CT is one of the highest contributor for medical radiation exposure to patients. Some common CT dose descriptors and dose optimizations methods are briefly described in this presentation.
College Call Girls Pune Mira 9907093804 Short 1500 Night 6000 Best call girls...
CT radiation dose concepts and radiation dose optimization- Avinesh Shrestha
1.
2. Topics
Radiation Units
CT Dose Descriptors
Factors Affecting Dose
Strategies for reducing dose
9/21/2021 2
3. Concern
Because of increasing use of multislice helical CT, CT must be considered a high-dose
procedure. U.S. Public Health Service data suggest that 17% of all x-ray examinations are CT,
yet CT accounts or 50% of total patient effective dose.
9/21/2021 3
4. CT, 3%
NM, 26%
IF, 3%
CR, 68%
9/21/2021 4
CT, 49%
NM, 26%
IF, 14%
CR, 11%
Medical exposure of patients, early 1980’s Medical exposure of patients, 2006
6. RADIATION QUANTITIES AND THEIR UNITS:
Exposure
Exposure is a measure of the amount of ionization produced in a specific mass of air by x
rays or gamma radiation.
The unit of exposure is the Roentgen (R), the conventional unit, or the coulomb per
kilogram (C/kg), the SI unit.
One Roentgen produces 2.58 × 10−4 C/kg of air at standard temperature and pressure.
9/21/2021 6
7. Absorbed Dose
The absorbed dose, or radiation dose as it is popularly referred to, is the amount of energy
absorbed per unit mass of material (patient).
The old unit of absorbed dose is the rad (r), which is equal to an energy absorption of 100 erg/g of
absorber.
The SI unit of absorbed dose is the Gray (Gy), so named in honor of Louis Harold Gray, a British
radiobiologist who devised ways to measure the absorbed dose.
One Gray is equal to 1 joule per kilogram (J/kg)
1 r is equal to 0.01 Gy or 100 r is equal to 1 Gy. 9/21/2021 7
8. Equivalent Dose
In recognition of the health effects of x-rays, another conversion factor is applied to the absorbed
dose.
It is the product of the absorbed dose and a radiation weighting factor (wR).
The unit for the equivalent dose is also the rem or the Sv.
Equivalent Dose= Absorbed dose x Radiation weighting factor
100 rem = 1 Sv
9/21/2021 8
9. Effective dose
The quantity effective dose, E, is used to quantify the risk from partial-body exposure to that from
an equivalent whole-body dose.
It is used to take into account the type of radiation and the radio sensitivity of different tissues.
It extrapolates the risk of partial body exposure to patients from data obtained from whole body
doses to Japanese atomic bomb survivors. This unit is a weighted average of organ doses.
Old unit of effective dose is the rem, the SI unit is the Sievert (Sv), where 1 Sv = 100 rem.
The effective dose relates exposure to risk. As noted by Brenner (2006), “…relevant organ doses for
CT examinations are on the order of 15 mSv or less.” Brenner also notes that “…there is good
epidemiologic evidence of increased cancer risk for children exposed to acute doses of 10 mSv (or
more) and for adults exposed to acute doses of 50 mSv (or more).”
Effective dose = Absorbed dose x Radiation weighting factor x Tissue Weighting Factor
9/21/2021 9
11. Radiation Weighting Factor (WR)
Accounts for relative hazard from various forms of radiation
alpha = 20
beta = 1
gamma/x-ray = 1
neutron = 10
9/21/2021 11
12. Tissue Weighting Factor
The tissue weighting factor (WT) is a relative measure of the risk of stochastic effects that might result
from irradiation of that specific tissue. It accounts for the variable radio sensitivities of organs and
tissues in the body to ionizing radiation.
9/21/2021 12
2007
14. Dose Geometry
In conventional radiography, the skin of the entrance plane receives 100% of the radiation, and
the percentage falls rapidly as the x-ray beam is attenuated by the patient’s tissue.
