BLOOD FLOW METER
Prepared by Abhijith Prabha , L6A ,ECE Department.
INTRODUCTION
• Blood flow is the one of the important physiological
parameter and the most difficult to measure accurately.
• The average velocities of blood flow vary over a wide range
depending on diameter of blood vessel.
• There are many techniques for measuring the blood flow and
velocity.
• They are categorized into
1.invasive(surgical).
2.non invasive(through the skin).
Need for blood flowmeter
• Inspection for block in blood flow.
• Testing artificial blood vessels during organ transplantation.
• During Fistula creation in dialysis.
Typical values of blood flow
Type of blood
vessels
Total cross-section
area
Blood velocity in
cm/s
Aorta 3–5 cm2 40 cm/s
Capillaries 4500–6000 cm2 0.03 cm/s
Vena cava inferior
and superior
14 cm2 15 cm/s
Introduction
• Most widely used blood flow meters are:
Electromagnetic blood flow meter
Ultrasonic blood flow meters
NMR blood flow meter
Laser Doppler blood flow meter
ELECTROMAGNETIC BLOOD
FLOWMETER
ELECTROMAGNETIC BLOOD FLOWMETER
• Measures instantaneous pulsatile flow of blood
• Works based on the principle of electromagnetic induction
• The voltage induced in a conductor moving in a magnetic
field is proportional to the velocity of the conductor
• The conductive blood is the moving conductor
Principle of EM blood flow meter.
ELECTROMAGNETIC BLOOD FLOWMETER
• A permanent magnet or electromagnet positioned around the blood
vessel generates a magnetic field perpendicular to the direction of the
flow of the blood.
• Voltage induced in the moving blood column is measured with stationary
electrodes located on opposite sides of the blood vessel and
perpendicular to the direction of the magnetic field.
• This method requires that the blood vessel be exposed so that the flow
head or the measuring probe can be put across it.
• The Induced emf
• Where
• B = magnetic flux density, T
• L = length between electrodes, m
• u = instantaneous velocity of blood, m/s
Fig: Typical flow transducer
• The electromagnetic flow-transducer is a tube of non-magnetic
material to ensure that the magnetic flux does not bypass the
flowing liquid and go into the walls of the tube.
Types of EM Flowmeters
• basically, all modern flowmeters consist of a
generator of AC, a probe assembly , a series of
capacitance coupled amplifiers, a demodulator, a DC
amplifier and a suitable recording device.
• Basing shape of the energizing current waveform for
the electromagnet 2 types of EM Flowmeters are :
A. Sine wave flometer.
B. Square wave flowmeter.
Sine wave Flowmeters
• Probe magnet is energized with a sine wave and the induced
voltage will also be sinusoidal.
• Since the flow of blood acts as a secondary terminal of a
transformer w.r.t probe magnet, an additional artifact voltage
induced is called transformer voltage.
• This voltage is 900 out of phase with the original signal
corresponding to flow of blood .
• A method for eliminating transformer voltage by using a gated
amplifier(amplify signal only during flow induced voltage is
maximum).
• This type of instrument is known as ‘gated sine wave
flowmeter’ ,
Square wave flowmeter
• Probe magnet is energized with a square wave and induced
voltage is also square wave.
• It is easier to control magnitude and wave shape of energizing
current.
• Separation of transformer voltage is easy .
• For the measurement action square wave is amplitude
modulated by variation in blood flow.
Block diagram of square wave flowmeter
Block diagram
Transducer
• consist of an electromagnet, a pair of electrodes.
• Electrodes may be in contact with either flowing blood or
outer surface of the blood vessel
Preamplifier
• The induced voltage pick up by the electrodes, is given to a
low noise differential amplifier through a capacitive coupling
• Must have a very high CMRR and input impedance
Block diagram
 Gating circuit
• It helps to remove spurious voltages generated during magnet
current reversal
• The gating action is controlled by the circuit which provides
an excitation current to the electromagnet
 Band pass filter
• It is an active RC band pass amplifier , which selectively pass
through it the amplified square wave signal
• Peak response is kept for 400Hz
Block diagram
 Detector
