QUALIFICATION OF HPLC SYSTEM AND LC-MS SYSTEM
2
QUALIFICATION OF HPLC SYSTEM
Introduction
Equipment qualification is a formal process that provides documented
evidence that an instrument is fit for its intended use.
The entire qualification process consists of four parts:
 Design qualification (DQ)
 Installation qualification (IQ)
 Operational qualification (OQ)
 Performance qualification (PQ)
3
Design qualification (DQ)
• Design qualification (DQ) describes the user requirements and defines the
functional and operational specifications of the instrument.
Design elements Examples
Intended use Analysis of drug compounds and impurities
User requirement specification
for the HPLC analysis
 Analysis up to 100 samples / day
 Automated over-night analysis
 Limit of quantitation: 0.1%
 Automated confirmation of peak identity and purity with
diode-array detection
 Automated compound quantitation and printing of report
4
Design elements Examples
Functional specifications
 Autosampler - 100 samples, 0.5 µl to 5 ml sample volume
 Column compartment - 15 to 60˚C, peltier controlled
 Detector - UV/Vis Diode-array, 190-900 nm
 Computer - System control, data acquisition for signals and
spectra, peak integration and quantitation, spectral evaluation
for peak purity and compound confirmation.
Operational specifications
Detector: Baseline noise < 5 X 10-5 AU
Sampler: Precision injection Volume
sample carry over: <0.5%
Maintenance
Vendor must deliver maintenance procedure and recommend
schedule
Training Vendor must provide familiarization and training
5
Installation qualification (IQ)
Before installation
 Check the site for the fulfillment of the manufacturer’s recommendations
(utilities such as electricity, and environmental conditions such as humidity and
temperature).
 Allow sufficient shelf space for the equipment, SOPs, operating manuals and
software.
During installation
 Compare equipment as received, with purchase order (including software,
accessories, spare parts)
 Check documentation for completeness (operating manuals, maintenance
instructions, standard operating procedures for testing, safety and validation
certificates) 6
 Check equipment for any damage
 Install hardware (computer, equipment, fittings and tubings for fluid
connections, columns in HPLC, power cables, and instrument control
cables)
 Switch on the instruments and ensure that all modules power up and
perform an electronic self-test.
 Run test sample and compare chromatogram with reference chromatogram
 Prepare an installation report
Installation qualification (IQ)
7
Performance qualification (PQ)
Parameter Procedure User Limit
Leak testing
Flow test by volume or
weight/time
± 5 %
Baseline drift ASTM Method <2 × 10-3 AU
Baseline noise ASTM Method <5 x 10-5 AU
Precision of injection volume
6 x injection of caffeine
standard, RSD of peak areas
0.3 % RSD
Precision of flow rate
6 x injection of caffeine
standard, RSD of retention
times
0.5 % RSD
8
ASTM Method → American Society for Testing and Material Method
Parameter Procedure User Limit
Detector linearity Inject 5 standards (caffeine solution)
>1.5 AU, 5%
RSD
Temperature accuracy
comparison with external measuring
device
± 1 °C
Temperature precision monitoring temperature over 20 min ± 0.25 °C
Auto sampler carry
over
Injection of blank solvent after large
concentration
< 0.5 %
Mobile phase
composition accuracy
Step gradients from 4 to 7 % B, step
heights relative to 100%, with acetone
tracer
± 1 %
9
 BASELINE NOISE AND DRIFT
• Water is pumped at a flow rate of 1ml/min
• UV signal is recorded at 254nm
• To calculate noise the measuring signal is split into 20 intervals for 1 min
each.
• For each interval chromeleon calculates a regression based on measured
values, using the method of least square, then determine maximum
distance of two data points above and below the line.
• Limit should be between <2 x 10-3 AU
10
11
Figure: Long-term noise measurement
Figure: Short-term noise measurement
For drift
• Chromeleon calculates a regression line from all data points with in a
based on the method of least square.
• The slope of the straight line is expressed in absorbance units/hrs can
be provide an estimation of detector drift.
• Limit should be between <5 x 10—5 AU.
12
Figure: Drift measurement
 DETECTOR LINEARITY
• A series of 5 traceable standards (caffeine solution of concentration about 0.00035 to
0.35mg/ml) are injected.
• The detector linearity is calculated by determining the peak area vs concentration.
• %RSD can also be calculated for checking the detector linearity.
• limit should be in between >1.5 AU, 5% RSD.
• Traceable caffeine standard is used to determine the wavelength accuracy.
• Caffeine is trapped in the flow cell and a programmable timetable is used to determine
the wavelength maxima (205nm) and minima (273nm).
• The wavelength accuracy is determined as the absolute difference between the
measured and certified wavelength values.
 WAVELENGTH ACCURACY
13
 TEMPERATURE ACCURACY
• Measuring points are used to check the temperature accuracy of the column
compartment.
• The check is performed with column oven sequence.
• The achieved temperature is measured with external calibrated
thermometer.
• The achieved temperatures are compared to the set values.
• The difference indicates the temperature accuracy and the limit should be in
between ± 1ºc
 TEMPERATURE PRECISION
• Monitor temperature for 20 minutes and limit should be in between ±0.25 º
14
 AUTO SAMPLER CARRY OVER
• After a highly concentrated sample, a sample containing only solvent is
injected.
• Ideally only the signal for the solvent is displayed in the chromatogram.
• However, if a signal for the sample is displayed, this indicates the carry
over by the autosampler.
• It Should be less than 0.5%
15
 GRADIENT MOBILE PHASE COMPOSITION ACCURACY
• It is important for accurate quantitative analysis.
• Acetone tracer is used to determine gradient mobile phase accuracy,
stability and linearity.
• Make 6 compositions of methanol and acetone or caffeine in concentration
of 0%, 20%, 40%, 60%, 80% and 100% (20% increment).
• Linear ramp down from 100% to 0% is performed where the composition
linearity is determined between ranges of 95, 75 and 25%.
• All compositions accuracies are calculated as the absolute difference
between the mean composition at each set point and the theoretical
composition.
16
17
Solnent channel A – Methanol
Solvent channel B - Caffeine
QUALIFICATION OF LC-MS SYSTEM
18
The day-to-day performance of LC-MS or LC-MS/MS depends on its
calibration, tuning, system suitability test, and final overall validation.
Calibration Parameters
 Calibration parameters are instrument parameters whose values do not vary
with the type of experiment such as,
1. Peak width
2. Peak shape
3. Mass assignment
4. Resolution versus sensitivity
19
1. Peak width
• Peak width depends on the mass resolution.
• A resolution of 1 mass unit is sufficient to distinguish ions in most
qualitative/quantitative small molecule applications.
• Unit resolution → when the peak width at half-height is about 0.6 to 0.8 mass
unit.
• The profile scan of ions on a typical benchtop LC-MS has a bandwidth of about 1
mass unit (Figure 1).
Figure 1: Band width in a typical benchtop LC-MS 20
2. Peak Shape and Profile Scan
• In a typical benchtop LC-MS, abundance measurements are collected at 0.10-
m/z increments.
• When these data are presented in a mass spectrum, a single line can be shown.
• The height and position are derived from the profile scan.
21
3. Mass Assignment
• It is performed using specific MS calibrants.
• Calibrants should be well-characterized reference materials.
• LC-MS qualitative analysis → Use reference materials