In conventional radiography, it is common for the exit dose to be only 1% (or 1/100) that of the
entrance dose.
In CT, the difference between the dose at the center and the dose at the periphery is not nearly
as great as that of conventional radiography.
9/21/2021 14
15. Dose Geometry
Dose is more uniform in CT than in general radiography for two reasons.
The beam is heavily filtered as it exits the x-ray tube. This means that fewer low-energy photons
remain. Because the beam is “harder” than that used in conventional radiography, a lower
percentage of the beam will be absorbed or scattered as it passes through patient tissue.
CT exposure comes from all directions, creating a more uniform exposure.
The uniformity of the dose decreases as the scan field of view and patient thickness increase.
Therefore, body scans are less uniform than head scans.
The central dose for a body scan is approximately one-third to one-half that of the peripheral dose
however, organ doses are clearly higher for children compared with (larger) adults.
9/21/2021 15
18. Comparison of Dose From CT With Dose
From Conventional Radiographic Studies
9/21/2021 18
19. Z Axis Variations
In addition to the variations within the scan plane, variations along the length, or z axis, of the
patient, are described by the z axis dose distribution (or radiation profile)
In general, the tails will contribute approximately 25% to 40% additional dose to the entire study.
Therefore, if a single slice of the chest delivered a dose of 4 cGy, the entire dose from 30 contiguous
slices would range from approximately 5 cGy (i.e., 4 cGy + 1 cGy) to 5.6 (i.e., 4 cGy + 1.6 cGy). As
compared with the dose from a single slice, the overall radiation dose will be higher when multiple
scans are performed.
So, total dose is the central slice radiation dose, plus the scatter overlap (or tails). This is called the
multiple scan average dose (MSAD).
Another type of radiation dose measurement in CT is the computed tomography dose index (CTDI).
This allows an estimate of the MSAD to be accomplished with a single scan.
9/21/2021 19
21. Phantoms for CT Dose Measurement
There are two standard polymethyl methacrylate (PMMA) dosimetry phantoms; the body phantom is
32 cm in diameter and 15 cm long, and the head phantom is 16 cm in diameter and 15 cm long.
The head phantom also serves as a pediatric torso phantom.
9/21/2021 21
22. Phantoms for CT Dose Measurement
9/21/2021 22
The computed tomography dose index (CTDI) is measured using either a 16-cm or 32-cm
diameter polymethyl methacrylate (PMMA) phantom.
27. Multiple Scan Average Dose (MSAD)
The average cumulative dose to the central slices from a series of slices with constant spacing is
referred to as the multiple-slice average dose.
The MSAD represents the dose to a specific section location resulting from the scan at that location as
well as contributions from adjacent scan locations since the penumbra from adjacent sections may
contribute to the dose received by the section of interest.
The MSAD may be 1.25 – 1.4 times (25 to 40% additional dose to the entire study) the single slice dose
depending upon scanner geometry, collimator design, slice spacing and the position of the slice within
the series of slices 9/21/2021 27
28. MSAD
9/21/2021 28
A series of seven scans spaced (bed indexing) apart along
the z-axis gives seven bell shaped dose distribution
curves (top). When the doses from this series are
summed, the resultant total dose appears as the bottom
curve. The total dose curve has peaks where the bell-
shaped curves overlap. The dotted line through the total
dose curve is the multiple scan average dose.
29. Computed Tomography Dose Index
CTDI100
The CTDI was originally designed as an index, not as a direct dosimetry method for patient dose
assessment.
CT Dose Index is the dose a slice gets from the complete series of contiguous slices
The CTDI is what manufacturers report to the U.S. Food and Drug Administration (FDA) and
prospective customers regarding the doses typically delivered for their machines.
The basic CTDI100 measurement involves the use of a 100-mm-long cylindrical (“pencil”) chamber,
approximately 9 mm in diameter, inserted into either the center or a peripheral hole of a PMMA
phantom.
