DOPPLER EFFECT
Course: Ultrasound
Victor Ekpo
MSc Medical Physics programme, College of Medicine, University of Lagos, 2017.
OUTLINE
• Introduction
• Physics of Doppler Effect
• Conditions for Doppler Effect
• Classifications
• Types
• Modes
• Applications
• Artifacts in Doppler Imaging
• Conclusion
2
INTRODUCTION
• Ultrasound is sound waves with frequency greater than
20kHz (Medical Ultrasound range: >2MHz)
• Sound waves travelling through a medium can be
reflected, refracted, absorbed or transmitted.
• Reflection occurs at the point of acoustic impedance
mismatch.
3
MOVING SOUND
• One or both the sound source and receiver may be
stationary or moving.
• If the sound source moves towards the listener, the
sound is perceived to have a higher frequency/pitch, and
a lower frequency as it moves away from the
listener.
4
DOPPLER EFFECT
Doppler effect is the change in frequency of
sound (or any wave) due to the relative motion* of
the source and receiver.
Doppler Shift (Δf) = Reflected Frequency – Transmitted frequency
Doppler angle ( θ ) = Angle between the direction of source – direction of sound
5
CONDITIONS FOR DOPPLER SHIFT
The Doppler effect will NOT occur:
• If the source and observer both move in the same direction at the
same speed.
• Where one source/listener is at the centre of a circle while the
other is moving on it with uniform speed.
• If the source/receiver is at 900 to the receiver/source.
6
7
Longer wavelengths
Lower frequency
Shorter wavelengths
Higher frequency
TOWARDS AWAY FROM
8
Sound waves reflected from a moving object are compressed
(higher frequency) when moving towards the listener/
transducer,
…and expanded (lower frequency) when moving away from
the listener/transducer compared to the incident sound
wave frequency.
9
The velocity of sound (c) is given by the equation,
Speed, c =
𝑊𝑎𝑣𝑒𝑙𝑒𝑛𝑔𝑡ℎ (λ)
𝑃𝑒𝑟𝑖𝑜𝑑 (𝑇)
= Wavelength (λ) * frequency (f)
λo =
𝑐
𝑓 𝑜
Consider a sound source travelling at a velocity 𝒗 𝒔 towards a detector.
After time t, the new wavelength,
λ1 =
𝑐 − 𝒗 𝒔
𝑓 𝑜
10
PHYSICS OF DOPPLER EFFECT
𝒗 𝒔
c
With a shortened wavelength, the new frequency f1 is given by:
f1 =
𝑐
λ1
=
𝑐
(c−vs) / fo
f1 = fo
𝑐
( 𝑐 − 𝒗 𝒔
)
Δf = f1 - fo = fo
𝑐
( 𝑐 − 𝒗 𝒔
) - fo
= fo
𝑐
( 𝑐 − 𝒗 𝒔
) - 1
= fo
𝑐 −( 𝑐 −𝑣𝑠 )
( 𝑐 − 𝒗 𝒔
)
Δf = fo
𝑣 𝑠
( 𝑐 − 𝒗 𝒔
)
If c >> vs , then
Δf = fo
𝑣 𝑠
𝑐
11
A similar equation can be written for a detector moving towards a source,
Δf = fo
𝑣 𝑑
𝑐
It can be shown that for a source moving away,
λ1 =
𝑐 + 𝒗 𝒔
𝑓 𝑜
Δf = fo
−𝑣 𝑠
𝑐
OR fo
−𝑣𝑑
𝑐
For a 2-way motion where a transducer emits ultrasound of frequency f,
and it gets reflected back by moving blood of velocity v (and c>>v), then:
Δf = 2f
𝑣
𝑐
12
If direction of motion is oblique, then:
Δf = (2vf Cos θ) /c
where:
Δf = Doppler shift
f = transducer frequency
v = velocity of source/receiver
c = velocity of sound
θ = Doppler angle 13
PHYSICS OF DOPPLER EFFECT (contd.)
• Doppler shift, Δf increases with:
• increasing source velocity v, and
• increasing transducer frequency, f
• Δf is maximum at θ = 0o and minimum at θ = 90o.
• To experience Doppler effect, θ ≠ 90o
14
Δf = (2vfCos θ) /c
Doppler Shift = 2 x Velocity of Blood x Transducer Freq x Cos θ
Propagation Velocity of Sound
FACTORS AFFECTING DOPPLER SHIFT
If the object is moving perpendicular (θ=90o) to the ultrasound beam, there is no change
in frequency or wavelength.
By measuring the frequency shift, Doppler Ultrasonography can be used to determine
any or all of these 4:
• Speed
• Direction
• Flow (Laminar or Turbulent)
• Location
15
Δf = (2vfCos θ) /c
USES OF DOPPLER METHOD
With Doppler, US can be used for static and dynamic imaging.
The Doppler method can be used for:
• detection and characterization of blood flow,
• detection of foetal heartbeat,
• detection of air emboli,
• blood pressure monitoring, and
• localization of blood vessel occlusions.
16
17
CLASSIFICATIONS OF DOPPLER ULTRASONOGRAPHY
OUTPUTS/IMAGING MODESTYPES FUNCTIONS
PULSED
WAVE
DOPPLER
CONTINUOUS
WAVE
DOPPLER
COLOUR
DOPPLER
POWER
DOPPLER
SPECTRAL
DOPPLER
DUPLEX
MODE
TRIPLEX
MODE
TRANSCRANIAL
DOPPLER
TISSUE
DOPPLER
OTHERS
TYPES
• Continuous Wave Doppler
• Pulsed Wave Doppler
18
CONTINUOUS WAVE DOPPLER
• It is a continuous wave of ultrasound being sent into
the body.
• Measures mainly the velocity of moving blood.
19
CW: TRANSDUCER PROPERTIES
• CW uses a relatively narrow-band high-Q transducer, to
preserve velocity information.
20
CW: TRANSDUCER PROPERTIES
• The piezoelectric elements of the transducer are divided into 2:
• Transmitter: continuously transmitting US, and
• Receiver: detecting reflected echoes.
Doppler Shift = Received Freq – Transmitted Freq
21
22
Transmitter and Receiver are angled
against each other.
The Transmitter produces sinuosoidal US
waves of the form Cos 2πfot.
The received signal is of form:
Cos [ 2π (fo + fD) t ]
The Doppler signal is thus of form:
Cos 2πfDt , and can be recovered via
frequency demodulation of the
received signal.
Fig: CW Doppler Scan Geometry (angled Tx and Rx)
WALL FILTER: CW DOPPLER
• The Doppler signal contains very low frequency signals
(clutter) from moving specular reflectors, such as blood vessel
walls.
