EMG involves detecting and recording electrical signals from muscle contractions. A successful EMG requires knowledge of anatomy, physiology, and technique. The equipment includes an EMG machine, needle, cables, and electrodes. Either concentric or monopolar needles can be used. A typical EMG examines insertional activity, spontaneous activity at rest, and motor unit action potentials. Abnormal spontaneous activities include fibrillation potentials, positive sharp waves, complex repetitive discharges, and myotonic discharges. Motor unit analysis assesses morphology, stability, and firing characteristics to determine neuropathic or myopathic disorders.
Basic MEP techniques and understanding for Intraoperative neuromonitoring of the motors tracts during Brain and Spinal surgeries to prevent postoperative complications.
Basic MEP techniques and understanding for Intraoperative neuromonitoring of the motors tracts during Brain and Spinal surgeries to prevent postoperative complications.
Late response are the most helpful findings in some of the diseases affecting the peripheral nerves, (e.g GBS, Radiculopathies, ). How to assess these responses while performing Nerve Conduction Studies, is the most technical and theoretical consideration.... Here we go with the same things in the stated slides
what is RNS and what the techniques to perform this test in the lab. Its significance in the evaluation and diagnosis of NMJ disorders like MG, LEMBS etc..
This presentation is an introduction to the principles of Nerve Conduction Study and is entirely sourced from the book by David C Preston and Barbara E Shapiro: Electromyography and Neuromuscular disorders, 3rd Edition
Anomalous Innervations in (EMG/NCS) by MurtazaMurtaza Syed
Anomalous Innervation.
These are the sort of normal variants which can be found in any normal subject or can concomitantly be found or superimposed in pathological cases. Identifying these anomalies helps out interpreting and making correct diagnosis and to avoid any misinterpretation.
This presentation describes the common conditions, anatomy and the ideal ways to do and perform nerve conduction studies in lower limbs. It is nicely depicted with self explanatory pictures.
Late response are the most helpful findings in some of the diseases affecting the peripheral nerves, (e.g GBS, Radiculopathies, ). How to assess these responses while performing Nerve Conduction Studies, is the most technical and theoretical consideration.... Here we go with the same things in the stated slides
what is RNS and what the techniques to perform this test in the lab. Its significance in the evaluation and diagnosis of NMJ disorders like MG, LEMBS etc..
This presentation is an introduction to the principles of Nerve Conduction Study and is entirely sourced from the book by David C Preston and Barbara E Shapiro: Electromyography and Neuromuscular disorders, 3rd Edition
Anomalous Innervations in (EMG/NCS) by MurtazaMurtaza Syed
Anomalous Innervation.
These are the sort of normal variants which can be found in any normal subject or can concomitantly be found or superimposed in pathological cases. Identifying these anomalies helps out interpreting and making correct diagnosis and to avoid any misinterpretation.
This presentation describes the common conditions, anatomy and the ideal ways to do and perform nerve conduction studies in lower limbs. It is nicely depicted with self explanatory pictures.
Basics of electro myo graphy study (EMG)Joe Antony
Electromyography (EMG) is a vital technique in the field of bioelectrical signal analysis. It involves capturing muscle activity through surface or needle electrodes for diagnostic purposes. EMG signals can be analyzed to detect various muscle conditions, such as myopathic or neuropathic lesions, using numerical parameters. The spatial frequency bandwidth of surface EMG signals is crucial for detailed muscle activity reconstruction, with appropriate inter-electrode distances being essential for accurate mapping. In the context of neuro-monitoring, EMG plays a role in intra-operative detection of adverse events and predicting postoperative outcomes, especially when used complementarily with other modalities like motor evoked potentials. Overall, EMG serves as a valuable tool for understanding muscle function, diagnosing muscle disorders, and enhancing neuro-monitoring practices
Nerve conduction studies (NCSs) have become a simple and reliable test of peripheral nerve function.
With adequate standardization, the method now provides a means of not only objectively identifying the lesion but also precisely localizing the site of maximal involvement.
Electrical stimulation of the nerve initiates an impulse that travels along motor or sensory nerve fibers.
The assessment of conduction characteristics depends on the analysis of compound evoked potentials recorded from the muscle in the study of motor fibers and from the nerve itself in the case of sensory fiber
ELECTRICAL STIMULation of the nerve CATHODE AND ANODE: Surface electrodes, usually made of silver plate, come in different sizes, commonly in the range of 0.5 to 1.0 cm in diameter.
