R E S P I R A T I O N R A T E
M E A S U R E M E N T
BASICS OF
BIOMEDICAL
INSTRUMENTATION
DISCUSSION ABOUT
2
Function of
Respiration
system
Types of
measurement
Coding Task
3
RESPIRATION RATE
MEASUREMENT
FUNCTION OF RESPIRATION SYSTEM
❖ The function of respiration system are to
supply oxygen and remove carbon dioxide
from the tissues.
❖ The action of breathing is controlled by a
muscular action causing the volume of the
lung to increase and decrease.
4
TYPES OF MEASUREMENT
5
Displacement
method
Thermistor
method
Impedance
pneumography
Co2 Method
Apnoea
detectors
DISPLACEMENT METHOD
• The transducer is hold by an elastic band which goes around the chest.
• The respiratory movements results in a corresponding resistance changes of the
strain gauge.
• It is connected as one arm of a wheat stone bridge circuit.
• Its output varies with chest expansion. This output corresponds to respiration
activity.
6
THERMISTOR METHOD
▸ Generally there is a temperature difference between
inspired and expired air.
▸ This temperature is sensed by placing thermistor
infront of nostrils.
▸ Thermistor is placed by suitable stand. Thermistor is
connected with the bridge circuit.
▸ So unbalance signal is amplified to get the respiratory
activity.
7
IMPEDANCE PNEUMOGRAPHY
8
V = I (R ± R)
IMPEDANCE PNEUMOGRAPHY
• This is the indirect method of measurement.
• Impedance across the chest of a patient changes as respiration occurs.
• 50Hz ac signal is produced by oscillator circuit and it is given to chest of patient
through electrodes.
• The signal voltage applied to the amplifier block the voltage drop across the
resistance.
Where,
V  O/PVoltage
I  Current through the chest
R  Chest impedance without respiration
R  Change of chest impedance due to respiration.
9
V = I (R ± R)
CO2 METHOD
• Respiration rate can be measured by measuring CO2 in expired air.
• This CO2 method is based on the absorption property of infrared rays
• When infrared rays are passed through the expired air which contains certain amount
of CO2 some of the radiation are absorbed by it. So there is a loss of heat energy
associated with the rays.
• The detector changes the loss in heating effect of the rays into an electrical signal. It is
used to get the average respiration rate.
10
11
▸ Two infrared sources are available in this setup.
▸ The beam from one infrared source falls on the Test cuvette side.
▸ The beam from another infrared source falls in the reference cuvette side.
▸ The detector has 2 identical portions. These portions are separated by a thin,
flexible metal diaphragm.
▸ The detector is filled with a sample of pure CO2. Because of the absorption
of CO2 in the test cuvette, the beam falling on the test side of the detector is
weaker than that falling on the reference side is heated more than that on the
test side. So the diaphragm is pushed slightly to the test side of the detector.
This diaphragm forms one plate of a capacitor.
▸ The ac signal appears across the detector is amplified and recorded using
recorder. The amplified output is integrated and shown in this method. It is
used for continuous monitoring the respiration rate.12
APNOEA DETECTORS
• Apnoea is the stoppage of breathing.
• It can be occurred at the conditions like head injury, drug over dose, etc.
• Apnoea can also occur in premature babies during first weeks of life because of their
immature nervous system.
• If Apnoea persists for prolonged period, then brain function can be severely damaged. So
Apnoea patients are closely monitored.
• Apnoea monitor gives audio signal and visual signal (flashing of light), when no inspiration
occurs for a particular period of time.
13
14
▸ Motion channel block differentiates motion (disturbances) and the respiration based on the frequency.
▸ The frequency below 1.5Hz is identified as respiration.
▸ The frequency above 1.5Hz is identified as motion.
▸ High frequency signal above the threshold is sensed by positive detector.
▸ The frequency below the threshold is sensed by negative detector.
▸ Output of motion channel is connected with comparator circuit. It compares the amplitude of motion
and respiration. Based on the output corresponding lamp will glow.
▸ If there is no respiration, then Schmitt trigger circuit gives the output to switch on the alarm.
▸ Apnoea period selector circuit contains low frequency alarm oscillator and tone oscillator and audio
amplifier.
▸ Apnoea period selector circuit drives the alarm circuit.
▸ The output of the alarm circuit is connected with the speaker.
▸ So when there is no respiration for a period of 10s or 20s, then audio signal through the speaker and
visual signal through the flash light is delivered.
15
16

