SlideShare a Scribd company logo
Method Development and
Validation : Laser Scattering
Particle Size Distribution (PSD)
Analyzer
Md. Saddam Nawaz
FDA guidance: when should particle size be
measured?
International Conference on Harmonization;
Guidance on Q6A Specifications:
Test Procedures and Acceptance Criteria for New Drug
Substances and New Drug Products: Chemical Substances*
3.3.1 New Drug Substances
(b) Particle size: For some new drug substances intended for use in solid
or suspension drug products, particle size can have a significant effect on
dissolution rates, bioavailability, and/or stability. In such instances, testing
for particle size distribution should be carried out using an appropriate
procedure, and acceptance criteria should be provided.
Decision Tree #3 provides additional guidance on when particle size
testing should be considered.
FDA guidance: when should particle size be
measured?
Method Development
Goal: Reproducible method that tracks product
performance
Choose measurement approach (dry vs. wet)
Lock down RI
Vary measurement settings that can influence result
 Dry: measurement duration, concentration, air pressure
 Wet: sampler selection, dispersion, duration,
concentration, energy (mixing + ultrasound)
Test method (reproducibility)
 Meet ISO, USP or internal guidelines
 Check COV at d10, d50, d90
Goals
Reproducible method that tracks product
performance
Use structured approach for any decision/choice that
may influence result
Have data to support selections made
Document process so others (in future) understand
the decisions
Method development: available guidance
for laser diffraction measurements
ISO13320-1: Section 6.4
 Dv50 - 5 different readings: COV < 3%
 Dv10 and Dv90: COV < 5%
 “Below 10μm, these maximum values should be doubled.”
USP <429>
 Provides reproducibility ranges
 Dv50 or any central value: <10%
 Dv10, Dv90 or any non-central value: <15%
 “Below 10μm, these maximum values should be doubled.”
EP 2.9.31 provides similar advice to USP<429>
Method development: Sub-Sampling
Sampling from Drums
Technique: Grab Sampling
PLACE SPATULA INTO POWDER
EXTRACT SMALL AMOUNT FOR ANALYSIS
ACCEPTABLE FOR NARROW DISTRIBUTIONS
SEGREGATE LARGE AND SMALL WHEN POLYDISPERSE
LARGE PARTICLES PERCOLATE UPWARD
SMALL PARTICLES GRAVITATE DOWNWARD
EASY METHOD
MOST USED METHOD
Grab Sampling from Bottle
When a powder is stored in a container,
it can be mixed by rolling and tumbling
the container. The container should not
be more than half to two-thirds full. It is
important to perform this action before
“grabbing” a sample with a spatula.
Then pull sample with a spatula…..
Technique: Coning & Quartering
Technique: Chute Riffling
Technique: Rotary Riffling
The best method of representative splitting of powders
is the ROTARY RIFFLER.
Lab Sample Dividers
Sampling Technique Error Levels
Method development: Wet Method
Choosing the Solvent (Dispersent)
When powders are dispersed in liquid the
solvent must meet the following criteria:
Negligible reactivity with powder
Does not swell or shrink particles
 Have refractive index (RI) different than the sample
 Be free from bubbles & particles
 Have suitable viscosity to enable recirculation
 Be compatible with materials in the analyzer
Choosing the Solvent (Dispersant)
Dispersion Definitions
How to check for wetting
If the particles float on top and do not penetrate the water surface,
they are not wetted. This is usually a bad sign.
If the particles break through surface and sink, they are a) wetted or
b) so big that gravity is more important than surface tension. If it is
case you are in luck!
If the going gets tough…
If the going gets tough…
Try a series of options.
Here I make a series of suspensions and check them
by eye, then measure.
Common Surfactants
Igepal CA630 Ethoxylated octyl phenol
Nonionic
Triton X100 Octylphenoxypolyethoxy
ethanol Nonionic
Tween 20 Polyoxyethylene sorbitan
Nonionic
Aerosol-OT Dioctyl ester Na Sulfo
succinic acid Anionic
Sodium Hexametaphosphate
(NaPO3)6
Sodium Hexametaphosphate
(NaPO3)6
Increasing Sodium Hexametaphosphate Concentration
Refractive Index
Refractive Index is defined by two components – REAL and
IMAGINARY
RI = n + ik
Where:
n = the real component, which is the ratio of the velocity of
light in a vacuum to the velocity of light in the material =
c/vp
c = speed of light in vacuum
Vp = speed of light in particle (liquid, air)
k = the extinction coefficient of the material
i = √ - 1
Determine RI
Real component via literature or web search, Becke
line, etc.
Measure sample, vary imaginary component to see
if/how results change
Recalculate using different imaginary components,
choose
value that minimizes R parameter error calculation
RI: Imaginary Component
For transparent particles use 0 for the
imaginary component
For slightly opaque materials use 0.01
or 0.1
For opaque materials use 1.0 or higher
Values can exceed 1.0 (see below*)
RI Choosing for Mixtures
Determine RI experimentally
Concentration
Detection limit and Range
Assessment of the detection limit of the laser
diffraction technique is not required as part of
method validation.
The range of the instrument should ideally
cover the size range of the sample. For
modern laser diffraction instrumentation with
