering.pptx
MandarPadhye5
 
VETOMAC_presentation_template.pptx
VETOMAC_presentation_template.pptxVETOMAC_presentation_template.pptx
VETOMAC_presentation_template.pptx
KarimullaShaik81
 

Similar to Quality assurance in pet (20)

Qualification of high performance liquid chromatography
Qualification of high performance liquid chromatographyQualification of high performance liquid chromatography
Qualification of high performance liquid chromatography
 
Hrt II
Hrt IIHrt II
Hrt II
 
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjQualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
 
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxQualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
 
qualification of FTIR and DSC
qualification of FTIR and DSCqualification of FTIR and DSC
qualification of FTIR and DSC
 
Commissionin of medical linear accelerator
Commissionin of medical linear acceleratorCommissionin of medical linear accelerator
Commissionin of medical linear accelerator
 
Different experimental techniques for solar flux
Different experimental techniques for solar fluxDifferent experimental techniques for solar flux
Different experimental techniques for solar flux
 
Quality assurance
Quality  assuranceQuality  assurance
Quality assurance
 
Optimal fuzzy rule based pulmonary nodule detection
Optimal fuzzy rule based pulmonary nodule detectionOptimal fuzzy rule based pulmonary nodule detection
Optimal fuzzy rule based pulmonary nodule detection
 
Task group report 135 cyberknie
Task group report 135 cyberknie Task group report 135 cyberknie
Task group report 135 cyberknie
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments
 
Emm unit i
Emm unit   iEmm unit   i
Emm unit i
 
Metrology and Quality Engineering.pptx
Metrology and Quality Engineering.pptxMetrology and Quality Engineering.pptx
Metrology and Quality Engineering.pptx
 
Qc in xray
Qc in xrayQc in xray
Qc in xray
 
QUALITY CONTROL PDF.pptx
QUALITY CONTROL PDF.pptxQUALITY CONTROL PDF.pptx
QUALITY CONTROL PDF.pptx
 
VETOMAC_presentation_template.pptx
VETOMAC_presentation_template.pptxVETOMAC_presentation_template.pptx
VETOMAC_presentation_template.pptx
 
Qualification of GC & FTIR
Qualification of GC & FTIRQualification of GC & FTIR
Qualification of GC & FTIR
 
14184864.ppt
14184864.ppt14184864.ppt
14184864.ppt
 
LTC Laser Profilometry
LTC Laser ProfilometryLTC Laser Profilometry
LTC Laser Profilometry
 
Dfma aaaaaaaaa
Dfma aaaaaaaaaDfma aaaaaaaaa
Dfma aaaaaaaaa
 

Recently uploaded

CAD CAM DENTURES IN PROSTHODONTICS : Dental advancements
CAD CAM DENTURES IN PROSTHODONTICS : Dental advancementsCAD CAM DENTURES IN PROSTHODONTICS : Dental advancements
CAD CAM DENTURES IN PROSTHODONTICS : Dental advancements
Naveen Gokul Dr
 
Cytoskeleton and Cell Inclusions - Dr Muhammad Ali Rabbani - Medicose Academics
Cytoskeleton and Cell Inclusions - Dr Muhammad Ali Rabbani - Medicose AcademicsCytoskeleton and Cell Inclusions - Dr Muhammad Ali Rabbani - Medicose Academics
Cytoskeleton and Cell Inclusions - Dr Muhammad Ali Rabbani - Medicose Academics
MedicoseAcademics
 
Unit 4 Pharmaceutical Organic Chemisty 3 Quinoline
Unit 4 Pharmaceutical Organic Chemisty 3 QuinolineUnit 4 Pharmaceutical Organic Chemisty 3 Quinoline
Unit 4 Pharmaceutical Organic Chemisty 3 Quinoline
AarishRathnam1
 

Recently uploaded (20)

