SlideShare a Scribd company logo
Use of Biological
Variation as an
Uncertainty
Component
Evaluation Methods of Uncertainty of
Measurement for Medical Testing
https://consultglp.com
by
Yeoh Guan Huah
GLP Consulting, Singapore
Outline
• A review of the uncertainty of measurement
(UM) calculation methods by component-by-
component approach
• Necessity to consider non-analytical
uncertainty components in addition to the
analytical uncertainty in the laboratory
• Biological variation – what and why
• How to incorporate biological variation in the
overall evaluation of UM.
Introduction
• The ultimate use of analysis data with its
associated uncertainty of measurement (UM)
must be fit for intended purpose
• Data generated by medical laboratory are used
for any signaling which may differentiate health
from disease, based on some pre-set reference
ranges for diagnostic decision
• Amongst other applications, these data are also
adopted for clinical monitoring of a patient upon
treatment.
Introduction
• Both ISO 15189 and ISO/IEC 17025 accreditation standards
have given uncertainty of measurement a new
perspective.
• They have recognized that apart from the analytical
uncertainty during the testing process, there are other
non-analytical uncertainty components to be considered
as well.
• For industrial, environmental and commodity testing,
ISO/IEC 17025 stresses the importance of taking
representative samples from the lot (population) for
analysis. Hence, sampling uncertainty is incorporated as
another uncertainty contributor, if applicable.
• For clinical application of a test result, the in vivo biological
variability of the measurand as an uncertainty component
cannot be ignored for clinical interpretation.
Reported result with uncertainty : X + U (units)
• Measurement uncertainty U is expressed
as:
• Test result : X + U
• and, U = k x u
• where k is a coverage factor (usually k = 2 or
1.96 for 95% confidence), and
• u, the combined standard uncertainty
expressed as standard deviation
• U is also called expanded uncertainty,
assumed rightly a normal probability
distribution of data
5
Example:
Consider an expression for a
measurand (urea in serum):
“5.0 + 0.2 mmol/L with 95%
confidence”
The + 0.2 mmol/L is the
uncertainty ( U ) of the supposed
5.0 mmol/L serum urea
measurement.
Its combined standard
uncertainty (u ) therefore, is
0.2/2 or 0.1 mmol/L
Basic methods for combining uncertainty
components
• Calculated from a sum and/or a difference of independent
measurements with associated standard uncertainties
• Let: R = X + Y or R = X – Y
• Then, (uR)2 = (uX)2 + (uY)2
• where uR , uX and uY are the respective analytical
standard uncertainty (as standard deviation)
• Therefore, combined standard uncertainty uR = (uR)2
• Note: X and Y are of same units and involved in addition
& subtraction conditions
Worked example No. 1
• Anion gap (AG) is calculated based on the results of serum (plasma)
ions: sodium, potassium, chloride and bicarbonate, in the form:
• AG = (Na+ + K+) – (Cl- + HCO3-)
Ion,
mmol/L Test result
Standard
deviation, u
Na+ 140 1.2
K+ 3.9 0.1
Cl- 102 1.3
HCO3- 24 1.2
Therefore,
AG = (140+3.9) – (102 + 24) = 17.9 mmol/L
𝐶𝑜𝑚𝑏𝑖𝑛𝑒𝑑 𝑆𝐷 𝑅= (1.2)2+(0.1)2+(1.3)2+(1.2)2
= 2.1 mmol/L
Report:
Anion Gap (AG) = 18 + 4 mmol/L
with a coverage factor of 2 (95% confidence)
Basic methods for combining uncertainty components
• From multiplication and/or division of independent measurements
• Using the Propagation Law of Standard Deviation,
• Let:
𝑢(𝑦)2 =
𝑖=1
𝑛
𝜕𝑓
𝜕𝑥𝑖
2
𝑢(𝑥𝑖)2
• R =
𝑃×𝑉
𝑇
with standard
deviations uP , uV , uT , then,
• 𝑐 𝑃 =
𝜕𝑅
𝜕𝑃
=
𝑉
𝑇
• 𝑐 𝑉 =
𝜕𝑅
𝜕𝑉
=
𝑃
𝑇
• 𝑐 𝑇 =
𝜕𝑅
𝜕𝑇
= −
𝑃×𝑉
𝑇2
𝑢 𝑅
2 = 𝑐 𝑃
2
𝑢 𝑃
2 + 𝑐 𝑉
2
𝑢 𝑉
2 + 𝑐 𝑇
2
𝑢 𝑇
2
= (
𝑉
𝑇
)2 𝑢 𝑃
2 + (
𝑃
𝑇
)2 𝑢 𝑉
2 + (−
𝑃𝑉
𝑇2)2 𝑢 𝑇
2
Dividing both sides by 𝑅2 = (
𝑃×𝑉
𝑇
)2 gives:
(
𝑢 𝑅
𝑅
)2
= (
𝑢 𝑃
𝑃
)2
+ (
𝑢 𝑉
𝑉
)2
+ (
𝑢 𝑇
𝑇
)2
, or
𝑢 𝑅
𝑅
= (
𝑢 𝑃
𝑃
)2+ (
𝑢 𝑉
𝑉
)2 + (
𝑢 𝑇
𝑇
)2
𝒊. 𝒆. , 𝑪𝑽 𝑹 = 𝑪𝑽 𝑷
𝟐
+ 𝑪𝑽 𝑽
𝟐
+ 𝑪𝑽 𝑻
𝟐
Worked example No. 2
• In a creatinine clearance test, the following calculation applies:
• C = (U x V ) / (P x T )
Parameter Unit Value u
U mmol/L 10.0 0.25
V ml 1500 15
P mmol/L 0.1 0.01
T secs 86400 0
By calculation,
C = (10.0 mmol/L x 1500 ml)/(0.1 mmol/L x 24 hrs x 3600 sec)
= 1.74 ml/sec
and,
(
𝑢 𝑅
1.74
)2
=(
0.25
𝑈10.0
)2
+(
15
1500
)2
+(
0.01
0.1
)2
+(
0
86400
)2
= 0.0107
Therefore,
uR = 1.74 x 0.0107 = 0.18 ml/sec
Report:
Creatinine clearance C = 1.74 + 0.36 ml/sec with the
coverage Factor of 2 for 95% confidence
Non-analytical uncertainty contributors
• In medical testing, there are many
potential uncertainty contributors
which can significantly affect test
results, apart from the analytical
process itself, such as:
• Improper practice during specimen
collection
• Poor specimen handling during transit
to the laboratory
• Patient related factors like biological
variability and the presence of any drug
residues
Analytical uncertainty component
• The laboratory’s analytical quality comes from two factors for total
analytical error as its analytical goal:
• Intermediate precision (random error; imprecision)
• Analytical bias (systematic error)
• Intermediate precision or imprecision (ua) is estimated from its
long-term combined (pooled) standard deviation after a series of
analysis in replicates by different analysts within the laboratory on
different occasions on a stable internal quality control serum
sample.
• Analytical bias is evaluated by studying any significant difference
between the average test result against the target or certified value
of the internal QC sample through the Student’s t-test. If there is a
significance difference between the mean result and the certified
value, then the standard uncertainty of bias (ub) is estimated.
Why do we consider the biological variation?
• There are many clinical uses of the test results reported.
• As a test result must be ‘fit for purpose’, there is a need to assess its clinical
goal in addition to the analytical goal.
• For some measurands, an analytical goal may not be physiologically or clinically
relevant. Then, the method imprecision study would be sufficient.
• However most medical laboratory need to identify if UM information when
reported could significantly affect clinical interpretations and patient
management.
• Hence, there is a necessity to have additional non-analytical uncertainty
component in the overall estimation of UM to set such goal.
• This additional uncertainty component is the intra-individual biological
variation of the measurand.
Components of biological variation (BV)
• Actually, there are two types of components of biological
variation, namely:
