This document discusses ultrasound physics and image optimization. It begins with an overview of basic principles, instrumentation, and image optimization techniques. It then describes how ultrasound works, including the generation of sound waves, their interaction with tissues through reflection, refraction, interference and absorption. This determines image quality. Instrumentation components like the transducer, transmitter, receiver and display are explained. Factors affecting the ultrasound beam like frequency, aperture, pulse length and coupling medium are also covered.
Ultrasound Physics Made easy - By Dr Chandni WadhwaniChandni Wadhwani
History of ultrasound, Principle of Ultrasound.
Ultrasound wave and its interactions
Ultrasound machine and its parts, Image display, Artifacts and their clinical importance
what is Doppler ultrasound, Elastography and Recent advances in field of ultrasound.
Safety issues in ultrasound.
Ultrasound Physics Made easy - By Dr Chandni WadhwaniChandni Wadhwani
History of ultrasound, Principle of Ultrasound.
Ultrasound wave and its interactions
Ultrasound machine and its parts, Image display, Artifacts and their clinical importance
what is Doppler ultrasound, Elastography and Recent advances in field of ultrasound.
Safety issues in ultrasound.
An overview of Doppler Effect in Ultrasonography - the medical imaging of the body using Ultrasound.
Includes Colour Doppler, Power Doppler, Spectral Doppler, Continuous Wave Doppler, Pulsed Wave Doppler, and comparisons with other Radiographic imaging modalities.
Training Material inherited form Philips Basics of Ultrasonography. Covers the fundamentals of Ultrasound Waveform, Piezoelectric Effect, Phased Echo Concept, Goal of Ultrasound, Ultrasound Image Construction process, Types of Resolution, Probe Internals, The Doppler Effect, Spectrum Waveform and concept, Color Doppler, Components of Ultrasound.
Usg transducer and basic principles of ultrasound Doppler, this slide describe the basic physics of ultrasound transducer and Doppler , must know thing is given in this presentaion. Good review for radiology resident. Thanks.
Definition of Side lobes and the principle behind its production during ultrasound imaging. Side lobes artifact and its result on image. Explanation of harmonic imaging, its production and the techniques use to eliminate fundamental frequency to produce optimal harmonic images.
Beginning with basic physics of ultrasound, in the presentation how an ultrasound image is constructed is tried to be revealed by investigation of the wave propagation through the tissue.
An overview of Doppler Effect in Ultrasonography - the medical imaging of the body using Ultrasound.
Includes Colour Doppler, Power Doppler, Spectral Doppler, Continuous Wave Doppler, Pulsed Wave Doppler, and comparisons with other Radiographic imaging modalities.
Training Material inherited form Philips Basics of Ultrasonography. Covers the fundamentals of Ultrasound Waveform, Piezoelectric Effect, Phased Echo Concept, Goal of Ultrasound, Ultrasound Image Construction process, Types of Resolution, Probe Internals, The Doppler Effect, Spectrum Waveform and concept, Color Doppler, Components of Ultrasound.
Usg transducer and basic principles of ultrasound Doppler, this slide describe the basic physics of ultrasound transducer and Doppler , must know thing is given in this presentaion. Good review for radiology resident. Thanks.
Definition of Side lobes and the principle behind its production during ultrasound imaging. Side lobes artifact and its result on image. Explanation of harmonic imaging, its production and the techniques use to eliminate fundamental frequency to produce optimal harmonic images.
Beginning with basic physics of ultrasound, in the presentation how an ultrasound image is constructed is tried to be revealed by investigation of the wave propagation through the tissue.
PRESENTATION IS COMPACT AND INFORMATIVE. HAS FLOWCHARTS AND DIAGRAMS. REFERENCE IS FROM LATEST ARTICLES AND STANDARD TEXTBOOKS. SERVES A GREAT DEAL TO BRUSH UP THE THEORETICAL KNOWLEDGE .
PRESENTATION IS COMPACT AND INFORMATIVE. HAS FLOWCHARTS AND DIAGRAMS. REFERENCE IS FROM LATEST ARTICLES AND STANDARD TEXTBOOKS. SERVES A GREAT DEAL TO BRUSH UP THE THEORETICAL KNOWLEDGE .
PRESENTATION IS COMPACT AND INFORMATIVE. HAS FLOWCHARTS AND DIAGRAMS. REFERENCE IS FROM LATEST ARTICLES AND STANDARD TEXTBOOKS. SERVES A GREAT DEAL TO BRUSH UP THE THEORETICAL KNOWLEDGE .
PRESENTATION IS COMPACT AND INFORMATIVE. HAS FLOWCHARTS AND DIAGRAMS. REFERENCE IS FROM LATEST ARTICLES AND STANDARD TEXTBOOKS. SERVES A GREAT DEAL TO BRUSH UP THE THEORETICAL KNOWLEDGE .