size ranges between 20nm and 2000μm this
normally presents no problem for most
pharmaceutical samples.
Specificity
Specificity, in terms of whether or not the technique
is appropriate to the material under analysis, should
be addressed as part of method development and
does not have to be revisited for method validation.
Laser diffraction provides a good method for
assessing small changes in the size distribution in this
regard. However, it is difficult to differentiate
between the different components within
pharmaceutical dosage forms using the technique.
Robustness
Measurement Duration
b) Repeatability / Sample
Measurement Stability
Measurement Duration
The purpose of this exercise is to determine a suitable
measurement duration, which will then be used for the
repeatability, reproducibility and intermediate precision
exercises.
A cycle of ten measurements should be performed for durations
of 2, 5, 7, 10 and 15 seconds.
For particles with a median size (D(v,0.5)) of >10 micron the
Relative Standard Deviation (RSD) should be <3%
For particles with a median size <10 micron the RSD must be
<6%. This reflects the Fact that smaller particles are more
difficult to disperse.
Measurement Duration
Variation of lactose D(v,0.5) with measurement duration.
Repeatability /
Sample Measurement Stability
In order to determine whether samples are stable over the
period of analysis and not subject to agglomeration, de-
agglomeration or dissolution, it is necessary to monitor the
particle size distribution at known time points.
Measurements could be taken after 1, 3, 5, 7 and 10 minutes.
At least five repeat measurements should be obtained at each
time point.
For material that has a D(v,0.5) greater than 10 microns an
acceptable time point is defined when the following conditions
are met: D(v,0.5) RSD <3%, D(v,0.1) and D(v,0.9) RSD <5%.
For material that has a D(v,0.5) less than 10 microns an
acceptable time point is defined when the following conditions
are met: D(v,0.5) RSD <6%, D(v,0.1) and D(v,0.9) RSD <10%
(ISO13320).
Repeatability /
Sample Measurement Stability
Variation of lactose size over time
(mean of 5 repeat measurements are shown)
Air Pressure (Dry measurement) /
Ultrasound (wet measurement)
A suitable pressure is one at which dispersion of the particles is
achieved but milling of the particles is not occurring.
Often both dispersion and milling occur simultaneously (which
leads to a broadening of the distribution)
By measuring the same amount of sample (ideally sampled to
single shot size by a spinning riffler) at different pressures, the
pressure at which maximum dispersion is achieved without
milling can be ascertained.
The use of ultrasound in wet measurements is very similar. ISO
13320 recommends that ultrasound can be used to assist
dispersion. Too much ultrasound can fracture friable
pharmaceutical particles, although this is extremely unusual.
Ideally measurements should be taken before, during and after
ultrasound to examine what effect sonication has on the
robustness of the measurement.
Effect of Ultrasound
(wet measurement)
Pump and Stir Rates
The pump and stir speeds used during
a measurement should be examined as
part of method development. The
chosen conditions should be capable of
suspending all the material without
causing air entrainment (a particular
problem if surfactants are being used).
Pump and Stir Rates
Effect of varying the stirrer rate on the result obtained for a typical lactose.
Linearity / Obscuration
It is suggested
that
obscurations of
5, 10, 15,
20 and 25% are
investigated in
the same way
that the
measurement
duration test
was
performed, with
the acceptable
RSDs being
similarly
specified.
Reproducibility
Variation in the results obtained for seven separate scooped sampled lactose samples.
Reproducibility
Intermediate Precision
(REPRODUCIBILITY by 2nd Analyst)
Lactose results obtained for a second analyst.
Method Development: Dry Method
Method Development: Dry
First get sampling right & determine RI
Measure at 3 different pressures (low, medium, high)
Determine optimum pressure based on good
dispersion while not breaking particles
Can also compare dry vs. wet measurements
Adjust other settings to optimize sample concentration &
duration
Ideally measure all of powder placed into the sampler
 Segregation can occur on vibrating tray
 Constant mass flow rate important for stable concentration
during measurement
Once settings chosen, test reproducibility
FDA Recommendations
The FDA’s guidance regarding the validation of
particle size analysis methods states that the
validation concepts associated with other analytical
methodologies, such as HPLC, do not transfer to
techniques such as laser diffraction. Instead, analysts
are only required to assess the intermediate precision
and robustness in order to show that the selected
method provides a reproducible method of controlling
product quality
Conclusions
Must have representative sample
Powders: select air pressure
Suspensions: wet, disperse
Check accuracy w/microscope
Investigate system settings:
Concentration, agitation, ultrasound
Design for maximum precision
Follow guidelines in standards
“Quality cannot be tested into
products; i.e., quality should be
built in by design”- ICH
THANK YOU FOR YOUR
ATTENTION!
Q and A