More Related Content

What's hot

MEASUREMENT OF BIO POTENTIAL USING TWO ELECTRODES AND RECORDING PROBLEMS
MEASUREMENT OF BIO POTENTIAL  USING TWO ELECTRODES AND RECORDING PROBLEMSMEASUREMENT OF BIO POTENTIAL  USING TWO ELECTRODES AND RECORDING PROBLEMS
MEASUREMENT OF BIO POTENTIAL USING TWO ELECTRODES AND RECORDING PROBLEMSBharathasreejaG
 
Biomedical Instrumentation
Biomedical InstrumentationBiomedical Instrumentation
Biomedical Instrumentationlinsstalex
 
Transducers for bio medical
Transducers for bio medicalTransducers for bio medical
Transducers for bio medicalSLIET
 
ECG machines -Operation and Maintenance
ECG machines -Operation and MaintenanceECG machines -Operation and Maintenance
ECG machines -Operation and Maintenanceshashi sinha
 
Instrumentation amplifier
Instrumentation amplifierInstrumentation amplifier
Instrumentation amplifiermohdabuzar5
 
Biopotentials
BiopotentialsBiopotentials
Biopotentialsstooty s
 
Sensors for Biomedical Devices and systems
Sensors for Biomedical Devices and systemsSensors for Biomedical Devices and systems
Sensors for Biomedical Devices and systemsGunjan Patel
 
Electromagnetic blood flow meter
Electromagnetic blood flow meterElectromagnetic blood flow meter
Electromagnetic blood flow meterYuga Aravind Kumar
 
Medical Instrumentation- EMG,ERG,EOG,PCG
Medical Instrumentation- EMG,ERG,EOG,PCGMedical Instrumentation- EMG,ERG,EOG,PCG
Medical Instrumentation- EMG,ERG,EOG,PCGPoornima D
 
Respiratory Rate Measurement
Respiratory Rate MeasurementRespiratory Rate Measurement
Respiratory Rate Measurementgoverdhan765
 
Signal conditioning unit
Signal conditioning unit Signal conditioning unit
Signal conditioning unit BharathasreejaG
 
Blood gas analyser ppt
Blood gas analyser   pptBlood gas analyser   ppt
Blood gas analyser pptBena Ruth
 
Flow measurement in Blood flow
Flow measurement in Blood flowFlow measurement in Blood flow
Flow measurement in Blood flowVinothini S
 

What's hot (20)

Telemetry
TelemetryTelemetry
Telemetry
 
MEASUREMENT OF BIO POTENTIAL USING TWO ELECTRODES AND RECORDING PROBLEMS
MEASUREMENT OF BIO POTENTIAL  USING TWO ELECTRODES AND RECORDING PROBLEMSMEASUREMENT OF BIO POTENTIAL  USING TWO ELECTRODES AND RECORDING PROBLEMS
MEASUREMENT OF BIO POTENTIAL USING TWO ELECTRODES AND RECORDING PROBLEMS
 
Biomedical Instrumentation
Biomedical InstrumentationBiomedical Instrumentation
Biomedical Instrumentation
 
Ecg Signal Processing
Ecg Signal ProcessingEcg Signal Processing
Ecg Signal Processing
 
ECG
ECGECG
ECG
 
Telemetry system
Telemetry systemTelemetry system
Telemetry system
 
Transducers for bio medical
Transducers for bio medicalTransducers for bio medical
Transducers for bio medical
 
ECG machines -Operation and Maintenance
ECG machines -Operation and MaintenanceECG machines -Operation and Maintenance
ECG machines -Operation and Maintenance
 
