SlideShare a Scribd company logo
GOOD REGULATORY PRACTICES
Validation Of Utility
Submitted by: Sonal Sharma
Submitted to: Dr. Navneet Sharma
1
Amity Institute of Pharmacy
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
TABLE OF CONTENT-
1. Introduction
2. Validation of HVAC system
3. Cleaning validation
4. Validation of water system
5. Validation of steam system
6. Validation of compressed air systems
7. bibiliography
2
Validation of HVAC
system
• HVAC stands for heating
ventilation and air conditioning
system.
• The HVAC system also known as
the air handling unit (AHU) system
are placed in the clean rooms for
thermal control.
3
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
Test performed for validation of HVAC system:
1. ACPH
2. Filter integrity test
3. Non-viable particle count
4. Air flow visualization
5. Recovery period study
6. Temperature / relative humidity/ differential pressure monitoring
7. Air borne viable particle monitoring
4
Air changes per hour
• If the supplied air by HVAC per hour
is equal to the volume of the room/
area, then it will be one ACPH.
• To determine the average filter face
velocity and uniformity, and the
average room airflow velocity and
uniformity within a clean room.
• Test apparatus: ANEMOMETER
• Anemometer is placed
approximately 6 inches beneath
the filter surface and take velocity
of filter from 5 corners.
5
V1 V2
V3 V4
V5
6
• Get average velocity ( V ) = V1 + V2 + V3 + V4 + V5
5
• Calculate the area of filter (A) : L ( Length of filter) x W ( width of filter )
• Calculate total air volume supplied ( T ): A (in square feet ) x V ( in minutes )
• Calculate total volume of room : L (Length) x W (width) x H ( height in feet)
• Calculate ACPH by: T x 60
Room volume
• As per ISO 14644 recommended ACPH for clean rooms are as:
ISO CLASS ACPH
ISO 5 240 – 600
ISO 6 150 – 240
ISO 7 60 – 150
ISO 8 5 - 60
Filter integrity test:
• Filter integrity test is carried out to
assure the integrity of HEPA (high
efficiency particle airflow system)
filters against filter failure due to
damage while transportation,
installation or operation.
• Test apparatus: AEROSOL
GENERATOR & PHOTOMETER
7
• Through aerosol generator we generate
particle of PAO (Poly alpha Olefin) of 0.5 to
0.7 micron and supply it to AHU.
• Scan the downstream aerosol concenteration
by keeping photometer at HEPA surface.
• Scanning rate should be approx. 10 CFM
(cubic feet/ minutes)
• Acceptance criteria: NMT(not more then)
0.01% of upstream aerosol concenteration.
Non - viable particle count
• To determine the concentration of non-viable
airborne particles, equal to and greater than the
specified sizes, at designated sampling locations.
• Test apparatus: LSAPC ( light scattering air born
particle counter )
8
• Prior to testing all relevant aspects of the clean room that contributes to its integrity are
complete and function in accordance with its performance qualification.
• Sampling shall be done at rest condition only.
• Sampling location shall determine as per total area of clean room.
• For areas larger then 1000 m² , NL = 27 x A NL = Minimum number of locations
1000 A = area of clean room in m².
• Sampling location must be evenly distributed to entire room.
• Through particle counter get particle size of 0.5 & 5.0µ ( per meter cube of air).
Area in Number of sampling
locations
Area in Number of sampling
locations
2 1 32 8
4 2 36 9
6 3 52 10
8 4 56 11
10 5 64 12
24 6 68 13
28 7 72 14
9
Sampling location chart:
The minimum sample volume should be atleast 2 liters. Air with a minimum sampling
time for 1 minute for each sample at each location.
Air flow visualization
• To confirm that the air flow direction and its
Uniformity confirm to the design and
performance specification.
• Test Apparatus: Video camera, Glycerin with
Rose water, Fogger (Previously TTC (Titanium
Tetra Chloride) was used to
produce smoke/fog).
• Generate fog with the help of fogger and expose
the fog at supply end of the system.
10
• Observe the smoke pattern and ensure that, it should be unidirectional, smoke must
move from supply to return duct. & from positive pressure area to negative pressure
area.
• Picturise the direction of fog by the help of a Video.
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
Recovery Period Study
• This test is performed to determine the amount of time necessary for the clean room and
its system to reach a steady state in means of environmental condition (temperature and
RH) & cleanliness level after a brief particle generation event within the clean space.
• Test Apparatus: PARTICLE COUNTER & DIGITAL HYGROMETER
• To verify the recovery of particles present in clean room generate and introduce particles
of size 0.5 & 5.0 micron through HVAC system by PAO and evaluate the recovery
through particle counter after starting the HVAC system.
• To verify recovery of environmental condition, increase the temperature and humidity of
area by spraying hot water upto 75%.After starting the HVAC system, monitor the
humidity and temperature to get stabilized and area comes within specified
11
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
Temperature/Relative Humidity/ Differential Pressure
Monitoring
• To demonstrate the capability of the HVAC System to consistently maintain
Differential Pressures, Temperatures and Relative Humidity in the control areas.
• After completion of all said activities in qualification, area should monitor for
Temperature, Relative Humidity and Pressure differential for at least for 8 hrs. with
interval of 1 hr. ± 15 minute.
• This should be lead by monitoring the same on daily basis, at least twice a day.
12
Air Borne Viable Particle Count
• To check and ensure that the
controlled area meets the desired
level of area classification with
respect to viable counts.
• Test Apparatus: PRE-
INCUBATED MEDIA PLATES.
• Viable particle count monitoring-
shall done for at least 3 days for
qualification and as per
schedule after that.
13
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
CLEANING VALIDATION
• Cleaning validation is documented evidence that an approved cleaning
procedure is being used that will make equipment suitable for processing of
required medicinal product.
• Cleaning validation should be performed in order to confirm the
effectiveness of any cleaning procedure for all product contact equipment.
CLEANING PRINCIPLE : Pharmaceutical products can be contaminated by
other pharmaceutical products or APIs, by cleaning agents, by micro-
organisms or by other material (e.g. air-borne particles, dust, lubricants, raw
materials, intermediates, auxiliaries etc.). In many cases, the same equipment
may be used for processing different products. To avoid contamination of the
following pharmaceutical product, adequate cleaning procedures are essential.
14
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
TYPE OF CLEANING :
• TYPE A (MINOR): This type of cleaning take place between two batches
of same product. For minor cleaning, cleaning validation is not required,
since cross contamination is not an issue.
• TYPE B (MAJOR): This type of cleaning take place between two
different products.
HARD TO CLEAN AREA:
• Area/spots in any equipment which were really hard to reach and clean
define as hard to clean area.
• Hard to clean area should consider as worst case and must capture
in sampling plan. 15
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
SELECTION OF WORST CASE THROUGH PRODUCT
MATRIX
• The Worst-Case product for cleaning should be selected based on the APIs
used in the formulation.
• Following approach should be considered for selection of worst case molecule:
1. Aqueous Solubility (Lower aqueous solubility = higher worst case)
2. Therapeutic Dose (Lower Therapeutic dose = higher worst case)
3. LD50 (Lower LD50 = higher worst case)
16
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
S. no. Product
name
Active
material
Strength
( in mg)
Solubility in water LDD
(in mg)
Batch size
(liters)
1 ABC XXXX 50 Slightly soluble in water 400 3000
XXXX 125 Practically insoluble in
water
500
2 DEF XXXX 50 Slightly soluble in water 400 2000
XXXX 125 Practically insoluble in
water
500
3 GHI XXXX 50 Slightly soluble in water 400 5000
XXXX 125 Practically in soluble in
water
500
4 JKL XXXX 4 Freely soluble in water 10 3000
XXXX 2.5 Freely soluble in water 700
5 MNO XXXX 20 Practically insoluble in
water
100 5000
XXXX 10 Insoluble in water 200
XXXX 5 Practically insoluble in
water
40
6 PQR XXXX 540 Insoluble in water 2000 5000
17
PRODUCT MATRIX
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
Sr.
no.
Equipment Make Equipment id Material of
construction
Surface
area
(cm²)
10% Extra
of Total
Surface
area (cm²)
1 Manufacturing
vessel
2 Storage vessel
3 Scoops
4 Filling &
sealing machine
5 Conveyor belt
6 Material
transfer pipes
7 Filter press
total
18
EQUIPMENT TRIAL
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
SAMPLING CRITERIA
• Before sampling the surface would be clean visually.
• SWAB (Direct Surface Sampling): When samples were taken through
direct surface of the equipment with the help of swab sticks.
• Rinse Sample: Rinse sample shall be collected from the hard-to-access
locations of the equipments. Equipment surface shall be rinsed with the
given quantity of rinse volume with purified water and the sample shall be
collected from the rinsate.
19
SEQUENCE OF
SAMPLING
1. Microbial Swab
2. Chemical Swab
3. Chemical Rinse
METHODOLOGY TO GET
SWAB SAMPLE
• Swab sample are taken with the
help of swab sticks, which are
initially soaked in Purified water
and after sampling put
into mother solvent.
20
• To get swab sample select an area of 5x5 or 10x10 cm with the help of
sampler, and apply swab stick in above shown manner.
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
CALCULATION (NOEL)
NO OBSERVED EFFECT LEVEL
• Is the amount of drug/substance in mg that does not have any effect on the
human health.
• NOEL is calculated by using LD 50
(Lethal Dose 50 is the amount of a material, given all at once, which cause the death
of 50% of a group of animals) of the drug.
NOEL = (LD 50 × 70 kg)
2000
Where: LD 50 = Lethal Dose, 70 kg = Average Adult weight & 2000 is a constant.
NOEL is determined to calculate the MACO in cleaning validation.
21
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
CALCULATION (MACO)
MAXIMUM ALLOWABLE CARRY OVER
• Mathematically calculated quantity of residue from a previous product
when carried out over into a different product that can cause potential
harm to the patient.
• Three criteria to calculate MACO:
1. Therapeutic Daily Dose criteria
2. NOEL criteria
3. PPM criteria
22
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
• THERAPEUTIC DAILY DOSE CRITERIA:
MACO = TDD(p) × MBS(n)
SF x TDD(n)
Where: TDD(p) = Therapeutic Daily Dose of previous product
MBS(n) = Minimum Batch Size of next product
TDD(n) = Therapeutic Daily Dose of next product
SF = Safety Factor (1000 for Oral, 100 for external Preparation,& 10 for
sterile preparations)
23
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
• NOEL CRITERIA:
MACO = NOEL (P) × MBS (n)
SF × TDD(n)
Where: NOEL(p) = No Observed Effect Level of previous product
MBS(n) = Minimum Batch Size of next product
TDD(n) = Therapeutic Daily Dose of next product
SF = Safety Factor (1000 for Oral, 100 for external Preparation,& 10 for
sterile preparations)
24
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
• PPM CRITERIA:
If MACO calculations result in unacceptably high or irrelevant carryover
figures, or toxicological data for intermediates are not known, the approach of
a general limit may be suitable. User preferable may choose to have such an
upper limit as a policy. The general limit is often set as an upper limit for the
maximum concentration (MAXCONC) of a contaminating substance in a
subsequent batch.
MACOppm = MAXCONCs x MBS
Where: MAXCONC = General limit for maximum allowed concentration
(kg/kg or ppm) of"previous" substance in the next batch.
MBS = Minimum batch size for the next products),(where
MACO can end up).
25
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
MICROBIAL/ENDOTOXINS CRITERIA
• Samples should be analyzed for microbial aspects in all the three runs
during cleaning validation. The following should be acceptance criteria for
Microbiological monitoring during cleaning validation / verification:
26
