2. Acknowledgements
Addisa Ababa University
Jimma University
Hawassa University
Haramaya University
University of Gondar
American Society for Clinical Pathology
Center for Disease Control and Prevention-
Ethiopia
3. Outline
Hematology Whole Blood Cell Analyzers
Instrumental Principles
Initial Set-up
Daily Start-up
Calibration
Quality Control
Histograms
Flagging
Problem Solving – Troubleshooting
Case Study
4. Objectives
At the end of this lesson, the students will be able to:
Describe the general characteristics of automated
haematology analyzers
State the different principles of automated analyzers
Explain the process of calibration, parameters and
situations which necessitate calibration
Describe the use of quality control
5. Cont’d
Discuss detection of errors (flagging) and the remedial
actions with automated haematology analyzers
Describe key aspects of automated haematology
analyzers, including histogram use and interpretation
6. What is Automated
Hematology?
• Is the performance of Hemotology Lab
investigations by using Automated Analyzer as
opposed to manual procedures.
• There are various Hematology analyzers
depending on operating principles and the
parameters they performanaly.
7. What do Automated Analyzers
Perform?
Counting of WBCs, RBCs and Platelets
Measurement of Hemoglobin
Calculation of Hematological Indices (Absolute
Values)
Some can perform Differential counts (3 part and
5 part)
Some can also indicate abnormalities of RBCs,
Platelets and WBCs (Flags)
8. Advantages of Automated
Analysers
They analyze and produce results within a very
short time.
If properly used, they produce Precise and
Accurate results.
They significantly increase the number of
patients to be served, making more efficient use
of laboratory resources.
9. CBC Performed on an Automated
Hematology Cell Analyzer
Well mixed EDTA sample is used
CBC is a group of tests (WBC, RBC, Hgb, Hct,
Red Cell Indices, Platelet Count, and automated
differential)
Tests are performed simultaneously (usually in
less than a minute)
When performance limits of automated
hematology analyzer is exceeded, manual
method of cell counting and blood smear review
must occur
11. CELL-DYN 1800 Technical
Specifications
System
Features
Utilizes impedence resistance to measure
human cells;18 parameters measured
Open-sampling system with throughput of up
to 60 samples per hour; 30 µl aspiration
volume
Accurate 3-part white blood cell differential
analysis
Uses only 3 reagents, including cyanide free
lyse that are individually packaged and
monitored to eliminate waste and improve
operator safety
Q.C. built in system- Levey-Jennings,
Westgard Rules and X B
12. Post-Draw
Stability
48 hours (5 DIFF)
Throughput 60 samples / hour
Sample
Volume 30 µL (whole blood)
Sample
Identification Numeric keyboard
Flags
User definable high/low flagging of
patient results
Number of
Reagents 3 reagents only
CELL –DYN 1800
Technical Specifications
13. Precision
(Reproducibility)
Parameter CV% (95% confidence limit)
WBC less than or equal to 2.5%
RBC less than or equal to 1.7%
HGB less than or equal to 1.2%
MCV less than or equal to 1.5%
PLT less than or equal to 6.0%
CELL-DYN 1800
Performance Specifications
14. Optical Detection Principle
In the optical or hydrodynamic focusing method
of cell counting and cell sizing, laser light is used
A diluted blood specimen passes in a steady
stream through which a beam of laser is focused
As each cell passes through sensing zone of flow
cell, it scatters focused lights
Scattered light is detected by a photodetector and
converted to an electrical pulse
Number of pulses generated is directly
proportional to the number of cells passing
through the sensing zone in a specific time period
15. Electrical Impedance Principle
Utilizes non-conductive properties of blood cells
As blood cell passes through orifice of aperture it
displaces its own volume
Increased resistance between electrodes results in an
electrical pulse
RBCs and platelets counted together, separated by
pulse heights
Hydrodynamic focusing forces cells to pass single file
through aperture
23. Initial Set-Up
“Out of the Box”
For the vendor’s service engineer to do:
Check instrument for visual damage
Check for any loose parts or connections
Make sure all computer boards are properly
sealed
Check the socket to verify proper voltage outlet
25. Cont’d
For the vendor’s service engineer
Plug instrument power cord into (voltage
stabiliser) electrical supply
Confirm the correct voltage on instrument
Main power supply
Photometric voltage
Any other voltage supply that is pertinent to
instrument functions
26. Cont’d
For the vendor’s service engineer
Permit instrument to stabilise/equilibrate
Let all components reach proper temperature
Set in any parameters that may be required
Ranges
Temperatures
Adjustment for altitude if necessary
27. Ensure the Instrument is
Functioning Properly
Check the reagent containers for:
Sufficient quantity
Not beyond expiration date
No precipitates, turbidity, particulate matter, or
unusual color
Proper connections between the instrument and
the reagent containers
1
28. Cont’d
Check the waste container for:
Sufficient capacity
Proper connections
Perform daily startup per instrument manual
In addition to verifying daily startup results, verify
acceptable
Reproducibility
Carryover
Control Results
3
2
29. Calibration
Calibration fine tunes your Haematology
analyser and provides the most accurate results
possible
In the normal process of tracking data for an
extended period of time, your laboratory can
make a specific decision to recalibrate a given
parameter.