• Intra-individual (random fluctuation of analytes around the
setting point of each individual), sometimes termed as “within-
subject” BV;
• Inter-individual (overall variation from the different person’s
setting point), sometimes termed as “between-subject” BV.
• The laboratory’s analytical method, instrument and
reagents do not make difference in BV estimates.
Some applications of biological variation
 Evaluating the clinical significance of changes in consecutive test
results from an individual in monitoring treatment progress
 Setting quality specifications for analytical performance (usually
targeting the analytical goal as a percentage of BV) to satisfy the
general needs of diagnosis and monitoring
 Validating new procedures in a laboratory
 Assessing the usefulness of population-based reference values
 Determining which type of specimen (e.g. plasma, serum, 24-hr
urine, first-morning urine) is optimal for analyzing a specific
constituent
Relationship of analytical imprecision and biological
variation
• There are 3 levels of analytical goal for long-term imprecision
(CVa) based on intra-individual biological variation (CVi):
• Optimum: CVa < 0.25 x CVi
• Desirable: CVa < 0.50 x CVi
• Minimum: CVa < 0.75 x CVi
• For analytes with unavailable CVi data, other criteria may be
used, such as relative performance in inter-laboratory
comparison studies or proficiency testing programs, certified
reference interval, professional clinical opinion, etc.
Relationship of analytical imprecision and biological
variation
• If test results are interpreted using reference or clinical decision
values determined by a different method, bias should be considered
as a component of uncertainty to be estimated.
• The inter-individual biological variation CVg is to be incorporated.
• Like imprecision, we have to ensure the three levels of analytical
goal for bias (CVb) based on biological variation are met:
• Optimum:
• Desirable:
• Minimum:
𝐶𝑉𝑏 ≤ 0.125 × 𝐶𝑉𝑖
2
+ 𝐶𝑉𝑔
2
𝐶𝑉𝑏 ≤ 0.250 × 𝐶𝑉𝑖
2
+ 𝐶𝑉𝑔
2
𝐶𝑉𝑏 ≤ 0.375 × 𝐶𝑉𝑖
2
+ 𝐶𝑉𝑔
2
Biological variation database
• There are study findings which indicate that the factors
inherent to the subject, such as sex, age, race,
geographical location, do not produce changes in BV
results
• Practically all the currently available estimates of
biological variation (more than 300 analytes) have been
compiled in a database:
https://www.westgard.com/biodatabase1.htm
• The intra-individual (or within-subject) coefficients of
variation (in %) from the database can be used to
evaluate the clinical significance of changes in two
consecutive results from a patient, in addition to the
analytical variation.
Summation of analytical uncertainty &
biological variation
• If the combined estimate of coefficient of variation of
reported test result (CVT) covers both the CV of analytical
imprecision (CVa) and that of within-subject biological
variation (CVi) for a comparison of two consecutive test
results from a patient, we then have:
• CVT
2 = CVa
2 + CVi
2
• Note: If test results are clinically interpreted by comparison
with reference or previous values produced by the same
analytical method, analytical bias component is not required.
Worked example No. 3
• Plasma alkaline phosphatase (ALP) activity for a patient:
• 1st day’s result : 95 U/L; 3rd day’s result : 108 U/L
• UM for ALP : CVa = 1.45% at QC sample mean : 87 U/L
• ALP intro-individual BV (given by Westgard website): CVi = 6.45%
• Were these two sets of daily test results of any significant
different?
• Sum of analytical & BV as CV’s:
• CVT = (CVa
2 + CVi
2) = (1.452 + 6.452) = 6.61%
Worked example No. 3 (contd.)
• If the two test results are analytically and biologically different, the “critical difference”
or Reference Change Value (RCV) is to be:
• 1.96 x 2 x (CVa
2 + CVi
2) or 2.77 x (CVa
2 + CVi
2) for 95% confidence
• That means the two results need to differ by 2.77 x 6.61% = 18.3%
• Now, the first day’s result was 95 U/L, thus
• 95 U/L + (95 U/L x 18.3%) = 95 + 17.4 = 112 U/L
• and, 95 U/L - (95 U/L x 18.3%) = 95 - 17.4 = 77.6 U/L
• So, the critical range of expanded uncertainty is [77.6 U/L, 112 U/L]
• The second result would have to be at least 112 U/L for there to be 95% confidence that
it was both analytically and biologically different from the first result. We can therefore
conclude that the test results of 95 U/L and 108 U/L are analytically different, but
probably not biologically different.
A note on
Reference
Change Value
RCV
• Some references recommend to use a one-tail test with a
probability risk (α) of 5% error, instead of a two-tail test.
• In that case, the Reference Change Value has a formula:
• 1.645 x 2 x (CVa
2 + CVi
2) or 2.33 x (CVa
2 + CVi
2)
• By using this new formula, the two results need to differ by
2.33 x 6.61% = 15.4%
• Now, the first day’s result was 95 U/L, thus the upper critical
limit is:
• 95 U/L + (95 U/L x 15.4%) = 95 + 14.7 = 109.7 U/L
• Same conclusion could be made based on the new RCV.
ααα/2α/2
The choice of 1-tail or 2-
tail probability depends
on the alternative
hypothesis H1
How do we get the factor 1.96 x 2 or 1.645 x 2?
• A measurement result, X, is associated with a standard deviation +sX
• The duplicate result, Y, on the same sample is also associated with the
same standard deviation +sX
• When an average value is calculated for X and Y, we have a total
variance of
• sX
2 + sX
2 = 2sX
2
• Therefore, the pooled standard deviation is 2 x sX
• The choice of coverage factor k = 1.96 or 1.645 depend on either 2-tail
or 1-tail hypothesis testing, assuming a normal distribution with an
error probability α = 0.05 or 95% confidence.
Conclusion
• It is a good practice to maintain a list of measurands with their
respective RCV (reference change value) to define healthy and
non-healthy situations.
• Remember that the RCV is applied for the target analyte in each
pair of consecutive results from the same patient
• The laboratory information system (LIMS) can be set to flag off
the second report with a predefined signal after the numeric
result of the target analyte is keyed in.
• Most important challenge for the medical personnel is how to
explain the UM concept to requesting doctors in a clinically
meaningful way.
References
• “Uncertainty of Measurement in Quantitative
Medical Testing – A Laboratory Implementation
Guide” by Australasian Association of Clinical
Biochemists, November 2004
• “Application of Biological Variation – A Review” by
Ricos C. et al, Biochemia Medica 2009, 19(3):250-9
• “Desirable Biological Variation Database
specifications” by James Westgard,
https://www.westgard.com/biodatabase1.htm