More Related Content

What's hot

Method-Validation-HPLC-case-study
Method-Validation-HPLC-case-studyMethod-Validation-HPLC-case-study
Method-Validation-HPLC-case-study
Shreekant Deshpande
 
Development of dissolution proceduers
Development of dissolution proceduersDevelopment of dissolution proceduers
Development of dissolution proceduers
Ahmed Elewa
 
HPLC Trouble Shotting
HPLC  Trouble ShottingHPLC  Trouble Shotting
HPLC Trouble Shotting
Manish Kumar
 
General considerations and method development in ce,
General considerations and method development in ce,General considerations and method development in ce,
General considerations and method development in ce,
ChowdaryPavani
 
Analytical Method Validation as per ICH vs USP
Analytical Method Validation as per ICH vs USPAnalytical Method Validation as per ICH vs USP
Analytical Method Validation as per ICH vs USP
Kushal Shah
 
Hplc detectors
Hplc detectorsHplc detectors
Hplc detectors
Irfan Ahmed
 
Auto titrator T70 presentation
Auto titrator T70 presentationAuto titrator T70 presentation
Auto titrator T70 presentation
Tanzeem Khan
 
Dissolution - Selection of Dissolution Media
Dissolution - Selection of Dissolution MediaDissolution - Selection of Dissolution Media
Dissolution - Selection of Dissolution Media
Sagar Savale
 
Ion pair chromatography for pharmacy students
Ion pair chromatography for pharmacy studentsIon pair chromatography for pharmacy students
Ion pair chromatography for pharmacy students
abhishek rai
 
Particle Size Analysis by Laser Diffraction Method.
Particle Size Analysis by Laser Diffraction Method. Particle Size Analysis by Laser Diffraction Method.
Particle Size Analysis by Laser Diffraction Method.
AshviniTanpure
 
Columns in HPLC
Columns in HPLCColumns in HPLC
Columns in HPLC
Varad Bende
 
Degradation Analysis Using LC-MS/MS
Degradation Analysis Using LC-MS/MSDegradation Analysis Using LC-MS/MS
Degradation Analysis Using LC-MS/MS
Macleodes pharmaceutical Ltd. Daman
 
Calibaration and validation of hplc
Calibaration and validation of hplcCalibaration and validation of hplc
Calibaration and validation of hplc
ANANT NAG
 
Analytical method development and validation for simultaneous estimation
Analytical method development and validation for simultaneous estimationAnalytical method development and validation for simultaneous estimation
Analytical method development and validation for simultaneous estimation
Professor Beubenz
 
Stability indicating rp hplc method development and validation for simultaneo...
Stability indicating rp hplc method development and validation for simultaneo...Stability indicating rp hplc method development and validation for simultaneo...
Stability indicating rp hplc method development and validation for simultaneo...
Rajasekhar
 
Supercritical fluid chromatography
Supercritical fluid chromatographySupercritical fluid chromatography
Supercritical fluid chromatography
Sagar Savale
 
Method Validation - Limit of Detection, Quantitation limits and Robustness
Method Validation - Limit of Detection, Quantitation limits and RobustnessMethod Validation - Limit of Detection, Quantitation limits and Robustness
Method Validation - Limit of Detection, Quantitation limits and Robustness
labgo
 
Chromatographic Column chemistry
Chromatographic Column chemistryChromatographic Column chemistry
Chromatographic Column chemistry
kiranreddy munnangi
 
Theory of chromatographic separations
Theory of chromatographic separationsTheory of chromatographic separations
Theory of chromatographic separations
prashik shimpi
 
Stability indicating analytical method development and validation for estimat...
Stability indicating analytical method development and validation for estimat...Stability indicating analytical method development and validation for estimat...
Stability indicating analytical method development and validation for estimat...
SriramNagarajan18
 

What's hot (20)

Method-Validation-HPLC-case-study
Method-Validation-HPLC-case-studyMethod-Validation-HPLC-case-study
Method-Validation-HPLC-case-study
 
Development of dissolution proceduers
Development of dissolution proceduersDevelopment of dissolution proceduers
Development of dissolution proceduers
 
HPLC Trouble Shotting
HPLC  Trouble ShottingHPLC  Trouble Shotting
HPLC Trouble Shotting
 
General considerations and method development in ce,
General considerations and method development in ce,General considerations and method development in ce,
General considerations and method development in ce,
 
Analytical Method Validation as per ICH vs USP
Analytical Method Validation as per ICH vs USPAnalytical Method Validation as per ICH vs USP
Analytical Method Validation as per ICH vs USP
 
Hplc detectors
Hplc detectorsHplc detectors
Hplc detectors
 
Auto titrator T70 presentation
Auto titrator T70 presentationAuto titrator T70 presentation
Auto titrator T70 presentation
 
Dissolution - Selection of Dissolution Media
Dissolution - Selection of Dissolution MediaDissolution - Selection of Dissolution Media
Dissolution - Selection of Dissolution Media
 
Ion pair chromatography for pharmacy students
Ion pair chromatography for pharmacy studentsIon pair chromatography for pharmacy students
Ion pair chromatography for pharmacy students
 
Particle Size Analysis by Laser Diffraction Method.
Particle Size Analysis by Laser Diffraction Method. Particle Size Analysis by Laser Diffraction Method.
Particle Size Analysis by Laser Diffraction Method.
 
Columns in HPLC
Columns in HPLCColumns in HPLC
Columns in HPLC
 
Degradation Analysis Using LC-MS/MS
Degradation Analysis Using LC-MS/MSDegradation Analysis Using LC-MS/MS
Degradation Analysis Using LC-MS/MS
 
Calibaration and validation of hplc
Calibaration and validation of hplcCalibaration and validation of hplc
Calibaration and validation of hplc
 
Analytical method development and validation for simultaneous estimation
Analytical method development and validation for simultaneous estimationAnalytical method development and validation for simultaneous estimation
Analytical method development and validation for simultaneous estimation
 
Stability indicating rp hplc method development and validation for simultaneo...
Stability indicating rp hplc method development and validation for simultaneo...Stability indicating rp hplc method development and validation for simultaneo...
Stability indicating rp hplc method development and validation for simultaneo...
 
