Dehradun Call Girls Service ❤️🍑 8854095900 👄🫦Independent Escort Service Dehradun
medical -laboratory- quality- control --
1.
2.
3. What is Quality Control?
Quality control in the medical laboratory
it is statistical process used to monitor and evaluate the
analytical process that produces patient results.
QC results are used to validate whether the instrument is
operating within pre-defined specifications, means that
patient results are reliable, and can be used for diagnosis,
prognosis and treatment
4. QUALITY: A DEGREE OF EXCELLENCE
• QUALITY CONTROL (INTER.)
+
QUALITYASSESSMENT
• (EXTR.)
QUALITY ASSURANCE
20
5. QCv/sQA
Quality Control -QC refers to operational techniques that must
be included during assay run to verify that the requirementseach
for Quality are met wit h
Quality Assurance - QA refers to all those planned and systematic
activities to provide confidence that the results given out by the
laboratory are correct
The aim of QC is simply to ensure that the results generated by the
test are correct.
However quality assurance is concerned with much more: that the
right test is carried out on the right specimen and that the right
result and right interpretation is delivered to the right person at
the right time
23
•
6. QUALITY ASSURANCE (QA)
• The purpose of QA is the maintenance of the overall
quality of patient results
• All factors that
the test
effect the test results from the time
Post-analytic
•
Pre-analytic Analytic
• •
Result accuracy
Analytical errors
Assay repeat rates
Specimen collection
Specimen transport
Specimen quality
24
Result reporting
Record keeping for
patient and QC
7.
8. How to implement a QC program?
– Establish written policies and procedures
– Assign responsibility for monitoring and reviewing
– Train staff
– Obtain control materials
– Collect data
– Set target values (mean, SD)
– Establish Levey-Jennings charts
– Routinely plot control data
– Establish and implement troubleshooting and
corrective action protocols
– Establish and maintain system for documentation
11. Quantitative QC - Module 7 11
Calibrators
A substance with a specific
concentration.
Used to adjust instrument,
kit, test system in order to
standardize the assay
Calibrators are used to set
(calibrate) the measuring
points on a scale.
Sometimes called a standard,
although usually not a
true standard and not a control
Controls
Known concentration of
the analyte
Use 2 or three levels of
controls
A substance similar to
patients’ samples that
has an established
concentration.
Controls are used to
Ensure the procedure is
working properly.
1 2 3 4 5
1 2 3 4 5
12. Quantitative QC - Module 7 12
Control Materials
ASSAYED
Target value predetermined
Verify and use
UNASSAYED
Target value not predetermined
Full assay required before using
“IN-HOUSE”
In-house pooled sera
Full assay, validation
13. Quantitative QC - Module 7 13
Choosing Control Materials
• values cover medical decision points
• similar to the test sample
• controls are usually available in high, normal, and low ranges
14. Quantitative QC - Module 7 14
Preparation and Storage of
Control Material
• adhere to manufacturer’s
instructions
• keep adequate amount
of same lot number
• store correctly
CONTROL
CONTROL
15. Quantitative QC - Module 7 15
Steps in Implementing Quantitative QC
• obtain control material
• run each control 20
times over 30 days
• calculate mean and +/-1,2,3
Standard Deviations
Mean
1SD
1SD
2SD
3SD
2SD
3SD
16. Quantitative QC - Module 7 16
Measurement of Variability
Variability is a normal occurrence when a control is
tested repeatedly
Affected by:
Operator
technique
Environmental
conditions
Performance
characteristics
of the
measurement
The goal is to differentiate between
variability due to chance from that due
to error
17. Control Charts
• A common method to compare the values observed for
control
control
Simple
materials with their known values is the use of
charts
graphical displays in which the observed values are
•
plotted versus the time when the observations are made
Known values are represented by an acceptable range of
values
•
• When plotted points falls
is performing properly
When point s falls out side
developing
within the control limit- method
• control limit - problem may be
73
18. Quantitative QC - Module 7 18
Accuracy and Precision
Accuracy closeness of measurements to the true value
Precision amount of variation in the measurements
Bias The difference between the expectation of a test
result and an accepted reference value
19. 19
Measures of Central Tendency
Mode the value which occurs with the greatest
frequency
Median the value at the center or midpoint of the
observations
Mean the calculated average of the values
20. Quantitative QC - Module 7 20
Measures of Central Tendency
Although variable, sets of data are distributed
around a central value
Measurement
F
r
e
q
u
e
n
c
y
21. Quantitative QC - Module 7 21
Not all central values are the same
Mode
Median
Mean
Measurement
F
r
e
q
u
e
n
c
y
22. Quantitative QC - Module 7 22
Symbols Used in Calculations
∑ is the sum of (add data points)
n = number of data points
x1 - xn = all of the measurements
(1 through n)
__
X represents the mean
23. Quantitative QC - Module 7 23
Standard Deviation (SD)
1
n
)
x
(x 2
1
SD
SD is the principle measure of
variability used in the laboratory
Standard Deviation – Statistical Formula
24. Quantitative QC - Module 7 24
Standard Deviation and Probability
For a set of data with a normal
distribution, a random
measurement will fall within:
+ 1 SD 68.3% of the time
+ 2 SD 95.5% of the time
+ 3 SD 99.7% of the time
68.2%
95.5%
99.7%
Frequency -3s- 2s -1s Mean +1s +2s +3s
X
25. Quantitative QC - Module 7 25
Coefficient of Variation
The coefficient of variation (CV) is the SD expressed
as a percentage of the mean.