More Related Content

What's hot

Total Quality Management (TQM) by Dr Anurag Yadav
Total Quality Management (TQM) by Dr Anurag YadavTotal Quality Management (TQM) by Dr Anurag Yadav
Total Quality Management (TQM) by Dr Anurag Yadav
Dr Anurag Yadav
 
Laboratory Quality Control .ppt
Laboratory Quality Control .pptLaboratory Quality Control .ppt
Laboratory Quality Control .ppt
ABRARAHMED767665
 
Quality management
Quality managementQuality management
Quality management
Tapeshwar Yadav
 
Laboratory accreditation by iso 15189
Laboratory accreditation by iso 15189Laboratory accreditation by iso 15189
Laboratory accreditation by iso 15189
CHRISTIAN MEDICAL COLLEGE AND HOSPITAL
 
Harmonization of Laboratory Indicators, 09 03-2017
Harmonization of Laboratory Indicators, 09 03-2017Harmonization of Laboratory Indicators, 09 03-2017
Harmonization of Laboratory Indicators, 09 03-2017
Ola Elgaddar
 
Total Quality Management in Medical Laboratories
Total Quality Management in Medical LaboratoriesTotal Quality Management in Medical Laboratories
Total Quality Management in Medical Laboratories
CHRISTIAN MEDICAL COLLEGE AND HOSPITAL
 
Internal quality control in clinical laboratories hematology(2)
Internal quality control in clinical laboratories hematology(2)Internal quality control in clinical laboratories hematology(2)
Internal quality control in clinical laboratories hematology(2)
NAZAR ABU-DULLA
 
Presentation on ISO 15189 documentation
Presentation on ISO 15189 documentationPresentation on ISO 15189 documentation
Presentation on ISO 15189 documentation
Global Manager Group
 
Quality Control In Clinical Laboratory
Quality Control In Clinical LaboratoryQuality Control In Clinical Laboratory
Quality Control In Clinical Laboratory
Dr. Rajesh Bendre
 
causes of laboratory errors
causes of laboratory errorscauses of laboratory errors
causes of laboratory errors
Shahad Bakhdar
 
Automation
AutomationAutomation
Automation
Tapeshwar Yadav
 
Quality Assurance of Laboratory Test Results based on ISO/IEC 17025
Quality Assurance of Laboratory Test Results based on ISO/IEC 17025Quality Assurance of Laboratory Test Results based on ISO/IEC 17025
Quality Assurance of Laboratory Test Results based on ISO/IEC 17025
PECB
 
Quality control in the medical laboratory
Quality control in the medical laboratoryQuality control in the medical laboratory
Quality control in the medical laboratory
Adnan Jaran
 
1.qualitycontrol final
1.qualitycontrol final1.qualitycontrol final
1.qualitycontrol final
Ishah Khaliq
 
Quality control
Quality controlQuality control
Quality control
Tapeshwar Yadav
 
Quality Control in Laboratory
Quality Control in LaboratoryQuality Control in Laboratory
Quality Control in Laboratory
Manal Elsayed CPPS, CPHQ, CLSSBB, FISQua, DTQM
 
External Quality Assurance of Serological Testing
External Quality Assurance of Serological TestingExternal Quality Assurance of Serological Testing
External Quality Assurance of Serological Testing
Arjuna Samaranayaka
 
Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods
Mostafa Mahmoud
 
Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014
Adel Elazab Elged
 
Haemoglobin quality control by maintaining levey jennings chart
Haemoglobin quality control by maintaining levey jennings chartHaemoglobin quality control by maintaining levey jennings chart
Haemoglobin quality control by maintaining levey jennings chart
Dr Rashmi Sood
 

What's hot (20)

Total Quality Management (TQM) by Dr Anurag Yadav
Total Quality Management (TQM) by Dr Anurag YadavTotal Quality Management (TQM) by Dr Anurag Yadav
Total Quality Management (TQM) by Dr Anurag Yadav
 
Laboratory Quality Control .ppt
Laboratory Quality Control .pptLaboratory Quality Control .ppt
Laboratory Quality Control .ppt
 
Quality management
Quality managementQuality management
Quality management
 
Laboratory accreditation by iso 15189
Laboratory accreditation by iso 15189Laboratory accreditation by iso 15189
Laboratory accreditation by iso 15189
 
Harmonization of Laboratory Indicators, 09 03-2017
Harmonization of Laboratory Indicators, 09 03-2017Harmonization of Laboratory Indicators, 09 03-2017
Harmonization of Laboratory Indicators, 09 03-2017
 
Total Quality Management in Medical Laboratories
Total Quality Management in Medical LaboratoriesTotal Quality Management in Medical Laboratories
Total Quality Management in Medical Laboratories
 
Internal quality control in clinical laboratories hematology(2)
Internal quality control in clinical laboratories hematology(2)Internal quality control in clinical laboratories hematology(2)
Internal quality control in clinical laboratories hematology(2)
 
Presentation on ISO 15189 documentation
Presentation on ISO 15189 documentationPresentation on ISO 15189 documentation
Presentation on ISO 15189 documentation
 
Quality Control In Clinical Laboratory
Quality Control In Clinical LaboratoryQuality Control In Clinical Laboratory
Quality Control In Clinical Laboratory
 
causes of laboratory errors
causes of laboratory errorscauses of laboratory errors
causes of laboratory errors
 
Automation
AutomationAutomation
Automation
 
Quality Assurance of Laboratory Test Results based on ISO/IEC 17025
Quality Assurance of Laboratory Test Results based on ISO/IEC 17025Quality Assurance of Laboratory Test Results based on ISO/IEC 17025
Quality Assurance of Laboratory Test Results based on ISO/IEC 17025
 
Quality control in the medical laboratory
Quality control in the medical laboratoryQuality control in the medical laboratory
Quality control in the medical laboratory
 
1.qualitycontrol final
1.qualitycontrol final1.qualitycontrol final
1.qualitycontrol final
 
Quality control
Quality controlQuality control
Quality control
 
Quality Control in Laboratory
Quality Control in LaboratoryQuality Control in Laboratory
Quality Control in Laboratory
 
External Quality Assurance of Serological Testing
External Quality Assurance of Serological TestingExternal Quality Assurance of Serological Testing
External Quality Assurance of Serological Testing
 
Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods Validation of lab instruments and quantitative test methods
Validation of lab instruments and quantitative test methods
 
Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014Laboratory Internal Quality Control presentation master revision, 2014
Laboratory Internal Quality Control presentation master revision, 2014
 
Haemoglobin quality control by maintaining levey jennings chart
Haemoglobin quality control by maintaining levey jennings chartHaemoglobin quality control by maintaining levey jennings chart
Haemoglobin quality control by maintaining levey jennings chart
 

