The document provides information about various medical devices used to record electrical activity in the body, including electrocardiograms (ECGs), electroencephalograms, electromyograms, and electrooculograms. It then discusses the history and development of the ECG, including early experiments in the late 18th century. The document outlines the basic anatomy and function of the heart, including the components that receive and pump blood. It describes the electrical conduction system of the heart and the patterns of excitation. The document explains the various waves that make up the ECG, including the P, QRS, T waves, and relates them to electrical and muscular events in the heart. It provides diagrams of heart anatomy and the instrumentation used
The property of automaticity of the sinus node is responsible foe the impulse initiation and travels along the cardiac tissue as depolarizations which result in its contraction. So, when activated, the heart is a concentrated locus of time varying potentials in the body. These voltage fluctuations can be measured by the placement of electrodes on the surface of the body. This forms the basis of electrocardiography. In this presentation we will see the basics, the lead systems and the principles behind recording of ECG.
Biatrial enlargement is diagnosed when criteria for both right and left atrial enlargement are present on the same ECG.
The diagnosis of biatrial enlargement requires criteria for LAE and RAE to be met in either lead II, lead V1 or a combination of leads.
Denoising of ECG -- A discrete time approach using DWTIJERA Editor
This paper is about denoising of ECG signal using DWT transform. In this paper, ECG signals are denoised using DWT transform.Ecg signals are taken and noise at different frequencies are generated which are superimposed on this original ecg signal.High frequency noise is of 4000 hertz and power line interference is of 50 hertz.Decomposition of noisy signal is achieved through wavelet packet .wavelet packets are reconstructed and appropriate wavelet packets are combined to obtain a signal, very similar to original ecg signal.This technique results in the minimization of mean square error in the filtered signals.
The property of automaticity of the sinus node is responsible foe the impulse initiation and travels along the cardiac tissue as depolarizations which result in its contraction. So, when activated, the heart is a concentrated locus of time varying potentials in the body. These voltage fluctuations can be measured by the placement of electrodes on the surface of the body. This forms the basis of electrocardiography. In this presentation we will see the basics, the lead systems and the principles behind recording of ECG.
Biatrial enlargement is diagnosed when criteria for both right and left atrial enlargement are present on the same ECG.
The diagnosis of biatrial enlargement requires criteria for LAE and RAE to be met in either lead II, lead V1 or a combination of leads.
Denoising of ECG -- A discrete time approach using DWTIJERA Editor
This paper is about denoising of ECG signal using DWT transform. In this paper, ECG signals are denoised using DWT transform.Ecg signals are taken and noise at different frequencies are generated which are superimposed on this original ecg signal.High frequency noise is of 4000 hertz and power line interference is of 50 hertz.Decomposition of noisy signal is achieved through wavelet packet .wavelet packets are reconstructed and appropriate wavelet packets are combined to obtain a signal, very similar to original ecg signal.This technique results in the minimization of mean square error in the filtered signals.
Denoising of ECG -- A discrete time approach using DWT.IJERA Editor
This paper is about denoising of ECG signal using DWT transform. In this paper, ECG signals are denoised using DWT transform.Ecg signals are taken and noise at different frequencies are generated which are superimposed on this original ecg signal.High frequency noise is of 4000 hertz and power line interference is of 50 hertz.Decomposition of noisy signal is achieved through wavelet packet .wavelet packets are reconstructed and appropriate wavelet packets are combined to obtain a signal, very similar to original ecg signal.This technique results in the minimization of mean square error in the filtered signals.
Electrocardiography: is the recording of the electrical impulses that are generated in the heart. These impulses initiate the contraction of cardiac muscles.
Saudi Arabia stands as a titan in the global energy landscape, renowned for its abundant oil and gas resources. It's the largest exporter of petroleum and holds some of the world's most significant reserves. Let's delve into the top 10 oil and gas projects shaping Saudi Arabia's energy future in 2024.
