Calibratio
n for Hematology
Analyzer
Y
ustinusAlexanderAgungY
uwono
Semarang, 5August 2018
WHAT IS YOUR CURRENT PRACTICE?
IS THERE A NEED TO CALIBRATE SYSMEX HEMATOLOGY
ANALYZER ?
YES
Definition
Calibration involves any adjustments made
to an instrument to correct the results
recovered so that they match “truth”, which
is defined by standards or reference
procedures.
🠜This procedure ensures that results for blood
specimens analysed on all instruments of a
given model will show worldwide homology.
Why Calibration Necessary?
Calibrator
WBC/RBC count calibration ICSH reference method for enumeration of erythrocytes and leucocytes
Platelet count calibration ICSH refernece method for platelet method
HGB calibration ICSH reference method on hemoglobinometry
HCT calibration ICSH reference method for the packed cell method
1. A Reference material whose value is used for the independent variable in a
calibration function.
2. SCS-1000 hematology calibrator is used for WBC, RBC, PLT, HGB and HCT
calibration.
3. The calibration value assignment of the SCS-1000 is traceable to International
conventional reference measurement procedures.
ICSH : International Council for Standardization in Haematology
Reference : Metrological Traceability of Values Assigned to Sysmex SCS-1000 Hematology Calibrator . SJI.
Vol12 no2 (2002)
QC vs Calibration
QC is Not Calibration!
• QC is checking to see if the analyzer is producing
correct results
– checking the instrument’s calibration and other analytical
processes
• Calibration is “setting” the analyzer to give correct
results
– Use calibration material/calibrator, not QC material
What is Frequency of Calibration?
Comply with
1. Manufacturer’s recommendations
2. Laboratory quality policy
3. Regulatory body regulations
4. Accreditation bodies requirements.
Calibration Requirement Recommended by Sysmex
The manufacturer recommendations (in the operator manual)
are as following :
1. During installation of analyzer at customer site.
2. After a major Preventive maintenance or change of major components
3. When QC data begin to reflect an unusual trend or are outside of the
acceptable limits and cannot be corrected by maintenance troubleshooting the
instrument
4. The laboratory's established schedule requires more frequent calibration or
calibration verification
5. The Sysmex service representative advised to do so
6. Regulatory standards require periodic calibration verification.
Record of Calibration
Calibration must be documented and the
following record must be maintained:
1. Calibration Report with calibrator data
and QC data and printouts
2. Certificate of calibration (optional)
• Remarks:
1. QC must be performed after calibration to verify that the calibration is
successful
2. Calibration report and certificate of calibration (optional) must be issued
by service organisation.
3. Standard template for commissioning report (include calibration) and
calibration report can be download from NPP.
Commisioning Report
Calibration Report
Spot the Difference
Low accuracy and low precision (poor repeatability)
Stone age man missed the bull's-eye and the 3 attempts were
not near each other.
Low accuracy but high precision
Robin Hood's Merry Man missed the bull's-eye but the 3
attempts were near each other.
Higher accuracy but low precision
Native American's 3 attempts were near the bull's-eye, but
were not near each other.
High accuracy and high precision
Olympic archer hit the bull's-eye 3 times!
• Definition
A system of maintaining standards in
manufactured products by testing a sample of the
output against the specification
• Purpose
Ensure that reported results are accountable
Quality Control
• A quality control should be performed:
– Before any start of operation
– At least every 8 hours during operation
– After replenishment of components
– After maintenance
– If there is any doubt about the accuracy of the
analysis value
Quality Control
• Eightcheck 3WP
– Poch-100i, KX-21, XP-Series
• e-Check(XS)
– XS-Series
• e-Check(XT/XE)
– XT-Series, XE-Series
• XN-Check
– XN-Series
• XN-Check BF
– XN-Series
• XN-L Check
– XN-L
Sysmex Quality Control Material
XN CHECK Level 1:
• New platelet componet can be stained
by PLT-F stain.
*Required for monitoring PLT-F
XN CHECK L2 and L3:
• do not include stainable thrombocyte
components, the cell concentration does
not allow that (technical limitation)
Expected PLT (103/µl) Concentrations
• Low (L1)
• Normal (L2)
• High (L3)
: 35-65  approx. 80*
: 190-250
: 400-700
XN-Check
*HE/ISSUE 42, 30 Sept 2016
QC Material Packing Open Stability
Eightcheck 3WP 1.5 ml 7 days
e-Check (XS) 1.5 ml 14 days
e-Check (XT/XE) 3.0 ml 7 days
XN-Check 3.0 ml 7 days
XN-Check BF 3.0 ml 30 days
XN-L Check 3.0 ml 14 days
Summary – Quality Control
Commonly used QC method:
– Levey-Jennings chart
– QC Rules
– Westgard Rules
Internal QC Method
Levey-Jennings Chart: Example
Rules:
•
•
•
•
•
12s
13s
22s
R4s
13s/22s/R4s
• Warning Rule
• Rejection Rule
QC Rules
Variables affecting Control results
Analyzer
Reagents
Controls
Mechanical
Temperature
Optical
Maintenance
Reagents
Standard
Reference / Calibration
Status
Reconstitution
Handling
Freezing
What do you do when your QC run is out?
• First do not release any results
• Do NOT automatically repeat the control!
• Investigate and determine the type of error (random
or systematic)
• Look at your QC/QA Records
– Instrument function and condition
• Any sign of instrument deterioration (frequency and length of
downtime & why, unscheduled PM), temperature checks
– Calibration and Calibration verification
• Lot #, expiry date, date (& why) of calibration
– Previous control runs (any bias observed)
• Lot#, expiry date of current run
– Reagent Storage, Reconstitution and Handling
20180805-Hematology-Calibration-1.pptx
20180805-Hematology-Calibration-1.pptx