Parameter Non sterile formulation
equipments
Sterile formulation
equipment
Total CFU - Swab NMT 50 CFU /25cm² NMT 5 CFU /25cm²
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
ANALYTICAL METHOD VALIDATION
Any instrumental analytical procedures used to analyze cleaning validation
samples need to be specified and sufficiently sensitive to determine the low
levels of residues typically found in samples.The methods used to analyze
samples that allow the equipment to be released for manufacture of another
product shall be validated to ensure that it meets following requirements, but
not limited to:
• Specificity
• Linearity
• Precision
• Robustness
• LOQ and LOD Determination
• Sample recovery
• Stability of analytical solution 27
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
HOLD TIME STUDIES
• The objective of hold time study is for establishing time limit between
equipment cleaning and reuse. It is done to ensure that the equipment
remains clean till the nest use.
• Equipment Holding Studies prior to cleaning/ Un cleaned equipment hold
time study: The interval between the end of production and the beginning
of the cleaning process shall be established through equipment holding
studies prior to cleaning.
• Cleaned Equipment Hold Time Study : The interval between the initiation
of production and the completion of the cleaning process
shall be established.
28
Water system validation
• Water system validation is very
important in pharmaceuticals.
Water is used in the
manufacturing process in very
large quantities during different
stages of manufacturing. During
manufacturing, water is
continuously generated and
distributed and can not be tested
and analyzed for each and every
quantity used in production.
• To provide consistent quality
standards, It needs to validate for
treatment, generation, storage, and
distribution. So, there is a total of
3 phases for water system
validation.
29
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
3 Types of Water System Validation
• Water System Validation Phase 1
• Water System Validation Phase 2
• Water System Validation Phase 3
PHASE I: INVESTIGATION PHASE
• Duration: Phase 1 typically lasts 2-4 weeks.
• Focus: It primarily focuses on rigorous monitoring and testing of the system to
ensure it operates continuously without any failures or deviations from expected
performance.
• Goals: The main goals are to establish initial operating procedures, verify the
quality of incoming water, develop maintenance procedures, and demonstrate that
the system can consistently produce the required quantity and quality of water. 30
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
PHASE II: SHORT-TERM CONTROL
• Duration: Phase 2 also lasts 2-4 weeks.
• Focus: This phase continues intensive monitoring, but it places a stronger
emphasis on refining and finalizing the standard operating procedures (SOPs)
established in Phase 1. It also allows for the use of water for manufacturing
purposes.
• Goals: Phase 2 aims to demonstrate consistent operation within established
ranges, ensure the system consistently produces the required water quality and
quantity when following the refined SOPs, and verify that the system works
effectively in actual manufacturing processes.
31
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
Phase III: Long-Term Control
• Duration: Phase 3 is the longest, running for approximately one year.
• Focus: Unlike the previous phases, Phase 3 is more about extended and long-
term performance. It takes into account seasonal variations and involves less
frequent testing.
• Goals: The primary objectives in Phase 3 are to prove that the system can
reliably perform over an extended period, evaluate its performance in different
seasons, and establish a routine maintenance plan based on the lessons learned
in the earlier phases. Sampling and testing become less frequent but are based
on established procedures from Phases 1 and 2.
32
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
VALIDATION OF STEAM SYSTEMS
• Pure steam is used in various operations in pharmaceuticals but its use in
sterilization is very common in pharmaceutical sterile manufacturing.
• At the time of performance qualification of pure steam generation system, the
sample shall be taken from each steam user point and analyzed for three
consecutive days. The purified water system must be qualified before starting
the qualification of pure steam.
• The sampling of Pure Steam:
Sampling for Bacterial Endotoxin Test and chemical tests should be done
separately. De-pyrogenated tubes or bottles should be used for taking the sample
for bacterial endotoxin test. Allow the steam to drain for minimum one minute.
33
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
Open the cap of the bottle and fill the bottle with steam condensate by holding the
bottle in the holder. Gloves should wear into the hands while sampling the pure
steam. Tighten the cap of the bottle and mark with the sampling information. If
the sample is not analyzed within 2 hours of sampling, store the sample at 2-8 °C.
• Analysis of Pure Steam: Pure steam should be analyzed for following tests.
1. Non-condensable Gases: Non-condensable gases are air and carbon dioxide
those do not condense with the steam. These are generated due to their presence
in the purified water that continuously circulates in the water distribution system.
Non-condensable gases should not be more than 3.5%.
2. Steam Dryness Value: Dry steam has more energy than the wet steam. Wet
steam has water with it and does not have heat energy as dry steam. Dryness of
steam is determined by the latent heat. Dryness of the pure steam should not be
less than 90%. High moisture content can cause the loss in energy of steam and
that may cause the longer sterilization time. 34
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
3. pH: Steam condensate is analyzed for pH value at 25 ° C. It should be
between 5-7.
4. Conductivity: Conductivity should be tested with calibrated conductivity
meter at 20 °C. Conductivity should not be more than 1.3 µS/cm.
5. Microorganisms: Steam condensate is tested for microbial contamination
using pour plate method. There should not any microbial contamination in the
steam condensate.
6. Endotoxin Test: Determine the endotoxin in the pure steam condensate and
it should not be more than 0.25 EU/ml as in water for injection.
35
COMPRESSED AIR
• Definition : Air kept under a
pressure that is greater then
atmospheric pressure.
• Compressed air serves as an
important medium to transfer of
energy in industrial process and to
operate air cylinders for
automation.
• Used as:
 Cleaning and sterilization
 Packaging
 Tablet coating
 Pneumatic conveying
36
Main parts of compressed air
system:
1. Compressor
2. Reciever’s tank
3. Air dryer
4. Moisture seperator
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
Parameters To Specify The Quality Of Compressed Air:
• Dew point
• Moisture content
• Viable count
• Oil content
• Particulate matter
DEW POINT TEMPERATURE OR SATURATION TEMPERATURE
• Temperature at which water vapour begins to condense.
• Gas of unknown water vapour concentration is passed over a temperature-controlled
surface.
• The surface is cooled until condensate forms.
• The temperature at which condensate forms is called the "Dew Point temperature”.
37
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
• Dew point temperatures in compressed air range from ambient down to -112 °F.
• Dew point temperatures in nitrogen gas range from -40°F.
Moisture content
• All atmospheric air contains some water vapour which will begin to condense into
liquid water in the compressed air or gas system.
• The condensed moisture must be removed by a separator and trap.
• Moisture in compressed air used in a manufacturing plant causes problems.
• Before test air flow rate is adjusted to 100ml/min.
• Drain the air for one minute before sampling. 38
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
• Spectroscopy methods are involved to detect the water vapour content in the gaseos
systems.
• Moisture content should be less than <0.01% for Compressed AirSystem
• Moisture content should be less than < 67ppm v/v for Nitrogen Gas System.
Particulate matter
• The impurities that nitrogen/compressed air contain contaminates the product and might
result in change of colour, taste and reduced shelf life.
• Particle purity class is provided by ISO 8573-1.
• Laser particle counter (LPC): This test can be used on-site, which is a big advantage for
troubleshooting the source of contamination. It provides a count of particles at all three
size ranges specified by ISO 8573 and can be used to measure purity down to Class 1. 39
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
• Filter collection with microscopy: This method collects particulate on a filter,
which is later examined under a microscope to provide a particle count. It is
necessary to know the pressure and flow rate for sampling to get an accurate
estimate of particulate concentrations.
• Mass concentration: This method is typically used for air of Class 6 or above. It
provides a total mass of particulate in a certain volume of air, reported in mg/m³. It
does not provide particle counts or size distribution.
• Compressed Air has the Particulate Matter to a level less than 0.02 mg/m³.
• Nitrogen has the particulate matter of < 5 mg/m³.
40
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
Oil mist testing:
• Oil may be present in a sample as an aerosol (liquid droplets suspended in
air) or a vapor. Testing methods depend on the level of contamination and
the purity class requirement for the sample.
• Heavier concentrations of oil aerosols are typically measured using a
membrane system that separates the oil from the air. Oil concentrations are
determined by weighing the amount of oil captured by the membrane.
• Oil vapor and organic solvent content testing (usually used for purity
classes 1 and 2) is usually done using gas chromatography.
• Oil content of less than 0.01 mg/m³ is allowed for compressed air and
nitrogen both. 41
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
Viable count
• Analytic testing can be used to determine what types of microbes are present and in
what quantities. Testing is usually focused on microbes that cause particular problems,
such as foodborne illnesses or respiratory diseases. These may include bacteria such as
Listeria, E.Coli and Salmonella and fungal species such as Candida and various molds
and yeasts. Microbial testing may be either qualitative or quantitative.
• Qualitative testing simply reports the presence or absence of specific microbes.
• Quantitative testing tells us the number of microbes present in a specified volume. ISO
8573-7 requires quantitative results and provides guidance on validated sampling
methods.
• Microorganisms can be identified using a variety of analytical methods, including
visual macroscopic and microscopic identification and molecular methods.
42
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
• The sample is usually cultured, using a petri dish to allow organisms to grow into a
colony. Identification may be made by looking at both macroscopic characteristics
(colony shape, color, borders, etc.) and microscopic characteristics (physiology,
Gram staining reaction, etc.).
• Identification may be completed or confirmed using molecular methods. These
methods include genetic testing and other forms of molecular analysis that
“fingerprint” organisms based on various characteristics.
• For Compressed Air System, Viable count should be <100CFU/m3 (colony
forming unit).
• For Nitrogen Gas system, Viable count should be <1CFU/m3.(colony forming unit)
43
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
BIBLIOGRAPHY
1. https://www.youtube.com/watch?v=rNGiUvVjM80
2. https://www.pharmaguideline.com/2012/07/purified-water-system-
validation.html
3. https://www.who.int/docs/default-source/medicines/norms-and-
standards/guidelines/production/trs1019-annex3-gmp-validation.pdf
4. https://www.researchgate.net/publication/40542934_Validation_Of_Pharmace
utical_Water_System_-_A_Review
5. https://fluidairedynamics.com/blogs/articles/compressed-air-testing-make-sure-
your-air-is-clean-dry-and-oil-free 44
GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ
45
THANK YOU