Electrical safety
Electrical safetyElectrical safety
Electrical safety
 
Respiratory activity sensor
Respiratory activity sensorRespiratory activity sensor
Respiratory activity sensor
 
Instrumentation amplifier
Instrumentation amplifierInstrumentation amplifier
Instrumentation amplifier
 
Biopotentials
BiopotentialsBiopotentials
Biopotentials
 
Sensors for Biomedical Devices and systems
Sensors for Biomedical Devices and systemsSensors for Biomedical Devices and systems
Sensors for Biomedical Devices and systems
 
Electromagnetic blood flow meter
Electromagnetic blood flow meterElectromagnetic blood flow meter
Electromagnetic blood flow meter
 
Medical Instrumentation- EMG,ERG,EOG,PCG
Medical Instrumentation- EMG,ERG,EOG,PCGMedical Instrumentation- EMG,ERG,EOG,PCG
Medical Instrumentation- EMG,ERG,EOG,PCG
 
Respiratory Rate Measurement
Respiratory Rate MeasurementRespiratory Rate Measurement
Respiratory Rate Measurement
 
Signal conditioning unit
Signal conditioning unit Signal conditioning unit
Signal conditioning unit
 
Blood gas analyser ppt
Blood gas analyser   pptBlood gas analyser   ppt
Blood gas analyser ppt
 
Flow measurement in Blood flow
Flow measurement in Blood flowFlow measurement in Blood flow
Flow measurement in Blood flow
 
Signal conditioning
Signal conditioningSignal conditioning
Signal conditioning
 

Similar to Respiration rate measurement

Pulse Oxymetry , Inspired & Expired Gas Monitoring
Pulse Oxymetry ,  Inspired & Expired Gas MonitoringPulse Oxymetry ,  Inspired & Expired Gas Monitoring
Pulse Oxymetry , Inspired & Expired Gas MonitoringAleenaGigiYU
 
Capnography vs plethysmography
Capnography vs plethysmographyCapnography vs plethysmography
Capnography vs plethysmographySourav Mondal
 
Equipment Technology CEC
Equipment Technology CECEquipment Technology CEC
Equipment Technology CECsleepcaredotcom
 
Pulse oximetry capnography
Pulse oximetry capnography Pulse oximetry capnography
Pulse oximetry capnography kshama_db
 
MONITORINGpulseoxy44554545654543354454.pptx
MONITORINGpulseoxy44554545654543354454.pptxMONITORINGpulseoxy44554545654543354454.pptx
MONITORINGpulseoxy44554545654543354454.pptxrijjorajoo
 
psgbasics03-04-2017-
psgbasics03-04-2017-psgbasics03-04-2017-
psgbasics03-04-2017-SyidaAhmat
 
basic of polysomnography presentation.pptx
basic of polysomnography presentation.pptxbasic of polysomnography presentation.pptx
basic of polysomnography presentation.pptxshaitansingh8
 
Respiratory diagnostic studies and nursing responsibilities
Respiratory diagnostic studies and nursing responsibilitiesRespiratory diagnostic studies and nursing responsibilities
Respiratory diagnostic studies and nursing responsibilitiesRuma SEN
 
Reactive airway obstruction in children detection management_2018_pmm
Reactive airway obstruction  in children detection management_2018_pmmReactive airway obstruction  in children detection management_2018_pmm
Reactive airway obstruction in children detection management_2018_pmmParthiv Mehta
 
Pulse oximetry & capnography
Pulse oximetry & capnographyPulse oximetry & capnography
Pulse oximetry & capnographySapan Jena
 
08 capnometry and pulse oximetry
08 capnometry and pulse oximetry08 capnometry and pulse oximetry
08 capnometry and pulse oximetryDang Thanh Tuan
 
08 capnometry and pulse oximetry
08 capnometry and pulse oximetry08 capnometry and pulse oximetry
08 capnometry and pulse oximetryDang Thanh Tuan
 

Similar to Respiration rate measurement (20)