• A phase sensitive detector is used to recover the signal
• It also helps in the rejection of interfering voltages at
frequencies below the carrier frequency
Low pass filter and output stage
• Demodulated signal is given to an RC LPF , which provides a
uniform frequency response and a linear phase shift
• Followed by an integrator provide output corresponding to the
mean flow
Block diagram
Magnet current drive
• The square wave input to the power amplifier stage which
supplies current to the electromagnet is fed from a free
running multivibrator
Zero flow reference line
• it establish the signal corresponding to zero-flow before
measurement.
Ultrasonic blood flowmeter
Ultrasonic blood flowmeter
• An ultrasonic flow meter is a type of flow meter that
measures the velocity of a fluid with ultrasound to calculate
volume flow.
• Using ultrasonic transducers, the flow meter can measure the
average velocity along the path of an emitted beam of
ultrasound.
• By averaging the difference in measured transit time between
the pulses of ultrasound propagating into and against the
direction of the flow or by measuring the frequency shift from
the Doppler effect.
Fig : ultra sonic flow meter
(a)Doppler-shift flow-velocity meters
• It is an non invasive method.
• It is based on the analysis of echo signals from
erythrocytes(RBCs) in blood.
• The incident ultrasound is scattered by the blood cells and
scattered wave is received by the second receiver.
• The frequency shift of the scattered wave gives idea about
velocity of scatterers.
• The Doppler frequency shift is a measure of size and direction
of the flow velocity.
Doppler-shift flow-velocity meters
5 MHz
exciter
Rx Tx
R F
amplifierdetector
LPF
Audio
amplifier
Zero
crossing
detector
LPF
Blood vessel
Fig : block diagram
output
(b)Range gated pulsed Doppler flow meter
• Baker(1970) stated that recording blood flow using Doppler
shift are sometimes misleading and inaccurate.
• These difficulties can be overcome , if the ultrasonic source is
pulsed.
• If the returning signal is range gated ,then the diameter and
velocity of blood stream can measure together.
Receiver
limiter
Phase detector LPF Sample and amplifier
Power amp Pulse amp
Sample
pulse gen.
LPF
Freq division
Master Oscillator
(4.5-5.5 MHz)
Rx element
Tx element
Audible
Doppler
output
Fig : block diagram of pulsed doppler flow meter.
(c) Flow measurement by Doppler Imaging
• Doppler ultrasound is not only used for measurement of the
absolute value of blood velocity and volume, but it also helps
to visualize blood flow.
• The probe this imaging equipment is mechanically coupled to
position resolver, which gives electrical output.
• Imaging is done by moving the probe through the skin, and
developing a 3D image using a computer.
• Thus, it is possible to construct anterior-posterior , lateral and
cross-sectional scans of blood vessel.
Flow measurement by Doppler Imaging
Figs : Flow measurement by Doppler Imaging
Flow measurement by Doppler Imaging
• Using this recording of mean blood flow, peak flow, reverse
peak flow etc. of cardiac cycle are possible.
• This method is also helpful for taking measurements from
brain , which is difficult to access.
NMR Blood Flowmeter
NMR Blood Flowmeter
• Non invasive method by Nuclear magnetic resonance
principle.
• For flow measurement , behaviour of the 2 Hydrogen atoms
of water is studied.(Since blood is approximately 83% water).
• Due to magnetic moment of the H atom , the nucleus behaves
as a micro miniature magnet which can affected by externally
applied magnetic field.
• the hydrogen nuclei orient themselves to produce a net
parallel alignment to a steady magnetic field.
NMR Blood Flowmeter
• Orientation of H atoms will be change according to the status
of blood flow.
• This orientation will results in the magnetization of H atoms.
• Therefore, magnitude of magnetization can be related to
either velocity of blood or flow rate.
• A crossed coil configuration is used to detect the level of
magnetization in body.
• The voltage induced in coil(NMR signal) is proportional to
velocity and area of vessel carrying blood.
• NMR flow meters are limited in their application to the
measurement in limbs.
Laser Doppler blood flowmeter
Fig : schematic diagram of laser doppler blood flow meter
THE END

