More Related Content Similar to Pharmaceutical Calibration.pptx calibration topic Similar to Pharmaceutical Calibration.pptx calibration topic (20) More from ManoharKumar81 (20) Pharmaceutical Calibration.pptx calibration topic2. TABLE OF CONTENTS
2
Calibration
Qualification
o DesignQualification[Dq]
o InstallationQualification[Iq]
o operationalQualification[oq]
o PerformanceQualification[Pq]
Validation
organizationFor Validation
ValidationMasterPlan
ValidationProtocol
3. Typesof Validation
o ProspectiveValidation
o ConcurrentValidation
o RetrospectiveValidation
o Re-Validation
StreamliningValidationoperations
Calibrationv/sValidation
UserRequirementSpecifications
AcceptanceTesting
o FactoryAcceptanceTest
o SiteAcceptanceTest
Calibration AndPreventiveMaintenance
Calibration of AnalyticalBalance 3
4. ○ CALIBRATION
• “Calibration ofaninstrumentis theprocessofdeterminingitsaccuracy.
Theprocessinvolvesobtainingareadingfromtheinstrumentand
measuringitsvariationfromthereadingobtainedfromastandard
instrument.”
• Calibration of an instrument also involves adjusting its precision and
accuracy so that its readings come in accordance with the established
standard.
• Thisis importantforjustifyingtheprocessesofQualificationand
Validation.
• Theinstrumentorequipmentwiththeknownaccuracyis knownas
standards.All theotherinstrumentsaremeasuredagainstthisstandard.
Itis importanttoknowthatthestandardsvaryfromonecountrytothe
otherdependinguponthetypeofindustry.
• CalibrationAchieves2Mainobjectives—
a)Itcheckstheaccuracyofaninstrument
b)Itdeterminesthetraceabilityofthemeasurement
4
6. • WhatIs InstrumentCalibration?
o Instrumentcalibrationcanbedefinedastheprocessof
comparingthemeasurementsmadebytheinstrumenttobe
calibratedagainstaknownmeasurementofeitherstandards
oraninstrumentknowntobemakingmeasurementsthat
exceedtheacceptablelimitsofaccuracyandprecision.
o Usually,calibrationlabspreferastandardwith10timesthe
accuracy; however,mostregulatingorganizationsand
authoritiesalsoaccepta3:1accuracyratio.
6
9. • CommonlyUsedCalibration MethodsandProcedures:
Therearedifferentwaysthatareusedtocalibrateaninstrument.These
methodsarechosenbasedonthedesiredresultsofthecalibrationand
regulatoryauthorities’requirements,likeFDA guidelines.Let uslookatthree
suchprocedures:
• StandardCalibration:This methodis mostlypreferredforcalibrating
instrumentsthatarenon-criticaltoqualityorarenotrequiredforaccreditation
andlicensepurposes.Usetraceablestandardsanddocumentitsperformance.
• CalibrationwithData:Proceduresforcalibrationswithdataaresimilarto
thatofaccreditedcalibration.Theonlyexceptionbeingthattheseprocedures
arenotaccreditedtotheISо standard.Moreover,theyarenotaccompaniedby
dataonmeasurementuncertainties.
• ISo 17025AccreditedCalibration:Thishastobethestrictestmethodof
calibration.Generally,it requires ameasurementreportwhichhasthedetails
ofthemeasurementsthataremadeagainstastandardof‘asfound’(before
calibrationis started)and‘as left’ (oncethecalibrationis completed).If the
calibrationis donebyacalibrationserviceprovider,theymustissuea
certificate ofthesame.
9
10. ImportanceofRegularCalibration:
10
• Calibrationis responsiblefordefiningtheaccuracyofanymeasurement
anditsqualitythatis recordedbyanyinstrument.
Whenyoustartworkingwithanyinstrument,itmustbecalibrated
well,thusassuringyouofaccurateresults.However,overaperiodof
timeyouwill startobservinga‘drift’. Calibrationminimizessuch
uncertaintiesbyassuringtheaccuracyofthetestequipment.
• Whenyouregularlycalibrateyourequipment,youcaneliminatethedrift
atitsbuddingstageinsteadofallowingittogrowtill itaffectsthe
measurementsinsignificantways.
• Calibrationhelpsinquantifyingandcontrollingerrorsanduncertainties
withinvariousmeasurementprocessestoanacceptablelevel.