Similar to Biological variation as an uncertainty component

Clinical lab qc sethu
Clinical lab qc sethuClinical lab qc sethu
Clinical lab qc sethu
subramaniam sethupathy
 
2. Quality Control Notes
2. Quality Control Notes2. Quality Control Notes
2. Quality Control Notes
Leah Molai
 
Quality control
Quality controlQuality control
Quality control
SKYFALL
 
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
Aamir Ijaz Brig
 
Data analysis ( Bio-statistic )
Data analysis ( Bio-statistic )Data analysis ( Bio-statistic )
Data analysis ( Bio-statistic )
Amany Elsayed
 
Qc and qa
Qc and qaQc and qa
Qc and qa
Zahid Chughtai
 
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Randox
 
Quality Control in Pathological Laboratory
Quality Control in Pathological LaboratoryQuality Control in Pathological Laboratory
Quality Control in Pathological Laboratory
sanarehman8159
 
Control of analytical quality using stable control materials postgrad
Control of analytical quality using stable control materials postgradControl of analytical quality using stable control materials postgrad
Control of analytical quality using stable control materials postgrad
MedhatEldeeb2
 
SHE, Quality, and Ethics in Medical Laboratories - PCLP
SHE, Quality, and Ethics in Medical Laboratories - PCLPSHE, Quality, and Ethics in Medical Laboratories - PCLP
SHE, Quality, and Ethics in Medical Laboratories - PCLP
AlAcademia Tsr
 
Bioequivalence of Highly Variable Drug Products
Bioequivalence of Highly Variable Drug ProductsBioequivalence of Highly Variable Drug Products
Bioequivalence of Highly Variable Drug Products
Bhaswat Chakraborty
 
Basics of laboratory internal quality control, Ola Elgaddar, 2012
Basics of laboratory internal quality control, Ola Elgaddar, 2012Basics of laboratory internal quality control, Ola Elgaddar, 2012
Basics of laboratory internal quality control, Ola Elgaddar, 2012
Ola Elgaddar
 
Bioequivalence of Highly Variable Drug Products
Bioequivalence of Highly Variable Drug ProductsBioequivalence of Highly Variable Drug Products
Bioequivalence of Highly Variable Drug Products
Bhaswat Chakraborty
 
Quality control.pptx
Quality control.pptxQuality control.pptx
Quality control.pptx
aditisikarwar2
 
Variability of clinical chemistry laboratory results
Variability of clinical  chemistry laboratory resultsVariability of clinical  chemistry laboratory results
Variability of clinical chemistry laboratory results
AdetokunboAjala
 
quality control in clinical laboratory
quality control in clinical laboratory quality control in clinical laboratory
quality control in clinical laboratory
DrmanarEmam
 
Cliniudorohdkdihkshckshsbdjhjdidgjdisybsjdoh
CliniudorohdkdihkshckshsbdjhjdidgjdisybsjdohCliniudorohdkdihkshckshsbdjhjdidgjdisybsjdoh
Cliniudorohdkdihkshckshsbdjhjdidgjdisybsjdoh
ssuser8691ad
 
Quality management system BY M .REJA
Quality management system BY M .REJAQuality management system BY M .REJA
Quality management system BY M .REJA
MamunReja1
 
Validity andreliability
Validity andreliabilityValidity andreliability
Validity andreliability
nuwan udugampala
 
Bioanalytical method validation usfda
Bioanalytical method validation usfdaBioanalytical method validation usfda
Bioanalytical method validation usfda
gopinathannsriramachandraeduin
 

Similar to Biological variation as an uncertainty component (20)

Clinical lab qc sethu
Clinical lab qc sethuClinical lab qc sethu
Clinical lab qc sethu
 
2. Quality Control Notes
2. Quality Control Notes2. Quality Control Notes
2. Quality Control Notes
 
Quality control
Quality controlQuality control
Quality control
 
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
Quality Control for Quantitative Tests by Prof Aamir Ijaz (Pakistan)
 
Data analysis ( Bio-statistic )
Data analysis ( Bio-statistic )Data analysis ( Bio-statistic )
Data analysis ( Bio-statistic )
 
Qc and qa
Qc and qaQc and qa
Qc and qa
 
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
Basic QC Statistics - Improving Laboratory Performance Through Quality Contro...
 
Quality Control in Pathological Laboratory
Quality Control in Pathological LaboratoryQuality Control in Pathological Laboratory
Quality Control in Pathological Laboratory
 
Control of analytical quality using stable control materials postgrad
Control of analytical quality using stable control materials postgradControl of analytical quality using stable control materials postgrad
Control of analytical quality using stable control materials postgrad
 
SHE, Quality, and Ethics in Medical Laboratories - PCLP
SHE, Quality, and Ethics in Medical Laboratories - PCLPSHE, Quality, and Ethics in Medical Laboratories - PCLP
SHE, Quality, and Ethics in Medical Laboratories - PCLP
 
Bioequivalence of Highly Variable Drug Products
Bioequivalence of Highly Variable Drug ProductsBioequivalence of Highly Variable Drug Products
Bioequivalence of Highly Variable Drug Products
 
Basics of laboratory internal quality control, Ola Elgaddar, 2012
Basics of laboratory internal quality control, Ola Elgaddar, 2012Basics of laboratory internal quality control, Ola Elgaddar, 2012
Basics of laboratory internal quality control, Ola Elgaddar, 2012
 
Bioequivalence of Highly Variable Drug Products
Bioequivalence of Highly Variable Drug ProductsBioequivalence of Highly Variable Drug Products
Bioequivalence of Highly Variable Drug Products
 
Quality control.pptx
Quality control.pptxQuality control.pptx
Quality control.pptx
 
Variability of clinical chemistry laboratory results
Variability of clinical  chemistry laboratory resultsVariability of clinical  chemistry laboratory results
Variability of clinical chemistry laboratory results
 
quality control in clinical laboratory
quality control in clinical laboratory quality control in clinical laboratory
quality control in clinical laboratory
 
Cliniudorohdkdihkshckshsbdjhjdidgjdisybsjdoh
CliniudorohdkdihkshckshsbdjhjdidgjdisybsjdohCliniudorohdkdihkshckshsbdjhjdidgjdisybsjdoh
Cliniudorohdkdihkshckshsbdjhjdidgjdisybsjdoh
 
Quality management system BY M .REJA
Quality management system BY M .REJAQuality management system BY M .REJA
Quality management system BY M .REJA
 
Validity andreliability
Validity andreliabilityValidity andreliability
Validity andreliability
 
Bioanalytical method validation usfda
Bioanalytical method validation usfdaBioanalytical method validation usfda
Bioanalytical method validation usfda
 

More from GH Yeoh

Implementing decision rule made simple
Implementing decision rule made simpleImplementing decision rule made simple
Implementing decision rule made simple
GH Yeoh
 
Common mistakes in measurement uncertainty calculations
Common mistakes in measurement uncertainty calculationsCommon mistakes in measurement uncertainty calculations
Common mistakes in measurement uncertainty calculations
GH Yeoh
 
Worked examples of sampling uncertainty evaluation
Worked examples of sampling uncertainty evaluationWorked examples of sampling uncertainty evaluation
Worked examples of sampling uncertainty evaluation
GH Yeoh
 