• A "Wall Filter" selectively removes these other low
frequencies.
23
BEAT: CW DOPPLER (contd.)
• The Doppler ultrasound signal is amplified to an audible
sound level (called Beat).
• Doppler Beat can be heard through a speaker or a
headphone.
24
CW DOPPLER (contd.)
• The audio indicates the spread of velocities involved in a
heart beat.
• It forms the basis of the ultrasound stethoscope for
foetal heartbeat monitoring.
25
Fig: Foetal Heart Rate (FHR) Doppler Detector
Some also display the Heart Rate in Beats Per Minute (BPM)
Uses the principle of Continuous Wave Doppler. 26
BEATS: CW DOPPLER (contd.)
• The higher the pitch, the greater the velocity.
• The harsher the sound, the greater the turbulence.
27
Δf = (2vfCos θ) /c
28
Fig: Block Diagram of CW Doppler System
CW DOPPLER (contd.)
• Output can also be displayed on a Spectrum analyser,
constructing a pixel grayscale.
• Quadrature Detection (type of signal processing) permits
determination of the direction of flow of blood in CWD, if
used.
29
CW DOPPLER SPECTRAL DISPLAY
• The pixel greyscale represents the magnitude of the short
time Fourier Transform of the Doppler signal.
• A pixel in a Doppler spectrum represents the proportion
of red blood cells (RBCs) in the field of view that was
moving at a particular velocity at a particular time.
30
LIMITATIONS OF CW DOPPLER
• Poor Spatial Resolution: resulting from the large area of overlap
between the transmitter and receiver beams.
• Range Ambiguity: Measures velocity, but not location.
• CW cannot be used to distinguish the flow of overlapping vessels
at different depths.
• Lack of TGC (time gain compensation).
31
ADVANTAGES OF CW DOPPLER
• For measuring fast flow (high velocities) without aliasing.
• Good for assessing deep-lying vessels.
• It is inexpensive.
32
PULSED WAVE DOPPLER
• It transmits a sequence of short pulses, rather than a
continuous sinusoidal wave.
• PW allows both velocity and depth information to be
obtained.
33
PW DOPPLER (contd.)
• One single group of array elements is used for both receiving and
transmitting.
• PW Doppler allows measurement from a small, specific blood volume
(region of interest), which is defined by a sample volume.
• After the pulse has traveled forth and back (travel time T), an electronic
range gate is opened for a short period of time to receive the echoes.
34
OPTIMIZING PW
• Doppler Shift measurement: Achieved using longer
spatial pulse length (SPL)/high Q-factor transducer.*
• Depth Selection: Achieved with an electronic time gate
circuit (or range gate) to reject all echo signals, except
those falling within the determined gate window.
35
SAMPLE AND HOLD OPERATION
As the echo from each successive transmission is received, a
single sample at the expected arrival time of echoes from the
range gate is acquired and held until the echo from the next
pulse is received.
36
Fig: Sample-and-
Hold Operation of
PW Doppler
37
If the reflectors are moving, the signal received from the range gate will
change with each subsequent pulse, and the sample-and-hold operation will
construct a staircase signal (as above).
PW DOPPLER (contd.)
• The second pulse should be transmitted no sooner than the
expected arrival time of the echoes from the range gate that arise
from the first pulse*.
• The pulse travel time T determines the shortest possible time
interval between two successive transmit pulses.
38
Pulse repetition Frequency (PRF) is the number of pulses that an ultrasound
system transmits into the body each second.
The PRF of the transmitted pulses is the effective sampling frequency.
PRF = 1/T
The maximum detectable frequency shift (Nyquist Limit) is determined by
the value of one-half PRF.
fmax = ½ PRF
39
NYQUIST LIMIT
CONDITION TO AVOID ALIASING:
The PRF must be at least twice the Nyquist Limit.
PRF ≥ 2 fmax
40
NYQUIST LIMIT
fmax = ½ PRF
PRF = 2 fmax
Otherwise, at high blood velocities, aliasing will occur.
Fig A: Aliasing as shown in spectral Doppler effect
41
Aliased signals wrap around to negative amplitude,
masquerading as reversed flow and underestimating velocity.
A
PW MAX. VELOCITY
• Recall Doppler frequency is given by:
Substituting Maximum PW Doppler Frequency fmax for Δf, we get:
fmax = (2 vmax f Cosθ ) /c
PRF/2 = (2 vmax f Cosθ ) /c
42
Δf = (2vfCos θ) /c
vmax = cPRF
4fCosθ
PW MAX VELOCITY
• vmax represents the maximum velocity of reflector (e.g.
RBCs) that can be measured without aliasing.
• For a range gate positioned at depth z,
z = ½ λ
PRF = c/2z
43
vmax = ___c2____
8 z f Cosθ
vmax = cPRF
4fCosθ
PW MAX. VELOCITY
• The maximum velocity that can be accurately
determined by Pulsed Doppler increases with PRF,
lower operating frequency, and increasing angle.
44
vmax = ___c2____
8 z f Cosθ
vmax = c PRF
4 f Cosθ
DISADVANTAGES OF PW DOPPLER
PW cannot accurately measure fast blood velocities due to aliasing.
To avoid aliasing:
• increase the PRF
• increase Doppler angle
• reduce the depth
• reduce the transducer frequency
• change the baseline
• use CW instead.
45
vmax = ___c2____
8 z f Cosθ
46
Fig: Block Diagram of a Pulsed Wave Doppler System
MULTIGATED (MG) PULSED WAVE DOPPLER
• This is a variation of Pulsed Wave Doppler.
• Whereas PW Doppler system can only provide information from a particular
sample, MG PW Doppler obtains information from several depths
simultaneously using multiple gates.
• After demodulation, the received signal is directed to a number of parallel
processing chains - each with slightly different range of gate settings. 47
MG PW DOPPLER (CONTD.)
• Velocity distribution profile across the vessel cross-section can be
determined.
• It is a useful diagnostic tool for presence of plaques and stenosis.
• Orientation and location of a desired vessel still remain a problem, which
can be solved by Duplex Scanning.
48
MODES OF DOPPLER IMAGING
To display Doppler information, one of 3 modes may be chosen:
• Color Doppler
• Power Doppler
• Spectral Doppler
49
(a)
Power Doppler 50
(b)
Colour Doppler
(c)
Spectral Doppler
COLOUR DOPPLER (CD or CCD)
• Color (flow) scanning involves displaying Colour Doppler data
on real time (B-mode) grayscale images.
• The superimposition is such that tissue volumes containing
• no detectable flow are displayed in greyscale,
• while those in motion are in colour (usually red and blue).