• LC-MS quantitative analysis → Use a combination of reference materials
and actual analyte standards as calibrants.
4. Resolution versus Sensitivity
• Mass resolution is a compromise between ion intensity and peak width.
• In general, as the resolution is increased, the ion intensity decreases.
22
Calibration
• Comparison of a measurement standard or instrument of known accuracy
with another standard or instrument to eliminate deviations by adjustment.
• Manual, semiautomatic, and automatic LC-MS calibration require
introduction of the solution of the calibrant (calibration solution) into the
MS at a steady rate while the procedure is running.
• Calibration solution is introduced directly (infused) into the MS from a
syringe pump or through a loop injector connected to the LC pump.
• MS must be calibrated at least once every three months.
23
Tuning Parameters
• Tuning parameters are instrument parameters whose values can vary with the type of
experiment.
• Manual, semiautomatic, and automatic tuning procedures require the introduction of a
tuning solution of the analyte of interest into the MS at a steady rate. This can be done
in following three different ways:
1. By introducing the solution directly from the syringe pump (direct infusion)
→ This method is good for tuning for experiments at a low flow rate.
2. By introducing the sample from the syringe pump into the effluent of the LC
by using a tee union
3. By injecting the sample into the effluent of the LC by using a loop injection
valve [flow injection analysis (FIA)]
→ second and third methods are useful for experiments at a higher flow rate.
24
• Optimized parameter which affect the signal quality, change from
instrument model to instrument model & from brand to brand.
• Examples are Source, temperature, ionization voltages, gases (nebulization,
desolvation, and collision), ion path potentials (lens, multipoles, or stacked
rings), collision energy, solution, or mobile-phase flow rate.
• The potentials, RF values, and gas pressure affect the declustering,
focusing, fragmentation, and efficiency of ion transmission.
25
System Suitability Testing
• Tests allows the determination of system performance by analysis of a
defined solution prior to running the analytical batch.
• System suitability should test the
- Entire analytical system
- Chromatographic performance
- Sensitivity of the mass spectrometer for the compounds of interest.
26
Validation
• Final step to guarantee that a LC-MS system performs as expected.
• Validation includes instrument calibration, tuning, testing, and checking of the
documentation for completeness, correctness, and compliance with SOPs.
• Validation consists of four separate steps:
1. Validation of the instrument and the computer controlling it (computer system
validation or CSV)
2. Validation of the analytical method running on that equipment
3. System suitability testing, to test the equipment and the method together to
confirm expected performance
4. QA/QC review of sample analysis data collected on such a system
27
MS CALIBRATION PARAMETER
• This is the window used by the calibration software to search for the most
intense peak.
• Increasing this window gives a greater chance of incorrect peak matching.
• It is important to ensure that the correct peak is located in the peak search
range; otherwise, a deviation in the calibration may arise.
• If an incorrect peak is located, the search peak range should be adjusted to
locate only the correct ion.
28
1. Peak search range
2. Peak threshold
• All peaks in the acquired spectrum below the intensity threshold
value (measured usually as a percentage of the most intense peak in
the spectrum) will not be used in the calibration procedure.
3. Peak Width
• This is a measure used to set the resolution, usually specified at 50%
of maximum intensity.
29
Calibration solution
• Calibrants are required to calibrate the mass scale of any mass
spectrometer.
• Calibrants commonly used in electron ionization (EI) and chemical
ionization (CI), such as perfluorocarbons, are not applicable in the ESI
mode.
• The main calibrants used or still in use to calibrate ESI-MS can be divided
into the following categories:
polymers, perfluoroalkyl triazines, proteins, alkali metal salt clusters, poly
ethers, water clusters, and acetate salts.
30
1. Polymers
• polypropylene glycols (PPGs) and polyethylene glycols (PEGs) are
the preferred calibrant for many small molecule applications.
• PPG calibration solutions produce mostly singly charged ions over the
entire instrument mass range in both positive and negative ion mode.
• The PPG ions in general used for calibration in positive mode are: 59.0
(solvent and fragment ion), 175.1 (fragment ion), 616.5, 906.7, 1254.9,
1545.1, 2010.5, and 2242.6
31
32
2. Perfluoroalkyl Triazines
• Perfluoroalkyl triazines such as Ultramark 1621 have mainly been used
as the calibrant for FAB and ESI-MS calibration.
• This standard is very sticky and is very difficult to remove from the ion
source. For this reason, Ultramark 1621 calibration solution should not be
used at flow rates above 10 µL/min, to avoid system contamination.
• Ultramark 1621 is in general used together with other calibrants to cover
the entire MS range.
• Calibration solutions containing caffeine, L-methionyl-arginyol-
phenylalanylalanine acetate H2O (MRFA) and Ultramark 1621 are
commonly employed for ESI-MS calibration.
33
34
caffeine: m/z 195; MRFA: m/z 524; and Ultramark 1621: m/z 1022, 1122, 1222, 1322,
1422, 1522, 1622, 1722, 1822.
3. Proteins
• ESI-MS calibration was initially performed using solutions of gramicidin
S, cytochrome c, or myoglobin as calibrant.
• Proteins produce multiply charged species in ESI.
• When proteins are used as calibrants for low-resolution LC-MS and LC-
MS/MS instruments, it is not possible to resolve individual isotope peaks.
35
4. Alkali Metal Salt Clusters
• It cover a wide m/z range and are used to calibrate mass spectrometers in
ESI mode.
• Cesium iodide solutions produce singly and doubly charged species from
m/z 133 up to m/z 3510 or higher.
• Cesium iodide calibration solutions are not very commonly used due to
the following drawbacks:
1. sample suppression
2. persistence in the ESI source
3. possible cation attachment
4. The large spacing 260 amu between peaks.
36
• A mixture of sodium iodide and rubidium iodide calibration solution is
able to cover the instrument’s full mass range from 23 to 3920.
• The peak at 23 is sodium, the 85 peak is rubidium, and the others are
clusters.
37
5. Polyethers
• Polyethers such as polyethylene oxide (PEO) and polypropylene oxide
(PPO) have been used for ESI-MS calibration.
• sodium attachment is frequently observed, due to traces of sodium in
solvents and glassware.
• Macrocyclic polyethers are also used as ESI-MS calibration.
• Derivatized polyethers such as polyether sulfate have been investigated for
both positive and negative-ion calibration.
• Lauryl sulfate ethoxy-lates were also used as calibrants for negative-ion
ESI.
• Polyether amines and quaternary ammonium salts were used as positive-
ion calibration solutions.
38
39
6. Acetate salts
• Sodium acetate and sodium trifluoracetate clusters were used
and produce useful reference peaks for both positive and negative
ESI.
• 0.5% acetic acid in ammonium acetate solutions can be used for
calibration in ESI-MS.
Reference
• Chung Chow Chan et al, Analytical method validation and instrument
performance verification, 2004, page no.173-184 and 197-216.
• Dr. Ludwig Huber qualification of high performance liquid
chromatography system, Bio pharm journal, vol 11 page number 1 to 9.
40
T H A N K Y O U
41