• Further,ithelpsinimprovingtheaccuracyofthemeasuringdevice,
whichinturnimprovesthequalityoftheendproduct.
• In short,regularcalibrationallowspharmaceuticalcompaniestohave
confidenceintheirresultswhichtheycanrecord,monitorandcontrol.
12. • Qualification ofanalytical instrumentationis essentialfor
accurateandprecisemeasurementofanalyticaldata.If the
instrumentationis notqualified,ensuringthattheresults
indicatedaretrustworthy,all otherworkbasedupontheuse
ofthatinstrumentationis suspect.
• Qualificationofinstrumentsis notasingle,continuous
processbutinsteadresultsfrommanydiscreteactivities.For
convenience,theseactivitieshavebeengroupedinto4phases
ofqualification.Thesephasesaredescribedbelow:
DesignQualification(DQ)
Installation Qualification (IQ)
operationalQualification(oQ)
PerformanceQualification(PQ)
12
13. DesignQualification (DQ):
• It is thedocumentedverificationthattheproposeddesignof
thefacilities,systemsandequipmentis suitableforthe
intendedpurpose.
• DQshouldbeperformedwhennewequipmentis being
purchased,orwhenexistingequipmentis beingusedfora
newapplication.DQservesastheprecursortodefiningthe
equipmentInstallationQualification(IQ) andoQ protocols.
• Thepurposeis toensurethatall therequirementsforthefinal
systemshavebeenclearlydefinedatthestart.
Inotherwords,
“Hasitbeendesignedandselectedcorrectly?”
13
15. InstallationQualification(IQ):
• It is documentedevidencethatthepremises,supportingutilities,the
equipmenthavebeenbuiltandinstalledincompliancewithdesign
specifications
• Itverifiesthattheequipmenthasbeeninstalledinaccordancewith
manufacturersrecommendationinapropermannerandplacedinan
environmentsuitableforitsintendedpurpose.
• It involvestheco-ordinateeffortsofthevendor,theoperatingdepartment
andtheprojectteam.
• ThepurposeofI.Q is tochecktheinstallationsite/environment,confirms
equipmentspecificationsandverifiestheconditionofinstalledequipment;
andalsotoensurethatall aspects(staticattributes)ofthefacilityor
equipmentareinstalledcorrectlyandcomplywiththeoriginaldesign.
In otherwords,
“Hasitbeenbuiltorinstalledcorrectly?”
• InI.Q, connecteachunit(Electricalsystem,Flowlinesystem)andconfirm
thattheconnectionsarecorrect. 15
16. • IQ checkitems:
Equipmentdesignfeatures(i.e. materialofconstructioncleanability,
etc.)
Installationconditions(wiring, utility, functionality, etc.)
Calibration, preventativemaintenance,cleaningschedules.
Safetyfeatures.
Supplierdocumentation,prints,drawingsandmanuals.
Softwaredocumented.
Sparepartslist.
Environmentalconditions(suchascleanroomrequirements,
temperature,andhumidity).
• AnyproblemsidentifiedinI.Q mustbeinvestigatedandappropriate
actionsmustbetaken.All suchactionsmustbedocumentedand
approvedbyhigherauthority.
16
18. • oQ checкitems:
Processcontrol limits(time,temperature,pressure,linespeed,setup
conditions, etc.)
Softwareparameters.
Rawmaterialspecifications
Processoperatingprocedures.
Materialhandlingrequirements.
Processchangecontrol.
Training.
Potentialfailuremodes,actionlevelsandworst-caseconditions.
Theuseofstatisticallyvalidtechniquessuchasscreeningexperimentsto
optimizetheprocesscanbeusedduringthisphase.
• Anyproblemsidentifiedino.Q mustbeinvestigatedand
appropriateactionsmustbetaken.All suchactionsmustbe
documentedandapprovedbyhigherauthority.
18
19. Performancequalification:
• AftertheIQ andoQ havebeenperformed,theinstrument’scontinued
suitabilityforitsintendeduseis provedthroughperformance
qualification.
• Itreferstoestablishingbyobjectiveevidencethattheprocess,under
anticipatedconditions,consistentlyproducesaproductwhichmeetsall
predeterminedrequirements.
• PQ shouldalwaysbeperformedunderconditionsthataresimilar to
routinesampleanalysis.PQ shouldbeperformedonadaily basisor
whenevertheequipmentis beingused.