9/21/2021 29
30. CTDI100
The CTDI100 describes the measurement of the dose distribution, D(z), along the z-axis, from a single
circular rotation of the scanner with a nominal x-ray beam width of nT, where N is the number of
simultaneously acquired slices of thickness T, where the primary and scattered radiation are measured
over a 100-mm length, and where center of the x-ray beam is at z = 0.
The 100-mm-long thin cylindrical device is long enough to span the width of 14 contiguous 7-mm CT
slices. This provides a better estimate of MSAD for thin slices than that of the single-slice method.
When this method is used it is referred to as the CTDI100.
9/21/2021 30
31. 9/21/2021 31
Dose that scanning of slice 1 gives to all slices
=
Dose that slice 1 gets from scanning of all slices
(A) Dose to slice 1 from scanning of slice 1. (B)
Dose to slice 1 from scanning of slice 2 equals
dose to slice 2 from scanning of slice 1. (C) Dose
to slice 1 from scanning of slice 3 equals dose to
slice 3 from scanning of slice 1. (D) Dose to slice
1 from scanning of all slices equals dose to all
slices from scanning of slice 1, or total area under
dose profile, which is measured using long
ionization chamber.
32. CTDIw
The dose for body scans are not uniform across the scan field of view—the dose at the periphery of
the slice is higher than the central dose. The CTDIW adjusts for this by providing a weighted average
of measurements at center and the peripheral slice locations (i.e., the x and y dimensions of the
slice).
CTDI100 measurements are made for both the center CTDI100,center) and periphery
(CTDI100,periphery). Combining the center and peripheral measurements using a 1/3 and 2/3
weighting scheme provides a good estimate of the average dose to the phantom, giving rise to the
weighted CTDI, CTDI
CTDIW = (1/3) (CTDI100)center+ (2/3) (CTDI100)periphery
9/21/2021 32
33. CTDIVOL
The CTDIvol radiation dose parameter takes the process a step further by taking account the exposure
variation in the z direction.
In helical CT scanning, the CT dose is inversely related to the helical pitch used, that is, dose ∝
1/pitch
Because of this relationship, the CTDIW is converted to the volume CTDI (CTDIVOL) using,
CTDIVOL = CTDIW/pitch
For helical sequences the CTDIVOL is now the preferred expression of radiation dose in CT dosimetry.
9/21/2021 33
35. Limitations of CTDIVOL
CTDIvol is a dose index and it’s concepts were meant to enable medical physicists to compare the
output between different CT scanners and were not originally intended to provide patient-specific
dosimetry information.
The dose received by a patient from a CT scan is dependent on both patient size and scanner radiation
output.
CTDIVOL provides information regarding only the scanner output. It does not address patient size, and
hence does not estimate patient dose.
9/21/2021 35
36. Limitations of CTDIVOL
The 32-cm-diameter PMMA body phantom corresponds to a person with a 119 cm (47”) waistline—a
large individual, indeed.
For smaller patients, the actual doses are larger than the CTDIvol for the same technique factors—thus,
the CTDIvol tends to underestimate dose to most patients.
For this limitation, researchers have shown that patient size conversion factors can be used with the
CTDIVOL reported on a given CT scanner to produce size specific dose estimates (SSDEs).