More Related Content

Similar to Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj

INSTRUMENTAL ANALYSIS INTRODUCTION
INSTRUMENTAL ANALYSIS INTRODUCTIONINSTRUMENTAL ANALYSIS INTRODUCTION
INSTRUMENTAL ANALYSIS INTRODUCTIONHamunyare Ndwabe
 
HPLC APPLICATION RECENT DEVELOPMENT STATIONARY PHASE MATERIAL Dr suman presen...
HPLC APPLICATION RECENT DEVELOPMENT STATIONARY PHASE MATERIAL Dr suman presen...HPLC APPLICATION RECENT DEVELOPMENT STATIONARY PHASE MATERIAL Dr suman presen...
HPLC APPLICATION RECENT DEVELOPMENT STATIONARY PHASE MATERIAL Dr suman presen...Dr. Suman Pattanayak
 
analytical method validation and validation of hplc
analytical method validation and validation of hplcanalytical method validation and validation of hplc
analytical method validation and validation of hplcvenkatesh thota
 
Item 2. Verification and Validation of Analytical Methods
Item 2. Verification and Validation of Analytical MethodsItem 2. Verification and Validation of Analytical Methods
Item 2. Verification and Validation of Analytical MethodsSoils FAO-GSP
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validationSai Praveen Reddy
 
Dissolution-Method validation _ppt_slide
Dissolution-Method validation _ppt_slideDissolution-Method validation _ppt_slide
Dissolution-Method validation _ppt_slideBhanu Prakash N
 
Calibaration and validation of hplc
Calibaration and validation of hplcCalibaration and validation of hplc
Calibaration and validation of hplcANANT NAG
 
System suitability testing
System suitability testingSystem suitability testing
System suitability testingVrushali Tambe
 
Quantitative and Qualitative analysis of HPLC and GC
Quantitative and Qualitative analysis of HPLC and GCQuantitative and Qualitative analysis of HPLC and GC
Quantitative and Qualitative analysis of HPLC and GCDr. Nimra Khan
 
Validation of Analytical method.ppt
Validation of Analytical method.pptValidation of Analytical method.ppt
Validation of Analytical method.pptPriyanka Yadav
 
Method-Validation-HPLC-case-study
Method-Validation-HPLC-case-studyMethod-Validation-HPLC-case-study
Method-Validation-HPLC-case-studyShreekant Deshpande
 
Good chromatography practices - requisites and guidelines
Good chromatography practices - requisites and guidelinesGood chromatography practices - requisites and guidelines
Good chromatography practices - requisites and guidelinessuprajakotam
 
Calibration and validation of analytical instruments
Calibration and validation of analytical instrumentsCalibration and validation of analytical instruments
Calibration and validation of analytical instrumentsSolairajan A
 
Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Mostafa Mahmoud
 
Quality of Analytical Procedures
Quality of Analytical ProceduresQuality of Analytical Procedures
Quality of Analytical ProceduresBrendon Naicker
 
Qualification of instruments
Qualification of instrumentsQualification of instruments
Qualification of instrumentsNavdha Soni
 

Similar to Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj (20)

Qualification of high performance liquid chromatography
Qualification of high performance liquid chromatographyQualification of high performance liquid chromatography
Qualification of high performance liquid chromatography
 
INSTRUMENTAL ANALYSIS INTRODUCTION
INSTRUMENTAL ANALYSIS INTRODUCTIONINSTRUMENTAL ANALYSIS INTRODUCTION
INSTRUMENTAL ANALYSIS INTRODUCTION
 
HPLC APPLICATION RECENT DEVELOPMENT STATIONARY PHASE MATERIAL Dr suman presen...
HPLC APPLICATION RECENT DEVELOPMENT STATIONARY PHASE MATERIAL Dr suman presen...HPLC APPLICATION RECENT DEVELOPMENT STATIONARY PHASE MATERIAL Dr suman presen...
HPLC APPLICATION RECENT DEVELOPMENT STATIONARY PHASE MATERIAL Dr suman presen...
 