• PQconsiderationsinclude:
Actualproductandprocessparametersandproceduresestablishedin oQ.
Acceptability oftheproduct.
AssuranceofprocesscapabilityasestablishedinoQ.
Process repeatability, longtermprocessstability.
20
20. • Theobjectiveis toensurethattheinstrumentis performingwithin
specifiedlimits.ThePQ representsthefinalqualificationofequipment
orsystem.
• It isusedtoestablishandorconfirm;
1.Definitionofperformancecriteria andtestprocedures.
2.Selectionofcritical parameters,withpredefinedspecifications.
3.Determinationofthetestintervals,e.g.,
(a)- Everyday.
(b)- Everytimethesystemis used.
(c) - Before,betweenandafteraseriesofruns.
4.Definecorrectiveactionsonwhattodoif thesystemdoesnotmeet
theestablishedcriteria.
20
22. ScopeofPerformanceQualification.
22
• Accordingtoregulatorydocuments,like FDA guidelines,thescopeofPQ
is somewhatlimited.Whileequipmentvalidationteststheability
individuallyforeachpieceofequipment,PQ verifiestheperformanceof
equipment,systemsandfacilitiesasawhole.
• Itrepresentsthefinalqualification,includinganyrequalificationofthe
systemandequipmentthatyouuseinyourbusiness.
• Typically,thescopeofPQextendstoincludethefollowingscenarios:
Newsystemsbeingdeliveredandoperatedforthefirst time
Existing systemsinuse(aspartofaregularmaintenanceschedule)
Systemsthathavebeenmodifiedtoanydegree
Equipment/systemswhichhavebeenusedmorethantheynormally
wouldbe
Afterasystemhasbeenexpandedinordertoincreaseitscapacity
23. FrequencyofPerformingPerformance
Qualification
23
• TheobjectiveofPQ is toprovidequalityassurancethatthesystemis
capableofbeingsubsequentlyvalidated.GMPandothersuch
guidelinesmightnotspecifythefrequencyofperformingPQ,sothe
scheduleorfrequencyyouchoosedependsonalotoffactors.
• Thiswill typicallybeoneormoreofthefollowing:
•Everyday
•Eachtimetheequipmentorsystemis used
•Before,after,orevenduring,aseriesofoperations
•otherperiodicschedule,orasneeded
24. VALIDATIoN:
24
• Validationis anintegralpartofqualityassurance;it involvesthe
systematicstudyofsystems,facilities andprocessesaimedat
determiningwhethertheyperformtheirintendedfunctions
adequatelyandconsistentlyasspecified.
• Avalidatedprocessis onewhichhasbeendemonstratedtoprovide
ahighdegreeofassurancethatuniformbatcheswill beproduced
thatmeettherequiredspecifications andhasthereforebeen
formallyapproved.
• Validationin itselfdoesnotimproveprocessesbutconfirmsthat
theprocesseshavebeenproperlydevelopedandareundercontrol.
25. • Definitions :
• AccordingtoISo:
“Validationis theconfirmationbyexaminationandtheprovisionof
objectiveevidencethattheparticularrequirementsforaspecificintended
usearefulfilled.”
• AccordingtotheUS FoodandDrugAdministration(FDA), thegoalof
validationis to:
“Establishdocumentedevidencewhichprovidesahighdegreeof
assurancethataspecificprocesswill consistentlyproduceaproduct
meetingitspredeterminedspecificationsandqualityattributes.”
• AccordingtoEuropeancommission:
“ActionprovidinginaccordancewiththeprinciplesofGMP,thatany
procedure,process,equipment,material,activityorsystemactuallylead
totheexpectedresults.”
25
26. Adequatevalidationis beneficialtothemanufacturerin manyways:
• Itdeepenstheunderstandingofprocesses;decreasestherisk of
preventingproblemsandthusassuresthesmoothrunningoftheprocess.
• Itdecreasestheriskofdefectcosts.
• It decreasestherisk ofregulatorynoncompliance.
• Afullyvalidatedprocessmayrequirelessin-processcontrolsandend
producttesting.