9/21/2021 36
37. Limitations of CTDI
9/21/2021 37
Both patients scanned with the same CTDIvol
Patient dose will be higher for the smaller patient
CTDIvol = 20 mGy CTDIvol = 20 mGy
120 kVp at 200 mAs 120 kVp at 200 mAs
32 cm
Phantom
32 cm
Phantom
38. Limitations of CTDI
9/21/2021 38
Smaller patient scanned with a lower CTDIvol
Patient doses will be approximately equal
CTDIvol = 10 mGy CTDIvol = 20 mGy
120 kVp at 100 mAs 120 kVp at 200 mAs
32 cm
Phantom
32 cm
Phantom
39. Size Specific Dose Estimate (SSDE)
AAPM report 204, 2011 describes a method to calculate SSDE using CTDIvol
Conversion factors based on patient size (e.g., AP or lateral width, effective diameter) are provided to
estimate patient dose for a patient of that size
However, SSDE is still not the exact patient dose, as factors such as scan length and patient
composition may differ from the assumptions used to calculate SSDE
SSDE is not dose to any specific organ, but rather the mean dose in the center of the scanned volume
9/21/2021 39
40. Size Specific Dose Estimate (SSDE)
9/21/2021 40
This figure shows data taken from AAPM
Report 204. This curve shows the conversion
factors as a function of the effective diameter
of the patient, relative to the 32-cm-diameter
CTDIvol phantom. The data from this curve
can be used to adjust the displayed CTDIvol
when the effective diameter of the patient is
known.
43. Size Specific Dose Estimate (SSDE)
9/21/2021 43
32 cm
Phantom
32 cm
Phantom
Patients have equivalent SSDE
CTDIvol = 10 mGy
SSDE = 13.2 mGy
CTDIvol = 20 mGy
SSDE = 13.2 mGy
120 kVp at 100 mAs 120 kVp at 200 mAs
44. Why Use CTDIvol?
CTDIvol is a useful index to track across patients and protocols for quality assurance purposes
CTDIvol provides information about the amount of radiation used to perform the study
CTDIvol can be used as a metric to compare protocols across different practices and scanners when
related variables, such as resultant image quality, are also taken in account
The ACR Dose Index Registry (DIR) allows comparison across institutions of CTDIvol for similar exam
types (e.g., routine head exam)
9/21/2021 44
45. Dose length product (DLP)
CTDI in any form is an estimate of average radiation dose only in the irradiated volume for single
slice, so another indicator called Dose length product (DLP) is in use.
The product of CTDIVOL and the length of CT scan along the z-axis of the patient, L, is the dose
length product (DLP)
DLP = CTDIVol x L
46. Estimates of Effective Dose from DLP
It has been shown that the DLP is
approximately proportional to
effective dose (E).
The slope of the E versus DLP
relationship is called the k value, and
there are different k values for
various CT examinations, as given in
Table 11-5
Effective Dose= DLP x K
9/21/2021 46
48. Factors Affecting Dose
1. Radiation Beam Geometry
2. Filtration
3. Detector Efficiency
4. Slice Width and Spacing
5. Pitch
6. Scan Field Diameter
7. Exposure Technique
8. Patient Size and Body Part Thickness
9. Repeat Scans
10. Collimation
11. Localization Scans
12. Number of Detectors
9/21/2021 48
49. Radiation Beam Geometry
Theoretically, a rotation arc of only 180° is all that is required to satisfy most construction
algorithms. Most scanners use a 360° tube arc to compensate for radiation beam divergence and
patient motion.
The extra scanning information improves image quality but increases radiation dose.
Additionally, overscanning, which is the process of using more than a 360° tube arc, is
sometimes used
Overscanning is also necessary for interpolation of projections in MDCT. Overscans will increase
the radiation dose.
9/21/2021 49
50. Filtration
Filtration affects the radiation dose by removing some of the low-energy x-rays.
These low-energy x-rays are quickly absorbed by the patient.
Photons are needed to penetrate the patient and strike the detectors, but some of them must be
absorbed to produce radiographic contrast.
Adding metal filters to the beam permits selective removal of x-rays with low energy and reduces
the radiation dose while maintaining contrast at an acceptable level.
9/21/2021 50
51. Detector Efficiency
Detector absorption efficiency affects radiation dose to the patient.
Less-efficient detectors will require a higher radiation exposure to produce an adequate image.
Solid state detectors are 90% to 100% efficient, whereas the xenon gas detectors used in older
model scanners are significantly less efficient (60% to 87%) at absorbing x-rays.
9/21/2021 51
52. Slice Width and Spacing
Decreasing slice thickness and using contiguous slices will increase the dose because of the increased
amount of scatter radiation to adjacent slices.
