This document discusses the stability testing of drug-eluting stents. It outlines Conor's reservoir technology where the drug is formulated with a bio-resorbable polymer and deposited in wells on the stent struts. Tests discussed for ensuring stability include assays to measure drug content, identification and quantification of impurities and degradation products, in vitro drug release kinetics, polymer molecular weight and content, residual solvents, and particulate matter. The document recommends a stability protocol that includes testing samples from multiple batches at various time points under long-term and accelerated conditions to establish a shelf life for the drug-eluting stents.
3. Why Drug-Eluting Stents?
First Bare Metal Stent was implanted 1987
Set the stage for stents to be regulated by
CDRH (with review of CMC by CDER)
First DES approved in US (Cypher™) in 2003
Drug/polymer matrix is applied to a scaffold
that props the artery open.
Drug slowly elutes further reducing restenosis
In some cases, the remaining metal is
endothelialized in the arterial wall
4. Stent Reservoir Technology
Drug is formulated with bio-resorbable polymer
Solution is deposited in wells in the stent struts as a
solution
Solvent is evaporated, leaving drug/polymer matrix
Isolates the matrix from the stresses of insertion and
expansion.
Drug is slowly released as the polymer decomposes
(hydrolysis of ester linkages) and is absorbed
Directionality as well as the rate of drug release can
be controlled
7. Assay
HPLC method – composite of 5 stents
Content on stent is <<1 mg (sum of
Isomers B&C)
Sample volume is low to keep
concentration at acceptable level
Sample directly injected without further
dilution
8. Impurities/Degradation Products
Thousands of isomers possible
HPLC – Size Exclusion Chromatography
(SEC)
Same sample solution as for assay
Impurity “monomers” are indistiguishable from
parent drug
Based on response of reference standard and
assay value, monomers are determined by
difference (average RRF)
Oligomers are separated and are quantitated
by normal area response calibration (RRF)
9. In Vitro Release
Desirable traits:
Predicts the release of the drug in vivo
would be optimal and highly desirable
Selected test should be simple, reliable,
and reproducible in order to be used for
QC purposes
USP apparatus/special equipment
10. In Vitro Release (continued)
Define robustness of the method:
Release media composition
pH
Dips/flow rate
Temperature
Physical factors, e.g. stent orientation, stent diameter/
length
Demonstrate discrimination – determine lots that
should be rejected
As least 3 specification-sampling times covering
initial, middle, and terminal elution
Develop IVIVC, if possible
11. Conor Elution Method
USP 4 Apparatus, closed-loop recirculating
system
Media: Buffered Tween 20 / ACN, 50 mL/stent
7-hour release method
Method development performed to have in vitro
curve resemble in vivo release
Complete profile being determined in
development, with 3 specified timepoints
chosen: early (burst), middle, >80% release
12. In Vitro Release Curve
Cummulative Sirolimus Released (% )
100.00%
90.00%
80.00%
70.00%
60.00%
50.00%
40.00%
30.00%
20.00%
10.00%
0.00%
0 2 4 6 8
Time (hr)
13. Residual Solvent
Solvent extraction
GC with FID
Solvent is difficult to remove below
~10% (w/w) relative to drug content
Solvent can potentially affect release
kinetics
Tested on release and stability
14. Polymer MW/Content
GPC Method with polystyrene stds / UV detection
Raw material controlled by Lactide/Glycolide ratio
and Inherent Viscosity (IV) of PLGA polymer
MW testing on raw material being performed to
develop correlation to IV
Product sterilization performed by e-beam, which
reduces MW by ~30%
Limited stability testing of MW on finished product
Polymer content controlled by drug/polymer ratio in
formulation.
15. Particulates
Sources: Formulation, stent platform, stent delivery
system, packaging, and environment.
Simulated use conditions should be employed in
testing
From Draft DES Guidance: “It is highly
recommended that particulate matter generation
over time be evaluated at each time point in the
stability protocol (instead of only at t=0 and
t=proposed expiration date). In the event that the
particle counts continually increase with aging or fail
to meet the acceptance criteria at the proposed
expiration date, additional data will be available to
support a shorter expiration date for the DES. “
Particle characterization and other considerations,
see draft guidance.
16. Stability Protocol
For establishment of shelf-life/expiration dating and
to support product stability for the duration of clinical
studies
Protocol can include appropriate bracketing and
matrixing designs as detailed in Guidance Document
Q1D. The protocol design should include the
extremes (in terms of both stent dimensions and total
drug load) as well as an intermediate size,
bracketing different stent designs.
“We recommend that stability testing include
samples from a minimum of three finished product
batches for each size tested.”
18. Stability Protocol (continued)
ICH recommended conditions including accelerated
are recommended.
Depending on the polymer being used, accelerated
may not be appropriate (Tg).
The following tests are recommended:
Appearance
Assay/drug content
Impurities/degradation products
In vitro drug release
Particulate matter
Sterility/package integrity at release, annually, &
expiration
19. Long Term (25 °C/60%RH) Stability
Testing Protocol
Time Points (months)
Tests Acceptance 0 3 6 9 12
Criteria*
Appearance X X X X X
Assay (drug X X X X X
content)
Impurities X X X X X
Individual
Total
Drug Release X X X X X
Particulate X X X X X
Matter
Endotoxins X X
Time Points (months)
*Same as regulatory specifications
20. Accelerated (40 °C/75%RH)
Stability Testing Protocol
Time Points (months)
Tests Acceptance 0 1 3 6
Criteria*
Appearance X X X X
Assay (drug content) X X X X
Impurities X X X X
Individual
Total
Drug Release X X X X
Time Points (months)
*Same as regulatory specifications
21. Stability Protocol (continued)
Real time, RT data should be used to establish a DES
shelf life. Depending on the quality of the data,
reasonable extrapolation of accelerated data may be
considered to assign shelf life.
22. Summary/Conclusion
Some testing is specific to the dosage form:
polymer characterization (MW, Tg),
particulate matter, residual solvents, elution
Review the draft guidance. You may have
specific circumstances in which a scientific
justification can be made for doing something
different, minimizing samples & studies.
Release and stability testing of Drug-Eluting
Stents similar to that of drugs with a different
controlled-release mechanism