Validationshouldthusbeconsideredin thefollowingsituations:
• Totallynewprocess;
• Newequipment;
• Processandequipmentwhichhavebeenalteredtosuitchanging
priorities; and
• Processwheretheend-producttestis poorandanunreliableindicatorof
productquality. 27
27. Scopeofvalidation
• Validationrequiresanappropriateandsufficientinfrastructureincluding:
– organization,documentation,personnelandfinances
• Involvementofmanagementandqualityassurancepersonnel
• Personnelwithappropriatequalificationsandexperience
• Extensivepreparationandplanningbeforevalidationis performed
• Validationshouldbeperformed:
– fornewpremises,equipment,utilitiesandsystems,andprocessesandprocedures;
– atperiodicintervals;and
– whenmajorchangeshavebeenmade.
• Validationinaccordancewithwrittenprotocols.
• Validationoveraperiodoftime,e.g.atleastthreeconsecutivebatches(full
productionscale)todemonstrateconsistency.(Worstcasesituationsshouldbe
considered.)
• Significantchanges(facilities,equipment,processes)- shouldbevalidated
• Risk assessmentapproachusedtodeterminethescopeandextentofvalidation
needed 28
30. • Departmentsresponsible
o Site validationcommittee:- Developsitemastervalidationplan.
o Manufacturingdepartment:- Preparesthebatchesasthoughtheir
routineproductionbatches.
o Qualityassurance:- Ensurecomplianceandthatdocumentation,
proceduresarein place.Approvesprotocolsandreports.
o Qualitycontrols:- Performtestingcontractsvalidationtestingand
reviewsprotocolandreportasneeded.
o Researchanddevelopment:- Dealswithproductdesign.
o Engineeringdepartment:- Installation, qualityandcertifyplant,
facilities, equipmentandsupportsystems.
30
33. • Departmentinteraction:
once the validation team has been constituted and mission
have been formalized, the team will interact with different
departmentswhichare:
Researchanddevelopmentdepartment
Engineering department,
Productiondepartment,
Maintenancedepartment,
Qualitycontroldepartment,
Qualityassurancedepartment.
33
34. VALIDATIoN MASTER PLAN
34
• Thevalidationmasterplanshouldprovideanoverviewoftheentire
validationoperation,itsorganizationalstructure,itscontentand
planning.
• Themainelementsofitbeingthemlist/inventoryoftheitemstobe
validatedandtheplanningschedule.
• All validationactivitiesrelatingtocritical technicaloperations,relevant
toproductandprocesscontrolswithinafirmshouldbeincludedinthe
validationmasterplan.
Itshouldcompriseall prospective,concurrentandretrospective
validationsaswellasrevalidation.
• TheValidationMasterPlanshouldbeasummarydocumentandshould
thereforebebrief,conciseandclear.
• Itshouldnotrepeatinformationdocumentedelsewherebutshouldrefer
toexistingdocumentssuchaspolicydocuments,SoP’s andvalidation
protocolsandreports.
35. • Theformatandcontentshouldinclude:
Introduction:validationpolicy, scope,locationandschedule.
organizationalstructure: personnelresponsibilities.
Plant/process/productdescription:rationalforinclusions or
exclusionsandextentofvalidation.
Specific processconsiderationsthatarecritical andthose
requiringextraattention.
Keyacceptancecriteria.
Documentationformat.
ReferencetotherequiredSoPs.
Timeplansofeachvalidationprojectandsub-project.
List ofproducts/processes/systemstobevalidated,summarizedin
amatrixformat,validationapproach.
Re-validationactivities,actualstatusandfutureplanning
35
36. VALIDATIoN PRoTocoL
36
• Awrittenplanstatinghowvalidationwillbeconducted,includingtest
parameters,productcharacteristics,productionandpackaging
equipment,anddecisionpointsonwhatconstitutesacceptabletest
results.
• This document should give details of critical steps of the manufacturing
process that should b
e measured, the allowable range of variability and
themannerinwhichthesystemwillbetested.
• Thevalidationprotocolprovidesasynopsisofwhatis hopedtobe
accomplished.
• Theprotocolshouldlist theselectedprocessandcontrolparameters,
statethenumberofbatchestobeincludedinthestudy,andspecifyhow
thedata,onceassembled,will betreatedforrelevance.Thedateof
approvalbythevalidationteamshouldalsobenoted.
• Inthecasewhereaprotocolis alteredormodifiedafteritsapproval,
appropriatereasoningforsuchachangemustbedocumented.
39. TYPES oF VALIDATIoN
39
Prospectivevalidation
• Itis definedastheestablisheddocumentedevidencethatasystemdoes
whatitpurportstodobasedonapre-plannedprotocol.