analytical method validation and validation of hplc
analytical method validation and validation of hplcanalytical method validation and validation of hplc
analytical method validation and validation of hplc
 
Item 2. Verification and Validation of Analytical Methods
Item 2. Verification and Validation of Analytical MethodsItem 2. Verification and Validation of Analytical Methods
Item 2. Verification and Validation of Analytical Methods
 
Analytical method validation
Analytical method validationAnalytical method validation
Analytical method validation
 
HPLC validation.ppt
HPLC validation.pptHPLC validation.ppt
HPLC validation.ppt
 
Dissolution-Method validation _ppt_slide
Dissolution-Method validation _ppt_slideDissolution-Method validation _ppt_slide
Dissolution-Method validation _ppt_slide
 
Calibaration and validation of hplc
Calibaration and validation of hplcCalibaration and validation of hplc
Calibaration and validation of hplc
 
Validation my ppt
Validation my pptValidation my ppt
Validation my ppt
 
System suitability testing
System suitability testingSystem suitability testing
System suitability testing
 
Quantitative and Qualitative analysis of HPLC and GC
Quantitative and Qualitative analysis of HPLC and GCQuantitative and Qualitative analysis of HPLC and GC
Quantitative and Qualitative analysis of HPLC and GC
 
Validation of Analytical method.ppt
Validation of Analytical method.pptValidation of Analytical method.ppt
Validation of Analytical method.ppt
 
Method-Validation-HPLC-case-study
Method-Validation-HPLC-case-studyMethod-Validation-HPLC-case-study
Method-Validation-HPLC-case-study
 
Good chromatography practices - requisites and guidelines
Good chromatography practices - requisites and guidelinesGood chromatography practices - requisites and guidelines
Good chromatography practices - requisites and guidelines
 
Calibration and validation of analytical instruments
Calibration and validation of analytical instrumentsCalibration and validation of analytical instruments
Calibration and validation of analytical instruments
 
Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods
 
Chap3 1
Chap3 1Chap3 1
Chap3 1
 
Quality of Analytical Procedures
Quality of Analytical ProceduresQuality of Analytical Procedures
Quality of Analytical Procedures
 
Qualification of instruments
Qualification of instrumentsQualification of instruments
Qualification of instruments
 

More from Pratik434909

sem 2 presentationnhh.pptxgggffufjffudjdjd
sem 2 presentationnhh.pptxgggffufjffudjdjdsem 2 presentationnhh.pptxgggffufjffudjdjd
sem 2 presentationnhh.pptxgggffufjffudjdjdPratik434909
 
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxQualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxPratik434909
 
tab-220603133234-655fd96e.pdf
tab-220603133234-655fd96e.pdftab-220603133234-655fd96e.pdf
tab-220603133234-655fd96e.pdfPratik434909
 
Mpharm pratik....pptx
Mpharm pratik....pptxMpharm pratik....pptx
Mpharm pratik....pptxPratik434909
 
Mpharm pratik.pptx
Mpharm pratik.pptxMpharm pratik.pptx
Mpharm pratik.pptxPratik434909
 
PILOT PLANT pratik.pptx
PILOT PLANT pratik.pptxPILOT PLANT pratik.pptx
PILOT PLANT pratik.pptxPratik434909
 
pratik new..........pptx
pratik new..........pptxpratik new..........pptx
pratik new..........pptxPratik434909
 

More from Pratik434909 (8)

sem 2 presentationnhh.pptxgggffufjffudjdjd
sem 2 presentationnhh.pptxgggffufjffudjdjdsem 2 presentationnhh.pptxgggffufjffudjdjd
sem 2 presentationnhh.pptxgggffufjffudjdjd
 
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxQualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
 
tab-220603133234-655fd96e.pdf
tab-220603133234-655fd96e.pdftab-220603133234-655fd96e.pdf
tab-220603133234-655fd96e.pdf
 
Mpharm pratik....pptx
Mpharm pratik....pptxMpharm pratik....pptx
Mpharm pratik....pptx
 
Mpharm pratik.pptx
Mpharm pratik.pptxMpharm pratik.pptx
Mpharm pratik.pptx
 
PILOT PLANT pratik.pptx
PILOT PLANT pratik.pptxPILOT PLANT pratik.pptx
PILOT PLANT pratik.pptx
 
pratik qms.pptx
pratik qms.pptxpratik qms.pptx
pratik qms.pptx
 
pratik new..........pptx
pratik new..........pptxpratik new..........pptx
pratik new..........pptx
 

Recently uploaded

Chapter 19_DDA_TOD Policy_First Draft 2012.pdf
Chapter 19_DDA_TOD Policy_First Draft 2012.pdfChapter 19_DDA_TOD Policy_First Draft 2012.pdf
Chapter 19_DDA_TOD Policy_First Draft 2012.pdfParomita Roy
 
VIP Call Girls Service Mehdipatnam Hyderabad Call +91-8250192130
VIP Call Girls Service Mehdipatnam Hyderabad Call +91-8250192130VIP Call Girls Service Mehdipatnam Hyderabad Call +91-8250192130
VIP Call Girls Service Mehdipatnam Hyderabad Call +91-8250192130Suhani Kapoor
 
VIP Call Girl Amravati Aashi 8250192130 Independent Escort Service Amravati
VIP Call Girl Amravati Aashi 8250192130 Independent Escort Service AmravatiVIP Call Girl Amravati Aashi 8250192130 Independent Escort Service Amravati
VIP Call Girl Amravati Aashi 8250192130 Independent Escort Service AmravatiSuhani Kapoor
 
NATA 2024 SYLLABUS, full syllabus explained in detail
NATA 2024 SYLLABUS, full syllabus explained in detailNATA 2024 SYLLABUS, full syllabus explained in detail
NATA 2024 SYLLABUS, full syllabus explained in detailDesigntroIntroducing
 
Dubai Call Girls Pro Domain O525547819 Call Girls Dubai Doux
Dubai Call Girls Pro Domain O525547819 Call Girls Dubai DouxDubai Call Girls Pro Domain O525547819 Call Girls Dubai Doux
Dubai Call Girls Pro Domain O525547819 Call Girls Dubai Douxkojalkojal131
 