Supercritical fluid chromatography
Supercritical fluid chromatographySupercritical fluid chromatography
Supercritical fluid chromatography
 
Method Validation - Limit of Detection, Quantitation limits and Robustness
Method Validation - Limit of Detection, Quantitation limits and RobustnessMethod Validation - Limit of Detection, Quantitation limits and Robustness
Method Validation - Limit of Detection, Quantitation limits and Robustness
 
Chromatographic Column chemistry
Chromatographic Column chemistryChromatographic Column chemistry
Chromatographic Column chemistry
 
Theory of chromatographic separations
Theory of chromatographic separationsTheory of chromatographic separations
Theory of chromatographic separations
 
Stability indicating analytical method development and validation for estimat...
Stability indicating analytical method development and validation for estimat...Stability indicating analytical method development and validation for estimat...
Stability indicating analytical method development and validation for estimat...
 

Similar to Method Development and Validation : Laser diffraction particle size analyzer (PSD)

Particle size distribution and its methods
Particle size distribution and its methodsParticle size distribution and its methods
Particle size distribution and its methods
sony mohammed
 
Analytical Method Validations & Detailed Method Validation Parameters
Analytical Method Validations & Detailed Method Validation ParametersAnalytical Method Validations & Detailed Method Validation Parameters
Analytical Method Validations & Detailed Method Validation Parameters
BussaSravanKumar
 
Physical Pharmaceutics Assignment 01
Physical Pharmaceutics Assignment 01Physical Pharmaceutics Assignment 01
Physical Pharmaceutics Assignment 01
MrHotmaster1
 
Analytical Method Development and validation of UV-Visible spectroscopy
Analytical Method Development and validation of UV-Visible spectroscopyAnalytical Method Development and validation of UV-Visible spectroscopy
Analytical Method Development and validation of UV-Visible spectroscopy
Imdad H. Mukeri
 
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdfANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
BALASUNDARESAN M
 
INDUSTRIAL-PHARMACY-LAB-MANUAL.pdf
INDUSTRIAL-PHARMACY-LAB-MANUAL.pdfINDUSTRIAL-PHARMACY-LAB-MANUAL.pdf
INDUSTRIAL-PHARMACY-LAB-MANUAL.pdf
subbu686728
 
Apt lab manual
Apt lab manualApt lab manual
Apt lab manual
Dr Prasanthi Boddu
 
Milling part 1 2020
Milling part 1 2020Milling part 1 2020
Milling part 1 2020
Hussein Alkufi
 
MRK2231-01_OD_broch_Viscosizer_TD_v2
MRK2231-01_OD_broch_Viscosizer_TD_v2MRK2231-01_OD_broch_Viscosizer_TD_v2
MRK2231-01_OD_broch_Viscosizer_TD_v2Rajib Ahmed
 
Dissolution procedure development and validation, USP 1092
Dissolution procedure development and validation, USP 1092Dissolution procedure development and validation, USP 1092
Dissolution procedure development and validation, USP 1092
Md. Saddam Nawaz
 
Stability studies
Stability studiesStability studies
Stability studies
Srikanth Nagabiru
 
Methods of particle size measrement (032)
Methods of particle size measrement (032)Methods of particle size measrement (032)
Methods of particle size measrement (032)
Hafiz Luqman Khalil
 
Analytical methods,cleaning validation
Analytical methods,cleaning validationAnalytical methods,cleaning validation
Analytical methods,cleaning validationSai Vivek Kosaraju
 
Gravimetric method
Gravimetric methodGravimetric method
Gravimetric method
Mohammad Mohsin Khan
 
Herbal drugs and fingerprints
Herbal drugs and fingerprintsHerbal drugs and fingerprints
Herbal drugs and fingerprintsSpringer
 
Micromeritics
MicromeriticsMicromeritics
Micromeritics
AparajitaVarshneya1
 
validation of analytical method used in cleaning
validation of analytical method used in cleaningvalidation of analytical method used in cleaning
validation of analytical method used in cleaning
KUNDLAJAYALAKSHMI
 

Similar to Method Development and Validation : Laser diffraction particle size analyzer (PSD) (20)

Particle size distribution and its methods
Particle size distribution and its methodsParticle size distribution and its methods
Particle size distribution and its methods
 
Analytical Method Validations & Detailed Method Validation Parameters
Analytical Method Validations & Detailed Method Validation ParametersAnalytical Method Validations & Detailed Method Validation Parameters
Analytical Method Validations & Detailed Method Validation Parameters
 
Physical Pharmaceutics Assignment 01
Physical Pharmaceutics Assignment 01Physical Pharmaceutics Assignment 01
Physical Pharmaceutics Assignment 01
 
Analytical Method Development and validation of UV-Visible spectroscopy
Analytical Method Development and validation of UV-Visible spectroscopyAnalytical Method Development and validation of UV-Visible spectroscopy
Analytical Method Development and validation of UV-Visible spectroscopy
 