Also, to maintain image quality at the same level, additional radiation is needed for thinner slices.
Multiple slice examinations using overlapping slices will produce a higher overall dose, whereas
gapped slices will produce a lower overall dose.
To have the same x-ray intensity reach all of the detectors in an MDCT system, the collimators must
be opened slightly more, allowing the x-ray penumbra to fall outside the active detectors, referred to
as overbeaming
This contributes to the patient dose but is not used to create an image. The effect is more
pronounced for smaller collimator openings. 9/21/2021 52
53. Slice Width and Spacing
9/21/2021 53
To have the same x-ray intensity at each of the
detectors in an MDCT scanner, the collimators
must be opened slightly more so that the x-ray
penumbra falls outside the detectors. This
contributes to the radiation dose to the patient
but dose not contribute to the image created.
54. Pitch
Helical scans performed with a pitch of 1 deliver approximately the same dose as that of
conventional axial CT studies—provided the kVp and mAs values are the same for each mode.
The pitch has a direct influence on patient radiation dose because as pitch increases, the time
that any one point in space spends in the x-ray beam is decreased.
It is generally expressed as 0.5, 1.0, 1.5, 2.0
pitch = 1 - coils of the helix (radiation beam) are in contact
pitch < 1 - coils of the helix (radiation beam) overlap. Increased patient dose
pitch > 1 - coils of the helix (radiation beam) are separated. Reduced patient dose
9/21/2021 54
57. Scan Field Diameter
Holding all technical factors constant, a scan of the head phantom will result in a higher radiation
dose than that of the body phantom.
The primary difference in results between the head and body phantoms is size.
The smaller object always absorbs the higher dose and that the difference is at least a factor of two.
Thus, smaller patients would be expected to absorb much higher amounts of radiation than larger
patients.
This effect is primarily attributed to the fact that total exposure is made up of both entrance
radiation and exit radiation.
For smaller patients, the patient has less tissue to attenuate the beam, which results in a much
more uniform dose distribution. Conversely, for a larger patient, the exit radiation is much less
intense as a result of its attenuation through more tissue.
9/21/2021 57
58. Exposure Technique
As in general radiography, the technique used to create the CT image affects radiation exposure to
the patient.
The higher the mAs and kVp settings used to create the image, the higher the dose to the patient.
The relationship between mAs and dose is linear.
Assuming that the minimum dose to obtain acceptable image quality has been determined. If this
dose is halved by halving the mAs, a noise increase of 41% can be expected.
Kvp also affects the radiation dose, although the effect is not linear. With the mAs kept constant,
changing from 120 kVp to 140 kVp increases the radiation dose approximately 30% to 45%.
9/21/2021 58
59. Effective Milliamperage-Seconds
The effective mAs is a term used for multislice CT scanners that denotes the mAs per slice. This is
given by the following relationship:
Effective mAs = True mAs/pitch
This expression simply implies that, to keep the effective mAs constant, as the pitch increases, the
true mAs must be increased as well
For example, increasing the pitch from, say, 1 to 2, increases the mAs per rotation from 100 to 200
while the relative CTDIvol remains the same
9/21/2021 59
60. Patient Size and Body Part Thickness
Large patients or thick body parts require radiographic techniques that increase the radiation
dose to avoid an unacceptable level of image noise.
In addition, the patient size and body composition may affect the degree of scatter radiation.
9/21/2021 60
61. Repeat Scans
Areas of the patient that are rescanned to visualize various stages of intravenous contrast
enhancement or for other technical or clinical reasons receive additional radiation.
9/21/2021 61
62. Collimation
Lead collimators are used near the x-ray tube to control the size of the beam striking the patient.
If the beam is not controlled to match the detector size, there will be additional scatter radiation to
degrade the image; this scenario will result in a higher radiation dose to the patient.
Collimators may also be used near the detectors for scatter rejection and aperture use.