More Related Content

What's hot

Phonocardiography(PCG)
Phonocardiography(PCG)Phonocardiography(PCG)
Phonocardiography(PCG)
nishanthi p
 
Biotelemetry
BiotelemetryBiotelemetry
BiotelemetrySamuely
 
Respiration rate measurement
Respiration rate measurement Respiration rate measurement
Respiration rate measurement
BharathasreejaG
 
Biomedical Instrumentation
Biomedical InstrumentationBiomedical Instrumentation
Biomedical Instrumentation
linsstalex
 
Sensors for Biomedical Devices and systems
Sensors for Biomedical Devices and systemsSensors for Biomedical Devices and systems
Sensors for Biomedical Devices and systems
Gunjan Patel
 
Biomedical instrumentation PPT
Biomedical instrumentation PPTBiomedical instrumentation PPT
Biomedical instrumentation PPT
abhi1802verma
 
Origin of biopotentials
Origin of biopotentialsOrigin of biopotentials
Origin of biopotentials
RAMESHBABUA3
 
Transducers for bio medical
Transducers for bio medicalTransducers for bio medical
Transducers for bio medicalSLIET
 
Topic 3 Bioelectrodes-sensors-transducers
Topic 3 Bioelectrodes-sensors-transducersTopic 3 Bioelectrodes-sensors-transducers
Topic 3 Bioelectrodes-sensors-transducers
Ghansyam Rathod
 
Carrier and chopper amplifiers
Carrier and chopper amplifiers Carrier and chopper amplifiers
Carrier and chopper amplifiers
BharathasreejaG
 
Transducers
TransducersTransducers
Transducers
Sachin Kumar
 
Ventilators - Biomedical Instrumentation
Ventilators - Biomedical InstrumentationVentilators - Biomedical Instrumentation
Ventilators - Biomedical Instrumentation
Girish Kumar K
 
Biomedical Instrumentation
Biomedical InstrumentationBiomedical Instrumentation
Biomedical Instrumentation
Manipal University Jaipur
 
Transducers
Transducers Transducers
Transducers
Arpit Sola
 
Electrical safety
Electrical safetyElectrical safety
Electrical safety
Yassir Ali Hassan
 
Cardiac pacemaker
Cardiac pacemakerCardiac pacemaker
Cardiac pacemaker
Rupon Bhowmik
 
Application of different medical electrodes
Application of different medical electrodesApplication of different medical electrodes
Application of different medical electrodes
Sikkim Manipal Institute Of Technology
 
Photoelectric transducers and its classification
Photoelectric transducers and its classificationPhotoelectric transducers and its classification
Photoelectric transducers and its classification
karoline Enoch
 
Ultrasonic Diathermy
Ultrasonic Diathermy Ultrasonic Diathermy
Ultrasonic Diathermy
Akkash Rao
 

What's hot (20)

Phonocardiography(PCG)
Phonocardiography(PCG)Phonocardiography(PCG)
Phonocardiography(PCG)
 
Biotelemetry
BiotelemetryBiotelemetry
Biotelemetry
 
Respiration rate measurement
Respiration rate measurement Respiration rate measurement
Respiration rate measurement
 
Biomedical Instrumentation
Biomedical InstrumentationBiomedical Instrumentation
Biomedical Instrumentation
 
Sensors for Biomedical Devices and systems
Sensors for Biomedical Devices and systemsSensors for Biomedical Devices and systems
Sensors for Biomedical Devices and systems
 
Biomedical instrumentation PPT
Biomedical instrumentation PPTBiomedical instrumentation PPT
Biomedical instrumentation PPT
 
Origin of biopotentials
Origin of biopotentialsOrigin of biopotentials
Origin of biopotentials
 
Transducers for bio medical
Transducers for bio medicalTransducers for bio medical
Transducers for bio medical
 
Topic 3 Bioelectrodes-sensors-transducers
Topic 3 Bioelectrodes-sensors-transducersTopic 3 Bioelectrodes-sensors-transducers
Topic 3 Bioelectrodes-sensors-transducers
 
Carrier and chopper amplifiers
Carrier and chopper amplifiers Carrier and chopper amplifiers
Carrier and chopper amplifiers
 
Transducers
TransducersTransducers
Transducers
 
Ventilators - Biomedical Instrumentation
Ventilators - Biomedical InstrumentationVentilators - Biomedical Instrumentation
Ventilators - Biomedical Instrumentation
 
Biomedical Instrumentation
Biomedical InstrumentationBiomedical Instrumentation
Biomedical Instrumentation
 
Transducers
Transducers Transducers
Transducers
 
Electrical safety
Electrical safetyElectrical safety
Electrical safety
 
Transducers
TransducersTransducers
Transducers
 
Cardiac pacemaker
Cardiac pacemakerCardiac pacemaker
Cardiac pacemaker
 
Application of different medical electrodes
Application of different medical electrodesApplication of different medical electrodes
Application of different medical electrodes
 
Photoelectric transducers and its classification
Photoelectric transducers and its classificationPhotoelectric transducers and its classification
Photoelectric transducers and its classification
 
Ultrasonic Diathermy
Ultrasonic Diathermy Ultrasonic Diathermy
Ultrasonic Diathermy
 

Similar to blood flow measurement

BMI blood flowmeter.pptx
BMI blood flowmeter.pptxBMI blood flowmeter.pptx
BMI blood flowmeter.pptx
chetanharihar2
 
BLOOD FLOW METER.pdf
BLOOD FLOW METER.pdfBLOOD FLOW METER.pdf
BLOOD FLOW METER.pdf
ravikumarbme
 