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdfANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
ANALYTICAL METHOD VALIDATION (PHARMACEUTICAL INDUSTRY).pdf
 
INDUSTRIAL-PHARMACY-LAB-MANUAL.pdf
INDUSTRIAL-PHARMACY-LAB-MANUAL.pdfINDUSTRIAL-PHARMACY-LAB-MANUAL.pdf
INDUSTRIAL-PHARMACY-LAB-MANUAL.pdf
 
Apt lab manual
Apt lab manualApt lab manual
Apt lab manual
 
Milling part 1 2020
Milling part 1 2020Milling part 1 2020
Milling part 1 2020
 
Unit 9 hygiene calculations sampling issues compliance
Unit 9 hygiene calculations sampling issues complianceUnit 9 hygiene calculations sampling issues compliance
Unit 9 hygiene calculations sampling issues compliance
 
MRK2231-01_OD_broch_Viscosizer_TD_v2
MRK2231-01_OD_broch_Viscosizer_TD_v2MRK2231-01_OD_broch_Viscosizer_TD_v2
MRK2231-01_OD_broch_Viscosizer_TD_v2
 
Dissolution procedure development and validation, USP 1092
Dissolution procedure development and validation, USP 1092Dissolution procedure development and validation, USP 1092
Dissolution procedure development and validation, USP 1092
 
Stability Studies
Stability StudiesStability Studies
Stability Studies
 
Stability studies
Stability studiesStability studies
Stability studies
 
Methods of particle size measrement (032)
Methods of particle size measrement (032)Methods of particle size measrement (032)
Methods of particle size measrement (032)
 
Representative sampling
Representative samplingRepresentative sampling
Representative sampling
 
Analytical methods,cleaning validation
Analytical methods,cleaning validationAnalytical methods,cleaning validation
Analytical methods,cleaning validation
 
Gravimetric method
Gravimetric methodGravimetric method
Gravimetric method
 
Herbal drugs and fingerprints
Herbal drugs and fingerprintsHerbal drugs and fingerprints
Herbal drugs and fingerprints
 
Micromeritics
MicromeriticsMicromeritics
Micromeritics
 
validation of analytical method used in cleaning
validation of analytical method used in cleaningvalidation of analytical method used in cleaning
validation of analytical method used in cleaning
 

Recently uploaded

Superficial & Deep Fascia of the NECK.pptx
Superficial & Deep Fascia of the NECK.pptxSuperficial & Deep Fascia of the NECK.pptx
Superficial & Deep Fascia of the NECK.pptx
Dr. Rabia Inam Gandapore
 
NVBDCP.pptx Nation vector borne disease control program
NVBDCP.pptx Nation vector borne disease control programNVBDCP.pptx Nation vector borne disease control program
NVBDCP.pptx Nation vector borne disease control program
Sapna Thakur
 
Are There Any Natural Remedies To Treat Syphilis.pdf
Are There Any Natural Remedies To Treat Syphilis.pdfAre There Any Natural Remedies To Treat Syphilis.pdf
Are There Any Natural Remedies To Treat Syphilis.pdf
Little Cross Family Clinic
 
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
i3 Health
 
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness JourneyTom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
greendigital
 
TEST BANK for Operations Management, 14th Edition by William J. Stevenson, Ve...
TEST BANK for Operations Management, 14th Edition by William J. Stevenson, Ve...TEST BANK for Operations Management, 14th Edition by William J. Stevenson, Ve...
TEST BANK for Operations Management, 14th Edition by William J. Stevenson, Ve...
kevinkariuki227
 
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
Oleg Kshivets
 
Surat @ℂall @Girls ꧁❤8527049040❤꧂@ℂall @Girls Service Vip Top Model Safe
Surat @ℂall @Girls ꧁❤8527049040❤꧂@ℂall @Girls Service Vip Top Model SafeSurat @ℂall @Girls ꧁❤8527049040❤꧂@ℂall @Girls Service Vip Top Model Safe
Surat @ℂall @Girls ꧁❤8527049040❤꧂@ℂall @Girls Service Vip Top Model Safe
Savita Shen $i11
 
Report Back from SGO 2024: What’s the Latest in Cervical Cancer?
Report Back from SGO 2024: What’s the Latest in Cervical Cancer?Report Back from SGO 2024: What’s the Latest in Cervical Cancer?
Report Back from SGO 2024: What’s the Latest in Cervical Cancer?
bkling
 
heat stroke and heat exhaustion in children
heat stroke and heat exhaustion in childrenheat stroke and heat exhaustion in children
heat stroke and heat exhaustion in children
SumeraAhmad5
 
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 UpakalpaniyaadhyayaCharaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Dr KHALID B.M
 
HOT NEW PRODUCT! BIG SALES FAST SHIPPING NOW FROM CHINA!! EU KU DB BK substit...
HOT NEW PRODUCT! BIG SALES FAST SHIPPING NOW FROM CHINA!! EU KU DB BK substit...HOT NEW PRODUCT! BIG SALES FAST SHIPPING NOW FROM CHINA!! EU KU DB BK substit...
HOT NEW PRODUCT! BIG SALES FAST SHIPPING NOW FROM CHINA!! EU KU DB BK substit...
GL Anaacs
 
