SlideShare a Scribd company logo
1 of 48
Download to read offline
Vineeth.C
CUAPRAD012
1
• Millenium 120 Leaf HDTM MLC
• PerfectPitch TM 6DoF Couch
• RGSC Gating
• aS1200 MV Imager
• Photon Energies :- 6,10,15 MV
& 6,10 MV FFF
• Electron Energies :-
6,9,12,15,18 MeV & 6,9 MeV
HDTSe
2
New TrueBeam machine is installed in the existing bunker of Siemens Primus Plus with a
maximum photon energy of 15 MV
according to AERB guidelines , though maximum energy remains the
same, additional shielding of 30 cm concrete was added on the
primary as well as secondary wall of area with maximum occupancy
3
The acceptance testing implies the verification process of the
machine based on manufacture’s guidelines for a very small subset of
beam data whereas commissioning is a process where a full set of
data is acquired that will be used for patient treatment
Manufacturers often provide guidelines and
tolerance limits for acceptance testing of a
machine through their acceptance testing
procedure
4
Energy
(MV)
6 10 15 6FFF 10FFF
Dmax(cm)
(±0.15)
1.60 2.4 2.9 1.50 2.34
PDD at
10cm (±1.0)
67.2 74.1 77.4 64.3 71.8
Min D.R
(MU/Min) 5 5 20 400 400
Max D.R
(MU/Min) 600 600 600 1400 2400
e- energy
MeV
6 9 12 15 18
Depth of
ionisation1
90% (cm,±0.1)
1.71 2.68 3.77 4.67 5.58
80% (cm,±0.07) 1.90 2.95 4.15 5.2 6.09
50% (cm,±0.1) 2.32 3.52 4.91 6.19 7.41
Radial &
Transverse
flatness2
±5% ±4.5% ±4.5% ±4.5% ±4.5%
1Depth of ionization applies to the 15 X15 cm2 applicator field size, using a water phantom at 100 SSD
2Flatness is defined as the maximum variation from the mean electron ionization delivered within the
central 80% FWHM region measured for 10x10 cm2 through 25 x 25 cm2 fields.
5
▪ Varian procedure for isocenter verification “Isolock”
▪ MV images –
1. 63 different gantry/collimator angles
2. 13 different couch angles
▪ Central axis s-ray beam variation :-
the gantry and/or collimator positions – 0.5 mm
gantry,couch, and/or collimator positions – 0.75 mm
▪ EPID-based Winston Lutz (WL) test :- within 1 mm tolerance
6
❑ Congruence between optical and radiation fields
Symmetric Fields
- 5x5 cm2 : 0.5mm
-10x10 cm2 : 1mm
-20x20 cm2 :1mm Tolerance Limits = 2mm
-30x30 cm2 :1mm
Asymmetric fields
- defined by over travel option of the collimator - 1mm
❑Overlapping between parallel opposed radiation fields
- defined by jaws for 10x10 cm2 Tolerance = 3mm
- defined by MLC for 10x10 cm2 0.5 mm
7
▪ Satisfactory acceptance testing simply assures that the accelerator satisfies all
agreed-upon specifications and pertinent safety requirements
▪ Commissioning includes comprehensive measurements of dosimetric parameters
▪ also includes
- entry of beam data into a treatment planning system
- testing of its accuracy
- development of operational procedures
-training of all concerned with the operation of the accelerator
8
9
i. PTW PinPoint® Chamber.
ii. Advanced Markus® Electron
Chamber
iii. PTW Farmer® Ionization
Chamber
iv. IBA CC13 Ionization
Chamber
v. IBA RAZOR™ Nano
Chamber
10
Dmax (cm)
6 MV 1.59
10 M V 2.42
15 MV 2.9
6 MV FFF 1.43
10 MV FFF 2.36
PHOTONS
11
100×
Dmax −Dmin
Dmax +Dmin
12
SYMMETRY
The ratio between measured values for each pair of symmetrical points
(with respect to beam axis) for a range of field sizes and gantry orientations
must lie between .98 and 1.02 within the central 80% flattened beam area.
Field Size
(cm2)
6 MV 10 MV 15 MV
Flatness Symmetry Flatness Symmetry Flatness Symmetry
I P C P I P C P I P C P I P C P I P C P I P C P
05x05 104.3 104.2 100.2 100.3 103.8 104.1 100.2 100.3 103.8 104.3 100.5 100.4
10X10 105.0 105.3 100.4 100.6 104.8 105.1 100.2 100.5 104.2 104.5 100.5 100.5
20x20 104.3 104.4 100.4 100.6 103.5 103.2 100.4 100.6 103.4 103.5 100.7 100.6
30x30 104.2 104.1 100.5 100.9 103.6 103.7 100.5 100.5 104.2 103.9 100.7 100.4
40x40 105.1 105.0 100.6 100.8 104.0 103.6 100.7 100.6 104.3 104.4 100.6 100.5
Flatness for field sizes 5x5 cm2 to 30x30 cm2 : ± 3% & for field size >30x30 cm2: ± 5%
Symmetry = ±2% 13
14
❖ In-line, cross-line and diagonal beam profiles were measured for all available beam
energies for various recommended field sizes at Dmax, 5 cm, 10 cm, 20 cm and 30 cm
depth
❖ Beam profile data were first smoothened by median filter and then corrected for the
central axis discrepancy
❖ Beam profiles ware normalize to 100% at the central axis to their corresponding field
size
PENUMBRA
Penumbra for flat beam defined as the lateral separation of (20% - 80%) isodose on either
side of beam profile normalized to 100% at the central axis
F S
10 X 10 cm2
In Plane Cross Plane
Left (mm) Right (mm) Left (mm) Right (mm)
6X 5.55 5.57 5.93 6.01
10X 7.08 7.11 7.21 7.13
15X 7.21 7.29 7.5 7.4
15
16
FFF beam analysis done
as per AERB Task Group
Recommendations
All Profiles Above 30 X 30 Field Size are
acquired in half profiles
17
Measured Diagonal Profiles Calculated Diagonal Profiles
18
19
20
According to TRS-398
TPR20,10 = 1.2661 × PDD20,10 − 0.0595
TIME
TPR20,10
6 MV 6 MV FFF 10 MV 10 MV
FFF 15 MV
10AM 0.6771 0.6381 0.7485 0.7157 0.7724
2PM 0.6754 0.6386 0.7482 0.7154 0.7730
6PM 0.6762 0.6386 0.7483 0.7144 0.7727
RAV 0.6762 0.6384 0.7483 0.7152 0.7727
21
ENERGY (MV) DLG (CM)
6 0.0470
10 0.0540
15 0.0560
6 FFF 0.0390
10 FFF 0.0490
22
Energy
With Couch
[MR(nC)]
Without Couch
[MR (nC) ]
Couch Trans.
Factor
Couch
Trans.
Factor
Thin Thick Thin Thick Thin Thick Avg
6MV 14.3 14.13 14.6 14.6 0.9795 0.9678 0.9736
10MV 15.98 15.84 16.22 16.22 0.9852 0.9766 0.9809
15MV 16.97 16.82 17.16 17.16 0.9889 0.9802 0.9846
6MV
FFF
13.7 13.52 14.01 14.01 0.9779 0.9650 0.9714
10MV
FFF
15.39 15.25 15.64 15.64 0.9840 0.9751 0.9795
23
X-Ray
contamination
24
Applicator
/ Field Size
(cm2)
Energy
(MeV)
Flatness (mm) Symmetry (%)
In Plane Cross Plane
In Plane Cross Plane
Lt Rt Lt Rt
10x10
6 8.5 8.4 9.2 8.7 100.42 100.73
9 7.2 6.9 7.2 7.4 100.7 100.62
12 7.6 8.0 7.3 7.7 101.7 100.93
15 7.6 7.6 7.3 7.8 100.96 100.95
18 7.6 7.6 7.5 7.9 100.67 100.72
25x25
6 9.0 8.2 8.9 8.7 101.37 100.52
9 6.8 6.2 6.6 6.6 101.6 100.82
12 6.9 6.3 6.6 6.6 101.53 100.4
15 7.2 6.8 6.7 6.8 101.41 100.97
18 7.5 7.2 6.9 6.8 100.91 100.82
Tolerance :- should not exceed 103% of the central axis value 25
Energy (MeV)
Penumbra (mm)
In Plane Cross Plane
Lt Rt Lt Rt
6 9.8 10.0 9.6 10.0
9 9.6 9.7 9.6 9.8
12 9.6 9.7 9.6 9.8
15 9.8 10.0 9.9 10.0
18 10.1 10.0 10.1 10.2
26
27
28
Measured PDD data
is calculated and
fitted using EMC in
Eclipse Treatment
planning systrem
Open Field Depth Dose - 6 Mev
29
30
Leakage Radiation Measurement
31
32
Energy Avg. % Transmission Max. %
Transmission
X Jaw Y Jaw X Jaw Y Jaw
6 MV 0.41 0.45 0.60 0.70
10 MV 0.48 0.53 0.66 0.76
15 MV 0.5 0.57 0.7 0.83
6MV FFF 0.24 0.26 0.39 0.47
10MV
FFF 0.22 0.24 0.34 0.39
33