PRESENTATION IS COMPACT AND INFORMATIVE. HAS FLOWCHARTS AND DIAGRAMS. REFERENCE IS FROM LATEST ARTICLES AND STANDARD TEXTBOOKS. SERVES A GREAT DEAL TO BRUSH UP THE THEORETICAL KNOWLEDGE .
PRESENTATION IS COMPACT AND INFORMATIVE. HAS FLOWCHARTS AND DIAGRAMS. REFERENCE IS FROM LATEST ARTICLES AND STANDARD TEXTBOOKS. SERVES A GREAT DEAL TO BRUSH UP THE THEORETICAL KNOWLEDGE .
PRESENTATION IS COMPACT AND INFORMATIVE. HAS FLOWCHARTS AND DIAGRAMS. REFERENCE IS FROM LATEST ARTICLES AND STANDARD TEXTBOOKS. SERVES A GREAT DEAL TO BRUSH UP THE THEORETICAL KNOWLEDGE .
PRESENTATION IS COMPACT AND INFORMATIVE. HAS FLOWCHARTS AND DIAGRAMS. REFERENCE IS FROM LATEST ARTICLES AND STANDARD TEXTBOOKS. SERVES A GREAT DEAL TO BRUSH UP THE THEORETICAL KNOWLEDGE .
PRESENTATION IS COMPACT AND INFORMATIVE. HAS FLOWCHARTS AND DIAGRAMS. REFERENCE IS FROM LATEST ARTICLES AND STANDARD TEXTBOOKS. SERVES A GREAT DEAL TO BRUSH UP THE THEORETICAL KNOWLEDGE .
PRESENTATION IS COMPACT AND INFORMATIVE. HAS FLOWCHARTS AND DIAGRAMS. REFERENCE IS FROM LATEST ARTICLES AND STANDARD TEXTBOOKS. SERVES A GREAT DEAL TO BRUSH UP THE THEORETICAL KNOWLEDGE .
PRESENTATION IS COMPACT AND INFORMATIVE. HAS FLOWCHARTS AND DIAGRAMS. REFERENCE IS FROM LATEST ARTICLES AND STANDARD TEXTBOOKS. SERVES A GREAT DEAL TO BRUSH UP THE THEORETICAL KNOWLEDGE .
PRESENTATION IS COMPACT AND INFORMATIVE. HAS FLOWCHARTS AND DIAGRAMS. REFERENCE IS FROM LATEST ARTICLES AND STANDARD TEXTBOOKS. SERVES A GREAT DEAL TO BRUSH UP THE THEORETICAL KNOWLEDGE .
PRESENTATION IS COMPACT AND INFORMATIVE. HAS FLOWCHARTS AND DIAGRAMS. REFERENCE IS FROM LATEST ARTICLES AND STANDARD TEXTBOOKS. SERVES A GREAT DEAL TO BRUSH UP THE THEORETICAL KNOWLEDGE .
PRESENTATION IS COMPACT AND INFORMATIVE. HAS FLOWCHARTS AND DIAGRAMS. REFERENCE IS FROM LATEST ARTICLES AND STANDARD TEXTBOOKS. SERVES A GREAT DEAL TO BRUSH UP THE THEORETICAL KNOWLEDGE .
PRESENTATION IS COMPACT AND INFORMATIVE. HAS FLOWCHARTS AND DIAGRAMS. REFERENCE IS FROM LATEST ARTICLES AND STANDARD TEXTBOOKS. SERVES A GREAT DEAL TO BRUSH UP THE THEORETICAL KNOWLEDGE .
TEST BANK for Operations Management, 14th Edition by William J. Stevenson, Ve...kevinkariuki227
TEST BANK for Operations Management, 14th Edition by William J. Stevenson, Verified Chapters 1 - 19, Complete Newest Version.pdf
TEST BANK for Operations Management, 14th Edition by William J. Stevenson, Verified Chapters 1 - 19, Complete Newest Version.pdf
Lung Cancer: Artificial Intelligence, Synergetics, Complex System Analysis, S...Oleg Kshivets
RESULTS: Overall life span (LS) was 2252.1±1742.5 days and cumulative 5-year survival (5YS) reached 73.2%, 10 years – 64.8%, 20 years – 42.5%. 513 LCP lived more than 5 years (LS=3124.6±1525.6 days), 148 LCP – more than 10 years (LS=5054.4±1504.1 days).199 LCP died because of LC (LS=562.7±374.5 days). 5YS of LCP after bi/lobectomies was significantly superior in comparison with LCP after pneumonectomies (78.1% vs.63.7%, P=0.00001 by log-rank test). AT significantly improved 5YS (66.3% vs. 34.8%) (P=0.00000 by log-rank test) only for LCP with N1-2. Cox modeling displayed that 5YS of LCP significantly depended on: phase transition (PT) early-invasive LC in terms of synergetics, PT N0—N12, cell ratio factors (ratio between cancer cells- CC and blood cells subpopulations), G1-3, histology, glucose, AT, blood cell circuit, prothrombin index, heparin tolerance, recalcification time (P=0.000-0.038). Neural networks, genetic algorithm selection and bootstrap simulation revealed relationships between 5YS and PT early-invasive LC (rank=1), PT N0—N12 (rank=2), thrombocytes/CC (3), erythrocytes/CC (4), eosinophils/CC (5), healthy cells/CC (6), lymphocytes/CC (7), segmented neutrophils/CC (8), stick neutrophils/CC (9), monocytes/CC (10); leucocytes/CC (11). Correct prediction of 5YS was 100% by neural networks computing (area under ROC curve=1.0; error=0.0).