• Thisvalidationusuallycarriedoutpriortodistributioneitherofanew
productoraproductmadeunderarevisedmanufacturingprocess.
• Performedonatleastthreesuccessiveproduction-size (Consecutive
batches).
• Theobjectiveoftheprospectivevalidationis toproveordemonstratethat
theprocesswillworkinaccordancewithvalidationprotocolprepared
forthepilotproductiontrials.Prospectivevalidationshouldnormallybe
completedpriortothedistributionandsaleofthemedicinalproduct.
• InProspectiveValidation,thevalidationprotocolis executedbeforethe
processis putintocommercialuse.
40. concurrent validation
• It is aprocesswherecurrentproductionbatchesareusedtomonitor
processingparameters.
• ConcurrentValidationmeansestablishingdocumentedevidencea
processdoeswhatitis supposedtobasedondatageneratedduring
actualimplementationoftheprocess.
• Itis importantinthesecaseswhenthesystemsandequipmenttobe
usedhavebeenfullyvalidatedpreviously.
• It is similartoprospective,excepttheoperatingfirmwill sell the
productduringthequalificationruns,tothepublicatitsmarketprice,
andalsosimilartoretrospectivevalidation.
• This validationinvolves in-processmonitoringofcritical processing
stepsandproducttesting.Thishelpstogenerateanddocumented
evidencetoshowthattheproductionprocessis inastateofcontrol.
40
43. sTREAMLINING VALIDATIoN oPERATIoNs
43
• Thebestapproachtoavoidingneedlessandexpensivetechnicaldelays
is toworkinparallel. Thekeyelementsatthisimportantstageofthe
overallprocessaretheAPI, analyticaltestmethods,andthedrug
product(pharmaceuticaldosageform).Anintegratedandparallelway
ofgettingthesethreevitallyimportantfunctionstoworktogetheris
depictedin Figurebelow.
• Figureshowsthattheuseofasingleanalyticalmethodstesting
functionis animportanttechnicalbridgebetweentheAPI andthedrug
productdevelopmentfunctionsasthelattertwomovethroughthe
variousstagesofdevelopment,clinicalstudy,processdevelopment,
andprocessvalidationandintoproduction.
45. cALIBRATIoN v/sVALIDATIoN
45
• Calibrationandvalidationaretwoprocessesin manufacturingto
guaranteethequalityoftheproductorrelatedapparatus.
• Withthecalibration,themeasurementsarecomparedwithanaccepted
referencemeasurement,toassuretheconsideredmeasurements
complywiththerequirements.
• Withthevalidation,theperformance,quality,andotheroperating
parametersofasystemaretestedtoverifythattheycomplywiththe
requirements.
46. cALIBRATIoN v/sVALIDATIoN
cALIBRATIoN
• calibrationis ademonstration
that,aparticular Instrumentor
deviceproducesresultswithin
specifiedlimitsbycomparisons
withthoseproducedbyareference
ortraceablestandardoveran
appropriaterangeof
measurements.
• In calibrationperformanceofan
instrumentordeviceis comparing
againstareference standard.
46
VALIDATIoN
• Validationis adocumented
programthatprovideshighdegree
ofassurancethataspecific
process, equipment,methodor
systemconsistently producesa
resultmeetingpre-determined
acceptancecriteria.
• Nosuchreference standardsare
usinginvalidation program.
47. cALIBRATIoN v/sVALIDATIoN
cALIBRATIoN
• Calibrationensuresthatinstrument
ormeasuringdevicesproducing
accurateresults.
• Shall b
e performed periodically, to
identify the ‘drift’ of the measuring
device or equipment and make them
accurate.
• Shall b
e performed as per calibration
SoP.
VALIDATIoN
• Validationprovidesdocumented
evidencethataprocess, equipment,
methodorsystemproducesconsistent
results(in otherwords,itensuresthat
uniformsbatchesareproduced).
• Nosuchrequirements.Shall be
performedwhenchangesor
modificationshappentotheexisting
systemoroncerevalidationperiodis
reached.
• Shall beperformedaspervalidation
protocol.
47
48. UserRequirementsspecification(URs).
• This focusesonthe‘‘what’’ratherthanthe‘‘how.’’
48
• TheUserRequirementsSpecificationdescribesthebusinessneedsforwhat
usersrequirefromthesystem.
• UserRequirementsSpecificationsarewrittenearlyinthevalidationprocess,
typicallybeforethesystemis created.Theyarewrittenbythesystemownerand
end-users,withinputfromQualityAssurance.