Basic calculations of measurement uncertainty in medical testing
Basic calculations of measurement uncertainty in medical testingBasic calculations of measurement uncertainty in medical testing
Basic calculations of measurement uncertainty in medical testing
GH Yeoh
 
A note and graphical illustration of type II error
A note and graphical illustration of type II errorA note and graphical illustration of type II error
A note and graphical illustration of type II error
GH Yeoh
 
Concept of Limit of Detection (LOD)
Concept of Limit of Detection (LOD)Concept of Limit of Detection (LOD)
Concept of Limit of Detection (LOD)
GH Yeoh
 
Ensuring the validity of results
Ensuring the validity of resultsEnsuring the validity of results
Ensuring the validity of results
GH Yeoh
 
Presenting our training courses
Presenting our training coursesPresenting our training courses
Presenting our training courses
GH Yeoh
 

More from GH Yeoh (8)

Implementing decision rule made simple
Implementing decision rule made simpleImplementing decision rule made simple
Implementing decision rule made simple
 
Common mistakes in measurement uncertainty calculations
Common mistakes in measurement uncertainty calculationsCommon mistakes in measurement uncertainty calculations
Common mistakes in measurement uncertainty calculations
 
Worked examples of sampling uncertainty evaluation
Worked examples of sampling uncertainty evaluationWorked examples of sampling uncertainty evaluation
Worked examples of sampling uncertainty evaluation
 
Basic calculations of measurement uncertainty in medical testing
Basic calculations of measurement uncertainty in medical testingBasic calculations of measurement uncertainty in medical testing
Basic calculations of measurement uncertainty in medical testing
 
A note and graphical illustration of type II error
A note and graphical illustration of type II errorA note and graphical illustration of type II error
A note and graphical illustration of type II error
 
Concept of Limit of Detection (LOD)
Concept of Limit of Detection (LOD)Concept of Limit of Detection (LOD)
Concept of Limit of Detection (LOD)
 
Ensuring the validity of results
Ensuring the validity of resultsEnsuring the validity of results
Ensuring the validity of results
 
Presenting our training courses
Presenting our training coursesPresenting our training courses
Presenting our training courses
 

Recently uploaded

THEMATIC APPERCEPTION TEST(TAT) cognitive abilities, creativity, and critic...
THEMATIC  APPERCEPTION  TEST(TAT) cognitive abilities, creativity, and critic...THEMATIC  APPERCEPTION  TEST(TAT) cognitive abilities, creativity, and critic...
THEMATIC APPERCEPTION TEST(TAT) cognitive abilities, creativity, and critic...
Abdul Wali Khan University Mardan,kP,Pakistan
 
The binding of cosmological structures by massless topological defects
The binding of cosmological structures by massless topological defectsThe binding of cosmological structures by massless topological defects
The binding of cosmological structures by massless topological defects
Sérgio Sacani
 
Compexometric titration/Chelatorphy titration/chelating titration
Compexometric titration/Chelatorphy titration/chelating titrationCompexometric titration/Chelatorphy titration/chelating titration
Compexometric titration/Chelatorphy titration/chelating titration
Vandana Devesh Sharma
 
Remote Sensing and Computational, Evolutionary, Supercomputing, and Intellige...
Remote Sensing and Computational, Evolutionary, Supercomputing, and Intellige...Remote Sensing and Computational, Evolutionary, Supercomputing, and Intellige...
Remote Sensing and Computational, Evolutionary, Supercomputing, and Intellige...
University of Maribor
 
Describing and Interpreting an Immersive Learning Case with the Immersion Cub...
Describing and Interpreting an Immersive Learning Case with the Immersion Cub...Describing and Interpreting an Immersive Learning Case with the Immersion Cub...
Describing and Interpreting an Immersive Learning Case with the Immersion Cub...
Leonel Morgado
 
如何办理(uvic毕业证书)维多利亚大学毕业证本科学位证书原版一模一样
如何办理(uvic毕业证书)维多利亚大学毕业证本科学位证书原版一模一样如何办理(uvic毕业证书)维多利亚大学毕业证本科学位证书原版一模一样
如何办理(uvic毕业证书)维多利亚大学毕业证本科学位证书原版一模一样
yqqaatn0
 
NuGOweek 2024 Ghent programme overview flyer
NuGOweek 2024 Ghent programme overview flyerNuGOweek 2024 Ghent programme overview flyer
NuGOweek 2024 Ghent programme overview flyer
pablovgd
 
Micronuclei test.M.sc.zoology.fisheries.
Micronuclei test.M.sc.zoology.fisheries.Micronuclei test.M.sc.zoology.fisheries.
Micronuclei test.M.sc.zoology.fisheries.
Aditi Bajpai
 
Basics of crystallography, crystal systems, classes and different forms
Basics of crystallography, crystal systems, classes and different formsBasics of crystallography, crystal systems, classes and different forms
Basics of crystallography, crystal systems, classes and different forms
MaheshaNanjegowda
 
Eukaryotic Transcription Presentation.pptx
Eukaryotic Transcription Presentation.pptxEukaryotic Transcription Presentation.pptx
Eukaryotic Transcription Presentation.pptx
RitabrataSarkar3
 
3D Hybrid PIC simulation of the plasma expansion (ISSS-14)
3D Hybrid PIC simulation of the plasma expansion (ISSS-14)3D Hybrid PIC simulation of the plasma expansion (ISSS-14)
3D Hybrid PIC simulation of the plasma expansion (ISSS-14)
David Osipyan
 
Bob Reedy - Nitrate in Texas Groundwater.pdf
Bob Reedy - Nitrate in Texas Groundwater.pdfBob Reedy - Nitrate in Texas Groundwater.pdf
Bob Reedy - Nitrate in Texas Groundwater.pdf
Texas Alliance of Groundwater Districts
 
Applied Science: Thermodynamics, Laws & Methodology.pdf
Applied Science: Thermodynamics, Laws & Methodology.pdfApplied Science: Thermodynamics, Laws & Methodology.pdf
Applied Science: Thermodynamics, Laws & Methodology.pdf
University of Hertfordshire
 
Shallowest Oil Discovery of Turkiye.pptx
Shallowest Oil Discovery of Turkiye.pptxShallowest Oil Discovery of Turkiye.pptx
Shallowest Oil Discovery of Turkiye.pptx
Gokturk Mehmet Dilci
 
The use of Nauplii and metanauplii artemia in aquaculture (brine shrimp).pptx
The use of Nauplii and metanauplii artemia in aquaculture (brine shrimp).pptxThe use of Nauplii and metanauplii artemia in aquaculture (brine shrimp).pptx
The use of Nauplii and metanauplii artemia in aquaculture (brine shrimp).pptx
MAGOTI ERNEST
 
waterlessdyeingtechnolgyusing carbon dioxide chemicalspdf
waterlessdyeingtechnolgyusing carbon dioxide chemicalspdfwaterlessdyeingtechnolgyusing carbon dioxide chemicalspdf
waterlessdyeingtechnolgyusing carbon dioxide chemicalspdf
LengamoLAppostilic
 