KDIGO 2024 guidelines for diabetologists
KDIGO 2024 guidelines for diabetologistsKDIGO 2024 guidelines for diabetologists
KDIGO 2024 guidelines for diabetologists
د.محمود نجيب
 
Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...
Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...
Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...
VarunMahajani
 
New Drug Discovery and Development .....
New Drug Discovery and Development .....New Drug Discovery and Development .....
New Drug Discovery and Development .....
NEHA GUPTA
 
Couples presenting to the infertility clinic- Do they really have infertility...
Couples presenting to the infertility clinic- Do they really have infertility...Couples presenting to the infertility clinic- Do they really have infertility...
Couples presenting to the infertility clinic- Do they really have infertility...
Sujoy Dasgupta
 
BRACHYTHERAPY OVERVIEW AND APPLICATORS
BRACHYTHERAPY OVERVIEW  AND  APPLICATORSBRACHYTHERAPY OVERVIEW  AND  APPLICATORS
BRACHYTHERAPY OVERVIEW AND APPLICATORS
Krishan Murari
 
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptxPharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
Dr. Rabia Inam Gandapore
 
Physiology of Chemical Sensation of smell.pdf
Physiology of Chemical Sensation of smell.pdfPhysiology of Chemical Sensation of smell.pdf
Physiology of Chemical Sensation of smell.pdf
MedicoseAcademics
 
Evaluation of antidepressant activity of clitoris ternatea in animals
Evaluation of antidepressant activity of clitoris ternatea in animalsEvaluation of antidepressant activity of clitoris ternatea in animals
Evaluation of antidepressant activity of clitoris ternatea in animals
Shweta
 

Recently uploaded (20)

Superficial & Deep Fascia of the NECK.pptx
Superficial & Deep Fascia of the NECK.pptxSuperficial & Deep Fascia of the NECK.pptx
Superficial & Deep Fascia of the NECK.pptx
 
NVBDCP.pptx Nation vector borne disease control program
NVBDCP.pptx Nation vector borne disease control programNVBDCP.pptx Nation vector borne disease control program
NVBDCP.pptx Nation vector borne disease control program
 
Are There Any Natural Remedies To Treat Syphilis.pdf
Are There Any Natural Remedies To Treat Syphilis.pdfAre There Any Natural Remedies To Treat Syphilis.pdf
Are There Any Natural Remedies To Treat Syphilis.pdf
 
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
New Directions in Targeted Therapeutic Approaches for Older Adults With Mantl...
 
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness JourneyTom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
Tom Selleck Health: A Comprehensive Look at the Iconic Actor’s Wellness Journey
 
TEST BANK for Operations Management, 14th Edition by William J. Stevenson, Ve...
TEST BANK for Operations Management, 14th Edition by William J. Stevenson, Ve...TEST BANK for Operations Management, 14th Edition by William J. Stevenson, Ve...
TEST BANK for Operations Management, 14th Edition by William J. Stevenson, Ve...
 
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...
 
Surat @ℂall @Girls ꧁❤8527049040❤꧂@ℂall @Girls Service Vip Top Model Safe
Surat @ℂall @Girls ꧁❤8527049040❤꧂@ℂall @Girls Service Vip Top Model SafeSurat @ℂall @Girls ꧁❤8527049040❤꧂@ℂall @Girls Service Vip Top Model Safe
Surat @ℂall @Girls ꧁❤8527049040❤꧂@ℂall @Girls Service Vip Top Model Safe
 
Report Back from SGO 2024: What’s the Latest in Cervical Cancer?
Report Back from SGO 2024: What’s the Latest in Cervical Cancer?Report Back from SGO 2024: What’s the Latest in Cervical Cancer?
Report Back from SGO 2024: What’s the Latest in Cervical Cancer?
 
heat stroke and heat exhaustion in children
heat stroke and heat exhaustion in childrenheat stroke and heat exhaustion in children
heat stroke and heat exhaustion in children
 
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 UpakalpaniyaadhyayaCharaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
Charaka Samhita Sutra sthana Chapter 15 Upakalpaniyaadhyaya
 
HOT NEW PRODUCT! BIG SALES FAST SHIPPING NOW FROM CHINA!! EU KU DB BK substit...
HOT NEW PRODUCT! BIG SALES FAST SHIPPING NOW FROM CHINA!! EU KU DB BK substit...HOT NEW PRODUCT! BIG SALES FAST SHIPPING NOW FROM CHINA!! EU KU DB BK substit...
HOT NEW PRODUCT! BIG SALES FAST SHIPPING NOW FROM CHINA!! EU KU DB BK substit...
 
KDIGO 2024 guidelines for diabetologists
KDIGO 2024 guidelines for diabetologistsKDIGO 2024 guidelines for diabetologists
KDIGO 2024 guidelines for diabetologists
 
Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...
Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...
Pulmonary Thromboembolism - etilogy, types, medical- Surgical and nursing man...
 
New Drug Discovery and Development .....
New Drug Discovery and Development .....New Drug Discovery and Development .....
New Drug Discovery and Development .....
 