Never adjust to a specific value for an individual
sample
30. Cont’d
For best performance, calibrate all the CBC
parameters
The WBC differential is calibrated at the factory, it
does not require calibration in the laboratory
31. When to calibrate
You should calibrate your instrument
At installation
After the replacement of any component that
involves dilution characteristics or the
primary measurements (such as the apertures)
When advised to do so by your service
representative
32. When to Check the Calibration
As dictated by your laboratory procedures
When controls begin to show evidence of
unusual trends
When controls exceed the vendor’s defined
acceptable limits
33. Reproducibility Check
Checks how close several results from the same
specimen are to each other (reproducibility)
Reproducibility checks also known as precision checks
Measurement is expressed as a percentage of deviation
from the mean which is called a Coefficient of Variation
or CV
Reproducibility is not a measure of accuracy, but true
accuracy is not possible unless an instrument is precise
Check instrument manual for procedure
34. Carryover Check
Carryover is the transfer of the previous sample
to the current sample
High to low carryover checks to verify the high
results of one sample do not affect the low
results of the next sample
If carryover check does not pass, consult the
instrument manual for how to troubleshoot
35. Precision
Parameter Result Tolerance
% LYM plus or minus 3.1%
% MID plus or minus 1.6%
% GRAN plus or minus 3.5%
CELL-DYN 1800
Performance Specifications
36. Daily Start-Up Procedures*
Daily cleaning
Background counts
Electronic checks
Check calibration
Run controls
Compare open and closed mode
sampling (use a normal patient
sample) * Check instrument
manual for procedures
Must be
within
specified
limits
37. Quality Control
Purpose of QC
Assures proper functionality of instrumentation
Means of assuring accuracy of unknowns
Monitors the integrity of the calibration
When controls begin to show evidence of unusual
trends
When controls exceed the vendor’s defined
acceptable limits
38. Quality Control Methods
Assayed or
unassayed stabilised
material (Commercial)
Patient replicates
Delta checks
39. Cont’d
Assayed stabilised material
Known values
Can be purchased in volume
Can be run over time
Easily illustrates trends and shifts
40. QC method: Assayed or
Unassayed Stabilised Material
Commercially available
Known values (assayed only)
Analyse low, normal and high control
Results stored in the instrument computer
(Pentra only)
Monitored with Levey-Jennings graphs
Easily illustrates trends and shifts
43. QC Method: Delta Checks
Compare a patient’s own leukocyte, haemoglobin, MCV,
and platelet values with previous results
If difference between the two is greater than
laboratory-set limits, current result is flagged for
review
44. What should be done if QC
Results are Unacceptable?
Verify instrument functioning
Check for shifts and trends
Check integrity of material
Troubleshoot
Repeat the assay
45. When a Control is Outside its
Expected Range
Ensure the control
Material was mixed and warmed properly
If not, mix it according to the package insert
Identification information was entered correctly
If using the numeric keypad, verify you
typed the correct information
Setup information (assigned values and expected
ranges) matches the control package insert for
the current lot number being used
If they do not match, change the control’s
information to match the package insert
1
47. When a Control is Outside its
Expected Range
If any of the problems existed, rerun the control;
otherwise, proceed to the next step
Rerun the control to ensure the problem was
not a statistical outlier
Ensure the control material was not
contaminated by running another vial or level of
control
4
3
2
48. QC Corrective Action
Log Sheet
1. Write the date on the corrective action log
sheet
2. Write the condition, such as “WBC high on
normal control”
3. Write the initials of the person who noticed the
condition
49. Cont’d
4. Write the date the control was run
5. Write the action that you performed, such as
“repeated-back in range”
5. Write the initials of the person who performed
the corrective action
51. General Histogram
Characteristics
Graphic representations of cell frequencies (Y-
axis) versus cell sizes (X- axis) in femtoliters
(fL)
Provide information about erythrocyte,
leukocyte, and platelet frequency and their
distributions about the mean, and also depict
presence of subpopulations
52. General Histogram
Characteristics
Provide means of comparing sizes of patient’s
cells with normal populations
Shifts in one direction or the other can be of
diagnostic importance
53. Histograms: Electrical
Impedance Method
RBC,PLT, and WBC data are plotted in the form
of a histogram
Cell number on Y-axis and cell size on X-axis
MCV (mean cell size of all red cell pulses) are
derived from the RBC histogram
HCT, MCH and MCHC are calculated by the
standard formulas
54. Histograms: Electrical
Impedance Method
From WBC histogram, three cell populations
(lymphocytes, mononuclear cells and