(a) Each beam limiting jaw (excluding MLC jaws) shall attenuate x-radiation such
that kerma rate due to transmitted radiation does not exceed 2% of the maximum
kerma rate measured along the central axis of the beam at NTD (normal treatment
distance) in a 10 x 10 cm radiation beam.
(b) For radiation fields of any size, the average kerma rate due to transmission
through the beam limiting jaws, including MLC jaws, shall not exceed 0.75% of the
maximum kerma rate on the central axis at NTD in a 10 x 10 cm radiation field.
34
Measurement:-
• SCD 100 cm at a depth of 5 cm in water phantom (RFA)
-Interleaf leakage measurements were performed separately for both banks
• CC13 ion chamber exactly placed between the two leaves.
• Projection of the leaf is properly verified and Meter readings are taken 10 cm
off-axis with fully closed MLC
• Increment of 1 cm taken in order to place the chamber exactly between the
two leaves and chamber moved to next consecutive position with the help of
myQA software.
- The same procedure is followed for intra leaf transmission except the
chamber is placed in center of leaf.
• Normalization - 10 × 10 cm2 open fields
• 600 MU/Min for FF beam & 1200 & 2400 MU/Min for 6 & 10 MV FFF beams
respectively
• 100 MU
ENERGY % TRANSMISSION
Avg. Max
6 MV 1.15 1.42
10 MV 1.42 1.71
15 MV 1.48 1.86
6 MV FFF 0.6 0.86
10 MV FFF 0.58 0.85
35
▪ plane circular surface of radius 2 m
centered on and normal to the
radiation beam axis at NTD and
outside the area of the maximum
radiation beam area
▪ Tolerance :- a maximum of 0.2% and
an average of 0.1%
Energy % Transmission
Max Min
6 MV 0.0158 0.0107
10 MV 0.0163 0.0061
15 MV 0.0206 0.0069
6 MV FFF 0.0074 0.0044
10 MV FFF 0.0094 0.0025
Neutron :-max = 0.0131 % tol= 0.05 %
avg = 0.0126 % 0.02 % 36
at 1 m from the path
of the electrons
between electron
gun to target &
reference axis other
than patient plane
Energy % Transmission
Max Avg
6 MV 0.0219 0.0129
10 MV 0.017 0.0050
15 MV 0.0172 0.0074
6 MV FFF 0.0080 0.0043
10 MV FFF 0.0034 0.0020
37
38
39
40
COV = (1/ X )*[(  (Xi- X)2 ) /(n-1)] ½
Set kV Output (mGy)
Avg X Σ(Xi-X)2 [Σ(Xi-X)2]/n-1 [[Σ(Xi-X)2]/n-1]1/2 COV
1 2 3
50 0.0615 0.0618 0.0613 0.0615 0.00000013 0.00000006 0.000251 0.00408
70 0.131 0.1313 0.1317 0.1313 0.00000024 0.00000012 0.000343 0.002613
80 0.173 0.1724 0.1722 0.1725 0.00000036 0.00000018 0.000422 0.002446
100 0.2618 0.2617 0.2613 0.2616 0.00000012 0.00000006 0.000247 0.000946
120 0.3599 0.3602 0.3584 0.3595 0.00000197 0.00000099 0.000994 0.002764
130 0.4106 0.4107 0.4116 0.411 0.0000006 0.0000003 0.000545 0.001327
140 0.4648 0.4654 0.4645 0.4649 0.00000021 0.00000021 0.000459 0.000987
41
Diameter of smallest hole clearly resolved
on monitor screen - 0.6mm or 0.023 inch
Recommended Performance standard that
must be resolved – 3 mm or 0.117 inch
Bar strip resolved on the monitor screen -
2.8 lp/mm
Recommended performance standard –
1.5 lp/mm must be resolved
42
Picket Fence Test
43
The picket fence test
consists of eight consecutive
leaf movements of 5 cm
wide rectangular fields
spaced at 5 cm intervals.
The field information is
contained in three test files,
which are run in sequence.
This test is used to verify the
leaf positioning accuracy
and also calibrates the
carriage positioning
accuracy
44
▪ Commissioning of Varian TrueBeamTM Linear Accelerator was successfully carried
out. Removal flattening filter alters various commissioning associated parameter as
Field flatness, Field penumbra, Beam quality, Surface dose,Transmission factor, off
axis energy variation and Homogeneity need to be redefined for FFF beam other
than FF beam.
45
▪ [1] Das, I.J., Cheng, C.W.,Watts, R.J., Ahnesjö, A.,
Gibbons, J., Li, X.A., Lowenstein, J., Mitra, R.K.,
Simon,W.E. and Zhu,T.C. (2008) Accelerator
Beam Data Commissioning Equipment and
Procedures: Report of the TG-106 of the Therapy
Physics Committee of the AAPM. Medical Physics,
35, 4186-4215.
http://dx.doi.org/10.1118/1.2969070
▪ [2] Nath, R., Biggs, P.J., Bova, F.J., Ling, C.C., Purdy,
J.A., van de Geijn, J. and Weinhous, M.S. (1994)
AAPM Code of Practice for Radiotherapy
accelerators: Report of AAPM Radiation Therapy
Task Group No. 45. Medical Physics, 21, 1093-
1121. (AAPM Report No. 47)
▪ [3] Sahani, G., Sharma, S.D., Sharma, P.K.,
Deshpande, D.D., Negi, P.S., Sathianarayanan,V.K.
and Rath, G.K. (2014) Acceptance Criteria for
Flattening Filter-Free Photon Beam from Standard
Medical Electron Linear Accelerator: AERB Task
Group Recommendations. Journal of Medical
Physics, 39, 206-211.
http://dx.doi.org/10.4103/0971-6203.144482
▪ [4] Khan, F.M. (2003) Physics of Radiation Therapy.
3rd Edition, Lippincott Williams & Wilkins.
▪ 5] Musolino, S.V. (2000) Absorbed Dose
Determination in External Beam Radiotherapy: An
International Code of Practice for Dosimetry
Based on Standards of Absorbed Dose to Water.
Health Physics, 81, 592-593. (IAEA TRS-398)
▪ [6] Boyer, A., Biggs, P., Galvin, J., Klein, E., LoSasso,
T., Low, D., Mah, K. and Yu, C. (2001) Basic
Applications of Multileaf Collimator: Report of
Task Group No. 50 Radiation Therapy Committee.
American Association of Physicists in Medicine
(AAPM TG-50, 10) (AAPM Report No. 72).
46
[7]Dosimetric Leaf Gap Measurement, Procedure
Recommended by Varian Medical System.
Eclipse 10 Inverse Planning Administration and
Physics rev. 6.1.1.
[8] Szpala, S., Cao, F. and Kohli, K. (2014) On Using
the Dosimetric Leaf Gap to Model the Rounded
Leaf Ends in VMAT/RapidArc Plans. Journal of
Applied Clinical Medical Physics, 15, 4484.
[9] Acceptance/quality assurance tests for
Medical linear accelerator, RPAD/ACC/QA/04
[10]Absorbed dose determination In external
Beam radiotherapy An International Code of
Practice for Dosimetry Based on Standards of
Absorbed Dose to Water -TRS 398
[11]IEC 60601-2-1 Particular requirements for the
basic safety and essential performance of
electron accelerators in range 1 MeV to 50 MeV
[12] IAEA TRS-381 The use of plane-parallel
ionization chambers in high-energy electron and
photon beams. An international code of practice
for dosimetry
[13] K. R.Winston and W. Lutz,“Linear accelerator
as a neurosurgical tool for
stereotactic radiosurgery,” Neurosurgery 22, 454–
464 (1988).
[14] Commissioning and Acceptance Testing of
the existing linear accelerator upgraded to
volumetric modulated arc therapy - Ekambaram
Varadharajana and Velayudham
Ramasubramanian
[15] Shende, R., et al. (2016) Commissioning of
TrueBeamTM Medical Linear Accelerator:
Quantitative and Qualitative Dosimetric Analysis
and Comparison of Flattening Filter (FF) and
Flattening Filter Free (FFF) Beam. International
Journal of Medical Physics, Clinical Engineering
and Radiation Oncology, 5, 51-69.
47
48