Abu Dhabi Call Girls O58993O4O2 Call Girls in Abu Dhabi`
Abu Dhabi Call Girls O58993O4O2 Call Girls in Abu Dhabi`Abu Dhabi Call Girls O58993O4O2 Call Girls in Abu Dhabi`
Abu Dhabi Call Girls O58993O4O2 Call Girls in Abu Dhabi`dajasot375
 
VIP College Call Girls Gorakhpur Bhavna 8250192130 Independent Escort Service...
VIP College Call Girls Gorakhpur Bhavna 8250192130 Independent Escort Service...VIP College Call Girls Gorakhpur Bhavna 8250192130 Independent Escort Service...
VIP College Call Girls Gorakhpur Bhavna 8250192130 Independent Escort Service...Suhani Kapoor
 
Cosumer Willingness to Pay for Sustainable Bricks
Cosumer Willingness to Pay for Sustainable BricksCosumer Willingness to Pay for Sustainable Bricks
Cosumer Willingness to Pay for Sustainable Bricksabhishekparmar618
 
Captivating Charm: Exploring Marseille's Hillside Villas with Our 3D Architec...
Captivating Charm: Exploring Marseille's Hillside Villas with Our 3D Architec...Captivating Charm: Exploring Marseille's Hillside Villas with Our 3D Architec...
Captivating Charm: Exploring Marseille's Hillside Villas with Our 3D Architec...Yantram Animation Studio Corporation
 
WAEC Carpentry and Joinery Past Questions
WAEC Carpentry and Joinery Past QuestionsWAEC Carpentry and Joinery Past Questions
WAEC Carpentry and Joinery Past QuestionsCharles Obaleagbon
 
NO1 Famous Amil Baba In Karachi Kala Jadu In Karachi Amil baba In Karachi Add...
NO1 Famous Amil Baba In Karachi Kala Jadu In Karachi Amil baba In Karachi Add...NO1 Famous Amil Baba In Karachi Kala Jadu In Karachi Amil baba In Karachi Add...
NO1 Famous Amil Baba In Karachi Kala Jadu In Karachi Amil baba In Karachi Add...Amil baba
 
VIP Kolkata Call Girl Gariahat 👉 8250192130 Available With Room
VIP Kolkata Call Girl Gariahat 👉 8250192130  Available With RoomVIP Kolkata Call Girl Gariahat 👉 8250192130  Available With Room
VIP Kolkata Call Girl Gariahat 👉 8250192130 Available With Roomdivyansh0kumar0
 
Kieran Salaria Graphic Design PDF Portfolio
Kieran Salaria Graphic Design PDF PortfolioKieran Salaria Graphic Design PDF Portfolio
Kieran Salaria Graphic Design PDF Portfolioktksalaria
 
PORTAFOLIO 2024_ ANASTASIYA KUDINOVA
PORTAFOLIO   2024_  ANASTASIYA  KUDINOVAPORTAFOLIO   2024_  ANASTASIYA  KUDINOVA
PORTAFOLIO 2024_ ANASTASIYA KUDINOVAAnastasiya Kudinova
 
shot list for my tv series two steps back
shot list for my tv series two steps backshot list for my tv series two steps back
shot list for my tv series two steps back17lcow074
 
VIP Call Girls Service Bhagyanagar Hyderabad Call +91-8250192130
VIP Call Girls Service Bhagyanagar Hyderabad Call +91-8250192130VIP Call Girls Service Bhagyanagar Hyderabad Call +91-8250192130
VIP Call Girls Service Bhagyanagar Hyderabad Call +91-8250192130Suhani Kapoor
 
How to Be Famous in your Field just visit our Site
How to Be Famous in your Field just visit our SiteHow to Be Famous in your Field just visit our Site
How to Be Famous in your Field just visit our Sitegalleryaagency
 

Recently uploaded (20)

Chapter 19_DDA_TOD Policy_First Draft 2012.pdf
Chapter 19_DDA_TOD Policy_First Draft 2012.pdfChapter 19_DDA_TOD Policy_First Draft 2012.pdf
Chapter 19_DDA_TOD Policy_First Draft 2012.pdf
 
VIP Call Girls Service Mehdipatnam Hyderabad Call +91-8250192130
VIP Call Girls Service Mehdipatnam Hyderabad Call +91-8250192130VIP Call Girls Service Mehdipatnam Hyderabad Call +91-8250192130
VIP Call Girls Service Mehdipatnam Hyderabad Call +91-8250192130
 
VIP Call Girl Amravati Aashi 8250192130 Independent Escort Service Amravati
VIP Call Girl Amravati Aashi 8250192130 Independent Escort Service AmravatiVIP Call Girl Amravati Aashi 8250192130 Independent Escort Service Amravati
VIP Call Girl Amravati Aashi 8250192130 Independent Escort Service Amravati
 
NATA 2024 SYLLABUS, full syllabus explained in detail
NATA 2024 SYLLABUS, full syllabus explained in detailNATA 2024 SYLLABUS, full syllabus explained in detail
NATA 2024 SYLLABUS, full syllabus explained in detail
 
Dubai Call Girls Pro Domain O525547819 Call Girls Dubai Doux
Dubai Call Girls Pro Domain O525547819 Call Girls Dubai DouxDubai Call Girls Pro Domain O525547819 Call Girls Dubai Doux
Dubai Call Girls Pro Domain O525547819 Call Girls Dubai Doux
 
young call girls in Pandav nagar 🔝 9953056974 🔝 Delhi escort Service
young call girls in Pandav nagar 🔝 9953056974 🔝 Delhi escort Serviceyoung call girls in Pandav nagar 🔝 9953056974 🔝 Delhi escort Service
young call girls in Pandav nagar 🔝 9953056974 🔝 Delhi escort Service
 
Abu Dhabi Call Girls O58993O4O2 Call Girls in Abu Dhabi`
Abu Dhabi Call Girls O58993O4O2 Call Girls in Abu Dhabi`Abu Dhabi Call Girls O58993O4O2 Call Girls in Abu Dhabi`
Abu Dhabi Call Girls O58993O4O2 Call Girls in Abu Dhabi`
 
VIP College Call Girls Gorakhpur Bhavna 8250192130 Independent Escort Service...
VIP College Call Girls Gorakhpur Bhavna 8250192130 Independent Escort Service...VIP College Call Girls Gorakhpur Bhavna 8250192130 Independent Escort Service...
VIP College Call Girls Gorakhpur Bhavna 8250192130 Independent Escort Service...
 