More Related Content

What's hot

Computer system validations
Computer system validationsComputer system validations
Computer system validations
Amruta Balekundri
 
Cfr 21 part 11
 Cfr 21 part 11 Cfr 21 part 11
Cfr 21 part 11
Ashish Chaudhari
 
GMP compliances of Audit
GMP compliances of AuditGMP compliances of Audit
GMP compliances of Audit
Megha bhise
 
Good Automated Laboratory Practices
Good Automated Laboratory PracticesGood Automated Laboratory Practices
Good Automated Laboratory Practices
Swapnil Fernandes
 
Batch Review And Batch Release.pptx
Batch Review And Batch Release.pptxBatch Review And Batch Release.pptx
Batch Review And Batch Release.pptx
AbhishekJadhav189260
 
QUALIFICATION OF HPLC , HPLC
QUALIFICATION OF HPLC , HPLCQUALIFICATION OF HPLC , HPLC
QUALIFICATION OF HPLC , HPLC
suriyapriya kamaraj
 
GLP
GLPGLP
Validation of pharaceutical water system and pure steam
Validation of pharaceutical water system and pure steamValidation of pharaceutical water system and pure steam
Validation of pharaceutical water system and pure steam
Jp Prakash
 
21 cfr part 210 and 211
21 cfr part 210 and 21121 cfr part 210 and 211
21 cfr part 210 and 211
Bhanu Chava
 
Gamp 5 overview by jaya prakash ra
Gamp 5 overview by jaya prakash raGamp 5 overview by jaya prakash ra
Gamp 5 overview by jaya prakash ra
JAYA PRAKASH VELUCHURI
 
Validation qualification
Validation qualificationValidation qualification
Validation qualification
Devipriya Viswambharan
 
Validation of hvac
Validation of hvacValidation of hvac
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
RavichandraNadagouda
 
Audits in pharma industries
Audits in pharma industriesAudits in pharma industries
Audits in pharma industries
Nikita Amane
 
GHTF
GHTFGHTF
GHTF
KDivya11
 
Computerized system validation
Computerized system validationComputerized system validation
Computerized system validation
Devipriya Viswambharan
 
cGMP Guidelines According to Schedule M
cGMP Guidelines According to Schedule McGMP Guidelines According to Schedule M
cGMP Guidelines According to Schedule M
ANKUSH JADHAV
 