COLLEGE BUS MANAGEMENT SYSTEM PROJECT REPORT.pdfKamal Acharya
The College Bus Management system is completely developed by Visual Basic .NET Version. The application is connect with most secured database language MS SQL Server. The application is develop by using best combination of front-end and back-end languages. The application is totally design like flat user interface. This flat user interface is more attractive user interface in 2017. The application is gives more important to the system functionality. The application is to manage the student’s details, driver’s details, bus details, bus route details, bus fees details and more. The application has only one unit for admin. The admin can manage the entire application. The admin can login into the application by using username and password of the admin. The application is develop for big and small colleges. It is more user friendly for non-computer person. Even they can easily learn how to manage the application within hours. The application is more secure by the admin. The system will give an effective output for the VB.Net and SQL Server given as input to the system. The compiled java program given as input to the system, after scanning the program will generate different reports. The application generates the report for users. The admin can view and download the report of the data. The application deliver the excel format reports. Because, excel formatted reports is very easy to understand the income and expense of the college bus. This application is mainly develop for windows operating system users. In 2017, 73% of people enterprises are using windows operating system. So the application will easily install for all the windows operating system users. The application-developed size is very low. The application consumes very low space in disk. Therefore, the user can allocate very minimum local disk space for this application.
Water scarcity is the lack of fresh water resources to meet the standard water demand. There are two type of water scarcity. One is physical. The other is economic water scarcity.
Hybrid optimization of pumped hydro system and solar- Engr. Abdul-Azeez.pdffxintegritypublishin
Advancements in technology unveil a myriad of electrical and electronic breakthroughs geared towards efficiently harnessing limited resources to meet human energy demands. The optimization of hybrid solar PV panels and pumped hydro energy supply systems plays a pivotal role in utilizing natural resources effectively. This initiative not only benefits humanity but also fosters environmental sustainability. The study investigated the design optimization of these hybrid systems, focusing on understanding solar radiation patterns, identifying geographical influences on solar radiation, formulating a mathematical model for system optimization, and determining the optimal configuration of PV panels and pumped hydro storage. Through a comparative analysis approach and eight weeks of data collection, the study addressed key research questions related to solar radiation patterns and optimal system design. The findings highlighted regions with heightened solar radiation levels, showcasing substantial potential for power generation and emphasizing the system's efficiency. Optimizing system design significantly boosted power generation, promoted renewable energy utilization, and enhanced energy storage capacity. The study underscored the benefits of optimizing hybrid solar PV panels and pumped hydro energy supply systems for sustainable energy usage. Optimizing the design of solar PV panels and pumped hydro energy supply systems as examined across diverse climatic conditions in a developing country, not only enhances power generation but also improves the integration of renewable energy sources and boosts energy storage capacities, particularly beneficial for less economically prosperous regions. Additionally, the study provides valuable insights for advancing energy research in economically viable areas. Recommendations included conducting site-specific assessments, utilizing advanced modeling tools, implementing regular maintenance protocols, and enhancing communication among system components.
Event Management System Vb Net Project Report.pdfKamal Acharya
In present era, the scopes of information technology growing with a very fast .We do not see any are untouched from this industry. The scope of information technology has become wider includes: Business and industry. Household Business, Communication, Education, Entertainment, Science, Medicine, Engineering, Distance Learning, Weather Forecasting. Carrier Searching and so on.
My project named “Event Management System” is software that store and maintained all events coordinated in college. It also helpful to print related reports. My project will help to record the events coordinated by faculties with their Name, Event subject, date & details in an efficient & effective ways.
In my system we have to make a system by which a user can record all events coordinated by a particular faculty. In our proposed system some more featured are added which differs it from the existing system such as security.
Final project report on grocery store management system..pdfKamal Acharya
In today’s fast-changing business environment, it’s extremely important to be able to respond to client needs in the most effective and timely manner. If your customers wish to see your business online and have instant access to your products or services.
Online Grocery Store is an e-commerce website, which retails various grocery products. This project allows viewing various products available enables registered users to purchase desired products instantly using Paytm, UPI payment processor (Instant Pay) and also can place order by using Cash on Delivery (Pay Later) option. This project provides an easy access to Administrators and Managers to view orders placed using Pay Later and Instant Pay options.
In order to develop an e-commerce website, a number of Technologies must be studied and understood. These include multi-tiered architecture, server and client-side scripting techniques, implementation technologies, programming language (such as PHP, HTML, CSS, JavaScript) and MySQL relational databases. This is a project with the objective to develop a basic website where a consumer is provided with a shopping cart website and also to know about the technologies used to develop such a website.
This document will discuss each of the underlying technologies to create and implement an e- commerce website.