Current Meters
Current MetersCurrent Meters
Current Meters
marvinnbustamante1
 
ii unit
 ii unit ii unit
ii unit
johny renoald
 
Electromagnetic flowmeters
Electromagnetic flowmetersElectromagnetic flowmeters
Electromagnetic flowmeters
AJAL A J
 
Medical Instrumentation- body parameters measurement
Medical Instrumentation- body parameters measurementMedical Instrumentation- body parameters measurement
Medical Instrumentation- body parameters measurement
Poornima D
 
Doppler Physics
Doppler PhysicsDoppler Physics
Doppler Physics
Sahil Chaudhry
 
Instrumentation in oil refinery
Instrumentation in oil refineryInstrumentation in oil refinery
Instrumentation in oil refinery
Bhavesh Solanki
 
Blood Pressure. NOTES docx.pdf
Blood Pressure. NOTES docx.pdfBlood Pressure. NOTES docx.pdf
Blood Pressure. NOTES docx.pdf
MukisaIvan
 
pdf
pdfpdf
Transcranial doppler ultrasound, principles and procuderes
Transcranial doppler ultrasound, principles and procuderesTranscranial doppler ultrasound, principles and procuderes
Transcranial doppler ultrasound, principles and procuderes
Nitish kumar
 
Doppler us (2)
Doppler us (2)Doppler us (2)
Doppler us (2)
KamalEldirawi
 
mel705-27.ppt
mel705-27.pptmel705-27.ppt
mel705-27.ppt
DrPalpandiMurugesan
 
Doppler echocardiography
Doppler echocardiographyDoppler echocardiography
Doppler echocardiography
sruthiMeenaxshiSR
 
Doppler echocardiography
Doppler echocardiographyDoppler echocardiography
Doppler echocardiography
Sruthi Meenaxshi
 
Blood Flow Sensors and Meters with their working
Blood Flow Sensors and Meters with their workingBlood Flow Sensors and Meters with their working
Blood Flow Sensors and Meters with their working
jafarfarwa
 
Ec 6001 medical electronics
Ec 6001  medical electronicsEc 6001  medical electronics
Ec 6001 medical electronics
Benesh Selvanesan
 
Doppler physics.pptx
Doppler physics.pptxDoppler physics.pptx
Doppler physics.pptx
VANI PUSHPA MUDAVATH
 
Meas of current.pptx
Meas of current.pptxMeas of current.pptx
Meas of current.pptx
AnkitKachhap4
 

Similar to blood flow measurement (20)

BMI blood flowmeter.pptx
BMI blood flowmeter.pptxBMI blood flowmeter.pptx
BMI blood flowmeter.pptx
 
BLOOD FLOW METER.pdf
BLOOD FLOW METER.pdfBLOOD FLOW METER.pdf
BLOOD FLOW METER.pdf
 
Current Meters
Current MetersCurrent Meters
Current Meters
 
ii unit
 ii unit ii unit
ii unit
 
Electromagnetic flowmeters
Electromagnetic flowmetersElectromagnetic flowmeters
Electromagnetic flowmeters
 
Medical Instrumentation- body parameters measurement
Medical Instrumentation- body parameters measurementMedical Instrumentation- body parameters measurement
Medical Instrumentation- body parameters measurement
 
Blood Flow Meter
Blood Flow MeterBlood Flow Meter
Blood Flow Meter
 
Doppler Physics
Doppler PhysicsDoppler Physics
Doppler Physics
 
Instrumentation in oil refinery
Instrumentation in oil refineryInstrumentation in oil refinery
Instrumentation in oil refinery
 
Blood Pressure. NOTES docx.pdf
Blood Pressure. NOTES docx.pdfBlood Pressure. NOTES docx.pdf
Blood Pressure. NOTES docx.pdf
 
pdf
pdfpdf
pdf
 
Transcranial doppler ultrasound, principles and procuderes
Transcranial doppler ultrasound, principles and procuderesTranscranial doppler ultrasound, principles and procuderes
Transcranial doppler ultrasound, principles and procuderes
 
Doppler us (2)
Doppler us (2)Doppler us (2)
Doppler us (2)
 
mel705-27.ppt
mel705-27.pptmel705-27.ppt
mel705-27.ppt
 
Doppler echocardiography
Doppler echocardiographyDoppler echocardiography
Doppler echocardiography
 
Doppler echocardiography
Doppler echocardiographyDoppler echocardiography
Doppler echocardiography
 