The Role of Regulatory GMP Audit in Pharmaceutical Companies.
The Role of Regulatory GMP Audit in Pharmaceutical Companies.The Role of Regulatory GMP Audit in Pharmaceutical Companies.
The Role of Regulatory GMP Audit in Pharmaceutical Companies.
Jitendra Sonawane
 
Water system validation by- Akshay kakde
Water system validation by- Akshay kakdeWater system validation by- Akshay kakde
Water system validation by- Akshay kakde
Akshay Kakde
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
Sagar Savale
 

What's hot (20)

Computer system validations
Computer system validationsComputer system validations
Computer system validations
 
Cfr 21 part 11
 Cfr 21 part 11 Cfr 21 part 11
Cfr 21 part 11
 
GMP compliances of Audit
GMP compliances of AuditGMP compliances of Audit
GMP compliances of Audit
 
Good Automated Laboratory Practices
Good Automated Laboratory PracticesGood Automated Laboratory Practices
Good Automated Laboratory Practices
 
Batch Review And Batch Release.pptx
Batch Review And Batch Release.pptxBatch Review And Batch Release.pptx
Batch Review And Batch Release.pptx
 
QUALIFICATION OF HPLC , HPLC
QUALIFICATION OF HPLC , HPLCQUALIFICATION OF HPLC , HPLC
QUALIFICATION OF HPLC , HPLC
 
GLP
GLPGLP
GLP
 
Validation of pharaceutical water system and pure steam
Validation of pharaceutical water system and pure steamValidation of pharaceutical water system and pure steam
Validation of pharaceutical water system and pure steam
 
21 cfr part 210 and 211
21 cfr part 210 and 21121 cfr part 210 and 211
21 cfr part 210 and 211
 
Gamp 5 overview by jaya prakash ra
Gamp 5 overview by jaya prakash raGamp 5 overview by jaya prakash ra
Gamp 5 overview by jaya prakash ra
 
Validation qualification
Validation qualificationValidation qualification
Validation qualification
 
Validation of hvac
Validation of hvacValidation of hvac
Validation of hvac
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 
Audits in pharma industries
Audits in pharma industriesAudits in pharma industries
Audits in pharma industries
 
GHTF
GHTFGHTF
GHTF
 
Computerized system validation
Computerized system validationComputerized system validation
Computerized system validation
 
cGMP Guidelines According to Schedule M
cGMP Guidelines According to Schedule McGMP Guidelines According to Schedule M
cGMP Guidelines According to Schedule M
 
The Role of Regulatory GMP Audit in Pharmaceutical Companies.
The Role of Regulatory GMP Audit in Pharmaceutical Companies.The Role of Regulatory GMP Audit in Pharmaceutical Companies.
The Role of Regulatory GMP Audit in Pharmaceutical Companies.
 
Water system validation by- Akshay kakde
Water system validation by- Akshay kakdeWater system validation by- Akshay kakde
Water system validation by- Akshay kakde
 
Cleaning validation
Cleaning validationCleaning validation
Cleaning validation
 

Similar to VALIDATION OF UTILITY ( HVAC , CLEANING VALIDATION,)

HVAC validation.pdf
HVAC validation.pdfHVAC validation.pdf
HVAC validation.pdf
Dr.Noor Nabi Lodhi
 
HVAC
HVACHVAC
Performance qualification of High performance liquid chromatography
Performance qualification of High performance liquid chromatography Performance qualification of High performance liquid chromatography
Performance qualification of High performance liquid chromatography
Atchaya Thalapathy
 
Sterile Drug Products produced by cGMP
Sterile Drug Products produced by cGMPSterile Drug Products produced by cGMP
Sterile Drug Products produced by cGMP
Iqra Shafeeq
 
Notes on validation
Notes on validationNotes on validation
Notes on validation
Dhruv Desai
 
HVAC training 2020
HVAC  training 2020HVAC  training 2020
HVAC training 2020
Time Pharmaceutical P.Ltd
 
STERILE Prpn.pptx
STERILE Prpn.pptxSTERILE Prpn.pptx
STERILE Prpn.pptx
Anuradha Mishra
 
HVAC system
 HVAC system HVAC system
HVAC system
Henisha Patel
 
STANDARD OPERATING PROCEDURES FOR PARENTERAL DOSAGE FORM PREPARATION
STANDARD OPERATING PROCEDURES FOR PARENTERAL DOSAGE FORM PREPARATIONSTANDARD OPERATING PROCEDURES FOR PARENTERAL DOSAGE FORM PREPARATION
STANDARD OPERATING PROCEDURES FOR PARENTERAL DOSAGE FORM PREPARATION
AVIJIT BAKSHI
 
validation HPLC and HPTLC D-A star.pptx
validation HPLC and HPTLC D-A star.pptxvalidation HPLC and HPTLC D-A star.pptx
validation HPLC and HPTLC D-A star.pptx
DevaPundkar
 
Validation of hvac
Validation of hvacValidation of hvac
Validation of hvac
AshwinDigarse
 
Sterile process validation
Sterile process validationSterile process validation
Sterile process validation
Sagar Savale
 
Control on Cleanroom Environmental Monitoring (Pharmaceutical)
Control on Cleanroom Environmental Monitoring (Pharmaceutical)Control on Cleanroom Environmental Monitoring (Pharmaceutical)
Control on Cleanroom Environmental Monitoring (Pharmaceutical)
Srinath Sasidharan
 
Qualification of High Performance Liquid Chromatography(HPLC)
Qualification of High Performance Liquid Chromatography(HPLC)Qualification of High Performance Liquid Chromatography(HPLC)
Qualification of High Performance Liquid Chromatography(HPLC)
PRAVADA
 
One slider for qualification and validation of depyrogenation and sterilizati...
One slider for qualification and validation of depyrogenation and sterilizati...One slider for qualification and validation of depyrogenation and sterilizati...
One slider for qualification and validation of depyrogenation and sterilizati...
Palash Das
 
Presentation on HVAC
Presentation on HVAC Presentation on HVAC
Presentation on HVAC
Avishek Sen Sarma
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments
PUNEET NIRMAL
 
Validation utility system
Validation utility systemValidation utility system
Validation utility system
ArpitSuralkar
 
Pharmaceutical Engineering Downflow Booth Testing
Pharmaceutical Engineering Downflow Booth TestingPharmaceutical Engineering Downflow Booth Testing
Pharmaceutical Engineering Downflow Booth TestingEric Ferguson
 

Similar to VALIDATION OF UTILITY ( HVAC , CLEANING VALIDATION,) (20)

HVAC validation.pdf
HVAC validation.pdfHVAC validation.pdf
HVAC validation.pdf
 
HVAC
HVACHVAC
HVAC
 
Performance qualification of High performance liquid chromatography
Performance qualification of High performance liquid chromatography Performance qualification of High performance liquid chromatography
Performance qualification of High performance liquid chromatography
 
Sterile Drug Products produced by cGMP
Sterile Drug Products produced by cGMPSterile Drug Products produced by cGMP
Sterile Drug Products produced by cGMP
 
Notes on validation
Notes on validationNotes on validation
Notes on validation
 
HVAC training 2020
HVAC  training 2020HVAC  training 2020
HVAC training 2020
 
STERILE Prpn.pptx
STERILE Prpn.pptxSTERILE Prpn.pptx
STERILE Prpn.pptx
 
HVAC system
 HVAC system HVAC system
HVAC system
 
HVAC PPT.pptx
HVAC PPT.pptxHVAC PPT.pptx
HVAC PPT.pptx
 
STANDARD OPERATING PROCEDURES FOR PARENTERAL DOSAGE FORM PREPARATION
STANDARD OPERATING PROCEDURES FOR PARENTERAL DOSAGE FORM PREPARATIONSTANDARD OPERATING PROCEDURES FOR PARENTERAL DOSAGE FORM PREPARATION
STANDARD OPERATING PROCEDURES FOR PARENTERAL DOSAGE FORM PREPARATION
 
validation HPLC and HPTLC D-A star.pptx
validation HPLC and HPTLC D-A star.pptxvalidation HPLC and HPTLC D-A star.pptx
validation HPLC and HPTLC D-A star.pptx
 