Pulse Oxymetry , Inspired & Expired Gas Monitoring
Pulse Oxymetry ,  Inspired & Expired Gas MonitoringPulse Oxymetry ,  Inspired & Expired Gas Monitoring
Pulse Oxymetry , Inspired & Expired Gas Monitoring
 
CAPNOMETER.pptx
CAPNOMETER.pptxCAPNOMETER.pptx
CAPNOMETER.pptx
 
Capnography vs plethysmography
Capnography vs plethysmographyCapnography vs plethysmography
Capnography vs plethysmography
 
Capnography
Capnography Capnography
Capnography
 
Equipment Technology CEC
Equipment Technology CECEquipment Technology CEC
Equipment Technology CEC
 
Pulse oximetry capnography
Pulse oximetry capnography Pulse oximetry capnography
Pulse oximetry capnography
 
Capnography in emergency room
Capnography in emergency roomCapnography in emergency room
Capnography in emergency room
 
MONITORINGpulseoxy44554545654543354454.pptx
MONITORINGpulseoxy44554545654543354454.pptxMONITORINGpulseoxy44554545654543354454.pptx
MONITORINGpulseoxy44554545654543354454.pptx
 
psgbasics03-04-2017-
psgbasics03-04-2017-psgbasics03-04-2017-
psgbasics03-04-2017-
 
Polysomnography
PolysomnographyPolysomnography
Polysomnography
 
basic of polysomnography presentation.pptx
basic of polysomnography presentation.pptxbasic of polysomnography presentation.pptx
basic of polysomnography presentation.pptx
 
Respiratory diagnostic studies and nursing responsibilities
Respiratory diagnostic studies and nursing responsibilitiesRespiratory diagnostic studies and nursing responsibilities
Respiratory diagnostic studies and nursing responsibilities
 
photoplethysmogram
photoplethysmogram photoplethysmogram
photoplethysmogram
 
CAPNOGRAPHY
CAPNOGRAPHYCAPNOGRAPHY
CAPNOGRAPHY
 
Reactive airway obstruction in children detection management_2018_pmm
Reactive airway obstruction  in children detection management_2018_pmmReactive airway obstruction  in children detection management_2018_pmm
Reactive airway obstruction in children detection management_2018_pmm
 
Pulse oximetry & capnography
Pulse oximetry & capnographyPulse oximetry & capnography
Pulse oximetry & capnography
 
08 capnometry and pulse oximetry
08 capnometry and pulse oximetry08 capnometry and pulse oximetry
08 capnometry and pulse oximetry
 
08 capnometry and pulse oximetry
08 capnometry and pulse oximetry08 capnometry and pulse oximetry
08 capnometry and pulse oximetry
 
Pulse Oxymetry
Pulse OxymetryPulse Oxymetry
Pulse Oxymetry
 
ETCO2 PPT.pptx
ETCO2 PPT.pptxETCO2 PPT.pptx
ETCO2 PPT.pptx
 

More from BharathasreejaG

IMPEDANCE MATCHING CIRCUIT AND FILTERS
IMPEDANCE MATCHING CIRCUIT AND FILTERSIMPEDANCE MATCHING CIRCUIT AND FILTERS
IMPEDANCE MATCHING CIRCUIT AND FILTERSBharathasreejaG
 
ARTIFACTS ON ECG RECORDING
ARTIFACTS ON ECG RECORDINGARTIFACTS ON ECG RECORDING
ARTIFACTS ON ECG RECORDINGBharathasreejaG
 
Blood pressure measurement
Blood pressure measurementBlood pressure measurement
Blood pressure measurementBharathasreejaG
 
Temperature measurement
Temperature measurement Temperature measurement
Temperature measurement BharathasreejaG
 
Carrier and chopper amplifiers
Carrier and chopper amplifiers Carrier and chopper amplifiers
Carrier and chopper amplifiers BharathasreejaG
 
Types of bio amplifiers
Types of bio amplifiers Types of bio amplifiers
Types of bio amplifiers BharathasreejaG
 
Bio signal characteristics and recording modes
Bio signal characteristics and recording modesBio signal characteristics and recording modes
Bio signal characteristics and recording modesBharathasreejaG
 