Blood Flow Sensors and Meters with their working
Blood Flow Sensors and Meters with their workingBlood Flow Sensors and Meters with their working
Blood Flow Sensors and Meters with their working
 
Ec 6001 medical electronics
Ec 6001  medical electronicsEc 6001  medical electronics
Ec 6001 medical electronics
 
Doppler physics.pptx
Doppler physics.pptxDoppler physics.pptx
Doppler physics.pptx
 
Meas of current.pptx
Meas of current.pptxMeas of current.pptx
Meas of current.pptx
 

Recently uploaded

Courier management system project report.pdf
Courier management system project report.pdfCourier management system project report.pdf
Courier management system project report.pdf
Kamal Acharya
 
road safety engineering r s e unit 3.pdf
road safety engineering  r s e unit 3.pdfroad safety engineering  r s e unit 3.pdf
road safety engineering r s e unit 3.pdf
VENKATESHvenky89705
 
Student information management system project report ii.pdf
Student information management system project report ii.pdfStudent information management system project report ii.pdf
Student information management system project report ii.pdf
Kamal Acharya
 
The Benefits and Techniques of Trenchless Pipe Repair.pdf
The Benefits and Techniques of Trenchless Pipe Repair.pdfThe Benefits and Techniques of Trenchless Pipe Repair.pdf
The Benefits and Techniques of Trenchless Pipe Repair.pdf
Pipe Restoration Solutions
 
CME397 Surface Engineering- Professional Elective
CME397 Surface Engineering- Professional ElectiveCME397 Surface Engineering- Professional Elective
CME397 Surface Engineering- Professional Elective
karthi keyan
 
DESIGN A COTTON SEED SEPARATION MACHINE.docx
DESIGN A COTTON SEED SEPARATION MACHINE.docxDESIGN A COTTON SEED SEPARATION MACHINE.docx
DESIGN A COTTON SEED SEPARATION MACHINE.docx
FluxPrime1
 
ethical hacking in wireless-hacking1.ppt
ethical hacking in wireless-hacking1.pptethical hacking in wireless-hacking1.ppt
ethical hacking in wireless-hacking1.ppt
Jayaprasanna4
 
NO1 Uk best vashikaran specialist in delhi vashikaran baba near me online vas...
NO1 Uk best vashikaran specialist in delhi vashikaran baba near me online vas...NO1 Uk best vashikaran specialist in delhi vashikaran baba near me online vas...
NO1 Uk best vashikaran specialist in delhi vashikaran baba near me online vas...
Amil Baba Dawood bangali
 
power quality voltage fluctuation UNIT - I.pptx
power quality voltage fluctuation UNIT - I.pptxpower quality voltage fluctuation UNIT - I.pptx
power quality voltage fluctuation UNIT - I.pptx
ViniHema
 
Gen AI Study Jams _ For the GDSC Leads in India.pdf
Gen AI Study Jams _ For the GDSC Leads in India.pdfGen AI Study Jams _ For the GDSC Leads in India.pdf
Gen AI Study Jams _ For the GDSC Leads in India.pdf
gdsczhcet
 
Architectural Portfolio Sean Lockwood
Architectural Portfolio Sean LockwoodArchitectural Portfolio Sean Lockwood
Architectural Portfolio Sean Lockwood
seandesed
 
Final project report on grocery store management system..pdf
Final project report on grocery store management system..pdfFinal project report on grocery store management system..pdf
Final project report on grocery store management system..pdf
Kamal Acharya
 
Event Management System Vb Net Project Report.pdf
Event Management System Vb Net  Project Report.pdfEvent Management System Vb Net  Project Report.pdf
Event Management System Vb Net Project Report.pdf
Kamal Acharya
 
J.Yang, ICLR 2024, MLILAB, KAIST AI.pdf
J.Yang,  ICLR 2024, MLILAB, KAIST AI.pdfJ.Yang,  ICLR 2024, MLILAB, KAIST AI.pdf
J.Yang, ICLR 2024, MLILAB, KAIST AI.pdf
MLILAB
 
Immunizing Image Classifiers Against Localized Adversary Attacks
Immunizing Image Classifiers Against Localized Adversary AttacksImmunizing Image Classifiers Against Localized Adversary Attacks
Immunizing Image Classifiers Against Localized Adversary Attacks
gerogepatton
 
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdfAKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
SamSarthak3
 
Design and Analysis of Algorithms-DP,Backtracking,Graphs,B&B
Design and Analysis of Algorithms-DP,Backtracking,Graphs,B&BDesign and Analysis of Algorithms-DP,Backtracking,Graphs,B&B
Design and Analysis of Algorithms-DP,Backtracking,Graphs,B&B
Sreedhar Chowdam
 