CONCLUSIONS: 5YS of LCP after radical procedures significantly depended on: 1) PT early-invasive cancer; 2) PT N0--N12; 3) cell ratio factors; 4) blood cell circuit; 5) biochemical factors; 6) hemostasis system; 7) AT; 8) LC characteristics; 9) LC cell dynamics; 10) surgery type: lobectomy/pneumonectomy; 11) anthropometric data. Optimal diagnosis and treatment strategies for LC are: 1) screening and early detection of LC; 2) availability of experienced thoracic surgeons because of complexity of radical procedures; 3) aggressive en block surgery and adequate lymph node dissection for completeness; 4) precise prediction; 5) adjuvant chemoimmunoradiotherapy for LCP with unfavorable prognosis.
Explore natural remedies for syphilis treatment in Singapore. Discover alternative therapies, herbal remedies, and lifestyle changes that may complement conventional treatments. Learn about holistic approaches to managing syphilis symptoms and supporting overall health.
Report Back from SGO 2024: What’s the Latest in Cervical Cancer?bkling
Are you curious about what’s new in cervical cancer research or unsure what the findings mean? Join Dr. Emily Ko, a gynecologic oncologist at Penn Medicine, to learn about the latest updates from the Society of Gynecologic Oncology (SGO) 2024 Annual Meeting on Women’s Cancer. Dr. Ko will discuss what the research presented at the conference means for you and answer your questions about the new developments.
These simplified slides by Dr. Sidra Arshad present an overview of the non-respiratory functions of the respiratory tract.
Learning objectives:
1. Enlist the non-respiratory functions of the respiratory tract
2. Briefly explain how these functions are carried out
3. Discuss the significance of dead space
4. Differentiate between minute ventilation and alveolar ventilation
5. Describe the cough and sneeze reflexes
Study Resources:
1. Chapter 39, Guyton and Hall Textbook of Medical Physiology, 14th edition
2. Chapter 34, Ganong’s Review of Medical Physiology, 26th edition
3. Chapter 17, Human Physiology by Lauralee Sherwood, 9th edition
4. Non-respiratory functions of the lungs https://academic.oup.com/bjaed/article/13/3/98/278874
Prix Galien International 2024 Forum ProgramLevi Shapiro
June 20, 2024, Prix Galien International and Jerusalem Ethics Forum in ROME. Detailed agenda including panels:
- ADVANCES IN CARDIOLOGY: A NEW PARADIGM IS COMING
- WOMEN’S HEALTH: FERTILITY PRESERVATION
- WHAT’S NEW IN THE TREATMENT OF INFECTIOUS,
ONCOLOGICAL AND INFLAMMATORY SKIN DISEASES?
- ARTIFICIAL INTELLIGENCE AND ETHICS
- GENE THERAPY
- BEYOND BORDERS: GLOBAL INITIATIVES FOR DEMOCRATIZING LIFE SCIENCE TECHNOLOGIES AND PROMOTING ACCESS TO HEALTHCARE
- ETHICAL CHALLENGES IN LIFE SCIENCES
- Prix Galien International Awards Ceremony
Title: Sense of Smell
Presenter: Dr. Faiza, Assistant Professor of Physiology
Qualifications:
MBBS (Best Graduate, AIMC Lahore)
FCPS Physiology
ICMT, CHPE, DHPE (STMU)
MPH (GC University, Faisalabad)
MBA (Virtual University of Pakistan)
Learning Objectives:
Describe the primary categories of smells and the concept of odor blindness.
Explain the structure and location of the olfactory membrane and mucosa, including the types and roles of cells involved in olfaction.
Describe the pathway and mechanisms of olfactory signal transmission from the olfactory receptors to the brain.
Illustrate the biochemical cascade triggered by odorant binding to olfactory receptors, including the role of G-proteins and second messengers in generating an action potential.
Identify different types of olfactory disorders such as anosmia, hyposmia, hyperosmia, and dysosmia, including their potential causes.
Key Topics:
Olfactory Genes:
3% of the human genome accounts for olfactory genes.
400 genes for odorant receptors.