More Related Content

What's hot

Pre analytic and postanalytic test management
Pre analytic and postanalytic test managementPre analytic and postanalytic test management
Pre analytic and postanalytic test managementVarsha Shahane
 
Automated blood cell counter by yasin arafat
Automated blood cell counter by yasin arafatAutomated blood cell counter by yasin arafat
Automated blood cell counter by yasin arafatYasin Arafat
 
Automation in clinical hematology
Automation in clinical hematologyAutomation in clinical hematology
Automation in clinical hematologyTitto Rahim
 
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...Adel Elazab Elged
 
Automation in cytology.
Automation in cytology.Automation in cytology.
Automation in cytology.Manan Shah
 
Post analytical variables dr kamlesh patel
Post analytical variables dr kamlesh patelPost analytical variables dr kamlesh patel
Post analytical variables dr kamlesh patelwww.jaailab.com
 
Calibration and Quality controls of automated hematology analyzer
Calibration and Quality controls of automated hematology analyzerCalibration and Quality controls of automated hematology analyzer
Calibration and Quality controls of automated hematology analyzerPranav S
 
Demonstration of le cells
Demonstration of le cellsDemonstration of le cells
Demonstration of le cellsSHRUTHI VASAN
 
Raju automation Msc MLT PATHOLOGY
Raju automation Msc MLT PATHOLOGYRaju automation Msc MLT PATHOLOGY
Raju automation Msc MLT PATHOLOGYrajusehrawat
 
Automated cell counters: principle and types
Automated cell counters: principle and typesAutomated cell counters: principle and types
Automated cell counters: principle and typesSivaranjini N
 
Quality Control In Clinical Laboratory
Quality Control In Clinical LaboratoryQuality Control In Clinical Laboratory
Quality Control In Clinical LaboratoryDr. Rajesh Bendre
 
Serum protein electrophoresis & their clinical importance
Serum protein electrophoresis & their clinical importanceSerum protein electrophoresis & their clinical importance
Serum protein electrophoresis & their clinical importanceDr.M.Prasad Naidu
 