CONGENITAL HYPERTROPHIC PYLORIC STENOSIS by Dr M.KARTHIK EMMANUEL
CONGENITAL HYPERTROPHIC PYLORIC STENOSIS  by Dr M.KARTHIK EMMANUELCONGENITAL HYPERTROPHIC PYLORIC STENOSIS  by Dr M.KARTHIK EMMANUEL
CONGENITAL HYPERTROPHIC PYLORIC STENOSIS by Dr M.KARTHIK EMMANUEL
 
High Purity 99% PMK Ethyl Glycidate Powder CAS 28578-16-7
High Purity 99% PMK Ethyl Glycidate Powder CAS 28578-16-7High Purity 99% PMK Ethyl Glycidate Powder CAS 28578-16-7
High Purity 99% PMK Ethyl Glycidate Powder CAS 28578-16-7
 
Anti viral drug pharmacology classification
Anti viral drug pharmacology classificationAnti viral drug pharmacology classification
Anti viral drug pharmacology classification
 
CAD CAM DENTURES IN PROSTHODONTICS : Dental advancements
CAD CAM DENTURES IN PROSTHODONTICS : Dental advancementsCAD CAM DENTURES IN PROSTHODONTICS : Dental advancements
CAD CAM DENTURES IN PROSTHODONTICS : Dental advancements
 
Tips and tricks to pass the cardiovascular station for PACES exam
Tips and tricks to pass the cardiovascular station for PACES examTips and tricks to pass the cardiovascular station for PACES exam
Tips and tricks to pass the cardiovascular station for PACES exam
 
Cytoskeleton and Cell Inclusions - Dr Muhammad Ali Rabbani - Medicose Academics
Cytoskeleton and Cell Inclusions - Dr Muhammad Ali Rabbani - Medicose AcademicsCytoskeleton and Cell Inclusions - Dr Muhammad Ali Rabbani - Medicose Academics
Cytoskeleton and Cell Inclusions - Dr Muhammad Ali Rabbani - Medicose Academics
 
Stereochemistry & Asymmetric Synthesis.pptx
Stereochemistry & Asymmetric Synthesis.pptxStereochemistry & Asymmetric Synthesis.pptx
Stereochemistry & Asymmetric Synthesis.pptx
 
Bangalore whatsapp Number Just VIP Brookefield 100% Genuine at your Door Step
Bangalore whatsapp Number Just VIP Brookefield 100% Genuine at your Door StepBangalore whatsapp Number Just VIP Brookefield 100% Genuine at your Door Step
Bangalore whatsapp Number Just VIP Brookefield 100% Genuine at your Door Step
 
Storage of Blood Components- equipments, effects of improper storage, transpo...
Storage of Blood Components- equipments, effects of improper storage, transpo...Storage of Blood Components- equipments, effects of improper storage, transpo...
Storage of Blood Components- equipments, effects of improper storage, transpo...
 
NDCT Rules, 2019: An Overview | New Drugs and Clinical Trial Rules 2019
NDCT Rules, 2019: An Overview | New Drugs and Clinical Trial Rules 2019NDCT Rules, 2019: An Overview | New Drugs and Clinical Trial Rules 2019
NDCT Rules, 2019: An Overview | New Drugs and Clinical Trial Rules 2019
 
Top 10 Most Beautiful Chinese Pornstars List 2024
Top 10 Most Beautiful Chinese Pornstars List 2024Top 10 Most Beautiful Chinese Pornstars List 2024
Top 10 Most Beautiful Chinese Pornstars List 2024
 
Hemodialysis: Chapter 1, Physiological Principles of Hemodialysis - Dr.Gawad
Hemodialysis: Chapter 1, Physiological Principles of Hemodialysis - Dr.GawadHemodialysis: Chapter 1, Physiological Principles of Hemodialysis - Dr.Gawad
Hemodialysis: Chapter 1, Physiological Principles of Hemodialysis - Dr.Gawad
 
Face and Muscles of facial expression.pptx
Face and Muscles of facial expression.pptxFace and Muscles of facial expression.pptx
Face and Muscles of facial expression.pptx
 