Cheap Rate Call girls Kalkaji 9205541914 shot 1500 night
Cheap Rate Call girls Kalkaji 9205541914 shot 1500 nightCheap Rate Call girls Kalkaji 9205541914 shot 1500 night
Cheap Rate Call girls Kalkaji 9205541914 shot 1500 night
 
Cosumer Willingness to Pay for Sustainable Bricks
Cosumer Willingness to Pay for Sustainable BricksCosumer Willingness to Pay for Sustainable Bricks
Cosumer Willingness to Pay for Sustainable Bricks
 
Captivating Charm: Exploring Marseille's Hillside Villas with Our 3D Architec...
Captivating Charm: Exploring Marseille's Hillside Villas with Our 3D Architec...Captivating Charm: Exploring Marseille's Hillside Villas with Our 3D Architec...
Captivating Charm: Exploring Marseille's Hillside Villas with Our 3D Architec...
 
WAEC Carpentry and Joinery Past Questions
WAEC Carpentry and Joinery Past QuestionsWAEC Carpentry and Joinery Past Questions
WAEC Carpentry and Joinery Past Questions
 
NO1 Famous Amil Baba In Karachi Kala Jadu In Karachi Amil baba In Karachi Add...
NO1 Famous Amil Baba In Karachi Kala Jadu In Karachi Amil baba In Karachi Add...NO1 Famous Amil Baba In Karachi Kala Jadu In Karachi Amil baba In Karachi Add...
NO1 Famous Amil Baba In Karachi Kala Jadu In Karachi Amil baba In Karachi Add...
 
VIP Kolkata Call Girl Gariahat 👉 8250192130 Available With Room
VIP Kolkata Call Girl Gariahat 👉 8250192130  Available With RoomVIP Kolkata Call Girl Gariahat 👉 8250192130  Available With Room
VIP Kolkata Call Girl Gariahat 👉 8250192130 Available With Room
 
Kieran Salaria Graphic Design PDF Portfolio
Kieran Salaria Graphic Design PDF PortfolioKieran Salaria Graphic Design PDF Portfolio
Kieran Salaria Graphic Design PDF Portfolio
 
young call girls in Vivek Vihar🔝 9953056974 🔝 Delhi escort Service
young call girls in Vivek Vihar🔝 9953056974 🔝 Delhi escort Serviceyoung call girls in Vivek Vihar🔝 9953056974 🔝 Delhi escort Service
young call girls in Vivek Vihar🔝 9953056974 🔝 Delhi escort Service
 
PORTAFOLIO 2024_ ANASTASIYA KUDINOVA
PORTAFOLIO   2024_  ANASTASIYA  KUDINOVAPORTAFOLIO   2024_  ANASTASIYA  KUDINOVA
PORTAFOLIO 2024_ ANASTASIYA KUDINOVA
 
shot list for my tv series two steps back
shot list for my tv series two steps backshot list for my tv series two steps back
shot list for my tv series two steps back
 
VIP Call Girls Service Bhagyanagar Hyderabad Call +91-8250192130
VIP Call Girls Service Bhagyanagar Hyderabad Call +91-8250192130VIP Call Girls Service Bhagyanagar Hyderabad Call +91-8250192130
VIP Call Girls Service Bhagyanagar Hyderabad Call +91-8250192130
 
How to Be Famous in your Field just visit our Site
How to Be Famous in your Field just visit our SiteHow to Be Famous in your Field just visit our Site
How to Be Famous in your Field just visit our Site
 

Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj

  • 1. Dr. Rajnikant B. Mardia Senior Assistant Professor Department of Pharmaceutical Chemistry Kamlesh Jivani M.Pharm sem – II ID No. 21MQPOS005 Dept. Pharmaceutical Quality assurance Faculty of Pharmacy, DDU, Nadiad Guided by, Prepared by, QUALIFICATION OF HPLC SYSTEM AND LC-MS SYSTEM
  • 3. Introduction Equipment qualification is a formal process that provides documented evidence that an instrument is fit for its intended use. The entire qualification process consists of four parts:  Design qualification (DQ)  Installation qualification (IQ)  Operational qualification (OQ)  Performance qualification (PQ) 3
  • 4. Design qualification (DQ) • Design qualification (DQ) describes the user requirements and defines the functional and operational specifications of the instrument. Design elements Examples Intended use Analysis of drug compounds and impurities User requirement specification for the HPLC analysis  Analysis up to 100 samples / day  Automated over-night analysis  Limit of quantitation: 0.1%  Automated confirmation of peak identity and purity with diode-array detection  Automated compound quantitation and printing of report 4
  • 5. Design elements Examples Functional specifications  Autosampler - 100 samples, 0.5 µl to 5 ml sample volume  Column compartment - 15 to 60˚C, peltier controlled  Detector - UV/Vis Diode-array, 190-900 nm  Computer - System control, data acquisition for signals and spectra, peak integration and quantitation, spectral evaluation for peak purity and compound confirmation. Operational specifications Detector: Baseline noise < 5 X 10-5 AU Sampler: Precision injection Volume sample carry over: <0.5% Maintenance Vendor must deliver maintenance procedure and recommend schedule Training Vendor must provide familiarization and training 5
  • 6. Installation qualification (IQ) Before installation  Check the site for the fulfillment of the manufacturer’s recommendations (utilities such as electricity, and environmental conditions such as humidity and temperature).  Allow sufficient shelf space for the equipment, SOPs, operating manuals and software. During installation  Compare equipment as received, with purchase order (including software, accessories, spare parts)  Check documentation for completeness (operating manuals, maintenance instructions, standard operating procedures for testing, safety and validation certificates) 6
  • 7.  Check equipment for any damage  Install hardware (computer, equipment, fittings and tubings for fluid connections, columns in HPLC, power cables, and instrument control cables)  Switch on the instruments and ensure that all modules power up and perform an electronic self-test.  Run test sample and compare chromatogram with reference chromatogram  Prepare an installation report Installation qualification (IQ) 7
  • 8. Performance qualification (PQ) Parameter Procedure User Limit Leak testing Flow test by volume or weight/time ± 5 % Baseline drift ASTM Method <2 × 10-3 AU Baseline noise ASTM Method <5 x 10-5 AU Precision of injection volume 6 x injection of caffeine standard, RSD of peak areas 0.3 % RSD Precision of flow rate 6 x injection of caffeine standard, RSD of retention times 0.5 % RSD 8 ASTM Method → American Society for Testing and Material Method
  • 9. Parameter Procedure User Limit Detector linearity Inject 5 standards (caffeine solution) >1.5 AU, 5% RSD Temperature accuracy comparison with external measuring device ± 1 °C Temperature precision monitoring temperature over 20 min ± 0.25 °C Auto sampler carry over Injection of blank solvent after large concentration < 0.5 % Mobile phase composition accuracy Step gradients from 4 to 7 % B, step heights relative to 100%, with acetone tracer ± 1 % 9
  • 10.  BASELINE NOISE AND DRIFT • Water is pumped at a flow rate of 1ml/min • UV signal is recorded at 254nm • To calculate noise the measuring signal is split into 20 intervals for 1 min each. • For each interval chromeleon calculates a regression based on measured values, using the method of least square, then determine maximum distance of two data points above and below the line. • Limit should be between <2 x 10-3 AU 10
  • 11. 11 Figure: Long-term noise measurement Figure: Short-term noise measurement
  • 12. For drift • Chromeleon calculates a regression line from all data points with in a based on the method of least square. • The slope of the straight line is expressed in absorbance units/hrs can be provide an estimation of detector drift. • Limit should be between <5 x 10—5 AU. 12 Figure: Drift measurement
  • 13.  DETECTOR LINEARITY • A series of 5 traceable standards (caffeine solution of concentration about 0.00035 to 0.35mg/ml) are injected. • The detector linearity is calculated by determining the peak area vs concentration. • %RSD can also be calculated for checking the detector linearity. • limit should be in between >1.5 AU, 5% RSD. • Traceable caffeine standard is used to determine the wavelength accuracy. • Caffeine is trapped in the flow cell and a programmable timetable is used to determine the wavelength maxima (205nm) and minima (273nm). • The wavelength accuracy is determined as the absolute difference between the measured and certified wavelength values.  WAVELENGTH ACCURACY 13
  • 14.  TEMPERATURE ACCURACY • Measuring points are used to check the temperature accuracy of the column compartment. • The check is performed with column oven sequence. • The achieved temperature is measured with external calibrated thermometer. • The achieved temperatures are compared to the set values. • The difference indicates the temperature accuracy and the limit should be in between ± 1ºc  TEMPERATURE PRECISION • Monitor temperature for 20 minutes and limit should be in between ±0.25 º 14
  • 15.  AUTO SAMPLER CARRY OVER • After a highly concentrated sample, a sample containing only solvent is injected. • Ideally only the signal for the solvent is displayed in the chromatogram. • However, if a signal for the sample is displayed, this indicates the carry over by the autosampler. • It Should be less than 0.5% 15
  • 16.  GRADIENT MOBILE PHASE COMPOSITION ACCURACY • It is important for accurate quantitative analysis. • Acetone tracer is used to determine gradient mobile phase accuracy, stability and linearity. • Make 6 compositions of methanol and acetone or caffeine in concentration of 0%, 20%, 40%, 60%, 80% and 100% (20% increment). • Linear ramp down from 100% to 0% is performed where the composition linearity is determined between ranges of 95, 75 and 25%. • All compositions accuracies are calculated as the absolute difference between the mean composition at each set point and the theoretical composition. 16
  • 17. 17 Solnent channel A – Methanol Solvent channel B - Caffeine
  • 19. The day-to-day performance of LC-MS or LC-MS/MS depends on its calibration, tuning, system suitability test, and final overall validation. Calibration Parameters  Calibration parameters are instrument parameters whose values do not vary with the type of experiment such as, 1. Peak width 2. Peak shape 3. Mass assignment 4. Resolution versus sensitivity 19
  • 20. 1. Peak width • Peak width depends on the mass resolution. • A resolution of 1 mass unit is sufficient to distinguish ions in most qualitative/quantitative small molecule applications. • Unit resolution → when the peak width at half-height is about 0.6 to 0.8 mass unit. • The profile scan of ions on a typical benchtop LC-MS has a bandwidth of about 1 mass unit (Figure 1). Figure 1: Band width in a typical benchtop LC-MS 20
  • 21. 2. Peak Shape and Profile Scan • In a typical benchtop LC-MS, abundance measurements are collected at 0.10- m/z increments. • When these data are presented in a mass spectrum, a single line can be shown. • The height and position are derived from the profile scan. 21