Haemoglobin quality control by maintaining levey jennings chart
Haemoglobin quality control by maintaining levey jennings chartHaemoglobin quality control by maintaining levey jennings chart
Haemoglobin quality control by maintaining levey jennings chartDr Rashmi Sood
 
Quality control in clinical laboratory
Quality control in clinical laboratoryQuality control in clinical laboratory
Quality control in clinical laboratorydrgomi basar
 

What's hot (20)

reticulocyte.pptx
reticulocyte.pptxreticulocyte.pptx
reticulocyte.pptx
 
Pre analytic and postanalytic test management
Pre analytic and postanalytic test managementPre analytic and postanalytic test management
Pre analytic and postanalytic test management
 
Automation in hematology part 1
Automation in hematology part 1Automation in hematology part 1
Automation in hematology part 1
 
Automated blood cell counter by yasin arafat
Automated blood cell counter by yasin arafatAutomated blood cell counter by yasin arafat
Automated blood cell counter by yasin arafat
 
Automation in clinical hematology
Automation in clinical hematologyAutomation in clinical hematology
Automation in clinical hematology
 
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
External Quality Control Lecture MD General 2014 Course, Clin Path Ain Shams ...
 
Automation in cytology.
Automation in cytology.Automation in cytology.
Automation in cytology.
 
quality control in pathology
quality control in pathologyquality control in pathology
quality control in pathology
 
Post analytical variables dr kamlesh patel
Post analytical variables dr kamlesh patelPost analytical variables dr kamlesh patel
Post analytical variables dr kamlesh patel
 
Quality Control
Quality ControlQuality Control
Quality Control
 
Calibration and Quality controls of automated hematology analyzer
Calibration and Quality controls of automated hematology analyzerCalibration and Quality controls of automated hematology analyzer
Calibration and Quality controls of automated hematology analyzer
 
Demonstration of le cells
Demonstration of le cellsDemonstration of le cells
Demonstration of le cells
 
Automation in Hematology part 2
Automation in Hematology part 2Automation in Hematology part 2
Automation in Hematology part 2
 
CPK Test
CPK TestCPK Test
CPK Test
 
Raju automation Msc MLT PATHOLOGY
Raju automation Msc MLT PATHOLOGYRaju automation Msc MLT PATHOLOGY
Raju automation Msc MLT PATHOLOGY
 
Automated cell counters: principle and types
Automated cell counters: principle and typesAutomated cell counters: principle and types
Automated cell counters: principle and types
 
Quality Control In Clinical Laboratory
Quality Control In Clinical LaboratoryQuality Control In Clinical Laboratory
Quality Control In Clinical Laboratory
 
Serum protein electrophoresis & their clinical importance
Serum protein electrophoresis & their clinical importanceSerum protein electrophoresis & their clinical importance
Serum protein electrophoresis & their clinical importance
 
Haemoglobin quality control by maintaining levey jennings chart
Haemoglobin quality control by maintaining levey jennings chartHaemoglobin quality control by maintaining levey jennings chart
Haemoglobin quality control by maintaining levey jennings chart
 
Quality control in clinical laboratory
Quality control in clinical laboratoryQuality control in clinical laboratory
Quality control in clinical laboratory
 

Similar to 20180805-Hematology-Calibration-1.pptx

Internal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labInternal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labAnkit Raiyani
 
Hematology Chemistry Coagulation
Hematology Chemistry CoagulationHematology Chemistry Coagulation
Hematology Chemistry CoagulationNeilbenReyGuisando
 
L25 b met_complete_blood_count
L25 b met_complete_blood_countL25 b met_complete_blood_count
L25 b met_complete_blood_countShakeel Mujahid
 
Hema II Chapter11_ Automation in Hematologya.ppt
Hema II Chapter11_ Automation in Hematologya.pptHema II Chapter11_ Automation in Hematologya.ppt
Hema II Chapter11_ Automation in Hematologya.pptPravin Amabade
 