Couples presenting to the infertility clinic- Do they really have infertility...
Couples presenting to the infertility clinic- Do they really have infertility...Couples presenting to the infertility clinic- Do they really have infertility...
Couples presenting to the infertility clinic- Do they really have infertility...
 
BRACHYTHERAPY OVERVIEW AND APPLICATORS
BRACHYTHERAPY OVERVIEW  AND  APPLICATORSBRACHYTHERAPY OVERVIEW  AND  APPLICATORS
BRACHYTHERAPY OVERVIEW AND APPLICATORS
 
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptxPharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
Pharynx and Clinical Correlations BY Dr.Rabia Inam Gandapore.pptx
 
Physiology of Chemical Sensation of smell.pdf
Physiology of Chemical Sensation of smell.pdfPhysiology of Chemical Sensation of smell.pdf
Physiology of Chemical Sensation of smell.pdf
 
Evaluation of antidepressant activity of clitoris ternatea in animals
Evaluation of antidepressant activity of clitoris ternatea in animalsEvaluation of antidepressant activity of clitoris ternatea in animals
Evaluation of antidepressant activity of clitoris ternatea in animals
 

Method Development and Validation : Laser diffraction particle size analyzer (PSD)

  • 1. Method Development and Validation : Laser Scattering Particle Size Distribution (PSD) Analyzer Md. Saddam Nawaz
  • 2. FDA guidance: when should particle size be measured? International Conference on Harmonization; Guidance on Q6A Specifications: Test Procedures and Acceptance Criteria for New Drug Substances and New Drug Products: Chemical Substances* 3.3.1 New Drug Substances (b) Particle size: For some new drug substances intended for use in solid or suspension drug products, particle size can have a significant effect on dissolution rates, bioavailability, and/or stability. In such instances, testing for particle size distribution should be carried out using an appropriate procedure, and acceptance criteria should be provided. Decision Tree #3 provides additional guidance on when particle size testing should be considered.
  • 3. FDA guidance: when should particle size be measured?
  • 4. Method Development Goal: Reproducible method that tracks product performance Choose measurement approach (dry vs. wet) Lock down RI Vary measurement settings that can influence result  Dry: measurement duration, concentration, air pressure  Wet: sampler selection, dispersion, duration, concentration, energy (mixing + ultrasound) Test method (reproducibility)  Meet ISO, USP or internal guidelines  Check COV at d10, d50, d90
  • 5. Goals Reproducible method that tracks product performance Use structured approach for any decision/choice that may influence result Have data to support selections made Document process so others (in future) understand the decisions
  • 6. Method development: available guidance for laser diffraction measurements ISO13320-1: Section 6.4  Dv50 - 5 different readings: COV < 3%  Dv10 and Dv90: COV < 5%  “Below 10μm, these maximum values should be doubled.” USP <429>  Provides reproducibility ranges  Dv50 or any central value: <10%  Dv10, Dv90 or any non-central value: <15%  “Below 10μm, these maximum values should be doubled.” EP 2.9.31 provides similar advice to USP<429>
  • 9. Technique: Grab Sampling PLACE SPATULA INTO POWDER EXTRACT SMALL AMOUNT FOR ANALYSIS ACCEPTABLE FOR NARROW DISTRIBUTIONS SEGREGATE LARGE AND SMALL WHEN POLYDISPERSE LARGE PARTICLES PERCOLATE UPWARD SMALL PARTICLES GRAVITATE DOWNWARD EASY METHOD MOST USED METHOD
  • 10. Grab Sampling from Bottle When a powder is stored in a container, it can be mixed by rolling and tumbling the container. The container should not be more than half to two-thirds full. It is important to perform this action before “grabbing” a sample with a spatula. Then pull sample with a spatula…..
  • 11. Technique: Coning & Quartering
  • 13. Technique: Rotary Riffling The best method of representative splitting of powders is the ROTARY RIFFLER.
  • 17. Choosing the Solvent (Dispersent) When powders are dispersed in liquid the solvent must meet the following criteria: Negligible reactivity with powder Does not swell or shrink particles  Have refractive index (RI) different than the sample  Be free from bubbles & particles  Have suitable viscosity to enable recirculation  Be compatible with materials in the analyzer
  • 18. Choosing the Solvent (Dispersant)
  • 20. How to check for wetting If the particles float on top and do not penetrate the water surface, they are not wetted. This is usually a bad sign. If the particles break through surface and sink, they are a) wetted or b) so big that gravity is more important than surface tension. If it is case you are in luck!
  • 21. If the going gets tough…
  • 22. If the going gets tough… Try a series of options. Here I make a series of suspensions and check them by eye, then measure.
  • 23. Common Surfactants Igepal CA630 Ethoxylated octyl phenol Nonionic Triton X100 Octylphenoxypolyethoxy ethanol Nonionic Tween 20 Polyoxyethylene sorbitan Nonionic Aerosol-OT Dioctyl ester Na Sulfo succinic acid Anionic
  • 25. Sodium Hexametaphosphate (NaPO3)6 Increasing Sodium Hexametaphosphate Concentration
  • 26. Refractive Index Refractive Index is defined by two components – REAL and IMAGINARY RI = n + ik Where: n = the real component, which is the ratio of the velocity of light in a vacuum to the velocity of light in the material = c/vp c = speed of light in vacuum Vp = speed of light in particle (liquid, air) k = the extinction coefficient of the material i = √ - 1
  • 27. Determine RI Real component via literature or web search, Becke line, etc. Measure sample, vary imaginary component to see if/how results change Recalculate using different imaginary components, choose value that minimizes R parameter error calculation