Authoring a personal GPT for your research and practice: How we created the Q...
Authoring a personal GPT for your research and practice: How we created the Q...Authoring a personal GPT for your research and practice: How we created the Q...
Authoring a personal GPT for your research and practice: How we created the Q...
Leonel Morgado
 
EWOCS-I: The catalog of X-ray sources in Westerlund 1 from the Extended Weste...
EWOCS-I: The catalog of X-ray sources in Westerlund 1 from the Extended Weste...EWOCS-I: The catalog of X-ray sources in Westerlund 1 from the Extended Weste...
EWOCS-I: The catalog of X-ray sources in Westerlund 1 from the Extended Weste...
Sérgio Sacani
 
在线办理(salfor毕业证书)索尔福德大学毕业证毕业完成信一模一样
在线办理(salfor毕业证书)索尔福德大学毕业证毕业完成信一模一样在线办理(salfor毕业证书)索尔福德大学毕业证毕业完成信一模一样
在线办理(salfor毕业证书)索尔福德大学毕业证毕业完成信一模一样
vluwdy49
 
8.Isolation of pure cultures and preservation of cultures.pdf
8.Isolation of pure cultures and preservation of cultures.pdf8.Isolation of pure cultures and preservation of cultures.pdf
8.Isolation of pure cultures and preservation of cultures.pdf
by6843629
 

Recently uploaded (20)

THEMATIC APPERCEPTION TEST(TAT) cognitive abilities, creativity, and critic...
THEMATIC  APPERCEPTION  TEST(TAT) cognitive abilities, creativity, and critic...THEMATIC  APPERCEPTION  TEST(TAT) cognitive abilities, creativity, and critic...
THEMATIC APPERCEPTION TEST(TAT) cognitive abilities, creativity, and critic...
 
The binding of cosmological structures by massless topological defects
The binding of cosmological structures by massless topological defectsThe binding of cosmological structures by massless topological defects
The binding of cosmological structures by massless topological defects
 
Compexometric titration/Chelatorphy titration/chelating titration
Compexometric titration/Chelatorphy titration/chelating titrationCompexometric titration/Chelatorphy titration/chelating titration
Compexometric titration/Chelatorphy titration/chelating titration
 
Remote Sensing and Computational, Evolutionary, Supercomputing, and Intellige...
Remote Sensing and Computational, Evolutionary, Supercomputing, and Intellige...Remote Sensing and Computational, Evolutionary, Supercomputing, and Intellige...
Remote Sensing and Computational, Evolutionary, Supercomputing, and Intellige...
 
Describing and Interpreting an Immersive Learning Case with the Immersion Cub...
Describing and Interpreting an Immersive Learning Case with the Immersion Cub...Describing and Interpreting an Immersive Learning Case with the Immersion Cub...
Describing and Interpreting an Immersive Learning Case with the Immersion Cub...
 
如何办理(uvic毕业证书)维多利亚大学毕业证本科学位证书原版一模一样
如何办理(uvic毕业证书)维多利亚大学毕业证本科学位证书原版一模一样如何办理(uvic毕业证书)维多利亚大学毕业证本科学位证书原版一模一样
如何办理(uvic毕业证书)维多利亚大学毕业证本科学位证书原版一模一样
 
NuGOweek 2024 Ghent programme overview flyer
NuGOweek 2024 Ghent programme overview flyerNuGOweek 2024 Ghent programme overview flyer
NuGOweek 2024 Ghent programme overview flyer
 
Micronuclei test.M.sc.zoology.fisheries.
Micronuclei test.M.sc.zoology.fisheries.Micronuclei test.M.sc.zoology.fisheries.
Micronuclei test.M.sc.zoology.fisheries.
 
Basics of crystallography, crystal systems, classes and different forms
Basics of crystallography, crystal systems, classes and different formsBasics of crystallography, crystal systems, classes and different forms
Basics of crystallography, crystal systems, classes and different forms
 
Eukaryotic Transcription Presentation.pptx
Eukaryotic Transcription Presentation.pptxEukaryotic Transcription Presentation.pptx
Eukaryotic Transcription Presentation.pptx
 
3D Hybrid PIC simulation of the plasma expansion (ISSS-14)
3D Hybrid PIC simulation of the plasma expansion (ISSS-14)3D Hybrid PIC simulation of the plasma expansion (ISSS-14)
3D Hybrid PIC simulation of the plasma expansion (ISSS-14)
 
Bob Reedy - Nitrate in Texas Groundwater.pdf
Bob Reedy - Nitrate in Texas Groundwater.pdfBob Reedy - Nitrate in Texas Groundwater.pdf
Bob Reedy - Nitrate in Texas Groundwater.pdf
 
Applied Science: Thermodynamics, Laws & Methodology.pdf
Applied Science: Thermodynamics, Laws & Methodology.pdfApplied Science: Thermodynamics, Laws & Methodology.pdf
Applied Science: Thermodynamics, Laws & Methodology.pdf
 
Shallowest Oil Discovery of Turkiye.pptx
Shallowest Oil Discovery of Turkiye.pptxShallowest Oil Discovery of Turkiye.pptx
Shallowest Oil Discovery of Turkiye.pptx
 
The use of Nauplii and metanauplii artemia in aquaculture (brine shrimp).pptx
The use of Nauplii and metanauplii artemia in aquaculture (brine shrimp).pptxThe use of Nauplii and metanauplii artemia in aquaculture (brine shrimp).pptx
The use of Nauplii and metanauplii artemia in aquaculture (brine shrimp).pptx
 
waterlessdyeingtechnolgyusing carbon dioxide chemicalspdf
waterlessdyeingtechnolgyusing carbon dioxide chemicalspdfwaterlessdyeingtechnolgyusing carbon dioxide chemicalspdf
waterlessdyeingtechnolgyusing carbon dioxide chemicalspdf
 
Authoring a personal GPT for your research and practice: How we created the Q...
Authoring a personal GPT for your research and practice: How we created the Q...Authoring a personal GPT for your research and practice: How we created the Q...
Authoring a personal GPT for your research and practice: How we created the Q...
 
EWOCS-I: The catalog of X-ray sources in Westerlund 1 from the Extended Weste...
EWOCS-I: The catalog of X-ray sources in Westerlund 1 from the Extended Weste...EWOCS-I: The catalog of X-ray sources in Westerlund 1 from the Extended Weste...
EWOCS-I: The catalog of X-ray sources in Westerlund 1 from the Extended Weste...
 