Stimulating electrodes consist of a cathode, or negative pole, and an anode, or positive pole, so called because they attract cations and anions.
As the current flows between them, negative charges that accumulate under the cathode, by making inside the axon relatively more positive than outside, depolarize the nerve or cathodal depolarization.
Conversely, positive charges under the anode hyperpolarize the nerve
TYPES OF STIMULATOR
Most commercially available stimulators provide a probe that mounts the cathode and the anode at a fixed distance, usually 2 to 3 cm apart.
The intensity control located in the insulated handle, though bulky, simplifies the operation for a single examiner.
The ordinary banana plugs connected by shielded cable also serve well as stimulating electrodes.
The use of a large diameter electrode for stimulation lowers current density in the skin, causing less pain, although the exact site of nerve activation becomes uncertain.
A monopolar stimulation with a small cathode placed on the nerve trunk and a large anode over the opposite surface in the same limb.
The use of a subcutaneously inserted needle as the cathode reduces the current necessary to excite the nerve compared to surface stimulation.
A surface electrode located on the skin nearby or a second needle electrode inserted in the vicinity of the cathode serves as the anode.
The maximum current during such stimulation causes neither electric nor heat damage to the tissue.
RECORDING OF MUSCLE AND NERVE POTENTIAL
Surface and Needle Electrodes:
Surface electrodes with a larger recording radius serve better than needle electrodes in assessing a compound muscle action potential (CMAP).
Its onset latency indicates the conduction time of the fastest motor fibers, and amplitude, the number of available motor axons.
A needle electrode, despite its small recording radius has its place in identifying the activity from a small muscle when surface recording fails.
Its use also improves segregation of a target activity from neighboring discharges after proximal stimulation, which tends to excite unwanted neighboring nerves simultaneously
Amplifier system
The electrodes convert bioelectric signal resulting from muscle or
Neuromuscular monitoring, also known as train of four monitoring, is a technique used during recovery from the application of general anesthesia to objectively determine how well a patient's muscles are able to function. It involves the application of electrical stimulation to nerves and recording of muscle response using, for example, an acceleromyograph. Neuromuscular monitoring is typically used when neuromuscular-blocking drugs have been part of the general anesthesia and the doctor wishes to avoid postoperative residual curarization (PORC) in the patient, that is, the residual paralysis of muscles stemming from these drugs.
Normal Labour/ Stages of Labour/ Mechanism of LabourWasim Ak
Normal labor is also termed spontaneous labor, defined as the natural physiological process through which the fetus, placenta, and membranes are expelled from the uterus through the birth canal at term (37 to 42 weeks
2024.06.01 Introducing a competency framework for languag learning materials ...Sandy Millin
http://sandymillin.wordpress.com/iateflwebinar2024
Published classroom materials form the basis of syllabuses, drive teacher professional development, and have a potentially huge influence on learners, teachers and education systems. All teachers also create their own materials, whether a few sentences on a blackboard, a highly-structured fully-realised online course, or anything in between. Despite this, the knowledge and skills needed to create effective language learning materials are rarely part of teacher training, and are mostly learnt by trial and error.
Knowledge and skills frameworks, generally called competency frameworks, for ELT teachers, trainers and managers have existed for a few years now. However, until I created one for my MA dissertation, there wasn’t one drawing together what we need to know and do to be able to effectively produce language learning materials.
This webinar will introduce you to my framework, highlighting the key competencies I identified from my research. It will also show how anybody involved in language teaching (any language, not just English!), teacher training, managing schools or developing language learning materials can benefit from using the framework.
Macroeconomics- Movie Location
This will be used as part of your Personal Professional Portfolio once graded.
Objective:
Prepare a presentation or a paper using research, basic comparative analysis, data organization and application of economic information. You will make an informed assessment of an economic climate outside of the United States to accomplish an entertainment industry objective.
This slide is special for master students (MIBS & MIFB) in UUM. Also useful for readers who are interested in the topic of contemporary Islamic banking.
Model Attribute Check Company Auto PropertyCeline George
In Odoo, the multi-company feature allows you to manage multiple companies within a single Odoo database instance. Each company can have its own configurations while still sharing common resources such as products, customers, and suppliers.