(IQC) - Internal Quality Control.pptx
(IQC) - Internal Quality Control.pptx(IQC) - Internal Quality Control.pptx
(IQC) - Internal Quality Control.pptxDr.Rajeev Ranjan
 
Seminar by prakash on validation
Seminar by prakash on validationSeminar by prakash on validation
Seminar by prakash on validationprakash64742
 
High-throughput and Automated Process Development for Accelerated Biotherapeu...
High-throughput and Automated Process Development for Accelerated Biotherapeu...High-throughput and Automated Process Development for Accelerated Biotherapeu...
High-throughput and Automated Process Development for Accelerated Biotherapeu...KBI Biopharma
 
Quality management system BY M .REJA
Quality management system BY M .REJAQuality management system BY M .REJA
Quality management system BY M .REJAMamunReja1
 
From Screening to QC: Development Considerations for Octet Methods
From Screening to QC: Development Considerations for Octet MethodsFrom Screening to QC: Development Considerations for Octet Methods
From Screening to QC: Development Considerations for Octet MethodsKBI Biopharma
 
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjQualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjPratik434909
 
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxQualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxPratik434909
 
QA for IHC and ISH USE.pdf
QA for IHC and ISH USE.pdfQA for IHC and ISH USE.pdf
QA for IHC and ISH USE.pdfTrungTonNguyn1
 
Validation of bevacizumab elisa ich q2 ver3,0 dt14.03
Validation of bevacizumab elisa   ich q2 ver3,0 dt14.03Validation of bevacizumab elisa   ich q2 ver3,0 dt14.03
Validation of bevacizumab elisa ich q2 ver3,0 dt14.03krishgen
 
Quality Control of Blood components.ppt
Quality Control of Blood components.pptQuality Control of Blood components.ppt
Quality Control of Blood components.pptpreciousbellaphemelo
 
Laboratory Quality Control .ppt
Laboratory Quality Control .pptLaboratory Quality Control .ppt
Laboratory Quality Control .pptABRARAHMED767665
 
1.qualitycontrol final
1.qualitycontrol final1.qualitycontrol final
1.qualitycontrol finalIshah Khaliq
 

Similar to 20180805-Hematology-Calibration-1.pptx (20)

ICP QC protocol
ICP  QC  protocolICP  QC  protocol
ICP QC protocol
 
Internal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation labInternal quality control (IQC) in coagulation lab
Internal quality control (IQC) in coagulation lab
 
Hematology Chemistry Coagulation
Hematology Chemistry CoagulationHematology Chemistry Coagulation
Hematology Chemistry Coagulation
 
L25 b met_complete_blood_count
L25 b met_complete_blood_countL25 b met_complete_blood_count
L25 b met_complete_blood_count
 
Hema II Chapter11_ Automation in Hematologya.ppt
Hema II Chapter11_ Automation in Hematologya.pptHema II Chapter11_ Automation in Hematologya.ppt
Hema II Chapter11_ Automation in Hematologya.ppt
 
(IQC) - Internal Quality Control.pptx
(IQC) - Internal Quality Control.pptx(IQC) - Internal Quality Control.pptx
(IQC) - Internal Quality Control.pptx
 
Seminar by prakash on validation
Seminar by prakash on validationSeminar by prakash on validation
Seminar by prakash on validation
 
Quality assurance in Hematology laboratory
Quality assurance in Hematology laboratoryQuality assurance in Hematology laboratory
Quality assurance in Hematology laboratory
 
High-throughput and Automated Process Development for Accelerated Biotherapeu...
High-throughput and Automated Process Development for Accelerated Biotherapeu...High-throughput and Automated Process Development for Accelerated Biotherapeu...
High-throughput and Automated Process Development for Accelerated Biotherapeu...
 
