This document provides an overview of the user software for the Maglumi X3 analyzer. It contains 3 parts:
Part 1 describes the homepage which provides shortcuts to status information, error logs, and maintenance records.
Part 2 covers status monitoring, including statuses for consumables, waste, environment parameters, and definitions of status indicators.
Part 3 outlines the settings section, including assay parameters, dilution and reflex rules, user management, alarms, maintenance schedules and more.
4. FOCUS MAKES PROFESSION P4
Homepage
Shortcut to
consumable
status
Shortcut to
waste status
Shortcut to environmental
parameter
Shortcut to
maintenance
record
Error log
7. FOCUS MAKES PROFESSION P7
02
PART TWO
Status
Icon color changes from green to yellow or red when status is abnormal
8. FOCUS MAKES PROFESSION P8
02
PART TWO
Status-Cuvette
Renew the
cuvette
quantity in
engaged
cuvette box
after setting
9. FOCUS MAKES PROFESSION P9
02
PART TWO
Definitions
Remaining volume of system liquid is detected by the liquid level sensor and has its status color-coded.
10. FOCUS MAKES PROFESSION P10
02
PART TWO
Definitions
Remaining volume of starters are detected by the liquid level sensor and has its status color-coded.
11. FOCUS MAKES PROFESSION P11
02
PART TWO
Analyzer Status-Disk & SPS
Disk Status: Check the status
of the cuvettes on board
12. FOCUS MAKES PROFESSION P12
02
PART TWO
Definitions
Liquid level of waste liquid are detected by the liquid level sensor and has its status color-coded.
13. FOCUS MAKES PROFESSION P13
02
PART TWO
Definitions
The quantity and status of waste cuvette and empty cuvette box, the status are color-coded.
Reset the counting
after empty the
cuvette dustbin.
14. FOCUS MAKES PROFESSION P14
02
PART TWO
Environment Param.- Temperature, Voltage & Humidity
Check the temperature in different unit.
The temperature, voltage and humidity status are color-coded.
16. FOCUS MAKES PROFESSION P16
03
PART THREE
Setting
Test priority on
samples or assays
Serial number,
software &
firmware version
Global initialization Local initialization
Lis Setting
17. FOCUS MAKES PROFESSION P17
03
PART THREE
Assay Param.
Select whether reagents of the same
Lot-No. will share the calibration data.
Calculate the results exceeding the linear
range; the upper bound of extended
calculation can be set.
18. FOCUS MAKES PROFESSION P18
03
PART THREE
Assay Param.- Master Curve
10-points master curve.
No changing the RLU or conc.
Value, or it will affect the
reagent performance.
21. FOCUS MAKES PROFESSION P21
03
PART THREE
Auto Dilution
Auto Dilution setting
Select Assay
Select Dilution Ratio
Input Threshold
Click Add button
22. FOCUS MAKES PROFESSION P22
03
PART THREE
Reflex Assay
Start an associated assay when the test results
of the current assay are within the reflex range.
Reflex assay setting
Select Assay
Select reflex assay
Input low & high value
Click Add button
23. FOCUS MAKES PROFESSION P23
03
PART THREE
Profile Def.
Use several assays to create a profile, which is displayed in the Profile Selection Area.
Use one button to allocate several assays to one sample for faster sample registration.
Profile definition
Input profile name
Select assays
Click Add button to
create the profile
24. FOCUS MAKES PROFESSION P24
03
PART THREE
Result format
Define number of decimal digits corresponding to measurement ranges
27. FOCUS MAKES PROFESSION P27
03
PART THREE
Users Mgmt
User management: add, edit and delete username, password and user rights
Administrator User : Right to use all functions of
the software.
Normal User : can’t modify or delete test result,
calibration or master curve.
28. FOCUS MAKES PROFESSION P28
03
PART THREE
Loudspeaker
Define when loudspeaker alarm and select alarm music
29. FOCUS MAKES PROFESSION P29
03
PART THREE
Alarm Shielding
Check the box to shield the alarm
30. FOCUS MAKES PROFESSION P30
03
PART THREE
General Setting
Emergency mode enables testing
with a non-working barcode reader