9/21/2021 62
63. Localization Scans
The localization scan performed before scanning, which is often referred to as the scout
image, delivers a very low dose.
The radiation dose for the scout image is much lower than that used to produce cross-
sectional slices.
9/21/2021 63
64. Number of Detectors
In a study comparing the dose from multislice CT scanners, Moore et al. (2006) demonstrated
that the measured radiation dose is inversely proportional to the number of detector rows.
They showed that there is a trend that as the number of detector rows increases from 4 to 8 to
16 detector rows, the dose decreases with both standard and near identical technique
9/21/2021 64
65. STRATEGIES FOR REDUCING DOSE
1. Use CT Only When Clinically Indicated
2. Beam Filter
3. Adjusting mAs
4. Automatic Tube Current Modulation
5. Avoid Increasing kVp
6. Increased Pitch
7. Limit the Use of Thin Slices
8. Limit Repeat Scans
9. Newer Reconstruction Methods 9/21/2021 65
66. Dose Reduction Strategies
General Strategies are based on the assumption that the CT scanner radiation dose level and image
quality fall within the manufacturer specification and other general quality criteria
Proper patient positioning on a CT scanner table is the first step to assure use of dose saving
options.
Higher noise images can occur when patients are not well centered in the scan field of view (FOV).
By positioning the body at isocenter, the need to increase mA to compensate for the noise is
eliminated.
66
67. Adjusting mAs
adjust mAs to suit individual patient size
Small bodies require a lesser dose, and large bodies require a greater dose.
Use ATCM
9/21/2021 67
68. Avoid Increasing kVp
Increasing the x-ray tube potential increases both the radiation dose and penetration of the x-rays
through the body.
In general, increases beyond 120 kVp should be avoided, except when imaging obese patients.
However, an increase in kVp could be accompanied by a reduction in tube current to offset the
increased dose.
There is superior enhancement of iodine at lower tube potential which improves the differentiation
of vascular pathology.
9/21/2021 68
69. Limit the Use of Thin Slices
Using a large number of thin adjacent CT slices results in 30% to 50% more radiation dose to
the patient than using fewer thicker slices to scan the same anatomy.
9/21/2021 69
70. Optimization of CT protocols
CT technique should be tailored to the patient and his/her body habitus.
CT technique should be monitored and controlled to ensure that the dose is as low as reasonably
achievable (ALARA principle):
Reduction of peak KvP: relative organ dose reductions.
Tube current adjustment (mAs): AEC software
Adjust pitch: high pitch, decrease in dose
Reduce the number of “multi-phasic” scans with contrast.
Limit the scan range. 70
71. Automatic Tube current modulation
ATCM is a generic name for any
technique aimed at optimizing dose
utilization by adjusting the tube
current in real-time to
accommodate differences in
attenuation due to patient
anatomy, shape, and size. The tube
current may be modulated as a
function of projection angle (Fig.
1a), longitudinal location along the
patients (Fig. 1b), or both (Fig. 1c).
Hence data acquired through body
part having less attenuation can be
acquired with substantially less
radiation without negatively
affecting the final image noise .
71
72. 9/21/2021 72
(AEC) system: a) patient size, b) z-axis (AEC), c)
rotational/diagonal (AEC), d) combined (AEC)
Source.
73. 9/21/2021 73
Positioning the patient at isocenter is essential.
Otherwise, the attenuation calculated based upon
an off-centered Topogram will not be accurate. If a
patient is centered too low and the tube is
positioned under the table (Fig. 5a), the patient
appears wider on the Topogram than he/she actually
is (Fig. 5b), thus misinforming the CARE Dose4D
algorithm regarding the patient width. Similarly,
centering too high will misinform CARE Dose4D as to
the patient size (Fig. 5c).