How to Make a Field invisible in Odoo 17Celine George
It is possible to hide or invisible some fields in odoo. Commonly using “invisible” attribute in the field definition to invisible the fields. This slide will show how to make a field invisible in odoo 17.
Exploiting Artificial Intelligence for Empowering Researchers and Faculty, In...Dr. Vinod Kumar Kanvaria
Exploiting Artificial Intelligence for Empowering Researchers and Faculty,
International FDP on Fundamentals of Research in Social Sciences
at Integral University, Lucknow, 06.06.2024
By Dr. Vinod Kumar Kanvaria
2. INTRODUCTION
• EMG is the detection, amplification, recording, processing,
analysis and interpretation of the electrical signal produced by
the contraction of a muscle of the electrical signal produced by
the contraction of a muscle
• No doubt that the needle EMG is the more challenging part of
the electrophysiologic examination.
• A successful study requires not only knowledge of anatomy and
physiology but also sound EMG technique and good patient
rapport.
3. EQUIPMENT
• In addition to the EMG machine, an EMG needle, needle cable, ground electrode, and
gloves are necessary to perform the needle EMG study.
• Either a concentric or monopolar EMG needle can be used. voltage is measured as the
difference between the active and reference recording electrodes.
• Concentric needle contains both the active and reference electrodes in the needle itself.
Shaft-reference electrode, whereas the active electrode runs as a very small wire
through the center of the needle and is exposed at the needle tip. The end of the
concentric needle is beveled, resulting in a recording area that has a “teardrop”
configuration.
• Monopolar needle is Teflon coated, and its exposed end serves as the active recording
electrode.An additional surface disc electrode is required as the reference electrode.
6. • With a concentric needle, the MUAP amplitude is slightly
smaller and the major spike rise time is shorter than those
obtained with a monopolar needle (likely a reflection of the
size and shape of the recording field of each needle).
• Disadvantage of monopolar- additional refrence electrode
needed, greater likelihood of electrode impedance mismatch
and increased electrical noise.
7. PATIENT PREPARATION
• Before beginning the needle EMG examination, it is
important to explain the procedure to the patient and allay
any patient fears.
• Afterward, the examiner should answer any questions from
the patient before proceeding to the needle EMG
examination. Good patient rapport both before and during
the study is essential.
9. • Once the correct needle placement has been established,
the first part of the examination is to assess insertional and
spontaneous activity at rest.
• sweep speed- 10 ms per division.
• sensitivity - 50 μV per division.
• Five to 10 brief insertions should be performed.
10. INSERTIONAL ACTIVITY
• When the needle is quickly moved
through muscle, there is a brief burst of
muscle fiber potentials, known as
insertional activity, which typically lasts
no longer than 300 ms after the needle
has stopped moving.
• Increased insertional activity is
defined as any activity other than
endplate potentials that last longer
than 300 ms after brief needle
movement,found in both neuropathic
and myopathic condition.
14. SPONTANEOUS ACTIVITY: NORMAL
• Endplate Noise- low-
amplitude, monophasic
negative potentials that
fire irregularly at 20–40
Hz and have a
characteristic “seashell”
sound on EMG.
Physiologically, they
represent MEPPs.
frequent association with
endplate spikes.
15. SPONTANEOUS ACTIVITY: NORMAL
• Endplate Spikes (“Nerve
Potentials”)-MFAP,fire
irregularly up to a frequency of
50 Hz.Biphasic, with an initial
negative deflection. They have
a cracking, buzzing, or
sputtering sound on EMG.
• The key features that
differentiate endplate spikes
from fibrillation potentials,
which are also brief spikes, are
their initial negative deflection
and their highly irregular firing
rate
16. SPONTANEOUS ACTIVITY: ABNORMAL MUSCLE
FIBER POTENTIALS
• Fibrillation Potentials- recognized by their single MFAP morphology: a
brief spike with an initial positive deflection, 1–5 ms in duration, and low in
amplitude (typically 10–100 μV).
• The firing pattern is very regular, with a rate usually 0.5–10 Hz, occasionally up
to 30 Hz.On EMG, single fibrillation potentials sound like “rain on the roof.”
• Signifies active denervation.