• RequirementsoutlinedintheURSareusuallytestedinthePerformance
QualificationorUser AcceptanceTesting.
• UserRequirementsSpecificationsarenotintendedtobeatechnicaldocument;
readerswithonlyageneralknowledgeofthesystemshouldbeableto
understandtherequirementsoutlinedintheURS.
• AURSdefinesclearly andprecisely, whatthecustomer(i.e.you)wantsthe
systemtodo,andshouldbeunderstoodbyboththecustomerandthe
instrumentvendor.
• TheURSis alivingdocument,andmustbekeptupdated,viaachangecontrol
procedure.
49. • Awell-writtenURs providesseveralspecificbenefits,asit:
Servesasareferenceagainstwhichoff-the-shelfcommercialproductsareselected,
evaluatedindetail,andanyenhancementsaredefined.
Reducesthetotalsystemeffortandcosts,sincecareful reviewofthedocument
shouldrevealomissions,misunderstandingsand/orinconsistenciesinthe
specificationandthismeansthattheycanbecorrectedeasilybeforeyoupurchase
thesystem.
Providestheinputtouseracceptancetestspecificationsand/orqualificationofthe
system.
• TheURs shouldinclude:
o Introduction–includingthescopeofthesystem,keyobjectivesfortheproject,and
theapplicableregulatoryconcerns
o ProgramRequirements–thefunctionsandworkflowthatthesystemmustbeable
toperform
o DataRequirements–thetypeofinformationthatasystemmustbeabletoprocess
o Life cycleRequirements–includinghowthesystemwillbemaintainandusers
trained
50
50. • GeneralGuidanceforWritingtheRequirements
Thefollowingguidelinesshouldbefollowedduringtheproductionofthe
specification:
Eachrequirementstatementshouldbeuniquelyreferencedandnolongerthan
250words.
TheURS shouldbeconsistentandrequirementstatementsshouldnotbe
duplicatedorcontradicted.
Specify requirementsandnotdesignsolutions.Thefocusshouldbeonwhatis
required,butnothowitis tobeachieved.
Eachrequirementshouldbetestable.This allowstheteststobedesignedassoon
astheURSis finalised.
Boththecustomerandthevendormustunderstandthedocument.Therefore,
jargonshouldbeavoidedwhereverpossibleandkeywordsaredefinedina
specific sectioninthedocument.
Requirementsshouldbeprioritisedasmandatoryordesirable.
TheURS shouldbemodifiablebutchangesshouldbeunderaformalcontrol
procedure.
AURS is correctif everystatedrequirementhasonlyoneinterpretationandis
metbythesystem.Unfortunately,thisis veryrare.
50
51. AccEPTANcE TEsTING
51
• ThequestionofwhatUSER ACCEPTANCE TESTING (UAT)is, becomesvery
importantasyoureachastateofhavingequipment/systemsdeliveredtoa
facility.
• Thesystem/equipmentis claimedtobeconstructedinamannerthathasbeen
definedbyaUserRequirement,andconstructedorfabricatedinamannerthat
meetstherequirementsoftheindustryandofyourcompanyspecifically.You
canill affordtoacceptasystemorpieceofequipmentthatdoesnotmeetthese
requirements.
• WhatisAcceptanceTesting?
UserAcceptanceTestingdefinespreciselyandclearlywhattheuserexpects
thesystemtodo.UATdocumentscontaininformationabouttheoperating
environment,therequireddataforprocessing,andthefunctionalitythatthe
systemshouldcarry out.
CompletedduringaFAT andaSAT
52. WhatisaFAT?
AFAT orFactoryAcceptanceTestis usuallypreformedatthevendorpriorto
shippingtoaclient.Thevendorteststhesysteminaccordance withtheclients
approvedtestplansandspecificationstoshowthatsystemis atapointtobe
installedandtestedonsite.
It’s anessentialaspectofthewholesystemlifecycle andshouldbeperformed
byexperiencedpersonnel.TimespentdoingaproperFAT willleadtofewer
problemswhentheequipmentis installed.
summary- It is thepartialcommissioningandqualificationofequipment
and/orsystemspriortotheirshipmentfromthefabricator’ssite.
