VARIOUS LABORATORY
INSTRUMENTS IN BIOCHEMISTRY
Presented by:
Dr.Kusum Bala Jain
2nd
Yr Resident
Dept. of Biochemistry
GMC Kota.
CENTRIFUGE
CENTRIFUGE
Centrifugation is a separation technique by which
particles of different shape,size and density are separated
based
upon their sedimentation rate by spinning under the
influence of centrifugal force.(extra gravitational force)
The instrument used to hold the sample and generate the
centrifugal force is called as centrifuge.centrifuge is made
up of rotor.
.A rotor can hold the samle tube and spin along its own
axis at different speeds to generate centrifugal force.Speed
Urinometer
•A Urinometer is  a  simple  piece  of  equipment  for 
determining urine specific gravity.
•A  typical  urinometer  is  composed  of  a  float,  a 
weight, and a stem. The float is an air-filled glass 
tube, ending in the weight and the stem.
•The  weight  is  a  bulb  filled  with  ball  bearings 
embedded in a red solid, probably a glue of some 
sort.  The  glass  stem  has  calibrated  graduations 
and numbers marked off to indicate specific gravity 
measurements.
•It is placed in a tube of urine, and where the 
meniscus of  the  urine  reaches  displays  the 
specific gravity of the urine.
•Correction for temperature has to be done.
•A  urinometer  is  typically  used  in  medical 
diagnostic labs
•It is the collective term for a set of laboratory techniques
 for the separation of mixtures. The mixture is dissolved 
in  a  fluid  called  the mobile phase, which  carries  it 
through  a  structure  holding  another  material  called 
the stationary phase. 
•The  various  constituents  of  the  mixture  travel  at 
different speeds, causing them to separate.
 
•The  separation  is  based  on  differential  partitioning 
between  the  mobile  and  stationary  phases.  Subtle 
differences in a compound's partition coefficient result in 
differential  retention  on  the  stationary  phase  and  thus 
changing the separation.
•Chromatography may be preparative or analytical.
 
•The purpose of preparative chromatography is to 
separate  the  components  of  a  mixture  for  more 
advanced use (and is thus a form of purification). 
•Analytical  chromatography  is  done  normally  with 
smaller amounts of material and is for measuring 
the relative proportions of analytes in a mixture. 
ELECTROPHORESIS
Factors affecting electrophoretic
mobility
Charge
Size
Shape
COLORIMETER
COLORIMETRYCOLORIMETRY
How Do We Use This Principle?How Do We Use This Principle?
Perform a Chemical Reaction with thePerform a Chemical Reaction with the
Element to be Analyzed that Results in aElement to be Analyzed that Results in a
colored Compound of that Element aftercolored Compound of that Element after
Absorbing Light.Absorbing Light.
Measure the AmountMeasure the Amount
of Light Absorbed.of Light Absorbed.
COLORIMETRYCOLORIMETRY
1.The Chemistry Involved.1.The Chemistry Involved.
2. The Length of Light Travel.2. The Length of Light Travel.
3. The Amount (Concentration) of3. The Amount (Concentration) of
Absorbing MaterialAbsorbing Material..
The Amount of Light AbsorbedThe Amount of Light Absorbed
Is Related To:Is Related To:
CONCENTRATION CAN BE COLORIMETRICALLYCONCENTRATION CAN BE COLORIMETRICALLY
DETERMINED IF:DETERMINED IF:
1.    Able  to  chemically  develop  a  color  with  that  substance  and  only  that         
substance
2.  The developed color obeys (follows) Beer’s Law over a  reasonable  range 
of concentrations
3.    The  developed  color  must  be  stable  for  reasonable  length  of  time, 
reproducible, and sensitive to small changes in concentration
4.    All  loss  of  transmitted  light  must  be  from  absorbance  by  substance 
measured (developed color)
5.  All of substance present in sample must be available for reaction with  color 
developing agent
6.  Able to measure amount of light absorbed
•A colorimeter is a device used in colorimetry. In 
scientific fields the word generally refers to the device 
that measures the absorbance of particular 
wavelengths of light by a specific solution.