Olfactory Membrane:
Located in the superior part of the nasal cavity.
Medially: Folds downward along the superior septum.
Laterally: Folds over the superior turbinate and upper surface of the middle turbinate.
Total surface area: 5-10 square centimeters.
Olfactory Mucosa:
Olfactory Cells: Bipolar nerve cells derived from the CNS (100 million), with 4-25 olfactory cilia per cell.
Sustentacular Cells: Produce mucus and maintain ionic and molecular environment.
Basal Cells: Replace worn-out olfactory cells with an average lifespan of 1-2 months.
Bowman’s Gland: Secretes mucus.
Stimulation of Olfactory Cells:
Odorant dissolves in mucus and attaches to receptors on olfactory cilia.
Involves a cascade effect through G-proteins and second messengers, leading to depolarization and action potential generation in the olfactory nerve.
Quality of a Good Odorant:
Small (3-20 Carbon atoms), volatile, water-soluble, and lipid-soluble.
Facilitated by odorant-binding proteins in mucus.
Membrane Potential and Action Potential:
Resting membrane potential: -55mV.
Action potential frequency in the olfactory nerve increases with odorant strength.
Adaptation Towards the Sense of Smell:
Rapid adaptation within the first second, with further slow adaptation.
Psychological adaptation greater than receptor adaptation, involving feedback inhibition from the central nervous system.
Primary Sensations of Smell:
Camphoraceous, Musky, Floral, Pepperminty, Ethereal, Pungent, Putrid.
Odor Detection Threshold:
Examples: Hydrogen sulfide (0.0005 ppm), Methyl-mercaptan (0.002 ppm).
Some toxic substances are odorless at lethal concentrations.
Characteristics of Smell:
Odor blindness for single substances due to lack of appropriate receptor protein.
Behavioral and emotional influences of smell.
Transmission of Olfactory Signals:
From olfactory cells to glomeruli in the olfactory bulb, involving lateral inhibition.
Primitive, less old, and new olfactory systems with different path
- Video recording of this lecture in English language: https://youtu.be/lK81BzxMqdo
- Video recording of this lecture in Arabic language: https://youtu.be/Ve4P0COk9OI
- Link to download the book free: https://nephrotube.blogspot.com/p/nephrotube-nephrology-books.html
- Link to NephroTube website: www.NephroTube.com
- Link to NephroTube social media accounts: https://nephrotube.blogspot.com/p/join-nephrotube-on-social-media.html
The prostate is an exocrine gland of the male mammalian reproductive system
It is a walnut-sized gland that forms part of the male reproductive system and is located in front of the rectum and just below the urinary bladder
Function is to store and secrete a clear, slightly alkaline fluid that constitutes 10-30% of the volume of the seminal fluid that along with the spermatozoa, constitutes semen
A healthy human prostate measures (4cm-vertical, by 3cm-horizontal, 2cm ant-post ).
It surrounds the urethra just below the urinary bladder. It has anterior, median, posterior and two lateral lobes
It’s work is regulated by androgens which are responsible for male sex characteristics
Generalised disease of the prostate due to hormonal derangement which leads to non malignant enlargement of the gland (increase in the number of epithelial cells and stromal tissue)to cause compression of the urethra leading to symptoms (LUTS
ARTIFICIAL INTELLIGENCE IN HEALTHCARE.pdfAnujkumaranit
Artificial intelligence (AI) refers to the simulation of human intelligence processes by machines, especially computer systems. It encompasses tasks such as learning, reasoning, problem-solving, perception, and language understanding. AI technologies are revolutionizing various fields, from healthcare to finance, by enabling machines to perform tasks that typically require human intelligence.
HOT NEW PRODUCT! BIG SALES FAST SHIPPING NOW FROM CHINA!! EU KU DB BK substit...GL Anaacs
Contact us if you are interested:
Email / Skype : kefaya1771@gmail.com
Threema: PXHY5PDH
New BATCH Ku !!! MUCH IN DEMAND FAST SALE EVERY BATCH HAPPY GOOD EFFECT BIG BATCH !
Contact me on Threema or skype to start big business!!
Hot-sale products:
NEW HOT EUTYLONE WHITE CRYSTAL!!
5cl-adba precursor (semi finished )
5cl-adba raw materials
ADBB precursor (semi finished )
ADBB raw materials
APVP powder
5fadb/4f-adb
Jwh018 / Jwh210
Eutylone crystal
Protonitazene (hydrochloride) CAS: 119276-01-6
Flubrotizolam CAS: 57801-95-3
Metonitazene CAS: 14680-51-4
Payment terms: Western Union,MoneyGram,Bitcoin or USDT.
Deliver Time: Usually 7-15days
Shipping method: FedEx, TNT, DHL,UPS etc.Our deliveries are 100% safe, fast, reliable and discreet.
Samples will be sent for your evaluation!If you are interested in, please contact me, let's talk details.