• typically are associated with neuropathic disorders (i.e., neuropathies,
radiculopathies, motor neuron disease), they also may be seen in some muscle
disorders (especially the inflammatory myopathies and dystrophies).
17.
18. SPONTANEOUS ACTIVITY: ABNORMAL MUSCLE
FIBER POTENTIALS
• Positive Sharp Waves- Positive sharp waves have a brief
initial positivity followed by a long negative phase.They
sound like a dull pop .The amplitude is variable (usually 10–
100 μV, occasionally up to 3 mV).
• Like fibrillation potentials, they have a regular firing
pattern, with a rate usually between 0.5 and 10 Hz,
occasionally up to 30 Hz.
19.
20. Complex Repetitive Discharges
• Result from depolarization of a single muscle fiber followed by ephaptic
spread to adjacent denervated fibers (i.e., direct spread from muscle
membrane to muscle membrane). If the depolarization spreads in a circus
movement whereby the original pacemaker muscle fiber is reactivated, a
recurrent discharge develops.
• On EMG, CRDs are recognized as high-frequency (typically 5–100 Hz),
multiserrated repetitive discharges with an abrupt onset and termination.
These discharges usually occur spontaneously.
• Characteristic machine-like sound on EMG. They occur in both chronic
neuropathic and myopathic disorders; they may arise in any setting in which
denervated muscle fibers lie adjacent to other denervated muscle fibers.
21.
22. Myotonic Discharges
• A myotonic discharge has characteristic waxing and waning of both
amplitude and frequency .The firing rate is generally between 20 and
150 Hz. Either a positive wave or a brief spike morphology.
• Myotonic discharges are characteristically seen in myotonic dystrophy,
myotonia congenita, and paramyotonia congenita. They may also occur
in other myopathies (acid maltase deficiency, polymyositis, myotubular
myopathy), hyperkalemic periodic paralysis, and, rarely, in denervation
of any cause.
• Myotonic discharges have a characteristic “revving engine” sound on
EMG, due to the waxing and waning of amplitude and frequency.
23.
24. SPONTANEOUS ACTIVITY: ABNORMAL MOTOR
UNIT POTENTIALS
• Fasciculation Potentials- single, spontaneous, involuntary discharge of an
individual motor unit. generally fire very slowly and irregularly, usually less than
1–2 Hz.
• The source generator of fasciculation potentials is the motor neuron or its axon,
prior to its terminal branches.
• On EMG, fasciculation potentials usually have the morphology of simple MUAPs,
or they can be complex and large if they represent a pathologic (i.e.,
reinnervated) motor unit.
• Fasciculations are associated with Motor neuron disease, such as amyotrophic
lateral sclerosis, can be seen in radiculopathies, polyneuropathies, and
entrapment neuropathies.
• On EMG, have the sound of “corn popping” in the microwave oven: they are dull,
irregular pops.
25.
26. Doublets, Triplets, and Multiplets
• Spontaneous MUAPs that fire in groups of
two are known as doublets. When they
fire in groups of three or multiple
potentials, they are known as triplets and
multiplets, respectively.
• These potentials have the same
significance as fasciculation potentials:
they represent the spontaneous
depolarization of a motor unit or its axon.
Doublets, triplets, and multiplets can be
seen in any situation where fasciculation
potentials occur (i.e., neuropathic
conditions), but they also are
characteristically seen in tetany from
hypocalcemia.
27. Myokymic Discharges
• Electrically, myokymic discharges are
rhythmic, grouped, spontaneous repetitive
discharges of the same motor unit (i.e.,
grouped fasciculations) .
• The firing frequency within the burst
typically is 5–60 Hz. The number of
potentials within a burst varies widely and
may change from burst to burst .The firing
frequency between bursts is much slower
(typically<2Hz) and produces a marching
sound on EMG. Changing to a longer sweep
speed often makes it easier to recognize the
bursting pattern of a myokymic discharge.
• Clinically, myokymia usually is recognized as
continuous involuntary quivering, rippling,
or undulating movement of muscle.
28.
29. Cramp Potentials
• Clinically, cramps are painful,
involuntary contractions of muscle that
tend to occur when a muscle is in the
shortened position and contracting.
High-frequency discharges of motor
axons and are not primarily a muscle
phenomenon.
• EMG characteristically shows several
normal appearing motor unit potentials
firing repetitively and sometimes
irregularly at high frequencies (usually
40–75 Hz).