granulocytes) are delineated on difference of
their cell size
Lymphocytes – 35 to 90 fL
Mononuclear cells – 90 to 160 fL
Granulocytes – 160 to 450fL
55. Histogram
WBC: Distribution with three individual peaks and
valleys at specific regions representing the
lymphocytes, monocytes, and granulocytes
All curves normally start and end at baseline
56. Normal Histogram
WBC: Trimodal distribution with individual
peaks and valleys at specific regions
LYMPHS = 35-90 fL MONOS = 90-160 fL GRANS = 160-450 fL
RBC: Unimodal, > 36 fL
PLT: Unimodal, 2-20 fL (fitted 0-70 fL)
All curves normally start and end at baseline
All curves normally represent Gaussian
distributions
57. Histograms
RBC, PLT, and WBC
plotted on histogram
X-Axis
Cell size in femtoliters
(fL)
Y-Axis
# of cells
58. WBC/Coulter Histogram as a
Quality Control Tool
Abnormality / Indicator Probable Cause Comment
WBC histogram (lymph peak) does
not start at baseline
Giant platelets, nRBC, Plt
clumping
Review smear,
correct for nRBC
Elevation of left portion of
granulocyte peak
Left Shift Review smear
Elevation of right portion of
granulocyte peak
Neutrophilia Review smear
Trail extending downward at
extreme left, or lymph peak not
starting at baseline
nRBC, Plt clumping,
unlysed RBC, cryoproteins,
parasites
Review smear
and for nRBC
Peak to left of lymph peak or lymph
peak widening towards left
nRBC
Review smear &
correct nRBC
Widening of lymph peak to right
Atypical lymphs, blasts,
plasma cells, hairy cells,
eosinophilia, basophilia
Review smear
Wider mono peak
Monocytosis, plasma cells,
eosinophilia, basophilia,
blasts
Review smear
59. RBC/Coulter Histogram as a
Quality Control Tool
Abnormality / Indicator Probable Cause Comment
Left of curve does not touch
baseline
Schistocytes and extremely
small red cells
Review smear
FBC and Platelet
histogram
Bimodal peak
Transfused cells,
therapeutic response
Review Smear
Right portion of curve extended Red cell autoagglutination
Review FBC &
Smear
Left shift of curve Microcytes
Review smear &
FBC
Right shift of curve Macrocytes
Review smear &
FBC
60. Platelet/Coulter Histogram as a
Quality Control Tool
Abnormality / Indicator Probable Cause Comment
Peak or spike at left end of
histogram (2-8 Fl)
Cytoplasmic
fragments
Review smear
Spike towards right end of
histogram
Schistocytes,
microcytes, giant
platelets
Review smear + FBC
( MCV & RDW)
( MPV & PDW)
Bimodal peak
Cytoplasmic
fragments
Review smear
61. Flagged/Abnormal Results
Questionable or abnormal results are flagged for
verification by additional means for WBCs,
RBCs, and/or platelets
Different types of flags exist for different
instruments - consult instrument manual
Smear should be reviewed when flags exist
according to standard operating procedures
62. Review of Standard Formulas
for RBC Parameter
HCT (l/l) = RBC(x10¹²/L) x MCV (fL)
MCH (pg/RBC) =
HGB (g/dL) x10
RBC (x 10¹²/L)
MCHC(g/dL) =
HGB (g/dL) x 10
HCT(l/l)
3
2
1
64. Automated Haematology-
Training Checklist
Locate important procedures in the appropriate manual
including:
Startup and shutdown procedures
QC and calibration procedures
Sample analysis procedures
Reagent replacement procedures
Printer paper and ribbon replacement procedures
Troubleshooting procedures
65. Cont’d
Identify basic instrument modules/components
and describe their functions
Correctly perform startup, clean cycle, and
shutdown procedures
Explain the importance of the startup, clean
cycle, and shutdown procedures and the
frequency with which they should be performed
Perform appropriate Quality Control checks
66. Cont’d
Review, delete and set up control files
Perform all preliminary procedures required for
calibration
Perform the CBC calibration and document the
results according to your laboratory’s protocol
Perform sample analysis in the primary,
secondary, predilute modes
67. Cont’d
Recognise the basic principles of parameter
derivation and state if a parameter is directly
measured, derived from a histogram, or
computed
Review data and recognise if a scatterplot is
normal or abnormal
Explain what a particular flag indicates
Recognise an instrument problem based on
abnormal sample results, abnormal startup or
system test results, abnormal control results,
and/or error messages
68. Vendors for Haematology
Whole Blood Analysers
Abbott (http://www.abbott.com)
Cell-Dyn 1200, Cell-Dyn 1700, Cell-Dyn 3200, Cell-
Dyn 4000
ABX Diagnostics (http://www.abx.com)
ABX Micros 60 and ABX Pentra 60 C+ and ABX
Pentra 80 ( Tanzania standard for ART laboratories)
Beckman-Coulter (http://www.beckmancoulter.com)
Ac.T series, STKS, Gen-S,MAXM,HmX
Sysmex (http://www.Sysmex.com)
SE- series; K-series; T-series
71. Knowledge Check Answer
2. What resolutions may be attempted to obtain reliable
results?
Answers:
Redraw another EDTA specimen and review results
Collect citrated blood for CBC or CBC/PLT (multiply
WBC and PLT by 1.11 to correct for anticoagulant
dilution)
Incubate blood 37 ºC for 10-30 minutes and reanalyze
Estimate WBC from smear and perform manual
differential
Perform a manual count to verify
Append appropriate comment to the result, “WBC
Clumps”