Quality management system BY M .REJA
Quality management system BY M .REJAQuality management system BY M .REJA
Quality management system BY M .REJA
 
From Screening to QC: Development Considerations for Octet Methods
From Screening to QC: Development Considerations for Octet MethodsFrom Screening to QC: Development Considerations for Octet Methods
From Screening to QC: Development Considerations for Octet Methods
 
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjjQualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
Qualification of HPLC & LCMS.pptxfjddjdjdhdjdjj
 
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfxQualification of HPLC & LCMS.pptdjdjdjdjfjkfx
Qualification of HPLC & LCMS.pptdjdjdjdjfjkfx
 
QA for IHC and ISH USE.pdf
QA for IHC and ISH USE.pdfQA for IHC and ISH USE.pdf
QA for IHC and ISH USE.pdf
 
Clinical lab qc sethu
Clinical lab qc sethuClinical lab qc sethu
Clinical lab qc sethu
 
Validation of bevacizumab elisa ich q2 ver3,0 dt14.03
Validation of bevacizumab elisa   ich q2 ver3,0 dt14.03Validation of bevacizumab elisa   ich q2 ver3,0 dt14.03
Validation of bevacizumab elisa ich q2 ver3,0 dt14.03
 
Quality Control of Blood components.ppt
Quality Control of Blood components.pptQuality Control of Blood components.ppt
Quality Control of Blood components.ppt
 
Laboratory Quality Control .ppt
Laboratory Quality Control .pptLaboratory Quality Control .ppt
Laboratory Quality Control .ppt
 
Journal review 06042018
Journal review 06042018Journal review 06042018
Journal review 06042018
 
1.qualitycontrol final
1.qualitycontrol final1.qualitycontrol final
1.qualitycontrol final
 

More from PUPUTPUJIANTI

7.17.11.2022_NILAI NILAI antikorupsi.ppt
7.17.11.2022_NILAI NILAI  antikorupsi.ppt7.17.11.2022_NILAI NILAI  antikorupsi.ppt
7.17.11.2022_NILAI NILAI antikorupsi.pptPUPUTPUJIANTI
 
080720-PAPARAN-RB.pptx
080720-PAPARAN-RB.pptx080720-PAPARAN-RB.pptx
080720-PAPARAN-RB.pptxPUPUTPUJIANTI
 
Nilai_dan_Prinsip_Anti_Korupsi.pptx
Nilai_dan_Prinsip_Anti_Korupsi.pptxNilai_dan_Prinsip_Anti_Korupsi.pptx
Nilai_dan_Prinsip_Anti_Korupsi.pptxPUPUTPUJIANTI
 
Pengembangan_Masyarakat.pptx
Pengembangan_Masyarakat.pptxPengembangan_Masyarakat.pptx
Pengembangan_Masyarakat.pptxPUPUTPUJIANTI
 
KELOMPOK 1_TINGKAT 3 D3 KEPERAWATAN_PENGKAJIAN CAB.pptx
KELOMPOK 1_TINGKAT 3 D3 KEPERAWATAN_PENGKAJIAN CAB.pptxKELOMPOK 1_TINGKAT 3 D3 KEPERAWATAN_PENGKAJIAN CAB.pptx
KELOMPOK 1_TINGKAT 3 D3 KEPERAWATAN_PENGKAJIAN CAB.pptxPUPUTPUJIANTI
 
Kelompok 1_Askep Komunitas pada Kelompok Anak Usia Sekolah_Pelayanan Primer (...
Kelompok 1_Askep Komunitas pada Kelompok Anak Usia Sekolah_Pelayanan Primer (...Kelompok 1_Askep Komunitas pada Kelompok Anak Usia Sekolah_Pelayanan Primer (...
Kelompok 1_Askep Komunitas pada Kelompok Anak Usia Sekolah_Pelayanan Primer (...PUPUTPUJIANTI
 
abortion-infographics.pptx
abortion-infographics.pptxabortion-infographics.pptx
abortion-infographics.pptxPUPUTPUJIANTI
 
ppt sempro dini (1).pptx
ppt sempro dini (1).pptxppt sempro dini (1).pptx
ppt sempro dini (1).pptxPUPUTPUJIANTI
 