Quality defects in TMT Bars, Possible causes and Potential Solutions.PrashantGoswami42
Maintaining high-quality standards in the production of TMT bars is crucial for ensuring structural integrity in construction. Addressing common defects through careful monitoring, standardized processes, and advanced technology can significantly improve the quality of TMT bars. Continuous training and adherence to quality control measures will also play a pivotal role in minimizing these defects.
About
Indigenized remote control interface card suitable for MAFI system CCR equipment. Compatible for IDM8000 CCR. Backplane mounted serial and TCP/Ethernet communication module for CCR remote access. IDM 8000 CCR remote control on serial and TCP protocol.
• Remote control: Parallel or serial interface.
• Compatible with MAFI CCR system.
• Compatible with IDM8000 CCR.
• Compatible with Backplane mount serial communication.
• Compatible with commercial and Defence aviation CCR system.
• Remote control system for accessing CCR and allied system over serial or TCP.
• Indigenized local Support/presence in India.
• Easy in configuration using DIP switches.
Technical Specifications
Indigenized remote control interface card suitable for MAFI system CCR equipment. Compatible for IDM8000 CCR. Backplane mounted serial and TCP/Ethernet communication module for CCR remote access. IDM 8000 CCR remote control on serial and TCP protocol.
Key Features
Indigenized remote control interface card suitable for MAFI system CCR equipment. Compatible for IDM8000 CCR. Backplane mounted serial and TCP/Ethernet communication module for CCR remote access. IDM 8000 CCR remote control on serial and TCP protocol.
• Remote control: Parallel or serial interface
• Compatible with MAFI CCR system
• Copatiable with IDM8000 CCR
• Compatible with Backplane mount serial communication.
• Compatible with commercial and Defence aviation CCR system.
• Remote control system for accessing CCR and allied system over serial or TCP.
• Indigenized local Support/presence in India.
Application
• Remote control: Parallel or serial interface.
• Compatible with MAFI CCR system.
• Compatible with IDM8000 CCR.
• Compatible with Backplane mount serial communication.
• Compatible with commercial and Defence aviation CCR system.
• Remote control system for accessing CCR and allied system over serial or TCP.
• Indigenized local Support/presence in India.
• Easy in configuration using DIP switches.
1. Department of Biomedical Engineering
Faculty of Engineering
University of Isfahan
Seyed Yahya Moradi s.yahyamoradi@mehr.ui.ac.ir
2013/5/10
2. Medical Devices
Electro cardio [kardio] gram [graph]
Is a tool that is used to assess the electrical and muscular functions of the heart.
Electro encephalo gram
Is a tool that measures and records the electrical activity of the brain.
Electro myo gram
Is a tool that is used to record the electrical activity of the muscles.
Electro oculo /nystagmo gram
Is a tool that measures normal eye movement and involuntary rapid eye movements.
1
3. History of ECG
1786 : Luigi Aloisio Galvani
On September 20th 1786 he wrote "I had dissected and prepared a frog in
the usual way and while I was attending to something else I laid it on a table
on which stood an electrical machine at some distance from its conductor
and was separated from it by a considerable space. Suddenly when one of
the persons present in there lightly touched the inner crural nerves of the
frog with the point of a scalpel, all the muscles of the legs seemed to
contract again and again as if they were affected by powerful cramps.“
2
4. 1856 : Rudolph von Koelliker and Heinrich Muller
They confirmed that an electrical current accompanies each heart beat by applying
a galvanometer to the base and apex of an exposed ventricle. They also applied a
nerve-muscle preparation similar to Matteucci's, to the ventricle and observed that
a twitch of the muscle occurred just prior to ventricular systole and also a much
smaller twitch after systole.
3
5. 1887 : Augustus Desiré Waller
1901-1905 : Willem Einthoven
By 1901 Willem Einthoven had completed work on the string galvanometer. String
galvanometers had been used to amplify signals across long-distance submarine
cables but his device was much more sensitive. The device consisted of a very thin
filament of conductive wire passing between very strong electromagnets.When a
current passed through the filament, the string moved because of the change in the
electromagnetic field. The electrical activity of the heart was recorded on
photographic paper from a light passing through the string The early machine
needed water for cooling because of the electromagnetic field , required 5 people
to operate, and weighed 600 pounds.In addition, the patient was required to keep
a leg and an arm in a bucket of water.