在线办理(salfor毕业证书)索尔福德大学毕业证毕业完成信一模一样
在线办理(salfor毕业证书)索尔福德大学毕业证毕业完成信一模一样在线办理(salfor毕业证书)索尔福德大学毕业证毕业完成信一模一样
在线办理(salfor毕业证书)索尔福德大学毕业证毕业完成信一模一样
 
8.Isolation of pure cultures and preservation of cultures.pdf
8.Isolation of pure cultures and preservation of cultures.pdf8.Isolation of pure cultures and preservation of cultures.pdf
8.Isolation of pure cultures and preservation of cultures.pdf
 

Biological variation as an uncertainty component

  • 1. Use of Biological Variation as an Uncertainty Component Evaluation Methods of Uncertainty of Measurement for Medical Testing https://consultglp.com by Yeoh Guan Huah GLP Consulting, Singapore
  • 2. Outline • A review of the uncertainty of measurement (UM) calculation methods by component-by- component approach • Necessity to consider non-analytical uncertainty components in addition to the analytical uncertainty in the laboratory • Biological variation – what and why • How to incorporate biological variation in the overall evaluation of UM.
  • 3. Introduction • The ultimate use of analysis data with its associated uncertainty of measurement (UM) must be fit for intended purpose • Data generated by medical laboratory are used for any signaling which may differentiate health from disease, based on some pre-set reference ranges for diagnostic decision • Amongst other applications, these data are also adopted for clinical monitoring of a patient upon treatment.
  • 4. Introduction • Both ISO 15189 and ISO/IEC 17025 accreditation standards have given uncertainty of measurement a new perspective. • They have recognized that apart from the analytical uncertainty during the testing process, there are other non-analytical uncertainty components to be considered as well. • For industrial, environmental and commodity testing, ISO/IEC 17025 stresses the importance of taking representative samples from the lot (population) for analysis. Hence, sampling uncertainty is incorporated as another uncertainty contributor, if applicable. • For clinical application of a test result, the in vivo biological variability of the measurand as an uncertainty component cannot be ignored for clinical interpretation.
  • 5. Reported result with uncertainty : X + U (units) • Measurement uncertainty U is expressed as: • Test result : X + U • and, U = k x u • where k is a coverage factor (usually k = 2 or 1.96 for 95% confidence), and • u, the combined standard uncertainty expressed as standard deviation • U is also called expanded uncertainty, assumed rightly a normal probability distribution of data 5 Example: Consider an expression for a measurand (urea in serum): “5.0 + 0.2 mmol/L with 95% confidence” The + 0.2 mmol/L is the uncertainty ( U ) of the supposed 5.0 mmol/L serum urea measurement. Its combined standard uncertainty (u ) therefore, is 0.2/2 or 0.1 mmol/L
  • 6. Basic methods for combining uncertainty components • Calculated from a sum and/or a difference of independent measurements with associated standard uncertainties • Let: R = X + Y or R = X – Y • Then, (uR)2 = (uX)2 + (uY)2 • where uR , uX and uY are the respective analytical standard uncertainty (as standard deviation) • Therefore, combined standard uncertainty uR = (uR)2 • Note: X and Y are of same units and involved in addition & subtraction conditions
  • 7. Worked example No. 1 • Anion gap (AG) is calculated based on the results of serum (plasma) ions: sodium, potassium, chloride and bicarbonate, in the form: • AG = (Na+ + K+) – (Cl- + HCO3-) Ion, mmol/L Test result Standard deviation, u Na+ 140 1.2 K+ 3.9 0.1 Cl- 102 1.3 HCO3- 24 1.2 Therefore, AG = (140+3.9) – (102 + 24) = 17.9 mmol/L 𝐶𝑜𝑚𝑏𝑖𝑛𝑒𝑑 𝑆𝐷 𝑅= (1.2)2+(0.1)2+(1.3)2+(1.2)2 = 2.1 mmol/L Report: Anion Gap (AG) = 18 + 4 mmol/L with a coverage factor of 2 (95% confidence)
  • 8. Basic methods for combining uncertainty components • From multiplication and/or division of independent measurements • Using the Propagation Law of Standard Deviation, • Let: 𝑢(𝑦)2 = 𝑖=1 𝑛 𝜕𝑓 𝜕𝑥𝑖 2 𝑢(𝑥𝑖)2 • R = 𝑃×𝑉 𝑇 with standard deviations uP , uV , uT , then, • 𝑐 𝑃 = 𝜕𝑅 𝜕𝑃 = 𝑉 𝑇 • 𝑐 𝑉 = 𝜕𝑅 𝜕𝑉 = 𝑃 𝑇 • 𝑐 𝑇 = 𝜕𝑅 𝜕𝑇 = − 𝑃×𝑉 𝑇2 𝑢 𝑅 2 = 𝑐 𝑃 2 𝑢 𝑃 2 + 𝑐 𝑉 2 𝑢 𝑉 2 + 𝑐 𝑇 2 𝑢 𝑇 2 = ( 𝑉 𝑇 )2 𝑢 𝑃 2 + ( 𝑃 𝑇 )2 𝑢 𝑉 2 + (− 𝑃𝑉 𝑇2)2 𝑢 𝑇 2 Dividing both sides by 𝑅2 = ( 𝑃×𝑉 𝑇 )2 gives: ( 𝑢 𝑅 𝑅 )2 = ( 𝑢 𝑃 𝑃 )2 + ( 𝑢 𝑉 𝑉 )2 + ( 𝑢 𝑇 𝑇 )2 , or 𝑢 𝑅 𝑅 = ( 𝑢 𝑃 𝑃 )2+ ( 𝑢 𝑉 𝑉 )2 + ( 𝑢 𝑇 𝑇 )2 𝒊. 𝒆. , 𝑪𝑽 𝑹 = 𝑪𝑽 𝑷 𝟐 + 𝑪𝑽 𝑽 𝟐 + 𝑪𝑽 𝑻 𝟐
  • 9. Worked example No. 2 • In a creatinine clearance test, the following calculation applies: • C = (U x V ) / (P x T ) Parameter Unit Value u U mmol/L 10.0 0.25 V ml 1500 15 P mmol/L 0.1 0.01 T secs 86400 0 By calculation, C = (10.0 mmol/L x 1500 ml)/(0.1 mmol/L x 24 hrs x 3600 sec) = 1.74 ml/sec and, ( 𝑢 𝑅 1.74 )2 =( 0.25 𝑈10.0 )2 +( 15 1500 )2 +( 0.01 0.1 )2 +( 0 86400 )2 = 0.0107 Therefore, uR = 1.74 x 0.0107 = 0.18 ml/sec Report: Creatinine clearance C = 1.74 + 0.36 ml/sec with the coverage Factor of 2 for 95% confidence
  • 10. Non-analytical uncertainty contributors • In medical testing, there are many potential uncertainty contributors which can significantly affect test results, apart from the analytical process itself, such as: • Improper practice during specimen collection • Poor specimen handling during transit to the laboratory • Patient related factors like biological variability and the presence of any drug residues
  • 11. Analytical uncertainty component • The laboratory’s analytical quality comes from two factors for total analytical error as its analytical goal: • Intermediate precision (random error; imprecision) • Analytical bias (systematic error) • Intermediate precision or imprecision (ua) is estimated from its long-term combined (pooled) standard deviation after a series of analysis in replicates by different analysts within the laboratory on different occasions on a stable internal quality control serum sample. • Analytical bias is evaluated by studying any significant difference between the average test result against the target or certified value of the internal QC sample through the Student’s t-test. If there is a significance difference between the mean result and the certified value, then the standard uncertainty of bias (ub) is estimated.
  • 12. Why do we consider the biological variation? • There are many clinical uses of the test results reported. • As a test result must be ‘fit for purpose’, there is a need to assess its clinical goal in addition to the analytical goal. • For some measurands, an analytical goal may not be physiologically or clinically relevant. Then, the method imprecision study would be sufficient. • However most medical laboratory need to identify if UM information when reported could significantly affect clinical interpretations and patient management. • Hence, there is a necessity to have additional non-analytical uncertainty component in the overall estimation of UM to set such goal. • This additional uncertainty component is the intra-individual biological variation of the measurand.