51
COLOUR DOPPLER
Colours are assigned, depending on:
• DIRECTION : motion towards (Positive Doppler shift) or
(Velocity Mode) away (Negative Doppler shift) from the transducer.
• FLOW (Variance Mode): Laminar or Turbulent flow
• VELOCITY MAGNITUDE: mapped to the colour intensity
52
53
LIMITATIONS OF COLOUR DOPPLER
• Clutter of slow-moving solid structures and noise can
overwhelm the smaller echoes from moving blood.
• Spatial resolution of the colour image is lower than
grayscale.
54
USES OF COLOUR DOPPLER
• Imaging the heart and major blood vessels in
applications for which mean flow velocity is a
diagnostically useful parameter.
• To recognize and localize vessels and vascular blockage.
55
POWER DOPPLER
• Power Doppler represents the total power in the
Doppler spectrum at each sample volume.
• Unlike colour Doppler, Power Doppler usually
uses a single colour.
56
POWER DOPPLER
1. Power Doppler uses the magnitude of return Doppler
signal strength / power /intensity / amplitude alone.
2. It ignores the angle and direction of flow (phase).
3. It adopts slower frame rates and thus has greater
sensitivity to motion of the patient, tissues, and
transducer. 57
Power α Intensity α Amplitude2
Fig: Power Doppler
58
59
60
ADVANTAGES OF POWER DOPPLER
• PW’s greater sensitivity to motion allows detection and interpretation of
very subtle and slow blood flow.
• Power Doppler provides a more continuous display of tortuous vessels.
• Helps to rule out vascular occlusions,
• Helps to differentiate between blood carrying vessels and other fluid
occurrences
• It is not prone to aliasing.
61
DISADVANTAGES OF POWER DOPPLER
• It is susceptible to flash artifacts, which are colour signals due to its greater
sensitivity to motion.
• No information on flow velocity and direction of flow.
• Evaluation of hemodynamic properties such as the pulsation of flow, is
limited, since, most often, the image frame rate is too low.
62
USES OF POWER DOPPLER
• Used for tumour imaging, where moving blood volume is a
diagnostically useful parameter.
63
SPECTRAL DOPPLER
• The Spectral Doppler uses the Doppler frequency shift of the
echo signal as a measure of flow velocity and direction of flow.
• The Doppler spectrum ascertains the distribution of flow
velocities and their frequencies of occurrence at a defined
sampling site as a function of time.
64
SAMPLING SITE: SPECTRAL DOPPLER
• The sampling site is determined by using B-mode, Color Doppler
or Power Doppler image display (Duplex imaging).
65
SPECTRAL DOPPLER
The spectra from within or directly behind a stenosis (numbers 2 to 4) show increased
velocities, turbulence and reverse flow.
These changes in spectrum provide information on flow in the blood vessels.
66
FLOW
• Laminar flow normally exists at the centre of large
smooth vessels.
• Turbulent flow occurs when the vessel is disrupted by
plaque and stenosis.
67
USES OF SPECTRAL DOPPLER
68
SPECTRAL DOPPLER
• Spectral Doppler provides a detailed qualitative, semi-
quantitative or quantitative evaluation of hemodynamic
changes in tissues.
DUPLEX SCANNING
• Duplex scanning combines 2D B-mode imaging and a Doppler
type (e.g. Colour Doppler).
• The duplex system allows estimation of the flow velocity directly
from the Doppler shift frequency.
69
Duplex mode : Combining B-mode Greyscale + Colour Doppler
70
TRIPLEX MODE
71
Combines 3 methods: B-Scale greyscale, Colour Doppler and Spectral Doppler
(SOFT) TISSUE DOPPLER IMAGING
• Used to image soft tissue using a low pass filter.
• A conventional Doppler system assumes blood flow is concentrated at
intermediate and high velocities, while scatterers moving at low
velocities correspond to soft tissue.
• Therefore to eliminate such low velocities, a high pass wall filter is
usually used.
72
(SOFT) TISSUE DOPPLER IMAGING
• Tissue Doppler Imaging replaces the high pass wall filter with a
low pass filter, thus allowing only low Doppler frequency signal,
corresponding to low velocity movements.
• Can be done in Pulsed Wave Doppler mode or Colour Doppler mode.
• Used in diagnosing regional abnormalities in ventricular wall motion.
73
TRANSCRANIAL DOPPLER ULTRASONOGRAPHY (TCD)
• TCD is a non-imaging Pulsed Wave Doppler technique measuring
local blood flow velocity and direction in the proximal portions
of large intracranial arteries.
• Used for evaluation and management of patients with risk of
cerebrovascular disease.
74
TCD: Stroke Prevention for Children with SCD
• In 2017, Dr. M. Adekunle and Prof. O. Akinyanju (of Sickle Cell
Foundation Nigeria) recommended annual screening of children
with Sickle Cell Disorder (SCD to assess risk of stroke.
• Children aged 2-16 years with Sickle Cell Disorder (SCD) have a
high risk of stroke, esp. children aged 2-8 years.
75
TCD: Stroke Prevention for Children with SCD
• Stroke occurs when part of the brain is damaged by inadequate
blood supply.
• Inadequate blood supply occurs due to cerebral bleeding
(bleeding from arteries) or thrombosis (blood clots), which
impede flow of blood in the arteries of the brain.
76
If the velocity in an artery exceeds 250cm/s, that area is at a higher risk of stroke
[Adams]. 77
Fig: Transcranial Doppler Ultrasound of blood vessels in the Circle of Willis of the brain
TCD: Stroke Prevention (contd.)
• Stroke can cause paralysis, seizures, loss of speech or reduced
intellectual capacity.
• Stroke prevention therapy can be carried out for affected children.
• Adekunle and Akinyanju recommend establishment of more TCD
centres across Nigeria.
78
79
KNOBOLOGY
Some ultrasound units have a knob for PW, CW, Colour, and Power Doppler modes.
(A) (B)
ARTIFACTS IN DOPPLER
• Aliasing
• Mirror Artifact
• Vibration Artifact (Bruit)
• Velocity Artifact
• Shadowing Artifact
80
ALIASING
• Aliasing is error due to insufficient sampling rate PRF
• It causes high velocity forward flow to appear as reverse flow.
• Occurs in Colour Doppler, but not Power Doppler.
81
82
Fig: Increasing PRF can solve Aliasing Artifact
Due to reverberations and mirroring at strong reflectors, images in Color or
Power Doppler show artifacts, similar to those observed in B-mode.
e.g. mirroring of hepatic vessels at the diaphragm.
83
Caused by tissue vibrations at the site of an arterial-venous fistula, occlusion or
stenosis. pulsating blood pressure is transmitted to adjacent tissue causing tissue
vibration.