• May be associated with a wide range of
neuropathic, endocrinologic, and
metabolic conditions.
30. Neuromyotonic Discharges
• Neuromyotonic discharges are high-frequency (150–250 Hz), decrementing,
repetitive discharges of a single motor unit that have a characteristic
“pinging” sound on EMG.
• They have highest frequency of any discharge. Represent the end of the
spectrum of abnormal spontaneous activity generated in motor nerve.
• Electrically, these syndromes are easily differentiated. Whereas myotonic
syndromes are associated with spontaneous discharges of muscle fibers (with
a positive wave or brief spike morphology), neuromyotonic disorders are
associated with involuntary spontaneous discharges of motor neurons or their
axons (with an MUAP morphology).
31.
32.
33.
34.
35. Analysis of Motor Unit Action Potentials
• MUAPs must be assessed for morphology (duration,
amplitude, phases), stability, and firing characteristics.
• The pattern of MUAP abnormalities allow a determination of
whether a disorder is primarily neuropathic or myopathic
and often helps determine the time course (acute vs.
chronic) and severity of the lesion.
37. MORPHOLOGY
• The analysis of MUAP morphology can be performed on either a
qualitative or a quantitative basis.
• To perform quantitative MUAP analysis, one must isolate 20 different
MUAPs for each muscle being studied and measure their individual
durations, amplitudes, and number of phases. From these values, the
mean duration, amplitude, and number of phases are calculated and
compared with a set of normal values for that particular muscle and age
group.
• MUAP morphology varies depending on the muscle being studied and
the patient’s age.
38. Duration
• Best reflects the number of muscle fibers within a motor unit .Typical
MUAP duration is between 5 and 15 ms.
• It is initial deflection from baseline to the final return of the MUAP to
baseline. It depends primarily on the number of muscle fibers within the
motor unit and the dispersion of their depolarizations over time.
• Duration lengthens as the number of fibers and the territory of a motor
unit increase; it varies directly with age (increased age, increased
duration) and inversely with temperature (decreased temperature,
increased duration) and depends on the individual muscle.
39. • Proximal and bulbofacial
muscles in general have
MUAPs of shorter
duration.
• Duration correlates with
pitch. Long-duration
MUAPs (low frequencies)
sound dull and thuddy,
whereas short-duration
MUAPs (higher
frequencies) sound crisp
and static-like.
40. Polyphasia, Serrations, and Satellite Potentials
• Polyphasia is a measure of synchrony, that is, the extent to which the muscle fibers
within a motor unit fire more or less at the same time.
• May be abnormal in both myopathic and neuropathic disorders. The number of phases
can be easily calculated by counting the number of baseline crossings of the MUAP and
adding one.
• Normally, MUAPs have two to four phases. However, increased polyphasia may be seen
in up to 5%–10% of the MUAPs in any muscle and is considered normal. The one
exception is the deltoid, where up to 25% polyphasia may be normal.
• Increased polyphasia beyond 10% in most muscles and 25% in the deltoid is always
abnormal. Through the speaker, polyphasic MUAPs are recognized as a high-frequency
“clicking” sound
41. Satellite potential
• satellite potentials- After denervation, muscle fibers often are
reinnervated by collateral sprouts from adjacent intact motor units.
• The newly formed sprout often is small, unmyelinated or thinly
myelinated, and therefore very slowly conducting.
• These satellite potentials are extremely unstable and may vary slightly
in their firing rate or may block and not fire at all .
• Over time, the sprout matures, and the thickness of the myelin and
consequently the conduction velocity increase.
42.
43. Amplitude
• MUAP amplitude varies widely among normal subjects. Most MUAPs have an
amplitude greater than 100 μV and less than 2 mV.
• Unlike duration, most muscle fibers of a motor unit contribute little to the
amplitude. MUAP amplitude reflects only those few fibers nearest to the
needle (only 2–12 fibers).
• Several factors are associated with increased amplitude, including
(1) the proximity of the needle to the motor unit,
(2) increased number of muscle fibers in the motor unit,
(3) increased diameter of muscle fibers (i.e., muscle fiber hypertrophy), and
(4) more synchronized firing of the muscle fibers.
Listening to the EMG, the amplitude of MUAPs is correlated not with pitch but
with volume.