We specializes in exporting high quality Research chemical, medical intermediate, Pharmaceutical chemicals and so on. Products are exported to USA, Canada, France, Korea, Japan,Russia, Southeast Asia and other countries.
Ozempic: Preoperative Management of Patients on GLP-1 Receptor Agonists Saeid Safari
Preoperative Management of Patients on GLP-1 Receptor Agonists like Ozempic and Semiglutide
ASA GUIDELINE
NYSORA Guideline
2 Case Reports of Gastric Ultrasound
3. What is ultrasound?
Ultrasound or ultrasonography is a medical imaging
technique that uses high frequency sound waves to
obtain cross sectional images of the body.
Also known as ‘pulse echo’ technique
6. Sound
Sound is a mechanical energy that travels through
matter, as a result of vibration of the particles of the
medium through which the sound wave is moving.
Vibration of particles Transport the energy through
medium pressure wave
Change in pressure sinusoidal waveform
9. Waves
Transverse waves Longitudinal waves
waves parallel to the
direction of energy
transfer
Waves perpendicular to
the direction of energy
transfer
10.
11. Sound waves
longitudinal waves (air)
Compression
Space where density and pressure is
elevated (positive)
Rarefaction
Space where density and pressure is
depressed (negative)
12. Unit of acoustic frequency(number of cycles in a unit of time) -
Hertz (Hz)
1 Hz = 1 cycle per second
High frequencies expressed in kHz and MHz
1 kHz = 1000Hz
1 MHz = 1,000kHz
13. Speed of sound
Determined by properties of medium
i.e. Its resistance to compression
C α-1 COMPRESSIBILITY
• FASTER IN SOLID
• SLOWER IN GAS
Propagation velocity of sound (c) = f x λ
18. Z1
Z2
The difference in acoustic impedance (z)
between the two tissues causes reflection
of the sound wave
19. angle of incidence α-1 amount of reflected sound
20.
21. soft tissue - air
interface
reflects almost the entire
beam,
transducer must be
directly coupled to the
patient skin without an
air gap
bone - tissue
interface
Bone transmits about
70% of sound energy
due to its impedance
difference - hence not
possible to image
through it
air/gas
interface
causes total reflection -
hence they cast a
shadow and structures
underneath cannot be
imaged
E.g Bowel wall can be
imaged and not its
lumen
22. Specular reflectors are smooth so
when incident beam strikes, the
reflected beam leaves in the same
angle as the incident beam
(bright fibrous structures like)
Diaphragm,
Bladder wall
Tendons
Diffuse reflectors tend to be
rough so when the incident beam
strikes, echoes from the interfaces
are scattered in all directions
kidney,
liver,
small blood vessels
walls of heart chambers
Specular reflectors Diffuse reflectors
23.
24. Interference
If two sound waves of the same wavelength cross each other, the pressure waves combine.
Constructive interference
waves in step/phase,
their amplitudes add up -
Destructive interference
waves out of phase, they
tend to cancel out each other
25. Refraction
When sound passes from a tissue with one acoustic
propagation velocity to a tissue with a higher or lower
sound velocity, there is a change in the direction of the
sound wave
Causes artefacts
26. Absorption
As the ultrasound pulse moves through matter, it continuously
loses energy - attenuation
There is absorption of the ultrasound energy by the material
and gets converted into HEAT.
Scattering and refraction interactions also remove some of the
energy from the pulse and contribute to its overall attenuation,
but absorption is the most significant
Reduction of intensity of beam as it traverses through matter.
27. Attenuation
Attenuation of sound energy influences the depth
in tissue, from which useful information can be
obtained Affects transducer selection and
operator controlled instrument settings
The deeper the wave travels in the body, the
weaker it becomes
The amplitude/strength of the wave decreases
with increasing depth
Decibel
29. frequencies a attenuation
Attenuation determines the efficiency with which ultrasound penetrates
a specific tissue
30. Instrumentation
1. Transmitter - energize transducer
2. Transducer
3. Receiver and processor - detect and amplify
backscattered energy
4. Display – presents the ultrasound image/data
5. Recording/Storage
31. Transmitter
Pulsed ultrasound
Produces precisely timed high amplitude voltage as
pulses to energize the transducer
Controls PRF
Pulse repetition frequency – rate of pulses emitted by
transducer
32. Pulse repetition frequency
Determines time interval between pulses
Determines depth
PRF – 1 to 10 kHz for diagnostic imaging
Pulse should travel to depth and return before next
pulse is send
Highest USG frequency permitting penetration to
depth of interest should be selected
33. Pulse echo principle
To determine the depth of the tissues.
T = 2D / c ---- principle of pulse echo
Generation of sound pulse
Reflection of pulse from targets
Detection of reflected echoes
Calculation and display of range of
targets
34.