More Related Content

What's hot

Brachytherapy dosimetry
Brachytherapy dosimetryBrachytherapy dosimetry
Brachytherapy dosimetrySabari Kumar
 
Seminar 2 .beam modification devices
Seminar 2 .beam modification devicesSeminar 2 .beam modification devices
Seminar 2 .beam modification devicesDr ABDUL WAHEED DAR
 
Beam modifying devices
Beam modifying devicesBeam modifying devices
Beam modifying devicesMariya Kikali
 
Quality Assurance in Radiotherapy and Dosimetry
Quality Assurance in Radiotherapy and DosimetryQuality Assurance in Radiotherapy and Dosimetry
Quality Assurance in Radiotherapy and Dosimetrymahbubul hassan
 
Linear Accelerator Acceptance, Commissioning and Annual QA
Linear Accelerator Acceptance, Commissioning and Annual QALinear Accelerator Acceptance, Commissioning and Annual QA
Linear Accelerator Acceptance, Commissioning and Annual QAMiami Cancer Institute
 
Beam modification devices
Beam modification devicesBeam modification devices
Beam modification devicesVibhay Pareek
 
challenges of small field dosimetry
challenges of small field dosimetrychallenges of small field dosimetry
challenges of small field dosimetryLayal Jambi
 
multiple filed arrangement in Radiotherapy, Medical College Kolkata
multiple filed arrangement in Radiotherapy, Medical College Kolkatamultiple filed arrangement in Radiotherapy, Medical College Kolkata
multiple filed arrangement in Radiotherapy, Medical College KolkataKazi Manir
 
Immobilization in radiotherapy
Immobilization in radiotherapy Immobilization in radiotherapy
Immobilization in radiotherapy prabhurt
 
Patient Positioning and Immobilization Devices In Radiotherapy Planning
Patient Positioning and Immobilization Devices In Radiotherapy PlanningPatient Positioning and Immobilization Devices In Radiotherapy Planning
Patient Positioning and Immobilization Devices In Radiotherapy PlanningSubhash Thakur
 
Beam modifying devices 1 Radiophysics
Beam modifying devices 1 RadiophysicsBeam modifying devices 1 Radiophysics
Beam modifying devices 1 RadiophysicsDrAyush Garg
 
Plan evaluation in Radiotherapy- Dr Kiran
Plan evaluation in Radiotherapy- Dr KiranPlan evaluation in Radiotherapy- Dr Kiran
Plan evaluation in Radiotherapy- Dr KiranKiran Ramakrishna
 
Treatment verification and set up errors
Treatment verification and set up errorsTreatment verification and set up errors
Treatment verification and set up errorssailakshmi pullookkara
 
Topic -Analysis of Patients specific Quality Assurance of IMRT/VMAT
Topic -Analysis of Patients specific Quality Assurance of IMRT/VMATTopic -Analysis of Patients specific Quality Assurance of IMRT/VMAT
Topic -Analysis of Patients specific Quality Assurance of IMRT/VMATRajeev Kumar Pandit
 

What's hot (20)

Gap correction
Gap correctionGap correction
Gap correction
 
Brachytherapy dosimetry
Brachytherapy dosimetryBrachytherapy dosimetry
Brachytherapy dosimetry
 