This is due to pulsating blood pressure at occlusions transmitted to adjacent
tissue, recognized as a colour mosaic. 84
• Due to changing Doppler angle
and transducer type used.
• Occurs only in Colour Doppler
(Doppler angle does not affect
Power Doppler)
• It can lead to Reverse-Flow and
Flow Acceleration artifacts.
85
• Convex/Sector Transducer & Linear Vessels: Curved transducer aperture
introduces Doppler angles, causing inaccurate depiction of blood vessel velocities.
• Linear Transducer & Curved Vessels: Produces similar effect as the above.
In the display of blood flow in the verterbral column in the Power Doppler mode,
due to strong reflectors (bones), a shadow is cast on the blood vessels, making
them look interrupted.
86
OTHER FORMS OF ANGIOGRAPHY (DSA vs CT vs MR)
• Digital Subtracted Angiography is the ‘gold standard’ in arterial
imaging.
• It uses X-ray based fluoroscopy.
• It produces its image by subtracting a pre-contrast image with an image
taken after injection of a contrast medium.
• It is considered invasive, as it involves use of catheter.
• It requires the patient to be still, and thus not favoured for heart
imaging.
87
OTHER FORMS OF ANGIOGRAPHY (DSA vs CT vs MR)
CT Angiography (CTA) has the advantage of producing 3D
images. It is less invasive and stressful for the patient.
MR Angiography (MRA) avoids ionizing radiation and
nephrotoxic contrast agents.
88
CTA vs MRA
Advantages of CTA over MRA include:
• Visualization of calcified plaques and
• Insusceptibility to metallic vascular clips.
• Lower cost: Iodinated contrast material for CT angiography is
less expensive than the dose of gadolinium required for MR
angiography.
89
CTA vs MRA
• Coronary Artery Disease: CTA offers better sensitivity and
specificity better than MRA.
• Arterial Stenosis of the Aortoiliac and Renal Arteries: No
statistically significant difference between CTA and MRA.
90
CTA vs MRA
• Aorto-iliac Arterial Disease: CTA shows higher sensitivity
(98.7%) than CCD (96.2%) in the assessment of aorto-iliac
arterial disease [A. Osama et al, 2012 ].
• Agreement between DSA and CCD was 96.1%
91
• Doppler Ultrasonography has a relatively high sensitivity in
comparison with digital subtraction angiography.
• It is cheaper, less-invasive, non-ionizing and offers minimal risk
of bio-effects.
92
CONCLUSION
REFERENCES
D. R. Dance, et al. Diagnostic Radiology, Physics. IAEA. 2014: Vienna.
M. A. Aweda. Principles of Doppler Imaging. Lagos University Teaching Hospital. Lagos. 2012.
K. O. Soyebi. An Introduction to Transcranial Doppler Imaging. Lagos University Teaching Hospital. Lagos.
2012.
J. T. Bushberg, et al. The Essential Physics of Medical Imaging. 2nd Ed. Philadelphia: Lippincott Williams &
Wilkins. 2002.
W. Huda, R. Sloan. Review of Radiologic Physics. 3rd ed. Philadelphia: Lippincott Williams & Wilkins. 2009.
M. Adekunle, O. Akinyanju. Prevention of Stroke in Children with Sickle Cell Anaemia. Sickle Cell Bulletin.
Vol 8. No. 1. Sickle Cell Foundation Nigeria. 2017.
Siemens AG. Principles of Ultrasound Imaging. 1999: Med USSE
J. M. Adams. Ultrasound’s Transcranial Doppler Imaging Checks for Risk of Stroke. 2016. Cincinnati
Children’s Hospital. http://www.blog.cincinnatichildrens.org/radiology/
W. R. Hendee, E. R. Ritenour. Medical Imaging Physics. 4th Ed. New York: Wiley-Liss. 2002.
A. Osama, et al. Role of Multi-Slice CT Angiography versus Doppler Ultrasonography and Conventional
Angiography in assessment of aorto-iliac arterial disease. Egyptian Journal of Radiology and Nuclear Medicine.
2012.
R. Herzig, et al. Comparison Of Ultrasonography, CT Angiography, and Digital Subtraction Angiography In Severe
Carotid Stenoses. European Journal of Neurology. 2004.
93
THANK YOU
94

Doppler Effect - Ultrasound

  • 1.
    DOPPLER EFFECT Course: Ultrasound VictorEkpo MSc Medical Physics programme, College of Medicine, University of Lagos, 2017.
  • 2.
    OUTLINE • Introduction • Physicsof Doppler Effect • Conditions for Doppler Effect • Classifications • Types • Modes • Applications • Artifacts in Doppler Imaging • Conclusion 2
  • 3.
    INTRODUCTION • Ultrasound issound waves with frequency greater than 20kHz (Medical Ultrasound range: >2MHz) • Sound waves travelling through a medium can be reflected, refracted, absorbed or transmitted. • Reflection occurs at the point of acoustic impedance mismatch. 3
  • 4.
    MOVING SOUND • Oneor both the sound source and receiver may be stationary or moving. • If the sound source moves towards the listener, the sound is perceived to have a higher frequency/pitch, and a lower frequency as it moves away from the listener. 4
  • 5.
    DOPPLER EFFECT Doppler effectis the change in frequency of sound (or any wave) due to the relative motion* of the source and receiver. Doppler Shift (Δf) = Reflected Frequency – Transmitted frequency Doppler angle ( θ ) = Angle between the direction of source – direction of sound 5
  • 6.
    CONDITIONS FOR DOPPLERSHIFT The Doppler effect will NOT occur: • If the source and observer both move in the same direction at the same speed. • Where one source/listener is at the centre of a circle while the other is moving on it with uniform speed. • If the source/receiver is at 900 to the receiver/source. 6
  • 7.
    7 Longer wavelengths Lower frequency Shorterwavelengths Higher frequency TOWARDS AWAY FROM
  • 8.
    8 Sound waves reflectedfrom a moving object are compressed (higher frequency) when moving towards the listener/ transducer, …and expanded (lower frequency) when moving away from the listener/transducer compared to the incident sound wave frequency.
  • 9.
  • 10.
    The velocity ofsound (c) is given by the equation, Speed, c = 𝑊𝑎𝑣𝑒𝑙𝑒𝑛𝑔𝑡ℎ (λ) 𝑃𝑒𝑟𝑖𝑜𝑑 (𝑇) = Wavelength (λ) * frequency (f) λo = 𝑐 𝑓 𝑜 Consider a sound source travelling at a velocity 𝒗 𝒔 towards a detector. After time t, the new wavelength, λ1 = 𝑐 − 𝒗 𝒔 𝑓 𝑜 10 PHYSICS OF DOPPLER EFFECT 𝒗 𝒔 c
  • 11.