•This device is most commonly used to determine the 
concentration of a known solute in a given solution by 
the application of the Beer-Lambert law, which states 
that the concentration of a solute is proportional to the 
absorbance.
1) Wavelength selection, (2) Printer button, (3) 
Concentration factor adjustment, (4) UV mode 
selector (Deuterium lamp), (5) Readout, (6) 
Sample compartment, (7) Zero control (100% T), 
(8) Sensitivity switch, (9)ON/OFF switch[2]
The essential parts of a colorimeter are:
•a light source (often an ordinary low-voltage 
filament    lamp)
•an adjustable aperture
•a set of colored filters
•a cuvette to hold the working solution
•a detector (usually a photocell) to measure the 
transmitted light
•a meter to display the output from the detector
Filters
•Changeable optics filters are used in the colorimeter to select the 
wavelength of light which the solute absorbs the most, in order to 
maximize  accuracy.  The  usual  wavelength  range  is  from  400  to 
700nanometers (nm).
•If  it  is  necessary  to  operate  in  the ultraviolet range  (below 
400 nm) then some modifications to the colorimeter are needed. 
In  modern  colorimeters  the  filament  lamp  and  filters  may  be 
replaced by several light-emitting diodes of different colors.
Cuvettes
•Main article: Cuvette
•In a manual colorimeter the cuvettes are inserted and removed 
by hand. An automated colorimeter (as used in an AutoAnalyzer) 
is fitted with a flowcell through which solution flows continuously.
Output
• The output from a colorimeter may be displayed
by an analogue or digital meter and may be shown
as transmittance (a linear scale from 0-100%) or
as absorbance (a logarithmic scale from zero to
infinity).
• The useful range of the absorbance scale is from
0-2 but it is desirable to keep within the range 0-1
because, above 1, the results become unreliable
due to scattering of light.
• In addition, the output may be sent to a
chart recorder, data logger, or computer.
SPECTROPHOTOMETER
PRINCIPLE
- Beer Lambert Law is concerned with light absorption in
relation to solution concentration and cell path length.
- It states that the intensity of a ray of monochromatic light
decreases exponentially as the concentration of the
absorbing medium increases.
- In other words, the more dissolved substance you have in
a solution, the more light that will be absorbed, and the
less light that will be transmitted through the solution.
Absorbance, Transmittance, and Reflection.  A spectrophotometer measures how light interacts with 
atoms or molecules in a sample.
Parts of a Spectrophotometer
• Lamp
• Prism
• Sample holder
• photomultiplier
• Display
How a Spectrophotometer Works
•White light hits grating or prism
•Light is split into colors of the rainbow
•Wavelength knob directs different colors   
toward sample
How a UV Spectrophotometer Works.
Similar to a VIS spectrophotometer, the
UV spec shines ultraviolet light or visible
light on a sample, and a detector
measures the amount of light that
passes through, or is absorbed by, the
sample.
Colors of Light in the Visible
Spectrum. Humans can see
light with wavelengths of about
350 to 700 nm.
How Concentration Affects Absorbance. If a sample has twice as many molecules as another, it
can absorb twice as much light. This is true at any wavelength. It is important to know a sample’s
wavelength of maximum light absorbance, so that the difference in absorbance due to concentration is
obvious.
FLUROMETER
•A fluorometer or fluorimeter is a device used to
measure parameters of fluorescence, its intensity and
wavelength distribution of emission spectrum after
excitation by a certain spectrum of light.
•These parameters are used to identify the presence
and the amount of specific molecules in a medium.
•Modern fluorometers are capable of detecting
fluorescent molecule concentrations as low as 1 part per
trillion.
•Fluorescence analysis can be orders of magnitude
more sensitive than other techniques.
Application:
•includechemistry/biochemistry, medicine, environmental
monitoring.
Principle
Fluorescence is the emission of visible light by a
substance that has absorbed light of a different
wavelength. The emitted photon has a longer
wavelength and lesser energy.This phenomenon is
called as fluorescence.
Quantification and Characterization of fluorescent
compounds by measuring intensity of fluorescence
using fluorimeter called as fluorimetry.