TheFAT protocolis aninspectionthatincludes bothstaticanddynamic
exhaustivetestingofsystemsormajorsystemcomponentstosupportthe
qualificationofequipmentorasystem.Thetestsmustverifythatall
functionality detailedin theUserRequirementsSpecification(URS) is
embodiedandperformsasspecified. It is writtenbythemanufacturersand
executedbytheclientorclientrepresentative.
52
53. WhatisasAT?
sAT is SiteAcceptanceTestofasystemtoensureitis testedin
accordancetoclientapprovedtestplansandspecifications&toshow
thesystemis installedproperlyandinterfaceswithothersystemsand
peripheralsinitsworkingenvironment.
summary- Inspectionand/ordynamictestingofthesystemsormajor
systemcomponentstosupportthequalification ofanequipment
systemconductedanddocumentedatthemanufacturingsite.
TheSATis relatedtotheFAT andalsoentailsinspectionanddynamic
testingofsystemsormajorsystemcomponentstosupportthe
qualificationofequipment.Thisis writtenbytheclientandverifiesthat
theinstalledfunctionalityoftheequipmentmeetsorexceedsthe
operationalrequirementsasspecifiedintheequipmentURS.
TheSATis executedoncompletionofall commissioningtasks;butprior
tothestartofInstallationQualificationexecution.
53
54. cALIBRATIoN AND PREVENTIVE
MAINTENANcE
54
• Regulations of the regulatory authorities like FDA and EU require that
all the firms have program for calibration and preventive maintenance
fortestaswellasmeasurementequipments.
• Preventive maintenance program is one of the most importance aspects
forGMPinspectionasitensurestheefficientGMPoperations.
• Anyequipmentseitheritis automaticormanuallyoperatedwill
performitsfunctionsproperlyandareusedformanufacturing,
processing,packaging,labelingorholdingofdrugproducts;itis
mandatorythatitwill betimelycalibrated,inspectedandcheckedfor
errorsaccordingtothewrittenprogramwhichis speciallydesignedto
assurethebestperformanceoftheequipments.
55. INTRoDUcTIoN
55
• Definitionofcalibration:- "It is asetofoperationthatperforms
underspecificconditionstoverifythevalues/dataobtainedby
comparisonoftwoinstrumentsormeasuringdevicesoneofwhichis a
standardofknownaccuracy(Traceabletonationalstandards).It is used
todetect,correlate,reportoreliminateanyofthediscrepancyin
accuracyofinstrumentsormeasuringdeviceswhenbeingcomparedto
thestandard."
• Calibrationisonetypeofcomparisonbutitis notanadjustments.
• DefinitionofPreventiveMaintenance:"It is acareorservice
providedbypersonneltomaintaintheequipmentorfacility in
satisfactorilyworkingconditionsbyprovidinginspections,detectionand
correctionoffailuresbeforetheyoccur.Basicallytheyareconductedto
keeptheinstrumentinworkingconditionsandtoextendthelifeofthe
instrument."
58. ○ softwareusedforcalibrationandPM
• LotsofsoftwareareusedforcalibrationandPMatindustrialscale.This
computerizedsystemhasmoreefficiency andguaranteedthebestresults.
○ calibrationsoftware
• QualityCalibrationManagementsystem(QCMS)is completeinstrument
anddesignedaccordingtotherequirementsof21CFR part11.This
softwareensuresregulatorycomplianceandalsotraceability.It will help
toimprovethereliabilityofplantandoptimizetheadministrativecosts.It
will helptoincreaseintheproductivityaswellasefficiency.
○ Preventivemaintenancesoftware
• RCMturbois verypopularPMsoftware.Traditionalapproachtook
yearstocompletejustonePMbutbyusingthisRCMturbo,anyonecan
quicklygoforPM.ThissoftwaredirectsusthroughEMEAprocess.But
thethingis itwillgothroughquicklyandefficiently.It is 100% reliable
andwill allowyoudorisk assessment. 59
59. cALIBRATIoN oF ANALYTIcAL BALANcE
59
Measurethepanpositionerrorofthebalance
Dailycalibration
Themaximumandminimumweightlimitsofthebalancearetaken
anddividedintofourtofivepartsandsinglestandardweightis
selectedin thosefourorfivedifferentregionsforthepurposeof
dailycalibration.
Thevariationallowedfromthestandardweightusedformeasuring
is NMT±0.1%ofstandardweight.
Driftcheckfromdaytodayis carriedoutusinganyparticular
weightandthedeviationallowedshouldnotbemorethan±1%of
thatweight.