He was the first to use
a mercury capillary
electrometer For
recording a human
electrocardiogram .
4
7. Basic Anatomy of the Heart
About the heart:
The heart is the hardest working muscle in the human body.
The heart weighs between 7 and 15 ounces (200 to 425 grams)
and is a little larger than the size of your fist. By the end of a
long life, a person's heart may have beat (expanded and
contracted) more than 3.5 billion times. In fact, each day, the
average heart beats 100,000 times, pumping about 2,000
gallons (7,571 liters) of blood. Your heart is located between
your lungs in the middle of your chest, behind and slightly to
the left of your breastbone (sternum)
The way a normal heart works :
The cardiovascular system, composed of the heart and
blood vessels, is responsible for circulating blood
throughout your body to supply the tissues with oxygen
and nutrients. The heart is the muscle that pumps blood
filled with oxygen and nutrients through the blood
vessels to the body tissues.
6
8. Components of heart
That receive blood from the body and pump out blood to it:
The atria receive blood coming back to the heart.
The ventricles pump the blood out of the heart.
Which compose a network of arteries and veins that carry blood throughout the body:
Arteries transport blood from the heart to the body tissues.
Veins carry blood back to the heart.
To prevent backward flow of blood:
Each valve is designed to allow the forward flow of blood and prevent backward flow.
Which stimulates contraction of the heart muscle.
7
11. The Heart Valves
Function of Heart Valves
■ Fibrous connective tissue prevents
enlargement of valve
openings and anchors valve flaps.
■ Valve closure prevents backflow of
blood during and aftercontraction
The tricuspid valve regulates blood flow between the right atrium and right ventricle.
The pulmonary valve controls blood flow from the right ventricle into the pulmonary
arteries, which carry blood to your lungs to pick up oxygen.
The mitral valve lets oxygen-rich blood from your lungs pass from the left atrium into
the left ventricle.
The aortic valve opens the way for oxygen-rich blood to pass from the left ventricle into
the aorta, your body's largest artery, where it is delivered to the rest of your body.
Four types of valves regulate blood
flow through your heart:
10
13. Dominant pacemaker of the heart, located in
upper portion of right atrium.
Direct electrical impulses between SA and AV
nodes.
Part of AV junctional tissue. Slows Conduction
before impulses reach ventricles.
Transmits impulses to bundle branches.
Located below AV node
Conducts impulses that lead to left ventricle
Conducts impulses that lead to right ventricle
which spreads impulses Rapidly throughout
ventricular walls.Located at terminals of
bundle branches.
Structure ,Function and Location of ESH
12
15. The P Wave
The first wave (p wave) represents atrial depolarisation. When the valves
between the atria and ventricles open, 70% of the blood in the atria falls
through with the aid of gravity, but mainly due to suction caused by the
ventricles as they expand.
Atrial contraction is required only for the final 30% and therefore a relatively
small muscle mass is required and only a relatively small amount of voltage is
needed to contract the atria.
14
16. After the first wave there follows a short period where the line is flat. This is
the point at which the stimulus is delayed in the bundle of His to allow the
atria enough time to pump all the blood into the ventricles.
As the ventricles fill, the growing pressure causes the valves between the
atria and ventricles to close. At this point the electrical stimulus passes from
the bundle of His into the bundle branches and Purkinje fibres. The amount
of electrical energy generated is recorded as a complex of 3 waves known
collectively as the QRS complex. Measuring the waves vertically shows
voltage. More voltage is required to cause ventricular contraction and
therefore the wave is much bigger
it is possible to break down the QRS complex into 3 distinct waves:
Q wave representing septal depolarisation
R wave representing ventricular depolarisation
S wave representing depolarisation of the Purkinje fibres.
The QRS Complex
15
17. The Q Wave
The picture below shows a small negative wave immediately before the large
QRS complex. This is known as a Q wave and represents depolarisation in the
septum.
Whilst the electrical stimulus passes through the bundle of His, and before it
separates down the two bundle branches, it starts to depolarise the septum
from left to right. This is only a small amount of conduction (hence the Q
wave is less than 2 small squares), and it travels in the opposite direction to
the main conduction (right to left) so the Q wave points in the opposite
direction to the large QRS complex.