• CV is used to monitor precision
• CV is used to compare methods
• CV ideally should be less than 5%
%
100
x
mean
SD
CV
27. Quantitative QC - Module 7 27
Statistics for Quantitative QC
assay control material at least 20
data points over a 20-30 day period
ensure procedural variation is
represented
calculate mean and + 1, 2 and 3 SD
28. Quantitative QC - Module 7 28
Draw lines for Mean and SDs
(calculated from 20 controls)
MEAN
+1SD
+2SD
-1SD
-2SD
-3SD
+3SD
Days
190.5
192.5
194.5
196.5
188.5
186.5
184.6
Chart name: Lot number:
30. Quantitative QC - Module 7 30
Number of Controls
Interpretation depends on number of controls
run with patients’ samples.
• Good: If one control:
– accept results if control is within ± 2SD
unless shift or trend
• Better: If 2 levels of controls
– apply Westgard multirule system
31. Quantitative QC - Module 7 31
Detecting error
• random error: variation in QC results with no
pattern- only a cause for rejection if outside
2SDs.
• systematic error: not acceptable, correct the
source of error
Examples:
– shift–control on one side of the mean 6 consecutive
days
– trend–control moving in one direction– heading
toward an “out of control” value
34. Westgard Rules
• “Multirule Quality Control”
• Uses a combination of decision criteria or control rules
• Allows determination of whether an analytical run is “in-
control” or “out-of-control”
(Generally used where 2 levels of control material are
analyzed per run)
12S rule
13S rule
22S rule
R4S rule
41S rule
10X rule
35. 12S Rule = A warning to trigger careful inspection of
the control data
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
“warning rule”
One of two control results falls
outside ±2SD
Alerts tech to possible problems
Not cause for rejecting a run
Must then evaluate the 13S rule
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
12S rule
violation
36. 13S Rule = Reject the run when a single control measurement
exceeds the +3SD or -3SD control limit
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
13S rule
violation
•If either of the two control
results falls outside of ±3SD,
rule is violated
•Run must be rejected
•If 13S not violated, check 22S
37. 22S Rule = Reject the run when 2 consecutive control
measurements exceed the same
+2SD or -2SD control limit
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
22S rule
violation
•2 consecutive control values for
the same level fall outside of
±2SD in the same direction,
or
•Both controls in the same run
exceed ±2SD
•Patient results cannot be
reported
•Requires corrective action
38. R4S Rule = Reject the run when 1 control
measurement exceed the +2SD and the other exceeds
the -2SD control limit
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
R4S rule
violation
One control exceeds the
mean by –2SD, and the
other control exceeds the
mean by +2SD
The range between the two
results will therefore exceed
4 SD
Random error has occurred,
test run must be rejected
39. Westgard – 41S Rule
Requires control data from previous runs
Four consecutive QC results for one level of control are
outside ±1SD, or
Both levels of control have consecutive results that are
outside ±1SD
40. 10x Rule = Reject the run when 10 consecutive control
measurements fall on one side of the mean
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24
Mean
Day
+1SD
+2SD
+3SD
-1SD
-2SD
-3SD
10x rule
violation
Requires control data from previous runs
Ten consecutive QC results for one level of control are on one side of the mean,
or
Both levels of control have five consecutive results that are on the same side of
the mean
42. When a rule is violated
• Warning rule = use other rules to inspect the
control points
• Rejection rule = “out of control”
–Stop testing
–Identify and correct problem
–Repeat testing on patient samples and
controls
–Do not report patient results until problem
is solved and controls indicate proper
performance
43. Quantitative QC - Module 7 43
When a rule is violated
• Warning rule = use other rules to
inspect the control points
• Rejection rule = “out of control”
–Stop testing
–Identify and correct problem
–Repeat testing on patient
samples and controls
–Do not report patient results
until problem is solved and
controls indicate proper
performance
44. External Quality Assessment
EQA results evaluate performance of the laboratory
against other laboratories participating in the same
Program
Different programs do this in different ways. Is used to test the
statistical significance of any difference between an individual lab’s
observed mean & the group mean
• When the difference is significant lab. is alerted
• EQA results should be formally documented within the lab and
should be available on request
90
45. Proficiency testing
Laboratories should all enroll and satisfactorily participate in a
performance evaluation/assessment program
If proficiency testing is not available, the
laborat ory must exercise an alt ernat ive performance
assessment system for determining the reliability of
analytic testing (sample splitting for inter-laboratory
testing)
If the lab has more than one method-system for
performing tests for an analyte, it must be checked
against each ot her at least twice a year for correlat ion of
patient results
92