Electrodes in medical field
Electrodes in medical fieldElectrodes in medical field
Electrodes in medical fieldBharathasreejaG
 
BASICS OF BIOMEDICAL INSTRUMENTATION
BASICS OF BIOMEDICAL INSTRUMENTATIONBASICS OF BIOMEDICAL INSTRUMENTATION
BASICS OF BIOMEDICAL INSTRUMENTATIONBharathasreejaG
 

More from BharathasreejaG (10)

IMPEDANCE MATCHING CIRCUIT AND FILTERS
IMPEDANCE MATCHING CIRCUIT AND FILTERSIMPEDANCE MATCHING CIRCUIT AND FILTERS
IMPEDANCE MATCHING CIRCUIT AND FILTERS
 
ARTIFACTS ON ECG RECORDING
ARTIFACTS ON ECG RECORDINGARTIFACTS ON ECG RECORDING
ARTIFACTS ON ECG RECORDING
 
Pulse rate measurement
Pulse rate measurement Pulse rate measurement
Pulse rate measurement
 
Blood pressure measurement
Blood pressure measurementBlood pressure measurement
Blood pressure measurement
 
Temperature measurement
Temperature measurement Temperature measurement
Temperature measurement
 
Carrier and chopper amplifiers
Carrier and chopper amplifiers Carrier and chopper amplifiers
Carrier and chopper amplifiers
 
Types of bio amplifiers
Types of bio amplifiers Types of bio amplifiers
Types of bio amplifiers
 
Bio signal characteristics and recording modes
Bio signal characteristics and recording modesBio signal characteristics and recording modes
Bio signal characteristics and recording modes
 
Electrodes in medical field
Electrodes in medical fieldElectrodes in medical field
Electrodes in medical field
 
BASICS OF BIOMEDICAL INSTRUMENTATION
BASICS OF BIOMEDICAL INSTRUMENTATIONBASICS OF BIOMEDICAL INSTRUMENTATION
BASICS OF BIOMEDICAL INSTRUMENTATION
 

Recently uploaded

CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxGaneshChakor2
 
Presiding Officer Training module 2024 lok sabha elections
Presiding Officer Training module 2024 lok sabha electionsPresiding Officer Training module 2024 lok sabha elections
Presiding Officer Training module 2024 lok sabha electionsanshu789521
 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptxVS Mahajan Coaching Centre
 
URLs and Routing in the Odoo 17 Website App
URLs and Routing in the Odoo 17 Website AppURLs and Routing in the Odoo 17 Website App
URLs and Routing in the Odoo 17 Website AppCeline George
 
APM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAPM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAssociation for Project Management
 
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptx
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptxContemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptx
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptxRoyAbrique
 
Accessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactAccessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactdawncurless
 
mini mental status format.docx
mini    mental       status     format.docxmini    mental       status     format.docx
mini mental status format.docxPoojaSen20
 
Hybridoma Technology ( Production , Purification , and Application )
Hybridoma Technology  ( Production , Purification , and Application  ) Hybridoma Technology  ( Production , Purification , and Application  )
Hybridoma Technology ( Production , Purification , and Application ) Sakshi Ghasle
 
PSYCHIATRIC History collection FORMAT.pptx
PSYCHIATRIC   History collection FORMAT.pptxPSYCHIATRIC   History collection FORMAT.pptx
PSYCHIATRIC History collection FORMAT.pptxPoojaSen20
 
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Sapana Sha
 
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991RKavithamani
 
Paris 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityParis 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityGeoBlogs
 
Alper Gobel In Media Res Media Component
Alper Gobel In Media Res Media ComponentAlper Gobel In Media Res Media Component
Alper Gobel In Media Res Media ComponentInMediaRes1
 
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxPOINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxSayali Powar
 
MENTAL STATUS EXAMINATION format.docx
MENTAL     STATUS EXAMINATION format.docxMENTAL     STATUS EXAMINATION format.docx
MENTAL STATUS EXAMINATION format.docxPoojaSen20
 