Uses of flurometer
To quantify amino acids and peptides by labelling
with an extrinsic fluor,acridine orange.
For localization of enzymes in cells and metals in
metalloproteins.
For conformation alanalysis of enzymes proteins
and nucleic acid.
In cell sorting and cell counting.
To detect malignant cells
To quantify catecholamine.quinidine,and
porphyrin.
SEMI AUTOANALYZER
Semi Autoanalyser
Here the samples and reagents are mixed and read
manually .Based on colorimetry principle.
Disadvantages are:
More amount of sample is needed
Time consuming
Need technical expertisation
Purpose of Autoanalyzers
An autoanalyzer sequentially measures
blood chemistry through series of steps of
• mixing,
• reagent reaction and
• colorimetric measurements.
consists of different module including:
•a sampler, pump, mixing coils, optional
sample treatments dialysis, distillation,
heating, etc,
•a detector, and data generator.
•Most continuous flow analyzers depend on
color reactions using a flow through
colorimeter
Principle of operation
In Segmented Flow Analyzers (SFA), the sample is
mixed with small reproducible volumes of the required
reagents
air bubbles are introduced into the flow, creating
about 20 - 100 segments of liquid for each sample
The sample / reagent mixture flows through mixing
coils (heated coils)  a color proportional to the
amount of analyte in each sample is developed
The samples with developed color flow through a
colorimeter to measure the color
It consists of
Sampler:
 Aspirates samples, standards, wash solutions into the
system
Proportioning pump:
 Mixes samples with the reagents so that proper
chemical color reactions can take place, which are then
read by the colorimeter
Dialyzer:
 The purpose of a dialyzer is to separate the analyte
from interfering substances such as protein, whose
large molecules do not go through the dialysis
membrane but go to a separate waste stream
 The analyte infuses through the diaphragm into a
separate flow path going on to further analysis
It consists of
Heating bath:
 Controls temperature (typically at 37 °C), as temp
is critical in color development
Colorimeter:
 Monitors the changes in optical density of the fluid
stream flowing through a tubular flow cell. Color
intensities proportional to the substance
concentrations are converted to equivalent
electrical voltages .
Recorder:
 Displays the output information in a graphical
form.
Block diagram
ELECTROLYTE ANALYZER
•It is a very sensitive in vitro assay technique used to
measure concentrations of antigens (for example,
hormone levels in the blood) by use of antibodies. As
such, it can be seen as the inverse of a
radiobinding assay, which quantifies an antibody by use
of corresponding antigens.
•Although the RIA technique is extremely sensitive and
extremely specific, requiring specialized equipment, it
remains among the least expensive methods to perform
such measurements. It requires special precautions and
licensing, since radioactive substances are used.
Radioimmunoassay (RIA)
Method:
To perform a radioimmunoassay, a known quantity of
an antigen is made radioactive, frequently by labeling it
with gamma-radioactive isotopes of iodine, such as
I-125, attached to tyrosine.
This radiolabeled antigen is then mixed with a known
amount of antibody for that antigen, and as a result,
these two specifically bind to one another. Then, a
sample of serum from a patient containing an unknown
quantity of that same antigen is added.
This causes the unlabeled (or "cold") antigen from the
serum to compete with the radiolabeled antigen ("hot")
for antibody binding sites.
As the concentration of "cold" antigen is
increased, more of it binds to the antibody,
displacing the radiolabeled variant, and reducing
the ratio of antibody-bound radiolabeled antigen
to free radiolabeled antigen.
The bound antigens are then separated from
the unbound ones, and the radioactivity of the
free antigen remaining in the supernatant is
measured using a gamma counter
Radioimmunoassay Procedure
Applications of
Radioimmunoassays
Endocrinology
Insulin, HCG, Vasopressin
Detects Endocrine Disorders
Physiology of Endocrine Function
Pharmacology
Morphine
Detect Drug Abuse or Drug Poisoning
Study Drug Kinetics
Applications of
Radioimmunoassays
Epidemiology
Hepatitis B
Clinical Immunology
Antibodies for Inhalant Allergens
Allergy Diagnosis
Oncology
Carcinoembryonic Antigen
Early Cancer Detection and Diagnosis
CHEMILUMINESCENCE
Emission of light with limited emission of heat
(luminescence), as the result of a chemical reaction.