HYDROPOWER - Hydroelectric power generation
HYDROPOWER - Hydroelectric power generationHYDROPOWER - Hydroelectric power generation
HYDROPOWER - Hydroelectric power generation
Robbie Edward Sayers
 
Quality defects in TMT Bars, Possible causes and Potential Solutions.
Quality defects in TMT Bars, Possible causes and Potential Solutions.Quality defects in TMT Bars, Possible causes and Potential Solutions.
Quality defects in TMT Bars, Possible causes and Potential Solutions.
PrashantGoswami42
 
COLLEGE BUS MANAGEMENT SYSTEM PROJECT REPORT.pdf
COLLEGE BUS MANAGEMENT SYSTEM PROJECT REPORT.pdfCOLLEGE BUS MANAGEMENT SYSTEM PROJECT REPORT.pdf
COLLEGE BUS MANAGEMENT SYSTEM PROJECT REPORT.pdf
Kamal Acharya
 

Recently uploaded (20)

Courier management system project report.pdf
Courier management system project report.pdfCourier management system project report.pdf
Courier management system project report.pdf
 
road safety engineering r s e unit 3.pdf
road safety engineering  r s e unit 3.pdfroad safety engineering  r s e unit 3.pdf
road safety engineering r s e unit 3.pdf
 
Student information management system project report ii.pdf
Student information management system project report ii.pdfStudent information management system project report ii.pdf
Student information management system project report ii.pdf
 
The Benefits and Techniques of Trenchless Pipe Repair.pdf
The Benefits and Techniques of Trenchless Pipe Repair.pdfThe Benefits and Techniques of Trenchless Pipe Repair.pdf
The Benefits and Techniques of Trenchless Pipe Repair.pdf
 
CME397 Surface Engineering- Professional Elective
CME397 Surface Engineering- Professional ElectiveCME397 Surface Engineering- Professional Elective
CME397 Surface Engineering- Professional Elective
 
DESIGN A COTTON SEED SEPARATION MACHINE.docx
DESIGN A COTTON SEED SEPARATION MACHINE.docxDESIGN A COTTON SEED SEPARATION MACHINE.docx
DESIGN A COTTON SEED SEPARATION MACHINE.docx
 
ethical hacking in wireless-hacking1.ppt
ethical hacking in wireless-hacking1.pptethical hacking in wireless-hacking1.ppt
ethical hacking in wireless-hacking1.ppt
 
NO1 Uk best vashikaran specialist in delhi vashikaran baba near me online vas...
NO1 Uk best vashikaran specialist in delhi vashikaran baba near me online vas...NO1 Uk best vashikaran specialist in delhi vashikaran baba near me online vas...
NO1 Uk best vashikaran specialist in delhi vashikaran baba near me online vas...
 
power quality voltage fluctuation UNIT - I.pptx
power quality voltage fluctuation UNIT - I.pptxpower quality voltage fluctuation UNIT - I.pptx
power quality voltage fluctuation UNIT - I.pptx
 
Gen AI Study Jams _ For the GDSC Leads in India.pdf
Gen AI Study Jams _ For the GDSC Leads in India.pdfGen AI Study Jams _ For the GDSC Leads in India.pdf
Gen AI Study Jams _ For the GDSC Leads in India.pdf
 
Architectural Portfolio Sean Lockwood
Architectural Portfolio Sean LockwoodArchitectural Portfolio Sean Lockwood
Architectural Portfolio Sean Lockwood
 
Final project report on grocery store management system..pdf
Final project report on grocery store management system..pdfFinal project report on grocery store management system..pdf
Final project report on grocery store management system..pdf
 
Event Management System Vb Net Project Report.pdf
Event Management System Vb Net  Project Report.pdfEvent Management System Vb Net  Project Report.pdf
Event Management System Vb Net Project Report.pdf
 
J.Yang, ICLR 2024, MLILAB, KAIST AI.pdf
J.Yang,  ICLR 2024, MLILAB, KAIST AI.pdfJ.Yang,  ICLR 2024, MLILAB, KAIST AI.pdf
J.Yang, ICLR 2024, MLILAB, KAIST AI.pdf
 
Immunizing Image Classifiers Against Localized Adversary Attacks
Immunizing Image Classifiers Against Localized Adversary AttacksImmunizing Image Classifiers Against Localized Adversary Attacks
Immunizing Image Classifiers Against Localized Adversary Attacks
 
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdfAKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
AKS UNIVERSITY Satna Final Year Project By OM Hardaha.pdf
 