Concept of Vouching. B.Com(Hons) /B.Compdf
Concept of Vouching. B.Com(Hons) /B.CompdfConcept of Vouching. B.Com(Hons) /B.Compdf
Concept of Vouching. B.Com(Hons) /B.CompdfUmakantAnnand
 

Recently uploaded (20)

Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1Código Creativo y Arte de Software | Unidad 1
Código Creativo y Arte de Software | Unidad 1
 
CARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptxCARE OF CHILD IN INCUBATOR..........pptx
CARE OF CHILD IN INCUBATOR..........pptx
 
Presiding Officer Training module 2024 lok sabha elections
Presiding Officer Training module 2024 lok sabha electionsPresiding Officer Training module 2024 lok sabha elections
Presiding Officer Training module 2024 lok sabha elections
 
TataKelola dan KamSiber Kecerdasan Buatan v022.pdf
TataKelola dan KamSiber Kecerdasan Buatan v022.pdfTataKelola dan KamSiber Kecerdasan Buatan v022.pdf
TataKelola dan KamSiber Kecerdasan Buatan v022.pdf
 
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions  for the students and aspirants of Chemistry12th.pptxOrganic Name Reactions  for the students and aspirants of Chemistry12th.pptx
Organic Name Reactions for the students and aspirants of Chemistry12th.pptx
 
Staff of Color (SOC) Retention Efforts DDSD
Staff of Color (SOC) Retention Efforts DDSDStaff of Color (SOC) Retention Efforts DDSD
Staff of Color (SOC) Retention Efforts DDSD
 
URLs and Routing in the Odoo 17 Website App
URLs and Routing in the Odoo 17 Website AppURLs and Routing in the Odoo 17 Website App
URLs and Routing in the Odoo 17 Website App
 
APM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across SectorsAPM Welcome, APM North West Network Conference, Synergies Across Sectors
APM Welcome, APM North West Network Conference, Synergies Across Sectors
 
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptx
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptxContemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptx
Contemporary philippine arts from the regions_PPT_Module_12 [Autosaved] (1).pptx
 
Accessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impactAccessible design: Minimum effort, maximum impact
Accessible design: Minimum effort, maximum impact
 
mini mental status format.docx
mini    mental       status     format.docxmini    mental       status     format.docx
mini mental status format.docx
 
Hybridoma Technology ( Production , Purification , and Application )
Hybridoma Technology  ( Production , Purification , and Application  ) Hybridoma Technology  ( Production , Purification , and Application  )
Hybridoma Technology ( Production , Purification , and Application )
 
PSYCHIATRIC History collection FORMAT.pptx
PSYCHIATRIC   History collection FORMAT.pptxPSYCHIATRIC   History collection FORMAT.pptx
PSYCHIATRIC History collection FORMAT.pptx
 
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111Call Girls in Dwarka Mor Delhi Contact Us 9654467111
Call Girls in Dwarka Mor Delhi Contact Us 9654467111
 
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
Industrial Policy - 1948, 1956, 1973, 1977, 1980, 1991
 
Paris 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activityParis 2024 Olympic Geographies - an activity
Paris 2024 Olympic Geographies - an activity
 
Alper Gobel In Media Res Media Component
Alper Gobel In Media Res Media ComponentAlper Gobel In Media Res Media Component
Alper Gobel In Media Res Media Component
 
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptxPOINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
POINT- BIOCHEMISTRY SEM 2 ENZYMES UNIT 5.pptx
 
MENTAL STATUS EXAMINATION format.docx
MENTAL     STATUS EXAMINATION format.docxMENTAL     STATUS EXAMINATION format.docx
MENTAL STATUS EXAMINATION format.docx
 
Concept of Vouching. B.Com(Hons) /B.Compdf
Concept of Vouching. B.Com(Hons) /B.CompdfConcept of Vouching. B.Com(Hons) /B.Compdf
Concept of Vouching. B.Com(Hons) /B.Compdf
 