[A] + [B] [◊] [→ → Products] + light
[A], [B]: reactants
[◊]: excited intermediate
For example, if [A] is luminol and [B] is hydrogen peroxide
in the presence of a suitable catalyst we have:
luminol + H2O2 →3-APA[◊] →3-APA + light
Where:
3-APA is 3-aminophthalate
3-APA[◊] is the excited state producing light as it decays
to a lower energy level.
CHEMILUMINISCENCE
Luminol and peroxidase
before adding H2O2
Chemiluminiscence after
addition H2O2
68
What is
AAS ?Atomic absorption spectroscopy is a
quantitative method of analysis that is
applicable to many metals and a few
nonmetals.
The technique was introduced in 1955 by
Walsh in Australia
IntroductionIntroduction
69
PRINCIPLEPRINCIPLE
A much larger number of the gaseous metal atoms will
normally remain in the ground state.
These ground state atoms are capable of absorbing radiant
energy of their own specific resonance wavelength.
If light of the resonance wavelength is passed through a flame
containing the atoms , then part of the light will be absorbed.
The extent of absorption will be proportional to the number of
ground state atoms present in the flame.
70
What is AAS ?
• An atomic absorption spectrophotometer
consists of a light source, a sample
compartment and a detector.
Light SourceLight Source DetectorDetector
SampleSample
CompartmentCompartment
72
Fuel and oxidant
flame
 Air – acetylene
Air- propane
Air- hydrogen
 Nitrous oxide – acetylene
Auxiliary
oxidant
Fuel
77
InstrumentationInstrumentation
Line
source Monochromator Detector
Read-outNebulizer
Schematic diagram of a flame spectrophotomer
Atomization
Background
The impact of a stream of high energy electrons
causes the molecule to lose an electron forming a
radical cation.
A species with a positive charge and one unpaired
electron
+ e
-
C H
H
H
H H
H
H
HC + 2 e
-
Molecular ion (M+
)
m/z = 16
BackgroundThe impact of the stream of high energy electrons
can also break the molecule or the radical cation
into fragments.
(not detected by MS)
m/z = 29
molecular ion (M
+
) m/z = 30
+ C
H
H
H
+ H
HH C
H
H
C
H
H
H C
H
H
C
H
H
H C
H
H
+ e
-
H C
H
H
C
H
H
H
m/z = 15
Background
Only cations are detected.
Radicals are “invisible” in MS.
The amount of deflection observed depends on the
mass to charge ratio (m/z).
Most cations formed have a charge of +1 so the amount
of deflection observed is usually dependent on the mass
of the ion.
Fragmentation Patterns
The impact of the stream of high energy electrons
often breaks the molecule into fragments, commonly
a cation and a radical.
Bonds break to give the most stable cation.
Mass Spectrometer can be combined with gas
chromatography to analyze mixtures of compounds.
GC separates the components of the mixture.
Each component is analyzed by the Mass
Spectrometer.
ARTERIAL BLOOD GAS ANALYZER
pH meter
pH meter
It is an electronic device used for measuring the pH which is
the concentration of Hydrogen ions in an aqueous solution.
The pH meters work in liquids though special probes are
sometimes also used to measure the pH of semi-solid
substances.
Typical pH meter consists of a special measuring probe
(a glass electrode) connected to an electronic meter that
measures and displays the pH reading.
pH Meter
A sample is placed in a cup and the glass probe at the
end of the retractable arm is placed in it.
The probe is connected to the main box.
There are two electrodes inside the probe that
measure voltage.
One is contained in liquid with fixed pH.
The other measures the acidity of the sample through
the amount of H+ ions.
pH Meter
A voltmeter in the probe measures the difference
between the voltages of the two electrodes.
The meter then translates the voltage difference into
pH and displays it on the screen.
Before taking a pH measurement the meter must be
calibrated using a solution of known pH.