Design and Analysis of Algorithms-DP,Backtracking,Graphs,B&B
Design and Analysis of Algorithms-DP,Backtracking,Graphs,B&BDesign and Analysis of Algorithms-DP,Backtracking,Graphs,B&B
Design and Analysis of Algorithms-DP,Backtracking,Graphs,B&B
 
HYDROPOWER - Hydroelectric power generation
HYDROPOWER - Hydroelectric power generationHYDROPOWER - Hydroelectric power generation
HYDROPOWER - Hydroelectric power generation
 
Quality defects in TMT Bars, Possible causes and Potential Solutions.
Quality defects in TMT Bars, Possible causes and Potential Solutions.Quality defects in TMT Bars, Possible causes and Potential Solutions.
Quality defects in TMT Bars, Possible causes and Potential Solutions.
 
COLLEGE BUS MANAGEMENT SYSTEM PROJECT REPORT.pdf
COLLEGE BUS MANAGEMENT SYSTEM PROJECT REPORT.pdfCOLLEGE BUS MANAGEMENT SYSTEM PROJECT REPORT.pdf
COLLEGE BUS MANAGEMENT SYSTEM PROJECT REPORT.pdf
 

blood flow measurement

  • 1. BLOOD FLOW METER Prepared by Abhijith Prabha , L6A ,ECE Department.
  • 2. INTRODUCTION • Blood flow is the one of the important physiological parameter and the most difficult to measure accurately. • The average velocities of blood flow vary over a wide range depending on diameter of blood vessel. • There are many techniques for measuring the blood flow and velocity. • They are categorized into 1.invasive(surgical). 2.non invasive(through the skin).
  • 3. Need for blood flowmeter • Inspection for block in blood flow. • Testing artificial blood vessels during organ transplantation. • During Fistula creation in dialysis.
  • 4. Typical values of blood flow Type of blood vessels Total cross-section area Blood velocity in cm/s Aorta 3–5 cm2 40 cm/s Capillaries 4500–6000 cm2 0.03 cm/s Vena cava inferior and superior 14 cm2 15 cm/s
  • 5. Introduction • Most widely used blood flow meters are: Electromagnetic blood flow meter Ultrasonic blood flow meters NMR blood flow meter Laser Doppler blood flow meter
  • 7. ELECTROMAGNETIC BLOOD FLOWMETER • Measures instantaneous pulsatile flow of blood • Works based on the principle of electromagnetic induction • The voltage induced in a conductor moving in a magnetic field is proportional to the velocity of the conductor • The conductive blood is the moving conductor
  • 8. Principle of EM blood flow meter.
  • 9. ELECTROMAGNETIC BLOOD FLOWMETER • A permanent magnet or electromagnet positioned around the blood vessel generates a magnetic field perpendicular to the direction of the flow of the blood. • Voltage induced in the moving blood column is measured with stationary electrodes located on opposite sides of the blood vessel and perpendicular to the direction of the magnetic field. • This method requires that the blood vessel be exposed so that the flow head or the measuring probe can be put across it.
  • 10. • The Induced emf • Where • B = magnetic flux density, T • L = length between electrodes, m • u = instantaneous velocity of blood, m/s
  • 11. Fig: Typical flow transducer • The electromagnetic flow-transducer is a tube of non-magnetic material to ensure that the magnetic flux does not bypass the flowing liquid and go into the walls of the tube.
  • 12. Types of EM Flowmeters • basically, all modern flowmeters consist of a generator of AC, a probe assembly , a series of capacitance coupled amplifiers, a demodulator, a DC amplifier and a suitable recording device. • Basing shape of the energizing current waveform for the electromagnet 2 types of EM Flowmeters are : A. Sine wave flometer. B. Square wave flowmeter.
  • 13. Sine wave Flowmeters • Probe magnet is energized with a sine wave and the induced voltage will also be sinusoidal. • Since the flow of blood acts as a secondary terminal of a transformer w.r.t probe magnet, an additional artifact voltage induced is called transformer voltage. • This voltage is 900 out of phase with the original signal corresponding to flow of blood . • A method for eliminating transformer voltage by using a gated amplifier(amplify signal only during flow induced voltage is maximum). • This type of instrument is known as ‘gated sine wave flowmeter’ ,
  • 14. Square wave flowmeter • Probe magnet is energized with a square wave and induced voltage is also square wave. • It is easier to control magnitude and wave shape of energizing current. • Separation of transformer voltage is easy . • For the measurement action square wave is amplitude modulated by variation in blood flow.
  • 15. Block diagram of square wave flowmeter
  • 16. Block diagram Transducer • consist of an electromagnet, a pair of electrodes. • Electrodes may be in contact with either flowing blood or outer surface of the blood vessel Preamplifier • The induced voltage pick up by the electrodes, is given to a low noise differential amplifier through a capacitive coupling • Must have a very high CMRR and input impedance