More from PUPUTPUJIANTI (8)

7.17.11.2022_NILAI NILAI antikorupsi.ppt
7.17.11.2022_NILAI NILAI  antikorupsi.ppt7.17.11.2022_NILAI NILAI  antikorupsi.ppt
7.17.11.2022_NILAI NILAI antikorupsi.ppt
 
080720-PAPARAN-RB.pptx
080720-PAPARAN-RB.pptx080720-PAPARAN-RB.pptx
080720-PAPARAN-RB.pptx
 
Nilai_dan_Prinsip_Anti_Korupsi.pptx
Nilai_dan_Prinsip_Anti_Korupsi.pptxNilai_dan_Prinsip_Anti_Korupsi.pptx
Nilai_dan_Prinsip_Anti_Korupsi.pptx
 
Pengembangan_Masyarakat.pptx
Pengembangan_Masyarakat.pptxPengembangan_Masyarakat.pptx
Pengembangan_Masyarakat.pptx
 
KELOMPOK 1_TINGKAT 3 D3 KEPERAWATAN_PENGKAJIAN CAB.pptx
KELOMPOK 1_TINGKAT 3 D3 KEPERAWATAN_PENGKAJIAN CAB.pptxKELOMPOK 1_TINGKAT 3 D3 KEPERAWATAN_PENGKAJIAN CAB.pptx
KELOMPOK 1_TINGKAT 3 D3 KEPERAWATAN_PENGKAJIAN CAB.pptx
 
Kelompok 1_Askep Komunitas pada Kelompok Anak Usia Sekolah_Pelayanan Primer (...
Kelompok 1_Askep Komunitas pada Kelompok Anak Usia Sekolah_Pelayanan Primer (...Kelompok 1_Askep Komunitas pada Kelompok Anak Usia Sekolah_Pelayanan Primer (...
Kelompok 1_Askep Komunitas pada Kelompok Anak Usia Sekolah_Pelayanan Primer (...
 
abortion-infographics.pptx
abortion-infographics.pptxabortion-infographics.pptx
abortion-infographics.pptx
 
ppt sempro dini (1).pptx
ppt sempro dini (1).pptxppt sempro dini (1).pptx
ppt sempro dini (1).pptx
 

Recently uploaded

Introduction and different types of Ethernet.pptx
Introduction and different types of Ethernet.pptxIntroduction and different types of Ethernet.pptx
Introduction and different types of Ethernet.pptxupamatechverse
 
Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxAsutosh Ranjan
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVHARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVRajaP95
 
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSMANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSSIVASHANKAR N
 
KubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlyKubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlysanyuktamishra911
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations120cr0395
 
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordCCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordAsst.prof M.Gokilavani
 
SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )Tsuyoshi Horigome
 
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...ranjana rawat
 
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...Call Girls in Nagpur High Profile
 
UNIT-III FMM. DIMENSIONAL ANALYSIS
UNIT-III FMM.        DIMENSIONAL ANALYSISUNIT-III FMM.        DIMENSIONAL ANALYSIS
UNIT-III FMM. DIMENSIONAL ANALYSISrknatarajan
 
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Dr.Costas Sachpazis
 
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsCall Girls in Nagpur High Profile
 
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINEMANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINESIVASHANKAR N
 
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).pptssuser5c9d4b1
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...Soham Mondal
 
Microscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxMicroscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxpurnimasatapathy1234
 
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls in Nagpur High Profile
 

Recently uploaded (20)

Introduction and different types of Ethernet.pptx
Introduction and different types of Ethernet.pptxIntroduction and different types of Ethernet.pptx
Introduction and different types of Ethernet.pptx
 
Coefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptxCoefficient of Thermal Expansion and their Importance.pptx
Coefficient of Thermal Expansion and their Importance.pptx
 
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANVI) Koregaon Park Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IVHARMONY IN THE NATURE AND EXISTENCE - Unit-IV
HARMONY IN THE NATURE AND EXISTENCE - Unit-IV
 
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLSMANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
MANUFACTURING PROCESS-II UNIT-5 NC MACHINE TOOLS
 
KubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghlyKubeKraft presentation @CloudNativeHooghly
KubeKraft presentation @CloudNativeHooghly
 
Extrusion Processes and Their Limitations
Extrusion Processes and Their LimitationsExtrusion Processes and Their Limitations
Extrusion Processes and Their Limitations
 
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete RecordCCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
CCS335 _ Neural Networks and Deep Learning Laboratory_Lab Complete Record
 
SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )SPICE PARK APR2024 ( 6,793 SPICE Models )
SPICE PARK APR2024 ( 6,793 SPICE Models )
 
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
(ANJALI) Dange Chowk Call Girls Just Call 7001035870 [ Cash on Delivery ] Pun...
 
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...Top Rated  Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
Top Rated Pune Call Girls Budhwar Peth ⟟ 6297143586 ⟟ Call Me For Genuine Se...
 
UNIT-III FMM. DIMENSIONAL ANALYSIS
UNIT-III FMM.        DIMENSIONAL ANALYSISUNIT-III FMM.        DIMENSIONAL ANALYSIS
UNIT-III FMM. DIMENSIONAL ANALYSIS
 
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
Sheet Pile Wall Design and Construction: A Practical Guide for Civil Engineer...
 
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur EscortsHigh Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
High Profile Call Girls Nagpur Meera Call 7001035870 Meet With Nagpur Escorts
 
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINEMANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
MANUFACTURING PROCESS-II UNIT-2 LATHE MACHINE
 
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
247267395-1-Symmetric-and-distributed-shared-memory-architectures-ppt (1).ppt
 
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
OSVC_Meta-Data based Simulation Automation to overcome Verification Challenge...
 
Microscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptxMicroscopic Analysis of Ceramic Materials.pptx
Microscopic Analysis of Ceramic Materials.pptx
 
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
★ CALL US 9953330565 ( HOT Young Call Girls In Badarpur delhi NCR
 
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service NashikCall Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
Call Girls Service Nashik Vaishnavi 7001305949 Independent Escort Service Nashik
 