35. Transducer
Device that converts one form of energy into another
Ultrasonic transducers convert an electric signal ultrasonic
energy reflected back from the tissues into electrical signal
Transducer functions as both transmitter and receiver
Used in both pulsed and continuous wave modes
36.
37. Piezoelectric element located near the face of
transducer.
Front and back faces coated with thin conducting film
Surfaces of crystal coated with gold or silver electrodes
Outside electrode grounded to protect the patient
and its surface is coated with water tight electric
insulator
Inside electrode rest against a thick backing block
that absorbs sound waves transmitted back into
transducer
Housing – strong plastic
An acoustic insulator of rubber or cork prevents the
sound from passing into the housing
38. Piezoelectricity
Discovered by Pierre and Jacques Curie in 1880
Greek word “piezein” - to press
Piezoelectric effect – application of an electric field to
certain materials causes a change in their physical
dimensions
Piezoelectric material - innumerable dipoles arranged
in a geometric pattern
Each dipole – positive charge at one end and
negative charge at other
39. Generates ultrasound waves
Crystal expands and contracts
Alignment of dipoles
Electric field
Generation of voltage
When crystal gets compressed
40. Serves as ultrasonic signal for display
Voltage amplified
Induces voltage between the electrodes
Forces dipoles to change orientation
Physical compression of crystal element
Transmits their energy to transducer
Echoes reflect back to transducer from each tissue
interface
Sound waves pass though the body
41. Compression force and associated voltage – piezoelectric effect
In response to a compressive force on a piezoelectric slab, a net
electric dipole moment detection of a voltage in response to
this strain
Reverse piezoelectricity - In response to the application of an
electric field, a deformation of the piezoelectric slab detected.
42. Piezoelectric material
Quartz – original transducer material
Ferroelectrics ( artificial piezoelectric material ) – barium titanate,
lead zirconate titanate
To produce polarization the ceramic is heated at high
temperature in a strong electric field
At high temperature, the dipoles are free to move, and the
electric field brings them into the desired alignment
The crystal is then gradually cooled while subjected to a constant
high voltage
43. Curie temperature
It is the temperature above which this polarization is
lost
PZT - 365˚C
Quartz - 573˚C
44. Resonant frequency
Frequency at which the wavelength = 2 x thickness of
piezoelectric disc
Constructive interference / resonance of waves emitted from
forwards and back face of disc occurs
Transducer most efficient as transmitter and receiver of sound
Depends on thickness of disc and material of disc
(determines velocity)
45. • The thicker the piezoelectric element, the lower the resonant
frequency
• Conversely thinner the element higher the operating frequency
• Range of frequencies produced by a transducer – bandwidth
• Broad band technology produces medical transducers that
contain more than one operating frequencies
46. 2.5 MHz deep abdomen, obstetric and
gynaecological imaging
3.5MHz general abdomen, obstetric and
gynaecological imaging
5.0MHz vascular, breast, pelvic imaging
7.5MHz breast, thyroid
10.0MHz breast, thyroid, superficial veins, superficial
masses, musculoskeletal imaging.
15.0MHz superficial structures, musculoskeletal
imaging.
47. Pulse duration/length
Transducer continues to vibrate for a short time after it is
stimulated ultrasound pulse will be several cycles long
Damping block is used to stop the crystal vibration so that
transducer is ready to receive the reflected waves from tissue
interfaces within the body
Damping materials used – tungsten and rubber powder
suspended in epoxy resin(backing block)
48.
49.
50. Q factor
Mechanical coefficient ( Q ) –
mean frequency : bandwidth
Greater Q –
narrower bandwidth,
lesser damping,
longer ring down time/ pulse duration
Lesser Q –
broader bandwidth,
heavier damping,
smaller ring down time/ pulse duration
51. Coupling agents
Commonly known as GEL
Fluid medium needed to provide a link between the transducer
and the patient
Composition –
Carbomer – 10gm
Propylene glycol – 75gm
Trolamine – 12.5gm
EDTA – 0.25gm
Distilled water – upto 550ml
52. Ultrasound beam characteristics
Unfocused beam – An ultrasound beam leaving a flat
crystal element has an initial cylindrical segment, followed
by a diverging conical portion
Frequency –
Higher the frequency, longer will be the cylindrical
segment or near field (Fresnel zone) and far field
(Fraunhofer zone) becomes less divergent
Lateral resolution – junction of near and far fields
Depth resolution improves at higher frequencies
As frequency is increased, greater absorption of sound
energy weakens beam intensity
53.
54. Near field / Fresnel zone
Interference of pressure waves near the
transducer results in great amplitude
variation
Far field / Fraunhofer zone
Farther from the transducer, the waves
diverge and amplitude decreases at steady
state
Higher frequency transducer, larger diameter - increases length
of near zone and decreases divergence of far zone, hence beam
becomes more directional
Crystal diameter –Increase in crystal diameter
increases the near zone length but worsens
lateral and depth resolution
55. Focused beam
Improves both lateral and depth resolution
Concave crystal face – Greater the concavity, nearer the beam
will focus relative to the crystal
Acoustic lenses – Shorter the focal length, the nearer the lens
will focus the beam relative to the lens face
56.