  • 13. Components of biological variation (BV) • Actually, there are two types of components of biological variation, namely: • Intra-individual (random fluctuation of analytes around the setting point of each individual), sometimes termed as “within- subject” BV; • Inter-individual (overall variation from the different person’s setting point), sometimes termed as “between-subject” BV. • The laboratory’s analytical method, instrument and reagents do not make difference in BV estimates.
  • 14. Some applications of biological variation  Evaluating the clinical significance of changes in consecutive test results from an individual in monitoring treatment progress  Setting quality specifications for analytical performance (usually targeting the analytical goal as a percentage of BV) to satisfy the general needs of diagnosis and monitoring  Validating new procedures in a laboratory  Assessing the usefulness of population-based reference values  Determining which type of specimen (e.g. plasma, serum, 24-hr urine, first-morning urine) is optimal for analyzing a specific constituent
  • 15. Relationship of analytical imprecision and biological variation • There are 3 levels of analytical goal for long-term imprecision (CVa) based on intra-individual biological variation (CVi): • Optimum: CVa < 0.25 x CVi • Desirable: CVa < 0.50 x CVi • Minimum: CVa < 0.75 x CVi • For analytes with unavailable CVi data, other criteria may be used, such as relative performance in inter-laboratory comparison studies or proficiency testing programs, certified reference interval, professional clinical opinion, etc.
  • 16. Relationship of analytical imprecision and biological variation • If test results are interpreted using reference or clinical decision values determined by a different method, bias should be considered as a component of uncertainty to be estimated. • The inter-individual biological variation CVg is to be incorporated. • Like imprecision, we have to ensure the three levels of analytical goal for bias (CVb) based on biological variation are met: • Optimum: • Desirable: • Minimum: 𝐶𝑉𝑏 ≤ 0.125 × 𝐶𝑉𝑖 2 + 𝐶𝑉𝑔 2 𝐶𝑉𝑏 ≤ 0.250 × 𝐶𝑉𝑖 2 + 𝐶𝑉𝑔 2 𝐶𝑉𝑏 ≤ 0.375 × 𝐶𝑉𝑖 2 + 𝐶𝑉𝑔 2
  • 17. Biological variation database • There are study findings which indicate that the factors inherent to the subject, such as sex, age, race, geographical location, do not produce changes in BV results • Practically all the currently available estimates of biological variation (more than 300 analytes) have been compiled in a database: https://www.westgard.com/biodatabase1.htm • The intra-individual (or within-subject) coefficients of variation (in %) from the database can be used to evaluate the clinical significance of changes in two consecutive results from a patient, in addition to the analytical variation.
  • 18.
  • 19.
  • 20. Summation of analytical uncertainty & biological variation • If the combined estimate of coefficient of variation of reported test result (CVT) covers both the CV of analytical imprecision (CVa) and that of within-subject biological variation (CVi) for a comparison of two consecutive test results from a patient, we then have: • CVT 2 = CVa 2 + CVi 2 • Note: If test results are clinically interpreted by comparison with reference or previous values produced by the same analytical method, analytical bias component is not required.
  • 21. Worked example No. 3 • Plasma alkaline phosphatase (ALP) activity for a patient: • 1st day’s result : 95 U/L; 3rd day’s result : 108 U/L • UM for ALP : CVa = 1.45% at QC sample mean : 87 U/L • ALP intro-individual BV (given by Westgard website): CVi = 6.45% • Were these two sets of daily test results of any significant different? • Sum of analytical & BV as CV’s: • CVT = (CVa 2 + CVi 2) = (1.452 + 6.452) = 6.61%
  • 22. Worked example No. 3 (contd.) • If the two test results are analytically and biologically different, the “critical difference” or Reference Change Value (RCV) is to be: • 1.96 x 2 x (CVa 2 + CVi 2) or 2.77 x (CVa 2 + CVi 2) for 95% confidence • That means the two results need to differ by 2.77 x 6.61% = 18.3% • Now, the first day’s result was 95 U/L, thus • 95 U/L + (95 U/L x 18.3%) = 95 + 17.4 = 112 U/L • and, 95 U/L - (95 U/L x 18.3%) = 95 - 17.4 = 77.6 U/L • So, the critical range of expanded uncertainty is [77.6 U/L, 112 U/L] • The second result would have to be at least 112 U/L for there to be 95% confidence that it was both analytically and biologically different from the first result. We can therefore conclude that the test results of 95 U/L and 108 U/L are analytically different, but probably not biologically different.
  • 23. A note on Reference Change Value RCV • Some references recommend to use a one-tail test with a probability risk (α) of 5% error, instead of a two-tail test. • In that case, the Reference Change Value has a formula: • 1.645 x 2 x (CVa 2 + CVi 2) or 2.33 x (CVa 2 + CVi 2) • By using this new formula, the two results need to differ by 2.33 x 6.61% = 15.4% • Now, the first day’s result was 95 U/L, thus the upper critical limit is: • 95 U/L + (95 U/L x 15.4%) = 95 + 14.7 = 109.7 U/L • Same conclusion could be made based on the new RCV. ααα/2α/2 The choice of 1-tail or 2- tail probability depends on the alternative hypothesis H1
  • 24. How do we get the factor 1.96 x 2 or 1.645 x 2? • A measurement result, X, is associated with a standard deviation +sX • The duplicate result, Y, on the same sample is also associated with the same standard deviation +sX • When an average value is calculated for X and Y, we have a total variance of • sX 2 + sX 2 = 2sX 2 • Therefore, the pooled standard deviation is 2 x sX • The choice of coverage factor k = 1.96 or 1.645 depend on either 2-tail or 1-tail hypothesis testing, assuming a normal distribution with an error probability α = 0.05 or 95% confidence.
  • 25. Conclusion • It is a good practice to maintain a list of measurands with their respective RCV (reference change value) to define healthy and non-healthy situations. • Remember that the RCV is applied for the target analyte in each pair of consecutive results from the same patient • The laboratory information system (LIMS) can be set to flag off the second report with a predefined signal after the numeric result of the target analyte is keyed in. • Most important challenge for the medical personnel is how to explain the UM concept to requesting doctors in a clinically meaningful way.
  • 26. References • “Uncertainty of Measurement in Quantitative Medical Testing – A Laboratory Implementation Guide” by Australasian Association of Clinical Biochemists, November 2004 • “Application of Biological Variation – A Review” by Ricos C. et al, Biochemia Medica 2009, 19(3):250-9 • “Desirable Biological Variation Database specifications” by James Westgard, https://www.westgard.com/biodatabase1.htm