Seminar 2 .beam modification devices
Seminar 2 .beam modification devicesSeminar 2 .beam modification devices
Seminar 2 .beam modification devices
 
Beam modifying devices
Beam modifying devicesBeam modifying devices
Beam modifying devices
 
Motion Management in Radiation Therapy
Motion Management in Radiation TherapyMotion Management in Radiation Therapy
Motion Management in Radiation Therapy
 
Quality Assurance in Radiotherapy and Dosimetry
Quality Assurance in Radiotherapy and DosimetryQuality Assurance in Radiotherapy and Dosimetry
Quality Assurance in Radiotherapy and Dosimetry
 
Linear Accelerator Acceptance, Commissioning and Annual QA
Linear Accelerator Acceptance, Commissioning and Annual QALinear Accelerator Acceptance, Commissioning and Annual QA
Linear Accelerator Acceptance, Commissioning and Annual QA
 
Beam modification devices
Beam modification devicesBeam modification devices
Beam modification devices
 
challenges of small field dosimetry
challenges of small field dosimetrychallenges of small field dosimetry
challenges of small field dosimetry
 
multiple filed arrangement in Radiotherapy, Medical College Kolkata
multiple filed arrangement in Radiotherapy, Medical College Kolkatamultiple filed arrangement in Radiotherapy, Medical College Kolkata
multiple filed arrangement in Radiotherapy, Medical College Kolkata
 
Flattening filter Free
Flattening filter FreeFlattening filter Free
Flattening filter Free
 
Isodose lines
Isodose linesIsodose lines
Isodose lines
 
Cobalt & linac
Cobalt & linacCobalt & linac
Cobalt & linac
 
Immobilization in radiotherapy
Immobilization in radiotherapy Immobilization in radiotherapy
Immobilization in radiotherapy
 
Beam Direction
Beam DirectionBeam Direction
Beam Direction
 
Patient Positioning and Immobilization Devices In Radiotherapy Planning
Patient Positioning and Immobilization Devices In Radiotherapy PlanningPatient Positioning and Immobilization Devices In Radiotherapy Planning
Patient Positioning and Immobilization Devices In Radiotherapy Planning
 
Beam modifying devices 1 Radiophysics
Beam modifying devices 1 RadiophysicsBeam modifying devices 1 Radiophysics
Beam modifying devices 1 Radiophysics
 
Plan evaluation in Radiotherapy- Dr Kiran
Plan evaluation in Radiotherapy- Dr KiranPlan evaluation in Radiotherapy- Dr Kiran
Plan evaluation in Radiotherapy- Dr Kiran
 
Treatment verification and set up errors
Treatment verification and set up errorsTreatment verification and set up errors
Treatment verification and set up errors
 
Topic -Analysis of Patients specific Quality Assurance of IMRT/VMAT
Topic -Analysis of Patients specific Quality Assurance of IMRT/VMATTopic -Analysis of Patients specific Quality Assurance of IMRT/VMAT
Topic -Analysis of Patients specific Quality Assurance of IMRT/VMAT
 

Similar to Varian TrueBeam Machine Commissioning Report

QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptxQUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptxSuryaSuganthan2
 
LINAC COMMSSN.ppt
LINAC COMMSSN.pptLINAC COMMSSN.ppt
LINAC COMMSSN.pptBBAdhikari
 
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfQUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfAngelina Protik
 
DSD-INT - SWAN Advanced Course - 02 - Setting up a SWAN computation
DSD-INT - SWAN Advanced Course - 02 - Setting up a SWAN computationDSD-INT - SWAN Advanced Course - 02 - Setting up a SWAN computation
DSD-INT - SWAN Advanced Course - 02 - Setting up a SWAN computationDeltares
 
Dose reduction technique in ct scan
Dose reduction technique in ct scanDose reduction technique in ct scan
Dose reduction technique in ct scanMohd Aiman Azmardi
 
UV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptUV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptJgdishrathi
 
Vibration monitoring of Gearboxes.pdf
Vibration monitoring of Gearboxes.pdfVibration monitoring of Gearboxes.pdf
Vibration monitoring of Gearboxes.pdfKamalPasha28
 
Data Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRA
Data Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRAData Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRA
Data Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRAPT. Siwali Swantika
 
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618PT. Siwali Swantika
 
First results from the full-scale prototype for the Fluorescence detector Arr...
First results from the full-scale prototype for the Fluorescence detector Arr...First results from the full-scale prototype for the Fluorescence detector Arr...
First results from the full-scale prototype for the Fluorescence detector Arr...Toshihiro FUJII
 
High speed board design considerations
High speed board design considerationsHigh speed board design considerations
High speed board design considerationsMohamad Tisani
 
LVTS - Image Resolution Monitor for Litho-Metrology
LVTS - Image Resolution Monitor for Litho-MetrologyLVTS - Image Resolution Monitor for Litho-Metrology
LVTS - Image Resolution Monitor for Litho-MetrologyVladislav Kaplan
 
FAST実験7:新型大気蛍光望遠鏡による極高エネルギー宇宙線観測報告
FAST実験7:新型大気蛍光望遠鏡による極高エネルギー宇宙線観測報告FAST実験7:新型大気蛍光望遠鏡による極高エネルギー宇宙線観測報告
FAST実験7:新型大気蛍光望遠鏡による極高エネルギー宇宙線観測報告Toshihiro FUJII
 
Effects of crystal size on the imaging performance using two collimator seri...
Effects of crystal size on the imaging performance  using two collimator seri...Effects of crystal size on the imaging performance  using two collimator seri...
Effects of crystal size on the imaging performance using two collimator seri...Sin-di Lee
 
FAST実験3:新型大気蛍光望遠鏡の試験観測報告
FAST実験3:新型大気蛍光望遠鏡の試験観測報告FAST実験3:新型大気蛍光望遠鏡の試験観測報告
FAST実験3:新型大気蛍光望遠鏡の試験観測報告Toshihiro FUJII
 

Similar to Varian TrueBeam Machine Commissioning Report (20)

QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptxQUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
QUALITY ASSURANCE IN LINAC AND CYBERKNIFE.pptx
 
LINAC COMMSSN.ppt
LINAC COMMSSN.pptLINAC COMMSSN.ppt
LINAC COMMSSN.ppt
 
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdfQUALITY-ASSURANCE---MAMMOGRAPHY.pdf
QUALITY-ASSURANCE---MAMMOGRAPHY.pdf
 
Quality assurance
Quality  assuranceQuality  assurance
Quality assurance
 
DSD-INT - SWAN Advanced Course - 02 - Setting up a SWAN computation
DSD-INT - SWAN Advanced Course - 02 - Setting up a SWAN computationDSD-INT - SWAN Advanced Course - 02 - Setting up a SWAN computation
DSD-INT - SWAN Advanced Course - 02 - Setting up a SWAN computation
 