Respiration rate measurement

  • 1. R E S P I R A T I O N R A T E M E A S U R E M E N T BASICS OF BIOMEDICAL INSTRUMENTATION
  • 4. FUNCTION OF RESPIRATION SYSTEM ❖ The function of respiration system are to supply oxygen and remove carbon dioxide from the tissues. ❖ The action of breathing is controlled by a muscular action causing the volume of the lung to increase and decrease. 4
  • 6. DISPLACEMENT METHOD • The transducer is hold by an elastic band which goes around the chest. • The respiratory movements results in a corresponding resistance changes of the strain gauge. • It is connected as one arm of a wheat stone bridge circuit. • Its output varies with chest expansion. This output corresponds to respiration activity. 6
  • 7. THERMISTOR METHOD ▸ Generally there is a temperature difference between inspired and expired air. ▸ This temperature is sensed by placing thermistor infront of nostrils. ▸ Thermistor is placed by suitable stand. Thermistor is connected with the bridge circuit. ▸ So unbalance signal is amplified to get the respiratory activity. 7
  • 9. IMPEDANCE PNEUMOGRAPHY • This is the indirect method of measurement. • Impedance across the chest of a patient changes as respiration occurs. • 50Hz ac signal is produced by oscillator circuit and it is given to chest of patient through electrodes. • The signal voltage applied to the amplifier block the voltage drop across the resistance. Where, V  O/PVoltage I  Current through the chest R  Chest impedance without respiration R  Change of chest impedance due to respiration. 9 V = I (R ± R)
  • 10. CO2 METHOD • Respiration rate can be measured by measuring CO2 in expired air. • This CO2 method is based on the absorption property of infrared rays • When infrared rays are passed through the expired air which contains certain amount of CO2 some of the radiation are absorbed by it. So there is a loss of heat energy associated with the rays. • The detector changes the loss in heating effect of the rays into an electrical signal. It is used to get the average respiration rate. 10
  • 11. 11
  • 12. ▸ Two infrared sources are available in this setup. ▸ The beam from one infrared source falls on the Test cuvette side. ▸ The beam from another infrared source falls in the reference cuvette side. ▸ The detector has 2 identical portions. These portions are separated by a thin, flexible metal diaphragm. ▸ The detector is filled with a sample of pure CO2. Because of the absorption of CO2 in the test cuvette, the beam falling on the test side of the detector is weaker than that falling on the reference side is heated more than that on the test side. So the diaphragm is pushed slightly to the test side of the detector. This diaphragm forms one plate of a capacitor. ▸ The ac signal appears across the detector is amplified and recorded using recorder. The amplified output is integrated and shown in this method. It is used for continuous monitoring the respiration rate.12
  • 13. APNOEA DETECTORS • Apnoea is the stoppage of breathing. • It can be occurred at the conditions like head injury, drug over dose, etc. • Apnoea can also occur in premature babies during first weeks of life because of their immature nervous system. • If Apnoea persists for prolonged period, then brain function can be severely damaged. So Apnoea patients are closely monitored. • Apnoea monitor gives audio signal and visual signal (flashing of light), when no inspiration occurs for a particular period of time. 13
  • 14. 14
  • 15. ▸ Motion channel block differentiates motion (disturbances) and the respiration based on the frequency. ▸ The frequency below 1.5Hz is identified as respiration. ▸ The frequency above 1.5Hz is identified as motion. ▸ High frequency signal above the threshold is sensed by positive detector. ▸ The frequency below the threshold is sensed by negative detector. ▸ Output of motion channel is connected with comparator circuit. It compares the amplitude of motion and respiration. Based on the output corresponding lamp will glow. ▸ If there is no respiration, then Schmitt trigger circuit gives the output to switch on the alarm. ▸ Apnoea period selector circuit contains low frequency alarm oscillator and tone oscillator and audio amplifier. ▸ Apnoea period selector circuit drives the alarm circuit. ▸ The output of the alarm circuit is connected with the speaker. ▸ So when there is no respiration for a period of 10s or 20s, then audio signal through the speaker and visual signal through the flash light is delivered. 15
  • 16. 16