Validation of hvac
Validation of hvacValidation of hvac
Validation of hvac
 
Sterile process validation
Sterile process validationSterile process validation
Sterile process validation
 
Control on Cleanroom Environmental Monitoring (Pharmaceutical)
Control on Cleanroom Environmental Monitoring (Pharmaceutical)Control on Cleanroom Environmental Monitoring (Pharmaceutical)
Control on Cleanroom Environmental Monitoring (Pharmaceutical)
 
Qualification of High Performance Liquid Chromatography(HPLC)
Qualification of High Performance Liquid Chromatography(HPLC)Qualification of High Performance Liquid Chromatography(HPLC)
Qualification of High Performance Liquid Chromatography(HPLC)
 
One slider for qualification and validation of depyrogenation and sterilizati...
One slider for qualification and validation of depyrogenation and sterilizati...One slider for qualification and validation of depyrogenation and sterilizati...
One slider for qualification and validation of depyrogenation and sterilizati...
 
Presentation on HVAC
Presentation on HVAC Presentation on HVAC
Presentation on HVAC
 
Qualification of analytical instruments
Qualification of analytical instruments Qualification of analytical instruments
Qualification of analytical instruments
 
Validation utility system
Validation utility systemValidation utility system
Validation utility system
 
Pharmaceutical Engineering Downflow Booth Testing
Pharmaceutical Engineering Downflow Booth TestingPharmaceutical Engineering Downflow Booth Testing
Pharmaceutical Engineering Downflow Booth Testing
 

Recently uploaded

The approach at University of Liverpool.pptx
The approach at University of Liverpool.pptxThe approach at University of Liverpool.pptx
The approach at University of Liverpool.pptx
Jisc
 
Supporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptxSupporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptx
Jisc
 
Students, digital devices and success - Andreas Schleicher - 27 May 2024..pptx
Students, digital devices and success - Andreas Schleicher - 27 May 2024..pptxStudents, digital devices and success - Andreas Schleicher - 27 May 2024..pptx
Students, digital devices and success - Andreas Schleicher - 27 May 2024..pptx
EduSkills OECD
 
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
Nguyen Thanh Tu Collection
 
Overview on Edible Vaccine: Pros & Cons with Mechanism
Overview on Edible Vaccine: Pros & Cons with MechanismOverview on Edible Vaccine: Pros & Cons with Mechanism
Overview on Edible Vaccine: Pros & Cons with Mechanism
DeeptiGupta154
 
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
siemaillard
 
How to Break the cycle of negative Thoughts
How to Break the cycle of negative ThoughtsHow to Break the cycle of negative Thoughts
How to Break the cycle of negative Thoughts
Col Mukteshwar Prasad
 
Basic phrases for greeting and assisting costumers
Basic phrases for greeting and assisting costumersBasic phrases for greeting and assisting costumers
Basic phrases for greeting and assisting costumers
PedroFerreira53928
 
How to Create Map Views in the Odoo 17 ERP
How to Create Map Views in the Odoo 17 ERPHow to Create Map Views in the Odoo 17 ERP
How to Create Map Views in the Odoo 17 ERP
Celine George
 
How libraries can support authors with open access requirements for UKRI fund...
How libraries can support authors with open access requirements for UKRI fund...How libraries can support authors with open access requirements for UKRI fund...
How libraries can support authors with open access requirements for UKRI fund...
Jisc
 
Additional Benefits for Employee Website.pdf
Additional Benefits for Employee Website.pdfAdditional Benefits for Employee Website.pdf
Additional Benefits for Employee Website.pdf
joachimlavalley1
 
How to Make a Field invisible in Odoo 17
How to Make a Field invisible in Odoo 17How to Make a Field invisible in Odoo 17
How to Make a Field invisible in Odoo 17
Celine George
 
Model Attribute Check Company Auto Property
Model Attribute  Check Company Auto PropertyModel Attribute  Check Company Auto Property
Model Attribute Check Company Auto Property
Celine George
 
Cambridge International AS A Level Biology Coursebook - EBook (MaryFosbery J...
Cambridge International AS  A Level Biology Coursebook - EBook (MaryFosbery J...Cambridge International AS  A Level Biology Coursebook - EBook (MaryFosbery J...
Cambridge International AS A Level Biology Coursebook - EBook (MaryFosbery J...
AzmatAli747758
 
Operation Blue Star - Saka Neela Tara
Operation Blue Star   -  Saka Neela TaraOperation Blue Star   -  Saka Neela Tara
Operation Blue Star - Saka Neela Tara
Balvir Singh
 
Chapter 3 - Islamic Banking Products and Services.pptx
Chapter 3 - Islamic Banking Products and Services.pptxChapter 3 - Islamic Banking Products and Services.pptx
Chapter 3 - Islamic Banking Products and Services.pptx
Mohd Adib Abd Muin, Senior Lecturer at Universiti Utara Malaysia
 
ESC Beyond Borders _From EU to You_ InfoPack general.pdf
ESC Beyond Borders _From EU to You_ InfoPack general.pdfESC Beyond Borders _From EU to You_ InfoPack general.pdf
ESC Beyond Borders _From EU to You_ InfoPack general.pdf
Fundacja Rozwoju Społeczeństwa Przedsiębiorczego
 
Synthetic Fiber Construction in lab .pptx
Synthetic Fiber Construction in lab .pptxSynthetic Fiber Construction in lab .pptx
Synthetic Fiber Construction in lab .pptx
Pavel ( NSTU)
 
special B.ed 2nd year old paper_20240531.pdf
special B.ed 2nd year old paper_20240531.pdfspecial B.ed 2nd year old paper_20240531.pdf
special B.ed 2nd year old paper_20240531.pdf
Special education needs
 
Home assignment II on Spectroscopy 2024 Answers.pdf
Home assignment II on Spectroscopy 2024 Answers.pdfHome assignment II on Spectroscopy 2024 Answers.pdf
Home assignment II on Spectroscopy 2024 Answers.pdf
Tamralipta Mahavidyalaya
 

Recently uploaded (20)

The approach at University of Liverpool.pptx
The approach at University of Liverpool.pptxThe approach at University of Liverpool.pptx
The approach at University of Liverpool.pptx
 
Supporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptxSupporting (UKRI) OA monographs at Salford.pptx
Supporting (UKRI) OA monographs at Salford.pptx
 
Students, digital devices and success - Andreas Schleicher - 27 May 2024..pptx
Students, digital devices and success - Andreas Schleicher - 27 May 2024..pptxStudents, digital devices and success - Andreas Schleicher - 27 May 2024..pptx
Students, digital devices and success - Andreas Schleicher - 27 May 2024..pptx
 
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
GIÁO ÁN DẠY THÊM (KẾ HOẠCH BÀI BUỔI 2) - TIẾNG ANH 8 GLOBAL SUCCESS (2 CỘT) N...
 
Overview on Edible Vaccine: Pros & Cons with Mechanism
Overview on Edible Vaccine: Pros & Cons with MechanismOverview on Edible Vaccine: Pros & Cons with Mechanism
Overview on Edible Vaccine: Pros & Cons with Mechanism
 
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaaa
 
How to Break the cycle of negative Thoughts
How to Break the cycle of negative ThoughtsHow to Break the cycle of negative Thoughts
How to Break the cycle of negative Thoughts
 
Basic phrases for greeting and assisting costumers
Basic phrases for greeting and assisting costumersBasic phrases for greeting and assisting costumers
Basic phrases for greeting and assisting costumers
 
How to Create Map Views in the Odoo 17 ERP
How to Create Map Views in the Odoo 17 ERPHow to Create Map Views in the Odoo 17 ERP
How to Create Map Views in the Odoo 17 ERP
 
How libraries can support authors with open access requirements for UKRI fund...
How libraries can support authors with open access requirements for UKRI fund...How libraries can support authors with open access requirements for UKRI fund...
How libraries can support authors with open access requirements for UKRI fund...
 