Dose reduction technique in ct scan
Dose reduction technique in ct scanDose reduction technique in ct scan
Dose reduction technique in ct scan
 
20 supsi workshop schinke
20 supsi workshop schinke20 supsi workshop schinke
20 supsi workshop schinke
 
UV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.pptUV-VIS SPECTRO final.ppt
UV-VIS SPECTRO final.ppt
 
Vibration monitoring of Gearboxes.pdf
Vibration monitoring of Gearboxes.pdfVibration monitoring of Gearboxes.pdf
Vibration monitoring of Gearboxes.pdf
 
Data Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRA
Data Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRAData Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRA
Data Teknis Gossen Metrawatt Ground Tester : GEOHM PRO & GEOHM XTRA
 
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618
Datasheet Metrel MI 3102H BT. Hubungi PT. Siwali Swantika 021-45850618
 
Mutt_Wind_Tunnel_Results_v2
Mutt_Wind_Tunnel_Results_v2Mutt_Wind_Tunnel_Results_v2
Mutt_Wind_Tunnel_Results_v2
 
First results from the full-scale prototype for the Fluorescence detector Arr...
First results from the full-scale prototype for the Fluorescence detector Arr...First results from the full-scale prototype for the Fluorescence detector Arr...
First results from the full-scale prototype for the Fluorescence detector Arr...
 
High speed board design considerations
High speed board design considerationsHigh speed board design considerations
High speed board design considerations
 
Accelerometers 2015
Accelerometers 2015Accelerometers 2015
Accelerometers 2015
 
LVTS - Image Resolution Monitor for Litho-Metrology
LVTS - Image Resolution Monitor for Litho-MetrologyLVTS - Image Resolution Monitor for Litho-Metrology
LVTS - Image Resolution Monitor for Litho-Metrology
 
Tomotherapy
TomotherapyTomotherapy
Tomotherapy
 
FAST実験7:新型大気蛍光望遠鏡による極高エネルギー宇宙線観測報告
FAST実験7:新型大気蛍光望遠鏡による極高エネルギー宇宙線観測報告FAST実験7:新型大気蛍光望遠鏡による極高エネルギー宇宙線観測報告
FAST実験7:新型大気蛍光望遠鏡による極高エネルギー宇宙線観測報告
 
Effects of crystal size on the imaging performance using two collimator seri...
Effects of crystal size on the imaging performance  using two collimator seri...Effects of crystal size on the imaging performance  using two collimator seri...
Effects of crystal size on the imaging performance using two collimator seri...
 
FAST実験3:新型大気蛍光望遠鏡の試験観測報告
FAST実験3:新型大気蛍光望遠鏡の試験観測報告FAST実験3:新型大気蛍光望遠鏡の試験観測報告
FAST実験3:新型大気蛍光望遠鏡の試験観測報告
 

Recently uploaded

Dehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service Dehradun
Dehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service DehradunDehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service Dehradun
Dehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service DehradunNiamh verma
 
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar SumanCall Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar SumanCall Girls Service Chandigarh Ayushi
 
No Advance 9053900678 Chandigarh Call Girls , Indian Call Girls For Full Ni...
No Advance 9053900678 Chandigarh  Call Girls , Indian Call Girls  For Full Ni...No Advance 9053900678 Chandigarh  Call Girls , Indian Call Girls  For Full Ni...
No Advance 9053900678 Chandigarh Call Girls , Indian Call Girls For Full Ni...Vip call girls In Chandigarh
 
Call Girl Gurgaon Saloni 9711199012 Independent Escort Service Gurgaon
Call Girl Gurgaon Saloni 9711199012 Independent Escort Service GurgaonCall Girl Gurgaon Saloni 9711199012 Independent Escort Service Gurgaon
Call Girl Gurgaon Saloni 9711199012 Independent Escort Service GurgaonCall Girls Service Gurgaon
 
Russian Call Girls Gurgaon Swara 9711199012 Independent Escort Service Gurgaon
Russian Call Girls Gurgaon Swara 9711199012 Independent Escort Service GurgaonRussian Call Girls Gurgaon Swara 9711199012 Independent Escort Service Gurgaon
Russian Call Girls Gurgaon Swara 9711199012 Independent Escort Service GurgaonCall Girls Service Gurgaon
 
💚😋Kolkata Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Kolkata Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Kolkata Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Kolkata Escort Service Call Girls, ₹5000 To 25K With AC💚😋Sheetaleventcompany
 
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in LucknowRussian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknowgragteena
 
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...Niamh verma
 
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.ktanvi103
 
VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Me
VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near MeVIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Me
VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Memriyagarg453
 
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service MohaliCall Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service MohaliHigh Profile Call Girls Chandigarh Aarushi
 
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591adityaroy0215
 
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meetpriyashah722354
 
Bangalore call girl 👯‍♀️@ Simran Independent Call Girls in Bangalore GIUXUZ...
Bangalore call girl  👯‍♀️@ Simran Independent Call Girls in Bangalore  GIUXUZ...Bangalore call girl  👯‍♀️@ Simran Independent Call Girls in Bangalore  GIUXUZ...
Bangalore call girl 👯‍♀️@ Simran Independent Call Girls in Bangalore GIUXUZ...Gfnyt
 
Hot Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In Ludhiana
Hot  Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In LudhianaHot  Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In Ludhiana
Hot Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In LudhianaRussian Call Girls in Ludhiana
 
Vip Kolkata Call Girls Cossipore 👉 8250192130 ❣️💯 Available With Room 24×7
Vip Kolkata Call Girls Cossipore 👉 8250192130 ❣️💯 Available With Room 24×7Vip Kolkata Call Girls Cossipore 👉 8250192130 ❣️💯 Available With Room 24×7
Vip Kolkata Call Girls Cossipore 👉 8250192130 ❣️💯 Available With Room 24×7Miss joya
 
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...Call Girls Noida
 

Recently uploaded (20)

Dehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service Dehradun
Dehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service DehradunDehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service Dehradun
Dehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service Dehradun
 
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar SumanCall Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
Call Girl Price Amritsar ❤️🍑 9053900678 Call Girls in Amritsar Suman
 
No Advance 9053900678 Chandigarh Call Girls , Indian Call Girls For Full Ni...
No Advance 9053900678 Chandigarh  Call Girls , Indian Call Girls  For Full Ni...No Advance 9053900678 Chandigarh  Call Girls , Indian Call Girls  For Full Ni...
No Advance 9053900678 Chandigarh Call Girls , Indian Call Girls For Full Ni...
 