16
18. The R Wave
The QRS complex is made up of three waves. These waves indicate the changing
direction of the electrical stimulus as it passes through the heart's conduction
system.
The largest wave in the QRS complex is the R wave
As you can see from the diagram, the R wave represents the electrical stimulus
as it passes through the main portion of the ventricular walls. The wall of the
ventricles are very thick due to the amount of work they have to do and,
consequently, more voltage is required. This is why the R wave is by far the
biggest wave generated during normal conduction.
17
19. The S Wave
You will also have seen a small negative wave following the large R wave.
This is known as an S wave and represents depolarisation in the Purkinje
fibres.
The S wave travels in the opposite direction to the large R wave because, as
can be seen on the earlier picture, the Purkinje fibres spread throughout the
ventricles from top to bottom and then back up through the walls of the
ventricles.
18
20. The T Wave
Both ventricles repolarise before the cycle repeats itself and
therefore a 3rd wave (t wave) is visible representing ventricular
repolarisation.
19
21. The ST Segment
There is a brief period between the end of the QRS complex and the beginning
of the T wave where there is no conduction and the line is flat. This is known
as the ST segment and it is a key indicator for both myocardial ischaemia and
necrosis if it goes up or down.
20
23. ECG paper
The output of an ECG recorder is a graph (or
sometimes several graphs, representing each of
the lead) with time represented on the x-axis
and voltage represented on the y-axis. A
dedicated ECG machine would usually print
onto graph paper which has a background
pattern of 1mm squares (often in red or green),
with bold divisions every 5 mm in both vertical
and horizontal directions.It is possible to change
the output of most ECG devices but it is standard
to represent each mV on the y axis as 1 cm and
each second as 25 mm on the x-axis (that is a
paper speed of 25 mm/s). Faster paper speeds
can be used, for example, to resolve finer detail
in the ECG. At a paper speed of 25 mm/s, one
small block of ECG paper translates into 40 ms.
Five small blocks make up one large block, which
translates into 200 ms. Hence, there are five
large blocks per second. A calibration signal may
be included with a record. A standard signal of
1 mV must move the stylus vertically 1 cm, that
is, two large squares on ECG paper.
22
25. Instrumentation Amplifier
AD524 ،AD521 ،AD625 ،AD624 ،AD620
An instrumentation amplifier is a type of differential amplifier that has been outfitted
with input buffer amplifiers, which eliminate the need for input impedance
matching and thus make the amplifier particularly suitable for use in measurement
and test equipment. Additional characteristics include very low DC offset, low drift,
low noise, very high open-loop gain, very high common-mode rejection ratio, and
very high input impedances
24
26. Filters
Low Pass Filter:
There are many types of low pass filter like Butterworth and Chebyshev
filters (in the project the Butterworth filter was used). Low pass filter used
to filter out signals at frequencies higher than the filters cut off frequency
which is required in the project (to reject frequencies higher than 100 Hz).
High Pass Filter:
A High Pass filter is a filter that passes high frequencies and attenuates low
frequencies. Also, high pass filters are useful to filter out the undesired low
frequency components from the main high frequency signals, they are
usually used in applications requiring the rejection of low frequency signals.
Notch Filter:
A notch filter is required to eliminate the 50Hz interference, this
interference comes from the power lines carrying this frequency in close
proximity to patient and equipment.
25
27. A microcontroller is an electronic device that
generally consists of a microprocessor, a memory unit in the form of random
access memory (RAM) and Read Only Memory (ROM), various timers and
input/output serial and parallel ports all integrated on a single chip. In some
advanced microcontroller systems, on chip ADC’s, DAC’s etc are also included.
examples of microcontrollers
26
28. ECG leads
Lead I : is connected to
the right arm and the
positive terminal to the
left arm.
Lead II : is connected
negative terminal of the
electrocardiograph which
connected to the right
arm and the positive
terminal to the left leg
and lead.
Lead III : is connected to
the left arm and the
positive terminal to the
left leg.
Limb leads
27
29. CHEST LEADS
4th Intercostal space to Septumright
of sternum
4th Intercostal space to Septumleft
of sternum
Directly between V2 and V4 Anterior
5th Intercostal space at Anteriorleft
midclavicular line
Level with V4 at left anterior Lateral axillary line
Level with V5 at left midaxillary line Lateral
28