    With a shortenedwavelength, the new frequency f1 is given by: f1 = 𝑐 λ1 = 𝑐 (c−vs) / fo f1 = fo 𝑐 ( 𝑐 − 𝒗 𝒔 ) Δf = f1 - fo = fo 𝑐 ( 𝑐 − 𝒗 𝒔 ) - fo = fo 𝑐 ( 𝑐 − 𝒗 𝒔 ) - 1 = fo 𝑐 −( 𝑐 −𝑣𝑠 ) ( 𝑐 − 𝒗 𝒔 ) Δf = fo 𝑣 𝑠 ( 𝑐 − 𝒗 𝒔 ) If c >> vs , then Δf = fo 𝑣 𝑠 𝑐 11
  • 12.
    A similar equationcan be written for a detector moving towards a source, Δf = fo 𝑣 𝑑 𝑐 It can be shown that for a source moving away, λ1 = 𝑐 + 𝒗 𝒔 𝑓 𝑜 Δf = fo −𝑣 𝑠 𝑐 OR fo −𝑣𝑑 𝑐 For a 2-way motion where a transducer emits ultrasound of frequency f, and it gets reflected back by moving blood of velocity v (and c>>v), then: Δf = 2f 𝑣 𝑐 12
  • 13.
    If direction ofmotion is oblique, then: Δf = (2vf Cos θ) /c where: Δf = Doppler shift f = transducer frequency v = velocity of source/receiver c = velocity of sound θ = Doppler angle 13 PHYSICS OF DOPPLER EFFECT (contd.)
  • 14.
    • Doppler shift,Δf increases with: • increasing source velocity v, and • increasing transducer frequency, f • Δf is maximum at θ = 0o and minimum at θ = 90o. • To experience Doppler effect, θ ≠ 90o 14 Δf = (2vfCos θ) /c Doppler Shift = 2 x Velocity of Blood x Transducer Freq x Cos θ Propagation Velocity of Sound FACTORS AFFECTING DOPPLER SHIFT
  • 15.
    If the objectis moving perpendicular (θ=90o) to the ultrasound beam, there is no change in frequency or wavelength. By measuring the frequency shift, Doppler Ultrasonography can be used to determine any or all of these 4: • Speed • Direction • Flow (Laminar or Turbulent) • Location 15 Δf = (2vfCos θ) /c
  • 16.
    USES OF DOPPLERMETHOD With Doppler, US can be used for static and dynamic imaging. The Doppler method can be used for: • detection and characterization of blood flow, • detection of foetal heartbeat, • detection of air emboli, • blood pressure monitoring, and • localization of blood vessel occlusions. 16
  • 17.
    17 CLASSIFICATIONS OF DOPPLERULTRASONOGRAPHY OUTPUTS/IMAGING MODESTYPES FUNCTIONS PULSED WAVE DOPPLER CONTINUOUS WAVE DOPPLER COLOUR DOPPLER POWER DOPPLER SPECTRAL DOPPLER DUPLEX MODE TRIPLEX MODE TRANSCRANIAL DOPPLER TISSUE DOPPLER OTHERS
  • 18.
    TYPES • Continuous WaveDoppler • Pulsed Wave Doppler 18
  • 19.
    CONTINUOUS WAVE DOPPLER •It is a continuous wave of ultrasound being sent into the body. • Measures mainly the velocity of moving blood. 19
  • 20.
    CW: TRANSDUCER PROPERTIES •CW uses a relatively narrow-band high-Q transducer, to preserve velocity information. 20
  • 21.
    CW: TRANSDUCER PROPERTIES •The piezoelectric elements of the transducer are divided into 2: • Transmitter: continuously transmitting US, and • Receiver: detecting reflected echoes. Doppler Shift = Received Freq – Transmitted Freq 21
  • 22.
    22 Transmitter and Receiverare angled against each other. The Transmitter produces sinuosoidal US waves of the form Cos 2πfot. The received signal is of form: Cos [ 2π (fo + fD) t ] The Doppler signal is thus of form: Cos 2πfDt , and can be recovered via frequency demodulation of the received signal. Fig: CW Doppler Scan Geometry (angled Tx and Rx)
  • 23.
    WALL FILTER: CWDOPPLER • The Doppler signal contains very low frequency signals (clutter) from moving specular reflectors, such as blood vessel walls. • A "Wall Filter" selectively removes these other low frequencies. 23
  • 24.
    BEAT: CW DOPPLER(contd.) • The Doppler ultrasound signal is amplified to an audible sound level (called Beat). • Doppler Beat can be heard through a speaker or a headphone. 24
  • 25.
    CW DOPPLER (contd.) •The audio indicates the spread of velocities involved in a heart beat. • It forms the basis of the ultrasound stethoscope for foetal heartbeat monitoring. 25
  • 26.
    Fig: Foetal HeartRate (FHR) Doppler Detector Some also display the Heart Rate in Beats Per Minute (BPM) Uses the principle of Continuous Wave Doppler. 26
  • 27.
    BEATS: CW DOPPLER(contd.) • The higher the pitch, the greater the velocity. • The harsher the sound, the greater the turbulence. 27 Δf = (2vfCos θ) /c
  • 28.
    28 Fig: Block Diagramof CW Doppler System
  • 29.
    CW DOPPLER (contd.) •Output can also be displayed on a Spectrum analyser, constructing a pixel grayscale. • Quadrature Detection (type of signal processing) permits determination of the direction of flow of blood in CWD, if used. 29
  • 30.
    CW DOPPLER SPECTRALDISPLAY • The pixel greyscale represents the magnitude of the short time Fourier Transform of the Doppler signal. • A pixel in a Doppler spectrum represents the proportion of red blood cells (RBCs) in the field of view that was moving at a particular velocity at a particular time. 30
  • 31.
    LIMITATIONS OF CWDOPPLER • Poor Spatial Resolution: resulting from the large area of overlap between the transmitter and receiver beams. • Range Ambiguity: Measures velocity, but not location. • CW cannot be used to distinguish the flow of overlapping vessels at different depths. • Lack of TGC (time gain compensation). 31
  • 32.
    ADVANTAGES OF CWDOPPLER • For measuring fast flow (high velocities) without aliasing. • Good for assessing deep-lying vessels. • It is inexpensive. 32
  • 33.
    PULSED WAVE DOPPLER •It transmits a sequence of short pulses, rather than a continuous sinusoidal wave. • PW allows both velocity and depth information to be obtained. 33
  • 34.