57. Receiver
Detects and amplifies weak signals
Compensates the differences in echo strength, which
results from attenuation by different tissue thickness
Compression of wider range of amplitudes returning
to transducer into a range that can be displayed to the
user
58. Time gain compensation
Amplitude of sound pulse diminishes as it travels through body
due to attenuation and echo pulse similarly attenuated as it
travels back to transducer
Selective amplification of signals received from greater depth
TGC is under operator control
59.
60.
61.
62.
63.
64.
65. Dynamic range
Equivalent to number of shades of gray
Ratio of the largest to the smallest echoes processed by
components of ultrasound device
Dynamic range decreases as signals pass through the imaging
system as operations such as TGC eliminate small and large
signals smaller change in echo amplitude that will effect the
gray scale brightness image contrast is enhanced for smaller
dynamic range
Echoes returning from tissues can have dynamic ranges of 100 to
150dB
Increasing dynamic range yields more shades of gray in
ultrasound signals
68. A mode
‘A’ - amplitude
Earliest incarnation of ultrasound system
Shows signals as spikes and height of the spike represents the
amplitude of the reflected signal
Only shows position of tissue interfaces
Ophthalmology studies – to detect optic nerve
69. M mode
‘M’ – motion
Detection of motion of a structure or structures over time along
scan line of interest
A B mode image is frozen on the screen and used to direct the
beam from a stationary transducer along a line of interest
Echoes are displayed as a line of moving bright dots
Displays echo amplitude and shows position of moving reflectors
Evaluation of rapid motion of cardiac valves and of cardiac
chamber and vessel walls
70.
71. B mode
‘B’ – brightness
Involves the pulse echo sequence with the buildup of
an image from innumerable echo blips
Brightness corresponds to received echo amplitude
Greatest intensity signal – white
Absence of signal – black
Signs of intermediate intensity – gray
72.
73. Static B mode ultrasound -
Image is compiled as sound beam is scanned
across patient
Larger field of view, restricted transducer
movements and single-depth focusing
• Real time ultrasound -
Image is automatically scanned in a
succession of frames sufficiently rapidly to
demonstrate the motion of tissues
Frame rate of 15-60/second
Assessment of both anatomy and motion
74. Gray scale imaging
Display great amplitudes of echoes arising from
tissues as varying shades of gray on monitor
Scan convertor –
Improve grayscale images
Enable freeze frame
Manipulate data and archive if necessary
Write, read and erase
Analog and digital scan convertors
75. 3D Ultrasound
Images obtained from a set
of 2D scans
3D scanners used for fetal,
gynaecological and cardiac
scanning
4D ultrasound is the term
used for real time 3D
imaging
76. To produce real-time
images
Mechanical scanners
Conventional /group of single
element transducers is mechanically
moved to form images in real time
Oscillating Rotating
electronic scanners
Other uses an array of transducers,
which do not move but are activated
electronically –
Linear Curvilinear Phased
77. Mechanical – Oscillating transducers
With unenclosed crystal –
single element caused to
oscillate through an angle,
which defines the field of view;
Sector scan
With enclosed crystal –
transducer enclosed within a
fluid filled container; use of an
electromagnet and trapezoidal
image produced
78.
79. Mechanical – Rotating transducers
Employs 3 or 4 crystals mounted symmetrically on a
rotating wheel
The wheel is driven by an electric motor to perform
circular motion in one direction only
The crystal elements are excited one at a time to
provide the ultrasound beam
Replaced by annular array transducers
80. Linear Array
Array element is pulsed sequentially to produce a
rectangular scan pattern
At any time, only one element or subset of elements
is in operation. The echoes are received before the
next element is excited
Used for small parts, vascular and
obstetric applications
81.
82. Curvilinear array
One element or a subset of
elements at a time scanned
sequentially
Produces trapezoid scan
pattern
83. Phased array
All elements are scanned at the same time
Ultrasound beam is caused to sweep back and forth
across the patient by electronically controlled steering
and focusing
Ideal for cardiac imaging
84. Transducer selection
For general purpose – convex with 3.5MHz
For obstetric purpose – convex or linear with 3.5MHz
For superficial structures – linear with 5MHz
For paediatric or thin people – 5MHz
85. Tissue harmonics imaging
Based on nonlinear propagation of ultrasound wave propagating
through tissue causing distortion and generating harmonics
It decreases “clutter” noise
It yields sharper overall images with improved contrast, axial and
lateral resolution
Used in obstetrics, cardiology and abdominal applications
86. Spatial Resolution
Ability to distinguish between two closely situated
structures from one another
Axial or lateral resolution
87. Axial (depth) resolution
Minimum reflector resolution along axis of ultrasound
beam
Determined by pulse length
Higher frequencies lesser pulse length higher
axial resolution
88. Lateral (horizontal) resolution
Minimal reflector resolution perpendicular to beam and parallel
to transducer
Determined by beam width
Beam width is a function of wavelength, piezoelectric
transducer diameter, distance from
transducer
Can be controlled by focusing the beam
89.