Call Girl Gurgaon Saloni 9711199012 Independent Escort Service Gurgaon
Call Girl Gurgaon Saloni 9711199012 Independent Escort Service GurgaonCall Girl Gurgaon Saloni 9711199012 Independent Escort Service Gurgaon
Call Girl Gurgaon Saloni 9711199012 Independent Escort Service Gurgaon
 
Russian Call Girls Gurgaon Swara 9711199012 Independent Escort Service Gurgaon
Russian Call Girls Gurgaon Swara 9711199012 Independent Escort Service GurgaonRussian Call Girls Gurgaon Swara 9711199012 Independent Escort Service Gurgaon
Russian Call Girls Gurgaon Swara 9711199012 Independent Escort Service Gurgaon
 
💚😋Kolkata Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Kolkata Escort Service Call Girls, ₹5000 To 25K With AC💚😋💚😋Kolkata Escort Service Call Girls, ₹5000 To 25K With AC💚😋
💚😋Kolkata Escort Service Call Girls, ₹5000 To 25K With AC💚😋
 
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service GuwahatiCall Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
Call Girl Guwahati Aashi 👉 7001305949 👈 🔝 Independent Escort Service Guwahati
 
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in LucknowRussian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
Russian Escorts Aishbagh Road * 9548273370 Naughty Call Girls Service in Lucknow
 
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
Call Girls Service Chandigarh Gori WhatsApp ❤7710465962 VIP Call Girls Chandi...
 
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
Call Now ☎ 9999965857 !! Call Girls in Hauz Khas Escort Service Delhi N.C.R.
 
VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Me
VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near MeVIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Me
VIP Call Girls Noida Jhanvi 9711199171 Best VIP Call Girls Near Me
 
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service MohaliCall Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
Call Girls in Mohali Surbhi ❤️🍑 9907093804 👄🫦 Independent Escort Service Mohali
 
#9711199012# African Student Escorts in Delhi 😘 Call Girls Delhi
#9711199012# African Student Escorts in Delhi 😘 Call Girls Delhi#9711199012# African Student Escorts in Delhi 😘 Call Girls Delhi
#9711199012# African Student Escorts in Delhi 😘 Call Girls Delhi
 
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
VIP Call Girl Sector 25 Gurgaon Just Call Me 9899900591
 
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real MeetCall Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
Call Girls Chandigarh 👙 7001035870 👙 Genuine WhatsApp Number for Real Meet
 
Bangalore call girl 👯‍♀️@ Simran Independent Call Girls in Bangalore GIUXUZ...
Bangalore call girl  👯‍♀️@ Simran Independent Call Girls in Bangalore  GIUXUZ...Bangalore call girl  👯‍♀️@ Simran Independent Call Girls in Bangalore  GIUXUZ...
Bangalore call girl 👯‍♀️@ Simran Independent Call Girls in Bangalore GIUXUZ...
 
Hot Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In Ludhiana
Hot  Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In LudhianaHot  Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In Ludhiana
Hot Call Girl In Ludhiana 👅🥵 9053'900678 Call Girls Service In Ludhiana
 
Vip Kolkata Call Girls Cossipore 👉 8250192130 ❣️💯 Available With Room 24×7
Vip Kolkata Call Girls Cossipore 👉 8250192130 ❣️💯 Available With Room 24×7Vip Kolkata Call Girls Cossipore 👉 8250192130 ❣️💯 Available With Room 24×7
Vip Kolkata Call Girls Cossipore 👉 8250192130 ❣️💯 Available With Room 24×7
 
Call Girl Lucknow Gauri 🔝 8923113531 🔝 🎶 Independent Escort Service Lucknow
Call Girl Lucknow Gauri 🔝 8923113531  🔝 🎶 Independent Escort Service LucknowCall Girl Lucknow Gauri 🔝 8923113531  🔝 🎶 Independent Escort Service Lucknow
Call Girl Lucknow Gauri 🔝 8923113531 🔝 🎶 Independent Escort Service Lucknow
 
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
Vip sexy Call Girls Service In Sector 137,9999965857 Young Female Escorts Ser...
 