44. • Of all MUAP parameters,
amplitude is most
dependent on needle
position. Only muscle
fibers very close to the
needle contribute to
amplitude, as opposed to
duration, wherein most
muscle fibers contribute.
Note change in amplitude
as the needle is moved to
different locations within
the same motor unit.
45. Major Spike
• The major spike is the largest positive-to-negative component of the MUAP and
usually occurs after the first positive peak.
• As the needle is moved closer to the MUAP, the major spike increases in amplitude
and its rise time shortens, indicating the proximity of the needle to the motor unit.
• MUAP parameters should be measured only when the needle is very close to the
motor unit.When the needle is close to the motor unit, the MUAP becomes “sharp.”
• The sharp sound represents the high-frequency component of the major spike,
occurring when the major spike rise time is less than 500 μs, indicating proper
needle placement.
46. Motor unit action potential (MUAP) morphology and
needle electromyography (EMG) position. The position of
the EMG needle influences the morphology of the
recorded MUAP. To properly assess MUAP parameters, the
major spike must be as steep as possible, indicating the
proximity of the needle to the motor unit. Note that
needle electrode position E3 has the shortest major spike
rise time and is the preferable position in which to assess
the MUAP. Also note that although MUAP amplitude
changes markedly with needle position (compare position
E1 with E3), duration is relatively unaffected.
47. STABILITY
• MUAPs usually are stable in morphology from potential to
potential.
• If there is impaired NMJ transmission, unstable MUAPs may
result.
• Any disorder associated with denervation may demonstrate
unstable MUAPs.
• During the process of early reinnervation, newly formed,
immature NMJs often fail to conduct NMJ transmission faithfully.
The result is variability in endplate transmission of transmission
across some of the muscle fibers within a motor unit.
48.
49. FIRING PATTERN (ACTIVATION, RECRUITMENT,
INTERFERENCE PATTERN)
• During muscle contraction, there are only two ways to increase muscle force:
either more motor units can fire, or the motor units which are firing can
increase their firing rate.
• First MUAP increases its firing rate, and then a second MUAP begins to fire,
and so forth.
• Normally, the ratio of firing frequency 5:1. Thus, by the time the first MUAP
firing frequency reaches 10 Hz, a second MUAP should begin to fire.
• During maximal contraction, multiple MUAPs normally overlap and create an
interference pattern in which no single MUAP can be distinguished.
50. Firing pattern analysis
• Are the number of different MUAPs firing appropriate for the
firing rate? That is, is the ratio of firing rate to MUAPs
approximately 5:1?
• Increasing force depends on two processes: activation and
recruitment.
• Activation refers to the ability to increase firing rate. This is
a central process. Poor activation may be seen in diseases of
the central nervous system (CNS) or as a manifestation of
pain, poor cooperation, or functional disorders.
51. Recruitment
• Recruitment refers to the
ability to add more MUAPs as
the firing rate increases.
Recruitment is reduced
primarily in neuropathic
diseases, although rarely it
may also be reduced in severe
end-stage myopathy.
• An incomplete interference
pattern may be due to either
poor activation or poor
recruitment.
52. Interference patterns. (A) Normal. (B)
Neuropathic. (C) Myopathic. (D) Central. In
each trace, the patient is asked to contract
maximally. In normal subjects, so many motor
unit action potentials (MUAPs) fire during
maximal contraction that differentiating
individual MUAPs are difficult. In neuropathic
recruitment, a reduced number of MUAPs fire
at a high frequency, resulting in an incomplete
interference pattern (often referred to as the
“picket fence” pattern, when only one MUAP
is firing). In myopathic recruitment, although
the number of MUAPs is normal, the
interference pattern consists of short-duration,
small-amplitude MUAPs, which fire with a
small amount of force. In central disorders, the
primary problem is the inability to fire faster
(i.e., decreased activation); although the
number of MUAPs is reduced, it is appropriate
for the level of firing.
53. “Early recruitment”
• In diseases in which there is dropout of individual muscle
fibers from a motor unit (e.g., myopathies or NMJ diseases
with block), the motor unit becomes smaller and
subsequently can generate less force.
• Because each motor unit generates less force, many motor
units must fire to generate even a small amount of force.
• This is known as “early recruitment,” which refers to the
inappropriate firing of many MUAPs to generate a small
amount of force.