Additional Benefits for Employee Website.pdf
Additional Benefits for Employee Website.pdfAdditional Benefits for Employee Website.pdf
Additional Benefits for Employee Website.pdf
 
How to Make a Field invisible in Odoo 17
How to Make a Field invisible in Odoo 17How to Make a Field invisible in Odoo 17
How to Make a Field invisible in Odoo 17
 
Model Attribute Check Company Auto Property
Model Attribute  Check Company Auto PropertyModel Attribute  Check Company Auto Property
Model Attribute Check Company Auto Property
 
Cambridge International AS A Level Biology Coursebook - EBook (MaryFosbery J...
Cambridge International AS  A Level Biology Coursebook - EBook (MaryFosbery J...Cambridge International AS  A Level Biology Coursebook - EBook (MaryFosbery J...
Cambridge International AS A Level Biology Coursebook - EBook (MaryFosbery J...
 
Operation Blue Star - Saka Neela Tara
Operation Blue Star   -  Saka Neela TaraOperation Blue Star   -  Saka Neela Tara
Operation Blue Star - Saka Neela Tara
 
Chapter 3 - Islamic Banking Products and Services.pptx
Chapter 3 - Islamic Banking Products and Services.pptxChapter 3 - Islamic Banking Products and Services.pptx
Chapter 3 - Islamic Banking Products and Services.pptx
 
ESC Beyond Borders _From EU to You_ InfoPack general.pdf
ESC Beyond Borders _From EU to You_ InfoPack general.pdfESC Beyond Borders _From EU to You_ InfoPack general.pdf
ESC Beyond Borders _From EU to You_ InfoPack general.pdf
 
Synthetic Fiber Construction in lab .pptx
Synthetic Fiber Construction in lab .pptxSynthetic Fiber Construction in lab .pptx
Synthetic Fiber Construction in lab .pptx
 
special B.ed 2nd year old paper_20240531.pdf
special B.ed 2nd year old paper_20240531.pdfspecial B.ed 2nd year old paper_20240531.pdf
special B.ed 2nd year old paper_20240531.pdf
 
Home assignment II on Spectroscopy 2024 Answers.pdf
Home assignment II on Spectroscopy 2024 Answers.pdfHome assignment II on Spectroscopy 2024 Answers.pdf
Home assignment II on Spectroscopy 2024 Answers.pdf
 

VALIDATION OF UTILITY ( HVAC , CLEANING VALIDATION,)