ROSE CASE SPINAL SBRT BY DR KANHU CHARAN PATRO
ROSE  CASE SPINAL SBRT BY DR KANHU CHARAN PATROROSE  CASE SPINAL SBRT BY DR KANHU CHARAN PATRO
ROSE CASE SPINAL SBRT BY DR KANHU CHARAN PATRO
 
ESC HF 2024 Spotlights Day-2.pptx heart failure
ESC HF 2024 Spotlights Day-2.pptx heart failureESC HF 2024 Spotlights Day-2.pptx heart failure
ESC HF 2024 Spotlights Day-2.pptx heart failure
 
Unveiling Alcohol Withdrawal Syndrome: exploring it's hidden depths
Unveiling Alcohol Withdrawal Syndrome: exploring it's hidden depthsUnveiling Alcohol Withdrawal Syndrome: exploring it's hidden depths
Unveiling Alcohol Withdrawal Syndrome: exploring it's hidden depths
 
Gallbladder Double-Diverticular: A Case Report المرارة مزدوجة التج: تقرير حالة
Gallbladder Double-Diverticular: A Case Report  المرارة مزدوجة التج: تقرير حالةGallbladder Double-Diverticular: A Case Report  المرارة مزدوجة التج: تقرير حالة
Gallbladder Double-Diverticular: A Case Report المرارة مزدوجة التج: تقرير حالة
 
parliaments-for-health-security_RecordOfAchievement.pdf
parliaments-for-health-security_RecordOfAchievement.pdfparliaments-for-health-security_RecordOfAchievement.pdf
parliaments-for-health-security_RecordOfAchievement.pdf
 
Report Back from SGO: What’s the Latest in Ovarian Cancer?
Report Back from SGO: What’s the Latest in Ovarian Cancer?Report Back from SGO: What’s the Latest in Ovarian Cancer?
Report Back from SGO: What’s the Latest in Ovarian Cancer?
 
Unit 4 Pharmaceutical Organic Chemisty 3 Quinoline
Unit 4 Pharmaceutical Organic Chemisty 3 QuinolineUnit 4 Pharmaceutical Organic Chemisty 3 Quinoline
Unit 4 Pharmaceutical Organic Chemisty 3 Quinoline
 