  • 22. 3. Mass Assignment • It is performed using specific MS calibrants. • Calibrants should be well-characterized reference materials. • LC-MS qualitative analysis → Use reference materials • LC-MS quantitative analysis → Use a combination of reference materials and actual analyte standards as calibrants. 4. Resolution versus Sensitivity • Mass resolution is a compromise between ion intensity and peak width. • In general, as the resolution is increased, the ion intensity decreases. 22
  • 23. Calibration • Comparison of a measurement standard or instrument of known accuracy with another standard or instrument to eliminate deviations by adjustment. • Manual, semiautomatic, and automatic LC-MS calibration require introduction of the solution of the calibrant (calibration solution) into the MS at a steady rate while the procedure is running. • Calibration solution is introduced directly (infused) into the MS from a syringe pump or through a loop injector connected to the LC pump. • MS must be calibrated at least once every three months. 23
  • 24. Tuning Parameters • Tuning parameters are instrument parameters whose values can vary with the type of experiment. • Manual, semiautomatic, and automatic tuning procedures require the introduction of a tuning solution of the analyte of interest into the MS at a steady rate. This can be done in following three different ways: 1. By introducing the solution directly from the syringe pump (direct infusion) → This method is good for tuning for experiments at a low flow rate. 2. By introducing the sample from the syringe pump into the effluent of the LC by using a tee union 3. By injecting the sample into the effluent of the LC by using a loop injection valve [flow injection analysis (FIA)] → second and third methods are useful for experiments at a higher flow rate. 24
  • 25. • Optimized parameter which affect the signal quality, change from instrument model to instrument model & from brand to brand. • Examples are Source, temperature, ionization voltages, gases (nebulization, desolvation, and collision), ion path potentials (lens, multipoles, or stacked rings), collision energy, solution, or mobile-phase flow rate. • The potentials, RF values, and gas pressure affect the declustering, focusing, fragmentation, and efficiency of ion transmission. 25
  • 26. System Suitability Testing • Tests allows the determination of system performance by analysis of a defined solution prior to running the analytical batch. • System suitability should test the - Entire analytical system - Chromatographic performance - Sensitivity of the mass spectrometer for the compounds of interest. 26
  • 27. Validation • Final step to guarantee that a LC-MS system performs as expected. • Validation includes instrument calibration, tuning, testing, and checking of the documentation for completeness, correctness, and compliance with SOPs. • Validation consists of four separate steps: 1. Validation of the instrument and the computer controlling it (computer system validation or CSV) 2. Validation of the analytical method running on that equipment 3. System suitability testing, to test the equipment and the method together to confirm expected performance 4. QA/QC review of sample analysis data collected on such a system 27
  • 28. MS CALIBRATION PARAMETER • This is the window used by the calibration software to search for the most intense peak. • Increasing this window gives a greater chance of incorrect peak matching. • It is important to ensure that the correct peak is located in the peak search range; otherwise, a deviation in the calibration may arise. • If an incorrect peak is located, the search peak range should be adjusted to locate only the correct ion. 28 1. Peak search range
  • 29. 2. Peak threshold • All peaks in the acquired spectrum below the intensity threshold value (measured usually as a percentage of the most intense peak in the spectrum) will not be used in the calibration procedure. 3. Peak Width • This is a measure used to set the resolution, usually specified at 50% of maximum intensity. 29
  • 30. Calibration solution • Calibrants are required to calibrate the mass scale of any mass spectrometer. • Calibrants commonly used in electron ionization (EI) and chemical ionization (CI), such as perfluorocarbons, are not applicable in the ESI mode. • The main calibrants used or still in use to calibrate ESI-MS can be divided into the following categories: polymers, perfluoroalkyl triazines, proteins, alkali metal salt clusters, poly ethers, water clusters, and acetate salts. 30
  • 31. 1. Polymers • polypropylene glycols (PPGs) and polyethylene glycols (PEGs) are the preferred calibrant for many small molecule applications. • PPG calibration solutions produce mostly singly charged ions over the entire instrument mass range in both positive and negative ion mode. • The PPG ions in general used for calibration in positive mode are: 59.0 (solvent and fragment ion), 175.1 (fragment ion), 616.5, 906.7, 1254.9, 1545.1, 2010.5, and 2242.6 31
  • 32. 32
  • 33. 2. Perfluoroalkyl Triazines • Perfluoroalkyl triazines such as Ultramark 1621 have mainly been used as the calibrant for FAB and ESI-MS calibration. • This standard is very sticky and is very difficult to remove from the ion source. For this reason, Ultramark 1621 calibration solution should not be used at flow rates above 10 µL/min, to avoid system contamination. • Ultramark 1621 is in general used together with other calibrants to cover the entire MS range. • Calibration solutions containing caffeine, L-methionyl-arginyol- phenylalanylalanine acetate H2O (MRFA) and Ultramark 1621 are commonly employed for ESI-MS calibration. 33
  • 34. 34 caffeine: m/z 195; MRFA: m/z 524; and Ultramark 1621: m/z 1022, 1122, 1222, 1322, 1422, 1522, 1622, 1722, 1822.
  • 35. 3. Proteins • ESI-MS calibration was initially performed using solutions of gramicidin S, cytochrome c, or myoglobin as calibrant. • Proteins produce multiply charged species in ESI. • When proteins are used as calibrants for low-resolution LC-MS and LC- MS/MS instruments, it is not possible to resolve individual isotope peaks. 35
  • 36. 4. Alkali Metal Salt Clusters • It cover a wide m/z range and are used to calibrate mass spectrometers in ESI mode. • Cesium iodide solutions produce singly and doubly charged species from m/z 133 up to m/z 3510 or higher. • Cesium iodide calibration solutions are not very commonly used due to the following drawbacks: 1. sample suppression 2. persistence in the ESI source 3. possible cation attachment 4. The large spacing 260 amu between peaks. 36
  • 37. • A mixture of sodium iodide and rubidium iodide calibration solution is able to cover the instrument’s full mass range from 23 to 3920. • The peak at 23 is sodium, the 85 peak is rubidium, and the others are clusters. 37
  • 38. 5. Polyethers • Polyethers such as polyethylene oxide (PEO) and polypropylene oxide (PPO) have been used for ESI-MS calibration. • sodium attachment is frequently observed, due to traces of sodium in solvents and glassware. • Macrocyclic polyethers are also used as ESI-MS calibration. • Derivatized polyethers such as polyether sulfate have been investigated for both positive and negative-ion calibration. • Lauryl sulfate ethoxy-lates were also used as calibrants for negative-ion ESI. • Polyether amines and quaternary ammonium salts were used as positive- ion calibration solutions. 38
  • 39. 39 6. Acetate salts • Sodium acetate and sodium trifluoracetate clusters were used and produce useful reference peaks for both positive and negative ESI. • 0.5% acetic acid in ammonium acetate solutions can be used for calibration in ESI-MS.
  • 40. Reference • Chung Chow Chan et al, Analytical method validation and instrument performance verification, 2004, page no.173-184 and 197-216. • Dr. Ludwig Huber qualification of high performance liquid chromatography system, Bio pharm journal, vol 11 page number 1 to 9. 40
  • 41. T H A N K Y O U 41