  • 1. GOOD REGULATORY PRACTICES Validation Of Utility Submitted by: Sonal Sharma Submitted to: Dr. Navneet Sharma 1 Amity Institute of Pharmacy
  • 2. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ TABLE OF CONTENT- 1. Introduction 2. Validation of HVAC system 3. Cleaning validation 4. Validation of water system 5. Validation of steam system 6. Validation of compressed air systems 7. bibiliography 2
  • 3. Validation of HVAC system • HVAC stands for heating ventilation and air conditioning system. • The HVAC system also known as the air handling unit (AHU) system are placed in the clean rooms for thermal control. 3
  • 4. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ Test performed for validation of HVAC system: 1. ACPH 2. Filter integrity test 3. Non-viable particle count 4. Air flow visualization 5. Recovery period study 6. Temperature / relative humidity/ differential pressure monitoring 7. Air borne viable particle monitoring 4
  • 5. Air changes per hour • If the supplied air by HVAC per hour is equal to the volume of the room/ area, then it will be one ACPH. • To determine the average filter face velocity and uniformity, and the average room airflow velocity and uniformity within a clean room. • Test apparatus: ANEMOMETER • Anemometer is placed approximately 6 inches beneath the filter surface and take velocity of filter from 5 corners. 5 V1 V2 V3 V4 V5
  • 6. 6 • Get average velocity ( V ) = V1 + V2 + V3 + V4 + V5 5 • Calculate the area of filter (A) : L ( Length of filter) x W ( width of filter ) • Calculate total air volume supplied ( T ): A (in square feet ) x V ( in minutes ) • Calculate total volume of room : L (Length) x W (width) x H ( height in feet) • Calculate ACPH by: T x 60 Room volume • As per ISO 14644 recommended ACPH for clean rooms are as: ISO CLASS ACPH ISO 5 240 – 600 ISO 6 150 – 240 ISO 7 60 – 150 ISO 8 5 - 60
  • 7. Filter integrity test: • Filter integrity test is carried out to assure the integrity of HEPA (high efficiency particle airflow system) filters against filter failure due to damage while transportation, installation or operation. • Test apparatus: AEROSOL GENERATOR & PHOTOMETER 7 • Through aerosol generator we generate particle of PAO (Poly alpha Olefin) of 0.5 to 0.7 micron and supply it to AHU. • Scan the downstream aerosol concenteration by keeping photometer at HEPA surface. • Scanning rate should be approx. 10 CFM (cubic feet/ minutes) • Acceptance criteria: NMT(not more then) 0.01% of upstream aerosol concenteration.
  • 8. Non - viable particle count • To determine the concentration of non-viable airborne particles, equal to and greater than the specified sizes, at designated sampling locations. • Test apparatus: LSAPC ( light scattering air born particle counter ) 8 • Prior to testing all relevant aspects of the clean room that contributes to its integrity are complete and function in accordance with its performance qualification. • Sampling shall be done at rest condition only. • Sampling location shall determine as per total area of clean room. • For areas larger then 1000 m² , NL = 27 x A NL = Minimum number of locations 1000 A = area of clean room in m². • Sampling location must be evenly distributed to entire room. • Through particle counter get particle size of 0.5 & 5.0µ ( per meter cube of air).
  • 9. Area in Number of sampling locations Area in Number of sampling locations 2 1 32 8 4 2 36 9 6 3 52 10 8 4 56 11 10 5 64 12 24 6 68 13 28 7 72 14 9 Sampling location chart: The minimum sample volume should be atleast 2 liters. Air with a minimum sampling time for 1 minute for each sample at each location.
  • 10. Air flow visualization • To confirm that the air flow direction and its Uniformity confirm to the design and performance specification. • Test Apparatus: Video camera, Glycerin with Rose water, Fogger (Previously TTC (Titanium Tetra Chloride) was used to produce smoke/fog). • Generate fog with the help of fogger and expose the fog at supply end of the system. 10 • Observe the smoke pattern and ensure that, it should be unidirectional, smoke must move from supply to return duct. & from positive pressure area to negative pressure area. • Picturise the direction of fog by the help of a Video.
  • 11. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ Recovery Period Study • This test is performed to determine the amount of time necessary for the clean room and its system to reach a steady state in means of environmental condition (temperature and RH) & cleanliness level after a brief particle generation event within the clean space. • Test Apparatus: PARTICLE COUNTER & DIGITAL HYGROMETER • To verify the recovery of particles present in clean room generate and introduce particles of size 0.5 & 5.0 micron through HVAC system by PAO and evaluate the recovery through particle counter after starting the HVAC system. • To verify recovery of environmental condition, increase the temperature and humidity of area by spraying hot water upto 75%.After starting the HVAC system, monitor the humidity and temperature to get stabilized and area comes within specified 11
  • 12. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ Temperature/Relative Humidity/ Differential Pressure Monitoring • To demonstrate the capability of the HVAC System to consistently maintain Differential Pressures, Temperatures and Relative Humidity in the control areas. • After completion of all said activities in qualification, area should monitor for Temperature, Relative Humidity and Pressure differential for at least for 8 hrs. with interval of 1 hr. ± 15 minute. • This should be lead by monitoring the same on daily basis, at least twice a day. 12
  • 13. Air Borne Viable Particle Count • To check and ensure that the controlled area meets the desired level of area classification with respect to viable counts. • Test Apparatus: PRE- INCUBATED MEDIA PLATES. • Viable particle count monitoring- shall done for at least 3 days for qualification and as per schedule after that. 13
  • 14. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ CLEANING VALIDATION • Cleaning validation is documented evidence that an approved cleaning procedure is being used that will make equipment suitable for processing of required medicinal product. • Cleaning validation should be performed in order to confirm the effectiveness of any cleaning procedure for all product contact equipment. CLEANING PRINCIPLE : Pharmaceutical products can be contaminated by other pharmaceutical products or APIs, by cleaning agents, by micro- organisms or by other material (e.g. air-borne particles, dust, lubricants, raw materials, intermediates, auxiliaries etc.). In many cases, the same equipment may be used for processing different products. To avoid contamination of the following pharmaceutical product, adequate cleaning procedures are essential. 14
  • 15. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ TYPE OF CLEANING : • TYPE A (MINOR): This type of cleaning take place between two batches of same product. For minor cleaning, cleaning validation is not required, since cross contamination is not an issue. • TYPE B (MAJOR): This type of cleaning take place between two different products. HARD TO CLEAN AREA: • Area/spots in any equipment which were really hard to reach and clean define as hard to clean area. • Hard to clean area should consider as worst case and must capture in sampling plan. 15
  • 16. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ SELECTION OF WORST CASE THROUGH PRODUCT MATRIX • The Worst-Case product for cleaning should be selected based on the APIs used in the formulation. • Following approach should be considered for selection of worst case molecule: 1. Aqueous Solubility (Lower aqueous solubility = higher worst case) 2. Therapeutic Dose (Lower Therapeutic dose = higher worst case) 3. LD50 (Lower LD50 = higher worst case) 16
  • 17. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ S. no. Product name Active material Strength ( in mg) Solubility in water LDD (in mg) Batch size (liters) 1 ABC XXXX 50 Slightly soluble in water 400 3000 XXXX 125 Practically insoluble in water 500 2 DEF XXXX 50 Slightly soluble in water 400 2000 XXXX 125 Practically insoluble in water 500 3 GHI XXXX 50 Slightly soluble in water 400 5000 XXXX 125 Practically in soluble in water 500 4 JKL XXXX 4 Freely soluble in water 10 3000 XXXX 2.5 Freely soluble in water 700 5 MNO XXXX 20 Practically insoluble in water 100 5000 XXXX 10 Insoluble in water 200 XXXX 5 Practically insoluble in water 40 6 PQR XXXX 540 Insoluble in water 2000 5000 17 PRODUCT MATRIX
  • 18. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ Sr. no. Equipment Make Equipment id Material of construction Surface area (cm²) 10% Extra of Total Surface area (cm²) 1 Manufacturing vessel 2 Storage vessel 3 Scoops 4 Filling & sealing machine 5 Conveyor belt 6 Material transfer pipes 7 Filter press total 18 EQUIPMENT TRIAL
  • 19. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ SAMPLING CRITERIA • Before sampling the surface would be clean visually. • SWAB (Direct Surface Sampling): When samples were taken through direct surface of the equipment with the help of swab sticks. • Rinse Sample: Rinse sample shall be collected from the hard-to-access locations of the equipments. Equipment surface shall be rinsed with the given quantity of rinse volume with purified water and the sample shall be collected from the rinsate. 19
  • 20. SEQUENCE OF SAMPLING 1. Microbial Swab 2. Chemical Swab 3. Chemical Rinse METHODOLOGY TO GET SWAB SAMPLE • Swab sample are taken with the help of swab sticks, which are initially soaked in Purified water and after sampling put into mother solvent. 20 • To get swab sample select an area of 5x5 or 10x10 cm with the help of sampler, and apply swab stick in above shown manner.
  • 21. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ CALCULATION (NOEL) NO OBSERVED EFFECT LEVEL • Is the amount of drug/substance in mg that does not have any effect on the human health. • NOEL is calculated by using LD 50 (Lethal Dose 50 is the amount of a material, given all at once, which cause the death of 50% of a group of animals) of the drug. NOEL = (LD 50 × 70 kg) 2000 Where: LD 50 = Lethal Dose, 70 kg = Average Adult weight & 2000 is a constant. NOEL is determined to calculate the MACO in cleaning validation. 21
  • 22. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ CALCULATION (MACO) MAXIMUM ALLOWABLE CARRY OVER • Mathematically calculated quantity of residue from a previous product when carried out over into a different product that can cause potential harm to the patient. • Three criteria to calculate MACO: 1. Therapeutic Daily Dose criteria 2. NOEL criteria 3. PPM criteria 22
  • 23. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ • THERAPEUTIC DAILY DOSE CRITERIA: MACO = TDD(p) × MBS(n) SF x TDD(n) Where: TDD(p) = Therapeutic Daily Dose of previous product MBS(n) = Minimum Batch Size of next product TDD(n) = Therapeutic Daily Dose of next product SF = Safety Factor (1000 for Oral, 100 for external Preparation,& 10 for sterile preparations) 23
  • 24. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ • NOEL CRITERIA: MACO = NOEL (P) × MBS (n) SF × TDD(n) Where: NOEL(p) = No Observed Effect Level of previous product MBS(n) = Minimum Batch Size of next product TDD(n) = Therapeutic Daily Dose of next product SF = Safety Factor (1000 for Oral, 100 for external Preparation,& 10 for sterile preparations) 24
  • 25. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ • PPM CRITERIA: If MACO calculations result in unacceptably high or irrelevant carryover figures, or toxicological data for intermediates are not known, the approach of a general limit may be suitable. User preferable may choose to have such an upper limit as a policy. The general limit is often set as an upper limit for the maximum concentration (MAXCONC) of a contaminating substance in a subsequent batch. MACOppm = MAXCONCs x MBS Where: MAXCONC = General limit for maximum allowed concentration (kg/kg or ppm) of"previous" substance in the next batch. MBS = Minimum batch size for the next products),(where MACO can end up). 25
  • 26. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ MICROBIAL/ENDOTOXINS CRITERIA • Samples should be analyzed for microbial aspects in all the three runs during cleaning validation. The following should be acceptance criteria for Microbiological monitoring during cleaning validation / verification: 26 Parameter Non sterile formulation equipments Sterile formulation equipment Total CFU - Swab NMT 50 CFU /25cm² NMT 5 CFU /25cm²