Quality assurance in pet

  • 1. Quality assurance in PET/CT Dr.Vishnukumar R, MD Nuclear Medicine Jawaharlal Institute of Post graduate Medical Education and Research (JIPMER), Puducherry, India.
  • 2.
  • 3.
  • 4. • Quality Assurance- General concept of taking actions to ensure that delivered products or services meet performance requirements. • QMS- Defines steps has to be taken to ensure that the desired level of care. • Quality control- Specific set of measurements focused on monitoring the performance of installed imaging equipment relative to image quality and dose on a periodic basis, for example, monthly or daily.
  • 8. • Defined actions regarding quality assurance (QA). • Documents illustrating correct use of the imaging equipment, test objects, phantoms and sources. • Detailing test modalities and procedures to follow in the case of abnormal results. • Records of all tests, calibrations and corrective actions performed. • Proper training of all the staff involved in handling PET, QC and QA.
  • 9.
  • 12. • Acceptance testing: – Done at the time of installation, end of warranty period and post service. – Serves to set reference values. – Sets optimal operating parameters. – Ensures buyer and regulatory bodies to meet their specifications. • Quality control: – Routine procedures done at different frequencies. – To ensure there is no deviation from the reference values.
  • 13. • Spatial resolution. • Sensitivity. • Scatter fraction. • Image quality.
  • 14. Spatial resolution • Purpose: – PMT and other components age over a period of time. • Materials: – 3 point sources 1mm in diameter. – 1MBq F18 – Source holder
  • 15. • Data acquisition: – 1cm represents centre of the axial FOV. – 20 cm: edge effects. • Pixel size- 1.5mm/pixel • 1,00,000 cpm • Filtered back-projection with a ramp filter and no further smoothing should be applied
  • 16. • Data analysis: – FWHM and FWTM will be calculated for all 3 positions (Radial, sagittal and transaxial). – Values should meet specifications quoted by vendor. – Ratio of FWHM to FWTM-1.8 (1.8-2.0) – Tolerance level
  • 17. Sensitivity • Aim: – To determine the rate of detected true coincidence events per unit of radioactivity concentration for a standard source configuration. • Materials: – Line source required (700±5mm) – 700 mm long Aluminum sleeves (5 sleeves). – 5 MBq F18
  • 18. • Data acquisition: – Centre of transaxial FOV and 10 cm from centre FOV. – 10000 true events per slice should be recorded. – After each recording increase wall thickness by adding next sleeve and complete all measurement associated with all 5 sleeves. • Data analysis: – Single slice rebinning should be used. – Count rate of each slice associated with each sleeve measurement has to be noted. – Measured count rate has to be decay corrected and attenuation corrected.
  • 19. • System sensitivity is obtained by • Tolerance level
  • 20. Scatter fraction • To estimate count losses due to random and scatter events. • However due to improvement in PET systems random event loss is negligible. • The scatter fraction (SF) is a measure of scatter coincidence data in the overall counts recorded.
  • 21. • Materials: – Polyethylene phantom. – Plastic line source of 3.2 mm in diameter. – Activity should not interfere with system dead time.
  • 22. • Data acquisition and processing: – Tomographic images are acquired. – Summed projection data of all projections within central 24 cm FOV will be considered. – Before summing, LOR in each projection will be normalized to central pixel. – SF will be obtained from Trapezoidal area and counts beyond 24 cm from central FOV
  • 24. Image quality • Different parameters like scanner sensitivity, uniformity, contrast and spatial resolution, etc,.. Determines image quality. • To provide an overall assessment of the imaging capabilities of the system under similar conditions as a clinical whole body scan. • To compare different systems, to check the reproducibility. • NEMA recommends to use a phantom mimicking patient chest.
  • 25. Materials • Body compartment- 18cm in length. • Six hollow spheres with internal diameters of 1.0, 1.3, 1.7, 2.2, 2.8 and 3.7 cm with wall thickness of not more that 1 mm. • Cylindrical insert to mimic lung- 5 cm in diameter, filled with low atomic number material. • It contains styroform beads and water. • For out of field radioactivity, another phantom similar to scatter phantom is used
  • 26. • Body compartment: – Fill with an F18 solution of 5.3 ± 0.27 kBq/mL radioactivity concentration. • Spheres: – 2.8 cm and 3.7 cm spheres are filled with water to mimic cold lesions. – 1.0, 1.3, 1.7 and 2.2 cm spheres are to be filled with an F18 solution that has either 4 or 8 times higher radioactivity concentration than the background. • Line source: – Filled with 116 MBq of F18 to mimic an out-of-field radioactivity with an effective concentration equal to the background radioactivity concentration.
  • 27.
  • 28. • Data acquisition: – Should mimic whole body protocol. – Scan speed- 100cm in 60 min • Analysis: – Choose a centre slice in which the hot and cold spheres are visualized with the highest contrast. – Circular ROI’s are drawn over hot and cold sphere areas. – ROI size should equal to size of the respective spheres. – 12 background ROI if 3.7 cm should be drawn 15 mm away from spheres. – ROIs of the same sizes as the smaller spheres (10, 13, 17, 22 and 28 mm) should be drawn concentric to each of the 37 mm ROIs on the background region. – 3 cm ROI is drawn over lung insert. – Similar set of ROI are drawn over 4 slices surrounding center slice.
  • 29. ROI
  • 30. Image contrast: – Contrast for hot sphere: – Contrast for cold sphere: Accuracy of attenuation and scatter correction: – Check accuracy in uniform medium. – Analysed in lung insert, if accurate values should be as close to 0.
  • 31. • Accuracy of radioactivity quantification: – Measure in background area. – Calculate true concentration after decay correction. – Calculate concentration from 12 background ROI.
  • 32. Summary • Quality assurance- process related; Quality control- product related. • Acceptance testing- done at installation, end of warranty period and post service. • Quality control- to ensure product works according to the set standards. • Acceptance testing- Spatial resolution, sensitivity, scatter correction and image quality.