74. Automatic Exposure Control (AEC)
Adapt the tube current to patient attenuation and reduce the dose about 20- 40% under the proper
specification of image quality
74
MANUFACTURER AEC TRADE NAME
GE Auto mA/ smart mA
TOSHIBA Sure exposure
SIEMENS Care dose 4D
PHILIP Dose Modulation (DOM)
75. Cardiac CT
ECG based tube current modulation is important dose reduction tool.
75
76. Advanced Iterative Reconstruction.
Iterative recon. accurately modulates the system geometry incorporating physical effect such as beam
spectrum, noise & scatter.
It Handles incomplete data set better than FBP.
Hence, no. of projection/view can be significantly reduced without compromising image quality.
There are a number of commercially-available algorithms (in 2019):
iterative reconstruction in image space (IRIS), Sinogram-affirmed iterative reconstruction (SAFIRE)-
Siemens
iDose4- Philips
adaptive iterative dose reduction (AIDR) - Toshiba
Adaptive statistical iterative reconstruction (AsIR) , Veo (model-based iterative reconstruction) – GE
76
79. 79
A–C, Low-dose CT scan obtained at 120 kVp, 3.75-mm
slice thickness, and CT dose index (CTDI) of 8 without
adaptive statistical iterative reconstruction (A) has more
image noise in liver than low-dose CT scan with adaptive
statistical iterative reconstruction (B) and routine-dose
CT scan (140 kVp; 3-mm slice thickness; CTDI, 22) (C). B
and C have nearly identical image quality.
81. Photon Counting Detector
In contrast to the conventional EIDs, which integrate the energy levels of all detected
photons, PCDs count the number of individual photons that exceed a specified energy level.
Signal from each interaction is individually processed
Improves the SNR due to elimination of electronic noise.
Benefit of SNR Improvement can be directly used to reduce patient dose.
81
82. Photon Counting Detector
82
Diagrams show detector types. A, In conventional energy-
integrating detector, an incident x-ray photon is converted into a
shower of visible light photons in a scintillator. Visible light hits an
underlying light sensor, where it generates positive and negative
electrical charges. B, In photon-counting detector, the x-ray
photon is absorbed in a semiconductor material, where it
generates positive and negative charges. Under the influence of a
strong electric field, the positive and negative charges are pulled
in opposite directions, generating an electrical signal.
83. 9/21/2021 83
Graph illustrates dose efficiency for low count
rates. A photon-counting detector has a
constant dose efficiency as a function of count
rate. For an energy-integrating detector, the
electronic noise becomes significant at very low
count rates, reducing dose efficiency.
92. Ten Steps to Wise and Gentle CT Protocols
1. Establish the appropriateness or justification guidelines for CT scanning.
2. Create clinical indication–specific CT protocols for different anatomic areas of the body. Such protocols
help tailor CT dose and image quality on the basis of specific clinical indications.
3. Ensure that the patient is positioned appropriately in the gantry iso-center.
4. Use AEC techniques to adapt tube current.
5. Use voltage adjustment on the basis of clinical indication and patient size.
9/21/2021 92
93. Ten Steps to Wise and Gentle CT Protocols
6. Select appropriate detector and section collimation on the basis of the clinical indication.
7. Avoid over-scanning beyond the anatomic area of interest.
8. Limit the use of multiphasic CT protocols. Multiphasic CT should be performed only when it is expected
to help in lesion detection or characterization.
9. Use IR techniques to improve image quality and allow dose reduction.
10. Assign a specific personnel team to be responsible for protocol review and dose monitoring.
9/21/2021 93
95. Conclusion
Although the absolute risk of performing CT is not definitively known, the team of
radiologists, physicists, technologists, and CT technology manufacturers must optimize
available technology to reduce the potential risks from ionizing radiation while maintaining
the ability to perform high-quality imaging that allows clinical questions to be answered
9/21/2021 95
97. Bushberg., 2012. Essential Physics of Medical Imaging, The. Lippincott, Williams & Wilkins, 3rd ed
Romans, L. Computed tomography for technologists, 2nd ed
Seeram, E. Computed tomography: Physical Principles, Clinical Applications, and Quality, 4th ed
Bushong, S. Radiologic science for technologists 11th ed
Goldman, Lee W. “Principles of CT: Radiation Dose and Image Quality.” Journal of Nuclear Medicine
Technology, Society of Nuclear Medicine, 1 Dec. 2007, tech.snmjournals.org/content/35/4/213.