90. Image optimization
Gain and TGC - Sensitivity controls
Focus – Lateral resolution
Frequency – Axial resoultion
High good axial resolution but poor penetration
Low less axial resolution but better penetration
Depth – decrease in depth increase in frame rate (temporal
resolution)
Sector width increase in frame rate
95. Gain –
Overall brightness of the image
Amplification of returning ultrasound signal received from all
depths
High gain high amplitude differentiation lost
Adjust to minimum required to obtain a complete range from low
(dark gray) to high (white) amplitude signals
Time Gain compensation –
Adjustment of image brightness at a selective depth
Corrects varying depths of intensities
Adjust TGC to decrease in near field and increase in far field (‘’
shape)
96. Focus –
Allows focus of ultrasound beam to area of interest
Optimizes ultrasound intensity in near and far field improved
spatial resolution
Depth –
Adjustment of the depth of field of view
Greater the depth, lesser resolution and frame rate
Set the depth to minimum required to visualize all structures of
interest
Frequency –
Increased frequency increase resolution
Sector width –
Wide sector frame rate and temporal resolution
97. • Dynamic range –
• 100dB of ultrasound information available but
monitor can display a much smaller range
• Allows compression of wide spectrum of
amplitudes
• Compressed signals then displayed on monitor
varying shades of gray
100. Thermal bioeffect
Mechanism - with sound wave attenuation, energy is lost as
heat
Longer use of ultrasound on the tissue longer the
ultrasound has attenuated in the tissue
More the tissue attenuates ultrasound more the bioeffects
More ultrasound attenuates at the surface and focal zone
Higher the frequency/power/PRF/pulse duration , more the
attenuation and thus the heating
101. Thermal indices
NCRP guidelines – in ultrasound examinations, no temperature
rise > 1°C
TI – indicator of relative potential for increasing tissue
temperature
TIS – Thermal index for soft tissue
TIB – Thermal index for bone, at the focus
TIC – Thermal index for bone, at surface
102. Mechanical bioeffects
Current USG systems can produce cavitation
Determines potential for bubble formation in vivo
Cavitation -
Stable
Inertial / Transient
103. Stable cavitation
A bubble forms as the peak low pressure region of an
ultrasound pulse passes through a nucleation site
This bubble grows and shrinks with the high pressure
and low pressure region of the ultrasound pulses
This growing and shrinking of the bubble causes fluid
to flow around the bubble - micro-streaming
Causes mechanical damage, sometimes cell lysis
104. Inertial/transient cavitation
Here, the bubbles formed in a similar way but under
high pressure or intensities, they expand very quickly
following which they collapse violently
This collapse produces high temperature and
pressure, which damages cells in the region of the
collapse
105.
106. Mechanical index
It is an approximation of mechanical effects
It is a calculated number based on ultrasound
parameters and facts about the machine that the
manufacturer built into the programming
Mechanical index less than 1.9, no mechanical effects
is seen
Editor's Notes
SOUND NEEDS A MEDIUM FOR PROPAGATION
CANNOT TRAVEL IN VACCUM
LONGITUDINAL PROPAGATION
PARTICLES VIBRATE PARALLEL TO THE DIRECTION OF PROPAGATION
In diagnostic ultrasound, only the longitudinal waves are important.
Reflection
Refraction
Though its not a EMR
ACOUSTIC IMPEDENCE IS THE OPPOSITION OF A MEDIUM TO A LONGITUDINAL WAVE MOTION
P = DENSITY
V = SPEED
Z = RAYLS (Kg / m3.s )
Reflection
Refraction
Though its not a EMR
Amount of reflection determined by angle of incidence between sound beam and reflecting surface
Water produces by far the least attenuation. This means that water is a very good conductor of ultrasound. Water within the body, such as in cysts and the bladder, forms "windows" through which underlying structures can be easily imaged.
Most of the soft tissues of the body have attenuation coefficient values of approximately 1 dB per cm per MHz, with the exception of fat and muscle.
Muscle has a range of values that depends on the direction of the ultrasound with respect to the muscle fibers.
Lung has a much higher attenuation rate than either air or soft tissue. This is because the small pockets of air in the alveoli are very effective in scattering ultrasound energy. Because of this, the normal lung structure is extremely difficult to penetrate with ultrasound.
Compared to the soft tissues of the body, bone has a relatively high attenuation rate. Bone, in effect, shields some parts of the body against easy access by ultrasound.