    PW DOPPLER (contd.) •One single group of array elements is used for both receiving and transmitting. • PW Doppler allows measurement from a small, specific blood volume (region of interest), which is defined by a sample volume. • After the pulse has traveled forth and back (travel time T), an electronic range gate is opened for a short period of time to receive the echoes. 34
  • 35.
    OPTIMIZING PW • DopplerShift measurement: Achieved using longer spatial pulse length (SPL)/high Q-factor transducer.* • Depth Selection: Achieved with an electronic time gate circuit (or range gate) to reject all echo signals, except those falling within the determined gate window. 35
  • 36.
    SAMPLE AND HOLDOPERATION As the echo from each successive transmission is received, a single sample at the expected arrival time of echoes from the range gate is acquired and held until the echo from the next pulse is received. 36
  • 37.
    Fig: Sample-and- Hold Operationof PW Doppler 37 If the reflectors are moving, the signal received from the range gate will change with each subsequent pulse, and the sample-and-hold operation will construct a staircase signal (as above).
  • 38.
    PW DOPPLER (contd.) •The second pulse should be transmitted no sooner than the expected arrival time of the echoes from the range gate that arise from the first pulse*. • The pulse travel time T determines the shortest possible time interval between two successive transmit pulses. 38
  • 39.
    Pulse repetition Frequency(PRF) is the number of pulses that an ultrasound system transmits into the body each second. The PRF of the transmitted pulses is the effective sampling frequency. PRF = 1/T The maximum detectable frequency shift (Nyquist Limit) is determined by the value of one-half PRF. fmax = ½ PRF 39 NYQUIST LIMIT
  • 40.
    CONDITION TO AVOIDALIASING: The PRF must be at least twice the Nyquist Limit. PRF ≥ 2 fmax 40 NYQUIST LIMIT fmax = ½ PRF PRF = 2 fmax Otherwise, at high blood velocities, aliasing will occur.
  • 41.
    Fig A: Aliasingas shown in spectral Doppler effect 41 Aliased signals wrap around to negative amplitude, masquerading as reversed flow and underestimating velocity. A
  • 42.
    PW MAX. VELOCITY •Recall Doppler frequency is given by: Substituting Maximum PW Doppler Frequency fmax for Δf, we get: fmax = (2 vmax f Cosθ ) /c PRF/2 = (2 vmax f Cosθ ) /c 42 Δf = (2vfCos θ) /c vmax = cPRF 4fCosθ
  • 43.
    PW MAX VELOCITY •vmax represents the maximum velocity of reflector (e.g. RBCs) that can be measured without aliasing. • For a range gate positioned at depth z, z = ½ λ PRF = c/2z 43 vmax = ___c2____ 8 z f Cosθ vmax = cPRF 4fCosθ
  • 44.
    PW MAX. VELOCITY •The maximum velocity that can be accurately determined by Pulsed Doppler increases with PRF, lower operating frequency, and increasing angle. 44 vmax = ___c2____ 8 z f Cosθ vmax = c PRF 4 f Cosθ
  • 45.
    DISADVANTAGES OF PWDOPPLER PW cannot accurately measure fast blood velocities due to aliasing. To avoid aliasing: • increase the PRF • increase Doppler angle • reduce the depth • reduce the transducer frequency • change the baseline • use CW instead. 45 vmax = ___c2____ 8 z f Cosθ
  • 46.
    46 Fig: Block Diagramof a Pulsed Wave Doppler System
  • 47.
    MULTIGATED (MG) PULSEDWAVE DOPPLER • This is a variation of Pulsed Wave Doppler. • Whereas PW Doppler system can only provide information from a particular sample, MG PW Doppler obtains information from several depths simultaneously using multiple gates. • After demodulation, the received signal is directed to a number of parallel processing chains - each with slightly different range of gate settings. 47
  • 48.
    MG PW DOPPLER(CONTD.) • Velocity distribution profile across the vessel cross-section can be determined. • It is a useful diagnostic tool for presence of plaques and stenosis. • Orientation and location of a desired vessel still remain a problem, which can be solved by Duplex Scanning. 48
  • 49.
    MODES OF DOPPLERIMAGING To display Doppler information, one of 3 modes may be chosen: • Color Doppler • Power Doppler • Spectral Doppler 49
  • 50.
    (a) Power Doppler 50 (b) ColourDoppler (c) Spectral Doppler
  • 51.
    COLOUR DOPPLER (CDor CCD) • Color (flow) scanning involves displaying Colour Doppler data on real time (B-mode) grayscale images. • The superimposition is such that tissue volumes containing • no detectable flow are displayed in greyscale, • while those in motion are in colour (usually red and blue). 51
  • 52.
    COLOUR DOPPLER Colours areassigned, depending on: • DIRECTION : motion towards (Positive Doppler shift) or (Velocity Mode) away (Negative Doppler shift) from the transducer. • FLOW (Variance Mode): Laminar or Turbulent flow • VELOCITY MAGNITUDE: mapped to the colour intensity 52
  • 53.
  • 54.
    LIMITATIONS OF COLOURDOPPLER • Clutter of slow-moving solid structures and noise can overwhelm the smaller echoes from moving blood. • Spatial resolution of the colour image is lower than grayscale. 54
  • 55.
    USES OF COLOURDOPPLER • Imaging the heart and major blood vessels in applications for which mean flow velocity is a diagnostically useful parameter. • To recognize and localize vessels and vascular blockage. 55
  • 56.
    POWER DOPPLER • PowerDoppler represents the total power in the Doppler spectrum at each sample volume. • Unlike colour Doppler, Power Doppler usually uses a single colour. 56
  • 57.
    POWER DOPPLER 1. PowerDoppler uses the magnitude of return Doppler signal strength / power /intensity / amplitude alone. 2. It ignores the angle and direction of flow (phase). 3. It adopts slower frame rates and thus has greater sensitivity to motion of the patient, tissues, and transducer. 57 Power α Intensity α Amplitude2
  • 58.
  • 59.
  • 60.
  • 61.
    ADVANTAGES OF POWERDOPPLER • PW’s greater sensitivity to motion allows detection and interpretation of very subtle and slow blood flow. • Power Doppler provides a more continuous display of tortuous vessels. • Helps to rule out vascular occlusions, • Helps to differentiate between blood carrying vessels and other fluid occurrences • It is not prone to aliasing. 61
  • 62.
    DISADVANTAGES OF POWERDOPPLER • It is susceptible to flash artifacts, which are colour signals due to its greater sensitivity to motion. • No information on flow velocity and direction of flow. • Evaluation of hemodynamic properties such as the pulsation of flow, is limited, since, most often, the image frame rate is too low. 62
  • 63.
    USES OF POWERDOPPLER • Used for tumour imaging, where moving blood volume is a diagnostically useful parameter. 63
  • 64.