  • 27. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ ANALYTICAL METHOD VALIDATION Any instrumental analytical procedures used to analyze cleaning validation samples need to be specified and sufficiently sensitive to determine the low levels of residues typically found in samples.The methods used to analyze samples that allow the equipment to be released for manufacture of another product shall be validated to ensure that it meets following requirements, but not limited to: • Specificity • Linearity • Precision • Robustness • LOQ and LOD Determination • Sample recovery • Stability of analytical solution 27
  • 28. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ HOLD TIME STUDIES • The objective of hold time study is for establishing time limit between equipment cleaning and reuse. It is done to ensure that the equipment remains clean till the nest use. • Equipment Holding Studies prior to cleaning/ Un cleaned equipment hold time study: The interval between the end of production and the beginning of the cleaning process shall be established through equipment holding studies prior to cleaning. • Cleaned Equipment Hold Time Study : The interval between the initiation of production and the completion of the cleaning process shall be established. 28
  • 29. Water system validation • Water system validation is very important in pharmaceuticals. Water is used in the manufacturing process in very large quantities during different stages of manufacturing. During manufacturing, water is continuously generated and distributed and can not be tested and analyzed for each and every quantity used in production. • To provide consistent quality standards, It needs to validate for treatment, generation, storage, and distribution. So, there is a total of 3 phases for water system validation. 29
  • 30. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ 3 Types of Water System Validation • Water System Validation Phase 1 • Water System Validation Phase 2 • Water System Validation Phase 3 PHASE I: INVESTIGATION PHASE • Duration: Phase 1 typically lasts 2-4 weeks. • Focus: It primarily focuses on rigorous monitoring and testing of the system to ensure it operates continuously without any failures or deviations from expected performance. • Goals: The main goals are to establish initial operating procedures, verify the quality of incoming water, develop maintenance procedures, and demonstrate that the system can consistently produce the required quantity and quality of water. 30
  • 31. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ PHASE II: SHORT-TERM CONTROL • Duration: Phase 2 also lasts 2-4 weeks. • Focus: This phase continues intensive monitoring, but it places a stronger emphasis on refining and finalizing the standard operating procedures (SOPs) established in Phase 1. It also allows for the use of water for manufacturing purposes. • Goals: Phase 2 aims to demonstrate consistent operation within established ranges, ensure the system consistently produces the required water quality and quantity when following the refined SOPs, and verify that the system works effectively in actual manufacturing processes. 31
  • 32. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ Phase III: Long-Term Control • Duration: Phase 3 is the longest, running for approximately one year. • Focus: Unlike the previous phases, Phase 3 is more about extended and long- term performance. It takes into account seasonal variations and involves less frequent testing. • Goals: The primary objectives in Phase 3 are to prove that the system can reliably perform over an extended period, evaluate its performance in different seasons, and establish a routine maintenance plan based on the lessons learned in the earlier phases. Sampling and testing become less frequent but are based on established procedures from Phases 1 and 2. 32
  • 33. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ VALIDATION OF STEAM SYSTEMS • Pure steam is used in various operations in pharmaceuticals but its use in sterilization is very common in pharmaceutical sterile manufacturing. • At the time of performance qualification of pure steam generation system, the sample shall be taken from each steam user point and analyzed for three consecutive days. The purified water system must be qualified before starting the qualification of pure steam. • The sampling of Pure Steam: Sampling for Bacterial Endotoxin Test and chemical tests should be done separately. De-pyrogenated tubes or bottles should be used for taking the sample for bacterial endotoxin test. Allow the steam to drain for minimum one minute. 33
  • 34. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ Open the cap of the bottle and fill the bottle with steam condensate by holding the bottle in the holder. Gloves should wear into the hands while sampling the pure steam. Tighten the cap of the bottle and mark with the sampling information. If the sample is not analyzed within 2 hours of sampling, store the sample at 2-8 °C. • Analysis of Pure Steam: Pure steam should be analyzed for following tests. 1. Non-condensable Gases: Non-condensable gases are air and carbon dioxide those do not condense with the steam. These are generated due to their presence in the purified water that continuously circulates in the water distribution system. Non-condensable gases should not be more than 3.5%. 2. Steam Dryness Value: Dry steam has more energy than the wet steam. Wet steam has water with it and does not have heat energy as dry steam. Dryness of steam is determined by the latent heat. Dryness of the pure steam should not be less than 90%. High moisture content can cause the loss in energy of steam and that may cause the longer sterilization time. 34
  • 35. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ 3. pH: Steam condensate is analyzed for pH value at 25 ° C. It should be between 5-7. 4. Conductivity: Conductivity should be tested with calibrated conductivity meter at 20 °C. Conductivity should not be more than 1.3 µS/cm. 5. Microorganisms: Steam condensate is tested for microbial contamination using pour plate method. There should not any microbial contamination in the steam condensate. 6. Endotoxin Test: Determine the endotoxin in the pure steam condensate and it should not be more than 0.25 EU/ml as in water for injection. 35
  • 36. COMPRESSED AIR • Definition : Air kept under a pressure that is greater then atmospheric pressure. • Compressed air serves as an important medium to transfer of energy in industrial process and to operate air cylinders for automation. • Used as:  Cleaning and sterilization  Packaging  Tablet coating  Pneumatic conveying 36 Main parts of compressed air system: 1. Compressor 2. Reciever’s tank 3. Air dryer 4. Moisture seperator
  • 37. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ Parameters To Specify The Quality Of Compressed Air: • Dew point • Moisture content • Viable count • Oil content • Particulate matter DEW POINT TEMPERATURE OR SATURATION TEMPERATURE • Temperature at which water vapour begins to condense. • Gas of unknown water vapour concentration is passed over a temperature-controlled surface. • The surface is cooled until condensate forms. • The temperature at which condensate forms is called the "Dew Point temperature”. 37
  • 38. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ • Dew point temperatures in compressed air range from ambient down to -112 °F. • Dew point temperatures in nitrogen gas range from -40°F. Moisture content • All atmospheric air contains some water vapour which will begin to condense into liquid water in the compressed air or gas system. • The condensed moisture must be removed by a separator and trap. • Moisture in compressed air used in a manufacturing plant causes problems. • Before test air flow rate is adjusted to 100ml/min. • Drain the air for one minute before sampling. 38
  • 39. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ • Spectroscopy methods are involved to detect the water vapour content in the gaseos systems. • Moisture content should be less than <0.01% for Compressed AirSystem • Moisture content should be less than < 67ppm v/v for Nitrogen Gas System. Particulate matter • The impurities that nitrogen/compressed air contain contaminates the product and might result in change of colour, taste and reduced shelf life. • Particle purity class is provided by ISO 8573-1. • Laser particle counter (LPC): This test can be used on-site, which is a big advantage for troubleshooting the source of contamination. It provides a count of particles at all three size ranges specified by ISO 8573 and can be used to measure purity down to Class 1. 39
  • 40. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ • Filter collection with microscopy: This method collects particulate on a filter, which is later examined under a microscope to provide a particle count. It is necessary to know the pressure and flow rate for sampling to get an accurate estimate of particulate concentrations. • Mass concentration: This method is typically used for air of Class 6 or above. It provides a total mass of particulate in a certain volume of air, reported in mg/m³. It does not provide particle counts or size distribution. • Compressed Air has the Particulate Matter to a level less than 0.02 mg/m³. • Nitrogen has the particulate matter of < 5 mg/m³. 40
  • 41. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ Oil mist testing: • Oil may be present in a sample as an aerosol (liquid droplets suspended in air) or a vapor. Testing methods depend on the level of contamination and the purity class requirement for the sample. • Heavier concentrations of oil aerosols are typically measured using a membrane system that separates the oil from the air. Oil concentrations are determined by weighing the amount of oil captured by the membrane. • Oil vapor and organic solvent content testing (usually used for purity classes 1 and 2) is usually done using gas chromatography. • Oil content of less than 0.01 mg/m³ is allowed for compressed air and nitrogen both. 41
  • 42. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ Viable count • Analytic testing can be used to determine what types of microbes are present and in what quantities. Testing is usually focused on microbes that cause particular problems, such as foodborne illnesses or respiratory diseases. These may include bacteria such as Listeria, E.Coli and Salmonella and fungal species such as Candida and various molds and yeasts. Microbial testing may be either qualitative or quantitative. • Qualitative testing simply reports the presence or absence of specific microbes. • Quantitative testing tells us the number of microbes present in a specified volume. ISO 8573-7 requires quantitative results and provides guidance on validated sampling methods. • Microorganisms can be identified using a variety of analytical methods, including visual macroscopic and microscopic identification and molecular methods. 42
  • 43. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ • The sample is usually cultured, using a petri dish to allow organisms to grow into a colony. Identification may be made by looking at both macroscopic characteristics (colony shape, color, borders, etc.) and microscopic characteristics (physiology, Gram staining reaction, etc.). • Identification may be completed or confirmed using molecular methods. These methods include genetic testing and other forms of molecular analysis that “fingerprint” organisms based on various characteristics. • For Compressed Air System, Viable count should be <100CFU/m3 (colony forming unit). • For Nitrogen Gas system, Viable count should be <1CFU/m3.(colony forming unit) 43
  • 44. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ BIBLIOGRAPHY 1. https://www.youtube.com/watch?v=rNGiUvVjM80 2. https://www.pharmaguideline.com/2012/07/purified-water-system- validation.html 3. https://www.who.int/docs/default-source/medicines/norms-and- standards/guidelines/production/trs1019-annex3-gmp-validation.pdf 4. https://www.researchgate.net/publication/40542934_Validation_Of_Pharmace utical_Water_System_-_A_Review 5. https://fluidairedynamics.com/blogs/articles/compressed-air-testing-make-sure- your-air-is-clean-dry-and-oil-free 44
  • 45. GRADE ‘A’ ACCREDITED BY NAAC AUUP HQ 45 THANK YOU