  • 17. Block diagram  Gating circuit • It helps to remove spurious voltages generated during magnet current reversal • The gating action is controlled by the circuit which provides an excitation current to the electromagnet  Band pass filter • It is an active RC band pass amplifier , which selectively pass through it the amplified square wave signal • Peak response is kept for 400Hz
  • 18. Block diagram  Detector • A phase sensitive detector is used to recover the signal • It also helps in the rejection of interfering voltages at frequencies below the carrier frequency Low pass filter and output stage • Demodulated signal is given to an RC LPF , which provides a uniform frequency response and a linear phase shift • Followed by an integrator provide output corresponding to the mean flow
  • 19. Block diagram Magnet current drive • The square wave input to the power amplifier stage which supplies current to the electromagnet is fed from a free running multivibrator Zero flow reference line • it establish the signal corresponding to zero-flow before measurement.
  • 21. Ultrasonic blood flowmeter • An ultrasonic flow meter is a type of flow meter that measures the velocity of a fluid with ultrasound to calculate volume flow. • Using ultrasonic transducers, the flow meter can measure the average velocity along the path of an emitted beam of ultrasound. • By averaging the difference in measured transit time between the pulses of ultrasound propagating into and against the direction of the flow or by measuring the frequency shift from the Doppler effect.
  • 22. Fig : ultra sonic flow meter
  • 23. (a)Doppler-shift flow-velocity meters • It is an non invasive method. • It is based on the analysis of echo signals from erythrocytes(RBCs) in blood. • The incident ultrasound is scattered by the blood cells and scattered wave is received by the second receiver. • The frequency shift of the scattered wave gives idea about velocity of scatterers. • The Doppler frequency shift is a measure of size and direction of the flow velocity.
  • 24. Doppler-shift flow-velocity meters 5 MHz exciter Rx Tx R F amplifierdetector LPF Audio amplifier Zero crossing detector LPF Blood vessel Fig : block diagram output
  • 25. (b)Range gated pulsed Doppler flow meter • Baker(1970) stated that recording blood flow using Doppler shift are sometimes misleading and inaccurate. • These difficulties can be overcome , if the ultrasonic source is pulsed. • If the returning signal is range gated ,then the diameter and velocity of blood stream can measure together.
  • 26. Receiver limiter Phase detector LPF Sample and amplifier Power amp Pulse amp Sample pulse gen. LPF Freq division Master Oscillator (4.5-5.5 MHz) Rx element Tx element Audible Doppler output Fig : block diagram of pulsed doppler flow meter.
  • 27. (c) Flow measurement by Doppler Imaging • Doppler ultrasound is not only used for measurement of the absolute value of blood velocity and volume, but it also helps to visualize blood flow. • The probe this imaging equipment is mechanically coupled to position resolver, which gives electrical output. • Imaging is done by moving the probe through the skin, and developing a 3D image using a computer. • Thus, it is possible to construct anterior-posterior , lateral and cross-sectional scans of blood vessel.
  • 28. Flow measurement by Doppler Imaging Figs : Flow measurement by Doppler Imaging
  • 29. Flow measurement by Doppler Imaging • Using this recording of mean blood flow, peak flow, reverse peak flow etc. of cardiac cycle are possible. • This method is also helpful for taking measurements from brain , which is difficult to access.
  • 31. NMR Blood Flowmeter • Non invasive method by Nuclear magnetic resonance principle. • For flow measurement , behaviour of the 2 Hydrogen atoms of water is studied.(Since blood is approximately 83% water). • Due to magnetic moment of the H atom , the nucleus behaves as a micro miniature magnet which can affected by externally applied magnetic field. • the hydrogen nuclei orient themselves to produce a net parallel alignment to a steady magnetic field.
  • 32. NMR Blood Flowmeter • Orientation of H atoms will be change according to the status of blood flow. • This orientation will results in the magnetization of H atoms. • Therefore, magnitude of magnetization can be related to either velocity of blood or flow rate. • A crossed coil configuration is used to detect the level of magnetization in body. • The voltage induced in coil(NMR signal) is proportional to velocity and area of vessel carrying blood. • NMR flow meters are limited in their application to the measurement in limbs.
  • 33. Laser Doppler blood flowmeter
  • 34. Fig : schematic diagram of laser doppler blood flow meter