    SPECTRAL DOPPLER • TheSpectral Doppler uses the Doppler frequency shift of the echo signal as a measure of flow velocity and direction of flow. • The Doppler spectrum ascertains the distribution of flow velocities and their frequencies of occurrence at a defined sampling site as a function of time. 64
  • 65.
    SAMPLING SITE: SPECTRALDOPPLER • The sampling site is determined by using B-mode, Color Doppler or Power Doppler image display (Duplex imaging). 65 SPECTRAL DOPPLER
  • 66.
    The spectra fromwithin or directly behind a stenosis (numbers 2 to 4) show increased velocities, turbulence and reverse flow. These changes in spectrum provide information on flow in the blood vessels. 66
  • 67.
    FLOW • Laminar flownormally exists at the centre of large smooth vessels. • Turbulent flow occurs when the vessel is disrupted by plaque and stenosis. 67
  • 68.
    USES OF SPECTRALDOPPLER 68 SPECTRAL DOPPLER • Spectral Doppler provides a detailed qualitative, semi- quantitative or quantitative evaluation of hemodynamic changes in tissues.
  • 69.
    DUPLEX SCANNING • Duplexscanning combines 2D B-mode imaging and a Doppler type (e.g. Colour Doppler). • The duplex system allows estimation of the flow velocity directly from the Doppler shift frequency. 69
  • 70.
    Duplex mode :Combining B-mode Greyscale + Colour Doppler 70
  • 71.
    TRIPLEX MODE 71 Combines 3methods: B-Scale greyscale, Colour Doppler and Spectral Doppler
  • 72.
    (SOFT) TISSUE DOPPLERIMAGING • Used to image soft tissue using a low pass filter. • A conventional Doppler system assumes blood flow is concentrated at intermediate and high velocities, while scatterers moving at low velocities correspond to soft tissue. • Therefore to eliminate such low velocities, a high pass wall filter is usually used. 72
  • 73.
    (SOFT) TISSUE DOPPLERIMAGING • Tissue Doppler Imaging replaces the high pass wall filter with a low pass filter, thus allowing only low Doppler frequency signal, corresponding to low velocity movements. • Can be done in Pulsed Wave Doppler mode or Colour Doppler mode. • Used in diagnosing regional abnormalities in ventricular wall motion. 73
  • 74.
    TRANSCRANIAL DOPPLER ULTRASONOGRAPHY(TCD) • TCD is a non-imaging Pulsed Wave Doppler technique measuring local blood flow velocity and direction in the proximal portions of large intracranial arteries. • Used for evaluation and management of patients with risk of cerebrovascular disease. 74
  • 75.
    TCD: Stroke Preventionfor Children with SCD • In 2017, Dr. M. Adekunle and Prof. O. Akinyanju (of Sickle Cell Foundation Nigeria) recommended annual screening of children with Sickle Cell Disorder (SCD to assess risk of stroke. • Children aged 2-16 years with Sickle Cell Disorder (SCD) have a high risk of stroke, esp. children aged 2-8 years. 75
  • 76.
    TCD: Stroke Preventionfor Children with SCD • Stroke occurs when part of the brain is damaged by inadequate blood supply. • Inadequate blood supply occurs due to cerebral bleeding (bleeding from arteries) or thrombosis (blood clots), which impede flow of blood in the arteries of the brain. 76
  • 77.
    If the velocityin an artery exceeds 250cm/s, that area is at a higher risk of stroke [Adams]. 77 Fig: Transcranial Doppler Ultrasound of blood vessels in the Circle of Willis of the brain
  • 78.
    TCD: Stroke Prevention(contd.) • Stroke can cause paralysis, seizures, loss of speech or reduced intellectual capacity. • Stroke prevention therapy can be carried out for affected children. • Adekunle and Akinyanju recommend establishment of more TCD centres across Nigeria. 78
  • 79.
    79 KNOBOLOGY Some ultrasound unitshave a knob for PW, CW, Colour, and Power Doppler modes. (A) (B)
  • 80.
    ARTIFACTS IN DOPPLER •Aliasing • Mirror Artifact • Vibration Artifact (Bruit) • Velocity Artifact • Shadowing Artifact 80
  • 81.
    ALIASING • Aliasing iserror due to insufficient sampling rate PRF • It causes high velocity forward flow to appear as reverse flow. • Occurs in Colour Doppler, but not Power Doppler. 81
  • 82.
    82 Fig: Increasing PRFcan solve Aliasing Artifact
  • 83.
    Due to reverberationsand mirroring at strong reflectors, images in Color or Power Doppler show artifacts, similar to those observed in B-mode. e.g. mirroring of hepatic vessels at the diaphragm. 83
  • 84.
    Caused by tissuevibrations at the site of an arterial-venous fistula, occlusion or stenosis. pulsating blood pressure is transmitted to adjacent tissue causing tissue vibration. This is due to pulsating blood pressure at occlusions transmitted to adjacent tissue, recognized as a colour mosaic. 84
  • 85.
    • Due tochanging Doppler angle and transducer type used. • Occurs only in Colour Doppler (Doppler angle does not affect Power Doppler) • It can lead to Reverse-Flow and Flow Acceleration artifacts. 85 • Convex/Sector Transducer & Linear Vessels: Curved transducer aperture introduces Doppler angles, causing inaccurate depiction of blood vessel velocities. • Linear Transducer & Curved Vessels: Produces similar effect as the above.
  • 86.
    In the displayof blood flow in the verterbral column in the Power Doppler mode, due to strong reflectors (bones), a shadow is cast on the blood vessels, making them look interrupted. 86
  • 87.
    OTHER FORMS OFANGIOGRAPHY (DSA vs CT vs MR) • Digital Subtracted Angiography is the ‘gold standard’ in arterial imaging. • It uses X-ray based fluoroscopy. • It produces its image by subtracting a pre-contrast image with an image taken after injection of a contrast medium. • It is considered invasive, as it involves use of catheter. • It requires the patient to be still, and thus not favoured for heart imaging. 87
  • 88.
    OTHER FORMS OFANGIOGRAPHY (DSA vs CT vs MR) CT Angiography (CTA) has the advantage of producing 3D images. It is less invasive and stressful for the patient. MR Angiography (MRA) avoids ionizing radiation and nephrotoxic contrast agents. 88
  • 89.
    CTA vs MRA Advantagesof CTA over MRA include: • Visualization of calcified plaques and • Insusceptibility to metallic vascular clips. • Lower cost: Iodinated contrast material for CT angiography is less expensive than the dose of gadolinium required for MR angiography. 89
  • 90.