20180805-Hematology-Calibration-1.pptx

  • 2. WHAT IS YOUR CURRENT PRACTICE?
  • 3. IS THERE A NEED TO CALIBRATE SYSMEX HEMATOLOGY ANALYZER ? YES
  • 4. Definition Calibration involves any adjustments made to an instrument to correct the results recovered so that they match “truth”, which is defined by standards or reference procedures. 🠜This procedure ensures that results for blood specimens analysed on all instruments of a given model will show worldwide homology. Why Calibration Necessary?
  • 5. Calibrator WBC/RBC count calibration ICSH reference method for enumeration of erythrocytes and leucocytes Platelet count calibration ICSH refernece method for platelet method HGB calibration ICSH reference method on hemoglobinometry HCT calibration ICSH reference method for the packed cell method 1. A Reference material whose value is used for the independent variable in a calibration function. 2. SCS-1000 hematology calibrator is used for WBC, RBC, PLT, HGB and HCT calibration. 3. The calibration value assignment of the SCS-1000 is traceable to International conventional reference measurement procedures. ICSH : International Council for Standardization in Haematology Reference : Metrological Traceability of Values Assigned to Sysmex SCS-1000 Hematology Calibrator . SJI. Vol12 no2 (2002)
  • 6. QC vs Calibration QC is Not Calibration! • QC is checking to see if the analyzer is producing correct results – checking the instrument’s calibration and other analytical processes • Calibration is “setting” the analyzer to give correct results – Use calibration material/calibrator, not QC material
  • 7. What is Frequency of Calibration? Comply with 1. Manufacturer’s recommendations 2. Laboratory quality policy 3. Regulatory body regulations 4. Accreditation bodies requirements.
  • 8. Calibration Requirement Recommended by Sysmex The manufacturer recommendations (in the operator manual) are as following : 1. During installation of analyzer at customer site. 2. After a major Preventive maintenance or change of major components 3. When QC data begin to reflect an unusual trend or are outside of the acceptable limits and cannot be corrected by maintenance troubleshooting the instrument 4. The laboratory's established schedule requires more frequent calibration or calibration verification 5. The Sysmex service representative advised to do so 6. Regulatory standards require periodic calibration verification.
  • 9. Record of Calibration Calibration must be documented and the following record must be maintained: 1. Calibration Report with calibrator data and QC data and printouts 2. Certificate of calibration (optional) • Remarks: 1. QC must be performed after calibration to verify that the calibration is successful 2. Calibration report and certificate of calibration (optional) must be issued by service organisation. 3. Standard template for commissioning report (include calibration) and calibration report can be download from NPP.
  • 12. Spot the Difference Low accuracy and low precision (poor repeatability) Stone age man missed the bull's-eye and the 3 attempts were not near each other. Low accuracy but high precision Robin Hood's Merry Man missed the bull's-eye but the 3 attempts were near each other. Higher accuracy but low precision Native American's 3 attempts were near the bull's-eye, but were not near each other. High accuracy and high precision Olympic archer hit the bull's-eye 3 times!
  • 13. • Definition A system of maintaining standards in manufactured products by testing a sample of the output against the specification • Purpose Ensure that reported results are accountable Quality Control
  • 14. • A quality control should be performed: – Before any start of operation – At least every 8 hours during operation – After replenishment of components – After maintenance – If there is any doubt about the accuracy of the analysis value Quality Control
  • 15. • Eightcheck 3WP – Poch-100i, KX-21, XP-Series • e-Check(XS) – XS-Series • e-Check(XT/XE) – XT-Series, XE-Series • XN-Check – XN-Series • XN-Check BF – XN-Series • XN-L Check – XN-L Sysmex Quality Control Material
  • 16. XN CHECK Level 1: • New platelet componet can be stained by PLT-F stain. *Required for monitoring PLT-F XN CHECK L2 and L3: • do not include stainable thrombocyte components, the cell concentration does not allow that (technical limitation) Expected PLT (103/µl) Concentrations • Low (L1) • Normal (L2) • High (L3) : 35-65  approx. 80* : 190-250 : 400-700 XN-Check *HE/ISSUE 42, 30 Sept 2016
  • 17. QC Material Packing Open Stability Eightcheck 3WP 1.5 ml 7 days e-Check (XS) 1.5 ml 14 days e-Check (XT/XE) 3.0 ml 7 days XN-Check 3.0 ml 7 days XN-Check BF 3.0 ml 30 days XN-L Check 3.0 ml 14 days Summary – Quality Control
  • 18. Commonly used QC method: – Levey-Jennings chart – QC Rules – Westgard Rules Internal QC Method
  • 21. Variables affecting Control results Analyzer Reagents Controls Mechanical Temperature Optical Maintenance Reagents Standard Reference / Calibration Status Reconstitution Handling Freezing
  • 22. What do you do when your QC run is out? • First do not release any results • Do NOT automatically repeat the control! • Investigate and determine the type of error (random or systematic) • Look at your QC/QA Records – Instrument function and condition • Any sign of instrument deterioration (frequency and length of downtime & why, unscheduled PM), temperature checks – Calibration and Calibration verification • Lot #, expiry date, date (& why) of calibration – Previous control runs (any bias observed) • Lot#, expiry date of current run – Reagent Storage, Reconstitution and Handling