Parakh A, Kortesniemi M, Schindera ST. CT radiation dose management: A comprehensive optimization
process for improving patient safety. Radiology. 2016;280(3):663–73.
Goo HW. CT radiation dose optimization and estimation: An update for radiologists. Korean Journal of
Radiology. 2012;13(1):1.
Kalra MK, Sodickson AD, Mayo-Smith WW. CT radiation: Key concepts for gentle and wise use.
RadioGraphics. 2015;35(6):1706–21.
9/21/2021 97
Reference
100. Question?
What are the dose descriptors used in CT?
What are the factors effecting CT dose?
How can you optimize radiation dose in CT?
9/21/2021 100
Editor's Notes
Exposure is reported in milliroentgens (mR), a much smaller unit (1 R = 1000 mR) or in microcoulombs/ kilogram (μC/kg) where 1 R = 258 μC/kg.
Any risk associated with radiation is related to the amount of energy absorbed, and this quantity is particularly important in radiation protection.
Relevant submultiples of the Gray are 1 centigray (cGy), which equals 1 r and 1 milligray (mGy), which equals 100 millirads (mrads).
The quality factor is 1 for the diagnostic x-rays that are used in CT
Old unit of effective dose is the rem, the SI unit is the Sievert (Sv), where 1 Sv = 100 rem.
To elaborate, consider an organ located on the proximal side of the body relative to the x-ray source. This organ will get roughly the same dose in both adult and child. As the x-ray source rotates, that same organ will be on the distal side of the body relative to the x-ray source; now that organ is partly shielded by the body tissue proximal to it, reducing the organ dose. But this dose-reducing, partial shielding will be much less for a thin individual, such as a child, compared with a thicker adult. Thus, organ doses for children are larger than those for adults.
To elaborate, consider an organ located on the proximal side of the body relative to the x-ray source. This organ will get roughly the same dose in both adult and child. As the x-ray source rotates, that same organ will be on the distal side of the body relative to the x-ray source; now that organ is partly shielded by the body tissue proximal to it, reducing the organ dose. But this dose-reducing, partial shielding will be much less for a thin individual, such as a child, compared with a thicker adult. Thus, organ doses for children are larger than those for adults.
The CTDI can only be calculated if slices are contiguous, that is, there are no overlapping or gapped slices. If there is slice overlap or gaps, the CTDI is multiplied by the ratio of slice thickness to slice increment. This would technically be the MSAD, because the CTDI conditions would no longer exist. Equipment manufacturers report CTDI doses for typical head and body imaging techniques. These are equivalent to the dose a patient receives if multiple adjacent slices are acquired.
With the pencil chamber located at the center (in the z-dimension) of the phantom and also at the center of the CT gantry, a single axial (also called sequential) CT scan is made (using no table translation)
Medical physicists usually use a special dosimeter called a pencil ionization chamber to measure the CTDI. This 100-mm-long thin cylindrical device is long enough to span the width of 14 contiguous 7-mm CT slices. This provides a better estimate of MSAD for thin slices than that of the single-slice method. When this method is used it is referred to as the CTDI100.
The nominal beam width refers to the beam width as reported by the scanner, not the actual measured beam width, which is generally slightly wider.
CTDIVOL provides information regarding only the scanner output. It does not address patient size, and hence does not estimate patient dose (McCollough 2011).
CTDIVOL provides information regarding only the scanner output. It does not address patient size, and hence does not estimate patient dose (McCollough 2011).
In SSCT, as slice thickness increases, the volume of tissue irradiated increases, and the dose may increase slightly to the adjacent slices.