Performance Testing in Computer Tomography using CATPHAN 600 Asim Gilani Radiation Protection Unit Medical Physics Department Dubai Hospital 08/07/09
CATPHAN 500 & 600
CATPHAN 500 & 600
Mounting
 
 
CATPHAN 600
CATPHAN 600
CTP 404 Slice Geometry and Sensitometry Module
CTP 404 Diameter = 150 mm Thickness = 25 mm 4 wire Ramps at an angle of 23 ° 5 Acrylic spheres 8 sensitometry samples 4 rods
CTP 404
CTP 404 Phantom position verification Patient alignment system Circular symmetry Scan width Sensitometry
Phantom Alignment Align phantom dots with the lasers
 
 
 
Phantom Alignment Scan 6 or 8 axial slices with minimum slice width Start 3 or 4 slices before the zero position
 
Phantom oriented to right side
Phantom oriented to left side
Phantom Tilted
 
Aligned Image
How to verify ? Switch ON grid to check the Phantom Alignment
 
 
CUATION Once the phantom is centered and aligned DO NOT touch or move the phantom by hand Only move or align it by table movements
Test 1: LASER Accuracy Align the phantom to lasers by moving the table Align all the dots to X, Y and Z LASERS as precisely as possible Scan axially with minimum slice thickness If the Image is same as previous one then LASERS are perfectly aligned OTHERWISE !
Measurement of Z Axis Misalignment
Z-axis Misalignment Measure the distance between center of the phantom and center of the slice Multiple by 0.42 to find the z-axis misalignment 4.99 x 0.42 = 2.1 mm
X and Y Misalignment Switch ON grid and measure distance of the center of the grid to center of the image in X and Y direction and multiply by 0.42
X and Y Misalignment X = 4.61 x 0.42 1.9 mm Y = 3.26 x 0.42 1.4 mm
Test 2: Patient Alignment System Take Scanogram or Scout Take an axial Scan at the crosswire Measure the accuracy of the scan
 
 
Test 3: Circular Symmetry Take an axial Scan of the CTP 404 Image must be circular not elliptical Measure diameter of the phantom in X & Y Measure distance between Teflon pins Additionally angle accuracy can be verified
 
Distance Accuracy
Angle Accuracy
Test 4: Slice Width Accuracy Axially scan CTP 404 with different slice widths Measure slice with at (FWHM) ‏ Full Width at Half Maximum Slice Thickness = FWHM * 0.42
 
 
6.00  *  0.42  =  2.52 mm
Test 5: Slice Sensitometry Scan the phantom with different slice widths Measure CT Numbers of Different samples
CTP 404
 
Material  CT #  Air PMP LDPE Water Polystyrene Acrylic Delrin Teflon - 1000 - 200 - 100 0 - 35 120 340 990
Table Movement  Select axial scan with minimum slice width Move table according to figure and scan Verify that you see the dot above
Water and kV Accuracy Axially scan at different kV's Verify that water CT number lies between -7 to + 7
Volume Verification The volume of five spheres ????
CTP 591 Bead Geometry Module
CTP 591 Diameter = 150 mm Thickness = 40 mm 2 pairs of coarse ramps (39 beads each)‏ 1 pair of precision ramp (25 Beads each)‏ 2 Isolated beads for MTF and SSP 1 Steel wire of 50  µm diameter
 
Slice Thickness Verification Make helical scan at different slice widths Verify that all slices are of same thickness
 
Pitch Verification Make helical scan with different Pitches available Verify the pitch by measuring the center to center difference in adjacent scan and subtracting by the slice width Turn grid ON for help
 
MTF Measurement The beads are for SSP and MTF measurements We need software for these measurements
CTP 528 High Resolution Module
CTP 528 Diameter = 150 mm Thickness = 40 mm Radial gauge from 1 to 21 lines per cm 2 mm thick aluminum 2 beads of 0.28 mm 20 mm above and below the center in Y-axis  & 2.5 mm and 10 mm past the center in Z-axis
 
High Resolution Measurements
7 lp / cm visible
The beads are used for MTF and SSP measurements HOW  ????????????
CTP 515 Low Contrast Module
CTP 515 Diameter = 150 mm Thickness = 40 mm 39 cylindrical rods 3 contrasts, 0.3 %, 0.5 % & 1.0 % 3, 5, 7 & 40 mm length 2 to 9 & 15 mm diameter
 
 
 
Contrast Verification (Measured Contrast) * (Smallest Diameter Discernible)    Constant (1.0)  *  (3.0)  =  3.0 (0.5)  *  (5.0)  =  2.5 (0.3)  *  (9.0)  =  2.7
Contrast Measurements Take ROI on 155 mm circles Note down the CT # Take ROI immediately around the circle Subtract the 2 ROI's It should be 10, 5 and 3
 
CTP 486 Uniformity Module
CTP 486 Diameter = 150 mm Thickness > 40 mm Uniform Module CT # between 0 to 2 % of water
 
Artifact Analysis Change Window Width and level and observe any artifacts
Questions
Suggestions
Thanks