    CTA vs MRA •Coronary Artery Disease: CTA offers better sensitivity and specificity better than MRA. • Arterial Stenosis of the Aortoiliac and Renal Arteries: No statistically significant difference between CTA and MRA. 90
  • 91.
    CTA vs MRA •Aorto-iliac Arterial Disease: CTA shows higher sensitivity (98.7%) than CCD (96.2%) in the assessment of aorto-iliac arterial disease [A. Osama et al, 2012 ]. • Agreement between DSA and CCD was 96.1% 91
  • 92.
    • Doppler Ultrasonographyhas a relatively high sensitivity in comparison with digital subtraction angiography. • It is cheaper, less-invasive, non-ionizing and offers minimal risk of bio-effects. 92 CONCLUSION
  • 93.
    REFERENCES D. R. Dance,et al. Diagnostic Radiology, Physics. IAEA. 2014: Vienna. M. A. Aweda. Principles of Doppler Imaging. Lagos University Teaching Hospital. Lagos. 2012. K. O. Soyebi. An Introduction to Transcranial Doppler Imaging. Lagos University Teaching Hospital. Lagos. 2012. J. T. Bushberg, et al. The Essential Physics of Medical Imaging. 2nd Ed. Philadelphia: Lippincott Williams & Wilkins. 2002. W. Huda, R. Sloan. Review of Radiologic Physics. 3rd ed. Philadelphia: Lippincott Williams & Wilkins. 2009. M. Adekunle, O. Akinyanju. Prevention of Stroke in Children with Sickle Cell Anaemia. Sickle Cell Bulletin. Vol 8. No. 1. Sickle Cell Foundation Nigeria. 2017. Siemens AG. Principles of Ultrasound Imaging. 1999: Med USSE J. M. Adams. Ultrasound’s Transcranial Doppler Imaging Checks for Risk of Stroke. 2016. Cincinnati Children’s Hospital. http://www.blog.cincinnatichildrens.org/radiology/ W. R. Hendee, E. R. Ritenour. Medical Imaging Physics. 4th Ed. New York: Wiley-Liss. 2002. A. Osama, et al. Role of Multi-Slice CT Angiography versus Doppler Ultrasonography and Conventional Angiography in assessment of aorto-iliac arterial disease. Egyptian Journal of Radiology and Nuclear Medicine. 2012. R. Herzig, et al. Comparison Of Ultrasonography, CT Angiography, and Digital Subtraction Angiography In Severe Carotid Stenoses. European Journal of Neurology. 2004. 93
  • 94.

Editor's Notes

  • #6 NEWTON’S 1ST LAW OF MOTION & ITS RELATION Doppler effect will not occur if both the source and receiver are both static, or both moving at the same speed in the same direction. Doppler effect can happen in light and aby other wave.
  • #7 This is explained in Slide 12. Moving is in acircle is same distance. Source/Receiver has to move away or towards.
  • #8 Sound emitted is continuous sound. So subsequent waves take longer/shorter to reach , thus appearing as a frequency change.
  • #11 Frequency of sound is determined by the source (and now detector) only. Speed is determined by the medium only. Wavelength is det. by both source and medium.
  • #17 Emboli: (sing. Embolus) : a mass, most commonly a blood clot, that becomes lodged in a blood vessel and obstructs it. Air embolus is air that escapes from lungs to blood vessels, blocking some artieries. Usually if divers emerge from a depth too quickly.
  • #20 But with quadrature detection, direction of flow is also possible.
  • #21 where fo is the centre frequency and the bandwidth is the width of the frequency distribution.
  • #24 The Doppler signal from the clutter is much greater than that from blood.
  • #25 Beats can be recorded to track spectral changes as a function of time to assess transient pulsatile flow.
  • #27 These are becoming very popular in the developed world, and cost just over $100.
  • #28 SEE FORMULA.
  • #30 Resolution is very low, due to use of high Q transducer.
  • #31 Fourier Transform is used to construct an image.
  • #32 Lack of TGC: Reflections created from RBC located at deeper depth will have a low amplitude than reflections from shallower depth.
  • #34 Uses the same crystals to send and receive the signal.
  • #36 Longer spatial pulse length give poorer axial resolution. Axial Resolution = SPL/2. Lower numerical values of axial resolution give better image quality.
  • #37 PW Doppler works with samples.
  • #39 This is determined by the depth of the range gate. The pulse travel time T determines the shortest possible time interval between two successive transmit pulses.
  • #40 Using Shannon’s sampling theorem, the maximum unaliased frequency is as given above. The PRF must be at least twice the sampling max Doppler freq shift.
  • #41 Using Shannon’s sampling theorem, the maximum unaliased frequency is as given above. The PRF must be at least twice the sampling max Doppler freq shift.
  • #42 Aliasing is an an error caused by an insufficient sampling rate (PRF) relative to the high-frequency Doppler signals generated by fast moving Corrected here by adjusting the baseline.
  • #43 V(max) represents maximum velocity that can be measured by PW Doppler effect
  • #47 Just like in CW Doppler, demodulation is used to obtain velocity information from the echo signal by separating the Doppler signal of frequency f from the received signal. Main diff is that PW includes a Sample and Hold & Range Delay
  • #48 This allows for a number of adjacent sample volumes to be positioned across a vessel Thus the problem of locating the vessel is greatly reduced
  • #52  Acquisition of image information is interleaved with that of flow information.
  • #53 Red and blue are used. But any colour scheme can be chosen.
  • #62 What are flash artifacts? Fluid occurences may be static. Blood vessels carry dynamic blood.
  • #63 What are flash artifacts?
  • #64 Since Doppler power is proportional to the concentration of moving blood cells in the sample volume, it is typically used for tumour imaging, where blood volume is a diagnostically useful parameter.
  • #66 increasing turbulence increases vascular resistance
  • #67 Systolic: heart contractions – during which blood is pumped into arteries Diastolic: expansion of each heartbeat
  • #68 The duplex system allows estimation of the flow velocity directly from the Doppler shift frequency. , since the velocity of sound and the transducer frequency are known, while the Doppler angle can be estimated from the B-mode image by the user and input into the scanner computer for calculation.
  • #70 The duplex system allows estimation of the flow velocity directly from the Doppler shift frequency. , since the velocity of sound and the transducer frequency are known, while the Doppler angle can be estimated from the B-mode image by the user and input into the scanner computer for calculation.
  • #85 Fistula is an opening caused by disease. Stenosis and occlusions are obstructions or constrictions.
  • #87 Fistula is an opening caused by disease. Stenosis and occlusions are obstructions or constrictions.
  • #91 Sensitivity: True positive rate. Ability to correctly identify those with disease Specificity: Tue negative rate. Ability to correctly identify those without disease.