  • 28. RI: Imaginary Component For transparent particles use 0 for the imaginary component For slightly opaque materials use 0.01 or 0.1 For opaque materials use 1.0 or higher Values can exceed 1.0 (see below*)
  • 29. RI Choosing for Mixtures
  • 32. Detection limit and Range Assessment of the detection limit of the laser diffraction technique is not required as part of method validation. The range of the instrument should ideally cover the size range of the sample. For modern laser diffraction instrumentation with size ranges between 20nm and 2000μm this normally presents no problem for most pharmaceutical samples.
  • 33. Specificity Specificity, in terms of whether or not the technique is appropriate to the material under analysis, should be addressed as part of method development and does not have to be revisited for method validation. Laser diffraction provides a good method for assessing small changes in the size distribution in this regard. However, it is difficult to differentiate between the different components within pharmaceutical dosage forms using the technique.
  • 34. Robustness Measurement Duration b) Repeatability / Sample Measurement Stability
  • 35. Measurement Duration The purpose of this exercise is to determine a suitable measurement duration, which will then be used for the repeatability, reproducibility and intermediate precision exercises. A cycle of ten measurements should be performed for durations of 2, 5, 7, 10 and 15 seconds. For particles with a median size (D(v,0.5)) of >10 micron the Relative Standard Deviation (RSD) should be <3% For particles with a median size <10 micron the RSD must be <6%. This reflects the Fact that smaller particles are more difficult to disperse.
  • 36. Measurement Duration Variation of lactose D(v,0.5) with measurement duration.
  • 37. Repeatability / Sample Measurement Stability In order to determine whether samples are stable over the period of analysis and not subject to agglomeration, de- agglomeration or dissolution, it is necessary to monitor the particle size distribution at known time points. Measurements could be taken after 1, 3, 5, 7 and 10 minutes. At least five repeat measurements should be obtained at each time point. For material that has a D(v,0.5) greater than 10 microns an acceptable time point is defined when the following conditions are met: D(v,0.5) RSD <3%, D(v,0.1) and D(v,0.9) RSD <5%. For material that has a D(v,0.5) less than 10 microns an acceptable time point is defined when the following conditions are met: D(v,0.5) RSD <6%, D(v,0.1) and D(v,0.9) RSD <10% (ISO13320).
  • 38. Repeatability / Sample Measurement Stability Variation of lactose size over time (mean of 5 repeat measurements are shown)
  • 39. Air Pressure (Dry measurement) / Ultrasound (wet measurement) A suitable pressure is one at which dispersion of the particles is achieved but milling of the particles is not occurring. Often both dispersion and milling occur simultaneously (which leads to a broadening of the distribution) By measuring the same amount of sample (ideally sampled to single shot size by a spinning riffler) at different pressures, the pressure at which maximum dispersion is achieved without milling can be ascertained. The use of ultrasound in wet measurements is very similar. ISO 13320 recommends that ultrasound can be used to assist dispersion. Too much ultrasound can fracture friable pharmaceutical particles, although this is extremely unusual. Ideally measurements should be taken before, during and after ultrasound to examine what effect sonication has on the robustness of the measurement.
  • 41. Pump and Stir Rates The pump and stir speeds used during a measurement should be examined as part of method development. The chosen conditions should be capable of suspending all the material without causing air entrainment (a particular problem if surfactants are being used).
  • 42. Pump and Stir Rates Effect of varying the stirrer rate on the result obtained for a typical lactose.
  • 43. Linearity / Obscuration It is suggested that obscurations of 5, 10, 15, 20 and 25% are investigated in the same way that the measurement duration test was performed, with the acceptable RSDs being similarly specified.
  • 44. Reproducibility Variation in the results obtained for seven separate scooped sampled lactose samples.
  • 46. Intermediate Precision (REPRODUCIBILITY by 2nd Analyst) Lactose results obtained for a second analyst.
  • 48. Method Development: Dry First get sampling right & determine RI Measure at 3 different pressures (low, medium, high) Determine optimum pressure based on good dispersion while not breaking particles Can also compare dry vs. wet measurements Adjust other settings to optimize sample concentration & duration Ideally measure all of powder placed into the sampler  Segregation can occur on vibrating tray  Constant mass flow rate important for stable concentration during measurement Once settings chosen, test reproducibility
  • 49. FDA Recommendations The FDA’s guidance regarding the validation of particle size analysis methods states that the validation concepts associated with other analytical methodologies, such as HPLC, do not transfer to techniques such as laser diffraction. Instead, analysts are only required to assess the intermediate precision and robustness in order to show that the selected method provides a reproducible method of controlling product quality
  • 50. Conclusions Must have representative sample Powders: select air pressure Suspensions: wet, disperse Check accuracy w/microscope Investigate system settings: Concentration, agitation, ultrasound Design for maximum precision Follow guidelines in standards
  • 51. “Quality cannot be tested into products; i.e., quality should be built in by design”- ICH THANK YOU FOR YOUR ATTENTION!