Varian TrueBeam Machine Commissioning Report

  • 2. • Millenium 120 Leaf HDTM MLC • PerfectPitch TM 6DoF Couch • RGSC Gating • aS1200 MV Imager • Photon Energies :- 6,10,15 MV & 6,10 MV FFF • Electron Energies :- 6,9,12,15,18 MeV & 6,9 MeV HDTSe 2
  • 3. New TrueBeam machine is installed in the existing bunker of Siemens Primus Plus with a maximum photon energy of 15 MV according to AERB guidelines , though maximum energy remains the same, additional shielding of 30 cm concrete was added on the primary as well as secondary wall of area with maximum occupancy 3
  • 4. The acceptance testing implies the verification process of the machine based on manufacture’s guidelines for a very small subset of beam data whereas commissioning is a process where a full set of data is acquired that will be used for patient treatment Manufacturers often provide guidelines and tolerance limits for acceptance testing of a machine through their acceptance testing procedure 4
  • 5. Energy (MV) 6 10 15 6FFF 10FFF Dmax(cm) (±0.15) 1.60 2.4 2.9 1.50 2.34 PDD at 10cm (±1.0) 67.2 74.1 77.4 64.3 71.8 Min D.R (MU/Min) 5 5 20 400 400 Max D.R (MU/Min) 600 600 600 1400 2400 e- energy MeV 6 9 12 15 18 Depth of ionisation1 90% (cm,±0.1) 1.71 2.68 3.77 4.67 5.58 80% (cm,±0.07) 1.90 2.95 4.15 5.2 6.09 50% (cm,±0.1) 2.32 3.52 4.91 6.19 7.41 Radial & Transverse flatness2 ±5% ±4.5% ±4.5% ±4.5% ±4.5% 1Depth of ionization applies to the 15 X15 cm2 applicator field size, using a water phantom at 100 SSD 2Flatness is defined as the maximum variation from the mean electron ionization delivered within the central 80% FWHM region measured for 10x10 cm2 through 25 x 25 cm2 fields. 5
  • 6. ▪ Varian procedure for isocenter verification “Isolock” ▪ MV images – 1. 63 different gantry/collimator angles 2. 13 different couch angles ▪ Central axis s-ray beam variation :- the gantry and/or collimator positions – 0.5 mm gantry,couch, and/or collimator positions – 0.75 mm ▪ EPID-based Winston Lutz (WL) test :- within 1 mm tolerance 6
  • 7. ❑ Congruence between optical and radiation fields Symmetric Fields - 5x5 cm2 : 0.5mm -10x10 cm2 : 1mm -20x20 cm2 :1mm Tolerance Limits = 2mm -30x30 cm2 :1mm Asymmetric fields - defined by over travel option of the collimator - 1mm ❑Overlapping between parallel opposed radiation fields - defined by jaws for 10x10 cm2 Tolerance = 3mm - defined by MLC for 10x10 cm2 0.5 mm 7
  • 8. ▪ Satisfactory acceptance testing simply assures that the accelerator satisfies all agreed-upon specifications and pertinent safety requirements ▪ Commissioning includes comprehensive measurements of dosimetric parameters ▪ also includes - entry of beam data into a treatment planning system - testing of its accuracy - development of operational procedures -training of all concerned with the operation of the accelerator 8
  • 9. 9
  • 10. i. PTW PinPoint® Chamber. ii. Advanced Markus® Electron Chamber iii. PTW Farmer® Ionization Chamber iv. IBA CC13 Ionization Chamber v. IBA RAZOR™ Nano Chamber 10
  • 11. Dmax (cm) 6 MV 1.59 10 M V 2.42 15 MV 2.9 6 MV FFF 1.43 10 MV FFF 2.36 PHOTONS 11
  • 13. SYMMETRY The ratio between measured values for each pair of symmetrical points (with respect to beam axis) for a range of field sizes and gantry orientations must lie between .98 and 1.02 within the central 80% flattened beam area. Field Size (cm2) 6 MV 10 MV 15 MV Flatness Symmetry Flatness Symmetry Flatness Symmetry I P C P I P C P I P C P I P C P I P C P I P C P 05x05 104.3 104.2 100.2 100.3 103.8 104.1 100.2 100.3 103.8 104.3 100.5 100.4 10X10 105.0 105.3 100.4 100.6 104.8 105.1 100.2 100.5 104.2 104.5 100.5 100.5 20x20 104.3 104.4 100.4 100.6 103.5 103.2 100.4 100.6 103.4 103.5 100.7 100.6 30x30 104.2 104.1 100.5 100.9 103.6 103.7 100.5 100.5 104.2 103.9 100.7 100.4 40x40 105.1 105.0 100.6 100.8 104.0 103.6 100.7 100.6 104.3 104.4 100.6 100.5 Flatness for field sizes 5x5 cm2 to 30x30 cm2 : ± 3% & for field size >30x30 cm2: ± 5% Symmetry = ±2% 13
  • 14. 14
  • 15. ❖ In-line, cross-line and diagonal beam profiles were measured for all available beam energies for various recommended field sizes at Dmax, 5 cm, 10 cm, 20 cm and 30 cm depth ❖ Beam profile data were first smoothened by median filter and then corrected for the central axis discrepancy ❖ Beam profiles ware normalize to 100% at the central axis to their corresponding field size PENUMBRA Penumbra for flat beam defined as the lateral separation of (20% - 80%) isodose on either side of beam profile normalized to 100% at the central axis F S 10 X 10 cm2 In Plane Cross Plane Left (mm) Right (mm) Left (mm) Right (mm) 6X 5.55 5.57 5.93 6.01 10X 7.08 7.11 7.21 7.13 15X 7.21 7.29 7.5 7.4 15
  • 16. 16
  • 17. FFF beam analysis done as per AERB Task Group Recommendations All Profiles Above 30 X 30 Field Size are acquired in half profiles 17
  • 18. Measured Diagonal Profiles Calculated Diagonal Profiles 18
  • 19. 19
  • 20. 20
  • 21. According to TRS-398 TPR20,10 = 1.2661 × PDD20,10 − 0.0595 TIME TPR20,10 6 MV 6 MV FFF 10 MV 10 MV FFF 15 MV 10AM 0.6771 0.6381 0.7485 0.7157 0.7724 2PM 0.6754 0.6386 0.7482 0.7154 0.7730 6PM 0.6762 0.6386 0.7483 0.7144 0.7727 RAV 0.6762 0.6384 0.7483 0.7152 0.7727 21
  • 22. ENERGY (MV) DLG (CM) 6 0.0470 10 0.0540 15 0.0560 6 FFF 0.0390 10 FFF 0.0490 22
  • 23. Energy With Couch [MR(nC)] Without Couch [MR (nC) ] Couch Trans. Factor Couch Trans. Factor Thin Thick Thin Thick Thin Thick Avg 6MV 14.3 14.13 14.6 14.6 0.9795 0.9678 0.9736 10MV 15.98 15.84 16.22 16.22 0.9852 0.9766 0.9809 15MV 16.97 16.82 17.16 17.16 0.9889 0.9802 0.9846 6MV FFF 13.7 13.52 14.01 14.01 0.9779 0.9650 0.9714 10MV FFF 15.39 15.25 15.64 15.64 0.9840 0.9751 0.9795 23
  • 25. Applicator / Field Size (cm2) Energy (MeV) Flatness (mm) Symmetry (%) In Plane Cross Plane In Plane Cross Plane Lt Rt Lt Rt 10x10 6 8.5 8.4 9.2 8.7 100.42 100.73 9 7.2 6.9 7.2 7.4 100.7 100.62 12 7.6 8.0 7.3 7.7 101.7 100.93 15 7.6 7.6 7.3 7.8 100.96 100.95 18 7.6 7.6 7.5 7.9 100.67 100.72 25x25 6 9.0 8.2 8.9 8.7 101.37 100.52 9 6.8 6.2 6.6 6.6 101.6 100.82 12 6.9 6.3 6.6 6.6 101.53 100.4 15 7.2 6.8 6.7 6.8 101.41 100.97 18 7.5 7.2 6.9 6.8 100.91 100.82 Tolerance :- should not exceed 103% of the central axis value 25
  • 26. Energy (MeV) Penumbra (mm) In Plane Cross Plane Lt Rt Lt Rt 6 9.8 10.0 9.6 10.0 9 9.6 9.7 9.6 9.8 12 9.6 9.7 9.6 9.8 15 9.8 10.0 9.9 10.0 18 10.1 10.0 10.1 10.2 26
  • 27. 27
  • 28. 28
  • 29. Measured PDD data is calculated and fitted using EMC in Eclipse Treatment planning systrem Open Field Depth Dose - 6 Mev 29
  • 30. 30
  • 32. 32
  • 33. Energy Avg. % Transmission Max. % Transmission X Jaw Y Jaw X Jaw Y Jaw 6 MV 0.41 0.45 0.60 0.70 10 MV 0.48 0.53 0.66 0.76 15 MV 0.5 0.57 0.7 0.83 6MV FFF 0.24 0.26 0.39 0.47 10MV FFF 0.22 0.24 0.34 0.39 33