QC of CT using Catphan 600

  • 1.
    Performance Testing inComputer Tomography using CATPHAN 600 Asim Gilani Radiation Protection Unit Medical Physics Department Dubai Hospital 08/07/09
  • 2.
  • 3.
  • 4.
  • 5.
  • 6.
  • 7.
  • 8.
  • 9.
    CTP 404 SliceGeometry and Sensitometry Module
  • 10.
    CTP 404 Diameter= 150 mm Thickness = 25 mm 4 wire Ramps at an angle of 23 ° 5 Acrylic spheres 8 sensitometry samples 4 rods
  • 11.
  • 12.
    CTP 404 Phantomposition verification Patient alignment system Circular symmetry Scan width Sensitometry
  • 13.
    Phantom Alignment Alignphantom dots with the lasers
  • 14.
  • 15.
  • 16.
  • 17.
    Phantom Alignment Scan6 or 8 axial slices with minimum slice width Start 3 or 4 slices before the zero position
  • 18.
  • 19.
  • 20.
  • 21.
  • 22.
  • 23.
  • 24.
    How to verify? Switch ON grid to check the Phantom Alignment
  • 25.
  • 26.
  • 27.
    CUATION Once thephantom is centered and aligned DO NOT touch or move the phantom by hand Only move or align it by table movements
  • 28.
    Test 1: LASERAccuracy Align the phantom to lasers by moving the table Align all the dots to X, Y and Z LASERS as precisely as possible Scan axially with minimum slice thickness If the Image is same as previous one then LASERS are perfectly aligned OTHERWISE !
  • 29.
    Measurement of ZAxis Misalignment
  • 30.
    Z-axis Misalignment Measurethe distance between center of the phantom and center of the slice Multiple by 0.42 to find the z-axis misalignment 4.99 x 0.42 = 2.1 mm
  • 31.
    X and YMisalignment Switch ON grid and measure distance of the center of the grid to center of the image in X and Y direction and multiply by 0.42
  • 32.
    X and YMisalignment X = 4.61 x 0.42 1.9 mm Y = 3.26 x 0.42 1.4 mm
  • 33.
    Test 2: PatientAlignment System Take Scanogram or Scout Take an axial Scan at the crosswire Measure the accuracy of the scan
  • 34.
  • 35.
  • 36.
    Test 3: CircularSymmetry Take an axial Scan of the CTP 404 Image must be circular not elliptical Measure diameter of the phantom in X & Y Measure distance between Teflon pins Additionally angle accuracy can be verified
  • 37.
  • 38.
  • 39.
  • 40.
    Test 4: SliceWidth Accuracy Axially scan CTP 404 with different slice widths Measure slice with at (FWHM) ‏ Full Width at Half Maximum Slice Thickness = FWHM * 0.42
  • 41.
  • 42.
  • 43.
    6.00 * 0.42 = 2.52 mm
  • 44.
    Test 5: SliceSensitometry Scan the phantom with different slice widths Measure CT Numbers of Different samples
  • 45.
  • 46.
  • 47.
    Material CT# Air PMP LDPE Water Polystyrene Acrylic Delrin Teflon - 1000 - 200 - 100 0 - 35 120 340 990
  • 48.
    Table Movement Select axial scan with minimum slice width Move table according to figure and scan Verify that you see the dot above
  • 49.
    Water and kVAccuracy Axially scan at different kV's Verify that water CT number lies between -7 to + 7
  • 50.
    Volume Verification Thevolume of five spheres ????
  • 51.
    CTP 591 BeadGeometry Module
  • 52.
    CTP 591 Diameter= 150 mm Thickness = 40 mm 2 pairs of coarse ramps (39 beads each)‏ 1 pair of precision ramp (25 Beads each)‏ 2 Isolated beads for MTF and SSP 1 Steel wire of 50 µm diameter
  • 53.
  • 54.
    Slice Thickness VerificationMake helical scan at different slice widths Verify that all slices are of same thickness
  • 55.
  • 56.
    Pitch Verification Makehelical scan with different Pitches available Verify the pitch by measuring the center to center difference in adjacent scan and subtracting by the slice width Turn grid ON for help
  • 57.
  • 58.
    MTF Measurement Thebeads are for SSP and MTF measurements We need software for these measurements
  • 59.
    CTP 528 HighResolution Module
  • 60.
    CTP 528 Diameter= 150 mm Thickness = 40 mm Radial gauge from 1 to 21 lines per cm 2 mm thick aluminum 2 beads of 0.28 mm 20 mm above and below the center in Y-axis & 2.5 mm and 10 mm past the center in Z-axis
  • 61.
  • 62.
  • 63.
    7 lp /cm visible
  • 64.
    The beads areused for MTF and SSP measurements HOW ????????????
  • 65.
    CTP 515 LowContrast Module
  • 66.
    CTP 515 Diameter= 150 mm Thickness = 40 mm 39 cylindrical rods 3 contrasts, 0.3 %, 0.5 % & 1.0 % 3, 5, 7 & 40 mm length 2 to 9 & 15 mm diameter
  • 67.
  • 68.
  • 69.
  • 70.
    Contrast Verification (MeasuredContrast) * (Smallest Diameter Discernible)  Constant (1.0) * (3.0) = 3.0 (0.5) * (5.0) = 2.5 (0.3) * (9.0) = 2.7
  • 71.
    Contrast Measurements TakeROI on 155 mm circles Note down the CT # Take ROI immediately around the circle Subtract the 2 ROI's It should be 10, 5 and 3
  • 72.
  • 73.
  • 74.
    CTP 486 Diameter= 150 mm Thickness > 40 mm Uniform Module CT # between 0 to 2 % of water
  • 75.
  • 76.
    Artifact Analysis ChangeWindow Width and level and observe any artifacts
  • 77.
  • 78.
  • 79.