  • 34. (a) Each beam limiting jaw (excluding MLC jaws) shall attenuate x-radiation such that kerma rate due to transmitted radiation does not exceed 2% of the maximum kerma rate measured along the central axis of the beam at NTD (normal treatment distance) in a 10 x 10 cm radiation beam. (b) For radiation fields of any size, the average kerma rate due to transmission through the beam limiting jaws, including MLC jaws, shall not exceed 0.75% of the maximum kerma rate on the central axis at NTD in a 10 x 10 cm radiation field. 34
  • 35. Measurement:- • SCD 100 cm at a depth of 5 cm in water phantom (RFA) -Interleaf leakage measurements were performed separately for both banks • CC13 ion chamber exactly placed between the two leaves. • Projection of the leaf is properly verified and Meter readings are taken 10 cm off-axis with fully closed MLC • Increment of 1 cm taken in order to place the chamber exactly between the two leaves and chamber moved to next consecutive position with the help of myQA software. - The same procedure is followed for intra leaf transmission except the chamber is placed in center of leaf. • Normalization - 10 × 10 cm2 open fields • 600 MU/Min for FF beam & 1200 & 2400 MU/Min for 6 & 10 MV FFF beams respectively • 100 MU ENERGY % TRANSMISSION Avg. Max 6 MV 1.15 1.42 10 MV 1.42 1.71 15 MV 1.48 1.86 6 MV FFF 0.6 0.86 10 MV FFF 0.58 0.85 35
  • 36. ▪ plane circular surface of radius 2 m centered on and normal to the radiation beam axis at NTD and outside the area of the maximum radiation beam area ▪ Tolerance :- a maximum of 0.2% and an average of 0.1% Energy % Transmission Max Min 6 MV 0.0158 0.0107 10 MV 0.0163 0.0061 15 MV 0.0206 0.0069 6 MV FFF 0.0074 0.0044 10 MV FFF 0.0094 0.0025 Neutron :-max = 0.0131 % tol= 0.05 % avg = 0.0126 % 0.02 % 36
  • 37. at 1 m from the path of the electrons between electron gun to target & reference axis other than patient plane Energy % Transmission Max Avg 6 MV 0.0219 0.0129 10 MV 0.017 0.0050 15 MV 0.0172 0.0074 6 MV FFF 0.0080 0.0043 10 MV FFF 0.0034 0.0020 37
  • 38. 38
  • 39. 39
  • 40. 40
  • 41. COV = (1/ X )*[(  (Xi- X)2 ) /(n-1)] ½ Set kV Output (mGy) Avg X Σ(Xi-X)2 [Σ(Xi-X)2]/n-1 [[Σ(Xi-X)2]/n-1]1/2 COV 1 2 3 50 0.0615 0.0618 0.0613 0.0615 0.00000013 0.00000006 0.000251 0.00408 70 0.131 0.1313 0.1317 0.1313 0.00000024 0.00000012 0.000343 0.002613 80 0.173 0.1724 0.1722 0.1725 0.00000036 0.00000018 0.000422 0.002446 100 0.2618 0.2617 0.2613 0.2616 0.00000012 0.00000006 0.000247 0.000946 120 0.3599 0.3602 0.3584 0.3595 0.00000197 0.00000099 0.000994 0.002764 130 0.4106 0.4107 0.4116 0.411 0.0000006 0.0000003 0.000545 0.001327 140 0.4648 0.4654 0.4645 0.4649 0.00000021 0.00000021 0.000459 0.000987 41
  • 42. Diameter of smallest hole clearly resolved on monitor screen - 0.6mm or 0.023 inch Recommended Performance standard that must be resolved – 3 mm or 0.117 inch Bar strip resolved on the monitor screen - 2.8 lp/mm Recommended performance standard – 1.5 lp/mm must be resolved 42
  • 44. The picket fence test consists of eight consecutive leaf movements of 5 cm wide rectangular fields spaced at 5 cm intervals. The field information is contained in three test files, which are run in sequence. This test is used to verify the leaf positioning accuracy and also calibrates the carriage positioning accuracy 44
  • 45. ▪ Commissioning of Varian TrueBeamTM Linear Accelerator was successfully carried out. Removal flattening filter alters various commissioning associated parameter as Field flatness, Field penumbra, Beam quality, Surface dose,Transmission factor, off axis energy variation and Homogeneity need to be redefined for FFF beam other than FF beam. 45
  • 46. ▪ [1] Das, I.J., Cheng, C.W.,Watts, R.J., Ahnesjö, A., Gibbons, J., Li, X.A., Lowenstein, J., Mitra, R.K., Simon,W.E. and Zhu,T.C. (2008) Accelerator Beam Data Commissioning Equipment and Procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM. Medical Physics, 35, 4186-4215. http://dx.doi.org/10.1118/1.2969070 ▪ [2] Nath, R., Biggs, P.J., Bova, F.J., Ling, C.C., Purdy, J.A., van de Geijn, J. and Weinhous, M.S. (1994) AAPM Code of Practice for Radiotherapy accelerators: Report of AAPM Radiation Therapy Task Group No. 45. Medical Physics, 21, 1093- 1121. (AAPM Report No. 47) ▪ [3] Sahani, G., Sharma, S.D., Sharma, P.K., Deshpande, D.D., Negi, P.S., Sathianarayanan,V.K. and Rath, G.K. (2014) Acceptance Criteria for Flattening Filter-Free Photon Beam from Standard Medical Electron Linear Accelerator: AERB Task Group Recommendations. Journal of Medical Physics, 39, 206-211. http://dx.doi.org/10.4103/0971-6203.144482 ▪ [4] Khan, F.M. (2003) Physics of Radiation Therapy. 3rd Edition, Lippincott Williams & Wilkins. ▪ 5] Musolino, S.V. (2000) Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water. Health Physics, 81, 592-593. (IAEA TRS-398) ▪ [6] Boyer, A., Biggs, P., Galvin, J., Klein, E., LoSasso, T., Low, D., Mah, K. and Yu, C. (2001) Basic Applications of Multileaf Collimator: Report of Task Group No. 50 Radiation Therapy Committee. American Association of Physicists in Medicine (AAPM TG-50, 10) (AAPM Report No. 72). 46
  • 47. [7]Dosimetric Leaf Gap Measurement, Procedure Recommended by Varian Medical System. Eclipse 10 Inverse Planning Administration and Physics rev. 6.1.1. [8] Szpala, S., Cao, F. and Kohli, K. (2014) On Using the Dosimetric Leaf Gap to Model the Rounded Leaf Ends in VMAT/RapidArc Plans. Journal of Applied Clinical Medical Physics, 15, 4484. [9] Acceptance/quality assurance tests for Medical linear accelerator, RPAD/ACC/QA/04 [10]Absorbed dose determination In external Beam radiotherapy An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water -TRS 398 [11]IEC 60601-2-1 Particular requirements for the basic safety and essential performance of electron accelerators in range 1 MeV to 50 MeV [12] IAEA TRS-381 The use of plane-parallel ionization chambers in high-energy electron and photon beams. An international code of practice for dosimetry [13] K. R.Winston and W. Lutz,“Linear accelerator as a neurosurgical tool for stereotactic radiosurgery,” Neurosurgery 22, 454– 464 (1988). [14] Commissioning and Acceptance Testing of the existing linear accelerator upgraded to volumetric modulated arc therapy - Ekambaram Varadharajana and Velayudham Ramasubramanian [15] Shende, R., et al. (2016) Commissioning of TrueBeamTM Medical Linear Accelerator: Quantitative and Qualitative Dosimetric Analysis and Comparison of Flattening Filter (FF) and Flattening Filter Free (FFF) Beam. International Journal of Medical Physics, Clinical Engineering and Radiation Oncology, 5, 51-69. 47
  • 48. 48